Inficon IRwin Methan Leak Detector Quick Start Manual
Read Online
Read the Inficon IRwin Methan Leak Detector Quick Start Manual online using the built-in PDF viewer.
Text of the manual
How to Perform GC Function Test on IRwin Methane Leak Detector
(for a factory set unit)
QUICK START GUIDE
1 of 3
1
Make sure your IRwin is set in Universal Mode (see
the How to set your IRwin for Universal Mode quick start
guide)
Select Function Test and press
2
3
4
5
Press
followed by
Press to perform a function test with GC Mix
1.0% CH4 50ppm C2H6 1000ppm C3H8
mimd66en1-03 (1910)
Press . Use to select GC in the drop-down
menu, press
How to Perform GC Function Test on IRwin Methane Leak Detector
6
Press to start the Function Test for GC
7
Expose the instrument to GC gas mix
(GC mix 1.0% CH4 50ppm C2H6 1000ppm C3H8)
8
The Function Test is in progress
9
Expose both probe and instrument to fresh air
2 of 3
10
The GC Analysis is in progress
mimd66en1-03 (1910)
Due to our continuing program of product improvements, specifications are subject to change without notice.
All trademarks are the property of the ir respective owners.
©2019 INFICON
mimd66en1-03 (1910)
How to Perform GC Function Test on IRwin Methane Leak Detector
11
The Function Test is successfully completed when
the graph shows detection of methane, ethane and
propane
12
A successful test will be confirmed by a green, ticked
circle. You can also save the GC report.
Use IRwin App to transfer your saved log files to your
computer and create pdf file of the test.
IRwin App is available on www.inficon.com
13
IRwin SXG
S/N: 92001853
S/W ver: 3.06.01
Function Test Result GC
10/10/2019
13:46:53
60s
120s
180s
45B Westmansgatan
Gas Concentration 1.6 %
CH4 (Methane)
C2H6 (Ethane)
C3H8 (Propane)
Address
Signature/Department
Comments
www.inficon.com
v3.21
003 012 12.8