Amico Overbed Table Operation & User’s Manual
Read Online
Read the Amico Overbed Table Operation & User’s Manual online using the built-in PDF viewer.
Text of the manual
Table of Contents
Introduction
3
Specifications
3
Warning / Caution / Note Definition
3
Operations Instructions
4
Single Top
4
Flip Top
4
Safe Usage Illustrations
5
Raising the Table
5
Lowering the Table
5
Flip Top Overbed Table Assembly
6-8
Parts List
6
Part Numbers
7
Part Numbers - Accessories
8
Preventative Maintenance
9
Checklist
9
Cleaning
9
Service Information
10
Single Top
10
Flip Top
10
Gas Spring
11
Outer Column Replacement
11
Caster Replacement
11
2 Amico Patient Care Corporation
Warning / Caution / Note Definition
The words WARNING, CAUTION and NOTE carry special meanings and should be carefully reviewed.
WARNING – Alerts the reader about a situation, which if not avoided, could result in death or serious injury. It
may also describe potentially serious adverse reactions and safety hazards.
CAUTION – Alerts the reader of a potentially hazardous situation, which if not avoided, may result in minor or
moderate injury to the user or damage to the equipment or other property. This includes special care necessary
for the safe and effective use of the device and the care necessary to avoid damage to a device that may occur
as a result of misuse or abuse.
NOTE – This provides special information to make maintenance easier or important instructions clearer.
Specifications
Specifications
Single Top
Flip Top
Base Length
28"
28"
Overall Table Top Length
32"
13.5"
Overall Table Top Width
17"
17"
Minimum Table Top Height
29"
29"
Maximum Table Top Height
44"
44"
Weight of Table
31 lbs
40 lbs
www.amico.com 3
1. Pull up on the table top to raise it.
2. To unlock and lower the table top, squeeze handle (A) upwards while guiding the table top
downwards.
Flip Top
1. Pull up on the table top to raise it.
2. To unlock and lower the table top, squeeze handle (A) upwards while guiding the table top
downwards.
3. Lift the secondary flip top surface all the way up and release. The flip top surface will now stay in its
extended position.
4. To fold away the flip top surface, raise the secondary top all the way up and slowly release to let it
flip down.
Operating Instructions
4 Amico Patient Care Corporation
Safe Usage Illustrations
Raising the Table
Gently pull up on the table top using both hands.
Lowering the Table
Gently lower the table by placing one hand on the table top while
using the other for controlling the release handle.
WEIGHT CAPACITY: 100 lbs evenly distributed over the main surface
WEIGHT CAPACITY: 10 lbs on the flip surface
NOTE: Amico overbed tables have a shock absorbent spring. When pressure is applied, the table will lower
slightly and spring back up when pressure is released.
WARNING: Ensure the overbed table is clear of the bed or any other piece of equipment before raising/lowering
the table or the other piece of equipment to ensure the table does not react.
WARNING: DO NOT use the table to:
• Sit on
• Lean on
• Get in/out of bed
• As a walker
WARNING: Never put children on the table
CAUTION: Do not apply more than 100 lbs, evenly distributed, to the main surface
CAUTION: Do not apply more than 10 lbs to the flip surface at any time
www.amico.com 5
Flip Top Overbed Table Assembly
Item
Part Name
Qty
A
Main Table Top
1
B
Flip Top
1
C
Butt Hinges
2
D
Gas Spring Lock Nut
1
E
Handle Bracket
1
F
Release Handle
1
G
Handle Support Pin
1
H
Retaining Clip
1
I
Support Bracket
1
J
Column Guide
2
K
Outer Column
1
L
Outer Column Cap
1
M
Flip Top Support
1
N
Gas Spring
1
O
Inner Column
1
P
Gas Spring Support Pin
1
Q
Retaining Clip
1
R
Base Plastic Cover
1
S
Base Frame
1
T
Front Casters
2
U
Back Casters
2
NOTE: Single Top model does not include items; “B”, “C” and “M”.
6 Amico Patient Care Corporation
Main Table Top - Laminate
1
A
OT-MAIN-TXX
Main Table Top - Thermofoil
1
B
OT-FLP-LXX
Flip Top - Laminate
1
B
OT-FLP-TXX
Flip Top – Thermofoil
1
C
OT-FLP-HINGE
Butt Hinges
2
D
N/A – PART OF SPRING
Gas Spring Lock Nut
1
E
OT-PSTN-BRKTG
Handle Bracket
1
F
OT-PLHDL-X
Release Handle
1
G
OT-HDL-PIN
Handle Support Pin
1
H
N/A – PART OF PIN
Retaining Clip
1
I
OT-TBL-SUPRTX
Support Bracket
1
J
OT-GUIDE
Column Guide
4
K
OT-OUTER-XXX
Outer Column
1
L
OT-COL-CAP
Outer Column Cap
1
M
OT-FLP-SUPRT-X
Flip Top Support
1
N
OT-SPRING-S
Gas Spring
1
O
OT-INR-COLF
Inner Column
1
P
OT-SPR-PIN
Gas Spring Support Pin
1
Q
N/A – PART OF PIN
Retaining Clip
1
R
N/A
Base Plastic Cover
1
S
OT-BASE-CHNLX
U BASE
1
S
OT-BASE-HX
H BASE
1
T & U
OT-CAS-LP
Casters – Front & Back
4
www.amico.com 7
Part Numbers - Accessories
Part Number
Part Name
OT-BSK-100609
Basket 10x6x9 with Adapter
OT-BSK-090910
Basket 9x9x10 with Adapter
OT-BSK-050504
Basket 5x5x4 with Adapter
OT-BSK-ADA
Basket Adapter - fits all baskets
OT-VANITY
Vanity Drawer and Mirror
8 Amico Patient Care Corporation
_____ All fasteners are secured
_____ Table top surface is intact – not damaged
_____ Optional flip top surface is intact – not damaged
_____ Release handle is locking and releasing properly
_____ Table is moving up and down when the table top is lifted or released
_____ All casters are secure and swivelling properly
_____ No debris or dirt in the casters
_____ Base is intact – not damaged
Serial No.
__________________
__________________
__________________
__________________
__________________
__________________
__________________
__________________
__________________
Completed By: ___________________________________________ Date: __________________
CAUTION: Do not apply any kind of lubricant to the gas spring. Lubricant will damage the gas spring.
Cleaning
Amico Patient Care recommends using a mild cleaning agent or detergent when cleaning this product.
When a mild cleaning agent or soap is not adequate, follow the instructions provided by the cleaning
product manufacturer for ammonia, bleach, isopropyl alcohol and ammonium chloride based cleaning
solutions.
Normal usage of Virox 5 RTU according to the manufacturer instruction will not alter the laminate surface.
Dampen a soft cloth with diluted cleaner and wipe the surface to remove dirt and fingerprints. Wipe the
surface with a soft dry cloth to remove any moisture.
CAUTION: DO NOT USE WD-40 or any other lubricant or damage to the table could occur.
www.amico.com 9
Primary Top Replacement
Required Tools: #3 Philips Screwdriver
1. Fully raise the table top.
WARNING: To avoid personal injury or damage to the table, support the table top before removing
the screws securing it to the top support.
2. Remove six #3 Philips screws under the table top holding the top to the support bracket.
3. Lift off the table top.
4. Reverse to install the new table top.
Flip Top
Top Replacement
Required Tools: #3 Philips Screwdriver; #2 Robertson Screwdriver
1. Fully raise the table top.
WARNING: To avoid personal injury or damage to the table, support the table top before removing
the screws securing it to the top support.
2. Remove two #2 Robertson screws connecting the lid-stay to the secondary flip top surface.
3. Remove six #2 Philips screws under the table top holding the top to the support bracket.
4. Lift off the table top.
5. Remove six #2 Robertson screws from flip top side of the hinges to fully detach the dual flip top
surface.
6. Reverse to install the new table top.
Service Information
10 Amico Patient Care Corporation
Required Tools: #3 Philips Screwdriver; Slotted (Flat) Screwdriver; 5/8 Socket or Wrench
1. Remove the release handle by prying out the retaining clip and removing the handle support pin.
2. Remove the nut holding the top end of gas spring to the support bracket.
3. Remove four screws from the support bracket and lift off the support bracket.
4. Pry out the retaining clip close to the bottom of the inner column and pull out the gas spring support
pin (this may require moving the gas spring from the top edge).
5. Remove gas spring assembly from the inner column.
6. Reverse to install the new gas spring.
NOTE: When reinstalling, the new gas spring will need to be compressed slightly in order to fasten
the support bracket to the outer column by means of the four screws pertaining to step 3.
Outer Column Replacement
Required Tools: #2 Philips Screwdriver
1. Remove the primary top.
2. Slide the column all the way down and remove the locking screw on the inner column.
3. Now remove the outer column cap by removing four screws.
4. Pull the outer column away from the base.
5. Remove the glides from the inside of the column.
6. Reverse to install a new outer column.
Caster Replacement
Required Tools: 13 mm wrench
1. Fully raise the table top.
2. While maintaining a secure grip on the table, carefully flip it over or turn it on its side to allow access
to the damaged caster.
3. Use the 13 mm wrench to remove the damaged casters and replace with new casters.
www.amico.com 11
Amico Patient Care Corporation | www.amico.com
122 East Beaver Creek Road, Richmond Hill
Ontario L4B 1G6, Canada
Toll Free Tel: 1.877.462.6426
Toll Free Fax: 1.866.440.4986
Tel: 905.764.0800
Fax: 905.764.0862
Email: [email protected]
APC-MANUAL-OBT 12.02.2015