Dickson KT8P0 Getting Started
Read Online
Read the Dickson KT8P0 Getting Started online using the built-in PDF viewer.
Text of the manual
Dickson KT8P Manual V6 52418
1
DICK
DICK
DICK
DICKSON
SON
SON
SON
KT8P0/2/3 & KT856
KT8P0/2/3 & KT856
KT8P0/2/3 & KT856
KT8P0/2/3 & KT856
Remote Sensing
Remote Sensing
Remote Sensing
Remote Sensing Temperature
Temperature
Temperature
Temperature Recorder
Recorder
Recorder
Recorder
GETTING STARTED
GETTING STARTED
GETTING STARTED
GETTING STARTED
7-day chart rotation
0 to 100oF
QUICK START
QUICK START
QUICK START
QUICK START
1.
1.1.
1. Remove the protective pen cap(s).
2.
2.2.
2. Connect Probe(s):
a. KT8P0/2/3: Make sure the KTC probe is connected firmly to the Replaceable
Sensor and that the Replaceable Sensor is plugged in firmly to the chart recorder.
b. KT856: Make sure both KTC probes are plugged in firmly to the chart recorder.
3.
3.3.
3. Insert batteries and plug in AC adapter (See Figure 3). KT8P0:4 AA batteries required;
KT8P2/3 & KT856: 4 AA batteries backup power only (See page 5 for backup life). Unit
will power on.
4.
4.4.
4. The instrument will move pen(s) to the appropriate readings.
5.
5.5.
5. Install the chart that matches the Dip Switch settings (See p.4).
Press the PEN
PEN
PEN
PEN HOME
HOME
HOME
HOME key to move the pen(s) to the outside of the chart. The pen(s) is
automatically raised off the chart.
Remove the old chart, place the new chart on the Chart Hub being certain that the edge
of the chart slides under the Chart Guide Clips located at the outside of the chart.
6.
6.6.
6. Set the appropriate time. There are two ways to adjust the chart and set the appropriate
time:
a. Set the chart time manually by inserting a coin into the groove in the chart hub
and turning clockwise until the correct hour (and day if applicable) on the chart is
referenced to the timing arrow (just to the right of the pen tip on the dial). Press
Pen
Pen
Pen
Pen Home
Home
Home
Home to move the pen(s) back onto the chart.
The following feature should be used for fine adjustments only.
b. To adjust the chart time, press and hold Adjust
Adjust
Adjust
Adjust----Up
Up
Up
Up and Adjust
Adjust
Adjust
Adjust----Down
Down
Down
Down buttons
located on the back of the unit next to the dip switch (see Figure 3).
The green LED will blink rapidly for about five seconds, then the LED will remain
solid green. While in this state the Adjust
Adjust
Adjust
Adjust----Up
Up
Up
Up button will move the chart
backward (counter clockwise) and the Adjust
Adjust
Adjust
Adjust----Down
Down
Down
Down button will move the chart
forward (clockwise).
Rotate the chart until the correct hour (and day if applicable) on the chart is
referenced to the timing arrow. Once you have set your chart, press the Pen
Pen
Pen
Pen
Home
Home
Home
Home button to exit Chart Adjust Mode. The unit will take one minute to exit the
Chart Adjust Mode once Pen
Pen
Pen
Pen Home
Home
Home
Home is pressed. Press Pen
Pen
Pen
Pen Home
Home
Home
Home to move the
pen(s) back onto the chart.
c.
Place the KT8 recorder on a flat vibration-free surface. Be sure it is in a vertical
position and level. For best performance and longevity, the location should be a
clean environment, free from dust and corrosive fumes. Do not exceed
temperature specifications. Wall Mount:
Wall Mount:
Wall Mount:
Wall Mount: Keyhole slots are provided on the KT8
recorder for wall mounting.
Dickson KT8P Manual V6 52418
2
DISPLAY SYMBOLS
DISPLAY SYMBOLS
DISPLAY SYMBOLS
DISPLAY SYMBOLS KT8
KT8
KT8
KT8PPPP
BUTTON OPERATION
BUTTON OPERATION
BUTTON OPERATION
BUTTON OPERATION
On/Off
On/Off
On/Off
On/Off
The On/Off key turns the unit on and off.
Pen
Pen
Pen
Pen Home (Down Arrow)
Home (Down Arrow)
Home (Down Arrow)
Home (Down Arrow)
If the pen(s) are located on the outside edge of the chart, press the Pen
Pen
Pen
Pen Home
Home
Home
Home to move pen(s) to
recording position. If the pen(s) are located on the chart, press the Pen
Pen
Pen
Pen Home
Home
Home
Home key to move the pen(s) to
the outside edge of the chart.
AAAAlarm
larm
larm
larm ((((Optional
Optional
Optional
Optional))))
Pen 1: = Red Pen
Pen 2: = Blue Pen
F
Channel Indicator
Low Battery Indicator
Unit of Measure
for channel
displayed: F, C
Current Reading or Flashing
“ALRN” when in alarm
condition.
Displays when
setting alarms.
Dickson KT8P Manual V6 52418
3
1.
1.1.
1. To set the alarm, make sure unit is on and press and hold the Alarm
Alarm
Alarm
Alarm button. The LED will blink red
rapidly for about five seconds, then the LED will turn solid green. Release the Alarm
Alarm
Alarm
Alarm button and the
LED will turn solid red. At this point the display will show “On” or “Off”.
Pressing either the Adjust
Adjust
Adjust
Adjust----Up
Up
Up
Up or Adjust
Adjust
Adjust
Adjust----Down
Down
Down
Down buttons, located on the back of the unit next to the dip
switch, will toggle the alarm on or off.
2.
2.2.
2. Pressing PEN
PEN
PEN
PEN HOME
HOME
HOME
HOME will scroll to the next alarm options.
Pen 1 alarm minimum
Pen 1 alarm maximum
If KT856 two pen model, the additional options are:
Pen 2 alarm minimum
Pen 2 alarm maximum
3.
3.3.
3. In order to set pen(s) alarm minimums and maximums, pressing Adjust
Adjust
Adjust
Adjust----Up
Up
Up
Up will increase the alarm
value, and pressing Adjust
Adjust
Adjust
Adjust----Dow
Dow
Dow
Downnnn will decrease the alarm value. There is acceleration if the Adjust
Adjust
Adjust
Adjust----Up
Up
Up
Up
button is held down.
Repetitively pressing the Pen
Pen
Pen
Pen Home
Home
Home
Home button will scroll through the five options until the AAAAlarm
larm
larm
larm button
is pressed to exit from alarm adjust. Each press of the Pe
Pe
Pe
Pennnn Home
Home
Home
Home or Alarm
Alarm
Alarm
Alarm button will store the new
settings. The unit will take one minute to exit Alarm Set Mode once the Alarm
Alarm
Alarm
Alarm button is pressed.
4.
4.4.
4. If the alarm is triggered, the LED will show as solid red and the alarm will sound. Press the Alarm
Alarm
Alarm
Alarm
button to silence the audible alarm.
Alarm
Alarm
Alarm
Alarm Relays
Relays
Relays
Relays
The SPST 24V 500mA relay contacts are Normally open and will close on alarm conditions when the
alarms are enabled. Relay contacts are always functional when the alarm is enabled. The relay will close
only during minimum and maximum alarm conditions. The KT856 has 2 sets of relay contacts which
operate independently. Relays will not operate under battery only power conditions.
PENS
PENS
PENS
PENS
The blue pen has a longer pen arm and records channel 2 temperature (depending on your dip switch
selection, see pg. 4). The red pen has a shorter arm and records channel 1 temperature. The pens are
offset to allow the red pen to glide under the blue pen. The blue pen indicates the correct time and the
red pen precedes it by 3/16 of an inch.
Dickson KT8P Manual V6 52418
4
DIP SWITCH SETUP
DIP SWITCH SETUP
DIP SWITCH SETUP
DIP SWITCH SETUP
To setup the KT8 recorder for your specific
application, you might need to change some of
the Dip Switches. The Dip Switches are located
on the back of the unit. (Figure 3)
Use a pen or small screw driver to flip the
switches. Remember to install the correct chart
to match the corresponding Dip Switch setting.
Dip Switches for Ranges
Dip Switches for Ranges
Dip Switches for Ranges
Dip Switches for Ranges::::
Reording Time
Reording Time
Reording Time
Reording Time
#1 down
#2 up
#1 down
#2 down
#1 up
#2 up
oF
#3 down
oC
#3 up
Temperature Ranges
Temperature Ranges
Temperature Ranges
Temperature Ranges
#4 down
#5 down
#6 down
#4 down
#5 down
#6 up
#4 down
#5 up
#6 up
#4 up
#5 down
#6 down
#4 down
#5 up
#6 down
#4 up
#5 up
#6 up
#4 up
#5 down
#6 up
#4 up
#5 up
#6 down
All the KT8 recorders feature usre selectable recording
times.
Note: Remember to install the correct chart to match
corresponding switch setting.
You can record in oF or oC with the KT8 by using dip
switch #3.
ooooF/C
F/C
F/C
F/C
24-hour all models
7-day all models
31-day KT8P2/3 & KT856
0 to +2000oF/C*
* A203 High Temp Probe required - Range not available
on Model KT800
0 to +100oF/C
-50 to +50oF/C
0 to +250oF/C
0 to +500oF/C
0 to +150oF/C
-20 to +20oF/C
0 to +1000oF/C*
Dickson KT8P Manual V6 52418
5
POWER
POWER
POWER
POWER
The KT8 operates on AC power with an optional 4 AA battery backup. Depending on chart
rotation selected and model; battery backup will last
KT8
KT8
KT8
KT8PPPP2/3 & KT856
2/3 & KT856
2/3 & KT856
2/3 & KT856
24 Hour Chart Rotation = 1 Day Backup
7 Day Chart Rotation = 2 Day Backup
31 Day Chart Rotation = 1 Week Backup
Note: Alarm and relays will not operate under battery only power.
KT8
KT8
KT8
KT8PPPP0000
At 70oF
24 Hour Chart Rotation = 1 Month
7 Day Chart Rotation = 6 Months
At 32oF
24 Hour Chart Rotation = 3 Weeks
7 Day Chart Rotation = 4 Months
Note: (Unit should not be placed in an environment below 32oF)
LED Indicators
LED Indicators
LED Indicators
LED Indicators
• AC Power with Battery Backup - Solid Green
• AC Power with Low Battery or No Battery - Blinks Red
• Battery Only - Blinks Green
• Battery Only (Low Battery) - Solid Red
Note: KT8P0 operates on battery only and will blink one time every minute to conserve
battery life.
PEN HOME ADJUST
PEN HOME ADJUST
PEN HOME ADJUST
PEN HOME ADJUST
If over time the pen location does not match the display, you may need to adjust the pen location
on the chart. A Pen Home Adjustment will not affect calibration of the sensor, but rather corrects
for the natural drift that occurs with mechanical moving parts over time.
1. While the unit is on, press and hold both Pen Home
Pen Home
Pen Home
Pen Home and On/Off
On/Off
On/Off
On/Off buttons until the LED is
solid green. Release the Pen Home
Pen Home
Pen Home
Pen Home and On/Off
On/Off
On/Off
On/Off buttons. The LED will flash amber and
green for one second then the LED will turn off.
Note:
Note:
Note:
Note: Press Pen Home
Pen Home
Pen Home
Pen Home first so the unit will not turn off
2. The pen(s) will move to the outer edge of the chart, and then the shorter red
red
red
red pen will
move to the outer chart ring (maximum temperature/RH line).
3. If the red pen tip does not line up with the outer chart ring, use the Adjust
Adjust
Adjust
Adjust----Up
Up
Up
Up and Adjust
Adjust
Adjust
Adjust----
Down
Down
Down
Down buttons (on the back of the unit) to move the red pen so that the pen tip is on top
of the outer chart ring.
4. Once set press On/Off
On/Off
On/Off
On/Off to save and exit Pen Home Adjust. If using KT856 continue to step
5 below.
5. Once the shorter red pen is set, press Pen Home
Pen Home
Pen Home
Pen Home to move the shorter red pen to the outer
edge of the chart. The longer blue pen will move to the out chart ring.
6. If the blue pen tip does not line up with the out chart ring, use the Adjust
Adjust
Adjust
Adjust----Up
Up
Up
Up and Adju
Adju
Adju
Adjustststst----
Down
Down
Down
Down buttons to move the blue pen so that the pen tip is on top of the outer ring.
7. Once the blue pen is set, pressing the Pen Home
Pen Home
Pen Home
Pen Home button will move the longer blue pen to
the outer edge of the chart and the shorter red pen will move to the outer chart ring.
Dickson KT8P Manual V6 52418
6
8. Each time PPPPen
en
en
en HHHHome
ome
ome
ome is pressed the two pens will switch positions (moving the most
recently adjusted position) until On/Off
On/Off
On/Off
On/Off is pressed to exit pen adjust mode.
Note: Pressing the On/Off button will exit the adjustment and return the unit to normal operation.
Each time the Pen Home or On/Off button is pressed, the adjustment currently in progress will be
stored. The unit will take one minute to exit Pen Adjust Mode once the On/Off button is pressed.
CALIBRATION
CALIBRATION
CALIBRATION
CALIBRATION
KT8P0/2/3:
The R400 Replaceable Sensor was carefully tested and calibrated before being shipped
from the factory. For greatest accuracy, we recommend replacing the sensor every 6-12
months (does not include N400 and N450).
To order a Calibrated Replacement Sensor Call customer service at (630) 543-3747 or go
to www.dicksondata.com.
When your new Replaceable Sensor arrives, simply turn off the recorder, remove and
discard the old sensor, plug in the new one and power the recorder back on. Your
recorder will continue to record temperature and humidity without interruption.
KT856:
The KT856 was carefully tested and calibrated before being shipped from the factory. For
greatest accuracy, we recommend factory re-calibration every 6-12 months. Call
customer service at (630) 543.3747.
USER CALIBRATION
USER CALIBRATION
USER CALIBRATION
USER CALIBRATION
If you have an accurate standard to compare against, the Replaceable Sensor or unit (KT856)
calibration can be adjusted at one point. This will not adjust the span and is not as accurate as
replacing the Replaceable Sensor with a newly calibrated one, or returning the KT856 for
recalibration.
1. To activate Calibration mode, turn the unit on and press and hold both the On/Off
On/Off
On/Off
On/Off and the
Adjust
Adjust
Adjust
Adjust----Down
Down
Down
Down button until the LED is solid green. The LED will then blink amber at which
point only the pen being adjusted will show on the display.
Note:
Note:
Note:
Note: Make sure to press the Adjust-Down button first so the unit will not turn off
2. To raise the unit of measurement, press the Adjust
Adjust
Adjust
Adjust----Down
Down
Down
Down button. To lower the unit of
measurement, press the Adjust
Adjust
Adjust
Adjust----Up
Up
Up
Up button. Pressing Pen home
Pen home
Pen home
Pen home switches between the red and
blue pens on a two pen unit and stores the current adjustment value.
3. When Calibration is complete, press the On/Off
On/Off
On/Off
On/Off button. The adjustment is stored in memory
even after you turn the unit off or if AC power fails.
Note:
Note:
Note:
Note: After two hours, if no buttons are pressed, the unit will time out of Calibration mode
and resume normal operation. If you wish to cancel the Calibration, simply enter Calibration
mode and toggle through the steps without adjusting displayed readings. Exit by pressing
the On/Off
On/Off
On/Off
On/Off button. You have now restored factory calibration settings.
Note:
Note:
Note:
Note: It is recommended that you use a controlled chamber when determining if Calibration
is necessary. Adjusting the pen locations in an open room is not recommended as humidity
can vary greatly within a very small area.
Dickson KT8P Manual V6 52418
7
NEED TO KNOW
NEED TO KNOW
NEED TO KNOW
NEED TO KNOW
The pen movement for the KT8P0 is slower than on the AC powered models in order to conserve
battery power. In 24-hour mode the pen(s) will move every 1 minute and 30 seconds. In 7-day
mode the pen(s) will move every 8 minutes and 30 seconds.
The recorder takes temperature readings every 50 seconds. An average of the reading taken
between pen(s) movements is what is drawn on the chart.
TROUBLE SHOOTING
TROUBLE SHOOTING
TROUBLE SHOOTING
TROUBLE SHOOTING
Why isn't the chart keeping time or running slow?
Why isn't the chart keeping time or running slow?
Why isn't the chart keeping time or running slow?
Why isn't the chart keeping time or running slow?
•
The chart may be hung up or restricted, possibly caused by a rip on the outer edge of the
chart or the chart hub, or the chart may be caught between the arm and pen(s) arm
platform.
•
Incorrect chart installed for the selected chart speed.
Why did the chart stop tu
Why did the chart stop tu
Why did the chart stop tu
Why did the chart stop turning?
rning?
rning?
rning?
•
Chart hung up or restricted, (ripped chart)
•
Unit may be "locked up", this can be confirmed by pressing any of the buttons on the
keypad, if the unit is locked up there will be no response to button presses and the Chart
Recorder may appear to be working, but the readings won't change, also the chart will
not rotate. Remove power and battery, and then re-power.
Why don't the display and chart match?
Why don't the display and chart match?
Why don't the display and chart match?
Why don't the display and chart match?
•
Dip switches set for a specific range, but using chart for another range, or vice versa.
•
Pen(s) not inserted on pen arm(s) all the way
•
To adjust pen(s) to match chart see “Pen Home Adjust” in the “Need to Know” section of
this manual.
Why does the display read PROB?
Why does the display read PROB?
Why does the display read PROB?
Why does the display read PROB?
•
Check the sensor pod and KTC probe(s). Make sure it is properly plugged into the unit.
•
Is the unit locked up? This can be confirmed by pressing any of the buttons on the
keypad, if the unit is locked up there will be no response to button presses and the unit
may appear to be working, but the readings won't change, also the chart will not rotate.
Remove power and battery, and then re-power.
Why does calibration seem to be off?
Why does calibration seem to be off?
Why does calibration seem to be off?
Why does calibration seem to be off?
•
What is the tolerance of the unit it's being compared to?
•
It's ok if the unit is within the sum of the two tolerances.
•
Has an outside calibration house attempted calibration? It may not have been adjusted
properly.
•
Walk through calibration adjustment, found in the manual
Why won't the battery back
Why won't the battery back
Why won't the battery back
Why won't the battery back----up work?
up work?
up work?
up work?
•
Are there good batteries in the Chart Recorder?
•
Keep in mind that battery backup varies significantly depending on temperature, pen
movement and chart rotation speed.
•
If there was a momentary power outage (Brown out), the unit may not have had enough
time to recognize this and switch to battery mode. (The Chart Recorder may lock up or
shut off in this situation) This situation may also occur if the unit is plugged into an outlet
that shares the circuit with other machinery that has phase motors or compressors that
cycle periodically. As these other machines cycle, they momentarily have a high current
draw, therefore drawing from the unit.
Dickson KT8P Manual V6 52418
8
Why won't the chart recorder respond to changes?
Why won't the chart recorder respond to changes?
Why won't the chart recorder respond to changes?
Why won't the chart recorder respond to changes?
•
Is the unit locked up? This can be confirmed by pressing any of the buttons on the
keypad, if the unit is locked up there will be no response to button presses and the unit
may appear to be working, but the readings won't change, also the chart will not rotate.
Remove power and battery, and then re-power.
Why won't the unit power up?
Why won't the unit power up?
Why won't the unit power up?
Why won't the unit power up?
•
Remove the batteries and power adapter for a minute or two, this will reset the unit. The
unit should respond when the adapter is plugged in.
WARRANTY
WARRANTY
WARRANTY
WARRANTY
Dickson warrants that this line of instruments will be free from defects in material and
workmanship under normal use and service for a period of twelve months after delivery.
This warranty does not cover routine calibration and battery replacement.
For Specifications and Technical Support go to www.DicksonData.com
FACTORY SERVICE AND RETURNS
FACTORY SERVICE AND RETURNS
FACTORY SERVICE AND RETURNS
FACTORY SERVICE AND RETURNS
Contact Customer Service (630.543.3747) for a Return Authorization (RA) Number before
returning any instrument. Please have the model number, serial number and a PO number ready
before calling.
Replaceable Sensor Certificate of Validation
Replaceable Sensor Certificate of Validation
Replaceable Sensor Certificate of Validation
Replaceable Sensor Certificate of Validation
Dickson ensures that the Replaceable Sensors / Instruments listed below were developed, tested
and validated together.
Replaceable Sensor Model: R400
Instrument Model: KT8P0, KT8P2, KT8P3
Operation
Operation
Operation
Operation: The Instruments and Replaceable Sensor models listed above were specifically
designed to work together. No other sensors will work with the Instrument models listed. No
other Instrument will work with the Replaceable Sensors listed. The Replaceable Sensor must be
connected to the Instrument for the Instrument to operate properly.
Calibration
Calibration
Calibration
Calibration: Only the Replaceable Sensor is calibrated. The sensor, and all calibration defaults
and adjustments are stored on the Replaceable Sensor. Accuracy is strictly controlled by the
sensor – no adjustments are made by the Instrument. Readings are sent from the Replaceable
Sensor to the Instrument for storage purposes only. Certificates of Calibration / NISTs are only
supplied for the Replaceable Sensor for this reason.
Recalibration
Recalibration
Recalibration
Recalibration: When a Replaceable Sensor is due for recalibration the existing Replaceable Sensor
can be replaced with a new Replaceable Sensor thus eliminating the need to return the entire
instrument for recalibration. Replaceable Sensors can be returned for recalibration as well.