DeltaOHM HD2013 Operating Manual
Read Online
Read the DeltaOHM HD2013 Operating Manual online using the built-in PDF viewer.
Text of the manual
Tipping bucket rain gauge
HD2013
www.deltaohm.com
English
Keep for future reference.
Companies / Brands of GHM
- 2 -
V1.0
TABLE OF CONTENTS
1
INTRODUCTION.................................................................................................... 3
2
TECHNICAL CHARACTERISTICS ............................................................................ 4
3
INSTALLATION ..................................................................................................... 6
3.1
ELECTRICAL CONNECTIONS ..................................................................................... 6
3.2
CONFIGURATION OF ANALOG OUTPUT ......................................................................... 9
3.3
TIPPING BUCKET................................................................................................ 10
3.4
INSTALLATION MODES ......................................................................................... 11
4
SERIAL COMMANDS............................................................................................ 14
5
SDI-12 PROTOCOL.............................................................................................. 20
6
MAINTENANCE.................................................................................................... 27
7
INSTRUMENT STORAGE ...................................................................................... 29
8
SAFETY INSTRUCTIONS...................................................................................... 29
9
ORDERING CODES .............................................................................................. 30
- 3 -
V1.0
1 INTRODUCTION
The HD2013 is a reliable and sturdy tipping bucket rain gauge, built entirely from cor-
rosion resistant materials in order to guarantee its durability. So as to ensure accurate
measurement even with low temperature climatic conditions or during and after pre-
cipitations of snow, a version with heating which is automatically activated around
+4 °C has been developed so that snow deposits and ice formations are prevented.
The rain gauge is formed by a metal base on which a tipping bucket is set. The rain
collector cone, fixed to the aluminium cylinder, channels the water inside the tipping
bucket: once the predefined level is reached, the calibrated bucked rotates under the
action of its own weight, discharging the water. During the rotation phase, the nor-
mally closed reed contact opens for a fraction of a second, sending an impulse to the
counter.
The quantity of rainfall measured is based on the count of the number of times the
bucket is emptied: the reed contacts, normally closed, open at the moment of the rota-
tion between one bucket’s section and the other. The number of impulses can be de-
tected and recorded by a data logger such as the HD2013-DB Delta OHM or by a
pulse counter.
The rain gauge is available with optional analog output (HD2013UA) at choice be-
tween 4…20 mA or 0…10 Vdc (to be specified when ordering) or with optional
SDI-12 digital output (HD2013UD).
The version with 4…20 mA output and the version with SDI-12 output require 7…30 Vdc
power supply. The version with 0…10 Vdc output requires 13…30 Vdc power supply. The
versions with analog or SDI-12 output are equipped with a backup battery that allows
maintaining the rainfall information even after short power supply outages (at least 10
seconds for HD2013UA, approx. 5 minutes for HD2013UD).
A removable filter for periodic cleaning and maintenance is inserted in the water col-
lector cone so as to prevent leaves or other elements blocking the end of the hole.
For a better water flow, the collecting cone is treated with a non-adherent product.
In the version with heating HD2013R, the heating circuit operates with 12 Vdc or
24 Vdc direct voltage (to be specified when ordering) and absorbs about 165W.
Heating is activated around +4 °C.
When submitting your order, upon request a bird dissuader, made of eight 3 mm
diameter spikes, 60 mm in height, can be installed on the rain gauge.
- 4 -
V1.0
2 TECHNICAL CHARACTERISTICS
HD2013R
HD2013
HD2013UA
analog output
option
HD2013UD
SDI-12 output
option
Power supply
Heating circuit:
12 or 24 Vdc ± 10%
(to be specified when
ordering)
---
7…30 Vdc
(4…20 mA output)
13…30 Vdc
(0…10 V output)
7…30 Vdc
Consumption
165 W
---
26 mA max. @ 18 Vdc
(4…20 mA version)
≈ 4 mA @ 18 Vdc
(0…10 V version)
≈ 100 µA @ 18 Vdc
in SDI-12 mode
Contact output
NC contact (opens during tipping).
With HD2013UA and HD2013UD options, the contact output is alternative to
the analog or SDI-12 output and the rain gauge is factory-configured
with analog (HD2013UA) or SDI-12 (HD2013UD) output.
Analog output
---
---
4…20 mA or 0…10 V
(to be specified
when ordering)
---
Digital output
---
---
---
SDI-12
Resolution
0.1 – 0.2 or 0.5 mm/tip
on request when ordering
Accuracy
+3.5…-2.5% in the interval 0…100 mm/h
(version with 0.2 mm @ 50 mm/h nominal resolution)
+ 2…-3,5% in the interval 0…200 mm/h
(version with 0.5 mm @ 50 mm/h nominal resolution)
The error refers to the calculation of the amount of rain using the resolution
stated in the rain gauge label.
If the amount of rain is calculated using the correction curves as a function of
the rainfall rate (fig. 1 and 2), the error is typically less than ± 2% in the in-
terval 0…200 mm/h.
If the HD2013-DB data logger is used, the measurement can be automatically
corrected according to the graphs in fig. 1 and 2.
With HD2013UA and HD2013UD options, the curve can be stored in the rain
gauge itself (on request when ordering).
Maximum rainfall rate
600 mm/h (version with 0.2 mm nominal resolution)
1000 mm/h (version with 0.5 mm nominal resolution)
Operating temperature
-20…+70 °C
+4…+70 °C
Heating intervention tem-
perature
+4 °C
---
Protection degree
IP 64
Collector area
400 cm2
Minimum section of the
wires of the connecting
cable
0.5 mm2 for the version without heating (HD2013)
2.5 mm2 for the version with heating (HD2013R)
- 5 -
V1.0
Fig. 1 – Normalized resolution (0.2 mm @ 50 mm/h) as a function of the rainfall rate
Fig. 2 – Normalized resolution (0.5 mm @ 50 mm/h) as a function of the rainfall rate
To correct the measurement depending on the rainfall rate, it is necessary to log, in
addition to the number of pulses, also the instants at which the pulses occur.
Example of measurement correction:
Let’s assume that a rain gauge with nominal resolution RN = 0.209 mm @ 50 mm/h
has generated 25 pulses at the frequency of 1 pulse every 50 seconds.
The rainfall rate can be estimated considering the nominal resolution RN and the inter-
val between two successive pulses: I = 0.209 x 3600 / 50 ≈ 15 mm/h.
From the graph in fig.1 we obtain the normalized corrected resolution: RCn = 0.1944 mm.
The corrected resolution of the rain gauge is: RC = RCn x RN / 0.2 = 0.2031 mm.
The amount of rain detected is 25 x 0.2031 = 5.07 mm.
Rainfall rate (mm/h)
Normalized resolution (mm)
Rainfall rate (mm/h)
Normalized resolution (mm)
- 6 -
V1.0
3 INSTALLATION
Upon request the rain gauge can be supplied already calibrated at 0.1 - 0.2 or 0.5 mm
of rain per tip of the bucket: the calibration value is shown on the instrument label.
The instrument must be installed in an open area, away from buildings, trees, etc.,
ensuring the space over it is free from all objects which could obstruct rain measure-
ments, and in an easily accessible position for the filter to be cleaned periodically.
Avoid installation in areas exposed to gusts of wind, turbulence (for example the top
of a hill) as these may distort the measurements.
The rain gauge can be installed on the ground or raised 500 mm above the ground.
Other sizes above the ground are available on request.
For ground installation, three adjustable support feet, so that the instrument can be
levelled correctly, and holes for fixing to the floor have been provided.
For raised installations, a collar to be fastened to the base of the instrument, and on
which the support staff must be inserted, is provided. The staff may end with either a
flange so that it can be fixed to the floor, or a tip to be driven into the ground. The
various fastening systems can be seen in the figures 6, 7, 8 and 9.
For the tipping device to function correctly and so for the measurement to be correct,
it is important that the instrument is placed perfectly levelled. The base of the rain
gauge is fitted with a bubble level.
For installation, unscrew the three screws at the base of the cylinder that supports the
water collector cone.
Attention: a heating resistor is fitted around the cone in the HD2013R version. To
disconnect the cone heater, unplug the connector on the upper side of the terminal
block protection cover (see fig.3).
3.1 ELECTRICAL CONNECTIONS
For the version without heating use a 2-wire cable with 0.5 mm2 minimum wires sec-
tion, for the version with heating use a 4-wire cable with 2.5 mm2 minimum wires
section.
Use a shielded cable over long distances. Slide the cable through the cable gland and
fasten it with the cable fastener located near the entry hole at the base of the rain
gauge.
Unscrew the terminal block protection cover and perform the connections as shown in
the following figures.
Connection of heating:
The heated version requires power (12 Vdc or 24 Vdc depending on the version sup-
plied) for the resistors: perform the connection as indicated at point 2 of fig.3.
If the connections are set correctly, the LED placed near the terminals will be lit up.
Connection of contact output:
The rain gauge contact output, indicated at point 1 in the figure 3, must be con-
nected to the input of the HD2013-DB rain gauge data logger (please see the
details in the manual of the data logger) or to a pulse counter or to a data logger.
To use the contact output in the versions with HD2013UA or HD2013UD option, set
- 7 -
V1.0
the SW1 switch down, towards the EXTERNAL indication (see the figures 4 and 5): the
analog or SDI-12 output is disabled.
Fig. 3 – Connections of contact output and heating
Connection of analog output (only with HD2013UA option):
Perform the connections as shown in figure 4. Set the SW1 switch up, towards the IN-
TERNAL indication (the contact output is disabled).
Fig. 4 – Connections of analog output
Connector of the
cone heater
RESET line
GND
Analog output positive
Power supply positive
SW1 switch
- 8 -
V1.0
The current analog output is active. The maximum value of the load resistance de-
pends on the power supply value:
RL max = 50 x (Vdc – 7) Ω
The minimum value of the load resistance for the version with voltage analog output is
10 KΩ.
In case of a measurement error, the analog output goes to the value 22 mA or 11 V.
The reset line allows the partial amount of rainfall measurement (amount of rainfall from
the last reset command) to be remotely reset. During the measurement, the reset line
must be connected to GND. To reset, disconnect the line from GND and apply a direct
voltage of at least 3 Vdc (and lower than or equal to the power supply) for at least 1 s,
then reconnect the line to GND.
Connection of SDI-12 digital output (only with HD2013UD option):
Perform the connections as shown in figure 5. Set the SW1 switch up, towards the IN-
TERNAL indication (the contact output is disabled).
Fig. 5 – Connections of SDI-12 digital output
The reset line is not used in the version with digital output: the reset is performed with
the appropriate SDI-12 extended command (aRES!, with a=address of the rain gauge).
Not connected
GND
SDI-12 output positive
Power supply positive
SW1 switch
To SDI-12 bus
- 9 -
V1.0
3.2 CONFIGURATION OF ANALOG OUTPUT
The analog output is proportional to the partial amount of rainfall (amount of rainfall
from the last reset command). The analog output full scale (20 mA or 10 V) can be as-
sociated with different values of rainfall. The association can be implemented via hardware,
by using the dip switches on the circuit board, or via software, by connecting the serial
communication port COM PC of the rain gauge to the PC.
The dip switches allow setting three rainfall values: 25, 100 or 250 mm (1, 5 or 10
inches if the set unit of measurement is “inches”), as shown in figure 6.
If both the dip switches are set to ON, the rainfall value considered is the one set via
serial port with the command CRW (see the table of the serial commands).
Fig. 6 – Dip switches
Warning: after reaching the full scale value, the measurement is reset and
the analog output returns to the initial scale value.
COM PC serial port
- 10 -
V1.0
3.3 TIPPING BUCKET
The tipping bucket is locked for the transport of the rain gauge. To unlock the bucket,
remove the holders as shown in fig.7.
Fig. 7 – Locking and unlocking the tipping bucket
The oscillation of the tipping bucket can be adjusted through the two threaded rods
located under the bucket, as illustrated in fig.8.
Fig. 8 – Adjustment of the tipping bucket
For the
transport
For the
operating
mode
- 11 -
V1.0
3.4 INSTALLATION MODES
Fig. 9 – Mechanical dimensions, installation modes
Fig. 10 – Base for ground fastening (HD2003.78)
- 14 -
V1.0
4 SERIAL COMMANDS
The rain gauges with HD2013UA or HD2013UD option allow editing or checking the
operating parameters of the instrument via serial commands. Proceed as follows:
• Connect the serial port COM PC of the rain gauge to the RS232 (via the RS27 ca-
ble) or USB (via the CP27 cable) port of the PC. If the CP27 cable is used, install
the corresponding USB drivers in your PC.
• In the PC, start a serial communication program (e.g. Hyperterminal), set the
baud rate to 9600 and the communication parameters to 8N2.
• Send the commands given below to edit or check the value of the rain gauge pa-
rameters. In the editing commands, the new value of the parameter follows imme-
diately (without spaces interposed) the command identifier.
The editing of the rain gauge parameters requires sending a password via the serial
command PWD. By default, the password is not set; therefore, it is sufficient to send the
command PWD without password to change the parameters. To protect the parameters
from unauthorized changes, it is advisable to set a password with the serial command
SUP.
The password remains active for 10 minutes, after which you must resubmit it.
In the serial commands described below, the column “PWD” indicates whether the exe-
cution of the command is password protected.
Password management
Command
Description
PWD
PWDx…x
Sends the password x…x (8 characters max.).
Reply of the instrument:
USER ACCESS OK if the password is accepted
WRONG PASSWORD if the password is not accepted
---
SUPx…x
Sets the string x…x (8 ASCII characters max.) as password.
Reply of the instrument:
PASS: x…x if the password is accepted
INVALID PASSWORD if the password is not accepted
Yes
Date and time
Command
Description
PWD
DSyyyy/mm/dd hh:mm:ss
Sets the date yyyy/mm/dd (year/month/day) and the
time hh:mm:ss (hour/minutes/seconds) in the rain gauge.
Reply of the instrument:
&| if date and time are accepted
? if date and time are not correct
Yes
DG
Reads the date (year/month/day) and the time
(hour/minutes/seconds) set in the rain gauge.
Reply of the instrument: yyyy/mm/dd hh:mm:ss
No
- 15 -
V1.0
General info
Command
Description
PWD
IR
Reads the manufacturer, the model, the firmware version and the serial
number of the rain gauge.
Reply of the instrument: v…vm…mf…fs…s, with:
v…v (8 characters) = name of the manufacturer (DELTAOHM)
m…m (6 characters) = model (HD2013)
f…f (3 characters) = firmware version
s…s (max. 13 characters) = serial number
No
Unit of measurement
Command
Description
PWD
UWn
Sets the unit of measurement of index n for the amount of rainfall:
n=0 ⇒ mm, n=1 ⇒ inches
Reply of the instrument:
&| if the unit of measurement is accepted
? if the index n is not correct
Yes
UR
Reads the unit of measurement of the amount of rainfall set in the rain
gauge.
Reply of the instrument:
0 if the unit of measurement is “mm”
1 if the unit of measurement is “inches”
No
TUWu
Sets the temperature unit of measurement (*): u=C ⇒ °C, u=F ⇒ °F
Reply of the instrument:
&| if the unit of measurement is accepted
? if the character u is not correct
Yes
TUR
Reads the temperature unit of measurement (*).
Reply of the instrument:
C if the unit of measurement is “°C”
F if the unit of measurement is “°F”
No
(*) The rain gauges with HD2013UA or HD2013UD option are equipped with a temperature
sensor which detects the internal temperature of the instrument (not the air temperature).
The temperature value can be read via the serial commands S1 / S2 or, in the rain gauges
with HD2013UD option, via the appropriate SDI-12 command described in the paragraph
“SDI-12 protocol”.
- 16 -
V1.0
Resolution
Command
Description
PWD
BWn
Sets the rain gauge resolution (amount of rainfall corresponding to an
emptying of the tipping bucket) of index n:
n=0 ⇒ 0.1 mm
n=1 ⇒ 0.2 mm
n=2 ⇒ 0.25 mm
n=3 ⇒ 0.5 mm
n=4 ⇒ 1.0 mm
n=5 ⇒ 0.254 mm (0.01 inches)
n=6 ⇒ 0.508 mm (0.02 inches)
n=7 ⇒ 1.016 mm (0.04 inches)
n=X ⇒ custom value (set via the CCW command)
Reply of the instrument:
&| if the resolution is accepted
? if the index n is not correct
Yes
BR
Reads the rain gauge resolution.
The reply of the instrument is the index n of the resolution according to
the list given in the description of the editing command BW.
No
CCWn…n
Sets n…n as custom value of the resolution. The value must be between
0.001 and 10.00 and is considered in the unit of measurement (mm or
inches) set in the instrument.
Example: CCW+0.42 sets 0.42 mm or 0.42 inches, depending on the
unit of measurement set, as custom value of the resolution.
Reply of the instrument:
&| if the value n…n is accepted
? if the value n…n is not correct
Warning: the command CCW sets the value but not the use of the cus-
tom resolution; to use the cutom resolution, the command BWX shoul
be sent (see the command BW described above).
Yes
CCR
Reads the custom value of the resolution. The value is considered in the
unit of measurement (mm or inches) set in the instrument.
No
Output contact type
Command
Description
PWD
TNWn
Sets the rain gauge output contact as normally open or normally closed:
n=0 ⇒ Normally OPEN, n=1 ⇒ Normally CLOSED
Reply of the instrument:
&| if the setting is accepted
? if the index n is not correct
Yes
TNR
Reads the type of the rain gauge output contact.
Reply of the instrument:
0 if the contact is normally OPEN
1 if the contact is normally CLOSED
No
- 17 -
V1.0
Analog output (only with HD2013UA option)
Command
Description
PWD
CPOE
Enables the offset of the analog output initial value: 4…20 mA or 2…10 V.
Reply of the instrument: &|
Yes
CPOD
Disables the offset of the analog output initial value: 0…20 mA or
0…10 V.
Reply of the instrument: &|
Yes
CPOR
Reads the enabling state of the offset of the analog output initial value.
Reply of the instrument:
0 if the offset is disabled (0…20 mA or 0…10 V)
1 if the offset is enabled (4…20 mA or 2…10 V)
No
CRWn…n
Sets n…n as the amount of rainfall corresponding to the analog output
full scale. The value must be between 0.01 and 9999.00 and is consid-
ered in the unit of measurement (mm or inches) set in the instrument.
Example: CRW+80.00 sets 80 mm or 80 inches, depending on the unit
of measurement set, as value corresponding to 20 mA or 10 V.
Reply of the instrument:
&| if the value n…n is accepted
? if the value n…n is not correct
Yes
CRR
Reads the amount of rainfall corresponding to the analog output full
scale. The value is considered in the unit of measurement (mm or inch-
es) set in the instrument.
No
Printing of the measurements (only with HD2013UA option)
Command
Description
PWD
S0
Stops the continuous printing of the measurements.
Reply of the instrument: &|
No
S1
Single printing of the measurement values, considered in the unit of
measurement set in the instrument.
Reply of the instrument:+ M1+M2+M3+M4, with:
M1 = total amount of rainfall (from when the instrument is powered)
M2 = partial amount of rainfall (from the last reset operation)
M3 = amount of rainfall in the current day
M4 = amount of rainfall in the previous day
Example: the reply +873.4+15.8+2.6+3.4 indicates that the amount of
rainfall is:
873,4 mm or inches from when the instrument is powered
15,8 mm or inches from the last reset operation
2,6 mm or inches today
3,4 mm or inches yesterday
Note: after reaching 99999.99, the total rainfall counter restarts from
zero.
No
S2
Continuous printing (once per second) of the measurement values, con-
sidered in the unit if measurement set in the instrument.
The reply of the instrument is in the same form described for the com-
mand S1.
No
- 18 -
V1.0
Reset
Command
Description
PWD
RES
Resets the partial rainfall counter.
Reply of the instrument: &|
Yes
HRS
Resets the electronic board (with consequent reset of all the measur-
ing counters).
Yes
Correction of the measurement
To increase the measurement accuracy, a measurement correction table as a function
of the rainfall rate can be stored in the rain gauge.
The correction is useful because during the emptying time of the tipping bucket, rain
is not collected from the bucket and the amount of rain in the absence of correction is
therefore underestimated. The correction to be made is greater the higher is the rain-
fall rate (greater number of cycles in the time unit).
The rain gauge allows you to set ten multiplication factors (numbered from o to 9) of
the nominal resolution, each corresponding to a range of rainfall rate.
By default, the multiplication factors are all set to 1. As option, you can request
that the correction factors of the specific rain gauge are already stored.
Command
Description
PWD
TE
Enables the measurement correction as a function of the rainfall rate.
Reply of the instrument: &|
Yes
TD
Disables the measurement correction as a function of the rainfall
rate.
Reply of the instrument: &|
Yes
TR
Reads the enabling state of the measurement correction as a func-
tion of the rainfall rate.
Reply of the instrument:
0 if the correction is disabled
1 if the correction is enabled
No
CTS
Starts the correction table writing procedure.
Reply of the instrument: CALIBRATION_STARTED
Yes
CTE
Ends the correction table writing procedure and saves the entered
values.
Reply of the instrument: CALIBRATION_ENDED
Yes
CTWi+r…r+k…k Set the multiplication factor of index i to the value k…k. The multipli-
cation factor is applied if the rainfall rate is between r…r (considered
in mm/h or in/h depending on the unit of measurement set in the in-
strument) and the rate value defined for the multiplication factor of
index i+1 (see the example below).
The multiplication factor must be between 0.500 and 2.000. The
rainfall rate must be between 0 and 7200.
Reply of the instrument:
&| if the parameters are correct
? if the parameters are not correct
Yes
- 19 -
V1.0
Command
Description
PWD
CTRi
Reads the setting of the multiplication factor of index i.
Reply of the instrument: r…rk…k, with:
r…r = rainfall rate corresponding to the multiplication factor
k…k = value of the multiplication factor
Example: +50+1.05 indicates that the multiplication factor is 1.05
and the corresponding rainfall rate is 50 (mm/h or in/h).
No
When you set the correction table, it is mandatory to enter in ascending order all the
ten multiplication factors. The rate corresponding to the multiplication factor of index i
must be greater or equal to the rate corresponding to the multiplication factor of index
i-1.
Example
To implement the following correction table:
Rainfall rate (mm/h)
0…25
25…50 50…75 75…90 90…120 120…150 150…180 180…220 220…260 260...
K
1
1.02
1.04
1.06
1.08
1.1
1.12
1.15
1.18
1.2
send, exactly in the order shown, the following commands:
CTS
CTW0+0+1
CTW1+25+1.02
CTW2+50+1.04
CTW3+75+1.06
CTW4+90+1.08
CTW5+120+1.1
CTW6+150+1.12
CTW7+180+1.15
CTW8+220+1.18
CTW9+260+1.2
CTE
Note that the sequence of the CTW commands must be preceded by the command
CTS and follwed by the command CTE.
- 20 -
V1.0
5 SDI-12 PROTOCOL
The rain gauge with HD2013UD option is equipped with an SDI-12 communication
interface compliant with the version 1.3 of the protocol, which allows the connection
to SDI-12 sensor networks.
The communication parameters of the protocol are:
• Baud Rate = 1200
• Data bits = 7
• Parity = Even (E)
• Stop bits = 1
Communication with the instrument is performed by sending a command in the follow-
ing format:
!
With = address of the instrument the command is sent to
= type of operation requested to the instrument
The instrument reply is as follows:
With = address of the replying instrument
= information sent by the instrument
= ASCII character Carriage Return
= ASCII character Line Feed
The table below shows the available SDI-12 commands. For consistency with the do-
cumentation of the SDI-12 standard, the instrument address is indicated in the table
with the letter a. The rain gauge leaves the factory with address preset to 0. The ad-
dress can be changed by using the proper SDI-12 command reported in the table.
SDI-12 Commands
Command
Instrument reply
Description
a!
a
Verifies the presence of the
instrument.
aI!
allccccccccmmmmmmvvvssssssss
with:
a = address of the instrument (1 character)
ll = SDI-12 compliant version (2 characters)
cccccccc = manufacturer (8 characters)
mmmmmm = instrument model (6 characters)
vvv = firmware version (3 characters)
ssssssss = serial number (8 characters)
⇒ Example of response:
013DeltaOhmHD201310013201518
with:
0 = instrument address
13 = SDI-12 version 1.3 compliant
DeltaOhm = manufacturer’s name
HD2013 = instrument model
100 = firmware version 1.0.0
13201518 = serial number
Requests for information
from the instrument.
- 21 -
V1.0
Command
Instrument reply
Description
aAb!
Where:
b = new
address
b
Note: if the b character is not an acceptable ad-
dress, the instrument responds with a instead of b.
Modification of the instru-
ment address.
?!
a
Request of the address of
the instrument. If more than
one sensor is connected to
the bus, a conflict occurs.
TYPE M COMMANDS (START MEASUREMENT)
Total amount of rainfall
aM!
atttn
with: ttt = number of seconds necessary for the in-
strument to make the measure available (3
characters)
n = number of detected variables (1 character)
Note: ttt = 000 means that datum is immediately
available.
Request of detection of the
total amount of rainfall
(from when the instrument
is powered).
aD0!
a+n+r…r
with:
n = number of emptying of the tipping bucket
r…r = total amount of rainfall in the unit of meas-
urement set in the instrument (mm or inches)
⇒ Example of response: 0+48+9.6
The instrument with address 0 has recorded 48
emptying of the tipping bucket and the total
amount of rainfall is 9.6 mm or inches depending
on the set unit.
Note: after reaching 99999.99, the total rainfall
counter restarts from zero.
Reads the total amount of
rainfall (from when the in-
strument is powered).
Partial amount of rainfall
aM1!
atttn
with: ttt = number of seconds necessary for the in-
strument to make the measure available (3
characters)
n = number of detected variables (1 character)
Note: ttt = 000 means that datum is immediately
available.
Request of detection of the
partial amount of rainfall
(from the last reset com-
mand).
aD0!
a+n+r…r
with:
n = number of emptying of the tipping bucket
r…r = partial amount of rainfall in the unit of meas-
urement set in the instrument (mm or inches)
⇒ Example of response: 0+12+2.4
The instrument with address 0 has recorded 12
emptying of the tipping bucket and the partial
amount of rainfall is 2.4 mm or inches depending
on the set unit.
Note: after reaching 99999.99, the partial rainfall
counter restarts from zero.
Reads the partial amount
of rainfall (from the last re-
set command).
- 22 -
V1.0
Command
Instrument reply
Description
Amount of rainfall in the current day
aM2!
atttn
with: ttt = number of seconds necessary for the in-
strument to make the measure available (3
characters)
n = number of detected variables (1 character)
Note: ttt = 000 means that data are immediately
available.
Request of detection of the
amount of rainfall in the cur-
rent day.
aD0!
a+n+r…r
with:
n = number of emptying of the tipping bucket
r…r = amount of rainfall in the current day in the
unit of measurement set in the instrument
(mm or inches)
⇒ Example of response: 0+2+0.4
The instrument with address 0 has recorded 2 emp-
tying of the tipping bucket and the amount of rain-
fall today is 0.4 mm or inches depending on the
set unit.
Reads the amount of rain-
fall in the current day.
Amount of rainfall in the previous day
aM3!
atttn
with: ttt = number of seconds necessary for the in-
strument to make the measure available (3
characters)
n = number of detected variables (1 character)
Note: ttt = 000 means that data are immediately
available.
Request of detection of the
amount of rainfall in the
previous day.
aD0!
a+n+r…r
with:
n = number of emptying of the tipping bucket
r…r = amount of rainfall in the previous day in the
unit of measurement set in the instrument
(mm or inches)
⇒ Example of response: 0+8+1.6
The instrument with address 0 has recorded 2 emp-
tying of the tipping bucket and the amount of rain-
fall yesterday was 1.6 mm or inches depending on
the set unit.
Reads the amount of rain-
fall in the previous day.
- 23 -
V1.0
Command
Instrument reply
Description
Rainfall rate
aM4!
atttn
with: ttt = number of seconds necessary for the in-
strument to make the measure available (3
characters)
n = number of detected variables (1 character)
Note: ttt = 000 means that data are immediately
available.
Request of detection of the
current rainfall rate.
aD0!
a+r…r
with:
r…r = rainfall rate in the unit of measurement set in
the instrument (mm/h or in/h)
⇒ Example of response: 0+7.3
The instrument with address 0 measures a rate of
7.3 mm/h or in/h depending on the set unit.
Reads the current rainfall
rate.
Power supply voltage and internal temperature
aM5!
atttn
with: ttt = number of seconds necessary for the in-
strument to make the measures available
(3 characters)
n = number of detected variables (1 character)
Note: ttt = 000 means that data are immediately
available.
Request of detecting the
power supply voltage and
the internal temperature
of the rain gauge.
aD0!
a+v…v+t…t
with: v…v = power supply voltage in V
t…t =internal temperature in the unit of mea-
surement set in the instrument (°C or °F)
⇒ Example of response: 0+15.3+0
The instrument with address 0 has 12.0 power
supply voltage and 15.3 °C internal temperature
(if °C is the set unit).
Reads the power supply
voltage and the internal
temperature of the rain
gauge.
TYPE C COMMANDS (START CONCURRENT MEASUREMENT)
Total amount of rainfall
aC!
atttn
with: ttt = number of seconds necessary for the in-
strument to make the measure available (3
characters)
n = number of detected variables (1 character)
Note: ttt = 000 means that datum is immediately
available.
Request of detection of the
total amount of rainfall
(from when the instrument
is powered).
aD0!
a+n+r…r
with:
n = number of emptying of the tipping bucket
r…r = total amount of rainfall in the unit of meas-
urement set in the instrument (mm or inch-
es)
Reads the total amount of
rainfall (from when the in-
strument is powered).
- 24 -
V1.0
Command
Instrument reply
Description
Partial amount of rainfall
aC1!
atttn
with: ttt = number of seconds necessary for the in-
strument to make the measure available (3
characters)
n = number of detected variables (1 character)
Note: ttt = 000 means that datum is immediately
available.
Request of detection of the
partial amount of rainfall
(from the last reset com-
mand).
aD0!
a+n+r…r
with:
n = number of emptying of the tipping bucket
r…r = partial amount of rainfall in the unit of meas-
urement set in the instrument (mm or inches)
Reads the partial amount
of rainfall (from the last re-
set command).
Amount of rainfall in the current day
aC2!
atttn
with: ttt = number of seconds necessary for the in-
strument to make the measure available (3
characters)
n = number of detected variables (1 character)
Note: ttt = 000 means that data are immediately
available.
Request of detection of the
amount of rainfall in the cur-
rent day.
aD0!
a+n+r…r
with:
n = number of emptying of the tipping bucket
r…r = amount of rainfall in the current day in the
unit of measurement set in the instrument
(mm or inches)
Reads the amount of rain-
fall in the current day.
Amount of rainfall in the previous day
aC3!
atttn
with: ttt = number of seconds necessary for the in-
strument to make the measure available (3
characters)
n = number of detected variables (1 character)
Note: ttt = 000 means that data are immediately
available.
Request of detection of the
amount of rainfall in the
previous day.
aD0!
a+n+r…r
with:
n = number of emptying of the tipping bucket
r…r = amount of rainfall in the previous day in the
unit of measurement set in the instrument
(mm or inches)
Reads the amount of rain-
fall in the previous day.
- 25 -
V1.0
Command
Instrument reply
Description
Rainfall rate
aC4!
atttn
with: ttt = number of seconds necessary for the in-
strument to make the measure available (3
characters)
n = number of detected variables (1 character)
Note: ttt = 000 means that data are immediately
available.
Request of detection of the
current rainfall rate.
aD0!
a+r…r
with:
r…r = rainfall rate in the unit of measurement set in
the instrument (mm/h or in/h)
Reads the current rainfall
rate.
Power supply voltage and internal temperature
aC5!
atttn
with: ttt = number of seconds necessary for the in-
strument to make the measures available
(3 characters)
n = number of detected variables (1 character)
Note: ttt = 000 means that data are immediately
available.
Request of detecting the
power supply voltage and
the internal temperature
of the rain gauge.
aD0!
a+v…v+t…t
with: v…v = power supply voltage in V
t…t =internal temperature in the unit of mea-
surement set in the instrument (°C or °F)
Reads the power supply
voltage and the internal
temperature of the rain
gauge.
TYPE R COMMAND (CONTINUOUS MEASUREMENTS)
aR0!
a+n+r…r
with:
n = number of emptying of the tipping bucket
r…r = total amount of rainfall in the unit of meas-
urement set in the instrument (mm or inch-
es)
Reads the total amount of
rainfall (from when the in-
strument is powered).
aR1!
a+n+r…r
with:
n = number of emptying of the tipping bucket
r…r = partial amount of rainfall in the unit of meas-
urement set in the instrument (mm or inches)
Reads the partial amount
of rainfall (from the last re-
set command).
aR2!
a+n+r…r
with:
n = number of emptying of the tipping bucket
r…r = amount of rainfall in the current day in the
unit of measurement set in the instrument
(mm or inches)
Reads the amount of rain-
fall in the current day.
aR3!
a+n+r…r
with:
n = number of emptying of the tipping bucket
r…r = amount of rainfall in the previous day in the
unit of measurement set in the instrument
(mm or inches)
Reads the amount of rain-
fall in the previous day.
- 26 -
V1.0
Command
Instrument reply
Description
aR4!
a+r…r
with:
r…r = rainfall rate in the unit of measurement set in
the instrument (mm/h or in/h)
Reads the current rainfall
rate.
aR5!
a+v…v+t…t
with: v…v = power supply voltage in V
t…t =internal temperature in the unit of mea-
surement set in the instrument (°C or °F)
Reads the power supply
voltage and the internal
temperature of the rain
gauge.
In addition to the above-mentioned commands, the rain gauge also implements the
corresponding commands with CRC, that require to add a 3-character CRC code at the
end of the reply before . The format of these commands is obtained from
the previous by adding the letter C: aMC!, aMC1!, aMC2!, aMC3!, aMC4, aMC5!, aCC!,
aCC1!, aCC2!, aCC3!, aCC4!, aCC5!, aRC0!, aRC1!, aRC2!, aRC3!, aRC4!, aRC5!
SDI-12 extended commands
In addition to the standard SDI-12 commands, the rain gauge with HD2013UD option
has extended commands that allow you to configure the instrument also via the
SDI-12 line, as an alternative to the configuration via the serial port COM PC. The ex-
tended commands are those already described in the paragraph “Serial commands”,
with the difference that they must be preceded by the address of the instrument they
are sent to, and must end with the character ! (exclamation mark).
For example, to set °F as temperature unit of measurement in the instrument with
address 0, send:
0PWDmypass!
Sends the password (assuming password=mypass)
0TUWF!
Sets °F
The reply to an extended command is always preceded by the address of the instru-
ment that replies.
- 27 -
V1.0
6 MAINTENANCE
Verify filter cleanliness periodically; check that there is no debris, leaves, dirt in the
lower filter or anything else that might obstruct the flowing of water. Check that the
tipping bucket contains no deposits of dirt, sand or any other obstruction. If neces-
sary, the surfaces can be cleaned with non aggressive detergent.
Fig. 13 – Exploded view of the upper part (version with heating)
1: knob
2: filter
3: filter support
4: filter
5: birds dissuader
6: tip
7: ring
8: housing
9: pin
10: funnel
11: nozzle
12: gasket
13: washer
14: nut
15: heater
- 28 -
V1.0
Fig. 14 – Exploded view of the lower part (version with heating)
1: magnet
2: bush
3: bearing
4: tipping bucket
5: spacer
6: plate
7: screw
8: guide
9: nut
10: screw
11: conveyor
23: heater
24: cable
25: lock nut
26: cable gland
27: junction
28: screw
29: bracket
30: spacer
31: pin
32: protection
33: washer
34: circuit
35: screw
36: support
37: cable clamp
38: screw
12: support
13: screw
14: support
15: spirit level
16: knob
17: washer
18: screw
19: filter
20: base
21: screw
22: screw
- 29 -
V1.0
7
INSTRUMENT STORAGE
Instrument storage conditions:
• Temperature: -30...+70 °C.
• Humidity: less than 90 %RH no condensation.
• In storage, avoid places where:
•
humidity is high;
•
the instrument is exposed to direct sun radiation;
•
the instrument is exposed to a high temperature source;
•
high vibration levels are present;
•
the instrument may be exposed to vapor, salt and/or corrosive gas.
8
SAFETY INSTRUCTIONS
General safety instructions
The instrument has been manufactured and tested in accordance with the safety
standard EN61010-1:2010 “Safety requirements for electrical equipment for meas-
urement, control and laboratory use” and has left the factory in perfect safety techni-
cal conditions.
The instrument proper operation and operating safety can be ensured only if all stan-
dard safety measures as well as the specific measures described in this manual are
followed.
The instrument proper operation and operating safety can be ensured only in the cli-
matic conditions specified in this manual.
Do not use the instruments in places where there are:
•
Corrosive or flammable gases.
•
Direct vibrations or shocks to the instrument.
•
High-intensity electromagnetic fields, static electricity.
Do not remove the cylindrical cover of the instrument before unplugging the
power cable of the heater.
Ensure that there is the system ground (Protective Earth) and the connecting cable is
in good condition.
User obligations
The instrument operator shall follow the directives and regulations below that refer to
the treatment of dangerous materials:
•
EEC directives on workplace safety.
•
National law regulations on workplace safety.
•
Accident prevention regulations.
- 30 -
V1.0
9
ORDERING CODES
HD2013
Tipping bucket rain gauge, area 400 cm2, for temperatures from
+4 °C to +70 °C; resolution on request when ordering 0.1 – 0.2 or
0.5 mm. Output contact normally closed.
HD2013R
Tipping bucket rain gauge, area 400 cm2, with heating for tempera-
tures from -20 °C to +70 °C. Resolution on request at the time of plac-
ing the order 0.1 – 0.2 or 0.5 mm. Output contact normally closed.
Power supply voltage: 12 Vdc or 24 Vdc ± 10% (to be specified when
ordering) / absorbed power 165W.
HD2013UA
Optional contact output and analog output at choice between 4…20
mA or 0…10 Vdc (to be specified when ordering) for HD2013 rain
gauge. Power supply 7…30 Vdc for 4…20 mA output, 13…30 Vdc for
0…10 Vdc output.
HD2013UD
Optional contact output and SDI-12 digital output for HD2013 rain
gauge. Power supply 7…30 Vdc.
HD2013.18
Bird dissuader.
HD2013.5K
Accessory kit for the installation of the rain gauge raised 500 mm
from ground (HD2013.5.1 + HD2013.64.1 + HD2013.66).
HD2013.5K.1 Accessory kit for the installation of the rain gauge raised 1 m from
ground (HD2013.5.3 + HD2013.64.1 + HD2013.66).
HD2003.75
Base with tip for the ground to support the rain gauge raised above
ground.
HD2003.78
Flat base for fastening the support of the rain gauge raised above
ground.
DELTA OHM metrology laboratories LAT N° 124 are ISO/IEC 17025 accredited by AC-
CREDIA for Temperature, Humidity, Pressure, Photometry / Radiometry, Acoustics and
Air Velocity. They can supply calibration certificates for the accredited quantities.
TERMS OF GUARANTEE
All DELTA OHM instruments are subject to accurate testing, and are guaranteed for 24 months from the
date of purchase. DELTA OHM will repair or replace free of charge the parts that, within the warranty
period, shall be deemed non efficient according to its own judgement. Complete replacement is excluded
and no damage claims are accepted. The DELTA OHM guarantee only covers instrument repair. The
guarantee is void in case of incidental breakage during transport, negligence, misuse, connection to a
different voltage than that required for the appliance by the operator. Finally, a product repaired or
tampered by unauthorized third parties is excluded from the guarantee. The instrument shall be returned
FREE OF SHIPMENT CHARGES to your dealer. The jurisdiction of Padua applies in any dispute.
The electrical and electronic equipment marked with this symbol cannot be disposed of in public
landfills. According to the Directive 2011/65/EU, the european users of electrical and electronic
equipment can return it to the dealer or manufacturer upon purchase of a new one. The illegal
disposal of electrical and electronic equipment is punished with an administrative fine.
This guarantee must be sent together with the instrument to our service centre.
IMPORTANT: Guarantee is valid only if coupon has been correctly filled in all details.
Instrument Code:
HD2013
HD2013R
Serial Number
RENEWALS
Date
Date
Inspector
Inspector
Date
Date
Inspector
Inspector
Date
Date
Inspector
Inspector
GHM GROUP – Delta OHM | Delta Ohm S.r.l. a socio unico
Via Marconi 5 | 35030 Caselle di Selvazzano | Padova | ITALY
Phone +39 049 8977150 | Fax +39 049 635596
www.deltaohm.com | [email protected]
The quality level of our instruments is the result of the constant development of the product. This may
produce some differences between the information written in this manual and the instrument you have
purchased. We cannot completely exclude the possibility of errors in the manual, for which we apologize.
The data, images and descriptions included in this manual cannot be legally asserted. We reserve the
right to make changes and corrections with no prior notice.
GHM GROUP – Delta OHM | Delta Ohm S.r.l. a socio unico
Via Marconi 5 | 35030 Caselle di Selvazzano | Padova | ITALY
Phone +39 049 8977150 | Fax +39 049 635596
www.deltaohm.com | [email protected]
V1.0
03/01/2019