KERN & Sohn GmbH

Ziegelei 1 D-72336 Balingen E-Mail: [email protected] Tel: +49-[0]7433- 9933-0 Fax: +49-[0]7433-9933-149 Internet: www.kern-sohn.com Operating Instructions Personal weighing scale, Handrail scale, Bariatric scale, Wheelchair scale, Transportation stretcher balance KERN MPS / MTS / MXS / MWS Type MPS 200K100M Type MPS 200K100NM Type MPS 200K100PM Type MPS 200K100PNM Type MTS 300K100M Type MTS 300K100NM Type MXS 300K100M Type MXS 300K100NM Type MWS 300K100M Type MWS 300K100NM Type MWS 400K100DM Type MWS 400K100DNM Type MWS 300K1LM Type MWS 300K1LNM Version 1.0 2017-02 GB MPS / MTS / MXS / MWS-BAPS-e-1710

MPS / MTS / MXS / MWS-BAPS-e-1710

2 GB KERN MPS/MTS/MXS/MWS Version 1.0 2017-02 Operating Instructions Personal weighing scale without column / with column, Handrail scale, Bariatric scale, Wheelchair scale/ Transportation stretcher balance Contents 1 Technical data ............................................................................................ 4 1.1 Tolerances altimeter .............................................................................................. 8 2 Declaration of conformity .......................................................................... 9 3 Basic directions (general information) ..................................................... 9 3.1 Intended use .......................................................................................................... 9 3.2 Inappropriate use ..................................................................................................10 3.3 Guarantee .............................................................................................................10 3.4 Monitoring the test substances ..............................................................................11 4 Basic safety directions ............................................................................ 11 4.1 Observing the directions included in the Operating Instructions ............................11 5 Transport and storage ............................................................................. 12 5.1 Check upon delivery .............................................................................................12 5.2 Packaging / return transport ..................................................................................12 6 Unpacking, installation and starting ....................................................... 13 6.1 Place of installation, place of use ..........................................................................13 6.2 Unpacking .............................................................................................................13 6.3 Installation and setting of scale .............................................................................14 6.3.1 Scope of delivery .................................................................................................................. 30 6.3.2 Installation direction for a model with wall bracket ............................................................... 30 6.4 Magnets display unit MWS ....................................................................................31 6.4.1 Transportation of balance .................................................................................................... 31 6.5 Mains socket .........................................................................................................32 6.6 Battery operation/ Rechargeable battery operation (optional)................................32 6.6.1 Battery operation .................................................................................................................. 33 6.6.2 Rechargeable battery operation (optional) ........................................................................... 35 6.7 Initial start-up ........................................................................................................37 6.8 Menu overview ......................................................................................................37 7 Operation .................................................................................................. 38 7.1 Display ..................................................................................................................38 7.2 Display view ..........................................................................................................39 7.3 Overview of keyboard ...........................................................................................40

MPS / MTS / MXS / MWS-BAPS-e-1710

3 8 Using scale ............................................................................................... 41 8.1 Weighing ...............................................................................................................41 8.1.1 Weighing with MWS ............................................................................................................. 41 8.2 Taring ...................................................................................................................42 8.3 HOLD function ......................................................................................................42 8.4 Mother/Child Function ...........................................................................................43 8.5 Determination of Body Mass Index........................................................................43 8.5.1 Classification of BMI values ................................................................................................. 44 8.6 PRE-TARE function ..............................................................................................44 8.6.1 PRE-TARE function with 5 memories .................................................................................. 45 8.7 Printing function ....................................................................................................46 8.7.1 Parameters of RS232 interface ............................................................................................ 46 9 Error messages ........................................................................................ 47 10 Service, maintenance, disposal .............................................................. 47 10.1 Cleaning ...............................................................................................................47 10.2 Cleaning / Disinfecting ..........................................................................................47 10.3 Sterilisation ...........................................................................................................47 10.4 Service, maintenance ...........................................................................................47 10.5 Disposal ................................................................................................................48 11 Troubleshooting ....................................................................................... 49 12 Adjusting ................................................................................................... 50 12.1 Adjustment key and seals .....................................................................................52 12.1.1 Menu overview in the service mode (adjustment switch in the adjustment position) .......... 54 12.2 Menu overview: .....................................................................................................56

MPS / MTS / MXS / MWS-BAPS-e-1710

4 1 Technical data KERN (Type) MPS 200K100NM/PNM MTS 300K100NM MXS 300K100NM Trademark MPS 200K100M/PM MTS 300K100M MXS 300K100M Display 6-position one Weighing range (max.) 200 kg 300 kg 300kg Display LCD with 25 mm high digits Recommended calibration weight, (class) 200 kg (M1) 300 kg (M1) 300 kg (M1) Signal rise time (typical) 2–3 s Warm-up time 10 min Operating temperature +5°C …. +35°C Storage temperature -20°C … +60°C Air humidity max. 80% (non-condensing) Power supply mains adapter 15 V / 300 mA (EN60601-1) operation with 6 x 1.5 V battery supply, AA type batteries operation time 50 h Auto-Off function after 3 min without load change (possibility of setting) Terminal (S x G x W) mm 210 x 110 x 50 Scale ready for operation (W x D x H) mm 275 x 295 x 58 with column: 275 x 460 x 1010 550x550x1060 550x550x61 Scale plate mm 275 x 295 x 60 550x550 550x550 Total weight kg (net) 4.1 6.6 21.8 15.0 Rechargeable battery operation (optional) Loading time: 14 h; operating time: 35 h; 7.2 V / 2000 mA Loading time: 14 h; operating time: 45h; 7.2 V / 2000 mA Loading time: 14 h; operating time: 50h; 7.2 V / 2000 mA

MPS / MTS / MXS / MWS-BAPS-e-1710

5 KERN (Type) MWS 300K1LNM MWS 300K100NM MWS 400K100DNM Trademark MWS 300K1LM MWS 300K100M MWS 400K100DM Display 6-position one Weighing range (max.) 300 kg 300 kg 300kg; 400kg Display LCD with 25 mm high digits Recommended calibration weight, (class) 300 kg (M1) 300 kg (M1) 400 kg (M1) Signal rise time (typical) 2 – 3 sec. Warm-up time 10 min; 10 min Operating temperature + 5° C …. + 35° C Storage temperature - 20°C … + 60°C Air humidity max. 80 % (non-condensing) Power supply mains adapter 15V / 300 mA ( EN60601-1) operation with 6 x 1.5 V battery supply, AA type batteries operation time 50 h Auto-Off function after 3 min without load change (possibility of setting) Terminal (B x T x H) mm 210 x 110 x 45 Scale ready for operation (W x D x H) mm 1500x860x68 1155x830x65 1255x1060x69 Scale plate mm 800x1200 910x740 1000x1000 Total weight kg (net) 42 28,6 42.2 Rechargeable battery operation (optional) Loading time:14 h; operating time: 45 h; 7.2 V / 2000 mA Loading time:14 h; operating time: 45h; 7.2 V / 2000 mA Loading time:14 h; operating time: 45h; 7.2 V / 2000 mA

MPS / MTS / MXS / MWS-BAPS-e-1710

6 KERN (Type) MPS 200K100M/PM MTS 300K100M MXS 300K100M Display 6-position one Weighing range (max.) 200 kg 300 kg 300kg Display LCD with 25 mm high digits Recommended calibration weight, (class) 200 kg (M1) 300 kg (M1) 300 kg (M1) Signal rise time (typical) 2–3 s Warm-up time 10 min Operating temperature +5°C …. +35°C Storage temperature -20°C … +60°C Air humidity max. 80% (non-condensing) Power supply mains adapter 15 V / 300 mA (EN60601-1) operation with 6 x 1.5 V battery supply, AA type batteries operation time 50 h Auto-Off function after 3 min without load change (possibility of setting) Terminal (S x G x W) mm 210 x 110 x 50 Scale ready for operation (W x D x H) mm 275 x 295 x 58 with column: 275 x 460 x 1010 550x550x1060 550x550x61 Scale plate mm 275 x 295 x 60 550x550 550x550 Total weight kg (net) 4.1 6.6 21.8 15.0 Rechargeable battery operation (optional) Loading time: 14 h; operating time: 35 h; 7.2 V / 2000 mA Loading time: 14 h; operating time: 45h; 7.2 V / 2000 mA Loading time: 14 h; operating time: 50h; 7.2 V / 2000 mA

MPS / MTS / MXS / MWS-BAPS-e-1710

7 KERN (Type) MWS 300K1LM MWS 300K100M MWS 400K100DM Display 6-position one Weighing range (max.) 300 kg 300 kg 300kg; 400kg Display LCD with 25 mm high digits Recommended calibration weight, (class) 300 kg (M1) 300 kg (M1) 400 kg (M1) Signal rise time (typical) 2 – 3 sec. Warm-up time 10 min; 10 min Operating temperature + 5° C …. + 35° C Storage temperature - 20°C … + 60°C Air humidity max. 80 % (non-condensing) Power supply mains adapter 15V / 300 mA ( EN60601-1) operation with 6 x 1.5 V battery supply, AA type batteries operation time 50 h Auto-Off function after 3 min without load change (possibility of setting) Terminal (B x T x H) mm 210 x 110 x 45 Scale ready for operation (W x D x H) mm 1500x860x68 1155x830x65 1255x1060x69 Scale plate mm 800x1200 910x740 1000x1000 Total weight kg (net) 42 28,6 42.2 Rechargeable battery operation (optional) Loading time:14 h; operating time: 45 h; 7.2 V / 2000 mA Loading time:14 h; operating time: 45h; 7.2 V / 2000 mA Loading time:14 h; operating time: 45h; 7.2 V / 2000 mA

MPS / MTS / MXS / MWS-BAPS-e-1710

8 1.1 Tolerances altimeter Measured value (cm) Tolerance (cm) 90 ± 0.5 100 ± 1.0 150 ± 1.0 200 ± 1.0

MPS / MTS / MXS / MWS-BAPS-e-1710

9 2 Declaration of conformity To view the current EC/EU Declaration of Conformity go to: www.kern-sohn.com/ce 3 Basic directions (general information) 3.1 Intended use The scales are used for determining the body weight of people in the standing posi- tion. The person being weighed shall stand in the middle of the weighing surface and re- main calm. The value may be read after the scales stabilizes. The scales are designed for continuous use. Only those people who can stand on the scales platform with their both legs may enter it. The scales platform is equipped in the anti-slippery surface which shall not be taken off or covered during the time the people are being weighed.

MPS / MTS / MXS / MWS-BAPS-e-1710

10 3.2 Inappropriate use Do not use the scales for dynamic weighing. Do not leave a permanent load on the weighing plate. This can damage the measur- ing equipment. Be sure to avoid impact shock and overloading the scale in excess of the prescribed maximum load rating (max.), minus any possible tare weight that is already present. This could result in damage of the scale. Never operate the scale in hazardous locations. The series design is not explosion- proof. Attention should be paid that flammable mixture may also be formed from anaesthesiological means that contain oxygen or laughing gas (nitrous oxide). Construction alterations may not be made to the scale. This can lead to incorrect weighing results, faults concerning safety regulations as well as to destruction of the scale. The scale may only be used in compliance with the described guidelines. Other are- as of application/planned use must be approved by KERN in writing. Do not use the scales for determining body weight in medicine. 3.3 Guarantee The guarantee shall become void in the event of the following: • non-observation of our guidelines in the Operating Instructions, • use outside the described applications, • alteration to or opening the device, • mechanical damage or damage caused by media, liquids, • usual wear and tear, • inappropriate erection or electric installation, • overloading of the measuring equipment, • scale falling down.

MPS / MTS / MXS / MWS-BAPS-e-1710

11 3.4 Monitoring the test substances The metrology features of the balance and any possible available adjusting weight must be checked at regular intervals within the scope of quality assurance. For this purpose, the responsible user must define a suitable interval as well as the nature and scope of this check. Information is available on KERN’s home page (www.kern- sohn.com) with regard to the monitoring of balance test substances and the test weights required for this. Test weights and balances can be adjusted quickly and at a reasonable price at KERN’s accredited DKD (Deutscher Kalibrierdienst) calibration laboratory (return to national standard). In the case of the scales for weighing people provided with the scale to determine a body size, it is recommended to carry out the check of its measuring accuracy be- cause determination of the human body size is always connected with a very large inaccuracy. 4 Basic safety directions 4.1 Observing the directions included in the Operating Instructions  Carefully read this operation manual before setup and commissioning, even if you are al- ready familiar with KERN balances.

MPS / MTS / MXS / MWS-BAPS-e-1710

12 5 Transport and storage 5.1 Check upon delivery Please check the packaging immediately upon delivery and the device during un- packing for any visible signs of external damage. 5.2 Packaging / return transport  Keep all parts of the original packaging for a possibly required return.  Only use original packaging for returning.  Prior to dispatch disconnect all cables and remove loose/mobile parts.  Reattach possibly supplied transport securing devices.  Secure all parts such as the weighing platform, power unit etc. against shifting and damage.

MPS / MTS / MXS / MWS-BAPS-e-1710

13 6 Unpacking, installation and starting 6.1 Place of installation, place of use The scale is designed in such a way that reliable weighing results can be achieved under normal application conditions. By selecting the correct location for your scale, you will be able to work quickly and precisely. Therefore, please observe the following when choosing a place of installation: - Place the scale on a firm, level surface; - Avoid extreme heat as well as temperature fluctuation caused by installing the scale next to a radiator or in the direct sunlight; - Protect the scale against direct draughts due to open windows and doors; - Avoid shaking during weighing; - Protect the scale against high humidity, vapours and dust; - Do not expose the device to extreme dampness for longer periods of time. In- admissible bedewing (condensation of air moisture on the device) can occur if a cold device is taken into a significantly warmer environment. In this case, please keep the device for approx. 2 hours at room temperature after it has been disconnected from mains supply; - Avoid static charge build-up on the scale and people to be weighed; - Avoid contact with water. Major display deviations (incorrect weighing results) are possible if electromagnetic fields occur (e.g. coming from mobile phones or radio equipment) as well as due to static charging and instable power supply. It is necessary then to change the scale location or remove disturbance source. 6.2 Unpacking Carefully remove individual scale parts or the whole scale from its packaging and po- sition the scale in its intended working location. When the mains adapter is used, be careful not to cause the danger of falling over the power cable.

MPS / MTS / MXS / MWS-BAPS-e-1710

14 6.3 Installation and setting of scale Personal weighing scale MPS with wall bracket: Scope of delivery:

MPS / MTS / MXS / MWS-BAPS-e-1710

15 Personal weighing scale MPS-PM with column: Scope of delivery: • Balance with display unit and tripod • Mains adapter • 4 screws

MPS / MTS / MXS / MWS-BAPS-e-1710

16 Assembly:  Remove cap (1)  Unscrew the screw (2)  Pull cable with Plug-in con- nection (3) through the sup- porting foot (4) and pull it out at the end (5)  Place supporting foot next to the balance

MPS / MTS / MXS / MWS-BAPS-e-1710

17  Introduce cable completely into the tripod tube (6)  Reinsert cap (1)  Screw-in screw (2) When screwing the screw en- sure that the plug-in connect- or inside the tripod foot is not squeezed.  Use the 4 screws to attach the tripod to the bottom of the balance  Level balance with foot screws until the air bubble of the water balance is in the pre- scribed circle.  Check levelling regularly.

MPS / MTS / MXS / MWS-BAPS-e-1710

18  Adjust foot screw of the tripod in a way that the tripod has a safe base and is not loose.

MPS / MTS / MXS / MWS-BAPS-e-1710

19 Mounting height measuring rod MSF 200: How to attach to KERN scale Use the 2 screws for screwing the bracket into the supplied insert nuts on the stand of the scales. Extend the height measuring rod and tighten it in the lower hole of the bracket with the help of the appertaining screw. The height measuring rod can be mounted in the same way at the back of the support stand.

MPS / MTS / MXS / MWS-BAPS-e-1710

21 Assembly: Fasten 3 corner elements to the platform, using 4 screws each time. Place the handrail on 3 corner elements and screw it. Fasten the terminal holder to the handrail with 3 screws. Remove the side rubber plugs at both sides of the display. Fasten the display to the holder with both handwheels. Adjust the display position with handwheels.

MPS / MTS / MXS / MWS-BAPS-e-1710

23 Wheelchair scale MWS and Transportation stretcher balance MWS-L: MWS MWS-L Scope of delivery:

MPS / MTS / MXS / MWS-BAPS-e-1710

24 Direction concerning installation of external column on MPS model without column, MXS and MWS models  Fasten the round plate to the aluminium profile with screws.  Fasten the wall bracket to the top of aluminium profile top with screws.  Remove the side rubber plugs at both sides of the display.  Fasten the display to the bracket with both handwheels.  Adjust the display position with handwheels.  Fasten the cable with cable clips.

MPS / MTS / MXS / MWS-BAPS-e-1710

25 Assembly of holding bracket set MWS-A02 at the MWS models 1 Railing 6 Screw 2 Handrail butts 7 Hexagonal socket wrench 3 Cross strut 8 Screw (for installation cross strut) 4 Support 9 Screw (for installation display unit) 5 Tapped bush

MPS / MTS / MXS / MWS-BAPS-e-1710

26 We recommend engaging a second person to assist you during installa- tion. Carefully remove the plastic cover from the weighing scale, ensuring that the weighing scale does not get scratched in the process. Mount all 4 handrail butts (2) on the weighing scale frame. Ensure that the handrail butt and ca- ble conduit are positioned to the right of the power unit supply point. (s. fig.) Take the two hexagonal socket wrench- es (7) and fix all handrail butts with the help of the screws 6 (3x) and the tapped bushes 5 (2x) as shown on the illustra- tion. Make sure that all screws are tight- ened properly.

MPS / MTS / MXS / MWS-BAPS-e-1710

27 Ensuring perfect fit, put the handrail con- taining three holes for the display unit on the handrail butt (1) and cable conduit. (see fig.) Using the holders 4 (2x) fasten the hand- rail to the handrail butts. Once again, use screws 6 (3x) and tapped bushes 5 (3x). Carry out same sequence of operations for second handrail. Use the two screws (8) to fasten the cross strut (3). Use the three screws to fasten the mounting plate to the handrail.

MPS / MTS / MXS / MWS-BAPS-e-1710

28 Take a screw driver and remove the plastic covers from both sides of the dis- play unit. Screw the display unit to the handle, us- ing the synthetic screws included in the delivery, The reading direction of the display unit is optional and can be adapted during the installation of the application. Display unit facing inward Display unit facing outward

MPS / MTS / MXS / MWS-BAPS-e-1710

29 Using the synthetic covers, seal both holes in the two handrail butts without cross struts. Once installation is complete, check all screws for tight fit. Otherwise the person to be weighed may suffer an injury. General direction concerning setting up the previously mentioned scales Place a personal weighing scale in the intended location and level it with the adjusta- ble rubber feet until the air bubble in the spirit level (located in the centre of the scale plate) is in the centre. When scales with large and heavy platforms are installed and transported (a scale plate folded upwards), take care not to drop a scale as this could cause its damage.

MPS / MTS / MXS / MWS-BAPS-e-1710

30 6.3.1 Scope of delivery Standard accessories: • Mains adapter • Operating Instructions 6.3.2 Installation direction for a model with wall bracket (personal weighing scale, bariatric scale, wheelchair scale, transportation stretcher scales)

MPS / MTS / MXS / MWS-BAPS-e-1710

31 6.4 Magnets display unit MWS The display unit of the MWS has two magnets on the rear side, use these magnets to fix the display unit on metallic surfaces. 6.4.1 Transportation of balance There exists the possibility to fix the display unit at the platform using the two mag- nets on the rear side, whereby the balance and the display unit can be transported together without problems (see fig. below).

MPS / MTS / MXS / MWS-BAPS-e-1710

32 6.5 Mains socket Power supply is carried out by means of the external mains adapter which also pro- vides separation between mains and a scale. The printed voltage value must be compliant with local voltage. It is allowed to use only admitted, original power supply adaptors of KERN company. 6.6 Battery operation/ Rechargeable battery operation (optional) (only models with battery operation and rechargeable battery operation) Connection CN 3 for rechargeable battery Connection CN 4 for batteries (AA x 6)

MPS / MTS / MXS / MWS-BAPS-e-1710

33 6.6.1 Battery operation On models where the back of the display unit is not directly accessible, remove the two black rotary knobs from both sides of the display unit in order to open the battery compartment and remove the display unit from the holder.  Lift-off the battery cover on the lower side of the balance  Carefully take out the battery holder (1)  Insert 6 batteries (AA). Ensure that the batteries are in- serted in the correct direction

MPS / MTS / MXS / MWS-BAPS-e-1710

34  Insert battery holder with the inserted batteries into the display unit Ensure that the cables are not squeezed  Close the battery cover If the batteries are run down, "LO" appears in the display. To turn off, press the button and immediately change the batteries. If the balance is not used for a longer time, take out the batteries and store them separately. Leaking battery liquid could damage the balance. OFF ON

MPS / MTS / MXS / MWS-BAPS-e-1710

35 6.6.2 Rechargeable battery operation (optional) When an optional rechargeable battery is used, proceed as follows: On models where the back of the display unit is not directly accessible, remove the two black rotary knobs from both sides of the display unit in order to open the battery compartment and remove the display unit from the holder.  Lift-off the battery cover on the lower side of the balance  Carefully take out the battery holder (1)  Carefully pull-out plug (2) from the connection CN 4 (3) 1

MPS / MTS / MXS / MWS-BAPS-e-1710

36  Carefully insert the rechargeable bat- tery block and insert plug into connec- tion CN 3 Ensure that the cables are not squeezed  Close the battery cover If the rechargeable battery is exhausted, „LO“ is displayed. The rechargeable battery is loaded via the provided plug-in power supply unit (loading time 14 h for a complete loading). If the balance is not used for a longer time, take out the recharge- able battery and store it separately. Leaking liquid could damage the balance.

MPS / MTS / MXS / MWS-BAPS-e-1710

37 6.7 Initial start-up To obtain accurate weighing results with electronic scales, the appropriate operating temperature is to be provided for them (refer to „Warm-up time”, section 1). During warm-up the scale must be connected to power supply and switched on (mains socket or batteries). The accuracy of the scale depends on the local acceleration of gravity. The value of acceleration of gravity is given on the rating plate. 6.8 Menu overview When the scale is switched on, hold the [→0←] key pressed for about 3 seconds until the display shows successively the „SETUP” and „A.OFF” symbols. Selection is carried out with the [TARE] and [HOLD] keys. Function Settings Description SEtuP A. oFF Automatic turning off, Auto Off func- tion 180 s Automatic turning off after 3 minutes 240 s Automatic turning off after 4 minutes 300 s Automatic turning off after 5 minutes oFF Automatic turning off turned off 120 s Automatic turning off after 2 minutes burr Acoustic signal on Acoustic signal turned on oFF Acoustic signal turned off End Leaving menu after pressing button HOLD

MPS / MTS / MXS / MWS-BAPS-e-1710

38 7 Operation 7.1 Display Type MPS 200K100M Type MPS 200K100PM Type MPS 200K100NM Type MPS 200K100PNM

MPS / MTS / MXS / MWS-BAPS-e-1710

39 7.2 Display view No. Display Description 1 [→0←] Scale zero display: If the scale does not show exactly zero value although the scale pan is unloaded, press the [→0←] key. After a short waiting time, the scale will be zeroed again. 2 [ο] Stabilisation display: If the display shows the stabilisation display [ο], the scale is in the stable condition. When the scale is in the unstable condition, the stabilisation display [ο] disappears. 3 It is illuminated when mains supply is via the mains adaptor. 4 BMI  Calculated value of the body mass index (BMI). 5 HOLD  Hold function / saving function is active. 6 PRE-TARE  Initially set tare value is active. 7 NET  The net weight is displayed. 8 WEIGHT  The present weight value is displayed.

MPS / MTS / MXS / MWS-BAPS-e-1710

40 7.3 Overview of keyboard Key Description ON/OFF Switching on/off the scale PRINT Data transmission via interface BMI Determination of Body Mass Index HOLD Hold function / determination of stable weighing value 0 The scale is reset to 0,0 kg display. It is possible to set max. up to 2% of maximal load for verified scales, and 2% or 100% of maximal load for common scales (possibility of selection in the menu) M 1-5 Memories 1–5 were called PRE-TARE Calling the tare function with set values TARE Taring the scale CLEAR Clearing the digits entered manually 0..9 Entering digits ENTER Using the entered digits

MPS / MTS / MXS / MWS-BAPS-e-1710

41 8 Using scale 8.1 Weighing  Switch on the scale with the [ON/OFF] key. The diagnostic scale self-check is performed and then the software version is displayed. The scale is ready for weighing when the „0,00 kg” weight display is shown. Direction: The [→0←] key makes it possible to zero the scale if necessary and at any time.  Place a person in the middle of the scale. Wait until the stability display (ο) is shown and then read the weighing result. Direction: If a person is heavier than the weighing range, the display will show the „Err” symbol (= overload). 8.1.1 Weighing with MWS Due to the great dimensions and the big weighing range especially suitable for weighing of immobile people on transportation stretchers, wheelchairs or overweight people in the adiposity range 8.1.1.1 Weighing with transportation stretcher or wheelchair  Place transportation stretcher/wheelchair with the person on the centre of the scales  Arrest the brakes of the transportation stretcher/of the wheelchair Do not leave the person unattended  When the person is lying/sitting quietly, read weighing value 1  Loosen the brakes and carefully pull off the transportation stretcher/wheelchair with the person  After that weigh the transportation stretcher/wheelchair without the person and subduct this weight from weighing value 1, from there results the person’s weight.

MPS / MTS / MXS / MWS-BAPS-e-1710

42 8.2 Taring The dead weight of any initial load used for weighing may be tared away by pressing the key, so that the following weighing shows the real weight of a person to be weighed.  E.g. when a rubber mat is put on the scale plate, the scale does not show 0 value.  To start the taring process, press the [TARE] key. Now internal weight saving is performed and value of 0.0 kg is displayed.  Place a person in the middle of the scale plate.  Then read the weight on the display. Direction: The scale can store only one tare value. When the scale is unloaded, the saved tare value is displayed with „negative” sign. To delete the saved tare value, unload the scale plate and then press the [TARE] key. 8.3 HOLD function The scale is provided with the integrated hold function (determination of average val- ue). It enables people to be weighed accurately although they are not still on the scale plate. Note: Determination of average value is not possible when a person moves too much.  Switch on the scale with the [ON/OFF] key. The diagnostic scale self-check is performed. The scale is ready for weighing when the „0.00 kg” weight display is shown.  Place a person in the centre of the scale plate.  Press the [HOLD] key. When the triangle is flashing on the display, the scale takes some measuring values and then the calculated average value is dis- played.  Press the [HOLD] key again to return the scale to the normal weighing mode.  Pressing the [HOLD] key makes it possible to activate the function at any time.

MPS / MTS / MXS / MWS-BAPS-e-1710

43 8.4 Mother/Child Function The “Mother/Child” function makes it possible to determine the body weight of infants and small children held in an adult’s arms.  Turn on the scales using the [ON/OFF] switch. The scales enters the self- diagnostic mode. The device is ready for operation as soon as “0.00 kg” is shown on the weight display.  Position the adult person in the middle of the scales platform. After showing the stability indicator, the scales displays the person’s weight. A triangle sym- bol is shown under the “WEIGHT” symbol.  Press the [TARE] button and the readout changes to “0.00 kg”.  Place the child in the arms of the adult person. After showing the stability indi- cator, the scales displays the child’s weight. The triangle symbol is now dis- played under the “NET” symbol.  Press the [TARE] button again and the readout again changes to “0.00 kg”.  After stepping off the scales, the total weight of the adult person and the child is displayed as a negative value.  Press the [TARE] button again. The saved tare value is cancelled and the scales can be used for weighing again. 8.5 Determination of Body Mass Index When you obtain a stable weight and display shows 0.0 kg, place a person in the middle of the scale plate. Wait until the weighing value is stable. Then press the BMI key. Now enter a body height. Please take note that reliable determination of BMI index is only possible for body height from 100 cm to 250 cm and weight > 10 kg. A body height entered as the last one is flashing on the display. Now you can enter a different value with the numerical keypad. Confirm the entered value with the ENTER key, and then a person’s BMI index will be displayed. When the BMI index value is displayed, it is presented with the arrow pointing the BMI symbol. To return to the weighing mode, press the BMI key once again and the arrow at the BMI symbol will disappear.

MPS / MTS / MXS / MWS-BAPS-e-1710

44 8.5.1 Classification of BMI values Classification of weight for adults over 18 years on the basis of Body Mass Index according to WHO, 2000 EK IV and WHO 2004 (WHO - World Health Organization). Category BMI (kg/m2) Risk of diseases accompanying overweight Underweight < 18.5 low Normal weight 18.5 – 24.9 average Overweight Preobesity I degree of obesity II degree of obesity III degree of obesity > 25.0 25.0 – 29.9 30.0 – 34.9 35.0 – 39.9 > 40 slightly increased increased high very high 8.6 PRE-TARE function When a tare weight (rubber mat, clothes, ...) is known, this value can be entered manually. If the PRE-TARE key is pressed shortly, the flashing display will be shown. The PRE-TARE function is active as long as the small arrow is pointing the ”PRE- TARE” symbol on the display. The value used as the last one will be displayed. If a different value is required, a new weight value can be entered with the numerical keypad. By pressing the ENTER key, the new value is confirmed and used. Then the entered value with minus sign will be shown on the display. When a person is placed on the scale plate, the display will show a weight value less the value entered previously. Pressing the PRE-TARE key again will return the scale to the normal weighing mode.

MPS / MTS / MXS / MWS-BAPS-e-1710

45 8.6.1 PRE-TARE function with 5 memories Owing to this function it is possible to store 5 Pre-Tare values (e.g. for different wheelchairs), and then call up them if necessary. Saving PRE-Tare values: To enable a later calling up values from the memory, they are to be entered into the memory first. It is carried out in the following way: The scale plate is unloaded, and the scale display is showing 0,0 kg. Put a weight, whose value is to be saved (e.g. an empty wheelchair), on the scale plate and wait until a stable weight display is shown. Press the M1-5 key repeatedly until the display will show the „ni” (M) symbol. Press the key with digit (1..5) shortly to indicate which number a value is to be saved under. The previously displayed weight value is flashing for 3 seconds. When the flashing is finished, press again the key with digit previously pressed and the weighing value is saved in the memory (short audible signal). By pressing the CLEAR key, the scale will return to the weighing mode without saving the value. The display will show the real value of the weight placed on the scale plate. When the weight is removed, the display will show 0,0 kg. Calling up the PRE-Tare value from memory: Press the PRE-Tare key repeatedly until the display will show the „ni” (M) symbol. Pressing the key with digit (1..5) will display the flashing weight value saved there. The small arrow, additionally shown on the display, is pointing to the „PRE-TARE” symbol. By pressing another key with digit (1..5), the appropriate also flashing weight value will be displayed. Press the ENTER key and the value will be accepted and shown on the display as the PRE-Tare value with minus sign. Now you can place e.g. a person in a wheelchair or on the transportation stretcher on the scale, and only a person‘s weight will be displayed. To return to the normal weighing mode when the scale plate is unloaded, press the PRE-Tare key shortly again. This will also result in disappearing the small arrow pointing the „PRE-TARE” symbol.

MPS / MTS / MXS / MWS-BAPS-e-1710

46 Printing Pre-Tare memory (refer also to section 8.6): To do this, press the PRE-Tare key repeatedly until the display will show the „ni” (M) symbol. Short pressing the PRINT key will activate printing of the values saved in 5 memo- ries. M1 0.0 kg M2 7.0 kg M3 10.0 kg M4 30.0 kg M5 50.0 kg 8.7 Printing function To use this function, you need the RS232 interface cable (available as accessories) which is connected with the round plug at the terminal back. Note: In medical applications, only the peripheral equipment meeting EN 60601-1 standard can be connected to the interface. When a scale is in the weighing mode, pressing the PRINT key will result in output of the particular data, presented below, via the interface. It is the standard way to output data, which cannot be changed. G 88.8 kg Gross weight T 2.0 kg Taring weight N 86.8 kg Net weight 180.0 cm Size of the person 24.4 BMI BMI value 8.7.1 Parameters of RS232 interface Set parameters of the scale interface on the connected device. It is not possible to change the scale parameters. Baud rate: 9600 bps Parity check: no Data length: 8 bits Stop bit: 1 bit Handshake: no or Xon/Xoff Data code: ASCII

MPS / MTS / MXS / MWS-BAPS-e-1710

47 9 Error messages The following messages can be shown on the display during switching on or using the scale. ERRL: Too small weight on the scale. ooooo: The scale plate was loaded during switching on the scale. Unload the scale plate. ERR: Overload, too large weight on the scale plate. 10 Service, maintenance, disposal 10.1 Cleaning Disconnect the unit from the mains power supply prior to cleaning, maintenance and repair activities. 10.2 Cleaning / Disinfecting Clean weighing platform (such as seat) as well as casing with household detergents or commercially available disinfectants, e.g. 70% isopropanol. We recommend a dis- infectant suitable for wiping disinfection. Please follow manufacturer’s instructions. Do not use abrasive or aggressive cleaners such as spirits or alcohol or similar as they might damage the high-quality surface. To prevent cross-contamination (fungal skin infection) please observe the following time intervals for disinfection: • Weighing plate before and after any measurement with direct skin contact • When required: o Display o Foil-covered keyboard Do not spray disinfectants onto appliance. Make sure that disinfectant does not penetrate the interior of the balance. Remove dirt immediately. 10.3 Sterilisation Sterilisation of the appliance is not allowed. 10.4 Service, maintenance The device may only be operated and maintained by trained service technicians who are authorised by KERN. Disconnect the scale from mains supply before its opening.

MPS / MTS / MXS / MWS-BAPS-e-1710

48 10.5 Disposal Disposal of packaging and device must be carried out by an operator according to valid national or regional law of the location where the device is used.

MPS / MTS / MXS / MWS-BAPS-e-1710

49 11 Troubleshooting The scale should be switched off for a short time following an interruption in the pro- gram sequence and disconnected from mains supply. It is then necessary to repeat the weighing process from the beginning. Interruption Possible cause Weight display is not illumi- nated. • The scale is not switched on. • The mains supply connection has been interrupted (mains cable not plugged in/faulty). • Check the fuse of the mains adapter / green LED is illuminated next to the fuse. • Mains failure. • Batteries are incorrectly inserted or discharged. • No batteries. Weight display changes continuously • Draught/air movement • Table/floor vibrations • The weighing plate is in contact with foreign matters or is installed incorrectly. • Electromagnetic fields/static charging (choose a different location for the scale, switch off an inter- fering device if possible). The weighing result is obviously incorrect • The scale display is not set to zero • Incorrect adjustment. • Great fluctuations in temperature. • Warm-up time was ignored. • Electromagnetic fields/static charging (choose a different location for the scale, switch off an inter- fering device if possible). Should other error messages occur, switch the scale off and then on again. If the er- ror message remains, inform the manufacturer.

MPS / MTS / MXS / MWS-BAPS-e-1710

50 12 Adjusting Observe stable environmental conditions. The warm-up time (refer to chapter 1) is required to ensure the scale stabilisation. Note: In the case of verified scales, adjusting is locked with the jumper. To carry out adjust- ing, the jumper is to be set in the adjusting position (central position). (refer to 12.2). Operation Display Turn the scale on using the [ON/OFF] key. Press and keep the [→0←] key pressed for about 3 seconds until the display shows successively the „SETUP” and „UNIT” symbol. SETUP   UNKT  Press the [TARE] key until the „CAL ib” symbol is displayed. CAL  KB Press the [HOLD] key. CAL U  Press the [TARE] key. The triangle  must be displayed at the top right side of the display. If not, press the [TARE] key. Press the [HOLD] key repeatedly until the „CAL 0” symbol is displayed. CAL U   CAL O  Press the [TARE] key, the display will show the present numerical value. Then press the [ENTER] key. 30770  CAL O 

MPS / MTS / MXS / MWS-BAPS-e-1710

51 Press the [HOLD] key. CAL 5  Press the [TARE] key. Enter the required calibration weight value (refer to chapter 1, „Technical data”): Select the item to be changed with the HOLD key and set its numerical val- ue with the [TARE] key. 200.0 Confirm by pressing the [ENTER] key. Place the calibration weight carefully in the centre of the scale plate, and the display will show a numerical value. Press the [ENTER] key. The adjustment pro- cess is started. 82077 When the adjustment is finished successfully, the scale is automatically switched over to the weighing mode again and the calibration weight value will be displayed. Remove the calibration weight.

MPS / MTS / MXS / MWS-BAPS-e-1710

52 12.1 Adjustment key and seals When verification of a scale is finished, the positions indicated on the scale are sealed. Verification of a scale without a seal is invalid. Position of seals: 1. Backside 2. Battery compartment

MPS / MTS / MXS / MWS-BAPS-e-1710

54 Position of adjustment switch: Position of adjustment switch Status 1. left side not documented 2. centre adjustment position - adjustment is possible 3. right side verification position – adjustment locking 12.1.1 Menu overview in the service mode (adjustment switch in the adjustment position) Overview is only used to check the set parameters by authorised verification offices. Changes may only be introduced in parameters of the automatic switching off function „A.OFF” and audible signal „BURR”. 1 2 3

MPS / MTS / MXS / MWS-BAPS-e-1710

55 Navigation in menu:  When the scale is switched on, press and hold the [→0←] key pressed for about 3 seconds until the display shows successively the „SETUP” and „UNIT” symbol.  Press the [TARE] key repeatedly until the required function is displayed.  Confirm the selected function with the [HOLD] key. The first parameter will be displayed. Select the required parameter with the [HOLD] key and confirm the selection with the [TARE] key. To exit the menu and save the settings, press the [TARE] key until the „END” symbol is displayed and then confirm with the [HOLD] key. The scale is automatically returned to the weighing mode. Selection is carried out with the [HOLD] and [TARE] key.

MPS / MTS / MXS / MWS-BAPS-e-1710

56 12.2 Menu overview: Function Settings Description SEtuP Unit on-off Weight unit: kg Grad 3000 d – 6000 d – 10000 d – 500 d – 1000 d – 1500 d - 2500 d - 2000 d The values of the scale elementary division, weigh- ing range (max) and elementary division (d) Ut.-d Full – S-Ut Selection: one-range scales (Full)- / multi-range scales (S-Ut) FillE Fast – Nor. - SLo Filter: fast - normal - slow Auto 0 0.25 d – 0.5 d – 1 d – 3 d - OFF Auto-Zero-Tracking Stab 0.25 d – 0.5 d – 1 d – 3 d - off Stabilisierungsbereich Orang 2 Pct – 100 Pct. Nullbereich: 2 % / 100 % Ould 9 d – 2 Pct. Überlastbereich: 9 d / 2 % CALib CAL-U – CAL–0 CAL-5 Justierung A.Off 120s/180s/240s/ 300s/off Automatische Abschaltfunktion burr on/off Acoustic signal default Resetting to factory settings End Leaving the menu after pressing button HOLD

MPS / MTS / MXS / MWS-BAPS-e-1710

57 Description: Weight unit: kg Scale divisions, weighing range (max.) and read-out (d) Selection of multi-range / single-range scale Single-range scale Multi-range scale Filter: fast / normal / slow Automatic zero tracking: 0,25 d/ 0,5 d/ 1 d/ 3 d/ OFF Stabilisation range: 0,25 d/ 0,5 d/ 1 d/ 3 d/ OFF Zero range: 2% / 100% Overload range: 9 d / 2% Adjusting Auto off function: 120 s / 180 s / 240 s / 300 sec/ OFF Audible signal: ON/OFF Restoring the factory settings (default settings) Exiting the menu