Inficon 122 20 Operating Instruction
Read Online
Read the Inficon 122 20 Operating Instruction online using the built-in PDF viewer.
Text of the manual
Test Leck for Sniffer
Leak Testing
O P E R A T I N G I N S T R U C T I O N
lina80e1-06 (1603)
Catalog No.
122 20 bis 122 23S
122 25, 122 25S
122 27 bis 122 43
122 75
123 22
Cleaning the outside
10
3.2
Repair
11
3.3
Returning the Product
11
3.4
Decommissioning
11
INFICON covers no responsibility nor warranty if the user or
third parties
•
disregard the information in this document
•
utilize the product not according to the defined use
•
make any kind of changes (modifications, alterations,
etc.) to the product.
Important remarks concerning operational safety and protec-
tion are emphasised as follows:
Notice:
Information on correct handling or use. Disregard may
lead to malfunctions or minor equipment damage.
The references to diagrams consist of the Fig. No. and the
item No. in that order. For example: (2/1) refers to item 1 in
the second figure, i.e. the Manometer
Indicates procedures that must be strictly observed to prevent hazards to
persons
Indicates procedures that must strictly be observed to prevent damage, or
destruction.
The test leak is designed as bench-top unit and used for cali-
brating and testing sniffing leak detectors. The safety regula-
tions have to be taken into account.
1.1.2
Validity
This manual applies to products with Cat. No.
Catalog number
Gas of test leak
122 20
TL R134a
122 20S
TL R134a
122 20SN
TL for special gases (non flammable)
122 20SF
TL for special gases (flammable)
122 21
TL R600a
122 21S
TL R600a
122 22
TL R404a
122 22S
TL R404a
122 23
TL R502
122 23S
TL R502a
122 25
TL R22
122 25S
TL R22a
122 27
TL R152a
122 27S
TL R152a
122 28
TL R407c
122 28S
TL R407c
122 29
TL R410a
122 29S
TL R410a
122 30
TL R401a
122 30S
TL R401a
122 31
TL R290
122 32
TL R744 (CO2)
122 33
Foming gas (10% hydrogen in Helium)
Cat. No.122 20SN and 122 40SN for special gases can be
ordered on special request only.
The Cat. No. is shown on the name plate on the back side of
the calibrated test leak.
1.2
Safety information
122 34
Halon 1301
122 35
TL HFO-1234yf
122 36
TL R32
122 37
S-TL 4 Helium
122 37S
S-TL 4 Helium
122 38
S-TL 5 Helium
122 39
S-TL 6 Helium
122 40
TL R134a
122 40SN
TL for special gases (non flammable)
122 40SF
TL for special gases (flammable)
122 41
TL R600a
122 42
TL R404a
122 43
TL R502
122 75
TL R744 (CO2)
Catalog number
Gas of test leak
Take into account the relevant regulations and safety measures
for the calibration gases used
( safety data sheets for the calibration gases).
The test leak is pressurized. Protect it from solar heat, tempera-
tures above 50 8C, and damage.
Calibration gas: Avoid eye-contact and inhalation in high con-
centrations.
Pass on the safety information to other users.
Flammable: Keep away from ignition sources.
Smallest deviation from calibration
± 10 %
pressure range
0.5 bar to 8 bar
temperature range
+15°C to +35°C
Storage temperature
0°C to +50°C
Service life (depending on the type of gas
and the leak rate)
1 - 5 years
Weight
approx. 3.5 kg
Fig. 1 Dimensional drawing
Leak rate, typical
Cat. No.
g/a
oz/y
R 22
2 … 5
0,07 … 0,18
122 25
R 22
2 … 5
0,07 … 0,18
122 25S
R 23
2 … 5
0,07 … 0,18
122 26
R 32
2… 8
0,07 …0,28
122 36S
R 134a
2 … 5
0,07 …0,18
122 20
R 134a
2 … 11
0,07 ... 0,39
122 20S
R 134a
10 … 14
0,36 … 0,49
122 40
R 152a
2 … 5
0,07 … 0,18
122 27
R 152a
2 … 8
0,07 … 0,28
122 27S
R 290
7 … 8
0,25 … 0,28
122 31
R 401a
2 … 5
0,07 … 0,18
122 30
R 401a
2 … 15
0,07 … 0,53
122 30S
R 404a
2 … 5
0,07 … 0,18
122 22
R 404a
10 … 15
0,35 … 0,53
122 22S
R 404a
13 … 17
0,46 … 0,60
122 42
R 407c
2 … 5
0,07…0,18
122 28
R 407c
2 … 15
0,07…0,53
122 28S
R 410a
2 … 5
0,07…0,18
122 29
R 410a
2 … 5
0,07…0,18
122 30S
R 502a
11 … 15
0,39…0,53
122 43
R 502a
2 … 5
0,07…0,18
122 23
R 502a
2 … 15
0,07…0,53
122 23S
R 600a
2 … 5
0,07…0,18
122 21
R 600a
2 … 20
0,07…0,71
122 21S
R 600a
14 … 18
0,49…0,63
122 41
* Test leaks are only suitable for specific gases. If you want
to have a test leak with a gas that is not on the list please con-
tact us. We will tell you if the test leak with the gas you reqest
can be build.
For additional information please see the safety data sheet of
the calibration gas and the quality test certificate that was
added to the test leak.
HFO-
1234yf
2 … 5
0,07…0,18
122 35
R 1301
2 … 17
0,07 … 0,60
122 34S
R744
2 … 3,5
0,07…0,12
122 32
R744
10 … 14
0,36…0,50
122 75
He
1,00 - 1,20 x 10-4 mbar l/s
122 37
He
1,00 - 1,20 x 10-4 mbar l/s
122 37S
He
2,00 - 6,00 x 10-5 mbar l/s
122 38
He
6,00 - 8,00 x 10-6 mbar l/s
122 39
10% H2,
90% He
1 - 1,2 x 10-4 mbar l/s
122 33
100% H2
1 - 1,1 x 10-4 mbar l/s
123 22
customer
specific*
2 … 5
122 20SF
customer
specific*
2 … 5
122 20SN
customer
specific*
10 … 14
122 40SF
customer
specific*
10 … 14
122 40SN
Gas type
Leak rate, typical
Cat. No.
g/a
oz/y
The calibrated test leak is installed in a rectangular housing.
The front view is shown in Fig. 2.
2.2
Handling
The calibration test leak is continually in operation.
Preconditions
Ambient temperature +15°C to +35°C
Displayed pressure range 0.5 bar to 8 bar
Notice:
p < 0.5 bar: Test leak no longer ready for operation.
p > 8 bar: Test leak too hot.
Pos.
1
2
3
4
Description
Manometer
Leak rate of the calibrated test leak.
(Example: Gas R152a, leak rate 2,64 g/a)
Calibration gas outlet
Date of calibration
Fig. 2 Front view
Flammable gases! The gases R152a, R290, R502, R600a and HFO-1234yf
may form explosive gas mixtures.
Care for sufficient ventilation and keep away from ignition source.
Keep probe in contact with the gas outlet until the measured
value is stable.
3
Maintenance
The calibrated test leak requires no maintenance.
When the calibration gas is used up (pressure < 0.5 bar) the
test leak has to be refilled and calibrated. Please contact your
nearest INFICON Service Center.
Notice:
If the pressure is < 0.5 bar a proper calibration of the leak
detector is not assured. Therefore an accurate
measurement of the test sample is not guaranteed.
3.1
Cleaning the outside
For cleaning the outside of the test leak, a slightly humid cloth
should normally do.
Do not use under any circumstances any aggressive or scou-
ring cleaning agents.
Fig. 3 Sniffing Probe
Defective products must not be opened.
Please contact your nearest INFICON-Service Center.
3.3
Returning the Product
3.4
Decommissioning
Calibration gas
Test leaks sent for repair may still contain calibration gas.
Take account of the national regulations for shipping the respective calibra-
tion gas.
Substances detrimental to the environment
Products, operating material, fluids, consumables etc. may have to be spe-
cially decommissioned.
For professional decommissioning of the product, please contact your nea-
rest INFICON-Service Center.
INFICON GmbH, Bonner Strasse 498, D-50968 Cologne, Germany
UNITED STATES TAIWAN JAPAN KOREA SINGAPORE GERMANY
FRANCE UNITED KINGDOM HONG KONG
Visit our website for contact information and sales offices. www.inficon.com
Document: lina80e1-06 (1603)