GIG DS404 Operation Manual
Read Online
Read the GIG DS404 Operation Manual online using the built-in PDF viewer.
Text of the manual
GfG Products For Increased Safety
Congratulations!
You decided in favour of a high-technology product from GfG.
A good choice!
Our detectors are characterised by reliability, safety, best performance and
economic efficiency.
They comply with national and international directives.
This manual will help you to operate the detector quickly and safely.
Please take note of the operational hints before putting into operation!
For any questions please feel free to contact us.
Your
GfG Gesellschaft für Gerätebau mbH
Klönnestraße 99
44143 Dortmund, Germany
Tel.: +49(0)231 – 564 00-0
Fax: +49(0)231 – 564 00-895
E-mail: [email protected]
Internet: www.gasmessung.de
Table of Contents
Page
INTRODUCTION
4
For your Safety
4
Application and Purpose
4
Installing and Connecting the Docking Station
4
Connections and Control Elements of the DS404
5
Detector Adapter
6
CO2 Absorption Filter
6
HANDLING
FEHLER! TEXTMARKE NICHT DEFINIERT.
Charging
7
Bump Test
7
Sensor Calibration
9
Data Storage/Data Transfer to a PC
11
SOFTWARE
12
Installation
12
Installation of the USB Driver
12
Configuration of the DS404
12
Connection
12
General
12
CO2 Filter
13
Gas Input 1
13
Test Times
13
Clock
14
Intervals
14
ANNEX
15
Care
15
Maintenance and Inspection
15
Troubleshooting
15
Accessories and Spare Parts
16
Technical Data
16
For your Safety
In accordance with § 3 of the German Equipment and Product Safety Act (GPSG), this manual describes the
intended use of the product and serves to avoid danger. It must be read and observed by all persons who
operate, service, maintain and inspect this product.
This detector can serve its intended purpose only if it is operated, serviced, maintained and inspected
according to the instructions given by GfG Gesellschaft für Gerätebau. The warranties given by GfG
Gesellschaft für Gerätebau mbH will be voided if the product is not operated, serviced, maintained and
inspected in accordance with GfG’s instructions.
The above does not alter statements regarding warranties and liabilities in the General Conditions of Sale
and Delivery of GfG mbH. Repairs may only be carried out by skilled or trained persons. Modifications and
changes to the product may only be carried out with the approval of GfG mbH. Unauthorised modifications
to the product result in the exclusion of any liability for possible damage. Only genuine GfG accessories may
be used together with the product. Only spare parts approved by GfG mbH may be used for repair work.
Note: When using the docking station you have to remember that all gas detectors that are calibrated with
the station will be influenced. Incorrectly entered test gas concentrations will result in all gas detectors being
incorrectly calibrated.
Application and Purpose
The Docking Station DS404 is an automatic test station with four test gas ports for carrying out function
tests ("bump tests") and for calibration of the zero point and sensitivity of the sensors of the Microtector II,
Microtector III und Polytector III series gas detectors.
Installing and Connecting the Docking Station
During commissioning, the hoses and leads for fresh air supply, test gas
supply, gas discharge, pressure switch, mains power supply and PC
interface or interface to a further docking station must be connected to
the test station.
The docking station must be installed on a firm and even surface. The
gases must be supplied pressure-free. For direct gas tapping from test
gas cylinders, use the gas connection fittings listed in section
"Accessories and Spare Parts". During operation, the gas is drawn in by
the pump of the docking station. Ensure also that the test gas can leave
the docking station freely without pressure! If CO2 sensors are also to be
tested or calibrated using the docking station, a CO2 absorption filter
should be installed in the fresh air supply line (see chapter CO2
absorption filter).
Ensure that the plug is installed on the T-piece of the fresh air supply
line or that it is connected to a further docking station.
Up to two DS404 or three DS400 docking stations can be operated with
one DS400-PS1 power supply unit. For this the DS404 or DS400 have to
be connected together using a double-sided jack plug.
During commissioning, the connections for gas supply, gas discharge,
pressure switch, mains power supply and PC interface or interface to a
further docking station must be made as shown in the figure in the
following chapter.
Connections and Control Elements of the DS404
Zero gas/fresh air
supply
Pressure switch –
channel 1
Gas exhaust pipe
Pressure switch –
channel 4
Test gas supply –
channel 4
SD card
Red LED
(on left side in the card slot)
Connection for
12 V DC power
supply unit
Ports for RS485
Green LED
(on the right in
the card slot)
Connection for
12 V DC power
supply unit
1 2 3 4
Display of active input
Three detector adapters are available for prepared docking stations. One version for devices without
screwed on electric pump. A second version for devices with screwed on electric pump (G400-MP2) and a
third version for the detectors Microtector III and Polytector III. The electric pump can remain on the
device for all tests. All three device adapters can be used at the same time by the docking station as
charging cradle. To replace the device adapter, the DS400 must be disconnected from the power supply.
CO2 Absorption Filter
CO2-free zero gas is required particularly for the testing and
calibration of the CO2 sensor zero point. As fresh air contains
approx. 500 ppm CO2, the disturbing CO2 can be removed using
a CO2 absorption filter. The CO2 is thereby absorbed by calcium
hydroxide (80 %) / sodium hydroxide (5 %) / H2O (15 %) in the
following reaction:
CO2 + 2 NaOH Na2CO3 + H2O
Na2CO3 + Ca(OH)2 CaCO3 + 2NaOH
The absorber removes 1000 ppm CO2 from approx. 2,500 l of air.
This corresponds to a service life of approx. 5,000 minutes of
zero gas pump operation with one docking station. The DS404
calculates the service life on the basis of the flow rate and gives a
warning to replace the filter in good time. The filter gradually
changes colour to pale blue.
As water is also produced during the absorption process due to the chemical reaction, depending on the CO2
content of the air, this can have a minor influence on the setting of the sensor zero point.
Handling
The docking station is operated by means of the control keys of the gas detector. The status and test data
are output to the detector display. If a gas detector is inserted into the docking station, the "Information",
"Function Test" or "AutoCal adjustment" options can be selected via softkeys on the gas detector.
Microtector II devices can remain in the docking station for operation. Devices of the Microtector III and
Polytector III series must be briefly removed from the charger to select the displayed operating options, in
order to be able to operate the concealed keys. After pressing the keys, the unit is placed back in the
charger. Under certain conditions, a "Firmware Update" can also be carried out using the docking station
when the gas detector is in the station. For this, there must be a newer version of a corresponding update
file on the SD card and the "Update" started by pressing the corresponding key.
In order to charge a gas detector, it must be switched off and placed in the docking station. If the detector is
placed into the docking station switched on, the necessary test or calibration will be started automatically
after 10 seconds. Within the 10 seconds you can override the automatic selection and start a function test or
calibration manually.
G400-DIC1D for
Microtector II
without a pump
G400-DIC2D for
Microtector II with
attached pump
DIC888/999-B
for Microtector III
and Polytector III
Turn the G450 or G460 off before placing it into the docking
station. The charging procedure starts automatically.
Yellow LED
Constantly lit: Normal charging
Flashing: Trickle charge
Green LED
ON: Voltage supply for charging module switched on
OFF: Detector is in the docking station and function test
or calibration is being performed
The charging process of an empty battery takes approx. 6 to
7.5 hours. After the normal charging process, the detector
automatically switches to trickle charge.
Function Test (Bump Test)
During the function (bump) test, the following points are checked:
Check visual alarm
Check audible alarm
Check response time of sensor for alarm 1 and for alarm 2, depending on test gas concentration
Depending on the configuration, check response time of sensor for t50 or t90
Automatic data storage on SD card
Setting of the clock
Setting of the function test interval
Turn the G450 or G460 on before placing it into the docking station. If the TEST button is pressed within the
first 10 seconds, the function test is started. If no button is pressed, the function test or calibration starts
automatically after 10 seconds, depending on which interval has expired. The effective time for the function
test is approx. 20 seconds, on the DS404 longer depending on the gas inlets used. During the test, the
relevant test gas is supplied to the sensors and then they are adequately purged with fresh air. The function
test interval is specified in accordance with T021/T023 and is used during each function test and calibration
to calculate the next date for function test or calibration.
The test progress and the function test report are shown on the display of the detector.
Function test starts after 10
seconds automatically
Function test running
Purging with fresh air
Talm = Response times for alarm 1
and alarm 2 of the respective
test gas
T50 = t50 response time of the
respective test gas
Test of audible and
visual alarm o.k.
Response time alarm 1 / 2 for CO
error – Response time too long
Response time t50 for CO error –
Response time too long
The following symbols signal the test result:
= In process
- = Sensor not tested
√ = Sensor successfully tested
= Sensor faulty
Result of the function test:
Check the proper performance of the test by inspecting the function test report. The result of the function
test is also indicated by a red and a green LED in the SD card slot of the docking station using the traffic
light principle. If the green LED is lit, the function test was successful and the detector is ready for use
again. If the red LED is lit or the display on the detector is red, however, an error has occurred which has to
be remedied before the detector can be used.
Function test successful:
Display is green + green LED lit in slot
Function test failed:
Display is red + red LED lit in slot
Display of the error which has occurred
(here Code 2: Flow error)
Possible error messages:
Code: 1 No gas defined – DS404 is not configured
Code: 2 Flow error – Fault in the zero or test gas supply
Code: 3 Aborted by detector – Detector problem / fault
Code: 4 Power error – Fault in supply voltage to TS400 power supply unit
Code: 5 Time too long for function test of the sensors
Code: 6 DS404 service necessary – The DS404 has to be serviced
Code: 7 No SD card – No SD card inserted
Code: 9 Parameter error – An error has been detected in the parameter memory
Code: 12 Firmware prior 3.44 – The firmware version on the detector is too old
Code: 13 Gas switch error – The gas switch signals a fault
Additional information can be called up with the "Info" key. These messages do not lead to the function test
being aborted.
1 = SD card
2 = Gas pressure
4 = CO2 filter
8 = Gas amount
If several errors occur at the same time, the error numbers are added:
3 = 2 + 1
(Gas pressure
+ SD card)
5 = 4 + 1
(CO2 filter + SD card)
6 = 4 + 2
(CO2 filter + gas pressure)
9 = 8 + 1
(Gas amount
+ SD card)
10 = 8 + 2
(Gas amount
+ gas pressure)
12 = 8 + 4
(Gas amount
+ CO2 filter)
Sensor Calibration (Calibration)
Turn the G460 on before placing it into the docking station. If the AUTOCAL button is pressed within the first
10 seconds, the calibration is started. If no button is pressed, the function test or calibration starts
automatically after 10 seconds, depending on which interval has expired.
During sensor calibration, the calibration is first performed with fresh air. On sensors for toxic and
inflammable gases, the zero point is set. The oxygen sensor is thereby set to 20.9 % v/v. The IR sensor for
carbon dioxide is calibrated to its zero point here using the CO2 absorber. Attention must be paid that
uncontaminated air is supplied to the docking station.
The AutoCal-Gas function is used to set the sensitivity of the sensors. Always use a suitable test gas or test
gas mixture for calibration with gas.
The sensors are purged after sensor calibration. The calibration takes approx. 3-4 minutes, on the DS404
longer depending on the gas inlets used. The calibration interval is specified in accordance with T021/T023.
The progress of the calibration as well as the zero report (air) and cal report (gas) are shown on the display
of the detector.
Function test starts after 10
seconds automatically
Purging
Sensor calibration
Purging
Sensor calibration
Result of sensor calibration with
fresh air (CO2-free) Reading
before => after
Results of sensor calibration with
test gas (gas mixture) Reading
before => after
The following symbols signal the calibration result:
-
= Sensor not calibrated
√
= Sensor successfully calibrated
= Error during calibration – not calibrated
Result of calibration:
Check the proper performance of the calibration by inspecting the zero and cal reports. The result of the
calibration is also indicated by a red and a green LED in the SD card slot of the docking station using the
traffic light principle. If the green LED is lit, the calibration was successful. The detector is now ready for use
again. If the red LED is lit or the display on the detector is red, however, an error has occurred which has to
be remedied before the detector can be used.
Calibration successful:
Display is green + green LED lit in slot
Calibration failed:
Display is red + red LED lit in slot
Display of the error which has occurred
(here Code 2: Flow error)
Possible error messages:
Code: 1 No gas defined – DS404 is not configured
Code: 2 Flow error – Fault in the zero or test gas supply
Code: 3 Aborted by detector – Detector problem / fault
Code: 4 Power error – Fault in supply voltage to TS400 power supply unit
Code: 5 Time too long for function test of the sensors
Code: 6 DS404 service necessary – The DS404 has to be serviced
Code: 7 No SD card – No SD card inserted
Code: 8 CO2 filter exhausted – CO2 filter exhausted
Code: 9 Parameter error – An error has been detected in the parameter memory
Code: 12 Firmware below 3.44 – The firmware version on the detector is too old
Code: 13 Gas switch error – The gas switch signals a fault
Additional information can be called up with the "Info" key. These messages do not lead to the abort of the
calibration.
1 = SD card
2 = Gas pressure
4 = CO2 filter
8 = Gas amount
If several errors occur at the same time, the error numbers are added:
3 = 2 + 1
(Gas pressure
+ SD card)
5 = 4 + 1
(CO2 filter + SD card)
6 = 4 + 2
(CO2 filter + gas pressure)
9 = 8 + 1
(Gas amount
+ SD card)
10 = 8 + 2
(Gas amount
+ gas pressure)
12 = 8 + 4
(Gas amount
+ CO2 filter)
Data Storage/Data Transfer to a PC
All information on the function tests and calibrations of the individual sensors are automatically stored on an
SD card (if inserted). The data transfer to a PC is effected automatically when the docking station is
connected to the PC. The data on the SD card can also be evaluated by means of a card reader, however,
without having the docking station connected to a PC. The data is transferred using the DS400 software.
Test data for storage both in the PC database and on the SD card:
Detector information:
Type of detector
Serial number
Software version
Sensor serial number
Gas type
Detection ranges
Alarm thresholds: A1, A2, A3, STEL, TWA
Battery capacity
Confidence beep ON/OFF
Docking station:
Date and time
Mode: Function test or sensor calibration
Test gas and gas concentration
Gas cylinder number
Test result
Function test:
Horn (audible alarm)
LED (visual alarm)
tAlarm1
tAlarm2
t50 or t90
Sensor calibration:
Zero point (ZP) before sensor calibration
Zero point (ZP) after sensor calibration
Sensitivity (CAL) before sensor calibration
Sensitivity (CAL) after sensor calibration
DS400 Serial No. and name
DS400 ID
DS404 firmware version
Install the program first with the SETUP program, which is started from the CD. When the CD is inserted,
the setup program starts automatically. If this does not work, select the relevant CD drive in the Explorer,
start the program manually by means of a double-click on the program icon and follow the instructions of
the program. You may either use the suggested directory name and drive designation or select a different
path. The program is then installed on the selected drive and directory. After installation, the program group
“GfG\DS(TS)400” and the icons for starting the program are created.
Installation of the USB Driver
Connect the RS485 USB interface cable to your PC. Windows recognises the new hardware and asks for the
driver. The driver is on the CD in directory USB_V206. Make Windows install the driver from this directory or
search for the driver on the CD.
Two driver components are installed. The first driver is the USB driver, the second driver the virtual COM
port. Depending on the PC configuration, a message will be displayed that the driver did not pass the
Windows logo test under Windows XP or Windows 7. This message can be ignored. After successful
installation you can see in the device manager (Start->System control->System->Device manager) which
COM port was reserved for the USB adapter. The designation of the COM port, e.g.: COM4, is required for
the setting in the software.
Configuration of the DS404
The docking station can be configured with the configuration program TS400Config.EXE from Version 3.5.3.
This includes in particular the setting of the test gas concentrations.
Connection
Click on the arrow in the port selection box to select the
correct COM port. Normally the software suggests the correct
port automatically. In field DS400/4 ID, select the DS404 with
which you wish to communicate. This is normally the ID "1".
Now click button "Connect". The connection to DS404 is now
established and the page "General" is displayed.
General
General information of the DS400 is entered and displayed
here.
DS400/4 alias name: Input of an alternative name for the
docking station.
ID: Change the ID of the connected DS404 –
only necessary if several DS404 are connected via patch cable.
In the field "Last service" you can see the date on which the
DS404 was last serviced.
The field "Next service" indicates when the DS404 is to be
serviced next.
Note: When this date is reached, the message "DS404 service necessary" appears on the display
of the detector. If the date is exceeded by 21 days, no further function tests or calibrations
can be carried out. The tests are aborted with the message “DS404 service necessary”.
The parameters for the CO2 absorption filter are set here.
Here you have to indicate whether no filter, a new filter, or a
used filter is connected. If you select No filter, the zero point
for CO2 will not be calibrated at the next G460 calibration.
The DS404 automatically counts the current cycle time of the
filter. When the filter reaches its warning threshold, the
message "CO2 filter" is displayed. The filter should be replaced
at the next opportunity. If the filter reaches the status Filter
exhausted, no further AutoCal adjustments will be carried out
and the message "CO2 filter exhausted!" appears on the
display of the gas detector.
Gas Input 1
The connected test gas is specified here. Function tests and
calibration are not possible without these inputs. Up to six
different gases of a test gas mixture can be entered here. Only
% vol. or ppm can be selected as units. For combustible
gases, the DS404 converts the test gas concentration
automatically from % vol. or ppm to % LEL.
Gas cylinder No.:
Enter the number or a name for the gas cylinder.
Gas hose length (m):
Input of the hose length to the gas cylinder necessary
This configuration program and the docking stations accept a wide range of test gas values. The gas
detectors require test gases with gas concentrations between 10 and 105 % of the respective measuring
range for calibration. Outside this range, no calibration is performed with test gas. The use of test gases
with gas concentrations between 30 and 70 % of the respective measuring range is recommended.
After selecting and setting new test gases and test gas concentrations, the connection to the docking station
must be interrupted by clicking the button "Disconnect". The connection must then be established again by
clicking the button "Connect". Now check in the file "Gas" whether the transmitted test gas concentrations
correspond to those of the connected test gas cylinder.
Test Times
The parameters for the function test and calibration can be
changed in the file Test times. The times should only be
changed, however, for special applications, e.g. in the case of
extreme hose lengths. Under certain circumstances, setting
the times too short may hinder the function test or the sensor
calibration. Setting CAL purging times too long consumes test
gas unnecessarily.
Function test time max. (sec):
Input of the maximum time (seconds) for the function test
Function test purging time (sec): Input of the sensor
purging time (seconds) with fresh air after the function test to
avoid false alarms after the function test
Zero point preparation (sec): Input of the sensor purging time (seconds) with fresh air before setting of
the zero point
Calibration preparation (sec): Input of the sensor purging time (seconds) before calibration with test gas
Calibration purging time (sec): Input of the sensor purging time with fresh air after calibration with
test gas
Only Alarm1 and Alarm2: Only the time for Alarm 1 and Alarm 2 is tested
Additional T50 test: Test of Alarm1, 2 and the response time t50
Additional T90 test: Test of Alarm1, 2 and the response time t90
Clock
The time and date for the DS404 are set here.
Note: Date and time are saved in the log file on the SD card
during the function test and during sensor calibration. Ensure
that the PC clock, and hence the clock of the DS404, is set
correctly. If the internal clock of the DS400/4 outputs a wrong
date, the calibration of function test will be aborted with the
error:
Code 6 DS404 service necessary.
Intervals
The intervals for function test and calibration are set under the
tab Intervals.
Transfer intervals:
The intervals are automatically transferred to the gas
detectors of the G400 Series. If these intervals are exceeded,
a message is displayed when the detector is switched on. This
message can be acknowledged. With the field "Transfer
intervals" you can determine whether or not the set intervals
are to be transferred at each function test or calibration.
Function test interval (days):
A check of the setting of the function test interval every working day is
recommended. In the field "Function test interval" you can set the number
of days after which the detector demands a function test. If the value is
set to 0 days, the detector does not demand a function test.
Calibration interval (days):
The lengths of the calibration intervals should not exceed a period of 6
months. In the field "Calibration interval" you can set the number of days
after which the detector demands calibration. If the value is set to 0 days,
the detector does not demand calibration.
The intervals for the function test and calibration are not a substitute for the annual inspection of the gas
detector by an expert, see also operating manual of the Microtector II gas detectors.
With the field "Set time in detector" you determine whether the date and time in the detector is to be set to
the date and time of the docking station.
With the field "Switch OFF detector after test" you determine whether or not the detector is to be switched
off 10 minutes after a function test.
Soiling of the detector housing can be removed using a cloth dampened with water. Do not use solvents or
cleaning agents!
Maintenance and Inspection
Regular inspection of the CO2 absorption filter and hose connections. An annual inspection by an expert is
recommended.
Troubleshooting
Code Fault / message
Cause
Remedy
Red and green LED in SD
card slot lit continuously
Error in program memory or error
in main memory
Switch detector off and on again
Contact GfG Service, if necessary
1
"No gas defined"
DS404 is not configured or error in
parameter memory
Configure DS404
Contact GfG Service, if necessary
2
"Flow error!"
Fault in the gas supply
Open valve of the test gas cylinder
Remedy interruption in the test gas
supply
Contact GfG Service, if necessary
3
"Aborted by detector!"
Fault in gas detector
Switch DS404 and detector off and on
again
Repeat procedure
4
"Power error!"
Fault in operating voltage of the
DS404
Replace plug-in power supply unit
Contact GfG Service, if necessary
5
"Time too long!"
Function test of the sensors
taking too long
Check gas supply
6
"DS404 service necessary!"
Service date has been exceeded by max.
21 days
Contact GfG Service
7
"No SD card!"
SD card not inserted or cannot be
written
Insert or replace SD card
8
"CO2 filter exhausted!"
The CO2 absorption filter is exhausted
Replace CO2 absorption filter and reset
in the configuration program
9
"DS404 parameter error!"
Error in parameter main memory
Switch detector off and on again
Reconfigure, if necessary
Contact GfG Service, if necessary
12
"Firmware below 3.44"
Detector software version too old
Carry out detector update
Contact GfG Service, if necessary
13
"Gas switch error"
The gas switch is faulty
Contact GfG Service
I 1
"DS404 service due"
Service date exceeded or
clock incorrect or not set
Contact GfG Service
I 2
INFO "Gas pressure!"
Gas pressure too low if external gas
pressure switch is installed
Check test gas cylinder and replace, if
necessary
I 8
INFO "Gas amount!"
Test gas cylinder almost empty
Replace test gas cylinder and reset with
the configuration program
Accessories and Spare Parts
Description
Order No.
CO2 absorption filter for zero gas filtration for CO2 sensors
1000220
Gas connection fitting, Type 2001, 2-stage pressure reducer for calibration gas supply by
pump for test gas cylinders of Type 2 AL, 8 AL or 6D, maximum inlet pressure: 0-1000
psi/70bar
1319216
Gas connection fitting, Type 2003, 2-stage pressure reducer for calibration gas supply by
pump for high-pressure cylinders; incl. fitting for high-pressure cylinders (M19x1.5LH)
maximum inlet pressure: 0-3000 psi/207bar
1319217
Pressure switch for high-pressure cylinders incl. T-piece for installation on the gas
connection fitting, Type 2003, pressure setting range: 65-300 psi/4.5-20.7 bar
1319218
USB-RS485 interface cable for data transmission PC<->DS400/DS404
1319221
Patch cable for data transmission and connection of several docking stations,
in various lengths:
1m / 3m
5m / 8m
10m
1319222 / 23
1319224 / 25
1319226
Plug-in power supply unit DS400-PS1 (European plug)
1450403
Jack plug 6.3 mm for connection of a pressure switch
1450404
Double-sided jack plug 6.3 mm for cascading of the DS400/DS404
1450405
Test and charging cradle G400-DIC1D (for detector without pump)
1450406
Test and charging cradle G400-DIC2D (for detector with pump)
1450407
Test and charging cradle DIC888/999 (for Microtector III and Polytector III)
1990223
SD memory card for archiving instrument-specific test data and test routines, different
storage capacities
1450408
Database software for automatic storage of instrument data, tests, test dates and test
routines.
1319203
Technical Data
Type designation:
DS404
Display and control
elements:
Display and 3 buttons on G450 or G460
Power supply:
12 V DC / 1.25 A (2 input/output jacks)
Gas
Connections:
1 zero gas inlet, 4 test gas inlets, 1 gas outlet (for hose di≈5 mm)
Delivery rate:
Flow rate: 0.5 to 0.6 l/min; drawn in by internal pump (pressure-free)
Pressure monitoring:
4 electric inputs for floating contacts (2 input/output jacks)
Time
Function test:
20 - 60 sec. (depending on setting and test gas concentration)
Calibration:
2 - 10 minutes (depending on sensor type and number of test gases)
Data
Storage medium:
2 GB SD card (formatted with FAT or FAT16 file system)
Communication:
RS-485 (2x RJ45 modular jacks)
Housing
Protection class:
IP20
Material:
Plastic
Dimensions:
125x169x225 mm (WxHxD)
Weight:
1500 g including G400-DIC1D
GfG Gesellschaft für Gerätebau mbH
Klönnestraße 99 – D-44143 Dortmund
Telefon: +49 (0)231 – 564 00-0
Telefax: +49 (0)231 – 564 00-895
Internet: www.gasmessung.de
E-Mail: [email protected]
Firmware Version 3.04
211-100.24_BA_DS404.doc
Last edited: 16. February 2018
Subject to change