Fourier MultiLogPRO Operation & User’s Manual
Text of the manual
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
MultiLogPRO
User Guide
July 2012
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Contents
Introduction .................................................................................................................. 1
Chapter 1: MultiLogPRO .............................................................................................. 2
General ................................................................................................................... 2
MultiLogPRO System Contents ............................................................................ 2
External Connections ........................................................................................... 2
Battery .................................................................................................................. 5
AC/DC Adaptor..................................................................................................... 5
Automatic Shut-off ................................................................................................ 5
Standalone Operation ............................................................................................. 5
Front Panel Layout ............................................................................................... 5
Input Modes.......................................................................................................... 7
Auto ID ....................................................................................................................... 7
8 Inputs ...................................................................................................................... 7
Quick-Start ........................................................................................................... 7
Working with the MultiLogPRO Menus ............................................................... 10
Turning MultiLogPRO On and Off ........................................................................... 10
Main Menu Display .................................................................................................. 11
Menu Buttons ........................................................................................................... 11
Menu Icons and Commands .................................................................................... 12
Graphic Display .................................................................................................. 14
The Cursor ............................................................................................................... 14
Zooming ................................................................................................................... 14
Measuring Timing Events ................................................................................... 15
Stopwatch ................................................................................................................ 16
Time and Speed ....................................................................................................... 16
Time, Speed and Acceleration ................................................................................ 18
Select Sensors Manually .................................................................................... 19
Load the Last Setup ........................................................................................... 20
Configure Your MultiLogPRO ............................................................................. 21
Input Mode ............................................................................................................... 21
Clear Memory .......................................................................................................... 21
Distance Sensor ....................................................................................................... 21
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
DO2 Sensor Calibration ........................................................................................... 22
Screen Contrast ....................................................................................................... 22
Connect to a Computer or to a TI Calculator .......................................................... 22
Internal Clock and Calendar ............................................................................... 23
Clear the Memory ............................................................................................... 23
Choose the Right Setup ..................................................................................... 23
Programming Rules and Limitations .................................................................. 25
Sampling Points ....................................................................................................... 25
Sampling Rate ......................................................................................................... 25
Continuous Sampling ............................................................................................... 26
Sensor Calibration ................................................................................................. 27
Hardware Offset Calibration ............................................................................... 27
MultiLogPRO Automatic Zero Calibration .......................................................... 27
pH Temperature Compensation ......................................................................... 28
DO2 Calibration .................................................................................................. 28
MultiLab Sensor Calibration ............................................................................... 28
Factory Calibration (No Calibration Required) ................................................... 28
Chapter 2: Troubleshooting Guide ............................................................................. 29
Technical Support: ..................................................................................................... 30
Chapter 3: Specifications ........................................................................................... 31
The MultiLogPRO Data Logger ............................................................................. 31
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Introduction
The MultiLogPRO is a powerful standalone 12-bit data
logger with a clear LCD graphic display and a 128K internal
memory. Recorded data is displayed in the form of graphs,
tables, meters or digital displays, and can be analyzed with a
number of pre-programmed analysis functions.
The MultiLogPRO’s internal memory stores experiment
notes and instructions for carrying out the experiment, which
can be edited or expanded at any time. These features
enable MultiLogPRO to function independently from a
computer - a perfect solution for when there is a shortage of
computers in the laboratory.
MultiLogPRO can record data from up to 8 sensors simultaneously; it is capable of
recording at rates of up to 21,000 samples per second, and of collecting up to
100,000 samples in its internal memory.
MultiLogPRO is very easy to use because all its functions are broken down into 8
icon menus, and its four buttons can browse every menu and execute any of the
commands. In addition, when a sensor is plugged in, the MultiLogPRO automatically
recognizes the type of sensor, and will record the data with the appropriate units.
A rechargeable battery powers the data logger, which shuts off automatically after 15
minutes have passed since the time of the last data recording, since the time the last
button was pressed, or since the time the last communication with the PC.
MultiLogPRO includes a built-in, easy to use, accurate Timer module that enables
the measuring of several types of timing events.
Together with the TI–83 Plus calculator, MultiLogPRO becomes a complete portable,
handheld data collection and analysis system.
The large variety of sensors with which MultiLogPRO is compatible makes the unit a
true digital Lab, which can measure scientific phenomena from the fields of Physics,
Chemistry, Biology and the environmental sciences.
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Chapter 1: MultiLogPRO
This chapter will focus on the MultiLogPRO data collection device and includes
sections on:
How to operate the MultiLogPRO keypad
How to set up MultiLogPRO
How to connect MultiLogPRO to your computer
General
MultiLogPRO System Contents
The following list indicates the items included with the MultiLogPRO.
1. The MultiLogPRO data logger
2. USB or Serial Communication cables (see your package list)
3. MultiLab software installation CD
4. AC-DC adaptor
External Connections
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Figure 1: MultiLogPRO external connections
1. Sensor Inputs
Sensor input/output (I/O) sockets are marked on the MultiLogPRO casing as I/O-1,
I/O-2, I/O-3 and I/O-4. These sockets are used to connect the sensors. Normally, all
four sockets can be used simultaneously.
To connect a sensor to the MultiLogPRO use one of the mini-din cables. Plug one
end of the cable into the data logger - arrow facing up, and the other end into the
sensor - arrow facing down.
If you are using one sensor only connect it to input 1. If you are using two sensors
connect them to inputs 1 and 2, and so on.
In order to connect more than 4 sensors at a time, use Fourier splitter cables (Part
Number DT011), which will enable the connection of up to 8 sensors simultaneously.
When a splitter cable is connected, it must be connected to the socket in the correct
numerical order (e.g. for 5 sensors, connect the splitter cable to I/O-1). One of the
fn
?
3. Serial
connection
socket
4. Power input
(DC 9–12V)
1. Sensor
inputs
1
st
input
2
nd
input
3
rd
input
4
th
input
2. USB
connection
socket
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
two splitter cables is marked with arrows - that is the main input (the lower I/O
number), the second line is marked with the letter S (split) - indicating that it is the
secondary input (the higher I/O number). Connect up to four input splitters to split the
MultiLogPRO’s inputs starting with I/O-4 (the splitters must be connected in order):
•
I/O-1 splits into I/O-1 and I/O-8
•
I/O-2 splits into I/O-2 and I/O-7
•
I/O-3 splits into I/O-3 and I/O-6
•
I/O-4 splits into I/O-4 and I/O-5
Note: Before connecting the mini-din cable to the data logger or the sensor
sockets, make sure that the mini-din plug is correctly positioned in front of the
socket. Connecting the cable in an awkward position might cause damage to the
cable pins.
2. Computer USB communication socket
For USB communication between logger and the computer, connect the USB Type B
plug to the MultiLogPRO and the USB Type A plug to the computer.
3. Computer serial communication socket
For Serial communication between logger and computer, connect the 9-pin Type D
cable between this port and one of the computer’s COM ports. At the back of the
computer you will find a 9-pin Type D male socket. If your computer is using this port
for the mouse, look for a 25-pin D shell male socket near the mouse. Use the RS-232
9 to 25-pin adaptor to connect the MultiLogPRO serial cable to the 25-pin port (the
adaptor is included with the MultiLogPRO).
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
4. External DC power supply socket
Plug in an AC/DC 9-12V adaptor whenever you want to save battery power, or to
charge the battery when necessary. Connecting external power to the MultiLogPRO
automatically charges the internal battery. The adaptor should meet the required
specifications (see section 0).
Battery
MultiLogPRO is equipped with a 7.2V NiCad rechargeable battery. Before you first
start working with MultiLogPRO, charge the unit for 10 to 12 hours while it is turned
off. Battery life is approximately 24 hours between charges.
If the data logger’s main battery runs out, the internal 3V Lithium battery backs up
the memory, so no data will be lost.
Note: Before storing the data logger make sure you have unplugged all the sensors
and pressed the OFF key.
AC/DC Adaptor
Output: Capacitor filtered 9 to 12 VDC, 400 mA
Female plug, center Negative
Automatic Shut-off
MultiLogPRO shuts off automatically after 15 minutes have passed since the time of
the last data recording, the time the last button was pressed, or the time the last
communication was made with the computer.
Standalone Operation
One way to program the MultiLogPRO is to use its keypad and screen (the other way
is to use the MultiLab software - see Error! Reference source not found.). The
keypad allows us to set all the parameters for data collection, while the LCD screen
displays the setting values.
Front Panel Layout
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Figure 2: MultiLogPRO front panel
LCD Display
Backward
Button
Forward
Button
On/Off
Buttons
Enter/Run
Button
Escape/Stop
Button
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Input Modes
MultiLogPRO has two Input Modes. The default mode is Auto ID and the
MultiLogPRO is in this mode when turned on for the first time. When using more than
4 sensors, when using third-party sensors, or if you’ve defined your own sensor,
switch to the 8 inputs mode. Select the system configuration menu in the MAIN
MENU and choose the mode you want (see page 21).
MultiLogPRO saves the current mode selection and will open in that mode the next
time you turn it on.
Auto ID
Automatic sensor identification – just plug in the sensors and the MultiLogPRO
prepares itself to collect data from these sensors.
While in this mode you can work with up to 4 sensors simultaneously.
8 Inputs
This mode enables data logging from up to 8 sensors simultaneously. Connect up to
four input splitters (DT011) to multiply the MultiLogPRO’s inputs starting with I/O-4
(the splitters must be connected in order):
•
I/O-1 splits into I/O-1 and I/O-8
•
I/O-2 splits into I/O-2 and I/O-7
•
I/O-3 splits into I/O-3 and I/O-6
•
I/O-4 splits into I/O-4 and I/O-5
In the splitter cable, one of the two wires is marked with arrows - that is the main
input (the lower I/O number). The second wire is marked with the letter S (split) -
indicating that it is the secondary input (the higher I/O number).
To learn how to switch to 8 inputs mode please refer to section 0 on page 21.
In this mode there is no automatic sensor identification. Select the sensors manually
either from the MultiLogPRO setup menu (see section 0 on page 15) or using the
MultiLab software (see section Error! Reference source not found. Error!
Bookmark not defined.).
Use this mode if you want to work with more than 4 sensors or if you are using third-
party sensors.
Quick-Start
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Before you first use MultiLogPRO, charge the unit for 10 to 12 hours while it is turned
off.
Turn on MultiLogPRO
Press the on button
. You will see the initialization screen. MultiLogPRO
performs a brief self-check and displays its status including battery level. At this point
you can load the last setup you used (see page 20). If you need a new setup wait
until you see the Main Menu screen:
Plug in the Sensors
Start with the first input on the right. If MultiLogPRO is in Auto ID mode (the default
mode) it will display the setup menu and will identify the sensors you’ve plugged in.
In–1: Microphone
In–2: Light
In–3:
In–4:
RATE = 10/s
SAMPLES = 500
DISPLAY = numeric
Note: Sensors must be added successively, starting with input-1. If a single sensor is
used it must be connected to I/O-1. If two sensors are used in an experiment, they
must be connected to I/O-1 and I/O-2.
The arrow indicates that MultiLogPRO is ready to accept a new sampling rate.
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
If the MultiLogPRO is in 8 input mode, you have to select the sensors manually (see
page 15).
Select Rate
Use the Forward
and Backward
arrow buttons to select the desired
rate, and then press the Enter button
.
In–1: Microphone
In–2: Light
In–3:
In–4:
RATE
= Every sec
SAMPLES = 500
DISPLAY = numeric
MultiLogPRO automatically switches to the next step in the setup process and the
arrow moves to the samples row.
Select Total Number of Samples
Use the Forward
and Backward
arrow buttons to select the number
of samples, and then press the Enter button
.
In–1: Microphone
In–2: Light
In–3:
In–4:
RATE
= Every sec
SAMPLES = 200
DISPLAY = numeric
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Choose Display
Use the Forward
and Backward
arrow buttons to select the type of
display, and then press the Enter button
.
In–1: Microphone
In–2: Light
In–3:
In–4:
SAMPLES = 200
DISPLAY = graphic
START
= ( > ).
Start Recording
Press the Forward arrow button
to start recording.
Or press the Enter button if you want to go back to the first item (Rate).
You can stop recording any time by pressing the Escape button
.
Working with the MultiLogPRO Menus
Turning MultiLogPRO On and Off
On
Turn MultiLogPRO on
Off
Turn MultiLogPRO off
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Note: Pressing OFF will not erase the sample memory. The data stored in the
memory will be kept for up to 10 years.
Main Menu Display
When turned on, MultiLogPRO opens with a system information window and then
displays the main menu:
MultiLogPRO has 9 menus. Use the Forward or Backward Arrow buttons to highlight
a menu and press the Enter button to select it. Then use the Arrow buttons to scan
the options. Press the Enter button to select an option. The MultiLogPRO
automatically executes the command.
Menu Buttons
Forward
Move to the next menu or to the next menu options
Backward
Move to the previous menu or menu options
Enter (Start)
Enter the selected menu or select the current menu
option and move to the next menu command or start
recording
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Escape
(Stop)
Return to the main menu or stop recording
Menu Icons and Commands
Start
Start recording
Press the Enter button to start recording
Setup
Setup MultiLogPRO in 4 steps:
Rate – Select recording rate
Samples – Select the total number of recording points
Display – Select the way MultiLogPRO will display the data (at a
rate of up to 10 samples per second):
Numeric - Displays the sensor values and the sample number
Meter - Displays all active sensors in a bar meter display
along with their values
Table - Displays the last 6 values of all the active sensors in a
table
Graphic - Displays a graphic representation of the sampled
sensors
At rates higher then 10/s the MultiLogPRO will display the data in
a graph at the end of the logging period.
Start – Press the Forward arrow to start recording
Function
Display statistics of the current data
Minimum – The minimum graph value
Maximum – The maximum graph value
Average – The graph average
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Open
Open a stored data in graphic display
Use the Forward and Backward Arrow buttons to browse the
stored files,
press the Enter button to open a file
Notes
Display experiment notes and instructions
Use the Arrow buttons to browse the notes.
You can use the MultiLab software to edit the notes or to write
new notes
Info
Display system information:
Number of experiments stored in MultiLogPRO’s internal
memory
Memory usage
MultiLogPRO version
Battery level (unplug any sensor and external power supply)
Current date and time
Note: When turning off the MultiLogPRO, the date and time is
reset. You restore the current date and time whenever the logger
communicates with MultiLab.
Timing
Enter to the timing module
You can select between the following timing events:
Stopwatch
Time and speed in one photogate
Time and speed between two photogates
Time, speed and acceleration between two photogates
Help
On-line help and specifications
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Configuration
Configure the MultiLogPRO:
Input Mode – Select Input Mode: Auto ID or 8 inputs
Clear memory – Delete the stored data files
DO2 – Press the Forward arrow button to enter the DO2 menu
Contrast – Use the arrow buttons to tune the screen contrast
Comm. – Press the Forward arrow button to connect to a
computer or the Backward arrow button to connect to TI
calculator
Graphic Display
MultiLogPRO will automatically use a graphic display in three cases:
If the recording rate is 10/s or less and you selected Graphic Display in the setup
menu
Once logging has ended for an experiment with a data recording rate of over 10/s
When opening a stored recording
The graphic display is available for recordings with up to 4 sensors simultaneously.
The Cursor
Use the Cursor in Graphic Display mode to read data values or to zoom in to the
area around a selected point. The cursor is displayed automatically after logging has
ended, or when opening a stored recording.
Use the Forward
and Backward
arrow buttons to
move the cursor.
MultiLogPRO displays the point coordinates at the bottom of the graph.
If there is more then one sensor, press the Enter button
to move
the cursor to another plot on the graph.
Zooming
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Position the cursor in the area you want to zoom into.
Press the two arrow buttons,
and
, simultaneously.
You will zoom in around the cursor in a 2:1 ratio.
Press the Escape button
to zoom out.
Measuring Timing Events
MultiLogPRO incorporates an accurate Timer module with resolution of 0.1ms. The
timer can measure several types of events triggered by Fourier Systems' photogates:
Time and speed with one photogate
Time and speed between two gates
Time, speed and acceleration between two gates
Stopwatch
To learn more about the different measurement types see page Error! Bookmark
not defined. and page Error! Bookmark not defined..
While in Timer mode MultiLogPRO does not store data. To save timing data in the
computer operate the Timer module via MultiLab.
To learn how to use the timer with MultiLab software please see section 4.9.
To use the photogates together with other sensors operate MultiLogPRO as a data
logger and then analyze the photogate data with the aid of MultiLab's Timing Wizard
(see page Error! Bookmark not defined.).
To perform a timing measurement:
Connect one or two photogates to MultiLogPRO
Use the Forward
and Backward
arrow buttons to navigate to
the Timing
menu
Press Enter
to display the timing menu:
_ _ _ _ _ T I M I N G _ _ _ _ _
M o d e : T i m e S p e e d
M e a s u r e : I n 1
C a r d w i d t h : 3 0 m m
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
G a t e s d i s t a n c e : 2 0 c m
S t a r t
( > )
The arrow indicates that MultiLogPRO is ready to accept the timing
mode
Use the arrow buttons to select the desired timing mode: Stopwatch,
Time Speed or Time Speed Acc
Stopwatch
_ _ _ _ _ T I M I N G _ _ _ _ _
M o d e : S t o p w a t c h
M e a s u r e :
_ _ _
C a r d w i d t h : _ _ _
G a t e s d i s t a n c e :
_ _ _
S t a r t ( > )
In this mode MultiLogPRO operates as a standard stopwatch with resolution of 0.01s.
Press Enter
four times, then press the Forward
arrow to
enter the stopwatch mode
Press the Forward
arrow to start measuring time
Press the Forward
arrow a second time to stop the watch
Press the Backward
arrow to reset the watch
Press Escape
to exit the stopwatch mode and return to the
timing menu
Press Escape
a second time to return to the main menu
Time and Speed
In this mode MultiLogPRO measures time and speed either in one gate or between
gates.
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
_ _ _ _ _ T I M I N G _ _ _ _ _
M o d e : T i m e S p e e d
M e a s u r e : I n 1
C a r d w i d t h : 3 0 m m
G a t e s d i s t a n c e : _ _ _
S t a r t ( > )
Press Enter
to select this mode. The arrow moves to the Measure row.
Use the Forward
and Backward
arrow buttons to select In1 if you want to
measure time and speed in one gate or In1 In2 if you want to measure time and
speed between two gates.
Time and speed at one gate
Time and speed between gates
MultiLogPRO measures the time
between blocking and unblocking the
photogate at input 1
MultiLogPRO measures the time
between blocking the photogate at input
1 and blocking the photogate at input 2
Time and speed at one gate
Time and speed between gates
1. Press Enter
, then use the arrow
buttons to select the card (the body
that blocks the gate) width in mm
between 0 to 59mm
1. Press Enter
twice, then use the
Forward
arrow button to select
the distance between the photogates
in cm between 0 to 99cm
2. Press Enter
twice, then use the
Forward
arrow button to enter to
a timing standby mode
2. Press Enter
, then use the
Forward
arrow button to enter to
a timing standby mode
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
3. Timing begins each time a body
blocks the photogate and ends when
unblocking
the
photogate.
MultiLogPRO
then
displays
the
elapsed time between entering and
leaving the gate and the speed of the
body
3. Timing begins each time a body
enters the first photogate (input 1)
and ends when leaving the second
photogate (input 2). MultiLogPRO
then
displays
the
elapsed
time
between entering the gates and the
body's average velocity
4. Use the arrow buttons to scroll
between the time and speed results
4. Use the arrow buttons to scroll
between the time and speed results
5. Press Escape
to return to the
timing menu
6. Press Escape
a second time to
return to the main menu
5. Press Escape
to return to the
timing menu
6. Press Escape
a second time to
return to the main menu
Time, Speed and Acceleration
In this mode MultiLogPRO measures the crossing time at the first gate, the time it
takes the body to move from one gate to the second gate and the crossing time at
the second gate and returns the time between gates, the average speed and the
average acceleration.
_ _ _ _ _ T I M I N G _ _ _ _ _
M o d e : T i m e S p e e d A c c
M e a s u r e : I n 1 – >
I n 2
C a r d w i d t h : 3 0 m m
G a t e s d i s t a n c e : 2 0 c m
S t a r t ( > )
Press Enter
twice, then use the arrow buttons to select the card
(the body that blocks the gate) width in mm between 0 to 59mm
Press Enter
, then use the Forward
arrow button to select the
distance between the photogates in cm between 0 to 99cm
Press Enter
, then use the Forward
arrow button to enter to a
timing standby mode
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Timing begins each time a body enters the first photogate (input 1) and
ends when leaving the second photogate (input 2). MultiLogPRO then
displays the elapsed time between entering the gates and the body's
average velocity and acceleration
Use the arrow buttons to scroll between the time, speed and
acceleration results
Press Escape
to return to the timing menu
Press Escape
a second time to return to the main menu
Select Sensors Manually
You must be in 8 input mode to be able to select the sensors manually (see page 7).
In the Main Menu screen, use the arrow buttons to select the setup
menu icon
.
Press the Enter button
to enter the setup menu:
In–1: Empty
In–2: Empty
In–3: Empty
In–4: Empty
RATE = 10/s
SAMPLES = 500
DISPLAY = numeric
Use the Forward
and Backward
arrows to select the
sensor in input 1 and then press the Enter button
. The arrow
indicator will move to the second input.
Repeat this procedure with all the sensors you plugged in.
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
After the fourth sensor has been selected, the screen will list the next
4 sensors:
In–5: Empty
In–6: Empty
In–7: Empty
In–8: Empty
RATE = 10/s
SAMPLES = 500
DISPLAY = numeric
You can press the Enter button in the last input if you want to go back to the first
input.
When you’ve finished selecting the sensors press the Escape button
. The arrow indicator will point to the Rate command and you
must complete the setup, as detailed in section 0.
Note: When you turn the MultiLogPRO off it will save the setup for the next session.
Load the Last Setup
When you turn MultiLogPRO on, once the self testing has been completed, it will
display:
_ _ _ I N I T I A L I Z A T I O N _ _ _
I n i t . D i s p l a y
I n i t . A D C
I n i t . R A M
B a t t . 9 9 %
E N T E R – l o a d s e t u p
Press the Enter button
to load the last setup.
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Configure Your MultiLogPRO
Use the System configuration menu to select the Input Mode, to clear the
MultiLogPRO’s memory, to change the screen contrast, or to calibrate the DO2
sensor.
In the MAIN MENU screen, select the System configuration icon
to display
the configuration screen. The default configuration is:
Input Mode: Auto ID
Clear memory ( > )
Distance 10m
DO2 calibration ( > )
Contrast ( < ) ( > )
Use the Forward
and Backward
arrows to select the mode and
then press the Enter button
to move to the next item. You can press the
Escape button
to leave the configuration menu at any time, saving the new
changes you made. Press Enter button in the last item (Contrast) if you want to go
back to the first item (Input Mode).
Input Mode
Select between Auto ID – automatic sensor identification and 8 inputs – manual
sensor selection (See also page 7).
The new configuration will be the default mode until the next time you change it.
Clear Memory
Press the Forward arrow button
if you want to delete all previous data files
from the MultiLogPRO.
Distance Sensor
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Press the Forward arrow button
to choose between the 2m and 10m
distance sensor you are using.
DO2 Sensor Calibration
If you need to calibrate a DO2 sensor press the Forward arrow button
to
enter the DO2 calibration screen:
_ _ _ _ _ D O 2 C A L _ _ _ _ _
U s e t h e f o l l o w i n g
m e n u t o c a l i b r a t e
t h e D O 2 e l e c t r o d e .
S a l i n i t y : 0 p p t
Use the arrow buttons to select between 0, 5, 10, 15, 20, 25, 30 and 35 ppt and
press the Enter button to confirm and move to calibrate altitude.
Use the arrow buttons to select between 0, 500, 1000, 1500, 2000, 2500, 3000 and
3500 ft and press the Enter button to confirm.
MultiLogPRO will display the calculated upper limit of the calibrated range and will
exit the DO2 calibration screen.
The new calibration parameters will be saved until the next time you change them.
Screen Contrast
Use the arrow buttons to adjust the LCD screen contrast.
Any contrast adjustment will be saved until the next time you change it.
Connect to a Computer or to a TI Calculator
Press the Forward arrow
button to connect to a computer or the Backward
arrow
button to connect to a TI calculator
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Internal Clock and Calendar
The internal clock is set the first time you use the Setup command from the MultiLab
software to program the MultiLogPRO. The clock is automatically updated to the
computer’s time and date whenever you connect your MultiLogPRO to a computer
and perform Setup in MultiLab.
When you shut off the MultiLogPRO, the internal clock and calendar is not saved. To
restore the clock and calendar, you must reconnect the MultiLogPRO to the
computer and Setup the logger again.
Clear the Memory
If you want to start recording and the MultiLogPRO’s internal memory is full you will
see this message at the bottom of the display:
In–1: Microphone
In–2: Light
In–3: Empty
In–4: Empty
SAMPLES = 200
DISPLAY = graphic
Mem full, clear = ( > )
Press the Forward arrow button
to clear the memory
In order to clear the MultiLogPRO’s memory when it is not full, use the Memory
clear command from the Configuration menu (see page 21), or clear the memory
from the Logger menu in the MultiLab software (see page Error! Bookmark not
defined.).
Choose the Right Setup
1. Sampling rate - The sampling rate should be determined by the frequency of
the phenomenon being sampled. If the phenomenon is periodic, sample at a
rate of at least twice the expected frequency. For example, sound recordings
should be sampled at the highest sampling rate – 20,800/sec, but changes in
room temperature can be measured at slower rates such as once per second or
even slower, depending on the speed of the expected changes.
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
There is no such thing as over-sampling. For extremely smooth graphs, the
sampling rate should be about 20 times the expected frequency.
Note: Sampling at a rate slower than the expected rate can cause frequency
aliasing. In such a case, the graph will show a frequency much lower than
expected. In Figure 3 below, the higher frequency sine wave was sampled at
1/3 of its frequency. Connecting the sampled points yielded a graph with a
lower, incorrect frequency.
Figure 3: Frequency Aliasing
Manual sampling - use this mode for:
Recordings or measurements that are not related to time.
Situations in which you have to stop recording data after each sample
obtained, in order to change your location, or any other logging
parameter (Note: During the experiment no changes can be made to
the MultiLogPRO’s configuration).
To start an experiment using manual data logging, set the RATE to manual and
press the Enter button
once to start the data recording, then press the
Enter button each time you want to collect a sample.
You can also perform manual logging via MultiLab (see page Error! Bookmark not
defined.).
2. Sampling Points - After you have chosen the sampling rate, choosing the
number of points will determine the logging period: Samples / Rate = Logging
time. You can also choose the duration of an experiment first, and then
calculate the number of samples: Samples = Logging time Rate.
Continuous
In the Continuous mode, MultiLogPRO does not save data, and can continue logging
indefinitely.
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
If MultiLogPRO is connected to the computer and the MultiLab software is running,
the data is automatically saved to the computer and displayed in a real-time graph.
To operate in Continuous mode select RATE equal to or less than 100/s and
SAMPLES = Continuous.
You can also select Continuous mode directly from the MultiLab software.
Note: MultiLogPRO must be set to a display mode other than graphic in order to
enable the Continuous mode.
Programming Rules and Limitations
The following are some rules and limitations you must take into account when
programming the MultiLogPRO, as MultiLogPRO integrates all programming
limitations automatically. MultiLogPRO will only allow the programming of settings
that comply with the rules below.
Sampling Points
Increasing the number of active inputs limits the number of
sampling points one can choose. The following condition must
be always satisfied: Samples Active Inputs < Memory.
MultiLogPRO’s memory is sufficient for 100,000 samples.
However, when sampling at rates faster than 100 samples per
second the memory can store only two experiments of 32,000
samples each.
Selection of 100,000 sampling points will create 2 files of 50,000
points each in the data logger’s memory
Sampling Rate
The number of sensors in use limits the maximum sampling rate:
Number of Sensors
Maximum Sampling Rate
1
20,800 samples per second
2
3,701 samples per second
3
2,631 samples per second
4
2,041 samples per second
5
1,667 samples per second
6
1,409 samples per second
7
1,220 samples per second
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
8
1,076 samples per second
Continuous Sampling
Continuous sampling is possible up to a maximum sampling rate of
100/s.
The data must be presented in a display mode other than graphical.
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Sensor Calibration
In this section you will learn how to increase the MultiLogPRO system accuracy using
calibration. Most of the sensors are linear, i.e. the output level of each sensor
changes according to the equation:
Where:
Y - Output of the sensor (voltage level changing from 0 to 5V)
X - Sensor input
a - Sensor gain
b - Sensor offset
The calibration process allows us to control the offset, and in some cases even the
gain, of a sensor. The MultiLogPRO system offers four types of calibration:
Hardware offset calibration
MultiLogPRO automatic zero calibration
MultiLab Sensor calibration
Factory calibration (no manual calibration required)
Hardware Offset Calibration
On some of the sensors there is a screw controlling the sensor offset. To calibrate
the sensor, rotate this screw until the sensor shows the correct measured value
(obtain the actual correct value from another source that is known to be accurate).
MultiLogPRO Automatic Zero Calibration
MultiLogPRO is able to automatically calibrate the sensor offset for all analog
sensors accurately, quickly, and for every new experiment conducted.
The calibration method is very simple. Whenever you plug in a sensor, the data
logger checks to see if the selected sensor measures a value within 2% of its zero
value. If so, MultiLogPRO sets that value as zero.
To enable this feature, make sure that the sensors are at their zero values when you
plug them in. To ensure the most accurate zero value, follow these instructions for
the following sensors:
Shorten the Voltage sensor plugs
Leave the Current sensor plugs open
Cover the Light, Photo Gate, and Microphone sensors
b
aX
Y
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Insert the pH sensor in a pH 7 solution
Unload the Force Transducer
Place the Accelerometer on a stationary surface
Place the Temperature probes in ice water
Place the Pressure sensor in a 1 ATM (1013 mb) chamber
pH Temperature Compensation
To compensate a pH sensor for temperature changes, plug the temperature sensor
into Input 1, and the pH sensor in Input 2. MultiLogPRO will then display the
compensated pH value.
DO2 Calibration
To calibrate your DO2 sensor with the right salinity and altitude parameters, use the
DO2 Calibration command from the Configuration menu (see page 21).
MultiLab Sensor Calibration
The sensors can also be calibrated using the Calibrate Sensors option in MultiLab’s
Logger menu. Please refer to section Error! Reference source not found. for more
information.
Factory Calibration (No Calibration Required)
All digital sensors that are essentially timers leave the factory fully calibrated, and do
not suffer from any accuracy degradation. An example of such a sensor is the Sonic
Ranger distance sensor, which measures the time passed from the transmission of a
sound pulse to its echo reception.
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
Chapter 2: Troubleshooting Guide
Symptom
Cause of Problem
I pressed the ON key on MultiLogPRO
and it did not turn on.
Try charging the battery, it
may be weak.
You are using an
unsuitable AC/DC adaptor
(see page 5).
The internal 3V disk battery is
dead.
MultiLab was unable to connect to
MultiLogPRO.
MultiLogPRO is turned off.
The communication cable is
not connected properly.
The communication cable is
not connected to the correct
computer COM Port.
There is no available
Serial COM Port.
The Logger USB driver was not
installed.
I downloaded unexpected data from the
MultiLogPRO.
There
may
be
up
to
100
experiments
stored
in
the
MultiLogPRO.
Performing
a
Download means loading one of
them in sequential order. You
have probably loaded an old
experiment
accidentally.
Try
using Selective download from
MultiLab
(see
page
Error!
Bookmark not defined.)
While running with an active trigger,
MultiLogPRO does not complete the data
logging.
The trigger condition has
not been fulfilled.
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
30
The trigger condition was fulfilled,
but the time period was too short
(less than 300S).
I started a new logging session and the
data
logger
stopped
immediately
afterwards.
First check if the sampling rate
was so fast that the experiment
began and ended in less than a
second. If this is not the reason,
recharge the 7.2 V battery of the
MultiLogPRO
and
try
again.
MultiLogPRO monitors the battery
power during the logging. If the
battery voltage is less then 5V,
the
logging
is
stopped
automatically.
MultiLogPRO’s memory is empty – there
are no stored files or experiment notes.
The internal 3 V disk
battery is dead. Replace
the battery.
Technical Support:
Fourier Help Desk:
E‐mail: [email protected]
Contact information:
1 866 771 6682 (toll‐free from within USA only)
1 708 478 5333
Hours of Operation:
Monday ‐ Friday, 9AM to 5PM (UTC ‐06:00)
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
31
Chapter 3: Specifications
The MultiLogPRO Data Logger
Input
4 Mini-Din sensor port interface
Auto ID mode:
Any combination of digital and analog inputs, with maximum of 2 digital
inputs and 4 analog inputs
Automatic sensor identification across 4 inputs
8 inputs mode:
Any combination of digital and analog inputs, with maximum of 2 digital
inputs and 8 analog inputs
Outputs
RS-232 computer Host Interface at 19,200 bps
USB 1.1 specification compliant
4 Digital Control Outputs
Sampling
Capacity: Up to 100,000 Samples
Analog sampling rate: From one sample per hour to 20,800 samples per
second
Digital sampling rate: >200 kHz
Resolution
12-bit (4096 Levels)
Timer Module
Time velocity and acceleration
Range: 0.0014 s to 6.5535 s
Resolution: 0.1 mS
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
32
Card width: 0 to 59 mm
Distance between gates: 0 to 99 cm
Stopwatch
Range: 0 to 99.99 s
Resolution: 10 mS
Man-machine Interface
Full keypad operation enabling manual programming of the MultiLogPRO
Graphic LCD
Display uses: displaying the logger setup, displaying experiment instructions and
notes and displaying the data measurements in graphs, meters, tables or digital
display
Features
Standalone operation - Working and sampling without connection to a computer.
Automatic or manual sensor identification
Saving and loading of last setup
Automatic self-test that reports system status to the user including battery level in
percentage of power remaining.
Trigger: programmable or manual
Automatic calibration of offset sensors
Backup memory battery: 3V replaceable Lithium battery saving samples data for
up to 10 years
Automatic data recovery if there is a power failure
Built in clock & calendar that keeps track of time and date for each data recording
Built in battery charger for charging the 7.2V internal battery
Automatic shut off after 15 minutes
Screen contrast control
Event recording
10 user defined sensors
Online notes editing
Post experiment graphical analysis
Power Supply
Fourier Systems Ltd.
16 Hamelacha St., POB 11681, Rosh Ha‘ayin 48091,
Tel: +972-3-901-4849, Fax: +972-3-901-4999
www.FourierEdu.com
EXPERIENCE SCIENCE!
33
Internal rechargeable 7.2V NiCad battery or external 9V to 12V DC input
Battery life: 750 mAH, 24 hours between charges
Operating temperature range: 0 to 50 °C
Dimensions: 182 x 100 x 28 mm
Weight: 450 gr
Standards compliance: CE, FCC