Fluke 5520A Service Getting Started
Read Online
Read the Fluke 5520A Service Getting Started online using the built-in PDF viewer.
Text of the manual
© 2003 Fluke Corporation, All rights reserved. Printed in USA
All product names are trademarks of their respective companies.
LIMITED WARRANTY AND LIMITATION OF LIABILITY
Each Fluke product is warranted to be free from defects in material and workmanship under
normal use and service. The warranty period is one year and begins on the date of shipment.
Parts, product repairs, and services are warranted for 90 days. This warranty extends only to the
original buyer or end-user customer of a Fluke authorized reseller, and does not apply to fuses,
disposable batteries, or to any product which, in Fluke’s opinion, has been misused, altered,
neglected, contaminated, or damaged by accident or abnormal conditions of operation or
handling. Fluke warrants that software will operate substantially in accordance with its functional
specifications for 90 days and that it has been properly recorded on non-defective media. Fluke
does not warrant that software will be error free or operate without interruption.
Fluke authorized resellers shall extend this warranty on new and unused products to end-user
customers only but have no authority to extend a greater or different warranty on behalf of Fluke.
Warranty support is available only if product is purchased through a Fluke authorized sales outlet
or Buyer has paid the applicable international price. Fluke reserves the right to invoice Buyer for
importation costs of repair/replacement parts when product purchased in one country is submitted
for repair in another country.
Fluke’s warranty obligation is limited, at Fluke’s option, to refund of the purchase price, free of
charge repair, or replacement of a defective product which is returned to a Fluke authorized
service center within the warranty period.
To obtain warranty service, contact your nearest Fluke authorized service center to obtain return
authorization information, then send the product to that service center, with a description of the
difficulty, postage and insurance prepaid (FOB Destination). Fluke assumes no risk for damage in
transit. Following warranty repair, the product will be returned to Buyer, transportation prepaid
(FOB Destination). If Fluke determines that failure was caused by neglect, misuse, contamination,
alteration, accident, or abnormal condition of operation or handling, including overvoltage failures
caused by use outside the product’s specified rating, or normal wear and tear of mechanical
components, Fluke will provide an estimate of repair costs and obtain authorization before
commencing the work. Following repair, the product will be returned to the Buyer transportation
prepaid and the Buyer will be billed for the repair and return transportation charges (FOB
Shipping Point).
THIS WARRANTY IS BUYER'S SOLE AND EXCLUSIVE REMEDY AND IS IN LIEU OF ALL
OTHER WARRANTIES, EXPRESS OR IMPLIED, INCLUDING BUT NOT LIMITED TO ANY
IMPLIED WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL, OR
CONSEQUENTIAL DAMAGES OR LOSSES, INCLUDING LOSS OF DATA, ARISING FROM
ANY CAUSE OR THEORY.
Since some countries or states do not allow limitation of the term of an implied warranty, or
exclusion or limitation of incidental or consequential damages, the limitations and exclusions of
this warranty may not apply to every buyer. If any provision of this Warranty is held invalid or
unenforceable by a court or other decision-maker of competent jurisdiction, such holding will not
affect the validity or enforceability of any other provision.
Fluke Corporation
P.O. Box 9090
Everett, WA 98206-9090
U.S.A.
Fluke Europe B.V.
P.O. Box 1186
5602 BD Eindhoven
The Netherlands
11/99
This Calibrator complies with IEC publication 1010-1 (1992-1), Safety Requirements for
Electrical Measuring, Control and Laboratory Equipment, and ANSI/ISA-S82.01-1994,
and CAN/CSA-C22.2 No. 1010.1-92. This manual contains information, warnings, and
cautions that must be followed to ensure safe operation and to maintain the Calibrator in
a safe condition. Use of this Calibrator in a manner not specified herein may impair the
protection provided by the Calibrator.
This Calibrator is designed for IEC 1010-1 Installation Category II use. It is not designed
for connection to circuits rated over 4800 VA.
Warning statements identify conditions or practices that could result in personal injury or
loss of life.
Caution statements identify conditions or practices that could result in damage to
equipment.
SYMBOLS MARKED ON THE CALIBRATOR
WARNING Risk of electric shock. Refer to the manual (see the Index for
references).
GROUND Ground terminal to chassis (earth).
Attention Refer to the manual (see the Index for references). This
symbol indicates that information about usage of a feature is contained in
the manual.
AC POWER SOURCE
The Calibrator is intended to operate from an ac power source that will not apply more
than 264V ac rms between the supply conductors or between either supply conductor
and ground. A protective ground connection by way of the grounding conductor in the
power cord is required for safe operation.
USE THE PROPER FUSE
To avoid fire hazard, use only the specified replacement fuse:
•
For 100 V or 120 V operation, use a 5A/250V time delay fuse (Fluke PN 109215).
•
For 220 V or 240 V operation, use a 2.5A/250V time delay fuse (Fluke PN 851931).
GROUNDING THE CALIBRATOR
The Calibrator uses controlled overvoltage techniques that require the Calibrator to be
grounded whenever normal mode or common mode ac voltages or transient voltages
may occur. The enclosure must be grounded through the grounding conductor of the
power cord, or through the rear panel CHASSIS GROUND binding post.
USE THE PROPER POWER CORD
Use only the power cord and connector appropriate for the voltage and plug
configuration in your country.
Use only a power cord that is in good condition.
Refer power cord and connector changes to qualified service personnel.
DO NOT OPERATE IN EXPLOSIVE ATMOSPHERES
To avoid explosion, do not operate the Calibrator in an atmosphere of explosive gas.
CHECK INSULATION RATINGS
Verify that the voltage applied to the unit under test does not exceed the insulation rating
of the UUT and the interconnecting cables.
DO NOT REMOVE COVER DURING OPERATION
To avoid personal injury or death, do not remove the Calibrator cover without first
removing the power source connected to the rear panel. Do not operate the Calibrator
without the cover properly installed. Normal calibration is accomplished with the cover
closed. Access procedures and the warnings for such procedures are contained in the
Service Manual. Service procedures are for qualified service personnel only.
DO NOT ATTEMPT TO OPERATE IF PROTECTION MAY BE IMPAIRED
If the Calibrator appears damaged or operates abnormally, protection may be impaired. Do
not attempt to operate the Calibrator under these conditions. Refer all questions of proper
Calibrator operation to qualified service personnel.
Table of Contents
Title
Page
Introduction ....................................................................................................... 1
Operation Overview .......................................................................................... 3
Local Operation............................................................................................. 3
Remote Operation (RS-232).......................................................................... 3
Remote Operation (IEEE-488)...................................................................... 3
Where to Go from Here..................................................................................... 4
Instruction Manuals........................................................................................... 5
5520A Getting Started Manual ..................................................................... 6
5520A Operators Manual.............................................................................. 6
5520A Operators Guide ................................................................................ 6
5520A Programmers Guide........................................................................... 6
5520A Service Manual.................................................................................. 6
Specifications .................................................................................................... 7
General Specifications .................................................................................. 8
DC Voltage Specifications............................................................................ 9
DC Current Specifications ............................................................................ 10
Resistance Specifications.............................................................................. 12
AC Voltage (Sine Wave) Specifications....................................................... 13
AC Current (Sine Wave) Specifications (cont)............................................. 16
Capacitance Specifications............................................................................ 17
Temperature Calibration (Thermocouple) Specifications............................. 18
Temperature Calibration (RTD) Specifications ............................................ 19
DC Power Specification Summary................................................................ 20
AC Power (45 Hz to 65 Hz) Specification Summary, PF=1......................... 20
Power and Dual Output Limit Specifications ............................................... 21
Phase Specifications...................................................................................... 22
Calculating Power Uncertainty ..................................................................... 23
Frequency Specifications .............................................................................. 24
Harmonics (2nd to 50th) Specifications........................................................ 25
AC Voltage (Sine Wave) Extended Bandwidth Specifications .................... 26
AC Voltage (Non-Sine Wave) Specifications............................................... 27
AC Voltage, DC Offset Specifications ......................................................... 28
AC Voltage, Square Wave Characteristics ................................................... 29
AC Voltage, Triangle Wave Characteristics (typical)................................... 29
AC Current (Sine Wave) Extended Bandwidth Specifications..................... 29
AC Current (Non-Sine Wave) Specifications ............................................... 30
AC Current, Square Wave Characteristics (typical) ..................................... 32
AC Current, Triangle Wave Characteristics (typical)................................... 32
If the 5520A Multi-Function Calibrator is operated in any way
not specified by this manual or other documentation provided
by Fluke, the protection provided by the Calibrator may be
impaired.
The 5520A Multi-Function Calibrator (referred to throughout this manual as either "the
5520A Calibrator" or "the Calibrator") is a fully programmable precision source of the
following:
•
DC voltage from 0 V to ±1000 V.
•
AC voltage from 1 mV to 1000 V, with output from 10 Hz to 500 kHz.
•
AC current from 100 µA to 20.5 A, with variable frequency limits.
•
DC current from 0 to ±20.5 A.
•
Resistance values from a short circuit to 1100 MΩ.
•
Capacitance values from 190 pF to 110 mF.
•
Simulated output for eight types of Resistance Temperature Detectors (RTDs).
•
Simulated output for eleven types of thermocouples.
0
•
1
2
3
4
5
6
7
8
9
ENTER
M
k
m
V
Hz
FIELD
EDIT
/
+
F
OPR
STBY
EARTH
SCOPE
EXGRD
MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
¡F
µ
n
p
W
dBm
sec
¡C
POWER
A
MULT
x
DIV
÷
MODES
MORE
5520A CALIBRATOR
HI
LO
TRIG
GUARD
TC
20A
NORMAL
AUX
SCOPE
OUT
V, , ,RTD
A, -SENSE, AUX V
20V PK MAX
20V PK MAX
nn030f.eps
Figure 1. 5520A Multi-Function Calibrator
Features of the 5520A Calibrator include the following:
•
Automatic meter error calculation.
•
X and D keys that change the output value to pre-determined cardinal values
for various functions.
•
Programmable entry limits that prevent invalid amounts from being entered.
•
Simultaneous output of voltage and current, up to an equivalent of 20.9 kW.
•
Pressure measurement when used with Fluke 700 Series pressure modules.
•
10 MHz reference input and output. Use this to input a high-accuracy 10 MHz
reference to transfer the frequency accuracy to the 5520A, or to synchronize one or
more additional 5520As to a master 5520A.
•
Simultaneous output of two voltages.
•
Extended bandwidth mode outputs multiple waveforms down to 0.01 Hz, and sine
waves to 2 MHz.
•
Variable phase signal output.
•
Standard IEEE-488 (GPIB) interface, complying with ANSI/IEEE Standards
488.1-1987 and 488.2-1987.
•
EIA Standard RS-232-C serial data interface for printing, displaying, or transferring
internally stored calibration constants, and for remote control of the 5520A.
•
Pass-through RS-232-C serial data interface for communicating with the Unit Under
Test (UUT).
The 5520A Calibrator may be operated at the front panel in the local mode, or remotely
using RS-232 or IEEE-488 ports. For remote operations, several software options are
available to integrate 5520A operation into a wide variety of calibration requirements.
Local Operation
Typical local operations include front panel connections to the Unit Under Test (UUT),
and then manual keystroke entries at the front panel to place the calibrator in the desired
output mode. The front panel layout facilitates hand movements from left to right, and
multiply and divide keys make it easy to step up or down at the press of a single key. You
can also review 5520A Calibrator specifications at the push of two buttons. The backlit
liquid crystal display is easy to read from many different viewing angles and lighting
conditions, and the large, easy-to-read keys are color-coded and provide tactile feedback.
Remote Operation (RS-232)
There are two rear-panel serial data RS-232 ports: SERIAL 1 FROM HOST, and
SERIAL 2 TO UUT (Figure 2). Each port is dedicated to serial data communications for
operating and controlling the 5520A during calibration procedures. For complete
information on remote operations, see Chapter 5 in the Operators Manual.
The SERIAL 1 FROM HOST serial data port connects a host terminal or personal
computer to the 5520A. You have several choices for sending commands to the 5520A:
you can enter commands from a terminal (or a PC running a terminal program), you can
write your own programs using BASIC, or you can run optional Windows-based software
such as 5500/CAL or MET/CAL. The 5500/CAL software includes more than 200
example procedures covering a wide range of test tools the 5520A can calibrate. (See
Chapter 6 in the Operators Manual for a discussion of the RS-232 commands.)
The SERIAL 2 TO UUT serial data port connects a UUT to a PC or terminal via the
5520A (see Figure 2). This “pass-through” configuration eliminates the requirement for
two COM ports at the PC or terminal. A set of four commands control the operation of
the SERIAL 2 TO UUT serial port. See Chapter 6 in the Operators Manual for a
discussion of the UUT_* commands. The SERIAL 2 TO UUT port is also used to
connect to the Fluke 700 series pressure modules.
Remote Operation (IEEE-488)
The 5520A rear panel IEEE-488 port is a fully programmable parallel interface bus
meeting standard IEEE-488.1 and supplemental standard IEEE-488.2. Under the remote
control of an instrument controller, the 5520A Calibrator operates exclusively as a
“talker/listener.” You can write your own programs using the IEEE-488 command set or
run the optional Windows-based MET/CAL software. (See Chapter 6 in the Operators
Manual for a discussion of the commands available for IEEE-488 operation.)
Unit Under Test
PC or Terminal
5520A
5520A
SERIAL 1 FROM HOST port
COM port
RS-232 Remote Operation using the
SERIAL 1 FROM HOST port
PC or Terminal
SERIAL 1 FROM HOST port
COM port
SERIAL 2
TO UUT port
RS-232 Remote Operation using the
SERIAL 1 FROM HOST and
SERIAL 2 TO UUT ports
nn031f.eps
Figure 2. RS-232 Remote Connections
Where to Go from Here
To locate specific information in the Operators Manual (provided as a pdf file on the CD-
ROM) refer to the following list:
•
Unpacking and setup: Chapter 2, “Preparing for Operation”
•
Installation and rack mounting: Chapter 2, “Preparing for Operation,” and the rack
mount kit instruction sheet
•
AC line power and interface cabling: Chapter 2, “Preparing for Operation”
•
Controls, indicators, and displays: Chapter 3, “Features”
•
Front panel operation: Chapter 4, “Front Panel Operation”
•
Cabling to a UUT (Unit Under Test): Chapter 4, “Front Panel Operation”
•
Remote operation (IEEE-488 or serial): Chapter 5, “Remote Operation”
•
Calibrating an Oscilloscope: Chapter 8, “Oscilloscope Calibration Options”
•
Accessories to the 5520A Calibrator: Chapter 9, “Accessories”
•
Performance Specifications: Chapter 1, “Introduction and Specifications”
The 5520A manual set provides complete information for operators and service or
maintenance technicians. The set includes:
•
5520A Getting Started Manual (PN 2075585)
•
5520A Operators Guide
•
English (PN 688754)
•
French (PN 688751)
•
German (PN 688762)
•
Italian (PN 690511)
•
Spanish (PN 688769)
•
Japanese (PN 688770)
•
Simplified Chinese (PN 688777)
•
5520A Programmers Guide (PN 688744)
•
5520A Operators Manual provided on CD-ROM (printed copy available for purchase
under PN 688739)
•
5520A Service Manual provided on CD-ROM ( printed copy available for purchase
under PN 688747)
5520A Getting Started Manual
This 5520A Getting Started Manual contains a brief introduction to the 5520A Manual
Set, instructions on how to get your calibrator prepared for operation and a complete set
of specifications.
5520A Operators Manual
The 5520A Operators Manual provides complete information for installing the 5520A
Calibrator and operating it from the front panel keys and in remote configurations. This
manual also provides a glossary of calibration, specifications, and error code information.
The Operators Manual includes the following topics:
•
Installation
•
Operating controls and features, including front panel operation
•
Remote operation (IEEE-488 bus or serial port remote control)
•
Serial port operation (printing, displaying, or transferring data, and setting up for
serial port remote control)
•
Operator maintenance, including verification procedures and calibration approach for
the 5520A
•
Oscilloscope calibration options
•
Accessories
5520A Operators Guide
The 5520A Operators Guide contains a summary of operating instructions, and a front
panel and rear panel feature reference. It is available in seven languages.
5520A Programmers Guide
The 5520A Programmers Guide contains a summary of remote commands and reference
information useful in determining system status using the status byte and related
registers.
5520A Service Manual
The 5520A Service Manual includes: theory of operation, performance testing,
maintenance, and calibration information.
The following tables list the 5520A specifications. All specifications are valid after
allowing a warm-up period of 30 minutes, or twice the time the 5520A has been turned
off. (For example, if the 5520A has been turned off for 5 minutes, the warm-up period is
10 minutes.)
All specifications apply for the temperature and time period indicated. For temperatures
outside of tcal ±5 °C (tcal is the ambient temperature when the 5520A was calibrated), the
temperature coefficient as stated in the General Specifications must be applied.
The specifications also assume the Calibrator is zeroed every seven days or whenever the
ambient temperature changes more than 5 °C. The tightest ohms specifications are
maintained with a zero cal every 12 hours within ±1 °C of use. (See “Zeroing the
Calibrator” in Chapter 4 of the Operators Manual.)
Also see additional specifications later in this chapter for information on extended
specifications for ac voltage and current. The dimensional outline for the 5520A
Calibrator is shown in Figure 3.
POWER
I
O
0
•
1
2
3
4
5
6
7
8
9
ENTER
M
k
m
V
Hz
FIELD
EDIT
/
+
F
OPR
EARTH
EXGRD
SCOPE
MENU
PREV
SHIFT
RESET
CE
SETUP
REF
NEW
TC
MEAS
¡F
µ
n
p
W
dBm
sec
¡C
A
MULT
x
DIV
÷
MODES
MORE
STBY
43.2 cm (17 in)
47.0 cm (18.5 in)
6.4 cm
(2.5 in)
For Cable
Access
17.8 cm
(7 in)
HI
LO
TRIG
GUARD
TC
20A
NORMAL
AUX
5520A CALIBRATOR
SCOPE
OUT
V, , ,RTD
A, -SENSE, AUX V
20V PK MAX
20V PK MAX
nn032f.eps
Figure 3. 5520A Calibrator Dimensional Outline
Twice the time since last warmed up, to a maximum of 30 minutes.
Settling Time
Less than 5 seconds for all functions and ranges except as noted.
Standard Interfaces
IEEE-488 (GPIB), RS-232, 5725A Amplifier
Temperature Performance
• Operating: 0 °C to 50 °C
• Calibration (tcal): 15 °C to 35 °C
• Storage: -20 °C to 70 °C [3]
Temperature Coefficient
Temperature Coefficient for temperatures outside tcal +5 °C is 0.1X/°C of
the 90-day specification (or 1-year, as applicable) per °C.
Relative Humidity [1]
• Operating: <80% to 30 °C, <70% to 40 °C, <40% to 50 °C
• Storage: <95%, non-condensing
Altitude
• Operating: 3,050 m (10,000 ft) maximum
• Non-operating: 12,200 m (40,000 ft) maximum
Safety
Complies with IEC 1010-1 (1992-1); ANSI/ISA-S82.01-1994;
CAN/CSA-C22.2 No. 1010.1-92
Analog Low Isolation
20 V
EMC
Designed to comply with FCC Rules Part 15; VFG 243/1991. If used in
areas with Electromagnetic fields of 1 to 3 V/m, resistance outputs have a
floor adder of 0.508 •. Performance not specified above 3 V/m. This
instrument may be susceptible to electro-static discharge (ESD) from direct
contact to the binding posts. Good static aware practices should be
followed when handling this and other pieces of electronic equipment.
Line Power [2]
• Line Voltage (selectable): 100 V, 120 V, 220 V, 240 V
• Line Frequency: 47 Hz to 63 Hz
• Line Voltage Variation: ±10% about line voltage setting
Power Consumption
5500A Calibrator, 300 VA; 5725A Amplifier, 750 VA
Dimensions
5500A Calibrator:
• Height: 17.8 cm (7 inches), standard rack increment, plus 1.5 cm (0.6
inch) for feet on bottom of unit;
• Width: 43.2 cm (17 inches), standard rack width
• Depth: 47.3 cm (18.6 inches) overall 5725A Amplifier:
• Height, 13.3 cm (5.25 inches), standard rack increment, plus 1.5 cm (0.6
inch) for feet on bottom of unit;
• Width, 43.2 cm (17 inches), standard rack width
• Depth, 63.0 cm (24.8 inches) overall.
Weight (without options)
5500A Calibrator, 22 kg (49 lb); 5725A Amplifier 32 kg (70 pounds)
Absolute Uncertainty Definition
The 5500A specifications include stability, temperature coefficient, linearity,
line and load regulation, and the traceability of the external standards used
for calibration. You do not need to add anything to determine the total
specification of the 5520A for the temperature range indicated.
Specification Confidence
Interval
99%
[1] After long periods of storage at high humidity, a drying out period (with the power on) of at least one week may be
required.
[2] For optimal performance at full dual outputs (e.g. 1000 V, 20A) choose a line voltage setting that is ± 7.5% from
nominal.
[3] The DC Current ranges 0 to 1.09999 A and 1.1 A to 2.99999 A are sensitive to storage temperatures above 50 °C.
If the 5520A is stored above 50 °C for greater than 30 minutes, these ranges must be re-calibrated. Otherwise, the
90 day and 1 year uncertainties of these ranges double.
DC Voltage Specifications
Absolute Uncertainty, tcal ± 5
°C
± (ppm of output + µV)
Stability
Range
90 days
1 year
24 hours, ± 1 °C
± (ppm output + µV)
Resolution
µV
Max
Burden
[1]
0 to 329.9999 mV
15 + 1
20 + 1
3 + 1
0.1
50 Ω
0 to 3.299999 V
9 + 2
11 + 2
2 + 1.5
1
10 mA
0 to 32.99999 V
10 + 20
12 + 20
2 + 15
10
10 mA
30 V to 329.9999 V
15 + 150
18 + 150
2.5 + 100
100
5 mA
100 V to 1000.000 V
15 + 1500
18 + 1500
3 + 300
1000
5 mA
Auxiliary Output (dual output mode only) [2]
0 to 329.999 mV
300 + 350
400 + 350
30 + 100
1
5 mA
0.33 V to 3.29999 V
300 + 350
400 + 350
30 + 100
10
5 mA
3.3 V to 7 V
300 + 350
400 + 350
30 + 100
100
5 mA
TC Simulate and Measure in Linear 10 µV/°C and 1 mV/°C modes [3]
0 to 329.999 mV
40 + 3
50 + 3
5 + 2
0.1
10 Ω
[1]
Remote sensing is not provided. Output resistance is < 5 mΩ for outputs ≥ 0.33 V. The AUX output has an output
resistance of < 1 Ω. TC simulation has an output impedance of 10 Ω ± 1 Ω.
[2]
Two channels of dc voltage output are provided.
[3] TC simulating and measuring are not specified for operation in electromagnetic fields above 0.4 V/m.
Noise
Range
Bandwidth 0.1 Hz to 10 Hz p-p
± (ppm output + floor)
Bandwidth 10 Hz to 10 kHz rms
0 to 329.9999 mV
0 + 1 µV
6 µV
0 to 3.299999 V
0 + 10 µV
60 µV
0 to 32.99999 V
0 + 100 µV
600 µV
30 to 329.9999 V
10 + 1 mV
20 mV
100 to 1000.000 V
10 + 5 mV
20 mV
Auxiliary Output (dual output mode only) [1]
0 to 329.999 mV
0 + 5 µV
20 µV
0.33 V to 3.29.999 V
0 + 20 µV
200 µV
3.3 V to 7 V
0 + 100 µV
1000 µV
[1]
Two channels of dc voltage output are provided.
DC Current Specifications
Absolute Uncertainty,
tcal ± 5 °C
± (ppm of output + µA)
Range
90 days
1 year
Resolution
Max
Compliance
Voltage
V
Max
Inductive
Load
mH
0 to 329.999 mA
120 + 0.02
150 + 0.02
1 nA
10
0 to 3.29999 mA
80 + 0.05
100 + 0.05
0.01 mA
10
0 to 32.9999 mA
80 + 0.25
100 + 0.25
0.1 mA
7
0 to 329.999 mA
80 + 2.5
100 + 2.5
1 mA
7
400
0 to 1.09999 A
160 + 40
200 + 40
10 mA
6
1.1 to 2.99999 A
300 + 40
380 + 40
10 mA
6
0 to 10.9999 A
(20 A Range)
380 + 500
500 + 500
100 mA
4
11 to 20.5 A [1]
800 + 750 [2]
1000 + 750 [2]
100 mA
4
[1] Duty Cycle: Currents < 11 A may be provided continuously. For currents >11 A, see Figure 1-4. The current may be
provided 60-T-I minutes any 60 minute period where T is the temperature in °C (room temperature is about 23 °C)
and I is the output current in amperes. For example, 17 A, at 23 °C could be provided for 60-17-23 = 20 minutes
each hour.
[2] Floor specification is 1500 µA within 30 seconds of selecting operate. For operating times > 30 seconds, the floor
specification is 750 µA.
Noise
Range
Bandwidth
0.1 Hz to 10 Hz p-p
Bandwidth
10 Hz to 10 kHz rms
0 to 329.999 µA
2 nA
20 nA
0 to 3.29999 mA
20 nA
200 nA
0 to 32.9999 mA
200 nA
2.0 µA
0 to 329.999 mA
2000 nA
20 µA
0 to 2.99999 A
20 µA
1 mA
0 to 20.5 A
200 µA
10 mA
DC Current Specifications (cont)
Ambient
0 °C
20 °C
10 °C
30 °C
40 °C
Current (Amps)
Minutes per Hour
Duty Cycle (%)
0%
10%
20%
30%
40%
60%
70%
80%
50%
5
10
15
20
25
30
35
40
45
50
0
11
12
13
14
15
16
17
18
19
20
nn326f.eps
Figure 4. Allowable Duration of Current >11 A
Absolute Uncertainty, tcal ± 5 °C
± (ppm of output + floor) [2]
ppm of output
Floor
Time & temp since ohms zero cal
Range
[1]
90 days
1 year
12 hrs ± 1 °C
7 days ± 5 °C
Resolution
Ω
Allowable
Current
[3]
0 to
10.9999 Ω
35
40
0.001
0. 01
0.0001
1 mA to 125 mA
11 Ω to
32.9999 Ω
25
30
0.0015
0.015
0.0001
1 mA to 125 mA
33 Ω to
109.9999 Ω
22
28
0.0014
0.015
0.0001
1 mA to 70 mA
110 Ω to
329.9999 Ω
22
28
0.002
0.02
0.0001
1 mA to 40 mA
330 Ω to
1.099999 kΩ
22
28
0.002
0.02
0.001
1 mA to 18 mA
1.1 kΩ to
3.299999 kΩ
22
28
0.02
0.2
0.001
100 µA to 5 mA
3.3 kΩ to
10.99999 kΩ
22
28
0.02
0.1
0.01
100 µA to 1.8 mA
11 kΩ to
32.99999 kΩ
22
28
0.2
1
0.01
10 µA to 0.5 mA
33 kΩ to
109.9999 kΩ
22
28
0.2
1
0. 1
10 µA to 0.18 mA
110 kΩ to
329.9999 kΩ
25
32
2
10
0.1
1 µA to 0.05 mA
330 kΩ to
1.099999 MΩ
25
32
2
10
1
1 µA to 0.018 mA
1.1 MΩ to
3.299999 MΩ
40
60
30
150
1
250 nA to 5 µA
3.3 MΩ to
10.99999 MΩ
110
130
50
250
10
250 nA to 1.8 µA
11 MΩ to
32.99999 MΩ
200
250
2500
2500
10
25 nA to 500 nA
33 MΩ to
109.9999 MΩ
400
500
3000
3000
100
25 nA to 180 nA
110 MΩ to
329.9999 MΩ
2500
3000
100000
100000
1000
2.5 nA to 50 nA
330 MΩ to
1100 MΩ
12000
15000
500000
500000
10000
1 nA to 13 nA
[1]
Continuously variable from 0 Ω to 1.1 G Ω.
[2]
Applies for 4-WIRE compensation only. For 2-WIRE and 2-WIRE COMP, add 5 µV per Amp of stimulus current to
the floor specification. For example, in 2-WIRE mode, at 1 kΩ, the floor specification within 12 hours of an ohms
zero cal for a measurement current of 1 mA is:
0.002 Ω + 5 µV / 1 mA = (0.022 + 0.005) Ω =0.007Ω.
[3] For currents lower than shown, the floor adder increases by Floor(new) = Floor(old) x Imin/Iactual. For example, a 50 µA
stimulus measuring 100 Ω has a floor specification of: 0.0014 Ω x 1 mA/50 µA = 0.028 Ω, assuming an ohms zero
calibration within 12 hours.
AC Voltage (Sine Wave) Specifications
NORMAL (Normal Output)
Range
Frequency
Absolute Uncertainty,
tcal ± 5 °C
± (ppm of output + µV) Resolution
Max
Burden
Max Distortion
and Noise
10 Hz to 5 MHz
Bandwidth
90 days
1 year
± (% output +
floor)
1.0 mV to
10 Hz to 45 Hz
600 + 6
800 + 6
1 µV
50 Ω
0.15 + 90 µV
32.999 mV
45 Hz to 10 kHz
120 + 6
150 + 6
0.035 + 90 µV
10 kHz to 20 kHz
160 + 6
200 + 6
0.06 + 90 µV
20 kHz to 50 kHz
800 + 6
1000 + 6
0.15 + 90 µV
50 kHz to 100 kHz
3000 + 12
3500 + 12
0.25 + 90 µV
100 kHz to 500 kHz
6000 + 50
8000 + 50
0.3 + 90 µV [1]
33 mV to
10 Hz to 45 Hz
250 + 8
300 + 8
1 µV
50 Ω
0.15 + 90 µV
329.999 mV
45 Hz to 10 kHz
140 + 8
145 + 8
0.035 + 90 µV
10 kHz to 20 kHz
150 + 8
160 + 8
0.06 + 90 µV
20 kHz to 50 kHz
300 + 8
350 + 8
0.15 + 90 µV
50 kHz to 100 kHz
600 + 32
800 + 32
0.20 + 90 µV
100 kHz to 500 kHz
1600 + 70
2000 + 70
0.20 + 90 µV [1]
0.33 V to
10 Hz to 45 Hz
250 + 50
300 + 50
10 µV
10 mA
0.15 + 200 µV
3.29999 V
45 Hz to 10 kHz
140 + 60
150 + 60
0.035 + 200 µV
10 kHz to 20 kHz
160 + 60
190 + 60
0.06 + 200 µV
20 kHz to 50 kHz
250 + 50
300 + 50
0.15 + 200 µV
50 kHz to 100 kHz
550 + 125
700 + 125
0.20 + 200 µV
100 kHz to 500 kHz
2000 + 600
2400 + 600
0.20 + 200 µV [1]
3.3 V to
10 Hz to 45 Hz
250 + 650
300 + 650
100 µV
10 mA
0.15 + 2 mV
32.9999 V
45 Hz to 10 kHz
125 + 600
150 + 600
0.035 + 2 mV
10 kHz to 20 kHz
220 + 600
240 + 600
0.08 + 2 mV
20 kHz to 50 kHz
300 + 600
350 + 600
0.2 + 2 mV
50 kHz to 100 kHz
750 + 1600
900 + 1600
0.5 + 2 mV
33 V to
45 Hz to 1 kHz
150 + 2000
190 + 2000
1 mV
5 mA,
0.15 + 10 mV
329.999 V
1 kHz to 10 kHz
160 + 6000
200 + 6000
except
0.05 + 10 mV
10 kHz to 20 kHz
220 + 6000
250 + 6000
20 mA for
0.6 + 10 mV
20 kHz to 50 kHz
240 + 6000
300 + 6000
45 Hz to
0.8 + 10 mV
50 kHz to 100 kHz
1600 + 50000
2000 +
50000
65 Hz
1.0 + 10 mV
330 V to
1020 V
45 Hz to 1 kHz
250 + 10000
300 +
10000
10 mV
2 mA,
except
0.15 + 30 mV
1 kHz to 5 kHz
200 + 10000
250 +
10000
6 mA for
45 Hz to
0.07 + 30 mV
5 kHz to 10 kHz
250 + 10000
300 +
10000
65 Hz
0.07 + 30 mV
[1] Max Distortion for 100 kHz to 200 kHz. For 200 kHz to 500 kHz, the maximum distortion is 0.9% of
output + floor as shown.
Note
•
Remote sensing is not provided. Output resistance is < 5 mΩ for outputs ≥0.33 V. The AUX output resistance is
< 1Ω. The maximum load capacitance is 500 pF, subject to the maximum burden current limits.
AC Voltage (Sine Wave) Specifications (cont)
AUX (Auxiliary Output) [dual output mode only] [1]
Absolute Uncertainty,
tcal ± 5 °C
± (% of output + µV)
Max Distortion
and Noise
10 Hz to
100 kHz
Bandwidth
Range
Frequency
90 days
1 year
Res-
olution
Max
Burden
± (% output +
floor)
10 Hz to 20 Hz
0.15 + 370
0.2 + 370
0.2 + 200 µV
20 Hz to 45 Hz
0.08 + 370
0.1 + 370
0.06 + 200 µV
45 Hz to 1 kHz
0.08 + 370
0.1 + 370
0.08 + 200 µV
1 kHz to 5 kHz
0.15 + 450
0.2 + 450
0.3 + 200 µV
5 kHz to 10 kHz
0.3 + 450
0.4 + 450
0.6 + 200 µV
10 mV to
329.999 mV
10 kHz to 30 kHz
4.0 + 900
5.0 + 900
1 µV
5 mA
1 + 200 µV
10 Hz to 20 Hz
0.15 + 450
0.2 + 450
0.2 + 200 µV
20 Hz to 45 Hz
0.08 + 450
0.1 + 450
0.06 + 200 µV
45 Hz to 1 kHz
0.07 + 450
0.09 + 450
0.08 + 200 µV
1 kHz to 5 kHz
0.15 + 1400
0.2 + 1400
0.3 + 200 µV
5 kHz to 10 kHz
0.3 + 1400
0.4 + 1400
0.6 + 200 µV
0.33 V to
3.29999 V
10 kHz to 30 kHz
4.0 + 2800
5.0 + 2800
10 µV
5 mA
1 + 200 µV
10 Hz to 20 Hz
0.15 + 450
0.2 + 450
0.2 + 200 µV
20 Hz to 45 Hz
0.08 + 450
0.1 + 450
0.06 + 200 µV
45 Hz to 1 kHz
0.07 + 450
0.09 + 450
0.08 + 200 µV
1 kHz to 5 kHz
0.15 + 1400
0.2 + 1400
0.3 + 200 µV
3.3 V to 5 V
5 kHz to 10 kHz
0.3 +1400
0.4 + 1400
100 µV
5 mA
0.6 + 200 µV
[1]
There are two channels of voltage output. The maximum frequency of the dual output is 30 kHz.
Note
•
Remote sensing is not provided. Output resistance is < 5 mΩ for outputs ≥0.33 V. The AUX output resistance is
< 1Ω. The maximum load capacitance is 500 pF, subject to the maximum burden current limits
AC Current (Sine Wave) Specifications
LCOMP off
Max
Inductive
Load
Range
Frequency
Absolute Uncertainty,
tcal ± 5 °C
± (% of output + µA)
Compliance
adder
± (µA/V)
Max
Distortion &
Noise 10 Hz
to 100 kHz
BW
± (% output
+
90 days
1 year
floor)
µH
29.00 µA to
10 Hz to 20 Hz
0.16 + 0.1
0.2 + 0.1
0.05
0.15 + 0.5 µA
200
329.99 µA
20 Hz to 45 Hz
0.12 + 0.1
0.15 + 0.1
0.05
0.1 + 0.5 µA
45 Hz to 1 kHz
0.1 + 0.1
0.125 + 0.1
0.05
0.05 + 0.5 µA
1 kHz to 5 kHz
0.25 + 0.15
0.3 + 0.15
1.5
0.5 + 0.5 µA
5 kHz to 10 kHz
0.6 + 0.2
0.8 + 0.2
1.5
1.0 + 0.5 µA
10 kHz to 30 kHz
1.2 + 0.4
1.6 + 0.4
10
1.2 + 0.5 µA
0.33 mA to
10 Hz to 20 Hz
0.16 + 0.15
0.2 + 0.15
0.05
0.15 + 1.5 µA
200
3.2999 mA
20 Hz to 45 Hz
0.1 + 0.15
0.125 + 0.15
0.05
0.06 + 1.5 µA
45 Hz to 1 kHz
0.08 + 0.15
0.1+ 0.15
0.05
0.02 + 1.5 µA
1 kHz to 5 kHz
0.16 + 0.2
0.2 + 0.2
1.5
0.5 + 1.5 µA
5 kHz to 10 kHz
0.4 + 0.3
0.5 + 0.3
1.5
1.0 + 1.5 µA
10 kHz to 30 kHz
0.8 + 0.6
1.0 + 0.6
10
1.2 + 0.5 µA
3.3 mA to
10 Hz to 20 Hz
0.15 + 2
0.18 + 2
0.05
0.15 + 5 µA
50
32.999 mA
20 Hz to 45 Hz
0.075 + 2
0.09 + 2
0.05
0.05 + 5 µA
45 Hz to 1 kHz
0.035 + 2
0.04 + 2
0.05
0.07 + 5 µA
1 kHz to 5 kHz
0.065 + 2
0.08 + 2
1.5
0.3 + 5 µA
5 kHz to 10 kHz
0.16 + 3
0.2 + 3
1.5
0.7 + 5 µA
10 kHz to 30 kHz
0.32 + 4
0.4 + 4
10
1.0 + 0.5 µA
33 mA to
10 Hz to 20 Hz
0.15 +20
0.18 + 20
0.05
0.15 + 50 µA
50
329.99 mA
20 Hz to 45 Hz
0.075 + 20
0.09 + 20
0.05
0.05 + 50 µA
45 Hz to 1 kHz
0.035 + 20
0.04 + 20
0.05
0.02 + 50 µA
1 kHz to 5 kHz
0.08 + 50
0.10 + 50
1.5
0.03 + 50 µA
5 kHz to 10 kHz
0.16 + 100
0.2 + 100
1.5
0.1 + 50 µA
10 kHz to 30 kHz
0.32 + 200
0.4 + 200
10
0.6 + 50 µA
0.33 A to
10 Hz to 45 Hz
0.15 + 100
0.18 + 100
0.2 + 500 µA
2.5
1.09999 A
45 Hz to 1 kHz
0.036 + 100
0.05 + 100
0.07 + 500 µA
1 kHz to 5 kHz
0.5 + 1000
0.6 + 1000
[3]
1 + 500 µA
5 kHz to 10 kHz
2.0 + 5000
2.5 + 5000
[4]
2 + 500 µA
1.1 A to
10 Hz to 45 Hz
0.15 + 100
0.18 + 100
0.2 + 500 µA
2.5
2.99999 A
45 Hz to 1 kHz
0.05 + 100
0.06 + 100
0.07 + 500 µA
1 kHz to 5 kHz
0.5 + 1000
0.6 + 1000
[3]
1 + 500 µA
5 kHz to 10 kHz
2.0 + 5000
2.5 + 5000
[4]
2 + 500 µA
3 A to
45 Hz to 100 Hz
0.05 + 2000
0.06 + 2000
0.2 + 3 mA
1
10.9999 A
100 kHz to 1 kHz
0.08 + 2000
0.10 + 2000
0.1 + 3 mA
1 kHz to 5 kHz
2.5 + 2000
3.0 + 2000
0.8 + 3 mA
11A to
45 Hz to 100 Hz
0.1 + 5000
0.12 + 5000
0.2 + 3 mA
1
20.5 A
100 Hz to 1 kHz
0.13 + 5000
0.15 + 5000
0.1 + 3 mA
[2]
1 kHz to 5 kHz
2.5 + 5000
3.0 + 5000
0.8 + 3 mA
[1]
Max Distortion for 100 kHz to 200 kHz. For 200 kHz to 500 kHz, the maximum distortion is 0.9% of
output + floor as shown.
[2]
Duty Cycle: Currents < 11 A may be provided continuously. For currents > 11 A, see Figure 1-4. The
current may be provided 60-T-I minutes any 60 minute period where T is the temperature in °C
(room temperature is about 23°C) and I is the output current in Amps. For example, 17 A, at 23°C could be
provided for 60-17-23 = 20 minutes each hour.
[3]
For compliance voltages greater than 1 V, add 1 mA/V to the floor specification from 1 kHz to 5 kHz.
[4]
For compliance voltages greater than 1 V, add 5 mA/V to the floor specification from 5 kHz to 10 kHz.
AC Current (Sine Wave) Specifications (cont)
LCOMP on
Range
Frequency
Absolute Uncertainty, tcal ± 5
°C
± (% of output + µA)
Max Distortion
& Noise, 10 Hz
to 100 kHz BW
Max
Inductive
Load
90 days
1 year
± (% output +
µA)
µH
29.00 µA to
10 Hz to 100 Hz
0.2 + 0.2
0.25 + 0.2
0.1 + 1.0
329.99 µA
100 Hz to 1 kHz
0.5 + 0.5
0.6 + 0.5
0.05 + 1.0
0.33 mA to
10 Hz to 100 Hz
0.2 + 0.3
0.25 + 0.3
0.15 + 1.5
3.2999 mA
100 Hz to 1 kHz
0.5 + 0.8
0.6 + 0.8
0.06 + 1.5
3.3 mA to
10 Hz to 100 Hz
0.07 + 4
0.08 + 4
0.15 + 5
400
32.999 mA
100 Hz to 1 kHz
0.18 + 10
0.2 + 10
0.05 + 5
33 mA to
10 Hz to 100 Hz
0.07 + 40
0.08 + 40
0.15 + 50
329.99 mA
100 Hz to 1 kHz
0.18 + 100
0.2 + 100
0.05 + 50
0.33 A to
10 Hz to 100 Hz
0.1 + 200
0.12 + 200
0.2 + 500
2.99999 A
100 to 440 Hz
0.25 + 1000
0.3 + 1000
0.25 + 500
3 A to 20.5 A
10 Hz to 100 Hz
0.1 + 2000 [2]
0.12 + 2000 [2]
0.1 + 0
400 [4]
[1]
100 Hz to 1 kHz
0.8 + 5000 [3]
1.0 + 5000 [3]
0.5 + 0
[1] Duty Cycle: Currents < 11 A may be provided continuously. For currents >11 A, see Figure 1-4. The current may
be provided 60-T-I minutes any 60 minute period where T is the temperature in °C (room temperature is about 23
°C) and I is the output current in amperes. For example, 17 A, at 23 °C could be provided for 60-17-23 = 20
minutes each hour.
[2] For currents >11 A, Floor specification is 4000 µA within 30 seconds of selecting operate. For operating times >30
seconds, the floor specification is 2000 µA.
[3] For currents >11 A, Floor specification is 1000 µA within 30 seconds of selecting operate. For operating times >30
seconds, the floor specification is 5000 µA.
[4] Subject to compliance voltages limits.
Range
Resolution
µA
Max Compliance Voltage
V rms
0.029 mA to 0.32999 mA
0.01
7
0.33 mA to 3.29999 mA
0.01
7
3.3 mA to 32.9999 mA
0.1
5
33 mA to 329.999 mA
1
5
0.33 A to 2.99999 A
10
4
3 A to 20.5 A
100
3
[1] Subject to specification adder for compliance voltages greater than 1 V rms.
Absolute Uncertainty, tcal ± 5 °C
± (% of output + floor)
Allowed Frequency or
Charge-Discharge Rate
Range
90 days
1 year
Res-
olution
Min and Max
to Meet
Specification
Typical
Max for
<0.5%
Error
Typical
Max for
<1% Error
0.19 nF to
0.3999 nF
0.38 + 0.01 nF
0.5 + 0.01 nF
0.1 pF
10 Hz to 10 kHz
20 kHz
40 kHz
0.4 nF to
1.0999 nF
0.38 + 0.01 nF
0.5 + 0.01 nF
0.1 pF
10 Hz to 10 kHz
30 kHz
50 kHz
1.1 nF to
3.2999 nF
0.38 + 0.01 nF
0.5 + 0.01 nF
0.1 pF
10 Hz to 3 kHz
30 kHz
50 kHz
3.3 nF to
10.9999 nF
0.19 + 0.01 nF
0.25 + 0.01 nF
0.1 pF
10 Hz to 1 kHz
20 kHz
25 kHz
11 nF to
32.9999 nF
0.19 + 0.1 nF
0.25 + 0.1 nF
0.1 pF
10 Hz to 1 kHz
8 kHz
10 kHz
33 nF to
109.999 nF
0.19 + 0.1 nF
0.25 + 0.1 nF
1 pF
10 Hz to 1 kHz
4 kHz
6 kHz
110 nF to
329.999 nF
0.19 + 0.3 nF
0.25 + 0.3 nF
1 pF
10 Hz to 1 kHz
2.5 kHz
3.5 kHz
0.33 µF to
1.09999 µF
0.19 + 1 nF
0.25 + 1 nF
10 pF
10 Hz to 600 Hz
1.5 kHz
2 kHz
1.1 µF to
3.29999 µF
0.19 + 3 nF
0.25 + 3 nF
10 pF
10 Hz to 300 Hz
800 Hz
1 kHz
3.3 µF to
10.9999 µF
0.19 + 10 nF
0.25 + 10 nF
100 pF
10 Hz to 150 Hz
450 Hz
650 Hz
11 µF to
32.9999 µF
0.30 + 30 nF
0.40 + 30 nF
100 pF
10 Hz to 120 Hz
250 Hz
350 Hz
33 µF to
109.999 µF
0.34 + 100 nF
0.45 + 100 nF
1 nF
10 Hz to 80 Hz
150 Hz
200 Hz
110 µF to
329.999 µF
0.34 + 300 nF
0.45 + 300 nF
1 nF
0 to 50 Hz
80 Hz
120 Hz
0.33 µF to
1.09999mF
0.34 + 1 µF
0.45 + 1 µF
10 nF
0 to 20 Hz
45 Hz
65 Hz
1.1 mF to
3.2999 mF
0.34 + 3 µF
0.45 + 3 µF
10 nF
0 to 6 Hz
30 Hz
40 Hz
3.3 mF to
10.9999 mF
0.34 + 10 µF
0.45 + 10 µF
100 nF
0 to 2 Hz
15 Hz
20 Hz
11 mF to
32.9999 mF
0.7 + 30 µF
0.75 + 30 µF
100 nF
0 to 0.6 Hz
7.5 Hz
10 Hz
33 mF to
110 mF
1.0 + 100 µF
1.1 + 100 µF
10 µF
0 to 0.2 Hz
3 Hz
5 Hz
[1] The output is continuously variable from 190 pF to 110 mF.
[2] Specifications apply to both dc charge/discharge capacitance meters and ac RCL meters. The maximum allowable
peak voltage is 3 V. The maximum allowable peak current is 150 mA, with an rms limitation of 30 mA below 1.1 µF
and 100 mA for 1.1 µF and above.
[3] The maximum lead resistance for no additional error in 2-wire COMP mode is 10 Ω.
Temperature Calibration (Thermocouple) Specifications
TC
Type
Range
°C
Absolute Uncertainty
Source/Measure
tcal ±5 °C
± °C [3]
TC
Type
Range
°C
Absolute Uncertainty
Source/Measure
tcal ±5 °C
± °C [3]
[1]
[2]
90 days
1 year
[1]
[2]
90 days
1 year
600 to 800
0.42
0.44
-200 to -100
0.37
0.37
800 to 1000
0.34
0.34
L
-100 to 800
0.26
0.26
1000 to 1550
0.30
0.30
800 to 900
0.17
0.17
B
1550 to 1820
0.26
0.33
-200 to -100
0.30
0.40
0 to 150
0.23
0.30
-100 to -25
0.17
0.22
150 to 650
0.19
0.26
N
-25 to 120
0.15
0.19
C
650 to 1000
0.23
0.31
120 to 410
0.14
0.18
1000 to 1800
0.38
0.50
410 to 1300
0.21
0.27
1800 to 2316
0.63
0.84
0 to 250
0.48
0.57
-250 to -100
0.38
0.50
R
250 to 400
0.28
0.35
-100 to -25
0.12
0.16
400 to 1000
0.26
0.33
E
-25 to 350
0.10
0.14
1000 to 1767
0.30
0.40
350 to 650
0.12
0.16
0 to 250
0.47
0.47
650 to 1000
0.16
0.21
S
250 to 1000
0.30
0.36
-210 to -100
0.20
0.27
1000 to 1400
0.28
0.37
-100 to -30
0.12
0.16
1400 to 1767
0.34
0.46
J
-30 to 150
0.10
0.14
-250 to -150
0.48
0.63
150 to 760
0.13
0.17
T
-150 to 0
0.18
0.24
760 to 1200
0.18
0.23
0 to 120
0.12
0.16
-200 to -100
0.25
0.33
120 to 400
0.10
0.14
-100 to -25
0.14
0.18
U
-200 to 0
0.56
0.56
K
-25 to 120
0.12
0.16
0 to 600
0.27
0.27
120 to 1000
0.19
0.26
1000 to 1372
0.30
0.40
[1] Temperature standard ITS-90 or IPTS-68 is selectable.
TC simulating and measuring are not specified for operation in electromagnetic fields above 0.4 V/m.
[2] Resolution is 0.01 °C
[3] Does not include thermocouple error
Temperature Calibration (RTD) Specifications
RTD
Range
°C
Absolute
Uncertainty
tcal ±5 °C
± °C [2]
RTD
Range
°C
Absolute
Uncertainty
tcal ±5 °C
± °C [2]
Type
[1]
90 days
1 year
Type
[1]
90 days
1 year
-200 to -80
0.04
0.05
-200 to -80
0.03
0.04
-80 to 0
0.05
0.05
-80 to 0
0.04
0.05
0 to 100
0.07
0.07
0 to 100
0.05
0.05
100 to 300
0.08
0.09
100 to 260
0.06
0.06
300 to 400
0.09
0.10
260 to 300
0.07
0.08
400 to 630
0.10
0.12
300 to 400
0.07
0.08
Pt 395,
100 Ω
630 to 800
0.21
0.23
400 to 600
0.08
0.09
-200 to -80
0.04
0.05
Pt 385,
500 Ω
600 to 630
0.09
0.11
-80 to 0
0.05
0.05
-200 to -80
0.03
0.03
0 to 100
0.07
0.07
-80 to 0
0.03
0.03
100 to 300
0.08
0.09
0 to 100
0.03
0.04
300 to 400
0.09
0.10
100 to 260
0.04
0.05
Pt 3926,
100 Ω
400 to 630
0.10
0.12
260 to 300
0.05
0.06
-200 to -190
0.25
0.25
300 to 400
0.05
0.07
-190 to -80
0.04
0.04
400 to 600
0.06
0.07
-80 to 0
0.05
0.05
Pt 385,
1000 Ω
600 to 630
0.22
0.23
0 to 100
0.06
0.06
PtNi 385,
-80 to 0
0.06
0.08
100 to 260
0.06
0.07
120 Ω
0 to 100
0.07
0.08
260 to 300
0.07
0.08
(Ni120)
100 to 260
0.13
0.14
300 to 400
0.08
0.09
Cu 427,
-100 to 260
0.3
0.3
400 to 600
0.08
0.10
10 Ω [3]
Pt 3916,
100 Ω
600 to 630
0.21
0.23
-200 to -80
0.03
0.04
-80 to 0
0.03
0.04
0 to 100
0.04
0.04
100 to 260
0.04
0.05
260 to 300
0.11
0.12
300 to 400
0.12
0.13
400 to 600
0.12
0.14
Pt 385,
200 Ω
600 to 630
0.14
0.16
[1] Resolution is 0.003 °C
[2] Applies for COMP OFF (to the 5520A Calibrator front panel NORMAL terminals) and 2- wire and 4-wire
compensation.
[3] Based on MINCO Application Aid No. 18
DC Power Specification Summary
Current Range
0.33 mA to
329.99 mA
0.33 A to
2.9999 A
3 A to
20.5 A
Voltage Range
Absolute Uncertainty, tcal ± 5 °C, ± (% of watts output) [1]
90
days
33 mV to 1020 V
0.021
0.019 [2]
0.06 [2]
1 year
33 mV to 1020 V
0.023
0.022 [2]
0.07 [2]
[1] To determine dc power uncertainty with more precision, see the individual “AC Voltage Specifications,” “AC
Current Specifications,” and “Calculating Power Uncertainty.”
[2] Add 0.02% unless a settling time of 30 seconds is allowed for output currents > 10 A or for currents on the highest
two current ranges within 30 seconds of an output current > 10 A.
AC Power (45 Hz to 65 Hz) Specification Summary, PF=1
Current Range
3.3 mA to
8.999 mA
9 mA to
32.999 mA
33 mA to
89.99 mA
90 mA
to 329.99 mA
Voltage Range
Absolute Uncertainty, tcal ± 5 °C, ± (% of watts output) [1]
90
days
33 to 329.999 mV
0.13
0.09
0.13
0.09
330 mV to 1020 V
0.11
0.07
0.11
0.07
1 year
33 to 329.999 mV
0.14
0.10
0.14
0.10
330 mV to 1020 V
0.12
0.08
0.12
0.08
Current Range [2]
0.33 A to
0.8999 A
0.9 A to
2.1999 A
2.2 A to
4.4999 A
4.5 A to
20.5 A
Voltage Range
Absolute Uncertainty, tcal ± 5 °C, ± (% of watts output) [1]
90
days
33 to 329.999 mV
0.12
0.10
0.12
0.10
330 mV to 1020 V
0.10
0.08
0.11
0.09
1 year
33 to 329.999 mV
0.13
0.11
0.13
0.11
330 mV to 1020 V
0.11
0.09
0.12
0.10
[1]
To determine ac power uncertainty with more precision, see the individual “DC Voltage Specifications” and “DC
Current Specifications” and “Calculating Power Uncertainty.”
[2]
Add 0.02% unless a settling time of 30 seconds is allowed for output currents > 10A or for currents on the highest
two current ranges within 30 seconds of an output current > 10A.
Power and Dual Output Limit Specifications
Frequency
Voltages
(NORMAL)
Currents
Voltages
(AUX)
Power
Factor
(PF)
dc
0 to ±1020 V
0 to ± 20.5 A
0 to ± 7 V
10 Hz to 45 Hz
33 mV to 32.9999 V
3.3 mA to 2.99999 A
10 mV to 5 V
0 to 1
45 Hz to 65 Hz
33 mV to 1000 V
3.3 mA to 20.5 A
10 mV to 5 V
0 to 1
65 Hz to 500 Hz
330 mV to 1000 V
33 mA to 2.99999 A
100 mV to 5 V
0 to 1
65 Hz to 500 Hz
3.3 V to 1000 V
33 mA to 20.5 A
100 mV to 5 V
0 to 1
500 Hz to 1 kHz
330 mV to 1000 V
33 mA to 20.5 A
100 mV to 5 V
1
1 kHz to 5 kHz
3.3 V to 1000 V [1]
33 mA to 2.99999 A
100 mV to 5 V [1]
1
5 kHz to 10 kHz
3.3 V to 1000 V [2]
33 mA to 329.99 mA
1 V to 5 V [2]
1
[1] In dual voltage output mode, voltage is limited to 3.3 V to 500 V in the NORMAL output.
[2] In dual voltage output mode, voltage is limited to 3.3 V to 250 V in the NORMAL output.
Notes
•
The range of voltages and currents shown in “DC Voltage Specifications,” “DC Current Specifications,”
“AC Voltage (Sine Wave) Specifications,” and “AC Current (Sine Wave) Specifications” are available in
the power and dual output modes (except minimum current for ac power is 0.33 mA). However, only
those limits shown in this table are specified. See “Calculating Power Uncertainty” to determine the
uncertainty at these points.
•
The phase adjustment range for dual ac outputs is 0° to ±179.99°. The phase resolution for dual ac
outputs is 0.01°.
1-Year Absolute Uncertainty, tcal ± 5 °C, (∆ Φ °)
10 Hz to
65 Hz
65 Hz to
500 Hz
500 Hz to
1 kHz
1 kHz to
5 kHz
5 kHz to
10 kHz
10 kHz to
30 kHz
0.10°
0.25°
0.5°
2.5°
5°
10°
Power Uncertainty Adder due to Phase Error
Phase
(Φ)
Watts
Phase
(Φ)
VARs
PF
10 Hz to
65 Hz
65 Hz to
500 Hz
500 Hz
to 1 kHz
1 kHz to
5 kHz
5 kHz to
10 kHz
10 kHz to
30 kHz
0°
90°
1.000
0.00%
0.00%
0.00%
0.10%
0.38%
1.52%
10°
80°
0.985
0.03%
0.08%
0.16%
0.86%
1.92%
4.58%
20°
70°
0.940
0.06%
0.16%
0.32%
1.68%
3.55%
7.84%
30°
60°
0.866
0.10%
0.25%
0.51%
2.61%
5.41%
11.54%
40°
50°
0.766
0.15%
0.37%
0.74%
3.76%
7.69%
16.09%
50°
40°
0.643
0.21%
0.52%
1.04%
5.29%
10.77%
22.21%
60°
30°
0.500
0.30%
0.76%
1.52%
7.65%
15.48%
31.60%
70°
20°
0.342
0.48%
1.20%
2.40%
12.08%
24.33%
49.23%
80°
10°
0.174
0.99%
2.48%
4.95%
24.83%
49.81%
100.00%
90°
0°
0.000
Note
1. To calculate exact ac watts power adders due to phase uncertainty for values not
shown, use the following formula:
(
)
(
)
Adder
Cos
Cos
%
(
(
)
=
−
+
100 1
Φ
∆Φ
Φ
) . For example:
for a PF of .9205 (Φ = 23) and a phase uncertainty of ∆Φ = 0.15, the ac watts power
adder is:
(
)
(
)
Adder
Cos
Cos
%
(
(23
)
=
−
+
=
100 1
15
23
011%
.
)
.
Calculating Power Uncertainty
Overall uncertainty for power output in watts (or VARs) is based on the root sum square (rss) of
the individual uncertainties in percent for the selected voltage, current, and power factor
parameters:
Watts uncertainty
U
U
U
U
2
power
voltage
current
PFadder
=
+
+
2
2
VARs uncertainty
U
U
U
U
2
VARs
voltage
current
VARsadder
=
+
+
2
2
Because there are an infinite number of combinations, you should calculate the actual ac power
uncertainty for your selected parameters. The method of calculation is best shown in the
following examples (using 90-day specifications):
Example 1 Output: 100 V, 1 A, 60 Hz, Power Factor = 1.0 (Φ=0), 1 year specifications
Voltage Uncertainty Uncertainty for 100 V at 60 Hz is 150 ppm + 2 mV, totaling:
100 V x 190 x 10
-6 = 15 mV added to 2 mV = 17 mV. Expressed in percent:
17 mV/100 V x 100 = 0.017% (see “AC Voltage (Sine Wave) Specifications”).
Current Uncertainty Uncertainty for 1 A is 0.036% + 100 µA, totaling:
1 A x 0.00036 = 360 µA added to 100 µA = 0.46 mA. Expressed in percent:
0.46 mA/1 A x 100 = 0.046% (see “AC Current (Sine Waves) Specifications”).
PF Adder Watts Adder for PF = 1 (Φ=0) at 60 Hz is 0% (see “Phase Specifications”).
Total Watts Output Uncertainty = U
2
power =
+
+
=
0017
0046
0
0 049%
2
2
.
.
.
Example 2 Output: 100 V, 1 A, 400 Hz, Power Factor = 0.5 (Φ=60)
Voltage Uncertainty Uncertainty for 100 V at 400 Hz is, 150 ppm + 2 mV, totaling:
100 V x 190 x 10
-6 = 15 mV added to 2 mV = 17 mV. Expressed in percent:
17 mV/100 V x 100 = 0.017% (see “AC Voltage (Sine Wave) Specifications”).
Current Uncertainty Uncertainty for 1 A is 0.036% + 100 µA, totaling:
1 A x 0.00036 = 360 µA added to 100 µA = 0.46 mA. Expressed in percent:
0.46 mA/1A x 100 = 0.046% (see “AC Current (Sine Waves) Specifications”).
PF Adder Watts Adder for PF = 0.5 (Φ=60) at 400 Hz is 0.76% (see “Phase Specifications”).
Total Watts Output Uncertainty = U
2
power =
+
+
=
0017
0 046
0 76
076%
2
2
.
.
.
.
VARs When the Power Factor approaches 0.0, the watts output uncertainty becomes unrealistic
because the dominant characteristic is the VARs (volts-amps-reactive) output. In these cases,
calculate the Total VARs Output Uncertainty, as shown in example 3:
Example 3 Output: 100 V, 1 A, 60 Hz, Power Factor = 0.174 (Φ=80)
Voltage Uncertainty Uncertainty for 100 V at 400 Hz is, 150 ppm + 2 mV, totaling:
100 V x 190 x 10
-6 = 15 mV added to 2 mV = 17 mV. Expressed in percent:
17 mV/100 V x 100 = 0.017% (see “AC Voltage (Sine Wave) Specifications”).
Current Uncertainty Uncertainty for 1 A is 0.036% + 100 µA, totaling:
1 A x 0.00036 = 360 µA added to 100 µA = 0.46 mA. Expressed in percent:
0.46 mA/1 A x 100 = 0.046% (see “AC Current (Sine Waves) Specifications”).
VARs Adder VARs Adder for Φ = 80 at 60 Hz is 0.02% (see “Phase Specifications”).
Total VARS Output Uncertainty = U
2
VARs =
0 017
0 046
0 03
0 058%
2
2
.
.
.
.
+
+
=
Additional Specifications
The following paragraphs provide additional specifications for the 5520A Calibrator ac
voltage and ac current functions. These specifications are valid after allowing a warm-up
period of 30 minutes, or twice the time the 5520A has been turned off. All extended
range specifications are based on performing the internal zero-cal function at weekly
intervals, or when the ambient temperature changes by more than 5 °C. (See the 5520A
Operators Manual, Chapter 4, “Front Panel Operation.”)
Frequency Specifications
Frequency
1-Year Absolute Uncertainty,
Range
Resolution
tcal ± 5°C
Jitter
0.01 Hz to 119.99 Hz
0.01 Hz
120.0 Hz to 1199.9 Hz
0.1 Hz
1.200 kHz to 11.999 kHz
1.0 Hz
2.5 ppm ± 5 µHz [1]
100 nS
12.00 kHz to 119.99 kHz
10 Hz
120.0 kHz to 1199.9 kHz
100 Hz
1.200 MHz to 2.000 MHz
1 kHz
[1] With REF CLK set to ext, the frequency uncertainty of the 5520A is the uncertainty of the external
10 MHz clock ±5 µHz. The amplitude of the 10 MHz external reference clock signal should be between
1 V and 5 V p-p.
Harmonics (2nd to 50th) Specifications
Fundamental
Frequency [1]
Voltages
NORMAL
Terminals
Currents
Voltages
AUX Terminals
Amplitude
Uncertainty
10 Hz to 45 Hz
33 mV to 32.9999 V
3.3 mA to 2.99999 A
10 mV to 5 V
Same % of
output as
45 Hz to 65 Hz
33 mV to 1000 V
3.3 mA to 20.5 A
10 mV to 5 V
the
equivalent
65 Hz to 500 Hz
33 mV to 1000 V
33 mA to 20.5 A
100 mV to 5 V
single output,
but twice the
500 Hz to 5 kHz
330 mV to 1000 V
33 mA to 20.5 A
100 mV to 5 V
floor adder.
5 kHz to 10 kHz
3.3 V to 1000 V
33 mA to
329.9999 mA
100 mV to 5 V
10 kHz to 30 kHz
3.3 V to 1000 V
33 mA to
329.9999 mA
100 mV to
3.29999 V
[1]
The maximum frequency of the harmonic output is 30 kHz (10 kHz for 3 V to 5 V). For example, if the
fundamental output is 5 kHz, the maximum selection is the 6th harmonic (30 kHz). All harmonic
frequencies (2nd to 50th) are available for fundamental outputs between 10 Hz and 600 Hz (200 Hz
for 3 V to 5 V).
Note
1. Phase uncertainty for harmonic outputs is 1 degree, or the phase uncertainty shown in
“Phase Specifications” for the particular output, whichever is greater. For example,
the phase uncertainty of a 400 Hz fundamental output and 10 kHz harmonic output is
10° (from “Phase Specifications”). Another example, the phase uncertainty of a 60
Hz fundamental output and a 400 Hz harmonic output is 1 degree.
Example of determining Amplitude Uncertainty in a Dual Output Harmonic Mode
What are the amplitude uncertainties for the following dual outputs?
NORMAL (Fundamental) Output: 100 V, 100 Hz
From “AC Voltage (Sine Wave) Specifications” the single output specification for
100 V, 100 Hz, is 0.015% + 2 mV. For the dual output in this example, the specification
is 0.015% + 4 mV as the 0.015% is the same and the floor is twice the value (2 x 2 mV).
AUX (50th Harmonic) Output: 100 mV, 5 kHz
From “AC Voltage (Sine Wave) Specifications” the auxiliary output specification for
100 mV, 5 kHz, is 0.15% + 450 mV. For the dual output in this example, the specification
is 0.15% + 900 mV as the 0.15% is the same and the floor is twice the value (2 x 450 mV).
AC Voltage (Sine Wave) Extended Bandwidth Specifications
Range
Frequency
1-Year Absolute
Uncertainty
tcal ± 5 °C
Max Voltage
Resolution
Normal Channel (Single Output Mode)
1.0 mV to 33 mV
0.01 Hz to 9.99 Hz
± (5.0 % of output +
Two digits, e.g., 25 mV
34 mV to 330 mV
0.5% of range)
Three digits
0.4 V to 33 V
Two digits
0.3 V to 3.3 V
500.1 kHz to 1 MHz
-10 dB at 1 MHz,
typical
Two digits
1.001 MHz to 2 MHz
-31 dB at 2 MHz,
typical
Auxiliary Output (Dual Output Mode)
10 mV to 330 mV
0.01 Hz to 9.99 Hz
± (5.0 % of output +
Three digits
0.4 V to 5 V
0.5% of range)
Two digits
AC Voltage (Non-Sine Wave) Specifications
Triangle Wave &
Truncated Sine
Range, p-p [1]
Frequency
1-Year Absolute Uncertainty,
tcal ± 5 °C,
± (% of output + % of range) [2]
Max Voltage
Resolution
Normal Channel (Single Output Mode)
2.9 mV
0.01 Hz to 10 Hz
5.0 + 0.5
Two digits on
each range
to 93 V
10 Hz to 45 Hz
0.25 + 0.5
45 Hz to 1 kHz
0.25 + 0.25
Six digits on
1 kHz to 20 kHz
0.5 + 0.25
each range
20 kHz to 100 kHz [3]
5.0 + 0.5
Auxiliary Output (Dual Output Mode)
93 mV
0.01 Hz to 10 Hz
5.0 + 0.5
Two digits on
each range
to 14 V
10 Hz to 45 Hz
0.25 + 0.5
Six digits on
45 Hz to 1 kHz
0.25 + 0.25
each range
1 kHz to 10 kHz
5.0 + 0.5
[1]
To convert p-p to rms for triangle wave, multiply the p-p value by 0.2886751. To convert p-p to rms for truncated
sine wave, multiply the p-p value by 0.2165063.
[2]
Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.
[3]
Uncertainty for Truncated Sine outputs is typical over this frequency band.
Square
wave Range
(p-p) [1]
Frequency
1-Year Absolute Uncertainty,
tcal ± 5 °C
± (% of output + % of range) [2]
Max Voltage
Resolution
Normal Channel (Single Output Mode)
0.01 Hz to 10 Hz
5.0 + 0.5
Two digits on each range
2.9 mV
10 Hz to 45 Hz
0.25 + 0.5
To
45 Hz to 1 kHz
0.25 + 0.25
Six digits on each range
66 V
1 kHz to 20 kHz
0.5 + 0.25
20 kHz to 100 kHz
5.0 + 0.5
Auxiliary Output (Dual Output Mode)
66 mV
0.01 Hz to 10 Hz
5.0 + 0.5
Two digits on each range
To
10 Hz to 45 Hz
0.25 + 0.5
14 V
45 Hz to 1 kHz
0.25 + 0.25
Six digits on each range
1 kHz to 10 kHz [3]
5.0 + 0.5
[1]
To convert p-p to rms for square wave, multiply the p-p value by 0.5000000.
[2]
Uncertainty is stated in p-p. Amplitude is verified using an rms-responding DMM.
[3]
Limited to 1 kHz for Auxiliary outputs ≥ 6.6 V p-p.
AC Voltage, DC Offset Specifications
Range [1]
(Normal Channel)
Offset Range [2]
Max
Peak
Signal
1-Year Absolute Offset
Uncertainty, tcal ± 5 °C [3]
± (% dc output + floor)
Sine Waves (rms)
3.3 mV to 32.999 mV
0 to 50 mV
80 mV
0.1 + 33 µV
33 mV to 329.999 mV
0 to 500 mV
800 mV
0.1 + 330 µV
0.33 mV to 3.29999 V
0 to 5 V
8 V
0.1 + 3300 µV
3.3 V to 32.9999 V
0 to 50 V
55 V
0.1 + 33 mV
Triangle Waves and Truncated Sine Waves (p-p)
9.3 mV to 92.999 mV
0 to 50 mV
80 mV
0.1 + 93 µV
93 mV to 929.999 mV
0 to 500 mV
800 mV
0.1 + 930 µV
0.93 mV to 9.29999 V
0 to 5 V
8 V
0.1 + 9300 µV
9.3 mV to 92.9999 V
0 to 50 V
55 V
0.1 + 93 mV
Square Waves (p-p)
6.6 mV to 65.999 mV
0 to 50 mV
80 mV
0.1 + 66 µV
66 mV to 659.999 mV
0 to 500 mV
800 mV
0.1 + 660 µV
0.66 mV to 6.59999 V
0 to 5 V
8 V
0.1 + 6600 µV
6.6 mV to 65.9999 V
0 to 50 V
55 V
0.1 + 66 mV
[1]
Offsets are not allowed on ranges above the highest range shown above.
[2]
The maximum offset value is determined by the difference between the peak value of the selected
voltage output and the allowable maximum peak signal. For example, a 10 V p-p square wave output
has a peak value of 5 V, allowing a maximum offset up to ±50 V to not exceed the 55 V maximum
peak signal. The maximum offset values shown above are for the minimum outputs in each range.
[3]
For frequencies 0.01 Hz to 10 Hz, and 500 kHz to 2 MHz, the offset uncertainty is 5% of output, ±1%
of the offset range.
AC Voltage, Square Wave Characteristics
Risetime
@ 1 kHz
Typical
Settling
Time
@ 1 kHz
Typical
Overshoot
@ 1 kHz
Typical
Duty Cycle Range
Duty Cycle Uncertainty
<1 µs
<10 µs to 1%
of final value
<2%
1% to 99%, <3.3 V p-p,
0.01 Hz to 100 kHz
± (0.8% of period +100 ns)
± (0.02% of period + 100
ns), 50% duty cycle
± (0.05% of period + 100
ns), other duty cycles from
10% to 90%
AC Voltage, Triangle Wave Characteristics (typical)
Linearity to 1 kHz
Aberrations
0.3% of p-p value, from 10% to 90% point
<1% of p-p value, with amplitude >50% of range
AC Current (Sine Wave) Extended Bandwidth Specifications
Range
Frequency
1-Year Absolute Uncertainty
tcal ± 5 °C
± (% of output + % of range)
Max
Current
Resolution
All current ranges, <330 mA
0.01 Hz to 10 Hz
5.0 + 0.5
2 digits
AC Current (Non-Sine Wave) Specifications
Triangle Wave &
Truncated Sine Wave
Range
p-p
Frequency
1-Year Absolute Uncertainty
tcal ± 5 °C
± (% of output + % of range)
Max
Current
Resolution
0.047 mA
0.01 Hz o 10 Hz
5.0 + 0.5
Two digits
to 0.92999 mA [1]
10 Hz to 45 Hz
0.25 + 0.5
45 Hz to 1 kHz
0.25 + 0.25
Six digits
1 kHz to 10 kHz
10 + 2
0.93 mA to
0.01 Hz to 10 Hz
5.0 + 0.5
Two digits
9.29999 mA [1]
10 Hz to 45 Hz
0.25 + 0.5
45 Hz to 1 kHz
0.25 + 0.25
Six digits
1 kHz to 10 kHz
10 + 2
9.3 mA to
0.01 Hz to 10 Hz
5.0 + 0.5
Two digits
92.9999 mA [1]
10 Hz to 45 Hz
0.25 + 0.5
45 Hz to 1 kHz
0.25 + 0.25
Six digits
1 kHz to 10 kHz
10 + 2
93 mA to
0.01 Hz to 10 Hz
5.0 + 0.5
Two digits
929.999 mA [1]
10 Hz to 45 Hz
0.25 + 0.5
45 Hz to 1 kHz
0.25 + 0.5
Six digits
1 kHz to 10 kHz
10 + 2
0.93 A to
10 Hz to 45 Hz
0.5 + 1.0
8.49999 A
45 Hz to 1kHz
0.5 + 0.5
1 kHz to 10 kHz
10 + 2
Six digits
8.5 A to 57 A [2]
45 Hz to 500 Hz
0.5 + 0.5
500 Hz to 1 kHz
1.0 + 1.0
[1] Frequency limited to 1 kHz with LCOMP on.
[2] Frequency limited to 440 Hz with LCOMP on
AC Current (Non-Sine Wave) Specifications (cont)
Square Wave
Range
p-p
Frequency
1-Year Absolute
Uncertainty,
tcal ± 5 °C,
± (% of output + % of range)
Max
Current
Resolution
0.01 Hz to 10 Hz
5.0 + 0.5
Two digits
0.047 mA to
10 Hz to 45 Hz
0.25 + 0.5
0.65999 mA [1]
45 Hz to 1 kHz
0.25 + 0.25
Six digits
1 kHz to 10 kHz
10 + 2
0.01 Hz to 10 Hz
5.0 + 0.5
Two digits
0.66 mA to
10 Hz to 45 Hz
0.25 +0.5
6.59999 mA [1]
45 Hz to 1 kHz
0.25 + 0.25
Six digits
1 kHz to 10 kHz
10 + 2
0.01 Hz to 10 Hz
5.0 + 0.5
Two digits
6.6 mA to
10 Hz to 45 Hz
0.25 + 0.5
65.9999 mA [1]
45 Hz to 1 kHz
0.25 + 0.25
Six digits
1 kHz to 10 kHz
10 + 2
0.01 Hz to 10 Hz
5.0 + 0.5
Two digits
66 mA to
10 Hz to 45 Hz
0.25 + 0.5
659.999 mA [1]
45 Hz to 1 kHz
0.25 + 0.5
1 kHz to 10 kHz
10 + 2
0.66 A to
10 Hz to 45 Hz
0.5 + 1.0
Six digits
5.99999 A [2]
45 Hz to 1 kHz
0.5 + 0.5
1 kHz to 10 kHz
10 + 2
6 A to 41 A [2]
45 Hz to 500 Hz
0.5 + 0.5
500 Hz to 1 kHz
1.0 + 1.0
[1] Frequency limited to 1 kHz with LCOMP on.
[2] Frequency limited to 440 Hz with LCOMP on.
AC Current, Square Wave Characteristics (typical)
Range
LCOM
P
Risetim
e
Settling Time
Overshoot
I < 6 A @ 400 Hz
off
25 µs
40 µs to 1% of final value
< 10% for < 1 V
Compliance
3 A & 20 A Ranges
on
100 µs
200 µs to 1% of final value
< 10% for < 1 V
Compliance
AC Current, Triangle Wave Characteristics (typical)
Linearity to 400 Hz
Aberrations
0.3% of p-p value, from 10% to 90% point