Endress+Hauser Proline Fieldcheck Operating Instructions Manual

Specifications

Brand: Endress+Hauser
Category: Measuring Instruments
Model: Proline Fieldcheck (Operating Instructions Manual)
File Type: PDF
File Size: —
Pages: 70

Read Online

Read the Endress+Hauser Proline Fieldcheck Operating Instructions Manual online using the built-in PDF viewer.

Text of the manual

Valid as of software version

V 1.02.XX Operating Instructions Proline Fieldcheck Flow simulator E + -

Table of contents

Endress+Hauser 3 Table of contents 1 Safety instructions . . . . . . . . . . . . . . . . 4 1.1 Designated use . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.2 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . 4 1.3 Notes on operating Fieldcheck in an environment with strong electromagnetic interference . . . . . . . . . . . . . 4 1.3.1 Radiated RF noise (EN 61326-1) . . . . . . . . . 4 1.4 Notes on safety conventions and icons . . . . . . . . . . . 5 2 Identification . . . . . . . . . . . . . . . . . . . . 6 2.1 Device designation . . . . . . . . . . . . . . . . . . . . . . . . . 6 2.1.1 Fieldcheck nameplate . . . . . . . . . . . . . . . . . 6 2.1.2 Simubox nameplate . . . . . . . . . . . . . . . . . . . 6 2.2 Scope of delivery . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 2.3 CE mark, Declaration of conformity . . . . . . . . . . . . . 7 2.4 Registered trademarks . . . . . . . . . . . . . . . . . . . . . . . 7 3 Wiring . . . . . . . . . . . . . . . . . . . . . . . . . 8 3.1 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 3.2 Connecting to Promag 10 . . . . . . . . . . . . . . . . . . . . 9 3.3 Connecting to Promag 50/53 compact version . . . 10 3.4 Connecting to Promag 50/53 remote version . . . . 11 3.5 Connecting to Promass 40/80/83 compact version 12 3.6 Connecting to Promass 80/83 remote version . . . . 13 3.7 Connecting to Prowirl 72 . . . . . . . . . . . . . . . . . . . 14 3.7.1 Prowirl 72 wiring diagram . . . . . . . . . . . . . 15 3.8 Assignment of Fieldcheck connections . . . . . . . . . . 16 3.9 Post-connection check . . . . . . . . . . . . . . . . . . . . . . 16 4 Operation . . . . . . . . . . . . . . . . . . . . . . 17 4.1 Display and operating elements . . . . . . . . . . . . . . . 17 4.1.1 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 4.1.2 Display icons . . . . . . . . . . . . . . . . . . . . . . . 19 4.2 Brief operating instructions on the function matrix . 20 4.2.1 General notes . . . . . . . . . . . . . . . . . . . . . . 20 4.3 Communication . . . . . . . . . . . . . . . . . . . . . . . . . . 21 4.3.1 FieldTool™ operating program . . . . . . . . . 21 5 Commissioning. . . . . . . . . . . . . . . . . . 22 5.1 Function check . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 5.2 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 5.2.1 Charging the battery . . . . . . . . . . . . . . . . . 22 6 Maintenance. . . . . . . . . . . . . . . . . . . . 23 7 Accessories. . . . . . . . . . . . . . . . . . . . . 24 7.1 Software history . . . . . . . . . . . . . . . . . . . . . . . . . . 24 8 Technical data . . . . . . . . . . . . . . . . . . 25 8.1 Technical data at a glance . . . . . . . . . . . . . . . . . . . 25 8.1.1 Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 8.1.2 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 8.1.3 Power supply . . . . . . . . . . . . . . . . . . . . . . . 25 8.1.4 Performance characteristics . . . . . . . . . . . . 25 8.1.5 Operating conditions (Environment) . . . . . . 25 8.1.6 Mechanical construction . . . . . . . . . . . . . . 26 8.1.7 Display and user interface . . . . . . . . . . . . . 26 8.1.9 Accessories . . . . . . . . . . . . . . . . . . . . . . . . 26 8.1.10 Documentation . . . . . . . . . . . . . . . . . . . . . 26 9 Description of device functions - General section . . . . . . . . . . . . . . . . . . 27 9.1 Notes on the description of device functions . . . . . . 27 9.1.1 Using the table of contents to locate a function description . . . . . . . . . . . . . . . . . . . . . . . . . 27 9.1.2 Using the graphic of the function matrix to locate a function description . . . . . . . . . . . 27 9.2 Function matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 9.3 Block USER INTERFACE . . . . . . . . . . . . . . . . . . . . 29 9.3.1 Group CONTROL . . . . . . . . . . . . . . . . . . . 29 9.4 Block BASIC FUNCTIONS . . . . . . . . . . . . . . . . . . . 31 9.4.1 Group SENSOR PARAMETER . . . . . . . . . . 32 9.4.2 Group OUTPUTS . . . . . . . . . . . . . . . . . . . . 33 9.4.3 Group PROCESS PARAMETER . . . . . . . . . 34 9.4.4 Group READ IN PARAMETERS . . . . . . . . . 35 9.5 Block SUPERVISION . . . . . . . . . . . . . . . . . . . . . . . 36 9.5.1 Group RESULTS . . . . . . . . . . . . . . . . . . . . 36 9.5.2 Group VERSION INFO . . . . . . . . . . . . . . . 38 10 Description of device functions - Fieldcheck and electromagnetic measuring devices . . . . . . . . . . . . . . . . 39 10.1 Block FUNCTION . . . . . . . . . . . . . . . . . . . . . . . . . 39 10.1.1 Group SIMULATION . . . . . . . . . . . . . . . . . 40 10.1.2 Group VERIFICATION . . . . . . . . . . . . . . . 44 11 Description of device functions - Fieldcheck and Coriolis measuring systems . . . . . . . . . . . . . . . . . . . . . . . . 49 11.1 Block FUNCTION . . . . . . . . . . . . . . . . . . . . . . . . . 49 11.1.1 Group SIMULATION . . . . . . . . . . . . . . . . . 50 11.1.2 Group VERIFICATION . . . . . . . . . . . . . . . 53 12 Description of device functions - Fieldcheck and vortex measuring systems . . . . . . . . . . . . . . . . . . . . . . . . 59 12.1 Block FUNCTION . . . . . . . . . . . . . . . . . . . . . . . . . 59 12.1.1 Group SIMULATION . . . . . . . . . . . . . . . . . 60 12.1.2 Group VERIFICATION . . . . . . . . . . . . . . . 63 Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 68

The Fieldcheck flow simulation system may only be used for the simulation of sensor signals with a

flowmeter of the Proline equipment generation from Endress+Hauser. The device is normally installed by service technicians who are familiar with the function and operating method of the equipment. The manufacturer accepts no liability for damages resulting from incorrect use. # Warning! Fieldcheck may not be used in hazardous areas without corresponding safety measures. 1.2 Operational safety • The manufacturer reserves the right to modify technical data without prior notice. Please contact your E+H distributor for information on modifications or updates of these Operating Instructions. • The measuring system complies with the general safety requirements in accordance with EN 61010, the EMC requirements of EN 61326/A1 and NAMUR Recommendation NE 21. 1.3 Notes on operating Fieldcheck in an environment with strong electromagnetic interference 1.3.1 Radiated RF noise (EN 61326-1) To operate Fieldcheck with a Proline flowmeter in the simulation and verification mode, the electronics and connection compartments must first be opened. In this mode, the radiated RF noise will not exceed the industrial limits (EN 61326-1). In a heavily disturbed (EMC) environment, you must expect that the operation of both Fieldcheck and the flowmeter may be temporarily affected. We recommend that - when using Fieldcheck - you take care to ensure effective ESD protection for the affected measuring point and the operator. If necessary, the measuring equipment can be reset: • Via the local controls • By interrupting the voltage supply • By restarting Fieldcheck.

Notes on safety conventions and icons

The devices are designed to meet state-of-the-art safety requirements, have been tested, and left the factory in a condition in which they are safe to operate. The devices comply with the applicable standards and regulations in accordance with EN 61010 “Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory Procedures”. They can, however, be a source of danger if used incorrectly or for other than the designated use. Consequently, always pay particular attention to the safety instructions indicated in these Operating Instructions by the following icons: # Warning! “Warning” indicates an action or procedure which, if not performed correctly, can result in injury or a safety hazard. Comply strictly with the instructions and proceed with care. " Caution! “Caution” indicates an action or procedure which, if not performed correctly, can result in incorrect operation or destruction of the device. Comply strictly with the instructions. ! Note! “Note” indicates an action or procedure which, if not performed correctly, can have an indirect effect on operation or trigger an unexpected response on the part of the device.

The Fieldcheck flow simulation system consists of the following components:

• The Fieldcheck flow simulator itself • The Simubox for correct connection to the respective measuring device 2.1.1 Fieldcheck nameplate F06-DXC10xxx-18-00-xx-xx-000 Fig. 1: Nameplate specifications for the Fieldcheck flow simulator (example) 1 Order code/serial number 2 Permitted ambient temperature range 2.1.2 Simubox nameplate F06-DXC10xxx-18-00-xx-xx-001 Fig. 2: Nameplate specifications for the Simubox (example) 1 Order code 2 Serial number FIELD CHECK ENDRESS+HAUSER Order Code: Ser.No.: 50098801 12345678901 0°C(+32°F)

Scope of delivery

The equipment supplied with Fieldcheck comprises the following components: • The Fieldcheck flow simulator with battery pack • A connecting cable for the service interface • A connecting cable for the Simubox • Two connecting cables for current and frequency output • An "RS 232" connecting cable for connecting the Fieldcheck to a computer (9-pin COM interface) • A line adapter • A softcase • Operating Instructions The Simuboxes for the respective flow measuring systems are available via separate order numbers. 2.3 CE mark, Declaration of conformity The devices are designed to meet state-of-the-art safety requirements, have been tested, and left the factory in a condition in which they are safe to operate. The devices comply with the applicable standards and regulations in accordance with EN 61010 “Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory Procedures”. The measuring system described in these Operating Instructions thus complies with the statutory requirements of the EC Directives. Endress+Hauser confirms successful testing of the device by affixing the CE mark. 2.4 Registered trademarks FieldTool™, Fieldcheck™ Registration-pending or registered trademarks of Endress+Hauser Flowtec AG, Reinach, Switzerland

When connecting Ex-certified devices, see the notes and diagrams in the Ex-specific supplement to

these Operating Instructions of the relevant measuring device. Fieldcheck may not be used in hazardous areas. Please do not hesitate to contact your E+H representative if you have any questions. 3.1 Overview Fieldcheck is connected to the following transmitter connections: • Via the Simubox to the transmitter amplifier • To the transmitter service plug • To the current or frequency output. For correct synchronisation with the measuring device we recommend that you abide by the following procedure. 1. Start the device to be tested. 2. First connect the cables to the flowmeter (→Page 9 to Page 14). 3. Next, connect the cables to the front of Fieldcheck (→Page 16). ! Note! Fieldcheck does not need an external power supply during operation. Power is provided by internal storage batteries. Make sure that the storage batteries are fully charged before beginning a simulation.

Connecting to Promag 10

" Caution! Connect the ground terminals on the cables to the uninsulated housing components or to the ground terminals in the cover of the connection compartment. 1. Remove cover of the electronics compartment (c). 2. Remove the local display (d) from the connection compartment cover. 3. Press the side latches (e) and fold down the connection compartment cover. 4. Pull the sensor signal cable plug out of the amplifier socket (f). 5. Insert Simubox cable plug into the amplifier socket (f). 6. Connect Fieldcheck to current and pulse output (g) of the transmitter. ! Note! The configuration of the outputs is dependent on the product type and the product version of the tested transmitter. To find out which terminal pairs are assigned to the outputs, see the “Wiring” chapter of the flowmeter Operating Instructions. 7. Connect FieldCheck to the service plug (h). F06-DXC10xxx-04-00-xx-xx-007 Fig. 3: Connecting Fieldcheck to Promag 5X (compact version) a Fieldcheck flow simulator b Simubox c Electronics compartment cover d Local display e Latches f Amplifier socket g Terminals for current/pulse output h Service plug-in jack 1 2 f h g Esc a b d c e

Connecting to Promag 50/53 compact version

" Caution! Connect the ground terminals on the cables to the uninsulated housing components or to the ground terminals in the cover of the connection compartment. 1. Remove cover of the electronics compartment (e). 2. Pull local display (f) down. 3. Pull the sensor signal cable plug out of the amplifier socket (g). 4. Insert Simubox cable plug into the amplifier socket (g). 5. Remove the cover of the connection compartment (h) from the transmitter housing. 6. Connect Fieldcheck to the current and frequency outputs of the transmitter (terminal pairs 22/23, 24/25 or 26/27). ! Note! The configuration of the outputs is dependent on the product type and the product version of the tested transmitter. To find out which terminal pairs are assigned to the outputs, see the “Wiring” chapter of the flowmeter Operating Instructions. 7. Connect Fieldcheck to the service plug (i). F06-DXC10xxx-04-00-xx-xx-000 Fig. 4: Connecting Fieldcheck to Promag 5X (compact version) a Fieldcheck flow simulator b Simubox c Stainless steel transmitter d Aluminium transmitter e Electronics compartment cover f Local display g Amplifier socket h Connection compartment cover I Service plug 27 25 23 21 2 1 26 24 22 20 L1 (L+) N (L-) E - + h c e d h e g f b a i

Connecting to Promag 50/53 remote version

" Caution! Connect the ground terminals on the cables to the uninsulated housing components or to the ground terminals in the cover of the connection compartment. 1. Remove the cover of the connection compartment (d). 2. Completely remove the terminal block (e, terminal nos. 4-8, 36+37) from the base. " Caution! If the terminal block is not removed, a correct simulation/verification is not guaranteed. 3. Insert Simubox cable plug into the amplifier socket (g). 4. Connect Fieldcheck to the current and frequency outputs of the transmitter (terminal pairs 22/23, 24/25 or 26/27). ! Note! The configuration of the outputs is dependent on the product type and the product version of the tested transmitter. To find out which terminal pairs are assigned to the outputs, see the “Wiring” chapter of the flowmeter Operating Instructions. 5. Connect Fieldcheck to the service plug (f). F06-DXC10xxx-04-00-xx-xx-001 Fig. 5: Connecting Fieldcheck to Promag 5X (remote version) a Fieldcheck flow simulator b Simubox c Transmitter wall-mount housing d Connection compartment cover e Terminal block f Service plug g Amplifier socket 1 2 22 23 20 21 24 25 26 27 41 42 6 5 7 8 4 37 36 c d e g f b a

Connecting to Promass 40/80/83 compact version

" Caution! Connect the ground terminals on the cables to the uninsulated housing components or to the ground terminals in the cover of the connection compartment. 1. Remove cover of the electronics compartment (e). 2. Remove local display. 3. Disconnect the excitation current cable from the excitation current socket (f). # Warning! The excitation current cable must be disconnected as otherwise the measuring tube could vibrate in the resonance frequency and become damaged! 4. Pull the sensor signal cable plug out of the amplifier socket (g). 5. Insert Simubox cable plug into the amplifier socket (g). 6. Remove the cover of the connection compartment (h) from the transmitter housing. 7. Connect Fieldcheck to the current and frequency outputs of the transmitter (terminal pairs 22/23, 24/25 or 26/27). ! Note! The configuration of the outputs is dependent on the product type and the product version of the tested transmitter. To find out which terminal pairs are assigned to the outputs, see the “Wiring” chapter of the flowmeter Operating Instructions. 8. Connect Fieldcheck to the service plug (i). F06-DXC10xxx-04-00-xx-xx-004 Fig. 6: Connecting Fieldcheck to Promass 8x (compact version) a Fieldcheck flow simulator b Simubox c Stainless steel transmitter d Aluminium transmitter e Electronics compartment cover f Exciting current socket g Amplifier socket h Connection compartment cover i Service plug 27 25 23 21 2 1 26 24 22 20 L1 (L+) N (L-) h c e d h e g f b a i

Connecting to Promass 80/83 remote version

" Caution! Connect the ground terminals on the cables to the uninsulated housing components or to the ground terminals in the cover of the connection compartment. 1. Remove the cover of the connection compartment (d). 2. Remove excitation current cable (h, terminal pair 41/42). # Warning! The excitation current cable must be removed as otherwise the measuring tube could vibrate in the resonance frequency and become damaged! 3. Completely remove the terminal block (e, terminals 4-12) from the base. " Caution! If the terminal block is not removed, a correct simulation/verification is not guaranteed. 4. Insert Simubox cable plug into the amplifier socket (g). 5. Connect Fieldcheck to the current and frequency outputs of the transmitter (terminal pairs 22/23, 24/25 or 26/27). ! Note! The configuration of the outputs is dependent on the product type and the product version of the tested transmitter. To find out which terminal pairs are assigned to the outputs, see the “Wiring” chapter of the flowmeter Operating Instructions. 6. Connect Fieldcheck to the service plug (f). F06-DXC10xxx-04-00-xx-xx-005 Fig. 7: Connecting Fieldcheck to Promass 8X (remote version) a Fieldcheck flow simulator b Simubox c Transmitter wall-mount housing d Connection compartment cover e Terminals for sensor signal cable f Service plug g Amplifier socket h Terminals for excitation current cable 1 2 22 23 20 21 24 25 26 27 41 42 6 5 7 8 4 9 10 11 12 c d e g f b a h

Connecting to Prowirl 72

" Caution! Connect the ground terminals on the cables to the uninsulated housing components or to the ground terminals in the cover of the connection compartment. 1. Remove cover of the electronics compartment (c). 2. Remove the local display module (d) from the retainer rails (e). 3. Fit the local display module (d) with its left side onto the right retainer rail (e) (this secures the local display module). (Continued on next page) F06-DXC10xxx-04-00-xx-xx-006 Fig. 8: Connecting Fieldcheck to Prowirl 72 a Fieldcheck b Simubox c Electronics compartment cover d Local display module e Local display module retainer rails f Fixing screw for cover of connection compartment g Connection compartment cover h Terminals for current output/power supply and pulse output (optional) i Plastic cover j Amplifier socket k Service plug b a e g f d c g i i j k h

Loosen the fixing screw (f) of the cover of the connection compartment (g) and fold down the

cover. 5. Connect Fieldcheck to the current output/power supply and to the pulse output if available. The terminals can be disconnected for this. " Caution! – The current output can only be tested for accuracy during the verification if it is connected to the current input of Fieldcheck. – The pulse output is connected to the frequency input of Fieldcheck. 6. Fold up the plastic cover (i). 7. Disconnect the signal cable plug from the amplifier socket (j). 8. Insert Simubox cable plug into the amplifier socket (j). 9. Connect Fieldcheck to the service plug (k). 3.7.1 Prowirl 72 wiring diagram F06-72xxxxxx-04-00-00-xx-000 Fig. 9: Assignment of terminals A = Power supply/current output B = Pulse/status output C = Ground terminal (only relevant for remote version) 1 2 3 4 A C + + - - B

Assignment of Fieldcheck connections

F06-DXC10xxx-04-00-xx-xx-003 Fig. 10: Fieldcheck connection diagram on the front panel 1 Connection to the frequency output of the device 2 Connection to the current output of the device 3 Connection to the internal service interface of the device 4 Connection to the Simubox 5 Connection to the plug power unit 3.9 Post-connection check • Fieldcheck can only complete the start-up procedure correctly if the simulation box and the service cable are correctly connected. • Current and frequency outputs which are incorrectly connected or mixed up result in an incorrect verification result. 1 2 3 4 5 DC SIM SRV I F

Display and operating elements

The local display enables you to read all important variables of the simulation and configure the device using the function matrix. F06-DXC10xxx-07-xx-xx-xx-000 Fig. 11: Display and operating elements 1 Liquid crystal display The backlit, liquid crystal display shows measured values, dialog texts, fault messages and information messages. The HOME position (operating mode) is the name given to the display during normal measuring operation. 2 Optical operating elements for "Touch Control" 3 Plus/minus keys – Enter numerical values, select parameters – Select different blocks, groups and function groups within the function matrix The following functions are activated by pressing the +/- keys (X) simultaneously: – Exit the function matrix step by step →HOME position – Press the +/- keys longer than 3 seconds →Direct return to the HOME position – Cancel data entry 4 Enter key – HOME position →Enter the function matrix – Save numerical values entered or altered settings Esc E + - 1 2 3 4 +7.000 +7.012 dm3/m dm3/m MP1 5 % V +7.000 +7.012 dm3/m dm3/m MP1 5 % V

The display area consists of three to four lines; this is where measured values are displayed, and/or

status variables (sequence information of a verification. F06-DXC10xxx-07-xx-xx-xx-001 Fig. 12: Typical display for normal operating mode (HOME position) 1 Main line: shows simulation values, e.g. volume flow in [dm3/s] 2 Additional line: shows up-to-date measured values of the tested measuring device] 3 Information line: shows the current variables for current and frequency if the corresponding outputs of the device are connected to Fieldcheck. This is only possible in the SIMULATION operating mode 4 “Information icons” field: icons representing additional information on the measured values are shown in this field. See figure for a full list of the icons and their meanings. 5 “Measured values” field: the current measured values appear in this field. 6 “Unit of measurement” field: the units of measure and time defined for the flow values appear in this field. 1 4 5 6 2 3 +7.000 +7.012 dm3/m dm3/m I = 17,12 mA F = 783 Hz V

The icons which appear in the field on the left make it easier to read and recognise error messages

locally. F06-DXC10xxx-07-xx-xx-xx-002 Fig. 13: Fieldcheck display icons 1 DC power supply 2 Charging status: battery fully charged 3 Charging status: battery 2/3 charged 4 Charging status: battery 1/3 charged (flashing in HOME position) 5 Simulation “continuous" 6 Simulation “step" 7 Simulation “profile" 8 Verification 9 Measured volume flow 10 Measured mass flow 1 2 3 4 5 9 6 7 8 V 10 m

Brief operating instructions on the function matrix

! Note! • See the general notes on Page 20. • Description of functions → Section 9 to Section 12 1. HOME position → F → Enter the function matrix 2. Select a block (e.g. USER INTERFACE) 3. Select a group (e.g. CONTROL) 4. Select a function group (e.g. BASIC CONFIGURATION) 5. Select a function (e.g. LANGUAGE) Change parameter / enter numerical values: P → Select or enter release code, parameters, numerical values F → Save your entries 6. Exit the function matrix: – Press Esc key (X) longer than 3 seconds → HOME position – Repeatedly press Esc key (X) → Return step by step to HOME position F06-x3xxxxxx-19-xx-xx-xx-000 Fig. 14: Selecting functions and configuring parameters (function matrix) 4.2.1 General notes The function matrix comprises a multiplicity of functions which, for the sake of clarity, are arranged on a number of menu levels (blocks, groups and function groups).

Comply with the following instructions when configuring functions:

You select functions as described on Page 20. ! Note! A detailed description of all the functions as well as a detailed overview of the function matrix are provided in Sections 9 to 12. 4.3 Communication Operation via a personal computer using the “FieldTool” operating software. 4.3.1 FieldTool™ operating program FieldTool™ is a universal service and configuration software package designed for the Proline measuring devices. The connection is made with a commercial serial cable (9-pole DSUB, 0- Modem) to the underside of Fieldcheck. FieldTool™ offers the user the following fields of application: • Configuring the equipment functions. • Archiving the verification data. • Printout of a verification certificate with all measuring results and I/O parameters. You can find more information on FieldTool™ in the following E+H document: System Information: SI 031D/06/en “FieldTool™"

Make sure that Fieldcheck is correctly connected to the measuring device that is to be simulated

(→Page 8 ff.). The measuring device must be working free of errors in order for a simulation or verification to be started. 5.2 Commissioning Fieldcheck is powered via an adapter or via the storage battery included in the delivery. Before the simulation, make sure that the storage batteries are fully charged. To ensure correct commissioning you are advised to abide by the processes as they are described on Page 8 ff. Fieldcheck switches itself on automatically as soon as one of the connecting cables is inserted. Fieldcheck performs a number of self-tests after power-on. As this procedure progresses, diverse messages which have to be acknowledged appear on the local display. You know that Fieldcheck is ready for operation when the HOME position appears on the display. If batteries are being used, the charge status of the battery pack also appears (Page 19 ff.). 5.2.1 Charging the battery The battery is charged automatically as soon as the power unit or the car-adapter cable is connected to Fieldcheck. The charging time is less than 3 hours if there is no simulation in progress. When the battery is fully charged it is switched to conservation charging. The charging time is considerably longer during a simulation. The charge status is displayed in the HOME position via the charge status icon (Page 19 ff.).

Fieldcheck requires no special maintenance.

" Caution! To prevent a low discharge of the battery in case of not using Fieldcheck for a longer period it is recommended to charge the device every quarter of a year. Exterior cleaning When cleaning the exterior of measuring devices, always use cleaning agents that do not attack the surface of the housing.

Various accessories can be supplied for Fieldcheck and they can be ordered separately from

Endress+Hauser. The E+H service organisation can provide detailed information on the order code of your choice. 7.1 Software history Accessory Description Order code FieldTool™ Configuration and service software for flowmeters in the field: – Commissioning, maintenance analysis – Configuring flowmeters – Service functions – Visualisation of process data – Troubleshooting – Controlling the “Fieldcheck™” tester/simulator Contact your E+H representative for more information. DXS10 - * * * * * Simubox MID for FieldCheck Serves communication between FieldCheck and electromagnetic flowmeters. 50097102 Simubox Coriolis for FieldCheck Serves communication between FieldCheck and Coriolis mass flowmeters. 50097101 Plug-in power unit 85260 VAC Power unit for charging the battery and the operation of FieldCheck at the mains 50095659 Car adapter 12 VDC Adapter for connecting to the car battery. 50095660 NiMH AccuPack Additional battery 50095661 Softcase Practical storage bag 50095662 Simubox/FieldCheck connecting cable For connecting the Simubox to FieldCheck 50095663 Proline measuring device/FieldCheck interface cable For connecting FieldCheck to the internal service interface of the flowmeter 50097100 FieldCheckcurrent or frequency output measuring cable For connecting FieldCheck to the current or frequency output of the flowmeter 50095664 FC/Tooling interface cable For connecting FieldCheck to the Tooling computer 50098800 Software version / date Changes to software Changes to documentation V 1.00.00 - - V 1.00.01/11.2001 Device release - V 1.00.02/08.2002 Software adaptation - V 1.01.00/07.2003 Software extension for Coriolis flow measuring systems Operating Instructions V 1.02.00/04.2004 Software extension for Prowirl 72 Operating Instructions

Technical data at a glance

8.1.1 Input Measured variable Current input, frequency input Measuring range 0...25 mA, 0...15000 Hz 8.1.2 Output Output signal Electromagnetic: simulation voltage Coriolis: 2 phase-shifted sinusoidal signals Galvanic isolation Yes 8.1.3 Power supply Electrical connections Input for power unit: 100-240 V AC (47-63 Hz) Prim., 12 V DC (2.5 A) Sec. Supply voltage 12 V DC, internal battery 7.2 V, 3800 mAh Power consumption 140 mA Charging current: max. 2A (max. 350 mA for simultaneous operation) 8.1.4 Performance characteristics Reference operating conditions +22 °C ± 2K Max. measured error for outputs Electromagnetic measuring device Voltage ±0.2 % of full scale (0.150 ... 20 mV) Coriolis measuring device: Phase difference ±0.2 % of full scale Max. measured error for inputs Current ± 5 µA (0 ... 25 mA) Frequency < 0.5 Hz (0 ...15 kHz) 8.1.5 Operating conditions (Environment) Ambient temperature range 0 ... 60°C Electromagnetic compatibility (EMC) In accordance with EN 61326/A1 and NAMUR Recommendation NE 21

270 x 130 x 60 mm (length x width x height)

Simubox (without cable) 120 x 60 x 30 mm (L x W x H) Softcase 400 x 180 x 300 mm (L x W x H) Weight 1.6 kg 8.1.7 Display and user interface Display elements Liquid crystal display: illuminated, for presenting different measured value and status variables Operating elements Local operation with three optical keys (O, S, F) Computer interface FieldTool™ • Permanent storage of verification data • Printing of certificates CE mark The measuring system is in conformity with the statutory requirements of the EC Directives. Endress+Hauser confirms successful testing of the device by affixing the CE mark. Other standards and guidelines EN 60529: Degrees of protection by housing (IP code) EN 61010: Protection Measures for Electrical Equipment for Measurement, Control, Regulation and Laboratory Procedures EN 61326/A1 (IEC 1326): Electromagnetic compatibility (EMC requirements) 8.1.8 Ordering information The E+H service organisation can provide detailed ordering information and information on the order codes on request. 8.1.9 Accessories Various accessories can be supplied for Fieldcheck and they can be ordered separately from Endress+Hauser (see →Page 24). The E+H service organisation can provide detailed information on the order codes of your choice. 8.1.10 Documentation • System Information SI 030D/06/en • Technical Information TI 059D/06/en

Description of device functions - General section

Endress+Hauser 27 9 Description of device functions - General section 9.1 Notes on the description of device functions The description of device functions is split into the following sections: • Description of device functions - General section (→Page 27 ff.) This introduces you to the function matrix and provides you with an overview of the device functions and a description of the device functions which apply to all applications. • Description of device functions - Fieldcheck and electromagnetic measuring devices (→Page 39 ff.) This provides you with a description of the device functions which only apply to the simulation and verification of electromagnetic flow measurement systems. • Description of device functions - Fieldcheck and Coriolis measuring systems (→Page 49 ff.) This provides you with a description of the device functions which only apply to the simulation and verification of Coriolis mass flow measuring systems. • Description of device functions - Fieldcheck and vortex measuring systems (→Page 59 ff.) This provides you with a description of the device functions which only apply to the simulation and verification of vortex measuring systems. There are various ways of locating the description of a function of your choice in the manual: 9.1.1 Using the table of contents to locate a function description The designations of all the cells in the function matrix are listed in the table of contents. You can use these unambiguous designations (such as USER INTERFACE, INPUTS, OUTPUTS, etc.) to choose whichever functions are applicable to a particular set of conditions. The page references show you exactly where to find the detailed descriptions of the functions in question. The table of contents can be found on Page 3. 9.1.2 Using the graphic of the function matrix to locate a function description This step-by-step, top-down approach starts with the blocks, the highest level, and works down through the matrix to the description of the function you need: 1. All blocks available, and their related groups are illustrated on Page 28. Select the block (or the group within the block) which you need for your application and use the page reference to locate the information corresponding to the next level. 2. The page in question contains a graphic illustration of the block with all its subordinate groups, function groups and functions.Select the function which you need for your application and use the page reference to locate the detailed function description.

Description of device functions - General section

Proline Fieldcheck 28 Endress+Hauser 9.2 Function matrix BLOCKS GROUPS FUNCTIONS FUNCTION Proline Promag: Page 39 ff. Proline Promass: Page 49 ff. Proline Vortex: Page 59 ff. → SIMULATION Electromagnetic measuring devices (Proline Promag) → Page 40 ff. Coriolis measuring devices (Proline Promass) → Page 50 ff. Vortex measuring devices (Proline Prowirl) → Page 60 ff. ↓ VERIFICATION Electromagnetic measuring devices (Proline Promag) → Page 44 ff. Coriolis measuring devices (Proline Promass) → Page 53 ff. Vortex measuring devices (Proline Prowirl) → Page 63 ff. USER INTERFACE → CONTROL → Page 29 ff. ↓ BASIC FUNCTIONS → Sensor data → Page 32 ff. OUTPUTS → Page 33 ff. ↓ PROCESS PARAMETER → Page 34 READ IN PARAMETERS → Page 35 SUPERVISION → RESULTS → Page 36 ff. VERSION INFO → Page 38 ff.

Description of device functions - General section

Endress+Hauser 29 9.3 Block USER INTERFACE 9.3.1 Group CONTROL Block Group Function group Function USER INTERFACE ⇒ CONTROL P. 29 ⇒ LANGUAGE P. 29 CONTRAST LCD P. 29 BACK LIGHT P. 29 FORMAT DATE P. 29 DATE P. 29 FORMAT TIME P. 29 TIME P. 30 Function description USER INTERFACE → CONTROL LANGUAGE Use this function to select the language for all texts, parameters and messages shown on the local display. Options: ENGLISH DEUTSCH CONTRAST LCD Use this function to optimise display contrast to suit local operating conditions. User input: 10...100% Factory setting: 50% BACK LIGHT Options: ON(AUTOMATIC) OFF Factory setting: ON(AUTOMATIC) ! Note! If the ON(AUTOMATIC) setting is selected, the lighting of the display is automatically switched off if the equipment is not used for 10 minutes. FORMAT DATE Use this function to select a date format. Options: DD.MM.YYYY MM/DD/YYYY Factory setting: DD.MM.YYYY DATE Use this function to enter the date. User input: DD.MM.YYYY (MM/DD/YYYY) FORMAT TIME Options: 24H 12H (AM/PM) Factory setting: 24H

Description of device functions - General section

Proline Fieldcheck 30 Endress+Hauser TIME Use this function to enter the time. User input: HH:MM Function description USER INTERFACE → CONTROL

Description of device functions - General section

Endress+Hauser 31 9.4 Block BASIC FUNCTIONS Block Group Function group Function BASIC FUNCTIONS ⇒ SENSOR PARAMETER P. 32 ⇒ K-FACTOR POS/NEG P. 32 ZERO POINT P. 32 NOMINAL DIAMETER P. 32 FLOW DIRECTION P. 32 SERIAL NUMBER P. 32 ⇓⇑ TAG NAME. P. 32 OUTPUTS P. 33 ⇒ OUTPUT P. 33 ASSIGN P. 33 CURRENT RANGE P. 33 VALUE 0_4 mA P. 33 VALUE 20 mA P. 33 ⇓⇑ OPERATION MODE P. 33 ASSIGN P. 33 START VALUE FREQUENCY P. 33 END FREQUENCY P. 33 VALUE FLOW P. 33 VALUE FHIGH P. 34 PULSE VALUE P. 34 OUTPUT SIGNAL P. 34 PROCESS PARAMETER P. 34 ⇒ DENSITY VALUE P. 34 STANDARD DENSITY P. 34 OPERATING DENSITY P. 34 REFERENCE DENSITY P. 34 APPLICATION P. 34 ⇓⇑ TYPE UNIT MEAS. P. 34 READ IN PARAMETERS P. 35 ⇒ READ IN PARAMETERS P. 35

Description of device functions - General section

Proline Fieldcheck 32 Endress+Hauser 9.4.1 Group SENSOR PARAMETER Use this function group to display the characteristic data of the connected sensor, such as calibration factor, nominal diameter, serial number, etc. Function description BASIC FUNCTIONS → SENSOR PARAMETER K-FACTOR POSITIVE/NEGATIVE Use this function to view the positive/negative K factor ZERO POINT Use this function to view the zero-point correction value for the sensor. ! Note! This function is not displayed when testing vortex measuring devices. NOMINAL DIAMETER Use this function to view the nominal diameter of the sensor. DIRECTION OF FLOW Use this function to view the set mounting direction of the sensor. ! Note! This function is not displayed when testing vortex measuring devices. SERIAL NUMBER Use this function to view the serial number of the connected measuring device. TAG NAME Use this function to view the device tag name of the connected measuring device.

Description of device functions - General section

Endress+Hauser 33 9.4.2 Group OUTPUTS Use this function group to view the configuration and settings of the current and frequency outputs. Function description BASIC FUNCTION → OUTPUTS OUTPUT Use this function to select the output Options: CURR. OUTPUT FREQUENCY OUTPUT ASSIGN Use this function to view the physical measured variable which is assigned to the current output of the measuring device (e.g. volume flow, mass flow, temperature) ! Note! This function is only available if CURRENT OUTPUT was selected in the OUTPUT function. CURRENT RANGE Use this function to view the set current range in the measuring device. ! Note! This function is only available if CURRENT OUTPUT was selected in the OUTPUT function. VALUE 0_4 mA Use this function to view the flow value, which is assigned to the 0_4 mA current in the measuring device. ! Note! This function is only available if CURRENT OUTPUT was selected in the OUTPUT function. VALUE 20 mA Use this function to view the flow value, which is assigned to the 20 mA current in the measuring device. ! Note! This function is only available if CURRENT OUTPUT was selected in the OUTPUT function. OPERATION MODE Use this function to view the set operating mode of the frequency output in the measuring device. ! Note! This function is only available if FREQUENCY OUTPUT was selected in the OUTPUT function. ASSIGN Use this function to view the physical measured variable which is assigned to the frequency output of the measuring device (e.g. volume flow, mass flow, temperature) ! Note! This function is only available if FREQUENCY OUTPUT was selected in the OUTPUT function. START FREQUENCY Use this function to view the set start frequency in the measuring device. ! Note! This function is only available if FREQUENCY OUTPUT was selected in the OUTPUT function. END FREQUENCY Use this function to view the set end frequency in the measuring device. ! Note! This function is only available if FREQUENCY OUTPUT was selected in the OUTPUT function. VALUE f-LOW Use this function to view the flow value, which is assigned to the start frequency in the measuring device. ! Note! This function is only available if FREQUENCY OUTPUT was selected in the OUTPUT function.

Description of device functions - General section

Proline Fieldcheck 34 Endress+Hauser 9.4.3 Group PROCESS PARAMETER VALUE f-HIGH Use this function to view the flow value, which is assigned to the end frequency in the measuring device. ! Note! This function is only available if FREQUENCY OUTPUT was selected in the OUTPUT (6900) function. PULSE VALUE Use this function to view the set pulse value in the measuring device. ! Note! This function is only available if FREQUENCY OUTPUT was selected in the OUTPUT (6900) function and the operating mode in the measuring device is set to PULSE. OUTPUT SIGNAL Use this function to view the output signal settings (ACTIVE/PASSIVE, POSITIVE/NEGATIVE). Function description BASIC FUNCTION → PROCESS PARAMETER DENSITY VALUE ! Note! This function only appears if a Promag 53-type electromagnetic flow measurement system is being tested. Use this function to view the density factor entered in the measuring device. With this value, the volume flow is converted to a mass flow during simulation. STANDARD DENSITY ! Note! This function only appears if a Coriolis flow measurement system is tested. Use this function to view the standard density entered in the measuring device. With this value, the mass flow is converted to the corrected volume flow during simulation. OPERATING DENSITY ! Note! This function only appears if a vortex flow measurement system is tested. Use this function to view the density at process conditions entered in the measuring device. With this value, the volume flow is converted to a mass flow or corrected volume flow during simulation. REFERENCE DENSITY ! Note! This function only appears if a vortex flow measurement system is tested. Use this function to view the density at reference conditions entered in the measuring device. With this value, the volume flow is converted to a mass flow or corrected volume flow during simulation. APPLICATION ! Note! This function only appears if a vortex flow measurement system is tested. Use this function to view the set application (liquid or gas/steam) in the measuring device. TYPE UNIT MEASURAND ! Note! This function only appears if a vortex flow measurement system is tested. Use this function to view the set unit type (volume flow, calculated mass flow or calculated corrected volume flow) in the measuring device. Function description BASIC FUNCTION → OUTPUTS

Description of device functions - General section

Endress+Hauser 35 9.4.4 Group READ IN PARAMETERS Function description BASIC FUNCTION → READ IN PARAMETERS READ IN PARAMETERS Use this function to transfer the complete data record from the measuring device to Fieldcheck, if, for example, you have changed parameters in the connected measuring device, Options: START CANCEL

Description of device functions - General section

Proline Fieldcheck 36 Endress+Hauser 9.5 Block SUPERVISION 9.5.1 Group RESULTS Block Group Function group Function SUPERVISION ⇒ RESULTS P. 36 ⇒ STORAGE LOC. VER. P. 36 INFO STORAGE LOC. P. 36 RESULTS VER. MID P. 36 RESULTS VER. COR. P. 37 RESULTS VER. VOR. P. 37 ⇓⇑ DELETE STORAGE P. 37 STORAGE LOCATION P. 37 VERSION INFO P. 38 ⇒ FIELDCHECK P. 38 ⇒ HW VERSION P. 38 SW VERSION P. 38 SERIAL NUMBER P. 38 CALIBRATION DATE P. 38 SIMUBOX P. 38 ⇒ TYPE P. 38 HW VERSION P. 38 SW VERSION P. 38 SERIAL NUMBER P. 38 CALIBRATION DATE P. 38 MANUFACT./ TEST DATE P. 38 Function description SUPERVISION → RESULTS STORAGE LOCATION VERIFICATION Use this function to define the storage location of the verification to be displayed. User input: 1...20 INFO STORAGE LOCATION Use this function to view information about the measuring device and the verification saved in the storage location. Display: Main line: measuring device type Additional line: serial number Information line: date of verification RESULTS VERIFICATION MAGMETER Use this function to recall the results of the verification. The result (PASSED, FAILED) is displayed in the information line. If the results of a linearity check are recalled, the list of deviation values also appears when confirming the selection. Options: FINAL RESULT ZERO POINT (with list of deviation values) AMPLIFIER (with list of deviation values) CURRENT OUTPUT (with list of deviation values) FREQ. OUT. (with list of deviation values) PULSE MEASUREMENT (with list of deviation values) SENSOR

Description of device functions - General section

Endress+Hauser 37 RESULTS VERIFICATION CORIOLIS Use this function to recall the results of the verification. The result (PASSED, FAILED) is displayed in the information line. If the results of a linearity check are recalled, the list of deviation values also appears when confirming the selection. Options: FINAL RESULT ZERO POINT (with list of deviation values) AMPLIFIER (with list of deviation values) DENSITY (with list of deviation values) TEMPERATURE (with list of deviation values) CURRENT OUTPUT (with list of deviation values) FREQUENCY OUTPUT (with list of deviation values) PULSE MEASUREMENT (with list of deviation values) SENSOR RESULTSVERIFICATI ON VORTEX Use this function to recall the results of the verification. The result (PASSED, FAILED) is displayed in the information line. If the results of a linearity check are recalled, the list of deviation values also appears when confirming the selection. Options: FINAL RESULT ZERO POINT (with measured value for the current output) AMPLIFIER (with list of deviation values) CURRENT OUTPUT (with list of deviation values) PULSE MEASUREMENT (with list of deviation values) SENSOR DELETE STORAGE Use this function to delete the content of the memory. Options: NO STORAGE LOCATION ALL STORAGE LOCATION Use this function to specify the storage location whose contents should be deleted. User input :1...20. This function does not appear unless the STORAGE LOCATION parameter was selected in the DELETE STORAGE function. Function description SUPERVISION → RESULTS

Description of device functions - General section

Proline Fieldcheck 38 Endress+Hauser 9.5.2 Group VERSION INFO Function group FIELDCHECK Function group SIMUBOX Function description SUPERVISION → VERSION → INFO FIELDCHECK HW VERSION Use this function to view the hardware version number of Fieldcheck. SW VERSION Use this function to view the software version number of Fieldcheck. SERIAL NUMBER Use this function to view the manufacturer serial number of Fieldcheck. CALIBRATION DATE Use this function to view the last calibration date of Fieldcheck. Function description SUPERVISION → VERSION-INFO → SIMUBOX TYPE Use this function to view the measuring principle of the Simubox HW VERSION Use this function to view the hardware version number of the Simubox. SW VERSION Use this function to view the software version number of the Simubox. SERIAL NUMBER Use this function to view the manufacturer serial number of the Simubox. CALIBRATION DATE Use this function to view the last calibration date of the Simubox -(electromagnetic and Coriolis Simuboxes). MANUFACTURING/TE ST DATE Use this function to view the date of manufacture of the Simubox (vortex and ultrasonic Simuboxes). or Use this function to view the last test date (vortex and ultrasonic Simuboxes ).

Description of device functions - Fieldcheck and electromagnetic measuring devices

Endress+Hauser 39 10 Description of device functions - Fieldcheck and electromagnetic measuring devices 10.1 Block FUNCTION Block Group Function group Function FUNCTION ⇒ SIMULATION P. 40 ⇒ CONFIGURATI ON P. 40 ⇒ OPERATION MODE P. 40 SIMUL SIGNAL P. 41 FLOW 1 P. 41 FLOW 2 P. 41 SIMUL. TIME 1 P. 41 ⇓⇑ ⇓⇑ SIMUL. TIME 2 P. 42 PROFILE TIME 1 P. 42 PROFILE TIME 2 P. 42 OPERATION P. 43 ⇒ SIMULATE P. 43 VERIFICATION P. 44 ⇒ OUTPUTS P. 45 ⇒ OUTPUTS P. 45 ⇓⇑ LIMIT VALUES P. 46 ⇒ DEVIATION FLOW P. 46 DEV.CURR.OU TPUT P. 46 DEV.FREQU.O UTPUT P. 46 ⇓⇑ PARAMETER P. 47 ⇒ FLOW 100% P. 47 MEAS. POINT 1/2 P. 47 ⇓⇑ OPERATION P. 48 ⇒VERIFICATION P. 48 VERIFICATE P. 48 STORAGE LOCATION P. 48

Description of device functions - Fieldcheck and electromagnetic measuring devices

Proline Fieldcheck 40 Endress+Hauser 10.1.1 Group SIMULATION Function group CONFIGURATION Function description FUNCTION → SIMULATION → CONFIGURATION OPERATION MODE Use this function to select the operation mode for the simulation. There are two freely adjustable flow signals of -12.5 m/s to + 12.5 m/s and the assigned duration of simulation is 1 second to 999 seconds. Options: CONTINUOUS The flow behaviour in the process has low dynamics. F06-DXC10xxx-05-xx-xx-xx-000 Q1 = Simulation value 1 STEP The flow rates in the process are very dynamic and change very quickly. This operation mode is also suitable for the simulation of flows outside the tolerance limits (Error simulation). Here very dynamic or disrupted conditions can also be simulated. F06-DXC10xxx-05-xx-xx-xx-001 Q1 = Simulation value 1, Q2 = Simulation value 2, T1 = Duration of simulation 1, T2 = Duration of simulation 2 PROFILE For the depiction of real sequences in the course of a process or of a filling procedure. The controlled rise or fall in one flow value to the next can be depicted. F06-DXC10xxx-05-xx-xx-xx-002.eps Q1 = Simulation value 1, Q2 = Simulation value 2, T1 = Duration of simulation 1, T2 = Duration of simulation 2, T3 = Profile time 1, T4 = Profile time 2 Factory setting: CONTINUOUS Q Q1 START STOP t Q Q1 Q2 T2 T1 START STOP t Q Q2 Q1 T2 T3 T4 T1 START STOP t

Description of device functions - Fieldcheck and electromagnetic measuring devices

Endress+Hauser 41 SIMULATION SIGNAL Use this function to select which measurement is to be simulated. Options: VOLUME FLOW MASS FLOW ! Note! • "Calculated mass flow" simulation is only possible with "Promag 53" type devices. • The value stored in the function DENSITY VALUE (DENSITY PARAMETER function group of the device) is used for the calculation. • If a current input is available in the device and the density is assigned to it, mass flow simulation is not possible Factory setting: VOLUME FLOW FLOW 1 Use this function to enter the flow value which is to be simulated. User input: 0.000...999999 (6-digit floating-point number, with sign) ! Note! The unit of the connected measuring device is adopted. The corresponding flow velocity must lie within the value range - 12.5 m/s to + 12.5 m/s. Unit: The unit configured in the measuring device is adopted Factory setting: 0.000 FLOW 2 Use this function to enter the second flow value which is to be simulated. ! Note! • This function is only available if the STEP or PROFILE setting was selected in the OPERATION MODE function. • More information is provided in the OPERATION MODE function. User input: 0.000...999999 (6-digit floating-point number, with sign) ! Note! The unit of the connected measuring device is adopted. The corresponding flow velocity must lie within the value range - 12.5 m/s to + 12.5 m/s. Unit: The unit configured in the measuring device is adopted Factory setting: 0.000 DURATION OF SIMULATION 1 Use this function to enter the duration of simulation signal 1 (flow). ! Note! • This function is only available if the STEP or PROFILE setting was selected in the OPERATION MODE function. • More information is provided in the OPERATION MODE function. User input: 1 ... 999 (max. 3-digit number ) Unit: s ( seconds ) Factory setting: 10 Function description FUNCTION → SIMULATION → CONFIGURATION

Description of device functions - Fieldcheck and electromagnetic measuring devices

Proline Fieldcheck 42 Endress+Hauser DURATION OF SIMULATION 2 Use this function to enter the duration of simulation signal 2 (flow). ! Note! • This function is only available if the STEP or PROFILE setting was selected in the OPERATION MODE function. • More information is provided in the OPERATION MODE function. User input: 1 ... 999 (max. 3-digit number ) Unit: s ( seconds ) Factory setting: 10 PROFILE TIME 1 Use this function to enter the period of time for the change from simulation signal 1 to simulation signal 2. ! Note! • This function is only available if the PROFILE setting was selected in the OPERATION MODE function. • More information is provided in the OPERATION MODE function. User input: 1 ... 99 (max. 2-digit number) Unit: s ( seconds ) Factory setting: 5 PROFILE TIME 2 Use this function to enter the period of time for the change from simulation signal 2 to simulation signal 1. ! Note! • This function is only available if the PROFILE setting was selected in the OPERATION MODE function. • More information is provided in the OPERATION MODE function. User input: 1 ... 99 (max. 2-digit number) Unit: s ( seconds ) Factory setting: 5 Function description FUNCTION → SIMULATION → CONFIGURATION

Description of device functions - Fieldcheck and electromagnetic measuring devices

Endress+Hauser 43 Function group OPERATION Function description FUNCTION → SIMULATION → OPERATION SIMULATE Use this function to control the simulation. Options: START STOP Factory setting: START Display during the simulation: Main line Symbol of the selected operating mode and the simulated value (with unit). Additional line The measured value sent back by the measuring device. Information line In the SIMULATION operating mode, the current values of the current and frequency output of the tested device are displayed (multimeter function). Pulses are displayed as frequency if there is more than 1 pulse/second at the frequency input of Fieldcheck. In the STOP status the simulation signal value “0” is always given. Affected settings for the simulation, however, are maintained.

Description of device functions - Fieldcheck and electromagnetic measuring devices

Proline Fieldcheck 44 Endress+Hauser 10.1.2 Group VERIFICATION In the VERIFICATION group, a measuring device can be checked for the following characteristic values. • Functionality and deviation in the flow measurement (based on the set point preset by Fieldcheck) • Functionality and deviation of the current and frequency outputs if these outputs are available in the measuring device. The specified set points correspond in the gradation to the measuring points when determining flow. Data bus outputs such as PROFIBUS or FOUNDATION FIELDBUS cannot be directly checked by Fieldcheck. • Static and dynamic characteristic values of the magnetic field and electrodes. Verification of the transmitter Checking the linearity of the amplifier: Simulated flow at four measuring points which are determined via the 100% full scale value (can be selected for a presumed fluid velocity between 4 m/s and 8 m/s). Checking a current output: Each current output can be verified in the configuration in the measuring device (CURRENT RANGE function (0-20 mA, 4-20 mA) in the same gradation (5% ... 100%) as the amplifier. For this purpose, the current output must be directly connected to the current input of the Fieldcheck. If more than one current output is available, verification takes place in several, menu-guided steps. Checking a frequency output (FREQUENCY operating mode, (not for "Promag 10" type devices): Each frequency output can be verified in the configuration in the measuring device (START VALUE FREQUENCY, END VALUE FREQUENCY, OUTPUT SIGNAL functions) in the same gradation ( 5% ... 100% ) as the amplifier. For this purpose, the frequency output must be directly connected to the frequency input of the Fieldcheck. If more than one frequency output is available, verification takes place in several, menu-guided steps. A frequency output cannot be checked on "Promag 10" type devices. Checking a frequency output (PULSE operating mode): Each pulse output available can be verified in the configuration in the measuring device (PULSE VALUE, PULSE WIDTH, OUTPUT SIGNAL functions). For this purpose, the flow specified by Fieldcheck must trigger at least one pulse within 30 s. Otherwise, 0 PULSE is specified. In this instance, the output is not completely tested in its components. Verification of the sensor Coil current test: For testing the sensor (for the electromagnetic measuring principle) the rise time of the coil current and further magnetic field-relevant characteristic variables are determined and compared with the corresponding set points. Exceeding the limit values leads to a FAILED message in the results presentation. " Achtung! Electrode integrity test (not for "Promag 10" type devices): When testing the sensor the electrode integrity test is carried out at the end of the verification. For this the measuring tube must be filled and the sensor signal cable must be inserted into the measuring amplifier socket. A corresponding message appears in the display and you are given the option to CONTINUE or CANCEL.

Description of device functions - Fieldcheck and electromagnetic measuring devices

Endress+Hauser 45 Evaluation of the results During verification, the current verification test section is displayed in the 'HOME position'.The results can be shown on the display once verification is complete. They can be filed in Fieldcheck under a storage location number of your choice (up to max. 20) together with the measuring device serial no. and the measuring device tag name. Via a connection with the FieldTool software, a protocol of the results can later be printed out and the data from the verification can be stored in an archive file system. Function group OUTPUTS During verification, the function and accuracy of a current output and a frequency output can be checked. Fieldcheck determines the current and frequency outputs available in the measuring device. Function description FUNCTION → VERIFICATION → OUTPUTS OUTPUTS Use this function to display the outputs determined in the flowmeter on the local display with the related terminals and to select the outputs to be tested. The outputs selected are displayed inversely. 1. Use the Pkeys to select the outputs which should be tested during the verification process. 2. Confirm the option selected with the F key. ! Note! • It is not possible to select two outputs of the same type. • Start the verification process in the VERIFICATE function (→Page 48). Two outputs of the same type are tested in two separate runs. After the 1st run, the message MORE OUTPUTS TO BE TESTED? appears on the display. If you want to test other outputs, Fieldcheck returns to the output selection area. 1. Select the output to be tested. ! Note! Having selected an output, make sure that the related terminal pair of the flowmeter is connected to the -Fieldcheck input before the verification process is started. The following combinations of current and frequency outputs are possible for PROline Promag: • I • I+I • I+F • I+I+F • I+F+F • I+P (only Promag 10) ! Note! • Fieldbus interfaces cannot be tested. • With this software-version current outputs run in passive mode cannot be tested.

Description of device functions - Fieldcheck and electromagnetic measuring devices

Proline Fieldcheck 46 Endress+Hauser Function group LIMIT VALUES You can specify the limit values for measured variables in this function group. If the deviation of a verification step (see OPERATION, P. 38) lies within the configured limit values the result is PASSED; if it lies outside of this range the result is FAILED. For a quick and unambiguous identification, the FAILED results are presented inversely on the display. ! Note! For a frequency output operated in the pulse mode the fixed error limit is: ± 1 pulse ± 0.1% of the pulse set number. Function description FUNCTION → VERIFICATION → LIMIT VALUES DEVIATION FLOW User input: 0.50 ... 10.0 Unit: % Factory setting: 0.5 DEVIATION CURRENT OUTPUT This function is only available if a current output was selected in the OUTPUTS function (→Page 45). User input: 0.02...10.0 Unit: mA Factory setting: 0.05 DEVIATION FREQUENCY OUTPUT This function is only available if a frequency output was selected in the OUTPUTS function (→Page 45). ! Note! This function is not available on "Promag 10" type devices. User input: 1.0...10.0 Unit: Hz Factory setting: 2.00

Description of device functions - Fieldcheck and electromagnetic measuring devices

Endress+Hauser 47 Function group PARAMETER Verification of the amplifier takes place along four measuring points MP1...MP4. Use this function group to enter the full scale value (100%) of the verification flow and the two selectable measuring points MP2 and MP3 (in % of the full scale value). ! Note! The value of MP1 is predetermined (5% of the full scale value MP4). Function description FUNCTION → VERIFICATION → PARAMETER FLOW 100% Use this function to enter the full scale value of the verification flow. User input: max 6-digit floating point number (without sign) ! Note! Only values between 4 and 8 m/s are permissible. The input limits MIN and MAX are displayed beyond the input line. Unit: The unit configured in the measuring device is adopted Factory setting: according 6m/s MEASURING POINT 2/3 Use this function to enter the measuring points MP2 and MP3 (in % of the full scale value). User input: MP2: integer between 10 and 90 MP3: integer between 15 and 95 ! Note! • The value of MP2 must be less than the value of MP3. • Both values are stored permanently. Unit: % Factory setting: MP2: 10 % MP3: 50%

Description of device functions - Fieldcheck and electromagnetic measuring devices

Proline Fieldcheck 48 Endress+Hauser Function group OPERATION Use this function group to verify the measuring point. Function description FUNCTION → VERIFICATION → OPERATION VERIFICATION Use this function to define the scope of the verification. ! Note! • A verification of the current and frequency outputs of the transmitter can only be achieved with the TRANSMITTER or TRANSM. + SENSOR option. • If SENSOR or TRANSM. + SENSOR is selected, the electrode integrity test is carried out at the end of the verification (→Page 44). Options: TRANSMITTER SENSOR TRANSM. + SENSOR Factory setting: TRANSMITTER VERIFICATE Use this function to control the verification. After completion of the verification, the question STORE DATA? appears in this function. Options: START The verification is started and the message VERIFICATION ACTIVE appears on the display.At the end of the verification, the message RESULT: PASSED/FAILED - STORE DATA? appears on the display. STOP Via the STOP option the active verification can be aborted. STORE DATA? The results of the verification can be saved using the STORE DATA? option. Factory setting: START STORAGE LOCATION Use this function to define where the results of the verification are to be stored. User input: 1...20 Factory setting: First available storage location ! Note! If there is no storage location available or the selected storage location is occupied, a corresponding warning is given before saving (“Overwrite storage location? Y/N”).

Description of device functions - Fieldcheck and Coriolis measuring systems

Endress+Hauser 49 11 Description of device functions - Fieldcheck and Coriolis measuring systems 11.1 Block FUNCTION Block Group Function group Function FUNCTION ⇒ SIMULATION P. 50 ⇒ CONFIGURATI ON P. 50 ⇒ OPERATION MODE P. 50 SIMUL SIGNAL P. 50 FLOW 1 P. 51 FLOW 2 P. 51 SIMUL. TIME 1 P. 51 ⇓⇑ ⇓⇑ SIMUL. TIME 2 P. 51 OPERATION P. 52 ⇒ SIMULATE P. 52 VERIFICATION P. 53 ⇒ APPLICAT./OU TP. P. 54 ⇒FLOW RANGE P. 54 TYPE UNIT FLOW P. 54 OUTPUTS P. 55 ⇓⇑ LIMIT VALUES P. 56 ⇒ DEVIATION FLOW P. 56 DEVIATION DENSITY P. 56 DEVIATION TEMPERATURE P. 56 DEVIATION CURRENT OUTPUT P. 56 DEVIATION FREQU. OUTPUT P. 56 ⇓⇑ PARAMETER P. 57 ⇒ FLOW 100% P. 57 MEAS. POINT 1/2 P. 57 ⇓⇑ OPERATION P. 58 ⇒VERIFICATION P. 58 VERIFICATE P. 58 STORAGE LOCATION P. 58

Description of device functions - Fieldcheck and Coriolis measuring systems

Proline Fieldcheck 50 Endress+Hauser 11.1.1 Group SIMULATION Function group CONFIGURATION Function description FUNCTION → SIMULATION → CONFIGURATION OPERATION MODE Use this function to select the operation mode for the simulation. There are two freely adjustable flow signals of -12.5 m/s to + 12.5 m/s and the assigned duration of simulation is 1 second to 999 seconds. Options: CONTINUOUS The flow behaviour in the process has low dynamics. F06-DXC10xxx-05-xx-xx-xx-000 Q1 = Simulation value 1 STEP The flow rates in the process are very dynamic and change very quickly. This operation mode is also suitable for the simulation of flows outside the tolerance limits (Error simulation). Here very dynamic or disrupted conditions can also be simulated. F06-DXC10xxx-05-xx-xx-xx-001 Q1 = Simulation value 1, Q2 = Simulation value 2, T1 = Duration of simulation 1, T2 = Duration of simulation 2 Factory setting: CONTINUOUS SIMULATION SIGNAL Use this function to select which measurement is to be simulated. Options: MASS FLOW VOLUME FLOW CORRECTED VOLUME FLOW (not for Promass 40) ! Note! • The option CORRECTED VOLUME FLOW is only available if the FIXED REFERENCE DENSITY option was selected in the CORRECTED VOLUME CALCULATION function in the tested device. • For the simulation of corrected volume flow, the system uses the values specified by the flowmeter in the REFERENCE DENSITY and UNIT REFERENCE DENSITY functions. Factory setting: MASS FLOW Q Q1 START STOP t Q Q1 Q2 T2 T1 START STOP t

Description of device functions - Fieldcheck and Coriolis measuring systems

Endress+Hauser 51 FLOW 1 Use this function to enter the flow value which is to be simulated. User input: 0.000...999999 (6-digit floating-point number, with sign) ! Note! The unit of the connected measuring device is adopted. Unit: The unit configured in the measuring device is adopted Factory setting: 0.000 FLOW 2 Use this function to enter the second flow value which is to be simulated. ! Note! • This function is only available if the STEP setting was selected in the OPERATION MODE function. • More information is provided in the OPERATION MODE function. User input: 0.000...999999 (6-digit floating-point number, with sign) ! Note! The unit of the connected measuring device is adopted. Unit: The unit configured in the measuring device is adopted Factory setting: 0.000 DURATION OF SIMULATION 1 Use this function to enter the duration of simulation signal 1 (flow). ! Note! • This function is only available if the STEP setting was selected in the OPERATION MODE function. • More information is provided in the OPERATION MODE function. User input: 1 ... 999 (max. 3-digit number ) Unit: s ( seconds ) Factory setting: 10 DURATION OF SIMULATION 2 Use this function to enter the duration of simulation signal 2 (flow). ! Note! • This function is only available if the STEP setting was selected in the OPERATION MODE function. • More information is provided in the OPERATION MODE function. User input: 1 ... 999 (max. 3-digit number ) Unit: s ( seconds ) Factory setting: 10 Function description FUNCTION → SIMULATION → CONFIGURATION

Description of device functions - Fieldcheck and Coriolis measuring systems

Proline Fieldcheck 52 Endress+Hauser Function group OPERATION Function description FUNCTION → SIMULATION → OPERATION SIMULATE Use this function to control the simulation. Options: START STOP Factory setting: START Display during the simulation: Main line Symbol of the selected operating mode and the simulated value (with unit). Additional line The measured value sent back by the measuring device. Information line In the SIMULATION operating mode, the current values of the current and frequency output of the tested device are displayed (multimeter function). Pulses are displayed as frequency if there is more than 1 pulse/second at the frequency input of Fieldcheck. In the STOP status the simulation signal value “0” is always given. Affected settings for the simulation, however, are maintained.

Description of device functions - Fieldcheck and Coriolis measuring systems

Endress+Hauser 53 11.1.2 Group VERIFICATION In the VERIFICATION group, a measuring device can be checked for the following characteristic values. • Functionality and deviation in the flow measurement (based on the set point preset by Fieldcheck) • Functionality and deviation of the current and frequency outputs if these outputs are available in the measuring device. The specified set points correspond in the gradation to the measuring points when determining flow. Data bus outputs such as PROFIBUS or FOUNDATION FIELDBUS cannot be directly checked by FieldCheck. • Characteristic values from internal tests for the sensor part. Verification of the transmitter Checking the linearity of the amplifier: Simulated flow at four measuring points which are determined via the 100% full scale value. The mass flow is verified using a constant density of 1.0kg/l. The presumed fluid velocity is 4...8m/s for the standard range and 0.4...0.8m/s for the low range. Checking a current output: Each current output can be verified in the configuration in the measuring device (CURRENT RANGE function ( 0-20 mA, 4-20 mA) in the same gradation (5% ... 100%) as the amplifier. For this purpose, the current output must be directly connected to the current input of the Fieldcheck. If more than one current output is available, verification takes place in several, menu-guided steps. Checking a frequency output (FREQUENCY operating mode) : Each frequency output can be verified in the configuration in the measuring device (START VALUE FREQUENCY, END VALUE FREQUENCY, OUTPUT SIGNAL functions) in the same gradation ( 5% ... 100% ) as the amplifier. For this purpose, the frequency output must be directly connected to the frequency input of the Fieldcheck. If more than one frequency output is available, verification takes place in several, menu-guided steps. Checking a frequency output (PULSE operating mode): Each pulse output available can be verified in the configuration in the measuring device (PULSE VALUE, PULSE WIDTH, OUTPUT SIGNAL functions). For this purpose, the flow specified by Fieldcheck must trigger at least one pulse within 30 s. Otherwise, 0 PULSE is specified. In this instance, the output is not completely tested in its components. Verification of the sensor circuits A series of internal tests are run through for checking the sensor connections and the closed-control loops of excitation current and frequency with plugs connected for the excitation current and amplifier. A PASSED or FAILED is output as a result. The test series are combined into four groups for the result output: • Excitation circuit • Frequency parameters • Temperature measuring points • Pick-up coils ! Note! One of the following options must be selected in the VERIFICATION function (→Page 58) to carry out the test series: • TRANSMITTER + SENSOR • SENSOR Evaluation of the results During verification, the current verification test section is displayed in the 'HOME position'. The results can be shown on the display once verification is complete. They can be filed in Fieldcheck under a storage location number of your choice (up to max. 20) together with the measuring device serial no. and the measuring device tag name.

Description of device functions - Fieldcheck and Coriolis measuring systems

Proline Fieldcheck 54 Endress+Hauser Via a connection with the FieldTool software, a protocol of the results can later be printed out and the data from the verification can be stored in an archive file system. Function group APPLICATION/OUTPUTS During verification, the function and accuracy of a current output and a frequency output can be checked. Fieldcheck determines the current and frequency outputs available in the measuring device. Function description FUNCTION → VERIFICATION → APPLICATION/OUTPUTS FLOW RANGE Use this function to select the amplifier verification range suitable for the application in question. Options: STANDARD RANGE: Flow range of the original factory calibration. LOW RANGE: Compared to the standard range, the flow range has dropped by a factor of 10. " Caution! If the LOW RANGE option is selected, the smallest permissible limit value (error limit) in the DEVIATION FLOW function (→Page 56) is higher than in the STANDARD RANGE option. Factory setting: STANDARD RANGE TYPE UNIT FLOW Use this function to select the unit type for the verification of the amplifier. Options: MASS FLOW CORRECTED VOLUME FLOW (not available for Promass 40) ! Note! The option CORRECTED VOLUME FLOW is only available if the FIXED REFERENCE DENSITY option was selected in the CORRECTED VOLUME CALCULATION function in the tested device. Factory setting: MASS FLOW

Description of device functions - Fieldcheck and Coriolis measuring systems

Endress+Hauser 55 OUTPUTS Use this function to display the outputs determined in the flowmeter on the local display with the related terminals and to select the outputs to be tested. 1. Use the Pkeys to select the outputs which should be tested during the verification process. 2. Confirm the option selected with the F key. ! Note! • It is not possible to select two outputs of the same type. • Start the verification process in the VERIFICATE function (→Page 58). Two outputs of the same type are tested in two separate runs. After the 1st run, the message MORE OUTPUTS TO BE TESTED? appears on the display. If you want to test other outputs, Fieldcheck returns to the output selection area. 1. Select the output to be tested. ! Note! Having selected an output, make sure that the related terminal pair of the flowmeter is connected to the -Fieldcheck input before the verification process is started. The following combinations of current and frequency outputs are possible for Proline Promass: • I • I+I • I+I+I • I+F • I+I+F • I+F+F ! Note! • Fieldbus interfaces cannot be tested. • With this software-version current outputs run in passive mode cannot be tested. Function description FUNCTION → VERIFICATION → APPLICATION/OUTPUTS

Description of device functions - Fieldcheck and Coriolis measuring systems

Proline Fieldcheck 56 Endress+Hauser Function group LIMIT VALUES You can specify the limit values for measured variables in this function group. If the deviation of a verification step (see OPERATION, P. 38) lies within the configured limit values the result is PASSED; if it lies outside of this range the result is FAILED. For a quick and unambiguous identification, the FAILED results are presented inversely on the display. Function description FUNCTION → VERIFICATION → LIMIT VALUES DEVIATION FLOW User input: 0.30...10.00 (for Promass 80/83 with sensors A, F, M) 0.50...10.00 (for Promass 80/83 with sensors E, H, I and Promass 40) 5.00...10.00 (if LOW RANGE is selected in the FLOW RANGE function, see Page 54) Unit: % Factory setting: 0.5 or 5.0 DEVIATION DENSITY User input: 0.50 ... 10.00 Unit: % Factory setting: 0.5 DEVIATION TEMPERATURE User input: 1.00...10.00 (for the units °C and K) 2.00...10.00 (for the units °F and R) Unit: The unit configured in the measuring device is adopted Factory setting: 2.0 (for the units °C and K) 4.0 (for the units °F and R) DEVIATION CURRENT OUTPUT This function is only available if a current output was selected in the OUTPUTS function (→Page 55). User input: 0.02...10.0 Unit: mA Factory setting: 0.05 DEVIATION FREQUENCY OUTPUT This function is only available if a frequency output was selected in the OUTPUTS function (→Page 55). User input: 1.0...10.0 Unit: Hz Factory setting: 2.00

Description of device functions - Fieldcheck and Coriolis measuring systems

Endress+Hauser 57 Function group PARAMETER Verification of the amplifier takes place along four measuring points MP1...MP4. Use this function group to enter the full scale value (100%) of the verification flow and the two selectable measuring points MP2 and MP3 (in % of the full scale value). ! Note! The value of MP1 is predetermined (5% of the full scale value MP4). Function description FUNCTION → VERIFICATION → PARAMETER FLOW 100% Use this function to enter the full scale value of the verification flow. User input: max 6-digit floating point number (without sign) ! Note! The permitted value range depends on the setting in the FLOW RANGE function (→Page 54). The input limits MIN and MAX are displayed beyond the input line. Unit: The unit configured in the measuring device is adopted Factory setting: according 6m/s MEASURING POINT 2/3 Use this function to enter the measuring points MP2 and MP3 (in % of the full scale value). User input: MP2: integer between 10 and 90 MP3: integer between 15 and 95 ! Note! • The value of MP2 must be less than the value of MP3. • Both values are stored permanently. Unit: % Factory setting: MP2: 10 % MP3: 50%

Description of device functions - Fieldcheck and Coriolis measuring systems

Proline Fieldcheck 58 Endress+Hauser Function group OPERATION Use this function group to specify the scope of the verification and to start/stop the verification. Function description FUNCTION → VERIFICATION → OPERATION VERIFICATION Use this function to define the scope of the test. ! Note! A verification of the current and frequency outputs of the transmitter can only be achieved with the TRANSMITTER or TRANSM. + SENSOR option. Options: TRANSMITTER: The meter electronics and the electronics of the current and frequency outputs selected are tested. Excitation current cable and sensor signal cable are plugged out. SENSOR The temperature measuring points, the closed-control loops of the excitation current and frequency and the pick-up coils are tested. Excitation current cable and sensor signal cable are plugged in. TRANSM. + SENSOR All test groups are run through. At the end of the transmitter test, you are requested to reconnect the excitation current cable and sensor signal cable. Factory setting: TRANSMITTER VERIFICATE Use this function to control the verification. After completion of the verification, the question STORE DATA? appears in this function. Options: START The verification is started and the message VERIFICATION ACTIVE appears on the display.At the end of the verification, the message RESULT: PASSED/FAILED - STORE DATA? appears on the display. STOP Via the STOP option the active verification can be aborted. STORE DATA? The results of the verification can be saved using the STORE DATA? option. Factory setting: START STORAGE LOCATION Use this function to define where the results of the verification are to be stored. User input: 1...20 Factory setting: First available storage location ! Note! If there is no storage location available or the selected storage location is occupied, a corresponding warning is given before saving (“Overwrite storage location? Y/N”).

Description of device functions - Fieldcheck and vortex measuring systems

Endress+Hauser 59 12 Description of device functions - Fieldcheck and vortex measuring systems 12.1 Block FUNCTION Block Group Function group Function FUNCTION ⇒ SIMULATION P. 60 ⇒ CONFIGURATI ON P. 60 ⇒ OPERATION MODE P. 60 SIMUL. SIGNAL P. 60 FLOW 1 P. 61 FLOW 2 P. 61 SIMUL. TIME 1 P. 61 ⇓⇑ ⇓⇑ SIMUL. TIME 2 P. 61 CONFIGURATI ON P. 62 SIMULATE P. 61 VERIFICATION P. 63 OUTPUTS P. 64 OUTPUTS P. 64 ⇓⇑ LIMIT VALUESP. 65 DEVIATION FLOW P. 65 DEV.CURR.OU TPUT P. 65 ⇓⇑ PARAMETER P. 66 FLOW 100% P. 66 MEAS. POINT 1/2 P. 66 ⇓⇑ OPERATION P. 67 VERIFICATION P. 67 VERIFICATE P. 67 STORAGE LOCATION P. 67

Description of device functions - Fieldcheck and vortex measuring systems

Proline Fieldcheck 60 Endress+Hauser 12.1.1 Group SIMULATION Please note the following: • In contrast to electromagnetic or Coriolis measuring devices, a gap-free flow range 0 - Qmax cannot be simulated with vortex flowmeters. Below a certain fluid speed (approx. 0.15 m/s for water), the amplitude of the vortex signal becomes so small that it is no longer possible to record the measured value. For this reason, when setting the simulated flow, the values 0, the smallest flow Qmin and the largest flow Qmax are displayed. • The flow value actually simulated can deviate slightly from the value entered in the FLOW 1/2 function. The value actually simulated is displayed on the first line in the HOME position of the Fieldcheck local display. Function group CONFIGURATION Function description FUNCTION → SIMULATION → CONFIGURATION OPERATION MODE Use this function to select the operation mode for the simulation. The following signals can be freely configured: • Operation mode GAS/STEAM: 0, Qmin...70 m/s • Operation mode LIQUID: 0, Qmin...9 m/s Options: CONTINUOUS The flow behaviour in the process has low dynamics. F06-DXC10xxx-05-xx-xx-xx-000 Q1 = Simulation value 1 STEP The flow rates in the process are very dynamic and change very quickly. This operation mode is also suitable for the simulation of flows outside the tolerance limits (Error simulation). Here very dynamic or disrupted conditions can also be simulated. F06-DXC10xxx-05-xx-xx-xx-001 Q1 = Simulation value 1, Q2 = Simulation value 2, T1 = Duration of simulation 1, T2 = Duration of simulation 2 SIMULATION SIGNAL Use this function to view the set simulation signal in the measuring device. Display: VOLUME FLOW CALCULATED MASS FLOW CALCULATED CORRECTED VOLUME FLOW ! Note! • For the simulation of a calculated mass flow, the system uses the value specified by the flowmeter in the OPERATING DENSITY function. • For the simulation of corrected volume flow, the system uses the values specified by the flowmeter in the OPERATING DENSITY and REFERENCE DENSITY functions. Q Q1 START STOP t Q Q1 Q2 T2 T1 START STOP t

Description of device functions - Fieldcheck and vortex measuring systems

Endress+Hauser 61 FLOW 1 Use this function to specify the first flow value which is to be simulated. User input: 0.000...99999 (6-digit floating-point number) ! Note! The flow limits are adopted depending on the option selected in the APPLICATION function of the flowmeter. Values smaller than Qmin cannot be entered (→Page 60). They have the following values: • GAS/STEAM: 0, Qmin...70 m/s • LIQUID: 0, Qmin...9 m/s Unit: The unit configured in the measuring device is adopted Factory setting: 0.000 FLOW 2 Use this function to enter the second flow value which is to be simulated. ! Note! • This function is only available if the STEP setting was selected in the OPERATION MODE function. • More information is provided in the OPERATION MODE function. • The flow limits are adopted depending on the option selected in the APPLICATION function of the flowmeter. Values smaller than Qmin cannot be entered (→Page 60). They have the following values: – GAS/STEAM: 0, Qmin...70 m/s – LIQUID: 0, Qmin...9 m/s User input: 0.000...99999 (6-digit floating-point number) Unit: The unit configured in the measuring device is adopted Factory setting: 0.000 DURATION OF SIMULATION 1 Use this function to enter the duration of simulation signal 1 (flow). ! Note! • This function is only available if the STEP setting was selected in the OPERATION MODE function. • More information is provided in the OPERATION MODE function. User input: 1 ... 999 (max. 3-digit number ) Unit: s ( seconds ) Factory setting: 10 DURATION OF SIMULATION 2 Use this function to enter the duration of simulation signal 2 (flow). ! Note! • This function is only available if the STEP setting was selected in the OPERATION MODE function. • More information is provided in the OPERATION MODE function. User input: 1 ... 999 (max. 3-digit number ) Unit: s ( seconds ) Factory setting: 10 Function description FUNCTION → SIMULATION → CONFIGURATION

Description of device functions - Fieldcheck and vortex measuring systems

Proline Fieldcheck 62 Endress+Hauser Function group OPERATION Function description FUNCTION → SIMULATION → OPERATION SIMULATE Use this function to control the simulation. Options: START STOP Factory setting: START Display during the simulation: Main line Symbol of the selected operating mode and the simulated value (with unit). Additional line The measured value sent back by the measuring device. Information line In the SIMULATION operating mode, the current values of the current and pulse output of the tested device are displayed (multimeter function). Pulses are displayed as frequency if there is more than 1 pulse/second at the frequency input of Fieldcheck. In the STOP status the simulation signal value “0” is always given. Affected settings for the simulation, however, are maintained.

Description of device functions - Fieldcheck and vortex measuring systems

Endress+Hauser 63 12.1.2 Group VERIFICATION In the VERIFICATION group, a measuring device can be checked for the following characteristic values. • Functionality and deviation in the flow measurement (based on the set point preset by Fieldcheck) • Functionality and deviation of the current output and optional pulse output if these outputs are available in the measuring device. The specified set points correspond in the gradation to the measuring points when determining flow. Fieldbus outputs such as PROFIBUS or FOUNDATION FIELDBUS cannot be directly checked by Fieldcheck. ! Note! The device can be tested at its mounting location; power is supplied by the switching unit connected to the current output. In order to test the current output for accuracy at the mounting location, it must be disconnected from the switching unit and connected to the Fieldcheck current input. Verification of the transmitter Checking the linearity of the amplifier: Simulated flow at four measuring points which are determined via the 100% full scale value (for a presumed fluid velocity of 8 m/s for liquids and 60 m/s for gases/steam). Checking the current output: The current output can be verified in the configuration in the measuring device (CURRENT RANGE function (0-20 mA, 4-20 mA) in the same gradation (5% ... 100%) as the amplifier. For this purpose, the current output must be directly connected to the current input of the Fieldcheck. Checking the pulse output: An existing pulse output can be verified in the configuration in the measuring device (PULSE VALUE, PULSE WIDTH, OUTPUT SIGNAL functions). For this purpose, the pulse output must be directly connected to the frequency input of the Fieldcheck. Evaluation of the results During verification, the current verification test section is displayed in the 'HOME position'. The results can be shown on the display once verification is complete. They can be filed in FieldCheck under a storage location number of your choice (up to max. 20) together with the measuring device serial no. and the measuring device tag name. Via a connection with the FieldTool software, a protocol of the results can later be printed out and the data from the verification can be stored in an archive file system.

Description of device functions - Fieldcheck and vortex measuring systems

Proline Fieldcheck 64 Endress+Hauser Function group OUTPUTS During verification, the function and accuracy of the current output and the optional pulse output can be checked. Fieldcheck determines the outputs available in the measuring device. Function description FUNCTION → VERIFICATION → OUTPUTS OUTPUTS Use this function to display the outputs determined in the flowmeter on the local display with the related terminals and to select the outputs to be tested. 1. Use the Pkeys to select the outputs which should be tested during the verification process. 2. Confirm the option selected with the Fkey. ! Note! Start the verification process in the VERIFICATE function (→Page 67). The following combinations are possible for Proline Prowirl: • I • I+P ! Note! Fieldbus interfaces cannot be tested.

Description of device functions - Fieldcheck and vortex measuring systems

Endress+Hauser 65 Function group LIMIT VALUES You can specify the limit values for measured variables in this function group. If the deviation of a verification step (see OPERATION, P. 38) lies within the configured limit values the result is PASSED; if it lies outside of this range the result is FAILED. For a quick and unambiguous identification, the FAILED results are presented inversely on the display. Function description FUNCTION → VERIFICATION → LIMIT VALUES DEVIATION FLOW (when measuring liquids) User input: 1.50...10.00 Unit: % Factory setting: 1.5 DEVIATION FLOW (when measuring gas or steam) User input: 2.0...10.00 Unit: % Factory setting: 2.0 DEVIATION CURRENT OUTPUT This function is only available if the current output was selected in the OUTPUTS function (→Page 64). User input: 0.02...1.00 Unit: mA Factory setting: 0.05

Description of device functions - Fieldcheck and vortex measuring systems

Proline Fieldcheck 66 Endress+Hauser Function group PARAMETER Verification of the amplifier takes place along four measuring points MP1...MP4. Use this function group to enter the two selectable measuring points MP2 and MP3 (in % of the full scale value). ! Note! The value of MP1 is predetermined (5% of the full scale value MP4). Function description FUNCTION → VERIFICATION → PARAMETER FLOW 100% Use this function to display the full scale value of the verification flow. In liquid applications, the 100% value usually corresponds to a fluid velocity of 8 m/s and a velocity of 60 m/s in gas/steam applications. MEASURING POINT 2/3 Use this function to enter the measuring points MP2 and MP3 (in % of the full scale value). User input: MP2: integer between 10 and 90 MP3: integer between 15 and 95 ! Note! • The value of MP2 must be less than the value of MP3. • Both values are stored permanently. • In demanding measuring applications, the usable measuring range of the vortex measuring device can be restricted to such an extent that the lower MP1 verification measuring point cannot achieve the 5% value. In this instance, the minimum which can be achieved of the measurable flow is determined and calculated as a % value of the maximum value MP 4 and displayed. MP2 and MP3 are then determined again as default values and displayed. However, they are not saved when the work with Fieldcheck is completed. Unit: % Factory setting: MP2: 10 % MP3: 50%

Description of device functions - Fieldcheck and vortex measuring systems

Endress+Hauser 67 Function group OPERATION Use this function group to specify the scope of the verification and to start/stop the verification. Function description FUNCTION → VERIFICATION → OPERATION VERIFICATION Use this function to define the scope of the test. Options: TRANSMITTER SENSOR TRANSMITTER + SENSOR Factory setting: TRANSMITTER VERIFICATE Use this function to control the verification. After completion of the verification, the question STORE DATA? appears in this function. Options: START The verification is started and the message VERIFICATION ACTIVE appears on the display.At the end of the verification, the message RESULT: PASSED/FAILED - STORE DATA? appears on the display. STOP Via the STOP option the active verification can be aborted. STORE DATA? The results of the verification can be saved using the STORE DATA? option. Factory setting: START STORAGE LOCATION Use this function to define where the results of the verification are to be stored. User input: 1...20 Factory setting: First available storage location ! Note! If there is no storage location available or the selected storage location is occupied, a corresponding warning is given before saving (“Overwrite storage location? Y/N”).

Accessories. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24

Ambient temperature range . . . . . . . . . . . . . . . . . . . . . . . 25 ASSIGN (current output) . . . . . . . . . . . . . . . . . . . . . . . . . 33 ASSIGN (frequency output) . . . . . . . . . . . . . . . . . . . . . . . 33 B BACK LIGHT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 BASIC FUNCTIONS (Block) . . . . . . . . . . . . . . . . . . . . . . . 31 Block BASIC FUNCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . 31 FUNCTION Coriolis measuring devices. . . . . . . . . . . . . . . . . . . 49 Electromagnetic measuring devices . . . . . . . . . . . . 39 Vortex measuring devices . . . . . . . . . . . . . . . . . . . 59 SUPERVISION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 USER INTERFACE . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Blocks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 C CALIBRATION DATE Fieldcheck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Simubox. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7, 26 Charging the battery Battery charging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Commissioning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 Communication FieldTool . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 CONFIGURATION (Function group) Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 50 Electromagnetic measuring devices . . . . . . . . . . . . . . . 40 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 60 Connection see Electrical connection CONTRAST LCD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 CONTROL (Group) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 CURRENT RANGE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 D DATE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 Declaration of conformity . . . . . . . . . . . . . . . . . . . . . . . . . . 7 DEL. STORAGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 DENSITY VALUE. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 Description of functions Sections 9 to 12 Design . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Designated use. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 DEVIATION CURRENT OUTPUT Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 56 Electromagnetic measuring devices . . . . . . . . . . . . . . . 46 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 65 DEVIATION DENSITY Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 56 DEVIATION FLOW Electromagnetic measuring devices . . . . . . . . . . . . . . . 46 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 65 DEVIATION FREQUENCY OUTPUT Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 56 Electromagnetic measuring devices . . . . . . . . . . . . . . . 46 DEVIATION MASS FLOW Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 56 DEVIATION TEMPERATURE Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 56 Device designation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Device functions General section . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 Test Coriolis measuring devices . . . . . . . . . . . . . . . . . . 49 Test electromagnetic measuring devices. . . . . . . . . . . . 39 Test vortex measuring devices. . . . . . . . . . . . . . . . . . . 59 dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 DIRECTION OF FLOW . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18, 26 Display and operating elements. . . . . . . . . . . . . . . . . . . . . 17 Display elements. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Display icons . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19 Documentation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 DURATION OF SIMULATION 1 Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 51 Electromagnetic measuring devices . . . . . . . . . . . . . . . 41 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 61 DURATION OF SIMULATION 2 Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 51 Electromagnetic measuring devices . . . . . . . . . . . . . . . 42 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 61 E Electrical connection Assignment of Fieldcheck connections. . . . . . . . . . . . . 16 Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8 Promag 50/53 (compact version) . . . . . . . . . . . . . . . . 10 Promag 50/53 (remote version) . . . . . . . . . . . . . . . . . 11 Promass 80/83 (compact version) . . . . . . . . . . . . . . . . 12 Promass 80/83 (remote version) . . . . . . . . . . . . . . . . . 13 Prowirl 72. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Electrode integrity test Electromagnetic measuring devices 44 Electromagnetic compatibility (EMC) . . . . . . . . . . . . . . . . 25 EMC - operation in environment with strong interference . . 4 EN 61010 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 END FREQUENCY. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Evaluation of the results Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 53 Electromagnetic measuring devices . . . . . . . . . . . . . . . 45 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 63 External cleaning . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23

FIELDCHECK (Function group). . . . . . . . . . . . . . . . . . . . . 38

FLOW 1 Coriolis measuring devices. . . . . . . . . . . . . . . . . . . . . . 51 Electromagnetic measuring devices . . . . . . . . . . . . . . . 41 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 61 FLOW 100% . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66 Coriolis measuring devices. . . . . . . . . . . . . . . . . . . . . . 57 Electromagnetic measuring devices . . . . . . . . . . . . . . . 47 FLOW 2 Coriolis measuring devices. . . . . . . . . . . . . . . . . . . . . . 51 Electromagnetic measuring devices . . . . . . . . . . . . . . . 41 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 61 FORMAT DATE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 FORMAT TIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 FUNCTION (Block) Coriolis measuring devices. . . . . . . . . . . . . . . . . . . . . . 49 Electromagnetic measuring devices . . . . . . . . . . . . . . . 39 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 59 Function group CONFIGURATION Coriolis measuring devices . . . . . . . . . . . . . . . . . . . 50 Electromagnetic measuring devices. . . . . . . . . . . . . 40 Vortex measuring devices. . . . . . . . . . . . . . . . . . . . 60 FIELDCHECK . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 LIMIT VALUES Coriolis measuring devices . . . . . . . . . . . . . . . . . . . 56 Electromagnetic measuring devices. . . . . . . . . . . . . 46 Vortex measuring devices. . . . . . . . . . . . . . . . . . . . 65 OPERATION (Group SIMULATION) Coriolis measuring devices . . . . . . . . . . . . . . . . . . . 52 Electromagnetic measuring devices. . . . . . . . . . . . . 43 Vortex measuring devices. . . . . . . . . . . . . . . . . . . . 62 OPERATION (Group VERIFICATION) Coriolis measuring devices . . . . . . . . . . . . . . . . . . . 58 Electromagnetic measuring devices. . . . . . . . . . . . . 48 Vortex measuring devices. . . . . . . . . . . . . . . . . . . . 67 OUTPUTS Coriolis measuring devices . . . . . . . . . . . . . . . . . . . 54 Electromagnetic measuring devices. . . . . . . . . . . . . 45 Vortex measuring devices. . . . . . . . . . . . . . . . . . . . 64 PARAMETER Coriolis measuring devices . . . . . . . . . . . . . . . . . . . 57 Electromagnetic measuring devices. . . . . . . . . . . . . 47 Vortex measuring devices. . . . . . . . . . . . . . . . . . . . 66 SIMUBOX . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Function groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Function matrix . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Brief operating instructions . . . . . . . . . . . . . . . . . . . . . 20 Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 G Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Group CONTROL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 OUTPUTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 PROCESS PARAMETER. . . . . . . . . . . . . . . . . . . . . . . . 34 READ IN PARAMETERS . . . . . . . . . . . . . . . . . . . . . . . 35 RESULTS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 SENSOR PARAMETER. . . . . . . . . . . . . . . . . . . . . . . . . 32 SIMULATION Coriolis measuring devices . . . . . . . . . . . . . . . . . . . 50 Electromagnetic measuring devices . . . . . . . . . . . . . 40 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . 60 VERIFICATION Coriolis measuring devices . . . . . . . . . . . . . . . . . . . 53 Electromagnetic measuring devices . . . . . . . . . . . . . 44 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . 63 VERSION INFO . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Groups . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20 Guidelines. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 H History (software) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 HOME position (display operating mode) . . . . . . . . . . . . . . 17 HW. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 I Identification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 INFO STORAGE LOCATION. . . . . . . . . . . . . . . . . . . . . . . 36 Input. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 K K-FACTOR POSITIVE/NEGATIVE . . . . . . . . . . . . . . . . . . 32 L LANGUAGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29 LIMIT VALUES (Function group) Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 56 Electromagnetic measuring devices. . . . . . . . . . . . . . . . 46 Vortex measuring devices. . . . . . . . . . . . . . . . . . . . . . . 65 Local display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18 M Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 MANUFACTURING/TEST . . . . . . . . . . . . . . . . . . . . . . . . 38 Measured error (inputs) . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Measured error (outputs) . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 MEASURING POINT 2/3 Coriolis measuring devices . . . . . . . . . . . . . . . . . . . 57 Electromagnetic measuring devices. . . . . . . . . . . . . . . . 47 Vortex measuring devices. . . . . . . . . . . . . . . . . . . . . . . 66 Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Mechanical construction . . . . . . . . . . . . . . . . . . . . . . . . . . 26 N Nameplate Fieldcheck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Simubox . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Nameplate specifications Fieldcheck . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Simubox . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 NOMINAL DIAMETER . . . . . . . . . . . . . . . . . . . . . . . . . . . 32

Operating conditions (Environment) . . . . . . . . . . . . . . . . . 25

OPERATING DENSITY . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 Operating elements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 OPERATION (Group SIMULATION) Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 52 Electromagnetic measuring devices . . . . . . . . . . . . . . . 43 Vortex measuring devices. . . . . . . . . . . . . . . . . . . . . . . 62 OPERATION (Group VERIFICATION) Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 58 Electromagnetic measuring devices . . . . . . . . . . . . . . . 48 Vortex measuring devices. . . . . . . . . . . . . . . . . . . . . . . 67 OPERATION MODE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 50 Electromagnetic measuring devices . . . . . . . . . . . . . . . 40 Vortex measuring devices. . . . . . . . . . . . . . . . . . . . . . . 60 Operational safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Ordering information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 Other standards. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 OUTPUT. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 OUTPUT SIGNAL . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 OUTPUTS (Function group) Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 54 Electromagnetic measuring devices . . . . . . . . . . . . . . . 45 Vortex measuring devices. . . . . . . . . . . . . . . . . . . . . . . 64 OUTPUTS (Group) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 P PARAMETER (Function group) Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . . 57 Electromagnetic measuring devices . . . . . . . . . . . . . . . 47 Vortex measuring devices. . . . . . . . . . . . . . . . . . . . . . . 66 Performance characteristics . . . . . . . . . . . . . . . . . . . . . . . . 25 Power consumption. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 PROCESS PARAMETER (Group) . . . . . . . . . . . . . . . . . . . . 34 PROFILE TIME 1 Electromagnetic measuring devices . . . . . . . . . . . . . . . 42 PROFILE TIME 2 Electromagnetic measuring devices . . . . . . . . . . . . . . . 42 Promag 50/53 (compact version) Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . 10 Promag 50/53 (remote version) Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . 11 Promag 80/83 (compact version) Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . 12 Promag 80/83 (remote version) Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Prowirl 72 Electrical connection . . . . . . . . . . . . . . . . . . . . . . . . . . 14 PULSE VALUE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 R Radiated RF noise (EN 61326-1). . . . . . . . . . . . . . . . . . . . . 4 READ IN PARAMETERS (Group) . . . . . . . . . . . . . . . . . . . 35 REFERENCE DENSITY. . . . . . . . . . . . . . . . . . . . . . . . . . . 34 Reference operating conditions . . . . . . . . . . . . . . . . . . . . . 25 Registered trademarks. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Remote configuration . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26 RESULTS (Group). . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 RESULTS VERIFICATION CORIOLIS . . . . . . . . . . . . . . . . 37 RESULTS VERIFICATION MAGMETER . . . . . . . . . . . . . . 36 RESULTS VERIFICATION VORTEX . . . . . . . . . . . . . . . . . 37 S Safety conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Safety instructions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Scope of delivery. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 Sensor circuit verification (Coriolis measuring devices) . . . 53 SENSOR PARAMETER (Group) . . . . . . . . . . . . . . . . . . . . 32 Sensor verification Electromagnetic measuring devices . . . . . . . . . . . . . . . 44 SERIAL NUMBER Fieldcheck. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Simubox . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 SERIAL NUMBER (Function) Measuring device . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 SIMUBOX (Function group) . . . . . . . . . . . . . . . . . . . . . . . 38 SIMULATE Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 52 Electromagnetic measuring devices . . . . . . . . . . . . . . . 43 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 62 SIMULATION (Group) Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 50 Electromagnetic measuring devices . . . . . . . . . . . . . . . 40 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 60 SIMULATION SIGNAL Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 50 Electromagnetic measuring devices . . . . . . . . . . . . . . . 41 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 60 Software history . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 STANDARD DENSITY . . . . . . . . . . . . . . . . . . . . . . . . . . . 34 START FREQUENCY . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 STORAGE LOCATION . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 58 Electromagnetic measuring devices . . . . . . . . . . . . . . . 48 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 67 STORAGE LOCATION VERIFICATION . . . . . . . . . . . . . . 36 SUPERVISION (Block) . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 SW. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38 T TAG NAME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32 Technical data at a glance . . . . . . . . . . . . . . . . . . . . . . . . . 25 TIME . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30 Transmitter verification Coriolis measuring devices . . . . . . . . . . . . . . . . . . . . . 53 Electromagnetic measuring devices . . . . . . . . . . . . . . . 44 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 63 TYPE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 38

User interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26

USER INTERFACE (Block). . . . . . . . . . . . . . . . . . . . . . . . . 29 V VALUE 0_4 mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 VALUE 20 mA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 VALUE f- . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33–34 VERIFICATE (Function) Coriolis measuring devices. . . . . . . . . . . . . . . . . . . . . . 58 Electromagnetic measuring devices . . . . . . . . . . . . . . . 48 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 67 VERIFICATION (Function) Coriolis measuring devices. . . . . . . . . . . . . . . . . . . . . . 58 Electromagnetic measuring devices . . . . . . . . . . . . . . . 48 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 67 VERIFICATION (Group) Coriolis measuring devices. . . . . . . . . . . . . . . . . . . . . . 53 Electromagnetic measuring devices . . . . . . . . . . . . . . . 44 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . 63 VERSION INFO (Group) . . . . . . . . . . . . . . . . . . . . . . . . . . 38 Vortex measuring devices . . . . . . . . . . . . . . . . . . . . . . . . . 66 W Wiring see Electrical connection Z ZERO POINT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32