Satelec X-Mind DC Installation Manual
Read Online
Read the Satelec X-Mind DC Installation Manual online using the built-in PDF viewer.
Text of the manual
Installation manual / Manuel d’installation
Installationshandbuch / Installatiehandleiing
Intraoral X-ray system at constant potential
INSTALLATION & MAINTENANCE MANUAL
This manual should always be kept in proximity to the device
“X-Mind DC” X-RAY SYSTEM
page
6
2.
IDENTIFICATION TAGS
page
7
3.
OVERALL DIMENSIONS
page
8
4.
INSTALLATION SPECIFICATIONS
page
12
5.
INSTALLATION
page
15
5.1 UNPACKING
page
15
5.2 ASSEMBLING THE WALL PLATE
page
16
5.3 ASSEMBLING THE BRACKET
page
18
5.4 ASSEMBLING THE PANTOGRAPH-TYPE ARM
page
20
5.5 CONNECTION TO THE FEEDING TERMINAL BOARD
page
21
5.6 ASSEMBLING THE TUBEHEAD
page
24
5.7 BALANCING THE PANTOGRAPH-TYPE ARM
page
27
5.8 ASSEMBLING THE “X-Mind DC” TIMER
page
30
5.9 TIMER ELECTRIC CONNECTION
page
32
6.
CONTROL PANEL
page
34
7.
SYSTEM CONFIGURATION
page
36
8.
CHANGING THE CONFIGURATION
page
38
9.
START-UP
page
40
10. CHECKING THE INSTALLATION
page
41
11. CHECKING THE EXPOSURE FACTORS
page
43
12. DIAGNOSTICS
page
44
13. CALIBRATION
page
45
13. ERROR MESSAGES
page
46
14. FUSE REPLACEMENT
page
48
15. MAINTENANCE
page
49
16. CLEANING OUTER SURFACES
page
50
17. SHOULD A REPAIR BECOME NECESSARY
page
51
18. DISPOSAL
page
51
19. ATTACHMENTS
page
52
SUMMARY
Before beginning to use the “X-Mind DC” X-ray system, instructions contained herein should be
carefully read and followed so as to obtain the best possible performance.
Always pay close attention to the CAUTION, WARNING, and PLEASE NOTE messages when
operating the system.
LEGEND
CAUTION
The word CAUTION refers to events which might compromise the operator's personal safety
or cause injuries to people.
WARNING
The word WARNING refers to events which might compromise the radiographic system’s
performance.
PLEASE NOTE
The words PLEASE NOTE refer to special indications to facilitate maintenance or make
important information clearer.
PRELIMINARY INFORMATION
The installer is responsible for the installation, with regards to the system safety and
operation.
For safe and reliable installation of the “X-Mind DC” X-ray system it is advisable to:
9 Check that the voltage mentioned in the rating plate matches the line voltage
9 Install the “X-Mind DC” X-ray system according to the procedures described in this
manual
9 Provide the user with any information regarding the use of the “X-Mind DC” X-ray system
pursuant to the manual
9 Certify the work done by a Statement of Compliance
9 Return the duly filled in warranty certificate to “Satelec S.A.S”: if this is not done, the
warranty is not valid
INSTALLER INFORMATION
The “X-Mind DC” X-ray system (Fig. 1) consists of:
c TUBEHEAD
d SPACER CONE
e PANTOGRAPH-TYPE ARM
f TIMER
g WALL PLATE
h BRACKET
OPTIONAL
− long 12” (31 cm) cone
− short 8” (20 cm) cone
− long cone 12” (31 cm) with a rectangular section sized 44x35mm
− second “CONTROL BUTTON” with extension cable
− RX signaling lamp for external use: X-Mind LIGHT
− remote control button: X-Mind ECB
Fig. 1
1. “X-Mind DC” X-RAY SYSTEM
f
g
e
h
c
d
The identification tags on the tubehead, on the timer and on the cone indicate the model name,
the serial number, the manufacturing date and the symbols of the main technical
characteristics.
ID TAG OF THE
“X-Mind DC”
TUBEHEAD
ID TAG OF THE
“X-Mind DC”
TIMER
ID TAG OF THE LONG 12” CONE
ID TAG OF THE SHORT 8” CONE
ID TAG OF THE RECTANGULAR CONE
GRADUATED SCALE TAG ON TUBEHEAD
Pictograms used
This symbol guarantees that the X-ray system complies with the regulations
contained in the European Directive EEC 93/42 regarding Medical Devices
The degree of protection against direct and indirect electric contacts is B type
Refer to Manual's instructions
N
“NEUTRAL" power supply wire
L
“PHASE" power supply wire
T
Ground cable
2. IDENTIFICATION TAGS
Fig. 1A, 1B, 1C give the overall dimensions of the possible supply conditions:
Â
SHORT BRACKET (optional)
length 41 cm
-
16.2”
Â
STANDARD BRACKET
length 82,5 cm
-
32.5”
Â
LONG BRACKET (optional)
length 110 cm
-
43.5”
SHORT BRACKET 41 cm
Fig. 1A
STANDARD BRACKET 82.5 cm
Fig. 1B
LONG BRACKET 110 cm
Fig. 1C
3. OVERALL DIMENSIONS
Fig. 2, 3A, 3B, 3C e 4 show the typical dimensions of the X-ray system:
STANDARD BRACKET 82.5 cm
Fig. 2
STANDARD BRACKET 82.5 cm
Fig. 3A
Fig. 3B
Fig. 3C
SHORT BRACKET 41 cm
LONG BRACKET 110 cm
Prior to installing the X-ray system the Office Owner must ascertain that:
the environment, the electrical system and the power supply comply with the
requirements specified, otherwise he must provide the required adjustment works.
1.
ENVIRONMENT REQUIREMENTS
a. The installation environment must be of appropriate width.
Using the sizes and overall dimensions provided (refer to §3 “OVERALL DIMENSIONS”),
ensure that no obstacles are present while positioning the X-ray system
b. The environment must not be subject to explosion risks and must not be pressurized
c. While operating, the ambient temperature must range within +5°C and +40°C
d. The storage temperature must range within -15°C and +50°C
e. The relative humidity must range within 25 % and 75 %
2.
SUPPORTING WALL REQUIREMENTS
a. The X-ray system supporting wall must be able to stand 200 kg tear at every point
PLEASE NOTE
The nature and consistency of the wall must be checked and if required, seek the
support of a brickwork expert. Walls of uncertain consistency must be reinforced with
a buried counter plate or with a sandwich type system.
3.
ELECTRICAL SYSTEM REQUIREMENTS
a. The electrical system must comply with regulations in force
b. The electrical system must be able to supply the power and voltage required in the
Manufacturer's rating plate of the X-ray system as shown in Chart 1
CHART 1
MANUFACTURER’S RATING PLATE
230 V ± 15%
115 V ± 15%
RATED LINE VOLTAGE
230 V
115 V
MINIMUM VOLTAGE
196 V
98 V
MAXIMUM LINE VOLTAGE
264 V
132 V
FREQUENCY
50/60 Hz
50/60 Hz
ELECTRICAL
SYSTEM
POWER ABSORBED DURING EXPOSURE
850 W
850 W
4. INSTALLATION SPECIFICATIONS
ELECTRICAL LINE REQUIREMENTS
a.
The electrical line must be "single-phase" type
b.
It is essential that a 16A – 250V, magnetothermal differential switch be fitted
upstream of the X-ray system, with differential protection In<= 30mA
(refer to §20 “ATTACHMENTS”)
c.
The power cords of the timer and the tubehead connection conductors with must be
two-pole + ground, and in proportion to the length of the power cord, as shown in
Chart 2
CHART 2
POWER SUPPLY VOLTAGE
MINIMUM
CONDUCTOR
SECTION
MAXIMUM
LINE
LENGTH
1.5 mm2
10 meters
196V to 264V
2.5 mm2
20 meters
1.5 mm2
10 meters
98V to 132V
2.5 mm2
20 meters
PLEASE NOTE
For longer lines, the cord section must be increased in proportion.
d.
The communication cables (C11, C12 – C21, C22) between the timer and X-Ray must
be two-pole, twisted and shielded with >0.25 mm2 section (example: Belden 9501
type)
e.
The cables (S11, S12 – S21, S22) connecting the timer and the signal lamps located
outside the surgery must be two-pole > 0.5 mm2
f.
The electrical line characteristics must comply with Chart 3
CHART 3
LINE VOLTAGE
230 V ± 15%
115 V ± 15%
MAXIMUM LINE VOLTAGE DROP
3%
3%
ELECTRICAL
LINE
MAXIMUM LINE APPARENT RESISTANCE
0.5 ohm
0.5 ohm
Prior to installing the X-ray system, it is advisable that all the electrical connections be in
place.
Timer electric connections
CAUTION
According to the relevant standard, the timer must be installed in a position allowing the
operator to always control the X-ray exposure.
On the timer installation wall, suitable conduits for the following electric cables must be
provided, pursuant to the electrical installation diagram (refer to §20 “ATTACHMENTS”):
a. Timer electrical cables
(refer to §5.9 “TIMER ELECTRIC CONNECTIONS”)
b. Cables connecting the timer and the tubehead
(refer to §5.9 “ TIMER ELECTRIC CONNECTIONS ”)
c.
Cables connecting the timer and the RX signal lamp/s for external use:
X-Mind LIGHT (OPTIONAL)
d. Cables connecting the timer and the remote control button:
X-Mind ECB (IF PROVIDED)
Tubehead electrical connections
On the wall plate installation wall, a suitable conduit for the cable connecting the timer
and tubehead must be provided
(refer to §5.5 “CONNECTION TO THE POWER SUPPLY TERMINAL BOARD”)
The “ X-MindDC X-ray system must be installed by professionally trained technicians,
who must be able to certify their work with a Statement of Compliance.
WARNING
Prior to installing the X-ray system, it is advisable to ensure that all needed requirements
have been met (refer to §4 “INSTALLATION SPECIFICATIONS”)
The components of the “ X-MindDC ” radiographic system are duly packed within a carton box,
as shown in the following sketch (Fig. 5):
D
DOCUMENTS: INSTRUCTION MANUAL, INSTALLATION & MAINTENANCE MANUAL, AND WARRANTY CARD
c X-MindDC TUBEHEAD PACKAGING
d X-MindDC TIMER + RX SIGNALING LAMP FOR EXTERNAL USE X-Mind LIGHT (OPTIONAL) + REMOTE
CONTROL BUTTON X-Mind ECB (OPTIONAL) PACKAGING
e PANTOGRAPH-TYPE ARM PACKAGING
f WALL PLATE + BRACKET PACKAGING
PLEASE NOTE
Prior to installation, duly check all components.
PLEASE NOTE
The carton and the polystyrene foam can be completely recycled and can be disposed of by
authorized recycling companies.
PLEASE NOTE
The original packaging should be stored in the event the equipment is returned for repairs.
5. INSTALLATION
5.1 UNPACKING
To fix the wall plate, DO NOT use plastic or rubber anchor screws.
For cement or solid/hollow brick walls, use metal anchor screws Ø12 (NOT included).
ASSEMBLY INSTRUCTIONS (Fig. 6)
1. Take the wall plate out of the packaging (refer to Fig. 5) and take the drilling template
c
2. Position the drilling templatec on the X-ray system installation wall, at the required
height (130 cm from the base is the suggested height)
3. Secure the template with adhesive tape
4. Check the holes for vertical and horizontal alignment, using a plumb line
5. Mark the wall plate anchor holes
6. Mark the holes for the electric cables connecting the timer to the tubehead
PLEASE NOTE
To prevent any flaking in the white coat and to control the center distances between the
holes, it is advisable to start drilling with a Ø7 bit, increasing this size gradually.
7. Drill the wall plate anchor holes
8. If required, drill the holes for the electric cables connecting the timer to the tubehead
9. Remove the template c and insert the suitable anchor screws, according to the wall’s
characteristics d
10. Unscrew the screw f and remove the plug e from the wall plate
11. Withdraw the sliding cover g
12. Apply the wall plate h to the wall and insert the screws i with the appropriate washers,
then tighten them alternately
13. Ensure that the wall plate is solidly attached to the wall
PLEASE NOTE
If the wall is not perfectly plumbed, insert a suitable shim between the wall and the wall
plate, in order to prevent warping.
5.2 ASSEMBLING THE WALL PLATE
The brackets come in the following lengths: 41 cm - 82,5 cm - 110 cm
The 82.5 cm and 110 cm brackets are provided with a stop key c (Fig. 7A and 7B) to prevent
the electrical cable from twisting.
PLEASE NOTE
Generally, the stop key is installed so that the equipment position at rest is on the right side,
viewed from the front of the wall plate (Fig. 7A).
Should the position at rest be on the left side, the stop key must be rotated by 180° (Fig. 7B).
5.3 ASSEMBLING THE BRACKET
Fig. 7A
Fig. 7B
c
c
WALL
After fixing the plate to the wall, assemble the bracket, using the following procedure (Fig. 8):
1. Take the bracket out of the packaging (refer to Fig. 5)
2. Insert the bracket pin d into the wall plate c (upwards)
3. Insert the supporting rest e
Fig. 8
PLEASE NOTE
Prevent all foreign matter (ground, dust, cement, etc. ) from settling on the pin seat.
The pin must slide freely in its seat. If required, thoroughly clean and lubricate with “molikote
D" lubricant.
PLEASE NOTE
Carefully check that the bracket is parallel to the floor, using a water level.
e
c
d
ASSEMBLY INSTRUCTIONS (Fig. 9)
1. Take out the pantograph-type arm from the packaging (refer to Fig. 5)
2. Remove the bracket plug c by unscrewing the set screw d
3. Insert the bracket guard wafer e
4. Insert the pantograph group cable g into the washer f and then the pantograph pin h
Fig. 9
5. If required, clean the pin and the washer and lubricate with “molikote D" lubricant
6. Insert the electric cable into the bracket housing i
7. Assemble the guard wafer
8. Insert the cable into the bracket and push it until it reaches the pin outlet near the supply
terminal board
5.4 ASSEMBLING THE PANTOGRAPH-TYPE ARM
c
d
e
f
g
h
i
For electrical safety, it is essential that the ground conductors be correctly connected.
WARNING
The length of the electrical cable and of the communication cable is suitable for 82.5 cm
brackets.
DOT NOT tamper with the terminals arrangement when 41 cm long brackets are used.
The excess cable must be stored in the housing provided.
WARNING
In case of electric power supply type PHASE – PHASE the higher value must be taken using a
tester and entered in L.
5.5 CONNECTION TO THE FEEDING TERMINAL BOARD
OPERATING INSTRUCTIONS (Fig. 10)
1. Remove the terminal board cover c by unscrewing the set screw d
Fig. 10
c
d
2. Proceed with the electrical connection as shown in Fig. 11
Fig. 11
TERMINAL BOARD CONNECTION DIAGRAM
c
BROWN
L (line)
d
YELLOW GREEN
T (ground)
e
BLUE
N (neutral)
f
BLACK
COMMUNICATIONS
g
RED
COMMUNICATIONS
h
YELLOW GREEN
T (ground)
i
YELLOW GREEN
T (ground)
THE COMMUNICATIONS CABLES ARE NOT POLARIZED
3. Connect the pantograph cable shield to the grounding potential h
4. Connect the wall plate to the grounding potential i
5. Clamp the cables with the cable clamps provided
6. Reassemble the terminal board cover (refer to c Fig. 10)
c
d
e
f
g
h i
IMPORTANT !
1. Remove the tubehead from the packaging (refer to Fig. 5)
2. Check that all the rating data match the power supply voltage
3. Remove both guards from the pantograph-type arm by loosening the appropriate screws
(Fig. 12)
Fig. 12
4. Using a bit, detach the front coupling device (Fig. 13A)
5. Remove both guards (Fig. 13B)
Fig. 13A
Fig. 13B
5.6 ASSEMBLING THE TUBEHEAD
6. Insert the tubehead pin into the pantograph head (Fig. 14A) and insert the
support (Fig. 14B)
Fig. 14A
Fig. 14B
7. Check that during insertion, the pin of the rotation prevention device is correctly
inserted into the seat located on the pantograph head (Fig. 15)
Fig. 15
Couple the pantograph and tubehead connectors and insert them into their seats (Fig. 16)
Fig. 16
The pantograph-type arm must be adjusted only when the tubehead is assembled .
WARNING
To prevent damage to the internal mechanism while making adjustments and executing the
balancing tests, the adjustment key must not be left in place.
WARNING
The key provided must be carefully stored.
PLEASE NOTE
To reach the “X” adjustment screw “ARM A” must be put in vertical position.
To reach the “Y” adjustment screw “ARM B” must be put in horizontal position.
The adjustment key provided can be inserted only under the above conditions (Fig. 17)
5.7 BALANCING THE PANTOGRAPH-TYPE ARM
ARM B
Y
ARM A
Fig. 17
X
Procedure to be followed to correctly balance the pantograph-type arm (refer to Fig. 17):
1.
“ARM A” BALANCING
PLEASE NOTE
The pantograph-type arm is supplied with “ARM A” already preloaded.
“ARM B” is supplied unloaded for safety reasons.
2.
“ARM B” BALANCING
For correct balancing, proceed as follows:
1. “ARM A” VERTICAL
2. “ARM B” HORIZONTAL
3. INSERT THE ADJUSTMENT KEY IN “Y”
4. LOAD THE SPRING WITH 22 TURNS
5. WITHDRAW THE KEY
3.
CHECKING THE BALANCING
To check the balancing, proceed as follows:
BRING “ARM B” TO THE VARIOUS POSITIONS
if it does not maintain the correct position:
1. BRING “ARM B” BACK TO THE HORIZONTAL POSITION
2. INSERT THE ADJUSTMENT KEY IN “Y”
3. ROTATE THE ADJUSTMENT KEY BY HALF TURN:
- - CLOCKWISE IF IT TENDS TO COME DOWN
- COUNTERCLOCKWISE IF IT TENDS TO GO UP
4. WITHDRAW THE KEY
PLEASE NOTE
Repeat the test and adjustment until “ARM B” is steady and stable in all positions,
even when “ARM A” is completely extended.
4.
READJUSTMENT OF “ARM A”
If it does not maintain the required position:
1.
BRING “ARM A” TO THE VERTICAL POSITION AGAIN
2.
INSERT ADJUSTMENT KEY IN “X”
3. ROTATE ADJUSTMENT KEY BY HALF TURN:
- CLOCKWISE IF IT TENDS TO COME DOWN
- COUNTERCLOCKWISE IF IT TENDS TO GO UP
4.
WITHDRAW THE KEY
PLEASE NOTE
Repeat the test and adjustment until “ARM A” is steady and stable in all positions, even
when “ARM B” is completely extended.
Upon completion of balancing:
1. Insert the movable finish (A) between the pantograph-type arm guard and the metal frame (Fig.
18)
Fig. 18
2. Insert the pins c of the guards d into their slots; then position them and ensure that the
movable finish e is coupled to the guards (Fig. 19)
Fig. 19
A
c
e
d
d
Check that the cable conduits are present in the timer installation wall; check the
compliance
of
the
power
supply
with
the
installation
specifications
(refer to §.4 “INSTALLATION SPECIFICATIONS”).
CAUTION
Check that the rating data match the power supply voltage.
ASSEMBLY INSTRUCTIONS (Fig. 20)
1. Take the timer out of the packaging (refer to Fig. 5)
2. Take out the drilling template c
3. Mark the anchor holes on the wall using the drilling template
4. Drill using a Ø3mm bit, then drill again with a Ø6 mm bit to prevent any flaking of the
white coat
5. Insert the three screw anchors provided into the holes d
6. Open the timer by unscrewing the three screws e
5.8 ASSEMBLING THE “X-Mind®DC” TIMER
Fig. 20
7. Withdraw the 26-pole connector from its seat to release both timer guards (Fig. 21)
Fig. 21
8. Insert the electrical feeding cables into the hole (Fig. 22)
Fig. 22
9. Insert the connection cables coming from the tubeheads into the rectangular slot
10. Insert the cables of the RX signaling lamp for external use (IF INSTALLED) (Fig. 22)
11. Insert the cables of the remote control button (IF INSTALLED) (Fig. 22)
12. Apply the timer base to the wall, matching the three anchor screws with the holes
13. Insert and tighten the screws (refer to Fig. 20 f) with the appropriate washers (refer to
Fig. 20 g)
into the screw anchors
PLEASE NOTE
If the wall is uneven, insert some shims to prevent the timer from buckling.
PLEASE NOTE
Do not contaminate the timer with dust or rubble from drilling.
26-POLE CONNECTOR
FEEDING CABLE INLET HOLE
INLET HOLE FOR THE
TUBEHEAD(S)
RX SIGNALING LAMP(S) FOR
EXTERNAL USE CABLE
REMOTE CONTROL BUTTON(S)
Before proceeding with connections, the power supply must be cut off.
CAUTION
If the equipment is mounted on metal walls, these must be connected to the grounding
circuit.
WARNING
While performing the connection, always comply with the PHASE – NEUTRAL polarity.
WARNING
While stripping the cables, be careful that small copper wires do not fall on the printed circuit
and cause short circuits or malfunctioning.
OPERATING INSTRUCTIONS (Fig. 23) (refer to §19 “ATTACHMENTS”)
1.
Connect the power supply cable to the terminal board h
2.
Insert the three mains cables into the rack
3.
Fix them with the cable clamp i
4.
Connect the cables coming from tubehead n° 1 to the terminals XRAY1
5.
Connect the YELLOW-GREEN grounding cable to the “equipotential metal plate”
j
6.
Connect the cables coming from tubehead n° 2 to terminals XRAY2
7.
Connect the YELLOW-GREEN grounding cable to the “equipotential metal plate”
j
8.
Clamp the cables in the cable clamp k
9.
Connect the RX signaling lamps for external use (IF INSTALLED) (refer to §19 “ATTACHMENTS”)
10.
Connect the remote control buttons (IF INSTALLED) (refer to §19 “ATTACHMENTS”)
11.
Check the configuration on the dip switches (refer to §7 “SYSTEM CONFIGURATION”)
12.
Reconnect the 26-pole connector
13.
Close the timer with the three screws
14.
Mount the wall plate sliding cover and plug (refer to Fig. 6 eg)
15.
Reconnect the power supply
5.9 TIMER ELECTRIC CONNECTION
TUBEHEAD N° 1 CONTROL BUTTON and/or REMOTE CONTROL BUTTON
2
2
TUBEHEAD N° 2 CONTROL BUTTON and/or REMOTE CONTROL BUTTON
3A
3A
TUBEHEAD N° 1 RX SIGNALING LAMP FOR EXTERNAL USE
3B
3B
TUBEHEAD N° 2 RX SIGNALING LAMP FOR EXTERNAL USE
The electronic card terminal boards control the following functions:
4A
4A
RS232 COMMUNICATION FOR XRAY1
4B
4B
RS232 COMMUNICATION FOR XRAY2
5
5
TUBEHEAD N° 1 POWER SUPPLY
6
6
TUBEHEAD N° 2 POWER SUPPLY
7
7
TIMER POWER SUPPLY
h
i
j
k
1 2 3A-3B 4A-4B 5 6 7
The control panel is the user's interface, from which the operator can:
use the keys
 to select the operating parameters
 to change the exposure times displayed
 to store the new customized exposure times
use the “green LED” light indicators
 to check the timer operating conditions
6. CONTROL PANEL
3344
XX--MMiinndd
DDCC
MAIN SWITCH
CONTROL
BUTTON
X-RAY
KEY
DISPLAY
KEY
TO DECREASE
EXPOSURE
TIME
KEY
TO INCREASE
EXPOSURE
TIME
TUBEHEAD
TYPE
INDICATOR
X-RAY DISTANCE
INDICATOR
TUBEHEAD
SELECTION
X-RAY VOLTAGE
INDICATOR
MEMORY
SELECTION OF
PATIENT’S PHYSIQUE
TYPE
UPPER MAXILLARY
TEETH
LOWER MANDIBLE
TEETH
X-RAY CURRENT
INDICATOR
X-RAY
OUTPUT SIGNAL
OCCLUSAL EXAM
BITEWING EXAM
PAUSE INDICATOR
MALFUNCTION
INDICATOR
DIGITAL
X-RAY TECHNIQUE
CONVENTIONAL
X-RAY TECHNIQUE
KEY SWITCH
The “ X-Mind®DC ” X-ray system is factory configured for “standard mode” operation which
determines:
No. 2 “ X-Mind®DC ” tube heads
Ö
pressing the RX button
on the control
panel causes the LED 1 button to light up,
pressing the button again causes the LED 2 to
light up
** IF THE LONG 12” (31 cm) CONE IS USED **
An X-ray distance
SSD = 31cm
with the long 12” cone
Ö on the control panel
LED 12” is lit
** IF THE SHORT 8” (20 cm) CONE IS USED **
An X-ray distance
SSD = 20cm
with the short 8” cone
Ö on the control panel
8” LED is lit
with film type “D”
Ö
on the control panel
“D” LED is lit
no. 1 CONTROL BUTTON
to make the exposure
Ö
The timer includes a key with an extension
cable
The installed tube head has the following features:
work with
direct current at constant potential
the “DC” LED
on the control panel must be lit
work with
an X-ray voltage equal to 70kVp
the “70 kVp” LED
on the control panel must be lit
work with
an X-ray current equal to 8mA
the “8 mA” LED
on the control panel must be lit
The above configuration depends on the position of 8 mini-switches (dip-switch
) on
the timer electronic card:
DIP
SWICH
PARAMETER
ON
OFF
1
TUBEHEAD No. 1
INSTALLED
NOT INSTALLED
2
TYPE OF TUBEHEAD No. 1
dc
DIRECT CURRENT
AT CONSTANT POTENTIAL
ac
ALTERNATING CURRENT
SINGLE-PHASE
3
TUBEHEAD No. 2
INSTALLED
NOT INSTALLED
4
TYPE OF TUBEHEAD No. 2
dc
DIRECT CURRENT
AT CONSTANT POTENTIAL
ac
ALTERNATING CURRENT
SINGLE-PHASE
5
CONTROL BUTTON No. 2
INSTALLED
NOT INSTALLED
6
CONE
LONG (12”)
SHORT (8”)
7
NOT AVAILABLE
NOT AVAILABLE
8
NOT AVAILABLE
NOT AVAILABLE
7. SYSTEM CONFIGURATION
** IF THE LONG 12” (31 cm) CONE IS USED **
ON
OFF
1
■
tubehead no. 1
2
■
“dc” type tubehead
3
■
tube head .no. 2
4
■
“dc” type tubehead
5
■
control button no. 2
6
■
12” cone (31cm)
7
■
not available
8
■
not available
** IF THE SHORT 8” (20 cm) CONE IS USED **
ON
OFF
1
■
no. 1tubehead
2
■
“dc type” tubehead
3
■
no. 2tubehead
4
■
“dc” type tubehead
5
■
no. 2 control button
6
■
8” cone (20 cm)
7
■
not available
8
■
not available
The configuration may be changed if:
POSSIBLE MODIFICATION
HOW TO CARRY OUT THE MODIFICATIONS
− type “E” and “F” films are used
− a digital system is used
− use of a voltage equal to 60 kV
− use of a current equal to 4 mA
Ö
refer to USER’S MANUAL
§ 6 “OPERATING INSTRUCTIONS”
− the short 8” (20 cm) cone is used
− the long 12” (31 cm) cone is used
− a single tubehead is used
− use of a tubehead with “ac” technology
− no. 2 CONTROL BUTTON is used
Ö
by changing the dip switch position
THIS ,MAY ONLY BE DONE BY THE INSTALLER
This must be done by the installer.
WARNING
To make the changes operative, turn off the timer and then turn it on again.
To change the configuration, the timer’s dip switch positions (
)must be changed:
DIP
SWICH
PARAMETER
ON
OFF
1
TUBEHEAD No. 1
INSTALLED
NOT INSTALLED
2
TYPE OF TUBEHEAD No. 1
dc
DIRECT CURRENT
AT CONSTANT POTENTIAL
ac
ALTERNATING CURRENT
SINGLE-PHASE
3
TUBEHEAD No. 2
INSTALLED
NOT INSTALLED
4
TYPE OF TUBEHEAD No. 2
dc
DIRECT CURRENT
AT CONSTANT POTENTIAL
ac
ALTERNATING CURRENT
SINGLE-PHASE
5
No. 2 CONTROL BUTTON
INSTALLED
NOT INSTALLED
6
CONE
LONG (12”)
SHORT (8”)
7
NOT AVAILABLE
NOT AVAILABLE
8
NOT AVAILABLE
NOT AVAILABLE
8. CHANGING THE CONFIGURATION
CHANGING THE NUMBER OF TUBEHEADS INSTALLED
To change the number of tubeheads installed, move dip switch no. 1 or no. 3
a. If the tubehead is connected to the X-RAY 1 terminal board, put dipswitch no. 1 in the
ON position; if not, put it in the OFF position
b. If the tubehead is connected to the X-RAY 2 terminal board, put dip switch no. 3 in the
ON position; if not, put it in the OFF position
2.
CHANGING THE TYPE OF TUBE HEAD
The following timer connections may be made:
no. 2 TUBEHEADS USING ALTERNATING CURRENT “ac”: “X-Mind® AC”
or
no. 2 TUBEHEADS USING DIRECT CURRENT “dc”: “X-Mind® DC”
or
no. 1 TUBEHEAD USING ALTERNATING CURRENT “ac” + no. 1 TUBE HEAD USING DIRECT
CURRENT “dc”: “X-Mind® AC” + “X-Mind® DC”
To change the type of tubehead, move dip switch no. 2 or no. 4
a. If tubehead RX1 is “ac”, put dip-switch no. 2 in the OFF position, otherwise put it in the
ON position
b. If tub head RX2 is “ac”, put dip-switch no. 4 in the OFF position, otherwise put it in the
ON position
3.
CHANGING THE CONE
To replace the long 12” cone with the short 8” cone, move dip switch no. 6 to the OFF
position.
Check that the “8” LED lights up on the control panel.
PLEASE NOTE
After modification, the set exposure times are automatically changed.
To replace the short 8” cone with the long 12” cone,
move dip switch no. 6 to the ON position.
Check that the “12” LED lights up on the control panel.
PLEASE NOTE
After modification, the set exposure times are automatically changed.
4.
CHANGING THE REMOTE BUTTONS SETTINGS
To change the control buttons settings, move dip switch no. 5 to the ON position.
PLEASE NOTE
After modification, each tube head is controlled by its button.
When all connections are completed, the installer must check the system’s electrical and
operating safety.
For further information, see the USER'S MANUAL.
TURN ON THE TIMER to power up the X-ray system
a.
Bring the "MAIN SWITCH" located on the upper part of the timer to the
“I” position (ON)
b. Bring the “’KEY SWITCH” to the “I” position (ON)
 the green light turns on indicating that the system is powered up
 the LEDs of the set parameters automatically light up
 the exposure time is shown on the display
c. THE X-RAY SYSTEM IS NOW READY FOR USE
WARNING
If an error is detected when the system is turned on, the anomaly is indicated as follows:
• an intermittent beep sounds
• the MALFUNCTION INDICATOR LED
blinks intermittently
The error code (E ….) appears on the display
(refer to § 13 “ERROR MESSAGES”)
• All control panel functions are disabled
In this event, turn off the timer and then turn it back on.
If the error should recur, call the “Support Department”.
PLEASE NOTE
The exposure time and parameters displayed are the last which to be set before the timer was
turned off.
PLEASE NOTE
If installed, the RX signal lamp for external use, corresponding to the selected tube head, turns
on.
PLEASE NOTE
If the timer remains inactive for a few minutes, it switches to the stand-by mode.
Press any control panel key to restore it to operating mode.
9. START UP
CHECKING THE CONFIGURATION
On the control panel, check that all LEDs corresponding to the required configuration are
lit; if not, change them.
(refer to § 7 “SYSTEM CONFIGURATION” and § 8 “CHANGING THE CONFIGURATION” )
2.
CHECKING THE TIMER OPERATION
a. Check the correct operation of the control panel by selecting different exposure
times
b. Check the time on the display
c. Check that, when changing the selected tube head, the corresponding RX signal
lamp for external use (IF INSTALLED) turns on
3.
CHECKING THE EXPOSURE
a. Set an exposure time of 1 sec
b. Take the timer’s "CONTROL BUTTON"
c. Using the extension cable of the “CONTROL BUTTON”, maintain a safe distance of
at least 2 meters from the tube head, so as to be able to constantly check the X-
ray exposure
d.
Press the “X-RAY”
key and keep it depressed until the acoustic signal (beep)
stops, and the yellow X-RAY OUTPUT SIGNAL
LED turns off
e.
At the end of the exposure, the green PAUSE INDICATOR
LED blinks
intermittently
f. On the display, check the actual exposure time
(refer to § 11 4.“EXPOSURE TIME (sec)” )
4.
CHECKING THE TUBEHEAD OPERATION
Take several exposures on the tubehead installed and check that:
a. There are no errors
b. The LED of the selected tube head is lit
c. The "CONTROL BUTTON" LED is lit for the entire duration of the acoustic signal
10. CHECKING THE INSTALLATION
CHECKING THE POWER ABSORBED BY THE X-RAY SYSTEM
To check the power absorbed by the radiographic system a tester must be used, in the
mode ampmeter in “AC”
a. Connect the instrument to the power supply line
(refer to § 5.9 “TIMER ELECTRIC CONNECTION”)
b. Set an exposure time of approx. 3 sec on the timer
c. Take an exposure and read the current value on the display
PLEASE NOTE
The X-ray system complies with the requirements when
the absorbed current is ≤ 8A with 230V line voltage (±15%)
the absorbed current is ≤ 12.5A with 115V line voltage (±15%)
Otherwise, check the electrical system or call the “Support Department”.
6.
CHECKING THE ELECTRIC SYSTEM
To check the electric system, a tester must be used, in the mode voltmeter in “AC”
a. Connect the instrument to the L and N terminals on the timer
b. Measure the line voltage (V0)
c. Connect the instrument to the L and N terminals of the wall plate terminal board
(refer to § 5.5 “CONNECTION TO THE FEEDING TERMINAL BOARD”)
d. Set an exposure time of approx. 3 sec on the timer
e. Proceed with the exposure
f. Measure the line voltage during exposure (V1)
PLEASE NOTE
The electrical system complies with the requirements when the result of the
formula
(V0 – V1 ) / V0 is ≤ 0.03 (3%)
Otherwise, the electric system must be adjusted.
(refer to § 4 “INSTALLATION SPECIFICATIONS”
RADIOGRAPHIC VOLTAGE (kVp)
The radiographic high voltage is measured using a calibrated "non-invasive" instrument
with an initial delay of 50 m. sec.
Line Voltage = nominal V ±15%
Max Voltage Drop = 3 %
kVp = 60 - 70
mA = 4 - 8
SET TECHNICAL
VALUES
Set Exposure Time = 3.2 sec
The radiographic VOLTAGE is 60 kVp - 70 kVp ±10 %
2.
RADIOGRAPHIC CURRENT (mA)
The radiographic current is measured by connecting a milliampmeter inside the tube head
Line Voltage = Nominal V ±15%
Max Voltage Drop = 3 %
kVp = 60 - 70
mA = 4 - 8
SET TECHNICAL
VALUES
Set Exposure Time = 3.2 sec
The radiographic CURRENT is 4mA - 8mA ±10 %
3.
DOSE (mGy)
The dose in the air is measured with a "non-invasive" instrument, by positioning the
detector at an SSD (source-skin distance) of 31 cm (12”) or 20 cm (8”)
Line Voltage = Nominal V ±15 %
Max Voltage Drop = 3 %
kVp = 60 - 70
mA = 4 - 8
SET TECHNICAL
VALUES
Set Exposure Time = 1 sec
DOSE in the air =
SSD = SOURCE-SKIN DISTANCE = 31 CM (12”)
SSD = SOURCE-SKIN DISTANCE = 20 CM (8”)
70 kVp – 8 mA = 5.5 mGy/s ±30 %
70 kVp – 8 mA = 12 mGy/s ±30 %
60 kVp – 8 mA = 4.5 mGy/s ±30 %
60 kVp – 8 mA = 9 mGy/s ±30 %
70 kVp – 4 mA = 3 mGy/s ±30 %
70 kVp – 4 mA = 6 mGy/s ±30 %
60 kVp – 4 mA = 2.2 mGy/s ±30 %
60 kVp – 4 mA = 4.5 mGy/s ±30 %
4.
EXPOSURE TIME (sec)
The exposure time is measured with a “non-invasive” instrument.
Line Voltage = Nominal V ±15 %
Max Voltage Drop = 3 %
kVp = 60 - 70
mA = 4 - 8
SET TECHNICAL
VALUES
Set Exposure Time = 3.2 sec
The EXPOSURE TIME measured is 3.2 sec ±10 %
11. CHECKING THE EXPOSURE FACTORS
The “X-Mind DC” X-ray system allows the installer to set or display some of the timer’s functional
parameters:
1.
To set the parameters, the installer must:
a. Turn off the timer
b. Simultaneously press and keep the following keys depressed:
(45) LOWER MANDIBLE PREMOLAR + (43) LOWER MANDIBLE CANINE
c. Turn on the timer
d. When the message "inst” is displayed, the installer can set the
minimum exposure time:
1. select the tube head
2. turn off the timer
3. turn on the timer, by keeping the
key depressed
4. the message “SEC” is displayed for approx 1 sec
5. the current value of the lower set limit is displayed
6. to change the value, press the keys
and
7. to confirm, press the
key
e. To exit this mode, turn the timer off and then on again
2.
To display the parameters, the installer must:
a. Simultaneously press and keep depressed the
(17) UPPER MAXILLARY MOLAR + (47) LOWER MAXILLARY MOLAR keys
b. Press the key associated with the desired parameter
KEY
DISPLAYED PARAMETER
Example
U.M.
BITEWING
ANT
X-RAY SYSTEM
NOMINAL VOLTAGE
230
Volt
BITEWING
POST
LINE VOLTAGE
227
Volt
UPPER
INCISOR
MAXIMUM LINE VOLTAGE
VALUE DETECTED
238
Volt
LOWER
INCISOR
MINIMUM LINE VOLTAGE
VALUE DETECTED
215
Volt
OCCLUSAL
SOFTWARE VERSION
2.3
12. DIAGNOSTIC
During this operation there is X-ray output.
All the safety measures relevant to radioprotection must be observed.
If necessary, “ X-Mind DC ” can be calibrated.
The relevant procedure is as follows:
a. Switch off the timer
b. Switch on the timer and keep the mA
button depressed
c. The display shows the “tube” message
d. Select X-RAY with the
button to calibrate
e. Take the exposure:
Press and keep the “RX”
button depressed until the beeping stops
stops the yellow
“X-RAY OUTPUT SIGNAL” LED goes off
f. Once the exposure has been taken, if the display does not show “ERRORS”, the calibration
has been successfully completed, otherwise (refer to §14 “ERROR MESSAGES”)
13. CALIBRATION
The following chart contains a list of error messages that may appear while the “X-Mind® DC” X-
ray system is operating.
The chart also includes the causes of the error messages and what to do to resolve them.
Error
Message
Cause
Solution
RX1 TUBE HEAD
IS NOT CONNECTED OR IS OUT OF ORDER
CALL THE “SUPPORT DEPARTMENT”
RX2 TUBE HEAD
IS NOT CONNECTED OR IS OUT OF ORDER
CALL THE “SUPPORT DEPARTMENT”
CORRUPTED EEPROM DATA
CALL THE “SUPPORT DEPARTMENT”
EEPROM DATA NOT PROPERLY SAVED
CALL THE “SUPPORT DEPARTMENT”
LINE VOLTAGE VALUE NOT WITHIN THE SET
LIMITS
CALL THE “SUPPORT DEPARTMENT”
LINE VOLTAGE VALUE NOT INCLUDED WITHIN
THE ±15 % NOMINAL VALUE
CALL THE “SUPPORT DEPARTMENT”
THE “X-RAY” KEY
ALWAYS SEEMS TO BE DEPRESSED
ENSURE THAT IT IS NOT JAMMED
ANOMALY ON THE CONTROL PANEL
CALL THE “SUPPORT DEPARTMENT”
THE EXPOSURE
HAS BEEN PREMATURELY INTERRUPTED
KEEP THE “X-RAY KEY”
DEPRESSED TILL THE END OF THE EXPOSURE
ANOMALY IN THE TRIAC/RELAY
CALL THE “SUPPORT DEPARTMENT”
ANOMALY IN THE ELECTRONIC CIRCUIT
CALL THE “SUPPORT DEPARTMENT”
ANOMALY IN THE CONTROL CIRCUIT
CALL THE “SUPPORT DEPARTMENT”
INCORRECT DIP SWITCH CONFIGURATION
CALL THE “SUPPORT DEPARTMENT”
THE “CONTROL BUTTON” DOES NOT
CORRESPOND TO THE SELECTED TUBE HEAD
USE THE “CONTROL BUTTON” THAT CORRESPONDS
TO THE SELECTED TUBE HEAD
THE TUBEHEAD
DOES NOT WORK PROPERLY
CALL THE “SUPPORT DEPARTMENT”
THE TUBEHEAD
IS NOT IN THE CORRECT MODE
CALL THE “SUPPORT DEPARTMENT”
14. ERROR MESSAGES
THE TUB HEAD
HAS NOT COMPLETED THE EXPOSURE
REPEAT THE EXPOSURE OR
CALL THE “SUPPORT DEPARTMENT”
FREQUENCY
OR REGULATION PROBLEM
CALL THE “SUPPORT DEPARTMENT”
THE TUBEHEAD
IS NOT CALIBRATED
CALIBRATE OR
CALL THE “SUPPORT DEPARTMENT”
EEPROM DATA NOT PROPERLY SAVED
CALL THE “SUPPORT DEPARTMENT”
CORRUPTED EEPROM DATA
CALL THE “SUPPORT DEPARTMENT”
OVERVOLTAGE ERROR
CALL THE “SUPPORT DEPARTMENT”
ANODE VOLTAGE OUT OF TOLERANCE
CALL THE “SUPPORT DEPARTMENT”
ANODE CURRENT OUT OF TOLERANCE
CALL THE “SUPPORT DEPARTMENT”
CONTROL CONNECTOR
CALL THE “SUPPORT DEPARTMENT”
REFERENCE VOLTAGE PROBLEM
CALL THE “SUPPORT DEPARTMENT”
MAJOR ERROR
ALL X-RAY SYSTEM FUNCTIONS ARE DISABLED
CALL THE “SUPPORT DEPARTMENT”
The timer’s electronic equipment is protected by 4 fuses located on the electronic card.
To replace them, proceed as follows (Fig. 25):
1.
Cut off the power supply
2.
Temporarily remove the timer’s guard c by removing the adjusting d
3.
Determine the fuse to be replaced (Fig. 26)
4.
Remove the plastic protector (Fig. 27)
5.
Remove the fuse
6.
Replace it with a fuse of same type
POWER VOLTAGE OF THE X-RAY SYSTEM
230 V ± 15 %
115 V ± 15 %
Internal protection fuses (5x20) quick
8AF – 250V
12.5AF – 250V
7.
Insert the protection
Fig. 26
Fig. 27
8.
Close the timer’s guard
9.
Restore the power
15. FUSE REPLACEMENT
To guarantee the X-ray system’s safety, the following checks must be carried out:
CAUTION
The owner is responsible for arranging and respecting the maintenance schedule.
The X-ray system must undergo maintenance to check for correct operation every 12
months.
1. Cut the power
2. Release the spring of the “ARM B” of the pantograph type arm using the key provided
(refer to § 5.7 “BALANCING THE PANTOGRAPH-TYPE ARM”)
3. Remove the tube head
(refer to § 5.6 “ASSEMBLING THE TUBE HEAD”)
4. Remove the wall plate guard
(refer to § 5.2 “ASSEMBLING THE WALL PLATE”)
5. Remove the terminal board cover and disconnect the pantograph-type arm cable
6. Remove the bracket plug and remove the guard slab
(refer to § 5.4 “ASSEMBLING THE PANTOGRAPH-TYPE ARM")
7. Remove the pantograph-type arm and its cable from the bracket
8. Remove the bracket from the wall plate
(refer to § 5.3 “ASSEMBLING THE BRACKET”)
9. Check the wall plate’s vertical alignment – adjust if required
10. Check the six wall plate’s adjusting screws – tighten if required
11. Clean the old lubricating grease from the bracket shaft
12. Should the bracket shaft be damaged, install a new bracket
13. Clean the old lubricating grease from the bracket bush
14. If the bracket bush is damaged, install a new bracket
15. Grease the bracket shaft (use “molikote D” lubricant)
16. Lubricate the bracket bush with lubrication grease (use “molikote D” lubricant)
17. Install the bracket in the wall plate
(refer to § 5.3 “ASSEMBLING THE BRACKET")
18. Check the pantograph-type arm cable
If it is be damaged, send the pantograph-type arm to the manufacturer for repairs
19. Check the pantograph-type arm guards
20. Replace the damaged guards
21. Remove all old grease of the shaft
(refer to §. 5.4 “ASSEMBLING THE PANTOGRAPH-TYPE ARM”)
22. If the shaft is damaged, send the pantograph-type arm to the manufacturer for repairs
23. Lubricate the pantograph-type arm shaft with lubricant (use “molikote D” lubricant) and
reposition it in the bracket
24. Replace the pantograph-type arm cable in the bracket and the wall plate, connect it to the
terminal board and replace the terminal board cover
(refer to § 5.5 “CONNECTION TO THE TERMINAL BOARD”)
25. Position the guard slab in the bracket
26. Position the bracket plug
27. Position the plate guard
28. Check the tubehead’s electrical contact
29. If damaged, send the tube head to the manufacturer for repairs
30. Clean the old grease from the tube head assembly shaft
31. Lubricate the tubehead’s assembly shaft with a thin layer of lubricant(use “molikote D”
lubricant)
32. Reposition the tubehead
33. Load the spring of the pantograph-type arm “ARM B” using the key provided
(refer to § 5.7 “BALANCING THE PANTOGRAPH-TYPE ARM”)
34. Power up and check that the X-ray system is operating correctly
16. MAINTENANCE
Use a soft cloth with water and soap to clean the outer surfaces.
The spacer cone may be cleaned with cotton wool soaked in surgical alcohol.
In case of a malfunction, send the defective part to (USING THE ORIGINAL PACKAGING):
Satelec S.A.S.
Z.I. du Phare
B.P. 216
33708 MERIGNAC cedex - France
Tel. +33 (0) 556 34 06 07
Fax +33 (0) 556 34 92 92
E-mail: [email protected]
In the event of disposal, the components and packaging must be disposed of in an
environmentally correct manner.
In particular, the dielectric oil as well as the shielding lead must be disposed by means of
authorized companies specialized in the disposal of waste material.
17. CLEANING THE OUTER SURFACES
18. IF A REPAIR BECOMES NECESSARY
19. DISPOSAL
The manufacturer undertakes to supply, upon request, drawings, circuit diagrams, component
parts lists, instructions or other information needed by qualified technical personnel to perform
repairs on those parts of the “ X-Mind® DC ” radiographic system which may be repaired.
20. ACCESSORIES
SPARE PARTS TUBE HEAD X-MIND® DC
CODE
Q.TY PCS
1
DC XG (1) COPERTURA FORCELLA
29700979
8
2
AC/DC XG (2) SET GHIERE FORCELLA
29700980
5
3
AC/DC XG (3) FORCELLA
29700777
3
4
AC/DC XG (4) LIMITATORE DI ROTAZIONE
29700750
1
5
DC XG (5) GUARNIZIONE
27500959
1
6
DC XG (6) SCHEDE ARIA
29700981
9
7
DC XG (7) COPERTURA SCHEDA ARIA
29700982
11
8
DC XG (8) CAVO TESTA
39700385
1
A
CONO LUNGO 31cm (12”) Ø60
29700397
1
B
CONO CORTO 20cm (8”) Ø60
29700396
1
C
CONO QUADRO 31cm (12”) 44x35
29700615
1
SPARE PARTS X-MIND® LIGHT
CODE
Q.TY PCS
1
XG LIGHT (1) GUSCIO INF & SUP
29700977
4
2
XG LIGHT (2) PORTALAMPADA
33400372
1
3
XG LIGHT (3) LAMPADINA 230V 15W
33300352
1
3
XG LIGHT (3) LAMPADINA 115V 15W
33300390
1
4
XG LIGHT (4) SCHEDA 230V
39200354
3
4
XG LIGHT (4) SCHEDA 115V
39200353
3
SPARE PARTS X-MIND® ECB
CODE
Q.TY PCS
1
XG ECB (1) GUSCIO INF & SUP
29700978
4
2
SCHEDA COMANDO CON PULSANTE E LED
39700303
1
3
XG ECB (3) INTERRUTTORE A CHIAVE
35100298
1
SPARE PARTS TIMER X-MIND®
CODE
Q.TY PCS
1
AC/DC XG TIMER (1) DIMA
24200883
1
2
AC/DC XG TIMER (2) SCHEDA CONTROL
39200306
7
2
AC/DC XG TIMER (2) SCHEDA CONTROL US
39200307
7
3
AC/DC XG TIMER (3) CAVO SPIR
32000325
1
4
AC/DC XG TIMER (4) CAVO SPIR DOPPIO
32000359
1
5
AC/DC XG TIMER (5) CASSA PULSANTE
29700974
4
6
SCHEDA COMANDO CON PULSANTE E LED
39700303
1
7
AC/DC XG TIMER (7) SCHEDA LED
39200313
7
8
AC/DC XG TIMER (8) TASTIERA DEG
39200308
1
8
AC/DC XG TIMER (8) TASTIERA SAT
39200309
1
9
AC/DC XG TIMER (9) CASSA
29700975
1
10
AC/DC XG TIMER (10) COPERCHIO
29700976
4
11
AC/DC XG TIMER (11) BLOCK CHIAVE
39700305
2
12
DC XG TIMER (12) FILTRO
36200382
3
13
INTERRUTTORE BIPOLARE VERDE
35000044
1
14
PROTEZIONE INTERRUTTORE BIPOLARE
33500228
1
SPARE PARTS PANTOGRAPH TYPE ARM X-MIND®
CODE
Q.TY PCS
1
AC XG B (1) CAVO + CONNETTORE 3 POLI
39700384
1
2
DC XG B (2) CAVO + CONNETTORE 5 POLI
39700381
1
3
AC/DC XG B (3) GUSCI LATO TESTA
29700990
9
4
BRACCIO PANTOGRAFO TIRANTE ANTERIORE
29700460
4
5
BRACCIO PANTOGRAFO PORTATIRANTE
29700458
6
6
AC/DC XG B (6) TAPPI (set)
29700991
7
7
AC/DC XG B (7) COPERTURA EST
27500757
1
8
AC/DC XG B (8) COPERTURA INT
27500758
1
9
AC/DC XG B (9) GUSCI CENTRALI
29700993
8
10
BRACCIO PANTOGRAFO CHIAVE CARICAMENTO
22000543
1
11
AC/DC XG B (11) COPERTURA INT
27500758
1
12
AC/DC B (12) COPERTURA EST
27500757
1
13
AC/DC XG B (13) TIRANTE POST
29700992
7
14
BRACCIO PANTOGRAFO LUNETTA
22000177
1
15
AC/DC B (15) GUSCI LATO MENSOLA
29700994
8
SPARE PARTS WALL PLATE 350
CODE
Q.TY PCS
1
PIASTRA (1) COPERTURA
29700985
5
1
AC/DC XG PIASTRA (1) COPERTURA SAT
29700986
5
2
PIASTRA LUNETTA
22000176
1
3
PIASTRA (3) DIMA
24200984
1
PIASTRA MURO 350 DEG
29700383
COMPLETO
AC/DC XG PIASTRA MURO 350 SAT
29700987
COMPLETO
SPARE PARTS WALL PLATE 350 TOP
CODE
Q.TY PCS
1
PIASTRA (1) COPERTURA
29700985
5
1
AC/DC XG PIASTRA (1) COPERTURA SAT
29700986
5
2
PIASTRA LUNETTA
22000176
1
3
PIASTRA (3) DIMA TOP
24201002
1
PIASTRA MURO 350 TOP DEG
29700988
COMPLETO
AC/DC XG PIASTRA MURO 350 TOP SAT
29700989
COMPLETO
SPARE PARTS BRACKET
CODE
Q.TY PCS
MENSOLA 400
29700380
COMPLETO
MENSOLA 800
29700378
COMPLETO
MENSOLA 1100
29700379
COMPLETO
REF
SPARE PARTS CEILING BRACKET
CODE
Q.TY PCS
MENSOLA PLAFONE
29700382
COMPLETO
REF
SPARE PARTS UNIT BRACKET
CODE
Q.TY PCS
MENSOLA RIUNITO
29700381
COMPLETO
REF
SPARE PARTS TOP MOUNT BRACKET
CODE
Q.TY PCS
MENSOLA 400 TOP MOUNT
29700620
COMPLETO
MENSOLA 800 TOP MOUNT
29700621
COMPLETO
MENSOLA 1100 TOP MOUNT
29700622
COMPLETO
WISSELSTUKKEN RICHTARM X-MIND®
CODE
AANTAL
1
AC XG B (1) CAVO + CONNETTORE 3 POLI
39700384
1
2
DC XG B (2) CAVO + CONNETTORE 5 POLI
39700381
1
3
AC/DC XG B (3) GUSCI LATO TESTA
29700990
9
4
BRACCIO PANTOGRAFO TIRANTE ANTERIORE
29700460
4
5
BRACCIO PANTOGRAFO PORTATIRANTE
29700458
6
6
AC/DC XG B (6) TAPPI (set)
29700991
7
7
AC/DC XG B (7) COPERTURA EST
27500757
1
8
AC/DC XG B (8) COPERTURA INT
27500758
1
9
AC/DC XG B (9) GUSCI CENTRALI
29700993
8
10
BRACCIO PANTOGRAFO CHIAVE CARICAMENTO
22000543
1
11
AC/DC XG B (11) COPERTURA INT
27500758
1
12
AC/DC B (12) COPERTURA EST
27500757
1
13
AC/DC XG B (13) TIRANTE POST
29700992
7
14
BRACCIO PANTOGRAFO LUNETTA
22000177
1
15
AC/DC B (15) GUSCI LATO MENSOLA
29700994
8
WISSELSTUKKEN MUURPLAAT 350
CODE
AANTAL
1
PIASTRA (1) COPERTURA
29700985
5
1
AC/DC XG PIASTRA (1) COPERTURA SAT
29700986
5
2
PIASTRA LUNETTA
22000176
1
3
PIASTRA (3) DIMA
24200984
1
PIASTRA MURO 350 DEG
29700383
COMPLETO
AC/DC XG PIASTRA MURO 350 SAT
29700987
COMPLETO
WISSELSTUKKEN MUURPLAAT 350 TOP
CODE
AANTAL
1
PIASTRA (1) COPERTURA
29700985
5
1
AC/DC XG PIASTRA (1) COPERTURA SAT
29700986
5
2
PIASTRA LUNETTA
22000176
1
3
PIASTRA (3) DIMA TOP
24201002
1
PIASTRA MURO 350 TOP DEG
29700988
COMPLETO
AC/DC XG PIASTRA MURO 350 TOP SAT
29700989
COMPLETO
MENSOLA 400 TOP MOUNT
29700620
COMPLETO
MENSOLA 800 TOP MOUNT
29700621
COMPLETO
MENSOLA 1100 TOP MOUNT
29700622
COMPLETO
I 03106 • P2 - Version 2
17 av Gustave Eiffel • BP 216 • 33708 MERIGNAC cedex • France • Tel. +33 (0) 556 34 06 07 • Fax. +33 (0) 556 34 92 92
E.mail : [email protected] • Web : acteongroup.com
France
SATELEC ACTEON GROUP
17 av. Gustave Eiffel - B.P. 216 - F-33708 MERIGNAC cedex
Tel. +33 (0) 556 34 06 07
Fax. +33 (0) 556 34 92 92
e-mail : [email protected]
U.S.A.
ACTEON NORTH AMERICA
130 Gaither Drive, Suite 100,
MOUNT LAUREL NJ 08034 - U.S.A.
Tel. +1 856 222 9988
Fax. +1 856 222 4726
e-mail : [email protected]
GERMANY
ACTEON GERMANY
Industriestrasse 9 - D-40822 METTMANN
Tel. +49 21 04 95 65 10
Fax. +49 21 04 95 65 11
e-mail : [email protected]
SPAIN
ACTEON IBERICA
Avda Principal n°11 H - Poligono Industrial Can Clapers
E-08181 SENTMENAT (BARCELONA)
Tel. +34 93 715 33 66
Fax. +34 93 715 32 29
e-mail : [email protected]
U.K.
SATELEC (UK) LIMITED
Unit 1B - Steel Close - Eaton Cocon, St Neots
GB-CAMBS PE19 8TT
Tel. +44 1480 477 307
Fax. +44 1480 477 381
e-mail : [email protected]
MIDDLE EAST
ACTEON MIDDLE EAST
Numan Center - 1st Floor N°111 - Gardens Street
PO Box 468 - AMMAN 11953 - JORDAN
Tel. +962 6 553 4401
Fax. +962 6 553 7833
e-mail : [email protected]
CHINA
ACTEON CHINA
Office 401 - 12 Xinyuanxili Zhong Street - Chaoyang District - BEIJING
100027 - CHINA
Tel. +86 10 646 570 11/2/3
Fax. +86 10 646 580 15
e-mail : [email protected]
PHILIPPINES
ACTEON PHILIPPINES
3F King's Court II Bldg. - Corner Pasong Tamo & de la Rosa Street -
Makati City - 1200 METRO MANILA - PHILIPPINES
Tel. +632 811 2487
Fax. +632 811 2488
e-mail : [email protected]
THAILAND
ACTEON (THAILAND) LTD
23/81 Sorachai Bldg 20th floor - Sukumvit 63 Road - Klongtan Nua
Wattana - 10110 BANGKOK - THAILAND
Tel. +66 2 714 3295
Fax. +66 2 714 3296
e-mail : [email protected]
KOREA
ACTEON KOREA
8F Hanil B/D - 132-4 1Ga Bongrae-dong - JOONG-GU - SEOUL - KOREA
Tel. +82 2 753 41 91
Fax. +82 2 753 41 93
e-mail : [email protected]
INDIA
ACTEON INDIA
E-91, G.I.D.C. Electronic Estate - Sector 26 - GANDHINAGAR 382044
(Gujarat)
Tel. ++91 79 2323 8000
Fax. ++91 79 2646 2041
e-mail : [email protected]
SATELEC