Amico Alert-4 Operating & Maintenance Manual
Read Online
Read the Amico Alert-4 Operating & Maintenance Manual online using the built-in PDF viewer.
Text of the manual
Description of the Alarm
6
Shipment Details
6
The Alarm Back Box
6
The Frame/Module Assembly
6
Description of Modules
6-7
System Power Supply
6
LCD Module
6
Sensor Module
7
Installation Guide
7-11
The Alarm Box
7
For Local Sensors Only
7
Standing Pressure Test
7
Sensor
7-8
Local
7
Remote
8
Frame Assembly
8-9
System Power Supply
9
Sensor Module
10
Gas Display
10
Local
10
Remote
10
LCD Display Module
11
LCD Display Setup
11
Closing The Frame/Module Assembly
11
Configuration Guide
12-18
Network Setup
12
Changing IP Address
13-14
Email Setup
15
Text Setup
16
Connecting to Alarm
17-18
Direct Connection
17
Simple Unmanaged Network
17
Complex Managed Network
18
Recorded History Log
18
Web Audio
19
Programming Gas Locations
20
Amico Mobile Eco System App
21-22
Model Numbers
23
Spare Part Numbers
24-25
Sensors
24
Accessories/Misc.
24
Demand Check Valves
25
Dimensions
26
LCD Alarm
26
Alert-4 Sensor
26
Troubleshooting
27-28
Appendix A
29
Wiring Diagram: LCD Board
Appendix B
30
Wiring Diagram: Auto-Switch Power Supply
Appendix C
31
Wiring Diagram: LCD Display Module - Alarm Buzzer
Appendix D
32
Wiring Diagram: LCD Display Module - Local Sensor
Appendix E
33
Wiring Diagram: LCD Display Module - Remote Sensor
Appendix F
34
Technical Specifications
Appendix G
35
Wiring
4 Amico Corporation
User Responsibility
The information contained in this Installation, Operation and Maintenance Manual pertains only to the Alert-4
microprocessor based digital LCD Alarm. This product will perform as described in this manual when assembled,
operated, maintained and serviced in accordance with the installation instructions provided.
The alarm must be checked periodically. Parts that are broken, missing, worn, distorted or contaminated must be
replaced immediately. Should such repair or replacement become necessary, please contact Amico Corporation or their
distributors.
All alarms should not be repaired or altered without prior written or verbal approval from Amico Corporation or its
distributors. Failure to comply will void all warranty on the alarm.
Statements in this manual preceded by the words WARNING, CAUTION, DANGER and NOTE are of special significance.
Please read these sections carefully.
NOTE: Amico strongly recommends that alarms be checked annually by qualified staff.
WARNING: denotes steps which can prevent injury.
CAUTION: denotes steps which can prevent damage to equipment.
DANGER: denotes steps which can prevent electrical shock to
equipment or to prevent serious injury and/or death.
IMPORTANT INFORMATION
1.
Do not use impact screw driver for installation.
2.
When installation is completed, peel off the LCD screen protector.
3.
To protect from static electricity, ensure to discharge body static before installing the Medical Gas Alarm.
4.
Do not ground the shield drain wire inside the alarm panel back box.
5.
Make sure power supply is turned off while wiring.
The Amico Area LCD Alarm System (Alert-4) incorporates the latest microprocessor based technology for alarm and
surveillance systems. The alarm has been designed to provide user flexibility and reliability. This manual shall enable the
customer to install, use and maintain the alarm appropriately.
There is one “MUTE” (
) or “PUSH TO TEST” button located on the front face of the LCD panel. The button has three
functions: to silence an alarm that has occurred, to view high and low set points and to test the audio. When an audible
alarm is triggered, press the mute button to silence the alarm. To view the high and low set points or to test the audio,
press and hold the mute button for 20 seconds.
All Gases and Vacuums are displayed on the LCD screen for clear visibility to facilitate the monitoring functions of
hospital personnel. Under normal operation, the gas indicator will be in “GREEN - NORMAL“. If an alarm condition occurs,
the gas indicator will be in “RED - FAULT” and an audible alarm shall be continuous until silenced by pushing the “MUTE”
button.
•
Microprocessor based digital LCD and individual microprocessor on each sensor module
•
Ethernet capable for viewing an exact image of the alarm remotely or wireless anywhere in the building
•
Gas-specific sensors can be mounted locally or remotely, up to 2000 feet [609.6 m], utilizing #22 gauge
stranded, shielded twisted pair cable ONLY
•
DISS gas-specific sensor housed in a tamper-proof enclosure. The Sensor Module is housed in an anodized
aluminum and nickel-plated brass enclosure to act as an interference barrier
•
The Sensor Module is the smallest computer-calibrated temperature-compensated sensor in the industry
•
PSI, kPa, inHg, mmHg or BAR display (programmable)
•
Self diagnostic circuitry with error display for problem identification
•
Highly accurate Solid State Pressure piezoresistive transducer
•
Dry contacts for remote monitoring for a generic alarm condition
•
Modules are factory mounted on a hinged frame assembly for ease of installation and maintenance
•
Field programmable push buttons for adjustment of “HI” and “LOW” set-points
•
Available in 1 to 8 gases
•
Web Audio Enabled: web audio will trigger when alert occurs via ethernet
•
Unlimited records of history log to keep track of alert history for easy troubleshooting
•
Easy access via Amico Medgas App (Apple and Android)
6 Amico Corporation
Description of Modules
The Alert-4 Area Alarm is a high technology microprocessor based module:
COMMON TO ALL ALARMS
SYSTEM POWER SUPPLY
The System Power Supply has been pre-installed into the back box assembly.
The System Power Supply converts the AC voltage supply to the alarm into
two voltages: 5 VDC (regulated) required by the microprocessor hardware
and 15 VDC (unregulated) required by the buzzer and the LCD. This unit also
contains the main ON/OFF power switch, the transformer, the heat sink, the
main fuse and fuse cover, the rectifying circuitry, the terminal blocks and
the low voltage DC power cable for connecting this unit to the module. The
System Power Supply can be easily removed and reinstalled by unscrewing it
from the back box.
LCD MODULE
The LCD Module contains the LCD screen, microprocessor, buzzer and the
“MUTE” button. The function of the “MUTE” button is to silence an alarm that
has occurred. By holding the “MUTE” button for 20 seconds, the module will
display the high and low pressure set points. This module also contains
a fail-safe relay that de-energizes when the buzzer is activated. This relay
can be used with the Amico Remote Buzzer for applications requiring a
remote audible alarm, master alarm or a Building Management System.
Description of the Alarm
SHIPMENT DETAILS
When you receive an Alert-4 LCD series alarm from Amico Corporation, the package will consist of three main sections:
the Alarm Back Box, Sensors and the Frame/Module Assembly.
THE ALARM BACK BOX
The Alarm Back Box contains the auto-switchable System Power Supply with an ON/OFF switch, a built-in fuse and
terminal blocks (115 to 220 VAC - 50 to 60 Hz). The back box also incorporates the pipe stubs for applications that require
locally (in box) mounted sensors.
THE FRAME/MODULE ASSEMBLY
The Frame/Module Assembly consists of the frame and the LCD module. The hinged frame is designed to swing down
from the back box to facilitate installation and servicing of the alarm. This design will reduce installation time and
eliminate the risk of improper installation since all the modules are connected and tested at the factory.
AC Supply
115 to 220 VAC
DC Power Cable:
Connect to
Annunciator Module
G
N
L
L - Live
N - Neutral
G - Ground
Fuse (1 AMPw)
Ground
Toggle Switch
Description of Modules
SENSOR MODULE
The Sensor Module contains the transducer which converts the source of the pressure/vacuum
into a digital signal that is displayed on the LCD alarm. The sensor module shall be housed in an
anodized aluminum and nickel-plated brass enclosure to act as a barrier against interference
and it is temperature compensated. Each sensor is clearly labeled and color coded for the gas or
vacuum being monitored. The sensor module contains a gas-specific DISS fitting to ensure correct
connection of the proper sensor to the respective gas. Each sensor has been factory calibrated for
the specific gas shown on the sensor housing.
For Annual Test
•
Hold the MUTE button for twenty (20) seconds to display the current high and low
set points and audio level.
STEP 1: THE ALARM BOX
Install the back-box to the studs of the wall at the desired height. Ensure that the box is securely in place. The mounting
brackets are adjustable to suit the thickness of the wall. MAKE SURE the box is parallel, squared and flush with the
finished wall surface to ensure that the frame assembly will fit properly.
STEP 2: FOR LOCAL SENSORS ONLY
If the sensors are to be mounted locally (inside the back box), the pipe stubs must be connected to the pipeline. Using
silver-brazing techniques, connect each pipe stub to its appropriate gas or vacuum while ensuring that the bottom of the
pipe stub is wrapped with a damp cloth. If the DISS demand check-valve is installed prior to brazing, BE CAREFUL not to
damage the check-valve by overheating the lower portion of the copper pipe. When the brazing of pipe stubs has been
completed, the system can be pressure tested.
STEP 3: STANDING PRESSURE TEST
Perform a standing pressure test on the piping system as per NFPA-99 “Health Care Facilities” or CSA-Z7396.1 “Medical Gas
Pipeline Systems”. Inspect all joints for leaks and make certain each gas is piped to a correspondingly labeled gas service.
STEP 4: SENSOR
A. LOCAL (inside the back box)
i.
Locate the gas-specific sensor module to be installed.
ii. Install the provided demand check-valve with a sensor module to the pipe stub.
iii. The sensor module contains a gas-specific DISS fitting. Push the sensor module hex-nut and
nipple adapter up into the demand check-valve. With a wrench, tighten the nut so that it
makes a good seal.
Note: Maximum input pressure not to exceed more than 249 psi for pressure sensors and 30 inHg for vacuum sensors.
Installation Guide
8 Amico Corporation
Alert-3 sensor operating pressure range:
Mid Pressure
(0 to 99 psi)
-
Oxygen, Medical Air, Nitrous Oxide, Carbon Dioxide
High Pressure
(0 to 249 psi)
-
Nitrogen, Instrument Air
Vacuum
(0 to 30 inHg)
-
Vacuum, WAGD, AGSS
CAUTION: To protect from static electricity, ensure to discharge body static before installing the Medical Gas
alarm and sensors. Do not ground the sheild drain wire at sensor or inside alarm panel back box.
B: REMOTE (outside the back box)
i.
Connect a tee (supplied by others) to the pipeline with a 1/4" NPT female
connection that will accept the DISS Demand Check Valve.
ii. Locate the gas-specific sensor module to be installed.
iii. Thread the DISS Demand Check Valve into the correct gas pipeline.
iv. The sensor module contains a gas-specific DISS fitting. Push the sensor module
hex-nut and nipple adapter into the demand check-valve. With a wrench, tighten
the nut so that it makes a good seal.
v. Marrette in a junction box (supplied by others) the #22 gauge stranded, shielded twisted pair sensor cable to
the installation cable (supplied by others).
vi. In the presence of any electrical, magnetic, radio frequencies, wireless or other interference, cable installation
(supplied by others) running from the junction box (supplied by others) to the alarm MUST be placed in a
metallic conduit.
Installation Guide
P a g e : 1 2
2.
In the back box, there are
colour coded gas labels
located under the DISS
Demand check valves. Each
label identifies where each
sensor module is to be
placed.
3.
The sensor module contains
a gas specific DISS fitting.
Push the sensor module
hex-nut and nipple adapter
up into the demand check-
valve. With a wrench, tighten the nut so that it makes a good seal.
NOTE: Pressure on sensors are not to exceed 250psi
for Pressure sensors and 30” for Vacuum sensors
REMOTE (Outside the Back Box)
1.
Connect a Tee (supplied
by others) to the pipeline
with a 1/4" NPT female
connection that will accept
the DISS Demand check-
valve.
2.
Locate the gas specific
sensor module to be
installed.
3.
Thread the DISS Demand
check-valve into the
correct gas pipe line.
4.
The sensor module contains a gas specific DISS fitting. Push the
sensor module hex-nut and nipple adapter up into the demand
check-valve. With a wrench, tighten the nut so that it makes a
good seal.
i.
Attach the LCD screen to the back box assembly by using star head
screws (provided with frame in a plastic bag).
STEP 5: FRAME ASSEMBLY
ii. Attach the frame wire with 2 dome head screws (provided with frame
in a plastic bag). This will allow the frame assembly and back box to the
fastened securely together.
iii. Close the LCD screen with a back box by tightening two screws
provided with divider plate.
The microprocessor circuitry on the Alert-4 alarm contains sophisticated integrated semiconductors.
DO NOT TOUCH any of the components on the board. Static discharge can cause the modules to
malfunction or become damaged.
2.
Keep the shield drain wires as short as possible and taped to prevent from grounding, so they can not
touch the front panel circuit board when front panel is closed.
3.
Do not use impact driller or screwdrivers when assembling new frame.
STEP 6: SYSTEM POWER SUPPLY
CAUTION: TURN OFF POWER SWITCH before changing any modules and/or disconnecting any cables. Failure
to do so can cause the fuse to blow, damaging the circuitry.
1.
Ensure that the ON/OFF switch is in the OFF position.
2.
Through the top left side of the back box, bring in the AC power wires. Knockouts are provided for
making conduit connections to the box. All wiring is to be installed according to local and national
codes.
3.
Connect the AC power to the terminal blocks as shown in the wiring diagram (see Appendix B).
CAUTION:
1.
Verify that power has been switched off prior to working on the alarm
2.
Risk of electric shock. Disconnect power at the circuit breaker before removing power supply shield
iv. Loosen the screws from the sides frame section (2 screws provided).
v. Cover the frame and tighten the side screws.
10 Amico Corporation
STEP 7: SENSOR MODULE
A: GAS DISPLAY (on screen location)
The location of gases displayed on screen is dependant upon which sensor channel each individual gas is connected to.
The display below indicates which sensor channel corresponds to each location the gas will be displayed on the LCD screen.
B: LOCAL (inside the back box)
i.
The sensor module is provided with a 6" - 8" [0.1 m - 0.2 m] #22 gauge stranded,
shielded and twisted pair cable. One wire is red (positive) and the other wire is
black (negative). Connect the wires to the display module as shown in Appendix D.
Take the red wire from the sensor and attach it to terminal “Sensor +” on the display
module. Take the black wire from the sensor and attach it to terminal “Sensor -”. The
terminal block on the display module is clearly marked for proper connection of the
sensor wires.
ii. Repeat the above procedures with the remaining sensor modules.
C: REMOTE (outside the back box)
i.
The sensor module is provided with 6" - 8" [0.1m - 0.2m] #22 gauge stranded, shielded and twisted pair cable.
Connect the wires to a junction box (supplied by others) located near the sensor as per the wiring diagram
(see Appendix E).
ii. Connect a #22 gauge stranded, shielded and twisted pair cable ONLY, up to 2000 ft [609.6 m]. Knockouts are
provided throughout the alarm back box.
iii. Connect the red wire from the cable to the terminal on the display module marked ”Sensor +”. Connect the
black wire to terminal “Sensor -” as shown in the wiring diagram (see Appendix E).
iv. Repeat the above procedures with the remaining sensor modules using the wiring diagram.
NOTE:
•
When remote sensors are used, ONLY a #22 gauge stranded, shielded twisted pair cable must be used (BELDEN
#8451 or equivalent, supplied by others).
•
Do not ground the shield drain wire at sensor or inside the alarm panel back box
Installation Guide
1 Gas
2 Gases
3 Gases
4 Gases
5 Gases
6 Gases
7 Gases
8 Gases
STEP 8: LCD DISPLAY MODULE
If the dry contacts for a generic alarm is to be used for remote monitoring, connect the wires to
the appropriate terminals: COM (Common), NO (Normally Open) or NC (Normally Closed), using
the diagram in Appendix A.
See Appendix G for contact rating.
Once the sensors are connected and the power has been switched on, use the following steps
to setup the LCD Alarm.
STEP 9: LCD DISPLAY SETUP
i.
Press the Setup (B1), choose the language by pressing (B2) or (B3) and (B4) to select
the language.
ii. Volume level: 90, 80, 70, 60 - press CHANGE UP (B2)/DOWN(B3) and (B4) to select the
volume level.
iii. LCD brightness: 1-19 press CHANGE UP (B2)/DOWN (B3) and (B4) to select brightness.
iv. Select DST ( Daylight Saving Time) (ON/OFF)- press CHANGE UP/DOWN to change.
v. Select the DATE (YEAR/MONTH/DAY) and press CHANGE UP/DOWN to change YEAR, MONTH, DAY.
vi. Select TIME (HR/MIN) and press CHANGE UP/DOWN to change HR/MIN.
*Screen will prompt with detected gas sensor and display the gas type automatically (Example: OXYGEN).
vii. Select unit of measure: Gas (PSl, BAR, KPA) for Vacuum ( in HG, mmHg) press CHANGE UP(B2)/DOWN(B3) to
change, press CHANGE UP (B2)/DOWN (B3) to change and (B4) to select.
viii. Select LOW/HIGH alarm set points: LOW ALARM 40/ HIGH ALARM 60, press CHANGE UP (B2)/DOWN (B3) to
change and (B4) to select.
** Screen will display line pressure of the selected gas sensor
ix. CURRENT OFF SET: to re-calibrate the pressure reading, press CHANGE UP (B2)/DOWN (B3) and press (B4) to
select.
** Repeat steps until all sensors are scanned and configured**
** All gases should be displayed after setup is complete.
If any errors occur, repeat the steps above.
STEP 10: CLOSING THE FRAME/MODULE ASSEMBLY
i.
Close the frame panel by tightening the screws found on the frame panel to the back box. Ensure that the
screws are securely fastened to keep the LCD Alarm closed.
ii. Carefully place the front frame over the frame panel. Screw in the screws that were removed in Step 5, part ii.
The alarm shall now be ready for use.
CAUTION: To protect from static electricity, ensure to discharge body static before installing the Medical Gas
alarm and sensors.
Installation Guide
12 Amico Corporation
NETWORK SETUP
CAUTION: Have the information systems personal set up the network interface. Before making any changes
to the network setting notify information systems personals.
EQUIPMENT NEEDED TO SETUP THE NETWORK
•
PC with Ethernet connection
•
PC with web browser, (Internet Explorer, Google Chrome)
•
CAT 5 Ethernet cable ( Straight-Through)
•
SD Card
SETUP
•
Connect Alert-4 Area alarm to an Ethernet switch using a CAT 5 Ethernet cable
•
For direct connection to PC, connect the Area alarm to PC using cat-5 Ethernet cable
NOTE: It is best to use a switch instead of a hub because the device communicates at 10 Mbps. A switch is better able to
support this speed, improves network performance and keeps unnecessary traffic from being routed to the alarm.
•
Amico Alert-4 Area alarm will be set to factory default setting, the IP Address, Subnet Mask and Gateway as
following:
IP address:
192.168.1.100
Gateway:
192.168.1.1
Subnet Mask:
255.255.255.0
•
Static IP configuration needs to used to connect to the Hospital Network
•
Upon power-up, the device will immediately begin using the static IP configuration
•
Verify the green “LINK” LED illuminates at the Ethernet Port
Configuration Guide
CHANGING IP ADDRESS
Open the SD Card with the files provided by Amico Corporation. Open the file named “network” to change the network
IP address.
Change the default IP address to the desired IP address, Gateway and Subnet Mask; then save the file by clicking “File“
then “Save.“
When all files are saved in the SD card, insert the SD Card into the SD Card Slot on the LCD Alarm board (Refer to
Appendix A).
Configuration Guide
14 Amico Corporation
To load the network configuration file, press and hold the Reset button and the Setup button at the same time for two
seconds; then let go of the Reset button while still holding down the Setup button, until the network configuration file
are uploaded to the Area Alarm.
If the configured information does not appear on the screen, repeat the steps above. If the problem persists, contact
Amico Corporation for further assistance.
•
Once the information is visible on the LCD Alarm screen, leave the SD Card in the slot for approximately 1
minute in order for the information to be completely uploaded onto the alarm, and then proceed to remove
the card.
•
Once the card has been removed, restart the LCD Alarm to ensure that the configured network setting have
been saved onto the LCD Alarm.
BUILD DATE: OCT 24 2014
BUILD VERSION: 1515
GATEWAY: 192.168.1.1
MASK: 255.255.255.0
IP ADDRESS: 192.168.1.100
Configuration Guide
CAUTION: Have the information system personnel setup the email interface. Notify information systems
personnel before making any changes to the network setting.
SMTP server is required for electronic mail service.
The following parameters are needed to activate the email service. Information systems personnel will be able to provide
the necessary parameters.
Open the network configuration file on the SD card provided by Amico to input the SMTP server parameters.
IP_SMTP= (provide SMTP server IP address)
SMTP_PORT= (provide SMTP port number)
SMTP_USER= (provide SMTP user name)
SMTP_PASSWORD= (provide SMTP password)
EMAIL= (provide recipient email address)
DEVICE_ID= (location where device is installed)
To load the Network configuration, press and hold the Reset and Setup buttons at the same time for two seconds. Let
go of the Reset button while still holding the Setup button until the new configuration is uploaded to the panel. When
parameters are uploaded, the panel will display the new parameters in boot sequence.
CAUTION: When wiring source equipment or gas sensors to the panel, make sure that the CAT 5 cable is
unplugged or turn off the panel before wiring. If the panel is connected to the mail server it will send email
while wiring the terminals to the recipients.
NOTE: IP address has to be routable if connecting the alarm panel to the global network.
Configuration Guide
16 Amico Corporation
TEXT SETUP
CAUTION: Have the information system personnel setup the email interface. Notify information systems
personnel before making any changes to the network setting.
SMTP server is required for electronic text service.
The following parameters are needed to activate the text service. Information systems personnel will be able to provide
the necessary parameters.
Open the network configuration file on the SD card provided by Amico to input the SMTP server parameters.
IP_SMTP= (provide SMTP server IP address)
SMTP_PORT= (provide SMTP port number)
SMTP_USER= (provide SMTP user name)
SMTP_PASSWORD= (provide SMTP password)
EMAIL= (provide recipient phone number and SMS gateway address, e.g., [email protected])
DEVICE_ID= (location where device is installed)
To load the Network configuration, press and hold the Reset and Setup buttons at the same time for two seconds. Let
go of the Reset button while still holding the Setup button until the new configuration is uploaded to the panel. When
parameters are uploaded, the panel will display the new parameters in boot sequence.
CAUTION: When wiring source equipment or gas sensors to the panel, make sure that the CAT 5 cable is
unplugged or turn off the panel before wiring. If the panel is connected to the mail server it will send email
while wiring the terminals to the recipients.
Configuration Guide
CONNECTING TO ALARM
•
Start the web browser (Google Chrome, Internet Explorer)
•
Enter the device IP address eg: (http://192.168.1.1xx) in the browser’s address bar*.
NOTE: To find Alarm IP Address, press reset button on the back of the Alert-4 Area Alarm.
SIMPLE UNMANAGED NETWORK
DIRECT CONNECTION
Configuration Guide
Switch
18 Amico Corporation
COMPLEX MANAGED NETWORK
Configuration Guide
Computer
Router
Server
Switch
Recorded History Log
Must leave the SD card in the SD card slot to record the logs. Panel will automatically log the track of alerts with date and
time provided.
1.
To view the logs
2.
Remove the SD card from the SD card slot.
3.
Plug the SD card to a PC card reader and open the SD card.
4.
Open the file called “ERROR” by double clicking
5.
File will displays history logs.
When an alert occurs, the PC audio that is monitoring will turn on automatically. To Mute the audio, click the mute button
on the web page. If the PC audio doesn't turn on automatically, click mute and re-activate the alert to turn PC Audio ON.
Note: By muting the audio in the web page doesn't not silence audio at the alarm panel. To silence press mute button at
the panel.
20 Amico Corporation
Programming Gas Locations
Notepad File
Updated Display on LCD Alarm
1. Use the Notepad program to enter information for gas locations. Each line can hold up to a maximum of 16
characters. (Notepad is a generic text editor included with all versions of Microsoft Windows)
2. Two lines may be used per individual gas.
3. The order of the text must go in order of the gas; meaning the first two lines of text shall represent the 1st gas
location, the next two lines of text shall represent the 2nd gas location, and so forth.
•
Please refer to the diagram in the Installation Guide under step 7a (see page 9) to determine gas location on the LCD
Alarm screen.
4. Once all the text has been entered, save the file onto the SD Card with the file name: location.
5. Insert the SD Card into the SD Card Slot on the LCD Alarm board (Refer to Appendix A).
6. While the LCD Alarm is on, to load gas location “location.txt” file, press and hold the Reset and Setup buttons at the same
time for two seconds. Let go of the Reset button while still holding the Setup button until the file is uploaded to the
panel. When file is uploaded, the panel will display the location under the gas icons.
•
If the gas location text does not appear on the screen, repeat step #6. If the problem persists, contact Amico
Corporation for further assistance.
7. Once the text is visible on the LCD Alarm, leave the SD Card in the slot for approximately 1 minute in order for the
information to be completely uploaded onto the alarm, and then proceed to remove the card.
8. Once the card has been removed, restart the LCD Alarm to ensure that the locations have been saved onto the
LCD Alarm.
NOTE: Symbols and special characters cannot be displayed.
PATIENT ROOMS
101-110
PATIENT ROOMS
110-112
PATIENT ROOMS
112-119
PATIENT ROOMS
119-120
PATIENT ROOMS
120-122
PATIENT ROOMS
122-124
PATIENT ROOMS
124-128
PATIENT ROOMS
128-130
Amico Mobile Eco System App
Amico is pleased to introduce the latest technology for monitoring the Medical Gas System of a hospital on a mobile
phone. This App allows facilities to monitor the pipeline equipment in real time on an iPhone or Android phones.
The App will provide an exact and instant visual representation of the equipment in alarm condition, thereby eliminating
the need for nurses to call maintenance personnel in the event of a gas outage. The App will also help maintenance
personnel to localize the outage for a quicker resolution.
NOTE: If Alert-4 alarm is given with local IP address, the phone must be connected to local WiFi before connecting the
app to the Alert-4 alarm. If Alert-4 alarm is given with global IP address, connecting the phone to local WiFi is not needed.
Click and open Amico Mobile Eco System App on the phone
Home Screen
Add a Device Screen
AMICO MOBILE ECO SYSTEM APP
Download Now!
Home Screen with Devices Added
Click to Add
a Device
Type in Device Name
and Location
Type in Device
IP Address
Click to Select
Product Category
Click to Save
22 Amico Corporation
Amico Mobile Eco System App
Press and hold the device screen
to move the device
Device Manual
Display Exact Image of the Alarm
Click to Open
Manuals
Click to
Contact Amico
Slide Left to
Delete Device
Click to add
more devices
Re-order mode
ALERT-3 SENSOR MODULE
NOTE: Each Alert-3 Sensor comes with an A2P-PIPE
The “L” Defines the
Language:
English (NFPA)
= U
English (CSA/ISO)
= E
Spanish (NFPA/ISO) = S
French (CSA/ISO)
= F
The “L” Defines the Language:
English (NFPA)
=
U
English (CSA/ISO) =
E
French (CSA/ISO) =
F
Spanish (NFPA)
=
S
Remote/Local
= R
Conversion (Retrofit) = C
The ”X” Defines the Type of Gas:
Oxygen
=
O
Medical Air
=
A
MedVac
=
V
Nitrous Oxide
=
2
Nitrogen
=
N
Carbon Dioxide
=
C
WAGD (NFPA)
=
W
AGSS (CSA/ISO)
=
E
Instrument Air
=
I
”GAS” Defines the Type of Gas:
Oxygen
=
OXY
Medical Air
=
AIR
MedVac
=
VAC
Nitrous Oxide
=
N2O
Nitrogen
=
NIT
Carbon Dioxide
=
CO2
WAGD (NFPA)
=
WAG
AGSS (CSA/ISO)
=
AGS
Instrument Air
=
IAR
24 Amico Corporation
Spare Part Numbers
Model Number
Description
A3P-SENS-E-AIR
Sensor Module ISO-AIR Eng. Alert-3
A3P-SENS-E-CO2
Sensor Module ISO-CO2 Eng. Alert-3
A3P-SENS-E-AGS
Sensor Module ISO-AGS Eng. Alert-3
A3P-SENS-E-N2O
Sensor Module ISO-N2O Eng. Alert-3
A3P-SENS-E-NIT
Sensor Module ISO-NIT Eng. Alert-3
A3P-SENS-E-OXY
Sensor Module ISO-OXY Eng. Alert-3
A3P-SENS-E-VAC
Sensor Module ISO-VAC Eng. Alert-3
A3P-SENS-E-IAR
Sensor Module ISO-IAR Eng. Alert-3
A3P-SENS-U-AIR
Sensor Module USA-AIR Eng. Alert-3
A3P-SENS-U-OXY
Sensor Module USA-OXY Eng. Alert-3
A3P-SENS-U-VAC
Sensor Module USA-VAC Eng. Alert-3
A3P-SENS-U-WAG
Sensor Module USA-WAG Eng. Alert-3
A3P-SENS-U-IAR
Sensor Module USA-IAR Eng. Alert-3
Model Number
Description
A2P-POWER-V2
Power Supply Module Alert-2
A2P-BOXASS-3LCD
Alarm Back Box Assembly 3-Station Alert-2
A2X-BOX-ASY-3LCD
Alert-3 LCD Alarm Box without Power Supply
A4A-AREA-FRAME
LCD Alarm Frame Assembly for Alert-4 Area
A2P-PIPE-AIR
Pipe, Demand Check, Gas Specific Label AIR
A2P-PIPE-OXY
Pipe, Demand Check, Gas Specific Label OCY
A2P-PIPE-VAC
Pipe, Demand Check,Gas Specific Label VAC
A2P-PIPE-NIT
Pipe, Demand Check,Gas Specific Label NIT
A2P-PIPE-CO2
Pipe, Demand Check,Gas Specific Label CO2
A2P-PIPE-EVA
Pipe, Demand Check,Gas Specific Label WAG
A2P-PIPE-IAR
Pipe, Demand Check,Gas Specific Label IAR
A2P-PIPE
Pressure Module Pipe Assembly (Alert-2)
A3X-P-SW-PVCAP
LCD PVA Switch Cap-Gray
A3X-LCD-LABEL
LCD Alarm Front Label
A3X-LCD-LABEL-MUTE
Alert-3 Mute Label
SENSORS
ACCESSORIES/MISC.
Spare Part Numbers
Model Number
Description
S-DIS-DEMC-AIR
DISS Demand Check Valve 1/4" MNPT - AIR
S-DIS-DEMC-CO2
DISS Demand Check Valve 1/4" MNPT - CO2
S-DIS-DEMC-NIT
DISS Demand Check Valve 1/4" MNPT - NIT
S-DIS-DEMC-N2O
DISS Demand Check Valve 1/4" MNPT - N2O
S-DIS-DEMC-EVA
DISS Demand Check Valve 1/4" MNPT - EVA
S-DIS-DEMC-OXY
DISS Demand Check Valve 1/4" MNPT - OXY
S-DIS-DEMC-VAC
DISS Demand Check Valve 1/4" MNPT - VAC
S-DIS-KIT-OXY
DISS Demand Check, Nut and Nipple - OXY
S-DIS-KIT-AIR
DISS Demand Check, Nut and Nipple- AIR
S-DIS-KIT-VAC
DISS Demand Check, Nut and Nipple- VAC
S-DIS-KIT-N2O
DISS Demand Check, Nut and Nipple - N2O
S-DIS-KIT-NIT
DISS Demand Check, Nut and Nipple - NIT
S-DIS-KIT-WAG
DISS Demand Check, Nut and Nipple- WAG
S-DIS-KIT-AGS
DISS Demand Check, Nut and Nipple- AGS
S-DIS-KIT-IAR
DISS Demand Check, Nut and Nipple - IAR
S-DIS-KIT-CO2
DISS Demand Check, Nut and Nipple - CO2
DEMAND CHECK VALVES
26 Amico Corporation
Dimensions
LCD ALARM
1/2"-14 NPSM [13]
Wrench Flats
1
[25]
6" - 8" [0.1m - 0.2 m] #22 gauge
stranded, shielded and twisted pair
cable supplied
1 7/8
[48]
DISS
Connector
ALERT-4 SENSOR
Inch
[mm]
Front View
Side View
Top View
Bottom View
NOTE: LCD Alarm itself is 8 lbs. Each sensor is 1 lb.
1/2-14NPSM THD.
Brass 1 [25] HEX
1/4 [6] NPT
71/2
[191]
6 3/8
[162]
NOTE: DISS demand check
valves by 1/4 [6] NPT,
supplied with each sensor
Brazing Filler
Metal AWS
Spec. BAg-7
(Approximate, can vary)
1/2" Conduit
Connection
7/8
[21]
3/4
[19]
3/4
[19]
3/4
[19]
7 1/2
[191]
2 1/2
[64]
2 1/2
[64]
2 1/2
[64]
2 1/2
[64]
2 1/4
[57]
1 3/4
[44]
1
[25]
8 1/8
[207]
11 1/2
[292]
11 7/8
[302]
61/8
[156]
1 5/16
[33]
1 5/16
[33]
1 5/16
[33]
1 3/16
[30]
1 3/16
[30]
1 3/16
[30]
15/16
[24]
9 3/4
[248]
1 3/4
[44]
7 1/8
[182]
1 1/4
[32]
9
[229]
4 3/4
[121]
3 7/8
[100]
1 3/8
[36]
1 1/4
[32]
10 3/4
[272]
An error or “LOW ALARM”
LCD screen
The Microprocessor detected a fault and
has shut down
Turn power switch to OFF position. Wait for at least 5
seconds before turning ON the power. The program will
reset itself.
Faulty wire connection between the
sensor and LCD module
Check wiring diagram in Appendix D and Appendix E
No power on the alarm
AC power not available
a.
Ensure that the ON/OFF switch on the power
supply module is turned ON (see Appendix B).
b.
AC wiring not connected.
c.
Check the building electrical breaker to ensure
that the power is ON.
d.
Check the voltage at the terminal block above the
transformer. Ensure that 115 VAC to 220 VAC is
being supplied.
Fuse is blown
Check the fuse. The fuse is located on the upper-right
corner of the system power supply. Replace the fuse if it
is defective (see Appendix B and Appendix G).
DC power plug not connected to the LCD
module
a.
Ensure that the DC power plug is firmly in its
socket on the LCD module.
b.
Replace the System Power Supply unit if all the
above steps fail to resolve the problem.
Power light is ON, however
there is no display on LCD
screen
Power failure on screen
a.
Remove all transducers and reset power by
switching the power supply OFF, then ON.
b.
Replace the LCD module.
No audible alarm
DC power cable is disconnected or loose,
check ribbon cable
a.
Ensure that the DC power cable from the system
power supply is firmly connected to the LCD
module.
b.
Replace LCD module.
Audible signal will not
silence
Faulty display module
Disconnect the ribbon cable from the back of the faulty
display module and replace the LCD module.
Connection of the DC power cable from
system power supply to LCD module is
loose
Disconnect the DC power cable from the LCD module
and then reconnect. If audible alarm still persists,
replace the System Power Supply unit.
Faulty push button
Replace the LCD module.
Gas reading incorrect
Loose connection of DISS fittings
Ensure that the sensor module is properly connected to
the DISS demand check-valve
Sensor module is not properly wired to
the display module
Ensure that the sensor module is properly wired to the
LCD module by using wiring diagram in Appendix D or
Appendix E
Requires calibration
Re-calibrate pressure reading (see page 11, Step 9 - iv,
“CURRENT OFF SET”)
Defective sensor
Replace the sensor module
Defective LCD display
Replace the LCD module
28 Amico Corporation
Troubleshooting
Symptom
Cause
Corrective Action
Display shows
“NO SENSORS”
No sensor(s) are connected to the LCD
module
Make sure sensor module(s) are connected to LCD
module (see Appendix D and Appendix E)
Program not set up
Press setup and select button to program all connected
sensors (see page 11, Step 9)
Faulty sensors
Replace sensors
Network connectivity lost
Wrong network cable used
a.
Use Cat 5 or 6 (Straight-Through)
b.
Static IP address must be used to configure the
network switch to the correct port number
c.
To test connectivity, assign static IP address to a
PC as same subnet as panel. Connect the PC to the
panel. Open a shell prompt (Microsoft Windows
Command Prompt or MS-DOS prompt) on the start
menu, type “ping” followed by a space and then the
IP address of the panel then hit Enter. When panel is
responded with a ping, connection is successful.
Dynamic IP address
Wrong port number assigned to the
network switch
Email or Text notification
not send
SMTP email server parameters not
assigned, authentication errors
a.
Contact Hospital IT administrator for SMTP
parameters.
b.
Must use plain text authentication.
c.
TLS/SSL authentication not supported.
d.
Configure the exchange server to accept plain text
authentication from the panel
**Check the SMTP server logs for reported authentication
errors from the panel**
Alarm Logs not recorded
Wrong memory card is used. Must use
SD card between 2GB – 32GB
a.
Must use SD card between 2GB – 32GB
b.
SD card must be formatted to FAT32 system.
**Must leave an SD card to the SD card slot to record
alarm logs**
**Set the DATE & TIME appropriately**
SD card formatted incorrectly
FACTORY DEFAULT SETTING GAS
Mid Pressure
Hi
=
60 psi
Low
=
40 psi
Vacuum
Hi
=
32 inHg
Low
=
12 inHg
High Pressure
Hi
=
195 psi
Low
=
140 psi
(located on reverse side of board)
LCD Cable
Connector
SD Card
Four Push
Buttons
12V DC
Output
LCD Power
Dry Contacts
(to Building Management
System)
Buzzer
DC
Connector
Buzzer
Jumper
Ethernet
Port
S 8
S 7
S 6
S 5
S 4
S 3
S 2
S 1
Appendix A
WIRING DIAGRAM: LCD BOARD
CAUTION:
1.
Keep the shield drain wires as short as possible and taped to prevent from grounding, so they cannot
touch the front panel circuit board when front panel is closed.
2.
To protect from static electricity, ensure to discharge body static before installing the Medical Gas Alarm
and Sensors.
DC Connector
LCD Power
Mute Button
(located on reverse side of board)
Dry-Contacts
(Building
Management System)
Four Push
Buttons
12V DC
Output
LCD Cable
Connector
SD Memory Card
Slot (memory card
not provided)
Buzzer
Buzzer
Jumper
Ethernet
Port
Orange
(15 VDC)
Blue
(5 VDC)
Black
(Ground)
30 Amico Corporation
Appendix B
WIRING DIAGRAM: AUTO-SWITCH POWER SUPPLY
AC Supply
115 to 220 VAC
DC Power Cable:
Connect to
Annunciator Module
G
N
L
L - Live
N - Neutral
G - Ground
Fuse (1 AMPw)
Ground
Toggle Switch
CAUTION:
1.
Verify that power has been switched off prior to working on the alarm
2.
Risk of electric shock, disconnect power at the circuit breaker before removing power supply shield
Orange (15 VDC)
Blue (5 VDC)
Ground
Fuse (1 AMP)
AC Supply
115 to 220 VAC
ON/OFF
Toggle Switch
DC Power Cable:
Connect to the
female connector
on the LCD Module
L - Live
N - Neutral
G - Ground
Black (Ground)
WIRING DIAGRAM: LCD DISPLAY MODULE - ALARM BUZZER
COM
B4
Mute
/ tc
ele
S
hange DOWN
C
hange UP
C
Up
- t
e
S
B3
B2
B1
+12V
NO
NC
GRND
Reset
Button
S 8
S 7
S 6
S 5
S 4
S 3
S 2
S 1
NOTE:
Relays on the annunciator are fail
safe for Version 3.1 or newer.
NC COM NO 12V GND
NO COM NC 12V GND
Optional: To Amico Manifold Universal
Remote Buzzer
NO COM NC 12V GND
NC COM
NO
Optional: Abnormal Alarm
To Amico: Master Module
or Building Management
System
32 Amico Corporation
Appendix D
WIRING DIAGRAM: LCD DISPLAY MODULE - LOCAL SENSOR
DC Power Cable:
Connect to
Annunciator Module
COM
B4
Mute
/ tc
ele
S
hange DOWN
C
hange UP
C
Up
- t
e
S
B3
B2
B1
+12V
NO
NC
GRND
Reset
Button
S 8
S 7
S 6
S 5
S 4
S 3
S 2
S 1
OXYGEN
# 22 Gauge twisted
pair shielded cable
6”-8” [0.1m-0.2m] supplied
CAUTION:
To protect from static electricity,
ensure to discharge body static before
installing the Medical Gas Alarm and
sensors
NOTE:
Do not ground the shield drain wire at
sensor or inside alarm panel back box
Sensor
Module
6" - 8" [0.1m - 0.2 m]
#22 gauge stranded,
shielded twisted pair
cable supplied
Black
Red
Orange
(15 VDC)
Blue
(5 VDC)
Black
(Ground)
WIRING DIAGRAM: LCD DISPLAY MODULE - REMOTE SENSOR
Note:
For multiple sensors, a
multi-conductor twisted
pair cable can be used.
OXYGEN
# 22 Gauge twisted
pair shielded cable
6”-8” [0.1m-0.2m] supplied
Junction Box
(by others)
Sensor
Module
DISS
Oxygen
Pipeline
Remote location alarm
DC Power Cable:
Connect to
Annunciator Module
COM
B4
Mute
/ tc
ele
S
hange DOWN
C
hange UP
C
Up
- t
e
S
B3
B2
B1
+12V
NO
NC
GRND
Reset
Button
S 8
S 7
S 6
S 5
S 4
S 3
S 2
S 1
NOTE:
For multiple sensors, a multi-conductor
#22 Gauge stranded, shielded and
twisted pair cable ONLY must be used.
Appendix E
Sensor
Module
Oxygen
Pipeline
DISS
6" - 8" [0.1m - 0.2 m]
#22 gauge stranded,
shielded twisted pair
cable supplied
Marrette the sensor
cable in a junction box
(supplied by others) to
the installation cable
(supplied by others)
#22 gauge stranded, shielded twisted pair
cable ONLY must be used, up to a distance
of 2000 ft [609.6 m]. In the presence of any
electrical, magnetic radio, wireless or other
interference, the installation cable MUST
be placed in a metallic conduit
Red
Black
Orange
(15 VDC)
Blue
(5 VDC)
Black
(Ground)
34 Amico Corporation
TECHNICAL SPECIFICATIONS
Supply Voltage:
115 - 220 VAC, 50 - 60 Hz
Current Draw:
1 Amp. Max.
Fuse (1/4 * 1-1/4):
Fast Blow 1 Amp.
Cable requirement:
LCD Alarm to Remote Sensor:
Important:
Cable:
ONLY a #22 gauge stranded, shielded twisted pair cable must be used. (Belden # 8451
or equivalent.) In the presence of any electrical, magnetic, radio frequencies, wireless or
other interference, cable installation MUST be placed in metallic conduit.
Distance:
Maximum 2000 ft [609.6 m]
Signal:
30 VDC
-
1.0 Amps.
60 VDC
-
0.3 Amps.
125 VAC
-
0.5 Amps.
LCD Alarm to Master:
Distance:
Maximum 10,000 ft [3,000 m]
Cable:
Minimum #22 gauge stranded wire
Signal:
5 VDC
-
< 5 µA
LCD Generic Alarm:
Output:
Dry Contacts NC, open on Alarm
Rating:
30 VDC
-
1.0 Amps.
60 VDC
-
0.3 Amps.
125 VAC
-
0.5 Amps.
Appendix F
1. All wiring shall be protected from physical damage by raceways, cable trays or conduit in accordance with NFPA
70, National Electric Code or the Canadian Electrical Code.
2. All alarms are to be powered from the life safety branch of the emergency power system as required by
applicable standards.
3. Alarm panel wires should be directly connected to switches or sensor as required by applicable standards.
4. All wire runs should be made with color coded wire. Record color, signal and source of signal for each wire lead
to aid in connection of alarm finish components.
5. The alarm panel and remote sensors should not be installed near radio transmitters, electrical motors, electrical
control room, switch gear, CT scanners, MRI machines or high voltage lines
6. In the presence of any electrical, magnetic, radio frequencies, wireless or other interference, cable installation
MUST be placed in metallic conduits.
7. No solid wire should be used for connecting sensors or master alarms to source equipment
8. To protect from static electricity, ensure to discharge body static before installing the Medical Gas Alarm and
Sensors
9. Do not ground the shield drain wire at sensor or inside alarm panel back box
2.
Low Voltage wire type, size and other requirements
All low voltage wiring must meet the following criteria:
1. #22 AWG stranded, shielded twisted pair wire ONLY must be used, rated for 300V and 60°C (140°F) minimum.
(Belden 8451 or equivalent)
2. Marrette the sensor cable in a junction box (supplied by others) to the installation cable (supplied by others)
to protect from physical damage, radio frequencies and EMI
3. For multiple sensors, a multi-conductor #22 gauge stranded, shielded and twisted pair cable ONLY must be
used
The following rules along with references to this manual’s schematics clarify wiring requirements. Two conductor cables
(must be #22 gauge stranded, shielded and twisted pair cable type) are required for each Gas Sensor module to the Gas
Input board.
Appendix G
WIRING
Amico Corporation | 85 Fulton Way, Richmond Hill, ON L4B 2N4, Canada
71 East Industry Court, Deer Park, NY 11729, USA
Toll Free Tel: 1.877.462.6426 | Tel: 905.764.0800 | Fax: 905.764.0862
Email: [email protected] | www.amico.com