ADMINISTRATIVE INFORMATION
SERVICE NOTE CLASSIFICATION:
MODIFICATION AVAILABLE
ACTION
[[]] PERFORMANCE ENHANCEMENT
CATEGORY:
AGREEABLE TIME [x] SERVICE / RELIABILITY ENHANCEMENT
LOCATION [[]] CUSTOMER INSTALLABLE AVAILABLE UNTIL: Product's Support Life
CATEGORY:
[[]] ON-SITE
[x] SERVICE CENTER
AUTHOR: SM PRODUCT LINE: 24
ADDITIONAL INFORMATION:
© AGILENT TECHNOLOGIES, INC. 2003
PRINTED IN U.S.A.
February 12, 2003
Page 1 of 2
81110A-01B
S E R V I C E N O T E
Supersedes:
81110A-01A
81110A Pulse/Pattern Generator
Serial Numbers: DE38700473 / DE38702131
(All units with Firmware Rev. 01.02.00 or 01.10.00)
Keypad locks during manual operation
To Be Performed By: Agilent-Qualified Personnel
Parts Required:
P/N
Description
Qty.
n/a
2 MB SRAM memory card with Firmware Rev. 01.11.00
1
Page 2 of 2
81110A-01B
Situation:
While the unit is operated manually by using the instrument's keypad and pressing the keys very fast
after each other, the keypad may lock intermittently and keypad operation is only possible after
cycling the power.
Remote operation via the GP-IB interface is not affected.
Solution/Action:
The problem can be solved by a firmware update to Revision 01.11.00
The update will be done at Agilent Service Centers only, please contact Agilent for further
instructions.