INPUT MODULE GUIDE
2
8860
8861
This Guide describes the optional input mod-
ules, related cable connection procedures, and
their settings and specifications.
Notes for Installing the Input Units in Model 8861
When installing 4 or more Model 8946 and
Model 8958 at the same time
1
UNIT
4 5
8
1
UNIT
4 5
8
1
UNIT
4 5
8
OK
Do not install 4 units of Model 8946 to the UNIT 1-4 slots.
1
UNIT
4 5
8
4 units of Model 8946 can be
installed to the UNIT 1-4 slots.
Other than
Model 8958
Only a maximum of 3 units of Model 8946 can be
installed in the UNIT 1-4 slots.
When Model 8958 is not installed
OK
8946
8946
8958
8946
8946
8958
8946
8958
IMPORTANT: When Model 8958 16ch SCANNER UNIT and 4 or more units of Model 8946
4ch ANALOG UNIT are installed in Model 8861 at the same time, only a maximum of 3 units
of Model 8946 can be installed to the UNIT 1-4 slots of Model 8861.
Number of input units
1
UNIT
4 5
8
1
UNIT
4 5
8
5
6
7
8
1
2
3
4
Recommended positions for installing input units
Balanced installation of input units in the UNIT 1-4 and 5-8 slots is recommended in order to
effectively utilize the internal memory of Model 8861.
Number of input units
Please refer to the following flow chart to when installing input units:
As much as possible, install
an equal number of units to
the UNIT 1-4 and 5-8 slots.
Yes
Installation completed.
No
Up to 3 units of Model 8946
can be installed to the UNIT
1-4 slots.
As much as possible, install
an equal number of units to
the UNIT 1-4 and 5-8 slots.
Will model 8958 be installed?
Will 4 or more units Model 8946 be installed?
As much as possible, install
an equal number of units to
the UNIT 1-4 and 5-8 slots.
1
UNIT
4 5
8
1
UNIT
4 5
8
8958
1
UNIT
4 5
8
8958
8946
8946
Installation example
Yes
No
Installation example
Installation example
Structure of this Document.................................................................3
Chapter 1
Overview___________________________________ 5
1.1
Product Overview ................................................................5
1.2
Input Module Usage List .....................................................9
1.3
List of Input Modules, Cables, Probes and Clamp
Combinations ....................................................................11
Chapter 2
Connections _______________________________ 13
2.1
Installing Input Modules (Adding or Replacing) ................14
2.2
Connecting the Cables ......................................................16
2.2.1
Connection Preparations .......................................................16
2.2.2
Connecting to an Analog Unit
(Models 8936, 8938, 8946, 8956 and 8957) ..........................17
2.2.3
Connecting to the Model 8937 Voltage/Temp Unit ................18
2.2.4
Connecting to a Strain Unit (Models 8939 and 8960) ............20
2.2.5
Connecting to the Model 8940 F/V Unit .................................21
2.2.6
Connecting to the Model 8947 Charge Unit .........................24
2.2.7
Connecting to the Model 8958 16-Ch Scanner Unit ..............26
2.2.8
Connecting to the Model 8959 DC/RMS Unit ........................28
2.3
Connecting Clamps ...........................................................29
2.4
Connecting a Differential Probe ........................................31
2.5
Connecting Attenuating Probes ........................................35
2.6
Connecting Logic Probes ..................................................37
2.7
Supplying power from the Model 9687 Probe Power Unit 41
Chapter 3
Input Channel Settings ______________________ 43
3.1
Analog Unit Settings (Models 8936, 8946 and 8956) .......44
3.2
Model 8937 Voltage and Temperature Unit Settings ........45
3.2.1
Voltage Measurement ...........................................................45
3.2.2
Temperature Measurement ...................................................46
3.3
Model 8938 FFT Analog Unit Settings ..............................48
3.4
Strain Unit Settings (Models 8939 and 8960) ...................49
3.5
Model 8940 F/V Unit Settings ...........................................50
Measuring Frequency, 50/60 Hz and Rotation Rate ............. 50
3.5.2
Pulse Count Measurement ................................................... 52
3.5.3
Pulse Duty Measurement ...................................................... 53
3.5.4
Voltage Measurement ........................................................... 54
3.5.5
Current Measurement ........................................................... 55
3.6
Model 8947 Charge Unit Settings .....................................56
3.6.1
Voltage Measurement ........................................................... 56
3.6.2
Acceleration Measurement (Charge, Preamp) ..................... 57
3.7
Model 8957 High Resolution Unit Settings .......................59
3.8
Model 8958 16-Ch Scanner Unit Settings ........................60
3.8.1
Voltage Measurement ........................................................... 60
3.8.2
Temperature Measurement .................................................. 61
3.9
Model 8959 DC/RMS Unit Settings ..................................63
3.10 About Setting Contents .....................................................64
3.10.1
Measurement Range Setting ............................................... 64
3.10.2
Setting Input Coupling ........................................................... 65
3.10.3
Low-Pass Filter (LPF) Settings ............................................. 66
3.10.4
Anti-Aliasing Filter (AAF) Settings ......................................... 67
3.10.5
Digital Filter (Digital F) Settings ............................................ 67
3.10.6
Thermocouple (Sensor) Type Setting ................................... 68
3.10.7
Reference Junction Compensation Setting ........................... 68
3.10.8
Drift Correction Setting .......................................................... 69
3.10.9
Disconnect (Burn-Out) Detection Setting .............................. 69
3.10.10 Hold Setting .......................................................................... 70
3.10.11 Pull-Up Setting ...................................................................... 71
3.10.12 Threshold Setting .................................................................. 72
3.10.13 Response (Timing) Setting ................................................... 72
3.10.14 Sensor Sensitivity Setting ..................................................... 73
3.10.15 Probe Attenuation Selection ................................................. 75
3.10.16 Bridge Voltage Setting .......................................................... 75
3.10.17 Executing Zero Adjustment ................................................... 76
3.10.18 Executing Offset Cancellation ............................................... 76
3.10.19 Executing Auto-Balance ........................................................ 78
Chapter 4
Logic Input Settings_________________________79
4.1
Operation Overview ..........................................................79
4.2
Setting Procedure .............................................................80
Chapter 5
Specifications______________________________83
5.1
Logic Input Section ...........................................................83
5.1.1
Model 9327 Logic Probe ....................................................... 83
5.1.2
Model 9321-01 Logic Probe .................................................. 84
5.1.3
Model 9320-01 Logic Probe .................................................. 85
Analog Input Section .........................................................86
5.2.1
Model 8936 Analog Unit ........................................................86
5.2.2
Model 8937 Voltage/Temp Unit .............................................88
5.2.3
Model 8938 FFT Analog Unit .................................................90
5.2.4
Model 8939 Strain Unit ..........................................................91
5.2.5
Model 8940 F/V Unit ..............................................................92
5.2.6
Model 8946 4-Ch Analog Unit ...............................................94
5.2.7
Model 8947 Charge Unit ........................................................95
5.2.8
Model 8956 Analog Unit ........................................................97
5.2.9
Model 8957 High Resolution Unit ..........................................98
5.2.10
Model 8958 16-Ch Scanner Unit ...........................................99
5.2.11
Model 8959 DC/RMS Unit ...................................................101
5.2.12
Model 8960 Strain Unit ........................................................102
Index ________________________________ Index1
The following documents are provided with Model 8860 and 8861 Memory
HiCorders.
Refer to them as appropriate for your application.
Which input module and cables to use with the instrument depend on your mea-
surement application. Refer to this as appropriate for your application. (⇒ p. 3)
• In this document, the “instrument” means the Model 8860 or 8861 Memory
HiCorder.
• “Clamp” refers to one of our optional clamp-on sensor products.
Introduction
Document
Description
1 Quick Start Manual
Read this first.
It describes preparations for use, basic operating pro-
cedures and usage methods.
2 Input Module Guide
(This document)
To connect input modules and measurement
cables, and when making input channel set-
tings;
this Guide describes the optional input modules, relat-
ed cable connection procedures, and their settings
and specifications.
3 Instruction Manual
To obtain setting details;
this Manual describes details of the functions and op-
erations of the instrument, and its specifications.
4 Analysis Supplement
The supplement describes usage of the cal-
culation functions to analyze measurement
data.
The following symbols in this manual indicate the relative importance of cautions
and warnings.
Accuracy
We define measurement tolerances in terms of f.s. (full scale) values, with the
following meanings:
f.s. (maximum display value or scale length)
The maximum displayable value or scale length. This is usually the name of the
currently selected range.
In this instrument, the maximum displayable value is the range (V/div) times the
number of divisions (20) on the vertical axis.
Example: For the 1 V/div range, f.s. = 20 V
Safety Symbols
In the manual, the
symbol indicates particularly important informa-
tion that the user should read before using the instrument.
The
symbol printed on the instrument indicates that the user
should refer to a corresponding topic in the manual (marked with the
symbol) before using the relevant function.
Indicates DC (Direct Current).
Indicates AC (Alternating Current).
Indicates both DC (Direct Current) and AC (Alternating Current).
Indicates a grounding terminal.
Indicates that incorrect operation presents an extreme hazard
that could result in serious injury or death to the user.
Indicates that incorrect operation presents a significant hazard
that could result in serious injury or death to the user.
Indicates that incorrect operation presents a possibility of
injury to the user or damage to the instrument.
Indicates advisory items related to performance or correct
operation of the instrument.
Other Symbols
Indicates the prohibited action.
(⇒ p. )
Indicates the location of reference information.
Indicates quick references for operation and remedies for troubleshoot-
ing.
*
Indicates that descriptive information is provided below.
[ ]
Screen labels such as menu items, page titles, setting items, dialog titles
and buttons are indicated by square brackets [ ].
CURSOR
(Bold
characters)
Bold characters within the text indicate operating key labels.
Structure of this Document
3
Structure of this Document
Preparation
Reference Location
Overview of Input
Modules and Logic
Groups
About the overview of input mod-
ules and logic groups
"1.1 Product Overview" (⇒ p. 5)
Overview of usage, connections
and settings
"1.2 Input Module Usage List" (⇒ p. 9)
Input module connections
"1.3 List of Input Modules, Cables, Probes
and Clamp Combinations" (⇒ p. 11)
Specifications of input modules
and logic probes
"Chapter 5 Specifications" (⇒ p. 83)
Procedure
Reference Location
Connections
Connecting the input modules,
cables and logic probes to the
instrument
"Chapter 2 Connections" (⇒ p. 13)
Refer to the Quick Start Manual for other connections.
Turn Power On
Instrument Settings
(input channel
settings)
(When using an input module)
Set analog channels
"Chapter 3 Input Channel Settings" (⇒ p. 43)
(When using a logic probe)
Set logic channels
"Chapter 4 Logic Input Settings" (⇒ p. 79)
In addition to the above, perform necessary measurement settings such as measure-
ment criteria and trigger settings.
Refer to the Quick Start Manual for overviews of all operations.
Refer to the Instruction Manual for details of settings other than those related to input
channels.
Measure
Structure of this Document
4
Maximum Input Voltage*1 of input module and Maximum Rated Voltage
to Ground*2
This Instrument
Input Module
GND
400 VDC max
370 VAC/DC
H
L
370 VAC/DC
Model 8936 Analog Unit
Model 8938 FFT Analog Unit
Model 8959 DC/RMS Unit
*2
*2
*1
This Instrument
Input Module
GND
400 VDC max
300 VAC/DC
H
L
300 VAC/DC
Model 8956 Analog Unit
Model 8957 High Resolution
Unit
*1
*2
*2
This Instrument
Input Module
GND
30 Vrms or 60 VDC
30 Vrms or 60 VDC
H
L
30 Vrms or 60 VDC
Model 8937 Voltage/Temp
Unit
Model 8946 4-Ch Analog
Unit
*2
*2
*1
This Instrument
Input Module
GND
10 VDC max
30 Vrms or 60 VDC
H
L
30 Vrms or 60 VDC
Model 8939 Strain Unit
*2
*2
*1
This Instrument
Input Module
GND
DC10 V max
33 Vrms or 70 VDC
H
L
33 Vrms or 70 VDC
Model 8960 Strain Unit
*2
*2
*1
This Instrument
Input Module
GND
30 Vrms or 60 VDC
30 Vrms or 60 VDC
H
L
30 Vrms or 60 VDC
30 Vrms or 60 VDC
H
L
Connectors
BNC
Model 8940 F/V Unit
*2
*2
*1
*2
*1
This Instrument
Input Module
GND
Maximum applied elec-
tric charge (⇒ p. 95)
30 Vrms or 60 VDC
H
L
30 Vrms or 60 VDC
30 Vrms or 60 VDC
H
L
BNC
Miniature
Connector
Common
Model 8947 Charge Unit
*2
*1
*2
This Instrument
Input Module
GND
40 VDC max
33 Vrms or 70 VDC
+
−
33 Vrms or 70 VDC
Model 8958 16-Ch Scanner
Unit
*2
*1
*2
1.1 Product Overview
5
This chapter provides an overview of the optional input modules and input cables that can be used with this
instrument.
Refer to "Appendix 6 Disposing of the Instrument" in the Instruction Manual for a full list of options for this
instrument.
Also refer to "Maximum Input Voltage*1 of input module and Maximum Rated Voltage to Ground*2" (⇒ p. 4)
Overview
Chapter 1
1.1
Product Overview
Voltage Measurement
Input Module
8956 Analog Unit
8957 High Resolution Unit
8936 Analog Unit
8946 4-Ch Analog Unit
2 channels, 20 MS/s, 12-bit,
maximum input voltage: 400
VDC
High-speed type
2 channels, 2 MS/s, 16-bit,
maximum input voltage: 400
VDC
Supports high-precision mea-
surements. (Includes internal
anti-aliasing filter.)
2 channels, 1 MS/s, 12-bit,
maximum input voltage: 400
VDC
4 channels, 1 MS/s, 12-bit,
maximum input voltage: 30
Vrms/60 VDC
8938 FFT Analog Unit
8959 DC/RMS Unit
8940 F/V Unit
8937 Voltage/Temp Unit
2 channels, 1 MS/s, 12-bit,
maximum input voltage: 400
VDC
(Includes internal anti-alias-
ing filter.)
2 channels, 1 MS/s, 12-bit,
maximum input voltage: 400
VDC
Measures both normal and
RMS voltages.
2 channels, 1 MS/s, 12-bit,
maximum input voltage: 30
Vrms/60 VDC
In addition to voltage, also
measures current, frequen-
cy, rotation rate, pulse total-
ization and duty cycle.
2 channels, 1 MS/s, 12-bit,
maximum input voltage: 30
Vrms/60 VDC
Measures temperature in
addition to voltage.
8958 16-Ch Scanner Unit
8947 Charge Unit
16 channels, 20 S/s, 16-bit,
maximum input voltage: 40
VDC
In addition to voltage, also
supports temperature mea-
surement.
2 channels, 1 MS/s, 12-bit,
maximum input voltage: 30
Vrms/60 VDC
In addition to voltage, also
supports measurements with
piezoelectric
acceleration
sensors. (Includes internal
anti-aliasing filter.)
1.1 Product Overview
6
Refer to "1.3 List of Input Modules, Cables, Probes and Clamp Combinations" (⇒ p. 11) for combinations of cables and
probes to connect to an input module.
Input cables for voltage measurement with input modules
(except the Model 8958 16-Ch Scanner Unit)
9197 Connection Cord
9198 Connection Cord
9217 Connection Cord
For high voltage, maximum input volt-
age: 500 V (large alligator clips)
For low voltage, maximum input volt-
age: 300 V (small alligator clips)
Maximum input voltage: 300 V (for
BNC output)
9322 Differential Probe
9665 10:1 Probe
9666 100:1 Probe
For high voltage, maximum input volt-
age: (CAT II): 2000 VDC, 1000 VAC,
(CAT III): 600 VAC/DC
Applicable Modules:
• Model 8940 F/V Unit
• Input
modules
for
voltage
measurement (except the Model
8958 16-Ch Scanner Unit)
To connect to an input module, the fol-
lowing separate items are required:
• Model 9325 Power Cord
(when using the Model 8940)
• Model 9418-15 AC Adapter
• Model 9248 Power Cord
(when using the Model 9687)
Maximum input voltage: 1 kVrms
(1 MHz or less)
Applicable Modules:
• Model 8936 Analog Unit
• Model 8938 FFT Analog Unit
• Model 8956 Analog Unit
• Model 8957 High Resolution Unit
• Model 8959 DC/RMS Unit
Maximum input voltage: 5 kVpeak
(1 MHz or less)
Applicable Modules:
• Model 8936 Analog Unit
• Model 8938 FFT Analog Unit
• Model 8956 Analog Unit
• Model 8957 High Resolution Unit
• Model 8959 DC/RMS Unit
Temperature Measurement (Thermocouple Inputs)
Input Module
8937 Voltage/Temp Unit
8958 16-Ch Scanner Unit
2 channels, 4 kS/s, 12-bit
Thermocouple Types: K, E, J, T, N, R,
S, B
In addition to temperature, also sup-
ports voltage measurement.
16 channels, 20 S/s, 16-bit
Thermocouple Types: K, E, J, T, N, R,
S, B, W
In addition to temperature, also sup-
ports voltage measurement.
1.1 Product Overview
7
*1. Except for Models 9018-10 and 9132-10, a separate conversion cable or power supply is required when connecting to
an input module. Refer to "1.3 List of Input Modules, Cables, Probes and Clamp Combinations" (⇒ p. 11) for details.
*2. Up to 15 A when combined with the Model 8940 F/V Unit.
*3. Not applicable to CE Marking
Current Measurement
Input Module
Input modules for voltage measurement
(except the Model 8958 16-Ch Scanner Unit)
8940 F/V Unit
Current can be measured using a clamp.
2 channels, 1 MS/s, 12-bit, clamp input
In addition to current, also measures voltage,
frequency, rotation rate, pulse totalization and
duty cycle.
Refer to "Voltage Measurement" (⇒ p. 5) for in-
put modules.
Measures in [Voltage] mode.
Clamps compatible with the above input modules.
9018-10 Clamp-On Probe
9132-10 Clamp-On Probe*3
3273 Clamp-On Probe *1
3273-50 Clamp-On Probe*1
3274 Clamp-On Probe *1
3275 Clamp-On Probe *1
3276 Clamp-On Probe *1
9270 Clamp-On Sensor *1, *3
9271 Clamp-On Sensor *1, *3
9272 Clamp-On Sensor *1, *3
For AC
Outputs 0.2 V AC waveform of com-
mercial power line current.
9018-10: 10 to 500 A, 40 Hz to 3 kHz
9132-10: 20 to 1000 A, 40 Hz to 1 kHz
AC/DC
Supports high-accuracy observation
over broad ranges from DC to MHz.
3273: 15 A, DC to 50 MHz
3273-50: 30 A*2, DC to 50 MHz
3274: 150 A, DC to 10 MHz
3275: 500 A, DC to 2 MHz
3276: 30 A, DC to 100 MHz
For AC
Supports inverter secondary current
measurement.
9270: 20 A, 5 Hz to 50 kHz
9271: 200 A, 5 Hz to 50 kHz
9272: 20/200 A, 5 Hz to 10 kHz
For users of the following legacy products:
• Model 9018 Clamp-On Probe
• Model 9132 Clamp-On Probe
Usable with this instrument by connecting the Model 9199 Conversion
Adapter.
For measuring leakage current:
The Model 9657-10 Clamp-On Leak Sensor can also be used.
9277 Universal Clamp-On CT *1
9278 Universal Clamp-On CT *1
9279 Universal Clamp-On CT *1,*3
AC/DC
Low zero drift supports stable long-
term measurements.
9277: 20 A, DC to 100 kHz
9278: 200 A, DC to 100 kHz
9279: 500 A, DC to 20 kHz
9018-10
9132-10
3273 and 3276
3274 and 3275
9270 and 9271
9272
9277
9278
9279
1.1 Product Overview
8
Frequency, Rotation Rate, Commercial Line
Frequency (50/60 Hz), Pulse Count,
Pulse Duty Measurement
Acceleration Measurement
Input Module
8940 F/V Unit
8947 Charge Unit
2 channels, 1 MS/s, 12-bit
• Frequency: can be measured from the input pulse corresponding to
the measurement waveform.
• Rotation Rate: can be measured from the input pulse correspond-
ing to the measurement waveform.
• 50/60 Hz: Frequencies near 50/60 Hz can be measured from the
input pulse corresponding to the measurement waveform.
• Totalization: cumulative count of input pulses.
• Pulse Duty: measured as the percentage of High level of a single
pulse waveform.
2 channels, 1 MS/s, 12-bit
• Electric Charge: can be measured using a
voltage-output type acceleration pick-up
sensor.
• Preamp: can be measured using an accel-
eration
pick-up
sensor
with
built-in
preamp.
Also supports voltage measurement.
Electric Charge, Pressure, Torque,
Displacement Measurement
Input Module
8939 Strain Unit
8960 Strain Unit
2 channels, 1 MS/s, 12-bit
Vibration and displacement can be
measured using a strain gauge trans-
ducer. (conversion cable supplied)
2 channels, 200 kS/s, 16-bit
Vibration and displacement can be
measured using a strain gauge trans-
ducer. Supports high-precision mea-
surements. (Includes internal anti-
aliasing filter.) (conversion cable sup-
plied)
Logic Signal Measurement
Logic probes that can be connected to this instrument
9320-01 Logic Probe
9321-01 Logic Probe
9327 Logic Probe
Measures digital signals and on/off
switching of non-voltage contacts.
Detects the presence of AC and DC
voltages. Measures activation timing
of relay sequencing circuits.
Maximum input voltage: 250 Vrms
(HIGH range)
Measures faster signals than the Mod-
el 9320-01 Logic Probe.
For users of the following legacy products:
Models 9306, 9307, 9320, 9321 Logic Probes
Usable with this instrument by connecting the Model 9323 Conversion Cable.
1.2 Input Module Usage List
9
Voltage Measurement
Temperature Measurement
1.2
Input Module Usage List
To Perform This Measurement
Recommended Input
Module
Use to connect
Connection
Procedure
Setting
Procedure
Up to 400 V
Up to 2 channels per
module
Model 8936 Analog Unit
Model 9197 Connection
Cord (Up to 500 V)
Model 9198 Connection
Cord (Up to 300 V)
Model 9217 Connection
Cord (Up to 300 V)
(⇒ p. 17)
(⇒ p. 44)
For faster sampling
Model 8956 Analog Unit
(⇒ p. 17)
(⇒ p. 44)
To see high-precision
voltage values
Model 8957 High Resolu-
tion Unit
(⇒ p. 17)
(⇒ p. 59)
To perform FFT analysis
Model 8957 High Resolu-
tion Unit
(⇒ p. 17)
(⇒ p. 59)
Model 8938 FFT Analog
Unit
(⇒ p. 17)
(⇒ p. 48)
To also see RMS voltage
Model 8959 DC/RMS Unit
(⇒ p. 28)
(⇒ p. 63)
Up to 30 Vrms or
60 V DC
For additional channels
(four)
Model 8946 4-Ch Analog
Unit
(⇒ p. 17)
(⇒ p. 44)
To measure with high
sensitivity (500 µV/div)
Model 8937 Voltage/
Temp Unit
(⇒ p. 18)
(⇒ p. 45)
Model 8940 F/V Unit
(⇒ p. 21)
(⇒ p. 54)
Model 8947 Charge Unit
(⇒ p. 24)
(⇒ p. 56)
Up to 40 V
When slow sampling is
acceptable, but more
channels desired
Model 8958 16-Ch Scan-
ner Unit
Input Cable
(⇒ p. 26)
(⇒ p. 60)
When measured
voltage exceeds
the maximum in-
put for the Input
module
Up to 2000 V DC or 1000
V AC (depending on mea-
surement site (⇒ p. 31) )
Above Input Modules (ex-
cept the Model 8958 16-
Ch Scanner Unit)
Model 9322 Differential
Probe *1
(⇒ p. 31)
(⇒ p. 75)
Up to 1000 Vrms
(@1 MHz max.)
Model 8936 Analog Unit
Model 8956 Analog Unit
Model 8957 High Resolu-
tion Unit
Model 8938 FFT Analog
Unit
Model 8959 DC/RMS Unit
Model 9665 10:1 Probe
*2
(⇒ p. 35)
(⇒ p. 75)
Up to 5000 Vpeak
(@1 MHz max.)
Model 9666 100:1 Probe
*2
(⇒ p. 35)
(⇒ p. 75)
Overview of the above input modules (⇒ p. 5), description of cables (⇒ p. 6)
Procedures for settings unrelated to input channels are the same. Refer to the Instruction Manual for details.
*1. Voltage to ground complies with Model 9322 specifications. (⇒ p. 31)
*2. Voltage to ground complies with the specifications of the input module used. (⇒ p. 86)
To Perform This
Measurement
Recommended Input
Module
Use to connect
Connection
Procedure
Setting
Procedure
Remarks
To measure temperature
Model 8937 Voltage/
Temp Unit
Thermocouple
(⇒ p. 18)
(⇒ p. 45)
Up to 2 channels can be
measured per module.
To measure multi-channel
temperature
Model 8958 16-Ch Scan-
ner Unit
Thermocouple
(⇒ p. 26)
(⇒ p. 60)
Up to 16 channels can be
measured per module.
Overview of the above input modules (⇒ p. 6)
Procedures for settings unrelated to input channels are the same. Refer to the Instruction Manual for details.
1.2 Input Module Usage List
10
Current Measurement
Other Measurements
Logic Measurements
To Perform This
Measurement
Recommended
Input Module
Use to connect
Connection
Procedure
Setting
Procedure
Remarks
To read current values di-
rectly without complicated
settings such as scaling
Model 8940 F/V
Unit
Models 3273/ 3273-50
Clamp-On Probe
Models 9270 to 9272
Clamp-On Sensor
Models 9277 to 9279
Universal Clamp-On CT
(⇒ p. 21)
(⇒ p. 29)
(⇒ p. 50)
A conversion cable is
required for
connection (⇒ p. 11).
To measure using a
voltage measurement
module
Voltage Measure-
ment Input Module
Models 9018-10/ 9132-10
Clamp-On Probe
Models 3273 to 3276
Clamp-On Probe
Models 9270 to 9272
Clamp-On Sensor
Models 9277 to 9279
Universal Clamp-On CT
(⇒ p. 29)
Refer to the
voltage
measurement
description for
each module
Scaling is required.
Depending on the
clamp, a power supply
or sensor unit may be
required for
connection (⇒ p. 11).
Overview of the above input modules (⇒ p. 7), about Clamps (⇒ p. 7)
Procedures for settings unrelated to input channels are the same. Refer to the Instruction Manual for details.
To Perform This Measurement
Recommended Input Module
Use to connect
Connection
Procedure
Setting
Procedure
To measure electric charge, accelera-
tion, pressure, torque or displacement
using a strain sensor.
Model 8939 Strain Unit
Model 8960 Strain Unit
Conversion Cable (sup-
plied) Strain Gauge
Transducer
(⇒ p. 20)
(⇒ p. 49)
To measure acceleration
Model 8947 Charge Unit
Piezoelectric Sensor
(⇒ p. 24)
(⇒ p. 57)
To measure frequency
Model 8940 F/V Unit
Models
9198 Connection Cord
9217 Connection Cord
(⇒ p. 21)
(⇒ p. 50)
To measure pulse counts (totalization),
pulse duty and similar values
Overview of the above input modules (⇒ p. 8)
Procedures for settings unrelated to input channels are the same. Refer to the Instruction Manual for details.
To Perform This Measurement
Recommended Probe
Connection Procedure
Setting
Procedure
To measure digital signals and on/off
switching of non-voltage contacts
Model 9320-01 Logic Probe
(⇒ p. 37), (⇒ p. 39)
(⇒ p. 79)
To measure larger signals than the above
Model 9327 Logic Probe
(⇒ p. 37), (⇒ p. 39)
To measure the presence or absence of
AC or DC voltage
Model 9321-01 Logic Probe
(⇒ p. 37), (⇒ p. 40)
1.3 List of Input Modules, Cables, Probes and Clamp Combinations
11
O = Compatible, − = Incompatible, Δ = Compatible, but scaling required
1.3
List of Input Modules, Cables, Probes and
Clamp Combinations
Measurement
Parameter
Use to connect
Input Module
8936 8937 8938 8939 8940 8946 8947 8956 8957 8958 8959 8960
Voltage
Connection
Cables
9197
O
O *1
O
−
O *1 O *1 O *1
O
O
−
O
−
9198
O
O
O
−
O
O
O
O
O
−
O
−
9217
O
O
O
−
O
O
O
O
O
−
O
−
Differential
Probe
9322
O *3 O *3 O *3
−
O
*2,*9
O *3 O *3 O *3 O *3
−
O *3
−
Attenuating
Probes
9665
9666
O
−
O
−
−
−
−
O
O
−
O
−
Input Cable
−
−
−
−
−
−
−
−
−
O
−
−
RMS Values Same as above
−
−
−
−
−
−
−
−
−
−
O
−
FFT Analysis
(with AAF
installed)
Same as above
−
−
O
−
−
−
O
−
O
−
−
−
Current
Clamp-On
Sensors *8
9270 *10
9271 *10
9272 *10
Δ*6
Δ*6
Δ*6
−
O
*4,*9
Δ*6
Δ*6
Δ*6
Δ*6
−
Δ*6
−
Universal
Clamp-On
CTs *8
9277
9278
9279 *10
Δ*6
Δ*6
Δ*6
−
O
*4,*9
Δ*6
Δ*6
Δ*6
Δ*6
−
Δ*6
−
Clamp-On
Probes *8
3273
3273-50
Δ*7
Δ*7
Δ*7
−
O
*5,*9
Δ*7
Δ*7
Δ*7
Δ*7
−
Δ*7
−
3274
3275
3276
Δ*7
9018-10
Δ
Δ
Δ
−
Δ
Δ
Δ
Δ
Δ
−
Δ
−
9132-10*10
Δ
Δ
Δ
−
Δ
Δ
Δ
Δ
Δ
−
Δ
−
Acceleration
Piezoelectric Sensor
−
−
−
−
−
−
O
−
−
−
−
−
Temperature
Thermocouple
−
O
−
−
−
−
−
−
−
O
−
−
Frequency,
Pulse
Totalization,
Pulse Duty
Connection Cable
(Sensor)
−
−
−
−
O
−
−
−
−
−
−
−
Electric Charge,
Acceleration,
Pressure,
Torque,
Displacement
Strain Gauge
Transducer
−
−
−
O
−
−
−
−
−
−
−
O
*1. Although compatible, the 9198 is recommended instead.
To connect to the input module, the following separate items
are required:
*2. Model 9325 Power Cord , 9418-15 AC Adapter, or 9248
Power Cord (when using the Model 9687)
*3. Model 9418-15 AC Adapter or 9248 Power Cord (when
using the Model 9687)
*4. Model 9318 Conversion Cable or 9555 Sensor Unit
*5. Model 9319 Conversion Cable, 3272 Power Supply, 3269
Power Supply, or 9687 Probe Power Unit
*6. Model 9555 Sensor Unit*9
*7. Model 3272 , 3269 Power Supply, or 9687 Probe Power
Unit
*8. Set the instrument to [Voltage] measurement mode
when using combinations that do not include the
Model 8940, and with the combination of the 8940 and
Model 9018-10 or 9132-10. (⇒ p. 54)
*9. Up to six units can be used with the 9325
Up to eight clamps can be used
When the 9325 and clamps are used at the same
time, up to eight can be used
*10. Not applicable to CE Marking
About Scaling settings: Voltage acquired from the sensor
is converted to the corresponding physical measurement
units for display.
See "5.4 Converting Input Values (Scaling Function)" in
the Instruction Manual
This chapter describes the installation and connection of input modules to the
instrument. Refer to the Quick Start Manual for other connections.
1
Install the input module(s) in the instrument’s input module compartment.
(⇒ p. 14)
2
Connect the cables and sensors to the input module(s).
Connection items differ according to your application. Refer to the description of each input
module.
Also refer to the following sections when measuring voltage and current:
• Using Connection Cables (⇒ p. 16) (Voltage measurement)
• Using Differential Probes (⇒ p. 31) (Voltage measurement)
• Using Attenuating Probes (⇒ p. 35) (Voltage measurement)
• Using Clamps (⇒ p. 29) (Current measurement)
• Supplying the power from the Model 9687 Probe Power Unit (⇒ p. 41)
3
After making connections, make instrument settings (⇒ p. 43).
This Guide describes only the procedures for setting the input channels of each input mod-
ule.
Refer to the Quick Start Manual for an overview of all settings, and to the Instruction
Manual for the details of each setting.
1
Connect the logic probes to the LOGIC receptacles on the instrument
(⇒ p. 37).
2
After connecting, perform settings on this instrument (⇒ p. 79).
Connections
Chapter 2
Input modules for measurement (analog inputs)
• Model 8936 Analog Unit
(⇒ p. 17)
• Model 8937 Voltage/Temp
Unit
(⇒ p. 18)
• Model 8938 FFT Analog Unit (⇒ p. 17)
• Model 8939 Strain Unit
(⇒ p. 20)
• Model 8940 F/V Unit
(⇒ p. 21)
• Model 8946 4-Ch Analog Unit (⇒ p. 17)
• Model 8947 Charge Unit
(⇒ p. 24)
• Model 8956 Analog Unit
(⇒ p. 17)
• Model 8957 High Resolution
Unit
(⇒ p. 17)
• Model 8958 16-Ch Scanner
Unit
(⇒ p. 26)
• Model 8959 DC/RMS Unit
(⇒ p. 28)
• Model 8960 Strain Unit
(⇒ p. 20)
Measurements with logic probes (logic inputs)
2.1 Installing Input Modules (Adding or Replacing)
14
Input modules specified at the time the instrument is ordered are supplied prein-
stalled. Use the following procedures to add or replace input modules, or to
remove them from the instrument.
2.1
Installing Input Modules (Adding or
Replacing)
Preparations
• To avoid electric shock accident, before removing or replacing an input
module, confirm that the instrument is turned off and that the connection
cords are disconnected.
• The mounting screws must be firmly tightened or the input module may
not perform to specifications, or may even fail.
When an input module is not used
• To avoid the danger of electric shock, never operate the instrument with
an input module removed. To use the instrument after removing an input
module, install a blank panel over the opening of the removed module.
To avoid damaging input modules, do not touch the input module connectors on
this instrument.
Using the Model 8958 16-Ch Scanner Unit
In the following cases, the Model 8958 must be adjusted before use with this
instrument.
See "Scanner Unit Zero Position Adjustment" (⇒ p. 27)
• When installing the Model 8958 16-Ch Scanner Unit in this instrument
• When the zero position has shifted due to aging or environmental changes
When using the Model 8958 16-Ch Scanner Unit together with four Model 8946
4-Ch Analog Units at the same time, the 8946s cannot all be installed in UNIT
locations 1 to 4 in the Model 8861.
In this case, no more than three 8946s can be installed in UNIT locations 1 to 4.
Adding an Input Module
Adding an input module does not affect the Sheet Settings screen.
To display waveforms from an added input module, select the corresponding dis-
play channel on the Sheet Settings screen. Even when waveform display is
enabled ([On]), a waveform is not displayed unless it has been enabled for dis-
play on the Sheet Settings screen.
See "7.2.6 Assigning Display Channels to Graphs (Analog Channels)" in the Instruction
Manual
1
UNIT
4 5
8
8958
1
UNIT
4 5
8
1
UNIT
4 5
8
Installation examples (installing four Model 8946s, one 8958 and either additional
input modules or blank panels)
8946
8946
8946
8946
8958 8946
8958
OK
OK
2.1 Installing Input Modules (Adding or Replacing)
15
Required item: One Phillips-head screwdriver
1
Turn the instrument’s POWER switch Off.
2
With attention to the orientation of the input module,
insert it firmly all the way in.
Make certain that the labels on the input module’s panel face
the same direction as the labels on the right side of the instru-
ment.
3
Using the Phillips screwdriver, tighten the two input
module mounting screws.
When the instrument is positioned vertically as illustrated,
module number one is at the left, and channel number one is
at the bottom.
Information about the input modules installed in the instru-
ment can be verified on the initial screen that appears when
power is turned on, and in the [Config] menu on the System
screen.
Channel Configuration
Module No. (UNIT)
Channel
(CH)
Input Module
Compartment
UNIT1
CH1
CH2
CH3
CH4
UNIT3
UNIT2
UNIT4
CH1
CH2
8947
8946
CH1
CH4
CH16
8958
Installing an input module
Right Side
Removing an input module
1
Turn the instrument’s POWER switch Off.
2
Remove any cables or thermocouples connected to
the input module.
3
Remove the power cord.
4
Using the Phillips screwdriver, loosen the two input
module mounting screws.
5
Grasp the handle and pull the module out.
If the input module has no handle, pull it out by grasp-
ing the connectors.
Right Side
Required item: One Phillips-head screwdriver
Handle
(Example: 8956)
Blank panel
Install a blank panel.
Using the Phillips screwdriver, tighten the two mount-
ing screws.
Measurements made without a blank panel
installed may fail to meet specifications because
of temperature instability within the instrument.
If not installing another input module after removal
Right Side
2.2 Connecting the Cables
16
Read the following carefully before making connections.
For voltage measurement
Using cables with BNC connectors
When using the Model 8958 16-Ch Scanner Unit together with other
input modules
2.2
Connecting the Cables
2.2.1
Connection Preparations
About maximum input ratings of input modules and cables
• The maximum input voltage is the lesser rating of either the input module
or the cable to be used. Do not measure if this voltage would be
exceeded, as damage the instrument or a serious accident may result.
Example: When connecting the Model 9197 Connection Cord (500 Vmax)
to the Model 8936 Analog Unit (400 Vmax), the maximum input voltage for
measurement is 400 V DC.
For details of maximum input voltages, refer to "1.1 Product Overview"
(⇒ p. 5) and "Chapter 5 Specifications" (⇒ p. 83).
• When connecting clip-type test leads to live terminals, be very careful to
avoid accidentally shorting conductors together and causing a serious
accident.
• When measuring voltage in a power line, connection cables should only
be connected to the secondary side of a breaker, so the breaker can pre-
vent an accident if a short circuit occurs. Connections should never be
made to the primary side of a breaker, because unrestricted current flow
could cause a serious accident if a short circuit occurs.
• Do not permanently connect the device in an environment where voltage
surges exceeding the maximum input voltage may occur. Failure to
observe this precaution could result in damage to the device and per-
sonal injury.
• Do not use cables other than those specified by Hioki. The specified
cables include insulated BNC plugs to avoid electric shock. An uninsu-
lated BNC plug may cause electric shock or damage to the BNC jack.
• Do not connect a cable to the instrument while it is connected to the
object to be measured. Otherwise, an electric shock accident may occur.
Use only the specified connection cords. Using a non-specified cable may result
in incorrect measurements due to poor connection or other reasons.
When disconnecting the BNC connector, be sure to release the lock before pull-
ing off the connector. Forcibly pulling the connector without releasing the lock, or
pulling on the cable, can damage the connector.
Do not attempt to measure the same signal with the 8958 and another input
module at the same time. Channel switching noise from the 8958 may affect
the other input module.
2.2 Connecting the Cables
17
Be sure to read "2.2.1 Connection Preparations" (⇒ p. 16) before connecting.
2.2.2
Connecting to an Analog Unit
(Models 8936, 8938, 8946, 8956 and 8957)
Input modules
• Model 8936 Analog Unit
• Model 8956 Analog Unit
• Model 8946 4-Ch Analog Unit *1
• Model 8938 FFT Analog Unit
• Model 8957 High Resolution Unit
Maximum Input Voltage (⇒ p. 4)
Any of the following connects to a BNC jack on an input
module.
• Model 9197 Connection Cord
Maximum input voltage:
500 V (large alligator clips)
• Model 9198 Connection Cord
Maximum input voltage:
300 V (small alligator clips)
• Model 9217 Connection Cord
Maximum input voltage:
300 V (for BNC output)
If the voltage to be measured exceeds the maximum input
rating of the input module being used:
• Model 9665 10:1 Probe*2 (⇒ p. 35)
• Model 9666 100:1 Probe*2 (⇒ p. 35)
• Model 9322 Differential Probe*3 (⇒ p. 31)
*1. Use Model 9198 Connection Cords for
the Model 8946 4-Ch Analog Unit.
*2. Incompatible with the Model 8946 4-
Ch Analog Unit.
*3. The Model 9418-15 AC Adapter option
is required for connection (⇒ p. 33).
1
Connect the BNC plug on the cable to a
BNC jack on the input module.
Align the slots in the BNC plug with the
guide pins on the jack on the input module,
then push and twist the plug clockwise until
it locks.
2
Connect the cable clips to the measure-
ment object.
Disconnecting BNC connectors
Push the BNC plug, twist it counterclock-
wise, and pull it out.
Example: Model 8936 Analog Unit
Required item: One of the above cables
BNC jack
Connection Procedure
Lock
Input module connec-
tor guide pins
BNC plug slots
Connect to the measurement object
Connecting the cable
2.2 Connecting the Cables
18
For voltage measurement, be sure to read "2.2.1 Connection Preparations" (⇒ p. 16) before
connecting.
About Connections and Inputs
When measuring temperature
2.2.3
Connecting to the Model 8937 Voltage/Temp
Unit
Input module
Model 8937 Voltage/Temp Unit
Maximum Input Voltage (⇒ p. 4)
Voltage Measurement:
Any of the following connects to a BNC jack on an input
module.
Temperature Measurement:
Connect the thermocouple to the terminal block.
(Recommended cables (⇒ p. 19))
• Model 9198 Connection Cord
Maximum input voltage:
300 V (small alligator clips)
• Model 9217 Connection Cord
Maximum input voltage:
300 V (for BNC output)
If the voltage to be measured exceeds the maximum input
rating of the input module being used:
• Model 9322 Differential Probe(⇒ p. 31) *
∗The Model 9418-15 AC Adapter option
is required for connection (⇒ p. 33).
The ground (GND) sides of the voltage and temperature inputs for each channel
are common. Do not connect and apply both voltage and temperature inputs at
the same time. The measurement object could be damaged.
Voltage and temperature cannot both be measured simultaneously on one chan-
nel.
Observe the following to avoid electric shock.
• When measuring temperature with an uninsulated thermocouple at a
point that has a non-zero electric potential, be careful to avoid touching
the input module’s terminal block. The terminals may have voltage
present.
• The voltage and temperature inputs are electrically isolated from one
another.
• The push-button terminals on the Model 8937 Voltage/Temp Unit are for ther-
mocouples only. Do not connect anything except a specified thermocouple
(type K, J, E, T, N, R, S or B).
• Use a tool (screwdriver) to connect and disconnect thermocouples.
• Confirm the polarity of a thermocouple before connection. Correct values are
not displayed if the thermocouple is connected with reverse polarity.
2.2 Connecting the Cables
19
1
Connect the BNC plug on the cable to a
BNC jack on the input module.
Align the slots in the BNC plug with the guide
pins on the jack on the input module, then push
and twist the plug clockwise until it locks.
2
Connect the cable clips to the measure-
ment object.
Disconnecting BNC connectors
Push the BNC plug, twist it counterclock-
wise, and pull it out.
Required item: Connection cord
BNC jack
Connection Procedure: Voltage Measurement
Lock
Input module
connector guide pins
BNC plug slots
Connect to the measurement object
Connecting the cable
1
Strip insulation from the thermocouple
wires as shown at the left.
Stripping length: approx. 10 mm
2
Push the blade of a flat screwdriver into the
button on the terminal block of the input
module.
3
Insert each thermocouple wire into the
appropriate terminal hole while pressing
the button.
Confirm proper polarity.
4
Release the button.
The thermocouple is connected.
5
Attach to the measurement object.
To remove the thermocouple
Hold the button while pulling the thermo-
couple wire out.
Terminal Block
Connection Holes
Required item:
Thermocouple, flat-blade screwdriver (2.6-mm blade)
Recommended wire:
Compatible wire: Single-strand thermocouple wire,
0.4 to 1.2-mm diameter
Stripping length: 10 mm
Connection Procedure: Temperature Measurement
Outer Insulation
10 mm
25 mm
Thermocouple
single wires
Inner insulation
Connect to the measurement object
Connecting a Thermocouple
2
1
3
4
5
2.2 Connecting the Cables
20
2.2.4
Connecting to a Strain Unit (Models 8939
and 8960)
Input module
Model 8939 Strain Unit
(conversion cable supplied)
Maximum Input Voltage (⇒ p. 4)
Connect the strain gauge converter* to the input module
jack.
(Depending on the sensor, the supplied conversion cable
may be needed.)
A bridge box* is required for measurement using a strain
gauge.*
∗ The strain gauge converter, strain gauge and bridge box may be
common types available in the market.
When using a conversion cable
• Do not connect a sensor other than the Model 8939 Strain Unit to the supplied
conversion cable.
• When removing the conversion cable, be sure to release the lock before pull-
ing the plug out.
Connect to the measurement object
Connecting the Conversion Cable
1
Connect the supplied conversion cable to
a receptacle on the input module.
Align the tooth on the inside of the receptacle
with the slot in the conversion cable plug, and
insert the plug. Insert it all the way to the end of
the slot in the plug ring (shaded part in the illus-
tration at the left). When inserted, twist the ring
clockwise to lock.
2
Connect the strain gauge transducer to
the conversion cable.
3
Attach to the measurement object.
To remove the conversion cable
Push and twist the plug ring counterclock-
wise to unlock it, then pull the plug out.
Example: Connecting the strain gauge converter with the supplied conversion cable
Required item:
Conversion Cable (supplied), Strain Gauge Converter
Connection Procedure: Vibration and displacement measurement
Connect to the Strain Gauge Converter
Bridge voltage
8939:
2 V is supplied.
8960:
Selectable
(2 V/ 5 V/ 10 V)
Input voltage
+
-
+
Connector Pinout
Tooth inside
receptacle
-
Lock
Connector
guide pin
Plug ring
Slot
3
2
1
The grounds of the instrument and the Con-
nector Shell of the 8960 are not isolated.
2.2 Connecting the Cables
21
For voltage measurement, be sure to read "2.2.1 Connection Preparations" (⇒ p. 16) before
connecting.
For current measurement, read also "2.3 Connecting Clamps" (⇒ p. 29).
When measuring current
See below for the standards defining the underlined terms.
IEC61010-1 (JIS C1010-1)
IEC61010-2-031 (JIS C1010-2-31)
IEC61010-2-032 (JIS C1010-2-32)
For information about other clamps, see "2.3 Connecting Clamps" (⇒ p. 29).
2.2.5
Connecting to the Model 8940 F/V Unit
Input module
Model 8940 F/V Unit
Maximum Input Voltage (⇒ p. 4)
Frequency, Pulse Totalization and Duty, and Voltage
Measurement:
Any of the following connects to a BNC jack on an input
module.
Current Measurement:
Connect to the BNC and sensor jacks on the input mod-
ule.
(The connection procedure depends on the clamp being
used.)
• Model 9197 Connection Cord
Maximum input voltage:
500 V (large alligator clips)
• Model 9198 Connection Cord
Maximum input voltage:
300 V (small alligator clips)
• Model 9217 Connection Cord
Maximum input voltage:
300 V (for BNC output)
If the voltage to be measured exceeds the maximum input
rating of the input module being used:
• Model 9322 Differential Probe*1 (⇒ p. 31)
• Model 3273 or 3273-50*5Clamp-On Probe*2
• Model 3274, 3275, or 3276 Clamp-On Probe*3
• Model 9270, 9271, or 9272 Clamp-On Sensor*4
• Model 9277, 9278, or 9279 Universal Clamp-On CT*4
*1. The optional Model 9325 Power Cord ,
Model 9418-15 AC Adapter, or Model
9248 Power Cord (when using the
Model 9687) is required for connec-
tion. (⇒ p. 32)
The following items are required to
connect a clamp.
*2. Any of the following models:
9319 Conversion Cable, the 3272 or
3269 Power Supply, or Model 9248
Power Cord (when using the Model
9687)
*3. The Model 3272 or 3269 Power Supply
or Model 9248 Power Cord (when us-
ing the Model 9687)
*4. The Model 9318 Conversion Cable or
the Model 9555 Sensor Unit
*5. When the Models 8940 and 3273-50
are used together, the input rating of
the 3273-50 is 15 Arms.
• Connect the clamp using the Model 9318 or 9319 Conversion Cable
The grounds of this instrument and the clamp are not isolated. Pay care-
ful attention to the connections to avoid damage to the equipment and
serious injury.
• When using the Model 3273 or 3273-50 Clamp-On Probe
To avoid short circuits and serious injury when opening the sensor jaws
to measure, if the voltage on the conductor to be measured may exceed
safe voltage up to 300 V, the conductor should be well-insulated (with
1400 V AC withstand voltage), comply with overvoltage category I, pollu-
tion level 2 and have double (strength) insulation rated for 300 V opera-
tion. To maintain safety, do not clamp over bare conductors. The core
and shield cover are not insulated.
• To avoid electric shock when using a clamp, be careful not to damage the
insulation on conductors being measured.
2.2 Connecting the Cables
22
To maintain safety, always use the optional Model 9318 or 9319 Conversion
Cables when using one of theses clamp models: 3273, 3273-50, 9270 –
9272 or 9277 – 9279.
1
Connect the BNC plug on the cable to a
BNC jack on the input module.
Align the slots in the BNC plug with the guide
pins on the jack on the input module, then push
and twist the plug clockwise until it locks.
2
Connect the cable clips to the measure-
ment object.
Disconnecting BNC connectors
Push the BNC plug, twist it counterclock-
wise, and pull it out.
Required item: Connection Cables
BNC jack
Connection Procedure: Frequency, Pulse Totalization and Duty,
and Voltage Measurement
Lock
Input module
connector guide pins
BNC plug slots
Connect to the measurement object
Connecting the cable
1
2
1
Align the slot in the conversion cable plug
with the sensor receptacle on the input
module, then insert and twist the plug until
it locks.
2
Align the slot in the sensor cable plug with
the conversion cable receptacle, then
insert and twist the plug until it locks.
3
Clamp the sensor around the measure-
ment object.
To remove the conversion cable
Push and twist the plug to unlock it, then
pull it out.
Example: When connecting the Model 9272 Clamp-On Sensor
Required item:
Model 9318 Conversion Cable and 9272 Clamp-On
Sensor
Sensor connector
Connection Procedure: Current measurement (with Model 9318 Conversion Cable)
(for Model 9270 to 9272 Clamp-On Sensors and Model 9277 to 9279 Universal Clamp-On CTs)
Connect to the measurement object
Connecting the clamp and the
Mode 9318 Conversion Cable
Conversion Cable Recep-
tacle
2
1
Conversion Cable Plug
9272 Clamp-On Sensor
Clamp-On
Sensor Plug
3
2.2 Connecting the Cables
23
Connection Procedure: Current measurement (with Model 9319 Conversion Cable)
(for Model 3273 or 3273-50Clamp-On Probe)
1
Align the slots in the clamp-on probe plug
with the pins on the input module’s BNC
jack, press and twist.
2
Align the groove in the conversion cable
plug with the sensor receptacle on the
input module, then insert and twist the plug
until it locks.
3
Align the groove in the probe’s power
cable plug with the conversion cable
receptacle, and mate the connectors.
4
Clamp the sensor around the measure-
ment object.
To remove the conversion cable
Push and twist the plug to unlock it, then
pull it out.
Example: When using the Model 3273 Clamp-On Probe
Required item:
Models 9319 Conversion Cable and 3273 Clamp-On
Probe
BNC jack
Sensor Receptacle
Connect to the measurement object
Connecting the clamp and the
Model 9319 Conversion Cable
Conversion Cable
Receptacle
Measurement Signal Plug
Alignment Notch
1
2
4
3273 Clamp-On Probe
1
2
3
Conversion Cable Plug
Power Cable Plug
2.2 Connecting the Cables
24
Before measuring voltage and connecting to the BNC jack, be sure to read "2.2.1 Connec-
tion Preparations" (⇒ p. 16).
Connection Preparations
2.2.6
Connecting to the Model 8947 Charge Unit
Input module
Model 8947 Charge Unit
Maximum Input Voltage (⇒ p. 4)
Measurement using a piezoelectric acceleration sensor:
Connect to the BNC jack or miniature receptacle
(according to the particular sensor).
Voltage measurement*2:
Connect to the BNC jack on an input module.
Model 9198 Connection Cord (recommended)
• Preamplified acceleration sensor *1
(BNC connector, 2-mA, 15-V power supply)
• Charge-output type piezoelectric acceleration sensor *1
(Miniature #10-32 connector)
*1. A commercially available acceleration
sensor can be used.
*2. For voltage measurement, the connec-
tion procedure is described in "2.2.2
Connecting to an Analog Unit (Models
8936, 8938, 8946, 8956 and 8957)"
(⇒ p. 17).
• Never connect the cable to the instrument while connected to a measure-
ment object. An electric shock could result.
• The BNC jack for each channel and the miniature receptacles all share
common ground. To avoid short circuits, do not connect to two recepta-
cles at the same time.
• To avoid electric shock or damage to the measurement object, turn the
instrument off while connecting a sensor or probe to the BNC jack.
When the [Preamp] measurement mode is selected, internal power (15 V
@ 2 mA) is applied to the BNC jack when measurement starts.
About preamplified acceleration sensors
The sensor should be compatible with the specification (15 V @ 2 mA) of the
Model 8947 Charge Unit. Incompatible sensors are likely to be damaged.
2.2 Connecting the Cables
25
1
Connect the BNC plug from the sensor to
a BNC jack on the input module.
Align the slots in the BNC plug with the guide
pins on the jack on the input module, then push
and twist the plug clockwise until it locks.
2
Connect the sensor to the measurement
object.
Disconnecting BNC connectors
Push the BNC plug, twist it counterclock-
wise, and pull it out.
Required item:
Preamplified acceleration sensor
BNC jack
Connection Procedure: Using a preamplified acceleration sensor
When using a sensor/preamp that has no BNC plug
Use an adapter or conversion cable to connect to the BNC jack.
Lock
Input module
connector guide pins
BNC plug slots
Connect to the measurement object
Connecting a preamplified
acceleration sensor
1
2
1
Connect the miniature plug from the
charge-output acceleration sensor to the
receptacle on the input module, and twist
the plug clockwise until tight.
2
Connect the sensor to the measurement
object.
To disconnect the miniature connector,
twist the plug counterclockwise and pull it
out.
Required item:
Charge-output acceleration sensor
Miniature Connector
Connection Procedure: Using a charge-output acceleration sensor
With a charge-output sensor that does not have a (#10-32) miniature connector
Use an adapter or conversion cable to connect to the miniature connector.
Connect to the measurement object
Connecting the charge-output
acceleration sensor
1
2
2.2 Connecting the Cables
26
2.2.7
Connecting to the Model 8958 16-Ch
Scanner Unit
Input module
Model 8958 16-Ch Scanner Unit
Maximum Input Voltage (⇒ p. 4)
Connect to the terminal block on the input module.
Temperature Measurement: Thermocouple
Voltage Measurement: Input Cable
Recommended cables: Solid 0.14 to 1.5 mm2
Stranded 0.14 to 1 mm2
16 to 26 AWG
Connect to the measurement object
Connecting to the terminal block
1
Strip back the insulation from the thermo-
couple wires as shown at the left.
Stripping length: approx. 5 mm
2
Remove the terminal block cover from the
input module.
3
Loosen the screws in the input module’s
terminal block using the screwdriver.
4
Insert the thermocouple wires into the con-
nection holes.
Confirm proper polarity.
5
Tighten the screws in the terminal block
using the screwdriver to affix the thermo-
couple leads.
6
Replace the terminal block cover.
7
Attach the thermocouple to the measure-
ment object.
Terminal
Block
Example: Connecting a thermocouple for temperature measurement (the connection procedure is the
same for voltage measurement)
Connection Holes
Required item:
Thermocouple, flat-bladed screwdriver (supplied)
Connection Procedure (for both temperature and voltage measurements)
5 mm
Thermocouple
single wires
Screw
Thermocouple
Terminal
Block
Cover
The terminal block can be de-
tached from the Model 8958 16-
Ch Scanner Unit.
How to detach:
Loosen the screws at the top and
bottom of the terminal block (left
illustration).
When attaching, be careful not to
overtighten the screws.
With cover in
place
Screw
Terminal Block
Turn the cover over and place the terminal
block on it as shown above for stability while
making connections.
Terminal Block
Cover
If difficulties are encountered connecting input cables or thermocouples
1
2
3
4
6
7
To remove the thermocouple
Loosen the screws on the terminal block
and pull the wires out.
2.2 Connecting the Cables
27
The zero position of the Model 8958 16-Ch Scanner Unit needs to be adjusted in the following
cases. If the adjustment is not performed, the accuracy specification may not be satisfied.
Allow one hour warm-up after turning power on before adjusting.
• When the Model 8958 16-Ch Scanner Unit has just been installed in the instrument (a message
requesting adjustment appears when the instrument is turned on)
• When the zero position has shifted due to aging or environmental changes
Init Settings Screen
1
Short the + and – terminals of channels 1
and 9 on the Model 8958 16-Ch Scanner
Unit to be adjusted.
2
Press F7 [System] on the Opening screen,
or place the cursor on the Function menu
of the Waveform or Settings screen, and
press the F7 [System] key.
The System screen appears.
3
Press the SUB MENU keys to select the
[Init] menu item.
The Initialization (Init) Settings screen appears.
4
Using the CURSOR keys, place the cursor
on the [Adjust Scanner Unit] button, and
press the F1 [Execute] key.
The Adjustment dialog box appears.
5
Select the module to be adjusted, and
press the F7 [Execute] key.
To cancel, press the F8 [Cancel] key.
A processing message appears when executed,
and the 8958 16-Ch Scanner Unit is adjusted.
Adjustment is complete when “Completed nor-
mally” appears.
Waveform or Settings Screen
Function Menu
Supplied shorting
bar or tinned wire.
CH1
CH9
Opening Screen
Scanner Unit Zero Position Adjustment
Tighten the screws
securely.
2.2 Connecting the Cables
28
Be sure to read "2.2.1 Connection Preparations" (⇒ p. 16) before connecting.
2.2.8
Connecting to the Model 8959 DC/RMS Unit
Input module
Model 8959 DC/RMS Unit
Maximum Input Voltage (⇒ p. 4)
Any of the following connects to a BNC jack on an input
module.
• Model 9197 Connection Cord
Maximum input voltage:
500 V (large alligator clips)
• Model 9198 Connection Cord
Maximum input voltage:
300 V (small alligator clips)
• Model 9217 Connection Cord
Maximum input voltage:
300 V (for BNC output)
If the voltage to be measured exceeds the maximum input
rating of the input module being used:
• Model 9665 10:1 Probe (⇒ p. 35)
• Model 9666 100:1 Probe (⇒ p. 35)
• Model 9322 Differential Probe*1 (⇒ p. 31)
*1. The Model 9418-15 AC Adapter option
is required for connection (⇒ p. 33).
1
Connect the BNC plug on the cable to a
BNC jack on the input module.
Align the slots in the BNC plug with the guide
pins on the jack on the input module, then push
and twist the plug clockwise until it locks.
2
Connect the cable clips to the measure-
ment object.
Disconnecting BNC connectors
Push the BNC plug, twist it counterclock-
wise, and pull it out.
Required item: One of the above cables
BNC jack
Connection Procedure
Lock
Input module
connector guide pins
BNC plug slots
Connect to the measurement object
Connecting the cable
1
2
2.3 Connecting Clamps
29
Because it measures only voltage input, current is not measured directly by this
instrument.
Use a clamp that provides voltage output for current measurement.
Refer to "1.3 List of Input Modules, Cables, Probes and Clamp Combinations"
(⇒ p. 11) for compatible combinations of Hioki clamps and input modules.
Refer also to the clamp’s Instruction Manual for clamp-specific details.
Preparations for Using Clamps
2.3
Connecting Clamps
Connect the clamp-on sensors to the instrument first, and then to the
active lines to be measured. Observe the following to avoid electric shock
and short circuits.
• To avoid short circuits and potentially life-threatening hazards, never
attach the clamp to a circuit that operates at more than the maximum
rated voltage, or over bare conductors.
• Clamps should only be connected to the secondary side of a breaker, so
the breaker can prevent an accident if a short circuit occurs. Connec-
tions should never be made to the primary side of a breaker, because
unrestricted current flow could cause a serious accident if a short circuit
occurs.
• When the clamp sensor is opened, do not allow the metal part of the
clamp to touch any exposed metal, or to short between two lines, and do
not use over bare conductors.
To avoid electric shock when measuring live lines, wear appropriate pro-
tective gear, such as insulated rubber gloves, boots and a safety helmet.
• To prevent damage to the instrument and clamp, never connect or disconnect
a sensor while the power is on, or while the sensor is clamped around a con-
ductor.
• Be careful to avoid dropping the clamps or otherwise subjecting them to
mechanical shock, which could damage the mating surfaces of the core and
adversely affect measurement.
• Measurements are affected by the combined accuracy of this instrument and
the clamp.
• The measured value is displayed as voltage [V], although this can be
converted for display as current [A] using the Scaling function.
See "5.4 Converting Input Values (Scaling Function)" in the Instruction Manual
• In the following cases, scaling setup is necessary:
(1) When using a clamp with a voltage measurement input module other than
the Model 8940 F/V Unit
(2) When connecting a clamp to the Model 8940 F/V Unit without the Model
9318 or 9319 Conversion Cable
For details about connecting clamps to the Model 8940 F/V UNIT:
See "2.2.5 Connecting to the Model 8940 F/V Unit" (⇒ p. 21).
2.3 Connecting Clamps
30
1
Connect the probe to a BNC jack on the
input module.
2
Clamp the sensor around the measure-
ment object.
Disconnecting BNC connectors
Push the BNC plug, twist it counterclock-
wise, and pull it out.
Example: Connecting the Model 9018-10 Clamp-On Probe to the 8936 Analog Unit
Required item:
9018-10 Clamp-On Probe
BNC jack
Connection Procedure
LOAD
SOURCE
Conductor
Current flow direc-
tion arrow
The arrows on the clamp indicating the
direction of current flow should point toward
the load side.
9018-10 Clamp-On Probe
Attach the clamp around only one conductor.
Single-phase (2-wire) or three-phase (3-
wire) cables clamped together will not pro-
duce any reading.
OK
Lock
Input module
connector guide pins
BNC plug slots
Connect to the measurement object
Connecting Clamps
For connection details, refer to "2.2.5 Connecting to the Model 8940 F/V Unit" (⇒ p. 21).
1
2
2.4 Connecting a Differential Probe
31
Input modules that are compatible with the Model 9322 Differential Probe:
• Model 8940 F/V Unit*1
• Voltage measurement input modules other than the Model 8940 *2
*1. The Model 9325 Power Cord or Model 9418-15 AC Adapter is required for connection.
*2. The Model 9418-15 AC Adapter is required for connection.
Refer to the instruction manual for the 9322 Differential Probe.
When using an input module other than the Model 8940 F/V Unit
2.4
Connecting a Differential Probe
Note the following maximum input voltage and maximum rated voltage to
earth. If their voltages are exceeded, this device will be damaged and per-
sonal injury will result. Therefore, do not perform measurement in this
case.
• Maximum input voltage
(CAT II) 2000 V DC, 1000 V AC, (CAT III) 600 V AC/DC
• Maximum rated voltage to earth
When using the Grabber clip:
(CAT II) 1500 V AC/DC, (CAT III) 600 V AC/DC
When using the Alligator clip:
(CAT II) 1000 V AC/DC, (CAT III) 600 V AC/DC
Grabber clip
H
L
(CAT II)
2000 V DC, 1000 V AC
(CAT III)
600 V AC/DC
(CAT II)
1500 V AC/DC
(CAT III)
600 V AC/DC
Between H-L
H-to-ground
L-to-ground
H
L
(CAT II)
2000 V DC, 1000 V AC
(CAT III)
600 V AC/DC
(CAT II)
1000 V AC/DC
(CAT III)
600 V AC/DC
Between H-L
H-to-ground
L-to-ground
Alligator clip
• Turn the instrument off before connecting the AC adapter to the 9322 and
to AC power.
• Use only the specified Model 9418-15 AC Adapter (SA130A-1225V-S,
SINO AMERICAN). AC adapter input voltage range is 100 to 240 VAC
(with ±10% stability) at 50/60 Hz. To avoid electrical hazards and damage
to the instrument, do not apply voltage outside of this range.
2.4 Connecting a Differential Probe
32
1
Connect the power cord to the power jack
on the probe.
2
Connect the other end of the power cord
to the sensor receptacle on the 8940 F/V
Unit
3
Connect the BNC plug on the probe to a
BNC jack on the input module.
4
Connect an alligator clip or grabber clip to
the input lead of the probe.
5
Attach to the measurement object.
Disconnecting BNC connectors
Push the BNC plug, twist it counterclock-
wise, and pull it out.
Required item:
Model 9322 Differential Probe and 9325 Power Cord
9325 Power Cord
BNC jack
Connection Procedure: When connecting to the Model 8940 F/V Unit
(using the 9325 Power Cord)
Connect to the measurement object
Power jack
Input leads
Sensor connector
2
4
5
Lock
Input module
connector guide pins
BNC plug slots
BNC connector
3
1
Set the [Probe] selection on the Channel
Setting screen to [9322+9325].
When connecting to the Model 8940 F/V Unit
(using the 9418-15 AC Adapter)
Set the [Probe] selection on the Channel
Setting screen to [9322+9418].
Refer to "Connection Procedure: Measur-
ing voltage with the input module (using the
Model 9418-15 AC Adapter)" (⇒ p. 33) for
connection details.
2.4 Connecting a Differential Probe
33
1
Connect the BNC plug on the probe to a
BNC jack on the input module.
2
Connect the AC adapter to the power jack
on the probe, and plug it into an AC outlet.
3
Connect an alligator clip or grabber clip to
the input lead of the probe.
4
Connect to the measurement object.
Disconnecting BNC connectors
Push the BNC plug, twist it counterclock-
wise, and pull it out.
Required item:
9322 Differential Probe and 9418-15 AC Adapter
BNC jack
Connection Procedure: Measuring voltage with the input module
(using the Model 9418-15 AC Adapter)
Connect to the measurement object
Power jack
Input leads
1
2
3
4
Lock
Input module
connector guide pins
BNC plug slots
BNC connector
Example: Connecting to the Model 8936 Analog Unit
Connect the Model
9418-15 AC Adapter
Connect to an AC outlet
Set the [Probe] selection on the Channel
Setting screen to [1000:1].
2.4 Connecting a Differential Probe
34
Model 9687 Probe
Power Unit
1
Connect the BNC plug on the probe to a
BNC jack on the input module.
2
Connect the Model 9248 Power Cord to
the power jack on the probe, and plug it
into the Model 9687 Probe Power Unit.
3
Connect an alligator clip or grabber clip to
the input lead of the probe.
4
Connect to the measurement object.
Disconnecting BNC connectors
Push the BNC plug, twist it counterclock-
wise, and pull it out.
Required item:
Model 9322 Differential Probe and Model 9248 Power Cord
BNC jack
Connection Procedure:
Supplying power from the Model 9687 Probe
Power Unit (Using the Model 9248 Power Cord)
Connect to the measurement object
Power jack
Input leads
1
2
3
4
Lock
Input module
connector guide pins
BNC plug slots
BNC connector
Example: Connecting to the Model 8936 Analog Unit
Connect the Model
9248 Power Cord
Connect the Model 9687
Probe Power Unit
Set the [Probe] selection on the Channel
Setting screen to [1000:1].
When using the Model 8940, select
[9322+9325] or [9322+9418]. Either setting is
acceptable.
2.5 Connecting Attenuating Probes
35
The Model 9665 10:1 Probe and 9666 100:1 Probe can be connected to the
following input modules:
• Model 8936 Analog Unit
• Model 8938 FFT Analog Unit
• Model 8956 Analog Unit
• Model 8957 High Resolution Unit
• Model 8959 DC/RMS Unit
Refer to the probe’s instruction manual for details.
2.5
Connecting Attenuating Probes
Note the following maximum input voltage and maximum rated voltage to
earth. If their voltages are exceeded, this device will be damaged and per-
sonal injury will result. Therefore, do not perform measurement in this
case.
The measurement category (overvoltage category) must correspond with
that of the input module being used.
• Maximum input voltage
Model 9665 10:1 Probe
1000 Vrms (1 MHz max)
Model 9666 100:1 Probe 5000 Vpeak (1 MHz max)
• Maximum rated voltage to earth
The maximum rated voltage to ground of the input module being used
Calibrate the attenuating probe before measurement.
Procedure: "Calibration" (⇒ p. 36)
1
Connect the BNC plug on the probe to a
BNC jack on the input module.
2
Connect the ground lead to the probe.
3
Connect the hook clip to the probe.
4
Connect to the measurement object.
Disconnecting BNC connectors
Push the BNC plug, twist it counterclock-
wise, and pull it out.
Required item: 9665 10:1 Probe
BNC jack
Connection Procedure
Connect to the measurement object
Ground lead
1
2
3
Lock
Input module
connector guide pins
BNC plug slots
BNC connector
Example: Connecting the Model 9665 10:1 Probe to the 8936 Analog Unit
Hook Clip
4
2.5 Connecting Attenuating Probes
36
1
Connect the metal pins or lead wires to the
TRIG.OUT/CAL and GND I/O terminals.
While pressing the terminal button with the
screwdriver, insert the metal pin or lead wire. Re-
moving the screwdriver affixes the pin or wire.
2
Connect the hook clip of the probe to the
metal pin or lead wire in the TRIG.OUT/
CAL terminal, and the ground lead to the
GND terminal.
3
Select the [Ext Term] menu from the Sys-
tem screen, and select [Calibration] for the
[TRIG.OUT/CAL] setting.
4
Select the following for each settings
screen.
5
Press the instrument’s START key to begin
recording.
A 1-kHz, 5-V square wave is displayed and out-
put at the terminals.
6
Adjust the trimmer on the probe to obtain
the proper waveform on the screen.
Settings Screen
Setting Item
Settings
Status
Timebase
100μS/div
Shot (recording
length)
25 div
Channel
Range
1V/div
Trigger
Trigger Mode
Repeat
[Analog] page
Type
Level
Level
2.5V
Slope
↑
Calibration
Right Side
Waveforms indicating calibration
required
Calibrated waveform
Required item:
Flat-blade screwdriver (2.6-mm blade)
Connect the probe between the TRIG.OUT/CAL and GND
external I/O terminals using metal pins or two short leads
Metal pin example: Mac8
• DH-4-20 pin for surface mounting
• Terminal for hybrid ICs
CD-5-20, CF-1-15 and CA-7-3
(Attenuating probes do not connect directly to the external
I/O terminals.)
Ext Term Settings
Screen
5 V
500 μs
1
2
TRIG.OUT/CAL
Hook Clip
Ground Lead
2
Press the DISP key, and the F7 [System] key.
The System screen appears.
Press the SUB MENU keys to select the [Ext Term] (Ex-
ternal terminal) menu.
3
Attenuating probe
1
To Insert
3
Push
Pull
External I/O
Terminals
GND
(Any GND terminal
may be used)
Calibration
Trimmer
6
2.6 Connecting Logic Probes
37
Up to four logic probes can be connected to the LOGIC receptacles on the right
side of the instrument.
Each logic probe provides four logic input channels, so up to 16 logic waveforms
can be recorded.
The following logic probes are supported:
• Model 9327 Logic Probe
• Model 9321-01 Logic Probe
• Model 9320-01 Logic Probe
Refer to the instruction manual supplied with the logic probe for specific details.
Before connecting a logic probe to the measurement object
2.6
Connecting Logic Probes
Do not connect logic probes other than those supplied by Hioki to the logic
inputs.
For users of the following legacy models:
Models: 9306, 9307, 9320 and 9321
These can be used with this instrument by connecting the Model 9323 Conver-
sion Cable.
To avoid electric shock and short circuit accidents or damage to the instru-
ment, pay attention to the following:
• The ground pin in the LOGIC connector (plug) of the Model 9320-01 and
9327 Logic Probes (and legacy Models 9306 and 9320) is not isolated
from this instrument’s ground (common ground).
Use grounding-type polarized power cords for the measurement object
and this instrument, and obtain power from the same mains circuit.
Connecting to different mains circuits or using a non-grounding power
cord may cause damage to the measurement object or this instrument
because of current flow through the logic probes resulting from potential
difference between the grounds of the different wiring systems.
To avoid these problems, we recommend the following connection pro-
cedure:
Measurement
Object
Memory
HiCorder
Logic probe
Logic probe
Functional Earth Terminals
GND
Connect this instrument to the same
outlet as the measurement object us-
ing the (supplied) grounding polar-
ized power cord.
Connect the measurement object’s
ground to the GND terminal of this in-
strument. (Always obtain power from
the same mains circuit.)
About Functional Earth:
Memory
HiCorder
Measurement
Object
See "3.5 Grounding the Instrument’s Functional Earth" in the Quick Start Manual
2.6 Connecting Logic Probes
38
• Maximum logic probe input voltages are as follows. Do not measure if
the maximum voltage would be exceeded, as damage the instrument or
personal injury may result.
Model 9327 Logic Probe: +50 VDC
Model 9320-01 Logic Probe: +50 VDC
Model 9321-01 Logic Probe: 250 Vrms (HIGH range), 150 Vrms (LOW
range)
• Do not allow the metal tip of a logic probe to cause a short between con-
ductors on the measurement object. Never touch the metal tip of a probe.
Connection Procedure
Connect to the measurement object
See "Logic Probe Usage Procedures" (⇒ p. 39)
Right Side
Example: Connecting the Model 9327 Logic Probe
1
Connect the logic probe by aligning the
groves on the plug and a LOGIC recepta-
cle.
2
Connect to the measurement object.
Required item:
Model 9327 Logic Probe
LOGIC receptacles
When a logic probe is not connected, the corresponding logic waveform appears
at HIGH level on the waveform screen.
2.6 Connecting Logic Probes
39
Logic Probe Usage Procedures
Range Table
Range
Digital input (Threshold value)
Contact input (Detecting resistance value)
1.4 V
1.4 V±0.3 V
more than 1.5 kΩ
less than 500Ω
opened (Output L)
shorted (Output H)
2.5 V
2.5 V±0.4 V
more than 3.5 kΩ
less than 1.5 kΩ
opened (Output L)
shorted (Output H)
4.0 V
4.0 V±0.5 V
more than 2.5 kΩ
less than 8 kΩ
opened (Output L)
shorted (Output H)
Using Model 9327 and 9320-01 Logic Probes
Connect to the measurement
object
When measuring digital signals
(Digital input)
1
Connect the IC clip leads to the logic
probe.
2
Set the input selector to DIGITAL.
3
Connect the alligator clip to the circuit
ground.
4
Use the threshold value selector to select
the threshold value.
5
Connect the IC clips to the measurement
object.
When measuring the contact signal
(Contact input)
1 Connect the Alligator clip leads to the logic
probe.
2 Set the input selector to CONTACT.
3 Connect the alligator clips to the measure-
ment object.
IC clip leads
Alligator clip leads
1
3
3
4
Alligator clip lead
5
Connect to the measurement
object
GND
Input selector
1
2
Input selector
2
Connect to the
LOGIC receptacle
Connect to the
LOGIC receptacle
Threshold value selector
2.6 Connecting Logic Probes
40
For other models, refer to the specifications (⇒ p. 84).
The number of channels
4 (Isolated between channels and the instrument)
Range
LOW
HIGH
Input resistance
30 kΩ min
100 kΩ min
Sensitivity
Output: L
0 to 10 VAC
±(0 to 15) VDC
0 to 30 VAC
±(0 to 43) VDC
Output: H
60 to 150 VAC
±(20 to 150) VDC
170 to 250 VAC
±(70 to 250) VDC
Response time
↑
Less than 1 ms
Less than 1 ms
↓
Less than 3 ms
Less than 3 ms
with 100 VDC
with 200 VDC
Maximum input voltage
150 V rms
250 V rms
Maximum rated voltage
to earth
250 V rms
Dielectric strength
(between unit and channels, between channels)
Insulation resistance
More than 100 MΩ / 500 VDC
(between unit and channels, between channels)
Using the Model 9321-01 Logic Probe
Connect to the measurement
object
Input selector
Connect to the LOGIC receptacle
Clips
1
Set the input selector in accordance with
the measured voltage.
LOW range: On/off for 100 VAC and 24
VDC, etc.
HIGH range: On/off for 200 VAC, etc.
2
Connect the clips to the measurement
object.
• Inputs are non-polarized, so polarity can be disregarded.
• Inputs are isolated, so each channel can be connected to a point of indepen-
dent potential.
• This instrument detects absolute values, so DC of either polarity can be input.
• AC voltage is evaluated as a 50/60 Hz standard sine wave.
2.7 Supplying power from the Model 9687 Probe Power Unit
41
The optional Model 9687 Probe Power Unit can be used to supply power to our
optional clamp-on and differential probes.
Probe models compatible with the Model 9687 Probe Power Unit
• 3273, 3273-50, 3274, 3275, 3276 Clamp-On Probes
• 9322 Differential Probe (using Model 9248 Power Cord)
9687 Probe Power Unit Specifications
2.7
Supplying power from the Model 9687 Probe
Power Unit
The power output connectors on the Model 9687 are especially designed for our
clamp-on and differential probes. To avoid damage, do not use the power supply
for any purpose other than supplying power.
There are limits to how many probes can be used with a single Model 8860 or
8861 with a 9687 installed.
Avoid exceeding the quantities in the following table for all probes to be used.
When mixing probe types, the lower quantity of usable probes applies.
Probe
Usable Quantity
3273 Clamp-On Probe
8
3273-50, 3274, 3276 Clamp-On Probe
6
3275 Clamp-On Probe
5
9322 Differential Probe
8
Accuracy is specified at 23±5°C and 20 to 80% RH, 30 minutes after power on
No. of powered channels
8
Compatible probes
3273, 3273-50, 3274, 3275, 3276, 9322
Rated output voltage
±12 V
Rated output current
±3 A (total for all channels)
Operating temperature and
humidity
0 to 40°C (32 to 104°F), 20 to 85% RH (non-condensating)
Storage temperature and
humidity
-10 to 50°C (14 to 122°F), 20 to 90% RH (non-condensat-
ing)
Operating
environment
Compatible with Models 8860/ 8861
Dimensions
Adds approx. 18.2 mm (0.72”) (D) to dimensions of Models
8860/ 8861
Weight
Adds approx. 570 g (20.1 oz.) to the weight of Models 8860/
8861
Supported Models
Model 8860 Serial Nos. 051040422 and above
Model 8861 Serial Nos. 051040432 and above
Connect the power cables from the probes to
the output connectors of the Model 9687
Probe Power Unit.
Power Output
Connector
Connections
After turning power on, set the measurement criteria before measuring.
This chapter describes the input channel setting procedures for each input mod-
ule.
Refer to "3.10 About Setting Contents" (⇒ p. 64) for details of each setting.
Refer to the Instruction Manual for other settings.
Input Channel
Settings
Chapter 3
• Model 8936 Analog Unit
(⇒ p. 44)
• Model 8937 Voltage/Temp
Unit
(⇒ p. 45)
• Model 8938 FFT Analog Unit (⇒ p. 48)
• Model 8939 Strain Unit
(⇒ p. 49)
• Model 8940 F/V Unit
(⇒ p. 50)
• Model 8946 4-Ch Analog Unit(⇒ p. 44)
• Model 8947 Charge Unit
(⇒ p. 56)
• Model 8956 Analog Unit
(⇒ p. 44)
• Model 8957 High Resolution Unit(⇒ p. 59)
• Model 8958 16-Ch Scanner Unit (⇒ p. 60)
• Model 8959 DC/RMS Unit
(⇒ p. 63)
• Model 8960 Strain Unit
(⇒ p. 49)
Level Monitor
Shows the input level while making settings.
See "5.3 Monitoring Input Status" in the
Instruction Manual
Comments can be entered for each channel.
See "5.2 Adding Comments" in the Instruction
Manual
Select the module and channel number to be set.
The display color of acquired waveforms, zero posi-
tion, vertical magnification and display range can be
changed.
See "7.1 Making Input Waveform Display Settings
(Analog
Waveforms)"
in
the
Instruction
Manual
Set the scaling if you want to convert the measured
units for display.
Measured values (output voltage) from sensors can
be converted to physical values appropriate to the
measurement object.
See "5.4 Converting Input Values (Scaling Func-
tion)" in the Instruction Manual
Set up the input channel.
Display appearance depends on the particular input
module. Refer to the description of each module.
Make settings on the Channel Settings screen.
Measurement
Range
Indicates the
selected range.
1 Press the SET key.
2 Press the SUB MENU keys to select the [Channel] menu.
3 Press the SHEET/PAGE keys to select the [One Ch] page.
4 Use the CURSOR keys to move among the setting items,
and press the F1 to F8 keys to select a setting.
Displayed Range
Indicates the range displayed on the screen. To change the display position or
range, change the waveform display settings.
3.1 Analog Unit Settings (Models 8936, 8946 and 8956)
44
Set the input channel for voltage measurement when using the following input
modules:
• Model 8936 Analog Unit
• Model 8956 Analog Unit
• Model 8946 4-Ch Analog Unit
3.1
Analog Unit Settings (Models 8936, 8946 and
8956)
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item Description
1
Select the module No.
[Unit ]
Select the number of the input module to set up.
("Channel Configuration" (⇒ p. 15))
2
Select the channel No.
[Ch ]
Select the channel number on the input module.
3
Confirm the input module and mea-
surement mode
[Module]
[Mode]
Confirm the type of the selected module.
• Model 8936 Analog Unit.............. [Analog]
• Model 8956 Analog Unit.............. [H-Speed]
• Model 8946 4-Ch Analog Unit.....[4CH]
Confirm that the display is set for [Voltage].
4
Select the measurement range
[Range]
Select the appropriate measurement range. The setting val-
ue is the units per division on the vertical axis.
The RANGE/POSN knob selects the range (⇒ p. 64).
Refer to the specifications for available ranges (⇒ p. 86).
5
Select the input coupling method
[Coupling]
Select the coupling method for input signals (⇒ p. 65).
[ DC, AC or GND ]
6
Select the low-pass filter
[LPF]
To suppress high-frequency components, select a low-pass
filter within the input module (⇒ p. 66).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ] (in the Model 8936 case)
7
Select the probe attenuation
[Probe]
Select according to the cable or probe being used (⇒ p. 75).
[ 1:1, 10:1 , 100:1 or 1000:1 ]
8
Perform zero adjustment
[Zero-Adjust] Perform at least 30 minutes after turning power on
(⇒ p. 76).
Set up scaling
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
Example: Setting a channel on the Model 8936 Analog Unit
8
7
Execute to adjust input
signal offset.
(⇒ p. 76)
1
2
3
4
5
([One Ch] Page of Channel Setting Screen)
6
3.2 Model 8937 Voltage and Temperature Unit Settings
45
3.2
Model 8937 Voltage and Temperature Unit
Settings
3.2.1
Voltage Measurement
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item Description
1
Select the module No.
[Unit ]
Select the number of the input module to set up.
("Channel Configuration" (⇒ p. 15))
2
Select the channel No.
[Ch ]
Select the channel number on the input module.
3
Confirm the module
[Module]
Confirm that the display is set for [Volt/Temp].
4
Select the measurement mode
[Mode]
Select [Voltage].
5
Select the measurement range
[Range]
Select the appropriate measurement range. The setting val-
ue is the units per division on the vertical axis.
The RANGE/POSN knob selects the range (⇒ p. 64).
Refer to the specifications for available ranges (⇒ p. 88).
6
Select the input coupling method
[Coupling]
Select the coupling method for input signals (⇒ p. 65).
[ DC, AC or GND ]
7
Select the low-pass filter
[LPF]
To suppress high-frequency components, select a low-pass
filter within the input module (⇒ p. 66).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
8
Select the probe attenuation
[Probe]
Select according to the cable or probe being used (⇒ p. 75).
[ 1:1, 10:1, 100:1 or 1000:1 ]
9
Set the digital filter
[Digital F]
(Effective measurement ranges: 500 µV to 2 mV)
Set the digital filter function to be enabled or disabled. Set it
to ON to suppress noise components (⇒ p. 67).
[ Off or On ]
10
Perform zero adjustment
[Zero-Adjust] Perform at least one hour after turning power on
(⇒ p. 76).
Set up scaling
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
10
([One Ch] Page of Channel Setting Screen)
8
Execute to adjust input
signal offset. (⇒ p. 76)
1
2
3
5
6
4
9
Displayed when the measurement
range is 500 µV to 2 mV.
7
3.2 Model 8937 Voltage and Temperature Unit Settings
46
3.2.2
Temperature Measurement
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item Description
1
Select the module No.
[Unit ]
Select the number of the input module to set up.
("Channel Configuration" (⇒ p. 15))
2
Select the channel No.
[Ch ]
Select the channel number on the input module.
3
Confirm the module
[Module]
Confirm that the display is set for [Volt/Temp].
4
Select the measurement mode
[Mode]
Select [Temp].
5
Select thermocouple
[Sensor]
Select the type of thermocouple (⇒ p. 68).
[ K, J, E, T, N, R, S or B ]
6
Select the measurement range
[Range]
Select the appropriate measurement range. The setting val-
ue is the units per division on the vertical axis.
The RANGE/POSN knob selects the range (⇒ p. 64).
[ 10, 20, 50 or 100°C/div ]
7
Select the low-pass filter
[LPF]
To suppress high-frequency components, select a low-pass
filter within the input module (⇒ p. 66).
[ Off, 5Hz or 500Hz ]
8
Set the reference junction compensa-
tion
[RJC]
Select whether reference junction compensation should be
provided internally (⇒ p. 68).
[ Internal or External ]
9
Set drift correction
[Drift]
Set whether to cancel variations in the reference voltage. Set
this On to improve the temperature characteristic (⇒ p. 69).
[ Off or On ]
Set up scaling
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
([One Ch] Page of Channel Setting Screen)
8
1
2
3
5
6
4
9
The [Voltage] measurement mode is
selected upon system reset.
7
3.2 Model 8937 Voltage and Temperature Unit Settings
47
• The upper and lower limits of measurement input depend on the measure-
ment range.
Refer to "About Measurement Range, Upper and Lower Input Limits and the
Waveform Display:" (⇒ p. 64).
• When the input terminals of the Model 8937 Voltage/Temp Unit are exposed to
strong drafts:
Loss of thermal equilibrium of the input section may result in measurement
errors. To measure in such environments, reposition the equipment so that the
input terminals are protected from the drafts.
• When the ambient temperature changes rapidly:
Loss of thermal equilibrium may result in measurement errors. If this occurs,
allow about an hour for the equipment to acclimatize to the new temperature,
then take measurements after thermal equilibrium is achieved.
• In an environment with fluctuating ambient temperature, when recording for a
long time using a thermocouple with weak thermoelectromotive force (R, S or
B sensors):
Temperature characteristic can be improved by setting drift correction On.
• The [Voltage] measurement mode is selected upon system reset.
3.3 Model 8938 FFT Analog Unit Settings
48
3.3
Model 8938 FFT Analog Unit Settings
9
([One Ch] Page of Channel Setting Screen)
7
Execute to adjust input
signal offset.
(⇒ p. 76)
1
2
3
4
5
8
6
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item Description
1
Select the module No.
[Unit ]
Select the number of the input module to set up.
("Channel Configuration" (⇒ p. 15))
2
Select the channel No.
[Ch ]
Select the channel number on the input module.
3
Confirm the input module and mea-
surement mode
[Module]
[Mode]
Confirm that the display is set for [FFT].
Confirm that the display is set for [Voltage].
4
Select the measurement range
[Range]
Select the appropriate measurement range. The setting val-
ue is the units per division on the vertical axis.
The RANGE/POSN knob selects the range (⇒ p. 64).
Refer to the specifications for available ranges (⇒ p. 90).
5
Select the input coupling method
[Coupling]
Select the coupling method for input signals (⇒ p. 65).
[ DC, AC or GND ]
6
Select the low-pass filter
[LPF]
To suppress high-frequency components, select a low-pass
filter within the input module (⇒ p. 66).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
7
Select the probe attenuation
[Probe]
Select according to the cable or probe being used
(⇒ p. 75).
[ 1:1, 10:1, 100:1 or 1000:1 ]
8
Set the anti-aliasing filter
[AAF]
Turn this on to suppress anti-aliasing distortion, such as for
FFT analysis (⇒ p. 67).
[ Off or On ]
9
Perform zero adjustment
[Zero-Adjust] Perform at least 30 minutes after turning power on
(⇒ p. 76).
Set up scaling
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
3.4 Strain Unit Settings (Models 8939 and 8960)
49
3.4
Strain Unit Settings (Models 8939 and 8960)
([One Ch] Page of Channel Setting Screen)
1
2
3
4
7
5
6
8939
8960
7
8
8
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item Description
1
Select the module No.
[Unit ]
Select the number of the input module to set up.
("Channel Configuration" (⇒ p. 15))
2
Select the channel No.
[Ch ]
Select the channel number on the input module.
3
Confirm the input module and mea-
surement mode
[Module]
[Mode]
Confirm that the display is set for [Strain].
4
Select the measurement range
[Range]
Select the appropriate measurement range. The setting val-
ue is the units per division on the vertical axis.
(Units: [µ] Microstrain = 1/2 µV/V)
The RANGE/POSN knob selects the range (⇒ p. 64).
Refer to the specifications for available ranges (⇒ p. 91).
5
Select the bridge voltage
[Bridge]
Select the bridge voltage (⇒ p. 75).
[ 2V, 5V, 10V ]
6
Set the anti-aliasing filter
[AAF]
Turn this on to suppress anti-aliasing distortion, such as for
FFT analysis (⇒ p. 67).
[ Off or On ]
7
Select the low-pass filter
[LPF]
To suppress high-frequency components, select a low-pass
filter within the input module (⇒ p. 66).
8939: [ Off, 10Hz, 30Hz, 300Hz or 3kHz ]
8960: [ OFF, 5Hz, 10Hz, 100Hz, 1kHz ]
8
Execute auto-balance
[Auto-
Balance]
Sets the reference output level of the transducer to the spec-
ified zero position (⇒ p. 78).
Execute one hour after turning the power on, with the sensor
in the unloaded state.
Set up scaling
Set as occasion demands.
See
"Scaling Setting Examples" in "5.4 Converting Input Values
(Scaling Function)" in the Instruction Manual
(Model 8960 only)
(Model 8960 only)
3.5 Model 8940 F/V Unit Settings
50
Settings are related to the following measurements. One measurement type can
be set per channel.
• Measure frequency from the input pulse corresponding to the measurement
waveform (⇒ p. 50)
• Measure frequencies near 50/60 Hz from the input pulse corresponding to the
measurement waveform (⇒ p. 50)
• Measure rotation rate from the input pulse corresponding to the measurement
waveform (⇒ p. 50)
• Count of input pulses (⇒ p. 52)
• Measure the percentage of High level of a single pulse waveform (pulse duty)
(⇒ p. 53)
• Measure voltage (⇒ p. 54)
• Measure current (⇒ p. 55)
3.5
Model 8940 F/V Unit Settings
3.5.1
Measuring Frequency, 50/60 Hz and Rotation
Rate
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item Description
1
Select the module No.
[Unit ]
Select the number of the input module to set up.
("Channel Configuration" (⇒ p. 15))
2
Select the channel No.
[Ch ]
Select the channel number on the input module.
3
Confirm the module
[Module]
Confirm that the display is set for [F/V].
4
Select the measurement mode
[Mode]
Select the measurement mode.
[ Frequency, Rotation, 50/60 Hz ]
5
Select the measurement range
[Range]
Select the appropriate measurement range. The setting val-
ue is the units per division on the vertical axis.
The RANGE/POSN knob selects the range (⇒ p. 64).
Refer to the specifications for available ranges (⇒ p. 92).
6
Select the low-pass filter
[LPF]
To suppress high-frequency components, select a low-pass
filter within the input module (⇒ p. 66).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
([One Ch] Page of Channel Setting Screen)
8
1
2
3
5
4
9
Example: Measuring frequency
6
7
10
The [Frequency] measurement mode is
selected upon system reset.
3.5 Model 8940 F/V Unit Settings
51
7
Select the probe attenuation
[Probe]
Select according to the cable or probe being used
(⇒ p. 75).
[ 1:1, 10:1, 100:1, 1000:1(9322+9325) or 1000:1(9322+9418) ]
8
Set the threshold level
[Threshold] Set the threshold value (⇒ p. 72).
9
Set pull-up
[Pull-Up]
Enable (set On) the pull-up resistance when connecting to an
open-collector output signal. Disable it (Off) for normal mea-
surements (⇒ p. 71).
[ Off or On ]
10
Set the Hold function
[Hold]
Select whether to hold the last measured value until the next
frequency or rotation rate is determined (⇒ p. 70).
[ On, 10-ms Off or 1-s Off ]
Set up scaling
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
Difference between frequency measurement and 50/60 Hz frequency
measurement
The frequency measurement function consists of selecting a range from 50 mHz
to 5 kHz and observing changes in frequency.
The 50/60 Hz frequency measurement function consists of setting either 50 or
60 Hz as the zero position, and observing deviations from that frequency.
3.5 Model 8940 F/V Unit Settings
52
3.5.2
Pulse Count Measurement
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item Description
1
Select the module No.
[Unit ]
Select the number of the input module to set up.
("Channel Configuration" (⇒ p. 15))
2
Select the channel No.
[Ch ]
Select the channel number on the input module.
3
Confirm the module
[Module]
Confirm that the display is set for [F/V].
4
Select the measurement mode
[Mode]
Select [Count].
5
Select the measurement range
[Range]
Select the appropriate measurement range. The setting val-
ue is the units per division on the vertical axis.
(Units: c = count)
The RANGE/POSN knob selects the range (⇒ p. 64).
Refer to the specifications for available ranges (⇒ p. 92).
6
Select the low-pass filter
[LPF]
To suppress high-frequency components, select a low-pass
filter within the input module (⇒ p. 66).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
7
Select the probe attenuation
[Probe]
Select according to the cable or probe being used
(⇒ p. 75).
[ 1:1, 10:1, 100:1, 1000:1(9322+9325) or 1000:1(9322+9418) ]
8
Set the threshold level
[Threshold] Set the threshold level (⇒ p. 72).
9
Set pull-up
[Pull-Up]
Enable (set On) the pull-up resistance when connecting to an
open-collector output signal. Disable it (Off) for normal mea-
surements (⇒ p. 71).
[ Off or On ]
Set up scaling
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
([One Ch] Page of Channel Setting Screen)
8
1
2
3
5
4
6
7
9
The [Frequency] measurement mode is
selected upon system reset.
3.5 Model 8940 F/V Unit Settings
53
3.5.3
Pulse Duty Measurement
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item Description
1
Select the module No.
[Unit ]
Select the number of the input module to set up.
("Channel Configuration" (⇒ p. 15))
2
Select the channel No.
[Ch ]
Select the channel number on the input module.
3
Confirm the module
[Module]
Confirm that the display is set for [F/V].
4
Select the measurement mode
[Mode]
Select [Duty].
5
Select the low-pass filter
[LPF]
To suppress high-frequency components, select a low-pass
filter within the input module (⇒ p. 66).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
6
Select the probe attenuation
[Probe]
Select according to the cable or probe being used
(⇒ p. 75)
[ 1:1, 10:1, 100:1, 1000:1(9322+9325) or 1000:1(9322+9418) ]
7
Set the threshold level
[Threshold] Set the threshold level (⇒ p. 72).
8
Set pull-up
[Pull-Up]
Enable (set On) the pull-up resistance when connecting to an
open-collector output signal. Disable it (Off) for normal mea-
surements (⇒ p. 71).
[ Off or On ]
Set up scaling
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
([One Ch] Page of Channel Setting Screen)
7
1
2
3
4
5
6
8
The [Frequency] measurement mode is
selected upon system reset.
Pulse duty measurement
If a rising pulse (at 20 kHz or above) is applied during calculation (“dead time”),
the duty during the dead time cannot be measured. Duty is measured from the
pulse following the dead time.
Calculation (50 µs)
Waveform Acquisition
Waveform Acquisition
Ignored
3.5 Model 8940 F/V Unit Settings
54
3.5.4
Voltage Measurement
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item Description
1
Select the module No.
[Unit ]
Select the number of the input module to set up.
("Channel Configuration" (⇒ p. 15))
2
Select the channel No.
[Ch ]
Select the channel number on the input module.
3
Confirm the module
[Module]
Confirm that the display is set for [F/V].
4
Select the measurement mode
[Mode]
Select [Voltage].
5
Select the measurement range
[Range]
Select the appropriate measurement range. The setting val-
ue is the units per division on the vertical axis.
The RANGE/POSN knob selects the range (⇒ p. 64).
Refer to the specifications for available ranges (⇒ p. 92).
6
Select the input coupling method
[Coupling]
Select the coupling method for input signals (⇒ p. 65).
[ DC, AC or GND ]
7
Select the low-pass filter
[LPF]
To suppress high-frequency components, select a low-pass
filter within the input module (⇒ p. 66).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
8
Select the probe attenuation
[Probe]
Select according to the cable or probe being used
(⇒ p. 75).
[ 1:1, 10:1, 100:1, 1000:1(9322+9325) or 1000:1(9322+9418) ]
9
Set pull-up
[Pull-Up]
Enable (set On) the pull-up resistance when connecting to an
open-collector output signal. Disable it (Off) for normal mea-
surements (⇒ p. 71).
[ Off or On ]
10
Perform zero adjustment
[Zero-Adjust] Perform at least 30 minutes after turning power on
(⇒ p. 76).
Set up scaling
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
10
([One Ch] Page of Channel Setting Screen)
8
Execute to adjust input
signal offset. (⇒ p. 76)
1
2
3
5
6
4
9
The [Frequency] measurement mode
is selected upon system reset.
7
3.5 Model 8940 F/V Unit Settings
55
When using the Model 9018-10 and 9132-10 Clamp-On Probe, select [Voltage]
mode.
See "3.5.4 Voltage Measurement" (⇒ p. 54)
3.5.5
Current Measurement
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item Description
1
Select the module No.
[Unit ]
Select the number of the input module to set up.
("Channel Configuration" (⇒ p. 15))
2
Select the channel No.
[Ch ]
Select the channel number on the input module.
3
Confirm the module
[Module]
Confirm that the display is set for [F/V].
4
Select the measurement mode
[Mode]
Select [Current].
5
Select a sensor
[Sensor]
Select the clamp to be used.
[ 3273, 3273-50, 9270, 9271, 9272LO*, 9272HI*, 9277, 9278
or 9279 ]
* 9272HI: 200A-range, 9272LO: 20A-range
6
Select the measurement range
[Range]
Select the appropriate measurement range. The setting val-
ue is the units per division on the vertical axis.
The RANGE/POSN knob selects the range (⇒ p. 64).
Refer to the specifications for available ranges (⇒ p. 92).
7
Select the input coupling method
[Coupling]
Select the coupling method for input signals (⇒ p. 65).
[ DC, AC or GND ]
8
Select the low-pass filter
[LPF]
To suppress high-frequency components, select a low-pass
filter within the input module (⇒ p. 66).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
9
Perform zero adjustment
[Zero-Adjust] Perform at least 30 minutes after turning power on
(⇒ p. 76).
9
5
Execute to adjust input
signal offset. (⇒ p. 76)
1
2
3
6
7
4
The [Frequency] measurement mode is
selected upon system reset.
([One Ch] Page of Channel Setting Screen)
8
• Up to eight channels can be used simultaneously when measuring with the
Model 9318 and 9319 Conversion Cables.
• When using a clamp model that is not displayed in the sensor selection list,
measure in [Voltage] mode, and apply the scaling function.
See:"5.4 Converting Input Values (Scaling Function)" in the Instruction Manual
3.6 Model 8947 Charge Unit Settings
56
Make settings regarding input channels for measuring voltage or acceleration
(Charge and Preamp). One type can be measured per channel.
3.6
Model 8947 Charge Unit Settings
3.6.1
Voltage Measurement
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item Description
1
Select the module No.
[Unit ]
Select the number of the input module to set up.
("Channel Configuration" (⇒ p. 15))
2
Select the channel No.
[Ch ]
Select the channel number on the input module.
3
Confirm the module
[Module]
Confirm that the display is set for [Charge].
4
Select the measurement mode
[Mode]
Select [Voltage].
5
Select the measurement range
[Range]
Select the appropriate measurement range. The setting val-
ue is the units per division on the vertical axis.
The RANGE/POSN knob selects the range (⇒ p. 64).
Refer to the specifications for available ranges (⇒ p. 95).
6
Select the input coupling method
[Coupling]
Select the coupling method for input signals (⇒ p. 65).
[ DC, AC or GND ]
7
Select the low-pass filter
[LPF]
To suppress high-frequency components, select a low-pass
filter within the input module (⇒ p. 66).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
8
Select the probe attenuation
[Probe]
Select according to the cable or probe being used (⇒ p. 75).
[ 1:1, 10:1, 100:1 or 1000:1 ]
9
Set the anti-aliasing filter
[AAF]
Turn this on to suppress anti-aliasing distortion, such as for
FFT analysis (⇒ p. 67).
[ Off or On ]
10
Perform zero adjustment
[Zero-Adjust] Perform at least one hour after turning power on
(⇒ p. 76).
Set up scaling
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
10
([One Ch] Page of Channel Setting Screen)
8
Execute to adjust input
signal offset. (⇒ p. 76)
1
2
3
5
6
4
9
The [Charge] measurement mode is
selected upon system reset.
7
3.6 Model 8947 Charge Unit Settings
57
3.6.2
Acceleration Measurement (Charge, Preamp)
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item Description
1
Select the module No.
[Unit ]
Select the number of the input module to set up.
("Channel Configuration" (⇒ p. 15))
2
Select the channel No.
[Ch ]
Select the channel number on the input module.
3
Confirm the module
[Module]
Confirm that the display is set for [Charge].
4
Select the measurement mode
[Mode]
Select the measurement mode.
• [Charge]: When using a charge-output acceleration pick-
up sensor
• [Preamp]: When using an acceleration pick-up sensor
equipped with an internal preamp
5
Set sensor sensitivity
[Sensitivity] Set the value per m/s2 defined for the acceleration sensor to
be used (⇒ p. 73).
"Sensor Sensitivity Setting Examples" (⇒ p. 73)
6
Select the measurement range
[Range]
Select the appropriate measurement range. The setting val-
ue is the units per division on the vertical axis.
(Units: m/s2 = acceleration)
The RANGE/POSN knob selects the range.
Refer to the specifications for available ranges (⇒ p. 95).
7
Select the input coupling method
[Coupling]
Select the coupling method for input signals (⇒ p. 65).
[ AC or GND ]
8
Select the low-pass filter
[LPF]
To suppress high-frequency components, select a low-pass
filter within the input module (⇒ p. 66).
[ Off, 500Hz or 5kHz ]
9
Set the anti-aliasing filter
[AAF]
Turn this on to suppress anti-aliasing distortion, such as for
FFT analysis (⇒ p. 67).
[ Off or On ]
10
Perform zero adjustment
[Zero-Adjust] Perform at least one hour after turning power on
(⇒ p. 76).
Set up scaling
Set as occasion demands.
See
"To convert [m/s2] units to [G] for display" (⇒ p. 73) in this
manual,
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
10
([One Ch] Page of Channel Setting Screen)
9
1
2
3
6
7
4
5
The [Charge] measurement mode is
selected upon system reset.
8
3.6 Model 8947 Charge Unit Settings
58
Supplement ___________________________________________________
Before connecting sensors and probes
Range Setting
Auto-ranging cannot be used on channels for which the measurement mode is set
to [Charge] or [Preamp].
For more on auto-ranging, refer to "3.3.5 Automatic Range Setting (Auto-Ranging
Function)" in the Instruction Manual.
Measuring Timing
In the following cases, five or six seconds should be allowed for input stabilization:
• Measurement Mode: [Charge]
When switching the measurement range between one of the six high-sensitiv-
ity ranges and one of the six low-sensitivity ranges.
• Measurement Mode: [Preamp]
After selecting preamp mode and before starting a new measurement
To avoid electric shock or damage to the measurement object, turn the
instrument off before connecting a sensor or probe to the BNC jack.
When the [Preamp] measurement mode is selected, internal power (15 V @
2 mA) is applied to the BNC jack when measurement starts.
5 to 6 seconds
3.7 Model 8957 High Resolution Unit Settings
59
3.7
Model 8957 High Resolution Unit Settings
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item Description
1
Select the module No.
[Unit ]
Select the number of the input module to set up.
("Channel Configuration" (⇒ p. 15))
2
Select the channel No.
[Ch ]
Select the channel number on the input module.
3
Confirm the input module and mea-
surement mode
[Module]
[Mode]
Confirm that the display is set for [High Res].
Confirm that the display is set for [Voltage].
4
Select the measurement range
[Range]
Select the appropriate measurement range. The setting val-
ue is the units per division on the vertical axis.
The RANGE/POSN knob selects the range (⇒ p. 64).
Refer to the specifications for available ranges (⇒ p. 98).
5
Select the input coupling method
[Coupling]
Select the coupling method for input signals (⇒ p. 65).
[ DC, AC or GND ]
6
Select the low-pass filter
[LPF]
To suppress high-frequency components, select a low-pass
filter within the input module (⇒ p. 66).
[ Off, 5Hz, 50Hz, 500Hz, 5kHz or 50kHz ]
7
Select the probe attenuation
[Probe]
Select according to the cable or probe being used
(⇒ p. 75).
[ 1:1, 10:1, 100:1 or 1000:1 ]
8
Set the anti-aliasing filter
[AAF]
Turn this on to suppress anti-aliasing distortion, such as for
FFT analysis (⇒ p. 67).
[ Off or On ]
9
Perform zero adjustment
[Zero-Adjust] Perform at least 30 minutes after turning power on
(⇒ p. 76).
Set up scaling
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
9
([One Ch] Page of Channel Setting Screen)
7
Execute to adjust input
signal offset. (⇒ p. 76)
1
2
3
4
5
8
6
3.8 Model 8958 16-Ch Scanner Unit Settings
60
3.8
Model 8958 16-Ch Scanner Unit Settings
Before Setting
The zero position of the Model 8958 16-Ch Scanner Unit needs to be adjusted in
the following cases.
See "Scanner Unit Zero Position Adjustment" (⇒ p. 27)
• When the Model 8958 16-Ch Scanner Unit has just been installed in the
instrument (a message requesting adjustment appears when the instrument is
turned on)
• When the zero position has shifted due to aging or environmental changes
3.8.1
Voltage Measurement
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item Description
1
Select the module No.
[Unit ]
Select the number of the input module to set up.
("Channel Configuration" (⇒ p. 15))
2
Select the channel No.
[Ch ]
Select the channel number on the input module.
3
Confirm the module
[Module]
Confirm that the display is set for [Scanner].
4
Select the measurement mode
[Mode]
Select [Voltage].
5
Select the measurement range
[Range]
Select the appropriate measurement range. The setting val-
ue is the units per division on the vertical axis.
The RANGE/POSN knob selects the range (⇒ p. 64).
Refer to the specifications for available ranges (⇒ p. 99).
6
Set the digital filter
[Digital F]
Set the digital filter function to be enabled or disabled. Set it
to ON to suppress noise components (⇒ p. 67).
[ Off, 60Hz, 50Hz or 10Hz ]
Set up scaling
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
([One Ch] Page of Channel Setting Screen)
Execute to adjust input
signal offset. (⇒ p. 76)
1
2
3
5
4
6
3.8 Model 8958 16-Ch Scanner Unit Settings
61
3.8.2
Temperature Measurement
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item Description
1
Select the module No.
[Unit ]
Select the number of the input module to set up.
("Channel Configuration" (⇒ p. 15))
2
Select the channel No.
[Ch ]
Select the channel number on the input module.
3
Confirm the module
[Module]
Confirm that the display is set for [Scanner].
4
Select the measurement mode
[Mode]
Select [Temp].
5
Select thermocouple
[Sensor]
Select the type of thermocouple (⇒ p. 68).
[ K, J, E, T, N, R, S, B or W ]
6
Select the measurement range
[Range]
Select the appropriate measurement range. The setting val-
ue is the units per division on the vertical axis.
The RANGE/POSN knob selects the range (⇒ p. 64).
[ 10, 50 or 100°C/div ]
7
Set the digital filter
[Digital F]
Set the digital filter function to be enabled or disabled. Set it
to ON to suppress noise components (⇒ p. 67).
[ Off, 60Hz, 50Hz or 10Hz ]
8
Set the reference junction compensa-
tion
[RJC]
Select whether reference junction compensation should be
provided internally (⇒ p. 68).
[ Internal or External ]
9
Set disconnect detection
[Burn-Out]
Select thermocouple broken-wire detection (⇒ p. 69).
When enabled (On), the waveform trace of a broken wire
stays at the top of the screen and the cursor value is dis-
played as “+OVER” when a broken wire is present.
[ Off or On ]
Set up scaling
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
([One Ch] Page of Channel Setting Screen)
8
1
2
3
5
6
4
9
7
The [Voltage] measurement mode is
selected upon system reset.
3.8 Model 8958 16-Ch Scanner Unit Settings
62
• The upper and lower limits of measurement input depend on the measure-
ment range. Refer to "About Measurement Range, Upper and Lower Input
Limits and the Waveform Display:" (⇒ p. 64).
• When the input terminals of the Model 8958 16-Ch Scanner Unit are subject
to strong drafts:
Loss of thermal equilibrium of the input section may result in measurement
errors. To measure in such environments, install the supplied cover or relocate
the equipment out of the wind.
• When the ambient temperature changes rapidly:
Loss of thermal equilibrium may result in measurement errors. If this occurs,
allow about an hour for the equipment to acclimatize to the new temperature,
then take measurements after thermal equilibrium is achieved.
3.9 Model 8959 DC/RMS Unit Settings
63
3.9
Model 8959 DC/RMS Unit Settings
Setting Procedure
Use the CURSOR keys to move among items. Select a setting with the F1 to F8 keys.
Step
Setting Item Description
1
Select the module No.
[Unit ]
Select the number of the input module to set up.
("Channel Configuration" (⇒ p. 15))
2
Select the channel No.
[Ch ]
Select the channel number on the input module.
3
Confirm the module
[Module]
Confirm that the display is set for [DC/RMS].
4
Select the measurement mode
[Mode]
Select the measurement mode.
• [DC]: Voltage measurement
• [RMS]: RMS measurement
5
Select the measurement range
[Range]
Select the voltage per division on the voltage (vertical) axis.
The RANGE/POSN knob selects the range (⇒ p. 64).
Refer to the specifications for available ranges (⇒ p. 101).
6
Select the input coupling method
[Coupling]
Select the coupling method for input signals (⇒ p. 65).
[ DC, AC or GND ]
7
Select the low-pass filter
[LPF]
To suppress high-frequency components, select a low-pass
filter within the input module (⇒ p. 66).
[ Off, 5Hz, 500Hz, 5kHz or 100kHz ]
8
Select the probe attenuation
[Probe]
Select according to the cable or probe being used
(⇒ p. 75).
[ 1:1, 10:1, 100:1 or 1000:1 ]
9
Set the response speed
[Response]
Select the response speed (⇒ p. 72).
[ Fast, Medium or Slow ]
10
Perform zero adjustment
[Zero-Adjust] Perform at least 30 minutes after turning power on
(⇒ p. 76).
Set up scaling
Set as occasion demands.
See
"5.4 Converting Input Values (Scaling Function)" in the
Instruction Manual
10
([One Ch] Page of Channel Setting Screen)
8
Execute to adjust input
signal offset. (⇒ p. 76)
1
2
3
5
6
4
9
Example: Measuring RMS values
The [DC] measurement mode is se-
lected upon system reset.
7
3.10 About Setting Contents
64
Select the appropriate measurement range. The setting value is the units per
division on the vertical axis.
Use the RANGE/POSN knob to make the setting independently from the cursor
position.
Setting Item: [Range]
• Refer to "Chapter 5 Specifications" (⇒ p. 83) for the extents of the measure-
ment ranges.
• If the range is undetermined, auto-range selection can be enabled.
Refer to "3.3.5 Automatic Range Setting (Auto-Ranging Function)" in the
Instruction Manual
• When data has been stored in the instrument’s memory, an asterisk “∗”
appears beside the data value’s measurement range.
• When a portion of the input waveform is outside of the measurement range,
that portion is indicated by a different color on the display.
Voltage Axis Range
When variable auto adjustment is enabled [On] (default setting) and the variable
function is enabled, the variable setting becomes linked for measurement range
changes.
If you do not want the display area of the screen to change, disable variable auto
adjustment [Off].
See "12.2.7 Performing Variable Function Auto Adjustment" in the Instruction Manual
Temperature Measurement
About Measurement Range, Upper and Lower Input Limits and the Wave-
form Display:
The upper and lower limits of measurement input depend on the measurement
range.
Measurement is not available outside of the temperature range limits in the fol-
lowing tables:
3.10 About Setting Contents
3.10.1 Measurement Range Setting
Minimum Resolution
If the range is changed while measuring, acquired data is deleted and data
recording is restarted.
Model 8937 Voltage/Temp Unit
Measurement Range [°C/div]
10
20
50
100
Upper Limit [°C]
400
800
2000
4000
Lower Limit [°C]
-92
-184
-460
-920
Model 8958 16-Ch Scanner Unit
Measurement Range [°C/div]
10
50
100
Upper Limit [°C]
200
1000
2000
Lower Limit [°C]
-100
-200
-200
3.10 About Setting Contents
65
Example: Zero position = 50%, Magnification = ×1, Thermocouple = type T (using
Model 8973)
(Measurement input range of T = -200 to 400°C
Select the coupling method for input signals
Available selections depend on the input module and measurement mode.
Measurement range: 50°C/div
(-460 to 2000°C)
When T measurement input range exceeded
Measurement range: 20°C/div
(-184 to 800°C)
Lower limit of measurement range (-184°C)
exceeded
Upper limit of mea-
surement input range
exceeded
Lower limit of mea-
surement range ex-
ceeded
Upper limit of T
measurement input
Lower limit of T
measurement input
Lower limit of mea-
surement input range
exceeded
Thermocouple: T
Measurement
Range
3.10.2 Setting Input Coupling
Setting Item: [Coupling]
Selections
Description
DC
DC Coupling (Default setting)
Select this to acquire both DC and AC components of an input signal.
AC
AC Coupling
Select this to eliminate any DC component from an input signal.
Use this to measure only the ripple component superimposed on pulsating
currents.
GND
The input signal is disconnected. Zero position can be confirmed.
Input modules
Modes
Selections
8936
Voltage
DC
AC
GND
8937
Voltage
DC
AC
GND
Temp
DC
−
−
8938
FFT
DC
AC
GND
8940
Frequency, Rotation,
50/60Hz, Count, Duty
DC
−
−
Voltage
DC
AC
GND
Current
depends on the particular
clamp
8946
Voltage
DC
−
GND
8947
Voltage
DC
AC
GND
Charge, Preamp
−
AC
GND
8959
DC, RMS
DC
AC
GND
8956
Voltage
DC
AC
GND
8957
Voltage
DC
AC
GND
3.10 About Setting Contents
66
To suppress high-frequency components, select a low-pass filter within the input
module
Available selections depend on the input module and measurement mode.
* When the digital filter is enabled [On], only Off, 5 Hz and 500 Hz selections are available.
3.10.3 Low-Pass Filter (LPF) Settings
Setting Item: [LPF]
Selections
Description
Off
The low-pass filter is disabled. (Default setting)
5Hz
Applies a filter with 5-Hz cutoff frequency.
500Hz
Applies a filter with 500-Hz cutoff frequency.
5kHz
Applies a filter with 5-kHz cutoff frequency.
100kHz
Applies a filter with 100-kHz cutoff frequency.
Input modules
Modes
Selections
[Hz]
8936
Voltage
Off, 5, 500, 5 k, 100 k
8937 *
Voltage
Off, 5, 500, 5 k, 100 k
Temp
Off, 5, 500
8938
Voltage
Off, 5, 500, 5 k, 100 k
8939
Strain
Off, 10, 30, 300, 3 k
8940
Voltage
Off, 5, 500, 5 k, 100 k
8946
Voltage
Off, 5, 500, 5 k, 50 k
8947
Voltage
Off, 5, 500, 5 k, 100 k
Charge, Preamp
Off, 500, 5 k
8956
Voltage
Off, 5, 500, 5 k, 1 M
8957
Voltage
Off, 5, 50, 500, 5 k, 50 k
8959
DC, RMS
Off, 5, 500, 5 k, 100 k
8960
Strain
Off, 5, 10, 100, 1 k
3.10 About Setting Contents
67
Enable the anti-aliasing filter to remove aliasing distortion. The cutoff frequency
is automatically set according to the time axis range setting.
Noise can be removed by applying additional averaging to the measurement
data within the input module.
With the 8937 Voltage/Temp Unit, digital filtering is available only in the mea-
surement ranges from 500 µV to 2 mV/div.
The digital filter is not available in other ranges.
3.10.4 Anti-Aliasing Filter (AAF) Settings
Setting Item: [AAF]
Selections
Description
Off
The anti-aliasing filter is disabled. (Default setting)
On
The anti-aliasing filter is enabled.
(When the Recorder function or External sampling is used, the anti-
aliasing filter (AAF) is not available.)
To perform sampling for analysis without being affected by aliasing
distortion
We recommend using an input module that supports anti-aliasing filtering. (Sup-
porting input module models: 8938, 8947 and 8957)
Printing setting data (List Print)
Settings can be printed as follows:
(Example) When enabled: “AAF ON”
3.10.5 Digital Filter (Digital F) Settings
Setting Item: [Digital F]
Selections
Description
8937
Off
The digital filter function is disabled. (Default setting)
On
The digital filter function is enabled.
(Data refresh rate: Approx. 100 µs)
8958
Off
The digital filter function is disabled. (Default setting)
60Hz, 50Hz or
10Hz
The digital filter function is enabled. The data refresh rate
is affected by the filter setting.
See "5.2.10 Model 8958 16-Ch Scanner Unit" (⇒ p. 99)
Printing setting data (List Print)
Settings can be printed as follows:
Example: Enabled = “Digital F ON"
3.10 About Setting Contents
68
Set to match the type of thermocouple being used.
When connecting a thermocouple directly to the input module, select [Internal].
Reference junction compensation is performed within the input module.
When connecting through a reference junction device (e.g., a 0°C control tank),
select [External].
3.10.6 Thermocouple (Sensor) Type Setting
Setting Item: [Sensor]
8937 Voltage/Temp Unit
Selections
Measurement
Range
Selections
Measurement
Range
K
-200 to 1350°C
R
0 to 1700°C
J
-200 to 1100°C
S
0 to 1700°C
E
-200 to 800°C
B
300 to 1800°C
T
-200 to 400°C
N
-200 to 1300°C
8958 16-Ch Scanner Unit
Selections
Measurement
Range
Selections
Measurement
Range
K
-200 to 1350°C
R
0 to 1700°C
J
-200 to 1200°C
S
0 to 1700°C
E
-200 to 1000°C
B
400 to 1800°C
T
-200 to 400°C
W
0 to 2000°C
N
-200 to 1300°C
3.10.7 Reference Junction Compensation Setting
Setting Item: [RJC]
Selections
Description
Internal
Reference junction compensation is provided within the input mod-
ule. (Default setting)
(Measurement Accuracy: The sum of the accuracies of the temper-
ature measurement and the reference junction compensation.)
External
Reference junction compensation is not provided within the input
module.
(Measurement Accuracy: The accuracy of temperature measure-
ment only)
3.10 About Setting Contents
69
This function periodically corrects for drift of the reference potential (about once
per second) to improve thermal characteristics in thermocouple mode.
A broken thermocouple wire can be detected during temperature measurement.
Normally when a thermocouple wire breaks, measured values exhibit random
instability.
3.10.8 Drift Correction Setting
Setting Item: [Drift]
Selections
Description
Off
Drift correction is disabled. (Default setting)
On
Drift correction is enabled.
• Data refresh rate is about once per second when drift correction is enabled.
• Drift correction is available only while measuring. It does not affect the Level
Monitor on the Waveform and Channel Setting screens.
3.10.9 Disconnect (Burn-Out) Detection Setting
Setting Item: [Burn-Out]
Selections
Description
Off
Broken wires are not detected.
On
Broken wires are detected.
When a broken wire is present, the waveform trace stays at the top
of the screen and the cursor value is displayed as “+OVER”.
Wire breakage is detected by sensing a miniscule current flow
(about 200 nA) through the thermocouple. If the thermocouple
wires are long or composed of a high-resistance material, set
[Burn-Out] to [Off] to avoid measurement errors.
3.10 About Setting Contents
70
During measurement, the measured frequency value can be retained by the
Hold function until the frequency of the next cycle has been determined.
Also, if the value is not retained and the next value cannot be determined within
a specified period, one half of the last measured value can be displayed.
Example: Measuring the phenomena of stopping a rotating body
3.10.10 Hold Setting
Setting Item: [Hold]
Selections
Description
On
Retains the measured value until the next frequency value is deter-
mined.
10-ms Off
If the next frequency value is not determined within 10 ms, the last
value is divided in half. (Default setting)
1-s Off
If the next frequency value is not determined within 1 s, the last val-
ue is divided in half.
The on/off settings of the Hold and Pull-Up functions are not indicated on the
Waveform screen. The settings are indicated on the Channel Settings screen.
What is the difference between Hold On and Off?
When measuring frequency and rotation rate, results are displayed only after the
value of one cycle has been determined.
When Hold is On, the previously measured value is retained, and with Hold Off,
one half of the previous value is displayed if a new value cannot be determined
within the specified period.
Rotation Stops
Hold: On
The previously measured value is retained continuously
until the next cycle is determined. Because the next cycle
never occurs, the stopping condition is not detected.
Hold: Off
If the cycle is not detected within the specified period (10
ms or 1 s), half of the value of the previous signal is dis-
played.
As a result, the displayed value gradually approaches zero,
allowing the stopping condition to be estimated.
Rotation Stops
3.10 About Setting Contents
71
Example: Measuring frequency dropping to 0 Hz, and then increasing
The pull-up resistance function is used when connecting to an open-collector
output signal.
The input terminal is pulled up to +5 V.
Set Pull-Up to Off for normal measurements.
Actual Phenomena
Frequency drops from a certain cycle to 0 Hz, and then
increases from that point.
Hold: 10-ms Off
At frequencies below 100 Hz, the displayed value is suc-
cessively halved to 0 Hz.
However, until the frequency reaches 100 Hz again, the
halved signal is now added.
Hold: On
The previous value is retained until the 0-Hz event.
Hold: 1-s Off
At frequencies below 1 Hz, the displayed value is succes-
sively halved to 0 Hz.
However, until the frequency reaches 1 Hz again, the
halved signal is now added.
200
100
0
[Hz]
2
1
0
[Hz]
3.10.11 Pull-Up Setting
Setting Item: [Pull-Up]
Selections
Description
Off
Pull-up resistance is disable (Off).
On
Pull-up resistance is enabled. (for connection to open-collector out-
put)
3.10 About Setting Contents
72
Sets threshold value. Measures when the waveform crosses a preset threshold.
Values can also be entered by numeric keypad.
See "3.3.3 Entering Text and Numbers" in the Instruction Manual
Response can be set to three speeds: Fast, Medium and Slow.
Normally set to [Fast], this can be changed to [Medium] or [Slow] to stabilize the
display when measuring low frequencies, or when severe fluctuations are
present.
3.10.12 Threshold Setting
Setting Item: [Threshold]
Selections
Description
↑↑
Increases value by large steps.
↑
Increases value by small steps.
↓
Decreases value by small steps.
↓↓
Decreases value by large steps.
Measurement results may differ according to the threshold setting. To obtain the
correct measurement results, set the threshold level to suit the input waveform.
5
[V]
0
-5
Threshold
Threshold Setting Example
3.10.13 Response (Timing) Setting
Setting Item: [Response]
Selections
Description
Fast
Sets the response time to about 100 ms.
Medium
Sets the response time to about 800 ms.
Slow
Sets the response time to about 5 s.
3.10 About Setting Contents
73
Sets the value per m/s2 defined for the acceleration sensor to be used.
Setting Item: [Sensitivity]
Setting Range: 0.1 to 10 (pC/(m/s2)) [Charge], 0.1 to 10 (mV/(m/s2) [Preamp]
When using a sensor with specified value per G:
See Example 2 of the "Sensor Sensitivity Setting Examples" (⇒ p. 73)
To use a sensor outside of the setting range:
See "To Use a Sensor Outside of the Setting Range" (⇒ p. 74)
Sensor Sensitivity Setting Examples
Setting Example 1:
When using a sensor with sensitivity specified in units of m/s2
If the specified sensor sensitivity is 1.08 pC/(m/s2)
Sensor sensitivity setting value: 1.08
Setting Example 2:
When using a sensor with sensitivity specified in units of G
An acceleration sensor with sensitivity specified in units of G can be set by divid-
ing the specified sensitivity value by 9.8 (m/s2).
If the specified sensor sensitivity is 64.0 (pC/G): 64.0 / 9.8 = 6.53 (pC/(m/s2))
Sensor sensitivity setting value: 6.53
Set up scaling as follows.
3.10.14 Sensor Sensitivity Setting
The measurement range can be affected by the sensor sensitivity. So sensor
sensitivity should be set before setting the measurement range.
To convert [m/s2] units to [G] for display
This instrument measures charge value in units of m/s2. The scaling function can
be used to convert to charge value in units of G.
See "5.4 Converting Input Values (Scaling Function)" in the Instruction Manual
• Conversion ratio setting [Ratio]
Conversion ratio:
1/9.8 = 0.1020E+00
Offset:
0.0000E+00
Units:
G
• Two-point setting [2-Point]
Entry 1:
9.8000E+00
Physical value 1:
1.0000E+00
Entry 2:
0.0000E+00
Physical value 2:
0.0000E+00
Units:
G
3.10 About Setting Contents
74
To Use a Sensor Outside of the Setting Range
The scaling function can be employed to use a sensor outside of the setting
range.
1
On the Channel Settings screen, select the channel on which to use the sensor
outside of the setting range.
2
Move the cursor to [Sensitivity] to set the sensor sensitivity.
Enter a multiplier value that will produce a settable value (0.1 to 10) when multi-
plied by the actual sensitivity of the sensor.
3
Move the cursor to [Scaling], and set it to [On].
4
Set the conversion ratio to the sensitivity multiplier value.
Setting Example 1:
If the specified sensor sensitivity is 23.4 pC/(m/s2):
Set 1/2.34 as the sensor sensitivity multiplier, and 10 (pc/m/s2)) as the sensor
sensitivity.
Scaling is set as follows to display measured values multiplied by 1/2.34.
Setting Example 2:
If the specified sensor sensitivity is 0.05 pC/(m/s2)
Set the sensor sensitivity multiplier to 2, and the sensor sensitivity to 0.1 (pc/m/
s2)).
Scaling is set as follows to display measured values multiplied by 2.
• Conversion ratio setting [Ratio]
Conversion ratio:
10/23.4 = 0.4274E+00
Offset:
0.0000E+00
Units:
m/s2
• Two-point setting [2-Point]
Input 1:
2.3400E+01
Scale 1:
1.0000E+01
Input 2:
0.0000E+00
Scale 2:
0.0000E+00
Units:
m/s2
• Conversion ratio setting [Ratio]
Conversion ratio:
0.1/0.05 = 2.0000E+00
Offset:
0.0000E+00
Units:
m/s2
• Two-point setting [2-Point]
Input 1:
0.0500E+00
Scale 1:
0.1000E+00
Input 2:
0.0000E+00
Scale 2:
0.0000E+00
Units:
m/s2
3.10 About Setting Contents
75
Probe attenuation can be selected when measuring using a connection cable,
differential probe or attenuating probe. (Only for voltage measurement)
By matching the input channel probe setting to the attenuation ratio of a probe
connected to an analog input module, the voltage axis range is automatically
converted for direct reading of numerical values.
Each channel should be set to match the attenuation ratio of its input probe.
When using the Model 8940 F/V Unit with the 9322 Differential Probe:
When supplying power from the Model 9687 Probe Power Unit, either of the
above settings can be selected.
The factory default and system reset default setting is 1:1.
Set to suit the sensor to be used.
3.10.15 Probe Attenuation Selection
Setting Item: [Probe]
Selections
Description
1:1
Select this setting when measuring with the Model 9197, 9198 or
9217 cable connected to the input module.
10:1
Select this setting when measuring with the Model 9665 10:1 Probe
connected to the input module.
100:1
Select this setting when measuring with the Model 9666 100:1
Probe connected to the input module.
1000:1
Select this setting when using the 9322 Differential Probe.
9322+9325
Select this setting when using the Model 9325 Power Cord to con-
nect the 9322 Differential Probe.
9322+9418
Select this setting when using the Model 9418-15 AC Adapter to
connect the 9322 Differential Probe.
Some probes cannot be used with certain input modules.
See "1.3 List of Input Modules, Cables, Probes and Clamp Combinations" (⇒ p. 11)
3.10.16 Bridge Voltage Setting
Setting Item: [Bridge]
Selections
Description
2V
Sets the bridge voltage to 2 V (Bridge resistance = 120 Ω to 1 kΩ)
5V
Sets the bridge voltage to 5 V (Bridge resistance = 350 Ω to 1 kΩ)
10V
Sets the bridge voltage to 10 V (Bridge resistance = 350 Ω to 1 kΩ)
This setting should not exceed the allowable applied voltage for the strain gauge
type converter to be used. Otherwise, converter specifications will not be satis-
fied, and correct measurements will not be possible.
3.10 About Setting Contents
76
The input module can apply an internal offset to set the reference potential of the
instrument to zero volts. Adjustment applies to the currently selected range.
Before executing zero adjustment
• Turn power on and wait 30 minutes to allow the internal temperature of the
input module to stabilize. (Some input modules may require one hour warm-
up.)
• Zero adjustment cannot be performed while measuring.
• Key operations are not accepted while zero adjustment is executing.
To execute zero adjustment
Move the cursor to the [Zero-Adjust] button, and select F1 [Execute].
Zero adjustment can be performed from the CH SET dialog on the Waveform
screen.
The input value can be forced to display as zero volts. Input signal bias, such as
sensor emf, can be adjusted to display as zero volts. This function applies to
voltage and current measurement.
See " What is the difference between Offset Cancellation and Zero Adjustment?" (⇒ p.
77)
Before executing offset cancellation
Connect the probe or clamp to the measurement object and apply the input sig-
nal that is supposed to measure zero volts.
See "About input voltage during offset cancellation" (⇒ p. 77)
When also using scaling, execute offset cancellation before setting the scaling.
To execute offset cancellation
Move the cursor to the [Offset Cancel] button, and select F1 [Execute].
An asterisk "∗" appears on the [Offset Cancel] button when offset cancellation is
active.
To revert to the offset value prior to executing offset cancellation, select F2
[Reset].
3.10.17 Executing Zero Adjustment
Zero adjustment is not applicable to the following modules:
• The [Temp] mode of the Model 8937 Voltage/Temp Unit
• Model 8939 Strain Unit
• Modes other than [Voltage] and [Current] of the Model 8940 F/V Unit
• Model 8958 16-Ch Scanner Unit
In the following cases, zero adjustment should be executed again.
• After an input module has been removed or inserted
• After power has been turned off and on
• After complete reinitialization
• When ambient temperature has changed significantly
The zero position may have drifted.
When using the Model 8958 16-Ch Scanner Unit, zero position adjustment of the
module from the System screen may be needed.
See "Scanner Unit Zero Position Adjustment" (⇒ p. 27)
3.10.18 Executing Offset Cancellation
3.10 About Setting Contents
77
About input voltage during offset cancellation
A constant input voltage is required to execute offset cancellation.
If the input voltage is fluctuating, the offset to zero volts depends on the timing of
offset cancellation execution.
Offset cancellation
Offset cancellation is not reset even when power is turned off. It is reset when
the input module configuration is changed (after an input module has been
inserted or removed).
What is the difference between Offset Cancellation and Zero Adjust-
ment?
Offset cancellation adjusts for input signal bias if there is a potential difference
between the reference potential of the measurement object and the reference
potential (0 V) of this instrument, so that the reference potential of the measure-
ment object appears as zero volts.
Zero adjustment sets the input module’s internal bias so the reference potential
of the instrument is zero volts.
Offset cancellation cannot be executed in the following cases:
• When measuring other than voltage or current
• When the input voltage is more than ±10 divisions from zero
The maximum input voltage and maximum rated voltage to ground are
unaffected by offset cancellation of an input voltage to zero volts. There-
fore, careful attention is necessary during measurement.
When input voltage is constant
When input voltage is fluctuating
Offset voltage +V1 is set to display as
zero regardless of execution time.
The offset voltage set to display as
zero depends on the timing of offset
cancellation execution.
If “Warning: Offset Cancel failed.” appears
This appears when executing offset cancellation if the input voltage is more than
±10 divisions from zero volts.
To avoid the warning, set the input signal within ±10 divisions of zero volts.
Offset cancellation reference
potential is +V.
Waveform after executing
offset cancellation
Input Signal Waveform
3.10 About Setting Contents
78
Auto-balance sets the reference output level of a transducer to the specified zero
position.
It is applicable only to a strain module.
Before executing auto-balance
• Turn power on and wait 1 hour to allow the internal temperature of the input
module to stabilize.
• With the sensor connected to the input module, execute auto-balance under
stable input conditions.
• Auto-balance cannot execute during measurement.
• Key operations are not accepted while auto-balance is executing.
To execute auto-balance
Move the cursor to the [Auto-Balance] button, and select F1 [Execute].
Auto-balance can be performed from the CH SET dialog on the Waveform
screen (when the unit and channel numbers of the strain module is selected).
In the following cases, auto-balance should be executed again.
• After changing the measurement range
• After an input module has been removed or inserted
• After the strain gauge transducer has been replaced
• After power has been turned off and on
• After performing a system reset
• When ambient temperature has changed significantly (the zero position may
drift)
3.10.19 Executing Auto-Balance
If “Warning: Auto balance failed.” appears
The channel on which auto-balance failed is displayed.
Verify the following, and execute again:
• Is the sensor in a discharged state? (Make sure that it is not being subject to
vibration, etc.)
• Is the sensor connected correctly?
4.1 Operation Overview
79
After turning power on, set the measurement criteria before measuring.
This chapter describes the settings related to logic inputs.
Refer to "7.3 Displaying Logic Waveforms" in the Instruction Manual for setting
details.
Measurement status and trigger settings are the same as for analog inputs.
Refer to the related chapters in the Instruction Manual.
Logic Input
Settings
Chapter 4
4.1
Operation Overview
(Select the Function)
To record simultaneously with an analog signal, select a function that suits
the analog input.
When recording only logic signals, the function can be set freely.
Select the measurement status and
channels
(For Memory Function)
[Use Ch] page on Status Settings screen
Enable (set to On) the logic channels to be used.
See "4.2.1 Selecting Channels to Use" in the Instruction Manual
Set up input channels
[Logic] page on Channel Settings screen
Select whether to enable (On) or disable (Off) the input waveform for each
logic channel probe.
(Default setting: All Off)
Select waveform colors as desired.
See "7.3 Displaying Logic Waveforms" in the Instruction Manual
Set trigger criteria
[Logic] page on Trigger Settings screen
To apply a trigger event for measurements, set the logic trigger for each
channel.
See "6.8 Triggering by Logic Signals (Logic Trigger)" in "Chapter 6 Trigger
Settings" in the Instruction Manual
Waveform display settings
[Logic] page on Sheet Settings screen
Select the channels to appear on the waveform display.
(Default setting: All On)
Set the waveform display position and height for each channel as needed.
See "7.2 Setting the Screen Layout of the Waveform Screen (Sheet Set-
tings Screen)" and "7.3 Displaying Logic Waveforms" in the Instruction
Manual
Start Measurement
4.2 Setting Procedure
80
4.2
Setting Procedure
Display Settings Screen
Press the SET key.
The Settings screen appears.
Function setting is not needed.
Selecting channels to use (when displayed with the memory function)
Status Settings Screen
Logic Channel A to D
1
Press the SUB MENU keys to select the [Status]
menu item.
2
Press the SHEET/PAGE keys to select the [Use Ch]
page.
To record simultaneously with analog input:
See "4.2.1 Selecting Channels to Use" in the Instruction
Manual
3
Use the CURSOR keys to move the cursor to the Set-
ting column of the logic channel, and press the F2
[On] key.
1
2
3
LOGIC
Receptacles
CH D
CH C
CH B
CH A
Logic Channels
(Default setting: depends on the settings of the channels to be used)
4.2 Setting Procedure
81
3
4
Select the time axis range and recording length
Status Setting Screen (with memory function)
1
Press the SUB MENU keys to select the [Status]
menu item.
2
(When using the memory function)
Press the SHEET/PAGE keys to select the [Basic]
page.
3
(To record only logic inputs)
Use the CURSOR keys to move the cursor to [Time
base], and press an F key to select the time per divi-
sion for the horizontal axis.
(When the Timebase 2 is enabled [On], and only logic
inputs are assigned)
Set the sampling rate.
4
Use the CURSOR keys to move the cursor to [Shot]
(Recording Length), and press an F key to select the
number of divisions for the recording time.
About measurement configuration (timebase, recording
length, etc.)
See "Chapter 4 Measurement Configuration Settings" in
the Instruction Manual
1
2
3
Select the input channels
Channel Settings Screen
Probes
Logic Channels
1
Press the SUB MENU keys to select the [Channel]
menu item.
2
Press the SHEET/PAGE keys to select the [Logic]
page.
The [Logic] page of the Channel Setting screen appears.
3
Use the CURSOR keys to move the cursor to the
Setting column for each probe (1 to 4) of the logic
channel, and press the F2 [On] key.
(On = display waveform, Off = do not display waveform)
To change a waveform’s color, move the cursor to
the waveform display color setting column and
select the color with the F1 or F2 key.
See "7.3.1 Setting the Waveform Display" in the
Instruction Manual
If you want to enter comments for individual
channels:
Select the [Comment] page.
See "5.2 Adding Comments" in the
Instruction Manual
1
2
(Default setting: [Off])
Waveform Display
On/Off
Waveform Display
Color
4.2 Setting Procedure
82
2
Set the trigger conditions (if applying a trigger)
Trigger Settings Screen
1
Press the SUB MENU keys to select the [Trigger]
menu item.
2
Press the SHEET/PAGE keys to select the [Logic]
page.
The [Logic] page of the Trigger Settings screen appears.
3
Use the CURSOR keys to move to each channel
to be triggered, and set each as needed.
See "6.8 Triggering by Logic Signals (Logic Trigger)" in
the Instruction Manual
4
Set triggering as occasion demands.
See "Chapter 6 Trigger Settings" in the Instruction
Manual
1
Logic Channels
Set the trigger pattern
for each probe
Select AND/OR
trigger conditions.
Set the trigger filter width.
Prevents noise from causing
false triggers.
(Default setting: Logic trigger [Off])
3
Set the trigger
detection method.
Set the waveform display method (as occasion demands)
Sheet Settings Screen
1
Press the SUB MENU keys to select the [Sheet]
menu.
2
Press the SHEET/PAGE keys to select the [Logic]
page.
The [Logic] page of the Sheet Settings screen appears.
3
Use the CURSOR keys to move to a logic channel
to be displayed, and press the F keys to select the
setting contents for each item.
See "7.3.2 Setting the Display Position" and "7.3.3 Set-
ting the Display Height" in the Instruction Manual
Setting other display sheets
See "Chapter 7 Waveform Display Settings" in the
Instruction Manual
1
2
3
(Default setting: All logic channels displayed [On])
Set storage and printing as occasion demands, then start measuring.
Logic
Channels
Set display of each channel on or off.
Waveform display
positions can be
modified.
Display height can
be modified.
5.1 Logic Input Section
83
Refer to "Safety Information" for the “Measurement Category (Overvoltage Category)” in the Quick Start
Manual.
Specifications
Chapter 5
5.1
Logic Input Section
5.1.1
Model 9327 Logic Probe
Temperature and humidity
range for guaranteed accu-
racy
Period of guaranteed accu-
racy
23 ±5°C(73±9°F), 35 to 80% RH
1 year
The number of input chan-
nels
4 channels (Common ground between chassis and between channels)
Input type
Digital input / Contact input
Input type can be selected for each channels.
Open collector outputs can be directly measured using contact inputs.
Input resistance
1 MΩ±5% (Digital input: 0 to +5 V)
500 kΩ or more (Digital input: +5 to +50 V)
Pull-up resistance
2 kΩ (Contact inputs: internally connected through pull-up resistance to +5 V)
Detecting level
Response pulse width
100 ns or more
Maximum input voltage
0 to +50 VDC
Operating temperature and
humidity ranges
0 to 40°C (32 to 104°F), 80%RH or less (non-condensating)
Storage temperature and
humidity ranges
-10 to 50°C (14 to 122°F), 90%RH or less (non-condensating)
Operating environment
Altitude up to 2000 m (6562-ft.), indoors
Dimensions
Approx. 62W x 94H x 20D mm (2.44"W x 3.70"H x 0.79"D) (sans protrusions)
Connector cable length
Approx. 1500 mm (59")
Probe tip cable length
Approx. 300 mm (11.8")
Mass
Approx. 150 g (5.3 oz.) (including connector cable, excluding input leads)
Accessories
IC clip leads, Alligator clip leads, Carrying case, Instruction Manual
Digital input
Threshold value
Contact input
Detecting resistance value
1.4 V range
1.4 V±0.3 V
more than 1.5 kΩ
less than 500Ω
opened (Output L)
shorted (Output H)
2.5 V range
2.5 V±0.4 V
more than 3.5 kΩ
less than 1.5 kΩ
opened (Output L)
shorted (Output H)
4.0 V range
4.0 V±0.5 V
more than 25 kΩ
less than 8 kΩ
opened (Output L)
shorted (Output H)
5.1 Logic Input Section
84
5.1.2
Model 9321-01 Logic Probe
Temperature and humidity
range for guaranteed accu-
racy
Period of guaranteed accu-
racy
23 ±5°C(73±9°F), 35 to 80% RH
1 year
The number of input chan-
nels
4 channels (Isolated from chassis and between channels)
Input voltage range
LOW
HIGH
Input resistance
More than 30 kΩ
More than 100 kΩ
Detecting level
(Output: L)
(Output: H)
0 to 10 VAC
±(0 to 15) VDC
60 to 150 VAC
±(20 to 150) VDC
0 to 30 VAC
±(0 to 43) VDC
170 to 250 VAC
±(70 to 250) VDC
Response time
(Rising)
↑
(Falling)
↓
Less than 1 ms
Less than 3 ms
at 100 V DC
Less than 1 ms
Less than 3 ms
at 200 V DC
Maximum input voltage
150 V rms
250 V rms
Maximum rated voltage to
earth
250 V rms
Dielectric strength
2.3 kVAC for 1 minute (between each input channel and chassis, and between input
channels)
Insulation resistance
At least 100 MΩ at 500 V DC (between each input channel and chassis, and between
input channels)
Operating temperature and
humidity ranges
0 to 40°C (32 to 104°F), 80%RH or less (non-condensating)
Storage temperature and
humidity ranges
-10 to 50°C (14 to 122°F), 90%RH or less (non-condensating)
Operating environment
Altitude up to 2000 m (6562-ft.), indoors
Dimensions
Approx. 62W x 128H x 20D mm (2.44"W x 5.04"H x 0.79"D) (sans protrusions)
Connector cable length
Approx. 1500 mm (59")
Probe tip cable length
Approx. 1000 mm (39.37")
Mass
Approx. 320 g (11.3 oz.)
Accessories
Carrying case, Instruction Manual
• Absolute values are detected, so either polarity DC may be input. AC voltage is
evaluated as a 50/60-Hz standard sine wave.
• On/off judgment of the input signal is performed within the above detection
range. On/off judgment of input signals cannot be performed correctly with
input voltage between L and H detection levels.
5.1 Logic Input Section
85
5.1.3
Model 9320-01 Logic Probe
Temperature and humidity
range for guaranteed accu-
racy
Period of guaranteed accu-
racy
23 ±5°C(73±9°F), 35 to 80% RH
1 year
The number of input chan-
nels
4 channels (Common ground between chassis and between channels)
Input type
Digital input / Contact input
Input type can be selected for each channel.
Open collector outputs can be directly measured using contact inputs.
Input resistance
Digital input: 1 MΩ±5% (0 to +5 V), 500 kΩ or more (+5 to +50 V)
Contact inputs: Approx. 2 kΩ (input terminals are connected to +5 V through 2-kΩ resis-
tance)
Detecting level
Response time
Less than 500 ns
Maximum input voltage
0 to +50 VDC
Operating temperature and
humidity ranges
0 to 40°C (32 to 104°F), 80%RH or less (non-condensating)
Storage temperature and
humidity ranges
-10 to 50°C (14 to 122°F), 90%RH or less (non-condensating)
Operating environment
Altitude up to 2000 m (6562-ft.), indoors
Dimensions
Approx. 62W x 94H x 20D mm (2.44"W x 3.70"H x 0.79"D) (sans protrusions)
Connector cable length
Approx. 1500 mm (59")
Probe tip cable length
Approx. 300 mm (11.8")
Mass
Approx. 150 g (5.3 oz.) (including connector cable, excluding input leads)
Accessories
IC clip leads, Alligator clip leads, Carrying case, Instruction Manual
Digital input
Threshold value
Contact input
Detecting resistance value
1.4 V range
1.4 V±0.3 V
more than 1.5 kΩ
less than 500Ω
opened (Output L)
shorted (Output H)
2.5 V range
2.5 V±0.4 V
more than 3.5 kΩ
less than 1.5 kΩ
opened (Output L)
shorted (Output H)
4.0 V range
4.0 V±0.5 V
more than 25 kΩ
less than 8 kΩ
opened (Output L)
shorted (Output H)
5.2 Analog Input Section
86
Input module specifications are described here.
Measurement error caused by sig-
nal source impedance
• Errors occur when the signal source
impedance is greater than the input
impedance of the module.
• The input impedance of the Model
8936 Analog Unit is 1 MΩ. Errors of
about 0.1% occur when the signal
source impedance is 1 kΩ.
5.2
Analog Input Section
Measurement
Error
Es 1
Rin
Rs
Rin
+
-----------------------
–
[V]
Es: signal voltage
Rs: signal source impedance
Rin: input impedance
=
5.2.1
Model 8936 Analog Unit
Temperature and humidity
range for guaranteed
accuracy
Period of guaranteed
accuracy
23 ±5°C(73±9°F), 35 to 80% RH (when zero adjustment is executed 30 minutes after
power on)
1 year
Measurement ranges
5 m, 10 m, 20 m, 50 m, 100 m, 200 m, 500 m, 1, 2, 5, 10, 20 V/div
DC amplitude accuracy
±0.4%f.s.
Zero position accuracy
±0.1%f.s. (after zero adjustment)
Temperature characteristic
Gain: ±0.025%f.s./°C
Zero position: ±0.02%f.s./°C (after zero adjustment)
Frequency characteristic
DC coupling: DC to 400 kHz ±3 dB
AC coupling: 7 Hz to 400 kHz ±3 dB (low cut-off frequency: 7 Hz±20%)
Noise
450 μV p-p typ., 750 μV p-p max.
(sensitivity range, with input shorted) (S/N 2004-041018235 and later)
Common mode rejection
ratio
80 dB minimum (at 50/60 Hz and with signal source resistance 100 Ω maximum)
Low-pass filter
OFF, 5±50%, 500±50%, 5k±50%, 100k±50% (Hz) -3 dB
Input type
Unbalanced (input isolated from output)
Input coupling
DC, GND, AC
Input resistance
1 MΩ±1%
Input capacitance
30 pF±10 pF (at 100 kHz)
A/D resolution
12 bits
Voltage axis resolution
80 points/div (with 1X magnification)
Maximum sampling rate
1 MS/s (sampling period: 1 μs)
Input terminals
Insulated BNC terminal
Maximum input voltage
400 VDC
Maximum rated voltage to
earth
370 VAC/DC (between each input channel and chassis, and between input channels)
Operating temperature and
humidity ranges
Same as the host Memory HiCorder
Operating environment
Same as the host Memory HiCorder
Storage temperature and
humidity ranges
-10 to 50°C (14 to 122°F), 80% RH or less (non-condensating)
Effect of radiated radio-
frequency electromagnetic
field
±2%f.s. at 3 V/m max.
5.2 Analog Input Section
87
Effect of conducted radio-
frequency electromagnetic
field
±40%f.s. at 3 V max. (100 mV/div with 1 V DC input)
Dielectric strength
3.7 kVAC for 1 minute (between input module and chassis, and between input modules)
Dimensions
Approx. 170W x 20H x 148.5D mm (6.69”W x 0.79”H x 5.85”D) (sans protrusions)
Mass
Approx. 290 g (10.2 oz.)
Applicable Standards
Safety
EMC
EN 61010 Pollution degree 2,
Measurement category II (anticipated transient overvoltage 4000 V)
EN 61326 Class A
5.2 Analog Input Section
88
5.2.2
Model 8937 Voltage/Temp Unit
General specifications
Temperature and humidity
range for guaranteed accu-
racy
Period of guaranteed accu-
racy
23 ±5°C(73±9°F), 35 to 80% RH (when zero adjustment is executed 1 hour after power
on)
1 year
Common mode rejection ra-
tio
80 dB min. (at 50/60 Hz and with signal source resistance 100 Ω max.)
Input type
Unbalanced (input isolated from output)
A/D resolution
12 bits
Maximum sampling rate
1MS/s (However, update rate differs with temperature input.)
Maximum input voltage
30 V rms or 60 VDC (voltage and temperature inputs)
Maximum rated voltage to
earth
30 V rms or 60 VDC (voltage and temperature inputs)
Operating temperature and
humidity ranges
Same as the host Memory HiCorder
Operating environment
Same as the host Memory HiCorder
Storage temperature and
humidity ranges
-10 to 50°C (14 to 122°F), 80% RH or less (non-condensating)
Effect of radiated radio-
frequency electromagnetic
field
±2% f.s. at 3 V/m max. (in 5-mV/div range)
Effect of conducted radio-
frequency electromagnetic
field
±5% f.s. at 3 V max. (100 mV/div with 1 V DC input)
Dimensions
Approx. 170W x 20H x 148.5D mm (6.69”W x 0.79”H x 5.85”D) (sans protrusions)
Mass
Approx. 300 g (10.6 oz.)
Applicable Standards
Safety
EMC
EN 61010 Pollution degree 2,
Measurement category I (anticipated transient overvoltage 330 V)
EN 61326 Class A
Voltage input
Measurement ranges
500 μ, 1 m, 2 m, 5 m, 10 m, 20 m, 50 m, 100 m, 200 m, 500 m, 1, 2 V/div
DC amplitude accuracy
±0.4%f.s.
Zero position accuracy
±0.15%f.s. (after zero adjustment)
Temperature characteristic
Gain: ±0.02%f.s./°C, Zero position: ±0.03%f.s./°C
Frequency characteristic
DC to 400 kHz
dB (in 500 μV to 2 V/div range)
With digital filter ON: DC to 3 kHz
dB (Data update rate: 100 μs±20% (in 500 μV to 2
mV/div range))
Noise
75 μVp-p typ.,120 μVp-p max. with digital filter OFF (in 500 μV/div range)
20 μVp-p typ., 30 μVp-p max. with digital filter ON (in 500 μV/div range)
(S/N 2004-041135258 and later)
Input terminals
BNC terminal
Input resistance
1 MΩ±1%
Input capacitance
50 pF±20 pF (at 100 kHz)
Input coupling
DC, GND, AC
Low-pass filter
OFF, 5±50%, 500±50%, 5k±50%, 100k±50% (Hz) -3 dB
+1
−3
+1
−3
5.2 Analog Input Section
89
Temperature input
Measurement ranges
10, 20, 50, 100°C/div
Measurement input range
K: -200 to 1350°C
E: -200 to 800°C
J: -200 to 1100°C
T: -200 to 400°C
N: -200 to 1300°C
R: 0 to 1700°C
S: 0 to 1700°C
B: 300 to 1800°C
Temperature measurement
accuracy
K, E, J, T, N
R, S
B
±0.1% f.s.±1.0°C, ±0.1% f.s.±2.0°C (-200 to 0°C)
±0.1% f.s.±3°C
±0.1% f.s.±4°C (effective measurement range: 400 to 1800°C)
Reference junction compen-
sation
Selectable internal or external
Reference junction compen-
sation accuracy
±0.1% f.s.±1.5°C (with internal reference contact compensation and input terminal in
state of temperature equilibrium)
Temperature characteristic
±0.05%f.s./°C (sensor: K, E, J, T, N, with drift compensation mode OFF)
±0.25%f.s./°C(sensor:R,S,B, with drift compensation mode OFF)
±0.04%f.s./°C(all sensors, with drift compensation mode ON)
Frequency characteristic
DC to 1 kHz
dB
Data update rate: 250 μs±70% (with drift compensation mode OFF)
1 s±20% (with drift compensation mode ON)
Input terminals
2-terminal terminal block
Input resistance
5.1 MΩ±5%
Low-pass filter
OFF, 5±50%, 500±50% (Hz) -3 dB
+1
−3
5.2 Analog Input Section
90
5.2.3
Model 8938 FFT Analog Unit
Temperature and humidity
range for guaranteed
accuracy
Period of guaranteed
accuracy
23 ±5°C(73±9°F), 35 to 80% RH (when zero adjustment is executed 30 minutes after
power on)
1 year
The number of input chan-
nels
2 channels
Measurement ranges
5 m, 10 m, 20 m, 50 m, 100 m, 200 m, 500 m, 1, 2, 5, 10, 20 V/div
DC amplitude accuracy
±0.4%f.s.
Zero position accuracy
±0.1%f.s. (after zero adjustment)
Temperature characteristic
Gain: ±0.025%f.s./°C
Zero position: ±0.02%f.s./°C (after zero adjustment)
Frequency characteristic
DC coupling: DC to 400 kHz ±3 dB
AC coupling: 7 Hz to 400 kHz ±3 dB (low cut-off frequency: 7 Hz±20%)
Noise
500 μV p-p typ., 750 μV p-p max. (sensitivity range, with input shorted)
(S/N 2004-041132533 and later)
Common mode rejection
ratio
80 dB minimum (at 50/60 Hz and with signal source resistance 100 Ω maximum)
Low-pass filter
OFF, 5±50%, 500±50%, 5k±50%, 100k±50% (Hz) -3 dB
Anti-aliasing filter
Cutoff frequency (fc): 20, 40, 80, 200, 400, 800, 2 k, 4 k, 8 k, 20 k, 40 kHz
(These frequencies are automatically set when the anti-aliasing filter is set to ON) Atten-
uation characteristic: -66 dB min. at 1.5 fc
Input type
Unbalanced (input isolated from output)
Input coupling
DC, GND, AC
Input resistance
1 MΩ±1%
Input capacitance
30 pF±10 pF (at 100 kHz)
A/D resolution
12 bits
Voltage axis resolution
80 points/div (with 1X magnification)
Maximum sampling rate
1 MS/s (sampling period: 1 μs)
Input terminals
Insulated BNC terminal
Maximum input voltage
400 VDC
Maximum rated voltage to
earth
370 VDC (between each input channel and chassis, and between input channels)
Operating temperature and
humidity ranges
Same as the host Memory HiCorder
Storage temperature and
humidity ranges
-10 to 50°C (14 to 122°F), 80% RH or less (non-condensating)
Operating environment
Same as the Memory HiCorder in which the 8938 is installed
Effect of radiated radio-
frequency electromagnetic
field
±2%f.s. at 3 V/m max.
Effect of conducted radio-
frequency electromagnetic
field
±28%f.s. at 3 V max. (100 mV/div with 1 V DC input)
Dielectric strength
3.7 kVAC for 1 minute (between input module and chassis, and between input modules)
Dimensions
Approx.170W x 20H x 148.5D mm (6.69”W x 0.79”H x 5.85”D)(sans protrusions)
Mass
Approx.290g (10.2 oz.)
Applicable Standards
Safety
EMC
EN 61010 Pollution degree 2,
Measurement category II (anticipated transient overvoltage 4000 V)
EN 61326 Class A
5.2 Analog Input Section
91
5.2.4
Model 8939 Strain Unit
Temperature and humidity
range for guaranteed
accuracy
Period of guaranteed
accuracy
23 ±5°C (73±9°F), 35 to 80% RH (When executing auto-balancing one hour after power
on)
1 year
The number of input
channels
2 channels
Measurement ranges
20, 50, 100, 200, 500, 1000 με/div
DC amplitude accuracy
± (0.5%f.s. + 2 με)
Zero position accuracy
±0.5%f.s. (after auto-balancing)
Temperature characteristic
Gain: ±0.05%f.s./°C
Zero position: ±2 με/°C (20,50 με/div), ±0.1%f.s./°C (other ranges)
Frequency characteristic
DC to 20 kHz
dB
Appropriate adapter
Strain gauge adapter, Bridge resistance: 120 Ω to 1 kΩ
Bridge voltage
2 ± 0.05 V
Balancing
Electronic auto-balancing
Balance adjustment range
±10000 με max
Low-pass filter
OFF, 10±30%, 30±30%, 300±30%, 3k±30% (Hz) -3 dB
A/D resolution
12 bits
Maximum sampling speed
1 MS/s (sampling period: 1 μs)
Maximum input voltage
10 V (DC + AC peak)
Maximum rated voltage to
earth
30 Vrms or 60 VDC
Operational ranges for
temperature and humidity
Same as the host Memory HiCorder
Temperature and humidity
ranges for storage
-10 to 50°C (14 to 122°F), 80% RH or less (non-condensating)
Operating environment
Same as the host Memory HiCorder
Effect of radiated radio
frequency electromagnetic
field
±5%f.s. at 3 V/m max.
Effect of conducted radio-
frequency electromagnetic
field
±5%f.s. at 3 V max (20 με/div when +/− signal lines shorted together)
Dimensions
Approx. 170W x 20H x 148.5D mm (6.69”W x 0.79”H x 5.85”D )
Mass
Approx. 250 g (8.8 oz.)
Accessories
Conversion cable X 2 (Compatible sensor connector: PRC03-12A10-7M10.5 by TAJIMI)
Applicable Standards
Safety
EMC
EN 61010 Pollution degree 2,
Measurement category I (anticipated transient overvoltage 330 V)
EN 61326 Class A
+1
−3
5.2 Analog Input Section
92
5.2.5
Model 8940 F/V Unit
General Specifications
Temperature and humidity
range for guaranteed
accuracy
Period of guaranteed
accuracy
23 ±5°C(73±9°F), 35 to 80% RH (when zero adjustment is executed 30 minutes after
power on)
1 year
A/D resolution
12 bits
Vertical axis resolution
80 LSB/div
Measurement mode
Frequency measurement, Count, Duty, Voltage measurement, Current measurement
Maximum sampling period
1 μs
Low-pass filter
OFF, 5±50%, 500±50%, 5k±50%, 100k ±50% (Hz) -3 dB
Input coupling
DC, GND, AC (Fixed DC coupling except voltage and current measurement)
Pull up
ON/OFF (Constant OFF in current measurement) Pull up resistance: 10 kΩ
BNC terminal
Input resistance: 1 MΩ±1% (at pull-up OFF)
Input capacitance: 60 pF±20 pF (at 100 kHz)
Input type: Unbalanced (for Voltage, Frequency, Count, Duty)
Note: With the 3273 or 3273-50, the BNC connector and sensor connector are used to-
gether. GND is common with the Memory HiCorder in which the device is installed.
Sensor connector terminal
(Current measurement)
Possible to connect 8 ch
Note: With the 3273 or 3273-50, the BNC connector and sensor connector are used to-
gether. GND is common with the Memory HiCorder in which the device is installed.
Maximum input voltage
30 Vrms or 60 VDC
Maximum rated voltage to
earth
30 Vrms or 60 VDC (BNC)
Operational ranges for
temperature and humidity
Same as the host Memory HiCorder
Operating environment
Same as the host Memory HiCorder
Temperature and humidity
ranges for storage
-10 to 50°C (14 to 122°F) 80%RH or less (non-condensating)
Effect of radiated radio-
frequency electromagnetic
field
±5 %f.s. at 3 V/m max. (in 5-mV/div range)
Effect of conducted radio-
frequency electromagnetic
field
±28%f.s. at 3 V max. (100 mV/div with 1 V DC input)
Dimensions
Approx. 170 W x 20 H x 148.5 D mm(6.69"W x 0.79"H x 5.85"D)(sans protrusions)
Mass
Approx. 300 g (10.6 oz)
Applicable Standards
Safety
EMC
EN 61010 Pollution degree 2,
Measurement category I (anticipated transient overvoltage 330 V)
EN 61326 Class A
Options
9318 Conversion Cable (for 9270, 9271, 9272, 9277, 9278, 9279)
9319 Conversion Cable (for 3273, 3273-50)