Viessmann VitoAid Operating Instructions Manual
Read Online
Read the Viessmann VitoAid Operating Instructions Manual online using the built-in PDF viewer.
Text of the manual
for the system user
VIESMANN
Ventilator for continuous respiratory support
VITOAID
6152712 GB
6/2020
Please keep safe.
Please follow these safety instruc-
tions closely to prevent accidents
and material losses.
Safety instructions explained
Danger
This symbol warns against the risk
of injury.
!
Please note
This symbol warns against the risk
of material losses and environmen-
tal pollution.
Note
Details identified by the word "Note"
contain additional information.
Target group
These operating instructions are inten-
ded for system users.
Before using the device, the operator
must have fully read and understood
these operating instructions.
The ventilator may only be used under
the direction of a physician by trained
specialist medical personnel with knowl-
edge of mechanical ventilation.
The ventilator is not intended for use by
patients, children, individuals with
reduced physical, sensory or mental fac-
ulties, or persons lacking the appropriate
experience and knowledge.
Safety instructions for working with the device
Connecting the device
■The device may only be put into opera-
tion by trained specialist medical per-
sonnel with knowledge of mechanical
ventilation.
■The device may only be connected to
the power supply at permanently instal-
led 230 V standard sockets using the
factory-installed power cable with
standard plug.
■Connection via extension cables or
sockets switched by means of wall or
dimmer switches is not permissible.
■Modifications to the existing installation
may only be carried out by authorised
specialists.
Danger
Incorrectly performed connections
can lead to life threatening condi-
tions for the ventilated patient.
The breathing tube of the VentStar
Oxylog VE300 single-tube system
and the oxygen supply tube may
only be connected by trained spe-
cialist medical personnel with
knowledge of mechanical ventila-
tion.
Safety instructions
For your safety
6152712
Working on the device
■All settings and work on the device
must be performed as specified in
these operating instructions.
No other work on or modifications to
the device are permissible.
■Do not open the device.
■Do not remove casings.
■Do not modify or remove attachments.
Safety instructions for operating the device
Danger
Incorrect settings and use of the
wrong components can put the ven-
tilated patient at risk.
■Ventilation must only be provided
via the VentStar Oxylog VE300
single-tube system, 1.5 metres
(standard delivery) and HME filter.
■The device may only be operated
with an externally controllable oxy-
gen supply.
■Regularly check tube connections
during operation. Do not discon-
nect tube connections.
Danger
Contamination and dirt on the
device and on the tubes can harm
the ventilated patient.
■Check connections and tubes
daily for contamination and dirt.
■Immediately remove any fluids
and biological matter.
■Keep the air inlet grille on the
device clear at all times. Before
every use, check the air inlet grille
for contamination and dirt.
Permissible ambient conditions
!
Please note
Incorrect ambient conditions can
result in damage to the device and
put safe operation at risk.
■Only use the device in dry,
enclosed spaces and at ambient
temperatures between 5 °C and
35 °C.
■Do not use the device in MRT
environments.
■The device must not be used in
hyperbaric chambers.
General information on the safe use and handling of oxygen
Source: Bundesanstalt für Materialfor-
schung und -prüfung [German Federal
Institute for Materials Testing and
Research], Berlin
Safety instructions
For your safety (cont.)
6152712
General safety instructions
■Do not allow oxygen to flow into
enclosed rooms/spaces as this can
lead to an oxygen-enriched atmos-
phere and consequently may present a
fire hazard.
■Keep away from open flames and
sources of ignition.
Safety instructions for medical per-
sonnel and users
■Oxygen must only be handled by
trained or instructed personnel
■Secure oxygen bottles to prevent them
from tipping over or falling down.
■Always open oxygen valves slowly.
■Keep oxygen fittings and system com-
ponents free of oil and grease at all
times.
■On completion of work, close all oxy-
gen fittings and valves.
■Improper use of oxygen is prohibited,
e.g. using it instead of compressed air
or for cooling or cleaning clothing
■Under no circumstances should leak-
ing or faulty oxygen bottles, fittings or
system components be used.
Safety instructions
For your safety (cont.)
6152712
Standard delivery ...................................................................................
6
Intended use ..........................................................................................
6
Your device is preset at the factory ........................................................
7
2.
Device overview
Overview of connections and controls ...................................................
8
3.
Connecting VitoAid
Connecting the oxygen supply and breathing tube ...............................
9
Adjusting the oxygen supply .................................................................. 10
4.
Switching on and off
Switching on VitoAid .............................................................................. 11
Switching off VitoAid ..............................................................................
11
5.
Where to find the controls
Operating VitoAid ................................................................................... 12
■Screen displays .................................................................................. 12
6.
Switching ventilation on
and off
Switching on ventilation ......................................................................... 14
Switching off ventilation ......................................................................... 14
7.
Setting the ventilation
parameters
Selecting the ventilation mode ............................................................... 15
■Pressure-controlled ventilation (PCV) ................................................
15
■High-flow CPAP (continuous positive airway pressure) ...................... 15
Setting the positive end-expiratory pressure (PEEP) ............................
15
Setting the inspiratory pressure ............................................................. 15
Setting the respiratory rate ....................................................................
16
Setting the ratio of inspiratory time to expiratory time ............................ 16
8.
Setting alarm limits
Setting the lower limit for respiratory volume ......................................... 18
Setting the upper limit for respiratory volume ........................................ 18
9.
Further settings
Restoring factory settings ...................................................................... 19
10.
Checks
Checking values ....................................................................................
20
11.
Alarms and fault mes-
sages
Acknowledging alarms and fault messages ........................................... 21
12.
What to do if...
is shown on the display. .................................................................... 22
is shown on the display. .................................................................... 22
is shown on the display. .................................................................... 22
is shown on the display. .................................................................... 22
is shown on the display. .................................................................... 22
is shown on the display. .................................................................... 22
is shown on the display. .................................................................... 22
and a numeric code > are shown on the display. ........................... 23
13.
Maintenance
Cleaning the device surface ..................................................................
24
Cleaning the air inlet grille .....................................................................
24
14.
Service
VitoAid Hotline ....................................................................................... 25
Index
Index
6152712
Reference to other document containing
further information
1.
Step in a diagram:
The numbers correspond to the order in
which the steps are carried out.
Warning of material losses and environ-
mental pollution
Live electrical area
Pay particular attention.
■Component must audibly click into place.
or
■Acoustic signal
■Fit new component.
or
■In conjunction with a tool: Clean the sur-
face.
Dispose of component correctly.
Dispose of component at a suitable collec-
tion point. Do not dispose of component in
domestic waste.
Standard delivery
■VitoAid ventilator
■VentStar Oxylog VE300 single-tube system, length:
1.5 metres
Intended use
The VitoAid ventilator is designed to provide continu-
ous respiratory support to adult patients who require
mechanical ventilation.
The VitoAid ventilator may only be used in emergency
situations where the capacities of the existing supplies
of approved ventilators are not sufficient.
The ventilator is intended solely for use in care facili-
ties and hospitals for invasive and non-invasive posi-
tive pressure ventilation.
Intended use requires the following:
■Ventilation with external oxygen supply:
The ventilator is connected via a silicone tube to a
supply of medical oxygen with pressure control.
■Ventilation without external oxygen supply:
The tube connection for the oxygen supply to the
ventilator is closed (delivered condition).
Introductory information
Symbols
6152712
■The patient is supplied via the VentStar Oxylog
VE300 single-tube system, 1.5 metres (standard
delivery) and HME filter. The HME filter must be
positioned close to the patient.
■The supply tubes comply with the specification for
ventilators.
Usage for a purpose other than invasive and non-inva-
sive positive pressure ventilation shall be deemed
inappropriate.
Incorrect usage or operation of the device (e.g. open-
ing of the device) is prohibited and will result in an
exclusion of liability. Incorrect usage also occurs if the
components of the ventilator or the tubes connected to
it are modified from their intended function.
Your device is preset at the factory
The ventilator is preset at the factory for pressure-con-
trolled ventilation (PCV) with the following parameters:
You can change these settings individually at any time:
See page 15 onwards.
Ventilation parameters
■The positive end-expiratory pressure (PEEP) is set
to 10 millibars.
■The inspiratory pressure is set to 25 millibars.
■The respiratory rate is set to 15 breaths per minute.
■The ratio of inspiratory time to expiratory time is set
to 1 : 2.
Alarm limits
■The alarm limit for expiratory pressure equates to the
setting for positive end-expiratory pressure (PEEP).
■The alarm limit for inspiratory pressure is 45 milli-
bars.
■The alarm limit for the lower respiratory volume is set
to 0. This means that monitoring for this value is
deactivated.
■The alarm limit for the upper respiratory volume is
set to 999. This means that monitoring for this value
is deactivated.
Note
The alarm limits are only active if PCV (pressure-con-
trolled ventilation) is set: See page 15.
Power failure
All settings are retained if there is a power failure.
Introductory information
Intended use (cont.)
6152712
A Connection for breathing tube of the VentStar Oxy-
log VE300 single-tube system (standard delivery)
■Internal 7 of breathing tube: 22 mm
■Tube length: 1.5 metres
B Tube connection for oxygen supply
■Internal 7: 6 mm
C Transport handle
D ON/OFF switch
E Buzzer for acoustic signals
F Connection for factory use only: Do not connect
anything here!
G Status LED
H Operating unit with 2-part display and operating
buttons
K Air inlet grille: It is essential to keep this grille clear.
Danger
If it is not possible to draw in a sufficient amount
of air through the air inlet grille, the ventilation
parameters may deviate from the set values.
This could potentially harm the patient.
■Keep the air inlet grille clear at all times.
■Before using the device, always check the air
inlet grille for contamination and dirt.
■If required, clean the air inlet grille with a vac-
uum cleaner.
Device overview
Overview of connections and controls
6152712
A Connection for breathing tube
B Tube connection for oxygen supply
C Oxygen supply tube
■Internal 7: 6 millimetres
D VentStar Oxylog VE300 single-tube system with
HME filter
■Internal 7: 22 millimetres
■Tube length: 1.5 metres
1. Pull the protective caps off tube connections A
and B.
2. Push oxygen supply tube C onto tube connection
B.
3. Push the breathing tube of the VentStar Oxylog
VE300 single-tube system with expiration valve
and HME filter D onto tube connection A.
4. Secure both tubes, e.g. with tube clips, to prevent
accidental disconnection.
5. Secure the ventilator to prevent it from accidentally
tipping over.
Connecting VitoAid
Connecting the oxygen supply and breathing tube
6152712
The inspiratory oxygen concentration is adjusted via
the flow rate of the oxygen supply according to the
curve shown below.
Note
To measure the oxygen flow rate, a flow meter needs
to be integrated into the oxygen supply.
80
60
20
0
100
Oxygen flow rate in l/min
oxygen concentration in %
5
10
15
20
25
40
Inspiratory
Fig. 3
Curve for adjusting the oxygen supply
Connecting VitoAid
Adjusting the oxygen supply
6152712
■The breathing tube of the VentStar Oxylog VE300
single-tube system is correctly connected to the ven-
tilator.
For ventilation with external oxygen supply:
■The oxygen supply tube (if present) is correctly con-
nected to the ventilator.
■The flow rate for the oxygen supply is set to the
required respiratory air oxygen concentration: See
previous section.
1. Insert the power plug into a standard socket.
2. Switch on the ON/OFF switch (position "I").
After approx. 4 seconds, a value is shown on the
operating unit display.
This indicates that the ventilator is ready for opera-
tion.
Switching off VitoAid
Ventilation of the patient has ended.
1. Switch off the ON/OFF switch (position "0").
The display on the operating unit goes blank.
2. Pull the power plug out of the standard socket.
Switching on and off
Switching on VitoAid
6152712
All ventilator settings can be performed centrally on the
operating unit.
Fig. 4
"Status"
Display area for selecting information, e.g.
for the positive end-expiratory pressure
(PEEP)
"Service" Display area for showing the currently set
value
To scroll up through the display or increase a
set value.
To scroll down through the display or reduce
a set value.
Only if an alarm or device fault is present:
Briefly press the button:
■This switches off the acoustic alarm and
fault signal.
The alarm or device fault remains active.
Press the button for at least 2 seconds:
■This acknowledges an alarm or fault mes-
sage: The device's internal fault memory is
deleted.
■In the event of device faults, this will also
reset the operating unit's control electron-
ics.
All settings are retained.
For further information: See page 21.
Briefly press the button:
■To switch ventilation on or off.
To select a menu or confirm a changed set-
ting.
Press the button for at least 2 seconds:
■The numeric code is shown in the "Serv-
ice" display area.
You are now in the settings menu.
Note
To reset the ventilator to the factory settings, press
and
simultaneously for more than 2 seconds.
Screen displays
The display screen is split into two areas, the "Status"
area (left) and the "Service" area (right).
Different letters can be shown in the "Status" display
area. These letters indicate which values are currently
being shown or changed in the "Service" display area.
The level of information shown depends on the set
ventilation method. To set the ventilation method: See
page 15.
"Status"
Information displayed in "Service"
Ventilation method: See
page 15.
PCV
CPAP
■Ventilation is switched on:
Current expiratory pressure
■When making settings:
Positive end-expiratory pressure (PEEP) in millibars: See
page 15.
X
X
■Ventilation is switched on:
Current inspiratory pressure
■When making settings:
Set value for inspiratory pressure in millibars: See page 15.
X
—
Set respiratory rate: Number of set respiratory cycles per minute,
see page 16.
X
—
Where to find the controls
Operating VitoAid
6152712
Information displayed in "Service"
Ventilation method: See
page 15.
PCV
CPAP
Ratio of inspiratory time to expiratory time
Note
The values are shown without a decimal point and are relative to
the inspiratory time.
Example:
Displayed value: "17"
The ratio of inspiratory time to expiratory time is 1 : 1.7.
Example:
Displayed value: "05"
The ratio of inspiratory time to expiratory time is 1 : 0.5 = 2 : 1
For an overview of possible settings: See page 17.
X
—
Numeric code for faults or alarms: See page 22.
X
X
Ventilation is switched on:
Current respiratory volume in millilitres
X
—
When making settings:
Lower alarm limit for the respiratory volume in millilitres
X
—
When making settings:
Upper alarm limit for the respiratory volume in millilitres
X
—
Note
"- - -" in the "Service" display area indicates that inter-
nal data is being processed. This display disappears
automatically as soon as the process is complete.
Where to find the controls
Operating VitoAid (cont.)
6152712
■The ventilator is switched on.
■The patient can receive the appropriate supply via
the breathing tube.
For ventilation with external oxygen supply:
■Oxygen is supplied at the required flow rate.
Press
.
The patient is ventilated using the set ventilation
method and the set values.
Note
■If ventilation does not start, reset the control elec-
tronics.
To do so, press
for more than 2 seconds. All set
values are retained.
■You must wait at least 2 seconds before you can
switch ventilation off again.
Switching off ventilation
Ventilation is switched on.
1. Press
.
The patient is no longer ventilated.
The status LED illuminates for 30 seconds.
2. Shut off the oxygen supply, for example by closing
the relevant external valve.
Note
You must wait at least 2 seconds before you can
switch ventilation on again.
Switching ventilation on and off
Switching on ventilation
6152712
VitoAid supports PCV (pressure-controlled ventilation)
and high-flow CPAP (continuous positive airway pres-
sure).
Pressure-controlled ventilation (PCV)
The patient is ventilated in controlled mode in accord-
ance with the set values for inspiratory pressure, posi-
tive end-expiratory pressure (PEEP), respiratory rate
and the ratio of inspiratory time to expiratory time.
Spontaneous ventilation is neither detected nor assis-
ted here.
To activate pressure-controlled ventilation, set the
inspiratory pressure higher than the positive end-
expiratory pressure (PEEP).
High-flow CPAP (continuous positive airway pressure)
With this method, a continuous supply of air/oxygen is
administered via a suitable mask to generate a positive
end-expiratory pressure (PEEP), which provides respi-
ratory support to a spontaneously breathing patient.
To activate high-flow CPAP, set identical values for
the inspiratory pressure and the positive end-expira-
tory pressure (PEEP).
If ventilation is switched on, only the expiratory pres-
sure ( ) is shown on the display.
The other information is not relevant for this ventilation
method and is therefore not displayed: See page 12.
Note
With this ventilation method, monitoring of the alarm
limits is switched off.
Setting the positive end-expiratory pressure (PEEP)
The settings for the positive end-expiratory pressure
(PEEP) and the inspiratory pressure are interdepend-
ent. The difference between the two pressures must
not exceed 25 millibars.
Example:
You have set the positive end-expiratory pressure
(PEEP) to 15 millibars.
You cannot set the inspiratory pressure higher than
40 millibars.
If the inspiratory pressure was previously set to a
higher value, this setting will be automatically reset to
40 millibars.
1. Press
for longer than 2 seconds.
is shown in the "Service" display area.
2. Use
or
to select the letter in the "Status" dis-
play area.
3. Press
.
The value shown in the "Service" display area
flashes.
4. Use
or
to set the required pressure in milli-
bars in the "Service" display area.
The setting is made in millibars.
5. To accept the set value, press
again.
Note
The ventilator regulates the respiratory pressure to the
set value. If this value cannot be achieved for pres-
sure-controlled ventilation, an alarm is triggered.
Setting the inspiratory pressure
The settings for the positive end-expiratory pressure
(PEEP) and the inspiratory pressure are interdepend-
ent. The difference between the two pressures must
not exceed 25 millibars.
Example:
You have set the positive end-expiratory pressure
(PEEP) to 15 millibars.
Setting the ventilation parameters
Selecting the ventilation mode
6152712
You cannot set the inspiratory pressure higher than
40 millibars.
If the inspiratory pressure was previously set to a
higher value, this setting will be automatically reset to
40 millibars.
1. Press
for longer than 2 seconds.
is shown in the "Service" display area.
2. Use
or
to select the letter in the "Status" dis-
play area.
3. Press
.
The value shown in the "Service" display area
flashes.
4. Use
or
to set the required pressure in milli-
bars in the "Service" display area.
The setting is made in millibars.
Note
The maximum setting is 45 millibars.
5. To accept the set value, press
again.
Note
The ventilator regulates the respiratory pressure to the
set value. If this value cannot be achieved for pres-
sure-controlled ventilation, an alarm is triggered.
Setting the respiratory rate
The respiratory rate is set by specifying the number of
respiratory cycles per minute.
1 respiratory cycle comprises 1 inspiration process and
1 expiration process.
1. Press
for longer than 2 seconds.
is shown in the "Service" display area.
2. Use
or
to select the letter in the "Status" dis-
play area.
3. Press
.
The value shown in the "Service" display area
flashes.
4. Use
or
to set the required respiratory rate
(respiratory cycles per minute) in the "Service" dis-
play area.
5. To accept the set value, press
again.
Setting the ratio of inspiratory time to expiratory time
This setting enables you to specify the ratio of inspira-
tory time to expiratory time.
1. Press
for longer than 2 seconds.
is shown in the "Service" display area.
2. Use
or
to select the letter in the "Status" dis-
play area.
3. Press
.
The value shown in the "Service" display area
flashes.
Setting the ventilation parameters
Setting the inspiratory pressure (cont.)
6152712
to set the required ratio of inspiratory
time to expiratory time in the "Service" display
area, in accordance with the following table.
Note
The values are shown without a decimal point and
are relative to the inspiratory time.
Setting val-
ue
Ratio of inspiratory time to expira-
tory time
05
1 : 0.5 = 2 : 1
06
1 : 0.6 = 1.67 : 1
07
1 : 0.7 = 1.43 : 1
08
1 : 0.8 = 1.25 : 1
09
1 : 0.9 = 1.11 : 1
10
1 : 1
11
1 : 1.1
12
1 : 1.2
13
1 : 1.3
14
1 : 1.4
15
1 : 1.5
16
1 : 1.6
17
1 : 1.7
...
...
30
1 : 3
5. To accept the set value, press
again.
Setting the ventilation parameters
Setting the ratio of inspiratory time to… (cont.)
6152712
If the measured respiratory volume falls below the set
lower limit, an alarm is triggered: See page 21.
Note
■This alarm limit is only active if PCV (pressure-con-
trolled ventilation) is set: See page 15.
■To deactivate monitoring of the lower limit for respira-
tory volume, set the value to "0".
1. Press
for longer than 2 seconds.
is shown in the "Service" display area.
2. Use
or
to select the symbol in the "Status"
display area.
3. Press
.
The value shown in the "Service" display area
flashes.
4. Use
or
to set the required lower alarm limit for
respiratory volume in millilitres in the "Service" dis-
play area.
5. To accept the set value, press
again.
Setting the upper limit for respiratory volume
If the measured respiratory volume exceeds the set
upper limit, an alarm is triggered: See page 21.
Note
■This alarm limit is only active if PCV (pressure-con-
trolled ventilation) is set: See page 15.
■To deactivate monitoring of the upper limit for respi-
ratory volume, set the value to "999".
1. Press
for longer than 2 seconds.
is shown in the "Service" display area.
2. Use
or
to select the symbol in the "Status"
display area.
3. Press
.
The value shown in the "Service" display area
flashes.
4. Use
or
to set the required upper alarm limit
for respiratory volume in millilitres in the "Service"
display area.
5. To accept the set value, press
again.
Setting alarm limits
Setting the lower limit for respiratory volume
6152712
To reset the ventilator to the factory settings, press
and
simultaneously for more than 2 seconds.
The two display areas together show the following
combination of letters:
(= Parameter dELeted)
For factory settings: See page 7.
Further settings
Restoring factory settings
6152712
You can call up current respiratory pressures and the
set respiratory parameters on the display.
Use
or
to select the required letter, e.g. , in the
"Status" display area.
A list of the information and letters which can be dis-
played is given in the table on page 12.
Checks
Checking values
6152712
The alarm limits are only active if PCV (pressure-con-
trolled ventilation) is set: See page 15.
You can recognise alarms and faults on the ventilator
as follows:
■ and a numeric code are shown on the display.
■The Status LED illuminates.
■An acoustic signal sounds.
Danger
Where possible, ventilation of the patient contin-
ues during alarms and device faults. However,
the patient may not be ventilated with the opti-
mum parameters. The patient being ventilated
may be harmed as a result.
■Rectify the cause of the alarm or device fault
immediately.
■Acknowledge the alarm or fault message.
Switching off the acoustic signal
Briefly press
.
■The acoustic signal stops, but the alarm or fault mes-
sage remains active.
■ and the numeric code are shown on the display.
■The Status LED illuminates.
Acknowledging the alarm or fault message
1. Rectify the cause of the fault or the alarm: See
page 22.
2. To acknowledge the alarm or fault on the ventilator,
press
for at least 2 seconds.
■ and the numeric code are no longer displayed.
■The Status LED goes off.
■If not switched off beforehand, the acoustic sig-
nal stops.
Note
If you acknowledge the alarm or fault on the device
without rectifying the cause, the alarm or fault will
most likely occur again within a short time.
Alarms and fault messages
Acknowledging alarms and fault messages
6152712
During ventilation, the expiratory pressure falls below
the set value for positive end-expiratory pressure
(PEEP).
■Check the tube connections and tubes for leaks.
■Check the air inlet grille for contamination and dirt. If
necessary, clean this grille with a vacuum cleaner.
is shown on the display.
Cause
Remedy
During ventilation, the expiratory pressure does not
reach the set value for positive end-expiratory pressure
(PEEP).
■Check the tube connections and tubes for leaks.
■Check the air inlet grille for contamination and dirt. If
necessary, clean this grille with a vacuum cleaner.
is shown on the display.
Cause
Remedy
During ventilation, the set inspiratory pressure is not
reached.
■Check the tube connections and tubes for leaks.
■Check the air inlet grille for contamination and dirt. If
necessary, clean this grille with a vacuum cleaner.
is shown on the display.
Cause
Remedy
During ventilation, the set inspiratory pressure is ex-
ceeded.
■Check the patient.
■Check whether the tubes are bent or constricted.
is shown on the display.
Cause
Remedy
During ventilation, the inspiratory pressure exceeds
45 millibars.
■Check the patient.
■Check whether the tubes are bent or constricted.
■Check the setting for the inspiratory pressure. If nec-
essary, adjust this pressure.
is shown on the display.
Cause
Remedy
The measured respiratory volume has fallen below the
set lower limit.
Check the ventilation parameter settings. If necessary,
adjust these parameters: See page 15.
is shown on the display.
Cause
Remedy
The measured respiratory volume has exceeded the
set upper limit.
Check the ventilation parameter settings. If necessary,
adjust these parameters: See page 15.
What to do if...
is shown on the display.
6152712
There is a fault in the control electronics.
Press
for more than 2 seconds.
The control electronics are reset.
All set values are retained.
If the fault persists after resetting the control electron-
ics:
■Turn off the ON/OFF switch.
■Wait for approx. 5 seconds.
■Turn on the ON/OFF switch again.
If the fault persists after switching the ventilator off and
back on again:
■Pull the power plug out of the standard socket.
■Wait for approx. 5 seconds.
■Re-insert the power plug.
If the fault persists continuously:
Stop using the ventilator.
Contact the seller from whom you purchased the de-
vice.
What to do if...
and a numeric code > are shown on the display.
6152712
You can clean the surfaces of the device with a com-
mercially available domestic cleaning agent. Do not
use scouring agents.
Clean the surface of the operating unit with a microfi-
bre cloth.
Cleaning the air inlet grille
Danger
If it is not possible to draw in a sufficient amount
of air through the air inlet grille, the ventilation
parameters may deviate from the set values.
This could potentially harm the patient.
■Keep the air inlet grille clear at all times.
■Before using the device, always check the air
inlet grille for contamination and dirt.
■If required, clean the air inlet grille with a vac-
uum cleaner.
Maintenance
Cleaning the device surface
6152712
Call the VitoAid Hotline on the following number if you
have any questions or require maintenance or repair
work on your ventilator:
+49 6452 70 1661
Service
VitoAid Hotline
6152712
Contact your local contractor if you have any questions about your system or wish to arrange maintenance or
repair work. You can find local contractors on the internet at www.viessmann.de.
Viessmann Limited
Hortonwood 30, Telford
Shropshire, TF1 7YP, GB
Telephone: +44 1952 675000
Fax: +44 1952 675040
E-mail: [email protected]
Viessmann Werke GmbH & Co. KG
D-35107 Allendorf
Telephone: +49 6452 70-0
Fax: +49 6452 70-2780
www.viessmann.com
6152712
Subject to technical modifications.