Valco Cincinnati LLD Manual
Read Online
Read the Valco Cincinnati LLD Manual online using the built-in PDF viewer.
Text of the manual
Low Level Detection Alarm/
Low Level Detection Alarm/
Low Level Detection Alarm/
Low Level Detection Alarm/
Low Level Detection Alarm/
AutoRefill Systems
AutoRefill Systems
AutoRefill Systems
AutoRefill Systems
AutoRefill Systems
© 2011 Valco Cincinnati, Inc.
All Rights Reserved.
This manual is furnished with the Low Level Detection Alarm/AutoRefill Systems and may be used or
copied only in accordance with the terms of purchase.
No part of this manual may be reproduced, stored in a retrieval system, or transmitted, in any form or by
any means, electronic, mechanical, recording, or otherwise, without the prior written permission of Valco
Cincinnati, Inc. The information in this manual is furnished for educational purposes only, is subject to
change without notice, and should not be construed as a commitment by Valco Cincinnati, Inc.
This manual written and designed at Valco Cincinnati, Inc., 411 Circle Freeway Drive, Cincinnati, Ohio
45246.
Part Number: MC036
Printed in the USA
To order parts, please contact your closest Valco office by mail, phone, or Email:
Valco Cincinnati, Inc.
Valco Cincinnati, Inc.
Valco Cincinnati, Inc.
Valco Cincinnati, Inc.
Valco Cincinnati, Inc.
411 Circle Freeway Drive
Cincinnati, OH 45246
TEL:
(513) 874-6550
FAX:
(513) 874-3612
Email:
[email protected]
Web:
http://www.valcomelton.com
Valco Cincinnati Limited
Valco Cincinnati Limited
Valco Cincinnati Limited
Valco Cincinnati Limited
Valco Cincinnati Limited
Hortonwood 32
Telford, TFI 7YN, England
TEL:
(+44) 1952-677911
FAX:
(+44) 1952-677945
Email:
[email protected]
Web:
http://www.valco.co.uk
Valco Cincinnati GmbH
Valco Cincinnati GmbH
Valco Cincinnati GmbH
Valco Cincinnati GmbH
Valco Cincinnati GmbH
Storkower Strasse 6
D-15749 Gallun, Germany
TEL:
(+49) 337 648 700
FAX:
(+49) 337 648 7070
Email:
[email protected]
Melton S.L.U.
Melton S.L.U.
Melton S.L.U.
Melton S.L.U.
Melton S.L.U.
Pol. Industrial Agustinos
calle G, n. 34
31160 Orcoyen, Navarra, Spain
TEL: (34) 948-321-580
FAX: (34) 948-326-584
Web:
http://www.valcomelton.com
Declaration of Conformity
Declaration of Conformity
Declaration of Conformity
Declaration of Conformity
Declaration of Conformity
(According to EN 45014)
This Declaration of Conformity has been generated electronically and is legally binding without signature
Manufacturer:
Valco Melton
A division of Valco
Cincinnati, Incorporated
411 Circle Freeway Drive
Cincinnati, OH 45246
USA
Authorized Representatives in
Europe:
Valco Cincinnati, Ltd.
Unit 7-8
Hortonwood 32
Telford TF1 7YN
England
Melton S.L.U
Pol. Industrial
Agustinos
calle G, n. 34
31160 Orcoye
Navarra, Spain
declares that the product:
Product Name:
Low Level Detection
Alarm/AutoRefill Systems
complies with the following Council Directives:
Safety of Machinery:
2006/42/EC
Low Voltage Equipment:
2006/95/EC
EMC:
2004/108/EC
and conforms to the following standards:
Safety:
EN60204-1:2006
EN13849-1
Risk:
EN14121-1:2007
EMC Emissions:
EN61000-6-4:2007
EN61000-4-2
EMC Immunity:
EN61000-6-2:2005
EN61000-4-3
EN61000-4-4
EN61000-4-5
EN61000-4-6
EN61000-4-8
EN61000-4-11
Place and Date:
Cincinnati, Ohio USA
CE Mark first fixed 2006
Signature:
David H. Swedes,
Director of Engineering &
Manufacturing
Valco Cincinnati, Inc.
TABLE OF CONTENTS
TABLE OF CONTENTS
TABLE OF CONTENTS
TABLE OF CONTENTS
TABLE OF CONTENTS
Section 1 - Introduction ................................................. 1-1
About this Manual .................................................................................................... 1-1
Product Description................................................................................................. 1-1
Section 2 - Safety and Use ............................................. 2-1
Read Thoroughly Before Handling Equipment ..................................................... 2-1
Symbols .................................................................................................................... 2-1
Owner Responsibilities ........................................................................................... 2-3
Limitations of Use .................................................................................................... 2-4
Installation/Startup/Use Safety Information .......................................................... 2-4
Shut Down Safety Information ................................................................................ 2-5
Hot Melt Specific General Safety Information ....................................................... 2-6
What to Do if Contact with Hot Adhesive Occurs ...................................................................2-7
What to Do if Inhalation of Adhesive Fumes Occurs..............................................................2-7
What to Do if Adhesive-Related Fire or Explosion Occurs ....................................................2-8
Hose Safety Information.......................................................................................... 2-9
Section 3 - Installation .................................................... 1-1
LLD Alarm Installation ............................................................................................. 1-1
Mounting the Control Box ........................................................................................................1-1
Electrical Installation of the Control Box ................................................................................1-1
Mechanical Installation of the Probe .......................................................................................1-3
Electrical Installation of the Probe ...........................................................................................1-6
Testing the LLD Alarm System .................................................................................................1-7
The AutoRefill LLD Installation ................................................................................................1-8
Mechanical Installation of the Probe (non-pressurized tanks) ............................................ 1-11
Section 4 - Glue Source and Probe Installation ........... 4-1
LLD Alarm Assembly Drawings .............................................................................. 4-1
LLD AutoRefill System Assembly Drawings ......................................................... 4-6
Section 5 - Maintenance ................................................. 5-1
Introduction .............................................................................................................. 5-1
General Maintenance ............................................................................................... 5-1
Section 6 - Troubleshooting........................................... 6-1
Introduction .............................................................................................................. 6-1
Section 7 - Specifications............................................... 7-1
Valco Cincinnati, Inc.
Table of Contents
MC036 - Low Level Detection Alarm/AutoRefill Systems
Section 8 - part number list ............................................ 8-1
How to Order Parts .................................................................................................. 8-1
Example: Auto Pot Fill System Layout (080xx377) ............................................... 8-2
The LLD Alarm Assembly Parts List ...................................................................... 8-4
LLD Control Assembly (080xx545) ...........................................................................................8-4
Installation Kit (781xx258) ....................................................................................... 8-7
Probe Assemblies .................................................................................................... 8-7
The LLD AutoRefill Assembly Parts List................................................................ 8-8
LLD AutoRefill Control Assembly (098xx699) .........................................................................8-8
Pot Fill Kit (754xx330) ............................................................................................ 8-11
Section 9 - Warranty........................................................ 9-1
Warranty Information............................................................................................... 9-1
Cold Glue Equipment and Electronic Controls.......................................................................9-1
Hot Melt Units, Hoses, Valves, Guns, and Related Equipment ..............................................9-1
Section 10 - Service ...................................................... 10-1
SECTION 1 - INTRODUCTION
SECTION 1 - INTRODUCTION
SECTION 1 - INTRODUCTION
SECTION 1 - INTRODUCTION
SECTION 1 - INTRODUCTION
The Valco LLD alarm system provides solid-state electronic surveillance of a
gluing system’s glue supply. There are two types of LLD alarm systems
available. The first is the traditional alarm system, the LLD Alarm System,
that alerts the operator that the glue level is low. The operator must then refill
the glue manually.
The traditional LLD alarm system is activated when the glue supply in the
tank drops below a pre-established level. It contains no moving parts. The
LLD alarm system consists of the following components:
•
A low-level detection probe, which is inserted into the glue
supply tank (reservoir)
•
An electronics control box, which incorporates both a visual
and an audio annunciator.
The LLD alarm system offers a dual-level warning capability:
•
Warning level
Warning level
Warning level
Warning level
Warning level—The warning level is indicated by a flashing
light that alerts the operator that the glue level is low.
•
Low level (for non-pressurized drums and vats only)
Low level (for non-pressurized drums and vats only)
Low level (for non-pressurized drums and vats only)
Low level (for non-pressurized drums and vats only)
Low level (for non-pressurized drums and vats only)—The
low level is indicated by an audio alarm that alerts the
operator that the glue level is critically low. When the low-
level alarm sounds, the glue tank should be refilled
immediately.
The activation depth of both alarms can
be easily adjusted in the field.
The second type of LLD alarm system, the AutoRefill LLD System, does not
require manual refilling. Instead of an alarm being sounded, the low level
detection triggers the automatic refilling pump to refill the glue pot. This LLD
configuration is for non-pressurized drums and vats only.
About this
Manual
Valco Cincinnati has prepared this manual as an aid for installing, operating,
and servicing the low level detection (LLD) alarm system. It also contains a
list of part numbers for replacement parts.
Product
Description
SECTION 2 - SAFETY AND USE
SECTION 2 - SAFETY AND USE
SECTION 2 - SAFETY AND USE
SECTION 2 - SAFETY AND USE
SECTION 2 - SAFETY AND USE
Read
Thoroughly
Before
Handling
Equipment
Warning!
Warning!
Warning!
Warning!
Warning!
Read and follow all safety precautions, warnings,
cautions, and other recommendations in this manual.
OTHERWISE, DEATH, PERSONAL INJURY OR
EQUIPMENT DAMAGE COULD OCCUR.
Read this entire section before handling the equipment.
Symbols
The following symbols may be used on the equipment and/or in this manual.
This symbol represents a Caution
Caution
Caution
Caution
Caution or a Warning
Warning
Warning
Warning
Warning. Cautions draw
special attention to anything that could damage equipment or
cause the loss of data. Warnings draw special attention to
anything that could injure or kill the reader. Both Cautions and
Warnings are placed before the step they apply to.
This symbol represents a Hot Surface
Hot Surface
Hot Surface
Hot Surface
Hot Surface.
This symbol represents a Puncture Risk.
Puncture Risk.
Puncture Risk.
Puncture Risk.
Puncture Risk. It is usually used in
regard to nozzle cleaning appliances and other sharp
instruments that can cause puncture wounds and risk exposure
to bloodborne pathogens and other debris.
This symbol means that Working Gloves
Working Gloves
Working Gloves
Working Gloves
Working Gloves are required.
This symbol means that Goggles
Goggles
Goggles
Goggles
Goggles are required.
This symbol indicates a Shock Hazard
Shock Hazard
Shock Hazard
Shock Hazard
Shock Hazard. There is a presence of
non-insulated dangerous voltage within the product’s enclosure.
This voltage may cause electrical shock or fire.
Continued next page
Section 2 - Safety Information
MC036 - Low Level Detection Alarm/AutoRefill Systems
Symbols - Continued
This symbol indicates a Note
Note
Note
Note
Note. Notes point out something of
special interest or importance to the reader. They give tips,
hints, and information in addition to what is necessary for the
step preceding it.
This symbol indicates the need to Unplug/Disconnect All Power
Unplug/Disconnect All Power
Unplug/Disconnect All Power
Unplug/Disconnect All Power
Unplug/Disconnect All Power
Sources
Sources
Sources
Sources
Sources and to let them de-energize before attempting any type
of work or maintenance. Remember that there can still be energy
in equipment, cords, and wires even when unplugged/
disconnected.
This symbol indicates the need to Lock Out All Power Sources
Lock Out All Power Sources
Lock Out All Power Sources
Lock Out All Power Sources
Lock Out All Power Sources
and to let them de-energize before attempting any type of work
or maintenance. If power is not locked out, the person working
on the equipment may be injured or killed if someone
unknowingly switches on the power to the equipment.
Section 2 - Safety Information
MC036 - Low Level Detection Alarm/AutoRefill Systems
Owner
Responsibilities
The owner of the equipment is under obligation to manage all safety
information. Some examples include:
•
Examine all safety materials and documents as well as
jurisdictional laws and make certain all laws, recommendations,
and other safety/hazard laws, certification requirements, training,
and instructions are followed and kept current.
•
Maintain all safety materials including tags, labels, documents,
and MSDS information. Make certain they are distinct and can be
read/understood. Replace any that are dirty, worn, or unreadable.
•
Make sure all personnel who will handle, install, maintain,
operate, fix, and work around the equipment have ready access
to the safety information, training, and equipment according to
jurisdictional authorities.
The owner of the equipment is under obligation to make certain that all
instructions, requirements, and jurisdictional laws are met. Some examples
include:
•
Make sure there are regular inspections of equipment and safety
devices.
•
Have regular safety drills and inspections supervised by the
proper authorities.
•
Provide all required safety items, first aid equipment, and
training.
The owner of the equipment is under obligation to make certain that all
personnel who will handle, install, maintain, operate, fix, and work around the
equipment are qualified, trained, and up-to-date with all information regarding
the equipment. Some examples include:
•
Make sure all personnel have the proper safety training,
equipment, education, and abilities necessary for the job function
according to safety instructions and all jurisdictional laws and
regulations.
•
It is strongly advised that personnel receive first-responder
medical care training in case of burns, medical emergencies, or
other injuries. Training should be kept up to date.
•
Make sure all personnel understand and can follow safety
policies and procedures for the organization as well as for the
specific equipment.
•
Make sure that all personnel are consistently trained, evaluated,
free of alcohol and medications that may impair judgment and
reflexes, and are tested for banned substances according to
jurisdictional authorities.
Section 2 - Safety Information
MC036 - Low Level Detection Alarm/AutoRefill Systems
Limitations of
Use
Read this document and all information regarding the equipment before
handling the equipment. The intended use of the equipment is stated in
Section 1 of this manual.
Do not use this equipment for anything other than its intended use. Do not
modify, change, or alter the equipment in any way. If you are unsure of the
intended use and the limitations of use for the equipment, contact your Valco
Melton Representative before handling the equipment.
Installation/
Startup/Use
Safety
Information
Warning!
Warning!
Warning!
Warning!
Warning!
Only qualified personnel should install the equipment.
Valco Melton strongly recommends that a Valco Melton
Technician install all equipment. OTHERWISE, DEATH,
PERSONAL INJURY, OR DAMAGE TO EQUIPMENT
COULD OCCUR.
Warning!
Warning!
Warning!
Warning!
Warning!
The equipment should be installed so that it can be turned
off at a location away
away
away
away
away from the equipment in case of
injury, electrical problems, or malfunction. OTHERWISE,
DEATH OR PERSONAL INJURY COULD OCCUR.
Warning!
Warning!
Warning!
Warning!
Warning!
Properly route all electrical wires. Never tamper with
equipment. Only use approved and correct voltage, type
of current, fuses, and other power supplies. Replace worn
cords, hoses, etc. immediately. FAILURE TO OBSERVE
WARNING MAY RESULT IN DEATH, PERSONAL
INJURY, AND/OR EQUIPMENT DAMAGE.
Warning!
Warning!
Warning!
Warning!
Warning!
Poor ventilation, smoking, and open flames can cause
overheated hot melt to ignite. Adequate ventilation must
be provided. Smoking should be prohibited in the
immediate vicinity of the molten adhesive. Open flames
must be kept away from the area around molten adhesive.
OTHERWISE, DEATH, PERSONAL INJURY, OR
DAMAGE TO EQUIPMENT COULD OCCUR.
Warning!
Warning!
Warning!
Warning!
Warning!
Never use any Valco Melton equipment in an explosive
environment. Explosive environments include, but are not
limited to, solvent-based cleaners or adhesives, explosive
materials, radioactive materials, etc. OTHERWISE,
DEATH OR PERSONAL INJURY COULD OCCUR.
Warning!
Warning!
Warning!
Warning!
Warning!
Equipment will start automatically when remotely
controlled by triggering devices. Be sure to disable all
triggering devices, carefully release hydraulic pressure,
and disconnect air pressure before servicing or working
near guns, valves, and other triggered devices.
OTHERWISE, DEATH OR PERSONAL INJURY COULD
OCCUR.
Valco Melton hot melt units, cold glue units, controllers, inspection systems
and all related accessories have the following universal safety precautions
(this is not intended to be an exhaustive list; follow all instructions and safety
precautions for the specific type of equipment involved):
Section 2 - Safety Information
MC036 - Low Level Detection Alarm/AutoRefill Systems
Shut Down
Safety
Information
Valco Melton hot melt units, cold glue units, controllers, inspection systems
and all related accessories have the following universal safety precautions
(this is not intended to be an exhaustive list; follow all instructions and safety
precautions for the specific type of equipment involved):
Warning!
Warning!
Warning!
Warning!
Warning!
Purge the fluid pressure and the air pressure from the
system before disconnecting/disabling any part of the
system. OTHERWISE, DEATH OR PERSONAL INJURY
COULD OCCUR.
Warning!
Warning!
Warning!
Warning!
Warning!
Disconnect and lock out all power before maintenance
or other need to open the equipment. Only qualified
personnel should open and service the control.
OTHERWISE, DEATH OR PERSONAL INJURY
COULD OCCUR.
Warning!
Warning!
Warning!
Warning!
Warning!
Equipment may still be energized even if unplugged!
When making adjustments or performing checkout
procedures, stay clear of any moving mechanical parts
and do not touch exposed electrical equipment or
electrical connectors. OTHERWISE, DEATH OR
PERSONAL INJURY COULD OCCUR.
Warning!
Warning!
Warning!
Warning!
Warning!
Disconnect/disable all mechanical and/or electrical
devices that send activation signals to the gun(s),
valve(s), melter pump(s), etc. This includes pattern
controls, timers, input/output signals, etc. Only qualified
personnel should open and service the control.
OTHERWISE, DEATH OR PERSONAL INJURY
COULD OCCUR.
Warning!
Warning!
Warning!
Warning!
Warning!
Disable all triggering devices, relieve all residual
pressure (hydraulic and air) and allow adhesive to cool
before attempting to disconnect guns, hoses, valves,
etc. Only qualified personnel should open and service
the control. OTHERWISE, DEATH OR PERSONAL
INJURY COULD OCCUR.
Warning!
Warning!
Warning!
Warning!
Warning!
Never point an adhesive dispensing gun, valve, hose, air
hose, or anything else at yourself or another person.
OTHERWISE, DEATH OR PERSONAL INJURY
COULD OCCUR.
Section 2 - Safety Information
MC036 - Low Level Detection Alarm/AutoRefill Systems
Hot Melt
Specific
General Safety
Information
Warning!
Warning!
Warning!
Warning!
Warning!
Never
Never
Never
Never
Never process any polyurethane reactive (PUR) hot
melt or solvent-based material in a Valco Melton unit
unless you are certain that the unit is compatible and is
marked “PUR”! Read all instructions and MSDS sheets
carefully, following manufacturer’s instructions,
especially regarding heat levels. If you have any
question as to the compatibility of a Valco Melton unit for
PUR hot melt, call your Valco Melton Representative
before attempting to use the unit for PUR or solvent-
based materials. OTHERWISE, HAZARDOUS FUMES,
EXPLOSION, DEATH, OR PERSONAL INJURY
COULD OCCUR.
Valco Melton hot melt units have the following universal safety precautions in
in
in
in
in
addition to all other universal precautions previously mentioned
addition to all other universal precautions previously mentioned
addition to all other universal precautions previously mentioned
addition to all other universal precautions previously mentioned
addition to all other universal precautions previously mentioned (this is not
intended to be an exhaustive list; follow all instructions and safety
precautions for the specific type of equipment involved):
Warning!
Warning!
Warning!
Warning!
Warning!
Keep pump cover and electrical enclosures closed
except during setup, service, and checkout procedures.
OTHERWISE, DEATH OR PERSONAL INJURY
COULD OCCUR.
Warning!
Warning!
Warning!
Warning!
Warning!
People with respiratory problems (e.g., asthma,
bronchitis, etc.) should not work in the vicinity of molten
adhesive. RESPIRATORY PROBLEMS MAY BE
AGGRAVATED BY THE FUMES. Do not wear a face
mask when working around molten adhesive. THE
MASK MAY TRAP THE FUMES AND DEATH OR
PERSONAL INJURY COULD OCCUR.
Warning!
Warning!
Warning!
Warning!
Warning!
Keep hot melt hoses away from walkways and the
moving parts of hot melt systems. OTHERWISE,
PERSONAL INJURY OR EQUIPMENT DAMAGE
COULD OCCUR.
Warning!
Warning!
Warning!
Warning!
Warning!
Hot surfaces! Do not touch! Use extreme caution when
refilling the unit by hand. OTHERWISE, PERSONAL
INJURY COULD OCCUR.
Warning!
Warning!
Warning!
Warning!
Warning!
Wear protective gloves and goggles at all times around
all machinery, especially hot melt. OTHERWISE,
SERIOUS PERSONAL INJURY COULD OCCUR.
Warning!
Warning!
Warning!
Warning!
Warning!
Never use an open flame to heat hot melt components
or adhesive. OTHERWISE, DEATH, PERSONAL
INJURY, OR DAMAGE TO EQUIPMENT COULD
OCCUR.
Section 2 - Safety Information
MC036 - Low Level Detection Alarm/AutoRefill Systems
If hot adhesive comes in contact with the skin, , , , , do the following:
Warning!
Warning!
Warning!
Warning!
Warning!
Do not attempt to remove heated hot melt adhesive from
the skin. OTHERWISE, SEVERE PERSONAL INJURY
AND DEATH COULD OCCUR.
1.
Immediately immerse the contacted area in clean, cold water.
It is strongly recommended that a
source of clean, cold water be provided
near the hot melt work area.
2.
Cover the affected area with a clean, wet compress and call the
emergency medical response system (such as 911) immediately.
3.
Watch for and treat the subject for signs of shock while waiting
for professional help to arrive.
What to Do if Contact
with Hot Adhesive
Occurs
If adhesive fumes are inhaled, immediately follow these steps:
1.
Take the victim away from the immediate work area.
2.
Provide victim with fresh air.
3.
Call the emergency medical response system (such as 911)
immediately.
What to Do if
Inhalation of
Adhesive Fumes
Occurs
Section 2 - Safety Information
MC036 - Low Level Detection Alarm/AutoRefill Systems
If the hot melt adhesive ignites, promptly perform the following steps:
1.
Sound a fire alarm.
2.
Evacuate the immediate area.
3.
Turn off all local electrical equipment at the source.
4.
Leave the area immediately if conditions are unsafe.
IfIfIfIfIf you feel you can fight the fire safely
safely
safely
safely
safely, do one
one
one
one
one of the following:
- Smother the fire with a fire blanket.
- Aim a CO2 fire extinguisher at the base of the flames.
- Aim a dry-powder fire extinguisher at the base of the
flames.
What to Do if
Adhesive-Related
Fire or Explosion
Occurs
During the heating and melting process, the surface of the adhesive will be
exposed to air. The mixture of polymer fumes and air can catch fire if the hot
melt is overheated.
Warning!
Warning!
Warning!
Warning!
Warning!
Poor ventilation, smoking, and open flames can cause
overheated hot melt to ignite. Adequate ventilation must
be provided. Smoking should be prohibited in the
immediate vicinity of the molten adhesive. Open flames
must be kept away from the area around molten
adhesive. OTHERWISE, DEATH, PERSONAL INJURY,
OR DAMAGE TO EQUIPMENT COULD OCCUR.
Warning!
Warning!
Warning!
Warning!
Warning!
Exposed arcing may ignite the fume/air mixture. Shield
all electrical equipment from melt fumes to avoid
exposed arcing. OTHERWISE, PERSONAL INJURY OR
EQUIPMENT DAMAGE COULD OCCUR.
Warning!
Warning!
Warning!
Warning!
Warning!
Do not use a water extinguisher to extinguish the fire!
OTHERWISE, PERSONAL INJURY OR EQUIPMENT
DAMAGE COULD OCCUR.
Section 2 - Safety Information
MC036 - Low Level Detection Alarm/AutoRefill Systems
Do not
Do not
Do not
Do not
Do not use bindings, wire ties, or unapproved
fasteners around the hoses.
Do
Do
Do
Do
Do use approved wrapping (P/N KAP0434),
making sure the wrapping is slightly snug but
not tight.
Do not
Do not
Do not
Do not
Do not place hoses close together.
Do
Do
Do
Do
Do allow at least 2 inches (5.1 cm)
between hoses for proper ventilation.
Do not
Do not
Do not
Do not
Do not bend hoses sharply. Do not
Do not
Do not
Do not
Do not allow
kinks or indentations in the hoses.
Do
Do
Do
Do
Do use a minimum bend radius of 10
inches for a 20-inch diameter coil hose.
Do not
Do not
Do not
Do not
Do not use unapproved hooks to hang hoses.
Do not
Do not
Do not
Do not
Do not wrap hoses over or around objects.
Do
Do
Do
Do
Do use a hose hanging kit (P/N 781xx827).
Do not
Do not
Do not
Do not
Do not use the “one handed/one wrench”
technique to attach or remove hoses. Do
Do
Do
Do
Do
not
not
not
not
not wrench on any surface other than the
large hexagon swivel nuts.
Do
Do
Do
Do
Do use two hands and two wrenches to
tighten or loosen connections on hoses.
Do
Do
Do
Do
Do wrench only on large hexagon swivel
nuts.
Hose Safety
Information
Section 2 - Safety Information
MC036 - Low Level Detection Alarm/AutoRefill Systems
Do not
Do not
Do not
Do not
Do not allow hoses to rub against
objects or to come into contact with
sharp edges or points.
Do
Do
Do
Do
Do wrap the hoses in approved padding
(P/N 795xx549) if the hoses must be
installed where they will come into
contact with objects.
Do not
Do not
Do not
Do not
Do not use worn, damaged, or bent
hoses.
Do
Do
Do
Do
Do inspect all hoses regularly for
damage and/or wear and replace
damaged or worn hoses immediately.
Hose Safety Information -
Continued
SECTION 3 - INSTALLATION
3 - INSTALLATION
3 - INSTALLATION
3 - INSTALLATION
3 - INSTALLATION
This section contains two subsections:
1) The LLD Alarm Installation 2) The
AutoRefill LLD System. Please be sure
to use the correct section for your
application.
LLD Alarm
Installation
Before installing the LLD alarm system, carefully inspect the system for
damage incurred during shipment. If damage is discovered, contact your
shipper.
Mounting the Control
Box
The control box should be mounted:
•
In a location where the flashing light can be readily observed
by the operator
•
Close to both an electrical power outlet and to the glue tank
A 15-foot (5 M) probe cable and 7.5-
foot (2.5 M) power cable are supplied
with the LLD alarm system.
To install the control box, follow these steps:
1.
Using the upper and lower mounting flanges of the unit, mount
the control box to a stationary (non-vibrating) support.
The main power supply must be 115/230 VAC, 50/60 Hz, 2/1 A. The 115/230
power requirement is field-switchable. The power module accepts a standard
IEC line cord (supplied).
To connect the line cord, follow these steps:
1.
Attach the line cord clamp to the bottom surface of the control
box, using the two screws on either side of the power module.
2.
Insert the line cord into the power module
3.
Tighten the line cord clamp to prevent inadvertent loss of power.
The line cord supplied can either be
directly wired to a branch circuit or (with
the proper plug type) plugged into a
115 or 230 VAC outlet socket. The wire
colors of the supplied line cord are as
follows:
Electrical Installation
of the Control Box
Wire Type
International Line Cord Color
North American Line Cord Color
Line
Brown
Black
Neutral
Blue
White
Ground
Green/Yellow
Green
Section 3 - Installation
MC036 - Low Level Detection Alarm/AutoRefill Systems
Figure 3-1. Fuse Drawer
Shorting clip
230V (Europe) or 115VAC
(USA) line-to-neutral
230V (USA)
line-to-line
115V/230V
Main-Power-Supply
Voltage and Fuse
Selection
The fuse holder on the bottom panel of the power module contains the fuse
drawer, which allows for either 115VAC or 230VAC operation (Figure 3-1).
To replace or check mains fuses, follow these steps:
1.
Remove all electric power from the unit.
2.
Remove the line cord.
3.
Use a small screwdriver to open the hinged door of
the fuse holder.
4.
Remove the fuse drawer (ensure that the fuses in
the holder remain in place).
5.
The drawer is labeled “230V” and “115V.” Rotate the
drawer to the appropriate voltage.
6.
For line-to-neutral operation, use the shorting clip
(jumper) and one fuse. Shorting clip must be on left
side (when rotated to read the correct voltage). For
line-to-line operation, use two fuses.
7.
Replace the drawer in the fuse holder and close the
door of the fuse holder.
8.
The operating voltage shows through the window in
the fuse holder door. Ensure that voltage is correct.
9.
Replace the line cord.
Section 3 - Installation
MC036 - Low Level Detection Alarm/AutoRefill Systems
The low level detection probe can be installed only on the following two types
of tanks:
•
Pressurized
Pressurized
Pressurized
Pressurized
Pressurized—5-, 7-, and 12-gallon epoxy-coated tanks; 3-, 5-
, and 10-gallon stainless-steel tanks
•
Non-pressurized
Non-pressurized
Non-pressurized
Non-pressurized
Non-pressurized—55-gallon drums
Pressurized tanks use a single-probe
assembly (Figure 3-2). Non-pressurized
tanks use a three-probe assembly
(Figure 3-4).
The LLD alarm system can be mounted in tank systems that incorporate two
glue-feed supply pipes only
only
only
only
only if one of the supply pipes can be removed. A
“tee” is then placed in the remaining supply line to feed the two glue systems
(this supply technique will prove adequate in most cases). However, systems
that require two feed pipes will not accommodate the LLD alarm system.
To install the low level detection probe on epoxy-coated and stainless-steel
tanks, follow these steps:
1.
Completely release tank pressure by either disconnecting the air
hose from the input of the air regulator or adjusting the air
regulator knob counterclockwise so gauge reads 0 psi.
2.
Ensure that the level probe (Figures 3-2 and 3-3) is the correct
length for the tank assembly.
The new-style level probes are longer
than the old-style level probes. To use
a new-style level probe in an old-style
tank, cut the rod to the length indicated
in Figure 3-2. Otherwise, see the
dimensions in Figure 3-3.
3.
Remove the pipe plug from the top of the tank assembly.
4.
Add pipe sealant to pipe threads of probe.
5.
Insert probe into tank.
6.
Attach probe interconnect cable.
7.
Wire probe interconnect cable to LLD control box (Figures 3-7
and 3-8).
Mechanical
Installation of the
Probe
Pressurized Tanks
Section 3 - Installation
MC036 - Low Level Detection Alarm/AutoRefill Systems
Figure 3-2. Low-Level Detection Probe for Old-Style Pressurized Tanks
Figure 3-3. Low-Level Detection Probe for New-Style Pressurized Tanks
1/2 x 1/4 NPT reducer bushing (shown)
3/8 x 1/4 NPT reducer bushing (3-gallon only)
12-gallon
20.25 inches (514,4 mm)
5- and 7-gallon
13.15 inches (336,6 mm)
3-gallon
9.00 inches (228,6 mm)
12-gallon (21.0 inches)
5-gallon (17.7 inches)
3-gallon (10.2 inches)
Pressurized Tanks - Continued
Figure 3-4. Manual Pressure Release Safety Valve and Pop Safety Valve
Section 3 - Installation
MC036 - Low Level Detection Alarm/AutoRefill Systems
The probe assembly for non-pressurized tanks contains three probes (Figure
3-5):
•
Warning level probe
Warning level probe
Warning level probe
Warning level probe
Warning level probe—The warning level probe is the shortest
probe. The warning level probe should be installed 1 inch
above the low level and common probes.
•
Low level and common (ground) probes
Low level and common (ground) probes
Low level and common (ground) probes
Low level and common (ground) probes
Low level and common (ground) probes—The low level and
common probes, which are the same length, are the longest
probes. The low level and common probes should be
installed approximately
1 inch above the suction tube on the pump.
Probes can be shortened to suit any
drum.
To install the low level detection probe assembly on a 55-gallon drum, follow
these steps:
1.
Determine the glue level at which the alarm will be activated.
2.
Install the probe assembly into the inspection port of the drum
(Figure 3-6).
For systems incorporating a glue-
supply pump, ensure that the activation
depth of the low level alarm is not
below the pump inlet level. If it is, the
pump will run out of glue before the
alarm is given.
Non-Pressurized Tanks
Figure 3-5. Low Level Detection Probe for Non-Pressurized Tanks
Warning level probe
Low level probe
Common probe
Warning level probe
Low level probe
Common probe
Section 3 - Installation
MC036 - Low Level Detection Alarm/AutoRefill Systems
Figure 3-6. Low Level Detection Probe Assembly in Tank
Probe assembly
Non-Pressurized Tanks -
Continued
Electrical Installation
of the Probe
To connect the low level detection probe to the electronics control, follow
these steps:
1.
Insert the probe cable through a cord grip located on the bottom
panel of the control.
2.
Connect the probe cable to the connector labeled “PROBE IN”.
3.
Tighten the cord grip used by the probe cable.
Electrical relay connections are as follows:
•
Switched 115/230VAC output
Switched 115/230VAC output
Switched 115/230VAC output
Switched 115/230VAC output
Switched 115/230VAC output—This output can be used as
the voltage source (115 or 230VAC, depending on the alarm
system’s source voltage) for the glue supply pump. When the
alarm system is activated, a device connected to the
switched output will be turned off.
•
Dry contacts
Dry contacts
Dry contacts
Dry contacts
Dry contacts—A set of either normally opened or normally
closed contacts can be used to control a device when the
alarm system is activated.
Section 3 - Installation
MC036 - Low Level Detection Alarm/AutoRefill Systems
After the control box and the probe have been installed, an initial system test
should be conducted. To test the system, follow these steps:
1.
Pressurize system.
2.
Inspect for leakage around fittings.
3.
Disconnect probe from probe cable.
4.
Plug main line cord into main power supply.
5.
Turn the control switch to the “ON” position.
The flashing light should flash on and
off. The audio alarm should also sound
if the “PROBE IN” connector was wired
for a dual sensor.
6.
Connect the probe to the probe cable.
Warning devices should turn off.
Testing the LLD
Alarm System
Internal Connections
Item
Description
Part Number
41
Wire terminal
075xx244
43
Cable assy.
030xx425
57
Beacon, red, low profile
105xx036
Figure 3-7. Wiring Diagram for Internal Connections (LLD Alarm System)
Please see the Parts List Section for more information.
Section 3 - Installation
MC036 - Low Level Detection Alarm/AutoRefill Systems
External Customer
Connections
Figure 3-8. Wiring Diagram for External Customer Connections
(LLD Alarm System)
080xx545-01
The AutoRefill LLD
Installation
Before installing the AutoRefill LLD system, carefully inspect the system for
damage incurred during shipment. If damage is discovered, contact your
shipper.
Mounting the Control Box
The control box should be mounted:
•
Close to both an electrical power outlet and to the glue tank
A 15-foot (5 M) probe cable and 7.5-
foot (2.5 M) power cable are supplied
with the AutoRefill LLD system.
To install the control box, follow these steps:
1.
Using the upper and lower mounting flanges of the unit, mount
the control box to a stationary (non-vibrating) support.
Section 3 - Installation
MC036 - Low Level Detection Alarm/AutoRefill Systems
The main power supply must be 115/230 VAC, 50/60 Hz, 2/1 A. The 115/230
power requirement is field-switchable. The power module accepts a standard
IEC line cord (supplied).
To connect the line cord, follow these steps:
1.
Attach the line cord clamp to the bottom surface of the control
box, using the two screws on either side of the power module.
2.
Insert the line cord into the power module
3.
Tighten the line cord clamp to prevent inadvertent loss of power.
The line cord supplied can either be
directly wired to a branch circuit or (with
the proper plug type) plugged into a
115 or 230 VAC outlet socket. The wire
colors of the supplied line cord are as
follows:
Wire Type
International Line Cord Color
North American Line Cord Color
Line
Brown
Black
Neutral
Blue
White
Ground
Green/Yellow
Green
Electrical Installation of
the Control Box
Section 3 - Installation
MC036 - Low Level Detection Alarm/AutoRefill Systems
Figure 3-9. Fuse Drawer
Shorting clip
230V (Europe) or 115VAC
(USA) line-to-neutral
230V (USA)
line-to-line
115V/230V
The fuse holder on the bottom panel of the power module contains the fuse
drawer, which allows for either 115VAC or 230VAC operation (Figure 3-9).
To replace or check mains fuses, follow these steps:
1.
Remove all electric power from the unit.
2.
Remove the line cord.
3.
Use a small screwdriver to open the hinged door of
the fuse holder.
4.
Remove the fuse drawer (ensure that the fuses in
the holder remain in place).
5.
The drawer is labeled “230V” and “115V.” Rotate the
drawer to the appropriate voltage.
6.
For line-to-neutral operation, use the shorting clip
(jumper) and one fuse. Shorting clip must be on left
side (when rotated to read the correct voltage). For
line-to-line operation, use two fuses.
7.
Replace the drawer in the fuse holder and close the
door of the fuse holder.
8.
The operating voltage shows through the window in
the fuse holder door. Ensure that voltage is correct.
9.
Replace the line cord.
Main-Power-Supply
Voltage and Fuse
Selection
Section 3 - Installation
MC036 - Low Level Detection Alarm/AutoRefill Systems
Figure 3-10 shows an example of how the AutoRefill LLD System can be set
up. For additional information on all the parts and options available, see the
Parts List Section.
Figure 3-11 shows the wiring diagrams for the AutoRefill LLD System.
CAUTION!
CAUTION!
CAUTION!
CAUTION!
CAUTION!
The low level detection probe with automatic refill
feature can only be used on non-pressurized 55-gallon
drums!
Mechanical
Installation of the
Probe (non-
pressurized tanks)
Figure 3-10 . AutoRefill LLD System Setup Example
Section 3 - Installation
MC036 - Low Level Detection Alarm/AutoRefill Systems
Internal and External Customer Connections
Figure 3-11 . Wiring Diagram for External and Internal Connections (AutoRefill LLD System)
098xx699
Please see the Part Number List Section for more details.
Item
Description
Part Number
16
Terminal; crimp
075xx239
25
Beacon, red, low profile
105xx036
38
Wire, PVC, brown
540xx035
41
Wire terminal
075xx244
43
Cable assy
030xx425
52
LED assy, power on/off
107xx022
SECTION 4 - GLUE SOURCE AND PROBE INSTALLATION
SECTION 4 - GLUE SOURCE AND PROBE INSTALLATION
SECTION 4 - GLUE SOURCE AND PROBE INSTALLATION
SECTION 4 - GLUE SOURCE AND PROBE INSTALLATION
SECTION 4 - GLUE SOURCE AND PROBE INSTALLATION
This section contains two subsections:
1) The LLD Alarm Assembly Drawings
2) The AutoRefill LLD System
Assembly Drawings. Please be sure to
use the correct section for your
application.
This section contains the following tank assembly drawings for the traditional
LLD Alarm:
•
Stainless-steel 3-gallon tank assembly
•
Top view of stainless-steel 3-, 5-, and 10-gallon tank
assembly
•
Top view of OVS stainless-steel 3-gallon tank assembly
•
Epoxy-coated 5-gallon tank assembly
•
Top view of Epoxy-coated 5-gallon tank assembly
•
Top view of OVS stainless-steel 5-gallon tank assembly
•
Top view of OVS stainless-steel 10-gallon tank assembly
•
Epoxy-coated 7- and 12-gallon tank assembly
Figure 4-1. Stainless-Steel 3-Gallon Tank Assembly
Air hose
Air regulator assembly
Gauge
Tank body
Tank lid
Glue filter assembly
Female quick-dicsonnect
Male quick-dicsonnect
Pop relief valve
Pipe plug
(probe installation location)
LLD Alarm
Assembly
Drawings
Section 4 - Glue Source and Probe Installation
MC036 - Low Level Detection Alarm/AutoRefill Systems
Figure 4-2. Top View of Stainless-Steel 3-, 5-, and 10-Tank Assembly
Figure 4-3. Top View of OVS Stainless-Steel 3-Gallon Tank Assembly
LLD Alarm Assembly Drawings -
Continued
Section 4 - Glue Source and Probe Installation
MC036 - Low Level Detection Alarm/AutoRefill Systems
Figure 4-4. Epoxy-Coated 5-Gallon Tank Assembly
Glue filter assembly,
screen, and O-ring
Male quick-disconnect
Feed pipe
Tank lid
Tank lid gasket
Gauge
Female quick-disconnect
Air regulator assembly
Pop relief valve
Air hose
Hold-down clamp assembly
Tank body
Figure 4-5. Top View of Epoxy-Coated 5-Gallon Tank Assembly
LLD Alarm Assembly Drawings -
Continued
Section 4 - Glue Source and Probe Installation
MC036 - Low Level Detection Alarm/AutoRefill Systems
LLD Alarm Assembly Drawings -
Continued
Figure 4-7. Top View of OVS Stainless-Steel 10-Gallon Tank Assembly
Figure 4-6. Top View of Stainless-Steel OVS 5-Gallon Tank Assembly
Section 4 - Glue Source and Probe Installation
MC036 - Low Level Detection Alarm/AutoRefill Systems
LLD Alarm Assembly Drawings -
Continued
Figure 4-8. 7- and 12-Gallon Epoxy-Coated Tank Assembly
Female quick-disconnect
Glue filter assembly
Male quick-disconnect
Feed pipe
Gauge
Air regulator assembly
Tank lid assembly
Tank lid safety valve
Pop relief valve
Air hose
Tank
Section 4 - Glue Source and Probe Installation
MC036 - Low Level Detection Alarm/AutoRefill Systems
LLD AutoRefill
System
Assembly
Drawings
The LLD AutoRefill System can only be used in a non-pressurized drum.
The following is one example of a layout for the AutoRefill System. Valco
Cincinnati, Inc. can work with you to develop a layout that fits your unique
needs.
Figure 4-9. Non-Pressurized 55-Gallon Drum with the LLD AutoRefill System
SECTION 5 - MAINTENANCE
SECTION 5 - MAINTENANCE
SECTION 5 - MAINTENANCE
SECTION 5 - MAINTENANCE
SECTION 5 - MAINTENANCE
Introduction
This section contains guidelines and procedures for maintaining the LLD
Alarm/AutoRefill System. Your system will run longer and more consistently
if a preventative maintenance program is used.
General
Maintenance
Valco recommends the following general maintenance:
•
Keep the control box door closed except when working on
the control circuitry. This will prevent dust and dirt from
entering the unit.
•
Occasionally wipe the surface of the control box, using a
clean cloth dampened with a mild soap and water solution.
•
To prevent glue buildup, remove the probe from the glue
supply tank whenever the glue supply is replenished, and
wash off the probe.
SECTION 6 - TROUBLESHOOTING
6 - TROUBLESHOOTING
6 - TROUBLESHOOTING
6 - TROUBLESHOOTING
6 - TROUBLESHOOTING
This section contains instructions for troubleshooting the LLD alarm system.
Prob lem
Possible Cau se
Poss ible So lutio n
1.
Ann unciators are
eith er no t operating o r
are failin g to s ignal a
low level. The po wer
light is n ot on ,
indic atin g tha t the re is
no powe r to the u nit.
1a. Unit is n ot plugge d into power
sou rce.
1b. Unit is n ot tu rned on.
1c. Unit's fuse is b lown .
1d. Plant su pply is faulty.
1e. Th ere is a system -wiring
problem .
1 a. P lug unit into p ower sou rce.
1 b. T urn unit o n.
1 c. R eplace fu se.
1 d. C heck m ains s upply at the
contro l con nection.
1 e. C heck the system wiring .
2.
Ann unciators are
eith er no t operating o r
are failin g to s ignal a
low level. Ho weve r,
the p ower lig ht is on,
indic atin g tha t the re is
pow er to the unit.
2a. Th ere is a problem w ith th e
probe.
2b. Th ere is a problem w ith th e PC
board.
2 a. D isco nnect the prob e.
If the syste m o pera tes, it
m eans the prob e is not
a llowing th e circuit to to be
b roken. T herefore , the p roblem
is in th e probe. Rep lace the
p robe.
If the syste m d oes not o perate,
th e pro be m a y be wired
in correctly at th e co ntro l.
C orre ct the wiring.
2 b. R eplace P C bo ard.
3.
Ann unciators ope rate
con tinuo usly.
3a. Glu e le vel is low.
3b. Th e pro be a nd/o r cab le are
eith er no t connec ted or
sev ered.
3c. Th ere is encruste d glue o n
probe co ntac ts.
3d. Th ere is a system -wiring
problem .
3 a. R eplenish glue leve l.
3 b. C onnect or rep lace prob e
a nd/o r cab le.
3 c. C lean prob e.
3 d. C heck the system wiring .
4.
Only on e of the
annu nciators
operates .
4a. If h orn works an d ligh t doe s no t,
bulb m ay be burn ed o ut.
4b. Th ere is a system -wiring
problem .
4 a. R eplace b ulb.
4 b. C heck the system wiring .
Introduction
SECTION 7 - SPECIFICATIONS
SECTION 7 - SPECIFICATIONS
SECTION 7 - SPECIFICATIONS
SECTION 7 - SPECIFICATIONS
SECTION 7 - SPECIFICATIONS
These charts are basic guidelines for
the general systems available. Due to
configuration changes and options,
your specific system may vary slightly.
The Valco Installation Specialist can
answer any questions you may have
regarding your specific system setup.
LLD Alarm System:
Width:
7"
Height:
14.5"
Depth:
5.5"
LLD probe inputs:
Warning level: activates flashing light and relays
Low-limit level: activates audio alarm
LLD Alarm system outputs:
Switched 115/230VAC, 1/0.5A, 50/60Hz—switched off when
warning level (or lower) is reached
Dry contacts, form C (1 N.O. I.N.C.), 230VAC, 1A maximum
LLD AutoRefill System:
Width:
7"
Height:
14.5"
Depth:
5.5"
LLD probe inputs:
Low-limit level: activates automatic refilling of glue
LLD AutoRefill system outputs:
Switched 115/230VAC, 1/0.5A, 50/60Hz—switched off when
warning level (or lower) is reached
Dry contacts, form C (1 N.O. I.N.C.), 230VAC, 1A maximum
Valco Cincinnati, Inc.
SECTION 8 - PART NUMBER LIST
SECTION 8 - PART NUMBER LIST
SECTION 8 - PART NUMBER LIST
SECTION 8 - PART NUMBER LIST
SECTION 8 - PART NUMBER LIST
How to Order
Parts
To order parts, please contact your closest Valco office by mail, phone, or
Email:
Valco Cincinnati, Inc.
Valco Cincinnati, Inc.
Valco Cincinnati, Inc.
Valco Cincinnati, Inc.
Valco Cincinnati, Inc.
411 Circle Freeway Drive
Cincinnati, OH 45246
TEL:
(513) 874-6550
FAX:
(513) 874-3612
Email:
[email protected]
Web:
http://www.valcomelton.com
Valco Cincinnati Limited
Valco Cincinnati Limited
Valco Cincinnati Limited
Valco Cincinnati Limited
Valco Cincinnati Limited
Hortonwood 32
Telford, TFI 7YN, England
TEL:
(+44) 1952-677911
FAX:
(+44) 1952-677945
Email:
[email protected]
Web:
http://www.valco.co.uk
Valco Cincinnati GmbH
Valco Cincinnati GmbH
Valco Cincinnati GmbH
Valco Cincinnati GmbH
Valco Cincinnati GmbH
Storkower Strasse 6
D-15749 Gallun, Germany
TEL:
(+49) 337 648 700
FAX:
(+49) 337 648 7070
Email:
[email protected]
Germany:
England:
USA:
Spain:
Melton S.L.U.
Melton S.L.U.
Melton S.L.U.
Melton S.L.U.
Melton S.L.U.
Pol. Industrial Agustinos
calle G, n. 34
31160 Orcoyen, Navarra, Spain
TEL: (34) 948-321-580
FAX: (34) 948-326-584
Valco Cincinnati, Inc.
Section 8 - Part Number List
MC036 - Low Level Detection Alarm/AutoRefill Systems
Example: Auto
Pot Fill
System
Layout
(080xx377)
999XD080-96
Valco Cincinnati, Inc.
Section 8 - Part Number List
MC036 - Low Level Detection Alarm/AutoRefill Systems
Item
Description
Part Number
Quantity
1
Control assembly
098xx699
1
2
Sensor
485xx090
1
3
Cable assembly
030xx604
1
6
Mounting bracket for scanner
581xx126
1
7
Clamp assembly
578xx526
1
8
L-bar
578xx539
1
9
Mounting base
578xx527
1
10
Tool
578xx505
1
12
Air filter regulator
594xx082
1
13
Pot fill kit
754xx330
1
14
Valve cable
030xx882
1
16
Gauge
786xx001
1
17
Bushing, M-F
797xx045
1
Example: Auto Pot Fill System
Layout (080xx377) - Continued
Valco Cincinnati, Inc.
Section 8 - Part Number List
MC036 - Low Level Detection Alarm/AutoRefill Systems
LLD Control
Assembly (080xx545)
The LLD
Alarm
Assembly
Parts List
Valco Cincinnati, Inc.
Section 8 - Part Number List
MC036 - Low Level Detection Alarm/AutoRefill Systems
LLD Control Assembly (080xx545)
- Continued
Valco Cincinnati, Inc.
Section 8 - Part Number List
MC036 - Low Level Detection Alarm/AutoRefill Systems
Item
Description
Part Number
Quantity
1
Cabinet; LLD-CE
025xx656
1
2
Alarm, low level
010xx001
1
5
Cord grip
066xx003
1
6
Locknut
066xx005
1
7
Dust seal
066xx006
1
9
Harness anchor
067xx002
3
10
Ty-rap
067xx007
5
11
Fuse, 1 AMP SLO-BLO
085xx001
1
13
Latch
091xx011
1
14
Rivet
091xx030
2
16
Terminal; crimp
075xx239
2
22
PCB assy, LLD w/relay
151xx411
1
26
Adapter
795xx480
1
27
Screw
798xx295
2
29
Screw
798xx466
4
31
Hex nut
798xx687
2
33
Lock washer
798xx733
6
37
Hole plug
794xx038
1
38
Wire, PVC, brown
540xx035
22
39
Wire, PVC, white
540xx043
22
40
Wire, PVC, green/yellow
540xx039
6
41
Wire terminal
075xx244
1
42
Power entry module
086xx055
1
43
Cable assy
030xx425
1
49
Screw
798xx442
2
50
Lock washer
798xx772
2
52
LED assy, power on/off
107xx022
1
54
O-ring
745xx382
1
57
Beacon, red, low profile
105xx036
1
110
Label, Valco CE
781xx347
1
LLD Control Assembly (080xx545)
- Continued
Valco Cincinnati, Inc.
Section 8 - Part Number List
MC036 - Low Level Detection Alarm/AutoRefill Systems
Installation Kit
(781xx258)
Description
Part Number
Quantity
Fuse (package of 5)
085xx040
1
Cable, IEC 320/European leads
030xx422
1
European line-cord clamp
066xx116
1
Cable, IEC 320/U.S. plug
030xx493
1
U.S line-cord clamp
066xx120
1
Cord grip
066xx003
1
Dust seal
066xx006
1
Probe
Assemblies
Description
Part Number
Quantity
Probe assembly for 3-, 5-, 7-, and 12-gallon epoxy-coated tank
080xx525
1
Probe assembly for 3-gallon stainless-steel tank
080xx523
1
Probe assembly for 5-gallon stainless-steel tank
080xx524
1
Probe assembly for 10-gallon stainless-steel tank
080xx525
1
Probe assembly for 55-gallon drum
080xx018
1
Probe assembly for vats and tote bins
080xx019
1
Probe cable
030xx021
1
Pop safety valve
703xx003
1
Manual pressure-release safety valve
703xx004
1
Valco Cincinnati, Inc.
Section 8 - Part Number List
MC036 - Low Level Detection Alarm/AutoRefill Systems
The LLD
AutoRefill
Assembly
Parts List
LLD AutoRefill
Control Assembly
(098xx699)
Valco Cincinnati, Inc.
Section 8 - Part Number List
MC036 - Low Level Detection Alarm/AutoRefill Systems
LLD AutoRefill Control Assembly
(098xx699) - Continued
Valco Cincinnati, Inc.
Section 8 - Part Number List
MC036 - Low Level Detection Alarm/AutoRefill Systems
LLD AutoRefill Control Assembly
(098xx699) - Continued
Item
Description
Part Number
Quantity
1
Cabinet; LLD-CE
025xx656
1
5
Cord grip
066xx003
1
6
Locknut
066xx005
1
7
Dust seal
066xx006
1
9
Harness anchor
067xx002
3
10
Ty-rap
067xx007
5
11
Fuse, 1 AMP SLO-BLO
085xx001
1
13
Latch
091xx011
1
14
Rivet
091xx030
2
16
Terminal; crimp
075xx239
2
22
PCB assy, LLD w/relay
151xx597
1
25
Beacon, red, low profile
105xx036
1
29
Screw
798xx466
4
31
Hex nut
798xx687
2
33
Lock washer
798xx733
6
36
Wire, PVC, blue
540xx040
3
37
Hole plug
794xx038
1
38
Wire, PVC, brown
540xx035
22
41
Wire terminal
075xx244
1
42
Power entry module
086xx055
1
43
Cable assy
030xx425
1
49
Screw
798xx442
2
50
Lock washer
798xx772
2
52
LED assy, power on/off
107xx022
1
110
Label, Valco CE
781xx347
1
Valco Cincinnati, Inc.
Section 8 - Part Number List
MC036 - Low Level Detection Alarm/AutoRefill Systems
Pot Fill Kit
(754xx330)
Valco Cincinnati, Inc.
Section 8 - Part Number List
MC036 - Low Level Detection Alarm/AutoRefill Systems
Pot Fill Kit (754xx330) - Continued
Item
Description
Part Number
Quantity
1
Shut off valve
703xx499
2
2
Feed pipe
558xx312
2
3
Reducer coupling
797xx059
2
4
Swivel fitting - female
750xx001
3
5
Male swivel
750xx009
3
6
Swivel fitting - plug
750xx005
2
7
Flat washer
798xx768
2
8
Air tubing and fitting
753xx701
1
10
Swivel fitting - female
750xx015
1
12
Mounting block
580xx606
1
13
Thumb screw
798xx864
1
18
Male branch tee
797xx027
1
20
Adapter m-f
797xx961
1
21
Swivel fitting, NPT
799xx104
2
22
Hose
755xx234
10
23
Elbow
797xx905
1
24
Valve assy., 3-way
411xx757
1
25
Bushing
797xx045
1
26
Glue valve
705xx114
1
SECTION 9 - WARRANTY
SECTION 9 - WARRANTY
SECTION 9 - WARRANTY
SECTION 9 - WARRANTY
SECTION 9 - WARRANTY
Cold Glue Equipment
and Electronic
Controls
Warranty
Information
Hot Melt Units,
Hoses, Valves, Guns,
and Related
Equipment
Valco Cincinnati, Inc. warrants its equipment worldwide against defects in
material and workmanship as outlined in this section.
Liability of the company is limited to repair of the product, or replacement of
any part shown to be defective, and does not extend to defects caused by
accidents, misuse, abuse, neglect, tampering or deterioration by corrosion.
This warranty does not cover those items determined by Valco Cincinnati,
Inc. to be normal wear items such as seals, O-rings, diaphragms, springs,
etc.
Reconditioned equipment, unless specified otherwise at the time of purchase,
will be warranted as described above for a period of ninety (90) days from the
date of shipment by Valco Cincinnati.
Components purchased by Valco Cincinnati, Inc. from others for inclusion in
its products are warranted only to the extent of the original manufacturer’s
warranty. In no event shall Valco Cincinnati, Inc. be liable for indirect or
consequential damages arising out of the use of Valco Cincinnati products.
This warranty is conditioned upon the prepaid return of the equipment
claimed to be defective to Valco Cincinnati, Inc. for examination and
verification. If claimed defect is verified, repairs or replacements will be made
F.O.B. Cincinnati, Ohio, U.S.A. or ex-works Telford, U.K. If the inspection of
the equipment does not
not
not
not
not disclose any defect of workmanship or material, any
necessary repairs will be made at a reasonable charge and return
transportation will be charged.
This is the only authorized Valco Cincinnati, Inc. warranty and is in lieu of all
other expressed or implied warranties, representations or any other
obligations on the part of Valco Cincinnati, Inc.
The warranty for cold glue equipment and electronic controls for a period of
one (1) year from the date of shipment by Valco Cincinnati, Inc.
All hot melt components except cast-in heating elements are warranted for a
period of six (6) months from the date of shipment by Valco Cincinnati. Cast-
in heaters carry an additional, pro-rated warranty not to exceed three (3)
years from the date of shipment by ValcoMelton, a Valco Cincinnati, Inc.
company.
SECTION 10 - SERVICE
SECTION 10 - SERVICE
SECTION 10 - SERVICE
SECTION 10 - SERVICE
SECTION 10 - SERVICE
If a problem with your system persists, contact a ValcoMelton Technical
Support representative. If your need is urgent, we encourage you to contact
our corporate office in Cincinnati, Ohio, U.S.A. at (513) 874-6550. If the
problem cannot be resolved, Valco Cincinnati, Inc. will promptly arrange to
have a technical representative visit your facility. Any charges for a service
call will be quoted at that time. Any part that fails during the warranty period
shall be returned prepaid to Valco Cincinnati, Inc. by the customer for
disposition.
Upon request, ValcoMelton personnel
are available to repair or replace such
parts at the customer’s facility. Charges
for this service include travel time and
expenses.
If an equipment problem is the result of customer abuse, improper installation
or operation, all travel time, labor, parts, and expenses will be charged to the
customer.
If the responsibility for a problem cannot be absolutely determined, the
customer will be charged for travel time and expenses only. No charge will be
made for parts and labor.