Triplett 360PL Instruction Manual
Text of the manual
The Triplett Electrical Instrument Company
Bluffton, Ohio
With your purchase of a Model 630-PL Volt-Ohm-Milliammeler, you have
made a worthwhile investment, not only in a fine instrument, but backed
up by a company which has been making instruments for more than hall
a century. The Triplett Company stands behind your 630-PL and will
give all possible assistance in its use and maintenance.
TRIPLETT WARRANTY AND CONDITIONS OF SALE
The Triplett Electrical Instrument Company warrants instruments manufactured by
ii lo be free from defective material or factory workmanship and agrees to repair
or replace such instruments which under normal use and service, disclose the
defect lo be the fault of our manufacturing. Our obligation under this warranty is
limited to repairing or replacing any instrument or test equipment which proves to
be defective, when returned to us transportation prepaid, within ninety (90) days
lrom the date of otiginal purchase.
This warranty does not apply to any ol our products which have been repaired or
altered by unauthorized persons or service stations in any way so as, in our
judgment, to injure their stability or reliability or which have been subject to misuse,
negligence or accident or which have had the serial number altered, effaced, or
removed. Neither does this warranty apply to any of our products which have been
connected, installed, or adjusted otherwise than in accordance with the instructions
lumished by us. Accessories including all vacuum tubes and batteries not of our
manulacture used with this product are not covered by this warranty.
· The Triplett Electrical Instrument Company reserves the right to discontinue models
al any time, or change specifications or design, w.ithout notice and without incurring
any obligation.
Upon acceptance of the material covered by this invoice the purchaser agrees to
assume all liability for any damages and bodily injury which may result from the
use or misuse of the material by the purchaser, his employees, or others, and that
The Triplett Electrical Instrument Company shall incur no liability for direct or
consequential damage of any kind.
Parts will be made available for a maximum period of five (5) years after the
manufacture of this equipment has been discontinued. Parts include all materials,
charts, instructions, diagrams, accessories, et cetera, whic~ were furnished in the
standard or special models.
This warranty and conditicns of sale are in lieu of all others expressed or implied
and no
representative or person is authorized to assume for us any other liability
in co-:.:1ection with the sale ol our products.
The Triplett Electrical Instrument Company
Bluffton, Ohio
Printed in
Part
o. 84-89-011764-15
(For abov
erial
o. 29,500 and after April 30, 1963)
TABLE OF CONTENTS
3
Page
Familiarization . . . . . . . . . . . . . . . . . . . . . . . . . . . . .......... • • • 4
Operation
Measuring DC Volts . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Measuring AC Volts ................. ........ .. ...... 8
Measuring DC Resistance ... ......................... 10
Measuring DC Current . .... . ...... ............... . ... 12
Measuring Output Volts (DB) ........................ 14
Operation Chart ............... . ... . ............. ... . 16
Measuring Capacity .......... .. ........ . .. ........ . .. 18
Additional Measurements
Measuring Kilovolts ... . . ... .. ... ....... ... ...... ..... 20
Measuring High DC Current ....... ........... . ....... 21
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . • • • • • • 21
Additional Applications .... .. .......................... 22
Maintenance ...... . .... ..... .... .... .. ...... .. ... • • • • • 25
Interior View -
Parts Location ................. .. .... ... 26
Replaceable Parts .... ... ........................... • • . 27
Circuit Diagram ..... ..... . ... .. .. ............. .... . • . • 28
Data
DB Chart ........ ... ..... ..... ...................... 29
EIA Condenser Color Code .... . ... ...... .. ........... 30
EIA Resistor and Speaker Color Code .......... ...... . 31
EIA Wiring Color Code . ... .. ..... ... ...... . ... ... .... 31
EIA Transformer Color Code .. . ................. .. ... 32
The Model 630-PL Volt-Ohm-Milliammeter is a rugged
long scale multi-range instrument in a compact portable case.
Remove the test leads from the small envelope and notice
the two alligator clips enclosed. The clips slide over the ends
of the test prods and make very handy connectors.
Plug the small ends of the test leads into the panel jacks
marked COM and V-n-A. The jacks and plugs are espe-
cially made for trouble-free, low resistance connections.
Scales
Notice there are five scales on the meter. The top scale
(red) is used when measuring ohms. This scale is marked
from 0 to lK (at left side). With the switch knob turned to
nx 1, the ohms scale is read just as it is marked. With the
switch knob turned to nX 10, the numbers on ohms scale must
all be multiplied by 10. Likewise nx lK and OX lO0K mean
to multiply by 1000 and 100,000 respectively.
The second scale from the top (black) is used to read
all DC currents and voltages. The third from the top (red) is
used for all AC voltages except the 3 volt range-the latter
is read on the bottom red scale (marked 3 at full scale).
The lowest scale (black) is used for all decibel measure-
men ts. Notice the chart near the lower right hand corner of
the dial. This is used in conjunction with the DB scale as
explained on page 14.
P anel
Just below the meter is a small plastic screw. This is
rotated with a small coin or screw driver to adjust the meter
pointer to zero. This need be adjusted only occasionally
but for best accuracy the pointer should always be on zero
before making a measurement.
The large knob in the lower center of the panel is used
to select aH ranges. The markings are self explanatory.
Left of the knob is a recessed n ADJ control used when
making resistance measurements.
To the right of the knob is a recessed switch used for
reversing polarity on DC ranges.
Observe the jacks and note carefully the marking for
5
each. You will use the COM and V-n-A, jacks for most
measurements. The polarity reversing switch must be in the
"DC+, n, ACV" position !or all ohm and ACV measurements.
P R O T E C T I O N: A one ampere {use Is incorporated in the ohm circuits for
protectinq the ohm circuit when it is accidentally placed across high voltage. A
spare fuse is attached to unit inside the tester.
Accuracy
Your 630-PL is accurate to within 3 % of full scale
reading on all DC ranges except the 5000 volt range which
1s within 5 % , 4 % on all AC ranges except 5000 volts which
is 5 % , and 3 % of the scale length on all resistance ranges.
Precision, non-aging resistors insure lasting accuracy. All
units are calibrated at 77° F. AC ranges are calibrated on a
60 cycle sine wave. In choosing ranges always endeavor to
have the readings fall in the upper (or right hand) half of
the scale for greatest accuracy.
Ranges
The following ranges are self contained in your 630-PL:
DC Volts
2.5-10-50-250-1000-5000 at 20,000 Ohms
AC Volts
DC Microamperes
DC Milliamperes
DC Amperes
Ohms
Megohms
Output Volts(AC)
Decibels
per Volt/ 0-.250 at 10,000 Ohms per volt
0-3-10-50-250-1000-5000 at 5,000 Ohms
per Volt
0-100 at 250 Mv.
0-10-100-1000 at 250 Mv.
0-10 at 250 Mv.
0-1000-10,000 ..... ( 4.4-44 at center scale)
0-1-100 . . . . (4400-440,000 at center scale)
0-3-10-50-250-1000 at 5,000 Ohms per Volt
-20 to + 11, 21, 35, 49, 61, 75 on 600
Ohm line
Your 630-PL is well constructed but like any instrument
should be handled carefully. You will also want to keep the
panel clean as cleanliness and carefulness go hand in hand.
Measuring DC Volts (Fig. 1)
Rotate the selector switch to the appropriate range for
DC volts. Always start with the highest range if in doubt
as to the approximate voltage.
In choosing ranges, endeavor to have the readings fall
in the upper, or right hand, half of the scale for greatest
accuracy.
Plug the black test lead into the COM jack and the red
lead into the V-n-A jack as shown in Figure 1.
CAUTION: When measuring up to 5000 volts set the
selector switch on the 5000/ 1000 range, plug the red lead
into the jack marked 5000 DCV, and leave the black lead
in COM.
Connect the test prods ACROSS the voltage source. The
red lead is positive. Where polarity is difficult to determine,
the meter may read backwards. No damage will be done if
this occurs. Simply reverse the polarity switch.
The V-n-A jack will be positive when the polarity switch
is in the "DC+ " position.
Read all DC voltages on the top black meter scale.
Notice that the scales are not all marked exactly the same
as the range indicated by the knob position. Thus 0-2.5 volts
is read on the 0-250 scale (i. e. + by 100) on all readings,
the 0-1000 range is read on the 0-10 scale by adding two
zeros, and the 0-5000 range is read on the 0-50 scale by
adding two zeros.
The high sensitivity of 20,000 ohms per volt will allow
you to take measurements in low current circuits such as
grid and discriminator circuits.
Voltages as low as .005 volts, ( .250 volts full sc~e al 10,000 ohms per volt)
may be read by using 100 uA DC current range. Simply place the range selector
in the 100 uA position and read the voltage on the 0-250 scale (divide by 1000 for
volts).
CAUTION: For maximum safety do not handle tester or leads when connected
to high voltages. Make certain that no condensers are charged to a high volta ge.
NOTE: To measure above 1000 volts, the red lead must be plugged in o the
5000 DCV jack and selector switch set on the 5000/1000 DCV range.
For handy operation chart see pages 16 and 17.
50
DC
0
5fc°v°
~
0
,i. A
OUT Q PUT
DC-
COM-
V-Il:A
0
Figure I - Measuring DC Volts
1
Measuring AC Volts {Fig. 2)
Rotate the Selector switch to the appropriate range for
AC volts and set the polarity switch to ACV. Always start
with the highest range if in doubt as to the approximate
voltage.
In choosing ranges, endeavor to have the readings fall
in the upper, or right hand, half of the scale for greatest
accuracy.
Plug the black test lead into the COM jack and the red
lead into the V-n-A jack as shown in Figure 2.
CAUTION: When measuring up to 5000 volts set the
selector switch on the 5000/1000 range, plug the red lead
into the jack marked 5000 ACV, and leave the black lead
in COM.
Connect the test prods ACROSS the voltage source. As
there is no polarity on AC, the red and black leads may be
interchanged without causing the meter to read backwards.
Use the two lower red scales for AC volt readings. Note
that the bottom scale is to be used ONLY for one range,
namely the 0-3.
When using the 0-1000 AC volt range, read on the 0-10
scale by adding two zeros. When on the 5000 volt range
use the 0-50 scale by adding two zeros.
W i ih the special design of the two diodes in a modified
full wave bridge circuit, the 630-PL gives a wide frequency
response usable from 10 c p s to 500 Kc.
CAUTION: For maximum safety do not handle tester
or leads when connected to high voltages.
NOTE: To measure above 1000 volts, red lead must be
plugged into the 5000 ACV jack and the selector switch set
on the 5000/1000 ACV range.
For handy operation chart see pages 16 and 17.
i~
0
5fc°v°
"f<
0
l'A
OUT Q PUT
=:JII
Figure 2-Measuring AC Volts
9
Measuring DC Resistance (Fig. 3)
Rotate the selector switch to the appropriate range for
ohms determined from the following chart:
To Read
Set Switch to
0-100□- ohms ....... .. .. ... ........... nX 1
0-10,000 ............... .............. nx 10
0-1 Meg. . ................ .... ...... . nX lK
0-100 Meg ........................... nX l00K
Set the polarity switch to n.
Plug the black test lead into the COM jack and the red
lead into the V-n-A jack as shown in Figure 3.
Short the test prods together and adjust then ADJ knob
until the meter pointer reads 0 on the top red (n) scale.
(The 0 for the ohms scale is at the extreme right side of
the scale.)
Connect the test prods across the resistor as shown. If
the resistor is wired in a circuit, disconnect one end of the
resistor before taking the reading.
Each time an ohm range is changed, it is well to check
the 0 setting as outlined in paragraph 4 above.
The basic scale 0-lK (0-1000) is used for reading on all
ohm ranges. Simply multiply the scale numbers by 1, 10,
1000, or 100,000 as indicated by the selector switch setting.
It should be kept in mind that in the measurement of
resistance a current is passed through the unknown resistor.
Generally this current is so small as to be negligible. How-
ever on the 0-1000 range fairly high current is employed.
CAUTION: On XI ohm position reading approximately center scale (4.4 ohms)
the current drain from the J.S volt battery is 170 MA. It is desirable to make a
practice of usinq one of the hiqher ohm ranges for general continuity or circuit
testing. This will give you much longer battery life.
NOTE: In the ohms circuit the battery polarity at the leads is as indicated.
The red lead is positive.
Since the scale of an ohmmeter is non-linear, the accuracy of the readinq
cannot be expressed as a per cent of full scale. Ohmmeter accuracy is generally
referred to a linear scale such as the DC volt scale. Thus ±3% ohmmeter
accuracy means an allowable ± l.S divisions on the DC scale. For example 2
ohms could read from about 1.75 to 2.3 ohms and be within tolerance.
NOTE: Do not touch any metal parts of the circuit when using the high ohm
ranges. It is not dangerous but your body resistance can cause serious error.
,For handy operation chart see pages 16 and 17.
500
DC
0
5000
~
~V
*A
0
A
UNKNOWN
RESISTOR
OUT Q PUT
Figure 3-Measuring DC Resistance
11
Measuring DC Current (Fig. 4)
Rotate the selector switch to the appropriate range for
DC current. Always start with the highest range if in doubt
as to the approximate current.
In choosing ranges, endeavor to have the readings fall
in the upper, or right hand, half of the scale for greatesi
accuracy.
Plug the black test lead into the COM jack and the red
lead into the V-0-A jack as shown in Figure 4.
Connect the test prods in SERIES with the circuit to be
measured. Do not test directly across any potential circuits
as this may burn out the instrument and shunt. The red
lead is positive. Where polarity is difficult to determine, the
meter may read backwards. No damage will be done if
this occurs. Simply reverse the polarity switch.
Read all current ranges (including microamperes, mil-
liamperes, and amperes) on the upper black scale. When on
the 10 Ma. range use the 0-10 scale and read direct. On the
100 Ma. range again use the 0-10 scale by adding one zero.
CAUTION: Turn off the power before connecting the
meter to the circuit. Do not handle tester or leads in high
voltage circuits.
In using the 100 microampere range, the meter may read
differently than calculations would indicate. This is some-
times caused in low current circuits by a slight leakage of
voltage due to moisture. Other times a slight potential is
generated by soldering or joining of dissimilar metals. Even
the proximity of fumes or liquid acids and alkalies may react
with the metal parts of the circuit and generate slight cur-
rents. The fingers should not be permitted to touch the metal
parts of the prods or circuit as body resistance can also up-
set some circuits.
For handy operation chart see pages 16 and 17.
500
DC
0
5000
"!(
ACV
0
p.A
OUT Q PUT
OPEN
CIRCUIT
RED
Figure 4- Measuring DC Current
13
Measuring Output Volts (DB) (Fig. 5)
Output is generally measured in units called the decibel,
a terminology used to indicate power levels in amplifiers or
telephone work.
Do not confuse the DB with the VU (Volume Unit). The
VU is based on .001 watt dissipated in a 600 ohm line and
is measured with a meter having special ballistic charac-
teristics.
Rotate the selector switch to the appropriate AC volt
range and set the polarity switch to ACV. Refer to the small
chart on the meter dial for the range to use. Always start
with the highest range if in doubt as to the approximate
number of decibels.
Normally it is recommended to measure output by plug-
ging the black lead into the COM jack and the red lead
into the OUTPUT jack as shown in Figure 5.
Oftentimes a DC voltage is present in the circuit where
output is to be measured. The extra jack marked OUTPUT
with a .1 mfd condenser in series is provided to block the DC.
The condenser impedance is generally disregarded in
most measurements. Where no DC is present, this output
voltage can be read accurately by using the 630-PL as a
regular AC voltmeter (i. e. by plugging the red lead into the
V-n-A jack instead of OUTPUT.)
Connect the test prods across the plate circuit or speaker
voice coil.
Read all DB ranges on the bottom black scale using the
small chart on the meter dial. For example, when the selector
switch is set on the 3 AC volt range, the DB scale is direct
reading. When on the 10 AC volt range, add 10 to reading
on the DB scale, thus with the meter reading -2, the ap-
proximate DB reading is + 8 DB.
For a reference level of .006 watt in 500 ohms, subtract
( + )7 DB from the reading to obtain the absolute value of
decibels.
For handy operation chart see pages 16 and 17.
500
DC
0
5000 :1
ACV
•A-A
0
Jl.A
OUT Q PUT
600
OHM
LINE
When using the 5000 AC voltage range for
output measurements ii i1
necesaary to
::>lace a condenser capable of withstandini
this high voltage extemal to the tuter and
connected in series to the 5000 AC volt fack
mung the COM lack to complete the circuit.
This e'ICtemal condenser is not a furnished
part of the equipment.
Figure 5-Measuring Output Volts (DB)
15
0- 1000 Ma
DC
0-10
AMPS DC
OHMS
0-1000
0- 10,000
0- 1 Meg
0-100 Meg
DECIBELS
-20 to
+11
-10 to
+21
+ 4 to
+35
+18 to +49
+30
to +61
+44
to +1s
OPERATION CHART
Set Switch To
Selector
Polarity
Plug Red* Test Lead
In Jack Marked
Read
On
Each Scale
Div. Equals
100
p.A
2.5
DCV
10
DCV
50
DCV
250
DCV
1000/5000 DCV
1000/5000 DCV
3
ACV
10
ACV
50
ACV
250
ACV
1000/5000 ACV
1000/ 5000 ACV
100
µ.a
10
DCMA
100
DCMA
1000
DCMA
10
AMP
nxl
nXlO
nXIK
nXlO0K
3
ACV
10
ACV
50
ACV
250
ACV
1000/5000 ACV
1000/5000 ACV
either
either
either
either
either
either
either
V-n-A
V-n-A
V-n-A
V-n-A
V-n-A
V-n-A
5000 DCV
BLACK SCALE
0-250
DC---;-1000
0-250
DC---;-100
0- 10
DCXI
0-50
DCXl
0-250
DCXl
0- 10
DCXlO0
0-50
DCXlO0
0.005
0.05
0.2
1.0
5.0
20.0
100.0
Volt
Volt
Volt
Volt
Volts
Volts
Volts
-
-
- ------- ---------1--------
ACV
ACV
ACV
ACV
ACV
ACV
either
either
either
either
either
n
n
n
n
ACV
ACV
ACV
ACV
ACV
ACV
V-l-A
V-p-A
V,h-A
V-n-A
V-n-A
5000 ACV
V-n-A
V-n-A
V-n-A
V-n-A
V-n-A
V-n-A
V-n-A
V-n-A
V-n-A
t
O4tput
O11Jput
Output
Output
Output
5000 ACV
RED SCALE
0-3V
ACXl
0- 10
ACXl
0-50
ACXl
0- 250
ACXl
0- 10
ACXlO0
0-50
ACXlO0
BLACK SCALE
0-10
DCXlO
0- 10
DCXl
0- 10
DCXlO
0-10
DCXlO0
0- 10
DCXl
0.05
0.2
1.0
5.0
20.0
100.00
2.0
0.2
2.0
20.0
0.2
Volt
Volt
Volt
Volts
Volts
Volts
µ.a
Ma
Ma
Ma
Amp
Each Scale Division Equals (in ohms}
RED SCALE
I
0-4
I 4-5
I 5-10 I 10-20 I ?.0-50 I 50-200
OX !
OXIO
j . I
I .2
I .5
I 1
I 5
I ~-
I 1
I 2
I s
I 10
I so
I soo
I 100 I 200
I 500
I lK
I SK
I SOK
OXIK
OXIOOK
I lOK I 20K I SOK I IOOK I SOOK I SMea
BLACK SCALE
DB plus 0
DB plus 10
DB plus 24
DB plus 38
DB plus 50
DB plus 64
* Black test lead
plugged in "COM"
jack for all
measurements
17
Measuring Capacity (Fig. 6)
Your 630-PL can be used to measure capacity by the
arrangement shown in Figure 6. It is set up as an AC volt-
meter. See page 8.
Use the following chart to determine the AC voltage
range to use. ALWAYS start with the selector switch on the
250 volt range since, if the condenser is shorted, serious
damage may result to the meter when on a low range.
To
Measure
MFD
.0021
.004 I
.006 ~
.oos I
.010 J
Set
Selector
Switch to
3 ACV
Deflection
in
AC Volts
f
.45
I
.83
~
1.25
I 1.65
t 2.10
CAUTION: Do not attempt to use this test on electrolytic
condensers.
500
DC
X
0
5000
':"(
ACY
V·Il:A
0
/lA
OUT Q PUT
IIOV
60~
UNKNOWN
CAPACITOR
.5MFD
Figure 6-Measuring Capacity
19
For measuring the high voltage employed in television receivers and in other
applications, an external probe is available. See Figure 7. Probes for DC are
made in three ranges 0-10,000, 0-25,000, and 0-50,000 volts. Probes for AC are
also available in the lower ranges.
To use the Kilovolt probe set range selector switch on 2.5 volt DC position
if the DC probe is to be used. When the AC probe is used set range selector
switch on 3 Volt AC position and set the polarity switch to ACY. Pluq the Kilovolt
probe lead into the V-0-A jack on bottom riqht comer of the front panel. Use
the black standard lead for negative or common lead with it connected in the
COM jack located on the lower right side of front panel.
Kilovolt
Set Selec-
Probe
tor Switch
0-IOKV DC
2.5V DC
0-25KV DC
2.5V DC
0-50KV DC
2.5V DC
0-IOKV AC
3V AC
0-25KV AC
3V AC
Figure 7
Read on
Range
0-IOV DC
0-250V DC
0-50V DC
0-IOV AC
0-250V AC
Multiply
by
1000
100
1000
1000
100
USE EXTREME CAUTION
in measuring the high
voltages such as found
in
television
receivers.
Clip the common lead
to the circuit so you
will
have
only
one
probe in your hand.
21
Measuring High DC Current
External plug-in shunts are available to extend the DC
current ranges of your 630-PL from the self-contained 0-10
to 0-25 amps. External portable shunts up to 100 amperes
are also available. (See paragraph on accessories.)
Set the 630-PL selector switch to the 10 Ma. position
and plug the desired external shunt into the COM and V-n-A
jacks. Connect the line to be measured to the binding posts
on top of the shunts. The external portable shunts are too
large to plug into the panel and must be connected to the
panel jacks by the leads furnished with the shunts.
Accessories
The following accessories for your 630-PL are available
from your distributor.
Item
DC Hi-Voltage probe 0-10 Kv
AC Hi-Voltage probe 0-10 Kv
DC Hi-Voltage probe 0-25 Kv
AC Hi-Voltage probe 0-25-Kv
DC Hi-Voltage probe 0-50 Kv
Carrying cases
639, 639-N and 639-P
Plug-in external shunt 0-25 DC Amp.
Portable external shunt 0-50 DC Amp.
Portable external shunt 0-100 DC Amp.
Clamp-on Ammeter Adapter, Model 10
(Ranges 6 to 300 AC Amps)
Lead Assembly, No. 611
(Used in conjunction with Model 10)
Line Separator, Model 101
(Used in conjunction with Model 10)
Part No.
T-79-83
79-165
T-79-130
79-166
T-79-92
T-91-243
T-91-244
T-91-191
60A-211
79A-160
60A-218
In The Home
When your refrigerator motor fails to "kick out" the
starting winding, use the 630-PL, to measure the AC line
voltage. If the voltage is below about 100 volts, notify your
power company.
If your electric stove does not seem to heat quickly
enough, measure the voltage input to the stove with all
burners turned on and again with all burners turned off.
If the difference between these two voltages is 10 or 15 volts,
the power cable to the stove has defective connections or
is not of large enough current carrying capacity.
Blown fuses sometimes do not visibly indicate they are
burned out. With your 630-PL measure the voltage ahead
and behind the fuse. Voltage ahead of the fuse but no
voltage following indicates a blown, defective, or loose fuse.
Sometimes it is easier to remove the fuse and measure its
resistance. This should be substantially zero.
Your 630-PL is handy for locating trouble in desk and
floor lamps. Pull the plug from the wall socket and check for
a faulty cord, plug, switch, socket, or bulb by measuring
resistance on the n X 1 range. 100 watt 120 volt bulbs should
read 10 to 20 ohms. 50 watt 120 volt bulbs should read
20 to 40 ohms.
For the Radio Man
In addition to all common voltage, current, and resist-
ance measurements used in servicing radios, the high sensi-
tivity of your 630-PL is well adapted to measuring AFC,
A VC, bias, and FM discriminator voltages.
Measurements of the high voltage up to 27,000 volts D. C.
used in some television receivers for the picture tube can
be effected with the special high voltage probe shown on
page 20.
Considerable trouble is had with leakage in automobile
radio antennas (due to moisture). Your 630-PL with the
h~gh ohm range 0-100 meg. is ideal to check this leakage.
Disconnect the antenna from the receiver before making this
check.
High frequency measurements are possible since the
23
630-PL has a wide frequency response-usable from 10
cps to 500 Kc.
In The Industrial Plant
Your 630-PL will be a big help in checking voltage
drop caused by adding that extra machine on the already
overloaded line. Correcting this will often save time later
when a rush comes and the line "just happens" to burn up.
Measure the voltage at the machine first with the ma-
chine turned off and again with the machine in operation.
If the voltage is proper with the machine off but low with
the machine in operation, the circuit wiring or transformers
have too small a capacity. If the voltage is low even with
the machine off, the circuit is probably already overloaded
and the machine should be wired into another circuit.
Equipment using automatic electric controls can be
checked with the 630-PL. Faulty relay or control action is
often caused by low voltage applied to the relay or control.
This low voltage in turn, may be caused by burned or dirty
contacts on the control device. Use the nX 1 range to check
for high or unstable contact resistance.
When a phone on your dial telephone system fails,
measure the line current and the voltage to the particular
relay in question. If the voltage is proper, measure the con-
tact resistance of the relay contacts using the nX 1 scale on
your 630-PL. If this resistance is over a fraction of an ohm
or if the resistance seems to waver, clean and adjust the
relay contacts.
In The Garage
Fuses in the automobiles have a tendency to look per-
fectly good and yet not function due to corrosion under the
metal end cap. Measure voltage ahead and behind the fuse
to determine a defective unit. Or remove the fuse and
measure its resistance. Anything over a fraction of an ohm
is too high.
. Checking automobile wiring, light switches, heaters,
radios, etc., can be speeded up by simple use of your
630-PL.
In The Laboratory
Your 630-PL is built with all prec1s1on, non-aging
resistors. The specially designed switch and special banana
type plugs insure lasting accuracy. The meter with specially
finished and selected pivots and jewels and a well designed
stable magnet further makes the 630-PL a must for the
laboratory.
Special Applications
The unusually high range ohmmeter in your 630-PL
permits some indication of condenser leakage resistance.
Measure as a resistor, see page 10, using the highest range.
A good paper or mica condenser under 1 mfd. will indicate
at the 100 Meg mark or above. If a steady reading (taken
after the initial surge required to charge the condenser) of
less than 100 megohms is obtained, the condenser probably
has defective insulation. Good paper condensers over 1 mfd.
may read somewhat less than 100 megohms. Electrolytic
condensers should read above .1 megohm. In checking
electrolytic condensers, the black test lead ("com" jack)
should be connected to the positive terminal of the condenser.
Checks of insulation resistance for motors, generators,
telephone cables, power cables, etc., can be made on the
high ohmmeter range of your 630-PL. The actual value of
resistance may vary from a few megohms to over 100 meg,
depending on weather conditions and quality of insulation.
The best method, therefore, is to make periodic checks on
important cables or equipment and observe the trend in
readings. As the readings tend to be lower and lower, it
is time to start drying out the equipment or determine the
cause of deterioration. Dirt, mice, or foreign matter can
sometimes cause excessive leakage.
MAINTENANCE
25
:Battery Replacement
Two batteries are used for the ohmmeter circuits, a 1.5
volt Eveready No. 950 or equivalent and a 30 volt Eveready
No. 413 or equivalent.
When the meter pointer can no longer be adjusted to
zero (see page 10)' ohms on the nx 1, nx 10, or nx 1000
ranges, replace the 1.5 volt battery.
When the meter pointer can no longer be adjusted to
zero ohms on the nx 100,000 range, replace the 30 volt
battery.
To replace batteries, remove the four screws in the bot-
tom of the case and lift panel from the case. Simply remove
the old, and replace with a new battery being careful to
observe polarity.
Care
A little precaution in handling and caring for your
630-PL can pay big dividends in satisfaction. Treat your
tester like you would a fine watch.
Avoid placing your tester on a bench where machine
tools are used or severe vibration is encountered.
Do not place your 630-PL near the edge of a shelf or
bench where it can easily be knocked off.
When possible keep your 630-PL in a place of moder-
ate temperatures. A void places with extreme temperatures
or severe temperature changes.
In .ise, don't take chances on overloading the meter.
If in doubt as to the approximate reading always start with
the highest range.
Get in the habit of double checking the position of the
switch before making a measurement. The meter can be
burned out by applying voltage when the switch is set on
the current or ohms scale.
If the unit has not been in use for a long period of time, rotating the switch
in both directions several times will wipe the contacts clean for good contact.
Turn the selector switch to OFF when the unit is to be
carried. The meter is damped in this position to prevent wild
swinging of the pointer.
100 DC UA.
R9
10 DC MA.
RS
100 DC MA.
R7
1000 DC MA.
X
AC
RECTIFIER
ASSEM.
R5 R6
REdTIFIER
CIRCUIT
R4
10 ACV
R3
50 ACV
R2
250 ACV
RI
1000 ACV
R3O
5000 DCV
Ref. No.
Rl
R2
R3
R4
R6, R6
R7
R8
R9
Rll
RlO
R12
R13
R14
R16
R16
R17
R18
R19
R20
R21
R22}
R23 f
R27
R28
R29
R30
R31
C
B2
Bl
M
X
Sl
s2
Fl
REPLACEABLE PARTS 630-PL
27
DESCRIPTION
Part No. Req,
Resistor
Resistor
Resistor
Resistor
Resistor
Resistor
Resistor
Resistor
Resistor
Resistor
Resistor
Resistor
Resistor
Resistor
Resistor
Resistor
Resistor
Resigtor
Resistor
Resistor
Resistor
3.75 Meg.,
1 Meg.,
200K,
35K,
6K,
.25 ohms,
2.5 ohms,
25.12 ohms
423K,
5K,
13.6K,
4690
37.2,
3.7 ohms,
732,
6.5 ohms,
45K,
150K,
SOOK,
4 Meg.,
7.5 Meg.,
±1%, 1hW, film type T-15-2842
±1%, ½W, film type T-15-1358
±1%, ½W, film type T-15-1061
±1%, ½W, film type T-15-2843
±1%, ½W, film type
T-15-1009
±1 %, Wire type T-16-1528
±1 %, Wire type T-15-1976
± 1 % Wire type 15-2850
±1%, ½W, film type T-16-4120
±1 %, ½ W, film type T-16-1009
± 1 %,
film type T-15-1768
± 1%, ½W, film type 15-4118
± 1%, ½W, film type T-15-4116
±1%, Wire type T-15-3223
± 1 %, ½ W, film type T-15-4119
±1 %,
Wire type T-15-4117
±1 %, ½ W, film type T-15-1060
± 1 %, ½ W, film type T-15-1763
±1%, ½W, film type T-15-1137
±1%, 1hW film type T-16-1138
±1 %, ½ W, film type
T-15-1262
Resistor
20K,
±20%, Variable
16-31
(with mtg. nuts & knob)
Shunt Assem. 10 Amp
Resistor
20 Meg.,
Resistor
60 Meg.,
Resistor
5100,
Capacitor .1 mfd., 400WV
±1%, 2W
±1%, 2W
±5%, ½W
Battery
1.5 Burgess No. 2, NEDA No. 13
Battery
30V, Burgess, U20E, Eveready 413,
or NEDA No. 210
Meter
60 Micro-Amps, 250 Millivolts
Rectifier assembly
Switch
20 Pos., 3 deck, without resistors
Switch
20 pos., 3 deck, with resistors
Switch Slide, DPDT (Wirt SW726)
Knob
Scarlet, 2 5/16 long with clip
Case
with handle
Jack
contact
Clip
shunt retainin_g- No. C2416-012-27
Bumper
rubber
Leads
Banana Type
Front
Clear plastic with zero adj.
Fuse
1 Amp. Littlefuse, 3AG, 312001
90A-449
15-1674
15-2844
15-1411
T-43-69
f Available
} Locally
52-1294
2250A 30
22A 319
22-320
T-22-152
34 B-62
T-10-784
8944
T-2451-6
T-3236-10
T-79-127
T-10-1225
3207-15
• 630•,Pl SPECIAL
1121
DC 5000 AC
V-A-A
COM
11000
1000
250
110
10
2.11
XI
XIO
)CIK
XIOOK
D
C
V
0
M
M s
100 ) ~
100
~
,o l
1000
10
) A
0 3
0 10
0 50
o 2110
0 1000
11000
0,,
A
C
V
RESISTORS R58R6 ARE PART OF RECTIFIER CIRCUIT BUT MOUNTED
ON SWITCH
Figure 8-Circuit Diagram
DATA
29
DB CHART
0 DB at 1 Mw
RMS Volts with
Decibels with
Line Power
line Imped. of
500 ohm line
Mw
600 ohms
-20
.01
.0775
-10
.l
.245
-5
.316
.436
0
1.00
.775
+10
10.0
2.45
+ 15
31.6
4.36
+ 20
100
7.75
+ 30
1000
24.5
+ 40
10,000
7'1.5
+ so
100,000
245.
+ 60
1,000,000
775.
+ 70
10,000,000
2450
Note:
The range of audibility can be considered to lie from
70 db below the normal speech level to 70 db above the
same level, or a total range of 140 db.
EIA MICA CONDENSER COLOR CODE
MOLDED PAPER
MOLDED MICA
CERAMIC
Color
Multiplier
Tolerance
Multiplier
Tolerance
Multipler
Tolerance
Black
I
20%
I
20%
l
20% or 2.0µµfd . •
Brown
10
10
10
l o/c
Red
100
100
20% EIA
100
2i
Orange
1000
1000
3% EIA
1000
2. % EIA
Yellow
10,000
5%
10,000
10,000
or 0.5µµfd .•
Green
5% EIA
5%
Blue
Violet
0.01
0.25µµld .•
Gray
1001,
0.1
10% or l.0µµfd . •
White
Gold
0.1
0.1
Silver
So/co
10%
0.01
5% (JAN) 10%
• Capacitance
None
20%
less than lOµµfd .
MOLDED
PAPER
MOLDED
MICA
CE"RAMIC
TUBULAR
.!!!!!.!..
.!.£Q!.
~
6 COLOR RAOIAL LCAO
@aonoe 0
,1~11 ~I r ~
8LACK A B
-mJ}
l JABCO
.aecD EF
0
5 COLOR ~AOIAL LEAO
FLAT
JAlf TYPE
G D C
G D C
@aaoa 0
~
SUTTON SILVER MICA
~-
H ABC D
G
e
5 COLOR AXIAL LEAD
0
-m~n~
~
G O C
C
3•dOIGIT
FLAT COM1L. TYPES
OBSOLETE
RMA
SYSTEMS
HA II CO
5 COLOR STANO-OFF
BLACK BODY E
A B 3rd0tGIT
~ ~
~
6rfm}
AB C
DOT o o O iW
ABC
H4 8 CD
E O C .
~
SCOLOR
3 COLOR
A 8 C
4w
0
DISC
DISC
[
DOT~
SLANI(
FRONT
"~
8
-{coo9>~]
A 8 C
ED
REAR
'O'
o~~ ~
-B-
E 0
A 8 C 0
A 8 C
A BC,
(Courtesy Popular Electronics)
Ca}Xlcitance is given in µµId .
Colors-$ame value as on resistors except as indicated in tables.
COLORS
INDICATES
A
Brat iligit
B
Second digit
C
Multiplier
D
Tolerance
E & F
Voltage Rating in hundreds of volts
( (E) Rati!}gs less than 1000 volts. (E) & (F) Firat two digits of ratings lCXXJ volts
or more. Values of colors for (E) & (F) are same as in resistance values. (q) ii
class or characteristics of caJXJcitor. (H). (I) & (J) give temperature coefficient.
(G), (H). (I) & (J) are not listed in the tables.]
Color
Black
Brown
Red
Oranqe
Yellow
3reen
Blue
Example:
DATA
31
EIA RESISTOR COLOR CODE
A
8
C
0
Color
A
B
C
D
Number
0
I
2
3
4
5
6
8
A
C
Indicates
First number
Second number
Number of zeros
Tolerance
Color
Violet
Gray
White
Gold
Silver
None
Number
7
8
9
5°1,, tolerance
10°1,, tolerance
20°/,, tolerance
The resu)tinq value is in ohms.
A 250,000 ohm 200/n resistor.
A red
B green
C yellow
D
no color
EIA SPEAKER COLOR CODE
Voic-
Coil:
Gr;,en-finish
Black-start
Field Coils:
Black and red-start
Yellow and red-finish
Slate and red-tap (if any)
EIA WIRING COLOR CODE
B+
Ground
Plate
Grid
Cathode
High Heater
Low Heater
Screen Grid
AVC
Red
Black
Blue
Green
Yellow
Brown
-
Black
Orange
White
EIA TRANSFORMER COLOR CODE
I. F. Transformers:
Blue- plate lead
Red-"B" + lead
Green-grid (or diode) lead
Black- grid (or diode) return
NOTE: If the secondary of the i. f. t. is center-tapped, the
second diode plate lead is green-and-black striped, and black
is used for the center-tap lead.
Power Transformers:
1. Primary Leads .... Black
If tapped:
Common .... . .... Black
Tap-Black and Yellow
Striped Finish-Black and
Red Striped
2. High-Voltage Plate Wind-
ing . ............... Red
Center-Tap -
Red and
Yellow Striped
3. Rectifier Fil. Winding ...
Yellow
Center-Tap-Yellow and
Blue Striped
4. Fil. Winding No.I. Green
Center-Tap -
Green and
Yellow Striped
5. Fil. Winding No. 2. Brown
Center-Tap- Brown and
Yellow Striped
6. Fil. Winding No. 3. Slate
Center-Tap -
Slate are
Yellow Striped
A. F. Transformers:
Blue-plate (finish) lead of
primary
Red-"B" + lead (this ap-
plies whether the primary
is plain or center-tapped).
Brown- plate (start) lead on
center tapped primaries
(Blue may be used for
this lead if polarity is not
important.)
Green- grid (finish) lead to
secondary
Black-grid return ( this ap-
plies whether the second-
ary is plain or center-
tapped.)
Yellow-grid (start) lead on
center
tapped
second-
aries. ( Green may be used
for this lead if polarity is
not important.)
Note: These marking ap-
ply also to line-to-grid, and
tube-to-line transformers.