Trinity SLM102 Operation & User’s Manual
Read Online
Read the Trinity SLM102 Operation & User’s Manual online using the built-in PDF viewer.
Text of the manual
SLM102 – Operational Manual
TRINITY
[1]
USER’S MANUAL
SLM102
0.2 Class Multifunction Power and Energy meter
This document contains the latest technical information about SLM102 which
is a micro-controller based 0.2 Class Multifunction Power and Energy meter.
The unit is tested against latest "MTE" Standard Model PRS400.3 having
basic accuracy of 0.02%, traceable upto International Standards derived using
appropriate ratio techniques.
The product, SLM102 is sophisticated electronic equipment, and the user is
advised to read this User’s Manual carefully before attempting to install or
operate the equipment.
Published on:------
Document Version: 1.0
SLM102 – Operational Manual
TRINITY
[2]
Warranty statement
Trinity warrants to the original retail purchaser of the Trinity product enclosed with this
limited warranty statement that the product, if purchased new and used in the India
conforms to the manufacturer’s specifications and will be free from defects in
workmanship and materials for a period of one year from the date of original
purchase, unless expressly stated otherwise by Trinity, in a written format.
Should your Trinity product prove defective during the warranty period, please bring
the product securely packaged in its original container or an equivalent, along with
proof of the date of original purchase, to our Trinity Dealer or Factory. You are
responsible for all costs (shipping, insurance, travel time) in getting the product to the
service location. Trinity will, at its option, repair or replace on an exchange basis the
defective unit, without charge for parts or labor. When warranty service involves the
exchange of the product or of a part, the item replaced becomes Trinity property. The
replacement unit may be new or refurbished to the Trinity standard of quality, and at
Trinity’s option, the replacement may be another model of like kind and quality.
Trinity’s liability for replacement of the covered product will not exceed the original
retail selling price of the covered product. Exchange or replacement products or parts
assume the remaining warranty period of the product covered by this limited warranty.
What This Warranty Does Not Cover:
This warranty does not apply to refurbished or reconditioned products. This warranty
covers only normal use in India. This warranty does not cover damage to the Trinity
product caused by parts or supplies not manufactured, distributed or certified by
Trinity. This warranty is not transferable. This warranty does not cover third party
parts, components or peripheral devices added to the Trinity product after its
shipment from Trinity. Trinity is not responsible for warranty service should the Trinity
label or logo or the rating label or serial number be removed or should the product fail
to be properly maintained or fail to function properly as a result of misuse, abuse,
improper installation, neglect, improper shipping, damage caused by disasters such
as fire, flood, and lightning, improper electrical current, interaction with non-Trinity
products, or service other than by an Trinity Authorized Service.
The warranty and remedy provided above are exclusive and in lieu of all other
express or implied warranties including, but not limited to, the implied
warranties of merchantability or fitness for a particular purpose. In the event,
the remedies above fail, Trinity’s entire liability shall be limited to a refund of
the price paid for the Trinity product covered by this limited warranty. Except
as provided in this written warranty, neither Trinity Energy Systems Pvt. Ltd.
nor its affiliates shall be liable for any loss, inconvenience, or damage,
including direct, special, incidental, or consequential damages, resulting from
the use or inability to use the Trinity product, whether resulting from breach of
warranty or any other legal theory.
SLM102 – Operational Manual
TRINITY
[3]
Contents
Introduction ................................................................................................................ 4
Main Features Available in the Unit ...................................................................... 4
Technical Specifications ....................................................................................... 7
Installation and Commissioning ............................................................................. 11
3P4W Installation Mode ...................................................................................... 11
3P3W Installation Mode ...................................................................................... 13
Connection Scheme ............................................................................................ 14
RS485 Connection .............................................................................................. 15
Operational Details ................................................................................................... 16
Programming Mode ................................................................................................ 17
Setting CT-Primary and date according to RTC ................................................. 17
Setting PT Ratio and Time according to RTC .................................................... 18
Setting Log Duration, Demand Window and Unit Address ................................. 19
Selecting Demand On, Installation type and CT-Secondary .............................. 20
Setting Four Alarm Relays, Alarm On, Alarm Value and Alarm Hysteresis ....... 21
Setting Four Events, Event On, Event Value and Event Hysteresis .................. 23
Selecting Baud Rate for RS485 Port and Changing Password .......................... 24
Setting IP Address .............................................................................................. 24
Setting MAC Address .......................................................................................... 25
Resetting Demand Manually / Daily/Monthly ...................................................... 27
Resetting Energy ................................................................................................. 28
Resetting Minimum and Maximum Parameters .................................................. 28
Resetting KW Demand and KVA Demand ......................................................... 29
Resetting Log ...................................................................................................... 30
Resetting Events Logging ................................................................................... 30
Resetting All Parameters .................................................................................... 31
Run Mode ............................................................................................................... 32
Run Mode Displays ............................................................................................. 32
Communications ...................................................................................................... 47
MODBUS RTU on RS485 Port ........................................................................... 47
Ethernet TCP/IP Communication ........................................................................ 48
SLM102 connected as Mini Web Server to a web browser. ....................... 48
Data Logging and USB Port ................................................................................ 49
Installing the USB Software ................................................................................ 49
Installation of USB driver .................................................................................... 50
Handling the Logger Window .............................................................................. 53
Downloading the Logged data ............................................................................ 53
Exporting the Logged Data including Min/Max Events and Harmonics Data ..... 54
Graph Representation ......................................................................................... 55
Plotting a Graph for Logged data. ....................................................................... 55
Programming Mode Using USB Utility ................................................................ 57
Control Outputs ........................................................................................................ 59
Alarm Action ........................................................................................................ 59
SLM102 as a Demand Management .................................................................. 60
Appendix ................................................................................................................... 61
Address map of RS485 Port for various parameters of both 3P4W and 3P3W. 61
SLM102 – Operational Manual
TRINITY
[4]
Introduction
SLM102 is a micro-controller based 0.2 Class Multifunction Power and Energy meter
which not only measures a host of electrical parameters to display three rows of eight
digits, of Red Seven Segment LEDs display along with eighteen LED indicators. The
unit acts as a comprehensive load managing device due to its four numbers of
outputs relay contacts. These outputs are individually field programmable for both
the parameter on which to generate an alarm as well as the values on which to
activate an alarm and deactivate it.
The unit is meant for use in three phase four wire and three phase three wire
systems. The installation type, CT Ratios and PT Ratios are site selectable. The unit
can also be used as Load Manager with its four relay contacts as well as Demand
Controller. The method of Demand calculation i.e. Sliding window, Fixed window can
also be selected at site.
SLM102 has three ports for communication i.e. RS-485 for connection to
SCADA/EMS, USB 2.0 for downloading of logged data and programming of field
parameters, and Ethernet Port with TCP/IP protocol support for connection with
LAN/Internet. It has 2 MB of memory for data logging of all measured parameters,
min-max values and events.
SLM102 is a versatile meter, with all the features needed to implement a robust
electrical load management system. It can be configured to suit most control and
communication needs.
Main Features Available in the Unit
Exceeds class 0.2 revenue accuracy in Power & Energy Metering.
On board Ethernet port supporting TCT/IP protocol. Acts as a small web-
server. Also supports MODBUS TCP/IP.
RS485 port with MODBUS-RTU protocol.
Demand Measurement: Fixed or Sliding for KW or KVA.
Predictive Demand for Fix 30 minutes Window
Harmonics measurement: All six waveforms upto 29th including THD.
Bright Red LED display. Three rows of 8-digits each ensure unbroken display
even large energy register values.
Four relay outputs for Alarm & Trip. Limits are field programmable.
Event definition and logging.
Data logging of all parameters for load pattern analysis. 2 MB memory on-
board.
USB 2.0 port to download logged data.
USB 2.0 port based programming of field selectable parameters.
SLM102 – Operational Manual
TRINITY
[5]
Comprehensive and Accurate Measurement:
3Φ Voltage (L-N), 3Φ Voltage (L-L), 3Φ Current
KW (3Φ and Total), KVAR (3Φ and Total), KVA (3Φ and Total), PF (3Φ and
Total)
Bidirectional KWh, KVAh, KVARh, Frequency, THD and Demand & MD
Advance Measurement Features:
SLM102 also measures the minimum & maximum values and also allows definition
of events, so that the user can define the parameter of interest, event definition value
and related time.
Voltage min/max. -
3Φ L-N
Voltage min/max. - 3Φ L-L (applicable only for 3P3W)
Current min/max. - 3Φ.
Hz min/max.
Active Power min/max. - 3Φ & Σ system
Apparent Power min/max. - 3Φ & Σ system
Harmonic Measurement
SLM102 measures the odd harmonics for 3Φ voltages and 3Φ currents upto 29th,
including THDs for each waveform.
Relay Outputs Options
• Programmable relays for alarm or trip.
• Alarm relays are individually configurable for different parameters.
• Maximum Demand measurement in Sliding Window, or fixed window modes for
effective management of demand
Data Logging
• ON board 2 MB of Memory.
• Logs all Min. /Max. Values. (Voltage, Current, Watt, VA, System Watt, System VA,
PF and Frequency).
• Logs all electrical parameters, real-time and integrated (energies and demands)
and harmonics to get complete Energy system snap-shot.
• Events Logging: 4 Events can be programmed by the user for: Each event log
consists of parameter of interest, date and time of event occurred.
SLM102 – Operational Manual
TRINITY
[6]
Communications
• USB port for downloading of logged Data and Programming of field parameters.
• Supports MODBUS-RTU protocol on RS485
• Ethernet (TCP/IP). Also supports MODBUS-TCP/IP protocol
• Acts as a mini web-server
Display
• Easy user interface through four push buttons switches.
• Bright Red LED Display. Three rows of 8-digits each ensure unbroken display of
even large energy register values
• Two LEDs for Active and Reactive Energy (1000 impulse / KWh or KVARh)
SLM102
SLM102 – Operational Manual
TRINITY
[7]
Technical Specifications
True RMS Basic Parameters
Parameter
Mode
Type
Statistics
Symbol
3P4W
3P3W
Voltage
(Volts L-N)
Direct Voltage Input
PT Ratio
Range of Reading
Accuracy
: 0 to 300V AC
: Programmable
up to 66KV
: 0.0 to 66KV
: 0.2% of Reading
:(50 < V < 300)
Vr
√
-
Vy
√
-
Vb
√
-
Voltage
(Volts L-L)
Direct Voltage Input
PT Ratio
Range of Reading
Accuracy
: 0 to 450V AC
: Programmable
up to 66KV
: 0.0 to 66KV
: 0.5% of Reading
:(85 < V < 545)
Vry
√
√
Vbr
√
-
Vby
√
√
Current
(Amps
Ir, Iy, Ib)
Secondary Current
Input
CT Ratio
Range of Reading
Accuracy
: 1A or 5A
: Programmable
5A to 5000A
: 0.00 to 5000A
: 0.2% of Reading
(0.02 Ib< I < 1.2 Ib)
Ir
√
√
Iy
√
-
Ib
√
√
Line Frequency
45 to 55 Hz,
Accuracy
: 0.5% of Reading
Hz
√
√
Power
Parameter
Mode
Type
Statistics
Symbol
3P4W
3P3W
Active
Power (P)
Accuracy
: See Table-1
( 0.5 < COSǾ < 0.8 )
(P) R-Phase
√
√
(P) Y-Phase
√
-
(P) B-Phase
√
√
Reactive
Power (Q)
Accuracy
: See Table-1
( 0.5 < SINǾ < 0.8 )
(Q) R-Phase
√
√
(Q) Y-Phase
√
-
(Q) B-Phase
√
√
Apparent
Power (S)
Accuracy
: 0.3% of Reading
(0.01 Ib ≤ I < Imax)
(S) R-Phase
√
-
(S) Y-Phase
√
-
(S) B-Phase
√
-
System
Active
Power (P)
Accuracy
: See Table-1
( 0.5 < COSǾ < 0.8 )
∑-P
√
√
System
Reactive
Power (Q)
Accuracy
: See Table-1
( 0.5 < SINǾ < 0.8 )
∑-Q
√
√
Apparent
Power (S)
Accuracy
: 0.3% of Reading
(0.01 Ib ≤ I < Imax)
∑-S
√
√
Power
Factor
0.5 % for 0.5 < |PF| < 0.8,
(0.05 Ib < I < 1.2 Ib)
PF R-Phase
√
-
PF Y-Phase
√
-
PF B-Phase
√
-
System PF
√
√
SLM102 – Operational Manual
TRINITY
[8]
Table 1- Percentage Error Limits for Power
Value of Current
Power Factor
Percentage Error Limits
0.01 Ib ≤ I < 0.05 Ib
1.0
±0.4
0.05 Ib ≤ I ≤ Imax
1.0
±0.2
0.02 Ib ≤ I < 0.1 Ib
0.5 Lagging
±0.5
0.8 Leading
±0.5
0.1 Ib ≤ I ≤ Imax
0.5 Lagging
±0.3
0.8 Leading
±0.3
Energy
Parameter
Mode
Type
Statistics
3P4W
3P3W
Total Active
Energy (KWh)
Range of reading
Accuracy
: 0 to 9999999.9 MWh
: 0.2S as per IS14697:1999.
√
√
Total Reactive Energy
(KVARh)
Range of Reading
Accuracy
: 0 to 9999999.9 MVARh
: 0.2S as per IS 14697:1999.
√
√
Total Apparent Energy
(KVAh)
Range of Reading
: 0 to 9999999.9 MVAh
√
√
Demand
Parameter
Mode
Statistics
Symbol
3P4W
3P3W
Active Power
(KW) Demand : Sliding & Fixed Selectable
KWDMND
√
√
Apparent Power
(KVA) Demand : Sliding & Fixed Selectable
KVADMND
√
√
Power Quality
Parameter
Mode
3P4W
3P3W
Harmonics for Voltages(3rd to 29th )
√
√
Harmonics for Currents(3rd to 29th )
√
√
THD for each voltage
√
√
THD for each current
√
√
SLM102 – Operational Manual
TRINITY
[9]
Minimum-Maximum
Parameter
Mode
Type
Statistics
Symbol
3P4W
3P3W
VOLTS
Min-Max
Volts
(Volts L-N)
(Volts L-L)
Minimum Value of Volt
Vr
√
-
Vy
√
-
Vb
√
-
Vry
-
√
Vbr
-
-
Vby
-
√
CURRENT
Min-Max
Current
(Amps
Ir, Iy, Ib)
Minimum Current Value
Ir
√
√
Iy
√
-
Ib
√
√
HZ
Min.–Max.
Line Frequency
Minimum Line Frequency
Hz
√
√
WATT
Min.-Max.
Active Power
(P)
Minimum Active Power
(P)
R-Phase
√
√
(P)
Y-Phase
√
-
(P)
B-Phase
√
√
VA
Min.-Max.
Apparent Power (S)
Minimum Apparent Power
(S)
R-Phase
√
-
(S)
Y-Phase
√
-
(S)
B-Phase
√
-
SYSTEM
WATT
Min.-Max.
System Active
Power
(P)
Minimum System Active Power
∑-P
√
√
SYSTE
M
VA
Min.-Max.
System Apparent Power
(S)
Minimum System Apparent Power
∑-Q
√
√
DEMAND
Max. Demand
Maximum Demand Value.
KW
or
KVA
√
√
Predictive Demand
Fix-30 minutes Window
KW
or
KVA
√
√
SLM102 – Operational Manual
TRINITY
[10]
Input Specifications
Supply
:
Three Phases and Neutral of a 3P4W
system / three phases of a 3P3W system
Voltage
Direct Voltage Input
:
Up to 480V L-L and 300V L-N
PT Ratio
:
Programmable up to 66KV
Site selectable.
Range of Reading
:
1-66KV
Burden
:
0.5VA
Current
Secondary Current
Input
:
5A or 1A
Range of Reading
:
5A – 5000A
Burden
:
< 1.0VA
Overload
:
5A CT = 6A RMS Continuous
1A CT = 5A RMS Continuous
Power Supply:
Self Power. Unit has in-built 3 phase supply with an operating range of 50VAC -
480 VAC, 50-60 Hz.
Output Specifications
Programmable Relay Output
Relay outputs
:
4.
Relay rating
:
3A @ 230 VAC, Resistive load.
SLM102 – Operational Manual
TRINITY
[11]
Installation and Commissioning
The SLM102 supports two installation modes – 3P4W and 3P3W. The installation
processes of these modes are described below:
3P4W Installation Mode
Follow these steps to install / commission the unit.
1. Push the unit into the panel and mount using the clamps provided. The unit is
not required auxiliary supply as unit has in-built 3 phases supply with an
operating range of 50VAC - 480VAC, 50-60 Hz.
2. Connect the three phases with the phase sequence being R-Y-B to the
terminals marked U1, U2 and U3 respectively. Make sure that the three
phases coming to the unit come through control fuses of 1.0 Amp rating. This
will protect the electronics inside from damage due to severe over voltages or
phase faults in the system.
3. Connect the neutral to the terminal marked U4.
4. Connect the two wires from the R-phase CT to terminals marked M1 & L1
such that S1 from CT goes to M1 on the unit. Connect the two wires from the
Y-phase CT to terminals marked M2 & L2 such that S1 from CT goes to M2
on the unit. Connect the two wires from the B-phase CT to terminals marked
M3 & L3 such that S1 from CT goes to M3 on the unit.
5. Switch on the three phase supply as well as Auxiliary supply. The unit will
come alive and display power information such as factory location for about
three seconds and then, it will display the first page of Run Mode.
6. If there is very little current or no current in the CT circuits, the unit may show
PFs as 0.999 or 1.000. This will go away as soon as the current builds up in
the CTs, above 0.4 % of rated CT.
SLM102 – Operational Manual
TRINITY
[12]
7. Now, the unit needs to be programmed for the various parameters which are
field programmable. For this refer to next section “OPERATIONAL DETAILS”
8. The unit is ready for operation.
SLM102 – Operational Manual
TRINITY
[13]
3P3W Installation Mode
Follow these steps to install / commission the unit.
1. Push the unit into the panel and mount using the clamps provided. The unit is
not required auxiliary supply as it has in-built 3 phase supply with an operating
range of 50VAC - 480 VAC, 50-60 Hz.
2. Connect the three phases with the R phase to terminal U1, B phase to
terminal U3 and Y phase to terminal U4. Make sure that the three phases
coming to the unit come through control fuses of 1.0 amp rating. This will
protect the electronics inside from damage due to severe overvoltages or
phase faults in the system.
3. Connect the two wires from the R-phase CT to terminals marked M1 & L1
such that S1 from CT goes to M1 on the unit. Connect the two wires from the
B-phase CT to terminals marked M3 & L3 such that S1 from CT goes to M3
on the unit.
4. Switch on the three phase supply. The unit will come alive and display power
receiving information such as factory location for about three seconds and
then, it will display the first page of Run Mode.
5. If there is very little current or no current in the CT circuits, the unit may show
PFs as 0.999 or 1.000. This will go away as soon as the current builds up in
the CTs, above 0.4 % of rated CT.
6. Now, the unit needs to be programmed for the various parameters which are
field programmable. For this refer to next section “OPERATIONAL DETAILS”.
7. The unit is ready for operation.
SLM102 – Operational Manual
TRINITY
[16]
Operational Details
The load manager SLM102 is a versatile meter, with all the features needed to
implement with a robust electrical load management system. It can be configured to
suit most control and communication needs.
The unit is achieved by making as many parameters field programmable, as much
as possible. There are basically three Modes of operation:
1. Programming Mode
2. Programming Mode using USB Utility
3. Run Mode
For Programming Mode using USB Utility, refer to last section, ”Data Logging and
USB Port”.
After supplying power (50 VAC - 480 VAC) to the three phases, the unit displays
immediately power receiving information such as Factory name, RTC and then,
enters into first page of Run Mode with the following display.
Now, user will enter into Programming Mode to define for all the desire parameters
and can also operate the Run Mode by pressing keys provided with the following:
SLM102 – Operational Manual
TRINITY
[17]
Programming Mode
The unit is designed with a default password protection,”1947”. The password is also
changeable from 0 to 9 according to user’s desire (see in the next section,
programmable parameter, “Change Password”). The user will therefore enter a
password before entering into Programming Mode. Now, press
key for about 4
seconds and then, the password display will prompt with a four digits of zeroes of
which first zero will be blinking.
To enter the password, press
key one time to set for 1; press
key to shift into
second digit; Press
again nine times to set for 9, press
key again to shift into
3rd digit and press
key four times to set for 4 and then, press
key to shift into
4th digits and press
key seven times to set for 7. If any digit entered is wrong and
needs to enter again, press
key several times till the digit becomes bilking and
set as above steps. After entering the default password, press
key to enter into
Programming Mode (the display is shown in the next section).
Hence, if the unit is in Programming Mode and keys are not operated for about 4
minutes, unit will return into Run Mode.
Setting CT-Primary and date according to RTC
CT-Primary
The CT Primary should be set to give actual current values for CT operated meter
and is also freely programmable from 5A to 5000A.
Date
The date is RTC and can also be set with the format of date, month and year. i.e.,
DD/MM/YYYY format.
To set CT Primary and Date, proceed the following instructions.
SLM102 – Operational Manual
TRINITY
[18]
1. After entering the password as before, the unit will prompt with the following
display.
2. In the above first row, CT-Primary displays along with character ‘C’ blinking
that shows the parameter is selected for setting. Press
key that will start
blinking with the first digit of a value. Set CT-Primary by pressing
and
keys and press
key to confirm the setting so that the blinking character will
also return.
3. Press
key that starts blinking DATE(dd) in second row which shows
DATE is selectable. Press
key that blinks digit (i.e., 10) in third row. Set
DATE by pressing
and
keys as per current date. Press
key to
confirm the setting so that the blinking character will also return.
4. Press
key to set MONTH that will start blinking with character (MM) in
second row. press
key that will start blinking with digit in third row. Set
MONTH by pressing
and
keys. Press
key to confirm the setting
so that the blinking character will also return.
5. If the setting is completed press
key for about four seconds to return into
Run Mode. Otherwise, press
key to enter into next programmable
parameter.
Setting PT Ratio and Time according to RTC
PT Ratio
The PT-Ratio should be set so as to give actual voltage for PT-operated meter. The
PT Ratio is selectable from any one of the following: 1, 30, 60, 100, 200, 300 and
600.
Time
The Time is RTC and can also be set with the format of hh.mm.ss.
SLM102 – Operational Manual
TRINITY
[19]
To set PT Ratio and Time, proceed the following instructions.
1. In previous programmable parameters, press
key till the next display
enters such as shown below.
2. Set PT Ratio and TIME such as previous steps of CT Primary and DATE
setting.
Setting Log Duration, Demand Window and Unit Address
Log Duration
For data logging, the Log Duration can be specified from 1 to 60 minutes. SLM102
has the capability to log data on its own Non-volatile RAM. This data is logged in an
FIFO buffer. Once the buffer (23271 records) is full, the oldest data is discarded and
the newest data is filled into the buffer. This data can then be downloaded by using a
software utility provided by Trinity, into a PC, record by record.
Demand Window
The unit supports two types of demand window for the calculation of demand such
as Fixed window or Sliding window. In case of Sliding Window, the method is
selectable to either 30 minutes interval or 15 minutes interval. In case of Fixed
Window, the method is 30 minutes interval only and hence, Predictive Demand will
also display in Run Mode.
Unit Address
The unit address is for RS485 communication port only. For communicating in
between the unit and host, the same unit address should be set from 1 to 255
accordingly.
To set the Log Duration, Demand Window and Address, proceed the following
instructions.
1. In the previous programmable parameter, press
key till the next display
enters such as shown below
SLM102 – Operational Manual
TRINITY
[20]
2. In the above first row, Log Duration displays along with character ‘L’ blinking
that shows the parameter is selected for setting. Press
key that will start
blinking with first digit. Set Log Duration by pressing
and
keys and
press
key to confirm the setting so that the blinking character will also
return.
3. Press
key to set for seond row, Dimand Window that Window will start
blinking with its first character, ‘w’ and now set as above steps of Log duration.
Similalry set Unit Address also.
4. If the setting is completed press
key for about four seconds to return into
Run Mode. Otherwise, press
key to enter into next programmable
parameter.
Selecting Demand On, Installation type and CT-Secondary
Demand On
Both the two KW Demand and KVA Demand can be set according to user’s desire
for the calculation of Demand to either KVA or KW with an integration period of 30 or
15 minutes.
Installation Types
Unit supports two types of Electrical Installation such as 3P4W and 3P3W. User
should select the installation type accordingly.
CT-Secondary
The CT Secondary can be selected to either 1 or 5 and the user should select the
CT Secondary to give actual current value for CT operated meter.
To set the above parameters, proceed the following instructions
SLM102 – Operational Manual
TRINITY
[21]
1. In the previous programmable parameter display, press
key till the next
display enters such as shown below
2. In the above first row, DEMAND (DMD) displays along with character ‘D’
blinking that shows the parameter is selected for setting. Press
key that
will start blinking to either KW (W) or KVA(VA) . Set DEMAND by pressing
and
keys and press
key to confirm the setting so that the blinking
character will also return. While selecting the Demand to KVA or KW, LED
indicator will glow such as above solid circled parameters respectively.
3. Press
key to set for second row, INSTALLATION (INS) that the character
(i) will start blinking and set as steps above. Similalry set Unit Address also.
4. If the setting is completed press
key for about four seconds to return into
Run Mode. Otherwise, press
key to enter into next programmable
parameter.
Setting Four Alarm Relays, Alarm On, Alarm Value and Alarm
Hysteresis
Alarm Relay
The unit has four relay contacts, which are individually field programmable from 1 to
4. However, the Delay for each Alarm relay can be programmed from 5 to 120
seconds using USB Logger utility only. In case of Demand, the Delay is fixed with 2
seconds. This Delay programming is applicable only for USB utility. Refer last
section, “Programming Mode Using USB Logger Utility”. If user is not desired USB
programming utility, Alarm Delay will be 5 seconds by default only.
SLM102 – Operational Manual
TRINITY
[22]
Alarm On
There are seven alarm parameters: Avg. Volts(VOLT), Avg. Amps(AMPS), KVA(VA),
KW(VATT), KVAR(VAR), Demand(DMD), PF(PF) and None(NONE). These parameters
are programmable for each alarm relay according to user’s requirement. In case, any
Alarm parameter is not desired, NONE should be selected.
Alarm Value
Alarm values are freely programmable from 1 to 9999. In case of VOLT, the value is
freely programmable from 1 to 99999. The value indicates the level at which the
relay will close. E.g. If the Alarm On has been selected to KW, and the value is set to
450, the relay will close when the KW exceeds 450 and remains so for 1 minute.
Also, the switching ON logic for PF is reversed. For all parameters, the relay closed
when the parameter value exceeds that defined by Value for more than one minute.
In case of PF however, the relay will close when the PF falls below the Value and
remains so for more than one minute. Also, the PF Value is entered with a
multiplication factor of 1000. e.g. If Second Relay is programmed for which ALARM
ON is set to PF and ALARM VALUE is 955, then the second relay should close when
the PF value falls below 0.955.
Alarm Hysteresis
The last parameter called Hysteresis defines the opening point for the relay which
has closed in response to an alarm condition. This is specified as % of the Value.
Thus, setting Hysteresis of 2 implies that the relay which closes when the parameter
reaches Value, will open when the parameter falls below 0.98 x Value. This is done
to prevent frequent operation / nuisance operation of the relay near the Value.
To set the Alarm Parameters, proceed the following instructions.
1. In the previous programmable parameter display, press
key till the next
display enters such as shown below.
SLM102 – Operational Manual
TRINITY
[23]
2. In the above first row, Relay (RL) displays along with character ‘R’ blinking that
shows the Relay is selected for setting. Press
key that will start blinking
digit (i.e., 2). Set Relay by pressing
and
keys and press
key to
confirm the setting so that the blinking character will also return.
3. Press
key to set for Alarm Parameter that the character ‘d’ will start
blinking and press
key and set as steps above. While selecting the Alarm
parameters, LED indicator will glow such as above solid circled parameters
respectively.
4. Press
key to set for second row that blinks ‘V'. Press
key that the digit
start blinking. Set the Alarm Value such as steps above. Similalry set
Hysteresis also.
Setting Four Events, Event On, Event Value and Event Hysteresis
As previous Relays, unit has also four events which are individually field
programmable from 1 to 4. There are 16 Logs that can be stored for each event. So
that there are 64 logs for all four events. Each log stores event type and date-time of
event occurrence.
Event On
Same as “Alarm On”.
Event Value
Same as “Alarm Value”.
Event Hysteresis
Same as “Alarm Hysteresis”.
1. In the previous programmable parameter display, press
key till the next
display enters such as shown below.
SLM102 – Operational Manual
TRINITY
[24]
2. Set all the Event parameters such as same steps of Relays before .
Selecting Baud Rate for RS485 Port and Changing Password
Baud Rate
In case of RS485 connection, the baud rate is settable to either 9600 or 19200 for the
speed of downloading. User can download live data to a host using software utility.
Password
The numerical password used can be changed according to user’s desire from 0 to
9. In case, the PASSWORD is not desired, set zero only for all four digits. In the
PASSWORD prompt, just press
key only to enter into Programming Mode!
To set Baud Rate and password change, proceed the following instruction
1. In the previous programmable parameter display, press
key till the next
display enters such as shown below.
2. In the above first row, Baud Rate displays along with character ‘B’ blinking
that shows the parameter is selected. Press
key that will start blinking digit
(i.e., 9). Set Baud Rate by pressing
and
keys and press
key to
confirm the setting so that the blinking character will also return.
3. Press
key to enter into next row, PASSWORD that character ‘P’ will also
blink. Now, press
key that will start blinking first digit ‘0’ and set by
pressing
key only and then, press
key to shift to next digit and set by
pressing
key only. Similay, Set for rest digits also. If any set digit is
supposed to change again, press
key several times to return onto the
desired digit.
Setting IP Address
SLM102 – Operational Manual
TRINITY
[25]
The IP Address consists of four bytes. This IP address is necessary when PC is
connected to the LAN or Internet. IP Address is a location of a given computer for
network devices. Each byte can be set from 000 to 255.
1. In the previous programmable parameter display, press
key till the next
display enters such as shown below.
2. In the above first row, IP displays along with character ‘ip’ blinking that
shows the parameter is selected. Press
key that first byte starts blinking.
Set first byte by pressing
key only and press
key to shift to next IP
address. Set such as same steps for third IP address and fourth IP address
also.
3. If the setting is completed press
key for about four seconds to return into
Run Mode. Otherwise, press
key to enter into next programmable
parameter.
Setting MAC Address
User can enter 6 bytes of MAC address. In a local area network (LAN) or other
network, the MAC (Media Access Control) address is computer's unique hardware
number (on an Ethernet LAN, it's the same as Ethernet address). The MAC address
is used by the Media Access Control sub layer of the Data-Link Layer of
telecommunication protocols. Each byte can be set from 000 to 255.
1. In the previous programmable parameter display, press
key till the next
display enters such as shown below.
SLM102 – Operational Manual
TRINITY
[26]
2. In the above first row, MAC Address displays along with character ‘m’ blinking
that shows the parameter is selected. Press
key. The characters, MAC
ADDRESS shown above will be replaced by first byte and second byte such
as diplay below.
3. In the first row, first digit of the fist byte will start blinking. Now set by pressing
key only and press
key to shift to second byte. Set second byte by
pressing
key only and press
key to shift onto third byte in second row
and set it also such as same steps. Similarly, set fourth, fifth and sixth byte
also. After setting all six bytes, press
key to confirm that will return such as
first display.
4. If the setting is completed press
key for about four seconds to return into
Run Mode. Otherwise, press
key to enter into next programmable
parameter.
SLM102 – Operational Manual
TRINITY
[27]
Resetting Demand Manually / Daily/Monthly
Reset Demand can be specified such as whether the demand will be reset
automatically or manually. In case of automatic reset, there are two options: auto day
(auto dd) or auto month (auto mm). If user selects auto day then MD will reset for
every 24 hrs at 12:00 a.m. and if user selects auto month then MD will reset on 1st of
every month at 7:00 a.m.
To Reset Demand, proceed the following instructions.
1. In the previous programmable parameter display, press
key till the next
display enters such as shown below.
2. In the above first row, Demand Reset displays along with character ‘d’
blinking that shows the parameter is selected. Press
key that reset type,
e.g Auto Daily ‘A’ starts blinking. Select the reset types by pressing
and
key and then, press
key to confirm the setting.
3. If the setting is completed, press
key for about four seconds to return into
Run Mode. Otherwise, press
key to enter into next programmable
parameter.
SLM102 – Operational Manual
TRINITY
[28]
Resetting Energy
The three energies-KWh, KVAh and KVARh can be reset to zeroes according to
user’s requirement.
To reset Energies, proceed the following instructions
1. In the previous programmable parameter display, press
key till the next
display enters such as shown below.
2. In the above first row, Reset Energy displays along with character ‘R’ blinking
that shows the parameter is selected. Press
key that will blink to a
parameter in second row. Select the Energy by pressing
and
key
where LED indicator will glow onto KWh, KVAh and KVARh respectively.
Now, press
key to confirm the setting. Simultaneously, the unit will also
reset and return into Run Mode.
Resetting Minimum and Maximum Parameters
All the minimum and maximum parameter values can be reset.
To Reset Minimum and Maximum, proceed the following instructions.
1. In the previous programmable parameter, Reset Energy display, Select the
Minimum(min) by pressing
and
key such as same steps of Reset
Energy where LED indicator will glow onto KW(WATT), HZ, KVA(VA),
KVAR(VAR), VOLT, current(AMPS), Minimum(MIN) and Maximum(Max) such
as shown below.
SLM102 – Operational Manual
TRINITY
[29]
2. Now, press
key to confirm the setting. Simultaneously, the unit will also
restart and return into Run Mode.
Resetting KW Demand and KVA Demand
Both KW Demand and KVA Demand can be reset.
To reset both demand, proceed the following instructions.
1. In the previous programmable parameter, Reset Energy display, Select the
Demand(DEMAND) by pressing
and
key such as same steps of Reset
Energy where LED indicator will glow onto Demand (DMND) such as shown
below.
2. Now, press
key to confirm the setting. Simultaneously, the unit will also
restart and return into Run Mode.
SLM102 – Operational Manual
TRINITY
[30]
Resetting Log
The logged data can be reset.
To reset Log, proceed the following instructions.
1. In the previous programmable parameter, Reset Energy display, Select the
Log(LOG) by pressing
and
key such as same steps of Reset Energy
such as shown below.
2. Now, press
key to confirm the setting. Simultaneously, the unit will also
restart and return into Run Mode.
Resetting Events Logging
All the four Events under which the desired parameters have been logged can be
reset.
To reset Events Logging, proceed the following instructions.
1. In the previous programmable parameter, Reset Energy display, Select the
Demand(DEMAND) by pressing
and
key such as same steps of Reset
Energy where LED indicator will glow onto Demand (DMND) such as shown
below.
SLM102 – Operational Manual
TRINITY
[31]
2. Now, press
key to confirm the setting. Simultaneously, the unit will also
restart and return into Run Mode.
Resetting All Parameters
All the parameters can be reset.
To reset all parameter, proceed the following instructions.
1. In the previous programmable parameter, Reset Energy display, Select the All
(ALL) by pressing
and
key such as same steps of Reset Energy with
the following display.
2. Now, press
key to confirm the setting. Simultaneously, the unit will also
restart and return into Run Mode.
SLM102 – Operational Manual
TRINITY
[32]
Run Mode
In the Run Mode, the various parameters measured and calculated by the meter are
displayed on different pages; on a “seven segment LED” display. The pages will
automatically scroll one by one at every 4 to 5 seconds. Press enter key to freeze the
page. Freeze LED indicator will glow. In freeze condition, user can change the page
using up or down key.
The measured or calculated values displayed will auto scale to Mega from Kilo and to
Kilo from value, once the value crosses 10,000. e.g., Once Wh ≥ 10000.0, it will show
as 10.000 KWh and if value crosses 10000000 then it shows 1.000000 MWh. If the
user has selected a PT-Ratio (PT-Primary/PT-Secondary)>1.0, voltages will auto
scale to KV.
SLM102 shows measured and calculated values on following different pages:
Run Mode Displays
The first page shows phase to neutral voltage in three phases. The LED indicator
glows onto VOLT and PHASE which indicates that R-Phase, Y-Phase and B-Phase
display in first row, second row and third row respectively.
In case of 3P3W, the page will show phase to phase Voltage with R-phase in first
row and B-phase in third row.
SLM102 – Operational Manual
TRINITY
[33]
The second page shows individual current in three phases. The LED indicator glows
onto AMPS and PHASE which indicates that R-Phase, Y-Phase and B-Phase
display in first row, second row and third row respectively.
The third page shows Active Power (KW) in three phases. The LED indicator glows
onto WATT and PHASE which indicates that R-Phase, Y-Phase and B-Phase
display in first row, second row and third row respectively.
SLM102 – Operational Manual
TRINITY
[34]
The fourth page shows Apparent Power (KVA) in three phases. The LED indicator
glows onto VA and PHASE which indicates that R-Phase, Y-Phase and B-Phase
display in first row, second row and third row respectively. This page is not
applicable for 3P3W.
The fifth page shows Reactive Power (KVAR) in three phases. The LED indicator
glows onto VAR and PHASE which indicates that R-Phase, Y-Phase and B-Phase
display in first row, second row and third row respectively.
SLM102 – Operational Manual
TRINITY
[35]
The sixth page shows PF in three phases. The LED indicator glows onto PF and
PHASE which indicates that R-Phase, Y-Phase and B-Phase display in first row,
second row and third row respectively. This page is not applicable for 3P3W.
The seventh page shows system powers. The LED indicator glows onto WATT, VA
and VAR which indicates that Active Power (KW), Apparent Power (KVA) and
Reactive Power (KVAR) display in first row, second row and third row respectively.
SLM102 – Operational Manual
TRINITY
[36]
The eight page shows system PF with lagging PF and Frequency (Hz). The LED
indicator glows onto PF and HZ which indicates that PF and Frequency display in
first row and second row respectively.
The ninth page shows three Energies. The LED indicator glows onto Wh, VAh and
VARh which indicates that Active Energy (KWh), Apparent Energy (KVAh) and
Reactive Energy (KVARh) display in first row, second row and third row respectively.
The tenth page shows individual Voltage in three phases. The LED indicator glows
onto VOLT which indicates that R-Phase, Y-Phase and B-Phase display in first row,
second row and third row respectively. This page is not applicable for 3P3W.
SLM102 – Operational Manual
TRINITY
[37]
The eleventh page shows THD Voltage in three phases. The LED indicator glows
onto VOLT, THD and PHASE which indicate that THD voltage in R-Phase, Y-Phase
and B-Phase display in first row, second row and third row respectively.
The twelfth page shows THD current in three phases. The LED indicator glows onto
THD, AMPS and PHASE which indicate that THD current in R-Phase, Y-Phase and
B-Phase display in first row, second row and third row respectively.
The thirteenth page shows minimum voltage in three phases. The LED indicator
SLM102 – Operational Manual
TRINITY
[38]
glows onto VOLT, MIN and PHASE which indicate that minimum voltage in R-Phase,
Y-Phase and B-Phase display in first row, second row and third row respectively.
The fourteenth page shows minimum current in three phases. The LED indicator
glows onto AMPS, MIN and PHASE which indicate that minimum current in R-
Phase, Y-Phase and B-Phase display in first row, second row and third row
respectively.
The fifteenth page shows minimum Active Power in three phases. The LED indicator
glows onto WATT, PHASE and MIN which indicate that minimum WATT in R-Phase,
Y-Phase and B-Phase display in first row, second row and third row respectively.
SLM102 – Operational Manual
TRINITY
[39]
The sixteenth page shows minimum Apparent Power (KVA) in three phases. The
LED indicator glows onto VA, PHASE and MIN which indicate that minimum
Apparent Power in R-Phase, Y-Phase and B-Phase display in first row, second row
and third row respectively.
The seventeenth page shows minimum system of Active Power, Frequency (Hz) and
Apparent Power. The LED indicator glows onto Active Power (WATT), Frequency
(HZ) and Apparent Power (VA) which indicate that Active Power, Frequency and
Apparent Power display in first row, second row and third row respectively.
SLM102 – Operational Manual
TRINITY
[40]
The eighteenth page shows Maximum Voltage in three phases. The LED indicator
glows onto VOLT, MAX and PHASE which indicate that maximum voltage in R-
Phase, Y-Phase and B-Phase display in first row, second row and third row
respectively.
The nineteenth page shows minimum current in three phases. The LED indicator
glows onto AMPS, MIN and PHASE which indicate that minimum Current in R-
Phase, Y-Phase and B-Phase display in first row, second row and third row
respectively.
SLM102 – Operational Manual
TRINITY
[41]
The twentieth page shows maximum current in three phases. The LED indicator
glows onto AMPS, MAX and PHASE which indicate that maximum current in R-
Phase, Y-Phase and B-Phase display in first row, second row and third row
respectively.
The twenty-first page shows maximum Active Power in three phases. The LED
indicator glows onto WATT, PHASE and MAX which indicate that maximum Active
Power in R-Phase, Y-Phase and B-Phase display in first row, second row and third
row respectively.
SLM102 – Operational Manual
TRINITY
[42]
The twenty-second page shows maximum Apparent Power in three phases. The
LED indicator glows onto Apparent Power (VA), PHASE and MIN which indicate that
maximum Apparent Power in R-Phase, Y-Phase and B-Phase display in first row,
second row and third row respectively.
The twenty-third page shows maximum Active Power, Frequency, Apparent Power.
The LED indicator glows onto Active Power (WATT), HZ, VA and MAX which
indicate that maximum Active Power, Frequency and Apparent Power display in first
row, second row and third row respectively.
SLM102 – Operational Manual
TRINITY
[43]
The twenty-fourth page shows Sliding Active Power Demand for 15 minutes. In case
user defined demand is Apparent Power, the display will show “DMD-VA" in second
row. The LED indicator glows onto WAT and DMND which indicate that Demand
window is Active Power.
The twenty-fifth page shows maximum Active Power Demand occurrence with date
and time. The LED indicator glows onto WATT, DMND and MAX for Active Power
Demand. However, LED indicator will glow onto VA, DMND and MAX if user defined
demand is Apparent Power.
In fact, the demand is registered with following:
In case of fixed window selection, the demand is integrated over 30 minutes.
In case of sliding window, it is integrated over 15 or 30 minutes, as selected
by the user.
If the demand is calculated by the meter, based on sliding window technique,
then unlike utility installed demand meters, this demand figure does not reset
to zero at the end of the 30 minute integration window.
If the user has initially selected KVA as demand parameter, the figure in the
SLM102 – Operational Manual
TRINITY
[44]
Maximum Demand register is that of KVA demand. If later, the user changes
the parameter to KW, the Maximum Demand register does not change. It
continues to show the highest Demand reached to date, which was in KVA.
This is because there is only one Maximum Demand register.
The twenty-sixth page shows last maximum Demand occurrence and status of the
logging buffer. The demand displayed is the last maximum demand. If present
maximum demand is 400 KVA and reset it and then, maximum Demand will become
zero and Last maximum Demand will hold as 400 KVA.
The twenty-seventh page shows Predictive Demand. The LED indicator glows onto
VA and DMND which shows Predictive Demand is for VA. If user selectable
parameter is Watt (W), the LED indicator will glow onto WATT and DMND.
SLM102 – Operational Manual
TRINITY
[45]
The twenty-eighth page shows alarm relays programmed for desired parameters. It
also display the relays switch on with the futures 1, 2 and 3 whereas relay 4 is
switched off.
The twenty-ninth page shows four events logging under which event 1 stores 10
logs; event 2 stores 12 logs; event 3 stores 11 logs; and event 4 stores 16 logs. This
event logging can be seen to the USB data downloading. The maximum Log is 16 for
each event. Once the Log is overflow, the event will start logging from 1 and
OVERFLOW will show YES.
SLM102 – Operational Manual
TRINITY
[46]
The thirtieth page shows number of logs with RTC date and time. The maximum log
for all the parameters is 23271 records.
SLM102 – Operational Manual
TRINITY
[47]
Communications
In addition to the industry standard RS-485 communication port, several other
communication options are also available in SLM102. These options make it possible
for a user to select SLM102 to provide power and energy information into a variety of
existing or new control systems and communication networks. Each communication
option supports data transfer with external devices or applications. Metering
measurement, logged data, configuration and status data may be accessed via
communication.
SLM102 supports USB 2.0 for downloading of logged data, MODBUS-RTU on RS-
485 port and an Ethernet port supporting TCP/IP protocol, which can be used as a
small web-server and also in MODBUS-TCP/IP mode to connect to a SCADA
application.
MODBUS RTU on RS485 Port
The SLM102 support an isolated RS485 port, for half duplex multi-drop
communication using MODBUS-RTU protocol. This makes it easy to integrate
SLM102 to any EMS/SCADA application.
The station id for every meter is site selectable, and so is the baud rate. The data
which can be read using MODBUS query # 3 (Read Holding Registers) is provided in
an address map, with the applicable multiplication factors, vide Appendix in last
section.
Communication line parameters: 9600 or 19200/8/N/1.
SLM102 – Operational Manual
TRINITY
[48]
The register map is described in Appendix. All addresses are in decimal whose
parameters are unsigned long. If illegal address is sent in query or the host, tries to
read more than 64 bytes of data in one query exception message is generated.
Reserved values are for future use, which are transmitted as zeroes.
Ethernet TCP/IP Communication
Ethernet-TCP/IP connection, Multiple sockets Ethernet connectivity enables the unit
to respond to as many as 12 different requests. The SLM102 can act as a small
web-server. This on-board web-sever, offers quick and easy access to basic
measurement, without special software. Built-in web pages display a range of power,
energy and basic power quality information through any web enabled device. It also
supports MODBUS-TCP/IP protocol.
To access the Ethernet TCP/ IP communication, connect the RJ45 Ethernet cable
into the port provided at the bottom of the unit and enter your protocol in the address
bar and then, the web page will open such as shown below.
SLM102 connected as Mini Web Server to a web browser.
User can browse four pages such as Home, Individual parameters, Minimum and
Maximum parameters and system values.
SLM102 – Operational Manual
TRINITY
[49]
Data Logging and USB Port
SLM102 has a 2 MB non-volatile memory. This memory is used to store the minimum
and maximum values of important parameters, events as defined by the user and the
snapshot of the entire electrical system where all instantaneous and integrated
parameters along with Harmonic data are saved. Maximum numbers of such
electrical snapshots are 23271.
While the electrical snapshots as saved in the logs can give a good idea of the overall
behavior of the system, the Event Log feature captures time-stamped records of
important power system occurrences that can be aligned with corresponding
production effects to better understand and optimize energy use and cost.
The log maintains a time-stamped record of 26 parameters. One can monitor values
over a day, a week, a month or any period to record the highest and lowest values of
voltage, current, power factor including Event parameters.
SLM102 supports USB 2.0 for downloading logged data. Users are provided a CD for
USB software installation along with the purchase of SLM102. Before installing the
software, users are advised to read carefully Read Me document in the Logger disc
for a list of minimum system requirements to be installed, if uninstalled before. If your
system does not meet the minimum system requirements, installation of SLM102
software will not be possible.
Installing the USB Software
1. Insert the Logger setup CD into the CD-ROM drive of your computer after
windows is started.
2. Double click on the Logger and select Logger Setup or Setup and then,
double click on any of the setup to install. A wizard will welcome for Logger
Setup. To proceed the installation, click Next and then, the wizard will show
the following dialog box for the location.
SLM102 – Operational Manual
TRINITY
[50]
3. Before selecting the location, user can also check the disk space for each drive
clicking Disk Cost. The Logger Setup has also two options such as Everyone
and Just me of which Everyone is selected by default. User can select Just me
accordingly.
4. To confirm the installation, click Next in the next wizard and wait for some time
while proceeding the installation. After completing the installation, the wizard will
ask the Logger Setup has been installed successfully. Now click Close to exit the
wizard.
Installation of USB driver
Make sure that the USB cable is connected in between the PC and unit. If the USB
connection is not all right, remove the USB cable and connect again. If the USB is
connected for the first time, the wizard will show with following option.
SLM102 – Operational Manual
TRINITY
[51]
Select the option, Yes, This time only and click Next to proceed for the next
wizard such as shown below.
The wizard will ask if your hardware comes with an installation CD or floppy
disk, insert it now. Select the wizard option, Installation from a list or
SLM102 – Operational Manual
TRINITY
[52]
specific location and click Next to give the following dialog box where the
file is located.
In the above dialog box, select option Search for the best drive in these
location and click check box of Include this location in the search to
specify path of usb drivers and then, click Next to get the path where the file
is located.
SLM102 – Operational Manual
TRINITY
[53]
5. Now click OK, to complete the wizard.
6. Wait for sometime while completing the wizard and then the wizard will
prompt the new hardware gets completed. Now click Finish to complete
the installation.
Handling the Logger Window
The Logger window consists of the following: Downing logged data, harmonic data,
Graph representation onto Logger window and Exporting the Logged data in to a
excel sheet.
Downloading the Logged data
1. By default the utility will show Logger 1.0 on desktop. Double-click on the
Logger icon. Now the Logger will open with the following window.
SLM102 – Operational Manual
TRINITY
[54]
2. In the unit type list box, select the SLM102 and click OK and DOWNLOAD to
start downloading the logged data with the following window.
3. In this LOG DATA window, click HARMONICS DATA at the bottom of the
window to open the harmonics data window for all current and voltage.
Exporting the Logged Data including Min/Max Events and Harmonics
Data
The logged data can also be exported into excel sheet so that the Min-Max &event
logs including the Harmonics can also be obtained in it such as shown below.
SLM102 – Operational Manual
TRINITY
[55]
Graph Representation
The graph plots for various selected parameters of Logged data as well as system
data including harmonics distortion including date and time. Hence, the back ground
color, Axis colors and Label colors can also be selectable.
Plotting a Graph for Logged data.
The logged data can be plotted to a graph with the selected parameters, date and
time.
To plot a graph, proceed the following instructions,
1. At the top of the Logger Windows, click Graph to show a graph window.
2. At the bottom of the of the graph window, enter starting date and time in the
Start box, enter ending date and time in End box under Date and Time. There
are also two type of Scale such as Auto scale and Manual scale. In case of
Auto scale, it is already selected by default. In case of Manual scale, select
Manual scale.
3. To select for the parameters to be plotted, double-click in check box which will
display the parameter selected at right list such as shown below.
SLM102 – Operational Manual
TRINITY
[56]
4. If the parameters selected are not desired, double-click again in the check box
to remove for a selected parameter from the list.
5. To change the color of the selected parameters displayed at right side (e.g. Ir,
Iy, Ib and KW) double-click on it, and then the color palette will show such as
below.
6. Select the desired color and click Ok. Now, the selected parameter will plot
with the selected color. Similarly, each selected parameter can also be plotted.
Before plotting the parameters, the background color, axis color and label color
should be specified by clicking
,
and
so
as to represent along with the selected colors.
7. Now, click
to plot a graph for the selected parameters such as shown
below.
SLM102 – Operational Manual
TRINITY
[57]
8. The plotted graphs can be saved and printed by clicking
and
respectively. Hence, to return into previous Logger Windows, click
.
Programming Mode Using USB Utility
All the users’ programmable parameters in Programming Mode of unit can also be
programmed remotely using Logger Utility. Using utility, user can easily specify for all
the programmable parameters which is also more convenient than the programming
in unit. Make sure USB cable coming from the PC is connected to SLM102 where
the USB port is provided at the top of the unit.
To configure the Programming Mode using Logger utility, proceed the following
instructions
By default the utility will show Logger icon on desktop. Double-click on the Logger
icon. Now the Logger will open with the following window.
SLM102 – Operational Manual
TRINITY
[58]
Click down arrow and select SLM102(Prog) in list box that will open the page such
as shown below.
By default the cursor will blink in password box which shows user will enter
password and click
to enter into above Programming Mode. After configuring
each programmable parameter, click
respectively to confirm the settings. If
user selelcts Installation type and CT Ratio, the unit may restart. User will wait till the
unit enters into Run Mode and then user will again enter password. In case, a
message displays “Error In USB Communication” click OK and try again so that all
the programmable parameter can be configured.
SLM102 – Operational Manual
TRINITY
[59]
Control Outputs
The relay contacts provided in SLM102 are rated for 3A @ 230VAC. They are also
protected by snubbers against fast voltage transients which occur when inductive
loads are switched off. Thus, the following points are to be taken care of when using
these relay contacts:
1. Use 230VAC coils only in the contactors. DO NOT use 440V AC coils.
2. DO NOT switch small loads like electronic Hooters, small relays with 230VAC
coils etc. directly from the relay contact of SLM102. If done so, the small
leakage current from the snubbers will not allow these loads to be switched off
fully. Thus, the electronic hooters will give a low hum continuously, and the
small relays will switch on but not switch off. Use these relay contacts to
switch an Auxiliary contactor and put the load on the contactor contacts.
Alarm Action
Considering two alarm relays with previous section, ”Setting Four Alarm Relays,
Alarm On, Alarm Value and Alarm Hysteresis” in Programming Mode, user can
configure for a desired parameter for alarm control action such as shown below.
Relay 1
Relay 2
SLM102 – Operational Manual
TRINITY
[60]
All relays, from 1 to 4 are programmable with the following parameters, Avg. Volts,
Avg. Amps, KVA, KW, KVAR, Demand and PF. The value indicates the level at
which the relay will close. E.g. If the Alarm On has been selected to KW, and the
value is set to 450, the relay will close when KW exceeds 450 and remains so for 1
minute. All parameters have this response time of 1 minute, except Demand which
has a response time of 2 seconds. This is because Demand is already an integrated
parameter.
Also, the switching ON logic in case of PF is reversed. For all parameters, the relay
closed when the parameter value exceeds that defined by Value for more than one
minute. In case of PF however, the relay will close when the PF falls below the Value
and remains so for more than one minute. Also, the PF Value is entered with a
multiplication factor of 1000. e.g. when the second alarm is set for PF and alarm
Value is set to 955, the relay should close when the PF value falls below 0.955
The last parameter called Hysteresis defines the opening point for the relay which
has closed in response to an alarm condition. This is specified as % of the Value.
Thus, setting Hysteresis of 2 implies that the relay which closes when the parameter
reaches Value, will open when the parameter falls below 0.98 x Value. This is done
to prevent frequent operation / nuisance operation of the relay near the Value. E.g. If
the Value has been set to 400, and Hysteresis to 5%, the relay will close when the
parameter reaches 400 and then, the relay opens when it falls below 380. The
response times as outlined above for closing of relays also applies to opening of
relays. Thus, the relay will open at 380 only if the parameter falls below 380 and
remains so for 1 minute (2 seconds in case of Demand).
If all four relays are programmed to close on Demand and different values for closing
and opening are set by the user, SLM102 can be used effectively as a Maximum
Demand Controller with four relay contacts!
SLM102 as a Demand Management
Each relay can be programmed separately to generate alarm on a specific
parameter with its specific value. The value at which it should open is also
programmable. Thus, user could have the first relay programmed to close when the
avg. Voltage exceeds certain value, second when the Avg. Amps exceeds some
level, third for KW, or KVAR or PF or Demand and so on. Before programming the
relays, having a look into Run Mode for Predictive Demand value is true for Demand
Management.
This allows the user to deploy SLM102 as a true Demand manager, to generate
alarm signals for four different conditions!
SLM102 – Operational Manual
TRINITY
[61]
Appendix
Address map of RS485 Port for various parameters of both 3P4W
and 3P3W.
RS‐485 MAP FOR SLM102
If PF is Lead Side then PF and Var Value
are Negative.
1000
VOLT‐R
4
FLOAT
1002
AMPS‐R
4
FLOAT
1004
WATT‐R
4
FLOAT
1006
VA‐R
4
FLOAT
1008
VAR‐R
4
FLOAT
1010
PF‐R
4
FLOAT
1012
3rd Har. Volt‐R
4
FLOAT
1014
5th Har. Volt‐R
4
FLOAT
1016
7th Har. Volt‐R
4
FLOAT
1018
9th Har. Volt‐R
4
FLOAT
1020
11th Har. Volt‐R
4
FLOAT
1022
13th Har. Volt‐R
4
FLOAT
1024
15th Har. Volt‐R
4
FLOAT
1026
17rd Har. Volt‐R
4
FLOAT
1028
19th Har. Volt‐R
4
FLOAT
1030
21th Har. Volt‐R
4
FLOAT
1032
23rd Har. Volt‐R
4
FLOAT
1034
25th Har. Volt‐R
4
FLOAT
1036
27th Har. Volt‐R
4
FLOAT
1038
29th Har. Volt‐R
4
FLOAT
1040
VOLT ‐ THD‐R
4
FLOAT
1042
3rd Har. Amps‐R
4
FLOAT
1044
5th Har. Amps‐R
4
FLOAT
1046
7th Har. Amps‐R
4
FLOAT
1048
9th Har. Amps‐R
4
FLOAT
1050
11th Har. Amps‐R
4
FLOAT
1052
13th Har. Amps‐R
4
FLOAT
1054
15th Har. Amps‐R
4
FLOAT
1056
17th Har. Amps‐R
4
FLOAT
1058
19th Har. Amps‐R
4
FLOAT
1060
21th Har. Amps‐R
4
FLOAT
1062
23rd Har. Amps‐R
4
FLOAT
1064
25th Har. Amps‐R
4
FLOAT
1066
27th Har. Amps‐R
4
FLOAT
1068
29th Har. Amps‐R
4
FLOAT
1070
AMPS ‐THD‐R
4
FLOAT
1072
VRY
4
FLOAT
148
148
1200
VOLT‐Y
4
FLOAT
1202
AMPS‐Y
4
FLOAT
1204
WATT‐Y
4
FLOAT
1206
VA‐Y
4
FLOAT
1208
VAR‐Y
4
FLOAT
1210
PF‐Y
4
FLOAT
1212
3rd Har. Volt‐Y
4
FLOAT
1214
5th Har. Volt‐Y
4
FLOAT
SLM102 – Operational Manual
TRINITY
[62]
1216
7th Har. Volt‐Y
4
FLOAT
1218
9th Har. Volt‐Y
4
FLOAT
1220
11th Har. Volt‐Y
4
FLOAT
1222
13th Har. Volt‐Y
4
FLOAT
1224
15th Har. Volt‐Y
4
FLOAT
1226
17rd Har. Volt‐Y
4
FLOAT
1228
19th Har. Volt‐Y
4
FLOAT
1230
21th Har. Volt‐Y
4
FLOAT
1232
23rd Har. Volt‐Y
4
FLOAT
1234
25th Har. Volt‐Y
4
FLOAT
1236
27th Har. Volt‐Y
4
FLOAT
1238
29th Har. Volt‐Y
4
FLOAT
1240
VOLT ‐ THD‐Y
4
FLOAT
1242
3rd Har. Amps‐Y
4
FLOAT
1244
5th Har. Amps‐Y
4
FLOAT
1246
7th Har. Amps‐Y
4
FLOAT
1248
9th Har. Amps‐Y
4
FLOAT
1250
11th Har. Amps‐Y
4
FLOAT
1252
13th Har. Amps‐Y
4
FLOAT
1254
15th Har. Amps‐Y
4
FLOAT
1256
17th Har. Amps‐Y
4
FLOAT
1258
19th Har. Amps‐Y
4
FLOAT
1260
21th Har. Amps‐Y
4
FLOAT
1262
23rd Har. Amps‐Y
4
FLOAT
1264
25th Har. Amps‐Y
4
FLOAT
1266
27th Har. Amps‐Y
4
FLOAT
1268
29th Har. Amps‐Y
4
FLOAT
1270
AMPS ‐THD‐Y
4
FLOAT
1272
VYB
4
FLOAT
148
148
1400
VOLT‐B
4
FLOAT
1402
AMPS‐B
4
FLOAT
1404
WATT‐B
4
FLOAT
1406
VA‐B
4
FLOAT
1408
VAR‐B
4
FLOAT
1410
PF‐B
4
FLOAT
1412
3rd Har. Volt‐B
4
FLOAT
1414
5th Har. Volt‐B
4
FLOAT
1416
7th Har. Volt‐B
4
FLOAT
1418
9th Har. Volt‐B
4
FLOAT
1420
11th Har. Volt‐B
4
FLOAT
1422
13th Har. Volt‐B
4
FLOAT
1424
15th Har. Volt‐B
4
FLOAT
1426
17rd Har. Volt‐B
4
FLOAT
1428
19th Har. Volt‐B
4
FLOAT
1430
21th Har. Volt‐B
4
FLOAT
1432
23rd Har. Volt‐B
4
FLOAT
1434
25th Har. Volt‐B
4
FLOAT
1436
27th Har. Volt‐B
4
FLOAT
1438
29th Har. Volt‐B
4
FLOAT
1440
VOLT ‐ THD‐B
4
FLOAT
1442
3rd Har. Amps‐B
4
FLOAT
1444
5th Har. Amps‐B
4
FLOAT
1446
7th Har. Amps‐B
4
FLOAT
1448
9th Har. Amps‐B
4
FLOAT
1450
11th Har. Amps‐B
4
FLOAT
1452
13th Har. Amps‐B
4
FLOAT
1454
15th Har. Amps‐B
4
FLOAT
SLM102 – Operational Manual
TRINITY
[63]
1456
17th Har. Amps‐B
4
FLOAT
1458
19th Har. Amps‐B
4
FLOAT
1460
21th Har. Amps‐B
4
FLOAT
1462
23rd Har. Amps‐B
4
FLOAT
1464
25th Har. Amps‐B
4
FLOAT
1466
27th Har. Amps‐B
4
FLOAT
1468
29th Har. Amps‐B
4
FLOAT
1470
AMPS ‐THD‐B
4
FLOAT
1472
VBR
4
FLOAT
148
148
2000
SYS W
4
FLOAT
For Address 2010 and
2011 Alarm status, there
are four bytes. First Three
Bytes are zero. Fourth
Byte shows Relay
position. First Bit of this
byte shows First relay
status. If it is 1 then it
shows First relay is on.
Same as second, third and
fourth bit is used for
second, third and fourth
relay.
For Address 2014 and
2015. Unit is single Byte
Parameter. First byte
shows Unit value and
other three are Dummy
Bytes. Unit shows the unit
of wh, Varh and Vah. If
Unit =2 means value of
wh, Vah and Varh are in
Kilo. If Unit = 3 means
value of wh, vah and varh
are in Mega. If Unit = 1
means wh, vah and varh
are not in kilo or mega
range.
2002
SYS VA
4
FLOAT
2004
SYS VAR
4
FLOAT
2006
FREQUENCY
4
FLOAT
2008
SYS PF
4
FLOAT
2010
ALARM STATUS
4
CHAR * 4
2012
DEMAND
4
FLOAT
2014
Unit & DUMMY
4
CHAR*4
2016
SYS WH
4
FLOAT
2018
SYS VAH
4
FLOAT
2020
SYS VARH
4
FLOAT
2022
MIN SYS W
4
FLOAT
2024
SECOND & MINUTE
2
CHAR*2
2025
HOUR & DUMMY
2
CHAR*2
2026
DATE & MONTH
2
CHAR*2
2027
YEAR & DUMMY
2
0
2028
MIN SYS VA
4
FLOAT
2030
SECOND & MINUTE
2
CHAR*2
2031
HOUR & DUMMY
2
CHAR*2
2032
DATE & MONTH
2
CHAR*2
2033
YEAR & DUMMY
2
0
2034
MIN FREQUENCY
4
FLOAT
2036
SECOND & MINUTE
2
CHAR*2
2037
HOUR & DUMMY
2
CHAR*2
2038
DATE & MONTH
2
CHAR*2
2039
YEAR & DUMMY
2
0
2040
MAX SYS W
4
FLOAT
2042
SECOND & MINUTE
2
CHAR*2
2043
HOUR & DUMMY
2
CHAR*2
2044
DATE & MONTH
2
CHAR*2
2045
YEAR & DUMMY
2
0
2046
MAX SYS VA
4
FLOAT
2048
SECOND & MINUTE
2
CHAR*2
2049
HOUR & DUMMY
2
CHAR*2
2050
DATE & MONTH
2
CHAR*2
2051
YEAR & DUMMY
2
0
2052
MAX FREQ
4
FLOAT
2054
SECOND & MINUTE
2
CHAR*2
2055
HOUR & DUMMY
2
CHAR*2
2056
DATE & MONTH
2
CHAR*2
2057
YEAR & DUMMY
2
0
2058
MAX DEMAND
4
FLOAT
2060
SECOND & MINUTE
2
CHAR*2
SLM102 – Operational Manual
TRINITY
[64]
2061
HOUR & DUMMY
2
CHAR*2
2062
DATE & MONTH
2
CHAR*2
2063
YEAR & DUMMY
2
0
132
2064
PRD DEMAND
4
FLOAT
132
3000 MIN VOLT‐R
4
FLOAT
3002
SECOND & MINUTE
2
CHAR*2
3003
HOUR & DUMMY
2
CHAR*2
3004
DATE & MONTH
2
CHAR*2
3005
YEAR & DUMMY
2
0
3006 MIN AMPS‐R
4
FLOAT
3008
SECOND & MINUTE
2
CHAR*2
3009
HOUR & DUMMY
2
CHAR*2
3010
DATE & MONTH
2
CHAR*2
3011
YEAR & DUMMY
2
0
3012 MIN WATT‐R
4
FLOAT
3014
SECOND & MINUTE
2
CHAR*2
3015
HOUR & DUMMY
2
CHAR*2
3016
DATE & MONTH
2
CHAR*2
3017
YEAR & DUMMY
2
0
3018 MIN VA‐R
4
FLOAT
3020
SECOND & MINUTE
2
CHAR*2
3021
HOUR & DUMMY
2
CHAR*2
3022
DATE & MONTH
2
CHAR*2
3023
YEAR & DUMMY
2
0
3024 MAX VOLT‐R
4
FLOAT
3026
SECOND & MINUTE
2
CHAR*2
3027
HOUR & DUMMY
2
CHAR*2
3028
DATE & MONTH
2
CHAR*2
3029
YEAR & DUMMY
2
0
3030 MAX AMPS‐R
4
FLOAT
3032
SECOND & MINUTE
2
CHAR*2
3033
HOUR & DUMMY
2
CHAR*2
3034
DATE & MONTH
2
CHAR*2
3035
YEAR & DUMMY
2
0
3036 MAX WATT‐R
4
FLOAT
3038
SECOND & MINUTE
2
CHAR*2
3039
HOUR & DUMMY
2
CHAR*2
3040
DATE & MONTH
2
CHAR*2
3041
YEAR & DUMMY
2
0
3042 MAX VA‐R
4
FLOAT
3044
SECOND & MINUTE
2
CHAR*2
3045
HOUR & DUMMY
2
CHAR*2
3046
DATE & MONTH
2
CHAR*2
3047
YEAR & DUMMY
2
0
96
SLM102 – Operational Manual
TRINITY
[65]
96
3300 MIN VOLT‐Y
4
FLOAT
3302
SECOND & MINUTE
2
CHAR*2
3303
HOUR & DUMMY
2
CHAR*2
3304
DATE & MONTH
2
CHAR*2
3305
YEAR & DUMMY
2
0
3306 MIN AMPS‐Y
4
FLOAT
3308
SECOND & MINUTE
2
CHAR*2
3309
HOUR & DUMMY
2
CHAR*2
3310
DATE & MONTH
2
CHAR*2
3311
YEAR & DUMMY
2
0
3312 MIN WATT‐Y
4
FLOAT
3314
SECOND & MINUTE
2
CHAR*2
3315
HOUR & DUMMY
2
CHAR*2
3316
DATE & MONTH
2
CHAR*2
3317
YEAR & DUMMY
2
0
3318 MIN VA‐Y
4
FLOAT
3320
SECOND & MINUTE
2
CHAR*2
3321
HOUR & DUMMY
2
CHAR*2
3322
DATE & MONTH
2
CHAR*2
3323
YEAR & DUMMY
2
0
3324 MAX VOLT‐Y
4
FLOAT
3326
SECOND & MINUTE
2
CHAR*2
3327
HOUR & DUMMY
2
CHAR*2
3328
DATE & MONTH
2
CHAR*2
3329
YEAR & DUMMY
2
0
3330 MAX AMPS‐Y
4
FLOAT
3332
SECOND & MINUTE
2
CHAR*2
3333
HOUR & DUMMY
2
CHAR*2
3334
DATE & MONTH
2
CHAR*2
3335
YEAR & DUMMY
2
0
3336 MAX WATT‐Y
4
FLOAT
3338
SECOND & MINUTE
2
CHAR*2
3339
HOUR & DUMMY
2
CHAR*2
3340
DATE & MONTH
2
CHAR*2
3341
YEAR & DUMMY
2
0
3342 MAX VA‐Y
4
FLOAT
3344
SECOND & MINUTE
2
CHAR*2
3345
HOUR & DUMMY
2
CHAR*2
3346
DATE & MONTH
2
CHAR*2
3347
YEAR & DUMMY
2
0
96
96
3600 MIN VOLT‐B
4
FLOAT
3602
SECOND & MINUTE
2
CHAR*2
SLM102 – Operational Manual
TRINITY
[66]
3603
HOUR & DUMMY
2
CHAR*2
3604
DATE & MONTH
2
CHAR*2
3605
YEAR & DUMMY
2
0
3606 MIN AMPS‐B
4
FLOAT
3608
SECOND & MINUTE
2
CHAR*2
3609
HOUR & DUMMY
2
CHAR*2
3610
DATE & MONTH
2
CHAR*2
3611
YEAR & DUMMY
2
0
3612 MIN WATT‐B
4
FLOAT
3614
SECOND & MINUTE
2
CHAR*2
3615
HOUR & DUMMY
2
CHAR*2
3616
DATE & MONTH
2
CHAR*2
3617
YEAR & DUMMY
2
0
3618 MIN VA‐B
4
FLOAT
3620
SECOND & MINUTE
2
CHAR*2
3621
HOUR & DUMMY
2
CHAR*2
3622
DATE & MONTH
2
CHAR*2
3623
YEAR & DUMMY
2
0
3624 MAX VOLT‐B
4
FLOAT
3626
SECOND & MINUTE
2
CHAR*2
3627
HOUR & DUMMY
2
CHAR*2
3628
DATE & MONTH
2
CHAR*2
3629
YEAR & DUMMY
2
0
3630 MAX AMPS‐B
4
FLOAT
3632
SECOND & MINUTE
2
CHAR*2
3633
HOUR & DUMMY
2
CHAR*2
3634
DATE & MONTH
2
CHAR*2
3635
YEAR & DUMMY
2
0
3636 MAX WATT‐B
4
FLOAT
3638
SECOND & MINUTE
2
CHAR*2
3639
HOUR & DUMMY
2
CHAR*2
3640
DATE & MONTH
2
CHAR*2
3641
YEAR & DUMMY
2
0
3642 MAX VA‐B
4
FLOAT
3644
SECOND & MINUTE
2
CHAR*2
3645
HOUR & DUMMY
2
CHAR*2
3646
DATE & MONTH
2
CHAR*2
3647
YEAR & DUMMY
2
0
96
SLM102 – Operational Manual
TRINITY
[67]
P.O No.
: …………………………………………………………………
Customer
: …………………………………………………………………
Sr. No.
: …………………………………………………………………
Routine
and
function
tests
conducted
to
relevant
standards
and
our
Specifications/Literature/O & M Manual.
Traceability: tested against "MTE" Standard Model PRS400.3 having basic accuracy
of 0.02% traceable upto International Standards derived using appropriate ratio
techniques.
Result of Test
: …………………………………………………………………
Remarks
: …………………………………………………………………
Test engineer
: …………………………………………………………………
Date
: …………………………………………………………………