1445/JRB
OPERATOR’S MANUAL
MODEL 156A
CHARGED-PLATE MONITOR
上海汇分电子科技有限公司
上海市闸北区天目西路511号锦程大厦1408室
联系方式:400-660-9565
网站:www.19mro.com
Thank you for buying a Trek instrument. This instrument has been designed and
built to high standards to give you years of trouble-free service.
If you have any questions, please feel free to contact your Trek Representative at:
Write:
TREK, INC.
190 Walnut Street
Lockport, NY 14094-3710
Fax:
(716) 201-1804
Telephone:
In the United States:
1 800 FOR-TREK
International Customers:
1 (716) 438-7555
Internet:
www.trekinc.com
E-Mail:
[email protected]
We welcome any comments or suggestions you may have relative to the operation,
performance, and/or quality of this product.
Trek products are covered by United States and foreign patents, issued and pending.
Information in this publication supersedes that in all previously published material.
Specifications are subject to change without notice.
Copyright © 2014 TREK, INC. All rights reserved. Printed in the U.S.A.
Table of Contents
Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
Safety Precautions for the Model 156A . . . . . . . . . . . . . . . . . . . . . v
Preface. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Section I Introduction
General Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I -1
Modes of Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I -2
Incoming Inspection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I -2
Incoming Confidence Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . I -3
Section II Installation
Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II-1
Battery Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II-1
Battery Charger/Eliminator Connection . . . . . . . . . . . . . . . . . . . . . . II-1
Initial Setup Connections . . . . . . . . . . . . . . . . . . . . . . . . . . . . . II-2
Section III Operation
Front Panel Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . III-1
Rear Panel Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . III-3
Float Mode Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . III-4
Decay Mode Operation. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . III-4
Programming Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . III-5
Programming the START voltage . . . . . . . . . . . . . . . . . . . . . . III-5
Programming the STOP voltage . . . . . . . . . . . . . . . . . . . . . . . III-6
Monitoring the Plate Voltage using the Output Monitor. . . . . . . . . . . . . . III-7
i
Section IV Specifications
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . IV-1
Section V Maintenance
Safety Precautions for the Model 156A . . . . . . . . . . . . . . . . . . . . . . V-1
Battery Pack Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . V-1
Preventative Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . V-2
Customer Service Assistance . . . . . . . . . . . . . . . . . . . . . . . . . . . . V-2
Service Information. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . V-3
Appendixes
Appendix A Accessories
Appendix B Warranty Statement
Appendix C Sales and Service
ii
Table of Contents
Review the following safety precautions to maintain safety and prevent damage to
the instrument or equipment connected to it.
The safety features of this instrument may be ineffective if the equipment is not
operated in the manner stated in this manual.
Refer all maintenance procedures to qualified personnel.
Safety Precautions
Power
Observe polarity of connections and use appropriate battery eliminator ratings for
voltage and current.
Avoid Electric Overload
To avoid electric shock or fire hazard, do not apply a voltage to a terminal that is
outside the range specified for that terminal.
Avoid Electric Shock
To avoid electric shock, do not touch the charged-plate while the instrument is on.
Do Not Operate Without Covers
To avoid electric shock or fire hazard, do not operate this instrument with the covers
removed.
Fuses
There are no user serviceable fuses inside this unit, please refer all problems to
Trek, Inc. or an authorized Trek service center.
iii
·
Safety Precautions (cont.)
Indoor Use Only
This instrument is intended for indoor use only.
Do Not Operate in Wet or Damp Conditions
To avoid electric shock, do not operate this instrument in wet or damp conditions.
Do Not Operate in an Explosive Environment
To avoid injury or fire hazard, do not operate this instrument in an explosive
environment.
Product Protection Precautions
Use the Proper Power Source
Do not operate this instrument from a power source that is different than the voltage
specified on the serial number tag.
Provide Proper Ventilation
To prevent the instrument from overheating, provide proper ventilation.
Do Not Operate with Suspected Failures
If you suspect there is damage to this instrument, have it inspected by qualified
personnel.
iv
Safety
Safety Terms and Symbols
Terms in the Manual
These terms may appear in this manual:
Warning: Warning statements identify conditions or practices that could result in
injury or loss of life.
Caution: Caution statements identify conditions or practices that could result in
damage to this product or other equipment.
Symbols on the Product
These symbols may appear on the instrument:
Warning, risk of electric shock
Caution, refer to Operator’s Manual
CAT I
Installation category I (overvoltage category): Classification for the
operation of a unit using voltage systems or circuits with required
standardized limits for transient voltages. Category I pertains to
voltages supplied at the peripheral level, with smaller tolerances for
transient voltages as specified by the Low-Voltage Safety standard
(EN 61010-1).
This symbol refers to the compliance of the equipment to the
European Council (E.C.) standards.
v
Safety
Safety Precautions for the Model 156A
Please observe the following safety precautions when checking connections or
providing maintenance procedures on the Model 156A:
1. Refer all maintenance procedures to qualified personnel.
2. Prior to performing maintenance procedures, turn off the 156A and disconnect
the battery charger/eliminator from its power source. Failure to observe this
precaution could result in an electrical shock.
3.
Allow a cool-down period to reduce the danger of burns from heated parts such
as transistors and heat sinks.
4.
This instrument incorporates a rechargeable 7.2 V battery pack. Do not
substitute a non-rechargeable battery because risk of fire or explosion could
result. Please contact TREK, Inc. or an authorized service organization for
battery replacement information.
Please go to our web site: www.trekinc.com for a complete list of our sales and
service representatives and distributors located in the United States and throughout
the world.
vi
Safety
This manual provides user information for the Model 156A Charged-Plate Monitor
and contains the following chapters and appendixes:
• Introduction contains a brief product description and describes an incoming
confidence test for the unit.
• Installation contains initial setup information for the Model 156A.
• Operation contains a description of the product’s features and a detailed
explanation of proper operating procedures.
• Specifications states the requirements, behavior, and performance of the
instrument in a concise format.
• Maintenance contains proper care and preventative maintenance guidelines
for the instrument.
• Appendix A: Accessories describes other products that are useful with
the Model 156A.
• Appendix B: Warranty Statement describes terms and conditions of the
Trek Warranty Statement.
• Appendix C: Sales & Service contains Trek contact information.
(Please go to our web site: www.trekinc.com for a complete list of our sales
and service representatives and distributors located in the United States and
throughout the world.)
vii
Danger: Trek high-voltage generating equipment, including Trek amplifiers and
supplies are not designed, rated, or qualified to be operated in an environment or
atmosphere which contains combustible or explosive materials or gases which may
be ignited by electrical discharges.
This manual provides instructions to install and operate the Trek Model 156A
Charged-Plate Monitor. We recommend you take the time to read this manual to
take full advantage of the features and benefits of the instrument.
General Information
The Trek Model 156A Charged-Plate Monitor System is designed to provide
quantitative and accurate measurements for a variety of air ionization tests in any
location. The OUTPUT MONITOR connector on the rear panel is a buffered output
providing a low-voltage replica of the voltage on the ion collecting plate.
The Model 156A employs a revolutionary new charged-plate monitor design that
utilizes an ultra-high-impedance high-voltage follower to monitor the ion collecting
plate voltage. This technique provides very high accuracy and virtually infinite
impedance loading of the plate, while allowing the ion collecting plate to be charged
and monitored through the same small-diameter connection.
This technique also makes the measurement capacitance independent of the physical
size and shape of the ion collecting plate. Therefore, the size and shape of an ion
collecting plate, as well as the measurement capacitance, may be customer specified
to match a particular ESD-sensitive device within a manufacturing process, or it may
be made to conform to the ESD Association Standard ANSI/ESD-STM3.1.
I - 1
Modes of Operation
The Model 156A features three primary modes of operation: a negative (-) decay
mode, a positive (+) decay mode, and a float mode. Decay mode START and STOP
voltages are easy to program.
For DECAY mode operation, the ion collecting plate is precharged, either positively
or negatively, to a level above the programmed START voltage. The time required
to discharge the ion collecting plate from the programmed START voltage to the
programmed STOP voltage is measured by the Model 156A. The discharge is due to
‘incident ion flow’ on the ion collecting plate.
For the FLOAT mode operation, the plate is initially discharged to zero and then
allowed to float to the voltage level dictated by ‘incident ion flow’ on the ion
collecting plate.
The simplicity of the design of the 156A makes it very reliable and completely
portable. The 156A is powered by a rechargeable battery (supplied) that offers
8 hours of operation from a full charge. The unit can also be operated from a line
supply using a battery eliminator/charger.
Caution: This instrument incorporates a rechargeable 7.2 V battery pack. Do not
substitute a non-rechargeable battery because risk of fire or explosion could result.
Please contact TREK, Inc. or an authorized service organization for battery
replacement information.
Incoming Inspection
Visually inspect the instrument for physical damage such as dents, nicks, scratches,
broken fittings, etc. External damage may indicate more serious damage has
occurred within the instrument. In the event of damage, notify the factory or your
nearest authorized Trek Service Organization and request instructions. Attempts to
operate a damaged instrument may cause permanent damage to the instrument.
I - 2
Introduction
Warning: Hazardous voltages may be present on ion plate during Decay mode
operation.
Incoming Confidence Test
The 156A undergoes extensive checks and adjustments at the factory, and no initial
calibration should be required. However, you may wish to perform an incoming
confidence test as part of the incoming inspection on the instrument. An incoming
confidence test of this nature is intended to confirm that the instrument was not
damaged in transit.
We recommend that you familiarize yourself with the information in Section II and
Section III before performing this test.
Placing a charged object near the ion collecting plate while the instrument is in the
DECAY mode and measuring the decay time of the object would constitute a
reasonable incoming confidence test. A simple source for a charged object is an
ordinary pocket comb or a hard plastic screw driver. A pocket comb drawn through
the hair will usually produce a negative charge on the comb. A hard plastic handle of
a screw driver brushed against a pant leg will usually produce a positive charge on
the handle of the screw driver.
Warning: Do not turn on the 156A until instructed to do so. To do so before the
appropriate point in time could result in an electrical shock and/or damage to the
instrument.
1. Ensure the power is OFF.
2. Connect the ion collecting plate to the ION COLLECTING PLATE connector on
the rear panel.
3.
Connect the GROUND jack to earth ground using the banana-alligator cable
provided.
4.
Place the MODE switch in the +DECAY position.
5.
Turn on the power.
6.
Momentarily press down on the TEST/RESET switch.
7.
Move a negatively charged object (pocket comb) towards the ion collecting plate
until the value indicated on the PLATE VOLTAGE display decreases below the
programmed decay start voltage.
The decay timer will start when the plate voltage passes the programmed decay start
voltage.
Note: The decay start voltage is factory programmed for 1000 V. (See
‘Programming the START voltage’ on page III-4 for information on changing the
I - 3
Introduction
直到指示这样做的156A不要打开。这样做的时间的适当位置,可能会导致电击和/或损坏仪器之前。
Incoming Confidence Test (cont.)
8.
Continue to move the negatively charged object towards the ion collecting plate
until the value indicated on the PLATE VOLTAGE display passes through the
programmed decay stop voltage.
Note: The decay stop voltage is factory programmed for 100 V. (See ‘Programming
the STOP voltage’ on page III-5 for information on changing the decay stop
voltage.)
At the programmed stop voltage the decay timer will stop and the decay time will be
indicated on the DECAY TIME display.
This completes the incoming confidence test. Turn OFF the 156A.
I - 4
Introduction
Section II Installation
Mounting
The Model 156A is designed to be operated as a portable benchtop instrument.
There are no special mounting requirements.
Battery Operation
Initially charge the 156A for 3 hours before operating the instrument. The FULL
CHARGE indicator will illuminate when the battery is fully charged.
The 156A will operate for at least 8 hours with a full charge.
The LOW BATTERY indicator will illuminate when the battery is approximately
30 minutes from total discharge.
Caution: This instrument incorporates a rechargeable 7.2 V battery pack. Do not
substitute a non-rechargeable battery because risk of fire or explosion could result.
Please contact TREK, Inc. or an authorized service organization for battery
replacement information.
Battery Charger/Eliminator Connection
The Model 156A is designed to be battery operated or operated while connected to a
line voltage source. A battery charger/eliminator is available for all nominal line
voltages.
1. Ensure the power is OFF.
2. Plug the battery charger/eliminator to the CHARGER receptacle on the rear panel.
3. Plug the free end of the AC line cord into the power source.
II - 1
Figure 2-1: Battery Charger/Eliminator connector polarity.
Initial Setup Connections
Before operating the Model 156A, the following connections should be made:
Ground Connection
The Model 156A must be connected to a solid ground reference to ensure accurate
and repeatable measurements. All voltage measurements will be made with respect to
this ground reference point. The power to the unit should be OFF when connecting
the ground on the rear panel of the unit to a ground reference.
Ion Collecting Plate Connection
Ensure that the power to the unit is OFF when connecting the plate cable to the
ION COLLECTING PLATE receptacle on the rear panel.
Note: Hazardous voltages may exist in the plate receptacle if the power to the unit is
ON. See warning below under OUTPUT MONITOR CONNECTION.
Output Monitor Connection
Warning: Exercise extreme caution while connecting a BNC connector cable to the
OUTPUT MONITOR BNC receptacle. Do not confuse the OUTPUT MONITOR
BNC cable connector for the ION PLATE BNC cable connector. Inserting the
OUTPUT MONITOR connector into the ION PLATE connector could deliver a
voltage of 1100 volts to the BNC shell.
The OUTPUT MONITOR connector on the rear panel is a buffered output providing
a low-voltage replica of the voltage on the ion collecting plate.
The voltage at this output represents 1/200th the voltage on the ion collecting plate.
Connect a monitoring device, such as an oscilloscope, to this connector to monitor
the voltage on the ion collecting plate. The signal at this connector can also be used
as a feedback signal in a closed-loop system.
II - 2
Installation
Section III Operation
Front Panel Features
Figure 3-1: Front Panel Features
1.
FULL CHARGE Indicator: This indicator will illuminate when the battery is
fully charged. At least 8 hours of operation can be expected from a fully charged
battery.
2.
MODE Switch: This switch selects either the (+) DECAY, the (-) DECAY or the
FLOAT mode of operation. This switch is also used in combination with the
TEST/RESET switch to program the start and stop voltages.
3.
LOW BATTERY Indicator: This indicator will illuminate when the battery is
4.
POWER Switch: This switch turns the power ON and OFF and may enable
Warning: The Model 156A is battery operated and high-voltage may be present
at the ion collecting plate or ion plate connector whether or not the battery
charger/eliminator is connected to the rear panel when the POWER switch is
III - 1
CHARGED PLATE MONITOR SYSTEM
VOLTS
VOLTS
MODEL 156A
PLATE VOLTAGE
PLATE VOLTAGE
SECONDS
SECONDS
DECAY TIME
DECAY TIME
DECAY
FLOAT
RESET
TEST /
MODE
DECAY
FULL
CHARGE
BATTERY
LOW
POWER
4
3
2
1
initial high-voltage to the plate.
ON.
approximately 30 minutes from being fully discharged。
Front Panel Features (cont.)
Figure 3-2: Front Panel Features (cont.)
5.
6.
DECAY TIME Display: This digital display indicates the time in seconds for the
ion collecting plate to discharge from the programmed START voltage to the
7.
TEST/RESET Switch: This switch sets the plate voltage to greater than the
programmed START voltage when the instrument is in the (+) DECAY or
(-) DECAY mode. This switch sets the plate voltage to 0 volts when the
instrument is in the FLOAT mode.
This switch is also used in combination with the MODE switch to program the
START and STOP voltages.
III - 2
Operation
CHARGED PLATE MONITOR SYSTEM
VOLTS
VOLTS
MODEL 156A
PLATE VOLTAGE
PLATE VOLTAGE
SECONDS
SECONDS
DECAY TIME
DECAY TIME
DECAY
FLOAT
RESET
TEST /
MODE
DECAY
FULL
CHARGE
BATTERY
LOW
POWER
7
6
5
PLATE VOLTAGE Display: This digital display indicates the plate voltage in
volts.
programmed STOP voltage.
Rear Panel Features
Figure 3-3: Rear Panel Features
1.
OFFSET ADJUST: This adjustment minimizes plate voltage drift. The calibration is
2.
CHARGER/SUPPLY Receptacle: This receptacle is for connection of the battery
3.
4.
OUTPUT MONITOR Connector: This BNC provides a buffered, low-voltage replica
of the voltage on the ion collecting plate. The voltage at this output is 1/200th of the
voltage on the ion collecting plate.
Warning: Exercise extreme caution while connecting a BNC connector cable to the
OUTPUT MONITOR BNC receptacle. Do not confuse the OUTPUT MONITOR
BNC cable connector for the ION PLATE BNC cable connector. Inserting the
OUTPUT MONITOR connector into the ION PLATE connector could deliver a
voltage of 1100 volts to the BNC shell.
5.
ION COLLECTING PLATE Receptacle: This BNC receptacle is for connection of
Warning: The Model 156A is battery operated and high-voltage may be present at
the ion collecting plate or ion plate connector whether or not the battery
charger/eliminator is connected to the rear panel when the POWER switch is ON.
III - 3
Operation
OFFSET
ADJUST
200 TO 1
PLATE
12VDC / 1A
GROUND
MONITOR
COLLECTING
CHARGER/SUPPLY
OUTPUT
ION
5
4
3
1
2
factory set and should not need adjustment.
eliminator/charger.
GROUND Jack: This banana jack is for connection to a ground reference point.
the ion collecting plate.
FLOAT Mode Operation
In the FLOAT mode, the 156A monitors the voltage on the plate caused by air ion
imbalance.
1.
Place the ion collecting plate in the area to be monitored.
2.
Place the MODE switch in the FLOAT mode.
3.
Press down and release the TEST/RESET switch to apply 0 volts to the plate.
Upon releasing the TEST/RESET switch, the plate will “float” to a voltage as
dictated by air ion imbalances.
DECAY Mode Operation
In the positive (+) DECAY and negative (-) DECAY modes, the 156A monitors the
time taken for the ion collecting plate to decay from the programmed START voltage
to the programmed STOP voltage.
1.
Place the ion collecting plate in the area to be monitored.
2.
Place the MODE switch in either the positive (+) DECAY or negative (-)
DECAY mode:
In the positive (+) DECAY mode the ion collecting plate is charged to a positive
voltage. In the negative (-) DECAY mode the ion collecting plate is charged to a
negative voltage.
3.
Press and release the TEST/RESET switch.
The ion collecting plate will charge to a value slightly greater than the
programmed START voltage. The decay timer will start when the plate voltage
reaches the programmed START voltage. The decay timer will stop when the
plate voltage reaches the programmed STOP voltage.
III - 4
Operation
The Model 156A can be operated in any one of three modes, the (+) DECAY mode,
the (-) DECAY mode, and the FLOAT mode.
Note: When the instrument is first turned on, it will indicate the revision level of the
program in the DECAY TIME display. The unit will also flash the (previously)
programmed START voltage in the PLATE VOLTAGE display and the (previously)
programmed STOP voltage in the DECAY TIME display. The START and STOP voltages
are initially factory set to 1000 V and 100 V, respectively.
Programming the START voltage
1.
Ensure the power is OFF.
2.
Place the MODE switch to the FLOAT position.
3.
Press and hold down the TEST/RESET switch. Keep the switch down when turning
ON the POWER switch.
4.
After the POWER switch is turned on, the letter ‘S’, representing START
voltage, appears in the PLATE VOLTAGE
display.
5. Release the TEST/RESET switch. The
DECAY TIME display will indicate the
present START voltage.
6.
Place the MODE switch in the positive (+)
DECAY position to increase the START
voltage. The START voltage will
automatically increase in 1 volt increments.
Place the MODE switch in the negative (-)
DECAY position to decrease the START
voltage. The START voltage will automatically decrease in 1 volt increments.
Select the voltage value desired for testing. When the desired START voltage
parameter is obtained, place the MODE switch in the FLOAT position.
Note: The START voltage cannot be set above 1000 V or below the programmed
STOP voltage.
III - 5
Operation
DECAY
FLOAT
RESET
TEST /
MODE
DECAY
Programming Operation (cont.)
Programming the STOP voltage
1.
After the START voltage has been programmed, press and hold down the
TEST/RESET switch. The letter “E”, representing “END” (STOP) voltage, appears
in the PLATE VOLTAGE display.
2. Release the TEST/RESET switch. The DECAY TIME display will indicate the
present STOP voltage.
3. Place the MODE switch in the positive (+) DECAY position to increase the STOP
voltage. The STOP voltage will automatically increase in 1 volt increments.
Place the MODE switch in the negative (-) DECAY position to decrease the STOP
voltage. The STOP voltage will automatically decrease in 1 volt increments. Select
the voltage parameter desired for testing. When the desired START voltage is
obtained, place the MODE switch in the FLOAT position
Note: The STOP voltage cannot be set above the programmed START voltage
or below 0 V.
4.
Press and hold down the TEST/RESET switch. The letter “P”, representing
‘PROGRAM’, appears in the PLATE VOLTAGE display.
Release the TEST/RESET switch. The 156A stores the new settings for START and
STOP voltages and resets itself to begin testing for the newly chosen parameters.
The 156A will display the ‘revision level’ of the command program and then will
flash the new START voltage in the PLATE VOLTAGE display and the new STOP
voltage in the DECAY TIME display. The unit will then go into the normal
operating mode as set by the MODE switch.
III - 6
Operation
Monitoring the Plate Voltage using the Output Monitor
The OUTPUT MONITOR receptacle provides a low-voltage replica of the plate
voltage. The OUTPUT MONITOR voltage is 1/200th of the plate voltage. The
OUTPUT MONITOR can be connected to an oscilloscope for visual representation
of the ion tests.
For example, if the plate voltage is 500 V, the voltage at the OUTPUT MONITOR
receptacle would be:
500 V x 1 V/200 V = 2.5 V.
III - 7
Operation
Section IV Model 156A Specifications
DANGER: This instrument is not rated for use in an explosive environment.
DO NOT use it an explosive environment or an explosion may result.
PERFORMANCE
Monitored Voltage Range
0 to ±1100 V DC or peak AC.
Small Signal Bandwidth (-3 dB)
DC to 1 kHz (measured at 20 V p-p).
Large Signal Bandwidth
DC to 10 Hz (measured at
2000 V p-p).
Zero Stability
(referred to plate voltage)
Drift with Time
(no incident ion flow)
Less than 6 V/minute.
Drift with Temperature
Less than 10 mV/°C, noncumulative.
Decay Mode Thresholds
Start Voltage
Programmable from 1 to ±1000 V in
1 V increments.
Start Accuracy
Within ±1 V of programmed start
voltage.
IV - 1
Decay Mode Thresholds (cont.)
Stop Voltage
Programmable from 0 to ±999 V in
1 V increments.
Stop Accuracy
Within ±1 V of programmed stop voltage.
Plate Self-Discharge
Less than 12 V/minute.
(unpolarized)
FEATURES
Power On/Off
A two-position rocker switch.
Test/Reset Control
A momentary toggle switch used in
conjunction with the Mode Select Switch
to program the START and STOP
voltages.
+Decay and -Decay Modes
Set the plate voltage to a value greater
than the programmed start voltage and
reset the decay timer to zero.
Float Mode
Sets the plate voltage to 0 V ±2 V.
Voltage Monitor Output
A BNC providing a low voltage replica of
the plate voltage.
Scale Factor
1/200th of the plate voltage.
DC Accuracy
Better than 0.1% of full scale.
Offset Voltage
Less than ±10 mV.
Output Noise
Less than 10 mV rms (measured using the
true rms feature of the Hewlett Packard
Model 34401A digital multimeter).
Output Impedance
Less than 0.1 ohm.
IV - 2
Specifications
Plate Voltage Digital Panel Meter
3 ½ digit red LED display.
Range
0 to ±1100 V.
Resolution
1 volt.
Accuracy
Better than 0.1% of full scale, ±1 count.
Decay Time Digital Panel Meter
4-digit red LED display.
Range
0 to 9999 seconds.
Resolution
0.1 seconds, from 0 to 999.9 seconds;
1 second, from 1000 to 9999 seconds.
(The display will indicate “- - - -” when
the decay time exceeds 9999 seconds.)
Ion Collector Plate
Standard Option
Meets ESD Association Standard
ANSI/ESD-STM3.1 requirements for:
Dimensions
15.3 cm x 15.3 cm square, (6² x 6²
square).
Capacitance
20 pF ±2 pF.
Non-standard Option
Dimensions
Specify when ordering.
Capacitance
Specify when ordering.
Mode Select
A three-position toggle switch that selects
the +Decay, the -Decay, or the Float mode
of operation. This switch is also used in
combination with the Test/Reset Control
switch to program the start and stop
voltages.
IV - 3
Specifications
Optionally available for all nominal
line voltages. Contract Trek, Inc. or
an authorized service organization for
more information. NOTE: If line
impulses exceed 500 V when using
the battery eliminator/charger, the unit
may reset and require operator
assistance
Specifications
Output Connector
2.1 mm DC power plug.
Output Voltage
12 to 14 V DC.
Output Current
1 A.
Battery Operation
Rechargeable battery supplied.
Operating Time
8 hours with full charge.
Recharge Time
3 hours to full charge.
Other
Low battery indicator.
Full charge indicator.
Operating Conditions
Temperature
5 °C to 35 °C.
Relative Humidity
To 80%, noncondensing.
IV - 4
Specifications
83 mm H x 318 mm W x 280 mm D
(3.25" H x 12.5" W x 11" D).
Instrument Weight
2 kg (4.4 lb).
Voltage Monitor Connector
BNC coaxial connector.
(3 meter length, maximum.)
Ground Receptacle
Banana jack.
Battery Eliminator/
Charger Receptacle
2.1 mm DC power plug.
Cable from Instrument to
Floating Plate
Coaxial type.
Diameter
4.95 mm (0.195") or
2.6 mm (0.102").
Length
3 m (10 ft), nominal.
IV - 5
Specifications
Certification and Compliance
Certification
TREK, INC. certifies that each
Model 156A is tested and calibrated to
specifications using measurement
equipment traceable to the National
Institute of Standards and Technology or
traceable to consensus standards. A
Certificate of Calibration accompanies
each instrument when it is shipped from
the factory.
Low Voltage Safety Compliance
EN 61010-1
Overvoltage Category
CAT I: Peripheral level outputs (less than
60 volts).
Pollution Category
Degree 1: Operate in environments where
no pollution or only dry, nonconductive
pollution occurs.
IV - 6
Specifications
Section V Maintenance
Safety Precautions for the Model 156A
Please observe the following safety precautions when checking connections or
providing maintenance procedures on the Model 156A:
1. Refer all maintenance procedures to qualified personnel.
2.
Prior to performing maintenance procedures, turn off the 156A and disconnect
the battery charger/eliminator from its power source. Failure to observe this
precaution could result in an electrical shock.
3.
Allow a cool-down period to reduce the danger of burns from heated parts such
as transistors and heat sinks.
4.
This instrument incorporates a rechargeable 7.2 V battery pack which is not user
serviceable. Do not substitute a non-rechargeable battery because risk of fire or
explosion could result. Please contact TREK, Inc. or an authorized service
organization for battery replacement information.
Please go to our web site: www.trekinc.com for a complete list of our sales and
service representatives and distributors located in the United States and throughout
the world.
Battery Pack Maintenance
The Model 156A is equipped with a rechargeable battery pack. Periodic cycling
(complete charge and discharge) of the battery pack is required in order to maximize
the life and capacity of the battery pack. Trek, Inc. recommends cycling the battery
pack at least once a month. Cycling of the battery pack can be accomplished by
following the procedure below.
1) Plug the battery eliminator/charger into the Model 156A.
2) Charge the battery pack until the “Full Charge” indicator on the Model 156A is
illuminated.
3) Remove the charger from the Model 156A.
4) Turn the power to the Model 156A on.
V - 1
Battery Pack Maintenance (cont.)
5) Leave the Model 156A “ON” for more than 12 hrs.
6) Plug the battery eliminator/charger into the Model 156A.
7) Charge the battery pack until the “Full Charge” indicator on the Model 156A is
illuminated.
8) Cycling is completed.
Note: An apparent loss of capacity for the battery pack (“Memory Effect”) might
be noticed if the battery pack is left on charge, for a long period of time. This
effect can be reversed by cycling the battery pack 1-3 times.
Preventative Maintenance
Preventative maintenance consists of inspecting and cleaning the instrument.
Preventative maintenance performed on a regular basis may prevent instrument
failure and improve reliability.
INSPECTION: Visually inspect the instrument for loose or damaged controls and
connectors or other undesirable conditions.
CLEANING: Clean the 156A as operating conditions require.
Clean the exterior of the instrument with a damp cloth. The use of solvents may
damage the finish or plastic components. A small brush is effective in removing dirt
from the front and rear panel controls and connectors.
Customer Service Assistance
In the event that you require assistance on a maintenance item, direct your request for
assistance to the Customer Service Group at Trek, Inc. or an authorized Trek Service
Organization. Please go to our web site: www.trekinc.com for a complete list of our
sales and service representatives and distributors located in the United States and
throughout the world.
Telephone assistance is usually effective for obtaining additional maintenance
information which is beyond the scope of this manual. Troubleshooting advice
which is given over the telephone may be useful for solving the simpler malfunctions
or confirming that the system should be returned to the factory or to an authorized
Trek Service Organization.
V - 2
Maintenance
Note: The warranty is voided if the instrument is serviced within the warranty
period by anyone other than Trek, Inc. or one of its authorized service organizations.
In the event of a malfunction, and the instrument must be returned for repair:
1.
Notify the Customer Service Group at Trek, Inc., giving full details about the
difficulty, including the model number and serial number of the instrument.
The Customer Service Group will issue a return authorization number.
2.
Forward the instrument (prepaid), with the return authorization number
prominently displayed on the shipping container and the packing list, to
Trek, Inc.
The instrument may also be returned to an authorized Trek Service Organization.
Contact the service organization nearest to you for details.
Please go to our web site: www.trekinc.com for a complete list of our sales and
service representatives and distributors located in the United States and throughout
the world.
V - 3
Maintenance
Appendix A Accessories
Item
Part Number
Operator’s Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23103
Nine (9) ft. Banana to Banana stackable cord (2.7 meter.) . . . . . . . . . N9044
OPTIONAL ACCESSORIES
Item
Part Number
Ion Collecting Plate Tripod . . . . . . . . . . . . . . . . . . . . . . . . DK142
AC Adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . F5054R
Carrying Case. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43433
ION COLLECTING PLATES
Item
Part Number
STANDARD PLATES
Model 156AP C150x150-R3M
(6²x 6², Meets ESD Association Standard ANSI/ESD-STM3.1) . . . . . 17397
Model 156P C25x25-S3M (1² x 1²) . . . . . . . . . . . . . . . . . . . . 17375
SPECIAL PLATES
Contact TREK, INC. or an authorized sales organization
for applications that require a different size or shape ion collecting plate.
Accessories - 1
Appendix B Warranty Statement
New instruments sold by TREK, INC. (hereinafter called the “Company”) are warranted only
as stated below:
Subject to the exceptions and upon the conditions specified below, the Company agrees to
correct, either by repair, or in the Company’s sole discretion, by replacement, any defect of
material or workmanship which develops within one year from the date of original purchase
by the customer (user), provided that investigation and factory inspection by the Company
discloses that such defect developed under normal and proper use. Repair or replacement are
the exclusive remedies under this warranty (batteries are not included under this warranty).
The exceptions and conditions mentioned are as follows:
(a) The Company makes no warranty concerning components or accessories not
manufactured by itself.
(b) The Company is released from all obligations under this warranty in the event that repairs
or modifications are made by persons other than its own or authorized personnel, unless such
repairs by others are made with the prior written consent of the Company. In the event of a
failure, and the customer neglects to take prompt and reasonable actions to prevent further
damage, the Company cannot be responsible for consequent damage.
(c) THERE ARE NO OTHER WARRANTIES WHICH EXTEND BEYOND THOSE
EXPRESSLY PROVIDED FOR HEREIN AND THE AFORESAID WARRANTY AND
THE COMPANY’S OBLIGATIONS AND LIABILITIES THEREUNDER ARE IN LIEU
OF, AND CUSTOMER WAIVES ALL OTHER WARRANTIES AND GUARANTEES
AND ALL OTHER LIABILITIES THEREFORE, EXPRESS OR IMPLIED, ARISING BY
LAW OR OTHERWISE, INCLUDING WITHOUT LIMITATION ANY IMPLIED
WARRANTY OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE,
AND ALL OBLIGATIONS AND LIABILITIES WITH RESPECT TO LOSS OF USE,
REVENUE OR PROFIT, OR INDIRECT, CONSEQUENTIAL OR INCIDENTAL
DAMAGES OF ANY KIND AND FROM ANY CAUSE WHATSOEVER ARISING OUT
OF OR IN ANY WAY CONNECTED WITH THE MANUFACTURE, SALE, HANDLING,
REPAIR, MAINTENANCE OR REPLACEMENT OF SAID PRODUCTS.
(d) Representation and warranties made by any person, including dealers and representatives
of the Company, which are inconsistent or in conflict with the terms of this warranty
(including, but not limited to, the limitations of the liability of the Company as set forth
above), shall not be binding upon the Company unless reduced to writing and approved by an
officer of the Company.
(e) This warranty shall be governed by the laws of the State of New York.
Warranty - 3
Appendix C Sales & Service
Please go to our web site: www.trekinc.com for a complete list of our sales and
service representatives and distributors located in the United States and throughout
the world.
Trek Contact Information:
TREK, INC.
190 Walnut Street
Lockport, NY 14094-3710 USA
Fax:
(716) 201-1804
Telephone:
1 800 FOR-TREK
1 (716) 438-7555
Internet:
www.trekinc.com
E-Mail:
[email protected]
Trek Japan KK
Motobayashi Building
1-2-12 Nihonbashi-ningyocho, Chuo-ku
Tokyo, 103-0013, Japan
Fax:
81 3 6264 8693
Telephone:
81 3 6264 8692
Internet:
www.trekj.com
E-Mail:
[email protected]
Sales - 1
上海汇分电子科技有限公司
上海市闸北区天目西路511号锦程大厦1408室
联系方式:400-660-9565
网站:www.19mro.com