Tenma 72-7910 Manual 

Specifications

Brand: Tenma
Category: Test Equipment
Model: 72-7910
File Type: PDF
File Size: —
Pages: 57

Read Online

Read the Tenma 72-7910 Manual  online using the built-in PDF viewer.

Text of the manual

7. Set the Utility System   ....................................................................................... 46

8. Set the Measurement Parameters ..................................................................... 51 9. Set the Cursor Measurement ............................................................................. 52 10. System Message Examples and Troubleshooting ............................................... 53 Ⅶ. Maintenance & Cleaning   ..................................................................................... 55

72‐7610/72‐10510  oscilloscopes  are  designed  and  manufactured  under  the  safety

requirements  of  IEC61010‐1  safety  standards,  and  conforms  to  insulation  and overvoltage standards, CATII (150V or 300V depending on probe setting) and pollution degree II safety standards. Please follow precautions below in order to avoid personal injuries, product damages, or damages to any products connected with the instrument. To avoid possible danger, please operate according to directions below.  Only professionals can execute the maintenance procedure.  Prevent fire and personal injuries.  Use the dedicated power lines recognized by your country.  Do not plug or unplug when the probe or test lead has been connected to voltage source.  Ground product reliably to avoid electrical shock. The products are grounded via power earth cord. Prior to connecting the input or output terminals of the products, make sure the products are grounded correctly.  Correctly connect the oscilloscope probes. The probe earth cord and the ground are of the same potential. Please do not connect the cord with high voltage source.  Check the rated values of all  terminals. To avoid fire and electrical shock,  please check all rated values and remarks on products; before connecting the products, check rated values in the manual.  Do not open the front panel or back case during operation.  Use proper fuses. Only use the specified fuses of the same specifications.  Avoid circuit exposure. Please do not contact with exposed plugs or elements after powering on.  Stop operating when any product is suspected to be out of order and can ask the qualified technician to inspect.  Keep proper ventilation.  Please do not operate in the humid environment.  Please do not operate in inflammable and explosive environments.  Please keep product surface clean and dry. Symbols on products: such symbols may appear on products. High voltage Please check the manual Protective ground terminal

Please  order  all  the  accessories  (standard  and  optional  accessories)  from  the  local

distributors. Ⅲ. Product Overview These digital storage oscilloscope provides the users with simple front panel to operate various functions. Scale and position knobs of each channel provide visual operations that are in line with users’ habits. Users may be able to master it without spending plenty of time Attention: attention statements refer to conditions and actions that may cause damages to the product or other properties.

Apart  from  the  easy  operation,  the  oscilloscope  also  has  high  performance  index  and

powerful functions required for faster measurements. Faster signals can be observed with the  oscilloscope  via  500MS/s  (or  1GS/s)  real‐time  sampling  and  25GS/s  (or  50GS/s) equivalent  sampling.  Powerful  trigger  and  analysis  ability  make  it  easier  to  capture  and analyze waveforms. Clear LCD and mathematical operating functions make it easier for users to observe and analyze signal problems in a faster and clearer way. From the following parameter features, you can understand how the oscilloscope can satisfy your measurement requirements.  Two analog channels  High resolution LCD display, 320×240 (or 400×240) resolution  Supports plug‐and‐play USB storage equipment for computer communication  Automatic waveform and state settings  Waveform saving and replay feature  Delicate  window  extension  function  and  precise  analysis  on  waveform  details  and overview  Automatic measurement of 28 waveform parameters  Automatic cursor tracking measurement function  Unique functions of waveform recording and play‐back  Built‐in FFT software function  Multi‐waveform  mathematical  operation  function  (including:  addition,  subtraction, multiplication and division)  Edge, video, pulse width, alternating trigger and other functions  Multi‐language menu selection  Simplified Chinese and English help information display Ⅳ. Functions The oscilloscope is disk type. It is designed with the conventional user interface for any basic digital storage oscilloscope in the test and measurement industry. This section will cover the following as the beginning guide of the oscilloscope: 1. Get to Know the User Interface 2. General Inspections 3. Functional Inspections 4. Probe Compensation 5. Automatic Settings of Waveform Display

Model 72‐7610/72‐10510 User Manual 4 6. Introduction to the Vertical System 7. Introduction to the Horizontal System 8. Introduction to the Trigger System 1. Get to Know the User Interface Getting to understand different parts of the user interface is one of the most important tasks to know the operations. The oscilloscope user interface consists of:  Vertical control  Horizontal control  Trigger control  LCD display area  Function buttons (F1, F2, F3, F4, F5)  User menu area  Input terminal (CH1, CH2, EXT TRIG)  Multi‐purpose knob  USB interface Please get familiar with the user interfaces shown in Figure 1‐1 and Figure 1‐2. Figure 1‐1: 72‐7610 Panel Diagram ‐ 5.7‐inch TFT color LCD display

In  case  of  badly  damages  to  packing  box  or  foamed  plastic  mat,  you  should  replace

immediately. 2. Accessories inspection

Users  can  power  on  the  oscilloscope  with  voltage  supply  of  AC  100V  –  AC  240V  and

frequency of 45Hz – 440Hz. After powering on, in order to activate the best working status, press UTILITY menu after waiting 30 minutes, then press F1 for self‐calibration. On the next menu interface, press  F1 button again for “factory settings” as shown in Figure 1‐4. Figure 1‐4 Warning: please confirm that the oscilloscope has been safely grounded to avoid hazard. Oscilloscope accessing to signal 1) The oscilloscope has two input channels with an additional external trigger input channel. Please check the signal according to the following steps. Connect a probe to  CH1  input  terminal  and  set  the  attenuation  rate  switch  on  the  probe  as  10× (Figure 1‐5)

measurement  result  correctly  reflect  the  amplitude  of  measured  signals.  Probe

attenuation  coefficients  are  set  in  the  following  ways:  press  F4  to  make  the  menu display 10×. Figure 1‐6 Setting of Attenuation Coefficient on the oscilloscope 3) Connect the probe pin and ground clamp to the corresponding terminal of the probe compensation  signal.  Press  AUTO  key.  Within  several  seconds,  square  wave  display (1kHz, about 3V, peak‐to‐peak value) is as shown in Figure 1‐7. Check CH2 in the same way; press OFF function key to close CH1; press CH2 function key to turn on CH2; repeat step 2 and 3.

2.    Observe displayed waveform

Figure 1‐8a Probe Compensation and Correction If the screen displays “under‐compensation” or “over‐compensation” for waveform as shown  in  above  figures,  use  screwdriver  with  nonmetal  handle  to  adjust  the  variable capacitance on the probe until screen displays “correct compensation” for waveform as shown in the below figure. Over‐compensation Correct compensation Under‐compensation

Warning:  To  avoid  electrical  shock  when  measuring  high  voltage

with probe, please ensure that the insulation lead of the probe is in good  condition  and  do  not  contact  the  metal  part  of  the  probe when connecting to power supply. 5. Automatic Setting of Waveform Display The  oscilloscope  has  automatic  setting  function.  According  to  input  signals, automatically adjust the vertical deflection factor, scanning time base and trigger mode until the most appropriate waveform is displayed. In automatic setting, the frequency of the measured signal is required to be ≥ 50Hz and the duty ratio is > 1%. Apply automatic setting 1. Connect measured signal to signal input channel. 2.  Press  AUTO  key.  The  oscilloscope  will  automatically  set  its  vertical  deflection  factor, scanning time base and trigger mode. If further careful observation is required, adjustment can be conducted again after automatic setting until the waveform display reaches required optimum effect. 6. Introduction to the Vertical System As shown in the figure on next page, there are a series of keys and knobs in the vertical control area. Practices below will gradually guide you to get familiar with the controlling of the vertical system. Figure 1‐8b: Correct compensation

      Model 72‐10510 (Right)

1. Use the vertical position knob to move the vertical position of the waveform display on the LCD screen. By default, waveforms should be displayed at the center of the screen. 2. Users should see the waveform will move vertically upward and downward when the knob is rotated. 3. Users can also change the waveform scale by rotating the scale knob on the vertical system. Please notice that the “V/div” will be changed to indicate the corresponding vertical scale level. 4. Users can enable/disable the input channel by pressing the [CH1] and [CH2] button on the vertical system. (For 72‐10510) Attention: To  disable  the  input  channel  for  72‐7610,  please  press  [OFF]  button  on  the  vertical system. More vertical system setting will be explained in “Operations” section of this user manual. 7. Introduction to the Horizontal System As  shown  in  the  figure  on  next  page,  there  is  one  key  and  two  knobs  within  the horizontal control area. Practices below will gradually guide you to get familiar with the horizontal time base setting.

72‐7610 (Left)                        72‐10510 (Right)

1. Use horizontal SCALE knob to change horizontal time base scale settings and observe the state information changes. Turn horizontal SCALE knob to change “s/div” time base scale, and you can find corresponding changes in the time base scale display of corresponding channel in the status bar. Horizontal scanning rate steps up in the multiples of 1, 2 and 5 e.g. 1 ms /div‐>2 ms/div‐>5 ms/div‐>10ms /div…. etc. Note: the horizontal time base range of 72‐7610/72‐10510 varies between 2ns/div and 50s/div. Details are listed in technical specifications. 2. Use horizontal POSITION knob to adjust the signal horizontal position in the waveform window.  Horizontal  POSITION  knob  controls  the  signal  trigger  shift.  When  applied  to trigger shift, the waveform horizontal movement can be observed along with POSITION knob turning. More horizontal system setting will be explained in “Operations” section of this user manual. 8. Introduction to the Trigger System As  shown  in  Figure  1‐11,  one  knob  and  three  (or  two)  keys  are  in  the  trigger  menu control  area.  Practices  below  will  gradually  guide  you  to  be  familiar  with  the  trigger system setting.

72‐7610 (Left)        72‐10510 (Middle)        Trigger Menu Display (Right)

More trigger system settings will be explained in “Operations” section of this user manual. Ⅴ. Specifications Unless otherwise specified, all technical specifications are applicable to the probes with attenuation switch set as 10× and to the oscilloscope. The oscilloscope must first meet the following two conditions to satisfy those specification standards: The  oscilloscope  must  continuously  run  for  over  half  an  hour  in  the  operating temperature environment. A. Technical Specifications Sampling System Specifications Sampling rate Real‐time Equivalent Sampling rate 72‐7610 1GS/s 50GS/s 72‐10510 25MS/s Average N can be chosen among 2, 4, 8, 16, 32, 64, 128 and 256; N is the number of average samples. Note: The real time sampling rate of 72‐7610 is 250MS/s, without equivalent sampling. Input Channel Specifications Input coupling DC, AC and GND Input impedance 72‐7610 1±2%MΩ with 21±3pF 72‐10510 1±2%MΩ with 24±3pF Probe attenuation coefficient 1×, 10×, 100×, 1000×

Model 72‐7610/72‐10510 User Manual 13 Maximum input voltage 400V (DC+AC peak value, 1MΩ input impedance) Time delay between Channels (Typical) 150ps Horizontal System Specifications Waveform interpolation Sin (x) /x Horizontal resolution 25pixels/div Record length 2×512k sampling point Storage depth 25k Scanning scope 72‐7610 2ns/div‐50s/div 72‐10510 2ns/div‐50s/div Sampling rate and delay time accuracy ±50ppm (any interval≥1m) Measurement accuracy of time interval (△T) (full bandwidth) Single time: ± (1 sampling interval+50ppm×reading+0.6ns) >16 average values: ± (sampling interval+50ppm×reading+0.4ns) Vertical System Specifications Analog‐to‐digital converter (A/D) 8‐bit resolution, sampling at two channels at the same time Deflection coefficient (V/div) range (at input BNC) 72‐7610 2mV/div‐5V/div 72‐10510 1mV/div‐20V/div Vertical range ±10div (72‐7610 is ±5div) Selectable bandwidth limitation (Typical) 20MHz Low frequency response (AC coupling, ‐3dB) ≤10Hz (at BNC) DC gain accuracy (sampling or average sampling mode) 72‐7610: ±4% (When vertical sensitivity is 2mV/div or 5mV/div) ±3% (When vertical sensitivity is 10mV/div‐ 5V/div)

Model 72‐7610/72‐10510 User Manual 14 72‐10510: ±5% (When vertical sensitivity is 1mV/div or 2mV/div) ±4% (When vertical sensitivity is 5mV/div) ±3% (When vertical sensitivity is 10mV/div ‐20mV/div) DC measurement accuracy (average sampling mode) 72‐7610: When vertical position is 0 and N≥16: ± (4%×reading + 0.1div + 1mV) at 2mV/div and 5mV/div; ± (3%×reading + 0.1div + 1mV) from 10mV/div to 5V/div; When vertical position is not at 0 and N≥16: ±[3%×(reading + vertical position reading) + (1%×vertical position reading)]+0.2div) from 2mV/div to 200mV/div + 2mV The setting value from 200mV/div to 5V/div plus 50mV 72‐10510: When vertical position is 0 and N≥16: ± (5%×reading+0.1div+1mV) and selects 1mV/div or 2mV/div; ± (4%×reading+0.1div+1mV) and selects 5mV/div; ± (3%×reading+0.1div+1mV) and selects 10mV/div or 20mV/div; When vertical position is 0 and N≥16: ±(3%×(reading + vertical position reading) + (1%×vertical position reading)]+0.2div) The setting from 5mV/div to 200mV/div plus 2mV; the setting value from 200mV/div to 20V/div plus 50mV Measurement accuracy of voltage difference (△V) (average sampling mode) Under the same setting and environment conditions and after averaging the captured waveforms with a quantity of ≥16, the voltage difference (ΔV) between any two points on the waveform: ± (3%×reading+0.05div) Note: The range of deflection factor for 72‐10510 is 2mV/div‐10V/div (in the place of input BNC).

Model 72‐7610/72‐10510 User Manual 15 Bandwidth for Each Model Model Analog Bandwidth Real‐Time Bandwidth Rise Time 72‐7610 150MHz 100MHz 2.3ns 72‐10510 25MHz 25MHz 14ns Trigger System Specifications Trigger sensitivity ≤1div Range of trigger level Interior From the screen center ±5div EXT ±3V EXT/5* ±15V Trigger level accuracy (Typical) applicable for the signal with rising and falling time ≥20ns Interior ±(0.3div×V/div) (within±4div from the screen center) EXT) ±(6% setting value+40mV) EXT/5* ±(6% setting value +200mV) Pre‐trigger capacity Normal mode/scan mode, pre‐trigger/delay trigger, the pre‐trigger depth adjustable Hold‐off range 72‐7610 100ns ‐ 1.5s 72‐10510 80ns ‐ 1.5s Set the level to 50% (Typical). Operate under the condition of input signal frequency of ≥50Hz Edge Trigger Edge type Rise, fall, rise and fall Pulse Width Trigger Trigger mode (>, <, and =) positive pulse width and (>, <, and =) negative pulse width range Pulse width range 20ns ‐ 10s Video Trigger* Trigger sensitivity (video trigger, Typical) Interior 2div Vpp EXT) 400mV EXT/5* 2V Signal model and line/field frequency (video trigger type) Support standard NTSC and PAL, and the line ranges are respectively 1‐525 (NTSC) and 1‐625 (PAL).

Model 72‐7610/72‐10510 User Manual 16 Alternating Trigger CH1 trigger Edge, pulse width, and video CH2 trigger Edge, pulse width, and video Note: No EXT/5 function for 72‐10510. Measurements Cursor Cursor Parameters Voltage difference between cursors (△V), Time difference between cursors (△T), Reciprocal of △T (Hz) (1/△T) Track Voltage value and time value of point of waveform Automatic measurement Vpp, Vamp, Vmax, Vmin, Vtop, Vbase, Vmid, Average, Vrms, Overshoot, Preshoot, Frequency, Period, Rise Time, Fall Time, +Width, ‐Width, +Duty, ‐Duty, Delay Math operation +, ‐, ×, ÷ Store (waveform and setting) 20 groups of waveforms, and 20 kinds of settings FFT Window Hanning, Hamming, Blackman, Rectangle Sampling size 1024 points Lissajous figure Phase difference ±3 degrees Trigger frequency meter (72‐10510 only) Reading resolution 6 bits Trigger sensitivity ≤30Vrms Accuracy (typical) ±51ppm (+1 character) B. General Technical Specifications Display 72‐7610 72‐10510 Display types 5.7‐inch LCD 7‐inch LCD Display resolution 320 horizontal×240 vertical pixels (RGB) 400 horizontaB×240 vertical pixels (RGB) Display color Color Waveform luminance Adjustable (color) (72‐10510 only) Backlight intensity (typical) 300nit

F1.6AL  250V  (Fuse  is  inside  the  instrument

housing) Environment Specifications Temperature Range Operating Temperature Range: 0Ԩ~+40Ԩ Storage Temperature Range: ‐20Ԩ‐+60 Cooling Build‐in cooling fan Operating Humidity Range <35Ԩ: ≤90%RH 35Ԩ‐40Ԩ: ≤60%RH Operating Altitude Below 3,000m Mechanical Specifications 72‐7610 72‐10510 Size W 320mm 306mm H 150mm 147mm D 130mm 122mm Weight Excluding package 2.5kg 2.2kg Including package 4.0kg 3.3kg IP Protection IP 2x Calibration Interval The recommended calibration interval is one year.

system  controls  of  the  oscilloscope.  We  recommend  you  to  re‐read  the  “Functions”

section if you are still not familiar with the basic operation controls. In this section, you will learn different settings related to the oscilloscope: 1. Set the Vertical System (CH1, CH2, MATH, REF, OFF) 2. Set the Horizontal System (MENU or HORI MENU) 3.   Set the Trigger System (MENU, TRIG MENU, 50% and FORCE) 4. Set the Acquire System (ACQUIRE) 5. Set the Display System (DISPLAY) 6. Set the Storage and Load System (STORAGE) 7. Set the Utility System (UTILITY) 8. Set the Measurement Parameters (MEASURE) 9. Set the Cursor Measurement (CURSOR) 10. System Message Examples and Troubleshooting It is suggested to read this section carefully to understand multiple measuring functions and system operation methods of the oscilloscope. 1. Set the Vertical System Each channel for  the oscilloscope  can  be set independently on  the vertical system menu. After pressing  CH1 or  CH2 function key, the system shall display function menu of CH1 or CH2 channel. See Table 2‐1 below for description. Table 2‐1

Activate waveform invert function.    A

waveform will be displayed reversely. Off A waveform is displayed normally. Setting coupling channel You will learn from the below example to change input channel coupling. Suppose the signal being measured in CH1 is a sinusoidal signal containing DC component. Press F1 to select AC coupling. Then the DC component is blocked and the waveform display is shown in the figure below.

If  you  press  F1  to  select  DC  coupling,  both  the  DC  and  AC  components  of  the  signal

being measured in CH1 can pass. The waveform display is shown in the figure below. Figure 2‐2 Simultaneous Display of Signal DC and AC Components If you Press F1 to select grounding, to set CH1 to connect to the internal ground of the instrument, both DC and AC components of the input signal are blocked. The waveform display is shown in the figure below. Note: although no waveform is displayed on the screen in this mode, input signal is still connected with channel circuit. Figure 2‐3 Simultaneous Blocking of Signal DC and AC Components Setting channel bandwidth Limit You  will  learn  through  the  following  example  to  change  the  bandwidth  limit  of  the  input channel. Assuming that the input signal to CH1 is a 40MHz sinusoidal signal. If you press F2

to  set  bandwidth  limit  as  OFF,  CH1  is  with  full  bandwidth.  All  AC  and  DC  high‐frequency

components in the measured signal can pass through. The waveform is displayed as below. Figure 2‐4 Waveform Display When Bandwidth Limit is OFF If  you  press  F2  again  to  set  the  bandwidth  limit  as  ON,  the  noises  or  high‐frequency component over 20MHz in measured signal shall be attenuated. The waveform is displayed as below. Figure 2‐5 Waveform Display When Bandwidth Limit is ON Setting probe rate In order to cooperate with probe attenuation coefficient setting, it is required to set the probe  attenuation  coefficient  on  channel  function  menu.  If  the  probe  attenuation coefficient is 10:1, the probe coefficient on channel function menu shall be set as 10× and vice‐versa.

ON.  Please  see  Figure  2‐8  for  a  non‐inverted  waveform  and  Figure  2‐9  for  an  inverted

waveform.

Model 72‐7610/72‐10510 User Manual 23 Figure 2‐8 Vertical Channel Inverted Setting (Invert: OFF) Figure 2‐9 Vertical Channel Inverted Setting (Invert: ON) Mathematical operation function The oscilloscope supports mathematical operation function for the input signals. The figure below shows the match result for CH1 + CH2 as an example. Figure 2‐10 Mathematical Operations After pressing the MATH button at the menu area to active mathematic operations.

transform)  operations,  the  time  domain  (YT)  signal  can  be  converted  into  frequency

domain signal. It is convenient to observe the following signals by FFT operations:  Harmonic content and distortion in measurement system  Noise characteristics in DC power supply display  Vibration analysis Table 2‐3: FFT Menu Description Function Menu Settings Description Type FFT FFT mathematical operations Signal source (F2) CH1 CH2 Set to select CH1 or CH2 as waveform to be transform Window (F3) Hanning Hamming Blackman Rectangle Set to select Hanning, Hamming, Blackman, or Rectangle window function Vertical unit (F4) Vrms dBVrms Set vertical units as Vrms or dBVrms FFT operation skills Signal with DC component or deflection will lead to error or deviation of FFT waveform component. AC coupling mode can be selected to reduce DC component. In order to reduce random noise of repeated or single pulse event and aliasing frequency components, the average acquiring mode is set.

Assuming  that  YT  waveform  is  continuously  repeated,  the  oscilloscope  shall  conduct  FFT

conversion for time record with finite length. In this case, when the cycle is an integer, YT waveforms are of the same amplitude at starting and ending positions, without interrupting. However,  if  the  YT  waveform  cycle  is  not  an  integer,  the  waveform  amplitudes  shall  be different  at  starting  and  ending  positions,  and  high‐frequency  transient  interruption  shall occur at the joints. In frequency domain, this effect is called leakage. Therefore, in order to avoid the generation of leakage, by multiplying original waveform by a window function, the value at forced starting and ending positions is 0. See the table below for window function application: Table 2‐4 FFT Window Function FFT Window Features The most suitable measurement content Rectangle The best frequency resolution and the worst amplitude resolution. It is basically similar to the condition without window. Signal levels, before and after transient state and short pulse are basically equal. Constant amplitude sinusoid with very similar frequency possesses bandwidth random noise with slow spectrum change. Hanning Compared with rectangular window, it has better frequency resolution and worse amplitude resolution. Random noise of sine, cycle and narrow‐band. Hamming The frequency resolution of Hamming window is slightly superior to Hanning window’ s. Signal levels, before and after transient state or short pulse, are of large difference.

FFT  resolution:  defined  as  the  quotient  of  sampling  and  operation  point.

When the number of operation point is fixed, the lower the sampling rate is, the better the FFT resolution will be. Nyquist  frequency:  for  waveform  with  f  maximum  frequency,  at  least  2f sampling rate is required to reconstruct original waveform. It is also called Nyquist rule, where f refers to Nyquist frequency, while 2f refers to Nyquist sampling rate.

3.  To  load  this  waveform  back  to  Ref  A,  users  should  do  the  following  on  the  Load

function menu. Select the memory location 3 by using multi‐purpose knob Select DSO on the Disk selection Press F4, a loading progress bar will be shown on the screen. If loading is done, a white waveform should be displayed on the screen labelled Ref A. If no waveform is stored at the selected memory location, you will see a system message show “no data in the position” Users can turn off the Ref A reference waveform by pressing OFF F3 button on the load function menu. Attention: Loading the pre‐saved waveform into Ref B follows the same procedure as above. If waveform is stored to the USB device, 200 available memory locations can be selected. Please see “Set the Storage and Load System” for how to save a waveform to USB device. 2. Set the Horizontal System The horizontal control knob can be used to change the horizontal scale (time base) and trigger horizontal position (trigger position) in internal storage. The vertical midpoint in screen horizontal direction is the time reference point of waveform. Changing of horizontal scale can lead to expansion or shrink of waveforms relative to screen center. When the horizontal position changes, the position relative to waveform trigger point will be changed. The horizontal system can be controlled by the following panel buttons/knobs. Figure 2‐11 Horizontal System Interface 72‐7610 (Left)              72‐10510 (Right)

The  horizontal  scale  control  knob  is  used  to  change  the  time  base  shown  as  s/div  of

channel waveform. If the extension time base is turned on, the window width will be changed  by  adjusting  the  delay  scanning  time  base  after  turning  the  horizontal  scale knob. Horizontal menu button: users can press to enter the horizontal menu function. (See table below) Table 2‐6 HORI MENU / MENU Function Menu Settings Description Main time base (F1) —— Press to turn on the main time base —— Window extension (F3) —— Press  to  turn  on  the  extension  time base —— Trigger hold‐off (@) —— To adjust the trigger hold‐off time by multi‐purpose knob Window extension time base function is used to observe the waveform detail for a given window  of  the  waveform.  Users  can  active  this  function  by  pressing  F3  button  at  the horizontal function menu. Figure 2‐12 Screen Display under Window Extension Waveform horizontal extension Horizontally extended part of waveform Extension time base

During  trigger  hold‐off,  the  oscilloscope  will  not  be  triggered  for  any  new  event  until  the

hold‐off time is over. For example, if a train of pulse shall be triggered from the first pulse upon request, then the hold‐off time can be set as the pulse train width. As shown in Figure 2‐13. Figure 2‐13 Trigger Hold‐off for Synchronization of Complex Waveform The trigger setting decides when the waveform data will start to be collected and display on the  screen.  Once  trigger  is  correctly  set,  it  can  convert  unstable  display  into  meaningful waveform  display.  The  starting  trigger  conditions  of  data  acquisition  are  satisfied.  After  a signal  trigger  is  detected,  the  oscilloscope  will  continuously  collect  data  and  display waveform on the right of the trigger point.

Users  can  activate  the  alternative  trigger  by  pressing  [F2]  button  at  [TRIG  MENU]  for

72‐10510, by pressing [F2] button at [MENU] for 72‐7610; Edge Trigger Edge  trigger  mode  refers  to  the  triggering  of  the  threshold  on  the  input  signal  edge. When selecting “edge trigger”, trigger on the rise edge or fall edge of input signal shall be completed. Table 2‐7 (Page 1)

Set  external  trigger  input  channel  as  the

triggering signal source EXT/5 Set external trigger source (divided by 5) as trigger level range (Only for 72‐7610) AC Line Set AC power line as trigger source Alter Set  to  alternative  trigger  between  CH1 and CH2 as signal source Table 2‐7 (Page 2) Slope (F3) Rising Set trigger on the signal rising edge Falling Set trigger on the signal falling edge Rise/fall Set trigger on the signal both rising and falling edge Mode (F4) Automation Set to automatic trigger. The oscilloscope will continuously perform data acquisition without triggering signal. Normal Set to normal trigger. The oscilloscope will only perform data acquisition when there is triggering signal. Single Set to single trigger. The oscilloscope will only perform 1 cycle of the data acquisition when there is triggering signal. Trigger coupling (F5) AC Block the DC component of triggering signal DC Pass through DC and AC components of triggering signal High‐frequency rejection Reject high‐frequency component of the triggering signal (over 80kHz signals)

Model 72‐7610/72‐10510 User Manual 32 Low‐frequency rejection Reject low‐frequency component of the triggering signal (below 80kHz signals) Pulse Width Trigger For pulse width trigger, the trigger time shall be subject to pulse width of the triggering signal. You can capture abnormal pulse by setting pulse width conditions. Table 2‐8 (Page 1) Function Menu Settings Description Type Pulse width Source (F2) CH1 Set CH1 as triggering signal source CH2 Set CH2 as triggering signal source EXT Set external trigger input channel as triggering signal source EXT/5 Set external trigger source (divided by 5) as trigger level range (Only for 72‐7610) AC Line Set AC power line as trigger source Alter Set to alternating trigger between CH1 and CH2 as signal source Pulse width conditions (F3) > Trigger when pulse width is greater than the setting values; < Trigger when pulse width is less than the setting values; = Trigger when pulse width is equal to the setting values; Pulse width setting (@) Set the pulse width to 20ns~10s by using the multi‐purpose knob on the front panel Next page 1/2 —— Go to next page Table 2‐8 (Page 2) Function Menu Settings Description Type Pulse width Set the positive pulse width as trigger signal; Set the negative pulse width as

Trigger  coupling  will  automatic  preset  to  DC  coupling  if  the  video  trigger  is  active.  Video

trigger menus are shown in the table below: Table 2‐9 Video Trigger Setting Function Menu Settings Description Type Video Source (F2) CH1 Set CH1 as triggering signal source

Figure  2‐14  is  an  example  for  the  display  when  PAL  video  trigger  model  is  selected  as

standard and synchronization mode is all line synchronization. Figure 2‐14 Video Trigger: Line Synchronization Figure 2‐15 is an example for the display when PAL video trigger model is selected as standard and synchronization mode is set to field synchronization.

CH2  alternately.  Alternating  trigger  is  useful  for  observing  two  signals  with  different

frequencies. See the figure below for the display of triggered alternating waveform and Table 2‐11 for triggered alternating menu setting: Figure 2‐16 Observing Signals with Two Different Frequencies by Alternating Trigger Mode Table 2‐10 Alternating Trigger Mode Setting Functional menu Settings Description Type Edge Edge trigger mode is set. Trigger source Alternating CH1 and CH2 alternating trigger Slope Ascending Rise edge is set for slope trigger.

Force  trigger  can  be  understood  as  forcing  the  oscilloscope  to  refresh  the  data

acquisition process, so users can observe waveform capturing. To enable force trigger, please press FORCE button. One of the very useful applications for force trigger is in the “normal” and “single” triggering mode.  When  the  triggering  signal  does  not  present,  the  oscilloscope  will  stop  data acquisition.  You  can  press  the  force  trigger  to  re‐enable  the  data  acquisition  of  the oscilloscope to see the current input signal. Explanation of nouns 1. Trigger source: Trigger can be achieved from several signal sources: input channels (CH1 and CH2), external triggers (EXT and EXT/5) and AC line. ■ External trigger (EXT TRIG): Meaning to set the triggering source according to the given  external  connection  to  the  “EXT  TRIG”  input  channel.  This  can  be  a  significant application for the oscilloscope to perform data acquisition using CH1 and CH2 based on an external clock signal connects to the “EXT TRIG” input channel. External trigger signal connects to “EXT TRIG” input terminal is used for EXT and EXT/5 trigger source selection. When selecting EXT as trigger source, external trigger levels voltage range limited to ‐3V and +3V. Whereas, if EXT/5 is selected as trigger source, external trigger level voltage range can be expanded to ‐15V‐+15V. (EXT/5 is only available for 72‐7610.) ■ AC line trigger: Use AC line power as trigger source. This can be significant to observe the signals correlated to AC line power for applications in lighting equipment and power supply equipment research and development. 2.  Trigger  mode:  Deciding  the  oscilloscope  behavioral  pattern  under  no  trigger  event condition. Three trigger modes are available: auto, normal and single. ■  Auto  trigger:  Set  to  automatic  trigger  mode.  The  oscilloscope  will  continuously perform  data  acquisition  without  triggering  signal.  When  there  is  trigger  signal,  the oscilloscope will automatically synchronize with the trigger signal. Attention: when scanning waveform is set on 50ms/div or slower time base, no trigger signal is needed under “auto trigger” mode. ■  Normal  trigger:  Set  to  normal  trigger  mode.  The  oscilloscope  will  perform  data acquisition only when trigger conditions are satisfied. ■ Single trigger: Set to single trigger mode. The oscilloscope will perform one‐cycle data

Model 72‐7610/72‐10510 User Manual 37 acquisition only when trigger conditions are satisfied. 4. Set the Acquire System Users can change the data acquisition mode by pressing  ACQUIRE button as shown in the figure below. Figure 2‐17 Function Key of Sampling System While ACQUIRE button is pressed, the data acquisition setting menu for the oscilloscope will display. Table 2‐11 Acquire Menu Function Menu Settings Description Acquisition Mode (F1) Sampling Set to sampling acquisition mode Peak value Set to peak value acquisition mode Average Set to average acquisition mode Averages (@) 2~256 In average sampling mode, set average number by using the multipurpose knob (Averages can be set as 2, 4, 8, 16, 32, 64, 128, 256) Sampling (F3) Real‐time Set to Real‐time sampling Equivalence Set to Equivalent sampling Fast Acq (F4) ON Turn on fast acquiring. Waveform capturing rage up to 2k waveform/second. OFF Turn off fast acquiring

2.  Please  select  equivalent  sampling  mode  when  observing  high  frequency  periodic

signal. 3. Please select peak detection mode when hoping to observe the signal envelope for

4.  Please  select  average  sampling  mode  when  hoping  to  reduce  the  random  noise  in

displayed signals. 5. Set the Display System Users  can  change  the  display  setting  of  the  oscilloscope  by  pressing  the  DISPLAY  button. Please refer to the figure below. Figure 2‐20 Function Keys of Display System After pressing DISPLAY button, the function menu will change to the display setting for the oscilloscope. Table 2‐12 Display Menus Function Menu Settings Description Explanation of nouns Real‐time sampling mode: Data sampling with the system real‐time rate is used to observe the waveform within the rate. Equivalent sampling mode: Data  sampling  with  the  rate  higher  than  the  maximum.  This  is  used  to observe the periodical waveform at beyond the maximum sampling rate.

CH1  and  CH2  are  required  to  be  applied  simultaneously.  The  figure  below  shows  an

example for the X‐Y mode display. Figure 2‐21 Waveform Display in X Y Mode Attention: The following functions will be disabled at X Y display mode. Automatic measurement Cursor measurement 6. Set the Storage and Load System

                        Function Key of Sampling System (STORAGE)

Below  steps  will  guide  you  to  store  a  waveform  into  the  internal  memory  of  the oscilloscope: Press the Storage button and you will see the following menu. Table 2‐13 Waveform Storage Menu Function Menu Settings Description Type (F1) Waveform Select to store an acquired waveform Source (F 2) CH 1 Select waveforms from CH1 channel to be saved CH 2 Select waveforms from CH2 channel to be saved Dest (@) 1~20 Destination (Position of the internal memory) where the waveform will be saved. There are 20 sets of internal memory available to store waveform navigated by the multipurpose knob. Save —— Store waveform Next page —— Get to the next page

Any  waveform  storage  will  overwrite  previous  memory  space  at  the  same  memory

location. To  load  a  waveform  from  the  internal  memory,  please  refer  to  the  item,  “Load  a Reference Waveform” in this section. Below steps will guide you to store a waveform to USB device: Insert a USB disk into the “USB Host” interface. If the USB is connected successfully, you should see the system message “USB device install successfully” shown on the screen. Press the Storage button and you will see the following menu. Table 2‐14 Waveform Storage Menu (Page 1) Function Menu Settings Description Type (F1) Waveform Select to store an acquired waveform Source(F2) CH1 Select waveforms from CH1 channel to be saved CH2 Select waveforms from CH2 channel to be saved Dest (@) 1~20 Destination (Position of the internal memory) where the waveform will be saved. There are 20 sets of internal memory available to store waveform navigated by the multipurpose knob. Save —— Store waveform Next 1/2 [F5] —— Get to the next page Press F5 to go to next function menu Users should see below function menu. Table 2‐14 Waveform Storage Menus (Page 2) Function Settings Description

Use  the  multi‐purpose  knob  to  select  the  storage  memory  destination  (please  notice

that there will be 200). Press F4 to store the waveform. Remove the USB disk. Attention: Any  waveform  storage  will  overwrite  previous  memory  space  in  the  same  memory location. If “Long” length data is selected, the data cannot be re‐loaded back to display and can only be re‐load through software application. To load a waveform from the USB disk, please refer to “Load a Reference Waveform” in this section. Below steps will guide to store a setting to the internal memory of the oscilloscope: Press the Storage button and you will see the following menu. Table 2‐15 Waveform Storage Menu (Page 1) Function Menu Settings Description Type (F1) Waveform Select to store an acquired waveform Setup Select to store a setting Source (F2) CH1 Select waveforms from CH1 channel to be saved

Below  steps  will  guide  you  to  load  a  setting  from  the  internal  memory  of  the

oscilloscope: From the setting storage menu (Page 1) Use the multi‐purpose knob to select the storage memory location. Press F4 to load the pre‐stored setting.

Destination  (Position  of  the  USB  disk

memory)  where  the  bitmap  files  will  be saved.  There  are  200  sets  of  USB  disk

memory  available  to  store  bitmap  file

navigated by the multipurpose knob. Save [F4] —— Press to store a bitmap file —— Use  the  multi‐purpose  knob  to  select  the  storage  memory  destination  (please  notice that 200‐group data storage is available). Press F4 to store a bitmap file into a USB disk; Remove the USB disk. 7. Set the Utility System By pressing the UTILITY function button in the menu area, users can do different utility settings, where the button location is shown in the figure below. Figure 2‐23 Function Key of Sampling System (Function) There are three function menu pages for the utility system setting shown in Table 2‐17, 18, 19 below. Table 2‐17 Utility Menu Page 1 Function Menu Settings Description Self‐Adj (F1) Execute Press to execute self‐calibration function Close Leave self‐calibration Pass/Fail (F2) See Table 2‐22 Set waveform Pass/Fail operation

Model 72‐7610/72‐10510 User Manual 47 Recorder (F3) See Table 2‐20 Set waveform recording operation Language (F4) Language Press to change to different system language Next(1/3) (F5) Press to next Utility Function Menu page Table 2‐18 Utility Menu Page 2 Function Menu Settings Description Reset (F1) Reset to factory default settings Quick Correction (F2) On To turn on the channel fast correction function OFF To turn off the channel fast correction function Skin (F3) Change the skin for the function menu Grid Bright (@) 1%‐100% Adjust the grid brightness by using the multi‐purpose knob Next 2/3 —— Get to the next page Table 2‐19 Utility Menu Page 3 Function Menu Settings Description Version (F1) Press to view the system version LAN (F2) Press to enter the LAN address setting Notice that LAN is an option for 72‐7610 —— —— —— Frequency Counter * (F4) ON To turn ON frequency counter OFF To turn OFF frequency counter First 3/3 (F5) —— Back to page 1 The oscilloscope waveform recorder function is generally easy to operate. Users should set up Source and Operation according to the below table. Table 2‐20 Waveform Recorder Menu Function Menu Settings Description

After  pressing  the  F2  button  on  the  waveform  recorder  menu,  you  should  see  the

waveform recorder operation menu below. Table 2‐21 Waveform Recorder Operation Menu One of the very useful functions for the oscilloscope is the pass/fail detection. You can create a template of pixels area to compare with the input signal. Users can also Function Menu Setting Description Record (F1) Press to start waveform recording; The number of waveform being recorded is shown on the screen. Replay (F2) Press to start waveform replay Stop (F3) Press to stop waveform recording Return (F4) —— Go back to previous menu

You  can  enable  PASS/FAIL function  by pressing  F1  on PASS/FAIL function  menu. After

turning on this function, you can see the total number of waveform being processed, the number of waveform passed and the number of waveform failed on the LCD screen. Table 2‐22 Pass/Fail Function Menu Function Menu Settings Description Status (F1) ON Press to turn on PASS/FAIL function OFF Press to turn off PASS/FAIL function Source (F2) CH1 Select CH1 as the detection signal source CH2 Select CH2 as the detection signal source MATH Select MATH result as the detection signal source Ref A Select Ref A as the detection signal source Ref B Select Ref B as the detection signal source Output (F3) Pass Select the logical result as input signal within the template of pixels area Fail Select the logical result as input signal outside of the template of pixels area Pass/Halt Same as Pass output condition; Stop once detected. Fail/Halt Same as Fail output condition; Stop once detected. Template (F4) See Table 2‐23 Press to create template of pixels area Back (F5) / Return to Utility Function Menu After pressing the F4 on the Pass/Fail function menu, you can create a template of pixels to be compared. Table 2‐23 Template Setting Function Menu Settings Description

You  will  need  to  press  F1  to  confirm  creating  the  pixel  template  after  setting  the

horizontal and vertical length of the pixels. 8. Set the Measurement Parameters The oscilloscope supports up to 28 automatic waveform measurement parameters. You can view 5 measurement parameters at the same time the most on F1‐F5 function menu. To view measurement parameters, please press MEASURE on the control panel. Figure 2‐24 Function Key of Auto Measurement Parameters For each measurement parameter located at F1‐F5 on the function measurement, you can press F1‐F5 for the settings (Measurement Sub Menu) Table 2‐24 Measurement Sub Menu Function Menu Settings Description Back (F1) Return to measurement parameter

Model 72‐7610/72‐10510 User Manual 51 function menu Source (F2) CH1 Select CH1 as source CH2 Select CH2 as source Volt (F3) Select different voltage parameters Time (F4) Select different time parameters Parameters (F5) Press to display all the measurement parameters of the input source. Or, press to turn off parameter display. Voltage Parameters Below are the automatically measured voltage parameters for the oscilloscope and their definitions: Vpp: Peak‐to‐peak voltage. Vmax: Maximum voltage. Vmin: Minimum voltage. Vmid: Medium voltage. Vamp: The voltage amplitude from waveform top to bottom. Vtop: The voltage value from waveform top to GND (Ground). Vbase: The voltage value from waveform base to GND (Ground). Over‐shoot: The ratio of (Vmax – Vtop) to voltage amplitude (Vamp). Pre‐shoot: The ratio of (Vmin – Vbase) to voltage amplitude (Vamp). Average: The average signal amplitude within 1 cycle. Vrms: Root‐mean‐square value (effective value). Time Parameters Rise Time: The time of waveform amplitude rising from 10% to 90%. Fall Time: The time of waveform amplitude falling from 90% to 10%. +Width: The width of positive pulse at 50% amplitude. ‐Width: The width of negative pulse at 50% amplitude. Delay (rising edge): The delay time between rising edges. Delay (falling edge): the delay time between falling edges. +Duty: The ratio of positive pulse width to cycle. ‐Duty: The ratio of negative pulse width to cycle. 9. Set the Cursor Measurement Users can press the CURSOR button to enable the cursor measurement function of the oscilloscope, where the location of the CURSOR button can be found in the figure below.

At  the  Time  cursor  measurement,  users  can  perform  referring  to  Volt  cursor

measurement by using the multi‐purpose knob and SELECT button on 72‐7610. The Track cursor measurement is similar to the time cursor measurement except it will automatically track voltage level where the time cursor is pointing at the screen. (Not for 72‐10510). After activating the COARSE button, there will be quicker response for the cursor movement. 10. System Message Examples and Troubleshooting System Message Examples  “Operation at limit”: System will show “Operation at limit” when users try to operate beyond the system limit; For example, setting the horizontal scale to more than 50.00s.  “Saving” Progress Bar When users are saving a bitmap file or a waveform and setup file, system will show the

②  Check  if  the  probe  is  working  properly  by  connecting  the  probe  to  “Probe

compensation signal output” ③ Check if the probe is properly connected to the device under test ④ Try to acquire signal again  The voltage amplitude measured is 10 times higher or lower than the actual value: ‐ Check if the attenuation coefficient of the channel matches with the attenuation rate of the probe used.  The signal waveform is not stable: ①  Check  if  the  trigger  source  setting  in  the  trigger  menu  matches  with  the  actual channel input of the signal ② Check the trigger mode. Most of the signals can be triggered in edge trigger mode, while video signals can be triggered in video trigger mode. Only in correct trigger mode the waveform can be stably displayed. ③  Change  the  coupling  to  “high  frequency  rejection”  or  “low  frequency  rejection” display to filter out the high frequency or low frequency noise that disturbs the trigger  No waveform display after pressing RUN/STOP key ①  Check  if  the  trigger  level  exceeds  the  waveform  range  and  the  trigger  mode  is “Normal” or “Single”, if yes, manually set the trigger level to the center or change the trigger mode to AUTO. ②  Or  press  AUTO  button  next  to  the  RUN/STOP  button,  the  above  settings  will  be completed automatically.

②  In  “Average”  mode,  waveform  refreshing  is  slow  because  it  displays  the  average

value of the waveform to reduce the random noise in displayed signal. ③ Change to “Sample” mode or “Peak” mode for the desired display.  The waveform display do not smooth ① The horizontal time base division value may be too small, increase the horizontal time base division to improve the waveform smoothness; ② If the display type is “Vector”, the signal waveform may not be smooth between dots. This can be improved by setting the display type to “Dot”. Ⅶ. Maintenance & Cleaning A. General Maintenance Please do not store or place the instrument at any places where the LCD of the instrument is exposed to sunlight directly. Warning:  Please  do  not  stain  the  oscilloscope  or  probe  with  spray,  liquid  or  solvent  to avoid instrument or probe damage. B. Cleaning Check the instrument and probe frequently. Clean the surface of the instrument according to the following steps: 1. Please wipe the instrument and probe surface with soft cloth. Pay attention not to scratch the LCD screen. 2.  Wipe  the  instrument  with  wet  cloth  without  dripping  after  disconnecting  the  power supply. Use gentle detergent or clear water to clean. Do not use any abrasive chemical agent to avoid instrument or probe damage. Warning: Please make sure that the oscilloscope is completely dry before powering on again to avoid electrical short circuit or injury.

Model 72‐7610/72‐10510 User Manual 55 Tenma Test Equipment 405 S. Pioneer Blvd. Springboro, Oh