Tektronix CT-6 Instruction Manual
Read Online
Read the Tektronix CT-6 Instruction Manual online using the built-in PDF viewer.
Text of the manual
Copyright Tektronix, Inc. All rights reserved.
Tektronix products are covered by U.S. and foreign patents, issued and
pending. Information in this publication supercedes that in all previously
published material. Specifications and price change privileges reserved.
Tektronix Inc., P.O. Box 500, Beaverton, OR. 97077
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
Tektronix warrants that the products that it manufactures and sells will be free from defects
in materials and workmanship for a period of one (1) year from the date of shipment. If a
product proves defective during this warranty period, Tektronix, at its option, either will
repair the defective product without charge for parts and labor, or will provide a
replacement in exchange for the defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the
defect before the expiration of the warranty period and make suitable arrangements for the
performance of service. Customer shall be responsible for packaging and shipping the
defective product to the service center designated by Tektronix, with shipping charges
prepaid. Tektronix shall pay for the return of the product to Customer if the shipment is to
a location within the country in which the Tektronix service center is located. Customer
shall be responsible for paying all shipping charges, duties, taxes, and any other charges for
products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or
improper or inadequate maintenance and care. Tektronix shall not be obligated to furnish
service under this warranty a) to repair damage resulting from attempts by personnel other
than Tektronix representatives to install, repair or service the product; b) to repair damage
resulting from improper use or connection to incompatible equipment; c) to repair any
damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a
product that has been modified or integrated with other products when the effect of such
modification or integration increases the time or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX IN LIEU OF ANY OTHER
WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS
DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR
FITNESS FOR A PARTICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY
TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND
EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR BREACH OF
THIS WARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE
FOR ANY INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL
DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR
HAS ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
CT-6 Instruction Manual
i
Table of Contents
General Safety Summary
v
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contacting Tektronix
vii
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Getting Started
1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Accessories and Options
2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Standard Accessories
2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Options
2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Optional Accessories
2
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Installation
3
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Probe Holder
7
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Evaluation Board (Optional)
8
. . . . . . . . . . . . . . . . . . . . . . . . . . .
Operating Basics
9
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Interpreting the Current and Voltage Ratings
9
. . . . . . . . . . . . . .
Matching Impedances
10
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Measuring Current
11
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Injecting Current Signals
12
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Optimizing Performance at High Frequencies
13
. . . . . . . . . . . . .
Circuit Equivalent
13
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Selecting the Wire Size and Length
14
. . . . . . . . . . . . . . . . . . .
Positioning the Probe
14
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Ground Side and Signal Side Measurements
16
. . . . . . . . . . . . . . .
Using the Optional Evaluation Board
17
. . . . . . . . . . . . . . . . . . . .
Calibration Lines
18
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Evaluation Lines
18
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Probe Holder
18
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Specifications
21
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Performance Verification
25
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Equipment Required
25
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Preparation
26
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Mount the Current Probe on the Evaluation Board
26
. . . . . . .
Frequency Response and Sensitivity
27
. . . . . . . . . . . . . . . . . . . . .
Set up the Signal Generator with the Digital Oscilloscope
27
.
Measure Output and Calculate Sensitivity
27
. . . . . . . . . . . . . .
Measure Low End Frequency Response
27
. . . . . . . . . . . . . . .
Measure High End Frequency Response
28
. . . . . . . . . . . . . . .
CT-6 Test Record
29
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Table of Contents
ii
CT-6 Instruction Manual
Frequency Response and Sensitivity – Alternate Procedure
30
. . .
Set up the Network Analyzer
30
. . . . . . . . . . . . . . . . . . . . . . . .
Calibrate Network Analyzer with Evaluation Board
31
. . . . . .
Calibrate out Line Losses
31
. . . . . . . . . . . . . . . . . . . . . . . . . . .
Remove Delay
31
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Verify Frequency Response and Sensitivity
32
. . . . . . . . . . . .
CT-6 Test Record – Alternate Procedure
33
. . . . . . . . . . . . . . . . . .
Table of Contents
CT-6 Instruction Manual
iii
List of Figures
Figure 1: CT-6 current probe with probe holder and
BNC-SMA adapter
1
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 2: Using a single wire
5
. . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 3: Using two wires to measure differential
current
6
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 4: Using a square pin and jumper
6
. . . . . . . . . . . . . . .
Figure 5: Securing the CT-6 probe head to a test fixture
board
7
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 6: Signal measurement
11
. . . . . . . . . . . . . . . . . . . . . . . .
Figure 7: Signal injection
12
. . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 8: Circuit equivalent of added inductance and
capacitance
13
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 9: Positioning for best HF response
(20–25 gauge)
15
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 10: Coupling the probe to signal side or
ground side
16
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 11: Test Fixture board for the CT-6 current
probe
17
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 12: Securing the CT-6 probe head to the test
fixture board
19
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Figure 13: Typical frequency response (Bode plot)
22
. . . . . . .
Figure 14: Typical group delay (d/df)
22
. . . . . . . . . . . . . . . . .
Figure 15: Securing the current probe to the evaluation
board
26
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
CT-6 Instruction Manual
v
General Safety Summary
Review the following safety precautions to avoid injury and prevent
damage to this product or any products connected to it. To avoid
potential hazards, use this product only as specified.
Only qualified personnel should perform service procedures.
To Avoid Fire or Personal Injury
Connect and Disconnect Properly. Do not connect or disconnect probes
or test leads while they are connected to a voltage source.
Observe All Terminal Ratings. To avoid fire or shock hazard, observe all
ratings and markings on the product. Consult the product manual for
further ratings information before making connections to the product.
Do Not Operate Without Covers. Do not operate this product with
covers or panels removed.
Do Not Operate With Suspected Failures. If you suspect there is damage
to this product, have it inspected by qualified service personnel.
Do Not Operate in Wet/Damp Conditions.
Do Not Operate in an Explosive Atmosphere.
Keep Product Surfaces Clean and Dry.
General Safety Summary
vi
CT-6 Instruction Manual
Safety Terms and Symbols
Terms in This Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that
could result in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that
could result in damage to this product or other property.
Terms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you
read the marking.
WARNING indicates an injury hazard not immediately accessible as
you read the marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. These symbols may appear on the product:
CAUTION
Refer to Manual
Double
Insulated
Protective Ground
(Earth) Terminal
CT-6 Instruction Manual
vii
Contacting Tektronix
Product
Support
For application-oriented questions about a Tektronix
measurement product, call toll free in North
America:
1-800-TEK-WIDE (1-800-835-9433 ext. 2400)
6:00 a.m. – 5:00 p.m. Pacific time
Or contact us by e-mail:
[email protected]
For product support outside of North America,
contact your local Tektronix distributor or sales
office.
Service
Support
Contact your local Tektronix distributor or sales
office. Or visit our web site for a listing of
worldwide service locations.
http://www.tektronix.com
For other
information
In North America:
1-800-TEK-WIDE (1-800-835-9433)
An operator will direct your call.
To write us
Tektronix, Inc.
P.O. Box 500
Beaverton, OR 97077
CT-6 Instruction Manual
1
Getting Started
The CT-6 is a miniature AC current probe suited for probing
low-voltage, high-frequency circuits. Performance highlights include
the following:
250 kHz to 2 GHz typical bandwidth
5 mV/mA sensitivity
low insertion impedance
inductive pick up or injection of signals
For a complete list of specifications, see page 21.
Figure 1: CT-6 current probe with probe holder and BNC-SMA adapter
CT-6 Instruction Manual
Accessories and Options
Standard accessories accompany every CT-6 current probe. Both
standard and optional accessories may be ordered separately. Options
must be ordered at the time of purchase.
Standard Accessories
The following accessories ship with every CT-6 current probe:
Instruction Manual (071-0453-00)
BNC-SMA adapter (015-0572-00)
Probe holder (015-0682-00) with 2-56 screws (211-0034-00)
Marker bands for identifying multiple probes on the inputs of a
test instrument (016-1315-00)
Certificate of traceable calibration
Options
The following CT-6 options are available at the time of purchase:
Option C3 (three years calibration services)
Option D1 (calibration data)
Option D3 (three years calibration data)
Option R3 (extended 3-year repair/service contract)
Option TF (evaluation board included with the current probe)
Optional Accessories
The following optional accessories are recommended for the CT-6
current probe:
Evaluation board without the current probe (067-0349-00)
SMA–SMA barrel adapter (015-1012-00)
CT-6 Instruction Manual
3
Installation
Install the probe on the measurement instrument and in the circuit as
detailed in this section.
To measure or inject a signal using the CT-6 current probe, do these
steps:
1. Connect the SMA output connector of the CT-6 current probe to
the measurement instrument, or to the signal generator if
injecting a signal. Use the SMA-BNC adapter if necessary. Make
sure the instrument or signal generator has a 50 impedance.
WARNING. To avoid personal injury, use the probe on grounded
systems only. To avoid RF burn hazard, apply stimulus signal only
after connecting the probe to the circuit.
2. Power off the circuit under test.
3. Choose the point in the circuit where you are going to measure
current or inject a signal.
WARNING. To avoid personal injury or fire, do not exceed the
maximum bare wire voltage of 30 V. Avoid using the probe in circuits
where voltages exceed 30 V.
Maintain adequate spacing between the housing on the probe head
and other exposed circuit connections. The hole in the probe head is
insulated, but the painted housing on the probe head is at ground
potential.
4. Interrupt the circuit under test and route a conductor through the
hole in the probe head in one of the following ways:
a. For the best overall signal fidelity, use a short wire with a
small diameter. (See Figure 2.)
b. For convenience in measuring signals below 1 GHz you
can use a square pin and jumper. (See Figure 4.)
CT-6 Instruction Manual
c. For measuring the difference between two signal currents,
use two wires with opposing polarities. (See Figure 3.)
Fine, thinly insulated wire such as transformer wire is best
for this application.
5. To fixture the probe in place, use the accessory probe holder. See
page 7.
6. Reconnect the interrupted circuit keeping the lead length as short
as possible. To avoid applying too much heat with the soldering
iron, do these steps:
a. Solder one end of the wire to the positive (+) polarity side
the interrupted circuit under test.
b. Insert the wire into the current probe.
c. Solder the other end of the wire on the polarity dot side of
current probe. Use just enough heat to temporarily tack the
wire in place.
7. Power on the circuit under test and measure or inject the signal.
Stay within the ratings of the probe.
For information on the ratings of the probe and on how to optimize
performance, refer to Operating Basics on page 9 and Specifications
on page 21.
CT-6 Instruction Manual
5
Observe polarity of circuit, use short
wire, keep probe head close to the
circuit board
Polarity dot
Position probe head
to optimize signal
fidelity
Figure 2: Using a single wire
CT-6 Instruction Manual
Transformer wire, 26
gauge or smaller
idiff i2 i1
Figure 3: Using two wires to measure differential current
Polarity dot
Jumper
NOTE: This type of connection is not
suitable for frequencies above 1 GHz.
Figure 4: Using a square pin and jumper
CT-6 Instruction Manual
7
Probe Holder
To help stabilize the probe, you can use the probe holder provided to
secure the probe to a test fixture or prototype circuit board. See
Figure 5.
Probe holder
Horizontal
Vertical
Figure 5: Securing the CT-6 probe head to a test fixture board
To mount the probe holder, do these steps:
1. Temporarily place the probe and probe holder in the position you
intend to mount them.
2. Use the probe holder to mark holes on the test fixture or
prototype board.
3. Drill holes that are the appropriate diameter for the mounting
hardware you intend to use. The 2-56 screws and probe holder
supplied with the current probe are designed to mount in plated
CT-6 Instruction Manual
circuit board holes that are 0.080 inches in diameter and 0.360
inches center to center.
Evaluation Board (Optional)
Prepare the optional evaluation board for use by installing the five
rubber feet with the screws provided; install one foot at each corner
of the board and one in the middle.
CT-6 Instruction Manual
9
Operating Basics
For accurate measurements, it is important to understand the
electrical characteristics of the CT-6 current probe and how these
characteristics affect the way you use the probe. The following topics
are discussed in this section:
Interpreting Current and Voltage Ratings
Matching Impedances
Measuring Current
Injecting Signals
Optimizing Signal Fidelity at High Frequencies
Making Ground Side and Signal Side Measurements
Using the Test Fixture Board
Interpreting the Current and Voltage Ratings
Refer to the specifications on page 21 for the current ratings and
safety ratings of the probe. Definitions of the current ratings are as
follows:
Maximum continuous current refers to the maximum RMS
current that the current probe can continuously measure.
Maximum pulsed current refers to the maximum peak value of a
pulse the current probe can measure, regardless of the pulse
duration or frequency.
Ampere-second product refers to the maximum allowable pulse
duration for a given amplitude. For example, with a rating of 0.25
ampere-microsecond, a 250 mA peak pulse is allowed for 1 s
and a 125 mA peak pulse is allowed for 2 s.
CT-6 Instruction Manual
Effect of DC current refers to the amount of DC current that can
pass through the current probe before the low frequency cutoff
point doubles. High amounts of DC current also degrade the
ampere-second product for low frequency signals.
Stay within the safety rating for maximum bare-wire voltage
regardless of the amount of current you are measuring. Note the
following warning:
WARNING. To avoid personal injury or fire, do not exceed the
maximum bare wire voltage of 30 V. Avoid using the probe in circuits
where voltages exceed 30 V.
Maintain adequate spacing between the housing on the probe head
and other exposed circuit connections. The hole in the probe head is
insulated, but the painted housing on the probe head is at ground
potential.
Matching Impedances
The probe is designed to work with 50 impedance on the input of
the measurement instrument (or output of a signal generator) as well
as in the circuit under test. Even though the current probe can
measure signals from (or inject signals to) circuits with varying
impedances, the ideal impedance match is 50 . Improperly matched
impedances can cause aberrations and affect the probe sensitivity.
CT-6 Instruction Manual
11
Measuring Current
Electrically, the current probe acts like a transformer. When
measuring the AC current through a single conductor, the conductor
acts as the primary winding and the current probe acts as the
secondary winding. See Figure 6.
Circuit
under test
(primary)
Current
probe
(secondary)
To measurement
instrument
Signal from circuit
under test
50
50
(input Z)
Figure 6: Signal measurement
The probe converts current in the circuit under test (primary) to a
measurable voltage on the output (secondary). With matching 50
impedances, the sensitivity of the probe is 5 mV/mA.
To calculate the current in the circuit, divide the units of voltage
measured on the probe output by 5. For example, if you measure a
50 MHz signal that is 200 mVP-P, the actual current in the circuit is
200 5 = 40 mAP-P.
CT-6 Instruction Manual
Injecting Current Signals
When injecting AC current into a single conductor, the current probe
acts as the primary winding and the conductor acts as the secondary
winding of a transformer. See Figure 7.
To circuit
under test
Circuit under
test
(secondary)
Current
probe
(primary)
Signal from
generator
50
50
Figure 7: Signal injection
The probe converts a voltage signal (primary) to a current signal in
the circuit under test (secondary). With matching 50 impedances,
the sensitivity of the probe is 5 mV/mA.
To calculate how much voltage to apply to the probe to inject a
desired amount of current, multiple the units of desired current by 5.
For example, if you want to inject a 50 MHz signal that is 30 mAP-P,
the amount of voltage to apply is 30 × 5 = 150 mVP-P.
WARNING. To avoid personal injury, use the probe on grounded
systems only. To avoid RF burn hazard, apply stimulus signal only
after connecting the probe to the circuit.
CT-6 Instruction Manual
13
Optimizing Performance at High Frequencies
The CT-6 current probe preserves the fidelity of the signal without
the effects caused by the resistive elements of conventional passive
probes. This is because the current probe couples signals magnetical-
ly rather than electrically to the measurement instrument or from the
signal source.
At the high end of the current probe’s frequency range, however, the
added inductance of the current probe and the stray inductance and
capacitance of the coupling may, in some cases, affect the fidelity of
the signal.
The information and techniques presented in this section will help
you to understand and minimize inductive and capacitive effects at
high frequencies and optimize the performance of the probe.
Circuit Equivalent
The series equivalent of the current probe when placed in the circuit
with a 27 gauge wire is shown in Figure 8.
1
1 H
1.5 nH
Figure 8: Circuit equivalent of added inductance and capacitance
The higher the frequency of signal you are measuring, the more you
may want to consider this added inductance and capacitance and any
effect it has on the circuit under test.
CT-6 Instruction Manual
Selecting the Wire Size and Length
The gauge and length of wire used to route the test current through
the probe may affect the amount of stray capacitance and inductance.
The longer the wire, the more stray inductance there is and the
greater the impedance the probe adds to the circuit at higher
frequencies. Because of this effect, it is best to use the shortest
wire possible. A 1 cm long wire is ideal.
The larger the gauge of wire, the more stray capacitance there is
between the wire and the current probe. Stray capacitance affects
the circuit at high frequencies and adds aberrations to high-fre-
quency signals. Because of these effects, it is best to use the
smallest diameter wire possible. A 20-gauge bare wire is the
largest wire that the 0.032-inch opening in the probe can
accommodate.
Positioning the Probe
There is also stray capacitance associated with the case that
surrounds the head of the current probe when there is a relatively
large diameter wire in the probe (approximately 20 to 25 gauge).
When measuring high-frequency signals with larger diameter wire, it
is generally best to position the probe head perpendicular to the
circuit board under test. See Figure 9.
CT-6 Instruction Manual
15
Figure 9: Positioning for best HF response (20–25 gauge)
CT-6 Instruction Manual
Ground Side and Signal Side Measurements
For the most accurate measurements, couple the probe to either the
signal side of a 50 load or the ground side of a low impedance
load.
For the best impedance match with the probe output, take current
measurements on the signal side of the 50 circuit load.
If coupling to the signal side of a 50 load is impractical, take
current measurements on the ground side of a load that has a low
enough impedance to allow measurable signal current to pass
through the probe.
Figure 10 shows the signal and ground side points.
Signal Side
Ground Side
Current
probe
50
Current
probe
50
Figure 10: Coupling the probe to signal side or ground side
The optional evaluation board has separate signal side and ground
side connections to help assess the effects of these two techniques on
signal measurements.
CT-6 Instruction Manual
17
Using the Optional Evaluation Board
The evaluation board for the CT-6 current probe (Figure 11) is
available as a purchase option or optional accessory. This fixture
simplifies testing the probe for purposes of characterization or
verification.
Figure 11: Test Fixture board for the CT-6 current probe
CT-6 Instruction Manual
Calibration Lines
The Short, Load, and Feed Thru calibration lines (and Open Cal
evaluation line) are used in setting up the network analyzer used in
the procedures for Performance Verification on page 25.
Evaluation Lines
Evaluation lines enable you to connect and measure your test signals.
The fixture helps to isolate and characterize probe performance and
signal characteristics for open, ground side, and signal side
connections.
System Bandwidth Line (or Open Cal). The system bandwidth line
allows you to test the system bandwidth and to isolate and evaluate
the fidelity of the test signal through the particular square pins or
wire jumper you intend use.
Without a wire jumper or set of pins in place, you can use the line for
the “open” calibration of a network analyzer.
Ground side. The ground side line is for evaluating the performance of
the current probe with the test signal on the ground side of a 50
load. (See Figure 10 on page 16.)
Signal side. The signal side line is for evaluating the performance of
the current probe with the test signal on the signal side of a 50
load. Because the signal side technique is commonly used, additional
lines are provided in case the circuit board pads are damaged by the
repetitive soldering and unsoldering of signal jumpers.
To prepare a signal side line for use, unsolder the connector located
on the edge of the board and solder it to the pads at the next location.
Without a wire jumper in place, you can use a signal side line for the
“open” calibration of a network analyzer.
Probe Holder
For the ground side and signal side lines, the probe holder is included
to stabilize the probe head and increase the repeatability of
measurements. Figure 12 demonstrates how to use a probe holder to
secure the probe in two different ways. The holder uses two 2-56
screws in .080 inch diameter holes spaced 0.360 inches apart center
to center.
CT-6 Instruction Manual
19
Probe holder
Horizontal
Vertical
Figure 12: Securing the CT-6 probe head to the test fixture board
CT-6 Instruction Manual
21
Specifications
Tables 1 and 2 list the specifications for the CT-6 current probe. All
specifications are guaranteed unless noted as “typical.” Typical
specifications are provided for your convenience but are not
guaranteed. Specifications marked with the symbol have
corresponding checks in the Performance Verification section on
page 25.
Table 1: General specifications
Maximum bare wire voltage
30 V
Maximum continuous current (RMS)
120 mA
Maximum peak pulse current
6 A
Effect of direct current, typical
200 mA
Amp-second product, typical
0.25 amp-s into 50
Sensitivity
5 mV/mA ± 3% into 50 load at 50 MHz
(–20 dB ± 0.27 dB)
Frequency response
250 kHz to 1.5 GHz
Frequency response, typical
250 kHz to 2 GHz (See Figure 13)
Rise time, calculated
< 233 ps, 10% to 90%
Rise time, calculated typical
< 175 ps, 10% to 90%
Aberrations, typical
< 6% for 5 ns
Group delay, typical
5.2 ns for > 10 MHz (See Figure 14)
L/R time constant (droop), typical
< 0.4 s
Insertion impedance, typical
1.1 at 10 MHz1
1.3 at 100 MHz1
Characteristic line impedance of optional
evaluation board
50 ± 5%
1
Requires matching 50 termination on probe input and output
CT-6 Instruction Manual
dB
–20
100k
1M
10M
100M
1G
Hz
–23
–26
–29
–32
–35
–38
–41
–17
Figure 13: Typical frequency response (Bode plot)
10
100k
1M
10M
100M
1G
Hz
20
30
40
50
0
–10
–20
ns
Figure 14: Typical group delay (d/df)
CT-6 Instruction Manual
23
Table 2: Environmental and mechanical specifications
Temperature
Operating: 0C to +50C
Non-operating: –55C to +75C
Humidity
Operating: 0C to +50C at 0 to 90%
relative humidity
Non-operating: 0C to +60C at 0 to 90%
relative humidity
Altitude
Operating: 4,572 m
Non-operating: 15,240 m
Weight
17 g, stand alone
322 g , when packaged for domestic
shipment
381 g, when packaged for domestic
shipment with Option TF
Probe head (case) dimensions
Height: 4.3 mm
Width: 4.3 mm
Depth: 17.8 mm
Cable length
1 m
CT-6 Instruction Manual
25
Performance Verification
Use the procedures in this section to verify the performance of the
CT-6 current probe to the following warranted specifications:
Sensitivity
Frequency Response
Before beginning these procedures, photocopy the test record on
page 29 (or page 33 if using a network analyzer) and use it to record
test results. The recommended verification interval is one year.
Equipment Required
Table 3 lists the equipment required to perform these procedures.
The types and quantities of connectors may vary depending on the
specific equipment you use. For an alternate procedure that uses a
network analyzer, see page 30.
Table 3: Test equipment
Description
Minimum requirements
Example product
Digital sampling
oscilloscope
> 8 GHz bandwidth
Tektronix
11801B with SD24
sampling head
Signal Generator
250 kHz to 2.1 GHz
Marconi 2024
Gigatronics 6062A
SMA cable
3.5 mm, 1 meter
Tektronix
012-0649-00
Bare wire
20 gauge, 1 cm long
As per description
Evaluation Board with
probe holder
–
Tektronix
067-0349-001
1
The evaluation board comes with the current probe when ordered
as Option TF.
CT-6 Instruction Manual
Preparation
Turn on all test equipment and allow a 20-minute warm up period.
Mount the Current Probe on the Evaluation Board
1. Solder one end of a 1 cm, 20 gauge bare wire to a signal side line
on the evaluation board.
2. Slide the probe holder over the end of the current probe.
3. Turn the current probe so the polarity dot faces the circuit load.
4. Insert the bare wire through the hole in the current probe.
5. Secure the probe holder to the board with two 2–56 screws.
6. Carefully solder the bare wire to the evaluation board.
Face Polarity
dot to this
side
Signal source
Load
Figure 15: Securing the current probe to the evaluation board
CT-6 Instruction Manual
27
Frequency Response and Sensitivity
Use the following procedures to verify performance with a digital
oscilloscope and signal generator.
Set up the Signal Generator with the Digital Oscilloscope
1. Connect the output of the signal generator directly to the signal
input of the sampling head on the digital oscilloscope with an
SMA cable.
2. Connect a trigger signal to the oscilloscope from the generator.
3. Set the output frequency to 50 MHz and the output amplitude to
0.0 dBm.
4. Set time per division to 5 ns and volts per division to 100 mV per
division and trigger on the signal or use the AutoSet function to
automatically set up the oscilloscope.
5. Measure the the output of the signal generator on the oscilloscope
and save or record the value.
Measure Output and Calculate Sensitivity
1. Connect the SMA output of the CT-6 current probe to the signal
input of the digital oscilloscope.
2. Connect the output of the signal generator to the SMA connector
on the evaluation board.
3. Set time per division to 5 ns and volts per division to 10 mV per
division and turn averaging on.
4. Measure the probe output and calculate the sensitivity:
Sensitivity at 50 MHz = 20 log (Probe Output/ Probe Input)
Measure Low End Frequency Response
1. Connect the output of the signal generator directly to the signal
input of the digital oscilloscope with an SMA cable.
2. Set the output frequency to 250 kHz and the output amplitude to
0.0 dBm.
CT-6 Instruction Manual
3. Set the oscilloscope attenuation to 100 mV/div and time base to
1 s/div (or press AutoSet).
4. Measure the the output of the signal generator on the oscilloscope
and save or record the value.
5. Connect the SMA output of the CT-6 current probe to the signal
input of the digital oscilloscope.
6. Connect the output of the signal generator to the SMA connector
on the evaluation board.
7. Measure the probe output and calculate the sensitivity:
Sensitivity at 250 kHz = 20 log (Probe Output/ Probe Input)
8. Calculate the frequency response at 250 kHz and record results
on the test record:
(sensitivity at 50 MHz) – (sensitivity at 250 kHz) ≥–3 dB.
Measure High End Frequency Response
1. Connect the output of the signal generator directly to the signal
input of the digital oscilloscope with an SMA cable.
2. Set the output frequency to 1.5 GHz and the output amplitude to
0.00 dB.
3. Set the oscilloscope attenuation to 100 mV/div and time base to
200 ps/div (or press AutoSet).
4. Measure the the output of the signal generator on the oscilloscope
and save or record the value.
5. Connect the SMA output of the CT-6 current probe to the signal
input of the digital oscilloscope.
6. Connect the output of the signal generator to the SMA connector
on the evaluation board.
7. Measure the probe output and calculate the sensitivity:
Sensitivity at 1.5 GHz = 20 log (Probe Output/ Probe Input)
8. Calculate the frequency response at 1.5 GHz and record results
on the test record:
(sensitivity at 50 MHz) – (sensitivity at 1.5 GHz) ≥–3 dB.
CT-6 Instruction Manual
29
CT-6 Test Record
Identification
Number
Procedure performed by
Date
Tests
Low limit
Test
Result
High limit
Sensitivity @ 50 MHz
–20.27 dB
–19.73 dB
Frequency
res
se
@250 kHz
–3 dB
N/A
response
@1.5 GHz
–3 dB
N/A
CT-6 Instruction Manual
Frequency Response and Sensitivity – Alternate Procedure
To verify the performance of the probe using a network analyzer, use
the alternate procedure in this section.
Not all network analyzers are identical. This procedure uses the
network analyzer identified in Table 4. Adapt this procedure as
necessary to suit your particular analyzer. The types and quantities of
connectors may vary depending on the specific equipment you use.
Table 4: Test equipment required for alternative procedure
Description
Minimum requirements
Example product
Network analyzer
100 kHz to 3 GHz range
Hewlett-Packard
HP8753D with female
SMA connectors
SMA cables (2)
3.5 mm, 1 meter
Tektronix
012-0649-00
SMA–SMA barrel
adapter
Female-female, 3.5 mm
Tektronix
015-1012-00
Bare wire
20 gauge, 1 cm long
As per description
Evaluation Board
with probe holder
–
Tektronix
067-0349-001
1
The evaluation board comes with the current probe when ordered
as Option TF.
Set up the Network Analyzer
1. Preset the network analyzer to the factory default setting.
2. Set frequency range (100 kHz to 3 GHz minimum span).
3. Set frequency to log scale if desired.
4. Set for 3.5 mm diameter cables and select full two-port operation.
5. Connect one SMA cable to Port 1 of the network analyzer and the
other cable to Port 2.
CT-6 Instruction Manual
31
Calibrate Network Analyzer with Evaluation Board
1. Connect the SMA cable from Port 1 to the open line. Calibrate
forward reflection with Port 1 connected to the open, short, and
load lines in turn.
2. Connect the SMA cable from Port 2 to the open line. Calibrate
reverse reflection with Port 2 connected to the open, short, and
load lines in turn.
3. Compute list of calibration constants.
Calibrate out Line Losses
NOTE. Since the feedthrough line is slightly longer than the line
extending the SMA cable from Port 1, this calibration will not be
exact; however, up to frequencies around 2 GHz, the difference is
negligible.
1. Calibrate forward and reverse transmission and forward and
reverse matching with Port 1 and Port 2 connected to the
Feedthrough line on the evaluation board.
2. Select Omit Isolation and Done Cal functions (or equivalent) to
complete the forward and reverse calibrations.
Remove Delay
1. Display the Smith chart.
2. Look at S11 parameter.
3. Calibrate the port extensions on the short and open lines for
Port 1, and then repeat for Port 2.
4. Connect Port 1 and Port 2 to the feedthrough line and check for a
flat line.
5. Move zero reference to about 3 divisions from the top graticule
line.
6. Save the file.
CT-6 Instruction Manual
Verify Frequency Response and Sensitivity
1. Connect Port 1 to a signal side connector on the evaluation board.
2. Connect the male SMA connector of the current probe to the
SMA cable on Port 2 of the network analyzer using the
Female-female SMA adapter.
3. Connect the current probe to a signal side line on the evaluation
with a probe holder and a 1 cm, 20 gauge bare wire. Install the
current probe so that it is perpendicular to the evaluation board.
4. Set reference value to –20 dB.
5. Make a Bode plot of the S21 parameter.
6. Place markers at the following points on the curve
a. –3 dB at the low end of the frequency range
b. 50 MHz
c. –3 db at the high end of the frequency range
7. Note the frequencies measured at the – 3 dB points at the high
and low frequencies and record them on the test record.
8. Record the sensitivity (in dB) at 50 MHz on the test record.
CT-6 Instruction Manual
33
CT-6 Test Record – Alternate Procedure
Identification
Number
Procedure performed by
Date
Tests
Low limit
Test Result
High limit
Frequency
res
se
low frequency
< 250 kHz
250 kHz
response
(–3 dB point)
high frequency
1.5 GHz
> 1.5 GHz
Sensitivity
(–dB)
@50 MHz
–20.27 dB
–19.73 dB