Pooltest 9 Bluetooth

® Operation Manual ZI INST 7009B EN Palintest ® all-guide s.com

Who We Are

“Palintest aim to be the best global provider of water and environmental analysis equipment that enable critical water quality decisions to be made with confidence.” Palintest develop, manufacture and support a comprehensive range of advanced water and environmental testing equipment. From our United Kingdom headquarters we export to over 100 countries across the globe. Our products are used in a wide variety of applications to protect the environment and safeguard public health. To find out more about Palintest please visit www.palintest.com or contact your local Palintest partner. Palintest is a registered trademark of Palintest Ltd. Bluetooth and Bluetooth SMART are registered trademarks of the Bluetooth Special Interest Group (SIG).

Pooltest 9 Layout

5 Pooltest 9 Interface 5 Info Panel 5 Power Supply 7 Replacing Batteries 7 Start-up Screen 7 Mode Screen 7 Choose a Test 7 System Mode 7 Check Standard Mode 7 Log 8 Operator ID 8 Sample ID 9 Bluetooth 10 USB Interface 11 Units 12 Dilution Factor 12 DPD Range 12 System Lock 12 Water Balance 12 Temperature 13 Language 13 Set Time/Set Date/Date Format 13 Time Out 13 Back Light 13 LCD Contrast 14 Version 14 Selecting Test Parameters 15 Dilution Factor 15 Blanking Pooltest 9 Bluetooth 15 Reading with the Pooltest 9 Bluetooth 16 Timer 17 Changing Result Units of Expression 17 Follow-on Tests 18 Check Standard Mode 23 Enter Check Standard Values 23 Check Standard Measurement 23 Photometric Testing Hints and Tips Troubleshooting Technical Specifications Spares and Consumables Palintest Aqua Pal App and the Palintest Portal 6 Introduction 4 Quick Start 5 1 2 Chapter Page Chapter Page 3 Taking a Sample 6 4 Starting up the Instrument 7 5 System Mode 8 Analysing Samples 15 7 Test Methods 19 8 Calibration/Validation 23 9 Appendix 1 24 10 Appendix 2 25 11 Appendix 3 28 12 Appendix 4 29 13 Appendix 5 30 V1-02/15

The Palintest Pooltest 9 Bluetooth is a direct-

reading, waterproof photometer for determining key water quality parameters for swimming pools, hydrotherapy pools, hot tubs and spas. Designed for both portable and static use, the Pooltest 9 Bluetooth should always be used with genuine Palintest reagents for optimal performance. The fundamental operating technique applied to the Pooltest 9 Bluetooth is based on the principles of optical absorbance and scattering of visible light. Optical absorbance techniques are based on the use of Palintest (spectro)photometric reagents, creating visible colours with specific analytes upon reaction.The intensity of colour produced is measured with the Pooltest 9 Bluetooth and the data compared to the stored calibration data to deliver the final result. Optical scattering techniques produce small particles to scatter the source beam, the amount of scatter providing a result for the concentration of parameter under test. For more information regarding the science behind both photometric and turbidimetric analysis technology please visit www.palintest.com/know. The Pooltest 9 Bluetooth is provided with programmed methods for a comprehensive range of water quality parameters. Upon choosing a test the instrument automatically selects the required parameters for accurate analysis including wavelength and reaction time. On completion of some tests optional follow-on tests are available plus results can be converted to alternative units of expression e.g. mg/l to ppm, N or NH3. The Pooltest 9 Bluetooth offers a choice of connectivity to download all or selected results via Bluetooth 4.0 wireless connectivity or USB connection. Bluetooth 4.0 (also known as Bluetooth SMART or Low Energy) connection allows seamless data exchange using the Palintest Aqua Pal app, available for iOS and Android devices. See Appendix 5 for more details. USB connection via the port located at the rear of the instrument provides a choice of either ‘Hard Disk’ mode or serial communication mode. See Section 6.5 for more details. The Pooltest 9 Bluetooth offers a choice of either mains power via the USB port or using three 1.5V ‘AA’ batteries (supplied). The Pooltest 9 Bluetooth is delivered with a two year warranty and a full range of service, calibration and technical support. Additional resources are available at www.palintest.com relating to both products and applications.

for test selection

and text entry Navigation Keys LCD screen with backlight On/Off Light Cap USB Port Adaptive Cuvette Holder (do not insert finger!) 5 2 Quick Start 2.0 Quick Start 2.1 Pooltest 9 Bluetooth Layout 2.2 Pooltest 9 Interface The LCD screen features a selectable backlight with the screen separated into four clear, easy to read zones. 1 Mode or Test Identification. 2 Dialogue screen - prompts and choices will be displayed as a list. Select using up/down arrows. 3 Info Panel - displays status icons, date/time and Sample/Operator ID. 4 Action select - choices are displayed as a row. Use left/right arrows to select. 2.3 Info Panel Icons 1 2 3 4 Description Battery status Bluetooth connected Bluetooth on, not connected Upper/lower case text/number entry USB connected Hard Drive/COM port mode Icons

Taking a Sample

6 3.0 Taking a Sample The first critical step in any analysis is taking a representative sample.When selecting a sample point a number of care points are recommended as follows: • Ensure the sample point is safe to access and follow all relevant/required precautions • When sampling from a tap or outlet, remove any attachment and clean the tap/outlet with a dry cloth before allowing to run for 1 minute prior to sample collection • When sampling from a pool or spa take the sample as near as possible to the centre of the pool and at least 20cm below the surface of the water • Rinse any sample container repeatedly with the sample to prevent any cross contamination from previous samples • Once collected the sample must be processed immediately or as quickly as possible, especially for highly reactive species such as chlorine for example. The use of a portable field test kit makes this possible. However if the delay between sample collection and analysis is likely to be several hours chill the sample to preserve and prevent potential sample deterioration • Samples containing solid particles can interfere with photometric analysis. Either allow solids to settle and decant the clear liquid or filter the sample prior to analysis all-guide s.com

Starting up the Instrument

4.0 Starting up the Instrument 4.1 Power Supply The Pooltest 9 Bluetooth is designed to be powered either from alkaline batteries or via the USB port. When operating on battery power, the battery level is indicated on the Info Panel.A minimum voltage of 3.0V is needed to operate the photometer and a flashing battery symbol indicates a critically low battery. Change batteries immediately or switch to alternative USB power.The Pooltest 9 Bluetooth will automatically power down when power is no longer capable of providing acceptable performance. To power via the USB port, use the supplied cable connected either to the mains adaptor or a PC.The USB icon will appear when the USB connection is made and battery power will no longer be consumed. The Pooltest 9 Bluetooth has a back-up battery mounted internally to save instrument settings and data during power loss and instrument idle periods. 4.2 Replacing Batteries The battery compartment is located on the base of the instrument and secured by four screws. Remove the cover and install a complete set of new batteries, observing the correct polarity as indicated. Use 3 x 1.5V ‘AA’ alkaline batteries or equivalent. See Appendix 2 for more details. To avoid corrosion damage through leakage, remove batteries from the instrument if it is to be stored or left unused for a long period. 4.3 Start-up Screen The default start screen on power up is the ‘Choose a Test’ screen. To access the Mode menu press the left arrow key to highlight ‘Back’ and press OK on the keypad. To choose a test use the up/down arrows to scroll through the list and press OK on the desired parameter or use the direct access. 4.4 Mode Screen The Pooltest 9 has three operating modes as follows: Choose a Test The Choose a Test mode is the standard operating mode for taking photometer readings and is the default start-up screen on power up. See Section 8 for more information. System Mode Personalise your Pooltest 9 Bluetooth and manage stored data within the System mode. Options include setting Operator/Sample IDs, interrogating the result log and defining the instrument operating conditions. See Section 6 for further information. Check Standard Mode Validate performance of your Pooltest 9 Bluetooth using Palintest Check Standards. See Section 9 for more information.

5.0 System Mode

Personalise your Pooltest 9 Bluetooth and access the data log via the System Mode. Scroll up or down using the appropriate keys to see all available options. The options and available settings are as follows: 5.1 Log The Pooltest 9 Bluetooth has an internal data log for up to 100 data points.The data is stored automatically upon completion of the test and automatically overwrites the oldest result when the memory is full. The data log is unaffected by power on/off. Each data point consists of the following information and is stored in a comma-separated values (CSV) format: Selecting Log offers two choices: View - to view individual data points use the up/down keys. Data is stored in chronological order with the most recent result shown by default. Scroll through results using the up/down arrows. Select Back to return to the previous menu. Clear - the entire log can be deleted from the Pooltest 9 Bluetooth if the instrument is not locked (see System Lock). Selecting Clear produces the following screen: Choose Yes to delete all data, No to return to the previous screen. Data can be downloaded via either Bluetooth (see Bluetooth Log Transfer) or USB connection (see USB Interface). 5.2 Operator ID The Pooltest 9 Bluetooth offers the option to create up to 12 unique alphanumeric Operator IDs. Operator IDs are added to the result data automatically but deleting IDs does not affect the result log. To create a new Operator ID, select Operator ID and use the up/down keys to select a blank field. Select New and press OK. Alphanumeric characters are entered/edited using the 0-9 keys or the up/down keys. Press and hold the 1 key to toggle between upper case, lower case and numeric characters.

After entering a character, the cursor automatically

moves to the next position if no key is pressed. Alternatively press the right key. Up to 10 characters can be added for Operator IDs, including spaces. To edit characters use the left/right keys to select the desired character. Press and hold the left key to delete the character or change the character using the entry mode. When the Operator ID is correct press the OK key to create the ID and return to the Operator ID list.The new Operator ID will be displayed in the Operator list. Choose the Operator ID to be used by scrolling through the list and pressing the OK key on the desired choice. The instrument will return to the System menu. To modify or delete an existing Operator ID, highlight the ID and select Edit. Choose either Edit to modify the existing entry or Delete to remove it from the list. 5.3 Sample ID The Pooltest 9 Bluetooth offers the option to create up to 24 unique alphanumeric Sample IDs. Sample IDs are added to the result data automatically but deleting IDs does not affect the result log. To create a new Sample ID, select Sample ID and use the up/down keys to select a blank field. Select New and press OK. Alphanumeric characters are entered/edited using the 0-9 keys or the up/down keys. Press and hold the 1 key to toggle between upper and lower case characters. After entering a character, the cursor automatically moves to the next position if no key is pressed. Alternatively press the right key. Up to 10 characters can be added for Sample IDs, including spaces. To edit characters use the left/right keys to select the desired character. Press and hold the left key to delete the character or change the character using the entry mode.

When the Sample ID is correct press the OK key to

create the ID and return to the Sample ID list.The new Sample ID will be displayed in the Sample list. Choose the Sample ID to be used by scrolling through the list and pressing the OK key on the desired choice. The instrument will return to the System menu. To modify or delete an existing Sample ID, highlight the ID and select Edit. Choose either Edit to modify the existing entry or Delete to remove it from the list. 5.4 Bluetooth The Pooltest 9 Bluetooth features the latest Bluetooth 4.0 (also known as Bluetooth Low Energy or Bluetooth SMART) for wireless communication with external devices. The Palintest Aqua Pal app provides seamless data exchange with the Pooltest 9 Bluetooth, data trend analysis and user-defined action limits for key parameters. Additional data management functionality is provided by the Palintest Portal (www.palintestportal.com). Uploaded data can be shared with colleagues and customers within your User Group(s) and integrated into customised reports. See Appendix 5 for more information on the Aqua Pal app and the Palintest Portal. There are four options available in the Bluetooth menu: • Bluetooth Communications ON - activate the Bluetooth and make the Pooltest 9 Bluetooth visible for connection/pairing. • Bluetooth Communications OFF • Bluetooth LogTransfer - transfer historical log data to theAqua Pal app when connected to a remote device • Bluetooth Device ID - create a unique device name for the Pooltest 9 Bluetooth to discriminate between multiple connections Bluetooth Communication On Select this option to enable Bluetooth communications allowing the instrument to be paired with a suitable Bluetooth SMART enabled device. Visit www.palintest.com\know for more information regarding Bluetooth SMART devices. The Bluetooth icon is shown in the Info Panel when Bluetooth is enabled. Connection status is shown as follows: indicates Bluetooth is activated and connected to an external device indicates Bluetooth is activated but the Pooltest 9 Bluetooth is not connected to an external device Bluetooth Communication Off Selecting this option disables the Bluetooth communications module. The Bluetooth icon is not visible on the Info panel. Bluetooth Log Transfer Selecting this option transfers all or a selected group of results stored in the log to the paired mobile device. The Pooltest 9 Bluetooth will validate the paired connection and confirm readiness to transfer. Transferring selected data will require specification of the result log window e.g. from result 40 to result 100 to be transferred selectively. Press OK to transfer the data log. The data will transfer in series. Each data point is validated by the Aqua Pal app prior to upload of the next.

If the connection is lost the Pooltest 9 Bluetooth

will prompt for re-connection. If connection is not required or possible press Exit to disable Bluetooth and cancel the log transfer. The message ‘Log transfer is complete’ will be shown when all data points have been uploaded successfully. Press OK to return to the previous menu. Bluetooth Device ID A number of Pooltest 9 Bluetooth instruments may be available to connect to a remote device, although only one active connection is possible at any time. A user-defined Bluetooth Device ID ensures simple pairing between the desired Pooltest 9 Bluetooth and the Palintest Aqua Pal app. Creating and/or editing Bluetooth Device ID is identical to Operator and Sample ID creation. 5.5 USB Interface The waterproof USB interface provides both communication between the Pooltest 9 Bluetooth and a PC and an alternative mains power source via the adaptor. When connected the USB icon will appear, replacing the battery icon in the Info Panel, as power will be preferentially drawn from the external source. The USB data interface has a choice of two operating modes - Hard Drive and COM Port.The current status of the USB connection is shown on the Info panel when the USB lead is connected. Toggle between COM Port and Hard Disk mode in the System -> USB menu by selecting the desired option. The USB connection supports software update and data download through a simple ‘drag and drop’ approach when operated in Hard Disk mode. Hard Drive The instrument appears as a removable hard drive when connected to a PC in Hard Disk mode. Upon connection the remote drive will have the following files included: • Pooltest 9_***.afx.*** where * represent version numbers of software - this is the operating software for the Pooltest 9 Bluetooth • Log.txt - the data log file stored in a comma separated value (csv) format Operating software or calibration library can be updated by dragging a new version to the instrument - contact [email protected] for new software if this option is required.Any updates to operating software will be notified via the www.palintest.com\know portal. Downloading the result log is carried out by dragging the LOG.txt to the local desktop and opening the file with any program that can open CSV format files. For more information regarding extracting and opening result logs using the Hard Disk mode visit www.palintest.com/know. COM Port The instrument behaves as if connected to the PC serial port via RS232 when connected in COM Port mode, allowing remote control from an external software system and data upload. 11 5 System Mode all-guide s.com

This allows backwards compatibility with software

written for earlier models of Palintest instruments. In this mode, the PC requires installation of a USB virtual COM Port driver, available from www.palintest.com/know, and the availability of software operating as a terminal server. A large number of third-party software systems are available to provide data upload and remote control of testing using the COM port mode. Please contact your local Palintest representative for more details. 5.6 Units The Pooltest 9 Bluetooth offers the choice of result expressed in mg/l, ppm, mmol/l, µmol, g/l and µg/l. Changing the result units will not affect the result log. 5.7 Dilution Factor When samples are above the test range, indicated by >> on the result screen, a dilution procedure can be used. Setting Dilution Factor to On will prompt the Pooltest 9 Bluetooth to automatically request the dilution factor when carrying out a test. Change the dilution factor by using the up/down arrows or manually type the dilution factor. The instrument will automatically correct the result for the dilution and display the corrected result (which will also be stored in the result log). The maximum dilution factor permitted is 99. If the calculated result exceeds the available number of permitted characters >> will be displayed. NOTE: do not use sample dilution when measuring pH or alkalinity. 5.8 DPD Range The Pooltest 9 Bluetooth operates in a choice of DPD ranges, depending on the pool treatment programme. Two choices are available and only one range can be used at any time: 0-5 the ‘standard range’ of 0-5mg/l (ppm) Cl2. In this range the AP 011 and AP 031 DPD reagent packs are suitable 0-10 the ‘extended range’ of 0-10mg/l (ppm) Cl2. In this range the AP 013 DPD XF and AP 033 DPD XF/XT packs will be required to carry out the testing. The selected range of DPD will be indicated in the Test List e.g.Test 001 Chlorine-Free/10 indicates the extended range has been selected and the DPD XF reagent should be used. 5.9 System Lock To prevent unauthorised or inadvertent changes to the System settings or log deletion a four digit code can be used to lock several options. The default code is set to 6812. To change the System Lock code, follow the on-screen prompts to choose a memorable four digit number. When the System Lock is applied, the items accessible within the System Mode are limited until the unlock code is entered.Access is limited to viewing the result log, adjusting the backlight, contrast and Bluetooth settings. 5.10 Water Balance Two options are available for the Test 77 Water Balance scale: Palintest uses the simplified Palintest scale which assumes values for Total Dissolved Solids (TDS) and temperature and removes the need to measure these parameters. Langelier uses the results from a TDS/ temperature meter to calculate the Langelier Saturation Index (LSI) value. For more information, see both the test method on page 24 and visit www.palintest.com\know for more information regarding the importance of Water Balance and the relationship between the two scales.

Choose between ºC and ºF.The temperature setting

is used during the Langelier Saturation Index when selected from the Water Balance option. 5.12 Language Select the desired local language for operation, choosing from English, French, Spanish, German, Italian and Chinese (Mandarin).The selection of language will also adjust appropriate tests and units to local convention as required. 5.13 Set Time/Set Date/Date Format All test results are recorded automatically in the Data Log and appended with date and time (plus additional information). Date and time are stored on an internal clock, supported by a coin cell battery. To correct the time select SetTime from the System menu. Use the up/down keys to adjust the hour; press the right key to select and adjust the minutes. Press OK when the correct time is set. To correct the date select Set Date from the System menu. Use the up/down keys to adjust the day/month/year, using the left/right keys to select the field. Press OK when the correct date is set. The Date Format can be set to DD/MM/YYYY or MM/DD/YYYY as required.To select the required format highlight the desired choice and press OK. 5.14 Time Out When operating using battery power the Pooltest 9 Bluetooth provides automatic power-off as a power-saving measure.Three settings are provided: Normal 5 minutes Long 15 minutes Off (disables Time Out) The time intervals begin after the last key is pressed or activity takes place. Time Out is automatically disabled when the instrument is powered by USB supply and during a Bluetooth data log transfer. 5.15 Back Light The instrument display features a high intensity backlight to support use in low light conditions. The backlight is designed to use minimal energy but activating will naturally consume battery power more rapidly.The settings available are: Backlight Auto-Dim Backlight activates on any key press and dims after 15 seconds automatically. Backlight On Backlight is on permanently Backlight Off Backlight is off permanently

5.16 LCD Contrast

In addition to the Backlight, the default contrast setting for the display can be adjusted using the up/down keys when light conditions are difficult. The display provides a sequence of alternating squares to give visual indication of the correct settings to apply. When complete/acceptable press the OK key. 5.17 Version The serial number of the instrument and the software version are displayed.The instrument serial number will be required for technical support and servicing/warranty and can also be found on the case of the instrument.

6.0 Analysing Samples

The Pooltest 9 Bluetooth provides simple, accurate and reliable analysis of key pool, spa and other aquatic samples. Selecting the required parameter and performing the test are supported through on-screen prompts and comprehensive test instructions. The principle of photometric testing is based on the absorption or scattering of a measured intensity of incident light compared to the light intensity reaching the detector array. The light intensity is determined as Transmittance (%T) or Absorbance (A) and compared to calibration tables stored within the Pooltest 9 Bluetooth. The stored calibration tables convert %T or A to results in a variety of units (mg/l, ppm etc.) as defined in Section 6.6. Calibration tables are defined by Palintest based on the measurement of reference standards using Palintest reagents. To achieve the best quality results there are a small number of care points: 1 Always use the provided light cap to prevent ambient light affecting the results. 2 Ensure Sample and Blank cuvettes are clean, dry and inserted correctly into the sample chamber, using the allocated orientation mark to align. 3 Always blank the instrument with untreated sample prior to analysis. Additional guidance is provided in Appendix 1 Photometric Testing Hints and Tips 6.1 Selecting Test Parameters The Pooltest 9 Bluetooth offers a number of choices to select the parameter to test: Test Number Entry - use the numeric keypad to enter the unique Test Number to directly access any programmed calibration. Direct Entry - use the numeric keypad to directly access the 10 most common tests with a single key push. Choose a Test - the full list of test parameters is available by selecting Choose a Test and scrolling using the up/down keys.When the desired parameter is highlighted, press the OK key to access the method.Tests are presented in Test Number order. When the required test is selected the Pooltest 9 Bluetooth automatically selects the correct wavelength and sets additional method parameters as required. Test method protocols are defined in detail section 8, including the reagents and accessories that may be required. When a test is selected, the Pooltest 9 Bluetooth will display a number of screens and options to guide the user through the testing process, as described in the following pages. 6.2 Dilution Factor If selected in the System menu, the initial screen will request the defined Dilution Factor to apply to results. If Dilution Factor is not active this screen will not be shown. Results shown on the final screen have automatically been corrected for dilution prior to display. Corrected results will also be stored in the log. Note: if the System Lock is applied the Dilution Factor functionality cannot be switched on or off from the current status until the system is unlocked. 6.3 Blanking the Pooltest 9 Bluetooth Blanking the photometer is a key first step in photometric analysis, effectively removing the potential entrained sample colour and minor amounts of turbidity from calculation of analytical results. To blank the photometer, prepare a Blank cuvette using untreated sample i.e. sample that has not been reacted with any reagents. If the sample is to be diluted or physically treated (filtered for example) before analysis, use the same dilution/ treatment for the Blank cuvette.

When accessing the test method the Pooltest 9

Bluetooth will request the user to Insert Blank. Insert the Blank cuvette and press OK. The Pooltest 9 Bluetooth will determine the absorbance due to the sample colour at all wavelengths simultaneously and store in temporary memory for use in analysis. Upon successful blanking the Pooltest 9 Bluetooth will automatically move to the Insert sample stage of the analytical method. If the sample is too highly coloured to support effective blanking and subsequent analysis the error message “Error 9 is caused by the blank cuvette being too dark. Check the correct cuvette is being used”. Ensure the blank cuvette is being used, not the sample plus reagent cuvette. Sample colour can be reduced by dilution with clean water; the dilution selected should take account of the expected concentration of parameter under test. Blank results are stored in the temporary memory of the Pooltest 9 Bluetooth and will be used for all subsequent tests until: • The instrument is powered down (temporary memory is lost/deleted) • A new Blank reading is taken - this option is available on accessing any subsequent test at the base of the screen. Repeat the blanking process if the sample changes significantly or a new sample is under test • Some tests use a reversed blanking process where a coloured blank cuvette may be required. When changing between standard tests and ‘reverse blank’ test, a new blank will be requested by the Pooltest 9 Bluetooth. If the blank value generated in this step is not detectable an error message will be displayed “Error 7 is caused by too much ambient light.Try using the light cover provided with the instrument”. See Appendix 1 Photometric Testing Hints and Tips for more advice regarding effective blanking. 6.4 Reading Results with the Pooltest 9 Bluetooth Assuming a suitable blank has been recorded, the next step of the photometric analysis process is to carry out the reading step. Prepare a Sample cuvette following the method instructions provided in Chapter 8. all-guide s.com

Select Read at the base of the screen and press OK.

At the Insert sample prompt, insert the sample cuvette ensuring it is clean and dry and oriented correctly using the location mark. Press OK to begin the measurement process. The screen will display Reading... At the completion of the measurement process the result is displayed on screen. 6.5 Timer Many photometric methods require a reaction time to develop, the recommended time period being documented in the Palintest Phot Book and included as part of the method parameters programmed into the Pooltest 9 Bluetooth. Tests requiring a reaction time will have the option to select an automatic timer to count down the reaction time required.While in the Insert sample screen use the right key to move the cursor to highlight Timer. The programmed reaction time will be displayed. Press OK to Start the countdown. Three options will be displayed: Stop cancel the countdown timer Exit exits the countdown screen and returns to the Insert sample screen. The countdown will continue and the current time can be seen by selecting Timer.At the end of the countdown an audible alarm will sound to indicate the sample is ready to read. Select OK to read the sample manually. Exit and Read exits the countdown screen and automatically reads the sample at the completion of the allocated time period. 6.6 Changing Result Units of Expression Many chemical species have a number of alternative units that can be used for reporting results e.g. Phosphate can be expressed as PO4 or P for example.

Where alternative units of expression for results

are available the LM symbols will be displayed next to the current result units. Use the up/down arrows to change the units of expression as required.Values are modified automatically. Results stored in the log will be in the units selected on screen, changing the chemical species will add an entry to the log showing the updated result and species parameter selected. 6.7 Follow-on Tests A number of photometric methods have additional optional methods that can be applied, known as Follow-onTests e.g. Phot 008Total Chlorine follows Phot 007 Free Chlorine. Follow-on methods are usually either based on further reagent addition to the sample just measured for sequential parameters or used to correct for potential/known interferences. Follow-on tests are clearly defined in the Section 8 and, if available, are accessed via the Follow-on option located at the right hand side of the options. To access the Follow-on test, highlight Follow-on and press OK. The next method is automatically loaded and operated in the usual manner. NOTE: if no viable result is produced during the first stage of a sequential test method, the Follow-on option will be automatically removed.

The Pooltest 9 Bluetooth provides simple test

protocols for the key parameters required to effectively control swimming pools, hydrotherapy pools, spas and hot tubs. The tests available by number are as follows: Test 1 Chlorine-Free/5 or Chlorine-Free/10 Standard Range: 0-5.00mg/l (ppm) Cl2 (use DPD 1) Extended Range: 0-10.0mg/l (ppm) Cl2 (use DPD XF) Colour change: Colourless - Purple/Red 1 Rinse test cuvette with sample leaving two or three drops in the tube. 2 Add one DPD 1 or DPD XF tablet, crush tablet and then fill the test tube with sample to the 10ml mark. Mix to dissolve tablet fully and ensure no particles remain. 3 Take photometer reading immediately. The result may drift on standing. 4 Retain test solution if the Total Chlorine Follow-On Test is required. Test 2 Chlorine-Total/5 or Chlorine-Total/10 Standard Range: 0-5.00mg/l (ppm) Cl2 (use DPD 3) Extended Range: 0-10.0mg/l (ppm) Cl2 (use DPD XT) Colour change: Purple/Red from Test 1 increases in intensity Carry out this test on the solution remaining from the Free-Chlorine/5 test. 1 If any shock treatment chemicals have been added to the pool, add one DPD Oxystop tablet, crush and mix to dissolve. Stand for one minute before proceeding. This will prevent a false response caused by the shock treatment chemicals. 2 Add one DPD 3 or DPD XT tablet, crush and mix to dissolve. 3 Stand for two minutes to allow full colour development. 4 Take photometer reading after two minutes have elapsed. Note: To obtain Combined Chlorine residual subtract Free Chlorine result from Total Chlorine result: Combined Chlorine = Total Chlorine - Free Chlorine 19 7 Test Methods 7.0 Test Methods Test Test Notes No. Name 0 Transmittance Select wavelength manually using up/down arrows 1 Chlorine-Free Select range from the System Mode 2 Chlorine-Total Select range from the System Mode 3 Bromine-Total 4 Ozone 5 Copper 6 pH-Phenol Red 7 Total Alkalinity 8 Calcium Hardness 9 Cyanuric Acid 10 Iron/1 88 Calc. Hard. Salt Use when testing Calcium Hardness in a salt pool 77 Water Balance Choose Palintest or Langelier from the System Mode

Test 3 Bromine

Range: 0-10.0mg/l (ppm) Br2 Colour change: Colourless - Purple/Red 1 Rinse test cuvette with sample leaving two to three drops in the tube. 2 Add one DPD 1 tablet, crush tablet and then fill with sample to the 10ml mark. Mix to dissolve tablet fully and ensure no particles remain. 3 Take photometer reading immediately. Result may drift on standing. Test 4 Ozone Range: 0-2.00mg/l (ppm) O3 Colour change: Colourless - Purple/Red 1 Rinse test cuvette with sample leaving two to three drops in the tube. 2 Add one DPD 4 tablet, crush tablet and then fill with sample to the 10ml mark. Mix to dissolve tablet fully and ensure no particles remain. 3 Take photometer reading immediately. Result may drift on standing. Note: The ozone test also responds to chlorine and bromine. Pools using ‘ozone with chlorine’ treatment systems, or other combined treatments will produce a result at this stage representing the ozone plus other treatment chemicals.To determine ozone specifically requires an additional protocol as follows: Correction for Chlorine or Bromine 1 Fill a test cuvette with sample to the 10ml mark. Add one DPD Glycine tablet, crush and mix to dissolve. 2 Take a second clean test tube and add two to three drops of solution from the first tube.Add one DPD 4 tablet, crush and then make up to the 10ml mark with the solution from Step 1. Mix to dissolve tablet fully. 3 Take Photometer reading immediately. The DPD Glycine complexes ozone in the sample and the result displayed equates to chlorine or bromine present in the sample only. 4 Subtract the result obtained in Step 3 from the initial result to determine the ozone concentration specifically. Test 5 Copper-Total/5 Range: 0-5.00mg/l (ppm) Cu Colour change: Colourless - Purple 1 Fill test cuvette with sample to the 10ml mark. 2 Add one Coppercol No 1 tablet and one Coppercol No 2 tablet, crush and mix to dissolve. 3 Take photometer reading. Test 6 pH-Phenol Red Range: pH 6.5-8.5 Colour change: Yellow - Red 1 Fill test cuvette with sample to the 10ml mark. 2 Add one Phenol Red tablet, crush and mix to dissolve completely. 3 Take photometer reading. Test 7 Total Alkalinity Range: 0-500mg/l (ppm) CaCO3 Colour change: Yellow - Green - Blue 1 Fill test cuvette with sample to the 10ml mark. 2 Add one Alkaphot™tablet, crush thoroughly and mix. Ensure all particles have dissolved completely. 3 Allow to stand for one minute. 4 Take photometer reading. Note: If a thin yellow layer is observed at the bottom of the tube, the cuvette should be thoroughly remixed before measurement.

Test 8 Calcium Hardness

Range: 0-500mg/l (ppm) CaCO3 Colour change: Violet - Orange 1 Fill test cuvette with sample to the 10ml mark. 2 Add one Calcicol No 1 tablet, crush and mix to dissolve. 3 Add one Calcicol No 2 tablet, crush and mix to dissolve. 4 Stand for two minutes. 5 Take photometer reading. Note: this method is calibrated for standard pools. For saltwater/seawater pools, please use Test 34 Calc.Hard.Salt. Test 9 Cyanuric Acid Range: 0-200mg/l (ppm) CNA Colour change: Clear - Cloudy (turbidimetric method) Note: Cyanuric acid is also commonly referred to as ‘Chlorine Stabiliser’ or ‘Pool Conditioner’. 1 Fill test cuvette with sample to the 10ml mark. 2 Add one Cyanuric Acid tablet. DO NOT CRUSH. Allow to disintegrate for at least two minutes. A cloudy solution indicates the presence of cyanuric acid. 3 Crush any remaining undissolved tablet, mix fully and take photometer reading. Test 10 Iron/1 Range: 0-1.00mg/l (ppm) Fe Colour change: Colourless - Pink 1 Fill test cuvette with sample to the 10ml mark. 2 Add one Iron LR tablet, crush and mix to dissolve. 3 Stand for one minute. 4 Take photometer reading. Test 88 Calcium Hardness (Saltwater Pool Calibration) Range: 0-500mg/l (ppm) Colour change: Violet - Orange 1 Fill test cuvette with sample to the 10ml mark. 2 Add one Calcicol No 1 tablet, crush and mix to dissolve. 3 Add one Calcicol No 2 tablet, crush and mix to dissolve. 4 Stand for two minutes. 5 Take photometer reading. all-guide s.com

Test 77 Water Balance

To determine Water Balance Index, select either Palintest or Langelier from the System Mode prior to commencing. In both Palintest and Langelier test methods, three parameters are measured using the Pooltest 9 Bluetooth in a defined sequence with the result for Water Balance displayed at the test conclusion. Alkalinity (Alkaphot), Calcium Hardness (Calcicol) and pH (Phenol Red) tests are required. Prepare the test solutions as described in the individual parameter test methods and proceed as follows: Palintest Water Balance 1 Select Test 77 by pressing the Alkalinity key (no. 7) twice or from the ‘Choose a Test’ list. 2 Carry out the Alkalinity test then select ‘Follow-On’ from the results screen. 3 Carry out the Calcium Hardness test then select ‘Follow-On’ from the results screen. 4 Carry out the pH test then select ‘Follow-on’ from the results screen. 5 The Water Balance Index is calculated and is displayed. See the table below for more details of the results guidance Note: the Palintest Water Balance test has been calculated for an average heated swimming pool (82°F, 28°C). For unheated swimming pools, subtract 0.1 from the index value obtained. For higher temperature pools and spas, add 0.1 to the index value obtained. Langelier Saturation Index Water Balance To complete the Langelier method, a separate Total Dissolved Solids (TDS) meter is required to provideTDS and temperature values which are entered manually. Prepare the test solutions and proceed as follows: 1 Select Test 77 by pressing the Alkalinity key (no. 7) twice or from the ‘Choose a Test’ list. 2 Carry out the Alkalinity test then select ‘Follow-On’ from the results screen. 3 Carry out the Calcium Hardness test then select ‘Follow-On’ from the results screen. 4 Carry out the pH test then select ‘Follow-on’ from the results screen. 5 Manually enter the TDS reading (in mg/l or ppm) taken using the numerical keys. Select ‘OK’ when the figure on screen is correct. 6 Manually enter the temperature using the numerical keys and select ‘OK’. 7 The Langelier Saturation Index is calculated and is displayed. Index <-1.5 Corrosive <9.6 Highly Corrosive Increase pH to 7.5 - 7.8. Increase Calcium Hardness to at least 50mg/l. Increase Total Alkalinity to 100mg/l or higher as necessary. Retest Water Balance. -0.6 to -1.5 Corrosive 9.6 to 10.5 Corrosive -0.1 to -0.5 Corrosive 10.6 to10.9 Acceptable Balance Retest water regularly. 0.0 Ideal Balance 11.0 to11.2 Ideal Balance No action required. 0.1 to 0.5 Scale Forming 11.3 to11.6 Acceptable Balance Retest water regularly. 0.6 to 1.5 Scale Forming 11.7 to12.6 Scale Forming >1.6 Scale Forming >12.7 Highly Scale Forming Decrease pH to 7.2 - 7.5. Decrease Total Alkalinity to 150mg/l or lower as necessary. Retest Water Balance. Water Balance Condition Condition Water Balance Condition Recommendation LANGELIER PALINTEST Note: Always keep the pH,Total Alkalinity and Calcium Hardness levels within the limits recommended in pool operating instructions and chemical suppliers’ recommendations. Seek specialist advice if corrosion or scale formation is still apparent even though the water is shown to be in a balanced condition. Water Balance Guidance

Your Pooltest 9 Bluetooth is delivered with a

calibration certificate validating the performance of the instrument as it leaves Palintest. We recommend annual service and calibration of all photometric instruments in normal use. The Pooltest 9 Bluetooth also includes an automatic routine to validate analytical performance using certified Palintest Check Standards.Accessed via the Mode menu, the Check Standard Mode provides a field method of ensuring your instrument is operating within defined specifications and also a troubleshooting method for unexpected results. Every Palintest Check Standards set is supplied with certified values expressed as %T (Transmission), derived from traceable reference materials.Acceptable tolerances are defined on the certificate and are automatically specified within the Pooltest 9 Bluetooth. 8.1 Check Standard Mode Access Check Standard Mode from the Mode screen. Highlight Check Standard Mode and press OK. Two choices are offered: Enter Check use the up/down keys to Standard Values adjust the displayed values to match the certificate Check Standard insert the Check Standards Measurement in the defined order to generate a validation report 8.2 Enter Check Standard Values Each standard has two values assigned, for two individual wavelengths. Use the up/down keys to adjust the values to match the certificate, following the order defined on the display. Press OK when the correct value is shown and the prompt will forward to the next value. Upon completion the message Check Standard Values assigned successfully will be displayed. Press OK to return to the Check Standard Mode menu. 8.3 Check Standard Measurement Follow the on-screen prompts to insert the Check Standards in the defined order. The Pooltest 9 Bluetooth will automatically measure the Transmittance at the required wavelength. Upon completion of the sequence the results are displayed on screen with pass or fail status. If the Check Standard Mode reports a failure, see Appendix 2 Troubleshooting for guidance or contact your local Palintest supplier.

Appendix 1 - Photometric Testing Hints and Tips

Photometric analysis is a very powerful technique, providing accurate analysis of a wide range of critical swimming pool, spa, hydrotherapy and hot tub parameters. A complete guide to the science behind photometric (also known as colorimetric) analysis can be found in the Know portal at www.palintest.com/know/ Palintest has focused on simplifying the test methods and equipment used for this technique but there are still a number of ways to ensure the results you generate are as accurate as possible: 1 Always use genuine Palintest reagents when using the programmed test methods. Each parameter has a unique calibration which has been generated using Palintest reagents. Alternative reagents may follow the same general methodology but can differ substantially in formulation and colour generated thereby rendering the calibration and results inaccurate. 2 Always correct for the blank value - any inherent colour in the sample (which may not be visible to the naked eye) will offset the result if the blank step is omitted. If the sample colour is too intense for the photometer to blank use dilution with deionised water to reduce the intensity. Remember to dilute the sample to the same extent for analysis. 3 Always respect the reaction time specified within the instructions. Some methods produce instant colour whereas others require a reaction time to reach full development. Taking a reading before the reaction time has elapsed may lead to low results. 4 The presence of solids, either large or in the form of turbidity, can adversely affect the quality of results by preventing incident light from reaching the detector.The blanking step can reduce the impact of turbidity interference but large solid particles must be removed prior to analysis. Solids can be removed by filtration prior to analysis or, if the solids are settleable and will not lie in the optical path, allowing them to settle in the photometer cuvette can be acceptable. 5 Calibration curves relate transmission/absorbance to concentration to provide result data but not all calibration ranges are linear. Frequently at higher concentrations the curve ‘flattens’ leading to higher potential variability in results. If greater accuracy is required than can be achieved on neat samples then dilution can be used to improve performance. 6 Ensure the photometer cuvette is clean, has no droplets on the outside and is not excessively scratched. Good technique is to wipe the outside surface of the cuvette prior to inserting into the optical chamber to prevent contamination of the optical system. 7 Always use good quality, genuine Palintest cuvettes. Use the orientation mark to ensure repeatable positioning of the cuvette. 8 Maintain the cleanliness of the optical chamber by only inserting clean cuvettes. If the chamber becomes fouled or sample is spilled the base can be removed for cleaning access. Clean the optical chamber with a soft cloth. Do not use abrasive chemicals or scouring agents. 9 Always use the light cap provided to prevent ambient light affecting results.This is especially relevant when operating in strong sunlight or other light conditions. 10 Ensure your Pooltest 9 Bluetooth is operating effectively by using Palintest Check Standards and the Check Standard Mode (see Section 9 Calibration/Validation) and having the photometer serviced and calibrated at regular intervals. Calibration is recommended at 12 month intervals for normal usage and can be provided by your local Palintest distributor. 24

Appendix 2 - Troubleshooting

The Pooltest 9 Bluetooth features self-diagnostic software and hardware to optimise performance and battery life.The Info panel indicates the status of the Pooltest 9 Bluetooth and any specific fault conditions are defined and displayed on screen. Optical Errors I have an Error 9 message Error 9 is caused by the blank cuvette being too dark to allow the blanking step to be carried out. Check that the correct cuvette is being used i.e. ensure the sample cuvette is not being used for blanking. If the sample is too highly coloured or contains significant solids, dilute and repeat the blanking step. If the problem persists and the blank cuvette is not the issue, clean the optical chamber by removing the access cover and cleaning with a soft cloth. Do not use corrosive or abrasive chemicals. I have an Error 7 message Error 7 is caused by too much ambient light reaching the detector. Use the light cover provided with the instrument. Check Standard Issues How do I maintain my Check Standards? Check Standards are manufactured to precise values/tolerances, certified against traceable reference materials and provided in sealed cuvettes. Do not decant or remove the sealed cap from the Check Standard. Ensure the Check Standard cuvettes are clean and dry using lint-free cloths before inserting into the optical chamber. Insert the Check Standard aligning the orientation arrow towards the front of the optical chamber. Values assigned to calibration standards are defined at 20-25°C. Extremely high or low ambient temperatures can affect Check Standard results so ensure standards are at the defined temperature to effectively validate. Check Standards have a two year shelf-life, after which the colours will no longer be valid. Please dispose of the expired standards after this period according to the MSDS. My Check Standard validation has failed Photometers may fail the Check Standard validation step due to the requirement for service/calibration. Contact your local Palintest partner for service and support. Ensure the Check Standards are inserted correctly, using the orientation mark to align and inserted fully. Use the light cap to prevent any ambient light interference. Service/calibration is recommended at annual intervals in normal operation. Bluetooth Issues The Pooltest 9 Bluetooth features the latest Bluetooth SMART connectivity. I can’t connect the Pooltest 9 Bluetooth to my device Ensure your device is Bluetooth SMART ready. Previous versions of Bluetooth (also known as Bluetooth Classic) will not connect to the Pooltest 9 Bluetooth. Check your device specification or visit www.bluetooth.com to see the latest list of SMART ready devices. I can’t download my results to my connected device The Info Panel will show the connected status of the Pooltest 9 Bluetooth. Ensure the connected icon is displayed. If more than one remote device is running the Aqua Pal app, check that the correct device is connected. The connected device is indicated at the base of the Aqua Pal results screen. I can’t upload my data to the Palintest Portal Ensure you have a reliable internet connection to exchange data with the Palintest Portal. Once uploaded data can be shared within your secure user group and downloaded for report generation.

Appendix 2 - Troubleshooting

Bluetooth Error Messages The Pooltest 9 Bluetooth communicates seamlessly with the Palintest Aqua Pal app. In the event of any errors the Pooltest 9 Bluetooth will display either of the following messages: The Pooltest 9 Bluetooth is not receiving a response from the Palintest Aqua Pal app but the remote device is connected.This will appear 10s after a result transmission has started and no valid response has been received. Re-start the Aqua Pal app and select Retry. When the Bluetooth connection to the remote device is lost the following message is shown: Check the Bluetooth has not been inadvertently deactivated in the mobile device settings. Note: the photometer will not appear as a ‘paired device’ in the settings of a Bluetooth SMART device. Battery/Power Issues My batteries are running out too quickly Use good quality batteries and always replace the batteries completely when indicated on the Info Panel. The battery compartment is located underneath the Pooltest 9 Bluetooth and secured by four screws. Remove the battery cover and replace batteries as a set. Refit the battery cover ensuring the cover is tight enough to prevent water ingress. Do not overtighten as this will damage the screw housings. The Pooltest 9 Bluetooth has a number of power-saving features such as auto-dim of the backlight and automatic power down after inactivity (See Section 6).Activating these features will prevent power being used unnecessarily. Bluetooth can also be de-activated if not required. Using the USB port to provide power will automatically prevent battery power being consumed when mains or external power is available. My photometer will not switch on The Info Panel provides an ongoing indication of power available from the battery supply. When the voltage falls below 3.0V the Pooltest 9 Bluetooth will not switch on as the available power will not be sufficient to provide effective photometric testing. Use the USB cable to provide an alternative power supply. If the photometer still fails to switch on, contact your local Palintest partner for service support. My USB power supply is not working Ensure your PC is not operating in power save mode or the mains supply is not isolated. Replace the cable with an alternative to ensure the cable is not faulty. 26 all-guide s.com

Appendix 2 - Troubleshooting

USB Connection Issues I cannot download my data Check the USB mode is set to Hard Disk, not COM port mode. In Hard Disk mode the data can be ‘dragged and dropped’ as with a conventional memory stick and is available in CSV format. Opening CSV data files can be accomplished by a number of text editing or spreadsheet programs. Where do I find the COM port drivers? The latest drivers are available at www.palintest.com COM port drivers are provided for Windows operating systems (Windows Vista, XP and 7). Care and Maintenance The Pooltest 9 Bluetooth contains no user- serviceable parts internally. User maintenance is only recommended for cleaning of the optical chamber, changing batteries and validating performance using the Check Standard Mode. Cleaning the Optical Chamber The optical chamber has been designed to support removal and cleaning with a lint-free cloth as required by removal of the access cover. Do not use any of the following agents when cleaning this optical chamber: • Abrasive cloths • Corrosive chemicals • Any organic solvents Do not overtighten the screws on re-assembly to avoid damaging the access cover. Replacing the Batteries Remove the four retaining screws from the battery cover and gently prise the cover free. Replace all batteries at the same time. Ensure on replacing the battery cover that the gasket is correctly located to prevent any water ingress. Tighten the screws carefully but do not overtighten. Battery Cover Access Screws Access Cover Screws

Appendix 3 - Technical Specifications

28 Pooltest 9 Bluetooth Instrument Type Dual light source photometer offering direct-reading of pre-programmed test calibrations and Transmittance Optical System Optical Source Dual LED sources with optical filters Optical Detectors Silicon photodiodes Peak Wavelengths 550nm, 570nm, 600nm Wavelength Selection Automatic Bandwidth ±5nm Range 1 - 100%T Accuracy ± 1.0% T User Interface Display 320 x 240 pixel LCD with contrast adjustment Backlight Timed, on key press with auto-dim and off User Interface On-screen prompts available in English, French, Spanish, German, Italian, Turkish and Mandarin (Chinese). Keypad Numeric keypad with assignable Hotkeys. Four navigation keys and OK key Physical Size (W x L x H) 150 x 250 x 70mm Weight 975g IP Rating IP67 Power Supply Batteries 3 x 1.5v ‘AA’ batteries Lifetime 40 hours (typical use, backlight off, ‘AA’ alkaline cells) Mains 5V DC, 900mA delivered via USB port Power Management Auto-switch off (user selectable between 5-15 minutes on battery) or continuous operation Power Saving User control for Backlight and Bluetooth to minimise battery consumption Test Methods Tests Available Pre-programmed for Palintest tablet reagent and liquid reagent tests. Also operates in Transmittance mode Test Selection Test number entry or direct access from assigned keys Test Cuvettes 12-20mm OD with automatic cuvette centering Result Units g/l, mg/l, ppm, mmol/l, µmol/l, µg/l Blanking Automatic blanking at all wavelengths. Blank value stored in memory until power off or new blank recorded Connectivity USB USB Type B Connector.Waterproof connector available Wireless Palintest Bluetooth SMART profile Data Management Instrument Memory Non-volatile storage Memory Capacity Up to 100 data sets. Each data set includes date, time, Sample ID, Operator ID, method number, method name, result, units Sample IDs Up to 24 at any time Operator IDs Up to 12 at any time Data Download To computer via USB using Hard Disk or COM port mode.Wireless Bluetooth SMART download, either instantly or as a data batch, using a connected device running the Palintest Aqua Pal app. Optional Palintest Portal data management available Data Output Format Plain text Software Upload Software update by ‘drag and drop’ in USB Hard Disk Mode

Appendix 4 - Spares and Consumables

Name Starter Pack Replacement (50 tests) Pack (250 tests) Total Alkalinity (Alkaphot®) PM 188 AP 188 Bromine PM 060 AP 060 Calcium Hardness (Calcicol™) PM 252 AP 252 Chlorine - Free (DPD 1) PM 011 AP 011 Chlorine - Free, extended range (DPD XF) PM 013 AP 013 Chlorine - Free, Combined and Total (DPD 1 and DPD 3) PM 031 AP 031 Chlorine - Free, Combined and Total, extended range (DPD XF and DPD XT) PM 033 AP 033 Chlorine - Total (DPD 4) PM 041 AP 041 Chlorine - Total (DPD 3 only) - AP 031/1 Chlorine - Total, extended range (DPD XT only) - AP 033/1 Copper - Free, Combined and Total (Coppercol®) PM 186 AP 186 Cyanuric Acid PM 087 AP 087 Iron LR - AP 155 Ozone (DPD method) PM 056 AP 156 pH - Phenol Red PM 130 AP 130 Reagents DPD Glycine, 250 pack AT 056 EDTA Tablets, 250 pack AT 090 DPD Oxystop, 250 pack AP 017 Accessory Reagents PT 791 Pooltest 9 Bluetooth Check Standards Validation standards for Pooltest 9 Bluetooth® with certificate. Supplied in hard transport/storage case 06233 Pooltest 9 Professional Series Service and Calibration Check Standards/Service for Pooltest 9 Bluetooth PT 595/5 Photometer Cuvettes Pack of 5 glass cuvettes with caps. 20mm OD, 77mm tall with 10ml graduation. For Pooltest 25 Bluetooth® and Pooltest 9 Bluetooth® PT 501 Cuvette Rack Rack for PT 595. Holds 8 cuvettes PT 502 Crush/Stirring Rods Pack of 10 crush/stirring rods PT 663 Cuvette Brush For cleaning cuvettes, length 120mm PT 500 De-ion Pack Cuvettes and Racks

Appendix 5 - Palintest Aqua Pal App and the Palintest Portal

30 What is the Aqua Pal App? Generating reliable and accurate water test data is only part of the process of managing swimming pools, spas, hydrotherapy pools and hot tubs. Collating the data quickly and sharing with other members of the team can be the difference between effective control and failing to adhere to local regulatory standards. The Palintest Aqua Pal app is intrinsic in the process of storing, sharing and managing data produced by the Pooltest 9 Bluetooth and also allows for additional data to be added manually for the Big Data experience. The Aqua Pal app is available for both iOS and Android (version 4.3 or later) devices that are equipped with Bluetooth SMART connectivity. On first use the Aqua Pal app will require a registration step for both the app and the secure Palintest Portal.An internet connection is required to process the request as it will involve a second stage authentication via the provided email address. Once the sign up process is completed the Aqua Pal app and Palintest Portal are ready to go. Connecting to the Aqua Pal App Enable Bluetooth on the Pooltest 9 Bluetooth. The icon will appear on the Info Panel showing Bluetooth is on but not connected. Open the Aqua Pal app and log in using your registration credentials. Choose whether to use automatic or manual result addition. To connect the Pooltest 9 Bluetooth press the ‘+’ symbol in the top right hand corner of the device screen. This will produce the list of all locally advertising Bluetooth SMART devices.

Select the Pooltest 9 Bluetooth you wish to

connect (see also Section 6.4 Bluetooth Device ID - Adding Device Name) and Aqua Pal will link. The Info Panel will display the connected Bluetooth icon. Previously connected instruments are stored in the app for future selection. Once connected, data can either be uploaded automatically (if selected) or direct from the log (all or a selected group). Manual data entry is also possible for results generated with non-Bluetooth instruments or external devices such as TDS meters, flow meters or level sensors. Connecting with the Palintest Portal Also included with the Aqua Pal app is the secure Palintest Portal found at https://palintestportal.com Data stored in the Aqua Pal app can be uploaded to the Palintest Portal when an internet connection is available. Data stored within the portal can be shared with team members using the User Group functionality. Simply assign team members to the User Group and they can view data generated by logging in to the portal. Results stored within the portal can be filtered by date, sample ID, Operator ID and parameter. Data can be viewed graphically over time with the status (in spec, nearing limit, out of spec) being indicated by a simple colour key. Further Information The full guide to theAqua Pal app and the Palintest Portal are available at www.palintest.com and available for download from the Palintest Know portal. 31 13 Appendix 5 - Palintest Aqua Pal App and the Palintest Portal all-guide s.com

Palintest House, Kingsway,Team Valley,

Gateshead, NE11 0NS Tel: +44 (0)191 491 0808 Fax: +44 (0)191 482 5372 [email protected] (UK) [email protected] (International) Palintest Australia and Asia Pacific 1/53 Lorraine Street, Peakhurst Business Centre, Peakhurst, NSW 2210,Australia Tel: +61 1300 131516 Fax: +61 1300 131986 [email protected] Palintest USA 1455 Jamike Avenue (Suite 100), Erlanger, Kentucky, 41018, USA Tel: +1 859 341 7423 Toll Free: 800 835 9629 Fax: +1 859 341 2106 [email protected] Palintest China Room 1711, Fanli Mansion, 22 Chaowai Street, Chaoyang District, Beijing, 100020, PRC Tel: +86 10 6588 6200 Fax: +86 10 6588 8311 [email protected] ZI INST 7009B EN