RTI Chamber Adapter Operation & User’s Manual
Text of the manual
Chamber Adapter User's Manual - English - Version 1.6A
RTI article number: 4700.000784
Chamber Adapter
Ion chamber adapter for Piranha
The Chamber Adapter makes it possible
to use ion chambers with the Piranha
and Cobia.
RTI Group AB reserves all rights to make changes in Chamber Adapter and
the information in this document without notice.
RTI Group AB assumes no responsibility for any errors or consequential
damages that may result from the use or misinterpretation of any information
contained in this document.
Copyright © 2019 by RTI Group AB. All rights reserved.
Content of this document may not be reproduced for any other purpose than
supporting the use of the product without prior permission from RTI Group
AB.
RTI Electronics Inc.
33 Jacksonville Road, Building 1
Towaco, NJ 07082
USA
Phone: 800-222-7537 (Toll free)
Int. +1-973-439-0242
Fax: Int. +1-973-439-0248
E-mail
Sales: [email protected]
Support: [email protected]
Service: [email protected]
Web site: https://www.rtigroup.com
Contact Information -
United States
RTI Group AB
Flöjelbergsgatan 8 C
SE-431 37 MÖLNDAL
Sweden
Phone: Int. +46 31 7463600
E-mail
Sales: [email protected]
Support: [email protected]
Service: [email protected]
Web site: https://www.rtigroup.com
Contact Information -
World-Wide
Chamber Adapter User's Manual
2019-10-14/1.6A
Saftey precuations and intended use of the
Chamber Adapter
Always remember to make all connections when the power is off.
After power off the bias voltage drops slowly due to the high impedance of the Chamber Adapter.
This means that high voltage is present on the connector up to 1 minute after power off.
The Chamber Adapter is only intended for use together with the Piranha and Cobia.
The Chamber Adapter is only intended for use together with ion chambers that works on positive
bias level.
Do not connect an ion chamber that is not intended for +300 V bias voltage, to the Chamber
Adapter.
The Chamber Adapter is only intended to be used with ion chambers that has a LEMO connector or
with available adpaters from RTI Group. AB.
RTI Group AB takes no responsibility for misuse of the Chamber Adapter or use together with
products that the Chamber Adapter is not intended for.
RTI Group AB assumes no responsibility to customers not following these safety precautions.
Note! The Chamber Adapter is intended for service and quality control of diagnostic X-ray
equipment. It is not intended for use during or together with diagnostic examinations of patients.
Chamber Adapter User's Manual
2019-10-14/1.6A