906-0731-04 Rev. B Sept. 2012

Page 36 of 68 Battery Case Bottom Cover Kit 712-0731-09 / 712-EGL2-09 Contents: Instructions: STEP NO. DIRECTIONS 1 Remove the damaged cover by unscrewing the (4) 6-32 X 5/16 screws. 2 Remove backing from Bumper feet and place one at each corner of Battery case bottom cover. 3 Rotate cover to align with battery compartment then insert and tighten the (4) 6-32 X 5/16 screws. 4 Perform HiPot Testing then perform Functional Test using the RCS. ITEM QTY DESCRIPTION 1 Battery case bottom cover 4 Screw, Phillips, Pan Head, SS, 6-32 X 5/16 4 Bumper, Rubber, Foot, P/S, Ro ...

906-0731-04 Rev. B Sept. 2012

Page 54 of 68 Gas Output Kit 712-0731-16 Contents: Instructions: Step No. Directions 1 Remove the damaged fitting by unscrewing with a 1-inch deep socket wrench. 2 Place O-ring unto underside of outlet fitting. 3 Place fitting and O-ring unto Gas Output and tighten with a 1” deep socket wrench. (Do not cross thread). 4 Perform HiPot Testing then perform Functional Test using the RCS. ITEM QTY DESCRIPTION 1 Fitting, Patient, Outlet, SPM 1 O-Ring, Neoprene, 1/2" ID ...

906-0731-04 Rev. B Sept. 2012

Page 8 of 68 WARNING! Sensors: Before use, carefully read the LNCS® sensor directions for use. Use only Masimo oximetry sensors for SpO2 measurements. Other oxygen transducers (sensors) may cause improper performance. Tissue damage can be caused by incorrect application or use of an LNCS® sensor, for example by wrapping the sensor too tightly. Inspect the sensor site as directed in the sensor Directions for Use to ensure skin integrity and correct positioning and adhesion of the sensor. Do not damage LNCS® sensors. Do not use an LNCS® sensor with exposed op ...

906-0731-04 Rev. B Sept. 2012

Page 15 of 68 Troubleshooting and Repairs The device uses a comprehensive suite of alarms to alert the operator and guide their actions to resolve the alarm condition and assure patient safety. At the onset of an alarm, the screen displays the alarm name and then a series of context-sensitive help messages (see Figure 2 example). These messages serve to guide the operator by presenting suggestions as to the cause and resolution of a particular alarm. When multiple alarms occur they are prioritized and displayed based on the risk to the patient. Should the opera ...

impact Uni-Vent 731 Series user guide recommended for: SHR 1504, ERAD Ex Series, AK965, GUG-01SL-E, 15530-LCMB-0200 - ESCON Multiplexing Line Card Multiplexor.

⚠️ Download Safety

The Fan Service manual (68 pages) is completely safe to download (last scan date: 13/01/2026).

Related PDF Instructions for impact Uni-Vent 731 Series:

Fan Equipment: Frequently Requested

Have some Questions about impact Uni-Vent 731 Series Device?

Get valuable and actionable advices from other impact Uni-Vent 731 Series device users. Describe your problem or question, regarding the using of Uni-Vent 731 Series, in as much detail as possible, and other users will be able to answer you.