Jeio tech IST-3075 Operating Manual
Read Online
Read the Jeio tech IST-3075 Operating Manual online using the built-in PDF viewer.
Text of the manual
Model : IST-3075, IST-4075, IST-3075R, IST-4075R
Menual no. : XXXXXXXXXXX
Before using this product, read this entire Operator's Manual
carefully. Users should follow all of the Operational Guidelines
contained in this Manual and take all necessary safety
precautions while using this product. Failure to follow these
guidelines could result in potentially irreparable bodily harm
and/or property damage.
Thank you for purchasing Jeio Tech’s products.
We are doing best for customer’s satisfaction.
Quality Management System
Jeio Tech Co, Ltd. is dedicated to providing world-best
product quality and customer satisfaction. To ensure we
maintain this commitment we have developed and
implemented a total quality program, which conforms to
the requirements according to DIN EN ISO 9001:2000
for the design, development, production, sales and
servicing of biotechnology, environmental chemical
engineering related products, and reliable measuring
equipment
for
electric
and
electronics
(ovens,
incubators, constant temperature humidity chambers,
constant
temperature
baths,
refrigerating
bath
circulators, heat exchangers and shakers).
Visit our Web site at www.jeiotech.com to view a copy of our certificate.
Disclaimer
Jeio Tech Co., Ltd. is committed to a continuing program of product development and
improvement, and reserves the right to change information, such as specifications, appearance,
and dimensions, described in this document without notice.
Copyright
No part of this manual may be reproduced or transmitted in any form or by any means,
including photocopying, recording, or using information storage and retrieval systems, for any
purpose other than the purchaser's own use, without the express written permission of Jeio Tech
Co., Ltd.
©2014. All Rights Reserved. Jeio Tech Co., Ltd.
Any other product names and services identified in this manual are trademarks or registered
trademarks of their respective owners. No such use, or the use of any trade name, is intended to
convey endorsement or other affiliation with Jeio Tech Co., Ltd
1.1 How to use the manual .........................................................................................................................................1
1.2 Symbols used in this Manual ................................................................................................................................1
1.3 Exemption for responsibility .................................................................................................................................2
1.4 Warning statement .....................................................................................................................................................2
1.5 Caution statement ......................................................................................................................................................4
2
Functional Description ....................................................................................................................................................7
2.1 Introductions .................................................................................................................................................................7
2.2 Features ........................................................................................................................................................................7
2.2.1
High Performance ......................................................................................................................................7
2.2.2
Advanced Convenience ...........................................................................................................................8
2.2.3
Advanced Safety .........................................................................................................................................9
2.3 Structure .................................................................................................................................................................. 10
3
Installation ....................................................................................................................................................................... 14
3.1 Unpacking and Checking .................................................................................................................................. 14
3.2 Component ............................................................................................................................................................ 14
3.3 Preparation before installation ....................................................................................................................... 15
3.3.1
Space requirements ............................................................................................................................. 15
3.3.2 Environmental setting .................................................................................................................................... 17
3.4 Installation .............................................................................................................................................................. 18
3.4.1 Caution for Installation ................................................................................................................................ 18
3.5 Attaching Condensate Drain ........................................................................................................................... 19
3.6 Attaching/Detaching External Refrigeration (IST-3075, IST-4075 Only) ....................................... 19
3.7 Connection power ................................................................................................................................................. 21
4
Operation ......................................................................................................................................................................... 22
4.1 Controller .................................................................................................................................................................. 22
4.2 How to use the controller ................................................................................................................................ 23
4.2.1
Temperature control .............................................................................................................................. 25
4.2.2 Shaking Control ............................................................................................................................................... 26
4.2.3 Shaking Timer .................................................................................................................................................. 27
4.2.4 Alarm and Stop by force ............................................................................................................................. 28
4.3 General Settings .................................................................................................................................................. 31
4.4 System Settings ................................................................................................................................................... 32
4.4.1 Control Deviation Limit ................................................................................................................................ 32
4.4.2 External Start/Stop Key ................................................................................................................................ 33
4.4.3 Calibration .......................................................................................................................................................... 33
4.4.4 Auto Tuning ....................................................................................................................................................... 36
4.4.5 Operating Parameters .................................................................................................................................. 36
4.4.6 Reset ..................................................................................................................................................................... 37
4.5 Safety Tool ............................................................................................................................................................... 37
4.5.1 Over Temp. Limiter ......................................................................................................................................... 37
4.5.2 Glass Fuse .......................................................................................................................................................... 38
5
Maintenance ................................................................................................................................................................... 39
5.1 Inspection Period .................................................................................................................................................. 39
5.2 Cleaning .................................................................................................................................................................... 40
5.3 Air filter cleaning (IST-3075R/IST-4075R) ................................................................................................ 41
5.4 Fuse Replacement ............................................................................................................................................... 42
6 Troubleshooting ................................................................................................................................................................. 44
6.1 Stop during the operating ............................................................................................................................... 44
6.2 Another problem and solution .......................................................................................................................... 45
7
Accessories ...................................................................................................................................................................... 47
7.1 Mountable maximum quantity.......................................................................................................................... 47
7.1.1 Universal Platform + Flask Clamp .......................................................................................................... 47
7.1.2 Universal Platform + Plastic Flask Clamp............................................................................................ 47
7.1.3 Universal Platform + Funnel Clamp....................................................................................................... 48
7.1.4 Universal Platform + Microplate Holder ............................................................................................. 48
7.1.5 Universal Platform + Test Tube Rack .................................................................................................... 48
7.1.6 Spring Wire Rack + Flask ........................................................................................................................... 49
8.0 S/W ....................................................................................................................................................................................... 50
8.1 Monitoring Program installation ................................................................................................................... 50
8.2 Communication Protocol .................................................................................................................................. 52
8.2.1 Physical Layer ................................................................................................................................................... 52
8.2.2 Model System number................................................................................................................................. 52
8.2.3 Modbus Protocol Address Definition .................................................................................................... 52
8.2.4 Modbus Protocol Description ................................................................................................................... 54
9.0 Appendix ............................................................................................................................................................................ 59
9.1 Technical Specifications ..................................................................................................................................... 59
9.2 Circuit Diagrams.................................................................................................................................................... 61
9.2.1 IST circuit diagram ......................................................................................................................................... 61
9.2.2 External refrigeration system (IST-3075/IST-4075) ........................................................................... 61
9.2.3 Internal refrigeration (IST-3075R/IST-4075R) ...................................................................................... 61
9.3 Disposing of products .......................................................................................................................................... 62
9.4 Warranty ................................................................................................................................................................... 62
9.4.1 Terms of Warranty Service .......................................................................................................................... 62
9.4.2 Warranty exception ......................................................................................................................................... 63
9.4.3 Service and technical advice ...................................................................................................................... 63
1.1 How to use the manual
This manual is intended for individuals requiring information about the use of product. Use
this manual as a guide and reference for installing, operating, and maintaining your Jeio
Tech product. The purpose is to assist you in applying efficient, proven techniques that
enhance equipment productivity
1.2 Symbols used in this Manual
(1) The alert marks are for safety operation and protect user and instrument from Damage.
(2) Signal word panels are a method for calling attention to a safety messages or property
damage messages and designate a degree or level of hazard seriousness.
(3) Pay attention enough to the contents of alert marks.
Signal word panels
Uses
Indicates a hazardous situation which, if
not avoided, will result in death or
serious injury
Indicates a hazardous situation which, if
not avoided, could result in death or
serious injury.
Indicates a hazardous situation which, if
not avoided, may result in minor or
moderate injury.
Indicates a property damage message.
1.3 Exemption for responsibility
(1) The claim which is out of the quality guaranteed by the manufacturer is out of
manufacturer’s responsibility.
(2) The damage which is from unexpected fault or damage of user by Acts of God is out of
Manufacturer’s responsibility
1.4 Warning statement
Indicates a hazardous situation which, if not avoided,
could result in death or serious injury.
Please use the product in safety facility installing laboratory in case of accident.
Installed the product on durable and flat surface.
Please, make sure safety equipment with relevant provision before handling the sample
which may cause flammable or toxic gases.
Do not use the machine near to places where explosion can be happened due to
organic evaporating gases.
Explosive materials: Acid, Esther, Nitro compound
Inflammable materials: salt peroxides, inorganic peroxide, salt acids.
Do not use the machine at places where moisture is high and flooding can be
happened.
Please check and connect properly -the voltage, phase and capacity of power supply on
the ID plate before installation.
Be sure to install a separate power wiring and use a dedicated power supply.
Power supply must be properly grounded.
Abnormal grounded connection causes serious damage. Grounded connection must not be
on the water pipe and gas pipe.
Put off the power plug if some sounds and burning smell, smokes are happened. And
request the service
Stop the product operation and request service.
Do not assemble, repair, modify on your own.
The product may not work well and electric shock in the efficiency of the product. Also you
cannot get after service by warranty regulation
1.5 Caution statement
Indicates a hazardous situation which, if not
avoided, may result in minor or moderate injury.
Be sure to disconnect the power after turning off the power switch.
This is the safety regulation for next user.
Do not put heavy things on the power line. Do not put the machine on the line.
It may take off the wire coating and causes the electric shock or fire.
Do not touch it with wet hands and put the main plug correctly.
It may cause the electric shock or injuries.
Do not inject any liquid and inflammable things inside of product.
Do not let the product take any strong shock or vibration.
It causes abnormal operation or trouble. It may deteriorate the ability of the product and not
obtain correct results.
Do not install the stirrer neat machinery generating high frequency noise.
Please avoid installed from high frequency- welding machine, sewing machine, and mass SCR
controller
Do not sprinkle insecticide or flammable spray on the product. Use smooth cloths.
Cleaning with solvent can cause fire and deformity.
Please power off while product cleaning.
It may cause the electric shock or fire.
Wear protective gloves.
Wear eye protection.
No water
No corrosive
Electrical shock.
Flammable
2 Functional Description
2.1 Introductions
The cultivation environment such as temperature, pH, oxygen concentration, nutrition supply
have a significant impact on cell cultures of animals and plants or proliferation of
microorganisms.
This incubated shaker is to provide a suitable environment for the cultivation through
controlling constant temperature and shaking speed.
The main uses of this device are as below:
Cell culture
Extractions
Solubility studies
Hybridization
Plasmid purification
Protein expression
2.2
Features
2.2.1 High Performance
(1) Precise and fast temperature / shaking speed control is available by microprocessor PID
feedback control. PID feedback control ensures the same experimental environment
conditions for the user. The device’s deviation of set temperature and set rpm are each
under ± 0.1 ℃ (standard: 37 ℃), ± 1% (standard: set rpm) which are regarded as very
precise value. If the deviations go above a certain level, alarm will be activated to the
user.
(2) Control range of temperature and stirring speed is wide. Temperature control up to 80 ℃
and stirring speed control up to 500rpm are available to provide users wider
experimental conditions.
(3) Optimizing the structural design of the product enables stable shaking even with heavy
load. Low center of gravity minimizes noise and vibration.
(4) The accurate temperature control is available on the three temperature points. At the
three points, the user adjust the temperature as same as the external calibration sensor.
2.2.2 Advanced Convenience
(1) Touch LCD built user-friendly interface. LCD provides intuitive operation status check and
easy operation.
(2) Timer and operation time check of shaking performance is available. Setting and
checking the value is available up to 999 hours and 59 minutes.
(3) Convenient sample monitoring without affecting inside environment thanks to the three
side transparent acryl lid with bright LED lamps.
(4) Sturdy and light-weight acryl lid for effortless opening and closing the door.
(5) Not only the conventional RS-232 port but also USB port is provided so it is easy to
control the unit with PC.
(6) By providing an automatic power failure recovery, even if the product gets power back
after a momentary power failure, the auto run function automatically runs the product.
(7) This equipment is more convenient when mounting / desorption of the sample. When
shaking performance, operation always starts and stops at a specified location, the
platform is fixed to easily replace the sample.
2.2.3 Advanced Safety
(1) Smooth shaking start and stop mechanism minimizes the opportunity of reagent leakage.
(2) In the case of the over load, the unit adjust itself with controllable rpm. There is no move
cased from the vibration of the unit.
(3) There is an independent adjustable temperature limit device in addition to the main
controller in this device. If temperature rises more than the specified value due to equipment
error, the independent device protects the internal sample and equipment by blocking the
power supply of the temperature-related equipment.
(4) When shaking operation is not possible due to the obstacle interfering the system, over
current protection device stops the operation.
(1) Door cover and Handle
You can monitor the inner samples without opening the door during operation.
(2) Gas Spring
Door can stay opened for the user’ convenience when sample replacement.
(3) LED Lamp
Long lasting less electricity consumption LED is installed to check the inside of the chamber.
(4) Temperature Sensor
It is Pt 100Ω sensor that temperature and resistance values are shown as the linear format
and response speed is fast.
Temperature control is up to 80℃ by two 400W pin heaters.
(6) Over Temperature Limiter
Over temperature limiter is independently configured from the main controller. The inner
temperature of the chamber exceeds more than the set temperature, the device cut off the
power to the power switch for the secure condition. (Refer to ‘4.5 Safety Device’)
(7) Shaking Table
Shaking table is connected to a shaking system and Accessory such as Universal Platform
can be equipped on shaking table.
(8) Control Panel
There are independent displays for each Temperature, RPM, Timer and there are and
start/stop buttons and LED on/off buttons.
(9) Computer Interface
The unit can be controlled by PC through USB port and RS-232 port. In the case
of connecting both of the port, the USB port will be priority.
(10) Condensate Drain Barb
In case of the liquid leakage caused by spilled reagent or cracked flask inside of the chamber,
the liquid is designed to be drained out through the drain of the unit. The drain barb is
made of chromium plated copper and it is easy to connect 6mm ID soft-walled tubing to the
condensate drain. (Refer to ‘3.5 Attaching Condensate Drain’)
(11) Power Switch and Fuse
Power switch provide the main power to the device. Fuse protects the device from
instantaneous overcurrent. In case of fuse replacement, check the rated power. (Refer to 5.4
(12) Door Switch
There is a door limit switch between the door and the appliances mainframe’s upper parts. If
the door is opened, Shaking, Blower and Heater operation automatically stops. After 3
minutes at this status, warning sound alarms users to close the door. If a user does not shut
a door so that five minutes passes, and an alarm sound continuously rings. And it blocks off
the power to be authorized with a power switch, and Off gets a power switch done and all
blocks off 2 phase of the power supplied with to an each part of appliances, and
configuration does the safe state that only a Ground part is connected.
(13) External Refrigeration Port (IST-3075/IST-4075only)
Even if the model such as IST-3075/4075 which has no refrigerator, It can lower the
temperature in the chamber by receiving refrigerant from the external refrigerant device such
as chiller. The In/out ports are located on the back and bottom side of the device. These
brass push-to-connect fittings accept Ø8mm OD hard-wall tubing and provide access ports
for external refrigeration. (Refer to ‘3.6 Connection to External Refrigeration System’)
(14) Condenser Cover and Filter (IST-3075R/IST-4075R only)
Attachable condenser cover and filter prevent dust to the inside system.
(15) Foot
Foot fix the device for stable operation.
3.1 Unpacking and Checking
(1) Inspect the shipping container carefully for any damage.
(2) Remove the outer container.
(3) Before use, inspect the product carefully for any damage that may have occurred
during shipping.
(4) Report any damage to your local Jeio Tech office or the distributor.
3.2
Component
(1) After unpacking, check the components.
(2) In the case of omission of components, contact to Jeiotech.
Item
Figure
Quantity
Description
Main Body
1
Fuse
2
JEIOTECH
SOFTWARE CD
1
Preparation before installation
3.3.1 Space requirements
It is essential that the product to be situated in an area where there is sufficient space for the
product. Below figures show the minimum space requirements needed to properly operate and
maintain the product.
Dimensions
(mm)
Models
A
B
C
D
IST-3075(R)
440
785
510
860
IST-4075(R)
540
890
510
915
3.3.2 Environmental setting
The unit can be operated properly under the following environmental conditions for a
long time running without any problem.
No direct sunlight on the product
Ambient temperature: 5℃ ~ 40℃ (41℉~104℉)
Relative humidity not to exceed 80%
Altitude not to exceed 2000m (6,562 feet)
Connect the product to earth grounded terminals only.
•
Please install on the sturdy surface laboratory which is set safety facility and
make sure horizontal align correctly.
•
Do not use the Product near environments where flammable gas may leak.
•
Moving casters must be changed height adjustable foots so that the unit is not
moving.
3.4.1 Caution for Installation
(1) The device is recommended to use under the room temperature and humidity (30°C,
80%RH) do not install it near the Heat devices like a Heater.
(2) Have enough space for door opening. (refer to 3.3.1)
(3) Please install it on the sturdy surface laboratory and do not throw down or gives a big
shock.
(4) Locate it 1.5m away from any light devices and 30 CM away from the wall.
(5) Install it on the sturdy leveled surface to prevent abnormal turbulence and noise.
1.5m(5.0')
30.0cm
(12.0”)
30.0cm
(12.0”)
WALL
WALL
3.5 Attaching Condensate Drain
In case of the liquid leakage caused by spilled reagent or cracked flask inside of the chamber,
the liquid is designed to be drained out through the drain of the unit. The drain barb is made of
chromium plated copper and it is easy to connect 6mm ID soft-walled tubing to the condensate
drain.
3.6 Attaching/Detaching External Refrigeration (IST-3075, IST-4075 Only)
Many laboratories are built with central chilling systems. With this in mind Jeio Tech has
developed the IST-3075/4075 so that it can be connected to external refrigeration systems.
To connect the unit to an external refrigeration source use Ø8mm OD hard-walled tubing,
such as nylon or polyethylene.
Note: The ISS-3075/4075 controller does not have the ability to communicate with the
external refrigeration source. Use the following procedure to connect/disconnect the tubing
to the unit. Use the following procedure to connect/disconnect the tubing to the unit.
(1) Inserting tubing into fitting
Hold the tubing firmly and push it tightly into the fitting. Run the tubing to the external
refrigeration source.
(2) Removing from fitting
Grab the tubing tightly and push towards the incubated shaker as you firmly push the
collet release button with two fingers. Continue to push the collect release button, and
pull the tubing out of the fitting.
3.7 Connection power
JEIOTECH’s Shaking Incubators use a single-phase current.
Use a suitable plug as the picture bellow by Identification label.
Voltage is the 10% of applied voltage.
•
Connect the power with checking the voltage, Phase,
Capacity.
•
Use the ground power for the connection.
Do no use the double cap or a current tap socket causing a
damage on the cable and fire due to an overcurrent.
110/120VAC
220/230VAC
The controller is consist of the touch screen, LED, membrane. Once the power is supplied, below
display is appeared.
1
Power ON indicator
Power is supplied.
2
Temperature Control ON
indicator
Under the temperature controlling.
3
Shaking Control ON
indicator
Under the shaking controlling. In the case of the timer is set, it is
off.
4
Main Display
Main Display for the operationg
5
General Setting
General setting.
6
System Setting
Parameter setting is available. For change the parameter,
acknowledge the manual.
7
Quick Start/Stop
Touch button is the base. Start/Stop external button is installed on
the controller for the user’s convenience. Function setting is
available on the system setting. (Default: shaking start/stop)
8
LED Lamp
On/Off the LED lamp.
4.2 How to use the controller
Setting is available on the main display.
1
Set RPM
Display the setting RPM. Touch the screen, set the RPM.
2
Set Temperature
Display the setting temperature. Touch the screen, set the
temperature.
3
Set Timer
Display the shaking time. Touch the screen, set the timer. In the case
of no timer is set, the operating time is displayed. (999hours
59minutes)
4
Actual Temperature
Display the temperature of inside of chamber.
5
Actual RPM
Display the present shaking speed.
6
Back
From main display to home display. It is available when the unit is
stop the operating.
In the case of the power is off for the supply faulty reason and the
supply is back, the previous operating is back. It is auto restart
function. At the time Power-Ack button is appeared on the display. In
the case of the shaking was operating, the operating time is appeared
on the display with red words. Once touch the Power-Ack or change
the setting value, The Power Ack display is disappeared.
8
Silent
In the case of attention on the unit is required, the unit generates the
alarm. Touch the silent, the unit stops the alarm. Once the problem is
not eliminated, the unit generates alarm again.
9
Status Bar
The unit stops the platform on the same position. This
display indicates that the platform is on the right
position. If the platform is not on the right position,
the color will be changed as yellow.
The unit provide USB port and RS-232 port. The
display is on when the computer is connected.
Report Only :
/ Slave Mode :
Door is opened.
LED lamp is on.
Mute function. The setting is available by the General
Setting.
10
Lock
The lock function is to prevent the not proper setting change. Once
touch the button, the display is changed same as
. To
unlock it, touch twice of the button.
->
4.2.1 Temperature control
(1) Start temperature control
Touch the set temperature button on the main display. Setting display is appeared. Set the
temperature with touch pad and touch the start button. The temperature control starts. SV and PV is
displayed on the main display with green color.
(2) Stop the temperature control
Touch the set temperature button. Setting display is appeared. Touch the stop button. The
temperature control stops. The temperature control starts. SV and PV is displayed on the main
display with gray color.
(3) Change the set temperature
Touch the set temperature button. Setting display is appeared. Put the setting value on the touch
pad and touch the enter button. SV is changed.
Temperature setting is one digit and one decimal point available. Touch the panel and put the
numbers.
Move out from the setting mode with ESC button.
In the case of no touch for 10 seconds, the display changed to the main automatically.
Check the PV, RPM on the setting display is available. (right upper side)
The operating is maintained as previous during the setting.
Once the door is opened during the shaking operating, the heater and fan is stopped. Close the
door before door error generating, the previous operating is back.
4.2.2 Shaking Control
(1) Start shaking control
Touch the Set RPM Button under the shaking stop status. Setting mode is appeared. Set the value on
the display and touch the Start button. Shaking control starts. SV/PV rpm is changed to green color.
(2) Stop the shaking control.
Touch the Set RPM button under the shaking operating status. Setting mode is appeared. Touch the
Stop button. Shaking stops. SV/PV rpm is changed to gray color.
(3) Change the shaking speed
Touch Set RPM button. Setting mode is appeared.
Set the value on the display and touch the Enter button. The rpm is changed.
Move out from the setting mode with ESC button.
In the case of no touch for 10 seconds, the display changed to the main automatically.
Check the PV, RPM on the setting display is available. (right upper side)
The operating is maintained as previous during the setting.
Once the door is opened during the shaking operating, the heater and fan is stopped. Close the
door before door error generating, the previous operating is back.
In the case of change the sample, stop the shaking and check whether the platform is
completely stopped.
4.2.3 Shaking Timer
Timer is only available for shaking. Once the timer is set, the shaking stops after operating the set
time and beep is generated. Setting time is on the display once the timer is set and ‘---:--‘ is on the
display once the timer is not set. Remaining time is on the display once the timer is set and
accumulated time is on the display once the timer is not set.
(1) Timer ON
Touch the Set timer on the main display. Timer OFF is on the display once the timer is set and Timer
on is on the display once the timer is not set. Touch the Timer ON button and set the time on the
display. Touch the Enter button. Check the set time on the main display.
(2) Timer OFF
Touch the Set timer on the main display. Timer OFF is on the display once the timer is set and Timer
on is on the display once the timer is not set. Touch the Timer OFF button and touch the Enter. Time
display on the main display is changed to “---:--“.
(3) Change the setting time
Touch the Set timer on the main display. On the Timer ON status, put the time on the display. Touch
the Enter. Check the changed time on the main display.
Once the setting time is changed during the timer operating, original set time is changed. The
accumulated time is applied to the new time setting and the remaining time is calculated and
displayed. If you want whole new time setting, off the timer and set the timer again.
Timer setting is available for hour:minute. Touch the part you want to change and put the
numbers on it.
Move out from the setting mode with ESC button.
In the case of no touch for 10 seconds, the display changed to the main automatically.
On the timer set display, you can check the present chamber temperature and rpm. (right upper
side of the display)
Below 1 minutes is displayed as“mmm:ss”.
Timer is not counted during the door open.
4.2.4 Alarm and Stop by force
The unit senses the problem on the unit and warn to the user. The warning is visual-audial effect.
The alarm is divided into warning and fault.
4.2.4.1 Warnings
Warning generates visual-audial alarm but the unit is operating. Once the warning is generated,
Silent button is appeared on the main display. Once the silent button is touched, next 15 minutes
there is no alarm. If the problem is on-going after 15 minutes, the alarm is generated again.
The temperature difference between setting temperature and present
temperature is more than the range of the permission. The unit check it
once the present value reached to setting value. Once the gap is over the
permission range, the alarm is generated.
2
RPM Deviation
The rpm difference between setting rpm and present rpm is more than
the range of the permission. The unit check it once the present value
reached to setting value. Once the gap is over the permission range, the
alarm is generated.
3
Load Unbalance
Once the shaking system is vibrated by the present operating, the system
adjusts the shaking speed to the controllable stage. This is on the display.
4
Door Open
Once the door is opened, the temperature and rpm are not operated.
Once the door is opened, visual-audial alarm is generated. After 3
minutes of door opening, the alarm is effective. Once the door is closed,
alarm is eliminated and the operating is back.
Alarm is not effective if Sound Off is set on the General Setting.
Deviation Alarm is not effective once the present value reaches to the setting value. (set time, in
permission range)
Deviation alarm is initialized under the below situation.
-
Set the setting value
-
Get back to the work after stopping the control. (Door open, Stop operating)
Once the Deviation initialized, the unit check the present value. If the temperature not reaches
to the SV for 4 hours and the rpm not reaches to the SV for 10 minutes, Deviation check is
started.
The range of permission and the period of Deviation is changeable on the System Setting.
Once Load Unbalance is generated and the setting rpm is changed, the changed rpm is on the
display with red color.
1
Temperature Limit
Digital temperature regulator is active. The temperature of inside of
chamber is out of range. Stop the unit for unit and user’s safety.
Check the range of the temperature regulator.
2
Load Unbalance
Shaking system has severe vibrating so that the unit must be
stopped. Unbalanced sample attaching, heavy sample attaching, un-
flat ground are the causes. Shaking system is stopped.
3
Door Open
Once the door is opened for 5 minutes, the unit is stopped and
generates the alarm.
4
Sensor Fault
Temperature sensor is faulty so that the operating is stopped.
5
Over Current
Over current is generated so stop the unit. Main causes are over-load
and obstacle inside of the system.
Fault Alarm stops the unit and inform it to the user when the unit need forced stop for the
protection. Fault alarm are over-temp protection, Load Unbalance, Door Open, Sensor Fault, Over
Current.
4.3
General Settings
Set the general operating. Touch General Setting button and below setting menu is appeared. Touch
the ‘<’ or ‘>’ and explore the menu.
1
Shaking Direction
Set the shaking direction
• CW(clock wise), CCW(counter clock wise)
2
Temperature Unit
C°/F°
3
Lock
Use the Lock or not
4
System Sound
Mute for sound or not
5
LCD Brightness
Adjust the brightness of LCD
6
LCD Auto Dim
LCD Dim function (No input, LCD dim)
7
Dim Brightness
LCD Dim, and the brightness of the dim
8
Dim Time Out
Waiting time for LCD Dim mode operating
9
Light Brightness
Select the brightness of LED
10
Light Auto
Door opened, auto LED light on
11
Auto Restart
Auto restart function. The unit is back to the operating, before the
power out.(Time information is not back)
12
USB (RS232)
USB and RS-232 port is available for the PC. Select the mode when PC
is connected.
Report Only : PC received the information from the unit.
Slave Mode : PC control the unit.
Set the main variable that affects on the unit. Please read the manual carefully before setting the
system. Touch the System setting button on the home display, below clauses are arranged.
1
Control Deviation
Limit
Set the deviation limit range. Temperature setting and rpm setting are
available.
2
External
Start/Stop Key
Start/Stop membrane button is located on the control panel. Set about
this key
3
Calibration
Calibration on the temperature sensor and rpm to the user’s reference.
Calibration on the LCD touch coordinate is available.
4
Auto Tuning
Temperature PID parameter update. Optimum control on the main
operating temperature is available.
5
Operating
Parameters
Set the operating and error check
6
Reset
Set the parameter as initial factory value.
4.4.1 Control Deviation Limit
Set the deviation limit between setting value and actual value of the temperature and rpm. The time
of out of deviation limit range can be set either.
1
Temperature
Deviation Limit
Set the deviation limit between setting value and actual value. Set on
the key pad.
2
Temperature
Deviation Alarm
Delay
Once the deviation is out of range for alarm delay period, the alarm is
generated. Once the actual temp reached to the set temp, the alarm
is disappeared after alarm delay period check.
3
RPM
Deviation Limit
Set the deviation limit between setting value and actual value. Set on
the key pad.
4
RPM Deviation
Alarm Delay
Once the deviation is out of range for alarm delay period, the alarm is
generated. Once the actual rpm reached to the set rpm, the alarm is
disappeared after alarm delay period check.
4.4.2 External Start/Stop Key
External Start/Stop button is on the control penal as membrane button. It generates
start and stop. Set the applying range of it.
1
Only Shaker
The button only works for shaker’s start/stop.
2
Only Incubator
The button only works for incubator’s start/stop.
3
Shaker Incubator
The button only works for shaker and incubator’s start/stop.
Shaker and incubator are both available for the external membrane button. Once one of the
shaker/incubator is on the generating and touch the button, the stopped part is operated. Touch
the button again, both of mode are stopped.
4.4.3 Calibration
The temperature sensor and rpm reached to the user’s reference measure value, accurate experiment
circumstance can be made. Calibration for coordinate is available.
4.4.3.1 Temperature Calibration
Calibration for temperature sensor of main controller. The unit provides three points calibration.
Each point of three shows the uncalibrated and calibrated status. Touch the point you want to
change. Touch the one point of Calibration 1,2,3. Put the external sensor on the chamber and close
the door.
Touch the temperature on Set Temp. and put the desired temperature on it and operate the unit.
Wait until the actual temp reaches to the setting temp. Once the actual temp reaches to the setting
temp, touch the calibrated temp and put the temperature from the external sensor. Touch the
confirm for the saving. The unit stops the operating. The unit read the temp as the external sensor’s
value.
Once the door is opened or the temperature is out of range, the temperature control is
stopped.
The calibration is operating on various points so it limits the calibrating that is our of the logic.
The result of the calibration is on the uncalibrated and calibrated. Uncalibrated means the value
from the unit and the calibrated means the temp that calibrated. The calibrated value is from
the external sensor.
The calibration setting is available to be initialized on the System Setting Reset.
4.4.3.2 Shaking Speed Calibration
The user calibrates the unit’s rpm with external sensor. The unit is available for one point rpm
calibration. The procedure is similar with temperature calibration. The difference is that the shaking is
not stopped during the door opening because the door need to be opened to check the shaking
with external sensor. Touch the set rpm. Put the desired rpm and start the nut. Wait unit the actual
rpm reached to the set rpm. Touch the calibrated rpm and put the value from the external sensor.
Touch the confirm button to save it. The unit read the calibrated value from the next operating.
Be careful that the external sensor never touch the platform. Once the external sensor is
removed from the platform during the operating, it will be dangerous.
The calibration setting is available to be initialized on the System Setting Reset.
4.4.3.3 Touch Screen Calibration
Touch penal coordinate could be in the problem for the reason of the older unit and the problem in
the unit. The unit has a touch screen coordinate function. Touch the “+” on the screen three time s.
And repeat it two times.
4.4.4 Auto Tuning
Update the parameter of PID temperature control. Optimum temperature for the main using
temperature is available. The unit is set to use the temperature on proper range. Once the
temperature is not actual on the main points during the actual use on the field, operating auto
tuning will tune the temperature parameter and make the unit operates the temperature on the
actual temperature.
Touch the Set Temp. and put the desired value. Touch the START button and start the parameter
tuning. Touch the stop button during the tuning and stop the tuning. Once the tuning is complicated,
touch the ok button and save it.
4.4.5 Operating Parameters
Set the parameters for the other operating.
1
Position Sensor
Check the platform stop on the right position.
2
Stall Check
Check the belt when the platform is not working properly.
3
Defrost Use
Defrost is not for the IST model. Off the defrost use.
4
Load Unbalance
Once the vibration is generated, the shaking speed is
adjustable.
5
Threshold
Load Unbalance check the vibration on the unit and adjust the
rpm automatically. The vibration is on the range of not
controlling, the automatic adjusting is operated.
Initialize the changed value to the factory value.
1
Temperature Calibration
Initialize the calibration value.
2
RPM Calibration
Initialize the calibration value.
3
Auto Tuning
Initialize the calibration value.
4
Touch Screen
Initialize the calibration value.
5
Operating Parameters
Initialize the calibration value.
6
Factory Default
Initialize the calibration value of all parts.
Auto tuning parameter can be initialized to the factory value.
4.5 Safety Tool
4.5.1 Over Temp. Limiter
This model has the over temp limit and temp sensor in the main controller.
Over temp limit is located on the right down side of the unit. Set the over
temp limit over 15% than normal use temperature by a flat head screwdriver.
In case of that the temperature in the chamber over set temp, the operation
discontinues.
4.5.2 Glass Fuse
In the case of the over current generating, the fuse is cut and the power is down.
5.1 Inspection Period
Classification
Inspection period
Daily
Weekly
Monthly
Quarterly
Yearly
General
Power
Connection between machine and
power
●
Power code
●
Exterior cleanliness
●
Drain connection
●
Outside
refrigeration
system
connection (IST-3075/IST-4075)
●
Incubator
Set
temperature
and
actual
temperature difference
●
Auto Tuning
●
Temperature setting
Temperature correction
●
Actual temp. / Display temp. correction
●
Air filter clean (IST-3075R/IST-4075R)
●
PC Sync operation
●
Controller and timer operation
●
System
Connection
between
platform
and
accessories
●
Connection
between
platform
and
shaking table
●
Actual rpm/ Display rpm correction
●
Malfunction of shaking system
●
(1) After disconnect the power cord, clean the machine with soft and dried towel.
Regarding the un-removable point, clean the polluted area only by towel with alcohol solvent
(methanol, ethanol) which has low boiling point.
(2) Do not use Acid solution, sharp one, soapy water, detergent and hot water.
It makes the machine discolored. Rubber and plastic part can be change of shape and color.
Especially, do not use volatile matter.
In case of neutral detergent, clean it with the soft fabric and dry well.
(3) Do not put the water to exterior of the machine when you clean the surrounding (Especially,
socket and controller part, it can make short circuit problem.)
(4) After discussion the proper cleaning method to avoid any damage for the machine with Jeiotech,
if you want to clean the machine or remove the polluted area with not mentioned cleaning
method before the clean.
(5) The inside electric part of the machine should be handled by Jeiotech or person who is
delegated by Jeiotech.
(6) If the parts is required to replace, please use genuine parts only.
(7) Technical maintenance is not offered in case of abnormal trouble beyond the normal limit.
•
Do not soak the machine into water and spray the water
•
Do not clean whole body with chlorine bleaching, detergent with chlorine,
an abrasive, ammonia, steel sponge and alcohol solvent (methanol, ethanol)
which has low boiling point.
5.3 Air filter cleaning (IST-3075R/IST-4075R)
Step 1: Put power switch and turn the machine off.
Step 2: Loosen the bolt that located in edge of condenser cover
Step 3: Disassemble condenser cover
Step 4: Disassemble air filter which is combined with condenser cover.
Step 5: Vacuum or clean the air filter with the water.
Step 6: Vacuum or high pressure air clean to the condenser in the machine.
It makes better refrigeration effect.
Step 7: Assemble air filter in reverse order of Step 1~Step4
•
When air filter clean, please be careful to avoid fold the condenser pin
•
It is possible to decrease refrigeration effect.
5.4 Fuse Replacement
Replace the fuse which is located in next to power button (Refer to 2.3.(11)) if no operation of short
circuit breaker and No power.
There is extra fuse 2pcs in socket case inside. If you need additional fuse, ask it to sales team or seller.
Model
Voltage
Current
consumption(A)
Fuse (A)
Fust Cat. No.
IST-3075
120VAC, 60Hz
7.5
10
00CDE0005543
230VAC, 50/60Hz
4
8
00CDE0005544
IST-4075
120VAC, 60Hz
10.1
12
00CDE0005542
230VAC, 50/60Hz
5.5
8
00CDE0005544
IST-3075R
120VAC, 60Hz
7.5
10
00CDE0005543
230VAC, 50/60Hz
4
8
00CDE0005544
IST-4075R
120VAC, 60Hz
10.1
12
00CDE0005542
230VAC, 50/60Hz
5.5
8
00CDE0005544
Fuse is located in next to power switch. (Side of machine.)
Please replace the fuse according the below picture.
Caution to Electrical shock.
Before replace the fuse, turn the machine off and check
the power connection again.
If the power is connected, serious injury or death can be
occurred.
Step 1:
Press power switch and Power off
Step 2:
Connect (-) screw driver to fuse holder
Spin the fuse holder counterclock wise 50 ˚ then fuse holder is
disassembled.
Step 3:
Compare the fuse specification with the above table and replace it
if there is no problem.
Step 4:
Complete the replacement in reverse order of Step 2
퓨즈 교체
반시계방향으로
50도 회전
드라이버
삽입
6.1 Stop during the operating
Display
Message
Description
Solution
Temperature
Limit
Analog over temperature regulator
is active.
Check the set on the analog over
temperature regulator and the operating.
Restart after temperature is inside of the
range.
Load
Unbalance
Vibration is generating and the
normal shaking is not available. To
prevent the shaking system, stop
the shaking.
Check the unbalanced sample setting,
overload, un-flat ground.
Door Open
Door is opened over 5 minutes
during the operating.
Close the door and touch the start button
and restart the unit.
Sensor Fault
Temperature sensor is faulty so the
operating is stopped.
Contact the manufacturer or sales point.
Over
Current
Over current is generated so the
operating is stopped.
Check the overload, obstacle in the
shaking system.
6.2 Another problem and solution
Trouble
Probable Cause & Solution
The equipment is not
on
① Check the power plug connection to the socket
② Check if the circuit breaker is operated
③ When the fuse is short circuited, replace new one as enclosed.
④ Check the power failure
⑤ If the problem is not fixed with the above method, please request
A/S
Power on but
controller does not
operated
① Check the power plug connection to the socket
② Check the panel board circuit breaker and reset it then turn the
machine on
③ Request A/S if Display board or Main PCB has problem
① Request A/S if Power switch has problem.
NO The temperature of
incubator rise to setting
value
① Check the Temperature Control ON indicator is ON
② Check outdoor temperature is too high or low and maintain the right
room temperature.
③ Request A/S after check the difference between actual temperature
and display temperature.
④ Check the door switch irregularity
⑤ Request A/S if the heater is out of order
⑥ Request A/S if the freezer is out of order.
⑦ Check air filter condition and clean it according to 5.3
⑧ Auto Tuning if the temperature is unstable according to 4.4.4.
⑨ If the problem is not fixed with the above method, please request
A/S.
① If there is extremely intense vibration of the machine than usual
during operation, check the horizontality. If you need, please adjust
horizontality according 3.4.2
② If there is extremely intense vibration of the machine and shaking
speed is slow down, check the sample location and relocate the
samples harmoniously
③ In case of overloading, control the sample weight according to
9.1(Max load per speed).
④ Check door switch if shaking is not operated after shaking start
button press.
⑤ Check Timer close status if shaking is not operated after shaking
start button press, cancel timer set-up.
⑥ If the problem is not fixed with the above method, please request
A/S.
Noise
① Check fixed condition of platform or sample.
② Check the weight of sample is shifted to one side, relocate the
samples evenly.
③ If the problem is not fixed with the above method, please request
A/S.
7.1 Mountable maximum quantity
7.1.1 Universal Platform + Flask Clamp
Model
Flask Clamp
IST-3075 series
IST-4075 series
50ml
25
49
100ml
20
39
250/300ml
9
18
500ml
9
13
1,000ml
5
9
2,000ml
3
5
2,800ml
1
2
7.1.2 Universal Platform + Plastic Flask Clamp
Model
Flask Clamp
IST-3075 series
IST-4075 series
50ml
25
49
100/125ml
14
25
200ml
9
16
250ml
9
16
300ml
9
16
500ml
6
9
1,000ml
4
7
2,000ml
2
4
7.1.3 Universal Platform + Funnel Clamp
Model
Funnel Clamp
IST-3075 series
IST-4075 series
250ml
4
6
500ml
2
4
1,000ml
-
3
2,000ml
-
-
7.1.4 Universal Platform + Microplate Holder
Model
Type
IST-3075 series
IST-4075 series
Single
7
13
Tower
5
8
Flat A(large)
-
2
Flat B(small)
2
2
7.1.5 Universal Platform + Test Tube Rack
Model
Type
IST-3075 series
IST-4075 series
0°
2
4
15°
2
4
30°
1
3
45°
1
2
60°
1
1
7.1.6 Spring Wire Rack + Flask
Platform
Flask
IST-3075 series
IST-4075 series
50ml
16
25
100/125ml
9
16
250/300ml
4
9
500ml
4
5
1,000ml
2
4
2,000ml
1
2
2,800ml
1
2
※ About the right position for mounting maximum quantity of Accessories, please visit Jeio
Tech website or contact our office and distributor.
8.1 Monitoring Program installation
(1) Put the installation CD on the CD-ROM drive then the program is operated automatically.
Click the “Next” button go to the License Agreement.
(2) Check the details and click the “I agree”.
(3) Click the “Install” and install the program.
(4) Once the program is installed, below window is up and JeioTechSoft icon is generated
on the background.
Check the Run JeioTechSoft 2.0.0.5 or double click the icon then the program starts.
8.2 Communication Protocol
The unit communicates with PC through RS-232C port. Software save the duplex
communication, operating status, date. If you want to change on software, refer to
below reference.
Communicatioin Reference :
http://www.modbus.org/docs/Modbus_Application_Protocol_V1_1b3.pdf
8.2.1 Physical Layer
-
Communication port : RS-232C /USB
8.2.2 Model System number
ITEM
System Number
System
Model Number
Incubated
Shaker
1611H
Shaking incubator
IST-3075
1612H
Shaking incubator
IST-3075R
1613H
Shaking incubator
IST-4075
1614H
Shaking incubator
IST-4075R
8.2.3 Modbus Protocol Address Definition
modbus
function code
address
command
data
description
data length
word
byte
W/S
1
Beep
0x0001
BEEP_SYS_BOOT
(Communication Test)
1
2
0x0002
BEEP_KEY
(Communication Test)
1
2
R/I
2
MOD_SYS_NA
ME
x
Model name
1
2
R/I
3
MOD_SYS_VER
x
Firmware Version
1
2
R/I
4
MOD_SYS_PAR
AM
x
system parameter
20
40
W/S, R/I
25
MOD_SYS_SOU
ND
0/1
system sound
1
2
W/S, R/I
26
MOD_SYS_LAM
P
1 ~ 3
lamp bright
1
2
lamp auto onoff
1
2
W/S, R/I
28
MOD_AUTO_R
ESTART
0/1
system auto restart
1
2
W/M, R/I
29
MOD_TEMP_DE
VI
temp devi min ~
temp devi max
temperature deviation
4
8
W/S, R/I
33
MOD_TEMP_DE
VI_ALARM
0 ~ 3599
temp devi alarm time
1
2
W/S, R/I
34
MOD_RPM_DE
VI
rpm devi min ~
rpm devi max
rpm deviation
1
2
W/S, R/I
35
MOD_RPM_DE
VI_ALARM
0 ~ 3599
rpm devi alarm time
1
2
W/S, R/I
36
MOD_EXT_KEY
0 ~ 2
external key operation
1
2
W/M, R/I
37
MOD_SET_TEM
P
low temp lmt ~
high temp lmt
incubator set temp
4
8
W/S, R/I
41
MOD_SET_RPM
rpm min ~ rpm
max
shaker set rpm
1
2
W/M, R/I
42
MOD_SHAKER_
TIME
60L ~ 3599940L
shaker set time
2
4
W/S
44
MOD_SHAKER_
TIME_ONOFF
0 ~ 1
shaker timer set
1
2
W/S
45
MOD_INCUBAT
OR_RUN_STOP
0 ~ 1
incubator operation set
1
2
W/S
46
MOD_SHAKER_
RUN_STOP
0 ~ 1
shaker operation set
1
2
R/I
47
MOD_SYS_REP
ORT
X
system report
18
36
W/S
65
MOD_WARN_A
CK
1
system warning ack
1
2
W/S
66
MOD_SILENT_A
CK
1
system silent ack
1
2
W/S
67
MOD_AUTO_A
CK
1
auto run ack
1
2
8.2.4 Modbus Protocol Description
Command
Description
Beep
Beep generating
Modbus function
code
0
Write
data
1
System boot beep
W.S
Data
2
System key beep
MOD_SYS_NAME
Return the unit name
Refer to System
number
Modbus
function code
0
Return data
0
Upper
Lower
R.I
address
Model num
MOD_SYS_VER
Return the firmware
version
Modbus
function code
0
Return data
0
Upper
Lower
R.I
address
version
MOD_SYS_VER
Return the system
parameter
Modbus
function
code
0
Return
data
0 ~ 3
4 ~ 7
8 ~ 11
12 ~
15
high
temp
lmt
low
temp
lmt
temp_d
evi_ma
x
temp_d
evi_min
R.I
address
16
17
18
19
max
rpm
min
rpm
rpm_de
vi_max
rpm_de
vi_min
MOD_SYS_SOUND
Set the system sound
modbus function code
0
return
data
0
R.I
address
upper
lower
sound flag
modbus function code
0
r/w data
0
sound off
W.S
data
1
sound on
MOD_SYS_LAMP
Set the lamp
brightness
modbus function code
0
return
data
0
R.I
address
upper
lower
lamp bright
modbus function code
0
r/w data
1 ~ 3
W.S
data
1 : Min, 3 : Max
Automatic lamp on
setting
modbus function
code
0
return
data
0
R.I
addres
s
uppe
r
lower
lamp auto flag
modbus function
code
0
r/w
data
0
lamp auto
off
W.S
data
1
lamp auto
on
MOD_AUTO_RESTAR
T
Auto run setting
modbus function
code
0
return
data
0
R.I
addres
s
uppe
r
lower
auto run flag
modbus function
code
0
r/w
data
0
auto run
off
W.S
data
1
auto run on
MOD_TEMP_DEVI
temperature
deviation
modbus function
code
0
return
data
0 ~ 3
R.I
addres
s
temp devi
modbus function
code
0~3
value
temp devi min ~
max
W.M
value
MOD_TEMP_DEVI_ALAR
M
temp devi alarm
time
modbus function
code
0
return
data
0
R.I
addres
s
temp devi alarm
modbus function
code
0
value
0 ~ 3599(sec)
W.S
value
MOD_RPM_DEVI
rpm deviation
modbus function
0
return
0
rpm devi min ~
max
W.S
value
MOD_RPM_DEVI_ALAR
M
rpm devi alarm time
modbus function
code
0
return
data
0
R.I
addres
s
rpm devi alarm
modbus function
code
0
value
0 ~ 3599(sec)
W.S
value
MOD_EXT_KEY external key operation
modbus function code
0
return
data
0
R.I
address
ext key value
modbus function code
0
value
0
1
2
W.S
value
shaker incubator both
MOD_SET_TEMP
incubator set temp
modbus function
code
0
return
data
0 ~ 3
R.I
addres
s
set
temp
modbus function
code
0 ~ 3
value
high temp ~ low
temp
W.M
value
MOD_SET_RPM
shaker set rpm
modbus function
code
0
return
data
0
R.I
addres
s
set rpm
modbus function
code
0
value
rpm min ~ max
W.S
value
shaker set time
modbus function
code
0
return
data
0 ~ 1
R.I
addres
s
set
time
modbus function
code
0 ~ 1
value
0 ~
3599940L(min)
W.M
value
MOD_SHAKER_TIME_ONO
FF
shaker timer on/off
set
modbus function
code
0
data
0 : shaker timer
off
W.S
data
1 : shaker timer
on
MOD_INCUBATOR_RUN_S
TOP
incubator operation
set
modbus function
code
0
data
0 : incubator stop
W.S
data
1 : incubator start
MOD_SHAKER_RUN_STOP
incubator operation
set
modbus function
code
0
data
0 : shaker stop
W.S
data
1 : shaker start
MOD_SYS_RE
PORT
system report
modbus function
code
0
return
data
0 ~ 3
4 ~ 7
8
9
R.I
addre
ss
set
temp
act
temp
set
rpm
act
rpm
10 ~11
12 ~
13
14
~
15
16 ~
17
set
time
remai
n
time
syst
em
indi
cate
dum
my
system
indicate
4byt
e
MSB
30
29
28
27
26
25
24
dumm
y
dumm
y
dum
my
dum
my
usb_m
ode
temp
delay
auto
tune
sk over cur
0 :
report
Igonor
e
warn message clear
modbus function code
0
data
1: message clear
W.S
data
0: operation none
MOD_SILENT_ACK
Mute
modbus function code
0
data
1: beep stop
W.S
data
0: operation none
MOD_AUTO_ACK
auto run ack
modbus function code
0
data
1: auto run clear
W.S
data
0: operation none
9.1 Technical Specifications
MODEL
IST-3075
IST-3075R
IST-4075
IST-4075R
Temperature
Range
Amb.+5~80℃
Amb.-15 (Min.
+15)~80℃
Amb.+5~80℃
Amb.-15(Min.
+15)~80℃
Accuracy
±0.1℃ in flask at 37℃
Uniformity
±0.5℃ in flask at 37℃
Refrigerator (Hp)
-
1/6Hp
-
1/6Hp
Sensor Type
Pt 100Ω
Shaking
Motion
Orbital
Amplitude Size
(mm, dia.)
19.1
Speed Range
(RPM)
10 to 500
Accuracy
±1% of set speed (>100rpm) / ±1 (<100rpm)
Timer
1 min. to 999 hr 59 min.
Max. Load (kg)
10 at 500 rpm
15 at 400 rpm
14 at 500 rpm
21 at 400 rpm
Material
Internal
Stainless Steel
External
Epoxy powder coated steel
Platform
Anodized aluminum plate
Insulation
Polyethylene
Lid
PET
Heater
Incoloy, Fin Type (400W X 2EA)
Safety Device
CLS(Custom Logical Safe)-control system
Electronic temperature limiter, Hydraulic over-temperature limiter
Temperature deviation alarm (High/Low)
Communication Interface
USB, RS232
Dimension
Volume (L)
53
83
Platform (W×D)
350×350
450×450
440 x 785×510
540 x 890×510
Net Weight (Kg)
65
73
81
90
Electric requirements(230V)
50/60Hz, 4A
50/60Hz, 5.5A
50/60Hz, 4A
50/60Hz, 5.5A
Cat. No.
Electric requirements(110V)
60Hz , 7.5A
60Hz ,10.1A
60Hz , 7.5A
60Hz ,10.1A
Cat. No.
9.2 Circuit Diagrams
9.2.1 IST circuit diagram
9.2.2 External refrigeration system (IST-3075/IST-4075)
9.2.3 Internal refrigeration (IST-3075R/IST-4075R)
9.3 Disposing of products
Before you dispose product or the components
1. The equipment should be cleaned and decontaminated to protect workers
servicing the equipment, the environment or the public purchasing surplus
equipment because the incubated shaker can potentially be contaminated with
biological material, chemicals or radioisotopes. Check with your institution or
laboratory for individual policies and procedures for disposal of laboratory
equipment.
2. Please contact your local governing body for regulations regarding disposal
of electrical, electronic, metal (brass, aluminum, steel and stainless steel),
refrigeration and rubber components. Jeio Tech recommends the user find a
local scavenger or laboratory equipment recycler to properly dispose of the
unit and its components.
9.4 Warranty
9.4.1 Terms of Warranty Service
(1) Customer can get free warranty service for 2 years limited warranty from the date
of shipping date when the machine is broken while operating.
(2) When you ask for repairing, please check the below details first.
•
Date of purchase
•
Customer name / address / Phone number / E-mail
•
Fault status
7
9.4.2 Warranty exception
Customer can’t get free warranty service in case of as below.
•
If the product is broken due to the user’s fault.
•
If the product is broken due to improper operation or storage.
•
If the product is broken due to improper modify or repairing.
•
If the product is broken due to overuse of voltage or earthshock.
• If the product is broken without taking care of the “Notice” alerted
on the manual.
9.4.3 Service and technical advice
We, Jeiotech Co., Ltd. are doing best to give best support based on customer service system.
When we get the symptoms, fault states, contact number by customer, we offer after sales
service.
International Sales Head Office (Korea)
#1005, Byucksan Digital Valley 6-Cha, 481-4 Gasan-Dong, Geumcheon-Gu, Seoul 153-704,
Korea
Tel: +82 2 2627 3816
E-mail: [email protected]
The Americas (U.S.A. Branch)
19 Alexander Road, Unit #7. Billerica, MA 01821-5094, U.S.A.
Tel: +1 781 376 0700
E-mail: [email protected]
FAX: +1 781 376 0704
Europe (U.K. Branch)
Unit 3, Tower Industrial Park, Chalgrove, Oxfordshire, OX44 7XZ, United Kingdom
Tel: +44 1865 400321
E-mail: [email protected]
FAX: +44 1865 400736
China (Shanghai Branch)
B-705, Jingting Bldg. No.1000 Hongquan Rd, Minhang district,
Shanghai, P.R.CHINA 201103
Tel: +86-21-5108-9161, 5414-8389
Fax: +86-21-5168-5414
E-mail: [email protected]
South East Asia (Malaysia Branch)
No 57-59, Jalan Adenium 2G/6, Pusat Perniagaan Adenium, 48300 Bandar Bukit
Beruntung, Selangor Darul Ehsan , Malaysia
Tel: +60 3 6021 6880
E-mail: [email protected]
FAX: +60 3 6021 7880
Product is subject to change for improving quality and performance without an advance
notice.
The copyright of this manual belongs to Jeiotech.