Keysight Technologies X Series Manual 

Text of the manual

Keysight X-Series Signal Analyzers

This manual provides documentation for the following Analyzers: PXA Signal Analyzer N9030A

© Keysight Technologies, Inc.

2006-2015 No part of this manual may be reproduced in any form or by any means (including electronic storage and retrieval or translation into a foreign language) without prior agreement and written consent from Keysight Technologies, Inc. as governed by United States and international copyright laws. Trademark Acknowledgments Manual Part Number N9030-90018 Print Date January 2015 Published in USA Keysight Technologies 1400 Fountaingrove Parkway Santa Rosa, CA 95403 Warranty THE MATERIAL CONTAINED IN THIS DOCUMENT IS PROVIDED “AS IS,” AND IS SUBJECT TO BEING CHANGED, WITHOUT NOTICE, IN FUTURE EDITIONS. FURTHER, TO THE MAXIMUM EXTENT PERMITTED BY APPLICABLE LAW, KEYSIGHT DISCLAIMS ALL WARRANTIES, EITHER EXPRESS OR IMPLIED WITH REGARD TO THIS MANUAL AND ANY INFORMATION CONTAINED HEREIN, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE. KEYSIGHT SHALL NOT BE LIABLE FOR ERRORS OR FOR INCIDENTAL OR CONSEQUENTIAL DAMAGES IN CONNECTION WITH THE FURNISHING, USE, OR PERFORMANCE OF THIS DOCUMENT OR ANY INFORMATION CONTAINED HEREIN. SHOULD KEYSIGHT AND THE USER HAVE A SEPARATE WRITTEN AGREEMENT WITH WARRANTY TERMS COVERING THE MATERIAL IN THIS DOCUMENT THAT CONFLICT WITH THESE TERMS, THE WARRANTY TERMS IN THE SEPARATE AGREEMENT WILL CONTROL. Technology Licenses The hardware and/or software described in this document are furnished under a license and may be used or copied only in accordance with the terms of such license. Restricted Rights Legend If software is for use in the performance of a U.S. Government prime contract or subcontract, Software is delivered and licensed as “Commercial computer software” as defined in DFAR 252.227-7014 (June 1995), or as a “commercial item” as defined in FAR 2.101(a) or as “Restricted computer software” as defined in FAR 52.227-19 (June 1987) or any equivalent agency regulation or contract clause. Use, duplication or disclosure of Software is subject to Keysight Technologies’ standard commercial license terms, and non-DOD Departments and Agencies of the U.S. Government will receive no greater than Restricted Rights as defined in FAR 52.227-19(c)(1-2) (June 1987). U.S. Government users will receive no greater than Limited Rights as defined in FAR 52.227-14 (June 1987) or DFAR 252.227-7015 (b)(2) (November 1995), as applicable in any technical data. Safety Notices CAUTION A CAUTION notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in damage to the product or loss of important data. Do not proceed beyond a CAUTION notice until the indicated conditions are fully understood and met. WARNING A WARNING notice denotes a hazard. It calls attention to an operating procedure, practice, or the like that, if not correctly performed or adhered to, could result in personal injury or death. Do not proceed beyond a WARNING notice until the indicated conditions are fully understood and met.

Where to Find the Latest Information

Documentation is updated periodically. For the latest information about these products, including instrument software upgrades, application information, and product information, browse to one of the following URLs, according to the name of your product: http://www.keysight.com/find/pxa To receive the latest updates by email, subscribe to Keysight Email Updates at the following URL: http://www.keysight.com/find/MyKeysight Information on preventing analyzer damage can be found at: www.keysight.com/find/PreventingInstrumentRepair Is your product software up-to-date? Periodically, Keysight releases software updates to fix known defects and incorporate product enhancements. To search for software updates for your product, go to the Keysight Technical Support website at: http://www.keysight.com/find/techsupport

Table of Contents

1. Functional Tests Functional Test Versus Performance Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4 Contents of this Document. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5 Before Performing a Functional Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7 2. Displayed Average Noise Level (DANL) Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 3. Frequency Readout Accuracy Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17 4. Second Harmonic Distortion (SHD) Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21 5. Amplitude Accuracy at 50 MHz Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25 Testing Preamp Option (P03, P08, P13, P26, P43, P44, P50) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27 6. Frequency Response (Flatness) Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31 7. Frequency Response (Flatness), Preamp On Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37 8. Scale Fidelity Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43 9. BBIQ Input Frequency Response (Option BBA only) Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47

Keysight X-Series Signal Analyzers

PXA N9030A Functional Tests 1 Functional Tests Functional tests are tests of various instrument parameters that give a high degree of confidence that the analyzer is operating correctly. They are recommended as a check of analyzer operation for incoming inspection or after a repair. Measurement uncertainty analysis is not available for functional tests, and the analyzer is checked against limits that are wider than the published specifications. The functional tests are designed to test an analyzer operating within the temperature range defined by the analyzer specifications using a minimum set of test equipment. If a test does not pass, performance verification tests must be run to determine whether a problem exists.

4 Chapter 1

Functional Tests Functional Test Versus Performance Verification Functional Test Versus Performance Verification Functional tests use a minimum set of test equipment to check a much smaller range of parameters (and a limited number of data points for each parameter) than do performance verification tests. Functional tests use limits that are wider than the published specifications; measurement uncertainty analysis is not available for functional tests. NOTE If a functional test does not pass, you must run performance verification tests to determine whether a problem exists. Performance verification tests span a wide range of instrument parameters and provide the highest level of confidence that the instrument conforms to published specifications. These tests can be time consuming and require extensive test equipment.

Chapter 1 5

Functional Tests Contents of this Document Contents of this Document This chapter includes the following: — “Before Performing a Functional Test” on page 6 (what to do first). — “Test Equipment” on page 7 (a list of the equipment required for all of the tests). Subsequent chapters describe the following Functional Tests: — “Displayed Average Noise Level (DANL)” on page 11 — “Frequency Readout Accuracy” on page 15 — “Second Harmonic Distortion (SHD)” on page 19 — “Amplitude Accuracy at 50 MHz” on page 23 — “Frequency Response (Flatness)” on page 29 — “Frequency Response (Flatness), Preamp On” on page 35 — “Scale Fidelity” on page 41 Each functional test includes: — Test limits (pass/fail criteria) — A description of the test — The equipment required for the test — A figure showing how to connect the equipment — Step-by-step instructions — One or more tables in which to record the measurement results

6 Chapter 1

Functional Tests Before Performing a Functional Test Before Performing a Functional Test 1. Ensure that you have the proper test equipment. 2. Switch on the unit under test (UUT) and let it warm up (in accordance with warm-up requirements in the instrument specifications). 3. Allow sufficient warm-up time for the required test equipment (refer to individual instrument documentation for warm-up specifications). 4. Ensure that the analyzer’s frequency reference is set to Internal: a. Press the Input/Output, More, Freq Ref In keys. b. If the Freq Ref In softkey does not show Internal, press the Freq Ref In softkey and select Internal. 5. Following instrument warm-up, perform the auto align routine: Press System, Alignments, Align Now, All. NOTE Functional test accuracy depends on the precision of the test equipment used. Ensure that all of the test equipment is calibrated before running a functional test.

Chapter 1 7

Functional Tests Test Equipment Test Equipment The table below summarizes the test equipment needed to perform all of the functional tests. Alternate equipment model numbers are given in case the recommended equipment is not available. If neither the recommended nor the alternative test equipment are available, substitute equipment that meets or exceeds the critical specifications listed. Table 1-1 Analyzer Option Item Critical Specifications Recommended Keysight Model Alternate Keysight Model Adapters 503, 508, 513, 526 3.5 mm (f) to 3.5mm (f) (connector saver for source) Frequency: 10 MHz to 26.5 GHz VSWR: < 1.1:1 83059B 1250-1749 543, 544, 550 2.4 mm (f) to 2.4mm (f) (connector saver for source) Frequency: 10 MHz to 50 GHz VSWR: < 1.1:1 11900B 543, 544, 550 2.4 mm (f) to 3.5mm (f) Frequency: 10 MHz to 26.5 GHz VSWR: < 1.1:1 11901B All BNC (f) to SMA (m) Frequency: 40 MHz 1250-1200 All Type N (f) to Type N (f) Frequency: 10 MHz to 18 GHz VSWR: < 1.05:1 1250-1472 1250-0777 All Type N (m) to 3.5 mm (m) Frequency: 10 MHz to 18 GHz VSWR: < 1.1:1 1250-1743 All Type N (m) to 3.5 mm (f) Frequency: 10 MHz to 18 GHz VSWR: < 1.1:1 1250-1744 All Type N (f) to 3.5 mm (f) Frequency: 10 MHz to 18 GHz VSWR: < 1.1:1 1250-1745 All Type N (m) to BNC (f) Frequency: 10 MHz to 1 GHz VSWR: < 1.05:1 1250-1476 Attenuators All 10 dB Step Attenuator Frequency: 50 MHz Range: 0 to 70 dB 8495A 8496A All 10 dB Fixed Attenuator (2 required) Frequency: 50 MHz VSWR: < 1.2:1 8493C Option 010 8493A Option 010 or 8493B Option 010

8 Chapter 1

Functional Tests Test Equipment Preamp Options (Pxx) 30 dB Fixed Attenuator Accuracy: < 0.05 dB @ 50 MHz 11708A Cables All 3.5 mm (1 meter) Frequency: 10 MHz to 26.5 GHz VSWR: < 1.4:1 Loss: < 2.0 dB 11500E 8120-4921 543, 544, 550 2.4 mm (f) to 2.4 mm (m) (1 meter) Frequency: 10 MHz to 50 GHz VSWR: < 1.55:1 at 50 GHz Loss: < 6.0 dB 8120-6164 BBA Cable, BNC BNC male Length: 23 cm (9 in) max 10502A All Cable, BNC (3 required) 120 cm (48 in.) BNC cable 10503A Signal Source All Synthesized Sweeper Frequency: 10 MHz to 50 GHza Harmonic level: < − 40 dBc Amplitude range: 10 to − 20 dBm Frequency Accuracy: 0.02% PSGb 83630B, 83640B, 83650B BBA RF Signal Generator Frequency: 250 kHz to 40 MHz Amplitude: ~–10 dBm Flatness: < ±0.75 dB E4438C 33250A Power Meter All Power Meter Power Reference Accuracy: ±1.2% Compatible with power sensor E4418B E4419B All Power Sensor Frequency Range: 50 MHz to 3.66 GHz Amplitude Range: −70 to –20 dBm 8481D 8487D 8485D 503, 508, 513 Power Sensor Frequency Range: 50 MHz to 13.66 GHz Amplitude Range: −30 to +20 dBm 8481A 526 Power Sensor Frequency Range: 50 MHz to 26.5 GHz Amplitude Range: −30 to +20 dBm 8485A 543, 544, 550 Power Sensor Frequency Range: 50 MHz to 50 GHz Amplitude Range: −30 to +20 dBm 8487A P03 Low Power Sensor Frequency Range: 50 MHz to 3.6 GHz Amplitude Range: –70 to –20 dBm 8481D 8485D 8487D Table 1-1 Analyzer Option Item Critical Specifications Recommended Keysight Model Alternate Keysight Model

Chapter 1 9

Functional Tests Test Equipment Preamp other than P03 Low Power Sensor Frequency Range: 50 MHz to 50 GHzc Amplitude Range: –70 to –20 dBm 8487D 8485D Miscellaneous Equipment All Filter, 50 MHz Low Pass Cutoff Frequency: 50 MHz Rejection at 65 MHz: > 40 dB Rejection at 75 MHz: > 60 dB Rejection at 80 MHz: > 60 dB Frequency: 10 MHz to 26.5 GHz 0955-0306 526 Power Splitter, 3.5 mm Nominal Insertion Loss: 6 dB Tracking Between Ports: < 0.25 dB 11667B 543, 544, 550 Power Splitter, 2.4 mm Nominal Insertion Loss: 8.5 dB Tracking Between Ports: < 0.4 dB 11667C 503, 508, 513 Power Splitter Nominal Insertion Loss: 6 dB Tracking Between Ports: < 0.25 dB 11667A 11667B 503, 508, 513, 526 Termination, 50Ω Type N (m) Connector Frequency: 30 Hz to 26.5 GHz 909A Option 012 544, 545, 550 Termination, 50Ω 2.4 mm (f) Connector Frequency: 30 Hz to 50 GHz 85138B BBA Termination, 50Ω Connector: BNC male Frequency: 250 kHz to 40 MHz 11593A a. Maximum frequency equal to or greater than the maximum frequency of the PXA. b. PSG model numbers: E8244A, E8254A, E8247C Option 520, E8247C Option H31 or 540, E8257D Option 520, E8257D Option 550, E8257D Option 567, E8267D Option 544. c. Maximum frequency equal to or greater than the maximum frequency of the preamplifier option. Table 1-1 Analyzer Option Item Critical Specifications Recommended Keysight Model Alternate Keysight Model

10 Chapter 1

Functional Tests Test Equipment

Keysight X-Series Signal Analyzers

PXA N9030A Functional Tests 2 Displayed Average Noise Level (DANL) Test Limits (with 0 dB input attenuation) See Table 2-1 on page 13 for values. The Displayed Average Noise Level (DANL) of the signal analyzer is measured across a 10 kHz frequency span at several center frequencies. The analyzer input is terminated into a 50Ω load. A test is performed to assure the measurement is not performed in the presence of a residual response. The measurement is then averaged, and the result is normalized to a 1 Hz bandwidth. Figure 2-1 DANL Test Setup Item Critical Specifications (for this test) Recommended Keysight Model Termination, 50Ω Type-N(m)a Frequency: 10 MHz to 18 GHz 909A Option 012 Termination, 50Ω 2.4 mm (f)b Frequency: 10 MHz to 50 GHz 85138B a. for Options 503, 508, 513, and 526 b. for Options 543, 544, and 550

12 Chapter 2

Displayed Average Noise Level (DANL) Procedure Procedure 1. Configure the equipment as shown in Figure 2-1 2. Press Mode, Spectrum Analyzer, Mode Preset on the analyzer. 3. Set up the signal analyzer by pressing: FREQ Channel, Center Freq, 10, MHz Input/Output, RF Input, RF Coupling, select DC (except Options 543, 544, and 550) SPAN X Scale, Span, 10, kHz AMPTD Y Scale, –70, dBm AMPTD Y Scale, Attenuation, MechAtten, 0, dB BW, Res BW, 1, kHz BW, Video BW, 100, Hz Meas Setup, Average/Hold, Number, 20, Enter Trace/Detector, Trace Average Single 4. Press Restart, then wait for Average/Hold to display 20/20. 5. Press View/Display, Display, then press Display Line, On. 6. Rotate the knob and set the display line at the average amplitude of the displayed noise floor by visual inspection. 7. Confirm that the measurement is performed on the analyzer noise floor and not on a residual response within the displayed 10 kHz span. NOTE Ignore the residual response if one appears when taking the measurement. 8. Enter the value of the display line as the Measured Average Noise Level at 10 MHz column in Table 2-1. 9. Normalize the measured value to a 1 Hz BW by adding 30 dB to the measured value. NOTE The 30 dB value is added because the formula used to calculate the value of the noise power in a 1 Hz BW when measured with a 1 kHz BW is: Normalized Noise = 10 Log (BW 2/BW 1) where BW 2 is the 1 kHz BW we measure and BW 1 is 1 Hz BW to which we want to normalize. Therefore, 10 Log (1000) = 30 dB, so the noise floor will be 30 dB lower in a 1 Hz BW. 10.Enter the normalized value of the displayed average noise level in Table 2-1.

Chapter 2 13

Displayed Average Noise Level (DANL) Procedure 11.The value of the normalized displayed average noise should be less than the specification value. 12.Change the analyzer center frequency to the next value listed in Table 2-1. Press: FREQ Channel, Center Freq, [Table 2-1 Value], GHz 13.Repeat step 4 through step 12 to fill in the remainder of Table 2-1 for your analyzer frequency range. Table 2-1 Displayed Average Noise Level (DANL) Results Center Frequency Measured Average Noise Level (dBm) Normalized Average Noise Level/(1 Hz BW) (dBm) Test Limits (dBm) Options 503, 508, 513, 526 Test Limits (dBm) Options 543, 544, 550 10 MHz –153 –153 2 GHz –152 –152 6 GHz –146 –142 13 GHz –147 –145 20 GHz –140 –139 26.5 GHz –135 –137 30 GHz N/A –136 35 GHz N/A –131 40 GHz N/A –129 43.5 GHz N/A –126 50 GHz N/A –126

14 Chapter 2

Displayed Average Noise Level (DANL) Procedure

Keysight X-Series Signal Analyzers

PXA N9030A Functional Tests 3 Frequency Readout Accuracy Test Limits Frequency Readout Accuracy is equivalent to the following equation: See Table 3-1 on page 17 for actual values. The frequency readout accuracy is measured in several spans and center frequencies that allow both internal analyzer synthesizer modes and prefilter bandwidths to be tested. Frequency reference error is eliminated by using the same frequency standard for the analyzer and signal source. 0.25% span 5% RBW 2 Hz 0.5 horizontal resolution × + + × + × ( ) ± Item Critical Specification (for this test) Recommended Keysight Model Adapter, Type-N (m), to 3.5 mm (f)a Frequency: 10 MHz to 1.51 GHz VSWR: < 1.1:1 1250-1744 Adapter, 3.5 mm (f) to 3.5 mm (f)a Frequency: 10 MHz to 1.51 GHz VSWR: < 1.1:1 83059B Adapter, 2.4 mm (f) to 2.4 mm (f)b Frequency: 10 MHz to 1.51 GHz VSWR: < 1.1:1 11900B Cable, 2.4 mm, 1 meterb Frequency: 10 MHz to 1.51 GHz VSWR: < 1.4:1 8120-6164 Cable, 3.5 mm, 1 metera Frequency: 10 MHz to 1.51 GHz VSWR: < 1.4:1 11500E Cable, BNC, 120 cm Frequency: 10 MHz 10503A Synthesized Sweeper Frequency: 10 MHz to 1.51 GHz PSG a. for Options 503, 508, 513, 526 b. for Options 543, 544, 550

16 Chapter 3

Frequency Readout Accuracy Figure 3-1 Frequency Readout Accuracy Test Setup

Chapter 3 17

Frequency Readout Accuracy Procedure Procedure 1. Configure the equipment as shown in Figure 3-1. Confirm the analyzer’s built-in auto alignment has been performed within the past 24 hours. 2. On the synthesized sweeper, press PRESET, then set the controls as follows: FREQUENCY, 1505, MHz POWER LEVEL, –10, dBm RF, On 3. Set up the signal analyzer by pressing: Mode, Spectrum Analyzer Mode Preset Input/Output, More, Freq Ref In, External FREQ Channel, Center Freq, 1505, MHz SPAN X Scale, Span, 2990, MHz Trace/Detector, More, Detector, Sample Single 4. Press Restart. Press Peak Search on the analyzer. If the instrument is functioning correctly, the marker reading in the active function block will be between the values listed in Table 3-1. Record the marker value in the Marker Frequency Readout column in Table 3-1. 5. On the signal analyzer, change the span and center frequency as listed in Table 3-1. 6. Change the synthesized sweeper frequency to match the center frequency of the analyzer. 7. Repeat step 4 through step 6 until the Marker Frequency Readout column of Table 3-1 is complete. Table 3-1 Frequency Readout Accuracy Results Span (MHz) Center Frequency (MHz) Minimum Marker Frequency Readout Maximum 2990 1505 1495.9 MHz 1514.1 MHz 127.2 1505 1504.56 MHz 1505.44 MHz 54.1 1505 1504.8122 MHz 1505.1878 MHz 7.95 1505 1504.97240 MHz 1505.0276 MHz 0.106 1505 1504.999630 MHz 1505.000370 MHz 1.98 517.59 517.58316 MHz 517.59684 MHz 1.98 832.50 832.49316 MHz 832.50684 MHz

18 Chapter 3

Frequency Readout Accuracy Procedure

Keysight X-Series Signal Analyzers

PXA N9030A Functional Tests 4 Second Harmonic Distortion (SHD) Test Limits This test checks the second harmonic distortion of the signal analyzer by tuning to twice the input frequency and examining the level of the distortion product. A low pass filter is inserted between the source and the signal analyzer to prevent the source second harmonic from artificially raising the second harmonic product displayed on the analyzer. Applied Frequency Mixer Levela Distortion 40 MHz –15 dBm < –60 dBc a. Mixer Level = Input Level - RF Attenuation Item Critical Specifications (for this test) Recommended Keysight Model Adapter, Type-N (m) to BNC (f) Frequency: 10 MHz to 80 MHz VSWR: < 1.05:1 1250-1476 Adapter, BNC (f) to SMA (m) Frequency: 40 to 80 MHz 1250-1200 Adapter, 3.5 mm (f) to 3.5 mm (f)a Frequency: 10 MHz to 80 MHz VSWR: < 1.1:1 83059B Adapter, 2.4 mm (f) to 3.5 mm (f)b Frequency: 10 MHz to 80 MHz VSWR: < 1.1:1 11901B Cable, BNC 120 cm (2 required) Frequency: 10 to 40 MHz 10503A Filter, 50 MHz Low Pass Cutoff Frequency: 50 MHz Rejection at 65 MHz: > 40 dB Rejection at 75 MHz: > 60 dB 0955-0306 Synthesized Sweeper Frequency: 50 MHz Spectral Purity: Better than –30 dBc PSG a. for Options 503, 508, 513, 526

20 Chapter 4

Second Harmonic Distortion (SHD) Figure 4-1 Second Harmonic Distortion Test Setup b. for Options 543, 544, 550

Chapter 4 21

Second Harmonic Distortion (SHD) Procedure Procedure 1. Configure the equipment as shown in Figure 4-1 2. Press Mode, Spectrum Analyzer, Mode Preset on the signal analyzer and Preset the synthesized sweeper. 3. Set up the synthesized sweeper by pressing: Frequency, 40, MHz Amplitude, –5, dBm 4. Set up the signal analyzer by pressing: Input/Output, More, Freq Ref In, External FREQ Channel, Center Freq, 40, MHz SPAN X Scale, Span, 1, MHz AMPTD Y Scale, Attenuation, Mech Atten, 10, dB 5. On the analyzer, press Peak Search. 6. Adjust the synthesized sweeper amplitude for a signal analyzer display of –5 dBm ±0.1 dB. 7. On the signal analyzer, activate the marker delta function by pressing the Marker and Delta keys. 8. On the signal analyzer, press: FREQ Channel, Center Freq, 80, MHz Meas Setup, Average/Hold Number, 20, Enter Trace/Detector, Trace Average Single 9. Press Peak Search. Enter the displayed value under the Measured Second Harmonic Distortion (dBc) heading in Table 4-1. Table 4-1 Second Harmonic Distortion Results Applied Frequency (MHz) Measured Second Harmonic Distortion (dBc) Mixer Level (dBm) Specification (dBc) 40 –15 –60

22 Chapter 4

Second Harmonic Distortion (SHD) Procedure

Keysight X-Series Signal Analyzers

PXA N9030A Functional Tests 5 Amplitude Accuracy at 50 MHz Test Limits Amplitude Accuracy should remain within 1.13 dB of the measured source value across the range of source levels and changes in resolution bandwidth. The Preamp (option P03, P08, P13, P26, P43, P44, P50) should remain within ±1.3 dB of measured values. A synthesized sweeper is used as the signal source for the test. The source amplitude is varied using the signal source amplitude control. The attenuation and resolution bandwidth are varied on the signal analyzer. The source amplitude is measured by the power meter and signal analyzer at each setting, and the values compared. The difference between each pair of measurements indicates the amplitude accuracy. Item Critical Specifications Recommended Keysight Model Adaptera Type-N (m), to 3.5 mm (m) 1250-1743 Adaptera 3.5 mm (f) to 3.5 mm (f) 83059B Adapterb 2.4 mm (f) to 3.5 mm (m) 11901A Adapterb 2.4 mm (f) to 3.5 mm (f) 11901B Attenuator, 30 dB Accuracy: < 0.5 dB at 50 MHz 11708A Cable 3.5 mm, 1 meter 11500E Cable BNC, 120 cm 10503A Power Meter Compatible with power sensor E4418B Power Sensor Amplitude Range: −25 dBm to 10 dBm 8485A Power Splitter 3.5 mm (f) connectors 6 dB loss 11667B Synthesized Sweeper Typical Temperature Stability: 0.01 dBc/°C PSG a. for Options 503, 508, 513, 526 b. for Options 543, 544, 550

24 Chapter 5

Amplitude Accuracy at 50 MHz Figure 5-1 Amplitude Accuracy Test Setup

Chapter 5 25

Amplitude Accuracy at 50 MHz Procedure Procedure 1. Zero and calibrate the power meter. 2. Configure equipment as shown in Figure 5-1, with the power splitter connected directly to the signal analyzer input through the adapter. CAUTION To minimize stress on the test equipment connections, support the power sensor. 3. If the auto alignment for the analyzer has not been performed within the past 24 hours, press System, Alignments, Align Now, All to perform the auto alignment routine. 4. Press Mode, Spectrum Analyzer, Mode Preset on the analyzer. 5. Set up the synthesized sweeper by pressing: CW, 50, MHz Power Level, -4, dBm RF (On) 6. Set up the signal analyzer by pressing: Input/Output, More, Freq Ref In, External FREQ Channel, Center Freq, 50, MHz SPAN X Scale, 2, MHz AMPTD Y Scale, Attenuation, Mech Atten, 10, dB Input/Output, RF Input, RF Coupling, select DC (except Options 543, 544, 550) Sweep/Control, Sweep Setup, Swp Time Rules, SA - Accuracy Meas Setup, Average/Hold Number, 20, Enter Trace/Detector, Trace Average Single 7. Perform the following steps for each row listed in Table 5-1: a. Set the synthesized sweeper amplitude to the value listed in the Nominal Source Amplitude column in Table 5-1 b. Set the Mech Atten as indicated in the Attenuation column in Table 5-1 c. Set the Span as listed in the Span column of Table 5-1 d. Record the synthesized sweeper amplitude, as measured by the power meter, in the Power Meter Amplitude column of Table 5-1 e. On the signal analyzer, press Restart. f. Wait for the signal analyzer to finish averaging. g. Press Peak Search.

26 Chapter 5

Amplitude Accuracy at 50 MHz Procedure h. Record the signal amplitude, as measured by the analyzer in the Measured Amplitude column of Table 5-1 i. Calculate the signal amplitude accuracy error using the following equation, and record the results under the Amplitude Accuracy Error column: Amplitude Accuracy Error = Meas_amp − Power_meter Table 5-1 Amplitude Accuracy Results Nominal Source Amplitude (dBm) Attenuation (dB) Span (MHz) Measured Amplitude Meas_amp (dBm) Power Meter Amplitude Power_meter (dBm) Amplitude Accuracy Error (dB) Test Limit (dB) −4 10 2 ±1.13 dB −9 10 1 ±1.13 dB −14 10 0.5 ±1.13 dB −4 20 0.1 ±1.13 dB −14 20 0.1 ±1.13 dB −4 30 0.1 ±1.13 dB −14 30 0.1 ±1.13 dB

Chapter 5 27

Amplitude Accuracy at 50 MHz Testing Preamp Option (P03, P08, P13, P26, P43, P44, P50) Testing Preamp Option (P03, P08, P13, P26, P43, P44, P50) Instruments containing Options P03, P08, P13, P26, P43, P44, P50 must have the preamp function turned on and tested. Procedure 1. On the analyzer, press AMPTD Y Scale, More, Internal Preamp, Low Band. 2. Connect the equipment as shown in Figure 5-1 on page 24, using a 30 dB Pad between the adaptor and the signal analyzer input. 3. Set the synthesized sweeper amplitude to the value listed in the Nominal Source Amplitude column in Table 5-2 4. Set the signal analyzer input attenuation to 0 dB. 5. Set the Span as listed in Table 5-2 6. Record the synthesized sweeper amplitude, as measured by the power meter, in the Power Meter Amplitude column of Table 5-2 7. On the signal analyzer, press Restart. 8. Wait for the analyzer to finish averaging. 9. Press Peak Search. 10.Record the signal amplitude as measured by the analyzer in the measured amplitude column of Table 5-2 11.Calculate the signal amplitude accuracy using the following equation: Amplitude Accuracy Error = Meas_amp + 30 dB − Power_meter 12.Record the results under the Amplitude Accuracy Error column of Table 5-2 Table 5-2 Amplitude Accuracy Results (Option P03, P08, P13, P26, P43, P44, P50) Nominal Source Amplitude (dBm) Low- band Preamp Res BW (kHz) Span (kHz) Measured Amplitude Meas_amp (dBm) Power Meter Amplitude Power_meter (dBm) Amplitude Accuracy Error (dB) Test Limit (dB) −13 ON 1 106 ±1.30 dB

28 Chapter 5

Amplitude Accuracy at 50 MHz Testing Preamp Option (P03, P08, P13, P26, P43, P44, P50)

Keysight X-Series Signal Analyzers

PXA N9030A Functional Tests 6 Frequency Response (Flatness) Test Limits The frequency response test measures the signal analyzer’s amplitude error as a function of the tuned frequency. Measurements are made ranging from 50 MHz to the maximum frequency range of your analyzer. The signal source amplitude is measured with a power meter to eliminate error due to source flatness. The measured value is normalized to 50 MHz. Frequency Range Limit Relative to 50 MHz 20 Hz to 3.6 GHz ±1.5 dΒ > 3.6 GHz to 8.4 GHz ±2.5 dΒ > 8.4 GHz to 13.6 GHz ±3.0 dΒ > 13.6 GHz to 26.5 GHz ±3.5 dΒ > 26.5 GHz to 34.5 GHz ±3.5 dΒ > 34.5 GHz to 50 GHz ±4.2 dΒ Item Critical Specifications (for this test) Recommended Keysight Model Adapter, Type N (m) to 3.5 mm (m)a Frequency: 10 MHz to 18 GHz VSWR: < 1.1:1 1250-1743 Adapter, 3.5 mm (f) to 3.5 mm (f)a Frequency: 10 MHz to 26.5 GHz VSWR: < 1.1:1 83059B Cable, 3.5 mm, 1 metera (Opt 503, 508, 513, 526) Frequency: 10 MHz to 26.5 GHz VSWR: < 1.4:1 11500E Cable, 2.4 mm, 1 meterb (Opt 543, 544, 550) Frequency: 10 MHz to 50 GHz VSWR: < 1.55:1 8120-6164

30 Chapter 6

Frequency Response (Flatness) Figure 6-1 Frequency Response Test Setup Cables, BNC 120 cm Frequency: 10 MHz 10503A Power Meter Compatible with power sensor E4418B Power Sensor Frequency Range: 50 MHz to 26.5 GHz 8485A Power Sensorb Frequency Range: 50 MHz to 50 GHz 8487A Power Splittera Frequency Range: 50 MHz to 26.5 GHz Tracking between ports: < 0.25 dB 11667B Power Splitterb Frequency Range: 50 MHz to 50 GHz Tracking between ports: < 0.4 dB 11667C Synthesized Sweeper Frequency Range: 50 MHz to maximum frequency of PXA PSG a. for Options 503, 508, 513, 526 b. for Options 543, 544, 550 Item Critical Specifications (for this test) Recommended Keysight Model

Chapter 6 31

Frequency Response (Flatness) Procedure Procedure 1. Zero and calibrate the power meter and power sensor as described in the power meter operation manual. 2. Configure the equipment as shown in Figure 6-1. NOTE Connect the power splitter to the signal analyzer input using the appropriate adapter. Do not use a cable. 3. Assure the signal analyzer’s built-in auto alignment has been performed within the last 24 hours. 4. Press Mode, Spectrum Analyzer, Mode Preset on the signal analyzer, and press Preset on the synthesized sweeper. 5. Set up the synthesized sweeper by pressing: CW, 50, MHz Power level, –4, dBm 6. Set up the signal analyzer by pressing: Input/Output, More, Freq Ref In, External FREQ Channel, Center Freq, 50, MHz SPAN X Scale, Span, 50, kHz AMPTD Y Scale, Ref Level, 0, dBm 7. Adjust the synthesized sweeper output power for a power meter reading of –10 dBm ±0.1 dB. 8. On the signal analyzer, press Single. 9. Press the Peak Search key on the signal analyzer to position the marker on the peak of the signal. 10.Refer to Table 6-1, “Frequency Response (Flatness) Results.” Enter the amplitude of the signal displayed on the signal analyzer into the MeasAmp column of Table 6-1. 11.Enter the power meter reading into the PowerMeter column of Table 6-1. 12.Tune the synthesized sweeper and signal analyzer to the next frequency listed in Table 6-1. 13.Enter the power sensor calibration factor into the power meter. 14.For frequencies 3.6 GHz and above, press AMPTD Y Scale, then Presel Center to center the preselector filter for an optimum amplitude measurement. 15.Repeat step 7 through step 14 and complete the remainder of Table 6-1 for the frequency range of your analyzer.

32 Chapter 6

Frequency Response (Flatness) Procedure 16.Compute the measurement error (MeasError = MeasAmp − PowerMeter). 17.Compute the flatness error normalized to 50 MHz: (MeasError − MeasError @ 50 MHz) 18.Enter the computed flatness error value into the FlatNorm column of Table 6-1. 19.Compare the value of FlatNorm to the test limit. Table 6-1 Frequency Response (Flatness) Results Center Frequency Analyzer Amplitude Measamp Power Meter Measurement Powermeter Meas Error Measerror Flatness Normalized to 50 MHz FlatNorm Flatness Error Test Limits (dB) 50 MHz 0 Ref 1 GHz ±1.5 dB 2 GHz 1.5 dB 3 GHz ±1.5 dB 3.5 GHz ±1.5 dB 4 GHz ±2.5 dB 6 GHz ±2.5 dB 8 GHz ±2.5 dB 9 GHz ±3.0 dB 11 GHz ±3.0 dB 13 GHz ±3.0 dB 14 GHz ±3.5 dB 17 GHz ±3.5 dB 20 GHz ±3.5 dB 23 GHz ±3.5 dB 26 GHz ±3.5 dB 29 GHz ±3.5 dB 32 GHz ±3.5 dB 35 GHz ±4.2 dB 38 GHz ±4.2 dB 41 GHz ±4.2 dB 42.5 GHz ±4.2 dB

Chapter 6 33

Frequency Response (Flatness) Procedure 43.5 GHz ±4.2 dB 46.5 GHz ±4.2 dB 49.5 GHz ±4.2 dB Table 6-1 Frequency Response (Flatness) Results Center Frequency Analyzer Amplitude Measamp Power Meter Measurement Powermeter Meas Error Measerror Flatness Normalized to 50 MHz FlatNorm Flatness Error Test Limits (dB)

34 Chapter 6

Frequency Response (Flatness) Procedure

Keysight X-Series Signal Analyzers

PXA N9030A Functional Tests 7 Frequency Response (Flatness), Preamp On Test Limits The frequency response test, with preamplifier on, measures the signal analyzer’s amplitude error as a function of the tuned frequency. Measurements are made ranging from 50 MHz to the maximum frequency range of the preamp. The signal source amplitude is measured with a power meter to eliminate error due to source flatness. The measured value is normalized to 50 MHz. Frequency Range Limit Relative to 50 MHz 100 kHz to 3.6 GHz ±2.0 dΒ > 3.6 GHz to 8.4 GHz ±3.0 dΒ > 8.4 GHz to 13.6 GHz ±3.5 dΒ > 13.6 GHz to 26.5 GHz ±4.0 dΒ > 26.5 GHz to 34.5 GHz ±4.5 dΒ > 34.5 GHz to 50 GHz ±6.0 dΒ Item Critical Specifications (for this test) Recommended Keysight Model Adapter, Type N (m) to 3.5 mm (m)a Frequency: 10 MHz to 18 GHz VSWR: < 1.1:1 1250-1743 Adapter, 3.5 mm (f) to 3.5 mm (f)a Frequency: 10 MHz to 26.5 GHz VSWR: < 1.1:1 83059B Cable, 3.5 mm, 1 metera Frequency: 10 MHz to 26.5 GHz VSWR: < 1.4:1 11500E

36 Chapter 7

Frequency Response (Flatness), Preamp On Figure 7-1 Frequency Response Test Setup Cable, 2.4 mm, 1 meterb Frequency: 10 MHz to 50 GHz VSWR: < 1.55:1 8120-6164 Cables, BNC 120 cm Frequency: 10 MHz 10503A Power Meter Compatible with power sensor E4418B Attenuator, 30 dB Fixed Frequency: 50 MHz Accuracy: ±0.05 dB 11708A Power Sensora Frequency Range: 50 MHz to 26.5 GHz Amplitude Range: –65 dBm to –55 dBm 8485D Power Sensorb Frequency Range: 50 MHz to 50 GHz Amplitude Range: –65 dBm to –55 dBm 8487D Power Splittera Frequency Range: 50 MHz to 26.5 GHz Tracking between ports: < 0.25 dB 11667B Power Splitterb Frequency Range: 50 MHz to 50 GHz Tracking between ports: < 0.4 dB 11667C Synthesized Sweeper Frequency Range: 50 MHz to maximum frequency of the PXA PSG a. for Options P03, P08, P13, P26 b. for Options P43, P44, P50 Item Critical Specifications (for this test) Recommended Keysight Model

Chapter 7 37

Frequency Response (Flatness), Preamp On Procedure Procedure 1. Zero and calibrate the power meter and power sensor as described in the power meter operation manual. 2. Configure the equipment as shown in Figure 7-1. NOTE Connect the power splitter to the signal analyzer input using the appropriate adapter. Do not use a cable. 3. Assure the signal analyzer’s built-in auto alignment has been performed within the last 24 hours. 4. Press Mode, Spectrum Analyzer, Mode Preset on the signal analyzer, and press Preset on the synthesized sweeper. 5. Set up the synthesized sweeper by pressing: CW, 50, MHz Power level, –54, dBm 6. Set up the signal analyzer by pressing: Input/Output, More, Freq Ref In, External FREQ Channel, Center Freq, 50, MHz SPAN X Scale, Span, 50, kHz AMPTD Y Scale, More, Internal Preamp, Full Range AMPTD Y Scale, Attenuation, Mech Atten, 0, dB AMPTD Y Scale, Ref Level, –55, dBm 7. Adjust the synthesized sweeper output power for a power meter reading of –60 dBm ±0.1 dB. 8. On the signal analyzer, press Single. 9. Press the Peak Search key on the signal analyzer to position the marker on the peak of the signal. 10.Refer to Table 7-1, “Frequency Response (Flatness) Results.” Enter the amplitude of the signal displayed on the signal analyzer into the MeasAmp column of Table 7-1. 11.Enter the power meter reading into the PowerMeter column of Table 7-1. 12.Tune the synthesized sweeper and signal analyzer to the next frequency listed in Table 7-1. 13.Enter the power sensor calibration factor into the power meter. 14.For frequencies 3.6 GHz and above, press AMPTD Y Scale, then Presel Center to center the preselector filter for an optimum amplitude measurement.

38 Chapter 7

Frequency Response (Flatness), Preamp On Procedure 15.Repeat step 7 through step 14 and complete the remainder of Table 7-1 for the frequency range of your preamp. 16.Compute the measurement error (MeasError = MeasAmp − PowerMeter). 17.Compute the flatness error normalized to 50 MHz: (MeasError − MeasError @ 50 MHz) 18.Enter the computed flatness error value into the FlatNorm column of Table 7-1. 19.Compare the value of FlatNorm to the test limit. Table 7-1 Frequency Response (Flatness) Results Center Frequency Analyzer Amplitude Measamp Power Meter Measurement Powermeter Meas Error Measerror Flatness Normalized to 50 MHz FlatNorm Flatness Error Test Limits (dB) 50 MHz 0 Ref 1 GHz ±2.0 dB 2 GHz ±2.0 dB 3 GHz ±2.0 dB 3.5 GHz ±2.0 dB 4 GHz ±3.0 dB 6 GHz ±3.0 dB 8 GHz ±3.0 dB 9 GHz ±3.5 dB 11 GHz ±3.5 dB 13 GHz ±3.5 dB 14 GHz ±4.0 dB 17 GHz ±4.0 dB 20 GHz ±4.0 dB 23 GHz ±4.0 dB 26 GHz ±4.0 dB 29 GHz ±4.0 dB 32 GHz ±4.0 dB 35 GHz ±5.0 dB 38 GHz ±5.0 dB 41 GHz ±5.0 dB

Chapter 7 39

Frequency Response (Flatness), Preamp On Procedure 42.5 GHz ±5.0 dB 43.5 GHz ±5.0 dB 46.5 GHz ±5.0 dB 49.5 GHz ±5.0 dB Table 7-1 Frequency Response (Flatness) Results Center Frequency Analyzer Amplitude Measamp Power Meter Measurement Powermeter Meas Error Measerror Flatness Normalized to 50 MHz FlatNorm Flatness Error Test Limits (dB)

40 Chapter 7

Frequency Response (Flatness), Preamp On Procedure

Keysight X-Series Signal Analyzers

PXA N9030A Functional Tests 8 Scale Fidelity Test Limits The scale fidelity error will be ≤ ±1.0 dB with ≤ −10 dBm at the mixer. This test checks the scale fidelity of the instrument by maintaining a constant reference level and measuring signals of different amplitudes over most of the display range. This test sets the input attenuator to 10 dB and the Reference Level to 0 dBm. The external attenuator is set to 0 dB, and the amplitude of the source is adjusted to set the displayed signal at the reference level. The instrument’s internal marker is used to measure the reference amplitude. The Marker Delta function is activated and the RF input is reduced using the external precision step attenuator. Signal input levels from 0 dBm to −50 dBm are measured. Item Critical Specifications (for this test) Recommended Keysight Model Adapter, Type-N (m), to 3.5 mm (f)a Frequency: 50 MHz VSWR: < 1.1:1 1250-1745 Adapter, 3.5 mm (f) to 3.5 mm (f)a Frequency: 50 MHz VSWR: < 1.1:1 83059B Adapter, 2.4 mm (f) to 3.5 mm (f) (2 required)b Frequency: 50 MHz VSWR: < 1.1:1 11901B Attenuator, 10 dB Step Range: 0-50 dB Frequency: 50 MHz Accuracy: ±0.25 dB 8495A Attenuator, 10 dB fixed (2 required) Frequency: 50 MHz VSWR: < 1.2:1 8493C, option 010 Cable, 3.5 mm, 1 meter (2 required) Frequency: 50 MHz VSWR: < 1.4:1 11500E

42 Chapter 8

Scale Fidelity Figure 8-1 Scale Fidelity Setup NOTE Averaging is used for all measurements to improve repeatability and reduce measurement uncertainty. Cable, BNC 120 cm Frequency: 10 MHz 10503A Synthesized Sweeper Output Level Accuracy: 0 to –15 dBm: ±1.0 dB at 50 MHz PSG a. for Options 503, 508, 513, 526 b. for Options 543, 544, 550 Item Critical Specifications (for this test) Recommended Keysight Model

Chapter 8 43

Scale Fidelity Procedure Procedure 1. Configure the equipment as shown in Figure 8-1 2. Preset the Source and press Mode, Spectrum Analyzer, Mode Preset on the analyzer. 3. Set up the synthesized sweeper by pressing: Frequency, 50, MHz Amplitude, +5, dBm RF On/Off, On 4. Set up the signal analyzer by pressing: Input/Output, More, Freq Ref In, External FREQ Channel, Center Freq, 50, MHz SPAN X Scale, Span, 1, MHz AMPTD Y Scale, Ref Level, 0, dBm Meas Setup, Average/Hold Number, 10, Enter Trace/Detector, Trace Average Peak Search 5. Set the external 10 dB step attenuator to 0 dB. 6. Adjust the amplitude on the signal source until the marker amplitude on the analyzer reads –15 dBm ±0.2 dB. 7. On the analyzer, press the Single, Restart to trigger a 10 sweep average. 8. On the analyzer, activate the Marker Delta function by pressing Peak Search, Marker Delta. 9. Perform the following steps for each attenuator setting listed in the table below: a. Select the next External attenuator setting. b. Press the Restart key to trigger a 10 sweep average. c. Enter the delta marker value into Table 8-1. d. Check delta marker reading against the test limits.

44 Chapter 8

Scale Fidelity Procedure Table 8-1 Scale Fidelity Results External Attenuator Setting Minimum (dB) Marker Delta Value (dB) Maximum (dB) 0 N/A Reference N/A 10 11.0 9.0 20 21.0 19.0 30 31.0 29.0 40 41.0 39.0 50 51.0 49.0

Keysight X-Series Signal Analyzers

PXA N9030A Functional Tests 9 BBIQ Input Frequency Response (Option BBA only) Test Limits This procedure tests the amplitude response of each of the BBIQ inputs versus frequency. For each input, a reference measurement is made at 250 kHz and then subsequent measurements are made at several frequencies up to the maximum IQ bandwidth. The amplitude difference between the measurement at 250 kHz and subsequent frequencies is calculated and compared to the test limit. Frequency Range Limit Relative to 250 kHz Frequency Range Limit Relative to 250 kHz 250 kHz to 10 MHz (Standard) ± 1.75 dB 250 kHz to 25 MHz (Option B25) ± 1.75 dB 250 kHz to 40 MHz (Option S40 or SU1) ± 1.75 dB Item Critical Specifications Recommended Keysight Model RF Signal Generator Frequency: 250 kHz to 40 MHz Amplitude: ~–10 dBm Flatness: < ±0.75 dB E4438C Cable, BNC Connectors: BNC male Length: 23 cm (9 in) max 10502A Cable, BNC, 120 cm Connectors: BNC male Frequency: 250 kHz to 40 MHz 10503A 50 Ohm Termination Connector: BNC male Frequency: 250 kHz to 40 MHz 11593A

46 Chapter 9

BBIQ Input Frequency Response (Option BBA only) Figure 9-1 BBIQ Input Frequency Response Setup (I Input) Figure 9-2 BBIQ Input Frequency Response Setup (I Input) Figure 9-3 BBIQ Input Frequency Response Setup (Q Input) Figure 9-4 BBIQ Input Frequency Response Setup (Q Input)

Chapter 9 47

BBIQ Input Frequency Response (Option BBA only) Procedure Procedure 1. On the X-Series analyzer, press Mode, IQ Analyzer (Basic). Press Mode Preset. 2. Press Input/Output, I/Q (to change to I/Q inputs), I/Q (to bring up the I/Q input menu), I/Q Cable Calibrate. Follow the instructions on screen. Use the short BNC cable listed above to perform the calibration. This process has several steps and will take about 3 minutes to perform. 3. Press I/Q Path, I+jQ, I Setup, Differential (On). Press Return, Q Setup, Q Same as I (Yes). 4. Preset the RF Signal Generator and set the controls as follows: Frequency 250 kHz Amplitude –10 dBm RF On NOTE Once the RF Signal Generator’s Amplitude is set, do not change the RF Signal Generator’s Amplitude setting during the rest of this procedure. 5. Connect the longer BNC cable between the RF Signal Generator’s RF OUTPUT and the I input of the X-Series analyzer, see Figure 9-1 6. Connect the 50 Ohm termination to the I input of the X-Series analyzer, see Figure 9-1 7. Press Mode Preset on the X-Series analyzer. Press System, Show, System. Note if either of the options listed below are present: N9030A-B25 N9030A-S40 N9030A-B40 8. If none of the options listed in step 7 are present, the maximum IQ bandwidth is 10 MHz 9. If N9030A-B25 is present, but N9030A-S40 and N9030A-B40 are not present, the maximum IQ bandwidth is 25 MHz 10.If either N9030A-S40 or N9030A-B40 is present, the maximum IQ bandwidth is 40 MHz. 11.Press FREQ on the X-Series analyzer and set the Center Freq equal to one half the maximum IQ bandwidth determined above. For example, if the maximum IQ bandwidth is 40 MHz, set the Center Freq to 20 MHz. 12.Press SPAN on the X-Series analyzer and set the Span equal to the maximum IQ bandwidth determined above. 13.Press BW on the X-Series analyzer and set the Res BW to 100 kHz.

48 Chapter 9

BBIQ Input Frequency Response (Option BBA only) Procedure 14.Press Peak Search on the X-Series analyzer. Record the marker amplitude reading in Table 9-1 for the I Input at 250 kHz. 15.Set the RF Signal Generator to each of the remaining frequencies listed in Table 9-1, up to the maximum IQ bandwidth of the X-Series analyzer. At each frequency, do the following: a. Press Peak Search b. Record the marker amplitude reading in Table 9-1 for the I Input at the current RF Signal Generator frequency 16.Set the RF Signal Generator frequency to 250 kHz 17.Connect the BNC Cable between the RF Signal Generator RF Output and the I input on the X-Series analyzer, see Figure 9-2 18.Connect the 50 Ohm termination to the I input on the X-Series analyzer, see Figure 9-2 Table 9-1 BBIQ Input Frequency Response, I and I Inputs RF Signal Generator Frequency I Input I Input Test Limits Marker Amplitude Frequency Response Marker Amplitude Frequency Response 250 kHz 0.0 dB 0.0 dB ±1.75 dB 2 MHz ±1.75 dB 5 MHz ±1.75 dB 7.5 MHz ±1.75 dB 9.9 MHz ±1.75 dB Stop here if Maximum IQ Bandwidth is 10 MHz 12.5 MHz ±1.75 dB 15 MHz ±1.75 dB 17.5 MHz ±1.75 dB 20 MHz ±1.75 dB 22.5 MHz ±1.75 dB 24.9 MHz ±1.75 dB Stop here if Maximum IQ Bandwidth is 25 MHz 30 MHz ±1.75 dB 35 MHz ±1.75 dB 39.9 MHz ±1.75 dB

Chapter 9 49

BBIQ Input Frequency Response (Option BBA only) Procedure 19.Press Peak Search on the X-Series analyzer. Record the marker amplitude reading in Table 9-1 for the I Input at 250 kHz. 20.Set the RF Signal Generator to each of the remaining frequencies listed in Table 9-1, up to the maximum IQ bandwidth of the X-Series analyzer. At each frequency, do the following: a. Press Peak Search b. Record the marker amplitude reading in Table 9-1 for the I Input at the current RF Signal Generator frequency 21.Connect the BNC Cable between the RF Signal Generator RF Output and the Q input on the X-Series analyzer, see Figure 9-3 22.Connect the 50 Ohm termination to the Q input on the X-Series analyzer, see Figure 9-3 23.Set the RF Signal Generator frequency to 250 kHz. 24.Press Peak Search on the X-Series analyzer. Record the marker amplitude reading in Table 9-2 for the Q Input at 250 kHz. 25.Set the RF Signal Generator to each of the remaining frequencies listed in Table 9-2, up to the maximum IQ bandwidth of the X-Series analyzer. At each frequency, do the following: Press Peak Search Record the marker amplitude reading in Table 9-2 for the Q Input at the current RF Signal Generator frequency Table 9-2 BBIQ Input Frequency Response, Q and Q Inputs RF Signal Generator Frequency Q Input Q Input Test Limits Marker Amplitude Frequency Response Marker Amplitude Frequency Response 250 kHz 0.0 dB 0.0 dB ±1.75 dB 2 MHz ±1.75 dB 5 MHz ±1.75 dB 7.5 MHz ±1.75 dB 9.9 MHz ±1.75 dB Stop here if Maximum IQ Bandwidth is 10 MHz 12.5 MHz ±1.75 dB 15 MHz ±1.75 dB 17.5 MHz ±1.75 dB 20 MHz ±1.75 dB

50 Chapter 9

BBIQ Input Frequency Response (Option BBA only) Procedure 26.Set the RF Signal Generator frequency to 250 kHz 27.Connect the BNC Cable between the RF Signal Generator RF Output and the Q input on the X-Series analyzer, see Figure 9-4 28.Connect the 50 Ohm termination to the Q input on the X-Series analyzer, see Figure 9-4 29.Press Peak Search on the X-Series analyzer. Record the marker amplitude reading in Table 9-2 for the Q Input at 250 kHz. 30.Set the RF Signal Generator to each of the remaining frequencies listed in Table 9-2, up to the maximum IQ bandwidth of the X-Series analyzer. At each frequency, do the following: a. Press Peak Search b. Record the marker amplitude reading in Table 9-2 for the Q Input at the current RF Signal Generator frequency 31.In Table 9-1 and Table 9-2, calculate the Frequency Response for each input by subtracting the Marker Amplitude at 250 kHz for that input from the Marker Amplitude for each frequency greater than 250 kHz. 32.Compare the Frequency Response values to the Test Limits. 22.5 MHz ±1.75 dB 24.9 MHz ±1.75 dB Stop here if Maximum IQ Bandwidth is 25 MHz 30 MHz ±1.75 dB 35 MHz ±1.75 dB 39.9 MHz ±1.75 dB Table 9-2 BBIQ Input Frequency Response, Q and Q Inputs RF Signal Generator Frequency Q Input Q Input Test Limits Marker Amplitude Frequency Response Marker Amplitude Frequency Response Frequency Response = Marker Amplitude (Freq > 250 kHz) – Marker Amplitude (250 kHz)

amplitude accuracy test 23

option 1DS 27 amplitude linearity test 41 B BBIQ frequency response 45 D DANL test 11 displayed average noise level. See DANL 11 E equipment functional tests 7 warm-up time 6 F frequency readout accuracy test 15 frequency response (flatness) test 29 BBIQ 45 frequency response (flatness) test preamp on 35 functional testing performance verification 4 functional tests before performing 6 equipment list 7 introduction 3 vs performance verification tests 4 warm-up time 6 See also individual functional tests 3 P performance verification tests vs functional tests 4 S second harmonic distortion test 19 T tests. See functional tests

This information is subject to change without notice.

© Keysight Technologies, 2015 Published in USA, January 2015 N9030-90018 www.keysight.com