Keysight Technologies X Series Manual
Text of the manual
Keysight X-Series Signal Analyzers
This manual provides documentation for the following
Analyzers:
PXA Signal Analyzer N9030A
© Keysight Technologies, Inc.
2006-2015
No part of this manual may be
reproduced in any form or by any
means (including electronic storage
and retrieval or translation into a
foreign language) without prior
agreement and written consent from
Keysight Technologies, Inc. as
governed by United States and
international copyright laws.
Trademark Acknowledgments
Manual Part Number
N9030-90018
Print Date
January 2015
Published in USA
Keysight Technologies
1400 Fountaingrove Parkway
Santa Rosa, CA 95403
Warranty
THE MATERIAL CONTAINED IN THIS
DOCUMENT IS PROVIDED “AS IS,”
AND IS SUBJECT TO BEING
CHANGED, WITHOUT NOTICE, IN
FUTURE EDITIONS. FURTHER, TO
THE MAXIMUM EXTENT PERMITTED
BY APPLICABLE LAW, KEYSIGHT
DISCLAIMS ALL WARRANTIES,
EITHER EXPRESS OR IMPLIED WITH
REGARD TO THIS MANUAL AND
ANY INFORMATION CONTAINED
HEREIN, INCLUDING BUT NOT
LIMITED TO THE IMPLIED
WARRANTIES OF
MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE.
KEYSIGHT SHALL NOT BE LIABLE
FOR ERRORS OR FOR INCIDENTAL
OR CONSEQUENTIAL DAMAGES IN
CONNECTION WITH THE
FURNISHING, USE, OR
PERFORMANCE OF THIS
DOCUMENT OR ANY INFORMATION
CONTAINED HEREIN. SHOULD
KEYSIGHT AND THE USER HAVE A
SEPARATE WRITTEN AGREEMENT
WITH WARRANTY TERMS
COVERING THE MATERIAL IN THIS
DOCUMENT THAT CONFLICT WITH
THESE TERMS, THE WARRANTY
TERMS IN THE SEPARATE
AGREEMENT WILL CONTROL.
Technology Licenses
The hardware and/or software
described in this document are
furnished under a license and may be
used or copied only in accordance
with the terms of such license.
Restricted Rights Legend
If software is for use in the
performance of a U.S. Government
prime contract or subcontract,
Software is delivered and licensed as
“Commercial computer software” as
defined in DFAR 252.227-7014 (June
1995), or as a “commercial item” as
defined in FAR 2.101(a) or as
“Restricted computer software” as
defined in FAR 52.227-19 (June
1987) or any equivalent agency
regulation or contract clause. Use,
duplication or disclosure of Software
is subject to Keysight Technologies’
standard commercial license terms,
and non-DOD Departments and
Agencies of the U.S. Government will
receive no greater than Restricted
Rights as defined in FAR
52.227-19(c)(1-2) (June 1987). U.S.
Government users will receive no
greater than Limited Rights as
defined in FAR 52.227-14 (June
1987) or DFAR 252.227-7015 (b)(2)
(November 1995), as applicable in
any technical data.
Safety Notices
CAUTION
A CAUTION notice denotes a hazard. It
calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in damage to the
product or loss of important data. Do
not proceed beyond a CAUTION
notice until the indicated conditions
are fully understood and met.
WARNING
A WARNING notice denotes a hazard.
It calls attention to an operating
procedure, practice, or the like that,
if not correctly performed or adhered
to, could result in personal injury or
death. Do not proceed beyond a
WARNING notice until the indicated
conditions are fully understood and
met.
Where to Find the Latest Information
Documentation is updated periodically. For the latest information about these products, including instrument software
upgrades, application information, and product information, browse to one of the following URLs, according to the name
of your product:
http://www.keysight.com/find/pxa
To receive the latest updates by email, subscribe to Keysight Email Updates at the following URL:
http://www.keysight.com/find/MyKeysight
Information on preventing analyzer damage can be found at:
www.keysight.com/find/PreventingInstrumentRepair
Is your product software up-to-date?
Periodically, Keysight releases software updates to fix known defects and incorporate product enhancements. To search
for software updates for your product, go to the Keysight Technical Support website at:
http://www.keysight.com/find/techsupport
Table of Contents
1.
Functional Tests
Functional Test Versus Performance Verification . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Contents of this Document. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Before Performing a Functional Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Test Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
2.
Displayed Average Noise Level (DANL)
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
3.
Frequency Readout Accuracy
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
4.
Second Harmonic Distortion (SHD)
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
5.
Amplitude Accuracy at 50 MHz
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
Testing Preamp Option (P03, P08, P13, P26, P43, P44, P50) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
6.
Frequency Response (Flatness)
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31
7.
Frequency Response (Flatness), Preamp On
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 37
8.
Scale Fidelity
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
9.
BBIQ Input Frequency Response
(Option BBA only)
Procedure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 47
Keysight X-Series Signal Analyzers
PXA N9030A
Functional Tests
1
Functional Tests
Functional tests are tests of various instrument parameters that give a high
degree of confidence that the analyzer is operating correctly. They are
recommended as a check of analyzer operation for incoming inspection or after
a repair. Measurement uncertainty analysis is not available for functional tests,
and the analyzer is checked against limits that are wider than the published
specifications. The functional tests are designed to test an analyzer operating
within the temperature range defined by the analyzer specifications using a
minimum set of test equipment. If a test does not pass, performance
verification tests must be run to determine whether a problem exists.
4 Chapter 1
Functional Tests
Functional Test Versus Performance Verification
Functional Test Versus Performance Verification
Functional tests use a minimum set of test equipment to check a much smaller
range of parameters (and a limited number of data points for each parameter)
than do performance verification tests. Functional tests use limits that are
wider than the published specifications; measurement uncertainty analysis is
not available for functional tests.
NOTE
If a functional test does not pass, you must run performance verification tests
to determine whether a problem exists.
Performance verification tests span a wide range of instrument parameters and
provide the highest level of confidence that the instrument conforms to
published specifications. These tests can be time consuming and require
extensive test equipment.
Chapter 1 5
Functional Tests
Contents of this Document
Contents of this Document
This chapter includes the following:
— “Before Performing a Functional Test” on page 6 (what to do first).
— “Test Equipment” on page 7 (a list of the equipment required for all of the
tests).
Subsequent chapters describe the following Functional Tests:
— “Displayed Average Noise Level (DANL)” on page 11
— “Frequency Readout Accuracy” on page 15
— “Second Harmonic Distortion (SHD)” on page 19
— “Amplitude Accuracy at 50 MHz” on page 23
— “Frequency Response (Flatness)” on page 29
— “Frequency Response (Flatness), Preamp On” on page 35
— “Scale Fidelity” on page 41
Each functional test includes:
— Test limits (pass/fail criteria)
— A description of the test
— The equipment required for the test
— A figure showing how to connect the equipment
— Step-by-step instructions
— One or more tables in which to record the measurement results
6 Chapter 1
Functional Tests
Before Performing a Functional Test
Before Performing a Functional Test
1. Ensure that you have the proper test equipment.
2. Switch on the unit under test (UUT) and let it warm up (in accordance with
warm-up requirements in the instrument specifications).
3. Allow sufficient warm-up time for the required test equipment (refer to
individual instrument documentation for warm-up specifications).
4. Ensure that the analyzer’s frequency reference is set to Internal:
a. Press the Input/Output, More, Freq Ref In keys.
b. If the Freq Ref In softkey does not show Internal, press the Freq Ref
In softkey and select Internal.
5. Following instrument warm-up, perform the auto align routine:
Press System, Alignments, Align Now, All.
NOTE
Functional test accuracy depends on the precision of the test equipment
used. Ensure that all of the test equipment is calibrated before running a
functional test.
Chapter 1 7
Functional Tests
Test Equipment
Test Equipment
The table below summarizes the test equipment needed to perform all of the
functional tests. Alternate equipment model numbers are given in case the
recommended equipment is not available. If neither the recommended nor the
alternative test equipment are available, substitute equipment that meets or
exceeds the critical specifications listed.
Table 1-1
Analyzer
Option
Item
Critical Specifications
Recommended
Keysight Model
Alternate
Keysight
Model
Adapters
503, 508,
513, 526
3.5 mm (f) to 3.5mm (f)
(connector saver for source)
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
83059B
1250-1749
543, 544,
550
2.4 mm (f) to 2.4mm (f)
(connector saver for source)
Frequency: 10 MHz to 50 GHz
VSWR: < 1.1:1
11900B
543, 544,
550
2.4 mm (f) to 3.5mm (f)
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
11901B
All
BNC (f) to SMA (m)
Frequency: 40 MHz
1250-1200
All
Type N (f) to Type N (f)
Frequency: 10 MHz to 18 GHz
VSWR: < 1.05:1
1250-1472
1250-0777
All
Type N (m) to 3.5 mm (m)
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
1250-1743
All
Type N (m) to 3.5 mm (f)
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
1250-1744
All
Type N (f) to 3.5 mm (f)
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
1250-1745
All
Type N (m) to BNC (f)
Frequency: 10 MHz to 1 GHz
VSWR: < 1.05:1
1250-1476
Attenuators
All
10 dB Step Attenuator
Frequency: 50 MHz
Range: 0 to 70 dB
8495A
8496A
All
10 dB Fixed Attenuator
(2 required)
Frequency: 50 MHz
VSWR: < 1.2:1
8493C
Option 010
8493A
Option 010
or
8493B
Option 010
8 Chapter 1
Functional Tests
Test Equipment
Preamp
Options
(Pxx)
30 dB Fixed Attenuator
Accuracy: < 0.05 dB @ 50 MHz
11708A
Cables
All
3.5 mm (1 meter)
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.4:1
Loss: < 2.0 dB
11500E
8120-4921
543, 544,
550
2.4 mm (f) to 2.4 mm (m)
(1 meter)
Frequency: 10 MHz to 50 GHz
VSWR: < 1.55:1 at 50 GHz
Loss: < 6.0 dB
8120-6164
BBA
Cable, BNC
BNC male
Length: 23 cm (9 in) max
10502A
All
Cable, BNC
(3 required)
120 cm (48 in.) BNC cable
10503A
Signal Source
All
Synthesized Sweeper
Frequency: 10 MHz to 50 GHza
Harmonic level: < − 40 dBc
Amplitude range: 10 to − 20 dBm
Frequency Accuracy: 0.02%
PSGb
83630B,
83640B,
83650B
BBA
RF Signal Generator
Frequency: 250 kHz to 40 MHz
Amplitude: ~–10 dBm
Flatness: < ±0.75 dB
E4438C
33250A
Power Meter
All
Power Meter
Power Reference Accuracy: ±1.2%
Compatible with power sensor
E4418B
E4419B
All
Power Sensor
Frequency Range: 50 MHz to 3.66 GHz
Amplitude Range: −70 to –20 dBm
8481D
8487D
8485D
503, 508,
513
Power Sensor
Frequency Range: 50 MHz to
13.66 GHz
Amplitude Range: −30 to +20 dBm
8481A
526
Power Sensor
Frequency Range: 50 MHz to 26.5 GHz
Amplitude Range: −30 to +20 dBm
8485A
543, 544,
550
Power Sensor
Frequency Range: 50 MHz to 50 GHz
Amplitude Range: −30 to +20 dBm
8487A
P03
Low Power Sensor
Frequency Range: 50 MHz to 3.6 GHz
Amplitude Range: –70 to –20 dBm
8481D
8485D
8487D
Table 1-1
Analyzer
Option
Item
Critical Specifications
Recommended
Keysight Model
Alternate
Keysight
Model
Chapter 1 9
Functional Tests
Test Equipment
Preamp
other than
P03
Low Power Sensor
Frequency Range: 50 MHz to 50 GHzc
Amplitude Range: –70 to –20 dBm
8487D
8485D
Miscellaneous Equipment
All
Filter, 50 MHz Low Pass
Cutoff Frequency: 50 MHz
Rejection at 65 MHz: > 40 dB
Rejection at 75 MHz: > 60 dB
Rejection at 80 MHz: > 60 dB
Frequency: 10 MHz to 26.5 GHz
0955-0306
526
Power Splitter, 3.5 mm
Nominal Insertion Loss: 6 dB
Tracking Between Ports: < 0.25 dB
11667B
543, 544,
550
Power Splitter, 2.4 mm
Nominal Insertion Loss: 8.5 dB
Tracking Between Ports: < 0.4 dB
11667C
503, 508,
513
Power Splitter
Nominal Insertion Loss: 6 dB
Tracking Between Ports: < 0.25 dB
11667A
11667B
503, 508,
513, 526
Termination, 50Ω
Type N (m) Connector
Frequency: 30 Hz to 26.5 GHz
909A Option 012
544, 545,
550
Termination, 50Ω
2.4 mm (f) Connector
Frequency: 30 Hz to 50 GHz
85138B
BBA
Termination, 50Ω
Connector: BNC male
Frequency: 250 kHz to 40 MHz
11593A
a. Maximum frequency equal to or greater than the maximum frequency of the PXA.
b. PSG model numbers: E8244A, E8254A, E8247C Option 520, E8247C Option H31 or 540, E8257D
Option 520, E8257D Option 550, E8257D Option 567, E8267D Option 544.
c. Maximum frequency equal to or greater than the maximum frequency of the preamplifier option.
Table 1-1
Analyzer
Option
Item
Critical Specifications
Recommended
Keysight Model
Alternate
Keysight
Model
10 Chapter 1
Functional Tests
Test Equipment
Keysight X-Series Signal Analyzers
PXA N9030A
Functional Tests
2
Displayed Average Noise Level (DANL)
Test Limits (with 0 dB input attenuation)
See Table 2-1 on page 13 for values.
The Displayed Average Noise Level (DANL) of the signal analyzer is measured
across a 10 kHz frequency span at several center frequencies. The analyzer
input is terminated into a 50Ω load. A test is performed to assure the
measurement is not performed in the presence of a residual response. The
measurement is then averaged, and the result is normalized to a 1 Hz
bandwidth.
Figure 2-1
DANL Test Setup
Item
Critical Specifications
(for this test)
Recommended
Keysight Model
Termination, 50Ω Type-N(m)a
Frequency: 10 MHz to 18 GHz
909A Option 012
Termination, 50Ω 2.4 mm (f)b
Frequency: 10 MHz to 50 GHz
85138B
a. for Options 503, 508, 513, and 526
b. for Options 543, 544, and 550
12 Chapter 2
Displayed Average Noise Level (DANL)
Procedure
Procedure
1. Configure the equipment as shown in Figure 2-1
2. Press Mode, Spectrum Analyzer, Mode Preset on the analyzer.
3. Set up the signal analyzer by pressing:
FREQ Channel, Center Freq, 10, MHz
Input/Output, RF Input, RF Coupling, select DC (except Options
543, 544, and 550)
SPAN X Scale, Span, 10, kHz
AMPTD Y Scale, –70, dBm
AMPTD Y Scale, Attenuation, MechAtten, 0, dB
BW, Res BW, 1, kHz
BW, Video BW, 100, Hz
Meas Setup, Average/Hold, Number, 20, Enter
Trace/Detector, Trace Average
Single
4. Press Restart, then wait for Average/Hold to display 20/20.
5. Press View/Display, Display, then press Display Line, On.
6. Rotate the knob and set the display line at the average amplitude of the
displayed noise floor by visual inspection.
7. Confirm that the measurement is performed on the analyzer noise floor
and not on a residual response within the displayed 10 kHz span.
NOTE
Ignore the residual response if one appears when taking the measurement.
8. Enter the value of the display line as the Measured Average Noise Level at
10 MHz column in Table 2-1.
9. Normalize the measured value to a 1 Hz BW by adding 30 dB to the
measured value.
NOTE
The 30 dB value is added because the formula used to calculate the value of
the noise power in a 1 Hz BW when measured with a 1 kHz BW is:
Normalized Noise = 10 Log (BW 2/BW 1) where BW 2 is the 1 kHz BW we
measure and BW 1 is 1 Hz BW to which we want to normalize.
Therefore, 10 Log (1000) = 30 dB, so the noise floor will be 30 dB lower in a
1 Hz BW.
10.Enter the normalized value of the displayed average noise level in Table
2-1.
Chapter 2 13
Displayed Average Noise Level (DANL)
Procedure
11.The value of the normalized displayed average noise should be less than
the specification value.
12.Change the analyzer center frequency to the next value listed in Table 2-1.
Press: FREQ Channel, Center Freq, [Table 2-1 Value], GHz
13.Repeat step 4 through step 12 to fill in the remainder of Table 2-1 for your
analyzer frequency range.
Table 2-1
Displayed Average Noise Level (DANL) Results
Center Frequency
Measured Average
Noise Level
(dBm)
Normalized Average
Noise Level/(1 Hz BW)
(dBm)
Test Limits
(dBm)
Options 503,
508, 513, 526
Test Limits
(dBm)
Options 543,
544, 550
10 MHz
–153
–153
2 GHz
–152
–152
6 GHz
–146
–142
13 GHz
–147
–145
20 GHz
–140
–139
26.5 GHz
–135
–137
30 GHz
N/A
–136
35 GHz
N/A
–131
40 GHz
N/A
–129
43.5 GHz
N/A
–126
50 GHz
N/A
–126
14 Chapter 2
Displayed Average Noise Level (DANL)
Procedure
Keysight X-Series Signal Analyzers
PXA N9030A
Functional Tests
3
Frequency Readout Accuracy
Test Limits
Frequency Readout Accuracy is equivalent to the following equation:
See Table 3-1 on page 17 for actual values.
The frequency readout accuracy is measured in several spans and center
frequencies that allow both internal analyzer synthesizer modes and prefilter
bandwidths to be tested. Frequency reference error is eliminated by using the
same frequency standard for the analyzer and signal source.
0.25%
span
5%
RBW
2 Hz
0.5
horizontal resolution
×
+
+
×
+
×
(
)
±
Item
Critical Specification
(for this test)
Recommended
Keysight Model
Adapter, Type-N (m), to
3.5 mm (f)a
Frequency: 10 MHz to 1.51 GHz
VSWR: < 1.1:1
1250-1744
Adapter, 3.5 mm (f) to
3.5 mm (f)a
Frequency: 10 MHz to 1.51 GHz
VSWR: < 1.1:1
83059B
Adapter, 2.4 mm (f) to
2.4 mm (f)b
Frequency: 10 MHz to 1.51 GHz
VSWR: < 1.1:1
11900B
Cable, 2.4 mm,
1 meterb
Frequency: 10 MHz to 1.51 GHz
VSWR: < 1.4:1
8120-6164
Cable, 3.5 mm,
1 metera
Frequency: 10 MHz to 1.51 GHz
VSWR: < 1.4:1
11500E
Cable, BNC, 120 cm
Frequency: 10 MHz
10503A
Synthesized Sweeper
Frequency: 10 MHz to 1.51 GHz
PSG
a. for Options 503, 508, 513, 526
b. for Options 543, 544, 550
16 Chapter 3
Frequency Readout Accuracy
Figure 3-1
Frequency Readout Accuracy Test Setup
Chapter 3 17
Frequency Readout Accuracy
Procedure
Procedure
1. Configure the equipment as shown in Figure 3-1. Confirm the analyzer’s
built-in auto alignment has been performed within the past 24 hours.
2. On the synthesized sweeper, press PRESET, then set the controls as
follows:
FREQUENCY, 1505, MHz
POWER LEVEL, –10, dBm
RF, On
3. Set up the signal analyzer by pressing:
Mode, Spectrum Analyzer
Mode Preset
Input/Output, More, Freq Ref In, External
FREQ Channel, Center Freq, 1505, MHz
SPAN X Scale, Span, 2990, MHz
Trace/Detector, More, Detector, Sample
Single
4. Press Restart.
Press Peak Search on the analyzer. If the instrument is functioning
correctly, the marker reading in the active function block will be between
the values listed in Table 3-1. Record the marker value in the Marker
Frequency Readout column in Table 3-1.
5. On the signal analyzer, change the span and center frequency as listed in
Table 3-1.
6. Change the synthesized sweeper frequency to match the center frequency
of the analyzer.
7. Repeat step 4 through step 6 until the Marker Frequency Readout column
of Table 3-1 is complete.
Table 3-1
Frequency Readout Accuracy Results
Span
(MHz)
Center
Frequency
(MHz)
Minimum
Marker Frequency Readout
Maximum
2990
1505
1495.9 MHz
1514.1 MHz
127.2
1505
1504.56 MHz
1505.44 MHz
54.1
1505
1504.8122 MHz
1505.1878 MHz
7.95
1505
1504.97240 MHz
1505.0276 MHz
0.106
1505
1504.999630 MHz
1505.000370 MHz
1.98
517.59
517.58316 MHz
517.59684 MHz
1.98
832.50
832.49316 MHz
832.50684 MHz
18 Chapter 3
Frequency Readout Accuracy
Procedure
Keysight X-Series Signal Analyzers
PXA N9030A
Functional Tests
4
Second Harmonic Distortion (SHD)
Test Limits
This test checks the second harmonic distortion of the signal analyzer by
tuning to twice the input frequency and examining the level of the distortion
product. A low pass filter is inserted between the source and the signal
analyzer to prevent the source second harmonic from artificially raising the
second harmonic product displayed on the analyzer.
Applied Frequency
Mixer Levela
Distortion
40 MHz
–15 dBm
< –60 dBc
a. Mixer Level = Input Level - RF Attenuation
Item
Critical Specifications
(for this test)
Recommended
Keysight Model
Adapter, Type-N (m) to
BNC (f)
Frequency: 10 MHz to 80 MHz
VSWR: < 1.05:1
1250-1476
Adapter, BNC (f) to SMA
(m)
Frequency: 40 to 80 MHz
1250-1200
Adapter, 3.5 mm (f) to 3.5
mm (f)a
Frequency: 10 MHz to 80 MHz
VSWR: < 1.1:1
83059B
Adapter, 2.4 mm (f) to 3.5
mm (f)b
Frequency: 10 MHz to 80 MHz
VSWR: < 1.1:1
11901B
Cable, BNC 120 cm
(2 required)
Frequency: 10 to 40 MHz
10503A
Filter, 50 MHz Low Pass
Cutoff Frequency: 50 MHz
Rejection at 65 MHz: > 40 dB
Rejection at 75 MHz: > 60 dB
0955-0306
Synthesized Sweeper
Frequency: 50 MHz
Spectral Purity: Better than –30 dBc
PSG
a. for Options 503, 508, 513, 526
20 Chapter 4
Second Harmonic Distortion (SHD)
Figure 4-1
Second Harmonic Distortion Test Setup
b. for Options 543, 544, 550
Chapter 4 21
Second Harmonic Distortion (SHD)
Procedure
Procedure
1. Configure the equipment as shown in Figure 4-1
2. Press Mode, Spectrum Analyzer, Mode Preset on the signal analyzer and
Preset the synthesized sweeper.
3. Set up the synthesized sweeper by pressing:
Frequency, 40, MHz
Amplitude, –5, dBm
4. Set up the signal analyzer by pressing:
Input/Output, More, Freq Ref In, External
FREQ Channel, Center Freq, 40, MHz
SPAN X Scale, Span, 1, MHz
AMPTD Y Scale, Attenuation, Mech Atten, 10, dB
5. On the analyzer, press Peak Search.
6. Adjust the synthesized sweeper amplitude for a signal analyzer display of
–5 dBm ±0.1 dB.
7. On the signal analyzer, activate the marker delta function by pressing the
Marker and Delta keys.
8. On the signal analyzer, press:
FREQ Channel, Center Freq, 80, MHz
Meas Setup, Average/Hold Number, 20, Enter
Trace/Detector, Trace Average
Single
9. Press Peak Search. Enter the displayed value under the Measured Second
Harmonic Distortion (dBc) heading in Table 4-1.
Table 4-1
Second Harmonic Distortion Results
Applied Frequency
(MHz)
Measured Second Harmonic Distortion
(dBc)
Mixer Level
(dBm)
Specification
(dBc)
40
–15
–60
22 Chapter 4
Second Harmonic Distortion (SHD)
Procedure
Keysight X-Series Signal Analyzers
PXA N9030A
Functional Tests
5
Amplitude Accuracy at 50 MHz
Test Limits
Amplitude Accuracy should remain within 1.13 dB of the measured source
value across the range of source levels and changes in resolution bandwidth.
The Preamp (option P03, P08, P13, P26, P43, P44, P50) should remain within
±1.3 dB of measured values.
A synthesized sweeper is used as the signal source for the test. The source
amplitude is varied using the signal source amplitude control. The attenuation
and resolution bandwidth are varied on the signal analyzer. The source
amplitude is measured by the power meter and signal analyzer at each setting,
and the values compared. The difference between each pair of measurements
indicates the amplitude accuracy.
Item
Critical Specifications
Recommended
Keysight Model
Adaptera
Type-N (m), to 3.5 mm (m)
1250-1743
Adaptera
3.5 mm (f) to 3.5 mm (f)
83059B
Adapterb
2.4 mm (f) to 3.5 mm (m)
11901A
Adapterb
2.4 mm (f) to 3.5 mm (f)
11901B
Attenuator, 30 dB
Accuracy: < 0.5 dB at 50 MHz
11708A
Cable
3.5 mm, 1 meter
11500E
Cable
BNC, 120 cm
10503A
Power Meter
Compatible with power sensor
E4418B
Power Sensor
Amplitude Range: −25 dBm to 10 dBm
8485A
Power Splitter
3.5 mm (f) connectors
6 dB loss
11667B
Synthesized Sweeper
Typical Temperature Stability: 0.01 dBc/°C
PSG
a. for Options 503, 508, 513, 526
b. for Options 543, 544, 550
24 Chapter 5
Amplitude Accuracy at 50 MHz
Figure 5-1
Amplitude Accuracy Test Setup
Chapter 5 25
Amplitude Accuracy at 50 MHz
Procedure
Procedure
1. Zero and calibrate the power meter.
2. Configure equipment as shown in Figure 5-1, with the power splitter
connected directly to the signal analyzer input through the adapter.
CAUTION
To minimize stress on the test equipment connections, support the power
sensor.
3. If the auto alignment for the analyzer has not been performed within the
past 24 hours, press System, Alignments, Align Now, All to perform the
auto alignment routine.
4. Press Mode, Spectrum Analyzer, Mode Preset on the analyzer.
5. Set up the synthesized sweeper by pressing:
CW, 50, MHz
Power Level, -4, dBm
RF (On)
6. Set up the signal analyzer by pressing:
Input/Output, More, Freq Ref In, External
FREQ Channel, Center Freq, 50, MHz
SPAN X Scale, 2, MHz
AMPTD Y Scale, Attenuation, Mech Atten, 10, dB
Input/Output, RF Input, RF Coupling, select DC (except Options
543, 544, 550)
Sweep/Control, Sweep Setup, Swp Time Rules, SA - Accuracy
Meas Setup, Average/Hold Number, 20, Enter
Trace/Detector, Trace Average
Single
7. Perform the following steps for each row listed in Table 5-1:
a. Set the synthesized sweeper amplitude to the value listed in the
Nominal Source Amplitude column in Table 5-1
b. Set the Mech Atten as indicated in the Attenuation column in Table
5-1
c. Set the Span as listed in the Span column of Table 5-1
d. Record the synthesized sweeper amplitude, as measured by the
power meter, in the Power Meter Amplitude column of Table 5-1
e. On the signal analyzer, press Restart.
f. Wait for the signal analyzer to finish averaging.
g. Press Peak Search.
26 Chapter 5
Amplitude Accuracy at 50 MHz
Procedure
h. Record the signal amplitude, as measured by the analyzer in the
Measured Amplitude column of Table 5-1
i.
Calculate the signal amplitude accuracy error using the following
equation, and record the results under the Amplitude Accuracy
Error column:
Amplitude Accuracy Error = Meas_amp − Power_meter
Table 5-1
Amplitude Accuracy Results
Nominal
Source
Amplitude
(dBm)
Attenuation
(dB)
Span
(MHz)
Measured
Amplitude
Meas_amp
(dBm)
Power Meter
Amplitude
Power_meter
(dBm)
Amplitude
Accuracy
Error
(dB)
Test Limit
(dB)
−4
10
2
±1.13 dB
−9
10
1
±1.13 dB
−14
10
0.5
±1.13 dB
−4
20
0.1
±1.13 dB
−14
20
0.1
±1.13 dB
−4
30
0.1
±1.13 dB
−14
30
0.1
±1.13 dB
Chapter 5 27
Amplitude Accuracy at 50 MHz
Testing Preamp Option (P03, P08, P13, P26, P43, P44, P50)
Testing Preamp Option (P03, P08, P13, P26, P43, P44, P50)
Instruments containing Options P03, P08, P13, P26, P43, P44, P50 must have
the preamp function turned on and tested.
Procedure
1. On the analyzer, press AMPTD Y Scale, More, Internal Preamp, Low Band.
2. Connect the equipment as shown in Figure 5-1 on page 24, using a 30 dB
Pad between the adaptor and the signal analyzer input.
3. Set the synthesized sweeper amplitude to the value listed in the Nominal
Source Amplitude column in Table 5-2
4. Set the signal analyzer input attenuation to 0 dB.
5. Set the Span as listed in Table 5-2
6. Record the synthesized sweeper amplitude, as measured by the power
meter, in the Power Meter Amplitude column of Table 5-2
7. On the signal analyzer, press Restart.
8. Wait for the analyzer to finish averaging.
9. Press Peak Search.
10.Record the signal amplitude as measured by the analyzer in the measured
amplitude column of Table 5-2
11.Calculate the signal amplitude accuracy using the following equation:
Amplitude Accuracy Error = Meas_amp + 30 dB − Power_meter
12.Record the results under the Amplitude Accuracy Error column of Table
5-2
Table 5-2
Amplitude Accuracy Results (Option P03, P08, P13, P26, P43, P44, P50)
Nominal
Source
Amplitude
(dBm)
Low- band
Preamp
Res BW
(kHz)
Span
(kHz)
Measured
Amplitude
Meas_amp
(dBm)
Power Meter
Amplitude
Power_meter
(dBm)
Amplitude
Accuracy
Error
(dB)
Test Limit
(dB)
−13
ON
1
106
±1.30 dB
28 Chapter 5
Amplitude Accuracy at 50 MHz
Testing Preamp Option (P03, P08, P13, P26, P43, P44, P50)
Keysight X-Series Signal Analyzers
PXA N9030A
Functional Tests
6
Frequency Response (Flatness)
Test Limits
The frequency response test measures the signal analyzer’s amplitude error as
a function of the tuned frequency. Measurements are made ranging from 50
MHz to the maximum frequency range of your analyzer. The signal source
amplitude is measured with a power meter to eliminate error due to source
flatness. The measured value is normalized to 50 MHz.
Frequency Range
Limit Relative to 50 MHz
20 Hz to 3.6 GHz
±1.5 dΒ
> 3.6 GHz to 8.4 GHz
±2.5 dΒ
> 8.4 GHz to 13.6 GHz
±3.0 dΒ
> 13.6 GHz to 26.5 GHz
±3.5 dΒ
> 26.5 GHz to 34.5 GHz
±3.5 dΒ
> 34.5 GHz to 50 GHz
±4.2 dΒ
Item
Critical Specifications
(for this test)
Recommended
Keysight Model
Adapter, Type N (m) to 3.5 mm
(m)a
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
1250-1743
Adapter, 3.5 mm (f) to 3.5 mm
(f)a
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
83059B
Cable, 3.5 mm, 1 metera
(Opt 503, 508, 513, 526)
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.4:1
11500E
Cable, 2.4 mm, 1 meterb
(Opt 543, 544, 550)
Frequency: 10 MHz to 50 GHz
VSWR: < 1.55:1
8120-6164
30 Chapter 6
Frequency Response (Flatness)
Figure 6-1
Frequency Response Test Setup
Cables, BNC 120 cm
Frequency: 10 MHz
10503A
Power Meter
Compatible with power sensor
E4418B
Power Sensor
Frequency Range: 50 MHz to 26.5 GHz
8485A
Power Sensorb
Frequency Range: 50 MHz to 50 GHz
8487A
Power Splittera
Frequency Range: 50 MHz to 26.5 GHz
Tracking between ports: < 0.25 dB
11667B
Power Splitterb
Frequency Range: 50 MHz to 50 GHz
Tracking between ports: < 0.4 dB
11667C
Synthesized Sweeper
Frequency Range: 50 MHz to maximum
frequency of PXA
PSG
a. for Options 503, 508, 513, 526
b. for Options 543, 544, 550
Item
Critical Specifications
(for this test)
Recommended
Keysight Model
Chapter 6 31
Frequency Response (Flatness)
Procedure
Procedure
1. Zero and calibrate the power meter and power sensor as described in the
power meter operation manual.
2. Configure the equipment as shown in Figure 6-1.
NOTE
Connect the power splitter to the signal analyzer input using the appropriate
adapter. Do not use a cable.
3. Assure the signal analyzer’s built-in auto alignment has been performed
within the last 24 hours.
4. Press Mode, Spectrum Analyzer, Mode Preset on the signal analyzer, and
press Preset on the synthesized sweeper.
5. Set up the synthesized sweeper by pressing:
CW, 50, MHz
Power level, –4, dBm
6. Set up the signal analyzer by pressing:
Input/Output, More, Freq Ref In, External
FREQ Channel, Center Freq, 50, MHz
SPAN X Scale, Span, 50, kHz
AMPTD Y Scale, Ref Level, 0, dBm
7. Adjust the synthesized sweeper output power for a power meter reading
of –10 dBm ±0.1 dB.
8. On the signal analyzer, press Single.
9. Press the Peak Search key on the signal analyzer to position the marker on
the peak of the signal.
10.Refer to Table 6-1, “Frequency Response (Flatness) Results.” Enter the
amplitude of the signal displayed on the signal analyzer into the MeasAmp
column of Table 6-1.
11.Enter the power meter reading into the PowerMeter column of Table 6-1.
12.Tune the synthesized sweeper and signal analyzer to the next frequency
listed in Table 6-1.
13.Enter the power sensor calibration factor into the power meter.
14.For frequencies 3.6 GHz and above, press AMPTD Y Scale, then Presel
Center to center the preselector filter for an optimum amplitude
measurement.
15.Repeat step 7 through step 14 and complete the remainder of Table 6-1
for the frequency range of your analyzer.
32 Chapter 6
Frequency Response (Flatness)
Procedure
16.Compute the measurement error (MeasError = MeasAmp − PowerMeter).
17.Compute the flatness error normalized to 50 MHz:
(MeasError − MeasError @ 50 MHz)
18.Enter the computed flatness error value into the FlatNorm column of Table
6-1.
19.Compare the value of FlatNorm to the test limit.
Table 6-1
Frequency Response (Flatness) Results
Center
Frequency
Analyzer
Amplitude
Measamp
Power Meter
Measurement
Powermeter
Meas Error
Measerror
Flatness
Normalized
to 50 MHz
FlatNorm
Flatness Error
Test Limits
(dB)
50 MHz
0
Ref
1 GHz
±1.5 dB
2 GHz
1.5 dB
3 GHz
±1.5 dB
3.5 GHz
±1.5 dB
4 GHz
±2.5 dB
6 GHz
±2.5 dB
8 GHz
±2.5 dB
9 GHz
±3.0 dB
11 GHz
±3.0 dB
13 GHz
±3.0 dB
14 GHz
±3.5 dB
17 GHz
±3.5 dB
20 GHz
±3.5 dB
23 GHz
±3.5 dB
26 GHz
±3.5 dB
29 GHz
±3.5 dB
32 GHz
±3.5 dB
35 GHz
±4.2 dB
38 GHz
±4.2 dB
41 GHz
±4.2 dB
42.5 GHz
±4.2 dB
Chapter 6 33
Frequency Response (Flatness)
Procedure
43.5 GHz
±4.2 dB
46.5 GHz
±4.2 dB
49.5 GHz
±4.2 dB
Table 6-1
Frequency Response (Flatness) Results
Center
Frequency
Analyzer
Amplitude
Measamp
Power Meter
Measurement
Powermeter
Meas Error
Measerror
Flatness
Normalized
to 50 MHz
FlatNorm
Flatness Error
Test Limits
(dB)
34 Chapter 6
Frequency Response (Flatness)
Procedure
Keysight X-Series Signal Analyzers
PXA N9030A
Functional Tests
7
Frequency Response (Flatness), Preamp On
Test Limits
The frequency response test, with preamplifier on, measures the signal
analyzer’s amplitude error as a function of the tuned frequency. Measurements
are made ranging from 50 MHz to the maximum frequency range of the
preamp. The signal source amplitude is measured with a power meter to
eliminate error due to source flatness. The measured value is normalized to 50
MHz.
Frequency Range
Limit Relative to 50 MHz
100 kHz to 3.6 GHz
±2.0 dΒ
> 3.6 GHz to 8.4 GHz
±3.0 dΒ
> 8.4 GHz to 13.6 GHz
±3.5 dΒ
> 13.6 GHz to 26.5 GHz
±4.0 dΒ
> 26.5 GHz to 34.5 GHz
±4.5 dΒ
> 34.5 GHz to 50 GHz
±6.0 dΒ
Item
Critical Specifications
(for this test)
Recommended
Keysight Model
Adapter, Type N (m) to 3.5 mm
(m)a
Frequency: 10 MHz to 18 GHz
VSWR: < 1.1:1
1250-1743
Adapter, 3.5 mm (f) to 3.5 mm
(f)a
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.1:1
83059B
Cable, 3.5 mm, 1 metera
Frequency: 10 MHz to 26.5 GHz
VSWR: < 1.4:1
11500E
36 Chapter 7
Frequency Response (Flatness), Preamp On
Figure 7-1
Frequency Response Test Setup
Cable, 2.4 mm, 1 meterb
Frequency: 10 MHz to 50 GHz
VSWR: < 1.55:1
8120-6164
Cables, BNC 120 cm
Frequency: 10 MHz
10503A
Power Meter
Compatible with power sensor
E4418B
Attenuator, 30 dB Fixed
Frequency: 50 MHz
Accuracy: ±0.05 dB
11708A
Power Sensora
Frequency Range: 50 MHz to 26.5 GHz
Amplitude Range: –65 dBm to –55 dBm
8485D
Power Sensorb
Frequency Range: 50 MHz to 50 GHz
Amplitude Range: –65 dBm to –55 dBm
8487D
Power Splittera
Frequency Range: 50 MHz to 26.5 GHz
Tracking between ports: < 0.25 dB
11667B
Power Splitterb
Frequency Range: 50 MHz to 50 GHz
Tracking between ports: < 0.4 dB
11667C
Synthesized Sweeper
Frequency Range: 50 MHz to maximum frequency
of the PXA
PSG
a. for Options P03, P08, P13, P26
b. for Options P43, P44, P50
Item
Critical Specifications
(for this test)
Recommended
Keysight Model
Chapter 7 37
Frequency Response (Flatness), Preamp On
Procedure
Procedure
1. Zero and calibrate the power meter and power sensor as described in the
power meter operation manual.
2. Configure the equipment as shown in Figure 7-1.
NOTE
Connect the power splitter to the signal analyzer input using the appropriate
adapter. Do not use a cable.
3. Assure the signal analyzer’s built-in auto alignment has been performed
within the last 24 hours.
4. Press Mode, Spectrum Analyzer, Mode Preset on the signal analyzer, and
press Preset on the synthesized sweeper.
5. Set up the synthesized sweeper by pressing:
CW, 50, MHz
Power level, –54, dBm
6. Set up the signal analyzer by pressing:
Input/Output, More, Freq Ref In, External
FREQ Channel, Center Freq, 50, MHz
SPAN X Scale, Span, 50, kHz
AMPTD Y Scale, More, Internal Preamp, Full Range
AMPTD Y Scale, Attenuation, Mech Atten, 0, dB
AMPTD Y Scale, Ref Level, –55, dBm
7. Adjust the synthesized sweeper output power for a power meter reading
of –60 dBm ±0.1 dB.
8. On the signal analyzer, press Single.
9. Press the Peak Search key on the signal analyzer to position the marker on
the peak of the signal.
10.Refer to Table 7-1, “Frequency Response (Flatness) Results.” Enter the
amplitude of the signal displayed on the signal analyzer into the MeasAmp
column of Table 7-1.
11.Enter the power meter reading into the PowerMeter column of Table 7-1.
12.Tune the synthesized sweeper and signal analyzer to the next frequency
listed in Table 7-1.
13.Enter the power sensor calibration factor into the power meter.
14.For frequencies 3.6 GHz and above, press AMPTD Y Scale, then Presel
Center to center the preselector filter for an optimum amplitude
measurement.
38 Chapter 7
Frequency Response (Flatness), Preamp On
Procedure
15.Repeat step 7 through step 14 and complete the remainder of Table 7-1
for the frequency range of your preamp.
16.Compute the measurement error (MeasError = MeasAmp − PowerMeter).
17.Compute the flatness error normalized to 50 MHz:
(MeasError − MeasError @ 50 MHz)
18.Enter the computed flatness error value into the FlatNorm column of Table
7-1.
19.Compare the value of FlatNorm to the test limit.
Table 7-1
Frequency Response (Flatness) Results
Center
Frequency
Analyzer
Amplitude
Measamp
Power Meter
Measurement
Powermeter
Meas Error
Measerror
Flatness
Normalized
to 50 MHz
FlatNorm
Flatness Error
Test Limits
(dB)
50 MHz
0
Ref
1 GHz
±2.0 dB
2 GHz
±2.0 dB
3 GHz
±2.0 dB
3.5 GHz
±2.0 dB
4 GHz
±3.0 dB
6 GHz
±3.0 dB
8 GHz
±3.0 dB
9 GHz
±3.5 dB
11 GHz
±3.5 dB
13 GHz
±3.5 dB
14 GHz
±4.0 dB
17 GHz
±4.0 dB
20 GHz
±4.0 dB
23 GHz
±4.0 dB
26 GHz
±4.0 dB
29 GHz
±4.0 dB
32 GHz
±4.0 dB
35 GHz
±5.0 dB
38 GHz
±5.0 dB
41 GHz
±5.0 dB
Chapter 7 39
Frequency Response (Flatness), Preamp On
Procedure
42.5 GHz
±5.0 dB
43.5 GHz
±5.0 dB
46.5 GHz
±5.0 dB
49.5 GHz
±5.0 dB
Table 7-1
Frequency Response (Flatness) Results
Center
Frequency
Analyzer
Amplitude
Measamp
Power Meter
Measurement
Powermeter
Meas Error
Measerror
Flatness
Normalized
to 50 MHz
FlatNorm
Flatness Error
Test Limits
(dB)
40 Chapter 7
Frequency Response (Flatness), Preamp On
Procedure
Keysight X-Series Signal Analyzers
PXA N9030A
Functional Tests
8
Scale Fidelity
Test Limits
The scale fidelity error will be ≤ ±1.0 dB with ≤ −10 dBm at the mixer.
This test checks the scale fidelity of the instrument by maintaining a constant
reference level and measuring signals of different amplitudes over most of the
display range. This test sets the input attenuator to 10 dB and the Reference
Level to 0 dBm. The external attenuator is set to 0 dB, and the amplitude of the
source is adjusted to set the displayed signal at the reference level.
The instrument’s internal marker is used to measure the reference amplitude.
The Marker Delta function is activated and the RF input is reduced using the
external precision step attenuator. Signal input levels from 0 dBm to −50 dBm
are measured.
Item
Critical Specifications
(for this test)
Recommended
Keysight Model
Adapter, Type-N (m), to
3.5 mm (f)a
Frequency: 50 MHz
VSWR: < 1.1:1
1250-1745
Adapter, 3.5 mm (f) to 3.5
mm (f)a
Frequency: 50 MHz
VSWR: < 1.1:1
83059B
Adapter, 2.4 mm (f) to 3.5
mm (f)
(2 required)b
Frequency: 50 MHz
VSWR: < 1.1:1
11901B
Attenuator, 10 dB Step
Range: 0-50 dB
Frequency: 50 MHz
Accuracy: ±0.25 dB
8495A
Attenuator, 10 dB fixed (2
required)
Frequency: 50 MHz
VSWR: < 1.2:1
8493C, option 010
Cable, 3.5 mm, 1 meter (2
required)
Frequency: 50 MHz
VSWR: < 1.4:1
11500E
42 Chapter 8
Scale Fidelity
Figure 8-1
Scale Fidelity Setup
NOTE
Averaging is used for all measurements to improve repeatability and reduce
measurement uncertainty.
Cable, BNC 120 cm
Frequency: 10 MHz
10503A
Synthesized Sweeper
Output Level Accuracy: 0 to –15 dBm: ±1.0 dB
at 50 MHz
PSG
a. for Options 503, 508, 513, 526
b. for Options 543, 544, 550
Item
Critical Specifications
(for this test)
Recommended
Keysight Model
Chapter 8 43
Scale Fidelity
Procedure
Procedure
1. Configure the equipment as shown in Figure 8-1
2. Preset the Source and press Mode, Spectrum Analyzer, Mode Preset on
the analyzer.
3. Set up the synthesized sweeper by pressing:
Frequency, 50, MHz
Amplitude, +5, dBm
RF On/Off, On
4. Set up the signal analyzer by pressing:
Input/Output, More, Freq Ref In, External
FREQ Channel, Center Freq, 50, MHz
SPAN X Scale, Span, 1, MHz
AMPTD Y Scale, Ref Level, 0, dBm
Meas Setup, Average/Hold Number, 10, Enter
Trace/Detector, Trace Average
Peak Search
5. Set the external 10 dB step attenuator to 0 dB.
6. Adjust the amplitude on the signal source until the marker amplitude on
the analyzer reads –15 dBm ±0.2 dB.
7. On the analyzer, press the Single, Restart to trigger a 10 sweep average.
8. On the analyzer, activate the Marker Delta function by pressing Peak
Search, Marker Delta.
9. Perform the following steps for each attenuator setting listed in the table
below:
a. Select the next External attenuator setting.
b. Press the Restart key to trigger a 10 sweep average.
c. Enter the delta marker value into Table 8-1.
d. Check delta marker reading against the test limits.
44 Chapter 8
Scale Fidelity
Procedure
Table 8-1
Scale Fidelity Results
External Attenuator Setting
Minimum
(dB)
Marker Delta Value
(dB)
Maximum
(dB)
0
N/A
Reference
N/A
10
11.0
9.0
20
21.0
19.0
30
31.0
29.0
40
41.0
39.0
50
51.0
49.0
Keysight X-Series Signal Analyzers
PXA N9030A
Functional Tests
9
BBIQ Input Frequency Response
(Option BBA only)
Test Limits
This procedure tests the amplitude response of each of the BBIQ inputs versus
frequency. For each input, a reference measurement is made at 250 kHz and
then subsequent measurements are made at several frequencies up to the
maximum IQ bandwidth. The amplitude difference between the measurement
at 250 kHz and subsequent frequencies is calculated and compared to the test
limit.
Frequency Range
Limit Relative to 250 kHz
Frequency Range
Limit Relative to 250 kHz
250 kHz to 10 MHz (Standard)
± 1.75 dB
250 kHz to 25 MHz (Option B25)
± 1.75 dB
250 kHz to 40 MHz (Option S40 or SU1)
± 1.75 dB
Item
Critical Specifications
Recommended
Keysight Model
RF Signal Generator
Frequency: 250 kHz to 40 MHz
Amplitude: ~–10 dBm
Flatness: < ±0.75 dB
E4438C
Cable, BNC
Connectors: BNC male
Length: 23 cm (9 in) max
10502A
Cable, BNC, 120 cm
Connectors: BNC male
Frequency: 250 kHz to 40 MHz
10503A
50 Ohm Termination
Connector: BNC male
Frequency: 250 kHz to 40 MHz
11593A
46 Chapter 9
BBIQ Input Frequency Response (Option BBA only)
Figure 9-1
BBIQ Input Frequency Response Setup (I Input)
Figure 9-2
BBIQ Input Frequency Response Setup (I Input)
Figure 9-3
BBIQ Input Frequency Response Setup (Q Input)
Figure 9-4
BBIQ Input Frequency Response Setup (Q Input)
Chapter 9 47
BBIQ Input Frequency Response (Option BBA only)
Procedure
Procedure
1. On the X-Series analyzer, press Mode, IQ Analyzer (Basic). Press Mode
Preset.
2. Press Input/Output, I/Q (to change to I/Q inputs), I/Q (to bring up the I/Q
input menu), I/Q Cable Calibrate. Follow the instructions on screen. Use
the short BNC cable listed above to perform the calibration. This process
has several steps and will take about 3 minutes to perform.
3. Press I/Q Path, I+jQ, I Setup, Differential (On). Press Return, Q Setup, Q
Same as I (Yes).
4. Preset the RF Signal Generator and set the controls as follows:
Frequency 250 kHz
Amplitude –10 dBm
RF On
NOTE
Once the RF Signal Generator’s Amplitude is set, do not change the RF Signal
Generator’s Amplitude setting during the rest of this procedure.
5. Connect the longer BNC cable between the RF Signal Generator’s RF
OUTPUT and the I input of the X-Series analyzer, see Figure 9-1
6. Connect the 50 Ohm termination to the I input of the X-Series analyzer,
see Figure 9-1
7. Press Mode Preset on the X-Series analyzer. Press System, Show,
System. Note if either of the options listed below are present:
N9030A-B25
N9030A-S40
N9030A-B40
8. If none of the options listed in step 7 are present, the maximum IQ
bandwidth is 10 MHz
9. If N9030A-B25 is present, but N9030A-S40 and N9030A-B40 are not
present, the maximum IQ bandwidth is 25 MHz
10.If either N9030A-S40 or N9030A-B40 is present, the maximum IQ
bandwidth is 40 MHz.
11.Press FREQ on the X-Series analyzer and set the Center Freq equal to one
half the maximum IQ bandwidth determined above. For example, if the
maximum IQ bandwidth is 40 MHz, set the Center Freq to 20 MHz.
12.Press SPAN on the X-Series analyzer and set the Span equal to the
maximum IQ bandwidth determined above.
13.Press BW on the X-Series analyzer and set the Res BW to 100 kHz.
48 Chapter 9
BBIQ Input Frequency Response (Option BBA only)
Procedure
14.Press Peak Search on the X-Series analyzer. Record the marker amplitude
reading in Table 9-1 for the I Input at 250 kHz.
15.Set the RF Signal Generator to each of the remaining frequencies listed in
Table 9-1, up to the maximum IQ bandwidth of the X-Series analyzer. At
each frequency, do the following:
a. Press Peak Search
b. Record the marker amplitude reading in Table 9-1 for the I Input at
the current RF Signal Generator frequency
16.Set the RF Signal Generator frequency to 250 kHz
17.Connect the BNC Cable between the RF Signal Generator RF Output and
the I input on the X-Series analyzer, see Figure 9-2
18.Connect the 50 Ohm termination to the I input on the X-Series analyzer,
see Figure 9-2
Table 9-1
BBIQ Input Frequency Response, I and I Inputs
RF Signal
Generator
Frequency
I Input
I Input
Test Limits
Marker
Amplitude
Frequency
Response
Marker
Amplitude
Frequency
Response
250 kHz
0.0 dB
0.0 dB
±1.75 dB
2 MHz
±1.75 dB
5 MHz
±1.75 dB
7.5 MHz
±1.75 dB
9.9 MHz
±1.75 dB
Stop here if Maximum IQ Bandwidth is 10 MHz
12.5 MHz
±1.75 dB
15 MHz
±1.75 dB
17.5 MHz
±1.75 dB
20 MHz
±1.75 dB
22.5 MHz
±1.75 dB
24.9 MHz
±1.75 dB
Stop here if Maximum IQ Bandwidth is 25 MHz
30 MHz
±1.75 dB
35 MHz
±1.75 dB
39.9 MHz
±1.75 dB
Chapter 9 49
BBIQ Input Frequency Response (Option BBA only)
Procedure
19.Press Peak Search on the X-Series analyzer. Record the marker amplitude
reading in Table 9-1 for the I Input at 250 kHz.
20.Set the RF Signal Generator to each of the remaining frequencies listed in
Table 9-1, up to the maximum IQ bandwidth of the X-Series analyzer. At
each frequency, do the following:
a. Press Peak Search
b. Record the marker amplitude reading in Table 9-1 for the I Input at
the current RF Signal Generator frequency
21.Connect the BNC Cable between the RF Signal Generator RF Output and
the Q input on the X-Series analyzer, see Figure 9-3
22.Connect the 50 Ohm termination to the Q input on the X-Series analyzer,
see Figure 9-3
23.Set the RF Signal Generator frequency to 250 kHz.
24.Press Peak Search on the X-Series analyzer. Record the marker amplitude
reading in Table 9-2 for the Q Input at 250 kHz.
25.Set the RF Signal Generator to each of the remaining frequencies listed in
Table 9-2, up to the maximum IQ bandwidth of the X-Series analyzer. At
each frequency, do the following:
Press Peak Search
Record the marker amplitude reading in Table 9-2 for the Q Input at the
current RF Signal Generator frequency
Table 9-2
BBIQ Input Frequency Response, Q and Q Inputs
RF Signal
Generator
Frequency
Q Input
Q Input
Test Limits
Marker
Amplitude
Frequency
Response
Marker
Amplitude
Frequency
Response
250 kHz
0.0 dB
0.0 dB
±1.75 dB
2 MHz
±1.75 dB
5 MHz
±1.75 dB
7.5 MHz
±1.75 dB
9.9 MHz
±1.75 dB
Stop here if Maximum IQ Bandwidth is 10 MHz
12.5 MHz
±1.75 dB
15 MHz
±1.75 dB
17.5 MHz
±1.75 dB
20 MHz
±1.75 dB
50 Chapter 9
BBIQ Input Frequency Response (Option BBA only)
Procedure
26.Set the RF Signal Generator frequency to 250 kHz
27.Connect the BNC Cable between the RF Signal Generator RF Output and
the Q input on the X-Series analyzer, see Figure 9-4
28.Connect the 50 Ohm termination to the Q input on the X-Series analyzer,
see Figure 9-4
29.Press Peak Search on the X-Series analyzer. Record the marker amplitude
reading in Table 9-2 for the Q Input at 250 kHz.
30.Set the RF Signal Generator to each of the remaining frequencies listed in
Table 9-2, up to the maximum IQ bandwidth of the X-Series analyzer. At
each frequency, do the following:
a. Press Peak Search
b. Record the marker amplitude reading in Table 9-2 for the Q Input at
the current RF Signal Generator frequency
31.In Table 9-1 and Table 9-2, calculate the Frequency Response for each
input by subtracting the Marker Amplitude at 250 kHz for that input from
the Marker Amplitude for each frequency greater than 250 kHz.
32.Compare the Frequency Response values to the Test Limits.
22.5 MHz
±1.75 dB
24.9 MHz
±1.75 dB
Stop here if Maximum IQ Bandwidth is 25 MHz
30 MHz
±1.75 dB
35 MHz
±1.75 dB
39.9 MHz
±1.75 dB
Table 9-2
BBIQ Input Frequency Response, Q and Q Inputs
RF Signal
Generator
Frequency
Q Input
Q Input
Test Limits
Marker
Amplitude
Frequency
Response
Marker
Amplitude
Frequency
Response
Frequency Response = Marker Amplitude (Freq > 250 kHz) – Marker Amplitude (250 kHz)
amplitude accuracy test 23
option 1DS 27
amplitude linearity test 41
B
BBIQ frequency response 45
D
DANL test 11
displayed average noise level. See DANL 11
E
equipment
functional tests 7
warm-up time 6
F
frequency readout accuracy test 15
frequency response
(flatness) test 29
BBIQ 45
frequency response (flatness) test
preamp on 35
functional testing
performance verification 4
functional tests
before performing 6
equipment list 7
introduction 3
vs performance verification tests 4
warm-up time 6
See also individual functional tests 3
P
performance verification tests vs functional tests 4
S
second harmonic distortion test 19
T
tests. See functional tests
This information is subject to change without notice.
© Keysight Technologies, 2015
Published in USA, January 2015
N9030-90018
www.keysight.com