Using Source/Measure Unit as a Voltmeter
B2900A Precision Source/Measure Unit
Demo Guide
02 | Keysight | Using Source/Measure Unit as a Voltmeter – Demo Guide
Introduction
The Keysight B2900A Series Precision Source/Measure Unit (SMU) is a compact and
cost-effective bench-top SMU with the capability to output and measure both voltage
and current. The B2900A Series SMU enables you to make a wide range of current
versus voltage (IV) measurements more accurately and quickly than ever before. In
addition, the B2900A Series SMU comes with an intuitive graphical user interface (GUI)
and free PC-based application software that make it easy for you to begin making
productive measurements immediately.
This demonstration guide shows how easily you can use the Keysight B2900A Series
SMU as a voltmeter.
Required Instrument and Accessories
All All of the accessories required to perform the demos described in this demonstration
guide are provided in a demo kit that is included with Keysight B2902A/12A demo units.
The kit includes items such as a banana cable, a 1.1 kΩ resistor, etc.
Keysight B2902A/12A
Precision Source/Measure Unit
11059A Kelvin Probe Set
1.1 kΩ Resistor
U8201A Combo Test Lead Kit
03 | Keysight | Using Source/Measure Unit as a Voltmeter – Demo Guide
Figure 1. Connection diagram and basic condition
Concept
Figure 1 illustrates the connection diagram used in the demo to use the Keysight B2900A
Series SMU as a voltmeter. Since the low terminal of the channels in the B2900A Series
SMU is grounded internally in the initial state, you need to configure it floating.
In this demo, a 1.1 kΩ resistor combined with a current source is used as a DUT and mea-
sure voltage across the resistor. The channel 2 of the B2902A/12A is used as a current
source. However, you can also use your own current source for it
Setup
1. Connect the yellow banana plug of the 11059A to the Ch1 Low Force Terminal.
2. Connect the red banana plug of the 11059A to the Ch1 High Force Terminal.
3. Connect the black banana cable of the U8201A to the Ch2 Low Force Terminal.
4. Connect the red banana cable of the U8201A to the Ch2 High Force Terminal.
5. Clip the end of the 1.1 kΩ resistor with the black alligator clip of the U8201A.
6. Clip the same end with the black gold-plated tweezers of the 11059A.
7. Clip the other end of the 1.1 kΩ resistor with the red alligator clip of the U8201A
8. Clip the same end of 1.1 kΩ resistor which the red alligator clip is clipped with the red
gold-plated tweezers of the 11059A.
A -
-
V
Common
High Force
Low Force
DUT
High
Low
B2901/02/11/12A
1.1 kΩ
Current
Source
DUT
High
Low
A -
R
Channel 2
as Current Source
Black (11059A)
Red (11059A)
U8201A
Channel 1
DUT
1.1 kΩ
Black
Brown
Brown
Brown
Brown
Red (11059A)
Yellow (11059A)
A -
-
V
Common
High Force
Low Force
B2901/02/11/12A
1.1 kΩ
Current
Source
DUT
High
Low
A -
R
04 | Keysight | Using Source/Measure Unit as a Voltmeter – Demo Guide
LAB1: Use Source/Measure Unit as an Voltmeter
Objective
This demo illustrates the voltage
measurement function as a voltmeter
by sourcing almost 0 A current and
measuring the voltage across the resis-
tor biased by a current source using a
Source/Measure Unit.
Procedure
1. Configure the low terminal state to
FLOATING
2. Measure voltage using the SMU
(Channel 1) as a voltmeter
3. Enable the DUT
4. Confirm measured voltage
5. (Optional) Configuring the
measurement speed
6. (Optional) Configuring the
measurement range operation
Demonstration
In the default setting, the low terminal of the channels in the Keysight B2900A Series SMU is grounded
internally. However, the low terminal can be disconnected from the ground and kept floating. Configur-
ing the low terminal state to FLOATING, which enables you to connect the low terminal to any potential
up to ±250 V, and measuring voltage with sourcing 0 A from the channel, which makes it possible to
use the channel as a voltmeter.
1. Configure the low terminal state to FLOATING
1) If you aren’t on the top of the Function menu, press repeatedly to return to the top level.
2) Press , , and then press to open the Output Connection
dialogue.
3) Press and select to specify the channel which the Low terminal state is
configured for.
On the top level of the Function menu
In the middle level of the Function menu
Any tree structure can’t be seen
Some tree structure can be seen
(1) Press Config
(2) Press Source
(3) Press Connection
05 | Keysight | Using Source/Measure Unit as a Voltmeter – Demo Guide
4) Press and select , and then press to configure the Low terminal
state to FLOATING.
If the low terminal state of the channel is set to FLOATING, you can see the status
indicator on the GUI as below, although no indicator can be seen on being set to
GROUNDED.
2. Measure voltage using the SMU (Channel 1) as a voltmeter
2-1. Change View mode to Ch1 Single View
1) Press repeatedly until the
Channel 1 Single View is displayed
2-2. Configure the condition to source and measure
1) Press to edit the Source function, and then select to set the Source
function to Current source.
GROUNDED
FLOATING
06 | Keysight | Using Source/Measure Unit as a Voltmeter – Demo Guide
2) Rotate and press to edit the Current source range operation. Then Select
to set the Current source range operation to FIXED.
3) Rotate and press to edit the Current source range, and then select
to set it to 10 nA. (If you couldn’t find , select instead.)
4) Press to edit the Source value, and then enter 0.001 nA (1 pA) to set the
Source value to 1 pA.
5) Press to edit the Limit value, and then enter 2 V to set the Limit value to 2 V
for example.
6) Press to configure the Measurement parameter, and then select to
set the Measurement parameter to Voltage.
07 | Keysight | Using Source/Measure Unit as a Voltmeter – Demo Guide
2-3. Perform the measurement
1) Press for the channel 1 to switch on its output terminal.
2) Press to perform a measurement repeatedly. Now you can see the
measurement result on the GUI of the B2902/12A as below.
3. Enable the DUT
You can also use your own current source instead of the channel 2 of the B2902A/12A
3-1. Change View mode to Ch2 Single View.
1) Press repeatedly until the
Channel 2 Single View is displayed.
3-2. Configure the condition of the current source (Channel 2)
1) Press to edit the Source function, and then select to set the Source
function to Current source.
2) Press to edit the Source value, and then enter 1 mA to set the Source value
to 1 mA for example.
3) Press to edit the Limit value, and then enter 2 V to set the Limit value to 2 V
for example.
08 | Keysight | Using Source/Measure Unit as a Voltmeter – Demo Guide
4) Press for the channel 2 to switch on its output terminal and enable the DUT.
4. Confirm measured voltage
4-1. Change View mode to Ch1 Single View.
1) Press repeatedly until
Channel 1 Single View is displayed.
You can see about 1.1 V as measured value, since 1 mA is sourced to a 1.1 kΩ
resistor. If you press for the channel 2 to switch it off, you can see only the
offset voltage because the DUT is disabled.
Theoretically speaking, the measured voltage should be 1.1 V since 1mA is sourced
to a 1.1 kΩ resistor. However, it may be varied because the resistor has some error on
its value actually.
Current Source enabled
Current Source disabled
09 | Keysight | Using Source/Measure Unit as a Voltmeter – Demo Guide
5. (Optional) Configuring the measurement speed
In the default setting, the instrument selects the appropriate measurement speed and
range automatically to get the fine accuracy. However, you can also specify these pa-
rameters on the GUI of the B2900A Series SMU to meet a variety of the requirement to
the measurement conditions.
For example, let’s try to change the measurement speed to NORMAL to make a mea-
surement more carefully. If you select NORMAL, the aperture time is set to 1 PLC. Here,
PLC stands for power line cycle and the specified number of power line cycles is used
per a measurement.
1) Press to edit the Measurement speed, and then select to set the
Measurement speed to NORMAL. (If you can’t see in Assist keys, press
to change the keys shown in Assist keys.)
6. (Optional) Configuring the measurement range operation
The parameters which configure the measurement range operation can be displayed in
the Range Sub-panel in the Channel 1 Single View. In the default setting, the B2900A
Series SMU performs the voltage measurement using a 200 mV voltage minimum mea-
surement range with AUTO range operation. With AUTO range operation, the B2900A
Series SMU selects the proper range for the measurement with the specified minimum
measurement range so that you don’t need to take care about it. To know how to change
the measurement range setting, try to configure to use the 2 V voltage minimum mea-
surement range with AUTO range operation.
1) Rotate and press to edit the Voltage minimum measurement range, and
then select to set it to 2 V.
If you’d like to fix the measurement range, you can select FIXED range operation as below.
2) Rotate and press to edit the Voltage measurement range operation. Then
Select to set the Voltage measurement range operation to FIXED
10 | Keysight | Using Source/Measure Unit as a Voltmeter – Demo Guide
B2900 Precision Instrument Family
The B2900 family contains products that perform both precision sourcing and
precision measurement. www.keysight.com/find/b2900a
Conclusion
The Keysight B2900A Series Precision Source/Measure Unit (SMU) is a compact and
cost-effective bench-top SMU with the capability to output and measure both voltage
and current. Although it has the capability to make a wide range of current versus volt-
age (IV) measurements as its intrinsic function, the B2900A Series SMU can be used as a
voltmeter easily.
A personalized view into the information most relevant to you.
Keysight Infoline
www.keysight.com/find/Infoline
Keysight’s insight to best in class information management. Free access to
your Keysight equipment company reports and e-library.
KEYSIGHT
SERVICES
Keysight Services
www.keysight.com/find/service
Our deep offering in design, test, and measurement services deploys an
industry-leading array of people, processes, and tools. The result? We help
you implement new technologies and engineer improved processes that
lower costs.
Three-Year Warranty
www.keysight.com/find/ThreeYearWarranty
Keysight’s committed to superior product quality and lower total cost
of ownership. Keysight is the only test and measurement company with
three-year warranty standard on all instruments, worldwide. And, we provide
a one-year warranty on many accessories, calibration devices, systems and
custom products.
Keysight Assurance Plans
www.keysight.com/find/AssurancePlans
Up to ten years of protection and no budgetary surprises to ensure your
instruments are operating to specification, so you can rely on accurate
measurements.
Keysight Channel Partners
www.keysight.com/find/channelpartners
Get the best of both worlds: Keysight’s measurement expertise and product
breadth, combined with channel partner convenience.
For more information on Keysight
Technologies’ products, applications or
services, please contact your local Keysight
office. The complete list is available at:
www.keysight.com/find/contactus
Americas
Canada
(877) 894 4414
Brazil
55 11 3351 7010
Mexico
001 800 254 2440
United States
(800) 829 4444
Asia Pacific
Australia
1 800 629 485
China
800 810 0189
Hong Kong
800 938 693
India
1 800 11 2626
Japan
0120 (421) 345
Korea
080 769 0800
Malaysia
1 800 888 848
Singapore
1 800 375 8100
Taiwan
0800 047 866
Other AP Countries
(65) 6375 8100
Europe & Middle East
Austria
0800 001122
Belgium
0800 58580
Finland
0800 523252
France
0805 980333
Germany
0800 6270999
Ireland
1800 832700
Israel
1 809 343051
Italy
800 599100
Luxembourg
+32 800 58580
Netherlands
0800 0233200
Russia
8800 5009286
Spain
800 000154
Sweden
0200 882255
Switzerland
0800 805353
Opt. 1 (DE)
Opt. 2 (FR)
Opt. 3 (IT)
United Kingdom
0800 0260637
For other unlisted countries:
www.keysight.com/find/contactus
(BP-06-08-16)
DEKRA Certified
ISO9001 Quality Management System
www.keysight.com/go/quality
Keysight Technologies, Inc.
DEKRA Certified ISO 9001:2015
Quality Management System
Evolving
Our unique combination of hardware, software, support, and people can help
you reach your next breakthrough. We are unlocking the future of technology.
From Hewlett-Packard to Agilent to Keysight
This information is subject to change without notice.
© Keysight Technologies, 2016
Published in USA, August 18, 2016
5992-1709EN
www.keysight.com
11 | Keysight | Using Source/Measure Unit as a Voltmeter – Demo Guide