Stratos Multi E401N
Industrial Transmitters
www.knick.de
Read before installation.
Keep for future use.
Stratos Multi E401N
ii
Supplemental Directives
These supplemental directives explain how safety information is laid out in this document and what
content it covers.
Safety Chapter
This document's safety chapter is designed to give the reader a basic understanding of safety.
It illustrates general hazards and gives strategies on how to avoid them.
Safety Guide
The external safety guide is designed to give the reader a basic understanding of safety.
It illustrates general hazards and gives strategies on how to avoid them.
Warnings
Stratos Multi E401 was subjected to a risk assessment. Nevertheless, not all risks can be sufficiently
reduced. This document uses the following warnings to indicate hazardous situations:
Symbol
Category
Meaning
Remark
WARNING
Designates a situation that can lead to death or serious
(irreversible) injury
The warnings contain
information on how to
avoid the hazard.
CAUTION
Designates a situation that can lead to slight or moderate
(reversible) injury
Without NOTICE
Designates a situation that can lead to property or environ-
mental damage
Additional Safety Information
• Stratos Multi E401 safety guide
Symbols Used in this Document
Symbol
Meaning
Cross-reference to content within this document
Interim or final results in instructions for action
Item number in figure
Item number in text
Stratos Multi E401N
iii
Table of Contents
1 Safety.............................................................................................................................................
8
1.1
Intended Use..........................................................................................................................................................
8
1.2
Symbols and Markings on the Product........................................................................................................
8
1.3
Personnel Requirements....................................................................................................................................
8
1.4
Safety Training.......................................................................................................................................................
8
1.5
Installation and Commissioning.....................................................................................................................
9
1.6
Maintenance and Repair.................................................................................................................................... 10
1.7
Residual Risks......................................................................................................................................................... 10
2 Product.......................................................................................................................................... 11
2.1
Design and Function........................................................................................................................................... 11
2.2
Product Range and Options............................................................................................................................. 12
2.3
System Overview.................................................................................................................................................. 14
2.3.1
Typical Applications............................................................................................................................... 15
2.4
Package Contents and Product Identification........................................................................................... 17
2.4.1
Nameplate ................................................................................................................................................. 18
2.5
Symbols and Markings on the Display ......................................................................................................... 18
3 Commissioning............................................................................................................................. 20
3.1
Assembly ................................................................................................................................................................. 20
3.1.1
Dimension Drawings ............................................................................................................................. 21
3.1.2
Housing Mounting Options ................................................................................................................ 22
3.1.3
Pipe Mounting ZU0274......................................................................................................................... 23
3.1.4
Protective Hood for Wall and Pipe Mounting ZU0737 .............................................................. 24
3.1.5
Panel-Mount Kit ZU0738 ...................................................................................................................... 25
3.1.6
Blanking Plugs, Reduction Sealing Inserts, Multiple Sealing Inserts.................................... 26
3.2
Electrical Installation........................................................................................................................................... 27
3.2.1
Relay Contacts: Protective Wiring ..................................................................................................... 27
3.2.2
Installing Active and Passive Current Outputs ............................................................................. 29
3.2.3
Terminal Assignments........................................................................................................................... 30
3.3
Sensor Connection............................................................................................................................................... 31
3.3.1
Connecting a Memosens Sensor/Optical Oxygen Sensor (LDO)........................................... 31
3.3.2
Connecting an Analog Sensor/Second Memosens Channel .................................................. 32
3.3.3
Cable Preparation for SE655/SE656.................................................................................................. 33
3.4
Terminal Assignments of Measuring Modules........................................................................................... 34
3.5
Final Check During Commissioning .............................................................................................................. 35
4 Operation and Use........................................................................................................................ 36
4.1
Changing the User Interface Language....................................................................................................... 36
4.2
Keypad and Display............................................................................................................................................. 36
4.3
Menu Structure Overview................................................................................................................................. 39
4.4
Access Control ....................................................................................................................................................... 39
4.5
Operating States................................................................................................................................................... 40
4.6
Measurement Display......................................................................................................................................... 40
Stratos Multi E401N
iv
5 Parameter Setting ........................................................................................................................ 41
5.1
Operating Levels................................................................................................................................................... 41
5.2
Locking a Function............................................................................................................................................... 42
5.3
System Control...................................................................................................................................................... 43
5.3.1
Transfer Configuration .......................................................................................................................... 44
5.3.2
Parameter Sets ......................................................................................................................................... 45
5.3.3
Function Control...................................................................................................................................... 46
5.3.4
Calculation Blocks (TAN Option FW-E020)..................................................................................... 46
5.3.5
Time/Date .................................................................................................................................................. 46
5.3.6
Measuring Point Description.............................................................................................................. 46
5.3.7
Firmware Update (TAN Option FW-E106)....................................................................................... 46
5.3.8
Option Activation.................................................................................................................................... 47
5.3.9
Logbook ..................................................................................................................................................... 47
5.3.10 Measurement Recorder (TAN Option FW-E103) .......................................................................... 47
5.3.11 Buffer Table (TAN Option FW-E002).................................................................................................. 47
5.3.12 Concentration Table (TAN Option FW-E009)................................................................................. 48
5.3.13 Restore Factory Settings....................................................................................................................... 48
5.3.14 Passcode Entry......................................................................................................................................... 48
5.4
Parameter Setting, General............................................................................................................................... 48
5.4.1
Configuring the Measurement Display........................................................................................... 49
5.4.2
Display......................................................................................................................................................... 54
5.4.3
Measurement Recorder (TAN Option FW-E103) .......................................................................... 54
5.5
Inputs and Outputs.............................................................................................................................................. 54
5.5.1
Current Outputs....................................................................................................................................... 55
5.5.2
Relay Contacts.......................................................................................................................................... 57
5.5.3
PID Controller ........................................................................................................................................... 61
5.5.4
Control Inputs........................................................................................................................................... 62
5.6
Sensor Selection [I] [II] ........................................................................................................................................ 63
5.7
Process Variable pH ............................................................................................................................................. 64
5.7.1
Sensor Data ............................................................................................................................................... 66
5.7.2
Calibration Presettings.......................................................................................................................... 68
5.7.3
Temperature Compensation of Process Medium........................................................................ 69
5.7.4
Delta Function.......................................................................................................................................... 70
5.7.5
Messages.................................................................................................................................................... 71
5.8
Process Variable ORP........................................................................................................................................... 72
5.8.1
Sensor Data ............................................................................................................................................... 73
5.8.2
Calibration Presettings.......................................................................................................................... 74
5.8.3
Delta Function.......................................................................................................................................... 74
5.8.4
Messages.................................................................................................................................................... 75
5.9
Process Variable Conductivity (Contacting)................................................................................................ 76
5.9.1
Sensor Data ............................................................................................................................................... 77
5.9.2
Calibration Presettings.......................................................................................................................... 79
5.9.3
Temperature Compensation of Process Medium........................................................................ 80
5.9.4
Concentration (TAN Option FW-E009)............................................................................................. 80
5.9.5
USP Function ............................................................................................................................................ 81
5.9.6
Messages.................................................................................................................................................... 82
5.10 Process Variable Conductivity (Inductive)................................................................................................... 83
5.10.1 Sensor Data ............................................................................................................................................... 85
5.10.2 Calibration Presettings.......................................................................................................................... 86
5.10.3 Concentration (TAN Option FW-E009)............................................................................................. 87
5.10.4 USP Function ............................................................................................................................................ 87
5.10.5 Messages.................................................................................................................................................... 88
Stratos Multi E401N
v
5.11 Dual Conductivity Measurement.................................................................................................................... 89
5.12 Process Variable Oxygen.................................................................................................................................... 89
5.12.1 Sensor Data ............................................................................................................................................... 92
5.12.2 Calibration Presettings.......................................................................................................................... 95
5.12.3 Pressure Correction................................................................................................................................ 95
5.12.4 Salinity Correction .................................................................................................................................. 96
5.12.5 Messages.................................................................................................................................................... 96
5.13 Flow........................................................................................................................................................................... 97
5.14 HART Communication (TAN Option FW-E050).......................................................................................... 97
6 Calibration/Adjustment............................................................................................................... 98
6.1
Memosens Calibration/Adjustment .............................................................................................................. 98
6.2
Process Variable pH Calibration/Adjustment............................................................................................. 99
6.2.1
Calibration Methods ..............................................................................................................................100
6.2.2
Temperature Compensation during Calibration .........................................................................101
6.2.3
Calibration/Adjustment Methods.....................................................................................................101
6.2.4
Calibration Mode: Calimatic................................................................................................................101
6.2.5
Calibration Mode: Manual....................................................................................................................103
6.2.6
Calibration Mode: Product...................................................................................................................104
6.2.7
Calibration Mode: Data Entry .............................................................................................................105
6.2.8
ISFET Zero Point Calibration Mode...................................................................................................105
6.2.9
Calibration Mode: Temperature.........................................................................................................106
6.3
Process Variable ORP Calibration/Adjustment ..........................................................................................107
6.3.1
Calibration Mode: ORP Data Entry....................................................................................................107
6.3.2
Calibration Mode: ORP Adjustment .................................................................................................107
6.3.3
Calibration Mode: ORP Check.............................................................................................................109
6.3.4
Calibration Mode: Temperature.........................................................................................................109
6.4
Process Variable Conductivity (Contacting) Calibration/Adjustment ...............................................110
6.4.1
Temperature Compensation during Calibration .........................................................................110
6.4.2
Calibration/Adjustment Methods.....................................................................................................111
6.4.3
Calibration Mode: Automatic..............................................................................................................111
6.4.4
Calibration Mode: Manual....................................................................................................................112
6.4.5
Calibration Mode: Product...................................................................................................................113
6.4.6
Calibration Mode: Data Entry .............................................................................................................114
6.4.7
Calibration Mode: Temperature.........................................................................................................114
6.5
Process Variable Conductivity (Inductive) Calibration/Adjustment ..................................................115
6.5.1
Temperature Compensation during Calibration .........................................................................115
6.5.2
Calibration/Adjustment Methods.....................................................................................................116
6.5.3
Calibration Mode: Automatic..............................................................................................................116
6.5.4
Calibration Mode: Manual....................................................................................................................117
6.5.5
Calibration Mode: Product...................................................................................................................118
6.5.6
Calibration Mode: Zero Point..............................................................................................................119
6.5.7
Calibration Mode: Installation Factor...............................................................................................119
6.5.8
Calibration Mode: Data Entry .............................................................................................................120
6.5.9
Calibration Mode: Temperature.........................................................................................................120
6.6
Process Variable Oxygen Calibration/Adjustment ...................................................................................121
6.6.1
Calibration/Adjustment Methods.....................................................................................................122
6.6.2
Calibration Mode: In Air........................................................................................................................122
6.6.3
Calibration Mode: In Water..................................................................................................................123
6.6.4
Calibration Mode: Data Entry .............................................................................................................124
6.6.5
Calibration Mode: Product...................................................................................................................125
6.6.6
Calibration Mode: Zero Point..............................................................................................................126
6.6.7
Calibration Mode: Temperature.........................................................................................................127
Stratos Multi E401N
vi
7 Diagnostics ...................................................................................................................................128
7.1
Favorites Menu......................................................................................................................................................128
7.2
Diagnostic Functions ..........................................................................................................................................129
7.2.1
Overview of Diagnostic Functions....................................................................................................129
7.2.2
Messages....................................................................................................................................................129
7.2.3
Logbook .....................................................................................................................................................129
7.2.4
Device Information.................................................................................................................................130
7.2.5
Device Test.................................................................................................................................................131
7.2.6
Measuring Point Description..............................................................................................................131
7.2.7
Channel I/II Diagnostic Functions .....................................................................................................131
8 Maintenance .................................................................................................................................133
8.1
Overview of Maintenance Functions ............................................................................................................133
8.2
Channel I/II Maintenance Functions..............................................................................................................134
8.2.1
Sensor Monitor.........................................................................................................................................134
8.2.2
Maintenance Functions for Oxygen Sensors ................................................................................134
8.3
Manual Function Check .....................................................................................................................................134
8.3.1
Current Source .........................................................................................................................................134
8.3.2
Relay Test....................................................................................................................................................134
8.3.3
Controller Test ..........................................................................................................................................134
9 Decommissioning.........................................................................................................................135
9.1
Disposal....................................................................................................................................................................135
9.2
Returns.....................................................................................................................................................................135
10 Troubleshooting...........................................................................................................................136
10.1 Malfunction States...............................................................................................................................................136
10.2 Error Messages ......................................................................................................................................................137
10.3 Sensocheck and Sensoface...............................................................................................................................151
11 Accessories....................................................................................................................................153
11.1 Memory Card .........................................................................................................................................................153
12 Specifications................................................................................................................................155
12.1 Power........................................................................................................................................................................155
12.2 Inputs and Outputs (SELV, PELV).....................................................................................................................155
12.3 Device.......................................................................................................................................................................157
12.7 Interfaces.................................................................................................................................................................158
12.8 Measuring Functions...........................................................................................................................................159
12.8.1 pH .................................................................................................................................................................159
12.8.2 Cond.............................................................................................................................................................161
12.8.3 CondI ...........................................................................................................................................................162
12.8.4 Oxygen........................................................................................................................................................163
12.8.5 Temperature Compensation (Conductivity)..................................................................................165
12.8.6 Concentration Determination, Conductivity (TAN Option FW-E009)..................................165
12.9 Diagnostics and Statistics..................................................................................................................................166
Stratos Multi E401N
vii
13 Annex ............................................................................................................................................167
13.1 Channel II Wiring Examples ..............................................................................................................................168
13.1.1 pH Analog Wiring Examples................................................................................................................168
13.1.2 ORP Analog Wiring Example...............................................................................................................174
13.1.3 Contacting Conductivity Wiring Examples....................................................................................175
13.1.4 Inductive Conductivity Wiring Examples .......................................................................................182
13.1.5 Dual Conductivity Wiring Examples.................................................................................................184
13.1.6 Oxygen Wiring Examples .....................................................................................................................187
13.2 Buffer Tables...........................................................................................................................................................189
13.3 Calibration Solutions...........................................................................................................................................197
13.4 TAN Options ...........................................................................................................................................................199
13.4.1 pH Buffer Table: Entry of Individual Buffer Set (FW-E002)........................................................199
13.4.2 Current Characteristic (FW-E006) ......................................................................................................201
13.4.3 Concentration Determination (FW-E009).......................................................................................201
13.4.4 Calculation Blocks (FW-E020)..............................................................................................................206
13.4.5 HART (FW-E050).......................................................................................................................................209
13.4.6 Parameter Sets 1-5 (FW-E102) ............................................................................................................210
13.4.7 Measurement Recorder (FW-E103)...................................................................................................211
13.4.8 Firmware Update (FW-E106) ...............................................................................................................213
14 Basics.............................................................................................................................................215
14.1 Basics of PID Control............................................................................................................................................215
Index..............................................................................................................................................217
Stratos Multi E401N
8
1 Safety
The following safety instructions contain the necessary information for the safe use of the product.
If you have any questions, please contact Knick Elektronische Messgeräte GmbH & Co. KG using the
information provided on the back page of this document.
1.1 Intended Use
Stratos Multi is an industrial transmitter in 4-wire technology. In the field of liquid analysis, it can
measure pH values, ORP, conductivity (contacting and inductive), and oxygen content both dissolved
and in the gaseous phase.
Alongside a permanently installed measuring channel I for Memosens sensors, the modular
transmitter has a slot that can be equipped with any analog or digital measuring module (measuring
channel II). The transmitter can be extended with device-specific add-on functions called TAN options.
The defined rated operating conditions must be observed when using this product. These conditions
are set out in full in the Specifications chapter of the user manual as well as in parts in the installation
guide for the Stratos Multi.
Using the product improperly or for any purpose other than the product's intended purpose is not
permitted and may result in injury to persons or damage to objects or the environment.
1.2 Symbols and Markings on the Product
Attaching the CE marking to the product means that the product satisfies the applicable requirements
specified in the European Union harmonization legislation.
Special conditions and the product’s possible danger points. Read the user manual,
observe the specifications, and follow the instructions in the safety guide.
Reminder to read the documentation
Protection class II
1.3 Personnel Requirements
The operating company must ensure that personnel are sufficiently qualified in accordance with the
local and national codes and regulations that apply for the area in which the product is being used.
The personnel must be authorized by the operating company and instructed in handling this product.
Personnel
Jobs in the process industry
Technicians
Operation of machines and plants, failure elimination
Electricians
Installation, systematic troubleshooting and maintenance of machines and plants
System specialists
Setup and commissioning, parameter setting, calibration, and adjustment on
the administrator level of the Stratos Multi.
1.4 Safety Training
Upon request, Knick Elektronische Messgeräte GmbH & Co. KG will provide safety instruction and
product training during initial commissioning of the product. Further information is available from
the relevant Knick representatives.
Stratos Multi E401N
9
1.5 Installation and Commissioning
Observe all applicable local and national codes and standards for the installation of electrical
equipment. Information on installation is provided in the installation guide for the Stratos Multi.
Cable
Only use cables with a suitable temperature resistance.
Transmitter
Cable temperature resistance
Stratos Multi E401N
75 °C / 167 °F or higher
Mains Connection
The transmitter does not have a power switch. An appropriately arranged and accessible disconnect-
ing device for the transmitter must be present in the system installation. The disconnecting device
must disconnect all non-grounded, current-carrying wires. The disconnecting device must be labeled
in a way that enables the associated transmitter to be identified. The power line may carry dangerous
touch voltages. Touch protection must be ensured by proper installation.
Inputs and Outputs (SELV, PELV)
All inputs and outputs must be connected to SELV/PELV circuits.
Parameter Setting, Calibration, and Adjustment
Incorrect parameter setting, calibration, and adjustment can result in incorrect outputs.
For this reason, Stratos Multi must be commissioned by a specialist, who will also set the parameters,
calibrate and adjust the transmitter.
Relay Contacts
Make sure that the maximum ratings of the relay contacts are not exceeded even during switching.
Relay contacts are subject to electrical erosion. Especially with inductive and capacitive loads,
the service life of the relay contacts will be reduced.
Degree of Protection
The housing of the device is dust-tight and offers complete protection against contact as well as
protection against strong water jets.
• Europe: IP protection IP66/IP67
• USA: NEMA 4X Outdoor (with pressure compensation)
Memory Card
Dangerous touch voltage may be present in the terminal compartment when the device is opened.
Professional installation guarantees direct contact protection.
The memory card can be replaced during operation. When doing so, maintain sufficient distance from
the power lines and do not use tools.
Stratos Multi E401N
10
1.6 Maintenance and Repair
Maintenance
Maintenance work on the Stratos Multi and the sensors is only permitted in the function check (HOLD)
mode, since otherwise there is a risk of unexpected system behavior.
The function check (HOLD) mode is activated on the device during:
• Calibration (the selected channel only)
• Maintenance (current source, measuring points)
• Parameter setting on the operator and administrator levels
Repair
The Stratos Multi and measuring module cannot be repaired by the user. To request a repair, please
contact Knick Elektronische Messgeräte GmbH & Co. KG by visiting www.knick.de.
1.7 Residual Risks
The product has been developed and manufactured in accordance with generally accepted safety
rules and regulations. However, it is not possible to rule out all risks.
The effects of moisture, ambient temperature, chemicals, and corrosion can negatively impact the safe
operation of the product. Please observe the following instructions:
Knick Elektronische Messgeräte GmbH & Co. KG recommends that the Stratos Multi E401 is installed in
a weather-proofed area of the plant or that a cover is used for protection against the weather.
Stratos Multi E401N
11
2 Product
2.1 Design and Function
Stratos Multi is designed as follows:
• The 1- and 2-channel transmitter permits a free combination of the pH/ORP process variables,
conductivity (2-/4-electrode sensors, toroidal sensors), and oxygen, and can, for example,
simultaneously measure pH values and conductivity.
• The device has four current outputs (active/passive).
• Add-on product options can be enabled by entering a transaction number (TAN).
• The color TFT display has white backlighting and provides clear information on errors during
parameter setting or measurement in accordance with NAMUR recommendations.
• The user interface is multi-lingual with full-text menu navigation presently available in
the following languages: German, English, French, Italian, Spanish, Portuguese, and Chinese.
Function
TAN option
1 measuring channel
2 current outputs
Second measuring channel via add-on measuring module
3rd and 4th current outputs
X
Multiparameter: Process variables freely selectable from pH, ORP, oxygen,
conductivity (contacting/inductive)
Door contact
HART
X
3 freely assignable relay contacts
for NAMUR messages (failure, maintenance required, out of specification, function check),
setpoint alarm relay, controller, rinse contact parameter set, USP (for conductivity), Sensoface
Function check input (HOLD)
2 control Inputs
Flow measurement
Current Input
X
Stratos Multi E401N
12
2.2 Product Range and Options
Version
Possible combinations
1-channel (basic unit)
1x Memosens sensor
1x SE740 optical oxygen sensor
1x analog sensor via measuring module (MK module)
2-channel
2x Memosens sensors (1x via MK module)
1x Memosens sensor (MK module)
and 1x SE740 optical oxygen sensor
1x Memosens sensor
and 1x analog sensor via measuring module (MK module)
1x CC measuring module for dual conductivity measurement
The transmitters have the designations E401N for safe areas.
Product Line
Device (digital basic unit)
Order No.
Stratos E401N
E401N
Measuring module for analog sensors or 2-channel Memosens,
non-intrinsically safe
Order No.
pH value, ORP measurement
MK-PH015N
Oxygen measurement
MK-OXY046N
Contacting conductivity measurement (process-wetted)
MK-COND025N
Toroidal conductivity measurement
MK-CONDI035N
Dual conductivity measurement
MK-CC065N
Memosens multiparameter (for 2-channel version)
MK-MS095N
The following add-on functions (TAN options) can be enabled by entering a TAN:
Add-on function (TAN option)
Order No.
pH buffer table: Entry of individual buffer set
FW-E002
Current characteristic
FW-E006
Concentration determination for use with conductivity sensors
FW-E009
Oxygen measurement in low oxygen concentrations
FW-E015
Operation with double high-impedance pH sensors
FW-E017
Calculation blocks
FW-E020
HART
FW-E050
Current input
FW-E051
Current outputs 3 and 4
FW-E052
Parameter sets 1 to 5
FW-E102
Measurement recorder
FW-E103
Logbook, in conjunction with Data Card
FW-E104
Firmware update
FW-E106
Mounting accessories
Order No.
Pipe-mount kit
ZU 0274
Panel-mount kit
ZU 0738
Protective hood
ZU 0737
M12 socket for sensor connection
with Memosens cable/M12 connector
ZU 0860
Stratos Multi E401N
13
Operation with Analog Sensors
To enable operation with analog sensors, analog measuring modules are plugged in and must be
configured during initial start-up.
Operation with Optical Sensors
The SE740 digital optical oxygen sensor can be directly connected to Stratos Multi.
Parameter Sets
Two complete parameter sets (A, B) can be stored in the device. The control element for switching
between the parameter sets (optocoupler input OK1, softkey) is selected in the system control.
The currently activated set can be signaled by a relay contact.
Power Supply
Current is provided through a universal mains supply 80 ... 230 V AC, 45 ... 65 Hz/24 ... 60 V DC.
HART Data Transmission
Device identification, measured values, statuses and messages, calibration data, configurations of the
current loop, and HART variables are transmitted using HART
communication.➜HART (FW-E050), p. 209
Stratos Multi E401N
14
2.3 System Overview
1
2
3
4
5
6
7
8
9
10
Stratos Multi E401N
1 Input for Memosens sensors or SE740 optical oxygen
sensor
6 Current output 1: 0/4 … 20 mA/HART active or
passive (TAN option FW-E050 HART: 4 … 20 mA)
2 Power output 3/15/24 V for SE740 optical oxygen
sensor or external transmitter
7 Current outputs 2/3/4:
Active or passive (current outputs 3 and 4:
TAN option FW-E052)
3 Port for an analog MK module or Memosens
via MK-MS module
8 Relay contact K1: Messages, limits, rinse contact, ...
4 Optocoupler inputs OK1/OK2
OK1: Parameter set selection A/B, flow, …
OK2: Function check (HOLD)
9 Relay contacts 2/3: Controller or messages, limits,
rinse contacts, ...
5 Current input 0/4 … 20 mA for external pressure
transmitter (TAN option FW-E051)
10 Power input:
80 … 230 V AC/24 … 60 V DC < 15 VA/10 W
Stratos Multi E401N
15
2.3.1 Typical Applications
Memosens pH Measurement and PID Control
24 V
AC/DC
MS pH
Stratos Multi E401N
Alarm
Acid
Brine
PID control
Alarm
Acid
Brine
Stratos Multi E401N
16
Memosens Oxygen Measurement and Pressure Correction with External Pressure Transmitter
(with TAN Option FW-E051 “Current Input”)
MS Oxy
e.g. + 24 V
Pressure Transmitter
Stratos Multi E401N
Stratos Multi E401N
17
2.4 Package Contents and Product Identification
• Basic device Stratos Multi
• Bag containing small accessory parts (2x plastic closure, 1x hinge pin, 1x washer, 2x insertable
jumper, 1x reduction sealing insert, 1x multiple sealing insert, 2x blanking plug, 5x cable gland and
hex nut M20x1.5)
• Test Report acc. to EN 10204
• Installation Guide
• Safety Guide
Note: Check all components for damage upon receipt. Do not use damaged parts.
Measuring modules are not included in the basic unit’s package contents.
4
5
6
7
8
9
10
2
3
1
11
12
13
1 Front unit
8 Housing screw (4x)
2 Rear unit
9 Reduction sealing insert (1x)
3 Holes for cable glands
10 Multiple sealing insert (1x)
4 Plastic sealing plug (2x),
for sealing in case of wall mounting
11 Blanking plugs (2x)
5 Hinge pin (1x), insertable from either side
12 Cable gland (5x)
6 Washer (1x), for conduit mounting:
Place washer between housing and nut
13 Hex nut (5x)
7 Insertable jumper (2x)
Stratos Multi E401N
18
2.4.1 Nameplate
1
3
7
6
8
2
4
9
10
11
5
1 Name of manufacturer
6 Protection class II
2 Product name
7 Power supply
3 CE marking
8 Manufacturer’s address with barcode
4 Special conditions: Read the user man-
ual, observe the specifications, and fol-
low the instructions in the safety guide.
9 Permissible ambient temperature, protection class
10 Product number/Serial number/Production year and week
5 Reminder to read the documentation
11 Model designation
You can view the device type, serial number, and firmware, hardware, and bootloader versions of your
device in the Diagnostics menu: Menu Diagnostics Device Information
➜Device Information, p. 130
2.5 Symbols and Markings on the Display
Function check in accordance with NAMUR NE 107
Wrench symbol on orange background
The NAMUR “HOLD” contact is active. Current outputs as configured:
Currently measured value: The currently measured value appears at the current output.
Last measured value: The last measured value is held at the current output.
Fixed value: The current output supplies a fixed value.
Out of specification in accordance with NAMUR NE 107
Black question mark on yellow background
The NAMUR “Out of Specification” contact is active.
Error message: Diagnostics Message List
Failure in accordance with NAMUR NE 107
Flashing black cross symbol on red background
The NAMUR “Failure” contact is active.
Error message: Diagnostics Message List
Maintenance required in accordance with NAMUR NE 107
Oil can symbol on blue background
The NAMUR “Maintenance Required” contact is active.
Error message: Diagnostics Message List
The device is in calibration mode. Function check (HOLD) is active.
The device is in maintenance mode. Function check (HOLD) is active.
The device is in parameter setting mode. Function check (HOLD) is active.
The device is in diagnostics mode.
Stratos Multi E401N
19
Selectable parameter sets (A/B). Indicates which parameter set is currently active when a control element
for parameter set selection was selected:
Parameter Setting System Control Function Control
A “closed” Data Card (memory card) is located in the device. The memory card can be removed.
If you want to continue using the card, select “Open Memory Card” in the Maintenance menu.
There is an enabled Data Card (memory card) in the device.
Note: Select “Close Memory Card” in the Maintenance menu before removing the memory card.
There is an FW Update Card (memory card) in the device. You can save the current device firmware or
perform a firmware update from the memory card
Note: Check the parameter settings after updating.
Free firmware repair in the event of device errors. The TAN option FW-E106 is not required here.
General data cannot be stored on this card.
I
II
Designates the measuring channel for clear assignment of measured-value/parameter displays in the
case of identical process variables.
Channel I: Memosens sensor/SE740 optical oxygen sensor (LDO)
Channel II: Measuring module for analog sensor or second Memosens sensor
IIA
IIB
Channel IIA: First channel in MK-CC module
Channel IIB: Second channel in MK-CC module
CI
CII
Channel CI: Calculation block 1
Channel CII: Calculation block 2
To the left of a menu line that contains a further menu level.
Pressing Enter opens the submenu.
To the left of a menu line that, at administrator level, can be blocked from access at operator level.
To the left of a menu line that, at administrator level, was blocked from access at operator level.
When in measuring mode, Sensoface smileys indicate the quality of the sensor data:
Happy
Neutral
Sad
Wait; device is busy.
To the left of a Diagnostics menu item set as a “Favorite”.
Stratos Multi E401N
20
3 Commissioning
3.1 Assembly
Note: Use only a suitable Phillips head screwdriver to open the housing. Do not use sharp or pointed
objects.
1
2
4
3
5
6
6
Assembling the Housing
01. Install the cable glands (5) from the bag containing small accessory parts in the rear unit.
As required, insert sealing plugs (3) in the event of wall mounting.
➜Package Contents and Product Identification, p. 17
02. Insert blanking plugs in unused cable glands (4).
➜Blanking Plugs, Reduction Sealing Inserts, Multiple Sealing Inserts, p. 26
03. Push in the hinge pin (5) to connect the front unit (1) and rear unit (2).
04. Tighten the captive enclosure screws (6) on the front of the front unit (1) in diagonal sequence
using a Phillips head screwdriver.
05. Select mounting type and install.
✓Pipe mounting ➜Pipe Mounting ZU0274, p. 23
✓Panel mounting ➜Panel-Mount Kit ZU0738, p. 25
✓Protective hood ➜Protective Hood for Wall and Pipe Mounting ZU0737, p. 24
Stratos Multi E401N
21
3.1.1 Dimension Drawings
Note: All dimensions are given in millimeters [inches].
80 [3.15]
34 [1.34]
74 [2.91]
121,4 [4.78]
106,4 [4.19]
90 [3.54]
90 [3.54]
131 [5.16]
20,3 [0.80]
29 [1.14]
36 [1.42]
137,6 [5.42]
42 [1.65]
42 [1.65]
22 [0.87]
148 [5.83]
148 [5.83]
115,5 [4.55]
39,5 [1.56]
8,5 [0.33]
29 [1.14]
3
1
2
CAUTION! Ensure the suitable
IP protection.
IP66/IP67 is only achieved with correctly
installed cable glands.
Front View
Rear View
Bottom View
Side View
4
1 Holes for pipe mounting, 4x
3 Holes for cable glands or
½” conduit, ø 21.5 mm, 2x
2 Holes for wall mounting, 2x
Sealing with plastic sealing plugs
4 Cable glands, 5x
See also
➜Blanking Plugs, Reduction Sealing Inserts, Multiple Sealing Inserts, p. 26
Stratos Multi E401N
22
3.1.2 Housing Mounting Options
Note: All dimensions are given in millimeters [inches].
Knockouts in the rear unit enable different mounting options:
• Pipe mounting ➜Pipe Mounting ZU0274, p. 23
• Panel mounting ➜Panel-Mount Kit ZU0738, p. 25
• Protective hood ➜Protective Hood for Wall and Pipe Mounting ZU0737, p. 24
Cable glands for connecting sensors:
• 3 knockouts for M20x1.5 cable glands
➜Blanking Plugs, Reduction Sealing Inserts, Multiple Sealing Inserts, p. 26
• 2 knockouts for NPT 1/2″ or rigid metallic conduit
Installing Multiple Devices
Note: Install the hinge pin to prevent tensile strain on the measuring cables when replacing the front
unit. Imprecise measuring values may result if this is not done.
110
[4.33071]
There is a 100 mm hinge pin in the bag containing small accessory parts included in the package con-
tents ➜Package Contents and Product Identification, p. 17. The hinge pin connects the front and rear
units. For installation of multiple Stratos Multi devices, a minimum clearance of 110 mm [4.331 inches]
must be maintained to ensure that the middle front units can still be replaced.
Stratos Multi E401N
23
3.1.3 Pipe Mounting ZU0274
Note: All dimensions are given in millimeters [inches].
Ø 40 … 60 mm, □ 30 … 45 mm
40 [1.57] ... 60 [2.36]
1
2
1
2
3
3
4
4
1 For vertical or horizontal posts or pipes
3 Hose clamp with worm gear drive according to
DIN 3017, 2x
2 Pipe-mount plate, 1x
4 Self-tapping screw, 4x
Stratos Multi E401N
24
3.1.4 Protective Hood for Wall and Pipe Mounting ZU0737
Note: All dimensions are given in millimeters [inches].
146,9 [5.78]
189,9 [7.48]
54 [2.13]
74,2
72 [2.83]
The protective hood can only be used for wall or pipe mounting.
The package contents includes 4 M6 nuts for fastening the protective hood on the threaded rod of the
pipe-mount kit.
Stratos Multi E401N
25
3.1.5 Panel-Mount Kit ZU0738
Note: All dimensions are given in millimeters [inches].
Cutout 138 mm x 138 mm (DIN 43700)
22 [0.866]
1 [0.039] ...
138 [5.433]
76 [2.992]
30 [1.181]
<
41 [1.614]
1
2
3
4
5
1 Position of panel
4 Span piece, 4x
2 Circumferential seal, 1x
5 Threaded sleeve, 4x
3 Screw 60.0 x 4.0 mm, 4x
Stratos Multi E401N
26
3.1.6 Blanking Plugs, Reduction Sealing Inserts, Multiple Sealing Inserts
As delivered, each cable gland includes a standard sealing insert. Reduction and multiple sealing in-
serts are available for tight insertion of one or two thinner cables. The coupling can be tightly sealed
using a blanking plug. Handling is as shown below.
Cable gland WITH
standard sealing insert
To use the multiple sealing insert,
remove the standard sealing
insert.
Cable gland WITHOUT
standard sealing instert
24 A/F
24 A/F
Blanking
plug
Reduction sealing
insert
Multiple sealing insert
(orientation as shown)
Blanking plugs and reduction sealing
inserts are inserted into the cable gland
WITH standard sealing instert.
The multiple sealing insert is
inserted into the cable gland
WITHOUT standard sealing insert.
Using blanking plugs and reduction sealing insert
Using a multiple sealing insert
Insert the cable into the cable gland and fasten with a coupling nut.
CAUTION! Ensure the suitable IP protection. IP66/IP67 can only be achieved with correctly in-
stalled cable glands and screwed housing. Observe the cable diameters and tightening torques. Only
use original accessories and spare parts.
Stratos Multi E401N
27
3.2 Electrical Installation
WARNING! The transmitter does not have a power switch. An appropriately arranged and acces-
sible disconnecting device for the transmitter must be present in the system installation. The discon-
necting device must disconnect all non-grounded, current-carrying wires and be labeled such that the
associated transmitter can be identified.
Before commencing with the installation, make sure that all lines to be connected are de-energized.
CAUTION! Ensure the suitable IP protection. IP66/IP67 can only be achieved with correctly
installed cable glands and screwed housing. Observe the cable diameters and tightening torques.
Only use original accessories and spare parts.
Note: Strip the insulation from the wires using a suitable tool to prevent damage. Only use cables that
are rated for use at 75 °C / 167 °F or higher.
01. Wire the current outputs. Deactivate unused current outputs in the parameter settings or use
jumpers.
02. Wire the relay contacts and inputs if necessary.
03. Connect the power supply (for ratings, see specifications).
04. When measuring with analog sensors: Insert the measuring module into the module slot.
05. Connect the sensor.
06. Check whether all connections are correctly wired.
07. Before switching on the power supply, make sure its voltage is within the specified range.
08. Switch on the power supply.
CAUTION! Incorrect parameter settings, calibration, and adjustments can result in incorrect
outputs. For this reason, Stratos Multi must be commissioned by a specialist, who will also set the
parameters, calibrate and adjust the device.
See also
➜Channel II Wiring Examples, p. 168
➜Package Contents and Product Identification, p. 17
3.2.1 Relay Contacts: Protective Wiring
Relay contacts are subject to electrical erosion. Especially with inductive and capacitive loads,
the service life of the contacts will be reduced. For suppression of sparks and arcing, components such
as RC combinations, nonlinear resistors, series resistors, and diodes should be used.
NOTICE! Make sure that the maximum ratings of the relay contacts are not exceeded, even during
switching. ➜Power, p. 155
Information Concerning Relay Contacts
As delivered, the relay contacts are suitable for low signal currents (down to approx. 1 mA). If currents
above approx. 100 mA are switched, the gold plating is destroyed during the switching process.
After that, the contacts will not reliably switch low currents.
Configuration of relay contacts ➜Relay Contacts, p. 57
Wiring of relay contacts ➜Terminal Assignments, p. 30
Stratos Multi E401N
28
Typical AC Application with Inductive Load
1
2
3
1
2
3
1 Load
3 Contact
2 Typical RC combination, e.g., capacitor 0.1 μF, resistor 100 Ω/1 W
Typical DC Application with Inductive Load
1
2
3
1 Inductive load
3 Contact
2 Free-wheeling diode, e.g.,1N4007 (observe polarity)
Typical AC/DC Application with Capacitive Load
1
2
3
1 Capacitive load
3 Contact
2 Resistor e.g., 8 Ω/1 W at 24 V/0.3 A
Stratos Multi E401N
29
3.2.2 Installing Active and Passive Current Outputs
The current outputs directly supply current (0/4 ... 20 mA) to a load according to the selected process
variable.
Passive current outputs require an external supply voltage.
Note: Observe the specifications and connected loads.
Terminal Assignment Diagram
Active current outputs
Passive current outputs
19
Stratos Multi E401N
30
3.2.3 Terminal Assignments
The terminals are suitable for single or stranded wires up to 2.5 mm2.
Terminal
Connection
Sensor (Memosens or
other digital sensors)
1
3 V
BN = brown cable
2
RS485 A
GN = green cable
3
RS485 B
YE = yellow cable
4
GND
WH = white cable
5
Shield
CL = transparent cable
6
Power Out
Power supply output
to supply power to special sensors or external transmitters
Card
Memory card
Current inputs
0/4 mA … 20 mA
7
+ I-Input
8
- I-Input
Digital
control inputs
Optocoupler inputs
9
OK2
10
OK2
11
OK1
12
OK1
Relay contacts
REL 2, REL 3
13
Relay 3
For contact ratings, see ➜Specifications, p. 155
14
Relays 2/3
15
Relay 2
16
N.C.
No connection
Power supply
24 V to 230 V AC/DC
17
Power
Power supply input
18
Power
Power supply input
Current outputs
Out 1/2/3/4
(0)4 mA … 20 mA
Active
Passive
19
N.C., no connection
+ Out 1 for HART
20
+ Out 1 for HART
- Out 1 for HART
21
- Out 1 for HART
N.C., no connection
22
N.C., no connection
+ Out 2/3/4
23
+ Out 2
- Out 2
24
- Out 2
N.C., no connection
25
+ Out 3
- Out 3
26
- Out 3
N.C., no connection
27
+ Out 4
- Out 4
28
- Out 4
N.C., no connection
Relay contact
REL 1
29
Relay 1
For contact ratings, see ➜Specifications, p. 155
30
Relay 1
For connection of analog sensors: Insert measuring module.
See also
➜Power, p. 155
Stratos Multi E401N
31
3.3 Sensor Connection
3.3.1 Connecting a Memosens Sensor/Optical Oxygen Sensor (LDO)
Top view of terminals for Memosens/LDO sensor. The figure shows the opened device, rear side of the
front unit.
1
2
3
1 Standard sensor connection for digital sensors (Memosens sensor/LDO SE740 sensor)
2 Terminal plate with terminal assignments for digital sensor
3 Sensor connection for analog sensors or second Memosens sensor via measuring module
Note: The terminals are suitable for single or stranded wires up to 2.5mm2.
Memosens sensor
SE 740 optical oxygen sensor (LDO)
Terminal
Wire color
Wiring
Memosens cable
Terminal
Wire color
Wiring
M12 cable
1
Brown (BN)
+3V
1
-
2
Green (GN)
RS-485 A
2
Gray
RS-485 A
3
Yellow (YE)
RS-485 B
3
Pink
RS-485 B
4
White (WH)
GND
4
Brown
Shield
5
Transparent (CL) Shield
5
-
-
6
6
White
Power Out
01. Using an appropriate sensor cable, connect a Memosens sensor or the SE740 optical oxygen
sensor (LDO) to the RS-485 interface (2) of the Stratos Multi.
02. Then select a measuring function and configure the sensor:
From within measuring mode, press the left softkey: Menu.
✓The Menu selection opens.
03. Select Parameter Setting Sensor Selection [I] [II] .
Note: Function check (HOLD) is active.
04. Press enter to open the Sensor Selection [I] .
05. Select process variable, mode, and functionality, and confirm with enter.
Set further parameters with the left softkey: Back.
06. Return to measuring mode to exit parameter setting,
e.g., with the right softkey: Back to Meas.
Stratos Multi E401N
32
3.3.2 Connecting an Analog Sensor/Second Memosens Channel
CAUTION! Electrostatic discharge (ESD). The modules' signal inputs are sensitive to electrostatic
discharge. Take measures to protect against ESD before inserting the module and wiring the inputs.
Note: Strip the insulation from the wires using a suitable tool to prevent damage. Only use cables that
are rated for use at 75 °C / 167 °F or higher.
Measuring Modules for Connection of Analog Sensors: pH, ORP, Oxygen, Conductivity
➜Terminal Assignments of Measuring Modules, p. 34
1
2
3
4
01. Switch off the power supply to the device.
02. Open the device (loosen the 4 screws on the front).
03. Loosen screw (1) on the module cover (2) (“ESD shield”) and open the cover.
04. Insert the module into the module slot (3).
05. Attach the module plate sticker (4).
06. Connect the sensor and separate temperature probe, if necessary; see Wiring Examples.
NOTICE! Moisture ingress. The cable glands must be tightly sealed. Insert blanking plugs or sealing
inserts if necessary.
07. Check whether all connections are correctly wired.
08. Close the module cover (2), tighten screw (1).
09. Close the device by tightening the screws on the front.
10. Switch on the power supply.
11. Then select a measuring function and configure the sensor:
From within measuring mode, press the left softkey: Menu.
✓The Menu selection opens.
12. Select Parameter Setting Sensor Selection [I] [II] .
Sensor Selection
Sensor Selection
Back
Sensor Selection
Lock
Power Out
Off
I
II (Admin.)
I
II
Note: Function check (HOLD) is active.
13. Press enter to open the Sensor Selection [II] .
14. Select the module and mode, and confirm with enter.
Set further parameters with the left softkey: Back.
15. Return to measuring mode to exit parameter setting,
e.g., with the right softkey: Back to Meas.
Stratos Multi E401N
33
Measuring Module for Connection of a Second Memosens Sensor
If you want to measure two process variables using Memosens sensors, you must insert an
MK-MS095N Memosens module for the second channel.
01. Insert a Memosens module in the module slot and connect it up (see above).
02. Then select a measuring function and configure the sensor:
From within measuring mode, press the left softkey: Menu.
✓The Menu selection opens.
03. Select Parameter Setting Sensor Selection [I] [II] .
Note: Function check (HOLD) is active.
04. Press enter to open the Sensor Selection [II] .
05. Select module MK-MS.
06. Select process variable, mode, and functionality, and confirm with enter.
Set further parameters with the left softkey: Back.
07. Return to measuring mode to exit parameter setting,
e.g., with the right softkey: Back to Meas.
3.3.3 Cable Preparation for SE655/SE656
Preparing the Shield Connection for Toroidal Conductivity Sensors
3
5
4
2
1
01. Insert the special cable through the cable entry into the terminal compartment.
02. Remove the previously separated part of the cable insulation (1).
03. Turn the shielding mesh (2) over the cable insulation (1) (3).
04. Then guide the crimp ring (4) over the shielding mesh (5) and tighten it using pliers.
Special Cable Wire Assignment
01. Connect the sensors to the measuring device.
RTD (temp)
Coax red
send lo
Coax white
send hi
receive hi
blue
receive lo
sense (temp)
blue
red
green
red
white
yellow
green
RTD (temp)
shield
Stratos Multi E401N
34
3.4 Terminal Assignments of Measuring Modules
Installation of the measuring modules
➜Connecting an Analog Sensor/Second Memosens Channel, p. 32
The module plate (1) is attached to the measuring module underneath the module cover on the rear
of the device.
1
pH/ORP Measuring Module
Order code MK-PH015N
Oxygen Measuring Module
Order code MK-OXY046N
Module for Contacting Conductivity Measurement
Order code MK-COND025N
A
B
C
D
E
F
G
H
I
K
COND Sensor
Temp
RTD (GND)
RTD (SENSE)
RTD
SHIELD
Module for Inductive Conductivity Measurement
Order code MK-CONDI035N
Stratos Multi E401N
35
Module for Dual Conductivity Measurement
Order code MK-CC065N
Memosens Module
Order code MK-MS095N
3.5 Final Check During Commissioning
Only commission the Stratos Multi if all the questions below can be answered “yes”:
• Are the Stratos Multi and all its cables externally intact and strain-relieved?
• Are the cables routed without any loops or crossovers?
• Have all the wires been correctly connected in accordance with their terminal assignments?
• Was the tightening torque of the screw terminals correctly adhered to?
• Are all connectors firmly engaged?
• Are all cable glands installed, tight, and leak-proof?
• Is the device closed and correctly screwed together?
• Does the supply voltage (power supply) accord with the voltage indicated on the nameplate?
Stratos Multi E401N
36
4 Operation and Use
4.1 Changing the User Interface Language
Preconditions
• Stratos Multi is connected to the power. Measuring mode is shown on the display.
Steps
01. Press the Left softkey: Menu. The menu selection opens.
02. Press the Right softkey: Lingua. Press the right arrow key and set the language of the user inter-
face.
03. Confirm with enter.
Note: The user interface language can also be changed in the Parameter Setting menu.
Parameter Setting General Language ➜Parameter Setting, General, p. 48
4.2 Keypad and Display
Display
Stratos Multi features a 4.3'' TFT color graphic display. The Calibration, Maintenance, Parameter Set-
ting, and Diagnostics menus each have their own colors. The device is operated using plain text in var-
ious languages. Messages are output as icons and plain text.
Menu Selection
Back
Maintenance
Lingua
1
2
3
4
5
6
1 Calibration
4 Diagnostics
2 Maintenance (selected)
5 Softkey function: Selecting the language
3 Parameter Setting
6 Softkey function display
Stratos Multi E401N
37
Sensor Monitoring Details (Admin.)
Zero Point
Back
2
7
8
9
Slope
ORP Offset
Sensocheck Ref El
Sensocheck Glass El
Response Time
1
3
4
Back to Meas.
I
6
5
10
11
1 Function check (HOLD)
7 Softkey function display
2 Current menu heading
8 Softkey function display
3 Device condition (PAR for parameter setting)
9 Folder element
4 Failure is active
10 Selected element
5 Active parameter set
11 Measuring channel, e.g. channel I
6 Scrollbar
Display of Measuring Channels
I
II
IIA
CI
CII
IIB
1
2
3
4
5
6
1 Channel I
3 First channel in MK-CC module
5
Calculation block 1
2 Channel II
4 Second channel in MK-CC module
6
Calculation block 2
Stratos Multi E401N
38
Keypad
1
2
3
4
5
1 Left softkey:
Function as shown on left function display
4 enter:
Open menu, confirm inputs
2 Left/right arrow keys:
Menu selection: prior/next menu, move left/
right
5 Right softkey:
Function as shown on right function display
3 Up/down arrow keys:
Row selection from selection window,
increase/decrease a numeral
Entering Text and Numbers; Selecting Signs
01. Select a number using the left/right arrow keys.
02. Enter numbers or letters using the up/down arrow keys.
Change the sign, required.
03. Switch to signs using the left arrow key.
04. Set the sign value with the up or down arrow keys.
05. Confirm with enter.
Note: If you enter values outside the specified value range, an information window showing the per-
missible value range is shown.
Stratos Multi E401N
39
4.3 Menu Structure Overview
Measuring
2958
1989 Administrator Level
1246 Operator Level
Passcode:
(factory
setting)
Calibration
Maintenance
Parameter Setting Diagnostics
Menu
1147
Message list, log,
device and sensor
information
System control (time, firmware update, options,
passcode entry), sensor selection, general
(language, measurement display), inputs and
outputs, measuring channels I and II
Maintenance of
sensors and current
outputs
Calibration mode,
calibration points,
buffer set
1
2
3
4
5
6
Submenus:
1 Left softkey: Pressing Menu accesses the menu
selection.
4 Press enter to confirm, enter passcode.
2 Right softkey: Back to Meas. returns to measure-
ment.
5 Further submenus and menu items are shown.
3 Select a menu using the arrow keys.
6 Selected functions in the Diagnostics menu can
also be accessed in measuring mode by pressing
the right softkey (Favorites menu).
4.4 Access Control
Access to the device functions is regulated and limited by individually adjustable passcodes. This pre-
vents unauthorized modification of device settings or manipulation of the measurement results.
Passcodes are configured in Parameter Setting System Control ➜System Control, p. 43
Stratos Multi E401N
40
4.5 Operating States
While the parameters are being set, Stratos Multi is in the function check (HOLD) mode. The current
output and relay contacts behave in accordance with the parameter settings. In the function check
(HOLD) mode, measurement is not permitted since unexpected system behavior might put users at
risk.
Operating Mode
Current
Outputs
Contacts
Controller
(PID Controller)
Timeout1)
Measuring
-
Diagnostics
-
Calibration2)
-
Maintenance2)
Sensor Monitor
-
Current Source
-
Manual Controller
-
Parameter Setting2)
20 min
Rinse Function2)
3)
At end of rinse time
Active (output functions normally)
Manual control of the outputs
Last value or fixed default value
Depending on parameter setting
4.6 Measurement Display
The following settings are possible:
2, 4, 6, or 8 values
without measuring channel selection
Any display of measured values from the measuring channels
and the device possible
2 or 4 values with measuring channel selection
Any display of measured values from the measuring channels
Settings can be changed in the Measurement Display submenu:
Parameter Setting General Measurement Display
An overview of display options can be found in the Parameter Setting chapter.
➜Parameter Setting, General, p. 48
The right softkey: Back to Meas. returns you to measurements from any menu level. You may have to
confirm that the system is ready for measurement.
If required, the display can be configured to switch off after not having been used for a user-defined
period of time.
This setting can be changed in the Display submenu:
Parameter Setting General Display
Display auto-off can be configured as follows:
• No auto-off
• After 5 minutes
• After 30 minutes
1)
“Timeout” means that the device will return to measuring mode after 20 minutes without key activity.
2)
Function check (HOLD) is active.
3)
Rinse contact is active.
Stratos Multi E401N
41
5 Parameter Setting
CAUTION! Incorrect parameter settings, calibration, and adjustments can result in incorrect
outputs. For this reason, Stratos Multi must be commissioned by a specialist, who will also set the
parameters, calibrate and adjust the device.
Opening Parameter Setting
Menu Menu Selection Parameter Setting
Menu Selection
Back
Parameter Setting
Lingua
01. From within measuring mode, press the left softkey: Menu.
✓The Menu Selection opens.
02. Using the right arrow key, select the Parameter Setting menu and confirm with enter.
✓The Parameter Setting menu contains items for things such as inputs and outputs, sensor selec-
tion I and II, system control, and general parameter setting. Parameter setting is automatically
ended 20 minutes after the last registered keystroke, after which Stratos Multi returns to mea-
suring mode (timeout).
Note: Function check (HOLD) is active. The current output and relay contacts behave in accordance
with the parameter settings. Return to measuring mode to exit the function check, e.g., with the right
softkey: Back to Meas.
5.1 Operating Levels
There are three access levels in the Parameter Setting menu:
• Viewing level (all data)
• Operator level (operation data)
• Administrator level (all data)
Viewing Level
• Display of all settings
Settings cannot be changed in the Viewing level.
Operator Level
• Access to all functions that have been enabled at the Administrator level.
Locked functions are displayed in gray and cannot be edited.
Administrator Level
• Access to all settings, including passcode settings.➜Passcode Entry, p. 48
• Enable or lock functions for access from within Operator level. ➜Locking a Function, p. 42
Functions that can be locked for the Operator level are marked with the “lock” icon. The functions are
enabled or locked using the right softkey.
Stratos Multi E401N
42
5.2 Locking a Function
Example: Locking access to the configuration of relay contact K1 from the Operator level
01. Open Parameter Setting.
02. Select Administrator level.
03. Enter passcode (factory setting: 1989).
04. Select submenu:
Inputs and Outputs Relay Contacts Contact K1
Relay Contacts (Admin.)
Contact K1
Back
Controller
Lock
Contact K2
Contact K3
Off
05. Right softkey: Lock
✓The Contact 1 submenu is now marked with the “lock” icon. This function can no longer be
accessed from the Operator level.
The softkey function automatically changes to Unlock.
✓At the Operator level, the locked function is shown in gray.
Relay Contacts (Operator)
Back
Controller
Back to Meas.
Contact K2
Contact K3
Off
Contact K1
Stratos Multi E401N
43
5.3 System Control
Submenus
Description
Memory Card
This option is shown if a Data Card is inserted:
Settings for logbook and measurement recorder data recording. Decimal separation
can be set. The memory card can be formatted.
Transfer Configuration
If a Data Card is inserted, the measuring device’s configuration can be saved and
transferred to another measuring device. ➜Transfer Configuration, p. 44
Parameter Sets
Two parameter sets (A, B) are available in the device. If a Data Card is inserted, up to
five parameter sets can be saved on or loaded from the Data Card.
➜Parameter Sets, p. 45
Function Control
Allocation of functions for activation by softkey or optocoupler input OK1.
➜Function Control, p. 46
Calculation Blocks
TAN option FW-E020: Convert available process variables to new
variables.➜Calculation Blocks (TAN Option FW-E020), p. 46.
Time/Date
Define date and time format; input of date, time, and weekday. ➜Time/Date, p. 46
Meas. Point Description
Free input of a tag number and notes; can be retrieved in the Diagnostics menu.
➜Measuring Point Description, p. 46
Firmware Update
This option is shown if a FW Update Card is inserted. TAN option FW-E106: Firmware
update with FW Update Card. ➜Firmware Update (TAN Option FW-E106), p. 46
Enable Options
Activation of add-on options via TAN. The TAN is only valid for the Stratos Multi with
the associated serial number. ➜Option Activation, p. 47
Logbook
Select events to be logged (failure/maintenance required); can be retrieved in the Di-
agnostics menu. ➜Logbook, p. 47
Buffer Table
TAN option FW-E002: Specification of a buffer set.
➜pH Buffer Table: Entry of Individual Buffer Set (FW-E002), p. 199
Concentration Table
TAN option FW-E009: Specification of a concentration solution for conductivity mea-
surement. ➜Concentration Determination (FW-E009), p. 201.
Restore Factory Settings
Restore all parameters to factory settings. ➜Restore Factory Settings, p. 48
Passcode Entry
Change passcodes. ➜Passcode Entry, p. 48
Stratos Multi E401N
44
5.3.1 Transfer Configuration
All device settings can be saved on a memory card (Data Card):
Parameter Setting System Control Transfer Configuration
Note: The inserted Data Card is shown on the display.
• Select Save Configuration to write all the device settings (except passcodes) to the Data Card.
Backup file generated on the Data Card:
param/config.par
• Select Load Configuration to read all the device settings from the Data Card and apply them to the
device.
Transferring all Device Settings from One Device to Other Devices
Preconditions
• The devices all feature identical hardware.
• TAN options (add-on functions):
All required TAN options must be enabled before they can be transferred.
Steps
01. Parameter Setting System Control Transfer Configuration
02. Menu item Configuration: Save
03. Start the transfer with the right softkey: Execute.
✓The device settings are saved to the Data Card.
04. Switch to the Maintenance Open/Close Memory Card menu.
05. Terminate access to the memory card using the right softkey: Close.
06. Remove the Data Card.
✓You can transfer the device settings to other, identically equipped devices.
07. Insert the Data Card containing the device settings in the next device to be configured.
08. Parameter Setting System Control Transfer Configuration
09. Menu item Configuration: Load
10. Start the transfer with the right softkey: Execute.
✓The device settings are read from the Data Card and applied.
11. Switch to the Maintenance Open/Close Memory Card menu.
12. Terminate access to the memory card using the right softkey: Close.
13. Remove the Data Card.
See also
➜Memory Card, p. 153
Stratos Multi E401N
45
5.3.2 Parameter Sets
Stratos Multi provides two complete selectable parameter sets (A/B) for different process connections
or process conditions. The currently activated parameter set can be signaled by a relay contact.
➜Relay Contacts, p. 57.
Parameter set “B“ only permits setting of process-related parameters.
Parameter Setting System Control Parameter Sets
Save Parameter Set
The active parameter set is transferred to the Data Card.
Note: The parameter set saved on the Data Card is overwritten.
Load Parameter Set
A parameter set stored on the Data Card is transferred to the device.
Note: This overwrites the current parameter set in the device.
Up to five parameter sets can be stored on the Data Card with TAN option FW-E102.
➜Parameter Sets 1-5 (FW-E102), p. 210
Selecting Parameter Sets A/B
The control element for switching between parameter sets (optocoupler input OK1 or softkey) is se-
lected in:
Parameter Setting System Control Function Control
The currently active parameter set is indicated by the
or
icons.
Selection via a signal at optocoupler input OK1:
Power supply (PWR out)
12 ... 24 V AC/DC
12
11
Optocoupler input
Stratos Multi
0...2 V AC/DC: Parameter set A active
10…30 V AC/DC: Parameter set B active
Note: The selection has no effect when using parameter sets from a memory card. Switching between
parameter sets A and B is possible if they are saved in the device.
Stratos Multi E401N
46
5.3.3 Function Control
The following functions can be activated by softkey or optocoupler input OK1:
Input OK1:
• Parameter set selection
• Flow
• Function check
• Function check (channel)
Right softkey:
• Off
• Value rotation
• Parameter set selection
• Favorites menu
The selection can be changed in the Function Control submenu:
Parameter Setting System Control Function Control
5.3.4 Calculation Blocks (TAN Option FW-E020)
Calculation blocks convert existing process variables to new variables.
The menu is only shown if the TAN option is activated.
Parameter Setting System Control Calculation Blocks
See also
➜Calculation Blocks (FW-E020), p. 206
5.3.5 Time/Date
The time and date in the installed real-time clock are required for:
• Controlling calibration and cleaning cycles
• Displaying the time on the display
• Assigning times to the calibration data in the sensor head on digital sensors
• Diagnostic functions; logbook entries are given a time stamp, for example
Note: No automatic switchover from winter to summer time.
Settings can be changed in the Time/Date submenu:
Parameter Setting System Control Time/Date
5.3.6 Measuring Point Description
You can enter a tag number and notes (e.g. date of last maintenance).
Parameter Setting System Control Meas. Point Description
Display of the tag number in the Diagnostics menu ➜Measuring Point Description, p. 131
5.3.7 Firmware Update (TAN Option FW-E106)
The firmware update is carried out using TAN option FW-E106 and a FW Update Card.
➜Firmware Update (FW-E106), p. 213
The menu is only shown if the TAN option is activated and the FW Update Card has been inserted.
Parameter Setting System Control Firmware Update
Stratos Multi E401N
47
5.3.8 Option Activation
Add-on functions (TAN options) expand the device’s capabilities. The TAN options are device-specific.
When ordering a TAN option, you must therefore specify the serial number of the device in addition to
the relevant order code for this function. The manufacturer then supplies a TAN (transaction number)
to activate the add-on function. This TAN is only valid for the device with the stated serial number.
You can find your device’s serial number in:
Diagnostics Device Information
Overview of TAN options ➜Product Range and Options, p. 12
Activate TAN Option
01. Parameter Setting System Control Option Activation
02. Select the option to be enabled.
03. Set to “Active” using the arrow keys.
✓Enter the TAN at the prompt. The current serial number is shown.
04. Enter the TAN and confirm with OK.
✓The option is available.
Note: An activated TAN option can be deactivated and reactivated without having to re-enter the TAN.
5.3.9 Logbook
The logbook records the last 100 events with date and time and displays them on the device.
In addition, when using the Data Card and TAN option FW-E104, 20,000 entries or more can be stored
on the Data Card.
Parameter Setting System Control Logbook
• Select whether to log failure and/or maintenance required messages in the logbook.
• Delete the logbook entries
Displaying the Logbook Entries
The entries can be viewed in the Diagnostics menu ➜Logbook, p. 129
Menu Diagnostics Logbook
5.3.10 Measurement Recorder (TAN Option FW-E103)
With TAN option FW-E103: Delete the data stored on the measurement recorder
The menu is only shown if the TAN option is activated.
Parameter Setting System Control Measurement Recorder
See also
➜Measurement Recorder (FW-E103), p. 211
5.3.11 Buffer Table (TAN Option FW-E002)
The menu is only shown if the TAN option is activated.
Parameter Setting System Control Buffer Table
See also
➜pH Buffer Table: Entry of Individual Buffer Set (FW-E002), p. 199
Stratos Multi E401N
48
5.3.12 Concentration Table (TAN Option FW-E009)
The menu is only shown if the TAN option is activated.
Parameter Setting System Control Concentration Table
See also
➜Concentration Determination (FW-E009), p. 201
5.3.13 Restore Factory Settings
Allows the parameters to be reset to their factory settings:
Parameter Setting System Control Restore Factory Settings
NOTICE! All individual parameter settings will be overwritten with the factory settings.
5.3.14 Passcode Entry
Passcodes (factory setting)
Calibration
1147
Maintenance
2958
Operator Level
1246
Administrator Level
1989
The passcodes can be changed or deactivated in the Passcode Entry submenu:
Parameter Setting System Control Passcode Entry
Note: The passcode for the Administrator level cannot be deactivated.
Note: If you lose the administrator passcode, system access is locked! The manufacturer can generate
a rescue TAN. If you have any questions, please contact Knick Elektronische Messgeräte GmbH & Co.
KG using the information provided on the last page of this document.
5.4 Parameter Setting, General
Note: Function check (HOLD) is active.
Submenus
Description
Language
User interface language: German (factory setting), English, French, Italian, Spanish,
Portuguese, Chinese
Units/Formats
Temperature unit °C (factory setting) or °F.
Other units and formats depending on the selected process variable,
e.g., pressure in mbar, kPa, psi
Display format pH xx.xx or xx.x
Measurement Display
Values to be displayed (up to 8) ➜Configuring the Measurement Display, p. 49
Display
Display color, brightness, and display auto-off (factory setting: None) ➜Display, p. 54
Measurement Recorder
TAN option FW-E103: Logging of measured and additional values
➜Measurement Recorder (FW-E103), p. 211
Stratos Multi E401N
49
5.4.1 Configuring the Measurement Display
Parameter Setting General Measurement Display
01. Set the number of values to be displayed:
2 values (1 channel), 2 values (2 channels), 4 values (2 channels),
2 values, 4 values, 6 values, 8 values
02. As required, assign channels and variables to be displayed.
03. Confirm with enter.
Measurement Display, Example with 2 Values
Selection
Result
Selection of any two variables:
2 Values
I
I
Number
1st Value
2nd Value
pH Value
Temperature
pH Value
Temperature
2 Values (1 Channel)
2 Values (2 Channels)
4 Values (2 Channels)
4 Values
2 Values
Select number of values.
Confirm with enter.
2 Values
pH Value
Temperature
I
I
Number
1st Value
2nd Value
I
I
I
I
I pH Value
ORP
Temperature
rH Value
pH Voltage
Select first variable.
Confirm with enter.
2 Values
pH Value
Time
I
Number
1st Value
2nd Value
Current Output 1
Dig. Controller Output
Current Input
Date
Time
Select second variable.
Confirm with enter.
Set further parameters with the left softkey: Back.
End parameter setting with the right softkey: Back to
Meas..
I
Favorites Menu
Menu
pH
7.08
14:03
1
2
(1) First value
(2) Second value
Stratos Multi E401N
50
Measurement Display, Example with 2 Values (1 Channel)
Selection
Result
Select two variables in one measuring channel:
2 Values
pH Value
I
Number
Channel 1
1st Meas. Value
2nd Meas. Value
2 Values (1 Channel)
2 Values (2 Channels)
4 Values (2 Channels)
4 Values
2 Values
Select number of values and channels.
Confirm with enter.
2 Values (1 Channel)
Memosens pH/ORP
I
Number
Channel 1
1st Meas. Value
2nd Meas. Value
I Memosens pH/ORP
Analog Cond
II
pH Voltage
Assign a sensor to the channel.
Confirm with enter.
2 Values (1 Channel)
Memosens pH/ORP
I
I
I
Number
Channel 1
1st Meas. Value
2nd Meas. Value
I
I
I
I
I pH Value
Redox Voltage
Temperature
rH Value
pH Voltage
Select the first variable for channel I.
Confirm with enter.
2 Values (1 Channel)
Memosens pH/ORP
I
I
I
Number
Channel 1
1st Meas. Value
2nd Meas. Value
I
I
I
I Temperature
pH Voltage
rH Value
Ref. Impedance
Glass Impedance
I
Select the second variable for channel I.
Confirm with enter.
Set further parameters with the left softkey: Back.
End parameter setting with the right softkey: Back to
Meas..
I
Favorites Menu
Menu
pH
9.89
178 mV
1
2
(1) First value in channel I
(2) Second value in channel I
Stratos Multi E401N
51
Measurement Display, Example with 2 Values (2 Channels)
Selection
Result
Select two variables in two measuring channels:
2 Values
pH Value
I
Number
Channel 1
1st Meas. Value
2nd Meas. Value
2 Values (1 Channel)
2 Values (2 Channels)
4 Values (2 Channels)
4 Values
2 Values
Select number of values and channels.
Confirm with enter.
2 Values (2 Channels)
Memosens pH/ORP
Analog Cond
I
I
Number
Channel 1
1st Meas. Value
Channel 2
1st Meas. Value
pH Value
II
I Memosens pH/ORP
Analog Cond
II
Conductivity
II
Assign a sensor to the first channel.
Confirm with enter.
2 Values (2 Channels)
Memosens pH/ORP
Analog Cond
I
I
Number
Channel 1
1st Meas. Value
Undefined
Channel 2
1st Meas. Value
pH Value
I
I
I
I
I pH Value
ORP
Temperature
rH Value
pH Voltage
II
Select the variable for the first channel.
Confirm with enter.
2 Values (2 Channels)
Memosens pH/ORP
I
I
Number
Channel 1
1st Meas. Value
Channel 2
1st Meas. Value
pH Value
II Analog Cond
Analog Cond
Undefined
I Memosens pH/ORP
Analog Cond
II
Assign a sensor to the second channel.
Confirm with enter.
2 Values (2 Channels)
Memosens pH/ORP
Analog Cond
I
I
Number
Channel 1
1st Meas. Value
Channel 2
1st Meas. Value
II
pH Value
Conductivity
Temperature
Salinity
Resistivity
Conductance
II
II
II
II
II
Select the variable for the second channel.
Confirm with enter.
Set further parameters with the left softkey: Back.
End parameter setting with the right softkey: Back to
Meas..
I
Favorites Menu
Menu
pH
7.08
984
II
µS
cm
1
2
(1) First value in channel I
(2) Second value in channel II
Stratos Multi E401N
52
Measurement Display, Example with 4 (6, 8) Values
Selection
Result
Select any four (six, eight) variables
4 Values
Number
1st Value
2nd Value
3rd Value
4th Value
I
I
pH Value
Temperature
Conductivity
II
II
Resistivity
4 Values (2 Channels)
2 Values
4 Values
8 Values
6 Values
Select number of values.
Confirm with enter.
4 Values
Number
1st Value
2nd Value
3rd Value
4th Value
I
I
pH Value
Temperature
Conductivity
II
II Resistivity
I
I
I
I
I pH Value
ORP
Temperature
rH Value
pH Voltage
Select first variable.
Confirm with enter.
4 Values
Number
1st Value
2nd Value
3rd Value
4th Value
I
I
pH Value
pH Voltage
Conductivity
II
II Resistivity
pH Voltage
I
II
II
II
II
Conductivity
Temperature
Salinity
Resistivity
Select second variable.
Confirm with enter.
4 Values
Number
1st Value
2nd Value
3rd Value
4th Value
I
I
pH Value
Temperature
Conductivity
II
II Resistivity
pH Value
Conductivity
Temperature
Salinity
Resistivity
Conductance
II
II
II
II
II
Select third variable.
Confirm with enter.
4 Values
Number
1st Value
2nd Value
3rd Value
4th Value
I
I
pH Value
Temperature
Conductivity
II
II Resistivity
Salinity
Resistivity
II
II
II Conductance
Time
Date
Select fourth variable.
Confirm with enter.
Set further parameters with the left softkey: Back.
End parameter setting with the right softkey: Back to
Meas..
I
Favorites Menu
Menu
pH
I
6.40
II
II
178 mV
1,135 mS/cm
0.00 MΩcm
3
4
1
2
(1) First value
(2) Second value
(3) Third value
(4) Fourth value
Stratos Multi E401N
53
Measurement Display, Example with 4 Values (2 Channels)
Selection
Result
Select four variables in two measuring channels:
2 Values
pH Value
I
Number
Channel 1
1st Meas. Value
2nd Meas. Value
Channel 2
II Analog Cond
2 Values (1 Channel)
2 Values (2 Channels)
4 Values (2 Channels)
4 Values
2 Values
Select number of values and channels.
Confirm with enter.
4 Values (2 Channels)
Memosens pH/ORP
Analog Cond
I
Number
Channel 1
1st Meas. Value
2nd Meas. Value
Channel 2
II
I pH Value
I pH Voltage
I Memosens pH/ORP
Analog Cond
II
Assign a sensor to the first channel.
Confirm with enter.
4 Values (2 Channels)
Memosens pH/ORP
Analog Cond
I
I
Number
Channel 1
1st Meas. Value
2nd Meas. Value
Channel 2
1st Meas. Value
I
pH Value
I
I
I
I
I pH Value
ORP
Temperature
rH Value
pH Voltage
II
II
Select the first variable for the first channel.
Confirm with enter.
4 Values (2 Channels)
Memosens pH/ORP
Analog Cond
I
Number
Channel 1
1st Meas. Value
2nd Meas. Value
Channel 2
II
I
Undefined
I
pH Value
I
I
I
I Temperature
pH Voltage
rH Value
Ref. Impedance
Glass Impedance
I
Select the second variable for the first channel.
Confirm with enter.
4 Values (2 Channels)
Memosens pH/ORP
Analog Cond
I
I
Number
Channel 1
1st Meas. Value
Undefined
2nd Meas. Value
Channel 2
1st Meas. Value
I
pH Value
I Memosens pH/ORP
Analog Cond
II
II
pH Voltage
Assign a sensor to the second channel.
Confirm with enter.
Memosens pH/ORP
Analog Cond
I
I
Channel 1
1st Meas. Value
2nd Meas. Value
Channel 2
1st Meas. Value
I
pH Value
II
Conductivity
II
pH Voltage
2nd Meas. Value
II Resistivity
pH Value
Conductivity
Temperature
Salinity
Resistivity
Conductance
II
II
II
II
II
Select the first variable for the second channel.
Confirm with enter.
Stratos Multi E401N
54
Selection
Result
Memosens pH/ORP
Analog Cond
I
Channel 1
1st Meas. Value
2nd Meas. Value
Channel 2
1st Meas. Value
2nd Meas. Value
II
I pH Value
I pH Voltage
II Conductivity
II Resistivity
I
Conductivity
I
pH Value
pH Voltage
pH Value
Conductivity
Temperature
Salinity
Resistivity
Conductance
II
II
II
II
II
Select the second variable for the second channel.
Confirm with enter.
Set further parameters with the left softkey: Back.
End parameter setting with the right softkey: Back to
Meas..
I
Favorites Menu
Menu
pH
4.00
II
178 mV
1.135 mS/cm
0.00 MΩcm
1
2
3
4
(1) First value in channel I
(2) Second value in channel I
(3) First value in channel II
(4) Second value in channel II
5.4.2 Display
It is possible to change the display’s color and brightness.
The following settings are possible
Display Color
White,
NE107 (factory setting): If a NAMUR message is available for a measured value, the measured
value is backlit in accordance with the NAMUR color.
Brightness
Factory setting: 80 %
Auto-off
None (factory setting), after 5 min, after 30 min
Settings can be changed in the Display submenu:
Parameter Setting General Display
Note on Display Auto-off
The display switches off 5 or 30 minutes after the last keystroke is registered. Press any key to switch
the display back on.
5.4.3 Measurement Recorder (TAN Option FW-E103)
The measurement recorder logs measured values and additional values depending on its parameter
setting. The last 100 entries are graphically presented on the display of the Stratos Multi.
The menu is only shown if the TAN option is activated.
Parameter Setting General Measurement Recorder
See also
➜Measurement Recorder (FW-E103), p. 211
5.5 Inputs and Outputs
The following inputs and outputs are available:
• Four current outputs 0/4 ... 20 mA for transferring variables such as measured value or temperature
(factory setting); two of these outputs can be enabled via TAN; active or passive configuration
➜Current Outputs, p. 55
• Three freely configurable floating switching outputs, ➜Relay Contacts, p. 57
of which two can be used to control a PID controller ➜PID Controller, p. 61
• Two digital control inputs, OK1 and OK2 ➜Control Inputs, p. 62
Stratos Multi E401N
55
5.5.1 Current Outputs
The current outputs are deactivated ex works. Current outputs 3 and 4 need to be activated by TAN
(TAN option FW-E052).
Adjustable parameters for the current outputs
Usage
On, Off
Process Variable
Selection from all available process variables
Current Range
4 … 20 mA or 0 … 20 mA
Characteristic
Linear
Trilinear (input of additional vertex points required)
Function (input of a 50 % point required)
Logarithmic
Table (with TAN option FW-E006 “current characteristic”)
➜Current Characteristic (FW-E006), p. 201
Output
Output current range 4 … 20 mA or 0 … 20 mA
Start 0(4) mA
Start of span
End 20 mA
End of span
Output Filter
To smooth the current output, a low-pass filter with adjustable filter time constant can
be switched on. The filter acts on the current output only.
Function Check
Current output response in the Function Check mode: Output of the current measure-
ment, the last measured value, or a fixed value 0 … 22 mA.
Behavior during Messages
Failure
Current output response in the event of a failure message:
Off, 3.6 mA, 22 mA
Delay
Input of a delay of 0 … 600 s in the event of a failure message.
Settings can be changed in the Current Outputs submenu:
Parameter Setting Inputs and Outputs Current Outputs
Setting the Span: Start (0/4 mA) and End (20 mA)
0
5
7
14
[pH]
[mA]
4
20
5
7
[pH]
[mA]
4
20
1
2
1 Span pH 0 … 14
2 Span pH 5 … 7
Higher resolution in range of interest
Stratos Multi E401N
56
Characteristic Curves
Linear Characteristic
The process variable is represented by a linear output current curve.
Output 4 … 20 mA, span pH 0 … 14
Output 4 … 20 mA, span pH 5 … 9
Start
End
pH
0
20
mA
4
8
12
16
4
6
8
2
10
12
14
Start
End
pH
5
20
mA
4
8
12
16
6
7
8
9
Trilinear/Bilinear Curve
Requires the input of two additional vertex points.
Trilinear: Vertex points (1) and (2) are different values.
Bilinear: Vertex points (1) and (2) are the same values.
Output 4 … 20 mA, span pH 0 … 14
Output 4 … 20 mA, span pH 5 … 9
Start
End
pH
0
20
mA
4
8
12
16
4
6
8
2
10
12
14
1
2
Start
End
pH
20
mA
4
8
12
16
6
8
2
10
12
14
4
2
1
Function/Logarithmic Curve
Nonlinear output current characteristic: allows measurements over several decades, e.g. measuring
very low values with a high resolution and high values with a low resolution. Requires the input of the
value for 50 % output current.
Characteristic: Function with input of 50 % value (1)
Characteristic: Logarithmic
Output 4 … 20 mA, span 1 … 100 µS/cm
Output 4 … 20 mA, span 1 … 1000 µS/cm
Start
End
μS/cm
0
20.0
mA
4.0
8.0
12.0
16.0
20 30 40
10
50 60 70 80 90 100
1
Start
End
μS/cm
0
20.0
mA
4.0
8.0
12.0
16.0
200 300 400
100
500 600 700 800 900 1000
Stratos Multi E401N
57
Output Filter Time Interval
To smooth the current output, a low-pass filter with adjustable time interval can be switched on.
When there is a jump at the input (100 %), the output level is at 63 % after the time interval has been
reached. The time interval can be set from 0 to 120 sec. If the time interval is set to 0 sec, the current
output follows the input.
Note: The filter only acts on the current output, not on the display, the limit values or the controller!
Current During Function Check (HOLD)
Depending on the parameter setting, the current outputs switch to one of the following states:
• Current measurement
• Last measured value (factory setting)
• Fixed value
Message when the Current Range is Exceeded
As delivered, the “Failure” message is generated when the current range is exceeded
(< 3.8 mA or > 20.5 mA). This setting can be changed in the parameter settings for the respective
measuring channel in the Messages menu:
Parameter Setting [I] [II] [Sensor] Messages
5.5.2 Relay Contacts
Up to three free relay contacts K1 … K3 can be configured. Contacts K2 and K3 are occupied if the con-
troller is used. ➜PID Controller, p. 61
Notes on wiring ➜Relay Contacts: Protective Wiring, p. 27
Use of Relay Contacts
The following uses are possible:
• Off
• Failure
• Maintenance required
• Out of specification
• Function check
• Limit value
• Rinse contact
• Rinse contact (channel) (if using two channels)
• Parameter set B active
• USP output (with conductivity sensor only)
• Sensoface
• Sensoface (channel) (if using two channels)
The switching behavior can be adjusted (normally open or normally closed contact).
• N/O/normally open/make contact: the relay contact closes when it is activated.
• N/C/normally closed/break contact: the relay opens when it is activated.
ON/OFF delays can also be configured.
Settings can be changed in the Relay Contacts submenu:
Parameter Setting Inputs and Outputs Relay Contacts
Stratos Multi E401N
58
Usage: Failure
01. Open Inputs and Outputs Relay Contacts .
02. Select the desired contact with the up/down arrow keys and enter.
03. Usage: Failure
04. Set the contact parameters.
Failure is active
• if a value has exceeded or fallen below “Failure Limit Hi“ or “Failure Limit Lo“ respectively
• if the device’s measuring range limits are exceeded
• in the event of other failure messages
That means that the equipment no longer operates properly or that process parameters have reached
a critical value.
The relay contact is not activated for “Function check” (HOLD).
Usage: Maintenance Required
01. Open Inputs and Outputs Relay Contacts .
02. Select the desired contact with the up/down arrow keys and enter.
03. Usage: Maintenance Required
04. Set the contact parameters.
Maintenance required is active if messages appear that require maintenance. That means that the
equipment is still operating properly but should be serviced, or that process parameters have reached
a value requiring intervention. Typical example: The meter detected a worn sensor.
The relay contact is not activated for “Function check” (HOLD).
Usage: Out of Specification
01. Open Inputs and Outputs Relay Contacts .
02. Select the desired contact with the up/down arrow keys and enter.
03. Usage: Out of Specification
04. Set the contact parameters.
Out of specification is active
• if a value has exceeded or fallen below “Out of Specification Hi“ or “Out of Specification Lo“
respectively
• if the device has detected deviations from the permissible ambient or process conditions
• if faults are present indicating that the measurement uncertainty is probably greater than to be
expected under normal operating conditions
The relay contact is not activated for “Function check” (HOLD).
Stratos Multi E401N
59
Usage: Function Check
01. Open Inputs and Outputs Relay Contacts .
02. Select the desired contact with the up/down arrow keys and enter.
03. Usage: Function Check
04. Set the contact parameters.
Function check (HOLD) is active:
• during calibration (only the corresponding channel)
• during maintenance (current source, relay test)
• during parameter setting at the operator level or the administrator level
• during an automatic rinse cycle.
The output signal is temporarily frozen.
Usage: Limit
01. Open Inputs and Outputs Relay Contacts .
02. Select the desired contact with the up/down arrow keys and enter.
03. Usage: Limit
04. Set the contact parameters.
Contact K1 (Admin.)
Process Variable
Back
Usage
Limit
Hysteresis
Direction
Limit
pH Value
Minimum
pH 3.00
pH 0.10
Contact Type
Normally Open N/O
Back to Meas.
II
The hysteresis prevents small fluctuations in the measured value around the limit from constantly trig-
gering a switching operation.
The hysteresis is configurable and can be activated with an ON/OFF delay time.
In the measurement display, an icon is used to indicate if the limit has been exceeded.
Hysteresis +
Limit
1
0
Contact
Hysteresis -
Limit
Contact
1
0
Signal
Signal
Measured value falls below limit
Measured value exceeds limit
Effective direction:
Minimum
Effective direction:
Maximum
Stratos Multi E401N
60
Usage: Rinse Contact
Please note when configuring the “Rinse Contact” function
• The “function check” (HOLD) mode (e.g. during parameter setting) delays the execution of the
“rinse contact” function.
• Up to 3 rinse functions (contacts K1 ... K3) can be configured independently of each other.
• The individual rinse functions are not synchronized with each other.
Configuring the Rinse Contact
01. Open Inputs and Outputs Relay Contacts .
02. Select the desired contact with the up/down arrow keys and enter.
03. Usage: Rinse Contact
04. Select contact type (e.g. “N/O”).
05. Specify rinsing interval.
06. Specify rinse duration.
07. Specify rinse/meas. lead time.
Note: Function check (HOLD) is active during the defined “Lead Time...”
08. Logbook entry: off/on
Usage: USP Output
Can be activated when using a conductivity sensor and the USP function ➜USP Function, p. 81
01. Open Inputs and Outputs Relay Contacts .
02. Select the desired contact with the up/down arrow keys and enter.
03. Usage: USP Output
04. Assign USP channel.
05. Set the contact parameters.
Usage: Sensoface
Sensoface messages can be output via a relay contact.
If using two sensors, the Sensoface messages can be assigned to different contacts:
01. Open Inputs and Outputs Relay Contacts .
02. Select the desired contact with the up/down arrow keys and enter.
03. Usage: Sensoface Channel
04. Select channel.
Contact K1 (Admin.)
Channel
Back
Usage
Contact Type
ON Delay
OFF Delay
Sensoface Channel
Memosens pH
N/O..
I Memosens pH
Memosens Cond
II
0 s
0 s
05. Set the contact parameters.
06. Using the left softkey: Back, go to the Relay Contacts menu and configure the second contact.
See also
➜Sensocheck and Sensoface, p. 151
Stratos Multi E401N
61
5.5.3 PID Controller
The PID controller can be configured as a pulse length or pulse frequency controller.
Controller Type
Pulse length or pulse frequency controller; see below.
Controlled Variable
Depending on the connected sensors.
Setpoint and
Neutral Zone
Input of setpoint and neutral zone in percent of the respective controller variable.
Pulse Period or
Maximum Pulse Frequency
0 ... 600 seconds or
0 ... 180 per minute
(P) Controller Gain
Specified in percent
(I) Reset Time
0 ... 9999 seconds. 0 s = reset time (integral action) deactivated.
(D) Rate Time
0 ... 9999 seconds. 0 s = rate time (derivative action) deactivated.
Feed Time Alarm after
0 ... 9999 seconds
Behavior during HOLD
Y = constant oder Y = 0 %
Settings can be changed in the Relay Contacts submenu:
Parameter Setting Inputs and Outputs Relay Contacts Controller
As soon as the “PID Linear” controller is selected, contacts K2 and K3 are occupied by the controller.
The corresponding settings are displayed in the Controller Contacts K2/K3 submenu; see the table.
Pulse Length Controller
The pulse length controller is used to control a valve as an actuator. It switches the contact on for a
length of time depending on the controller output (Y). The cycle duration remains constant. The ON
time does not fall below the minimum of 0.5 s, even if the controller output assumes corresponding
values.
1_
0_
1_
0_
2
1
3
1 ON time (Y = 20 %)
3 Pulse period
2 ON time (Y = 80 %)
Stratos Multi E401N
62
Pulse Frequency Controller
The pulse frequency controller is used to control a frequency-controlled actuator (metering pump).
It varies the frequency with which the contacts are switched on. The maximum pulse frequency [Imp/
min] can be configured, and is dependent on the actuator. The ON time is constant. It is automatically
calculated from the configured maximum pulse frequency.
1_
0_
1_
0_
1
2
3
1 ON time
3 Pulse frequency (Y = 80 %)
2 Pulse frequency (Y = 20 %)
Adjustable Controlled Variables
Sensor type
Controlled variable
pH, ORP
pH value, ORP, temperature
Conductivity
Conductivity, temperature,
with TAN option FW-E009: Concentration (liquid)
Oxygen
Saturation %air, saturation %O2, temperature
See also
➜Basics of PID Control, p. 215
5.5.4 Control Inputs
Stratos Multi provides 2 digital optocoupler inputs (OK1, OK2).
The following functions (depending on the parameter setting) can be started via the control signals:
Input OK1: off, parameter set selection, flow, function check total, or function check channel
• The function of optocoupler input OK1 is defined in the System Control :
Parameter Setting System Control Function Control ➜Function Control, p. 46
Input OK2: off, function check total, or functional check channel.
• Selection in the menu Parameter Setting Inputs and Outputs Control Inputs Input OK2
The switching level for the control signal must be specified:
Parameter Setting Inputs and Outputs Control Inputs Contacts Input OK…
Input level: Active 10 ... 30 V or active < 2 V
Stratos Multi E401N
63
5.6 Sensor Selection [I] [II]
Note: Function check (HOLD) is active.
Stratos Multi is factory-set to a pH value measurement with Memosens sensor. This measuring func-
tion is also provided by the ORP measurement. The measuring function can be changed to measure
conductivity or oxygen in the Parameter Setting menu:
To prepare Stratos Multi for measurements, the operating mode of the used measuring channel must
be set:
Parameter Setting Sensor Selection [I] [II]
Sensor Selection [I] (measuring channel I): Memosens sensor or SE740 optical oxygen sensor (LDO)
Sensor Selection [II] (measuring channel II): Second Memosens sensor or analog sensor via measuring
module
If Memosens sensors are connected directly, the process variable can be set to “Auto”. In this case, the
sensor is automatically detected by the device, which sets itself to the correct process variable. This
does not apply to the MK-MS095N Memosens module.
NOTICE! Parameters that are dependent on process variables (e.g. measurement display, current out-
puts, contacts, ...) are set separately; this step is not automatic.
If “Auto” is not used with a Memosens sensor, and in general when using analog sensors, the operating
mode must be set to the used sensor. The parameters that are dependent on process variables can
then be set when a sensor is not connected.
Power Out
The output voltage of terminal 6 is selected in the Power Out menu item: 3.1 V, 14 V, 24 V. This output
voltage is used to supply power to special sensors or external transmitters. The voltage at terminal 6 is
automatically adjusted when using the SE740 optical oxygen sensor. The menu item is then not avail-
able.
Identifying a Memosens Sensor
A connected Memosens sensor is displayed as follows: sensor name, manufacturer, serial number, date
of last adjustment
All relevant and typical sensor parameters are automatically transferred to the Stratos Multi.
Stratos Multi E401N
64
5.7 Process Variable pH
Note: Function check (HOLD) is active.
Note: After changing the process variable or measuring mode, Stratos Multi retains its settings but
needs to be reconfigured.
Selecting a Memosens pH Sensor
Parameter Setting Sensor Selection [I] [II] Sensor Selection [I]
Selection of the Memosens pH sensor connected to the RS-485 interface (terminals 1 … 5):
Parameter:
Auto or pH
Mode:
Memosens
Functionality:
pH, ISFET, or pH/ORP (depending on sensor type)
Selecting a Second Memosens pH Sensor
Parameter Setting Sensor Selection [I] [II] Sensor Selection [II]
Selection of a second Memosens pH sensor connected to the MK-MS 095N measuring module:
Module:
MK-MS
Parameter:
pH
Mode:
Memosens
Functionality:
pH, ISFET, or pH/ORP (depending on sensor type)
Adjustable parameters for Memosens pH sensors Parameter Setting [I] [II] Memosens pH
Input Filter
Pulse Suppression
Enable/disable suppression of interference pulses.
Sensor Data
➜Sensor Data, p. 66
Sensoface
Enable/disable display of Sensoface messages and icons.
Sensor Monitoring Details
Option to enter individual limits for monitoring slope and
zero point.
Enable/disable Sensocheck sensor monitoring. Set whether
Sensocheck should generate failure or maintenance required
messages.
Option to enter individual values before a message is trig-
gered for response time, sensor wear, sensor operating time
and SIP counter; also CIP counter with pH/ORP sensors; also
operating point and leakage current with ISFET sensors.
Cal Presettings
Presetting of calibration mode and corresponding parameters, configuration of drift check
and calibration timer. ➜Calibration Presettings, p. 68
TC Process Medium
➜Temperature Compensation of Process Medium, p. 69
ORP/rH Value
With Memosens pH/ORP sensor:
Selection of reference electrode:
Ag/AgCl, KCl 1 mol, Ag/AgCl, KCl 3 mol, Hg,Tl/TlCl, KCl 3.5 mol, Hg/Hg2SO4, K2SO4 sat
Enable/disable ORP conversion to standard hydrogen electrode SHE.
Calculate rH with or without factor.
Delta Function
Display deviations from a preset value (delta value):
Output value = measured value – delta value ➜Delta Function, p. 70
Messages
Enable/disable messages for the separate process variables, or specify individual limits.
➜Messages, p. 71
Stratos Multi E401N
65
Selecting an Analog pH Sensor
Parameter Setting Sensor Selection [I] [II] Sensor Selection [II]
Selection of a pH or pH/ORP sensor connected to the MK-MS 015N measuring module:
Module:
MK-PH
Mode:
Analog
Adjustable parameters for analog sensors
Parameter Setting [II] Analog pH
Input Filter
Pulse Suppression
Enable/disable suppression of interference pulses.
Sensor Data
➜Sensor Data, p. 66
Sensor Type and Sensoface
Sensoface, temperature monitoring, and the details of sensor
monitoring can be set, depending on the sensor type.
Temperature Detection
Select temperature probe, set measuring and calibration
temperature.
Sensor Monitoring Details
Slope, zero point, set Sensocheck of reference and glass
electrodes, and select response time.
Cal Presettings
Presetting of calibration mode and calibration timer with corresponding parameters.
➜Calibration Presettings, p. 68
TC Process
Medium
➜Temperature Compensation of Process Medium, p. 69
ORP/rH Value
With pH/ORP sensor:
Selection of reference electrode.
Enable/disable ORP conversion to standard hydrogen electrode SHE.
Calculate rH with or without factor.
Delta Function
Display deviations from a preset value (delta value):
Output value = measured value – delta value ➜Delta Function, p. 70
Messages
Enable/disable messages for the separate process variables, or specify individual limits.
➜Messages, p. 71
Stratos Multi E401N
66
5.7.1 Sensor Data
Memosens Sensors
Memosens sensors provide relevant sensor data automatically.
Analog Sensors
The sensor type must be selected if using analog sensors:
Parameter Setting [II] Analog … Sensor Data
Sensor Data
Sensor Type
Standard
On
Back
Sensor Monitoring Details
Temperature Detection
Sensoface
II
Back to Meas.
01. In Temperature Detection , select the used temperature probe and whether the temperature is to
be measured automatically or manually during measurement and/or calibration.
Temperature Detection (Admin.)
Measuring Temp
Back
Temperature Probe
Cal Temperature
Pt 1000
Automatic
Automatic
II
Pt 100
Pt 1000
NTC 30k
NTC 8.55k
Balco 3 kΩ
Sensoface
The Sensoface icons provide the user with diagnostic information on wear and required maintenance
of the sensor. In measuring mode, an icon (happy, neutral, or sad smiley) is shown on the display to re-
flect the continuous monitoring of the sensor parameters.
You can configure the current outputs so that a Sensoface message generates a 22-mA error signal.
Parameter Setting Inputs and Outputs Current Outputs Current Output I…
Behavior during Messages
The Sensoface message can also be output via a relay contact:
Parameter Setting Inputs and Outputs Relay Contacts Contact K… ➜Usage: Sensoface, p. 60
Sensoface monitors the pH sensor on the basis of the following parameters:
Slope, zero point, glass impedance (if Sensocheck is enabled), response time, calibration timer, wear
Enabling/Disabling Sensoface
Sensoface is enabled and disabled in the Sensor Data submenu:
Parameter Setting [I] [II] [Sensor] Sensor Data
Note: After a calibration, a smiley is always displayed for confirmation, even if Sensoface is disabled.
Stratos Multi E401N
67
Adjusting Sensor Monitoring
01. Open Sensor Data Sensor Monitoring Details .
02. Open a sensor parameter, e.g., Slope .
03. Set Monitoring of the slope to automatic or individual.
04. If you select “Individual”: The nominal slope and the min/max limits can be entered.
05. In the Message menu item, select whether and how an exceeded limit is to be displayed:
Off
No message, but the parameter is still shown on the sensor diagram.
Failure
A failure message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
Mainte-
nance
A maintenance required message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
06. Set the sensor monitoring details for other sensor data, e.g. zero point, Sensocheck, response time,
sensor wear, or sensor operating time.
07. Using the left softkey: Back, confirm the sensor monitoring settings and adjust further parameters.
or
Using the right softkey: Back to Meas., confirm the sensor monitoring settings and end the func-
tion check (HOLD).
Stratos Multi E401N
68
CIP/SIP Counters
Note: CIP/SIP counters are available for the following pH sensor types:
SIP counter: Memosens pH, Memosens pH/ORP; CIP counter: Memosens pH/ORP
CIP/SIP cycles are used to clean or sterilize process-wetted parts in the process. Depending on the ap-
plication, either one chemical (alkaline solution, water) or several chemicals (alkaline solution, water,
acidic solution, water) are used.
• CIP temperature > 55 °C/131 °F
• SIP temperature > 115 °C/239 °F
The cleaning (cleaning in place) and sterilization (sterilization in place) cycles are counted to measure
the load on the sensor, e.g., in biotechnology applications.
Note: If measurements are generally taken at high temperatures (> 55 °C/131 °F), the counters should
be switched off.
When a CIP/SIP counter is switched on, a maximum number of cycles can be entered. As soon as the
counter has reached the specified value, a maintenance required message is triggered and the NAMUR
icon
is displayed.
Note: A CIP or SIP cycle is only entered into the logbook 2 hours after the start to ensure that the cycle
is complete.
Note: With Memosens sensors, an entry is also made in the sensor.
5.7.2 Calibration Presettings
The calibration presettings can be defined in the parameter settings or adjusted directly in the Calibra-
tion menu prior to the calibration.
Calibration mode: Presetting of calibration mode, e.g., Calimatic, manual, product calibration, data en-
try, temperature
If Calimatic automatic calibration is selected, the buffer set to be used must also be selected.
Calibration points: Selection of how many calibration points are to be used in the calibration
Drift check: Sets the sensitivity of the drift check (fine, standard, coarse)
Cal Presettings (Administrator)
Buffer Set
Back
Calibration Mode
Calibration Points
Drift Check
Calibration Timer
ORP Check
Calimatic
Knick CaliMat
Automatic
Standard
II
Knick CaliMat
Mettler-Toledo
Merck/Riedel
DIN19267
NIST Standard
Calibration Timer
When a preset calibration interval expires, the calibration timer generates a message text to indicate
the need for calibration. If “Auto” is selected, the interval is set to 168 h. If “Individual” is selected, you
can select a custom interval.
Calibration Timer (Admin.)
Adaptive Cal Timer
Back
Monitoring
Auto
Calibration Timer
168 h
Off
II
Off
On
Stratos Multi E401N
69
Note: If Sensoface is enabled, a neutral smiley is displayed once 80 % of the interval has expired. Once
the entire interval has expired, a sad smiley is shown, a maintenance required message is generated,
and the corresponding NAMUR icon
is displayed. If the current outputs are configured accordingly,
a 22-mA error signal is generated.
Adaptive calibration timer: The time until the next calibration is automatically shortened depending
on the temperature and pH value.
Old sensor = timer expires faster.
The following measuring conditions shorten the adaptive calibration timer interval:
• Temperatures above 30 °C/86 °F
• pH ranges below pH 2 or above pH 12
The message test is displayed in the Diagnostics menu:
Diagnostics Message List
The calibration timer is reset after each calibration.
Settings can be changed in the Cal Presettings submenu:
Parameter Setting [I] [II] [Sensor] Cal Presettings
5.7.3 Temperature Compensation of Process Medium
Note: If temperature compensation of the process medium is enabled, “TC” is shown on the display in
measuring mode.
Note: Function check (HOLD) is active. The current output and relay contacts behave in accordance
with the parameter settings. Return to measuring mode to exit the function check, e.g., with the right
softkey: Back to Meas.
The following are available for temperature compensation:
• Linear with input of a temperature coefficient TC
• Ultrapure water
• Table
Linear Temperature Compensation of Process Medium
If the medium’s pH value changes in linear fashion with the temperature, the temperature coefficient
TC can be determined for temperature compensation in %/K as follows:
TC = (pH25 – pHT) · 100 / (25 °C – T) [%/K]
TC
Temperature coefficient [%/K]
pH25
pH value at 25 °C
pHT
pH value at measuring temperature T
T
Measuring temperature [°C]
Stratos Multi E401N
70
Table
When using process media with a known pH value temperature response, the pH output value can be
corrected using a table. The percentage deviation from the measured value in % can be entered for
temperatures between 0 and 95 °C in steps of 5 °C. The pH output value is then corrected by the
corresponding percentage deviation from the measured value in %, depending on the measuring
temperature. Table values are linearly interpolated. If the temperature falls below or exceeds the
specified value (< 0 °C or > 95 °C), the last value in the table is used for calculation.
The table must be completed with the following values in steps of 5 °C:
((pH25 / pHT) – 1) • 100 [%]
pH25
pH value at 25 °C
pHT
pH value at measuring temperature T
Settings can be changed in the TC Process Medium submenu:
Parameter Setting [I] [II] … pH TC Process Medium
Note: If the delta function and TC correction are enabled at the same time, the TC correction is carried
out first and the delta value is then deducted.
5.7.4 Delta Function
Note: If the delta function is enabled, “∆” is shown on the display in measuring mode.
If a delta value is specified, the measuring system calculates the difference
output value = measured value – delta value
The delta value can be set using the “+” or “-” signs. If using a negative sign, the delta value is added to
the measured value.
The delta value is adjusted in the Delta Function submenu:
Parameter Setting [I] [II] [Sensor] Delta Function
All outputs are controlled by the output value; the displays show the output value.
Note: If the delta function and TC correction are enabled at the same time, the TC correction is carried
out first and the delta value is then deducted.
Stratos Multi E401N
71
5.7.5 Messages
All values determined by the measuring module or sensor can generate messages.
Messages can be configured for the following process variables:
• pH value
• ORP (with pH/ORP sensor)
• rH value (with pH/ORP sensor)
• Temperature
• pH voltage
Configuring Messages
Process variable limits for the monitoring range can be selected in the Messages submenu:
Menu Selection Parameter Setting [I] [II] [Sensor] Messages Messages [Process Variable] Monitoring
• Max. Device Limits: Messages are generated when the process variable is outside the measuring
range. The “failure” or “out of specification” icons are shown; the corresponding relay contact is
enabled. The current outputs can signal a 22-mA message (user-defined).
• Variable Limits: Upper and lower limits at which a message is generated can be defined for the
“failure” and “out of specification” messages.
Displaying Messages
Go to the Diagnostics menu if the “failure”, “maintenance required”, or “out of specification” icons are
flashing on the display.
Menu Selection Diagnostics Message List
All active messages are displayed in the “Message List” menu item with the following information: Er-
ror number, type (failure, maintenance required, out of specification), channel, message text.
You can scroll forward and backward using the up/down arrow keys.
The error message disappears from the display around 2 s after the error is cleared.
An overview of message texts with notes on troubleshooting can be found in the “Troubleshooting”
chapter. ➜Malfunction States, p. 136
Stratos Multi E401N
72
5.8 Process Variable ORP
Note: Function check (HOLD) is active.
Note: After changing the process variable or measuring mode, Stratos Multi retains its settings but
needs to be reconfigured.
Selecting a Memosens ORP Sensor
Parameter Setting Sensor Selection [I] [II] Sensor Selection [I]
Selection of a Memosens ORP sensor connected to the RS-485 interface (terminals 1 … 5):
Parameter:
Auto or pH
Mode:
Memosens
Functionality:
ORP
Selecting a Second Memosens ORP Sensor
Parameter Setting Sensor Selection [I] [II] Sensor Selection [II]
Selection of a second Memosens ORP sensor connected to the MK-MS 095N measuring module:
Module:
MK-MS
Parameter:
pH
Mode:
Memosens
Functionality:
ORP
Adjustable parameters for Memosens ORP sensors Parameter Setting [I] Memosens ORP
Input Filter
Pulse Suppression
Enable/disable suppression of interference pulses.
Sensor Data
➜Sensor Data, p. 73
Sensoface
Enable/disable display of Sensoface messages and icons.
Sensor Monitoring
Details
Option to enter individual limits for monitoring ORP offset.
Option to enter individual limits before a message for sensor
operating time and SIP counter is triggered.
Cal Presettings
Presetting of the calibration mode; configuration of the calibration timer and ORP
check. ➜Calibration Presettings, p. 74
ORP/rH Value
Selection of reference electrode.
Enable/disable ORP conversion to standard hydrogen electrode SHE.
If using a pH sensor connected via a module at the same time: Calculate rH with or
without factor.
Delta Function
Display deviations from a preset value (delta value):
Output value = measurement - delta value
Messages
Enable/disable messages for the separate process variables, or specify individual limits.
➜Messages, p. 75
Selecting an Analog ORP Sensor
Parameter Setting Sensor Selection [I] [II] Sensor Selection [II]
Selection of an ORP sensor connected to the MK-PH015N measuring module:
Module:
MK-PH
Mode:
Analog
When using an analog ORP sensor, the menus are the same as for an analog pH sensor:
Parameter Setting [II] Analog pH
Stratos Multi E401N
73
5.8.1 Sensor Data
Memosens sensors provide relevant sensor data automatically.
Sensoface
The Sensoface icons provide the user with diagnostic information on wear and required maintenance
of the sensor. In measuring mode, an icon (happy, neutral, or sad smiley) is shown on the display to re-
flect the continuous monitoring of the sensor parameters.
You can configure the current outputs so that a Sensoface message generates a 22-mA error signal.
Parameter Setting Inputs and Outputs Current Outputs Current Output I…
Behavior during Messages
The Sensoface message can also be output via a relay contact:
Parameter Setting Inputs and Outputs Relay Contacts Contact K… ➜Usage: Sensoface, p. 60
Enabling/Disabling Sensoface
Sensoface is enabled and disabled in the Sensor Data submenu:
Parameter Setting [I] [II] [Sensor] Sensor Data
Note: After a calibration, a smiley is always displayed for confirmation, even if Sensoface is disabled.
Adjusting Sensor Monitoring
01. Open Sensor Data Sensor Monitoring Details .
02. Open a sensor parameter, e.g., ORP Offset .
03. Set Monitoring of the ORP offset to automatic or individual.
04. If you select “Individual”: The nominal ORP offset and the min/max limits can be entered.
05. In the Message menu item, select whether and how an exceeded limit is to be displayed:
Off
No message.
Failure
A failure message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
Maintenance
A maintenance required message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
06. Set the sensor monitoring details for further sensor data such as sensor operating time or SIP
counter.
07. Using the left softkey: Back, confirm the sensor monitoring settings and adjust further parameters.
or
Using the right softkey: Back to Meas., confirm the sensor monitoring settings and end the func-
tion check (HOLD).
CIP/SIP Counters
Note: CIP/SIP counters are available for the following pH sensor types:
SIP counter: Memosens pH, Memosens pH/ORP; CIP counter: Memosens pH/ORP
CIP/SIP cycles are used to clean or sterilize process-wetted parts in the process. Depending on the ap-
plication, either one chemical (alkaline solution, water) or several chemicals (alkaline solution, water,
acidic solution, water) are used.
• CIP temperature > 55 °C/131 °F
• SIP temperature > 115 °C/239 °F
The cleaning (cleaning in place) and sterilization (sterilization in place) cycles are counted to measure
the load on the sensor, e.g., in biotechnology applications.
Stratos Multi E401N
74
Note: If measurements are generally taken at high temperatures (> 55 °C/131 °F), the counters should
be switched off.
When a CIP/SIP counter is switched on, a maximum number of cycles can be entered. As soon as the
counter has reached the specified value, a maintenance required message is triggered and the NAMUR
icon
is displayed.
Note: A CIP or SIP cycle is only entered into the logbook 2 hours after the start to ensure that the cycle
is complete.
Note: With Memosens sensors, an entry is also made in the sensor.
5.8.2 Calibration Presettings
The calibration presettings can be defined in the parameter settings or adjusted directly in the Calibra-
tion menu prior to the calibration.
Calibration mode: Presetting of the calibration mode, e.g., ORP data entry, ORP adjustment, ORP
check, temperature
Calibration timer: When a preset calibration interval expires, the calibration timer generates a mes-
sage text to indicate the need for calibration. If “Auto” is selected, the interval is set to 168 h. If “Individ-
ual” is selected, you can select a custom interval.
Note: If Sensoface is enabled, the smiley is “sad” as soon as the interval has expired. You can configure
the current outputs so that a Sensoface message generates a 22-mA error signal.
ORP check: Settings for test period in seconds and test difference in millivolts
Cal Presettings (Admin.)
Back
Calibration Mode
Drift Check
Calibration Timer
ORP Check
ORP Data Entry
Standard
II
Data Entry
ORP Data Entry
ORP Adjustment
ORP Check
Temperature
Settings can be changed in the Cal Presettings submenu:
Parameter Setting [I] [II] [Sensor] Cal Presettings
5.8.3 Delta Function
Note: If the delta function is enabled, “∆” is shown on the display in measuring mode.
If a delta value is specified, the measuring system calculates the difference
output value = measured value – delta value
The delta value can be set using the “+” or “-” signs. If using a negative sign, the delta value is added to
the measured value.
The delta value is adjusted in the Delta Function submenu:
Parameter Setting [I] [II] [Sensor] Delta Function
All outputs are controlled by the output value; the displays show the output value.
Note: If the delta function and TC correction are enabled at the same time, the TC correction is carried
out first and the delta value is then deducted.
Stratos Multi E401N
75
5.8.4 Messages
All values determined by the measuring module or sensor can generate messages.
Messages can be configured for the following process variables:
• ORP voltage
• Temperature
Configuring Messages
Process variable limits for the monitoring range can be selected in the Messages submenu:
Menu Selection Parameter Setting [I] [II] [Sensor] Messages Messages [Process Variable] Monitoring
• Max. Device Limits: Messages are generated when the process variable is outside the measuring
range. The “failure” or “out of specification” icons are shown; the corresponding relay contact is en-
abled. The current outputs can signal a 22-mA message (user-defined).
• Variable Limits: Upper and lower limits at which a message is generated can be defined for the
“failure” and “out of specification” messages.
Displaying Messages
Go to the Diagnostics menu if the “failure”, “maintenance required”, or “out of specification” icons are
flashing on the display.
Menu Selection Diagnostics Message List
All active messages are displayed in the “Message List” menu item with the following information: Er-
ror number, type (failure, maintenance required, out of specification), channel, message text.
You can scroll forward and backward using the up/down arrow keys.
The error message disappears from the display around 2 s after the error is cleared.
An overview of message texts with notes on troubleshooting can be found in the “Troubleshooting”
chapter. ➜Malfunction States, p. 136
Stratos Multi E401N
76
5.9 Process Variable Conductivity (Contacting)
Note: Function check (HOLD) is active.
Note: After changing the process variable or measuring mode, Stratos Multi retains its settings but
needs to be reconfigured.
Selecting a Memosens Conductivity Sensor
Parameter Setting Sensor Selection [I] [II] Sensor Selection [I]
Selection of a Memosens conductivity sensor connected to the RS-485 interface (terminals 1 … 5):
Parameter:
Auto or Conductivity
Mode:
Memosens
Functionality:
2-Electrode Sensor or 4-Electrode Sensor (depending on sensor type)
Selecting a Second Memosens Conductivity Sensor
Parameter Setting Sensor Selection [I] [II] Sensor Selection [II]
Selection of a second Memosens conductivity sensor connected to the MK-MS 095N measuring mod-
ule:
Module:
MK-MS
Parameter:
Conductivity
Mode:
Memosens
Functionality:
2-Electrode Sensor or 4-Electrode Sensor (depending on sensor type)
Adjustable parameters for Memosens conductivity sensors Parameter Setting [I] [II] Memosens Cond
Input Filter
Pulse Suppression
Enable/disable suppression of interference pulses.
Sensor Data
➜Sensor Data, p. 77
Sensoface
Enable/disable display of Sensoface messages and icons.
Sensor Monitoring
Details
Option to enter individual limits for monitoring cell
constants.
Disable Sensocheck sensor monitoring or select whether
Sensocheck should generate failure or maintenance
required messages.
Option to enter individual limits before a message for SIP
counter, CIP counter, and sensor operating time is
triggered.
Cal Presettings
Presetting of calibration mode with corresponding parameters.
➜Calibration Presettings, p. 79
TC Process Medium
➜Temperature Compensation of Process Medium, p. 80
Concentration
➜Concentration (TAN Option FW-E009), p. 80
TDS
Enable/disable TDS function
(“Total Dissolved Solids” that influence conductivity)
USP
Enable/disable USP function for monitoring ultrapure water and set the USP limit.
➜USP Function, p. 81
Messages
Enable/disable messages for the separate process variables, or specify individual limits.
➜Messages, p. 82
Stratos Multi E401N
77
Selecting an Analog Conductivity Sensor
Parameter Setting Sensor Selection [I] [II] Sensor Selection [II]
Selection of a conductivity sensor connected to the MK-COND025N measuring module:
Module:
MK-COND
Mode:
Analog
Analog 2-channel conductivity measurement with the MK-CC065 measuring module
➜Dual Conductivity Measurement, p. 89
Adjustable parameters for analog conductivity sensors Parameter Setting [II] Analog Cond
Input Filter
Pulse Suppression
Enable/disable suppression of interference pulses.
Sensor Data
➜Sensor Data, p. 77
Sensor Type
Select the used sensor type: 2-el.-sensor, 4-el.-sensor, SE600,
SE602, SE603, SE604, SE610, SE620, SE630.
Nominal Cell Constant
Enter when 2-el. or 4-el. sensor is selected.
Sensoface
Enable/disable display of Sensoface messages and icons.
Sensocheck
Enable or disable Sensocheck to generate failure or mainte-
nance required messages.
Temperature Detection
Set measuring and calibration temperature.
When 2-el. or 4-el. sensor is selected: Select temperature
probe.
Cal Presettings
Presetting of calibration mode with corresponding parameters.
➜Calibration Presettings, p. 79
TC Process Medium
➜Temperature Compensation of Process Medium, p. 80
Concentration
➜Concentration (TAN Option FW-E009), p. 80
TDS
Enable/disable TDS function
(“Total Dissolved Solids” that influence conductivity)
USP
Enable/disable USP function for monitoring ultrapure water and set the USP limit.
➜USP Function, p. 81
Messages
Enable/disable messages for the separate process variables, or specify individual limits.
➜Messages, p. 82
5.9.1 Sensor Data
Memosens sensors provide relevant sensor data automatically.
The sensor type must be selected if using analog sensors:
Parameter Setting [II] Analog … Sensor Data
Sensor Data
Sensor Type
2-El. Sensor
Back
1.000 /cm
Nom. Cell Constant
II
Back to Meas.
Temperature Detection
On
Sensoface
Off
Sensocheck
01. Select sensor type.
02. Enter the sensor’s nominal cell constant.
03. In Temperature Detection , select the used temperature probe and whether the temperature is to
be measured automatically or manually during measurement and/or calibration.
Stratos Multi E401N
78
Sensoface
The Sensoface icons provide the user with diagnostic information on wear and required maintenance
of the sensor. In measuring mode, an icon (happy, neutral, or sad smiley) is shown on the display to
reflect the continuous monitoring of the sensor parameters.
You can configure the current outputs so that a Sensoface message generates a 22-mA error signal.
Parameter Setting Inputs and Outputs Current Outputs Current Output I…
Behavior during Messages
The Sensoface message can also be output via a relay contact:
Parameter Setting Inputs and Outputs Relay Contacts Contact K… ➜Usage: Sensoface, p. 60
Sensoface monitors the conductivity sensor on the basis of the following parameters:
Cell constant, polarization (if Sensocheck is enabled)
With Memosens sensors, also: Number of CIP and SIP cycles compared to the specified “Sensor Moni-
toring Details”.
Enabling/Disabling Sensoface
Sensoface is enabled and disabled in the Sensor Data submenu:
Parameter Setting [I] [II] [Sensor] Sensor Data
Note: After a calibration, a smiley is always displayed for confirmation, even if Sensoface is disabled.
Adjusting Sensor Monitoring
Note: This function is not available for analog sensors.
01. Open Sensor Data Sensor Monitoring Details .
02. Open a sensor parameter, e.g., Cell Constant .
03. Set Monitoring of the cell constant to automatic or individual.
04. If you select “Individual”: The nominal cell constant and the min/max limits can be entered.
05. In the Message menu item, select whether and how an exceeded limit is to be displayed:
Off
No message.
Failure
A failure message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
Maintenance
A maintenance required message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
06. Set the sensor monitoring details for further sensor data, e.g., Sensocheck, sensor operating time,
or CIP/SIP counter.
07. Using the left softkey: Back, confirm the sensor monitoring settings and adjust further parameters.
or
Using the right softkey: Back to Meas., confirm the sensor monitoring settings and end the func-
tion check (HOLD).
Stratos Multi E401N
79
CIP/SIP Counters
Note: This function is not available for all sensor types.
CIP/SIP cycles are used to clean or sterilize process-wetted parts in the process. Depending on the ap-
plication, either one chemical (alkaline solution, water) or several chemicals (alkaline solution, water,
acidic solution, water) are used.
• CIP temperature > 55 °C/131 °F
• SIP temperature > 115 °C/239 °F
The cleaning (cleaning in place) and sterilization (sterilization in place) cycles are counted to measure
the load on the sensor, e.g., in biotechnology applications.
Note: If measurements are generally taken at high temperatures (> 55 °C/131 °F), the counters should
be switched off.
When a CIP/SIP counter is switched on, a maximum number of cycles can be entered. As soon as the
counter has reached the specified value, a maintenance required message is triggered and the NAMUR
icon
is displayed.
Note: A CIP or SIP cycle is only entered into the logbook 2 hours after the start to ensure that the cycle
is complete.
Note: With Memosens sensors, an entry is also made in the sensor.
Settings can be changed in the Sensor Monitoring Details submenu:
Parameter Setting [I] [II] [Sensor] Sensor Data Sensor Monitoring Details
5.9.2 Calibration Presettings
The calibration presettings can be defined in the parameter settings or adjusted directly in the Calibra-
tion menu prior to the calibration.
Calibration mode: Presetting of calibration mode, e.g., automatic, manual, product calibration, data
entry, temperature
Further options are available depending on the calibration mode.
Automatic:
• Select calibration solution (NaCl/KCl)
Product calibration:
• Product calibration with/without temperature compensation
Cal Presettings (Admin.)
Product Calibration
Back
Calibration Mode
Conductivity
Product
Wi
II
Conductivity
Automatic
Manual
Product
Data Entry
Temperature
Calibration presettings can be changed in the Cal Presettings submenu:
Parameter Setting [I] [II] … Cond Cal Presettings
Stratos Multi E401N
80
5.9.3 Temperature Compensation of Process Medium
Note: If temperature compensation of the process medium is enabled, “TC” is shown on the display in
measuring mode.
The following are available for temperature compensation:
• Off
• Linear (enter temperature coefficient TC)
• EN 27888 (natural waters)
• Ultrapure water (with different trace impurities)
Trace impurities in ultrapure water
NaCl
Neutral ultrapure water, for conductivity measurement in water processing downstream of gravel bed
filter
HCl
Acidic ultrapure water, for conductivity measurement downstream of cation filter
NH3
Ammoniacal ultrapure water
NaOH
Alkaline ultrapure water
Settings can be changed in the TC Process Medium submenu:
Parameter Setting [I] [II] … Cond TC Process Medium
5.9.4 Concentration (TAN Option FW-E009)
With TAN option FW-E009, the substance concentration in percent by weight (wt%) can be
determined for H2SO4, HNO3, HCl, NaOH, NaCl, and Oleum from the measured conductivity and
temperature values. A custom solution can also be specified.
The menu is only shown if the TAN option is activated.
Parameter Setting [I] [II] … Cond(I) Concentration
See also
➜Concentration Determination (FW-E009), p. 201
Stratos Multi E401N
81
5.9.5 USP Function
Monitoring Ultrapure Water in the Pharmaceutical Industry
The conductivity of ultrapure water in the pharmaceutical industry can be monitored online in
accordance with the guideline “USP” (U.S. Pharmacopeia), Annex 5, Section 645 “Water Conductivity”.
The conductivity is measured without temperature compensation and is compared with limits.
The water is usable without any further testing when the conductivity is below the USP limit.
Configuring the USP Function
The USP value can be configured as a process variable USP% for output (display, current output, limit,
measurement recorder)
Settings can be changed in the USP submenu:
Parameter Setting [I] [II] … Cond(I) USP
Reduced limit: The USP limit can be reduced down to 10 %.
Monitoring: Select whether and how an exceeded limit is to be displayed.
Off
No message, but the parameter is still shown in the Diagnostics menu.
Failure
A failure message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
Mainte-
nance
A maintenance required message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
USP Function: Specifying a Relay Contact
The USP function can also be assigned to a relay contact:
Parameter Setting Inputs and Outputs Relay Contacts Contact K… ➜Usage: USP Output, p. 60
Display of the USP Function in the Diagnostics Menu
Diagnostics [I] [II] … Cond(I) USP Function
Display of the USP limit, the reduced limit, and conductivity.
Stratos Multi E401N
82
5.9.6 Messages
All values determined by the measuring module or sensor can generate messages.
Messages can be configured for the following process variables:
• Conductivity
• Resistivity
• Concentration (with TAN option FW-E009)
• Temperature
• Salinity
Configuring Messages
Process variable limits for the monitoring range can be selected in the Messages submenu:
Menu Selection Parameter Setting [I] [II] [Sensor] Messages Messages [Process Variable] Monitoring
• Max. Device Limits: Messages are generated when the process variable is outside the measuring
range. The “failure” or “out of specification” icons are shown; the corresponding relay contact is en-
abled. The current outputs can signal a 22-mA message (user-defined).
• Variable Limits: Upper and lower limits at which a message is generated can be defined for the
“failure” and “out of specification” messages.
Displaying Messages
Go to the Diagnostics menu if the “failure”, “maintenance required”, or “out of specification” icons are
flashing on the display.
Menu Selection Diagnostics Message List
All active messages are displayed in the “Message List” menu item with the following information: Er-
ror number, type (failure, maintenance required, out of specification), channel, message text.
You can scroll forward and backward using the up/down arrow keys.
The error message disappears from the display around 2 s after the error is cleared.
An overview of message texts with notes on troubleshooting can be found in the “Troubleshooting”
chapter. ➜Malfunction States, p. 136
Stratos Multi E401N
83
5.10 Process Variable Conductivity (Inductive)
Note: Function check (HOLD) is active.
Note: After changing the process variable or measuring mode, Stratos Multi retains its settings but
needs to be reconfigured.
Selecting a Digital Toroidal Conductivity Sensor
Parameter Setting Sensor Selection [I] [II] Sensor Selection [I]
Selection of toroidal Memosens conductivity sensor connected to the RS-485 interface
(terminals 1 … 5):
Parameter:
Auto or Conductivity (Induct.)
Mode:
Memosens
Functionality:
CondI
Selection of an SE680 toroidal digital conductivity sensor connected to the RS-485 interface
(terminals 1 … 5):
Parameter:
Auto or Conductivity (Induct.)
Mode:
Other Digital
Functionality:
SE680K
Selecting a Second Digital Toroidal Conductivity Sensor
Parameter Setting Sensor Selection [I] [II] Sensor Selection [II]
Selection of a second Memosens toroidal conductivity sensor connected to the MK-MS 095N
measuring module:
Module:
MK-MS
Parameter:
Conductivity (Induct.)
Mode:
Memosens
Functionality:
CondI
Selection of a second SE680 digital toroidal conductivity sensor connected to the MK-MS 095N
measuring module:
Module:
MK-MS
Parameter:
Conductivity (Induct.)
Mode:
Other Digital
Adjustable parameters for digital or Memosens sensors for toroidal conductivity
Parameter Setting [I] [II] Digital/Memosens CondI
Input Filter
Pulse Suppression
Enable/disable suppression of interference pulses.
Sensor Data
➜Sensor Data, p. 85
Sensoface
Enable/disable display of the Sensoface icon.
With the SE680 sensor:
Sensocheck
Monitoring of primary and secondary coils.
Enable or disable Sensocheck to generate failure or main-
tenance required messages.
With Memosens:
Sensor Monitoring
Details
Option to enter individual limits for monitoring the cell
factor.
Sensocheck: Monitoring of primary and secondary coils.
Enable or disable Sensocheck to generate failure or main-
tenance required messages.
Option to enter individual limits before a message for SIP
counter, CIP counter, and sensor operating time is trig-
gered.
Stratos Multi E401N
84
Adjustable parameters for digital or Memosens sensors for toroidal conductivity
Parameter Setting [I] [II] Digital/Memosens CondI
Cal Presettings
Presetting of calibration mode with corresponding parameters.
➜Calibration Presettings, p. 86
TC Process Medium
Temperature compensation of process medium.
Concentration
➜Concentration (TAN Option FW-E009), p. 87
TDS
Enable/disable TDS function
(“Total Dissolved Solids” that influence conductivity)
USP
Enable/disable USP function for monitoring ultrapure water and set the USP limit.
➜USP Function, p. 87
Messages
Enable/disable messages for the separate process variables, or specify individual limits.
➜Messages, p. 88
Selecting an Analog Toroidal Conductivity Sensor
Parameter Setting Sensor Selection [I] [II] Sensor Selection [II]
Selection of a toroidal conductivity sensor connected to the MK-CONDI035N measuring module:
Module:
MK-CONDI
Mode:
Analog
Adjustable parameters for analog toroidal conductivity sensors
Parameter Setting [II] Analog CondI
Input Filter
Pulse Suppression
Enable/disable suppression of interference pulses.
Sensor Data
➜Sensor Data, p. 85
Sensor Type
Select the used sensor type: SE655, SE656, SE660, ISC40,
ISC40S, 5000 TC, other
If you select “Other”, please enter further sensor data.
Sensoface
Enable/disable display of Sensoface messages and icons.
Sensocheck
Monitoring of primary and secondary coils. Enable or dis-
able Sensocheck to generate failure or maintenance re-
quired messages.
Temperature Detection
Select temperature probe, set measuring and calibration
temperature.
Cal Presettings
Presetting of calibration mode with corresponding parameters.
➜Calibration Presettings, p. 86
TC Process Medium
Temperature compensation of process medium.
Concentration
➜Concentration (TAN Option FW-E009), p. 87
TDS
Enable/disable TDS function
(“Total Dissolved Solids” that influence conductivity)
USP
Enable/disable USP function for monitoring ultrapure water and set the USP limit.
➜USP Function, p. 87
Messages
Enable/disable messages for the separate process variables, or specify individual limits.
➜Messages, p. 88
Stratos Multi E401N
85
5.10.1 Sensor Data
Memosens sensors provide relevant sensor data automatically.
The sensor type must be selected if using analog sensors:
Parameter Setting [II] Analog … Sensor Data
Sensor Data
Sensor Type
Other
Back
F0031
Sensor Coding
II
Back to Meas.
On
Sensoface
Off
Sensocheck
1.980 /cm
Nom. Cell Factor
100.00
Transfer Ratio
01. Select sensor type
02. Enter the sensor coding, nominal cell factor, and transfer ratio.
03. In Temperature Detection , select the used temperature probe and whether the temperature is to
be measured automatically or manually during measurement and/or calibration.
Note: The sensor coding for unknown sensor types can be requested from Knick (see the back page of
this document for contact details)
Sensoface
The Sensoface icons provide the user with diagnostic information on wear and required maintenance
of the sensor. In measuring mode, an icon (happy, neutral, or sad smiley) is shown on the display to
reflect the continuous monitoring of the sensor parameters.
You can configure the current outputs so that a Sensoface message generates a 22-mA error signal.
Parameter Setting Inputs and Outputs Current Outputs Current Output I…
Behavior during Messages
The Sensoface message can also be output via a relay contact:
Parameter Setting Inputs and Outputs Relay Contacts Contact K… ➜Usage: Sensoface, p. 60
Sensoface monitors the toroidal conductivity sensor on the basis of the following parameters:
Cell factor, zero point, and, if Sensocheck is enabled: Primary/secondary coils and wires
With Memosens sensors, also: Number of CIP and SIP cycles compared to the specified
“Sensor Monitoring Details”.
Enabling/Disabling Sensoface
Sensoface is enabled and disabled in the Sensor Data submenu:
Parameter Setting [I] [II] [Sensor] Sensor Data
Note: After a calibration, a smiley is always displayed for confirmation, even if Sensoface is disabled.
Adjusting Sensor Monitoring
Note: This function is not available for analog sensors.
01. Open Sensor Data Sensor Monitoring Details .
02. Open a sensor parameter, e.g., Cell Constant .
03. Set Monitoring of the cell constant to automatic or individual.
04. If you select “Individual”: The nominal cell constant and the min/max limits can be entered.
Stratos Multi E401N
86
05. In the Message menu item, select whether and how an exceeded limit is to be displayed:
Off
No message.
Failure
A failure message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
Maintenance
A maintenance required message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
06. Set the sensor monitoring details for further sensor data, e.g., Sensocheck, sensor operating time,
or CIP/SIP counter.
07. Using the left softkey: Back, confirm the sensor monitoring settings and adjust further parameters.
or
Using the right softkey: Back to Meas., confirm the sensor monitoring settings and end the func-
tion check (HOLD).
CIP/SIP Counters
Note: This function is not available for all sensor types.
CIP/SIP cycles are used to clean or sterilize process-wetted parts in the process. Depending on the ap-
plication, either one chemical (alkaline solution, water) or several chemicals (alkaline solution, water,
acidic solution, water) are used.
• CIP temperature > 55 °C/131 °F
• SIP temperature > 115 °C/239 °F
The cleaning (cleaning in place) and sterilization (sterilization in place) cycles are counted to measure
the load on the sensor, e.g., in biotechnology applications.
Note: If measurements are generally taken at high temperatures (> 55 °C/131 °F), the counters should
be switched off.
When a CIP/SIP counter is switched on, a maximum number of cycles can be entered. As soon as the
counter has reached the specified value, a maintenance required message is triggered and the NAMUR
icon
is displayed.
Note: A CIP or SIP cycle is only entered into the logbook 2 hours after the start to ensure that the cycle
is complete.
Note: With Memosens sensors, an entry is also made in the sensor.
Settings can be changed in the Sensor Monitoring Details submenu:
Parameter Setting [I] [II] [Sensor] Sensor Data Sensor Monitoring Details
5.10.2 Calibration Presettings
The calibration presettings can be defined in the parameter settings or adjusted directly in the
Calibration menu prior to the calibration.
Calibration mode: Presetting of calibration mode, e.g., automatic, manual, product calibration, zero
point, installation factor, data entry, temperature
Further options are available depending on the calibration mode.
Automatic:
• Select calibration solution (NaCl/KCl)
Product calibration:
• Product calibration with/without temperature compensation
Stratos Multi E401N
87
Cal Presettings (Admin.)
Product Calibration
Back
Calibration Mode
Conductivity
Product
Wi
II
Conductivity
Automatic
Manual
Product
Zero Point
Data Entry
Calibration presettings can be changed in the Cal Presettings submenu:
Parameter Setting [I] [II] … Condl Cal Presettings
5.10.3 Concentration (TAN Option FW-E009)
With TAN option FW-E009, the substance concentration in percent by weight (wt%) can be deter-
mined for H2SO4, HNO3, HCl, NaOH, NaCl, and Oleum from the measured conductivity and temperature
values. A custom solution can also be specified.
The menu is only shown if the TAN option is activated.
Parameter Setting [I] [II] … Cond(I) Concentration
See also
➜Concentration Determination (FW-E009), p. 201
5.10.4 USP Function
Monitoring Ultrapure Water in the Pharmaceutical Industry
The conductivity of ultrapure water in the pharmaceutical industry can be monitored online in accor-
dance with the guideline “USP” (U.S. Pharmacopeia), Annex 5, Section 645 “Water Conductivity”. The
conductivity is measured without temperature compensation and is compared with limits. The water
is usable without any further testing when the conductivity is below the USP limit.
Configuring the USP Function
The USP value can be configured as a process variable USP% for output (display, current output, limit,
measurement recorder)
Settings can be changed in the USP submenu:
Parameter Setting [I] [II] … Cond(I) USP
Reduced limit: The USP limit can be reduced down to 10 %.
Monitoring: Select whether and how an exceeded limit is to be displayed.
Off
No message, but the parameter is still shown in the Diagnostics menu.
Failure
A failure message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
Mainte-
nance
A maintenance required message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
USP Function: Specifying a Relay Contact
The USP function can also be assigned to a relay contact:
Parameter Setting Inputs and Outputs Relay Contacts Contact K… ➜Usage: USP Output, p. 60
Display of the USP Function in the Diagnostics Menu
Diagnostics [I] [II] … Cond(I) USP Function
Display of the USP limit, the reduced limit, and conductivity.
Stratos Multi E401N
88
5.10.5 Messages
All values determined by the measuring module or sensor can generate messages.
Messages can be configured for the following process variables:
• Conductivity
• Resistivity
• Concentration (with TAN option FW-E009)
• Temperature
• Salinity
Configuring Messages
Process variable limits for the monitoring range can be selected in the Messages submenu:
Menu Selection Parameter Setting [I] [II] [Sensor] Messages Messages [Process Variable] Monitoring
• Max. Device Limits: Messages are generated when the process variable is outside the measuring
range. The “failure” or “out of specification” icons are shown; the corresponding relay contact is en-
abled. The current outputs can signal a 22-mA message (user-defined).
• Variable Limits: Upper and lower limits at which a message is generated can be defined for the
“failure” and “out of specification” messages.
Displaying Messages
Go to the Diagnostics menu if the “failure”, “maintenance required”, or “out of specification” icons are
flashing on the display.
Menu Selection Diagnostics Message List
All active messages are displayed in the “Message List” menu item with the following information: Er-
ror number, type (failure, maintenance required, out of specification), channel, message text.
You can scroll forward and backward using the up/down arrow keys.
The error message disappears from the display around 2 s after the error is cleared.
An overview of message texts with notes on troubleshooting can be found in the “Troubleshooting”
chapter. ➜Malfunction States, p. 136
Stratos Multi E401N
89
5.11 Dual Conductivity Measurement
Analog 2-Channel Conductivity Measurement with the MK-CC065N Measuring Module
The module can operate with two 2-electrode conductivity sensors, each with a Pt1000 temperature
probe.
Select the MK-CC065N measuring module:
Parameter Setting Sensor Selection [I] [II] Sensor Selection [II]
Module:
MK-CC
Mode:
Analog
Menu Parameter Setting [II] Analog Cond-Cond
Analog Cond-Cond (Admin.)
Back
Analog Cond
Back to Meas.
Analog Cond
II
IIA
IIB
Parameter setting ➜Process Variable Conductivity (Contacting), p. 77
2-Channel Conductivity Measurement with Memosens Sensors
A 2-channel conductivity measurement can also be performed with two Memosens sensors or one
Memosens and one analog sensor. In this case, a Memosens sensor is directly connected to the device,
a second conductivity sensor via the MK-COND025N or MK-MS095N modules.
Parameter setting ➜Process Variable Conductivity (Contacting), p. 76
Calculation Blocks (TAN Option FW-E020)
With TAN option FW-E020, “Calculation Blocks”, the measured conductivity values can be converted
into new variables. ➜Calculation Blocks (FW-E020), p. 206
5.12 Process Variable Oxygen
Note: Function check (HOLD) is active.
Note: After changing the process variable or measuring mode, Stratos Multi retains its settings but
needs to be reconfigured.
Note: Oxygen measurements in low oxygen concentrations require TAN option FW-E015.
Selecting a Memosens Oxygen Sensor
Parameter Setting Sensor Selection [I] [II] Sensor Selection [I]
Selection of the Memosens oxygen sensor connected to the RS-485 interface (terminals 1 … 5):
Parameter:
Auto or oxygen
Mode:
Memosens
Functionality:
Amperometric
Stratos Multi E401N
90
Selecting a Second Memosens Oxygen Sensor
Parameter Setting Sensor Selection [I] [II] Sensor Selection [II]
Selection of a second Memosens oxygen sensor connected to the MK-MS 095N measuring module:
Module:
MK-MS
Parameter:
Oxygen
Mode:
Memosens
Functionality:
Amperometric
Adjustable parameters for Memosens oxygen sensors Parameter Setting [I] [II] Memosens Oxy
Input Filter
Pulse Suppression
Enable/disable suppression of interference pulses.
Sensor Data
➜Sensor Data, p. 92
Measure in
Liquids, gases
Sensoface
Enable/disable display of Sensoface messages and icons.
Sensor Monitoring
Details
Option to enter individual limits for monitoring separate pa-
rameters.
Disable Sensocheck sensor monitoring or select whether
Sensocheck should generate failure or maintenance required
messages.
Option to enter individual limits before a message for response
time, sensor wear, sensor operating time, SIP counter is trig-
gered.
Cal Presettings
Presetting of the calibration mode and calibration timer.
➜Calibration Presettings, p. 95
Pressure Correction
Manual input of pressure during measurement and calibration.
With TAN option FW-E051: automatic pressure correction by external pressure transmit-
ter. ➜Pressure Correction, p. 95
Salinity Correction
Salinity, chlorinity, conductivity ➜Salinity Correction, p. 96
Messages
Enable/disable messages for the separate process variables, or specify individual limits.
➜Messages, p. 96
Selecting an SE740 Optical Oxygen Sensor (LDO)
Parameter Setting Sensor Selection [I] [II] Sensor Selection [I]
Selection of the SE740 optical oxygen sensor connected to the RS-485 interface (terminals 1 … 6):
Parameter:
Auto or oxygen
Mode:
Other digital
Functionality:
SE740
Note: The voltage at terminal 6 (Power Out) is automatically set to 14 V when using the SE740 optical
oxygen sensor. The Power Out menu item is not available.
Adjustable parameters for the SE740 optical oxygen sensor Parameter Setting [I] Digital Oxy
Input Filter
Pulse Suppression
Enable/disable suppression of interference pulses.
Sensor Data
➜Sensor Data, p. 92
Measure in
Liquids, gases
Sensoface
Enable/disable display of Sensoface messages and icons.
Sensor Monitoring
Details
Select whether a failure or maintenance required message is
triggered if the range of the Stern-Volmer constant and phase
angle is exceeded.
Disable Sensocheck sensor monitoring or select whether
Sensocheck should generate failure or maintenance required
messages.
Option to enter individual limits before a message for sensor
wear, sensor operating time, CIP counter, O2 measurement with
CIP/SIP, autoclaving counter is triggered.
Stratos Multi E401N
91
Adjustable parameters for the SE740 optical oxygen sensor Parameter Setting [I] Digital Oxy
Cal Presettings
Presetting of the calibration mode and calibration timer.
➜Calibration Presettings, p. 95
Pressure Correction
Manual input of pressure during measurement and calibration.
With TAN option FW-E051: automatic pressure correction by external pressure transmit-
ter. ➜Pressure Correction, p. 95
Salinity Correction
Salinity, chlorinity, conductivity ➜Salinity Correction, p. 96
Messages
Enable/disable messages for the separate process variables, or specify individual limits.
➜Messages, p. 96
Selecting an Analog Oxygen Sensor
Parameter Setting Sensor Selection [I] [II] Sensor Selection [II]
Selection of an oxygen sensor connected to the MK-OXY046N measuring module:
Module:
MK-OXY
Mode:
Analog
Adjustable parameters for analog oxygen sensor Parameter Setting [II] Analog Oxy
Input Filter
Pulse Suppression
Enable/disable suppression of interference pulses.
Sensor Data
➜Sensor Data, p. 92
Measure in
Liquids, gases
Sensor Type
Standard or other
Temperature Probe
NTC 22kΩ, NTC 30kΩ
Sensor Polarization
Automatic or individual
If you select “Individual”, separate values for polarization
can be entered during measurement and calibration.
Membrane Compensation
If “Other Sensor Type” is selected
Sensoface
Enable/disable display of Sensoface messages and icons.
Sensor Monitoring
Details
Option to enter individual limits for monitoring zero point
and slope.
Disable Sensocheck sensor monitoring or select whether
Sensocheck should generate failure or maintenance re-
quired messages.
Option to enter individual limits before a message for re-
sponse time is triggered.
Cal Presettings
Presetting of the calibration mode and calibration timer.
➜Calibration Presettings, p. 95
Pressure Correction
Manual input of pressure during measurement and calibration.
With TAN option FW-E051: automatic pressure correction by external pressure transmit-
ter. ➜Pressure Correction, p. 95
Salinity Correction
Salinity, chlorinity, conductivity ➜Salinity Correction, p. 96
Messages
Enable/disable messages for the separate process variables, or specify individual limits.
➜Messages, p. 96
Stratos Multi E401N
92
5.12.1 Sensor Data
Display Example for an Analog Oxygen Sensor
Sensor Data (Admin.)
Sensor Type
Back
Measure in
Liquids
Standard
I
Temperature Probe
Sensor Polarization
Polariz. During Meas
Polariz. During Cal
NTC 22kΩ
Auto
-675 mV
-675 mV
Liquids
Gases
01. Select whether to measure in liquids or gases.
02. If measuring in gases: Enter the relative humidity of the process medium.
03. If using an analog sensor: Select the sensor type and the used temperature probe.
04. If using an analog sensor: Select whether the polarization voltage should be selected automati-
cally or individually during measurement and calibration.
Note: The preset polarization voltage of -675 mV is appropriate for most measurements.
Settings can be changed in the Sensor Data submenu:
Parameter Setting [I] [II] ... Oxy Sensor Data
Sensoface
The Sensoface icons provide the user with diagnostic information on wear and required maintenance
of the sensor. In measuring mode, an icon (happy, neutral, or sad smiley) is shown on the display to re-
flect the continuous monitoring of the sensor parameters.
You can configure the current outputs so that a Sensoface message generates a 22-mA error signal.
Parameter Setting Inputs and Outputs Current Outputs Current Output I…
Behavior during Messages
The Sensoface message can also be output via a relay contact:
Parameter Setting Inputs and Outputs Relay Contacts Contact K… ➜Usage: Sensoface, p. 60
Sensoface monitors the oxygen sensor for slope, zero point, response time, and sensor wear. Senso-
face is displayed when Sensocheck has been enabled during parameter setting.
Enabling/Disabling Sensoface
Sensoface is enabled and disabled in the Sensor Data submenu:
Parameter Setting [I] [II] [Sensor] Sensor Data
Note: After a calibration, a smiley is always displayed for confirmation, even if Sensoface is disabled.
Stratos Multi E401N
93
Adjusting Sensor Monitoring
01. Open Sensor Data Sensor Monitoring Details .
02. Open a sensor parameter, e.g., Slope .
03. Set Monitoring of the slope to automatic or individual.
04. If you select “Individual”: The nominal slope and the min/max limits can be entered.
05. In the Message menu item, select whether and how an exceeded limit is to be displayed:
Off
No message, but the parameter is still shown on the sensor diagram.
Failure
A failure message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
Maintenance
A maintenance required message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
06. Set the sensor monitoring details for other sensor data, e.g. zero point, Sensocheck, response time,
sensor wear, or sensor operating time.
07. Using the left softkey: Back, confirm the sensor monitoring settings and adjust further parameters.
or
Using the right softkey: Back to Meas., confirm the sensor monitoring settings and end the func-
tion check (HOLD).
CIP/SIP Counters
Note: A CIP counter is available for the SE740 oxygen sensor.
An SIP counter is available for amperometric Memosens oxygen sensors.
CIP/SIP cycles are used to clean or sterilize process-wetted parts in the process. Depending on the
application, either one chemical (alkaline solution, water) or several chemicals (alkaline solution,
water, acidic solution, water) are used.
• CIP temperature > 55 °C/131 °F
• SIP temperature > 115 °C/239 °F
The cleaning (cleaning in place) and sterilization (sterilization in place) cycles are counted to measure
the load on the sensor, e.g., in biotechnology applications.
Note: If measurements are generally taken at high temperatures (> 55 °C/131 °F), the counters should
be switched off.
When a CIP/SIP counter is switched on, a maximum number of cycles can be entered. As soon as the
counter has reached the specified value, a maintenance required message is triggered and the NAMUR
icon
is displayed.
Note: A CIP or SIP cycle is only entered into the logbook 2 hours after the start to ensure that the cycle
is complete.
Note: With Memosens sensors, an entry is also made in the sensor.
Stratos Multi E401N
94
O2 Measurement with CIP/SIP
When using the SE740 optical oxygen sensor, the temperature can be monitored during the CIP/SIP
process. When the preset cut-off temperature is exceeded, the sensor automatically stops the oxygen
measurement to extend the life of the membrane. The measured oxygen value is frozen,
the temperature continues to be measured.
01. Open Sensor Monitoring Details O2 Measurement with CIP/SIP .
02. Monitoring: Automatic or individual
03. If you select “Individual”: Enter the maximum cut-off temperature.
04. In the Message menu item, select whether and how an exceeded limit is to be displayed:
Off
No message, but the parameter is still shown on the sensor diagram.
Failure
A failure message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
Maintenance
A maintenance required message is shown in off-limit conditions; the corresponding NAMUR icon
is displayed.
Settings can be changed in the Sensor Monitoring Details submenu:
Parameter Setting [I] [II] ... Oxy Sensor Data Sensor Monitoring Details
Autoclaving Counter
An autoclaving counter is available for the SE740 oxygen sensor.
When the autoclaving counter is switched on, a maximum number of cycles can be entered. As soon
as the autoclaving counter has reached the specified value, a maintenance required message is
triggered and the NAMUR icon
is displayed.
After each autoclaving process, you must manually increment the autoclaving counter in the
Maintenance menu.
Maintenance [I] Digital Oxy Sensor Monitor Autoclaving Cycles
Stratos Multi E401N
95
5.12.2 Calibration Presettings
The calibration presettings can be defined in the parameter settings or adjusted directly in the
Calibration menu prior to the calibration.
Calibration mode: Presetting of calibration mode, e.g., in air, in water, data entry, product calibration,
zero calibration, temperature
If you select “Product Calibration” mode, you must also select the measured value: Saturation %air,
concentration (gas), partial pressure
Calibration timer: When a preset calibration interval expires, the calibration timer generates a mes-
sage text to indicate the need for calibration. If “Auto” is selected, the interval is set to 720 h.
If “Individual” is selected, you can select a custom interval.
Cal Presettings (Admin.)
Product Calibration
Back
Calibration Mode
Calibration Timer
Product
Saturation
II
Sat. %Air
Conc. (Liquid)
Partial Pressure
Settings can be changed in the Cal Presettings submenu:
Parameter Setting [I] [II] … Oxy Cal Presettings
Note: If Sensoface is enabled, a neutral smiley is displayed once 80 % of the interval has expired. Once
the entire interval has expired, a sad smiley is shown, a maintenance required message is generated,
and the corresponding NAMUR icon
is displayed. If the current outputs are configured accordingly,
a 22-mA error signal is generated.
5.12.3 Pressure Correction
The pressure used for measurement or calibration can be manually specified (factory setting
1013 mbar).
With TAN option FW-E051, “Current Input”, an external pressure transmitter can be connected to the
current input (terminals 8 and 9). This enables automatic pressure correction. The start and end of the
current input can be set in the ranges from 0/4 ... 20 mA.
Settings can be changed in the Pressure Correction submenu:
Parameter Setting [I] [II] ... Oxy Pressure Correction
Setting Automatic Pressure Correction
(TAN Option FW-E051)
01. Open the Ext. Pressure Transmitter submenu.
02. Select “Absolute” or “Differential” pressure transmitter.
03. Select current input 0 … 20 mA or 4 … 20 mA.
04. Enter pressure values for current start and current end
05. Use the left softkey: Back to return to the Pressure Correction submenu.
06. Select an external or manual pressure correction in Pressure during Meas. and Pressure during Cal .
Stratos Multi E401N
96
5.12.4 Salinity Correction
The solubility of oxygen in water depends on its salinity. The correction is made by either directly
entering the salinity in g/kg, the chlorinity in g/kg, or the conductivity in μS/cm, and the temperature.
Settings can be changed in the Salinity Correction submenu:
Parameter Setting [I] [II] ... Oxy Salinity Correction
5.12.5 Messages
All values determined by the measuring module or sensor can generate messages.
Messages can be configured for the following process variables:
• Saturation %air
• Saturation %O2
• Concentration
• Partial pressure
• Temperature
• Process pressure
Configuring Messages
Process variable limits for the monitoring range can be selected in the Messages submenu:
Menu Selection Parameter Setting [I] [II] [Sensor] Messages Messages [Process Variable] Monitoring
• Max. Device Limits: Messages are generated when the process variable is outside the measuring
range. The “failure” or “out of specification” icons are shown; the corresponding relay contact is en-
abled. The current outputs can signal a 22-mA message (user-defined).
• Variable Limits: Upper and lower limits at which a message is generated can be defined for the
“failure” and “out of specification” messages.
Displaying Messages
Go to the Diagnostics menu if the “failure”, “maintenance required”, or “out of specification” icons are
flashing on the display.
Menu Selection Diagnostics Message List
All active messages are displayed in the “Message List” menu item with the following information: Er-
ror number, type (failure, maintenance required, out of specification), channel, message text.
You can scroll forward and backward using the up/down arrow keys.
The error message disappears from the display around 2 s after the error is cleared.
An overview of message texts with notes on troubleshooting can be found in the “Troubleshooting”
chapter. ➜Malfunction States, p. 136
Stratos Multi E401N
97
5.13 Flow
Stratos Multi can calculate flow for limit messages or to monitor an ion exchanger. A pulse generator is
connected to control input OK1 for this purpose.
Parameter Setting
Control input OK1 must first be assigned the “Flow” function.
01. Open System Control Function Control .
02. Input OK1: Select “Flow”.
03. 2x left softkey: Back
04. Open Inputs and Outputs Control Inputs Flow .
05. Enter the number of pulses per liter.
06. If required, enable monitoring of the minimum and maximum flow.
The flow measurement can process up to 100 pulses per second at the signal input of control input
OK1.
Flow monitoring with a connected external flow transmitter
Factory setting to generate a failure message
Minimum flow
5 liters/h
Maximum flow
25 liters/h
The flow messages can activate a relay contact and/or trigger a 22-mA message via a current output
(user-defined).
5.14 HART Communication (TAN Option FW-E050)
Note: In order to use the TAN option FW-E050 HART with Stratos Multi, current output I1 must be set
to 4 … 20 mA. HART communication is not possible below 4 mA.
See also
➜Current Outputs, p. 55
➜HART (FW-E050), p. 209
Stratos Multi E401N
98
6 Calibration/Adjustment
Menu Selection
Back
Calibration
Lingua
Note: During calibration, the function check (HOLD) operating state is active for each of the calibrated
measuring channels. The outputs respond as configured.
During calibration, Stratos Multi remains in calibration mode until stopped by personnel. When cali-
bration mode is exited, a confirmation prompt is displayed to ensure that the system is ready for oper-
ation again.
Adjustment
Adjustment involves the transfer of the calibration values obtained during calibration to Stratos Multi
or the digital sensor. These calibration values are entered in the adjustment record of the Stratos Multi
for the sensor, and directly in the sensor when using digital sensors:
Menu Selection Diagnostics I/II [Sensor] Cal/Adj Record [Process Variable]
NOTICE! Without adjustment, a measuring device will supply an imprecise or incorrect output value.
For it to measure properly, Stratos Multi must be adjusted. When using analog sensors, adjustment is
required following sensor replacement.
Assigning passcodes helps to ensure that only specialist personnel with access rights are allowed to
carry out calibrations and adjustments.
The passcodes can be changed or disabled:
Parameter Setting System Control Passcode Entry ➜Passcode Entry, p. 48
6.1 Memosens Calibration/Adjustment
Menu Selection Calibration [I] [II] Memosens …
Note: The calibration data are saved in the Memosens sensor. This means that Memosens sensors can
be cleaned, reconditioned, calibrated, and adjusted away from the measurement location, e.g. in a lab-
oratory. Sensors in the system are replaced on-site by adjusted sensors.
Stratos Multi E401N
99
6.2 Process Variable pH Calibration/Adjustment
Note: During calibration, the function check (HOLD) operating state is active for each of the calibrated
measuring channels. The outputs respond as configured.
• Calibration: Detect deviations without readjustment of calibration data
• Adjustment: Detect deviations with readjustment of calibration data
NOTICE! When using analog sensors, adjustment is required following sensor replacement.
pH Calibration/Adjustment Explanations
Each pH sensor has a separate zero point and a separate slope. Both values change as a result of aging
and wear. The voltage supplied by the pH sensor is corrected by Stratos Multi for the zero point and
the electrode slope of the pH sensor, and displayed as the pH value.
During calibration, the sensor’s deviation is initially determined (zero point, slope). The sensor is
immersed in buffer solutions with a precisely known pH value for this purpose. Stratos Multi measures
the voltages of the sensors and the temperature of the buffer solution, using this information to
calculate the sensor’s zero point and slope.
Calibration values determined during calibration
Zero Point
The pH value at which the pH sensor supplies the voltage 0 mV. The zero point is different for each
sensor and changes with age and wear.
Slope
The slope of a sensor is the voltage change per pH unit. With an ideal sensor, this is -59.2 mV/pH.
Temperature
The temperature of the process solution must be logged, since the pH measurement is tempera-
ture-dependent. Many sensors feature an integrated temperature probe.
There are limits that are calculated during calibration when monitoring glass and reference imped-
ances. The following limits apply to standard glass electrodes:
• Temperature range: 0 ... 80 °C/32 ... 176 °F
• Impedance range: 50 ... 250 MΩ (at 25 °C/77 °F)
Stratos Multi E401N
100
6.2.1 Calibration Methods
One-Point Calibration
The sensor is calibrated with just one buffer solution. A one-point calibration is useful and permissible
if the measured values are close to the sensor zero point, such that the change in sensor slope is of
negligible significance. The sensor’s zero point is subsequently adjusted, while the slope remains
unchanged.
Two-Point Calibration
The sensor is calibrated with two buffer solutions. This makes it possible to calculate the sensor’s zero
point and slope. The sensor’s zero point and slope are subsequently adjusted. A two-point calibration
is required in the following cases, for example:
• The sensor has been replaced
• The pH measured value covers a large range
• The pH measured value is far from the sensor zero point
• The pH value needs to be determined with high precision
• The sensor is subject to heavy wear
1
2
4
3
1 First point of first buffer solution
3 Recommended measuring range
2 Second point of second buffer solution
4 Result of ideal calibration at 25 °C/77 °F
Three-Point Calibration
The sensor is calibrated with three buffer solutions.
The zero point and slope are calculated using a line of best fit in accordance with DIN 19268. The sen-
sor’s zero point and slope are subsequently adjusted.
1
2
3
4
1 First point of first buffer solution
3 Third point of third buffer solution
2 Second point of second buffer solution
4 Rise