UPGRADING ETHERNET INTERFACE SOFTWARE

20 Planmeca Ethernet interface Installation manual 11. Wait at least a minute before disconnecting the power cable. The Flash chip in the Ethernet box will be programmed and this takes some time. It is safe to disconnect the power cable after link led goes off and on again. 12. Change the Computer’s IP settings back to normal. 13. If network cable connections were changed for the update, change them back to normal. ...

Planmeca Ethernet interface 1

INTRODUCTION Installation manual 1 INTRODUCTION The Planmeca Ethernet Interface is for use with the Planmeca Dixi3 digital intra X-ray system or with the Planmeca Proline XC X-ray with Dimax3 system and Planmeca Dimaxis Pro and Dimaxis Classic imaging softwares. This manual describes how to install and use the system. Please read this manual carefully before installing and using the system. Refer also to the Planmeca Intra and Planmeca Dixi3 digital intra X-ray unit’s and / or Planmeca Proline XC X- ray with Dimax3 user’s and installation manual. NOTE The Plan ...

INSTALLATION OF THE PLANMECA ETHERNET INTERFACE BOX TO THE PLANMECA PROLINE XC X-RAY

12 Planmeca Ethernet interface Installation manual The following window appears. Change your client name, Ethernet Interface IP address, Eth- ernet Interface port and Proline box port number. Confirm your selection by clicking “save” on the File pull-down menu. If a password has been set, it must include only capital letters. Do not use space or special characters. The Planmeca Ethernet Interface is now installed and you can enter the Dimaxis Pro or Dimaxis Classic program. NOTE Before installing the devi ...

UPGRADING ETHERNET INTERFACE SOFTWARE

18 Planmeca Ethernet interface Installation manual 4. Change the IP address to 10.1.0.x (for example 10.1.0.10) and the subnet mask to 255.255.255.0. Click OK. 5. a: Connect a Cross connected Cat. 5 UTP cable between the PC and the Ethernet box. OR b: Make sure the computer and the ethernet box are connected to same network. Before doing this ask the local network administrator that it is allowed to use these IP settings temporarily. 6. Open the Command Prompt from the Start/Accessories menu. 7. Go to the folder where you have the loader software and the ...

Planmeca Ethernet interface 21

TECHNICAL SPECIFICATIONS Installation manual 8 TECHNICAL SPECIFICATIONS Planmeca Ethernet interface Ethernet interface box dimensions 198 x 178 x 36 mm (7.80 x 7,00 x 1.42 in.) Ethernet cable RJ45 cable, max. 100 m (328 ft.) Electrical classification Class I, type BF Power supply Type: Mascot 9920/12/3617 Input voltage 100-240 VAC (50-60 Hz) Output voltage 12VDC...13.2VDC +-5% Max. output current> 1A Insulation voltage Primary-secondary4000 VAC / 5640 VDC Plug ø2.5mm ø5.5mm ...

UPGRADING ETHERNET INTERFACE SOFTWARE

20 Planmeca Ethernet interface Installation manual 11. Wait at least a minute before disconnecting the power cable. The Flash chip in the Ethernet box will be programmed and this takes some time. It is safe to disconnect the power cable after link led goes off and on again. 12. Change the Computer’s IP settings back to normal. 13. If network cable connections were changed for the update, change them back to normal. ...

Planmeca Ethernet interface 5

INSTALLATION OF THE PLANMECA ETHERNET INTERFACE BOX TO THE DIXI3 DIGITAL INTRA X-RAY Installation manual Open then the Internet Explorer, insert the Planmeca Ethernet Interface default IP address 192.168.0.135. Without this default IP Address you can not configurate your Ethernet Inter- face. In the appearing Ethernet Interface window select Configurations on your left-hand side. The following window enables you to change the factory default IP Address and TCP Port. Con- sult the local network administrator about the IP address settings you can use. After saving y ...

Planmeca Ethernet Interface user guide recommended for: KALEIDOLOOP, VSC-3, BMC58, AMI2-70, DAT 6000 SERIES.

⚠️ Download Safety

The Recording Equipment Installation manual (26 pages) is completely safe to download (last scan date: 19/03/2026).

Related PDF Instructions for Planmeca Ethernet Interface:

Recording Equipment Equipment: Frequently Requested

Have some Questions about Planmeca Ethernet Interface Device?

Get valuable and actionable advices from other Planmeca Ethernet Interface device users. Describe your problem or question, regarding the using of Ethernet Interface, in as much detail as possible, and other users will be able to answer you.