Desco TB-3048 Technical Bulletin
Text of the manual
TB-3048 March 2011 Page 1 of 5
Made in America
© 2010 DESCO INDUSTRIES INC.
Employee Owned
Digital Static Locator
Installation, Operation, and Maintenance
Description
Desco’s Digital Static Locator is a compact
electrostatic fieldmeter used for locating and measuring
static charges.
The Digital Static Locator measures static voltages
within ±20kV (20,000V) at a distance of 1". Results
are simultaneously displayed numerically and in bar
graph format. POWER on/off, ZERO adjustment, Ion
Balance (IB) and HOLD are all push button operation.
The HOLD button allows the display to retain the static
charge reading. This is especially useful where display
is difficult to see during measurement.
Two LED guide ring range lights help position the
fieldmeter at the right distance from a charged surface.
The conductive case and ground snap facilitate
grounding for accurate measurement. The circuitry of
the 19430 has been designed to make measurements
in areas using air ionization.
Packaging
1 Digital Static Locator
1 Plate Assembly
1 Ground Cord
1 Battery
Features
The 19430 has four push buttons. The buttons operate
by pressing once lightly. There is no need to press the
button repeatedly. Its effective life may be decreased by
repeated misuse.
Figure 1. Digital Static Locator Item #19430
TECHNICAL BULLETIN TB-3048
DESCO WEST - 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178 • Fax (909) 627-7449
DESCO EAST - One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Fax (781) 575-0172 • Web Site: Desco.com
The POWER button turns the Digital Static Locator on
and off. If the 19430 is left on for 5 minutes, it will
automatically turn itself off. To resume operations, press
the red POWER button once. To disable the AutoOff
feature and allow continuous operation, press the red
Power button and hold down for at least 3 seconds
when turning the Digital Static Locator on.
The HOLD button holds the display of a measured
charge. HOLD mode turns off the ring range lights. No
measurements are possible when in the HOLD mode.
The hold measurement feature enables the user to use
the fieldmeter in inaccessible areas. When the HOLD
button is pressed once during a measurement, the
digital numerical value and the bar graph will be held on
the display. This function allows the operator to move
the fieldmeter where it can be more easily read. The
HOLD mode is displayed above the battery capacity
display. To cancel HOLD mode and resume
measurement, press HOLD once more.
The IB button and prepares the fieldmeter for Ion
Balance (offset voltage) measurement.
The ZERO button sets the display to a zero static
charge. When the digital meter is turned on, the zero
adjustment is inoperative if it is in HOLD mode.
NOTE: For reliable measurement, the fieldmeter must
be grounded properly using the ground lead provided.
The plastic case of the 19430 Electrostatic Fieldmeter
is made of conductive resin. The grounding terminal
provides the reference potential for the electrical circuit.
This terminal must be grounded properly for accurate
measurement.
Operation For Static Charge Measurement
1. Turn the ring range lights and sensor toward a
surface to be evaluated.
2. Press the POWER button. A single beep for one
second indicates proper operation. The battery
capacity indicator and bar graph will be displayed.
3. The meter reading should be zero (0). Use the ZERO
button if it is not.
4. Hold the fieldmeter 1” away from the charged
surface. Two beams of light will form concentric
circles when the meter is the correct distance, 1”
away, from the surface. The measurement
accuracy is dependent on a stable ground reference
and the 1 inch measuring distance. It is also
dependent on the “aspect ratio”, relating the size of
the object to be measured to the measurement
distance. This aspect ratio should be at least 3 for
best accuracy, i.e. the object should be at least a
3 inch square when measuring at a 1 inch distance.
Accurate measurements may be made at other
measurement distances by scaling the meter range
and observing the proper aspect ratio. For example,
at a measurement distance of 3 inches, multiply the
meter reading by 3 to give a range of 0 to 60 kilovolts.
For accuracy, the object being measured at this
distance should be at least a 9 inch square.
5. A bar graph and a digital display on the 19430
displays the voltage of the charge. Press HOLD if you
want to hold the display of the measured charge.
6. To turn off meter: Press POWER if you have not
used the HOLD mode.
If you have pressed HOLD and
wish to retain the
meter reading for the next use,
press the POWER button to
turn off the 19430.
If you have pressed HOLD and
do not wish to retain the meter
reading for next use, press
HOLD then POWER.
If the 19430 is left on for five minutes, the unit will beep
five times then turn itself off automatically. To disable
this feature, simply press and hold the red power button
for at least three seconds when turning the 19430 on.
Three beeps and the A-OFF display indicates that the
feature has been disabled.
Results
A display of [+] or [-] next to the digital numeric value
represents the polarity of the static charge being
measured. In addition, the color of the bars in the bar
graph indicates the polarity of the static charge. Red
and blue bars indicate positive and negative charges
respectively.
During a static charge measurement, if the digital
display starts flashing along with a continuous beep,
stop taking the measurement. The static charge is
greater than the meter can measure. Attempts to
measure too high a charge may damage the sensor.
© 2010 DESCO INDUSTRIES INC.
Employee Owned
TB-3048 Page 2 of 5
DESCO WEST - 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178 • Fax (909) 627-7449
DESCO EAST - One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Fax (781) 575-0172 • Web Site: Desco.com
Operation For Ion Balance (Offset Voltage)
Measurement
NOTE: The Digital Static Locator Ion Balance
measurement feature should not be used to measure
the ion balance of static bars. It may be used to mea-
sure other ionizers for ion balance (offset voltage).
1. Depress the plate assembly release button on either
side of the fieldmeter to remove the charge plate
assembly.
2. Insert the plate assembly into the other end of the
fieldmeter (the end with the Desco logo) until it
clicks into place.
NOTE: Ion Balance plate assembly must not move
during measurement.
3. The conductive case must be grounded using the
supplied ground cord. It is imperative for accurate
ion balance and to drain any charge that has
developed prior to measurement.
4. To turn on the ion balance measuring circuit, press
the blue IB button.
5. Turn sensor toward an uncharged area.
6. The meter reading should be zero (0). Press the
Zero button if it is not.
7. Hold the fieldmeter centered in front of the ionizer at
any convenient, repeatable distance, such as 1 foot.
8. A bar graph and a digital display on the 19430
displays the ion balance information. Press HOLD if
you want to hold the display of the measured
charge.
9. To turn off the meter:
Press POWER if you have not used the HOLD
mode.
If you have pressed HOLD and wish to retain the
meter reading for the next use, press the POWER
button to turn off the 19430.
If you have pressed HOLD and do not wish to
retain the meter reading for next use, press HOLD
then POWER.
10. Depress the charge plate assembly release buttons
on either side of the fieldmeter to remove the charge
plate assembly. Store the plate on the opposite end,
pressing it until it clicks into place.
Figure 3.
© 2010 DESCO INDUSTRIES INC.
Employee Owned
TB-3048 Page 3 of 5
DESCO WEST - 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178 • Fax (909) 627-7449
DESCO EAST - One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Fax (781) 575-0172 • Web Site: Desco.com
If the 19430 is left on for five minutes, the unit will beep
five times then turn itself off automatically. This feature
can be disabled.
Troubleshooting Guide
Fault
Possible
Causes
Solution
Fieldmeter is on but
when turned toward a
charges object, the
number and the bar
graph do not change.
The meter is in HOLD
mode of operation.
Cancel the HOLD
mode by pressing the
HOLD button once.
The display cannot be
set to zero.
The meter is in HOLD
mode.
The true ZERO point
has shifted.
Cancel the HOLD
mode by pressing the
HOLD button once.
If the meter reads
0.20 or more for an
uncharged object, a
read adjustment
may be needed.
Contact your local
representative or
Desco Customer
Service for
adjustment, repair or
replacement.
When the POWER
button is pressed,
LCD display does not
appear.
No battery, low
battery capacity or
bad contact.
Malfunction of the
LCD display.
Replace the battery /
Make proper
connection.
Discontinue use and
contact your local
representative
or Desco Customer
Service for
adjustment, repair or
replacement.
A part of the LCD
display does not
appear.
Display shows
ERROR.
Alarm does not
sound.
Malfunction of the
LCD display.
Discontinue use and
contact your local
representative or
Desco Customer
Service for
adjustment, repair or
replacement.
LCD display
renewal rate: 5 times/s
Accuracy:
20KV ± 10 %
Ambient conditions: 50º-100ºF, (10º - 40ºC); 0 - 60
%RH (non-condensing)
Display features: Bar graph; red LCD positive
polarity voltage, blue LCD
negative polarity voltage
Bar graph
precision:
±0.1 kV for low range ±1.5 kV for
high (Hi) range; ±15 V for ion
balance range
Digital reading: Auto-ranging, three digits . = 0 to
±1.49 kV . = ±1.0 kV to
±20.0kV (Hi) = 0 to 200 (Ion
Balance)
Ion Balance (offset voltage) mode display: [IB] is
displayed
Battery capacity display indicator
HOLD switch to retain display after measurement
Error sign [Error] appears if the sensor is damaged
Alarm feature: Beep sound will be heard during
the following actions:
Power on: one beep
Power on with Auto Off feature disabled: [Depress
power button for more than 3 seconds] three beeps
Auto power-off: short beeps at 1 second interval for 5
seconds before power off
Over range: Continuous sound
Auto power-off: Power turns off automatically
after five minutes. [A.OFF] is
displayed whenever this feature
is disabled.
Power source: 9 V, 6F22Y manganese battery.
Battery life:
In excess of 30 hours.
Size:
4 7/8” (123mm) (L) × 2 7/8”
(73mm) (W) × 1” (25mm)(D)
Weight:
6 oz. (170g)
Case material: Conductive resin (ABS)
Specifications
Measuring range: 0 to ±1.49 kV (Lo range)
± 1.0 kV to ± 20.0 kV (Hi range)
(Range hysteresis: ± 1.0 kV to
± 1.5 kV)
0 to ± 200 V (Ion Balance
measuring range)
Measuring distance: 1"±1/64" (25mm ± 0.5mm); LEDs
guide for correct distance
(between charged surface and
fieldmeter)
Response time: <1 s
© 2010 DESCO INDUSTRIES INC.
Employee Owned
TB-3048 Page 4 of 5
DESCO WEST - 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178 • Fax (909) 627-7449
DESCO EAST - One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Fax (781) 575-0172 • Web Site: Desco.com
NOTE: The dark area in the battery indicator display is
not proportional to the remaining energy.
Replacement Procedure
The battery compartment on the back side of the
fieldmeter has a removable cover marked OPEN. Be
careful while connecting or disconnecting the battery.
To avoid breaking the contacts, do not pull the
connection plug hard.
1. Press down on OPEN and slide the cover down.
2. Remove the old battery (if present).
3. Connect the new battery and insert it into the battery
compartment.
4. Reattach the cover back to the battery compartment.
Before closing the cover, make sure that no parts of the
leads are outside the battery compartment.
Saftey Instructions
• DANGER indicates an immediately hazardous
situation which, if not avoided, will result in death or
serious injury.
• WARNING indicates a potentially hazardous situation
which, if not avoided, could result in death or serious
injury.
• CAUTION indicates a potentially hazardous situation
which, if not avoided, will result in property damage.
• NOTE indicates precautions necessary to avoid
potential equipment failure.
This equipment must be installed and maintained as
outlined in this manual. This equipment is not
constructed for classified (hazardous) environments. It
cannot be used where it will be exposed to ignitable or
corrosive materials.
The 19430 is intended for use in areas that are free of
water, oil, solvents and other conductive
contaminants. Exposure to such contaminants will
cause failure of the electrical insulation system in the
product. Formulation of dew on the fieldmeter must be
avoided.
Do not operate near corrosive fumes of acid/alkali or
corrosive gases such as chlorine. The 19430 must be
properly grounded to function accurately.
The 19430 contains a microcomputer chip. It should not
be used in an environment with excessive
electromagnetic noise. The internal parts of the 19430
should never come in contact with foreign
substances.
The 19430 has been calibrated for a measuring
distance of 1”± 1/64" (25mm ±0.5mm).
Important Notes and Cautions
1. Store and use the fieldmeter in a clean and dry
environment. Foreign substances should never enter
or touch the internal parts.
2. Vibration and impact may damage and/or affect the
sensitivity of measurement of the instrument.
3. Do not attempt to disassemble the fieldmeter.
4. The measuring distance between the fieldmeter and
a test object should be 1" ± 1/64" (25mm ± 1mm).
The calibration was carried out at this distance. The
specified accuracy is valid for this distance.
5. Grounding of the instrument is essential for accurate
ion balance measurements
6. To maintain the high insulation level of insulation of
the plate, the 19430 should be stored in a place
having less than 60 %RH. It should be kept inside a
plastic zipper bag with desiccant.
7. This fieldmeter is not an explosion-proof product. Do
not use this product in flammable atmosphere such
as paint-thinner, benzene, gasoline, propane, natural
gas etc.
8. The instrument should be re-calibrated once a year.
Contact Desco Customer Service when a
recalibration of the fieldmeter is necessary.
Fully Charged
Empty
Figure 2. Residual battery charge as displayed.
Battery Replacement
The Digital Static Locator uses 9 V, 6F22Y Manganese
battery. The life of the battery is about 30 hours. When
the POWER switch is on, a battery charge indicator
displays the state of the battery on the front panel.
© 2010 DESCO INDUSTRIES INC.
Employee Owned
TB-3048 Page 5 of 5
DESCO WEST - 3651 Walnut Avenue, Chino, CA 91710 • (909) 627-8178 • Fax (909) 627-7449
DESCO EAST - One Colgate Way, Canton, MA 02021-1407 • (781) 821-8370 • Fax (781) 575-0172 • Web Site: Desco.com
Limited Warranty
Desco expressly warrants that for a period of (1) year from the date of purchase an Desco Digital Static Locator will be free of defects in
material (parts) and workmanship (labor). Within the warranty period, a credit for purchase of a replacement Desco Digital Static Locator,
or, at Desco’s option, the Digital Static Locator will be repaired or replaced free of charge. Call our Customer Service Department at
909-664-9980 (Chino, CA) for a Return Material Authorization (RMA) and proper shipping instructions and address. Please include a copy
of your original packing slip, invoice, or other proof of date of purchase. Any unit under warranty should be shipped prepaid to the Desco
factory. Warranty replacements will take approximately two weeks.
If your unit is out of warranty, call our Customer Service Department at 909-664-9980 (Chino, CA) for a Return Material Authorization
(RMA) and proper shipping instructions and address. Desco will quote repair charges necessary to bring your unit up to factory standards.
Warranty Exclusions
THE FOREGOING EXPRESS WARRANTY IS MADE IN LIEU OF ALL OTHER PRODUCT WARRANTIES, EXPRESSED AND IMPLIED,
INCLUDING MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE WHICH ARE SPECIFICALLY DISCLAIMED. The
express warranty will not apply to defects or damage due to accidents, neglect, misuse, alterations, operator error, or failure to properly
maintain, clean or repair products.
Limit of Liability
In no event will Desco or any seller be responsible or liable for any injury, loss or damage, direct or consequential, arising out of the use
of or the inability to use the product. Before using, users shall determine the suitability of the product for their intended use, and users
assume all risk and liability whatsoever in connection therewith.