Using Source/Measure Unit as an Ammeter

B2900A Precision Source/Measure Unit Demo Guide

02 | Keysight | Using Source/Measure Unit as an Ammeter – Demo Guide

Introduction The Keysight B2900A Series Precision Source/Measure Unit (SMU) is a compact and cost-effective bench-top SMU with the capability to output and measure both voltage and current. The B2900A Series SMU enables you to make a wide range of current versus voltage (IV) measurements more accurately and quickly than ever before. In addition, the B2900A Series SMU comes with an intuitive graphical user interface (GUI) and free PC-based application software that make it easy for you to begin making productive measurements immediately. This demonstration guide shows how easily you can use the Keysight B2900A Series SMU as an ammeter. Required Instrument and Accessories All of the accessories required to perform the demos described in this demonstration guide are provided in a demo kit that is included with Keysight B2902A/12A demo units. The kit includes items such as a banana cable, a 1.1 kΩ resistor, etc. Keysight B2902A/12A Precision Source/Measure Unit 11059A Kelvin Probe Set 1.1 kΩ Resistor U8201A Combo Test Lead Kit

03 | Keysight | Using Source/Measure Unit as an Ammeter – Demo Guide

Figure 1. Connection diagram and basic condition Concept Figure 1 illustrates the connection diagram used in the demo to use the Keysight B2900A Series SMU as an ammeter. Since the current polarity of the SMU is opposite to an ammeter, the Low Force terminal of the channel has to be connected to the current output port of the device under test (DUT). In this demo, a 1.1 kΩ resistor combined with a power supply is used as a DUT, a current source circuit. The channel 2 of the B2902A/12A is used as a power supply. However, you can also use your own power supply for it. Setup 1. Connect the yellow banana plug of the 11059A to the Ch 1 Low Force Terminal. 2. Connect the red banana plug of the 11059A to the Ch1 High Force Terminal. 3. Clip the end of the 1.1 kΩ resistor with the black gold-plated tweezers of the 11059A. 4. Connect the black banana cable of the U8201A to the Ch2 Low Force Terminal. 5. Connect the red banana cable of the U8201A to the Ch2 High Force Terminal. 6. Clip the other end of the 1.1 kΩ resistor with the red alligator clip of the U8201A 7. Connect the red gold-plated tweezers of the 11059A to the black alligator clip of the U8201A. A - - V Common High Force Low Force DUT Current Out Low B2901/02/11/12A R 1.1 kΩ Power Supply DUT Current Out Low Red (11059A) Yellow (11059A) Black (11059A) Red (11059A) U8201A Channel 2 as Power Supply A - - V Common High Force Low Force B2901/02/11/12A R 1.1 kΩ Power Supply DUT Current Out Low Channel 1 DUT 1.1 kΩ

04 | Keysight | Using Source/Measure Unit as an Ammeter – Demo Guide

LAB1: Use Source/Measure Unit as an Ammeter Objective This demo illustrates the current mea- surement function as an ammeter by sourcing 0 V voltage and measuring the current from the resistor biased by a power supply using a Source/Measure Unit. Procedure 1. Configure the low terminal state to FLOATING 2. Measure current using the SMU (Channel 1) as an ammeter 3. Enable the DUT, the current soure circuit 4. Confirm measured current 5. (Optional) Configuring the measurement speed 6. (Optional) Configuring the measurement range operation Demonstration In the default setting, the low terminal of the channels in the Keysight B2900A Series SMU is grounded internally. However, the low terminal can be disconnected from the ground and kept floating. Configur- ing the low terminal state to FLOATING, which enables you to connect the low terminal to any potential up to ±250 V, and measuring current with sourcing 0 V from the channel, which makes it possible to use the channel as an ammeter. 1. Configure the low terminal state to FLOATING 1) If you aren’t on the top of the Function menu, press repeatedly to return to the top level. 2) Press , , and then press to open the Output Connection dialogue. 3) Press and select to specify the channel which the Low terminal state is configured for. On the top level of the Function menu In the middle level of the Function menu Any tree structure can’t be seen Some tree structure can be seen (1) Press Config (2) Press Source (3) Press Connection

05 | Keysight | Using Source/Measure Unit as an Ammeter – Demo Guide

4) Press and select , and then press to configure the Low terminal state to FLOATING. If the low terminal state of the channel is set to FLOATING, you can see the status indicator on the GUI as below, although no indicator can be seen on being set to GROUNDED. 2. Measure current using the SMU (Channel 1) as a ammeter 2-1. Change View mode to Ch1 Single View 1) Press repeatedly until the Channel 1 Single View is displayed 2-2. Configure the condition to source and measure 1) Press to edit the Source function, and then select to set the Source function to Voltage source. GROUNDED FLOATING

06 | Keysight | Using Source/Measure Unit as an Ammeter – Demo Guide

2) Press to edit the Source value, and then enter 0 V to set the Source value to 0 V. 3) Press to edit the Limit value, and then enter 100 mA to set the Limit value to 100 mA for example. 4) Press to configure the Measurement parameter, and then select to set the Measurement parameter to Current. 5) Rotate and press to edit the Voltage source range operation. Then Select to set the Voltage source range operation to FIXED. 6) Rotate and press to edit the Voltage source range, and then select to set it to 200 mV.

07 | Keysight | Using Source/Measure Unit as an Ammeter – Demo Guide

2-3. Perform the measurement 1) Press for the channel 1 to switch on its output terminal. 2) Press to perform a measurement repeatedly. Now you can see the measurement result on the GUI of the B2902/12A as below. 3. Enable the DUT, the current source circuit You can also use your own power supply instead of the channel 2 of the B2902A/12A 3-1. Change View mode to Ch2 Single View. 1) Press repeatedly until the Channel 2 Single View is displayed. 3-2. Configure the condition of the power supply (Channel 2) 1) Press to edit the Source function, and then select to set the Source function to Voltage source. 2) Press to edit the Source value, and then enter 1.1 V to set the Source value to 1.1 V for example. 3) Press to edit the Limit value, and then enter 100 mA to set the Limit value to 100 mA for example.

08 | Keysight | Using Source/Measure Unit as an Ammeter – Demo Guide

4) Press for the channel 2 to switch on its output terminal and enable the DUT. 4. Confirm measured current 4-1. Change View mode to Ch1 Single View. 1) Press repeatedly until Channel 1 Single View is displayed. You can see about 1 mA as measured value, since 1 V is sourced to a 1.1 kΩ resistor. If you press for the channel 2 to switch it off, you can see only the offset current because the DUT, the current source circuit is disabled. Theoretically speaking, the measured current should be 1 mA since 1.1 V is sourced to a 1.1 kΩ resistor. However, it may be varied because the resistor has some error on its value actually. Power Supply enabled Power Supply disabled

09 | Keysight | Using Source/Measure Unit as an Ammeter – Demo Guide

5. (Optional) Configuring the measurement speed In the default setting, the instrument selects the appropriate measurement speed and range automatically to get the fine accuracy. However, you can also specify these parameters on the GUI of the B2900A Series SMU to meet a variety of the requirement to the measurement conditions. For example, let’s try to change the measurement speed to NORMAL to make a mea- surement more carefully. If you select NORMAL, the aperture time is set to 1 PLC. Here, PLC stands for power line cycle and the specified number of power line cycles is used per a measurement. 1) Press to edit the Measurement speed, and then select to set the Measurement speed to NORMAL. (If you can’t see in Assist keys, press to change the keys shown in Assist keys.) 6. (Optional) Configuring the measurement range operation The parameters which configure the measurement range operation can be displayed in the Range Sub-panel in the Channel 1 Single View. In the default setting, the B2900A Series SMU performs the current measurement using a 1 μA current minimum measure- ment range with AUTO range operation. With AUTO range operation, the B2900A Series SMU selects the proper range for the measurement with the specified minimum mea- surement range so that you don’t need to take care about it. To know how to change the measurement range setting, try to configure to use the 10 μA current minimum mea- surement range with AUTO range operation. 1) Rotate and press to edit the Current minimum measurement range and then select to set it to 10 μA. If you’d like to fix the measurement range, you can select FIXED range operation as below. 2) Rotate and press to edit the Current measurement range operation. Then Select to set the Current measurement range operation to FIXED.

10 | Keysight | Using Source/Measure Unit as an Ammeter – Demo Guide

B2900 Precision Instrument Family The B2900 family contains products that perform both precision sourcing and precision measurement. www.keysight.com/find/b2900a Conclusion The Keysight B2900A Series Precision Source/Measure Unit (SMU) is a compact and cost-effective bench-top SMU with the capability to output and measure both voltage and current. Although it has the capability to make a wide range of current versus voltage (IV) measurements as its intrinsic function, the B2900A Series SMU can be used as an ammeter easily.

For more information on Keysight

Technologies’ products, applications or services, please contact your local Keysight office. The complete list is available at: www.keysight.com/find/contactus Americas Canada (877) 894 4414 Brazil 55 11 3351 7010 Mexico 001 800 254 2440 United States (800) 829 4444 Asia Pacific Australia 1 800 629 485 China 800 810 0189 Hong Kong 800 938 693 India 1 800 11 2626 Japan 0120 (421) 345 Korea 080 769 0800 Malaysia 1 800 888 848 Singapore 1 800 375 8100 Taiwan 0800 047 866 Other AP Countries (65) 6375 8100 Europe & Middle East Austria 0800 001122 Belgium 0800 58580 Finland 0800 523252 France 0805 980333 Germany 0800 6270999 Ireland 1800 832700 Israel 1 809 343051 Italy 800 599100 Luxembourg +32 800 58580 Netherlands 0800 0233200 Russia 8800 5009286 Spain 800 000154 Sweden 0200 882255 Switzerland 0800 805353 Opt. 1 (DE) Opt. 2 (FR) Opt. 3 (IT) United Kingdom 0800 0260637 For other unlisted countries: www.keysight.com/find/contactus (BP-9-7-17) DEKRA Certified ISO9001 Quality Management System www.keysight.com/go/quality Keysight Technologies, Inc. DEKRA Certified ISO 9001:2015 Quality Management System Evolving Since 1939 Our unique combination of hardware, software, services, and people can help you reach your next breakthrough. We are unlocking the future of technology. From Hewlett-Packard to Agilent to Keysight. myKeysight www.keysight.com/find/mykeysight A personalized view into the information most relevant to you. www.keysight.com/find/emt_product_registration Register your products to get up-to-date product information and find warranty information. Keysight Services www.keysight.com/find/service Keysight Services can help from acquisition to renewal across your instrument’s lifecycle. Our comprehensive service offerings—one- stop calibration, repair, asset management, technology refresh, consulting, training and more—helps you improve product quality and lower costs. Keysight Assurance Plans www.keysight.com/find/AssurancePlans Up to ten years of protection and no budgetary surprises to ensure your instruments are operating to specification, so you can rely on accurate measurements. Keysight Channel Partners www.keysight.com/find/channelpartners Get the best of both worlds: Keysight’s measurement expertise and product breadth, combined with channel partner convenience. This information is subject to change without notice. © Keysight Technologies, 2017 Published in USA, December 1, 2017 5992-1710EN www.keysight.com 11 | Keysight | Using Source/Measure Unit as an Ammeter – Demo Guide www.keysight.com/find/b2900a