Models RH300 and RH305 kit

Additional User Manual Translations available at www.extech.com

Extech Instruments RH300 Thank you for selecting the Extech RH300 Digital Psychrometer. This device measures Relative

Humidity, Air Temperature from the internal sensor (T1), and Temperature from the TP890 optional external probe (T2). The RH300 calculates Dew Point and Wet Bulb temperature based on T1 air temperature measurements and relative humidity measurements. The meter also calculates T1 minus T2 temperature and T2 minus Dew Point tem

Press for 1 second to power ON or OFF

Short press to freeze/unfreeze displayed readings Short press to step through T1 internal sensor temperature, Dew Point temperature, and T1 minus T2 (external probe) temperature Short press to step through T2 external probe (TP890) temperature, T2 temperature minus Dew Point temperature, and Wet Bulb temperature Short press to toggle temperature units (oC/oF) Short presses

Sensor Protective Cap

Turn the protective cap (top of meter), in the direction of the OPEN arrow to allow air into the sensor cavity. Turn the cap in the direction of the CLOSE arrow to protect the sensors. Always open the cap to take measurements and close when storing. Meter Power Press the power button for one second to switch the meter ON/OFF. The meter will perform a short self-test when switched ON. Air Temperatur

Extech Instruments RH300 Minimum (MIN) Maximum (MAX) Recording

Short press the MnMx/Reset button to view the minimum temperature and humidity readings recorded (MIN is shown). Short press the MnMx/Reset button again to view the maximum temperature and humidity readings recorded (MAX icon is shown). To exit the MIN/MAX mode, press the MnMx/Reset button again (MIN and MAX icons switch off). To clear the MIN/MAX memory, long press the MnMx/Reset button unti

Extech Instruments RH300 For the following verification and calibration procedures, you will need the 33% and 75% RH

reference bottles (supplied with the RH305 kit). To obtain reference bottles please contact Extech (www.extech.com). If the salt in the reference bottles is overly dry, please replace the bottles (RH300-CAL). Accuracy Verification Checking the 33% or 75% RH Calibration: 1. Insert the meter sensor into the 33% or 75% salt refere

Cleaning and storage

1. Clean the meter housing with the sensor cap closed, using a lightly damp cloth and mild detergent when necessary. Do not use solvents or abrasives. 2. Store the meter -- with the protective cap closed -- in an area with moderate temperature and humidity (refer to the operating/storage conditions specifications). Battery Replacement When the batteries need replacing, the symbol will appear on t

Range and Resolution Accuracy

Humidity 0.0 to 100.0% RH ±3% RH (10 to 90%) @ 23oC Temperature (internal sensor) -20 to 50oC (-4.0 to 122.0oF) ±1oC (±1.8oF) Temperature (optional external probe TP890) -20 to 70oC (-4.0 to 158.0oF) ±1oC (±1.8oF) Display Triple-reading LCD Sensor Types Relative Humidity: Precision capacitance sensor Temperature (internal): Thermistor Optional TP890 Temperature probe: Thermistor R

Extech Instruments Measuring Instruments Owners Instructions: