Text of Extech Instruments RH300 Operation & User’s Manual
Models RH300 and RH305 kit
Additional User Manual Translations available at www.extech.com
Extech Instruments RH300 Thank you for selecting the Extech RH300 Digital Psychrometer. This device measures Relative
Humidity, Air Temperature from the internal sensor (T1), and Temperature from the TP890
optional external probe (T2).
The RH300 calculates Dew Point and Wet Bulb temperature based on T1 air temperature
measurements and relative humidity measurements. The meter also calculates T1 minus T2
temperature and T2 minus Dew Point tem
Press for 1 second to power ON or OFF
Short press to freeze/unfreeze displayed readings
Short press to step through T1 internal sensor temperature, Dew Point
temperature, and T1 minus T2 (external probe) temperature
Short press to step through T2 external probe (TP890) temperature, T2
temperature minus Dew Point temperature, and Wet Bulb temperature
Short press to toggle temperature units (oC/oF)
Short presses
Sensor Protective Cap
Turn the protective cap (top of meter), in the direction
of the OPEN arrow to allow air into the sensor cavity.
Turn the cap in the direction of the CLOSE arrow to
protect the sensors. Always open the cap to take
measurements and close when storing.
Meter Power
Press the power button for one second to switch the meter ON/OFF. The meter will
perform a short self-test when switched ON.
Air Temperatur
Extech Instruments RH300 Minimum (MIN) Maximum (MAX) Recording
Short press the MnMx/Reset button to view the minimum temperature and humidity
readings recorded (MIN is shown). Short press the MnMx/Reset button again to view
the maximum temperature and humidity readings recorded (MAX icon is shown). To
exit the MIN/MAX mode, press the MnMx/Reset button again (MIN and MAX icons
switch off).
To clear the MIN/MAX memory, long press the MnMx/Reset button unti
Extech Instruments RH300 For the following verification and calibration procedures, you will need the 33% and 75% RH
reference bottles (supplied with the RH305 kit). To obtain reference bottles please contact Extech
(www.extech.com). If the salt in the reference bottles is overly dry, please replace the bottles
(RH300-CAL).
Accuracy Verification
Checking the 33% or 75% RH Calibration:
1. Insert the meter sensor into the 33% or 75% salt refere
Cleaning and storage
1. Clean the meter housing with the sensor cap closed, using a lightly damp cloth and mild
detergent when necessary. Do not use solvents or abrasives.
2. Store the meter -- with the protective cap closed -- in an area with moderate temperature
and humidity (refer to the operating/storage conditions specifications).
Battery Replacement
When the batteries need replacing, the
symbol will appear on t
Range and Resolution Accuracy
Humidity
0.0 to 100.0% RH
±3% RH (10 to 90%) @ 23oC
Temperature (internal sensor)
-20 to 50oC (-4.0 to
122.0oF)
±1oC (±1.8oF)
Temperature (optional external
probe TP890)
-20 to 70oC (-4.0 to
158.0oF)
±1oC (±1.8oF)
Display
Triple-reading LCD
Sensor Types
Relative Humidity: Precision capacitance sensor
Temperature (internal): Thermistor
Optional TP890 Temperature probe: Thermistor
R