Model 407114 Version 2.0 November 2005
2
Introduction
Congratulations on your purchase of the Extech CFM Thermo Anemometer.
This instrument measures Air Velocity, Air Flow (volume) and Temperature.
The RS-232 PC interface provides data acquisition functionality. Proper use of
this meter will provide years of reliable service.
Model 407114 Version 2.0 November 2005 3
Specifications
General Specifications
Circuit
Custom LSI microprocessor circuit
Display
Dual function 0.5" (13 mm) 4-digit LCD
Measurement units
Air Velocity: m/s, km/h, ft/min, knots, mph;
Air Flow: CMM (m3/min); CFM (ft3/min); Temp: °C & °F
Data hold
Freezes displayed reading
Sampling rate
1 reading per second approx.
Sensors
Air velocity/flow sensor: Conventional angled vane arms
with low-friction ball bearing.
Model 407114 Version 2.0 November 2005 4
Meter Description
1. Sensor plug. Can be inserted in the
sensor jack from only one direction.
Ensure proper orientation before
exerting force to plug in the sensor.
2. PC Interface jack (RS-232). For
use with the interface cable
provided in the optional 407001 PC
Data Acquisition kit.
3. LCD Display for Air Velocity, Air
Flow, and Temperature. Units of
measure and user alerts are also
displayed on the LCD.
4. Vane
Model 407114 Version 2.0 November 2005 5
Operation
Air Velocity Measurements
1. Connect the sensor to the sensor input jack on top of the meter.
2. Turn on the meter using the POWER button.
3. Select the VELOCITY function using the VEL/FLOW button. The LCD will display VEL
when the velocity mode is selected.
4. Select the desired temperature units using the C/F select button. The LCD will reflect
the current unit selection.
5. Select the desired air veloci
Model 407114 Version 2.0 November 2005 6
Air Flow Measurements (CMM / CFM)
1. Connect the sensor to the sensor input jack on top of the meter.
2. Turn on the meter using the Power button.
3. Select the FLOW mode using the FLOW/VELOCITY button. The LCD will display
FLOW when the Flow function has been properly selected.
4. Select the desired air flow units: CMM (cubic meters per minute) or CFM (cubic feet
per minute) using the UNIT button. The LCD will refl
Model 407114 Version 2.0 November 2005 7
Support line (781) 890-7440
Technical support: Extension 200; E-mail: [email protected]
Repair & Returns: Extension 210; E-mail: [email protected]
Product specifications subject to change without notice
For the latest version of this User’s Guide, Software updates, and other
up-to-the-minute product information, visit our website: www.extech.com
Extech Instruments Corporation, 285 Bear Hill Rd., Waltham, MA 02451
Model 407114 Version 2.0 November 2005 8
Useful Equations and Conversions
Area equation for rectangular or square ducts
Area equation for circular ducts
Cubic equations
NOTE: Measurements made in inches
must be converted to feet or meters before using the above formulae.
Unit of Measure Conversion Table
m/s
ft/min
knots
km/h
MPH
1 m/s
1
196.87
1.944
3.6
2.24
1 ft/min
0.00508
1
0.00987
0.01829
0.01138
1 knot
0.5144
101.27
1
1.8519
1.1523
1 km/h
0.2778
54.69