Haug EN SL MT Series Operating Instructions Manual
Read Online
Read the Haug EN SL MT Series Operating Instructions Manual online using the built-in PDF viewer.
Text of the manual
Table of contents
1
Notes on operating instructions ................................................................................. 5
1.1
Pictorial markings used ................................................................................................. 5
2
Safety ............................................................................................................................ 6
2.1
Intended use .................................................................................................................. 6
2.2
Danger sources ............................................................................................................. 7
2.3
Installer qualifications .................................................................................................... 7
2.4
Operator qualifications .................................................................................................. 7
3
Description of unit ....................................................................................................... 8
4
Installation .................................................................................................................... 9
4.1
Important installation instructions ................................................................................. 9
4.2
Setting up, connecting ................................................................................................ 10
5
Application ................................................................................................................. 11
5.1
Putting into operation .................................................................................................. 11
6
Remedy of defects ..................................................................................................... 12
6.1
Troubleshooting .......................................................................................................... 13
7
Maintenance and repairs ........................................................................................... 14
7.1
Replacing fuse ............................................................................................................. 14
7.2
Accessories .................................................................................................................. 14
8
Technical data ............................................................................................................ 15
8.1
Characteristics and specification ................................................................................ 15
8.2
Supply voltage ............................................................................................................. 15
8.3
Ambient conditions ...................................................................................................... 16
8.4
Housing ........................................................................................................................ 17
9
Disposal ..................................................................................................................... 18
Notes on operating instructions
Page 5
Keep in a safe place for future reference!
Types:
EN SL (MT) 230 V
01.7906.005
EN SL (MT) 115 V
01.7907.005
EN SL (MT) 100 V
01.7907.055
1
Notes on operating instructions
In these operating instructions, the power pack EN SL (MT) is also referred to
as "unit".
1.1
Pictorial markings used
In these operating instructions
NOT FOR USE by persons with pacemaker!
WARNING!
High voltage!
Danger of fatal accidents!
Do not open unit!
WARNING!
Only plug in/unplug coaxial connector when the unit is switched off!
CAUTION!
Important instructions!
On the unit
WARNING!
High voltage!
Danger of fatal accidents!
Do not open unit!
WARNING!
Only plug in/unplug coaxial connector when the unit is switched off!
NOT FOR USE!
Persons wearing heart pacemakers must maintain a safety distance of more
than 50 cm from the ionizing unit!
Make sure that you read and observe the operating instructions of the
connected ionizing units!
_____________________________________________________________________________________
The unit is operationally safe, provided that it is operated in accordance with
its intended use.
In case of misuse, dangers may result:
- for life and limb of the operator,
- for the unit and other assets.
Also note Chapter 4.1 (Important installation notes, page 9).
2.1
Intended use
CAUTION!
Do not install or use the unit in areas subject to explosion hazards!
The unit is intended exclusively for the high-voltage supply of HAUG ionizing
units. It generates an alternating high tension of approx. 7 – 8 kV. It is
intended, in connection with an ionizing unit, for the removal of electrostatic
charges from, for example, glass, paper, plastics etc.
For reasons of safety, unauthorized conversions and modifications of the unit
are not permitted.
The installation and operating conditions indicated in these Operating
Instructions must be adhered to.
Danger of fatal accidents!
Do not open unit!
WARNING!
Only plug in/unplug coaxial connector when the unit is switched off!
____________________________________________________________________________________
Defective high-voltage terminals and cables may lead to danger of electric
shocks. Shut down the unit immediately in case of visible damage and
suspected electrical defects.
2.3
Installer qualifications
The unit may be installed by trained electricians only. The above mentioned
person must have read the operating instructions and must follow the
instructions, notes and safety advice.
2.4
Operator qualifications
The unit may be maintained and put into operation by trained electricians
only or by authorized persons informed about the potential dangers. The
above mentioned persons must have read the operating instructions and
must follow the instructions, notes and safety advice.
Description of unit
Page 8
3
Description of unit
Figure 1
1.
Mains switch: Switch lights up green when the unit is switched on.
2.
Ground connection
Figure 2
3.
High-voltage terminals
4.
Check fuse (for replacement, refer to Chapter 7.1, page 14)
5.
Connection to supply voltage
1
2
3
4
5
The unit may be installed by trained electricians only. The above mentioned
person must have read the operating instructions and must follow the
instructions, notes and safety advice.
4.1
Important installation instructions
____________________________________________________________________________________
WARNING!
High voltage!
Danger of fatal accidents!
Do not open unit!
WARNING!
Only plug in/unplug coaxial connector when the unit is switched off!
____________________________________________________________________________________
The operation of the unit is not affected by the position in which it is installed.
However, we recommend installing the unit so that the high-voltage terminals
points downwards (to protect it from humidity, oil and dirt).
Do not place the unit on a surface generating or radiating heat. Avoid
installation positions exposed to direct sunlight.
Setting up, connecting
_____________________________________________________________________________________
WARNING!
High voltage!
Danger of fatal accidents!
Do not open unit!
WARNING!
Only plug in/unplug coaxial connector when the unit is switched off!
_____________________________________________________________________________________
1.
Before connecting always check that the unit is suitable for the local mains
voltage (the voltage is indicated on the name plate). Incorrect mains voltage
may result in damage to the unit.
2.
Set up unit at the required site of usage (if required, attach using the enclosed
holders).
3.
Ensure that the unit is switched off (for mains switch, refer to Figure 1, item 1,
page 8).
4.
Connect ionizing units to high-voltage terminals.
5.
Always connect unit to a socket with functioning ground wire. Always use the
supplied mains cable with its integrated ground wire.
6.
Put unit into operation.
The unit may be put into operation by trained electricians only or by persons
instructed in the potential dangers. The above mentioned persons must have
read the operating instructions and must follow the instructions, notes and
safety advice.
The unit is intended exclusively for the high-voltage supply of HAUG ionizing
units. It generates an alternating high tension of approx. 7 – 8 kV. It is
intended, in connection with an ionizing unit, for the removal of electrostatic
charges from, for example, glass, paper, plastics etc.
5.1
Putting into operation
Preconditions:
The power pack and the ionizing unit must be connected correctly.
1.
Switch on unit using the mains switch (refer to Figure 1, item 1, page 8).
Remedy of defects
Page 12
6
Remedy of defects
_____________________________________________________________________________________
WARNING!
High voltage!
Danger of fatal accidents!
Do not open unit!
WARNING!
Only plug in/unplug coaxial connector when the unit is switched off!
_____________________________________________________________________________________
CAUTION!
The units include a thermal cut out.
If the unit overheats at the ionizing unit due to short circuiting, the thermal cut
out will switch off the unit. The power pack can only be put back into
operation once the short-circuit has been removed and the unit has cooled
down (approx. 15 minutes)!
Any remedy of defects must be carried out by trained electricians only. The
above mentioned person must have read the operating instructions and must
follow the instructions, notes and safety advice.
In case of defects regarding the power pack and the ionizing unit, please
check for correct installation and fusing first (for replacement, refer to
Chapter 7.1 page 14).
Remedy of defects
Page 13
6.1
Troubleshooting
Faults
Measures
No ionization
Check mains voltage
Check fuse (for replacement, refer to
Chapter 7.1, page 14)
Check connection
Clean ionizing unit
Check ionizing unit for damage. If damaged,
immediately shut down and secure against
restarting.
If this does not remedy the fault, please return the unit with the ionizing unit to
HAUG for checking (see reverse).
Maintenance and repairs
Page 14
7
Maintenance and repairs
_____________________________________________________________________________________
WARNING!
High voltage!
Danger of fatal accidents!
Do not open unit!
_____________________________________________________________________________________
This unit does not include any parts which can be maintained or repaired by
the operator. HAUG only is authorized to repair or calibrate the unit.
Should the unit prove defective or if a defect is suspected, switch off unit
immediately and secure against subsequent reuse.
7.1
Replacing fuse
1.
Switch off unit.
2.
Determine and remove the cause for the blown fuse.
3.
Detach the fuse holder using a screwdriver and lift out.
4.
Replace fuse and reattach fuse holder.
Use the following fuses only:
Unit type
Fuse
EN SL (MT) 100 / 115 V
T 500 mA L 250; 5x20 mm
EN SL (MT) 230 V
T 250 mA L 250; 5x20 mm
The unit type and the rated voltage are indicated on the nameplate.
Only use fuses of the type indicated.
7.2
Accessories
Article
Order
number
One Point Ionizer
03.8510.200
Punkt – Elektrode XP
03.8513.200
Punkt – Elektrode XP (Set)
23.8513.100
U – Elektrode
05.8091.100
U – Elektrode (Set)
23.8091.100
Characteristics and specification
Reference temperature 23 °C
High-voltage terminals
2 HAUG High-voltage
terminals
High voltage
U = approx. 7 - 8 kV
Short-circuit current
Ik ≤ 5 mA
Cannot be used in pulsed mode
8.2
Supply voltage
CAUTION!
Always connect the PE conductor (green/yellow conductor) to the
protective earth of the mains!
Unit type
Nominal
value
Operating
range
Frequency
range
Power input
EN SL (MT)
100 VAC
±10 %
50 - 60 Hz
Pmax = 40 VA
EN SL (MT)
115 VAC
±10 %
50 - 60 Hz
Pmax = 40 VA
EN SL (MT)
230 VAC
±10 %
50 - 60 Hz
Pmax = 40 VA
Rated application range
+5 °C to +45 °C
Extreme range for storage and transport
-15 °C to +60 °C
Humidity:
Rated application range
Air pressure:
Rated application range
800 mbar to 1060 mbar
Vibrations:
Extreme range for storage and transport
max. 1.5 g (10 to 55 Hz), 1 h
Shock
max. 15 g in each direction
10
20
30
40
50
60
70
80
90
%RF
5 10 15 20 25 30 35 40 °C
Degree of protection
IP 20
Protection class
I
Overvoltage category
II
Pollution severity
2
Connection to supply voltage
Mains cable via socket for
non-heating apparatus
Dimensions:
Height
approx. 170 mm
Width
approx. 110 mm
Depth
approx 100 mm
Weight:
approx. 3,5 kg
Observe and maintain national and regional waste disposal regulations for
the disposal of the unit!
D – 0282
V 08-1
16.09.16
EN SL-MT v08-1gb.doc
made by
HAUG GmbH & Co.KG
Friedrich-List-Straße 18
D-70771 Leinfelden-Echterdingen
Telefon 07 11 / 94 98 - 0
Telefax 07 11 / 94 98 - 298
www.haug.de
E-Mail: [email protected]
HAUG Biel AG
Postfach
CH-2500 Biel/ Bienne 6
Johann-Renfer-Strasse 60
CH-2500 Biel/ Bienne 6
Telefon 0 32 / 3 44 96 96
Telefax 0 32 / 3 44 96 97
www.haug.de
E-Mail: [email protected]