The 4230 Bubbler Flow Meter Pocket Guide
is provided as a handy field reference. It is
not intended to replace the instruction
manual, but complements it by providing
condensed instructions. Study the manual
thoroughly before installing or operating the
flow meter.
COPYRIGHT © 1994 by
Teledyne Isco, Inc., 4700 Superior St.,
Lincoln, Nebraska, U.S.A. 68504
Phone: (402) 464-0231
Toll Free: (800) 228-4373
FAX: (402) 465-3022
www.isco.com
Part #60-3233-102
Revision H, May 11, 2006
The installation and use of this product may
require you to work in locations where you
could be seriously injured or even killed. Take
whatever precautions are necessary to ensure
your safety before entering the installation.
Never work alone or unsupervised. Install and
operate this product in accordance with all
applicable safety and health regulations, as
well as any appropriate local ordinances.
This product is often installed in confined
spaces. Examples of confined spaces are
manholes, pipelines, digesters, and storage
tanks. These places can be dangerous or fatal
if you are not suitably prepared. The primary
hazards are the presence of poisoned air, the
lack of ventilation, and the possibility of falls.
Other hazards may be present, as well. Work
in such places is governed by OSHA
1910.146, and may require a permit before
entering.
Material Safety Data Sheets (MSDS) for
chemical agents supplied or recommended for
use with this product are in the MSDS
Appendix in the back of the instruction
manual. These sheets provide information
about possible hazards from the chemicals.
Additional MSDS, covering various
proprietary agents (name-branded or
trademarked mixtures) that can also be used
with this product, are available from the
manufacturers of those agents.
Table of Contents
iii
Table of Contents
1.
Getting Started
1.1 Becoming Familiar with the 4230 Flow
Meter . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
1.1.1 Selection of a Power Source . . 1-2
1.1.2 Turning the Flow Meter On . . 1-3
1.1.3 Adjusting the Bubble Rate . . . 1-3
1.2 Checking the Installation . . . . . . . . . 1-4
1.2.1 Bubble Lines in High-Velocity
Streams . . . . . . . . . . . . . . . . . . 1-6
1.2.2 Clogging Bubble Lines . . . . . . 1-7
1.3 Options and Capabilities . . . . . . . . . . 1-8
1.3.1 Flow Metering Inserts . . . . . . 1-8
1.3.2 Connection to a Sampler . . . . 1-9
1.3.3 Adjusting Level . . . . . . . . . . . . 1-9
1.3.4 Data Acquisition and Storage
(Flowlink®) . . . . . . . . . . . . . . 1-10
1.3.5 Parameter Sensing with the 4230
Flow Meter . . . . . . . . . . . . . . 1-12
2.
Programming Introduction
2.1 Why Programming Is Necessary . . . . 2-1
2.1.1 Operation of the Keypad and
Display . . . . . . . . . . . . . . . . . . 2-2
2.2 Keypad Functions . . . . . . . . . . . . . . . 2-4
2.3 Programming Procedure . . . . . . . . . . 2-7
2.4 Description of Program Steps . . . . . 2-10
2.4.1 Step 1 - Operating Mode . . . 2-10
2.4.2 Step 2 - Flow Conversion Type . .
2-11
2.4.3 Step 3 - Adjust Level, Parameters
2-15
2.4.4 Step 4 - Reset Totalizer . . . . 2-17
2.4.5 Step 5 - Sampler Pacing . . . . 2-17
2.4.6 Step 6 - Sampler Enable . . . . 2-19
2.4.7 Step 7 - Alarm Dialout Mode 2-20
2.4.8 Step 8 - Printer . . . . . . . . . . . 2-21
2.4.9 Step 9 - Reports/History . . . . 2-22
2.5 Interpreting the Program Screens . 2-22
3.
Programming
3.1 Step 1 - Operating Mode . . . . . . . . . . 3-1
3.2 Step 2 - Flow Conversion . . . . . . . . . 3-36
3.2.1 Flow Metering Inserts . . . . . 3-45
3.2.2 Enter Maximum Head - All
Models . . . . . . . . . . . . . . . . . . 3-46
3.2.3 Programming the 4-20 mA
Outputs . . . . . . . . . . . . . . . . . 3-46
3.3 Step 3 - Port to Adjust . . . . . . . . . . . 3-48
3.4 Step 4 - Reset Totalizer . . . . . . . . . . 3-55
3.5 Step 5 - Sampler Pacing . . . . . . . . . 3-56
3.6 Step 6 - Sampler Enable . . . . . . . . . 3-61
3.7 Step 7 - Alarm Dialout Mode . . . . . . 3-67
3.8 Step 8 - Printer . . . . . . . . . . . . . . . . . 3-71
3.9 Step 9 - Reports/History . . . . . . . . . 3-75
4.
Maintenance
4.1 Care of the Flow Meter Case . . . . . . . 4-1
4.1.1 Care of the Case Seal . . . . . . . 4-2
4.1.2 Preventing Moisture Damage 4-2
4.2 Changing the Printer Paper and Ribbon
4-3
4.2.1 Installing the New Roll of Paper .
4-4
4.2.2 Ink Ribbon Replacement . . . . 4-6
4.3 Maintenance of the Bubble Line . . . . 4-7
4.3.1 Automatic Air Purge . . . . . . . 4-8
4.4 Maintenance of the Batteries . . . . . . 4-8
4.4.1 Charging the Nickel-Cadmium
Table of Contents
v
Battery . . . . . . . . . . . . . . . . . . 4-9
4.4.2 Five Station Battery Charger 4-9
4.4.3 Self-Discharge Characteristic . . .
4-10
4.4.4 Hazards of Overcharging . . . 4-11
4.4.5 Charging from Another Source . .
4-11
4.4.6 Memory Effects Temporary . 4-12
4.4.7 Internal Fuse . . . . . . . . . . . . 4-12
4.4.8 Using the Lead-Acid Battery 4-13
4.4.9 Using Other Types of Batteries . .
4-14
4.4.10 Attaching the Battery . . . . . 4-15
4.5 Regenerating the Desiccators . . . . . 4-16
4.5.1 Regenerating the Internal Case
Desiccant . . . . . . . . . . . . . . . . 4-16
4.5.2 Regenerating the External
Desiccant Cartridges . . . . . . 4-18
1.1 Becoming Familiar with the
4230 Flow Meter
The 4230 Flow Meter is a
microprocessor-equipped flow meter capable of
measuring flow rate in a wide variety of open
channels. The 4230 Flow Meter uses the
bubbler method for sensing level. A small air
pump and reservoir provide a source of
pressurized air. A transducer inside the flow
meter senses the air pressure necessary to force
a bubble from the end of a tube submerged in
the flow stream. As the liquid level increases,
the pressure required to release the bubble also
increases.
The flow meter converts the output from the
pressure transducer to a level reading. Built-in
level-to-flow rate tables convert the level to flow
rate. The level and flow rate appear on a
two-line, eighty character liquid crystal display
(LCD). These flow meters are equipped with an
internal printer to provide a continuous printed
record of level, flow rate, and other information
4230 Bubbler Flow Meter
1-2
that may be used for later reference. The Flow
Meters also have memory allocated that you
can set up to store level and other readings. You
must use Flowlink®, Teledyne Isco’s
proprietary data processing and acquisition
software to initialize (set up) and access this
memory.
1.1.1 Selection of a Power Source
Teledyne Isco offers a variety of power sources
to operate the flow meter. For
commercially-powered installations, there is
the AC Power Supply. Also available is a power
supply with a 1.2 Ah (Ampere-hour) standby
battery built in, the Battery-Backed Power
Supply. For applications where there is no AC
power connection available, Teledyne Isco offers
a 4 Ah nickel-cadmium battery
(Nickel-Cadmium Battery), or a 6 Ah
gelled-electrolyte battery (Lead-Acid Battery).
All these power sources mount on top of the
flow meter and are secured with rubber draw
catches. The cable with the two-pin connector
(12 volt) attaches to the +12 VDC connector on
the flow meter.
Battery life expectancy varies from about seven
to fourteen days between recharge cycles,
depending on the settings for the bubble rate
and the setting for the printer chart speed. For
very remote sites where changing the battery is
difficult, Teledyne Isco offers the Solar Panel
Battery Charger, used with the Lead-Acid
Battery. You may also power the 4230 Flow
Section 1 Getting Started
1-3
Meter with a deep-cycle R-V or marine battery.
You must mount these larger batteries
externally, as they are quite large. Teledyne
Isco offers a special connect cable for them.
1.1.2 Turning the Flow Meter On
After you have connected the flow meter to
power, turn the flow meter on with the ON
button located on the keypad. When the flow
meter is turned on, the air pump will run for
awhile, and then stop when the reservoir tank
is pressurized. If it does not stop, locate the
Bubble Rate Adjust Valve on the side of the
flow meter cabinet and turn it clockwise until it
just seats. Do not over-tighten or force the
valve; the needle-and-seat are fragile and can
easily be broken.
1.1.3 Adjusting the Bubble Rate
Locate the Bubble Rate Adjust valve on the
flow meter’s side. If it is locked, use the small
hex wrench supplied with the flow meter to
release the lock. (Refer to the manual for
details.) Attach a bubble line (identical in
diameter and similar in length to the one that
will be used) to the Bubble Line outlet. Place
the free end of the bubble line in a container
(preferably clear) of water. Set the container
beside the flow meter so you can watch the
bubbles coming out of the end. Adjust the valve
until a bubble rate of one bubble per second is
reached. Lock the valve handle with the hex
wrench supplied with the flow meter.
4230 Bubbler Flow Meter
1-4
Note
Teledyne Isco recommends setting the bubble rate in
the shop first, as this is easier with a small container of
clear water placed beside the flow meter. Using a
different length and diameter bubble line than the
installation will result in an inaccurate setting. In any
case, you should still recheck the bubble rate after
installation, particularly if the bubble line will be under
several feet of water in the flow stream.
1.2 Checking the Installation
At the job site it is worthwhile to check the
installation to see that it was made correctly
and to make sure nothing has changed which
could affect the accuracy of the measurements
made by the flow meter. If the flow meter or
bubble line is installed in a sewer or manhole
please read and observe the following warning:
DANGER
YOU CAN BE KILLED working in sewers or
manholes if you do not follow proper safety
procedures. Besides the risk of falling,
poisonous gases in most sewers can
overcome you quickly, without warning.
Safety information is provided in the 4230
Flow Meter Instruction Manual under the
heading of Safety Considerations. Please
read and follow these procedures; the life
you save may be your own.
Section 1 Getting Started
1-5
If the installation is temporary, make sure the
flow meter is installed so it is not difficult to
reach in case you need to change the
programming or chart paper. The unit should
not be a hazard to anyone entering the
manhole. You can lay the flow meter
horizontally or hang it vertically with a
suspension harness from the ladder. Make sure
the flow meter is suspended well above the
maximum level the flow stream can reach;
water drawn into the desiccator cartridge will
cause serious damage to the unit. If this is not
possible, attach a vinyl line to the open end of
the desiccator cartridge and route the line to a
place well above the maximum expected liquid
level.
If the installation is permanent, check to make
sure that any work going on in the area has not
caused any damage to the wiring or bubble
lines. Check the bubble line where it attaches to
the flow meter to see if flexing has cut the
silicone tubing used to connect the 1/8" bubble
line to the barbed fitting. A crack in the silicone
tubing will appear as a white line across the
middle of the tubing. The resulting leakage will
seriously affect the accuracy of the readings.
Check to make sure the bubble line is secure
from the flow meter to where it enters the flow
stream. Keep the bubble line as short as
possible between the flow meter and the
measuring point in the flow stream. Cut off any
excess length of tubing. Do not coil it up or let it
drift in the stream. Make sure the end of the
4230 Bubbler Flow Meter
1-6
bubble line is installed correctly (at a right
angle to the flow) and securely at the proper
measuring place in the primary device.
This point is usually below the “zero” level of
the primary device and a prescribed distance
upstream from the crest of the device. (This
distance cannot be called out specifically; it
depends on the primary measuring device.)
Refer to the Isco Open Channel Flow
Measurement Handbook for this information.
Flumes can be made with a specific location for
the bubble line. The Adjust Level/Parameters
programming step lets you install the bubble
line anywhere between ten feet above or below
the actual “zero” level of the primary device. In
practice, the installation is usually close to the
“zero” point.
Note
The 4230 Flow Meter cannot measure level below the
bubble line outlet. If you wish to measure accurately
down to the “zero” level of the device, install the bubble
line outlet one to two inches below the “zero” point of
the primary device, and adjust the level appropriately at
the flow meter. This way the liquid level will always be
higher than the bubble line outlet.
1.2.1 Bubble Lines in High-Velocity Streams
If the velocity of the flow stream is more than
five feet per second, and the bubble line outlet
is mounted at a right angle to the flow, the
measured level will tend to be lower than
actual. You can use the Adjust
Level/Parameters step to compensate for this
Section 1 Getting Started
1-7
difference when the level (and velocity) are at a
high point, but a reduction in the flow rate will
result in inaccurate settings. The best way to
overcome this problem is to use a stilling well.
You can also make a depression in the bottom of
the flow stream and place the bubbler there,
but flow streams with significant solids content
may fill the depression and restrict the bubbler.
Or, you can form a 90° bend about two inches
from the end of the bubble line and then place
this bend so it points downstream, parallel with
the flow.
1.2.2 Clogging Bubble Lines
In some flow streams, clogging of the bubble
line from algae, silt, or suspended solids may be
a problem. There are several possible solutions
you can try. You can increase the automatic
purge frequency. This is preferable to
increasing the bubble rate, and is highly
recommended where the flow meter is powered
from AC. Teledyne Isco has stainless steel
bubble line extensions available that may be
helpful, and for situations where algae tends to
grow in the lines, copper tubing is useful. The
copper salts formed over time in the flow
stream are algicidal agents and will inhibit the
growth of algae.
4230 Bubbler Flow Meter
1-8
1.3 Options and Capabilities
1.3.1 Flow Metering Inserts
For temporary monitoring purposes in 6", 8",
10", and 12" round sewer pipes, Teledyne Isco
offers flow metering inserts that have a built-in
bubbler and form a round orifice or v-notch weir
in the flow stream. An inflatable rubber
bladder keeps the insert firmly seated inside
the round pipe. The inserts should not be
considered permanent, as they do place a
restriction in the pipe and are subject to
clogging.
Figure 1-1 Flow Metering Insert Diagram
Bubble Line
Handle
Round Orifice
Air Line
Rubber Bladder
Bubbler
Section 1 Getting Started
1-9
1.3.2 Connection to a Sampler
A 4230 Flow Meter can control a sampler in a
flow-proportional sampling mode. This means
the sampler will take a sample after a specific
volume has passed through the flow stream,
rather than after a particular interval of time.
In this mode, the sampler and flow meter can
compensate for varying flow rates. You can use
the 4230 Flow Meter with Isco samplers or
samplers made by others.
Connecting an Isco Sampler to a 4230 Flow
Meter requires an Isco flow meter-to-sampler
connect cable. The cable is 25 feet long. If you
plan to use the sampler enable feature (see
Section 2 - Programming), make sure you use
the newer cable that has all six pins connected.
Older cables do not have a connection between
the F pins and this keeps sampler enabling
from working. If in doubt, use an ohmmeter or
continuity checker to test your cable.
1.3.3 Adjusting Level
Generally, you measure the level in the flow
stream with a gauge. Make this measurement
accurately, as the level setting on the flow
meter will determine the accuracy of all
subsequent level and flow rate measurements
made by the flow meter. When you are satisfied
that the flow stream has been measured
accurately, you can change the level reading on
the flow meter to show this reading.
First, make sure the bubble line is properly
installed and set at or below the “zero” point in
4230 Bubbler Flow Meter
1-10
the flow stream. Then set the level by selecting
Adjust Level/Parameters step in the flow meter
program. You may then adjust the level by
entering the measured level value through the
number keys. When the proper value is
displayed, enter the number into memory by
pressing the Enter/Program Step key.
Note
It is very important that you measure the level in the
stream accurately, as all subsequent calculations of flow
will be based on this measurement. If the level value
entered is incorrect, all subsequent flow calculations will
be incorrect also.
1.3.4 Data Acquisition and Storage
(Flowlink®)
As mentioned in previously, all 4230 Flow
Meters contain memory you can allocate to
store level, rainfall, and sample data. You can
interrogate this information for later
processing. To set up the flow meter’s memory
for data storage, you must use Flowlink,
Teledyne Isco’s proprietary data acquisition
and processing software. Programs from
Flowlink allow you to set up the flow meter for
memory initialization. data acquisition, and
interrogation from a different location through
an IBM PC® compatible computer. The
computer and the flow meter connect together
through a standard dialup telephone line with
modems. (The modem in the flow meter is an
optional accessory.) Multiple flow meters can be
initialized and interrogated by the same
Section 1 Getting Started
1-11
computer. You can also use Flowlink to access
the flow meter at the installation site with a
laptop computer plugged into the Interrogator
connector on the flow meter. Other programs in
Flowlink allow the processing of the data
retrieved from the flow meter. Refer to the
Flowlink manual for further information.
Note
Storage of data is not automatic in the 4230 Flow Meter.
You must use programs from Flowlink software to
initialize, partition, and size the memory, and also to
retrieve the stored data.
Figure 1-2 Recommended level measurement
method
Note
Note that you are actually measuring the distance from
the top center of the pipe to the surface of the flow
stream. The level (h), then is the distance from the top
center of the pipe, d, subtracted from the diameter of
the pipe, D. You must know the correct diameter of the
pipe.
D
d
h
D – d = h (level)
4230 Bubbler Flow Meter
1-12
1.3.5 Parameter Sensing with the 4230 Flow
Meter
The 4230 Flow Meter also has the capability of
displaying, recording, and (if Flowlink software
is used) storing data from parameter sensors.
The sensors available from Teledyne Isco for
the 4230 Flow Meter measure temperature, pH
(the relative acidity or alkalinity of a solution)
and D.O. (dissolved oxygen). The YSI Model 600
Multiparameter Sonde offers pH, D.O.,
temperature, and conductivity. You can also
measure and record rainfall with the Isco 674
Tipping Bucket Rain Gauge, which connects to
its own port on the flow meter.
All parameter probes require constant and
complete submersion in the flow stream for
proper operation. Dry operation can damage
the pH and D.O. probes. The pH and D.O.
probes are extremely sensitive devices and
require the use of the 201 pH Module and 270
D.O. Modules (signal amplifiers) between the
probes and the flow meter. The modules are not
interchangeable.
Note
The 270 D.O. module has been discontinued. Probes,
service kits, and accessories are still available to
maintain existing field units.
The probes can be installed in a variety of
places, but typical installation is in round pipes.
Teledyne Isco offers a series of rings that
provide simplified mounting for all probes
inside round pipes.
Section 1 Getting Started
1-13
Each probe snap-mounts to a specialized sensor
carrier.
• The sensor carrier then attaches to the
mounting rings by sliding tabs into
mating slots.
• The probe cable is routed from the
stream to the parameter module.
• The module is installed and its cable
routed to the flow meter’s Parameters
connector.
Figure 1-3 YSI 600 Multi-Parameter Sonde
(End cover removed to show
sensors)
Note
The 4230 Flow Meter has only one parameter sensing
port. You must select temperature, D.O. with
temperature, or pH with temperature. (D.O. and pH
require temperature monitoring for proper operation.) If
Temperature Sensor
pH Sensor
pH Reference
D.O. Sensor
Conductivity
(inside)
4230 Bubbler Flow Meter
1-14
you want to change your selection later, you can. You
will need the correct probe and module and you will
have to change the program in the flow meter.
Note
All brands or product names used in this text are
trademarks or registered trademarks, and are the
property of their respective companies or organizations.
2.1 Why Programming Is
Necessary
You must program the 4230 Flow Meter to
accurately monitor a flow stream. You must
also install the level measuring device, a
bubbler tube. The installation usually includes
a primary measuring device, a structure placed
in the stream that regulates flow. This section
describes programming the flow meter with the
keypad and display. There are nine program
steps that control the operation.
Teledyne Isco ships the flow meter with a
program installed called the default program.
This program is an example of the flow meter’s
capabilities. Note that the default program is to
test the unit at the factory. The flow meter’s
internal computer must always have something
programmed, so that becomes the default
program. Your flow situation will usually
require other programming. The text provided
4230 Bubbler Flow Meter
2-2
with each screen explains the reasons for the
various menu options.
2.1.1 Operation of the Keypad and Display
The display is a two-line, forty
character-per-line liquid crystal (LCD). It has a
backlight for easy viewing in low light
situations. The display has three different
operating modes, normal, programming, and
messages. In the normal mode, the display
shows such things as level, flow rate, total flow,
parameter measurement, etc. In the
programming mode, the top line of the display
shows each step as you work through the
program while the bottom line shows the
choices available for that step. In the message
mode, the display provides instructional
information, such as how to leave
programming, or what to do if you have entered
a number that is out-of-range.
Following is a “normal” display on the flow
meter. This is typical of what the flow meter
will display when it is in the normal operating
mode and you are not programming it.
An interpretation of the numbers on this
display would be as follows: Time and date are
replaced by pH/D.O. and temperature if you are
sensing parameters. The (X X) to the right of
the time indicates letters that may appear from
time to time on the 4230 Flow Meter.
0000004.78 CF 1.13 FT 16-MAR-04
1.03 CFS (X X) 8:25:37
Section 2 Programming Introduction
2-3
The letter C appears when the flow meter is
communicating with a remote computer
(Flowlink applications only). The letter Z
appears when the flow meter is doing an
auto-zero. The letter P appears when the flow
meter is purging the bubble line. The letters E
or D appear (Enable or Disable) when the
sampler enable function (step 6) is programmed
by condition. (Programmed by condition means
that the flow meter will enable the sampler
only when a certain condition or set of
conditions, sensed by the flow meter, are met.)
Following is a typical programming display on
the flow meter: (One of the items in the second
line will be flashing. The item flashing is the
selection currently held in memory.)
Following is a typical display providing
information:
Note
If you stop programming for more than two minutes, the
flow meter will time out, and whatever is on the display,
(message or program step) will revert to the “normal”
display, shown previously.
Total Flow Current Level Date (or pH/D.O.)
Flow Rate Time (temperature)
TOTALIZED VOLUME UNITS
• CF • • GAL • • M3 • • AF • • L • • MGAL •
CHANGES HAVE BEEN MADE IN STEP
PRESS '0' TO CONTINUE, PRESS '1' TO DISCARD
4230 Bubbler Flow Meter
2-4
The program consists of steps and substeps.
The steps are listed on the flow meter front
panel. Most steps contain several substeps.
Generally, you need to complete all the substeps
before stopping, or the flow meter will reject the
changes you made for that step after it times
out. There are some exceptions.
The flow meter keeps in memory any changes
that you made for the finished steps (all
substeps completed before stopping). Most
steps not finished when you stop will return to
the previous selection.
2.2 Keypad Functions
Use the keypad in response to messages on the
display. The following sections describe the
function of each key.
• OFF and ON–These two keys turn the
flow meter off and on.
• Go To Program Step–Pressing this key
lets you go directly to a particular step
without passing through all the steps of
the program. The display will ask you to
enter the number of the step you want to
program. Enter the number by pressing
one of the number keys. There are nine
program steps, so numbers from one to
nine are valid.
• Exit Program–Press this key when you
want to leave the programming mode and
return to the normal operating mode.
Section 2 Programming Introduction
2-5
• Clear Entry–Lets you return to the
previous entry for a program step if you
have changed the entry, but have not yet
pressed Enter.
• Enter/Program Step–This key has two
functions. One is to enter a program
selection into the flow meter’s memory
(Enter). The other is to step through the
program (Program Step).
• Print Program–Pressing this key
makes the flow meter print out a
complete list of the current program kept
in memory.
• Print Report–One of the functions of
the flow meter is to print periodic reports
of the activity recorded on the flow meter
at regular intervals. The contents of
these reports are defined in step 1. If you
set up the flow meter to generate these
reports, you can have a report printed at
any time by pressing this key. The report
will cover the time interval from the last
scheduled report up to the time you press
this key.
The flow meter will print the next report at
the next scheduled time. Note that if power
fails for five minutes or more, the flow meter
will print a report when power is restored
that will cover the interval between the last
report and the time that the power failed.
The next report will cover the time from the
power failure to next scheduled report time.
4230 Bubbler Flow Meter
2-6
• Chart Advance–Pressing this key
causes the paper chart to advance
through the printer at the fastest
possible speed. Nothing will be printed
while you press this key.
• Chart Reroll–You can unroll the chart
from the take-up roll to look at it by
pulling it out with your hands. Pressing
this key rewinds the chart onto the
take-up roll.
• Number keys–These keys let you enter
numeric values.
• Decimal Point–This key enters a
decimal point into a number when
programming. On flow meters equipped
with the modem, you can use this
character as a comma (delay) when
entering telephone dialout numbers.
• Arrow keys–These keys, referred to as
the left and right arrow keys let you
select a programming option by moving
across the menus shown on the second
line of the display.
• +/– key–This key lets you enter a plus or
minus to a quantity. Its most common use
is in entering values for the equation, a
method of flow conversion. On flow
meters equipped with the modem, use
this character as a dash when entering
dialout numbers.
Section 2 Programming Introduction
2-7
• Manual Purge key–This key lets you
purge the bubble line on the 4230 Flow
Meter at any time.
2.3 Programming Procedure
To begin programming, turn on the flow meter and
wait for the display to settle. Then either press the
Enter/Program Step key (generally referred to as
Enter) or the Go To Program Step key.
The display will change to two lines; the first
line describes the step you are programming,
and the second line shows the choices available.
One of the choices will be flashing. The flashing
indicates that this choice is the one in memory.
If you are satisfied with this choice, press
Enter, and the flow meter will advance to the
next step.
If you want a different choice from that
flashing, move across the display using the left
and right arrow keys. Each time you press the
right arrow key, the flashing selection will move
one position to the right. This will continue
until the last selection is flashing.
From time to time you will notice an arrow
pointing to the right edge of the display. This
indicates more choices are available beyond
what the display shows. Continue to press the
right arrow key and you can view these options.
After reaching the furthest option, the arrow
will move to the left side of the display,
indicating that there are options to the left.
These are the options you started with. If you
4230 Bubbler Flow Meter
2-8
want to go back, use the left arrow key until the
option reappears. When the desired selection is
flashing, press Enter. The display will then
automatically advance to the next step of the
program.
All of the program steps contain several
“substeps” that must be completed before you
advance to the next program step. Some steps,
like Reset Totalizer contain only a few substeps.
Some steps will require the entry of a numeric
value.
Program these steps by using the number and
decimal keys to enter the correct value.
Note
You can program most of the flow meter in the shop,
rather than at the job site, with the exception of step 3,
Adjust Level/Parameters. To set level you must make an
accurate measurement of the level in the flow stream
and then enter that value. This can only be done at the
job site.
If you are programming the flow meter for the
first time, generally you will press Enter, start
with step 1, and go on from there. If the flow
meter has been in use and you need to change
only a certain part of the program, you would
more likely use the Go To Program Step key.
With this key you can go directly to the
program step you need to change, saving time.
If you change an entry and do not like it, you
can make the display go back to the original
entry by pressing Clear Entry. If you have
already pressed Enter, however, the new value
Section 2 Programming Introduction
2-9
will be in memory. To change it, press Exit
Program and return to that step. If you are in
the middle of a step with multiple substeps, the
flow meter will display, “Changes have been
made in step; press 0 to continue or 1 to
discard.” If you press 1, the display will return
to normal and the last step you were working
on will return to its previous selection. (Any
step you completely change before you exit will
remain changed.)
You can re-enter the program with either Enter
or the Go To Program Step keys. If you become
confused while programming, the best
suggestion is to press Exit Program and start
over. Also remember that you can have the flow
meter print a complete list of your program
choices by exiting the program and by pressing
the Print Program key as soon as the display
returns to the normal operating condition,
displaying level and total flow, etc.
Following are the steps to program a 4230 Flow
Meter:
1. Operating Mode
2. Flow Conversion
3. Adjust Level/Parameters
4. Reset Totalizer
5. Sampler Pacing
6. Sampler Enable
7. Alarm Dialout
8. Printer
9. Reports/History
4230 Bubbler Flow Meter
2-10
2.4 Description of Program Steps
2.4.1 Step 1 - Operating Mode
Step 1, Operating Mode, determines how you
set up the flow meter. In this step there are two
choices, Program and Setup. Program advances
you to step 2, and from there on you correlate
the flow meter to the flow stream. Setup selects
various basic “housekeeping” features for the
flow meter. Here you determine the internal
clock, site identification, measurement setup,
hysteresis, report contents, operation of the
display backlight, and program lock. In
Program you select the units of measure the
flow meter will use for the display, calculations,
and reports.
Note
If you choose NOT MEASURED for any selection, the
flow meter will make no further reference to that value or
function for the rest of the program, and you will not be
able to activate that process or function later on unless
you reprogram step 1.
If there is a feature or option you need that does not
appear on your display when the manual says it should,
return to step 1 and make sure you have not
inadvertently it turned off in the Program or Setup
menus.
Note that selecting some features automatically
excludes others. For example, selection of pH or D.O.
excludes the other parameter, unless you use the YSI
600 Multi-Parameter Sonde, which measures pH, D.O.,
and conductivity at the same time.
Section 2 Programming Introduction
2-11
This method keeps program size manageable
and makes programming more efficient. By
turning off unneeded features early, you do not
have to keep de-selecting those features over
and over.
Consequently, choose carefully from the first
step. Study the program first, then fill out the
Programming Worksheet (in the back of the
manual), and then program the flow meter last.
2.4.2 Step 2 - Flow Conversion Type
Step 2, Flow Conversion Type, determines how
the flow meter calculates flow rate and total
flow. For a 4230 Flow Meter, flow rate is
calculated by knowing the measured level and
(usually) the characteristics of a structure
called a primary measuring device.
A primary measuring device is a structure
placed in a flow stream through which the
entire stream flows. These devices are made in
a number of styles and sizes, but they all have
one thing in common: For any type of primary
measuring device there is a known relationship
between the level in the flow stream ahead of
the device and flow rate through the device.
Consequently, after you measure level, the flow
meter can calculate flow rate and total flow
from the level, by consulting built-in look-up
tables.
Information about many common primary
measuring devices is provided in the Isco Open
Channel Flow Measurement Handbook. This
useful book provides formulas, flow rates at
4230 Bubbler Flow Meter
2-12
various levels, and values for maximum head,
as well as much interesting descriptive
material. This book is available free from
Teledyne Isco by returning the request card
shipped with the flow meter. If your installation
uses a nonstandard primary device, you should
consult the manufacturer of the device for flow
rates at given levels. The flow meter will then
calculate a flow conversion for such a device on
the basis of the manufacturer’s data you enter
as data points or an equation. In some
instances, a nonstandard primary device could
be supplied with a flow equation; you can enter
that equation into the flow meter and the flow
meter will calculate the flow rate from that
equation.
Flow Calculations Without Primary
Devices – Note that it is not always necessary
to have a primary measuring device. The 4230
Flow Meter can measure level and calculate
flow without any primary device being installed
in the flow stream. Sometimes the shape of the
flow stream itself forms the primary device.
The Manning formula uses the shape of a pipe
or channel and its slope to calculate flow in
open (non-pressurized) pipes.
An accessory is available for the 4230 Flow
Meter called the Flow Metering Insert. These
inserts, used in round pipes of 6", 8", 10", and
12", form a primary device inside the pipe by
restricting flow and measuring the level of the
liquid backed-up behind the insert. The
opening in the insert, either a smaller round
Section 2 Programming Introduction
2-13
opening or a V-notch that forms a weir, forms a
primary device.
The conversion types available are
WEIR/FLUME, MANNING, DATA POINTS,
EQUATION, and FLOW METERING
INSERTS.
Use Weir/Flume flow conversion when your
primary device is a weir or a flume. A weir is a
wall or dam across the flow stream. Water must
rise to the point where it flows over the top of
the wall. The measured level upstream of the
device and the appropriate formula are used to
calculate flow. Flumes differ from weirs in that
there is no wall or barrier, but instead a
restriction, typically a sharp narrowing or
change in the slope of the channel that restricts
the flow.
Again, the measured level of the stream at
some point ahead of the restriction is used by
the flow meter to calculate flow. In this flow
conversion mode, the flow meter uses internal
look-up tables for many common primary
measuring devices.
An Equation is used when you have a
non-standard primary device, or want to use
different values from those programmed into
the look-up tables of the flow meter. Equation
uses the standard flow equation:
Where Q equals flow rate; k1 and k2 are
constants; H is level (or head), and P1 and P2
Q
k1Hp1
k2Hp2
+
≡
4230 Bubbler Flow Meter
2-14
are the powers to which the two H terms are
raised. (Your equation may not have the second
term, in which case you would enter 0 for the
second constant, k2.) Most common primary
devices are supported in the flow meter’s
software, so generally you will not need this
option. But it is available for those needing to
enter their own values, or for those who have a
nonstandard primary device for which an
equation can correlate level and flow.
Manning Flow Conversion – uses the
Manning formula to calculate flow in open or
closed (nonpressurized) gravity-flow situations
based on slope, diameter, and roughness of the
pipe. There is no primary measuring device as
such. Instead the pipe, with considerations for
its slope and internal roughness, serves as the
primary device. The 4230 Flow Meter can
calculate flow in round pipes, rectangular,
U-shaped, or trapezoidal channels based on this
formula.
Data Point Flow Conversion (DATA
POINTS) – calculates flow based on a set of
user-entered data points for a flow stream.
Data consist of correlated level and flow
measurements for the stream. Like the
Equation method of flow conversion, this flow
conversion is most commonly used where the
primary measuring device is nonstandard, but
where tables of level and flow rate data are
available from the device manufacturer. The
4230 Flow Meter has space for four sets of data
with as many as fifty points per set. The flow
Section 2 Programming Introduction
2-15
meter then calculates flow from these data
tables using a three-point interpolation.
Flow Metering Insert Conversion – The
inserts are installed in upstream pipes and held
with compressed air pressurizing a bladder.
They are set from street level with a handle
that can be extended as far as sixteen feet. The
inserts contain a bubble line outlet and have an
opening in the face that is either round or
triangular. The 4230 Flow Meter reads the
upstream level (the water backs up behind the
insert) and calculates the flow through the
insert from this measured level.
2.4.3 Step 3 - Adjust Level, Parameters
Adjust Level, Parameters calibrates the sensors
that provide the flow meter with level and other
information. In this step you set the level in the
flow stream. First you measure the level, as
accurately as possible. Then you enter this
value with the numeric keys. Accuracy is
important. This measurement provides the
basis for all subsequent flow calculations in the
flow meter.
The flow meter also has an input port for
measurements other than level. This is the
Parameter Port. Here you can sense such
variables as temperature, pH (the acidity or
alkalinity of a solution) and D.O. (dissolved
oxygen) in the flow stream. You can have either
pH with temperature, D.O. with temperature,
or temperature alone. The port is not dedicated
to a particular sensor, except through
4230 Bubbler Flow Meter
2-16
programming. You can change the sensor. For
example, you can change from a pH probe to a
D.O. sensor if you change the programming.
Selection of either parameter will keep the
other from appearing later on the menus. Note,
however, that it is possible to measure several
different stream conditions including pH and
D.O. at the same time with the YSI 600 Sonde.
The YSI 600 Sonde – The YSI 600 Sonde is a
multi-purpose, water quality measurement
device. It is intended for use in research,
assessment, and regulatory compliance. The
sonde attaches to the modified RAIN GAUGE
connector on the 4230 Flow Meter. Flow meters
having only a 4-pin rain gauge connector will
not support the YSI Sonde. If you wish to
upgrade your flow meter to use this system,
contact the factory. Note that you can have both
the YSI 600 Sonde and the Rain Gauge
connected to the flow meter at the same time by
using a special Y-connect cable.
The YSI 600 can measure the following water
qualities: dissolved oxygen (D.O.), conductivity,
temperature, and pH. Conductivity
measurements made by the sonde can be used
to calculate specific conductivity, salinity, and
total dissolved solids. A brief description and
specifications for the YSI 600 are printed in
Section 4 (Accessories) in the manual. You may
also contact the factory or your Teledyne Isco
representative. More information on the sonde
is found in the YSI 600 Manual, shipped with
each YSI 600 Sonde.
Section 2 Programming Introduction
2-17
Figure 2-1 YSI 600 Sonde (cover removed)
2.4.4 Step 4 - Reset Totalizer
In this step you decide whether you want to
reset the flow meter’s internal flow totalizers. If
the installation is permanent you generally
won’t reset the totalizer. If you are using the
flow meter as a portable recording unit and are
moving it from one site to another, you
generally reset the totalizer between sites.
2.4.5 Step 5 - Sampler Pacing
It is common to use a flow meter with an Isco
sampler. Typically the flow meter signals the
sampler to take a sample after a certain volume
has passed. This might also occur after a
condition or set of conditions has either
changed or been met. This step allows you to
determine that control. There are several
possible options—DISABLE, CONDITIONAL,
VOLUME, and FLOWLINK. DISABLE will
Assembled Sonde
pH Glass
Sensor
pH
Reference
D.O.
Sensor
Temperature
Conductivity
(Inside)
4230 Bubbler Flow Meter
2-18
keep the sampler from receiving any flow
pulses from the flow meter. VOLUME allows
the flow meter to signal the sampler whenever
a specific flow volume has passed. FLOWLINK
(this option only appears if you are using
Flowlink software), allows the sampler to be
signalled from the flow meter as a result of
conditions determined by Flowlink.
Note
If you choose CONDITIONAL for sampler pacing and it
doesn’t seem to work properly for you, read the section
on hysteresis. Then check the hysteresis setpoints for
your conditions. (The defaults are all zero.)
You must also have the appropriate sensors to
measure temperature, dissolved oxygen, conductivity,
or pH; the flow meter cannot do this by itself, nor does it
occur automatically.
Flowlink is Teledyne Isco’s proprietary data
acquisition and management software.
Flowlink works with personal computers,
modems, and laptop computers to monitor flow
meters from a distance. Consult the factory for
more information.
VOLUME causes the flow meter to pace the
sampler after a specific volume has passed
through the flow stream.
CONDITIONAL allows pacing of the sampler
by the flow meter when a particular condition
has been met, or has changed. Among these
conditions are changes in level, flow rate,
temperature, rainfall, (if you are using the
optional rain gauge), dissolved oxygen, or pH.
Section 2 Programming Introduction
2-19
You can also use a pair of conditions, or if you
are using the YSI 600 sonde, you can select
multiple conditions from its sensors.
2.4.6 Step 6 - Sampler Enable
Sampler Enable means that in a combination
flow meter/sampler pair, the flow meter
controls the sampler’s ability to run its own
program. The difference between step 5,
sampler pacing, and step 6, sampler enable is
that in sampler pacing, the flow meter merely
sends flow pulses to the sampler from time to
time. The sampler counts these flow pulses to
determine when to take a sample (according to
its own programming).
With sampler pacing, the sampler is always
enabled. With sampler enabling, the flow meter
can actually stop operation of the sampler. The
sampler is still set up to run its own program,
but the inhibit/enable line from the flow meter
will determine when and whether the sampler
runs its program. This feature is useful for
storm water runoff monitoring applications,
where it may be necessary for the flow
meter/sampler pair to have to wait a long time
between storm intervals.
Again, changing or meeting a condition or set of
conditions triggers the enabling. The conditions
that can be used for sampler enabling are
similar to those used for sampler pacing: level,
flow rate, rainfall, temperature, dissolved
oxygen, pH, or a combination of these
conditions. The YSI 600 Sonde provides several
4230 Bubbler Flow Meter
2-20
measurements at the same time. You must
have the appropriate sensors for rainfall,
temperature, D.O., pH, and the YSI outputs.
2.4.7 Step 7 - Alarm Dialout Mode
This feature lets you program a 4230 Flow
Meter to signal a remote location through a
telephone line. This feature is useful for
transmitting alarm conditions or other
information to a remote location.
Note
You must have the optional modem installed to make
use of this program step. The menus will not even
appear unless the flow meter has a modem installed.
The Alarm Dialout feature is useful if you need
to signal a remote location when there is a
change of condition in the flow stream that
could constitute an alarm. You can program as
many as five different twenty-digit telephone
numbers into the flow meter in decreasing
order of importance. The modem is capable of
speech.
DISABLE inhibits this function altogether.
CONDITIONAL lets you program the flow
meter to signal these alarms for a variety of
reasons. You can use rainfall, time, level, flow
rate, dissolved oxygen, pH, rate-of-change, a
combination of conditions, or define the
operation through Flowlink software from
another computer. STORM lets you set the
alarm through a combination of rainfall and
time. You can also program the interval
Section 2 Programming Introduction
2-21
between calls and set up the system to reset the
alarm condition by dialing back from the
remote telephone.
2.4.8 Step 8 - Printer
The 4230 Flow Meter has a built-in printer. The
printer is also capable of plotting linear data
along with printing alphanumeric (letters and
numbers) messages. In this step you set the
speed for the chart to advance, from 1/2" to 4"
per hour.
Set chart speed according to the amount of
resolution you want to see on the chart. If there
is a great deal of activity on the chart, you
would generally choose a faster speed so the
marks are more “spread out” and are easier to
interpret. If there is little activity on the chart
and you want the flow meter to run for long
periods without having to change the chart
paper roll, you would probably pick a slower
speed.
The flow meter can plot three separate data
lines on the chart in addition to messages.
These lines may indicate such things as level,
flow rate, pH, dissolved oxygen, or
temperature. Note that you must have the
appropriate sensors to take advantage of the
availability of these plots. Rainfall is printed as
a bar graph.
The printer can plot over-ranges for the data
lines. The printer is in over-range if a data line
goes off the chart on the right side and then
4230 Bubbler Flow Meter
2-22
immediately starts over again plotting on the
left side.
2.4.9 Step 9 - Reports/History
This step lets you program the flow meter to
print out regular reports on the internal
printer. The reports are a summary of activity
recorded over a period of time. Included are
such items as maximum and minimum flow
rates, the time they were reached, sample
records, etc.
The flow meter will create two separate reports,
and you define what appears on them. Note
that the contents of the reports are defined in
Setup in step 1. Step 9 merely lets you turn
them off and on and set the timing. You can
define the start time, the interval between
reports, and other aspects.
History provides a record of changes made to
the flow meter’s program or operation events.
As many as 50 changes can be stored in the
flow meter’s memory at a time. The memory
can store up to 50 history items and 200 sample
events at a time.
2.5 Interpreting the Program
Screens
Following are the program screens as they
appear on the display of a 4230 Flow Meter.
Explanations of most of the screens will be
provided.
Section 2 Programming Introduction
2-23
Important
Some items that appear in these menus will have
parentheses (...) around them. This means that the item
may or may not appear on your flow meter. Choices you
make early in the program will make some options
unavailable later.
An example of this is the pH/D.O. (Dissolved
Oxygen) sensor option. Selection of either in
step 1 will keep the other from appearing in all
following menus. If your installation does not
use either sensor, you would select NOT
MEASURED for both, and no further reference
to either sensor would appear again for the rest
of the program.
As you work through the program, your
selections will rule out alternatives. This will
cause those alternatives not to appear later in
the program. Since there is no way of knowing
what program choices will be made for a
particular installation, it is necessary to
provide all the possible menus in the manual,
even though some of them will not appear on
your instrument.
Note
This list does not include all possible screens for the
unit, but does cover the screens found in a typical
programming sequence. Some diagnostic and error
screens are discussed in Section 5 in the instruction
manual.
3.1 Step 1 - Operating Mode
Turn on the machine. Wait for the display to
settle. Then press the Enter/Program Step
(Enter) key. The following will appear (step 1).
If the following menu does not appear, press
Exit Program, then Go To Program Step, then
press 1.
PROGRAM is always the default. If you press
Enter, the display will automatically advance to
the next display, which will ask you to select
units of measurement. If you select SETUP, the
following will appear:
If you press the right arrow key, the following
options will appear on the display:
SELECT OPTION
• PROGRAM • • SETUP •
SETUP OPTIONS: 'EXIT' TO QUIT
• SET CLOCK • • SITE ID • • MEASUREMENT SETUP •
SETUP OPTIONS: 'EXIT' TO QUIT
• STATUS • • ENABLE/ALARM HYSTERESIS •
4230 Bubbler Flow Meter
3-2
Then:
And:
Finally:
LANGUAGE may not appear on your unit. This
is intended primarily for export models, as they
are programmed in different languages.
Domestic models will contain English and
Spanish.
If you select SET CLOCK, the following will
appear:
Enter the year (four digits), the month (01-12),
the day (01-31), the hour (01-24), and the
minute (01-59).
If you select SITE ID, the following will appear:
You can select any suitable three-digit number
for the site identification.
If you select MEASUREMENT SETUP, the
following will appear. You will have to use the
SETUP OPTIONS: 'EXIT' TO QUIT
• OPTIONAL OUTPUTS • • REPORT SETUP •
SETUP OPTIONS: 'EXIT' TO QUIT
• LCD BACKLIGHT • •(LANGUAGE) • •PROGRAM LOCK •
SETUP OPTIONS: 'EXIT' TO QUIT
• (LANGUAGE) • • PROGRAM LOCK • • PROGRAM •
YEAR MONTH DAY HOUR MIN
XXXX XX XX XX XX
SITE ID: XXX
Section 3 Programming
3-3
right arrow key to bring all the options on
screen:
DO/PH READING INTERVAL refers to the
measurement of specific aspects of the flow
stream other than amount. The 4230 Flow
Meters support measurement of three different
characteristics: temperature, pH (the relative
acidity or alkalinity of a solution), and D.O.,
dissolved oxygen.
Note
If you are using the Isco D.O. sensor or are sensing D.O.
with the YSI Sonde, select as long an interval as is
practical for your application. The reasoning is that the
D.O. sensor is turned off between measurement
intervals and this turned-off period prolongs the life of
the sensor.
YSI 600 READING INTERVAL refers to the
YSI 600 sonde. The flow meter can measure
several different aspects of the stream at the
same time, including pH, D.O., temperature,
plus conductivity.
If you select PURGE INTERVAL, the following
will appear:
MEASUREMENT SETUP
• DO/PH READING INTERVAL • • YSI 600 READING INT-
MEASUREMENT SETUP
-ERVAL • • PURGE INTERVAL • • SUPER BUBBLE MODE •
PURGE INTERVAL
• 5 MIN • • 10 MIN • • 15 MIN • • 30 MIN • • 1 HR •
4230 Bubbler Flow Meter
3-4
PURGE INTERVAL refers to how often the
flow meter discharges a blast of air through the
bubble line. These periodic blasts are intended
to keep the end of the bubble line clear from
any obstruction that could clog it, preventing it
from releasing any bubbles. The option of
setting the purge interval is offered for the sake
of battery conservation.
If the bubble line tends to clog, characteristic of
dirty flow streams, select a more frequent
purge. If the stream is clean and the bubble line
does not clog, you could try a less frequent
purge. However, if your installation is
battery-powered, you should try to get by with
as few purges as possible to prolong the battery
life. Some experimentation may be necessary to
find the right setting.
If you select SUPER BUBBLE MODE, the
following will appear:
Super Bubble is an Isco feature that increases
the air supply to the bubble line when the flow
meter senses a rapidly rising liquid level. This
allows the level measurement to be more
accurate and to react more quickly. In
operation, Super Bubble resembles Purge in
that it forces a blast of air through the bubble
line; but the effect is different. Without Super
Bubble, there would be a period when no
bubbles come out of the bubble line when the
level rises suddenly. Until the bubble rate is
SUPER BUBBLE MODE
• ON • • OFF •
Section 3 Programming
3-5
re-established, the flow meter cannot
accurately measure the level. Super Bubble
helps maintain accurate level measurement by
filling the bubble line with air, ensuring that
the bubble rate will re-establish quickly when
the level stabilizes.
Operation of Super Bubble will cause an
increase in power consumption, because the air
pump must run. If your flow meter operates
from AC power, leave Super Bubble on. If your
installation is battery-powered and the flow is
generally stable (without sudden increases),
you can turn Super Bubble off. Again, some
experimentation may be necessary.
If you select DO/PH READING INTERVAL, the
following will appear:
Again, this selection is offered as a way to save
battery power. (and prolong the life of the D.O.
sensor). After selecting the appropriate
interval, press Enter. The Setup menu will
reappear. This time select YSI 600 READING
INTERVAL from the menu. The following
display will appear:
After selecting the appropriate parameter
reading interval, press Enter.The Setup menu
will reappear. This time select STATUS from
DO/PH READING INTERVAL
•CONTINUOUS••15 SEC••30 SEC••1 MIN••2 MIN••5 MIN•
YSI 600 READING INTERVAL
•CONTINUOUS••15 SEC••30 SEC••1 MIN••2 MIN••5 MIN•
4230 Bubbler Flow Meter
3-6
the Setup menu. Press Enter. The following will
appear:
• HW REV refers to the hardware revision
number.
• SW REV refers to the software revision
number.
• ID is an internal identification number
for the flow meter.
If you press Enter again, the flow meter will
display the system voltage:
This value should be from 10.5 to 13.5 (volts
DC). PUMP DUTY CYCLE tells you how much
the pump is running. This value is refreshed
every few minutes and should generally
average below 20%. If you see a value
consistently higher than this, check the bubble
line, look for leaks, or check the air system
inside the flow meter.
Note
If you do not press Enter after the first diagnostic menu
appears, the flow meter will automatically advance the
display through the next two screens and finally revert to
the Setup menu after a short time-out.
MODEL 4230 HW REV: XXXXXX SW REV X.XX
ID XXXXXXXXXXX
SUPPLY VOLTAGE: XX.X
PUMP DUTY CYCLE XX.X%
YSI SOFTWARE REV: XX.X
Section 3 Programming
3-7
Return to the Setup menu if the unit has not
already done so. This time select
ENABLE/ALARM HYSTERESIS from the
menu. Press Enter. The following will appear:
The HYSTERESIS menu lets you set the range
over which the level (or other condition) can
vary before the flow meter responds to the
change.
In the PROGRAM section there are several
steps that require a change in a condition to
make the flow meter carry out certain actions.
For example, step 6 programs the flow meter to
enable (activate) a sampler. In that step, you
select a condition (or set of conditions) that
must occur before the sampler is enabled. Enter
a value (level is an example) that must be met
before the enabling occurs.
But what if this value is met briefly and then
falls away? It is possible for a condition to vary
rapidly over a narrow range. Without
hysteresis, the flow meter would turn the
sampler off and on repeatedly, causing a
condition known as chattering, resulting in
very erratic operation of the sampler.
With hysteresis, you can enter a value that will
keep the flow meter from responding to small
changes in the enabling condition. Select a
value for hysteresis narrow enough to allow the
flow meter to respond to any significant change,
LEVEL ENABLE/ALARM
HYSTERESIS X.XXX FT (or meters)
4230 Bubbler Flow Meter
3-8
but broad enough to ignore minor changes that
could cause chattering.
Press Enter again and the following will
appear:
The next three menus may appear or not,
depending on selections you made earlier in
Program. They concern alarm/enable hysteresis
set points for parameter sensing—temperature,
pH, and D.O.
If you want to set hysteresis for any of these
items, enable them when you work through the
program section, then re-enter the Setup
section (Hysteresis) and they will appear. Note
that you can have temperature alone, or
temperature with either pH or D.O. You must
have temperature with either pH or D.O. You
cannot have pH and D.O. at the same time
unless you use the YSI 600 Sonde, and selection
of one will prevent the other from appearing on
the menus later.
The following will appear if you are measuring
temperature:
The following will appear if you are measuring
pH:
FLOW RATE ENABLE/ALARM
HYSTERESIS X.XXX CFS (Or other units of measure.)
TEMPERATURE ENABLE/ALARM
HYSTERESIS XX.XXX DEG F (or C)
pH ENABLE/ALARM
HYSTERESIS X.XXX pH
Section 3 Programming
3-9
The following will appear if you are measuring
D.O:
(or mg./l depending on the units selected in
Program.)
A set of menus similar to those shown above for
pH, D.O., and temperature will then appear for
the YSI 600 Sonde, if you have selected it. You
can set hysteresis for YSI-pH, YSI-D.O.,
YSI-Conductivity, and YSI-temperature, if
these parameters have been turned on in
previous program selections.
After all the HYSTERESIS menus have been
set, press Enter. The display will return to the
Setup menu. This time select OPTIONAL
OUTPUTS with the arrow key. Press Enter.
ALARM BOX refers to an external accessory
used to signal alarms from flow meter
measurements. See Section 4 in the manual for
more information about the Alarm Box. Note
that choice of SERIAL OUTPUT will eliminate
ALARM BOX as an option. Likewise, choice of
ALARM BOX will eliminate SERIAL OUTPUT
as an option.
If you select any of these OPTIONAL
OUTPUTS, the flow meter will request that you
turn them on or off. If you are running on
battery, select OFF for all unused outputs.
DO ENABLE/ALARM
HYSTERESIS X.XXX PPM
OPTIONAL OUTPUTS
• ANALOG OUTPUT • • SERIAL OUTPUT • • ALARM BOX •
4230 Bubbler Flow Meter
3-10
ANALOG OUTPUT refers to the flow meter’s
capability of managing associated equipment
through a 4-20 mA current loop. The 4-20 mA
current loop is a common method used to
control industrial processes that are variable
(rather than just fully off or on). At the lower
value (4 mA) the control is turned off (0%); at
20 mA the control is completely turned on
(100%). In between, rates range from 1 to 99%.
A typical application is a chlorinator, which
must vary in application of the chlorine gas as
the amount of water passing through the
system increases or decreases. Current ranges
other than 4-20 mA are also in use, although
they are less common than 4-20 mA. Examples
are 0-20 mA (supported by the flow meter on
the internal card only) and for longer current
loops, 10-50 mA (not supported by the flow
meter).
Teledyne Isco offers two different arrangements
for the 4-20 mA control circuit. You can have
either or both with the same flow meter. One
arrangement requires the use of an external
accessory, the 4-20 mA Output Interface (see
Section 4 in the manual). This module connects
to the flow meter and a source of AC power and
contains the circuitry necessary to create the
4-20 mA current loop. This accessory connects
to the flow meter through the Interrogator
connector.
The other 4-20 mA option is a board installed
inside the flow meter that contains circuitry for
three separate, isolated 4-20 mA current loops.
Section 3 Programming
3-11
If you use both the external converter and the
internal board, you can have a total of four
current loops controlled by the same flow meter.
The internal current loops are brought out to a
6-pin M/S connector in the slot where the
Modem connector is usually mounted.
Additional information for the internal analog
output board, including specifications for the
loops, is found in Section 4 of the manual.
CAUTION
Each 4-20 mA output represents a constant drain on
the flow meter of at least 16 mA whether activated or
not. While 4-20 mA applications are generally made
in installations with commercial power available,
Teledyne Isco suggests the following for those who
have a 4-20 mA output in a battery-powered
installation.
Use with battery powered flow meters only:
1. If the battery is continuously on charge (for
example with a Solar Panel Battery Charger).
2. If the battery is very large, such as a deep-cycle
or marine type battery, or an Isco 35 Ampere-hour
lead-acid battery.
3. Use only one 4-20 mA output.
4. Keep in mind that programming choices also
affect power consumption. Use “minimum” settings
on the flow meter wherever possible. (See Section 1,
Table 1-8, in the manual.)
Even with these circumstances, you may expect
significantly shorter charge life from your
battery. To determine the effect of this extra
current draw on battery life, please refer to the
4230 Bubbler Flow Meter
3-12
section How to Make Battery Calculations, at
the end of Section 1 in the manual.
The following menus determine the behavior of
the 4-20 mA current outputs. If you select
ANALOG OUTPUT (another term for the 4-20
mA Output) and the flow meter is equipped
with the optional internal board or the 4-20 mA
external accessory has been turned on,
RANGE, SMOOTHING, and MANUAL
CONTROL will appear:
Note
If you know for certain that you have the proper installed
hardware for internal 4-20 mA outputs and you do not
see the proper displays, return to the “normal” flow
meter display and press 4 - 2 - 0.
Note
If you do not have the proper hardware installed and you
press 4 - 2 - 0 and the number of analog outputs is not
zero, the external 4-20 mA converter will not work
properly. If this occurs return to the 4 - 2 - 0 option, and
at the prompt, enter 0 for the number of outputs
activated. This will restore the external 4-20 mA
converter capability.
MANUAL CONTROL will appear if you
continue moving to the right. “RANGE” will
appear with the ANALOG OUTPUT menu if
the optional internal 4-20 mA converter is
present in the flow meter.
ANALOG OUTPUT
• EXTERNAL 4-20 MA • • (RANGE) • • (SMOOTHING) •
Section 3 Programming
3-13
If you select RANGE, the following will appear:
This menu lets you select the current value for
zero-percent (baseline) compatible with your
equipment (internal 4-20 mA board only).
If you select SMOOTHING from the ANALOG
OUTPUT menu the following will appear:
The SMOOTHING option lets you stabilize
operation of the outputs by preventing a rapid
reaction to sudden sharp changes in the
condition being monitored that quickly return
to normal (transients). Selection of a smoothing
interval will prevent the equipment controlled
by the 4-20 mA loop from reacting too quickly,
too much, or operating erratically. A low-pass
filter algorithm is incorporated in the software.
If you select the MANUAL CONTROL option
from the ANALOG OUTPUT menu the
following will appear:
This option lets you control the operation of a
4-20 mA loop to check the operation of
equipment controlled by the loop at any level
from 0 to 100%. After connecting a 4-20 mA
output to a controlled device, you can program
the flow meter to put a specified current on a
specific analog output. If you are using the
OUTPUT RANGE
• 0 - 20 mA • • 4 - 20 mA •
SMOOTHING
• NONE • • 15 SEC • • 30 SEC • • 1 MIN •
MANUAL CONTROL (OUTPUT 0 = EXTERNAL)
OUTPUT 0 = 0.0 MA
4230 Bubbler Flow Meter
3-14
external 4-20 mA converter, the Analog Output
number will be zero.
Note
Selecting the MANUAL CONTROL option and
programming any one of the ports will prevent the
values transmitted by the other active 4-20 mA ports
from being updated until the test is completed. The
other ports will continue to transmit whatever value they
held at the start of the test. Exiting from the MANUAL
CONTROL menu at the end of the test will return all
active 4-20 mA ports to normal operation.
Programming for the conditions and values
that determine the operation of the 4-20 mA
loop (or loops) is done in step 2.
Returning to the OPTIONAL OUTPUTS menu,
you will see the SERIAL OUTPUT option. This
feature lets the flow meter transmit the most
recent values for all currently enabled ports as
ASCII text. You can then write a simple
program to retrieve this data periodically, or
you can do it interactively using a terminal
program.
Note
The information in the following section is provided for
those who can write their own software programs to
process the data transmitted from the Serial Output
port. Special cables may be required. Contact Teledyne
Isco technical support for more information.
Command Line: (Use the INTERROGATOR
connector.) The lines of text contain the port
values for each port that is turned on. The
Section 3 Programming
3-15
DATA command will use a special command
response protocol.
The following table provides the ASCII codes
for port types and standard units of measure.
Table 3-1 ASCII Output Codes
Code
Parameter
Units
DE
Description
String (enclosed in
quotes)
ID
Unit specific identifier
Unsigned long
MO
Model
String (enclosed in
quotes)
TI
Time since 1900
Days
BV
Battery Voltage
Volts
LE
Level
Meters
LSI
Level Signal Strength
0 - 100%
VE
Velocity
Meters per second
VSI
Velocity Signal Strength 0 - 100%
VSP
Velocity Spectrum
Strength
0 - 100%
FL
Flow
Cubic meters per
second
VO
Volume
Cubic meters
FV
Forward volume
Cubic meters
RV
Reverse volume
Cubic meters
SV
Sampler Enabled
Volume
Cubic Meters
RA
Rain (rolls over every
255 tips)
Tips
4230 Bubbler Flow Meter
3-16
CR
Current day’s rain (tips
since midnight)
Tips
PR
Previous day’s rain (tips
since midnight)
Tips
PH
pH
pH units
DO
Dissolved Oxygen
Milligrams per liter
TE
Temperature
Degrees Celsius
YPH
YSI 600 pH
pH units
YDO
YSI Dissolved Oxygen
Milligrams per
liters
YCO
YSI 600 Conductivity
Millisiemens per
centimeter
YSP
YSI 600 Specific
Conductance
Millisiemens per
centimeter
YSA
YSI 600 Salinity
Parts per
thousand
YTD
YSI 600 Total Dissolved
Solids
Milligrams per liter
YTE
YSI 600Temperature
Degrees Celsius
YSP
YSI 600 Specific
Conductance
Millisiemens per
centimeter
YCO
YSI 600 Conductance
Millisiemens per
centimeter
YSA
YSI 600 Total Salinity
Parts per
thousand
YTD
YSI 600 Total Dissolved
Solids
Milligrams per liter
Table 3-1 ASCII Output Codes
Code
Parameter
Units
Section 3 Programming
3-17
Note
The output string for a given flow meter will have values
only for those parameters it is currently measuring. The
order of the fields in this table is subject to change.
Additional data types may be inserted anywhere in the
list. Parsing routines for this output string should search
by type identifier instead of depending on the position in
the string. If an active port has an error flag set, the
serial output will insert ERROR for the value.
You can enter the command line by connecting
the interrogator cable with the interrogator
sense line shorted to ground. Then send a series
of ‘?’ (question marks) until the flow meter
transmits the unit’s banner and prompt. The
number of question marks necessary is a
function of the baud rate auto detection. At the
prompt, enter DATA and the flow meter
will respond with the appropriate ASCII output
string. You can send the DATA command as
YTE
YSI 600 Sonde
Temperature
Degrees Celsius
SS
Sampler Enable Status
Logical
B?
Bottle Number and
Time
Days
CS
Check sum (does not
include the check sum,
carriage return, and
line feed)
Unsigned long
Table 3-1 ASCII Output Codes
Code
Parameter
Units
4230 Bubbler Flow Meter
3-18
often as you want. Type ‘Q’ to leave the
command response interface.
In addition to the port values, the data includes
the flow meter’s current time, the bottle
number and time stamp of the three most
recent sample events, the previous day’s
rainfall total (midnight to midnight), the
current day’s rainfall total since midnight, and
a rainfall tips counter that rolls over every 255
tips. The port values appear in a
comma-separated values format. Each data
field is preceded by a two or three-character
type identifier. The table lists the type
identifiers. Note that the flow meter’s current
time and the sample event time stamp appear
as a number in standard spreadsheet format
(days since 1900). The supported baud rates are
9600, 4800, 2400, and 1200 (no parity, eight bits
and one stop bit).
Periodic Output: (Use a special RAIN GAUGE
connector cable—contact the factory for
assistance.) Note that the periodic output will
terminate during phone connection and when
the interrogator cable is connected.
CAUTION
It is important to use CHECKSUM if you plan to use
internal modems or the interrogator. The UART is
shared with these devices.
Section 3 Programming
3-19
If you select SERIAL OUTPUT from the
OPTIONAL OUTPUTS menu, the following
display will appear:
Selection of OFF from this menu will disable
this feature, and there will be no further
references to it.
The Serial Output data appears on the
Interrogator connector of the flow meter. You
should not use a standard interrogator cable for
this application, as the sense line in the
standard cable is shorted to ground.
Selection of ON from this menu will enable the
feature and cause the following display to
appear:
After you select the appropriate baud rate, the
program will advance to the following menu:
This menu lets you select how often the flow
meter transmits the ASCII text string.
Following is an example of a string showing all
options: (Note that five lines are shown here;
this is due to the confines of the text column. In
actual practice, there are no carriage returns in
the text string.)
DE, “Theresa Street”, 4250 ALPHA 4,ID,0721577657,
PERIODIC SERIAL OUTPUT
• ON • • OFF
SELECT BAUD RATE (N81)
• 9600 • • 4800 • • 2400 • • 1200 •
SERIAL OUTPUT INTERVAL
• 15 SEC • • 1 MIN • • 5 MIN • • 15 MIN •
4230 Bubbler Flow Meter
3-20
MO,4250,TI,35317.343715,BV,12.3,LE,0.1000,VE,0.12
25,FL,0.001555,VO,2.199325,FV,2.199325,RV,0.00000
0,SV,2.195539,SS,1,B0,35317.307384,B0,35317.26990
7,B0,35317.232593,CS,10819
If you select any of these outputs, the flow
meter will request that you turn them on or off.
If you are running on battery and do not need
these options, select OFF. Otherwise, select ON.
After the OPTIONAL OUTPUTS menus have
been set, press Enter. The display will return to
the SETUP menu.
The alarm box, also called the High-Low Alarm
Box is an Isco product that allows you to
operate control relays to signal alarms when
flow rate rises above or falls below a certain set
value. You can set both the high and low alarm
values from 1 to 99% of the controlling
condition. (See Section 4 in the manual for
more information about the alarm box.)
The mechanical totalizer is an
electromechanical counter that is installed in
the flow meter front panel just above the
display. It is an optional feature; it is
non-resettable and displays seven digits.
If you select any of these outputs, the flow
meter will request that you turn them on or off.
If you are running on battery and do not intend
to use either of these options, select OFF.
Otherwise, select ON. After the OPTIONAL
OUTPUTS menus have been set, press Enter.
The display will return to the SETUP menu.
Section 3 Programming
3-21
This time, select REPORT SETUP with the
arrow key. Press Enter.
This step lets you determine the contents of the
reports generated by the flow meter. The report
generator can create two different reports (A
and B) that can be identical or very different.
The reason for two reports is to allow the
summary of flow meter recording over different
time periods. For example you might generate
report A weekly, and report B monthly. At this
point we are only interested in selecting the
items the flow meter will include in each report.
Press Enter and the following will appear:
FLOW METER HISTORY is just off the screen,
to the right. If you select FLOW and press
Enter, the following will appear:
Select YES if you want LEVEL to appear in the
report, then press Enter.
Select YES if you want FLOW RATE to appear
in the report, then press Enter again. Then:
REPORT SETUP
• REPORT A • • REPORT B •
REPORT SETUP
• FLOW • • DO/PH • • YSI 600 • • SAMPLE HISTORY •
LEVEL IN REPORT
• YES • • NO •
FLOW RATE IN REPORT
• YES • • NO •
RAINFALL IN REPORT
• YES • • NO •
4230 Bubbler Flow Meter
3-22
Select YES if you want RAINFALL to appear in
the report. Note that you must have a rain
gauge connected to the flow meter to sense
rainfall occurrence. Press Enter. The following
will appear:
FLOW METER HISTORY is just off the screen.
This time select DO/PH. Press Enter. The
following will appear:
Select YES if you want DO/PH to appear in the
report. Note that you must have the
appropriate sensor connected to the flow meter
to sense parameters; the flow meter is capable
of sensing temperature, pH and temperature,
and D.O. (dissolved oxygen) and temperature.
Press Enter. The following will appear:
Select yes if you want TEMPERATURE to
appear in the report. Press Enter again and the
display will return to the REPORT SETUP
menu:
FLOW METER HISTORY is just off the screen.
REPORT SETUP
• FLOW • • DO/PH • • SAMPLE HISTORY • • FLOW MET...
PH OR DO IN REPORT
• YES • • NO •
TEMPERATURE IN REPORT
• YES • • NO •
REPORT SETUP
• FLOW • • DO/PH • • YSI 600 • • SAMPLE HISTORY •
Section 3 Programming
3-23
This time select YSI 600. The following display
will appear:
Press Enter again and the display will return to
the REPORT SETUP menu.
Select SAMPLE HISTORY. Press Enter. The
following will appear:
Select YES if you want SAMPLE HISTORY to
appear in the report. Press Enter again and the
display will return to the REPORT SETUP
menu:
This time select FLOW METER HISTORY.
Press Enter. The following will appear:
Select yes if you want HISTORY to appear in
the report. HISTORY is a list of the changes
that have been made to the flow meter’s
program.
YSI DATA IN REPORT
• YES • • NO •
REPORT SETUP
• FLOW • • DO/PH • • YSI 600 • • SAMPLE HISTORY •
SAMPLE HISTORY IN REPORT
• YES • • NO •
REPORT SETUP
• YSI 600• • SAMPLE HISTORY• • FLOW METER HISTORY•
FLOW METER HISTORY IN REPORT
• YES • • NO •
4230 Bubbler Flow Meter
3-24
Press Exit to leave the program. Then press
Enter and reselect SETUP. The Setup menu
will reappear:
SET CLOCK, SITE ID, MEASUREMENT
SETUP, PROGRAM LOCK, PROGRAM are
off-screen and can be accessed with the arrow
keys.
Select LCD BACKLIGHT with the arrow key.
Press Enter. The following will appear:
KEYPRESS TIMEOUT will cause the backlight
to be turned on whenever you press a key on
the keypad (other than On and Off). An
internal timer is started that will keep the
backlight on for approximately two minutes
after you press a key.
Each time you press a key, the timer is
restarted, so the backlight will never go off as
long as you continue to program the flow meter,
with keystrokes coming less than two minutes
apart. At the end of programming, the
backlight will go out, and will stay out until you
start to program again.
This feature is designed to conserve battery
power by de-energizing the backlight when it is
not needed. The backlight is still available if it
is necessary to program in a dark environment,
such as a manhole. We recommend using this
SETUP OPTIONS: 'EXIT' TO QUIT
• STATUS • • REPORT SETUP • • LCD BACKLIGHT •
LCD BACKLIGHT MODE
• KEYPRESS TIMEOUT • • CONTINUOUS • • OFF •
Section 3 Programming
3-25
selection if the flow meter is battery-powered,
but installed in an environment where the
lighting is poor.
CONTINUOUS will cause the backlight to be
lit continuously. Where the flow meter is
powered by an AC power supply, battery life
considerations do not intervene. If the backlight
makes the display easier to read, use it. Do not
use CONTINUOUS in any installation that is
battery-powered, as it will run the battery
down in a matter of hours.
OFF will keep the backlight feature turned off
under all circumstances. Select this option for
maximum battery life in installations where
there is sufficient ambient light to read the
display without the backlight feature.
Press Enter. The SETUP menu will return.
This time move the flashing cursor from LCD
BACKLIGHT to LANGUAGE.
When LANGUAGE appears on your display,
you may select an alternate language to
program the flow meter. The other language
depends on how the flow meter was ordered.
The following display will appear:
Select the language appropriate for your
application. The menus and the printed reports
will appear in the selected language. Press
Enter. The SETUP menu will reappear. This
LANGUAGE
• ENGLISH • • (second language, as ordered) •
4230 Bubbler Flow Meter
3-26
time select PROGRAM LOCK from the menu.
Press Enter:
PROGRAM LOCK keeps the program from
being changed. Select OFF while you are
programming, and then go back and select ON
if you need to lock the program. At that, we
suggest using the lock only if there are
compelling security reasons.
Further changes will require entry of the
password, which is the model number of the
flow meter, 4230. If you select ON, there is a
time-out before the lock engages.
If you continue to work through the rest of the
program, the lock will not engage until you are
done. But if you stop programming longer than
two minutes, the lock will engage, and you will
not be able to make any further program
changes.
Note
If you choose NOT MEASURED for any selection, the
flow meter will make no further reference to that function
for the rest of the program, and you will be unable to
activate that function later. If there is a feature you need
that does not appear when the manual says it should,
return to step 1 and make sure you have not
accidentally left it turned off.
If you program a parameter value as a
condition for sampler enabling, pacing, dialout,
etc., and then turn that parameter sensor off,
PROGRAM LOCK
• ON • • OFF •
Section 3 Programming
3-27
the flow meter will also remove that condition
from the program.
Consider all aspects of your program before you
make any changes!
Press Enter and the SETUP menu will
reappear. step 1 - Program:
PROGRAM will be flashing. (Note that
PROGRAM is always the default choice. That is
because you are more likely to need to make
changes in the PROGRAM section of the
software than in the SETUP section.)
Press Enter. The following will appear:
Selection of feet, inches, meters or mm depends
on your situation. You would select NOT
MEASURED if you were using the flow meter
for some other form of sensing only, such as pH,
or temperature. Press Enter. The following will
appear. You will have to press the right arrow
key several times to see all of the options
displayed below:
Pressing the right arrow key several times will
move other units onto the display:
SELECT OPTION
• PROGRAM • • SETUP •
UNITS OF LEVEL MEASUREMENT
• FT • • IN • • M • • MM • • NOT MEASURED •
FLOW RATE UNITS OF MEASURE
• GPS• •GPM• •GPH• •MGD• •CFS• •CFM• •CF-
FLOW RATE UNITS OF MEASURE
• CFH••CFD••LPS••M3S••M3M••M3H••M3D••AFD •>
4230 Bubbler Flow Meter
3-28
NOT MEASURED will also appear if you keep
moving with the right arrow key. This step
establishes the units of measure the flow meter
will use in all subsequent displays and
calculations.
This step determines the units the flow meter
will use to record the totalized flow that passes
by.
For the next several substeps you must have
the appropriate sensor attached to the flow
GPS
gallons per second
GPM
gallons per minute
GPH
gallons per hour
MGD
millions of gallons per day
CFS
cubic feet per second
CFM
cubic feet per minute
CFH
cubic feet per hour
CFD
cubic feet per day
LPS
liters per second
M3S
cubic meters per second
M3M
cubic meters per minute
M3H
cubic meters per hour
M3D
cubic meters per day
AFD
acre-feet per day.
TOTALIZED VOLUME UNITS
• GAL • • MGAL • • CF • • L • • M3 • • AF •
GAL
gallons
MGAL
millions of gallons
CF
cubic feet
L
liters
M3
cubic meters
AF
acre-feet
Section 3 Programming
3-29
meter’s Parameter Port or Rain Gauge Port (for
the Rain Gauge or YSI 600 Sonde). You can
only have D.O./temperature, pH/temperature or
temperature alone on a given flow meter, unless
you use the YSI 600 Sonde. The YSI 600 Sonde
provides multiple outputs simultaneously. You
can use the Rain Gauge with the YSI 600 sonde
if you use a special Y-connect cable.
Remember that programming for either pH or
D.O. will prevent the display of any references
to the other in later program steps. All sensors
but the rain gauge must be fully submerged in
the flow stream. The pH and D.O. probes must
be kept constantly wet.
The pH or D.O. probes cannot attach directly to
the flow meter, as their output signals are low.
You must also install the appropriate amplifier
box between the probes and the flow meter.
Note
The 270 D.O. module has been discontinued. Probes,
service kits, and accessories are still available to
maintain existing field units.
Their use in streams with intermittent flow
(such as storm drainage) is not recommended.
See Section 4 in the manual (Accessories) for
more information about the pH, D.O. probes,
and rain gauge. Note that the pH probe is a
consumable item, and will eventually need
replacement in any case.
RAIN GAUGE
• INCHES • • MM • • NOT MEASURED •
4230 Bubbler Flow Meter
3-30
You must have an Isco 674 Rain Gauge
connected to the flow meter through the Rain
Gauge Port to sense rainfall. MM = millimeters.
The rain gauge is factory-calibrated. If you do
not want to use a rain gauge, you would select
NOT MEASURED for this step.
Note
If you choose NOT MEASURED, no more references to
rainfall will appear in the rest of the program.
pH measurement determines the relative
acidity or alkalinity of a solution. You must
have an Isco pH Probe (or approved equivalent)
connected to the flow meter through the
parameter port to sense pH. pH measurements
range from 0 to 14 pH units, with solutions
below 7 considered acidic and solutions above 7,
alkaline. Pure water has a pH of 7. These
devices require periodic calibration for accurate
sensing of pH. See Section 4 (Accessories) in the
manual for more information about the pH
probe.
If you are not using a pH probe you would
select NOT MEASURED.
• If you choose NOT MEASURED for pH,
you will be unable to activate pH later in
the program.
pH UNITS OF MEASURE
• pH • • NOT MEASURED •
Section 3 Programming
3-31
• If you select pH, you will be unable to
activate and measure D.O. later in the
program.
• If your situation requires the
measurement of both parameters (pH
and D.O.) at the same time, or if you also
need to measure conductivity, use the
YSI 600 Multi-Parameter Sonde.
If you select NOT MEASURED for pH and
press Enter, the following will appear:
Measurement of dissolved oxygen is conducted
in studies of water quality in lakes and
streams. Some dissolved oxygen is necessary
for the survival of aquatic life in these waters.
You must have an Isco Dissolved Oxygen Probe
to sense dissolved oxygen. The probe attaches
to the module, which attaches to the Parameter
Port. These devices require periodic calibration
for accurate sensing. PPM = parts per million;
MG/L = milligrams per liter. If you are not
using the D.O. probe, you would select NOT
MEASURED for this step.
Note
Selection of NOT MEASURED will keep D.O. from
appearing on subsequent programming menus.
D. O. UNITS
• MG/L • • PPM • • NOT MEASURED •
4230 Bubbler Flow Meter
3-32
Selection of PPM or MG/L will keep references
to pH from showing up on subsequent menus.
This step sets up measurement of the
temperature of the flow stream. You must have
an Isco Temperature Probe attached to the flow
meter’s Parameter Port. The temperature probe
contains a thermistor and needs no further
calibration.
Measurement is in degrees Celsius or degrees
Fahrenheit. If you are not using the
temperature probe, select NOT MEASURED.
Note
Selection of NOT MEASURED will keep
TEMPERATURE from appearing in subsequent
programming menus.
Note that if you are using either the pH or D.O.
probe, temperature must be measured; the
NOT MEASURED option will not even appear.
The following menus concern the use of the YSI
600 Multi-Parameter Sonde. This probe allows
you to measure several different characteristics
of a flow stream at the same time. The YSI 600
Sonde attaches to the Rain Gauge connector on
the 4230 Flow Meter. Note that this connector
must be a special, modified connector with nine
pins.
TEMPERATURE UNITS
• DEG F • • DEG C • • (NOT MEASURED) •
Section 3 Programming
3-33
Note
The 4230 Flow meters with 4-pin Rain Gauge
connectors cannot support the YSI 600. It is necessary
to return the flow meter to the factory for modifications.
In addition to the connector, there are significant
internal modifications to the flow meter’s electronics and
software.
You can use both the YSI Sonde and a Rain
Gauge on flow meters that support the YSI
with a special Y-connect cable. Note that the
YSI 600 Sonde differs from the
previously-mentioned pH and D.O. probes. The
YSI 600 can measure pH and D.O. at the same
time, as well as temperature and conductivity.
If you are not using the YSI 600 sonde, select
NO in the following display and the flow meter
will advance to the next step. Otherwise, select
YES.
Note
If you select NO, you will be unable to activate the YSI
Sonde later in the program.
If no communication has been confirmed, the
following will appear:
YSI 600 CONNECTED
• YES • • NO •
YSI COMMUNICATIONS CHECK
• YES • • NO •
4230 Bubbler Flow Meter
3-34
NO is the default. If you select YES, the
following will appear:
CAUTION
Do not disconnect either the sonde or flow meter
power during a communications check. The memory
in the sonde can be damaged by a power failure
occurring during an update.
If the communication check is bad, the
following will appear:
Note
The flow meter cannot communicate at 600 baud. If
your sonde has been set up for 600 baud, you will get a
communications failure. Consult the YSI 600 Manual for
what to do in this case.
If the communications check is good, the
following will appear:
After you press ENTER, the flow meter will
advance to the following display:
WARNING! DO NOT DISCONNECT POWER
YSI COMMUNICATION CHECK. PLEASE WAIT...
COMMUNICATIONS CHECK FAILED
PRESS ENTER TO CONTINUE
COMMUNICATIONS RATE SET AT 2400 BAUD
PRESS ENTER TO CONTINUE
YSI 600 pH UNITS OF MEASURE
• pH • • NOT MEASURED •
Section 3 Programming
3-35
Note
Selection of NOT MEASURED from any of the YSI
menus will keep you from activating that function later in
the program.
If you wish to make use of the YSI 600 sonde’s
pH measurement capability, select pH. If you do
not, select NOT MEASURED.
If you wish to make use of the YSI 600 sonde’s
D.O. measurement capability, select MG/L.
Otherwise, select NOT MEASURED.
For any of the YSI CONDUCTIVITY options, if
you press Enter, the following display will
appear:
This value is provided because conductivity
rises (~2% / °C) with temperature. The default
setting is 1.91%. This value is the temperature
coefficient for pure KCl (potassium chloride) in
water. For other salts this value will be
somewhat inaccurate, but it does provide a
close approximation for solutions of many
common salts, such as NaCl (sodium chloride),
NH4Cl (ammonium chloride) and sea water. If
you use the value of 1.91, in most cases you will
YSI 600 D.O. UNITS OF MEASURE
• MG/L • • NOT MEASURED •
YSI 600 CONDUCTIVITY PARAMETER
•YSI SP COND••YSI SALINITY••YSI CONDUCTIVITY•>
TEMPERATURE COEFFICIENT
1.91%
4230 Bubbler Flow Meter
3-36
be able to identify that gross changes are
occurring in the ionic content of the stream.
If you move with the right arrow, the following
options will appear on the display:
T.D.S. stands for “total dissolved solids.” T.D.S.
are measured in parts per thousand (ppt).
Total dissolved solids are estimated by
multiplying conductivity by an empirical factor.
This factor can vary between 0.55 and 0.9
depending on the solubility of the ionic
components in the water and its temperature.
Select the temperature units appropriate for
your application.
3.2 Step 2 - Flow Conversion
METERING INSERTS will also appear.
WEIR/FLUME = weir or flume; EQUATION =
equation; MANNING = Manning; DATA
POINTS = data points.
YSI 600 CONDUCTIVITY PARAMETER
• YSI T.D.S. • • NOT MEASURED •
TDS SCALE FACTOR
0.75
YSI 600 TEMPERATURE UNITS
• °F • •°C • • NOT MEASURED •
FLOW CONVERSION TYPE
•WEIR/FLUME• •EQUATION• •MANNING• •DATA POINTS•>
Section 3 Programming
3-37
If you select WEIR/FLUME, the following
display will appear:
For detailed information on weirs and flumes,
refer to the Isco Flow Measurement Handbook,
available from Teledyne Isco. Consulting the
manufacturer of the specific weir or flume is
also worthwhile. Note that for weirs and
flumes, there is a preferred location for
installing the level measuring device. Proper
mounting of the level measuring device and
accurate measurement of the level in the flow
stream at the calibration point are essential for
accurate flow calculation.
If you select WEIR, the following display will
appear:
If you select V-NOTCH, the following will
appear:
If you select RECTANGULAR for weir, the
following will appear:
If you select YES, the following will appear:
TYPE OF DEVICE:
• WEIR • • FLUME •
SELECT TYPE OF WEIR:
• V-NOTCH • • RECTANGULAR • • CIPOLLETTI •
SELECT V-NOTCH WEIR ANGLE (IN DEGREES)
• 22.5 • • 30 • • 45 • • 60 • • 90 • • 120 •
END CONTRACTIONS ON RECTANGULAR WEIR:
• YES • • NO •
RECTANGULAR WEIR WITH END CONTRACTIONS
ENTER CREST LENGTH XX.XXX FEET (or meters)
4230 Bubbler Flow Meter
3-38
If you select CIPOLLETTI for weir, the
following will appear:
If you selected FLUME for the type of standard
device, the following display will appear:
Also available with the arrow key:
If you select PARSHALL for flume, the
following will appear:
If you press the right arrow key several times,
the sizes shown below will move onto the
screen:
If you select PALMER-BOWLUS for flume, the
following will appear:
If you press the right arrow key several times,
the sizes shown below will move onto the
screen:
CIPOLLETTI WEIR
ENTER CREST LENGTH XX.XXX FEET (or meters)
SELECT TYPE OF FLUME
• PARSHALL • • PALMER-BOWLUS • • LEOPOLD-LAGCO •
SELECT TYPE OF FLUME
• HS • • H • • HL • • TRAPEZOIDAL •
SELECT PARSHALL SIZE:
•1"• •2"• •3"• •6"• •9"• •1.0'• •1.5'• •2.0'•
SELECT PARSHALL SIZE:
• 3' • • 4' • • 5' • • 6' • • 8' • • 10' • • 12' •
SELECT PALMER-BOWLUS SIZE
•4"• •6"• •8"• •9"• •10"• •12"• •15"• •18"• •21"
SELECT PALMER-BOWLUS SIZE
• 24" • • 27" • • 30" • • 48" •
Section 3 Programming
3-39
If you select LEOPOLD-LAGCO for flume, the
following will appear:
If you press the right arrow key several times,
the sizes shown below will move onto the
screen:
If you select HS for flume, the following display
will appear:
If you select H for the type of flume, the
following will appear:
If you select HL for the type of flume, the
following will appear:
If you select TRAPEZOIDAL for flume, the
following will appear:
(This completes the section on WEIR/FLUME
flow conversions.)
Returning to step 2, SELECT FLOW
CONVERSION.
LEOPOLD-LAGCO FLUME SIZE
• 4"• •6"• •8"• •10"• •12"• •15"• •18"• •21" •
LEOPOLD-LAGCO FLUME SIZE
• 24" • • 30" •
HS FLUME SIZE
• 0.4' • • 0.5' • • 0.6' • • 0.8' • • 1.0' •
H FLUME SIZE
• 5'• •.75'• •1'• •2'• •2.5'• •3'• •4.5'•
HL FLUME SIZE
• 2.0' • • 2.5' • • 3.0' • • 3.5' • • 4.0' •
TRAPEZOIDAL SIZE
• LG 60 V • • 2" 45 WSC • • 12” 45 SRCRC •
4230 Bubbler Flow Meter
3-40
If you select EQUATION the following display
will appear:
This step allows you to enter an equation that
is appropriate for your flow situation. The
equation is expressed in the general form of Q =
k1HP1 + k2HP2, where Q = flow rate, k1 = a
constant, H = level or head, and P1 is the power
to which k1 and H are raised. k2 and P2 are a
second constant and power found in some
equations. If your equation has only one term,
you should enter 0 for the second constant.
Again returning to step 2, SELECT FLOW
CONVERSION, if you select MANNING, the
following display will appear:
TRAPEZOIDAL is also available, if you move to
the right with the right arrow key.
If you select ROUND PIPE for the Manning
flow conversion, the following displays will
appear:
Slope is entered as a dimensionless quantity,
delta Y/ delta X, not as percent slope. Or, as
otherwise expressed:
ΔX = Rise
For example:
1
= .01
ΔY
Run
100
ENTER EQUATION UNITS
Q = XXX.XXXH^X.XX + XXX.XXXH^X.XX
SELECT MANNING TYPE
• ROUND PIPE • • U-CHANNEL • • RECTANGULAR • • T–
MANNING ROUND PIPE
SLOPE = X.XXXXX ROUGH = X.XXXX
Section 3 Programming
3-41
Roughness coefficients are published in the Isco
Open Channel Flow Measurement Handbook.
You must know the material the pipe is made
of. The roughness coefficients are published for
all common materials in three grades:
minimum, normal and maximum. Then:
If you select U-CHANNEL for the Manning
flow conversion, the following displays will
appear:
(Slope and roughness are entered as for
ROUND PIPE previously.) Then:
If you select RECTANGULAR for the Manning
flow conversion, the following displays will
appear:
(Slope and roughness are entered as for
ROUND PIPE.) Then:
MANNING ROUND PIPE
DIAMETER = X.XXX FEET (or meters)
MANNING U-CHANNEL
SLOPE = X.XXXXX ROUGH = X.XXX
MANNING U-CHANNEL
WIDTH = X.XXX FEET (or meters)
MANNING RECTANGULAR
SLOPE = X.XXXXX ROUGH = X.XXX
MANNING RECTANGULAR
WIDTH = X.XXX FEET (or meters)
4230 Bubbler Flow Meter
3-42
If you select TRAPEZOID for the Manning flow
conversion, the following displays will appear:
(Slope and roughness are entered as for
ROUND PIPE.) Then:
Then:
Returning to step 2, FLOW CONVERSION
TYPE, if you select DATA POINTS, the
following displays will appear:
Then:
This allows you to enter data points that are in
different units than you are using. Then:
If you press the right arrow key a few times, the
following units of measure will appear on the
display:
MANNING TRAPEZOID
SLOPE = X.XXXXX ROUGH = X.XXX
MANNING TRAPEZOID
TOP WIDTH = X.XXX FEET (or meters)
MANNING TRAPEZOID
BOTTOM WIDTH = X.XXX FEET (or meters)
SELECT DATA SET
• ONE • • TWO • • THREE • • FOUR • • (NONE) •
LEVEL UNITS FOR DATA POINT ENTRY
• FT • • IN • • M • • MM •
FLOW RATE UNITS
• GPM• •GPS• •MGD• •CFS• •CFM• •M3S• •M3H• •M3D•
FLOW RATE UNITS
• LPS• •CFD• •GPH• •AFD• •CFH• •CFM• •M3M•
Section 3 Programming
3-43
DATA POINT flow conversion allows you to
enter measured level and flow rate values for a
number of different points. The 4230 Flow
Meter can accept as many as four sets of data
points with each set containing up to fifty
points.
The flow meter then performs a three-point
interpolation to calculate a flow rate
appropriate for the data entered. The common
use of data point flow conversion is with
unusual primary measuring devices,
specifically devices that the 4230 Flow Meter
does not support in its internal flow conversion
set.
The level-to-flow rate data for such devices is
usually available from the manufacturer. From
this the flow meter can create a conversion
based on the relationship between the level and
flow rate.
GPM
gallons per minute
GPS
gallons per second
MGD
million gallons per day
CFS
cubic feet per second
CFM
cubic feet per minute
M3S
cubic meters per second
M3M
cubic meters per minute
M3H
cubic meters per hour
M3D
cubic meters per day
LPS
liters per second
CFD
cubic feet per day
GPH
gallons per hour
AFD
acre-feet per day
CFH
cubic feet per hour.
4230 Bubbler Flow Meter
3-44
After the FLOW RATE UNITS menu has
appeared, the next menu is:
Then:
After you have entered the data set the
following will appear:
UNITS, SAVE will also appear if you move the
flashing cursor with the right arrow key. USE
will only appear after four points have been
entered. USE tells the flow meter that the set is
complete and can be used for the flow rate
calculation.
Select EDIT POINT if you need to change
either the level or the flow value for a
particular data point.
Select ADD POINT if you want to add another
point to a data set.
CLEAR erases an entire set of data points from
the memory.
PRINT will make the flow meter print out the
entire data set.
UNITS allows you to set or change the units of
measure used in the data set. Note that you can
only set UNITS if the set is empty, or you have
cleared it. You cannot change the units once you
SET X (1-4): (0) POINTS ENTERED
• ADD POINT • • (UNITS) •
SET 1 DATA POINT 1
ENTER: 0.00 (level units) 0.000 (units of volume)
SET X (1-4): XX (1-50) POINTS ENTERED
• (USE)• •EDIT POINT• •ADD POINT• •CLEAR• •PRINT•
Section 3 Programming
3-45
have entered data points into a set unless you
clear it and start over.
SAVE will save the data set as it is.
If you select either EDIT POINT or ADD
POINT, the following will appear:
3.2.1 Flow Metering Inserts
There is one more possible flow conversion for
the 4230 Bubbler Flow Meter. This option will
appear as FLOW METERING INSERTS. These
devices are metal inserts for 6", 8", 10", and 12"
diameter round pipe sewers. The inserts form a
primary measuring device inside the pipe when
they are placed.
The inserts are installed from street level to a
maximum depth of 16 feet with a handle made
up of snap-fit extension pipes. The inserts are
sealed into the upstream pipe by inflating a
rubber bladder. Each insert is supplied with
two circular weir plates that provide either a
round orifice or V-notch opening. Once
installed, the flow backs up behind the insert
and a bubbler tube in the bottom of the insert
measures level. The flow meter calculates flow
based on this measured level. If you select
METERING INSERTS for the FLOW
CONVERSION mode, the following menu will
appear:
SET X (1-4) DATA POINT XX (1-50)
ENTER: XX.XX (level units) XXX.XXX (volume)
SELECT WEIR/ORIFICE TYPE
• V-NOTCH • • ROUND •
4230 Bubbler Flow Meter
3-46
Then:
3.2.2 Enter Maximum Head - All Models
Before advancing from step 2 (Flow
Conversion) to step 3 (Adjust Ports), the flow
meter will request that you enter a value for
Maximum Head (Level) for the device or flow
conversion you are using.
For most standard measuring devices this
information is published or is available from
the device manufacturer. Note, however that
you should not arbitrarily use the largest value
available. Instead, use the value that is the
largest expected level for your actual situation,
even if this is less than the published
maximum.
The flow meter’s internal resolution and its
accuracy are based on the value you enter for
Maximum Head. The flow meter will display:
3.2.3 Programming the 4-20 mA Outputs
If you turn on any of the 4-20 mA outputs
(ANALOG OUTPUTS) in step 1- Setup,
programming the actual operation of the output
appears in step 2 - Select Flow Conversion.
SELECT FLOW INSERT SIZE
• 6" • • 8" • • 10" • • 12" •
FLOW RATE AT MAXIMUM HEAD
X.XXX CFS (or other units of measure)
Section 3 Programming
3-47
Note
If you do not turn on the 4-20 mA output(s) in step 1, the
menus determining its (their) operation will not even
appear later in the program. If you need this function
and cannot find the appropriate menus in step 2, return
to step 1, Setup and check to see that you have not
inadvertently switched the option off.
For each 4-20 mA output port turned on, the
flow meter will request entry of the type of data
that will drive the output, along with minimum
and maximum values. Here is an example of
what you might see for programming analog
output 1.
TEMPERATURE, DISSOLVED OXYGEN,
CONDUCTIVITY, SPECIFIC
CONDUCTANCE, SALINITY, and TOTAL
DISSOLVED SOLIDS may also appear as
driving conditions.
The actual choices available to you will depend
on what ports you have turned on previously
and what accessories (YSI, pH, DO probes, rain
gauge, etc.) you are using with your flow meter.
After you select one of the choices available to
you, the flow meter will request that you enter
minimum and maximum values for that choice:
Note that 4 MA in the second line of the display
could also be 0 MA if that is what you selected
DATA TYPE FOR ANALOG OUTPUT 1
• (OFF)••(LEVEL)••(FLOW RATE)••(VELOCITY)••(pH)•
ANALOG OUTPUT PORT 1
4 MA = X.X (units)
4230 Bubbler Flow Meter
3-48
for the current loop minimum in Setup. The
units are the units of measure appropriate for
the option you selected; for example, feet or
meters for level, degrees F or C for
temperature, mg./l for dissolved oxygen, etc.
After you have set the minimum value for the
port, the flow meter will request you to enter a
value for full-scale, or 100%:
This value causes the port to transmit 100% or
20 mA. For example if the data type selected for
this output were level, and the unit is
measuring level in a four-foot pipe, you would
enter a full-scale value of four feet. If the actual
level reading is currently two feet, the analog
output would read 12 mA (50% if the 4-20mA
current range is selected) or 10 mA (50% if the
0-20 mA current range is selected).
The flow meter will then request that you
repeat the process of defining the data type and
setting the minimum and maximum values for
any of the other analog ports you activated
previously in Setup.
3.3 Step 3 - Port to Adjust
This step lets you enter the measured level in
the flow stream. It also lets you calibrate the
pH (acidity or alkalinity), D.O. (dissolved
oxygen) parameter sensors, and the YSI 600
Multiple Parameter Sonde. Note that there is
no calibration step for the temperature sensor
ANALOG OUTPUT PORT 1
20 MA = X.X (units)
Section 3 Programming
3-49
because it doesn’t need calibration. When you
select step 3 the following will appear:
LEVEL will not show up if you are using the
flow meter only for parameter sensing.
Likewise, pH and/or D.O. and YSI 600 will not
show up on the display if you have locked them
out by programming selections you made in
step 1.
Remember that selection of either pH or D.O. in
step 1 will keep the other from appearing on
the display in this or subsequent programming
steps. If the parameter you want does not
appear in this menu, exit the program and
return to Setup. Check to see that you have not
accidentally locked out your choice with
selections made in the early part of the
program.
If you select NONE, the flow meter will
advance to the next step. If you select LEVEL,
the following will appear:
For this value, you must measure the level in
the flow stream. This is usually done with a
measuring stick. Generally, you should
measure the level upstream from the bubble
line outlet, or at the prescribed point in the
primary measuring device. The bubbler should
be installed in an area of stable flow.
PORT TO ADJUST
• NONE• •(LEVEL)• •(pH)• •(D.O.)• •(YSI 600)•
ENTER CURRENT LEVEL
X.XXX FEET (or other units of measure, as selected)
4230 Bubbler Flow Meter
3-50
You should remove the bubbler and mounting
ring to measure the level if either of the
following occur:
• If it is not possible to measure the level
upstream when the bubbler is installed,
(cramped quarters inside smaller pipes).
• If the bubbler and its mounting ring
create a visible “jump” in the stream.
This “jump” would be any noticeable
disturbance on the surface of the flow stream
where it passes over the bubbler and mounting
ring, and is typical of lower flows. See Section 3,
Installing the Bubble Line, (in the instruction
manual) for more information.
Enter this value with the number keys. Note
that LEVEL ADJUST must be done at the job
site, while most other programming can be
done in the shop. Unless you have reliable
information about the size of the channel, you
should measure that, also.
Note
It is very important to enter accurate measurements for
both the level in the stream and the dimension(s) of the
channel, as all calculations of flow will be based on
these measurements. If the values entered are
incorrect, even by relatively small amounts, all
subsequent flow calculations will be incorrect also.
For example, an error of only 1/4" for a 3" level
and 1/4" for a 10" diameter round pipe can result
in a combined error of over 14%!
Section 3 Programming
3-51
Errors in level measurement have a greater
effect on flow calculations at low liquid levels.
Dimensional errors are more significant at
higher levels.
If you select pH for port to adjust, the following
display will appear. (Note that pH will not
appear as an option unless you have selected it
in step 1.) If you want to measure pH and pH
does not appear in this step, you must go back
to step 1 and select pH instead of NOT
MEASURED.
You can perform a two- or three-point
calibration for pH with the pH sensor. Select
the calibration that best suits your stream’s
profile. If the pH in your stream in generally
below 7, you would probably select pH 4 & 7. If
the pH is generally above 7, you would probably
select pH 7 & 10. If your stream’s pH varies a
great deal, say from 3 to 12, your best choice
would be pH 4, 7, & 10. Then:
The flow meter will direct you to repeat this
process with the other standard buffers (7
and/or 10) to calibrate the pH sensor. If the
probe fails to provide the correct output with
any of the buffer solutions you will receive the
following message:
pH CALIBRATION
• pH 4 & 7 • • pH 7 & 10 • • pH 4, 7, & 10 •
RINSE PROBE AND PLACE IN 4.0 pH SOLUTION
PRESS ENTER WHEN STABLE X.XX pH
pH BUFFER/PROBE OUT OF RANGE
PRESS ENTER TO CONTINUE
4230 Bubbler Flow Meter
3-52
If you select D.O. for port to adjust, the
following will appear:
ALTITUDE is just off screen to the right. If you
select D.O. STANDARD for the calibration
method, the following display will appear:
If you select ABS (absolute) BAROMETRIC
PRESSURE the following will appear:
Note that absolute barometric pressure is
barometric pressure not corrected to sea level.
The barometric pressure published by the U.S.
Weather Bureau is corrected to sea level. If you
use their value, you must convert it to the
absolute pressure for your altitude. You should
use Weather Bureau barometric pressure only
if you are at sea level, or are able to correct the
Weather Bureau figure to absolute pressure at
your location. Enter the value for atmospheric
pressure.
Then the display will advance to the following:
DISSOLVED OXYGEN CALIBRATION
• D.O. STANDARD • • ABS BAROMETRIC PRESSURE • >
D.O. STANDARD
0.00 MG/L
ABS BAROMETRIC PRESSURE
X.XX mmHg
WRAP D. O. PROBE IN MOIST CLOTH
PRESS ENTER WHEN STABLE: X.XX MS/CM:
CALIBRATING...
PLEASE WAIT...
Section 3 Programming
3-53
If you select ALTITUDE for D.O., the following
will appear:
Select the appropriate units and press Enter.
Enter the altitude for your location. Then:
For more detailed information on the pH and
D.O. probes, see Section 4 in the instruction
manual.
Returning to step 3, Parameter to adjust, the
following will appear:
Note
If you are using the YSI Sonde and YSI 600 does not
appear on your display, return to step 1, Program, and
make sure you have selected YES from the YSI
CONNECTED menu.
If you select YSI 600, the following display will
appear:
UNITS FOR ALTITUDE ENTRY
• FT • • M •
ALTITUDE
ALTITUDE = X.XX FT (or meters)
WRAP D. O. PROBE IN MOIST CLOTH
PRESS ENTER WHEN STABLE: X.XXX MG/L
PARAMETER TO ADJUST
• NONE • • (LEVEL) • • (pH) • • (DO) • • (YSI 600) •
YSI 600 PARAMETER TO CALIBRATE
• NONE • • pH • • DO • • CONDUCTIVITY •
4230 Bubbler Flow Meter
3-54
If you select pH for the parameter, the following
will appear:
You can perform a two- or three-point
calibration for pH with the YSI sonde. The
menus that follow are similar to those in the
preceding section for the Isco pH sensor. Select
the calibration that best suits your stream’s
profile. When you complete the pH calibration
successfully, the following display will appear:
Returning to the YSI menu, if you select D.O:
ALTITUDE is just off screen to the right.
Programming for YSI 600 D.O. is essentially
the same as that described for the Isco D.O.
sensor on the preceding section, with the
exception that you always place the sensor in a
cup, rather than wrap a moist cloth around it as
is done for the Isco D.O. sensor.
If you select CONDUCTIVITY for the
parameter to calibrate, the following display
will appear:
MS/CM is milli-siemens per centimeter. The
siemen is the S.I. (Système Internationale)
YSI 600 pH CALIBRATION
• pH 4 & 7 • • pH 7 & 10 • • pH 4, 7, & 10 •
CALIBRATING...
PRESS ENTER TO CONTINUE
YSI 600 DISSOLVED OXYGEN CALIBRATION
• D.O. STANDARD • • ABS BAROMETRIC PRESSURE •>
CONDUCTIVITY CALIBRATION UNITS
• MS/CM • • PPT •
Section 3 Programming
3-55
name for the unit of conductance, which is also
the reciprocal of the ohm. The siemen was
formerly called the mho (ohm spelled
backwards), and that term is more familiar to
some. PPT is parts per thousand. Select the
standard most suitable for your application.
Then:
Then:
If you select PPT for the conductivity standard:
Then:
Then:
Note that there is no need to calibrate the YSI
600 temperature sensor, as it is self-calibrating.
CONDUCTIVITY STANDARD
X.XX MS/CM
PLACE PROBE IN X.XX MS/CM
PRESS ENTER WHEN STABLE: X.XX MS/CM
CALIBRATING...
PLEASE WAIT...
CONDUCTIVITY STANDARD
X.XX PPT
PLACE PROBE IN X.XX PPT
PRESS ENTER WHEN STABLE: X.XX MS/CM
CALIBRATING...
PLEASE WAIT...
4230 Bubbler Flow Meter
3-56
3.4 Step 4 - Reset Totalizer
This step lets you reset the flow meter’s
internal flow totalizer. Note that there is the
possibility of more than one totalizer. The 4230
Flow Meter can maintain a separate totalizer
for the time the sampler is enabled through the
sampler enabling feature (step 6).
If you select step 4, the following will appear: If
you select NO, the flow meter will advance to
the next step. If you select YES, the flow meter
will reset the internal totalizer to zero.
Then, if you have sampler enabling turned on
(step 6), the flow meter will ask you whether
you want to reset the totalizer for the total flow
during the time the sampler was enabled. (See
step 6 for an explanation of sampler enabling.)
(Other units of measure may appear.) Then:
3.5 Step 5 - Sampler Pacing
This step determines how the flow meter will
signal an associated automatic wastewater
sampler to take a sample. The flow meter and
sampler must be connected together with a
cable. The flow meter sends flow pulses to the
sampler. The sampler uses these flow pulses as
RESET TOTALIZER
• YES • • NO •
ENABLE TOTALIZER XX CF
PRESS 'ENTER
RESET SAMPLER ENABLE TOTALIZER
• YES • • NO •