Geotech Keck Interface Probe Installation And Operation Manual
Read Online
Read the Geotech Keck Interface Probe Installation And Operation Manual online using the built-in PDF viewer.
Text of the manual
Table of Contents
CHAPTER 1: SYSTEM DESCRIPTION..................................................................... 5
FUNCTION AND THEORY ................................................................................................ 5
SYSTEM COMPONENTS................................................................................................... 6
CHAPTER 2: SYSTEM INSTALLATION.................................................................. 7
CHAPTER 3: SYSTEM OPERATION ......................................................................... 8
KECK KIR DNAPL CORRECTION FACTORS........................................................ 9
CHAPTER 4: SYSTEM MAINTENANCE................................................................ 10
CHAPTER 5: SYSTEM TROUBLESHOOTING ..................................................... 11
TROUBLESHOOTING CONTINUED.................................................................................. 12
CHAPTER 6: SYSTEM SPECIFICATIONS............................................................. 13
CHAPTER 7: SYSTEM SCHEMATIC...................................................................... 15
CHAPTER 8: REPLACEMENT PARTS LIST......................................................... 16
THE WARRANTY ........................................................................................................ 19
EQUIPMENT RETURN POLICY............................................................................... 19
EQUIPMENT DECONTAMINATION....................................................................... 19
This uses the following conventions to present information:
An exclamation point icon indicates a WARNING
of a situation or condition that could lead to
personal injury or death. You should not proceed
until you read and thoroughly understand the
WARNING message.
WARNING
CAUTION
A raised hand icon indicates CAUTION information
that relates to a situation or condition that could lead
to equipment malfunction or damage. You should not
proceed until you read and thoroughly understand the
CAUTION message.
A note icon indicates NOTE information. Notes
provide additional or supplementary information
about an activity or concept.
Chapter 1: System Description
Function and Theory
The Keck Interface Probe is a portable instrument capable of accurately
measuring water and hydrocarbon levels and hydrocarbon layer thickness
in monitoring wells. The probe consists of a stainless steel and Teflon®
probe attached to a reel-mounted, Tefzel* coated engineer's tape. The
engineer's tape comes in standard or metric graduations and is accurate
to 1/100 of a foot. The probe has a float which detects hydrocarbon
levels and a pair of stainless steel contacts for sensing conductive fluids.
The probe has an audible buzzer signal and a visible light which are
activated when the instrument reaches water or product. The probe
features an auto shutoff circuit to prevent battery discharge. This auto
shutoff circuit allows the instrument 10 minutes of continuous operation
before the unit shuts off.
When the Keck Interface Probe is lowered down a well and contacts any
fluid with a specific gravity of .60 or greater, the float will activate the
visible and audible signals. If the fluid is non-conductive, the light will be
red and a solid tone will sound. If the fluid is water or other conductive
fluid, the conductive contacts will cause the light to turn green and the
tone will oscillate. If the probe is lowered through the water layer to a
non-conductive layer (such as DNAPL) the tone will change and the light
will return to red.
To avoid damage to tape and strain relief, do not over
tighten reel with probe in storage position.
Figure 1 – System Components
Probe
Assembly
Reel
Assembly
Frame
Assembly
Tape
Assembly
Tape Guard
Faceplate/
Electronics
Assembly
Hanger
Chapter 2: System Installation
The Keck Interface Probe Frame has a metal loop, the “hanger”, which
may be used to hang the probe assembly on the well casing. If the
Interface Probe frame will not hang on well head, use the tape guard to
prevent damage to the tape.
Figure 2 – Reel Frame
Chapter 3: System Operation
Step 1:
Test the battery. Press the "ON/TEST" button. If the
buzzer makes a loud signal and the light is visible, the battery is adequate
for normal operation. Pressing the "ON/TEST" button provides 10
minutes of operation.
Step 2:
Test the instrument operation. Remove the probe from
the retaining clip and raise the phenolic float. When the float switch
triggers, the visible signal will be red and the audible signal will emit a
steady tone. With the float still raised, moisten your fingers and touch the
float shaft and the conductive contact. If the battery is properly charged,
the visible signal will turn green and the tone will oscillate. Successful
completion of these tests will indicate that the unit is operating properly.
Measuring Fluid Levels
Step 1:
Lower the Interface Probe down the well until the signals
activate and then read the measurement at the top of the well casing or
other reference point. Record this figure as the first fluid level.
Step 2:
Continue to lower the probe until the signals change.
Record the measurement at the reference point as the second fluid level.
Step 3:
To determine the product layer thickness, subtract the
first reading from the second reading.
Measuring Water Levels in Product-Free Wells
If the probe immediately emits a green light, nonconductive fluids are
absent from the surface of the water.
To avoid damage to tape and strain relief, do not over
tighten reel with probe in storage position.
KECK KIR DNAPL CORRECTION FACTORS
2nd Reading
____’ + .08’
=
____’
(total depth)
MINUS
—
1st Reading
____’ - .08’
=
____’
(to product layer)
CORRECTED PRODUCT THICKNESS
=
____’
Example:
You find that a sinker (DNAPL) begins at 22.36 (1st
reading) and the total depth of your well is 23.61 (2nd reading). Use the
above formula to get your correct product thickness.
2nd Reading
23.61’ + .08’
=
23.69’
(total depth)
MINUS
—
1st Reading
22.36’ - .08’
=
22.28’
(to product layer)
CORRECTED PRODUCT THICKNESS
=
1.41’
Product thickness can be determined by adding .15’ to the
difference of your original measurements.
Chapter 4: System Maintenance
Battery Replacement
Replace the battery when the audible and visible signals become weak or
the unit does not operate.
Step 1:
Gently remove the battery tray.
Step 2:
Remove the old battery and replace it with a new one.
Cleaning
The probe can be cleaned with any detergent such as trisodium
phosphate (TSP), Alconox or Liquinox. If other detergents are used, take
care to select detergents that are compatible with Teflon®, polypropylene,
and stainless steel. The reel should not be submerged in any liquid, but
may be cleaned with a damp cloth.
If the float becomes covered with silt or mud, remove the float retaining
clip, slide the float off the shaft, and clean both the float and the float
shaft. Replace the float so the arrow is facing the top of the probe and
replace the retaining clip.
The conductive contact can be cleaned with detergent and a small brush.
Calibration
Be aware of the polarity (+, -) of the battery when placing
the new battery in the tray.
THE FLOAT SHAFT AT THE BOTTOM OF THE
PROBE IS NOT DESIGNED TO BE REMOVED.
Field calibration of the probe is not normally required.
Chapter 5: System Troubleshooting
Problem:
No signal (audible or visible) from battery test.
Solutions:
1) The battery is discharged.
•
Check or change battery.
2) The circuit is malfunctioning.
•
Consult the manufacturer
Problem:
No signal (audible or visible) from the probe.
Solutions:
1) The battery is discharged.
•
Check or change battery.
2) There is an open connection in the tape.
•
Replace tape and top cap. **
3) The float switch is malfunctioning.
•
Replace the probe. **
Problem:
No indication of product.
Solutions:
1) There is a short in the probe.
•
Replace the probe. **
2) Water is bridging at the bottom of the probe.
•
Clean the probe. Do not unscrew the float switch.
No indication of water.
1) The conductive contact is dirty.
•
Clean the contact.
2) There is an open connection in the tape.
•
Replace tape and top cap. **
3) The circuit is malfunctioning.
•
Consult the manufacturer.
Problem:
The signal (audible or visible) is intermittent.
1) There is an open connection in the tape.
•
Replace tape and top cap. **
2) There is a loose connection in the circuit or the probe.
•
Repair the connection. **
3) The float is damaged or missing.
•
Replace the float.
**For Technical Assistance, Call Geotech 1-800-275-5325
Chapter 6: System Specifications
Probe
Material:
Stainless Steel & Teflon®
Weight:
19.75 oz.
560 g
Diameter:
1.5 Inches
3.8 cm
Length:
8.9 Inches
22.6 cm
Conductivity Threshold
(detects water at) :
>6.7 µS
Minimum detectable hydrocarbon: .01 foot
Tape
Material:
Tefzel coated stainless steel
Length/Weight:
100 foot / 30 meters = 9 lbs
200 foot / 60 meters = 11 lbs
300 foot / 100 meters = 14 lbs
Accuracy:
100th of a foot/100’
Per Federal Specification
GGG-T-106E
Reel/Frame
Material:
Polypropylene & Aluminum
Size:
13” H x 11” W x 7” D
Unit
Battery:
9 volt alkaline
Self shut – off time:
10 minutes
Output tone:
500 Hz
Modulation (water detected):
3.5 Hz
Operating temperature range:
32 – 140 °F
0 – 60 °C
Storage temperature range:
-40 – 158 °F
-40 – 70 °C
Response time:
<10 milliseconds
Chapter 8: Replacement Parts List
Figure 5 – Frame, Reel, & Control Assembly Parts List
ITEM # PARTS DESCRIPTION
PARTS LIST
1
FRAME ASSEMBLY
52050027
2
REEL ASEMBLY, 100ft/30m
52050029
REEL ASEMBLY, 200ft/60m
52050030
REEL ASEMBLY, 300ft/100m
52050031
3
CONTROL ASSEMBLY
52050036
4
CLAMPING KNOB
12050524
5
RETAINING CLIP
12050001
6
BREAK SCREW
12050525
7
CLAMPING NUT 17500129
MANUAL
12050096
TAPE GUARD
12050060
CARRYING CASE (OPTIONAL)
12050059
For a period of one (1) year from date of first sale, product is warranted to be free
from defects in materials and workmanship. Geotech agrees to repair or replace,
at Geotech’s option, the portion proving defective, or at our option to refund the
purchase price thereof. Geotech will have no warranty obligation if the product is
subjected to abnormal operating conditions, accident, abuse, misuse,
unauthorized modification, alteration, repair, or replacement of wear parts. User
assumes all other risk, if any, including the risk of injury, loss, or damage, direct
or consequential, arising out of the use, misuse, or inability to use this product.
User agrees to use, maintain and install product in accordance with
recommendations and instructions. User is responsible for transportation
charges connected to the repair or replacement of product under this warranty.
Equipment Return Policy
A Return Material Authorization number (RMA #) is required prior to return of any
equipment to our facilities, please call 800 number for appropriate location. An
RMA # will be issued upon receipt of your request to return equipment, which
should include reasons for the return. Your return shipment to us must have this
RMA # clearly marked on the outside of the package. Proof of date of purchase
is required for processing of all warranty requests.
This policy applies to both equipment sales and repair orders.
FOR A RETURN MATERIAL AUTHORIZATION, PLEASE CALL OUR
SERVICE DEPARTMENT AT 1-800-833-7958 OR 1-800-275-5325.
Model Number: ________________
Serial Number: ________________
Date:
________________
Equipment Decontamination
Prior to return, all equipment must be thoroughly cleaned and
decontaminated. Please make note on RMA form, the use of equipment,
contaminants equipment was exposed to, and decontamination
solutions/methods used.
Geotech reserves the right to refuse any equipment not properly
decontaminated. Geotech may also choose to decontaminate equipment
for a fee, which will be applied to the repair order invoice.
Geotech Environmental Equipment, Inc
2650 East 40th Avenue Denver, Colorado 80205
(303) 320-4764 ● (800) 833-7958 ● FAX (303) 322-7242
email: [email protected] website: www.geotechenv.com
In Michigan Call
Geotech Environmental Equipment Inc.
1099 East Grand River Road, Suite 6, Williamston, Michigan 48895
(517) 655-5616 • (800) 275-5325 • FAX (517) 655-1157