Dickson HT300 Quick Start Manual
Read Online
Read the Dickson HT300 Quick Start Manual online using the built-in PDF viewer.
Text of the manual
Dickson warrants this line of instruments will be free from defects in material and
workmanship under normal use and service for a period of twelve months, after delivery.
This warranty does not cover routine calibration or battery replacement. For
Specifications and Technical Support, go to http://www.dicksondata.com
or contact us via telephone at 800.323.2448.
Factory Service and Returns
Contact Customer Service 630.543.3747 for a Return Authorization Number (RA) before
returning any instrument. Please have the model number, serial number and PO ready
before calling.
V4 06282016
DICKSON
DICKSON
DICKSON
DICKSON Bottle Logger
Bottle Logger
Bottle Logger
Bottle Logger HT300
HT300
HT300
HT300
Getting Started
Presets
•
Sample Rate = 1 Minute
•
Stop or Wrap When Full = Wrap
•
DicksonWareTM Unit of Measure = oF
Quick Start
Install DicksonWareTM
•
Version 17.23 or higher
Install Battery
•
Each end of the HT300 can be removed. The battery is located on the cap end
with the lanyard hole. Note the battery symbol etched on this end of the case.
•
Remove the cap carefully by twisting the cap counter clockwise.
•
Insert the battery positive side in first.
•
Replace the battery cap tightly.
•
It may take up to 5 seconds for the unit to power on.
•
Open DicksonWareTM software via the icon on your desktop.
NOTE: It is very important to attach the correct cap to the correct end of the
case. The cap end with the lanyard hole and spring must always go on the
battery end of the case.
Setup Logger
•
The USB port is located opposite end of the case. Note the USB symbol etched
on the USB access end of the case.
•
Remove the cap.
•
The USB port is just inside the logger opening.
•
Connect the USB cable to the logger and to an available working USB port on
your PC.
•
Your computer should detect the logger when it is powered on and connected to
the computer via the USB connection. Depending on your version of Windows,
the computer may go through an Add New Hardware routine. Click on the Setup
button. You will be prompted to select USB or Serial COM Port Connection.
Select USB then Click Continue. A Setup window will appear. All fields should be
automatically filled in, confirming that DicksonWareTM has recognized your
logger. Should all fields remain blank, refer to the troubleshooting section in
this manual
•
Once DicksonWareTM recognizes the logger, press the Clear button. This will
delete all data currently stored.
•
The logger is now ready to be placed in the desired logger location.
Software
(All of these features can be modified by click on the main Setup button.)
Setup (button)
Click this button first to establish communications between the logger and
DicksonWareTM software. You may be prompted to select the communication
method between USB or Serial COM port. You may save this setting so you will
not be prompted again. This setting is also changeable in
File/Preferences/Communications. A Setup window will appear with “all fields”
populated. This confirms that the software has recognized the logger. Should
the fields remain blank then communication has not been established, refer to
the Troubleshooting section of this manual
This tab provides you with the model and serial number of the logger, as well as
the option to set a custom “User ID” by clicking the active “Setup” button to the
right of the “User ID” field. This tab also includes the date the unit was
calibrated, calibration interval, and factory calibration date.
Samples (tab)
The majority of the setup process takes place in this section. Each field with an
active “Setup” button to the right, is a parameter that you can customize.
Sample Interval: Tells your logger how frequently to take and store readings.
This can be done in 1 or 10 second intervals. The dialog box for editing the
sample interval also indicates how long it will take for the logger memory to fill
up with new readings before it either stops saving new samples or begins to
wrap. “Sub ten second interval” should be enabled for desired sample intervals
under 10 seconds.
Stop or Wrap when Full: Determines what the logger should do when it is full.
The logger will either stop and discontinue logging, or continue logging by
wrapping the newest data over the oldest data.
Note: When changing logger settings (sample interval, stop/wrap, and start date
and time) the logger will automatically clear all stored data.
Channels (tab)
Click the Adjust button to the right of the temperature value for channel 1 to
change the name of the channel.
Alarms (tab)
Alarms are not active on this model.
Download (button)
From the main menu, click on the Download button to automatically extract all
logged data into a graph and table format.
Customized Graphed Data
DicksonWareTM calculates MIN, MAX and AVE of all data collected. Customize
data by eliminating unnecessary data points and customize MIN, MAX and AVE to
show only desired information.
Export Data
A snapshot of your graph or a real-time graph of points can be easily exported to
other programs such as Excel or PowerPoint.
Calibration
•
A Zero Adjust calibration can be performed on this logger. SW400
calibration software is required. Note: It is strongly recommended that a
higher accuracy NIST’d instrument should be used as the standard.
•
For more accurate calibration, return the instrument to Dickson for
calibration in our A2LA Certified lab. Contact Customer Service for a Return
Authorization Number before returning for calibration.
Need to Know
Logger Settings: When changing logger settings (sample interval, stop/wrap
and start date and time) the logger will automatically clear all stored data.
Fahrenheit/Celsius: The DicksonWareTM default is Fahrenheit. To change graph
view in DicksonWareTM to Celsius, go to File/Preferences to change unit of
measure.
IP Rating: The logger has an IP rating of 68 in up to 50 meters of water.
Downloading: Do not open the logger (remove caps) until the logger has
returned to ambient temperature.
Power:
Battery Type: Tadiran TLH-5955 high-energy lithium battery, Quantity
1, Dickson part R170
If the battery is low: Removing the battery will cause the logger to take
a dummy reading of -414F to mark the point where power was restored.
Replace the battery and clear the logger in DicksonWareTM before using.
Note: Power draw from the battery cannot be turned off. If
decommissioning a logger, remove the battery. To achieve longer
battery life during operation, use a less frequent sample rate and
disconnect the unit from the USB when not downloading.
Troubleshooting
Can the HT300 be used in an autoclave or incubator?
•
Yes it is perfect for both applications.
Why won’t the logger communicate?
•
Is the correct version of DicksonWareTM (v17.23 being used)? If not
contact customer service for the latest version.
•
Make sure the battery is good.
•
Confirm that USB is selected under File/Preferences/ Communications
•
Try another available USB port. If this is the first time attempting to
communicate on this PC, or if new software or programs have been
installed since the data logger was last used, try reinstalling the
software or installing the software on a different PC.
Why does my downloaded data show a -414 reading?
•
You will see a -414F reading if the battery was removed at any time.
This marks the point at which logging was resumed once the battery
was replaced.
Will I lose any saved logged data when replacing the battery?
•
No