Dewetron DEWE-3010-IPC Technical Reference Manual
Text of the manual
(217) 352-9330 | [email protected] | artisantg.com
-~ ARTISAN®
~I TECHNOLOGY GROUP
Your definitive source
for quality pre-owned
equipment.
Artisan Technology Group
Full-service, independent repair center
with experienced engineers and technicians on staff.
We buy your excess, underutilized, and idle equipment
along with credit for buybacks and trade-ins.
Custom engineering
so your equipment works exactly as you specify.
• Critical and expedited services
• Leasing / Rentals/ Demos
• In stock/ Ready-to-ship
• !TAR-certified secure asset solutions
Expert team I Trust guarantee I 100% satisfaction
All trademarks, brand names, and brands appearing herein are the property of their respective owners.
Find the Dewetron DEWE-3010 at our website: Click HERE
Technical Reference Manual
Contents
General Information, Safety Instructions
3
Notice
3
Safety Instructions
5
General System Information6
DEWE-3010 IPC
7
System specifications
7
Access to PCI Slots
8
Connectors
8
Instructions for battery exchange
1
2
System performance
1
3
Display / Graphic interface
1
4
Keyboard
1
4
Declaration of conformity
1
5
Printing version 1.3.6 / April 18, 2000
The information contained in this document is subject to change without notice.
DEWETRON elektronische Messgeraete Ges.m.b.H. (DEWETRON) shall not be liable for any
errors contained in this document. DEWETRON MAKES NO WARRANTIES OF ANY KIND WITH
REGARD TO THIS DOCUMENT, WHETHER EXPRESS OR IMPLIED. DEWETRON
SPECIFICALLY DISCLAIMS THE IMPLIED WARRANTIES OF MERCHANTABILITY AND
FITNESS FOR A PARTICULAR PURPOSE. DEWETRON shall not be liable for any direct,
indirect, special, incidental, or consequential damages, whether based on contract, tort, or any
other legal theory, in connection with the furnishing of this document or the use of the information
in this document.
Warranty Information:
A copy of the specific warranty terms applicable to your DEWETRON product and replacement
parts can be obtained from your local sales and service office.
Restricted Rights Legend:
Use austrian law for duplication or disclosure.
DEWETRON elektronische Messgeraete Ges.m.b.H.
Riesstrasse 146
A-8010 Graz / Austria
Printing History:
Version 1.3.6
Released April 18, 2000
Copyright © 1998 - 2000 DEWETRON elektronische Messgeraete Ges.m.b.H.
This document contains information which is protected by copyright. All rights are reserved.
Reproduction, adaptation, or translation without prior written permission is prohibited, except as
allowed under the copyright laws.
All trademarks and registered trademarks are acknowledged to be the property of their owners.
Printing version 1.3.6 / April 18, 2000
Page 3
Safety symbols in the manual:
Indicates hazardous voltages.
Calls attention to a procedure, practice, or condition that could cause bodily injury
or death.
Calls attention to a procedure, practice, or condition that could possibly cause
damage to equipment or permanent loss of data.
The following general safety precautions must be observed during all phases of operation,
service, and repair of this product. Failure to comply with these precautions or with
specific warnings elsewhere in this manual violates safety standards of design,
manufacture, and intended use of the product. DEWETRON Elektronische Messgeraete
Ges.m.b.H. assumes no liability for the customer's failure to comply with these
requirements.
Page 4
Safety instructions for all DEWETRON systems
• The DEWETRON data acquisition systems may only be installed by experts.
• Read your manual before operating the system.
• Observe local laws when using the instrument.
• Ground the equipment: For Safety Class 1 equipment (equipment having a protective earth
terminal), an non interruptible safety earth ground must be provided from the mains power source
to the product input wiring terminals or supplied power cable.
• DO NOT operate the product in an explosive atmosphere or in the presence of flammable gases
or fumes.
• DO NOT operate damaged equipment: Whenever it is possible that the safety protection features
built into this product have been impaired, either through physical damage, excessive moisture,
or any other reason, REMOVE POWER and do not use the product until safe operation can be
verified by service-trained personnel. If necessary, return the product to a DEWETRON sales and
service office for service and repair to ensure that safety features are maintained.
• Keep away from live circuits: Operating personnel must not remove equipment covers or shields.
Procedures involving the removal of covers or shields are for use by service-trained personnel
only. Under certain conditions, dangerous voltages may exist even with the equipment switched
off. To avoid dangerous electrical shock, DO NOT perform procedures involving cover or shield
removal unless you are qualified to do so.
• No modifications are allowed at the instrument. The fuse in the power module has to be replaced
by the same type. For continued protection against fire, replace the line fuse(s) only with fuse(s)
of the same voltage and current rating and type. DO NOT use repaired fuses or short-circuited
fuse holder labels and print on the power module may not be removed.
• DO NOT service or adjust alone. Do not attempt internal service or adjustment unless another
person, capable of rendering first aid and resuscitation, is present.
• DO NOT substitute parts or modify equipment: Because of the danger of introducing additional
hazards, do not install substitute parts or perform any unauthorized modification to the product.
Return the product to a DEWETRON sales and service office for service and repair to ensure
that safety features are maintained.
• Before opening the instrument (experts only) or exchanging the fuse in the power module
disconnect power!
• Don't touch internal wiring!
• Don't use higher supply voltage than specified!
• Use only original plugs and cables for harnessing.
• Install filler-panels in unused slots.
• The power-cable and -connector serve as Power-Breaker. The cable must not exceed 10 feet,
disconnect function must be possible without tools.
• Safety of the operator and the unit depend on following these rules.
Printing version 1.3.6 / April 18, 2000
Page 5
General System Information
DEWE-2010 and DEWE-3010 systems
• The system BIOS is protected by password. Any change in the BIOS may cause a system crash.
When the system is booting, do not press ESC-button on keyboard. This may clear the BIOS
settings and cause system faults.
• Any change in the file structure as deleting or adding files or directories might cause a system
crash.
• Before installing software updates contact DEWETRON or your local distributor. Use only
software packages which are released by DEWETRON. Further informations are also available
in the internet (http://www.dewetron.com).
• After power off the system wait at least 10 seconds before switching the system on again.
Otherwise the system may not boot correct. This prolongs also the life of all system components
Page 6
• Portable industrial PC
• 2 PCI slots full length and height
System specifications
Power supply:
■
AC power supply ACE-716A
(for details see 1.1 on next page)
❑ DC power supply NET-202 + Accu
(for details see 1.2 on next page)
Operating temperature:
Storage temperature:
Humidity (operating):
Vibration:
5 Hz - 200 Hz - 5 Hz:
5 Hz - 7 Hz:
7 Hz - 92 Hz:
92 Hz - 110 Hz:
110 Hz - 200 Hz:
Shock:
High intensity:
Crash safety:
Basic design:
0 °C to 50 °C (standard)
-20 °C to +60 °C
10 % to 80 %, non condensing
5 % to 95 %, rel. humidity
MIL-STD 810C 514.2 procedure X
12 min, 84 min per axis
1 in. displacement
2.5 g
0.0058 in. displacement
3.5 g
MIL-STD 810C
100 g 6 ms, operating, procedure IV
60 g 6 ms, operating, procedure III
30 g 11 ms, operating, procedure I
Physical characteristics:
Size (W x D x H): 1 4 . 6 x 10.2 x 6.7 in. (372 x 258 x 170 mm)
Weight:
t
y
p
.
12 lbs (6 kg), depending on configuration
Printing version 1.3.6 / April 18, 2000
Page 7
Access to PCI Slots
To open the DEWE-3010 IPC remove
the 8 big screws and the cover.
Connectors
4
5
Note: The location of the connectors might vary from system
to system and depends on system configuration
Connector overview:
1. Power supply connector
2. PS/2 mouse connector
3. PS/2 keyboard connector
4. Ethernet RJ45 connector
5. VGA connector
6. Ultra Wide SCSI controller interface
(option)
7. RS-232 connector (COM1)
8. RS-232 connector (COM2)
9. LPT Printer interface connector
9
Page 8
1. Power supply connector
1.1 AC power supply
Input:
Input range:
Input frequency:
Max. input current:
Output:
Output power:
Output voltages:
85 .. 135 VAC and 180 .. 265 VAC (auto selecting)
47 .. 440 Hz
5 A (115 VAC) or 3 A (230 VAC)
150 W
+5 V (max. 25 A)
+12 V (max. 10 A)
-5 V ( m a x . 1 A)
-12 V (max. 1 A)
The maximum load is 150 W, otherwise power supply
can be damaged.
The system fuse is located between connector and power switch. If the system doesn't work
after switching on, disconnect the power cord and check the fuse.
1.2 DC power supply
Input:
DC and battery input:
Max. input current:
Input power:
Input voltage priority:
Battery type:
Output:
Output power:
Output voltages:
Power LED:
dim green:
bright green:
orange:
red:
10 to 32 VDC
3.5 A (230 VAC), startup peak up to 25 A
160 W
1. DC voltage
2. Battery
NiCd, 24 VDC, 600 mAh
160 W
+5 V (max. 20 A)
+12 V (max. 5 A)
-5 V ( m a x . 0.3 A)
-12 V (max. 0.4 A)
The maximum load is 160 W, otherwise the power supply
can be damaged.
DC connected to system, system powered off
DC connected to system, system powered on
system powered on, working via accu
starts 30 sec. before accu is getting low together with beep signal
Printing version 1.3.6 / April 18, 2000
Page 9
1. P o w e r supply connector (cont.)
AC power supply connector
AC connector
Poweron/off
Fuse ( 2 5 0 V / 1AT
110 V / 3 15 A T)
DC power supply connector
1: +Vcc
2: GND
3: Not connected
2. P S / 2 mouse connector
The mouse / trackball connector is used to connect the trackball embedded in the keyboard or
an external PS/2 mouse. The connector meets standard PS/2 pin assignment.
3. P S / 2 keyboard connector
The keyboard connector is used to connect PS/2 keyboard to DEWE-3010 IPC system. The
connector meets standard PS/2 pin assignment.
4. Ethernet connector
The DEWE-3010 system supports 10baseT Ethernet with standard RJ45 connector.
5. V G A connector
The VGA connector offers the possibility to connect an external CRT to the system.
15 pin mini SUB-D connector
5 4 3 2 1
15 14 13 12 11
Schematic
Pin assignment
1: Red video
2: Green video / Sync on green
3: Blue video
4: -
5: -
6: Red video ground
7: Green video ground
8: Blue video ground
9:
10: Ground
11: Ground
12: Data line
13: H -Sync / HV-Sync
14: V -Sync
15: Clock line
Page 10
6. Ultra wide SCSI controller interface (option)
The Ultra-Wide-SCSI interface connector is located on the right side of the DEWE-3010 IPC.
External SCSI devices, i.e. CD-Writer, can be connected with appropriate cables.
68 pin HD-SUB-D connector
7. R S -232 interface connector (COM1)
8. R S -232 interface connector (COM2)
The two RS-232 interface connectors (male) are located on the left and right side of the
DEWE-3010 IPC. They are configured as standard RS-232 interface COM 1 and COM 2 and
can be used for mouse or other peripheral units.
9 pin SUB-D
connector (male)
1 2 3 4 5
6 7 8 9
Schematic
Pin assignment
1: DCD (Data Carrier Detector)
2: RD (Received Data)
3: TD (Transmitted Data)
4: DTR (Data Terminal Ready)
5: GND (Ground)
6: DSR (Data Set Ready)
7: RTS (Request To Send)
8: CTS (Clear To Send)
9: RI ( R i n g Indicator)
9. LPT printer interface connector
The printer interface connector (female) is located on the left side of the DEWE-3010 IPC It is
configured as standard LPT interface.
25 pin SUB-D connector (female)
13 12 11 10 9 8
7 6 5 4 3 2 1
25 24 23 22 21 20 19 18 17 16 15 14
Schematic
Pin assignment
Printing version 1.3.6 / April 18, 2000
Page 11
Instructions for battery exchange
The following part is only valid for systems with DC power supply
Battery exchange has to be done by qualified persons only!
Power down system and disconnect from power supply. Remove four screws
from bottom of the DEWE-3010 system to get access to the battery. Exchange
battery and close battery compartment again with the four screws.
Do not touch any internal wiring expect of the battery connection cable!
Remove this 4 screws for battery access
Page 12
Pentium III 1000 MHz
❑ Intel Celeron 1100 MHz
System Memory:
Memory type:
■
168 pin DIMM (PC133)
❑ 168 pin DIMM (PC100)
■ 1
O 2
O 3
O 4
Number of modules:
Size per module:
U 64 MB
❑ 128 MB
■ 256 MB
❑ 512 MB
Harddisk:
Interface:
E
-
I
D
E
Capacity:
❑
10 GB
■ 20 GB
❑ 60 GB
Floppy Disk Drive:
■
3.5" / 1.44 MB
CD-ROM:
2
4
x
Network connection:
❑
10BaseT Ethernet (RJ45)
■ 10/100BaseT Ethernet (RJ45)
Printing version 1.3.6 / April 18, 2000
Page 13
Display / Graphic interface
Display:
Size:
1
2
.
1
"
Resolution:
8
0
0
x 600 pixel
Touchscreen:
0
yes
II no
Graphic interface:
Bus system:
P
C
I
Video memory:
2
MB
CRT simultaneous: y e s
Keyboard
Keyboard:
8
3
/ 84 keys
Interface:
P
S
/
2
Trackball Interface: P S / 2
l • •
, . . . 0 2 4 4 t e L a l j , 1 1 1 P
.
_J
J
_ I _ I _ I
_J
J
J
_ J
Page 14
Declaration of conformity
Following products are in conformity with EN50081 (Generic emission standard for light and
heavy industry) and EN50082 (Generic emission standard for light and heavy industry):
DEWE-2000 / DEWE-2010
DEWE-3000 / DEWE-3010
DEWE-BOOK
DEWE-RACK
DEWE-DAQ modules
Note: enhanced EMC specifications are achieved when using DC power supply input.
Manufacturer:
DEWETRON Elektronische Messgeraete Ges.m.b.H.
Parkring 4
A-8074 Graz-Grambach / AUSTRIA
Tel.: ++43-316-3070-0
Fax: ++43-316-3070-90
e-mail: [email protected]
http://www.dewetron.com
Graz, Feb. 5th, 1998
l
n
g
.
Herbert Wierihigg
Managing directtor
Printing version 1.3.0 / October 29, 1999
Page 15
Artisan Technology Group is an independent supplier of quality pre-owned equipment
Gold-standard solutions
Extend the life of your critical industrial,
commercial, and military systems with our
superior service and support.
We buy equipment
Planning to upgrade your current
equipment? Have surplus equipment taking
up shelf space? We'll give it a new home.
Learn more!
Visit us at artisantg.com for more info
on price quotes, drivers, technical
specifications, manuals, and documentation.
Artisan Scientific Corporation dba Artisan Technology Group is not an affiliate, representative, or authorized distributor for any manufacturer listed herein.
We're here to make your life easier. How can we help you today?
(217) 352-9330 I [email protected] I artisantg.com