Atmos Cam 21 Operating Instructions Manual
Read Online
Read the Atmos Cam 21 Operating Instructions Manual online using the built-in PDF viewer.
Text of the manual
2018-11 Index: 07
Operating Instructions
English
MedizinTechnik GmbH & Co. KG
Ludwig-Kegel-Straße 16
79853 Lenzkirch
Germany
Phone: + 49 7653 689-0
Fax:
+ 49 7653 689-190
+ 49 7653 689-493 (Service Centre)
E-mail: [email protected]
Internet: www.atmosmed.de
1.0
Introduction .........................................................3
1.1
Notes on operating instructions ............................3
1.2
Intended use .........................................................4
1.3
Function ................................................................4
1.4
Explanation of symbols .........................................5
2.0
For your safety ....................................................6
3.0
Setting up and starting up .................................7
3.1
Construction ..........................................................7
3.2
Starting up ATMOS® Cam 21 / 31 .........................9
3.3
Digital monitor outlets ...........................................9
3.4
Starting up ATMOS® Strobo 21 LED ...................11
4.0
Cleaning and care .............................................13
4.1
General information on
cleaning and disinfection.....................................13
4.2
Recommended instrument disinfectants .............14
4.3
Recommended surface disinfectants ..................14
5.0
Maintenance and Service .................................15
6.0
Options, accessories and consumables ........16
7.0
Technical data ...................................................18
8.0
Checking / Disposal ..........................................19
8.1
Checking .............................................................19
8.2
Disposal ..............................................................19
Content
1.1 Notes on Operating Instructions
These operating instructions contain important notes on how to operate the ATMOS® Endoscopy and
Stroboscopy System, correctly and effectively. Their reading helps to avoid risks, and also to reduce
repair costs and down-times. This increases, amongst other things, the reliability and service-life of the
device.
These operating instructions serve not only for new operating personnel to be instructed in its use, but
also for use as a reference manual. Reproduction, even partial, is only permitted with written permission
from ATMOS.
These operating instructions must always be kept available near the device.
Care and period tests in conjunction with professional execution provide for operational safety and
readiness for use of your ATMOS® Endoscopy and Stroboscopy System and are therefore a must
besides regular cleaning.
Repair work and period tests may be carried out only by expert personnel authorised by ATMOS. By
applying only original spare parts you will have the guarantee that operational safety, readiness for work
and the value of your ATMOS® Endoscopy and Stroboscopy System will be preserved.
• The ATMOS® Endoscopy and Stroboscopy System bears CE marking CE according to the EC
Directive of the council for medical products 93/42/EEC and meets the basic requirements of
Appendix I of the directive.
• The product ATMOS® Endoscopy and Stroboscopy System complies with all applicable requirements
of the Directive 2011/65/EC restricting the use of certain hazardous substances in electrical and
electronic equipment (“RoHS”).
• The declaration of conformity and our general standard terms and conditions can be obtained on our
website at www.atmosmed.com.
• The quality management system applied at ATMOS has been certified according to international
standards EN ISO 13485.
• Prior to start-up please peruse chapter 2.0 „For your safety“, in order to be prepared for any possible
dangerous situations.
These operating instructions are valid for the following
devices:
Endoscopy Trolley
REF 950.0223.0
Stroboscopy Trolley
REF 950.0226.0
in connection with the options:
• ATMOS® Cam 21 / 31
• ATMOS® Strobo 21 LED
Packing / Storage
The ATMOS® Endoscopy and Stroboscopy System
is delivered in cardboard boxes. Please check the
consignment on any damage immediately on receipt. Please
contact your dealer in case of any complaints or prior to
returning the goods.
Please observe the following limit values during storage of
the ATMOS® Endoscopy and Stroboscopy System:
Ambient temperature:
5 °C to + 50 °C
Air humidity:
5 % to 85 %
Please observe the separate operating instructions for ATMOS® Cam 21 / 31 and ATMOS® Strobo 21 LED.
1.0 Introduction
1.2 Intended use
1.3 Scope of supply
Standard scope of supply:
Endoscopy Trolley
950.0223.0
Stroboscopy Trolley
950.0226.0
Further options, accessories and consumables can be
added individually (see chapter 6.0).
1.0 Introduction
Main function:
• Storage room and power supply system for diagnostic
devices
• Instrument storage
Application:
For application in the surrounding of the patient, for storage
in the surrounding of the patient and for storage of medical
products and therapeutic appliances.
Application restrictions: No
Sterility: unsterile
Active medical product
Please observe the separate operating instructions for ATMOS® Cam 21 / 31 and ATMOS® Strobo 21 LED.
Explanation of symbols
1.0 Introduction
!
■
●
→
~
Icons
Foot switch inlet
Interlink / Data
Signal input and output
Off (feed-in, power
connection)
On (feed-in, power
connection)
Alternating current
Ground wire connection
Control panel buttons of ATMOS® Cam 31
Endoscope
Usage:
Hysteroscopy
Typ. ø:
2.7 mm
Typ. angels:
0°, 30°, 45°
Xenon light source
Typical:
180 resp. 300 W
Colour
temperature:
5600...5800 K
Halogen light source
Typical:
150 resp. 250 W
Colour
temperature:
3200 K
LED light source
Typical:
2 W (e.g. ATMOS® LED
light source
→ see accessories chapter 6.0
Extra-light source
Stroboscopy for connection of an ATMOS®
Strobo 21 LED
Endoscope
Usage:
Hysteroscopy
Typ. ø:
4 mm
Typ. angels:
0°, 30°, 45°
Flexible Endoscope
Usage:
Nose, throat, larynx
examinations
Microscope
Usage:
Examination of ears
Warning, special diligent notice
Application part type BF
Fuse according to IEC 417/5016,
DIN 30600/0186
Potential equalisation
Short cuts / symbols
Please press where
dot indicates
Subnumeration
Numeration
General information
Follow the arrows
Activate the optional foot
switch
Replace
Check
Please read,
important information
Move, plug... in this
direction
Engage, check
correct fit
Turn, shift ... in this
direction
Y/C / FBAS
Signal output
Audio
IEEE 1394
DV port
Only valid for ATMOS® Cam 31 DV
Freeze (Storage)
click
• The ATMOS® Endoscopy and Stroboscopy System is
produced according to IEC 60601/EN 60601 and listed in
the following classes:
-
VDE Class of protection I
-
Class I Medical Devices Act (German law for medical
products)
• The device may only be connected to a properly installed
safety socket.
• The ATMOS® Endoscopy and Stroboscopy System is
designed in line with approved quality standards and
safety-related regulations. It should only be used in
accordance with regulations and under consideration of
the information stated in the operating instructions.
• Please do not perform any modifications of the ATMOS®
Endoscopy and Stroboscopy System without approval
by the manufacturer. Please observe general rules for
accident avoidance as well as for environment protection.
In order to guarantee correct function, please only use
original spare parts ordered from the manufacturer.
• The ATMOS® Endoscopy and Stroboscopy System may
only be used by skilled staff who have been authorised
by ATMOS and trained in its operation (IEC 60601-1 / EN
60601-1).
• The mains voltage indicated on the type plate must
correspond to the values of the supply network.
• Make sure prior to every application of the equipment that
it is technically safe and in proper condition. Damaged
cables must be replaced immediately!
• Correct configuration in assembly of country-specific
connections:
-
green/yellow: protective conductor (PE)
-
blue: neutral conductor (N)
-
black or brown: phase (L)
• Please observe that the mobile parts could cause injury
and contusion!
• The ambient conditions specified in the "Technical data"
(chapter 7.0) must be strictly observed!
• Switch off all function modules when you finish your daily
business.
• The ATMOS® Endoscopy and Stroboscopy System may
be operated only in rooms used for medical purposes, but
not in areas subject to explosion hazards.
• The ATMOS® Endoscopy and Stroboscopy System meets
the immunity to interference requirements of IEC 60601-
1-2 / EN 60601-1-2 „Electromagnetic Compatibility –
Medical Electrical Devices“.
• Please note that the insulation transformer has to be
installed in the base of the trolley if more than one
camera respectively stroboscopy module as well as
further consumer loads are connected.
• The ATMOS® Endoscopy and Stroboscopy System may
not be operated with devices not complying with the
requirements of standard EN 60601-1 „Medical Electrical
Equipment“ and EN 60601-1-2 „Electromagnetic
Compatibility“ (Medical Electrical Equipment).
• There are no warranty claims whatsoever on defects
which arise from the use of third party accessories or
consumables.
• ATMOS is not liable for personal injury and damage to
property if
-
no original ATMOS parts are being used,
-
the advice for use in these operating instructions is
not being observed,
-
assembly, new settings, alterations, extensions
and repairs have been carried out by personnel not
authorised by ATMOS.
• Also pay attention to the safety information in following
chapters.
2.0 For your safety
!
For your safety
Please feed the Velcro fastener through the slots in the
metal sheet.
Tighten the ATMOS® Strobo 21 LED from below on the left
side. The ATMOS® Cam 21 / 31 you tighten on the right side.
Screw the support for the LED light handle to the metal
sheet on the left and the support for the camera head to the
right.
Afterwards please tighten the monitor to the monitor
support.
3.0 Setting up and starting up
3.0 Setting up and starting up
The power supply at the monitor is plugged in and fixed
with the Velcro fastener below the metal sheet where
camera and stroboscope are positioned.
Connecting the SVHS cable
Connecting the DVI cable
3.2 Starting up ATMOS® Cam 21 / 31
3.0 Setting up and starting up
G
E
F
D
C
I
H
Backside
Nut for mains supply
Audio input
Data gateway
Interlink interface for ATMOS® Stroboscope
Connection for video output (Y/C)
Connection for video output (FBAS)
Connection for potential equalization
J
Connection for foot switch
A
B
M
E
D
C
H
I
J
G
K
Backside
Nut for mains
supply
Data gateway
Interlink interface for ATMOS® Strobo 21 LED
Connection for video output (Y/C)
Connection for video output (FBAS)
Forefront
Connection for camera
head
Connection for LED
light source
Control panel
Buttons for selecting
application part and
light source
Mains on/off switch
IEEE 1394 connection
Audio input
Connection for foot switch
Connection for potential equalization
Only valid for ATMOS® Cam 21 DV / 31 DV
L
DVI-I-monitor
output
SXGA monitor
output
F
Warning notice / safety note
The cameras ATMOS® Cam 31 DV and ATMOS® Cam 21 DV have a digital monitor outlet with SXGA signal. The
resolution is 1280 x 1024 pixels.
Please make sure that only monitors are connected which support the stated resolution.
Otherwise the camera and / or monitor could get damaged.
!
3.3 Digital monitor outlets
Please pay attention, just to switch on the devices after they have been completely wired.
If required, analogue peripheral video equipment (monitor, video printer, etc) must be adjusted to NTSC video norm. For further
details please look at the corresponding operating instructions.
i
Optional:
Cable
IEEE 1394
K
4
Only valid for ATMOS®
Cam 21/31 DV
Only valid for ATMOS® Cam
31
Y / C
3
...or
Connection for
video output (Y/C)
(S-VHS)
Connection for
video output
(FBAS) (VHS)
DVI-I-monitor
output
SXGA monitor
output (VGA)
M
D
C
L
...or
3.0 Setting up and starting up
3.4 Starting up ATMOS® Strobo 21 LED
3.0 Setting up and starting up
+
.....
3
B
1
2
5
4
K
A
C
F
Please pay attention, just to switch on the devices after they have been completely wired.
When applying the ATMOS® Cam 21 / 31 for video
stroboscopy please pay attention to connecting
the dongle connector (Ref 507.4781.0) with the
interlink connection of ATMOS® Cam 21 / 31
before startup of the devices!
ATMOS® Cam
21 / 31
ATMOS® Strobo
21 LED
ATMOS® Cam 31
DV Data
or
or
A
B
Forefront
Control panel
On / Off switch
Settings
Control knob
Connection for microphone
Connection for LED light
handle
Indication for permanent light
Indication for freeze image
Indication for slow motion
Display
K
F
Plug in the 5pin data cable at the "data" jack, with it camera
and stroboscope housing are connected with each other.
The 6pin interlink cable you plug in at the "interlink" jack.
6pin interlink
5pin data
Connect the audio cable to the audio channel of the
stroboscope with the black connector. Please do not
connect the red cable. Plug in the other end of the audio
cable at the audio channel of the camera housing. With it
both housings are directly connected to each other with 3
cables.
"Audio out"
ATMOS® Cam
"Audio in"
ATMOS® Strobo 21 LED
ATMOS® Strobo 21 LED
ATMOS® Cam 21/31
3.0 Setting up and starting up
4.1 General information on cleaning and
disinfection
• Switch all devices off prior to cleaning and disinfection!
• The described actions relating to cleaning and
disinfection resp. sterilisation do not substitute the
relevant instructions which must be adhered to prior to
operation!
• For disinfection, you may use all surface and instrument
disinfectants listed on page 14.
• Always observe the concentration specifications and
instructions by the respective manufacturer!
• Do not use
-
Disinfectants which contain organic or inorganic acids
or bases as they could cause corrosion damage.
-
Disinfectants containing chloramides, phenol
derivatives or anionic tensides, as these may cause
stress cracks in the material used for the housing of
the unit.
4.1.1 Cleaning the unit surface
• The surfaces of the ATMOS® Endoscopy and
Stroboscopy System are resistant against all surface
disinfectants listed in on page 14.
• Wipe the unit surface with a cloth moistened with a
cleaning or disinfecting solution.
• You may also use disinfectant sprays or disinfectant
tissues for cleaning and disinfection.
ª Please note that the alcohol contained in these
agents could corrode or cloud the protective covers if
employed on a long term basis.
• Clean the ATMOS® Endoscopy and Stroboscopy System
with a soft cloth. Alkalescent multi-purpose cleaner, e.g.
soap sud can also be used.
• Do not use any abrasives or solvents as these might
damage the surface of the appliance.
• For disinfection please use usual aldehyde surface
disinfectants in a watery solution. The surfaces of the
ATMOS® Endoscopy and Stroboscopy System are
resistant against all disinfectants listed in chapter
„Recommended surface disinfectants“.
• Please do not use any disinfectants or cleaning agents
which contain chlorine. By the use of disinfectants
containing alcohol ignitable gas mixtures might occur.
Disinfection sprays may cause corrosion at mechanical
parts.
4.0 Cleaning and care
4.2 Recommended instrument disinfectants
Disinfectant
Ingredients
(in 100 g)
Manufacturer
Gigasept FF
(Application concentrate)
succindialdehyde
11.0 g
Schülke & Mayr,
Norderstedt
dimethoxytetrahydrofurane
3.0 g
corrosion protection components
non-ionic surfactants and perfumes
Sekusept PLUS
(Application concentrate)
glucoprotamin
25.0 g
Henkel, Düsseldorf
non-ionic surfactants
Solvents, complexing agents
4.3 Recommended surface disinfectants
Disinfectant
Ingredients
(in 100 g)
Manufacturer
Incidin Plus
(Application concentrate)
glucoprotamin
26.0 g
Henkel, Düsseldorf
non-ionic surfactants
Solvents, complexing agents
Dismozon pur
(Granulate)
End of product 12/2014
Magnesium monoperoxyphthalate
Bode Chemie, Ham-
burg
Hexahydrate
80 g
Dismozon plus
(Granulate)
Magnesium monoperoxyphthalate
Bode Chemie, Ham-
burg
Hexahydrate
95.8 g
Green & Clean SK
(Application concentrate)
Alkyl dimethyl benzyl ammonium chloride
< 1 g
Metasys, Rum (Aus-
tria)
Dialkyl-dimethyl-ammonium chloride
When using disinfectants containing aldehyde and amine at the same object colour changes may occur.
4.0 Cleaning and care
Maintenance, repairs and period tests may only be
carried out by persons who have the appropriate technical
knowledge and are familiar with the product. To carry out
these measures the person must have the necessary test
devices and original spare parts.
ATMOS recommends: Work should be carried out by an
authorized ATMOS service partner. This ensures that repairs
and testing are carried out professionally, original spare
parts are used and warranty claims remain unaffected.
Period tests:
Observe the specifications for the individual components.
Exchanging castors:
When exchanging a rear castor please remove the cover
first. Then loosen the screws with an Allen key SW8. The
front castors can be unscrewed with a jaw spanner SW13.
Mounting the replacement castors is performed in reverse
order.
Exchanging drawers:
Drawers can only be exchanged as complete components.
Please observe the provided mounting instructions.
Exchanging shelves:
Loosen the screws with an Allen key SW4 and remove the
shelf from the column. Mounting is performed in reverse
order.
Exchanging the handle:
It is also possible to unscrew the handle with an Allen key
SW4 and remove it from the column. Mounting is performed
in reverse order.
5.0 Maintenance and Service
5.1 Sending in the device
• Remove and properly dispose of consumables.
• Clean and disinfect the product and accessories
according to the operating instructions.
• Place used accessories with the product.
• Fill in the form QD 434 „Delivery complaint / return
shipment“ and the respective decontamination
certificate.
This form is enclosed with each delivery and can be found
at www.atmosmed.com.
• The device must be well padded and packed in suitable
packaging.
• Place the form QD 434 „Delivery complaint / return
shipment“ and the respective decontamination
certificate in an envelope.
• Affix the envelope to the outside of the package.
• Send the product to ATMOS or to your dealer.
6.0 Options, accessories and consumables
Insulation transformer 100 VA with insulation monitor
950.0225.0
Metal quiver for storage of rigid scopes
508.0775.0
Quiver for storage of flexible scopes, max. 500 mm length
508.0790.0
Plastic quiver for used rigid scopes
443.0017.0
Quiver for used flexible scopes, max. 500 mm length
508.0795.0
Monitor support for endoscopy / stroboscopy
950.0224.0
Medical 17" TFT monitor
534.3010.0
ATMOS® Panel PC
507.3121.0
ATMOS® Panel-PC Touch
507.3122.0
ATMOS® Capture Suite lite license dongle
700.0045.0
ATMOS® Capture Suite license dongle
700.0046.0
ATMOS® Capture Suite HD license dongle
700.0047.0
ATMOS® Cam 21
507.4100.0
ATMOS® Cam 21 DV
507.4120.0
ATMOS® Cam 21 DV Data
507.5100.0
ATMOS® Cam 31 with image memory
507.5110.0
ATMOS® Cam 31 DV with image memory
507.5130.0
ATMOS® Cam 31 Data with image memory
507.5120.0
ATMOS® Cam 31 DV Data with image memory
507.5140.0
ATMOS® Cam 41 HD
507.5500.0
ATMOS® LS 21 LED
507.4600.0
ATMOS® LS 21 LED, warm white
507.4602.0
ATMOS® LS 31 LED
507.4800.0
ATMOS® Strobo 21 LED
507.4700.0
Trigger cable (Y-cable), LS 31 LED - Strobo 21 LED to 2016-09
507.4837.0
Trigger cable LS 31 LED - Strobo 21 LED
507.4838.0
ATMOS® Strobo 21 DATA cable USB
507.4784.0
SD Grabber
534.1204.0
HD Grabber
534.1205.0
There will be no warranty claims on defects, missing functions resp. poor image quality. If third party accessories (e.g. monitors),
which differ from the above mentioned articles are used.
6.0 Options, accessories and consumables
Optical instruments
0°
30°
45°
70°
90°
Laryngoscopes
Laryngoscope 70°, Ø 10 mm, working length 176 mm, autoclavable,
without light guide
950.0209.0
Laryngoscope 90°, Ø 10 mm, working length 174 mm,
autoclavable, without light guide
950.0210.0
Laryngoscope 70°, Ø 8 mm, working length 166 mm,
autoclavable, without light guide
950.0246.0
Tele-magnifying laryngoscope 70°, Ø 10 mm, working length 147.5 mm,
can be immersed, without light guide
950.0211.0
Tele-magnifying laryngoscope 90°, Ø 10 mm, working length 145 mm,
can be immersed, without light guide
950.0212.0
Ear endoscopes
Wide-angle optic, 0°, working length: 50 mm, Ø 4 mm, autoclavable
950.0213.0
Wide-angle optic, 30°, working length: 50 mm, Ø 4 mm, autoclavable
950.0214.0
Wide-angle optic, 0° working length: 34 mm, Ø 2,7 mm, can be immersed
incl. adapter for ear-speculum
950.0215.0
Nose / Pharynx endoscopes Ø 4 mm
Wide-angle optic, 0°, working length: 180 mm, Ø 4 mm, autoclavable
950.0216.0
Wide-angle optic, 30°, working length: 180 mm, Ø 4 mm, autoclavable
950.0217.0
Wide-angle optic, 45°, working length: 180 mm, Ø 4 mm, autoclavable
950.0218.0
Wide-angle optic, 70°, working length: 180 mm, Ø 4 mm, autoclavable
950.0219.0
Nose/Pharynx endoscopes wide angle Ø 2.7 mm
Wide-angle optic, 0°, working length: 110 mm, Ø 2.7 mm, autoclavable
950.0220.0
Wide-angle optic, 30°, working length: 110 mm, Ø 2.7 mm, autoclavable
950.0221.0
Flexible nasopharyngoscope
High resolution naso-pharyngoscope Ø 3,8 mm, working length 300 mm,
0°, angle of field of view 80°, depth of focus: 5 mm – infinite, angle: 125° /
125°
Scope of delivery includes aluminium transport case and leak tester
950.0222.0
Flexible video-nasopharyngoscope
ATMOS® Scope
950.0300.0
ATMOS® Scope Basic
950.0330.0
7.0 Technical data
Isolation transformer
Performance
600 VA
Protection class (EN 60601-1)
I
Operating time
Continuous operation
Endoscopy Trolley / Stroboscopy Trolley
Dimensions HxWxD
112.6 x 63.5 x 59.5 cm
Weight
19 kg
Ambient conditions for transport/
storage
• Temperature
-20 °C to +70 °C
• Humidity without condensation
5 to 95 %
• Air pressure
700 - 1060 hPa
Ambient conditions for operation
• Temperature
+15 °C to +40 °C
• Humidity without condensation
5 to 95 %
• Air pressure
700 - 1060 hPa
Classification according to Annex
IX, EC Directive 93/42/EEC
Class 1
CE marking
CE
GMDN code
11585
UMDNS code
11-585
ID No. (REF)
950.0223.0 Endoscopy Trolley
950.0226.0 Stroboscopy Trolley
Please observe the separate operating instructions for ATMOS® Cam 21 / 31 and ATMOS® Strobo 21 LED.
• The ATMOS® Endoscopy and Stroboscopy System does not contain any hazardous materials.
• The housing is recyclable.
• Device and accessories must be decontaminated prior to disposal.
• Pay attention to a careful separation of the different materials.
• Please observe national disposal regulations (e.g. waste incineration).
Disposal within the EC
The device described above is a high-quality medical product with a long service life. After its life cycle it must be disposed
of professionally. According to the EC directives (WEEE and RoHS) the device may not be disposed of in domestic waste.
Please observe existing national laws and rules for disposal of old devices in the respective country.
Disposal within the Federal Republic of Germany
In order to guarantee a proper disposal of your old device, please either pass on your old device to your specialised dealer
or send it to ATMOS MedizinTechnik for disposal.
Before disposal respectively before transport all parts, which came into contact with the patient must be
thoroughly cleaned, disinfected. The device surface must be disinfected.
Please observe the relevant instructions for disposal of modular devices.
8.1 Checking ATMOS® devices
Observe the specifications for the individual components.
Regular, thoroughly cleaning and disinfection of the devices and the application parts respectively the operation in line with the
operating instructions are assumed.
8.0 Checking / Disposal
ATMOS MedizinTechnik GmbH & Co. KG
Ludwig-Kegel-Straße 16
79853 Lenzkirch / Germany
Phone: +49 7653 689-0
[email protected]
www.atmosmed.com