Argus Copper Box v4 Manual
Text of the manual
ARGUS Copper Box
© by intec Gesellschaft für Informationstechnik mbH
D-58507 Lüdenscheid, Germany, 2020
All rights, including translation rights, reserved. No part of this work may be reproduced,
duplicated or disseminated in any form (print, photocopy, microfilm or any other method)
without written consent.
All rights are reserved. No one is permitted to reproduce or duplicate, in any form, the
whole or part of this document without intec´s permission.
ARGUS Copper Box
3
1
Introduction .........................................................................................4
2
Safety information ..............................................................................7
3
General technical data .....................................................................13
4
Configuring accesses ......................................................................17
5
Using the ARGUS Copper Box v4 ...................................................18
5.1
Selecting and activating the Copper Box .......................................18
5.2
Hotkeys / settings .............................................................................20
5.3
Changing socket selection ..............................................................22
5.4
Alarm tones .......................................................................................23
6
Auto test ............................................................................................25
7
Voltage measurement (U= and U~) .................................................35
8
Capacitive symmetry measurement (CSym) ..................................42
9
Capacitance measurement (C) ........................................................46
9.1
Calculating line length .....................................................................49
10
Isolation resistance measurement (Iso.) ........................................51
11
Resistance symmetry (RSym) .........................................................55
12
Loop resistance measurement (R) ..................................................62
12.1
Calculating line length .....................................................................65
13
DC current measurement (I=) ..........................................................66
14
Longitudinal conversion loss (LCL) ...............................................68
15
Near-end crosstalk (NEXT) ..............................................................71
15.1
Starting NEXT measurement ...........................................................73
16
Signature detection/termination detection .....................................75
16.1
Possible results ................................................................................77
17
Fast cable check ...............................................................................79
17.1
Possible results ................................................................................80
18
Remote kit control (Rem.) ................................................................81
19
Appendix ...........................................................................................84
A)
Abbreviations .....................................................................................84
B)
Software licenses ...............................................................................87
C)
Index ..................................................................................................88
ARGUS Copper Box
1 Introduction
ARGUS Copper Box v4
With the ARGUS Copper Box, intec offers an expansion for the xDSL multitesters
ARGUS 15x and ARGUS 16x. This USB box enables you to detect dangerous voltages
and currents early on and reliably assess the physical quality of the line – particularly when
DSL synchronisation is not possible or the data rate is low due to asymmetry, interference
sources or other mechanical line problems.
The box is connected to the ARGUS tester
simply via the USB host interface. You can
select the box and perform all
measurements easily and rapidly via the
ARGUS graphical user interface.
The ARGUS Copper Box is equipped with
four standard banana sockets, which are
colour-matched to the respective cables.
These sockets are designed for a four
millimeter (4 mm) thick, contact-protected
banana cable.
ARGUS Copper Box
5
Further features of the ARGUS Copper Box v4:
-
Multiple, preconfigured measurement profiles can be executed as an auto test.
-
Voltage measurement enables measurement of the supply voltage (e.g. SHDSL, ISDN-
BRI U, POTS), as well as external voltages caused e.g. by contact between two wires or
isolation faults.
-
With the aid of capacitive symmetry measurement, it is possible to detect asymmetry
in the subscriber line that can result in signal distortions or transmission errors.
-
Capacitance measurement detects interruptions as well as typical input capacitances
of connected devices and enables estimation of line lengths.
-
Isolation resistance measurement reveals damage to cable isolation, moisture
penetration or oxidised contact points.
-
With the aid of resistance symmetry measurement, it is possible to detect irregula-
rities in the subscriber line that can result in signal distortions or transmission errors.
-
Loop resistance measurement aids detection of short circuits and estimation of line
lengths.
-
DC current measurement can be used to detect emergency, external and normal
power feeds as well as breaks in the line.
-
Unbalance measurement (LCL): This measurement uses a frequency of 1 MHz to
measure asymmetry in a line pair.
-
NEXT measurement: Measurement of near-end crosstalk at a frequency of 1 MHz.
-
Signature detection is used to detect the signatures and test terminations (PPAs)
connected to the subscriber line.
-
The fast cable check verifies the correctness of a three-pole (Tip/Ring/Ground)
connection.
-
The integrated remote kit control lets you remotely set the line end to the desired state
either manually or automatically using a variety of electronic instrument kits.
ARGUS Copper Box
All measurements can be executed as highly accurate automatic TRG (Tip, Ring, Ground)
measurements.
With its light weight of less than 160 grams and sturdy plastic housing, the ARGUS Copper
Box is not only extremely compact but also insensitive to impacts, falls, and other
mechanical hazards. Despite its exceptional performance and high measuring voltage, the
Box can operate for exceptionally long periods, as it is supplied by the powerful Li ion
battery pack of the ARGUS tester.
Its compatibility enables the Box to be used with multiple ARGUS test instruments. Thus,
an installation team can expand multiple ARGUS meters independently of each other. It is
not necessary to return testers for servicing in order to expand their functionality. A special
protective rubber jacket forms the USB box and the ARGUS tester into a single unit.
The Copper Box can be kept up to date quickly and easily via firmware updates that
ARGUS executes automatically. This insures that the Copper Box and ARGUS firmware is
always optimally matched.
The standard scope of supply includes a high-quality connection lead (twisted), a ground
lead and this manual.
For further questions, please contact:
intec Gesellschaft für Informationstechnik mbH
Rahmedestr. 90
D-58507 Lüdenscheid
Tel.: +49 (0) 2351 / 9070-0
Fax: +49 (0) 2351 / 9070-70
www.argus.info/en
[email protected]
2 Safety information
ARGUS Copper Box
7
2 Safety information
General safety information:
The ARGUS Copper Box may only be operated using the accessories supplied with the
device. The use of other accessories can result in faulty measurements or even damage to
the ARGUS Copper Box and the connected equipment. Only use the ARGUS Copper Box
according to the instructions contained in this manual. Any other use can result in injuries to
persons and destruction of the ARGUS and/or the ARGUS Copper Box.
1.
Before connecting the ARGUS Copper Box to an access, make sure that no
dangerous voltages or currents are present for which the ARGUS Copper Box or its
accessories are not specified. Also keep in mind that the voltage can change over the
time that the device is connected.
2.
The ARGUS Copper Box is intended solely for use in telecommunications networks
with limited power. It is not intended to be used e.g. to measure mains voltages
(230 V/50 Hz).
Specifications:
DC voltage UDC (U=):
AC voltage UAC (U~):
Capacitance measurement C (C, Csym):
Isolation resistance measurement Iso (Iso):
Loop resistance R (R, Rsym):
DC current IDC (I=):
Longitudinal conversion loss, (LCL):
Near-end crosstalk (NEXT):
0.01 V to 200.0 V
0.01 V to 200.0 V (50 Hz sine)
0.01 nF to 8000.00 nF @ 8Hz
0.1 kΩ to 999.9 MΩ, (105 V, max. 2 mA)
0.1 kΩ to 40.0 MΩ, (8 V, max. 9 mA)
0.1 to 40.0 MΩ, (13 V, max. 15 mA)
0.1 mA to 499.9 mA
Attenuation for 1 MHz @120 Ω
Crosstalk attenuation for 1 MHz @120 Ω
3.
Use the ARGUS Copper Box only according to its intended purpose at all interfaces
and accesses. It is intended for the measurement and acquisition of physical quantities
in low-power telecommunication networks. The ARGUS Copper Box is not designed
for any other applications (e.g. measuring electronic components, determining the out-
put quantities of voltage sources, etc.).
4.
In each measurement, avoid any contact with electrically live components (sockets,
plugs, cables, adapters, etc.) under all circumstances.
5.
Voltages over 50 V AC and 120 V DC can cause death.
6.
The ARGUS Copper Box may only be used by trained personnel.
7.
The ARGUS Copper Box is not water-tight. Therefore, protect the ARGUS Copper Box
against water penetration.
8.
The USB connection lead of the ARGUS Copper Box may only be connected to
approved ARGUS testers (ARGUS 15x and ARGUS 16x). Each of these testers must
be enabled using an individual option key provided especially for this purpose. The
USB lead must not be extended.
2 Safety information
8
ARGUS Copper Box
9.
The electromagnetic compatibility (EMC) was tested according to the regulations
specified in our declaration of conformity. The ARGUS Copper Box is a class A device.
This device can cause radio interference in residential areas.
In this case, the user may be required to implement appropriate measures.
10. If the ARGUS Copper Box is operated under extreme conditions, an internal protective
mechanism (protection element) can be activated to protect the device, the Box and the
user. In this case, the Box may no longer be used, and may only be repaired by autho-
rized personnel. The device may not be opened by unauthorized personnel.
11.
To ensure dependable, long-term operation of the ARGUS Copper Box, always make
sure that it is optimally protected against high temperatures. The ARGUS Copper Box
may only be operated with the temperature range permitted for the ARGUS device
(-10 °C to +50 °C in battery operation, 0 °C to +40 °C in adapter operation).
2 Safety information
ARGUS Copper Box
9
Important instructions for use:
1.
Even a measurement on low-power telecommunications networks can present hazards
- particularly in the event of a fault; consequently, always first check the line to be tested
for a supply (DC voltage) and an external voltage (DC or AC voltage). If even one of
these conditions is present, safely de-energise this voltage and secure it against being
switched on again before conducting any further tests.
2.
Once ARGUS detects the ARGUS Copper Box, an auto test can be executed or each
separate measurement can be started and stopped individually. If you change the
measurement without first stopping, the measurement is stopped automatically.
All voltages >200V are outside the permissible limits.
3.
Never switch the ARGUS Copper Box from line to line while a measurement is running.
This would bypass the safety voltage measurement conducted before the start of each
measurement. Connecting the Box with a running resistance measurement to a line
with a regular supply voltage, for example, could destroy the Box even if this voltage is
within the specification for voltage measurement. In general, the ARGUS Copper Box
automatically checks before every measurement (except current measurement)
whether the line is carrying a voltage, and stops the measurement with a message as
necessary.
4.
The ARGUS Copper Box is equipped with an automatic “TRG-opt.” measurement. This
means that the measurement quantities can be measured between different wires,
such as a (Tip), b (Ring), Ground and optional socket (opt.). The ARGUS tester
displays for selection the pairings that the ARGUS Copper Box supports for the various
measurements. Except for measurement of isolation resistance, the ARGUS Copper
Box performs a continuous measurement when only one wire pair (e.g. Tip/Ring or Tip/
Ground or Ring/Ground) is selected, i.e. changing values appear immediately on the
display.
2 Safety information
10
ARGUS Copper Box
Special instructions for individual measurements
Voltage
Be sure to observe the permissible measuring range specified above when measuring
voltage.
Capacitance/capacitive symmetry:
Before measuring, make sure that the voltage thresholds given in the table below (see
page 11) are not exceeded during the measurement. Certain measurements can require a
longer settling time. An automatic “TRG-opt.” measurement can thus take some time.
Isolation resistance:
Before measuring, make sure that the voltage thresholds given in the table below (see
page 11) are not exceeded during the measurement. During the measurement, the
ARGUS Copper Box can apply a measuring voltage of up to 105 V max. 1 mA) to the line.
Certain measurements can require a longer settling time. An autotest can thus take some
time.
Loop resistance/resistance symmetry
Before measuring, make sure that the voltage thresholds given in the table below are not
exceeded during the measurement. During the measurement, the ARGUS Copper Box can
apply a measuring voltage of up ton 13 V (max. 15 mA) to the line.
DC current:
Be sure to observe the permissible measuring range specified above when measuring DC
current. Make sure you connect the ARGUS Copper Box in the circuit in series.
Longitudinal conversion loss/NEXT:
Before measuring, make sure that the voltage thresholds given in the table below are not
exceeded during the measurement. When measuring LCL or NEXT using the ARGUS
Copper Box, any type of additional measurement lead can falsify the measurement. Both
the length and the position of the measurement leads with respect to each other can
influence the measurement result significantly. The specification values apply only for the
Box itself and not for the system comprising the Box and the measurement leads.
Consequently, connecting the telecommunications line directly to the Box is recommended.
Signature detection/fast cable check/remote kit control
Before measuring, make sure that the voltage thresholds given in the table below are not
exceeded during the measurement.
2 Safety information
ARGUS Copper Box
11
Voltage thresholds
Measurement
UDC (V) to
UAC (Veff) to
R*
3.5
30
R-sym*
30
30
ISO-R*
5
30
C*
17
17
C-sym*
17
17
LCL
10
7
NEXT
10
7
Signature
3
3
Fast cable check**
5
30
Remote box**
3
3
* Measured on 200 kΩ load
** Loop limit value 10 kΩ
2 Safety information
12
ARGUS Copper Box
Return and environmentally compatible disposal
Current environmental law restricts the use of certain hazardous substances in electrical
and electronic devices, in particular the concentration or use of lead (Pb), cadmium (Cd),
mercury (Hg), hexavalent chromium [Cr(VI)], polybrominated biphenyls (PBB) and
polybrominated diphenyl ethers (PBDE).
We hereby confirm that, according to the assurances, marking and documentation of our
suppliers, ARGUS brand measuring products do not contain any substances in
concentrations, compounds or applications whose marketing is prohibited according to the
valid provisions of the RoHS Directive 2011/65/EU of the European Parliament and Council
dated 8 June 2011.
Our registration number issued by the EAR is: WEEE REG. no. DE 92829367.
Since October 2005, we have been marking all our measuring devices with this symbol, in
compliance with WEEE 2002/96/EC and the corresponding German statute ElektroG:
(
) (DIN EN 50419).
In other words, ARGUS and its accessories, may not be disposed of as household waste.
Please consult with our Service department with respect to the return of old devices.
3 General technical data
ARGUS Copper Box
13
3 General technical data
ARGUS user safety tested according to EN62368-2
RoHS conformity pursuant to the WEEE directive
The electromagnetic compatibility (EMC) was tested
according to the regulations specified in our declaration of
conformity.
CE marking
The ARGUS Copper Box complies with EC directives
2004/108/EC and 2009/C197/03. We will be happy to pro-
vide a detailed declaration of conformity on request.
Device specifications:
The ARGUS Copper Box is intended solely for measurements in
telecommunications networks with limited power.
Make sure that the inserted banana plugs of the measuring leads are completely
enclosed by the sockets of the Copper Box.
Dimensions/weight
Inputs/outputs
Height: 125 mm
Width: 74 mm
Depth: 22 mm
Weight: approx. 160 g
- 4 x 4 mm banana sockets (contact-protected)
- USB-A plug
Temperature range
Operating temperature (ARGUS in battery mode): -10 °C to +50 °C
Storage temperature: -20 °C to +60 °C
Humidity: up to 95 % relative, non-condensing
Miscellaneous
3 General technical data
14
ARGUS Copper Box
Measuring accuracies:
Voltage:
DC voltage measurement (U=):
Measuring range:
Resolution:
Accuracy:
0.01 V - 9.99 V
0.01 V
±(0.5 % + 2 digits)
10.0 V - 200.0 V
0.1 V
±(0.5 % + 2 digits)
AC voltage measurement (sine, 50 Hz) (U~):
Measuring range:
Resolution:
Accuracy:
0.01 V - 9.99 V
0.01 V
±(2 % + 2 digits)
10.0 V - 200.0 V
0.1 V
±(1.5 % + 2 digits)
Frequency: 10 Hz to 200 Hz; 0.2 Hz; ±(1.5 % + 2 digits); sine
When measuring voltages, the input resistance depends on the measuring range:
- Measuring range 1: 0 V - 25 V; Input resistance: 1.694 MΩ
- Measuring range 2: 25 V - 200 V; Input resistance: 1.05 MΩ
As soon as the total of all voltages exceeds 25 V, the measurement is switched over.
Capacitive symmetry measurement (CSym):
Measuring range:
Resolution:
Accuracy of capacity diffe-
rence:
10 nF - 4 µF
0.01 nF
0.1% of the capacity to
ground
Capacitive measurement, measuring frequency 8 Hz (C):
Measuring range:
Resolution:
Accuracy:*
0.01 nF - 9.99 nF
0.01 nF
±(4 % + 4 digits)
10.00 nF - 99.99 nF
0.01 nF
±(4 % + 4 digits)
100.0 nF - 999.9 nF
0.1 nF
±(3 % + 1 digit)
1000 nF - 8000 nF
1 nF
±(3 % + 1 digit)
* All specifications relate to a comparison measurement using film capacitors.
3 General technical data
ARGUS Copper Box
15
Isolation resistance measurement with 105 V, max. 2 mA (Iso.):
Measuring range:
Resolution:
Accuracy:
0.1 Ω - 99.9 Ω
0.1 kΩ
±(2 % + 1 digit)
100.0 kΩ - 999.0 kΩ
1 kΩ
±(2 % + 1 digit)
1.00 MΩ - 9.99 MΩ
10 kΩ
±(2 % + 1 digit)
10.0 MΩ - 99.9 MΩ
100 kΩ
±(5 % + 1 digit)
100.0 MΩ - 999.9 MΩ
100 kΩ
±(5 % + 1 digit)
Isolation resistance measurement with 8 V, max. 9 mA (Iso.):
Measuring range:
Resolution:
Accuracy:
0.1 Ω - 99.9 kΩ
0.1 kΩ
±(2 % + 1 digit)
100.0 Ω - 999.0 kΩ
1 kΩ
±(2 % + 1 digit)
1.00 MΩ - 9.99 MΩ
10 kΩ
±(2 % + 1 digit)
10.0 MΩ - 40.0 MΩ
100 kΩ
±(5 % + 1 digit)
Resistance difference (RSym):
Measuring range:
Resolution:
Accuracy:
10 Ω - 5 kΩ
0.1 Ω
± 0.2 % of RLoop ± 0.2 Ω
Loop resistance measurement with 13 V, max. 15 mA (Iso.):
Measuring range:
Resolution:
Accuracy:
0.1 Ω - 999.9 kΩ
0.1 Ω
±(1 % + 3 digits)
1.000 Ω - 9.999 kΩ
1 Ω
±(1 % + 1 digit)
10.00 Ω - 99.99 kΩ
10 Ω
±(1 % + 1 digit)
100.0 Ω - 999.9 kΩ
100 Ω
±(1 % + 1 digit)
1.000 MΩ - 9.999 MΩ
1 kΩ
±(2 % + 1 digit)
10.0 MΩ - 40.0 MΩ
10 kΩ
±(5 % + 1 digit)
DC current measurement (I=):
Measuring range:
Resolution:
Accuracy:
0.1 mA - 499.9 mA
0.1 mA
±(2.5 % + 3 digits)
3 General technical data
16
ARGUS Copper Box
Unsymmetrie balance at 1 MHz (LCL):
Measuring range:
Resolution:
Accuracy*:
0.1 dB - 55.0 dB
0.1 dB
± 1.5 dB
55.1 dB - 65.0 dB
0.1 dB
± 3 dB
The length of the measuring leads can significantly affect the measurement. Therefore,
always use the original accessories for measuring.
Near-end crosstalk at 1 MHz (NEXT):
Measuring range:
Resolution:
Accuracy:
0.1 dB - 65.0 dB
0.1 dB
± 1 dB
The length of the measuring leads can significantly affect the measurement. Therefore,
always use the original accessories for measuring.
Reference conditions (calibration):
- Temperature: +23 °C to 5 °C
- Humidity: 50 % ± 20 %, relative, non-condensing
- Frequency of measurement quantity: 50 Hz ± 5 Hz, sine
4 Configuring accesses
17
ARGUS Copper Box
4 Configuring accesses
This section describes how to select and configure the access type "Copper Test". First
though, note that ARGUS offers 100 user-configurable accesses, which can be configured
as DSL, ISDN or copper test accesses according to the customer's requirements.
Generally, some of these 100 accesses are pre-configured before delivery. At least one
access with the name "Copper tests" should appear in the list of accesses when you power
up ARGUS. This should be sufficient, as the Copper Box can be started on this access as
an individual test just like TDR or Line scope.
Switch on ARGUS.
The most recently used access is indi-
cated on the display with
.
ARGUS additionally displays a preview of
the selected access settings. The preview
window opens after 2 seconds.
You can also change the access name
"Copper tests", e.g. to in Copper Box,
TDR similar. See the ARGUS manual for
instructions on editing access names.
See section nocties of the ARGUS manual
for instructions on editing notices.
Creates an access, see
ARGUS manual.
Edits an access, see ARGUS
manual.
Toggles softkey assignment,
see ARGUS manual.
Switches to main menu.
Switches to Access, see
ARGUS manual.
5 Using the ARGUS Copper Box v4
18
ARGUS Copper Box
5 Using the ARGUS Copper Box v4
First connect the ARGUS Copper Box with the USB host interface of an ARGUS tester. You
may need to enter an option key in the ARGUS tester under Settings/Device/Software
options. Then connect the device to the measurement configuration as shown in the
connection examples.
Connection example 1 (DSLAM direction):
Connection example 2 (Modem direction):
5.1 Selecting and activating the Copper Box
The chapter "Configuring accesses", page 17, explains how to set up the connection type
"Copper test".
ARGUS in the status display.
The device always measures the DC
voltage at the line socket (pins 4/5) before
executing a copper test. This value
should warn of the presence of a voltage
at the line socket (e.g. TDR, Line scope)
when a copper test is executed.
Select one of the copper tests. In this
example Copper Box.
This is not the value that the
Copper Box measures between
the Tip and Ring wire.
ARGUS automatically switches
to the Copper Box auto test
configuration, page 27.
Copper Box
Continued on
next page
5 Using the ARGUS Copper Box v4
19
ARGUS Copper Box
ARGUS detects the Copper Box. This
can take a few seconds. If a conflict
occurs, update your firmware or consult
our Service Department.
When detection is complete, ARGUS
activates the Copper Box and starts
voltage measurement immediately.
ARGUS is in the Copper Box status
display.
The voltage between the Tip and Ring
wire is measured continuously here.
Shows the current measured
value.
The continuing measurement
is paused.
You can use the cursor
keys to select individual
tests, in this example
capacitive symmetry mea-
surement.
5 Using the ARGUS Copper Box v4
20
ARGUS Copper Box
Copper Box menu.
Depending on the currently selected test,
you can choose among the following
configurations (see in the sheet under),
here for example the hotkeys for
capacitive symmetry measurement
(Csym).
5.2 Hotkeys / settings
The hotkeys/settings can vary depending on the selected sockets.
Hotkey
Number key 0
Reset (only for active (min/max display))
Number key 1
Auto test profile, s. page 26
Number key 3
Polarity (Iso. only), s. page 52
Number key 4
Range limit (Iso. only), s. page 52
Number key 5
Measuring voltage (Iso. only), s. page 52
Number key 6
Wire types (R, RSym, C, CSym), s. page 31
Number key 7
QR code (only at the end of an auto test)
Number key 8
Susceptibility, s.page 43
Number key 9
Display mode, s. page 36
Input resistance, s.page 36
Press one after
another
and
Remote kit commands
commands, s. page 83
Remote kit s. page 81
Information bell, s. page 36
5 Using the ARGUS Copper Box v4
21
ARGUS Copper Box
You can find more information on the respective configurations in the chapter "Tests".
Press one after
another
and
Copper Box info. You can find more detailed information on the Copper
Box here, such as Copper Box type, software version, serial number,
etc.
Press one after
another
and
Save (only at the end of an auto test)
5 Using the ARGUS Copper Box v4
22
ARGUS Copper Box
5.3 Changing socket selection
By default, tests are conducted on the T/R
sockets.
Use socket selection to choose between
which wires you want to measure.
In the following example, a measurement
between sockets Tip and Ground are
selected.
You can select a variety of
socket combinations using
the cursor key.
Red
Tip (a-wire)
Black
Ring (b-wire)
Green
GND
Blue
opt. (optional)
White
not selected
5 Using the ARGUS Copper Box v4
23
ARGUS Copper Box
In this example, sockets Tip, Ring and
GND are selected.
Each measurement between two sockets (e.g. on Tip/Ring) is always performed
in real time.
5.4 Alarm tones
ARGUS generates various alarm tones in conjunction with the Copper Box, for instance as
soon as a fault occurs or a test has been concluded. The alarm tones must be activated in
the ARGUS settings, see Main Manual.
Short - short
Fault:
- Immediately when an external voltage is detected.
- In the absence of a loop or ground (RSym).
- In the absence of a ground/line or presence of a loop (CSym).
Short - long
Test complete:
- After every test that ended automatically.
- After every auto test - (not after every individual test here).
Short - short -
long
In the event of a fault at the end of a test.
Long
Prompts for user input.
6 Auto test
ARGUS Copper Box
25
6 Auto test
ARGUS executes various tests automatically, depending on the configured auto test pro-
file. Ten user-configurable profiles (including measurement assistance control) are avail-
able for auto test.
A voltage check is performed before every measurement (except current
measurement) in which the voltage limits given in the table (see page 11)
may not be exceeded.
The measuring ranges, resolutions and accuracies of the respective measurement may be
found in the chapter Technical data, see page 14.
Auto test settings
ARGUS in the status display.
open the profile
selection
The auto test is not yet started.
Hotkeys / settings auto test
See the respective chapters for
details on the individual
measurements. In the following,
the procedure is described using
an auto test example.
Auto test profile, s.
page 26
Wire types, s. page 31
Press one after
another
and
Remote kit command,
s. page 83
Remote kit, s. page 81
Press one after
another
and
Copper Box info, s.
page 21
6 Auto test
26
ARGUS Copper Box
Auto test profile selection
Select the profile to be editied and open it
with .
The auto test it configured under Auto
test configuration, see page 27.
6 Auto test
ARGUS Copper Box
27
Settings
Description
Auto test profiles:
Auto test/
auto test with
instrument kit/
Copper Box
auto test
All available Copper Box tests can be compiled in an automated test. Five
different profiles are available for saving.
Auto test
sequence
Sets the wait inter-
val until the next test. This
setting only applies for the
respectively selected test.
Range: 100 ms to 5000 ms.
Default: 100 ms
Adds further tests.
Deletes the
currently selected
test.
Selection of e.g. RSym
Different socket combinations
are available depending on
the selected test, see
page 29.
Confirm socket
selection with.
6 Auto test
28
ARGUS Copper Box
Selection of susceptibility to
interference, see also
page 59.
Select the start function:
- Automatic start
- Pause: manual loop
ARGUS waits for user input
to confirm that the loop is
connected.
The selected test, in this
example RSym, has been
added to the auto test.
Switches softkey
assignment
6 Auto test
ARGUS Copper Box
29
< -> > Shifts the marked test
one position to the right.
< <- > Shifts the marked test
one position to the left.
The added test and the
test sequence are saved.
Different socket combinations are available depending on the test.
U / R / Iso. /
C
Possibility 1: a1/T1, b1/R1
Possibility 2: a1/T1, GND
Possibility 3: b1/R1, GND
Possibility 4: a1/T1, b1/R1, GND
Possibility 5: a1/T1, opt.
Possibility 6: b1/R1, opt.
Possibility 7: a1/T1, b1/R1, opt.
Possibility 8: a1/T1, b1/R1, GND, opt.
I=
Possibility 1: a1/T1, b1/R1
Possibility 2: a1/T1, GND
Possibility 3: b1/R1, GND
Possibility 4: a1/T1, opt.
Possibility 5: b1/R1, opt.
CSym /
RSym
Possibility 1: a1/T1, b1/R1, GND
Possibility 2: a1/T1, b1/R1, opt.
Rem.
Possibility 1: short-circuit a/T-b/R
Possibility 2: exchange connect
Possibility 3: short-circuit a/T-b/R-GND
Possibility 4: open circuit
Possibility 5: activate tone generator
You must confirm the selected socket combination with
.
6 Auto test
30
ARGUS Copper Box
You can define the input resistance for voltage measurements. Determines
whether the voltage measurement is to be carried out with an input
resistance of 1 MΩ (high-ohm, standard) or low-ohm (200 kΩ).
Default: high-ohm (standard)
For R, Rsym, Iso., C and Csym measuremnets, you can select the
susceptibility which ARGUS sholud use for measuring.
Default: Reliable for external voltages
Notice
Allows you to enter a note, see main manual.
Profile name
Enter/modify name of editable auto test profile
Wire types:
Propagation
speed
To determine the correct distance, the calculation must take into account a
correction value dependent on the wire type that specifies the ratio of the
pulse propagation speed in the cable to the speed of light in a vacuum
(c0 = 299.792458 m/µs).
For cable types, the pulse time is also specified in V/2.
Minimum: 45.0 m/μs (VoP 30 %)
Maximum: 149.7 m/μs (VoP 99.9 %)
Default: 100.0 m/μs (VoP 66.7 %)
The choice of propagation speed as V/2 or VoP is saved.
Specific
resistance
Sets the specific resistance per kilometer
Range: 40 Ω/km to 400 Ω/km
Default: 353 Ω/km
Specific
capacitance
Sets the specific capacitance per kilometer
Range: 35 nF/km to 100 nF/km
Default: 48 nF/km
Name
Enter the name of the wire type.
Default: No wire type
Remote kit command
Remote kit
command
Commands s.page 85
Remote kit:
Remote kit
Remote kit type, s. page 83
6 Auto test
ARGUS Copper Box
31
List of preconfigured default wire types:
No. Name
Wire
diameter
(mm)
Line
resistance
(Ohm/km)
Mutual
capaci-
tance
(nf/km)
VoP
(%)
Note
1
PE 0.5 AWG24
0.5
80
39
66.0
Outdoor cable, Poly-
ethylene, air filled
2
PP 0.5 AWG24
0.5
80
39
66.0
Outdoor cable, Poly-
propylone, air filled
3
Pulp 0.4 AWG26 0.4
150
36
60.0
Outdoor cable, conductor
with pulp jacket, air filled
4
Pulp 0.5 AWG24 0.5
80
39
67.0
Outdoor cable, conductor
with pulp jacket, air filled
5
Pulp 0.6 AWG22 0.6
65
38
68.0
Outdoor cable, conductor
with pulp jacket, air filled
6
PIC 0.4 AWG26
0.4
150
36
64.0
Outdoor cable, conductor
with plastic jacket, air filled
7
PIC 0.5 AWG24
0.5
80
39
66.0
Outdoor cable, conductor
with plastic jacket, air filled
8
PIC 0.6 AWG22
0.6
65
38
67.0
Outdoor cable, conductor
with plastic jacket, air filled
9
PIC 0.9 AWG19
0.9
29
34
72.0
Outdoor cable, conductor
with plastic jacket, air filled
10
Jelly 0.9 AWG19 0.9
29
34
68.0
Outdoor cable, petroleum/
jelly filled cable
11
Filled 0.4
AWG26
0.4
150
36
58.0
Outdoor cable, filled
12
Filled 0.5
AWG24
0.5
80
39
60.0
Outdoor cable, filled
13
Filled 0.6
AWG22
0.6
65
38
62.0
Outdoor cable, filled
14
PTFE/Teflon 0.5
AWG24
0.5
80
39
67.0
Outdoor cable, conductor
with teflon jacket
15
Wire type 15
-
80
49
66.7
default, editable
16
Wire type 16
-
80
49
66.7
default, editable
17
Wire type 17
-
80
49
66.7
default, editable
18
Wire type 18
-
80
49
66.7
default, editable
19
Wire type 19
-
80
49
66.7
default, editable
20
Wire type 20
-
80
49
66.7
default, editable
6 Auto test
32
ARGUS Copper Box
Starting auto test
ARGUS in the status display.
The auto test is not yet started.
The default is the most recently used auto
test profile, in this example Copper box
auto test 1.
ARGUS displays the stored notice in the
preview.
Starts the auto test using the
selected auto test profile.
ARGUS executes the tests stored in the
profile in consecutive order.
During the tests, you can
select the respective results
using the cursor keys.
You can use the cursor keys
to select individual test
results, in this example
capacitance measurement.
The auto test has been executed.
Screen display for selection of a
wire type or auto test profile (in
this example Copper Box auto
test 1 and wire type 1).
6 Auto test
ARGUS Copper Box
33
When the auto test is complete, the
device automatically outouts a message
and prompts you to save the results.
After saving, ARGUS automatically swit-
ches to the view of the saved measured
values.
Press one
after another
and
ARGUS saves the current
measurement
Display of stored auto test values in
device.
ARGUS saves the result of the
auto test to the first free storage
slot in internal memory.
ARGUS returns to the last
selection menu without saving
results.
6 Auto test
34
ARGUS Copper Box
You can display the measuring result in
the form of a QR code using the number
key
.
The results can then be transferred to
other systems in csv format using a
camera or a suitable QR code reader.
exits QR code display and the
results are displayed as plain text.
7 Voltage measurement (U= and U~)
ARGUS Copper Box
35
7 Voltage measurement (U= and U~)
Voltage measurement enables detection of e.g. connected accesses (such as ISDN-BRI U
or POTS) and supply voltages (as for SHDSL using ZWR) and external voltages, such as a
connection to another line.
In AC voltage measurement, the frequency (10 - 200 Hz) is additionally displayed. This also
enables identification of call switching and coupled voltages as well as carrier or
power frequencies.
Note that unexpected or hazardous mixed voltages or hazardous voltage
peaks can occur. All voltages >200V are outside the permissible limits.
See the chapter Technical data page 14 for information on measuring ranges, resolutions
and accuracies of voltage measurement.
Setting Voltage measurement:
ARGUS in the status display.
The voltage measurement is not yet
started.
Starts voltage measurement
between sockets Tip and Ring.
Hotkeys/settings
7 Voltage measurement (U= and U~)
36
ARGUS Copper Box
Starting voltage measurement
ARGUS in the status display.
Continued on
next page
Voltage measurement starts automati-
cally when the Copper Box is activated. In
this example, the
- Current voltage in volts
- Current frequency of AC voltage in hertz
- Minimum voltage in volts
- Maximum voltage in volts
are measured.
Settings
Description
Display mode:
Determines whether the standard display (Normal display mode) or min./max display (incl.
min./max. display mode) is used.
Default: Normal display
Input resistance:
Sets the input resistance as high-ohm or low-ohm.
Default: high-ohm
Information bell:
Sets whether an acoustic signal sounds when a measured value is displayed.
Default: off
7 Voltage measurement (U= and U~)
ARGUS Copper Box
37
Display:
- Current DC voltage in volts
- Current AC voltage in volts
- Current frequency of AC voltage in hertz
Hotkeys/settings s. page 20
Selection of voltage measurement
By default, voltage measurements are
carried out using high-ohm values.
A change in the input resistance is stored
until ARGUS is restarted.
For high-ohm measurement, the input
resistance is at least 1 MΩ (depending on
the measuring range).
Voltage measurement can also be set to
a low-ohm range. Low-ohm measure-
ment is carried out using a 200 kΩ
resistor connected parallel to the input
impedance that is dependent on the
measuring range. Low-ohm measure-
ment permits discharge of coupled
voltage on lines (charged lines).
This load indicates the state (low-ohm,
high-ohm) of the source of any external
voltage.
7 Voltage measurement (U= and U~)
38
ARGUS Copper Box
The voltage measurement must be
restarted for the changes to take effect.
The example here shows a low-ohm
(loaded) measurement.
Switches the input
resistance.
To measure voltage using a different socket combination, see page 22.
The input resistance can only be
switched over during a two-pole
measurement.
7 Voltage measurement (U= and U~)
ARGUS Copper Box
39
Example measurements:
The measured DC voltage of 48.90 V can
indicate the POTS access of a PBX.
Typical POTS PBX voltages are 48 V,
32 V or 24 V DC.
The measured DC voltage of 96.2 V can
indicate an ISDN-BRI U access.
The voltage for an ISDN-BRI U access is
normally between 91 V and 99 V.
The measured voltage of 60.93 V could
indicate an POTS access.
The normal voltage range of an POTS
access is between 50 V and 72 V.
7 Voltage measurement (U= and U~)
40
ARGUS Copper Box
Ideally, no AC voltage should be
measured between individual wires.
Still, AC voltages can be coupled to the
wires directly via a line termination or
indirectly through the air.
The typical call-switching voltage can
occur transiently between Tip and Ring.
Display of the measurements between the
Tip (a-wire) and Ring (b-wire).
- Display of DC voltage in V
- Display of AC voltage in V
- Frequency of AC voltage in hertz
You can open the measur-
ing results of the wire pairs
using the cursor keys.
Opens the measuring results
for the wire pairs.
7 Voltage measurement (U= and U~)
ARGUS Copper Box
41
Example: AC voltage measurement
Incoming call
In this example the Copper Box is
connected so as to measure the voltage of
an POTS access.
A typical DC voltage is measured.
This example shows an incoming call. In
addition to the DC voltage, we can see a
call-switching voltage with the typical
frequency of 25 Hz.
8 Capacitive symmetry measurement (CSym)
42
ARGUS Copper Box
8 Capacitive symmetry measurement (CSym)
With the aid of capacitive symmetry measurement, it is possible to detect irregularities in
the wiring that can result in signal distortions or transmission errors. The line must be open.
In a real line, a network of serial and parallel capacitances is formed between the individual
reference points, as illustrated the following diagram.
The total capacitance Cab of the series connection of CaE and CbE is actually measured
between a (tip) and b (ring).
The measured value shows the mutual capacitance (Cm) value. This applies correspond-
ingly for measurements between a (tip) and ground (E) and between b (ring) and ground.
These measurements for the basis for calculating the actual values of Cab, CaE and CbE.
The calculated values are then used in determining symmetry.
Additionally, the absolute deviation of the two capacitances CaE and CbE with respect to
ground and the relative deviation are determined.
Relative deviation (in %) = 2x
Absolute deviation (in nF) = absolute difference (CaE-CbE)
The calculated values represent the actual values CaE, CbE and Cab for the individual
capacitances. As the capacitances with respect to ground CaE and CbE are particularly
important for the symmetry of a line, the absolute and percentage values are also given
here for the sake of clarity. The relative value should not exceed 1 % (recommendation).
The absolute deviation is of particular importance on extremely short lines, as even small
differences here can result in a higher-percentage deviation. It must be reviewed on a
case-by-case basis whether this is acceptable.
The line must be open. This can be achieved e.g. using the instrument kit
function "open circuit". A voltage check is performed before the measure-
ment in which the voltage limits given in the table (see page 11) may not be
exceeded.
See the chapter Technical data page 14 for information on capacitive symmetry
measurement.
8 Capacitive symmetry measurement (CSym)
ARGUS Copper Box
43
Capacitive symmetry measurement settings
ARGUS in the status display.
Capacitive symmetry measurement is not
yet started.
Hotkeys/settings of capacitive symmetry
measurement s. page 20 and in the sheet
under.
Under certain circumstances, the capacitive symmetry measurement can abort.
Possible causes include the presence of an external voltage, absence of a line or
the existence of a loop.
When you select the susceptibility
"insecure (fast)", the red warning
"ATTENTION: susceptible"
appears on the display.
If a wire type has been selected,
this is shown in the display.
8 Capacitive symmetry measurement (CSym)
44
ARGUS Copper Box
Starting symmetry measurement
ARGUS in the status display.
Depending on the line characteristics, the
measurement can take a few seconds.
Display:
When you select the susceptibility
"Insecure (fast)", a warning
appears in the display.
Starts capacitive symmetry
measurement.
- Cm (operating (mutual) capacitance)
(in nF)
- Cab (capacitance between Tip and Ring)
(in nF)
- CaE (capacitance between Tip and
Ground) (in nF)
- CbE (capacitance between Ring and
Ground) (in nF)
- ΔC (absolute deviation between CaE and
CbE) (in nF)
- Relative deviation (in %)
When a wire type is selected the
line length is calculated, see
page 45.
8 Capacitive symmetry measurement (CSym)
ARGUS Copper Box
45
When a wire type is selected the specific
capacitance and measured operating
capacitance are used to calculate the line
length.
Display:
Example measurements:
The capacitances CaE and CbE are very
close together. The relative deviation is
below 1 %. The wires are extremely
homogeneous.
The capacitances CaE and CbE deviate
greatly. The relative deviation is signifi-
cantly greater than the recommended
1 %.
The wires are extremely asymmetrical.
- Cm (operating (mutual) capacitance) (in nF)
- Line length in km
- Specific capacitance of wire type
9 Capacitance measurement (C)
46
ARGUS Copper Box
9 Capacitance measurement (C)
Capacitance measurement shows the typical input capacitance of connected devices or the
capacitance of the open line.
The line must be open. This can be ensured using the instrument kit
function "open circuit". Before the measurement, a voltage check is
performed in which the voltage limits given in the table (see page 11) may
not be exceeded.
The measuring ranges, resolutions and accuracies of the capacitance measurement may
be found in the chapter Technical data, see page 14.
Capacitance measurement settings
ARGUS in the status display.
Capacitance measurement is not yet
started.
Hotkeys/settings of capacitance mea-
surement see page 20 and in the sheet
under.
Once you have measured C-symmetry, you no longer need to measure
capacitance.
When you select the susceptibility
"Insecure (fast)", the red warning
"ATTENTION: susceptible"
appears on the display.
9 Capacitance measurement (C)
ARGUS Copper Box
47
Starting capacitance measurement
ARGUS in the status display.
In this example, a capacitance of approx.
48 nF is measured between each of the
sockets Tip/Ring and Ground.
For a specific capacitance of approx.
50 nF/km, this value indicates that the line
is approx. 1 km long and open.
No measurement is possible on a closed
line. ARGUS indicates this with the
message "Loop?".
Settings
Description
Wire types:
See page 30 for an explanation of wire types.
Susceptibility:
See page 43 for an explanation of susceptibility.
When you select the susceptibility
"Insecure (fast)", the red warning
"ATTENTION: susceptible"
appears on the display.
Starts capacitance
measurement
9 Capacitance measurement (C)
48
ARGUS Copper Box
Example measurements:
In this example, a capacitance of approx.
86.51 nF is measured between the
sockets Tip/Ring.
Ideally, the line should be in capacitive
balance. However, if the capacitance
between Tip and Ground and between
Ring and Ground differ by more than
2-3 %, this can indicate an asymmetry.
You can investigate this hypothesis by
measuring the symmetry (see Chapter 8),
which delivers a much more accurate
determination of symmetry.
Switches to the line length
parameters, see page 50.
9 Capacitance measurement (C)
ARGUS Copper Box
49
9.1 Calculating line length
Every telecommunication cable has a certain loop resistance and operating capacitance
per unit of length. As the type of cable laid is generally known, this value too is known. The
values are design-related and relatively constant, as the cables are laid in the earth and
subject to relatively minor temperature and humidity effects. However, a typical subscriber
line can consist of multiple segments. In these segments, the wire cross-sections, and thus
the technical specifications, can vary.
This can only be determined for a measurement between the Tip- and Ring-wires. The
measurements with respect to Ground can be subject to random influences and thus can-
not be specified with a general capacitance or resistance value per unit of length. The line
length calculations can then be compared with the line length, which is generally also
known, thus providing information about faults. Using the calculated value, you can roughly
compare the line lengths and conduct further fault-finding (e.g. with a TDR) in the event of
discrepancies.
A wire type must be selected to enable calculation of the line length using a C or R
measurement. See page 30 for these and further settings.
The wire type can be selected and configured during or after the measurement.
Starting capacitance measurement.
ARGUS in the status display.
In this example, capacitance measure-
ment is carried out with a measuring
frequency of 8 Hz and the values stored
for wire type 1.
It is important to select a cable type (see
page 31), the default is „No cable type“.
9 Capacitance measurement (C)
50
ARGUS Copper Box
In this example, a capacitance of approx.
22.30 nF is measured between Tip/Ring.
On the basis of the values stored under
wire type 1, this corresponds to a 455 m
line.
For a two-pole measurement the
measured value is displayed in real time.
Example measurements:
In this example, a capacitance of approx.
86 nF is measured between Tip/Ring.
For a capacitance of approx. 49 nF/km,
this value indicates that the line is approx.
1.77 km long.
Display:
10 Isolation resistance measurement (Iso.)
ARGUS Copper Box
51
10 Isolation resistance measurement (Iso.)
Isolation resistance measurement can indicate damage to cable isolation, moisture
penetration or oxidised contact points on the line.
A voltage check is performed before the measurement in which the voltage
limits given in the table (see page 11) may not be exceeded.
See the chapter Technical data page 14 for information on measuring ranges, resolutions
and accuracies of isolation resistance measurement.
Isolation resistance measurement settings
ARGUS in the status display.
Isolation resistance measurement is not
yet started.
Before the measurement, the selected
test parameters are displayed.
- Measuring voltage (in this example 105 V)
- Measuring current (in this example
max. 2 mA)
- Polarity (in this example standard)
Hotkeys/settings of Isolation resistance
measurement s. page 20 and in the sheet
under.
The configuration options can vary depending on the selected sockets.
Opens the Copper Box
menu, see page 20.
The polarity is inverted/
reversed, see page 52.
Starts the isolation
resistance measurement.
10 Isolation resistance measurement (Iso.)
52
ARGUS Copper Box
Settings
Description
Polarity:
Determines whether the polarity of the measurement is inverted. Standard is selected
such that e.g. a correctly connected passive test termination (PPA) is non-conducting. If
the PPA is reversed due to reversed connection, the display shows the value PPA
(~470 kΩ). If this occurs, reverse the polarity.
Default: Standard
Range limit:
Sets the limiting thresholds for isolation resistance measurement (see also page 14). The
measurement is only carried out up to the set threshold.
The threshold applies for isolation resistance measurement when a measurement voltage
of 105 V is selected.
Default: 1000 MΩ
Measurement voltage:
Determines whether isolation measurement is to be conducted at a high 105 V or low 8 V
voltage. With 8 V, it is possible e.g. to measure against a connected NTBA, so that this
does not become low-ohm (above approx. 10 V) and falsify the measurement.
Default: High voltage (105 V)
Susceptibility:
Selects the susceptibility with which ARGUS performs isolation resistance measurement.
Robust for ext.
voltage:
Insecure
(fast):
Default:
The measurement is reliable in the event of external voltages up to
17 V.
Using another measuring method, the measurement is extremely fast but
much more susceptible to interference.
Reliable for external voltages
Display mode:
Determines whether the standard display (Normal display mode) or min./max display
(incl. Min./max. display mode) is used.
Default: Normal display
10 Isolation resistance measurement (Iso.)
ARGUS Copper Box
53
Starting Isolation resistance measurement
ARGUS in the status display.
Isolation resistance measurement is not
yet started.
Current measuring polarity
Opens the Copper Box menu
Inverts the polarity,
see page 52.
Starts the isolation resistance
measurement.
When Isolation resistance
measurement is started, a
warning is displayed when a
measurement voltage of 105 V is
used.
ARGUS generates a voltage of
105 V at the start of the
measurement.
Special care is necessary due to
this generated voltage.
Stops the test.
10 Isolation resistance measurement (Iso.)
54
ARGUS Copper Box
Example measurements:
The example shows the measurement of
a resistance between Tip/Ring, Tip/
Ground and Ring/Ground of >1000 MΩ.
This indicates that the line is not
damaged. Depending on the require-
ment, a value of e.g >300 MΩ is
considered good. Values <300 MΩ and
>5 MΩ should be examined more closely.
Values under 5 MΩ indicate an isolation
fault.
In this example, the measurement shows
a markedly low resistance between Ring
and Ground. This indicates a fault.
Ideally, the values of Tip to Ground and
Ring to Ground should always be in
balance.
A value of 470 kΩ indicates a passive test
termination.
The value should be verified by reversing
the polarity with .
Fast „Isolation resistance measurement“,
e. g. of all pins at the demarcation point
with respect to each other and earth.
To save significant time in measuring, the
measuring range was set e. g. to „2 MΩ -
500 MΩ“ and susceptibility „uncertain
(fast)“ was selected. You can display val-
ues outside the measuring range using
(e. g. <2 MΩ or >500 MΩ).
When continuously measuring with the susceptibility „Reliable for external volt-
age“, the measured values cannot be updated directly after short-curcuiting the
line. The measurement can repeat up to four times before the display shows a
new value.
Starts a new measurement.
11 Resistance symmetry (RSym)
ARGUS Copper Box
55
11 Resistance symmetry (RSym)
This measurement determines the resistance symmetry of a line. Of primary importance for
resistance symmetry are the two resistances Ra and Rb, i.e. the resistances of the two
signal wires. To great a difference in resistance between the two wires can result in signal
distortions and transmission errors. In a real line, a network of serial resistances is formed
between the individual reference points, as illustrated the following diagram.
The measuring loops between Tip-Ring-Ground can be realised manually or
using an electronic instrument kit.
Ideal measuring configuration:
Values Ra, Rb and RE are not measured
directly, but instead are calculated from three
individual measurements (Rab, RaE, RbE).
The caculated values are then used in deter-
mining symmetry.
RE is dependent on the measuring current
here and is an auxiliary quantity. RE is gen-
erally significantly less (RE << Ra reasp. Rb)
than Ra or Rb.
The cable shield is generally connected to ground at the main distribution frame resp. at the
cable branch box. When the demarcation point is earthed, an electrical connection exists
via both the cable shield and the ground connection. Always make sure that the measuring
loops are properl connected to ground.
If the Tip-Ring-Ground measuring loops are not earthed, the measurement cannot be con-
ducted (e. g. clamping to painted, non-conducting termination frames with crocodile clips).
11 Resistance symmetry (RSym)
56
ARGUS Copper Box
Special features for RE (Examples):
1a. Measuring site, e. g. demarcation point:
The demarcation point is not earthed. The
Copper box (GND socket) is disconnected
from the shielding terminal of the demarca-
tion point. The measurement works when
the cbale shield os continuous. The dis-
played RE measurement is the R-cable
shield (Rshield). The measurement works
even if e. g. a shield terminal in the earth
cable becomes high-ohm due to an installa-
tion error or a lightning strike. This can then
be recognised from the high RE measured
value.
1b.However, if the cable shield is not contin-
uous (see image) or too high-ohm, the mea-
surement will no longer function
(RE>10 kΩ).
As a remedy, the reistance symmetry mea-
surement can be conducted using the build-
ing potential equalisation rail or an earthed
metal water or heating pipe. If this is not pos-
sible either, trained electricians can, after
checking the building configuration (TN-S,
TT), use the PE (protective ground) line as
an earthing point after verifying thar it is
voltage-free. In general, attention must be paid to ensuring secure contact, particularly to
oxidised or corroded earthing points.
The measurement does not work when the earthing is interrupted at one of the connection
points. However, it will work for a poor ground connection. Reason for cancellation of mea-
surement (RE >10 kΩ).
11 Resistance symmetry (RSym)
ARGUS Copper Box
57
2. Measureing site, e. g. demarcation point:
The demarcation point is not earthed. The
cable shield is interrupted at the site shown
in the illustration. However, the resistance
symmetry measurement can still work if the
cable shield at the demarcation point is con-
nected to one of the earthing points shown in
the picture. However, the RE measurement
is not Rshield, as the shield can have an
ground potential of varying quality due to the
lead sleeve in the ground (obsolete constuc-
tion method). The ground potential can also come from a network of building accesses
(other earthed demarcation points) in the branch cable: In the main cable area, the ground
potential can come from another cable branch box. In practice, a combination of the two
examples is quite common. All these effects lead to varying RE measurements.
3. TAE (residence)
Resistance symmetry is often faulty in the
last mile, e. g. due to a poor splice (rat-tail
splice, poorly crimped conductor connec-
tions, loose screw connection).
A measurement is possible if the last mile is
a connected to a buried, continuous shield
conductor this case is not illistrated). How-
ever, This is seldom the case. Thus, you
must either lay an auxiliary conductor, e. g.
from the potential equalistion point to the measurement, or use earthed harting pipes in the
residence.
If does not work, trained electricians can, after checking the bulidung mains configuration
(TN-S, TT), use the PE (protective ground) line as an earthing point after verifying that it
voltage-free. In general, attention must be paid to ensuring secure contact, particulary to
osidised or corroded earthing points.
4. If can intact conductor is used as an auxil-
ilary conductor instead of ground, RE=RH,
RH is then entirely on the order of magnitude
of Ra and Rb. A complete twisted pair (Tip-
Ring) can also be used for this.
11 Resistance symmetry (RSym)
58
ARGUS Copper Box
Calculation basis:
The following calculation basis is used for a 3-way measurement Rab, RaE, RbE, with R1,
R2, R3 as the total resistance of the respective line pairs.
Additionally, the absolute deviation of the two resistances Ra and Rb and the relative
deviation are determined.
Relative deviation (in %) = 2x
ΔR (absolute deviation in Ω) = absolute deviation (Ra - Rb)
A voltage check is performed before the measurement in which the voltage
limits given in the table (see page 11) may not be exceeded. A loop must be
created between Tip-Ring-Ground. This can be achieved using the instru-
ment kit function "Tip-Ring-Ground".
See the chapter Technical data page 14 for information on measuring ranges, resolutions
and accuracies of resistance symmetry measurement.
11 Resistance symmetry (RSym)
ARGUS Copper Box
59
Resistance symmetry measurement settings
ARGUS in the status display.
The measurement is not yet started.
Hotkeys/settings of resistance symmetry
measurement s. page 20 and in the sheet
under.
Under certain circumstances, the resistance symmetry measurement can abort.
Possible causes include the presence of an external voltage, absence of a line or
of a loop.
When you select the susceptibility
"insecure (fast)", the red warning
"ATTENTION: susceptible"
appears on the display.
If a wire type has been selected,
this is shown in the display (in this
example wire type 1).
Opens the Copper Box
menus, see page 20.
Starts the resistance
symmetry measurement.
Settings
Description
Wire types:
See page 30 for an explanation of wire types.
Susceptibility:
Selects the susceptibility with which ARGUS performs resistance symmetry
measurement.
Robust for external
voltage:
Insecure (fast):
Default:
The measurement is reliable in the event of external voltages up to
30 V.
Using another measuring method, the measurement is extremely fast
but much more susceptible to interference.
Robust for external voltages
11 Resistance symmetry (RSym)
60
ARGUS Copper Box
Starts resistance symmetry measurement
ARGUS in the status display.
The resistance symmetry measurement
is not yet started.
Depending on the line characteristics, the
measurement can take a few seconds.
Display:
Opens the Copper Box
menus, see page 20.
Starts the resistance symmetry
measurement.
- Rs (loop resistance) (in Ω)
- Ra (resistance of Tip-wire) (in Ω)
- Rb (resistance of Ring-wire) (in Ω)
- RE (resistance of Ground) (in Ω)
- ΔR (absolute deviation between Ra and Rb)
(in Ω)
- Relative deviation (in %)
11 Resistance symmetry (RSym)
ARGUS Copper Box
61
When a wire type is selected, the line
length is calculated from the specific
resistance.
Display:
Example measurements:
The resistances Ra and Rb are very
close together. The relative deviation is
below 1 %. The wires are extremely
homogeneous.
The resistances Ra and Rb deviate
greatly. The relative deviation is consider-
ably above the recommended 1 %. The
wires are extremely asymmetrical.
- Rs (sum of loop resistances) (in Ω)
- Ra (resistance of Tip-wire) (in Ω)
- Rb (resistance of Ring-wire) (in Ω)
- Line length in km
- Specific resistance of the selected wire
type
12 Loop resistance measurement (R)
62
ARGUS Copper Box
12 Loop resistance measurement (R)
Loop resistance measurement enables you e.g. to detect short-circuits and estimate line
lengths, among other things.
A voltage check is performed before the measurement in which the voltage
limits given in the table (see page 11) may not be exceeded. A loop must be
created. This can be ensured using the instrument kit function "short circ.
Tip-Ring-Ground".
Loop resistance measurement is not intended for single-wire deter-
minations. Use resistance symmetry measurement for this purpose, see
page 55.
See the chapter Technical data page 14 for information on measuring ranges, resolutions
and accuracies of loop resistance measurement.
Loop resistance measurement settings
ARGUS in the status display.
The loop resistance measurement is not
yet started.
Loop resistance is always measured
continuously.
Hotkeys/settings of loop resistance mea-
surement see page 20 and in the sheet
under.
When you select the susceptibility
"Insecure (fast)", the red warning
"ATTENTION: susceptible"
appears on the display.
Opens the Copper Box
menus, see page 20.
Starts loop resistance
measurement.
Selects whether an acoustic signal sounds when a measured value is displayed.
Default: off
12 Loop resistance measurement (R)
ARGUS Copper Box
63
Settings
Description
Wire types:
See page 30 for an explanation of wire types.
Susceptibility:
Selects the susceptibility with which ARGUS performs loop resistance measurement.
Robust for external
voltage:
Insecure (fast):
Default:
The measurement is reliable in the event of external voltages up to
3.5 V DC and 30 V AC.
Using another measuring method, the measurement is extremely fast
but much more susceptible to interference.
Robust for external voltages
Display mode:
Determines the normal display (current measured value) or min./max display (min./max.
values) is used.
Default: Normal display
12 Loop resistance measurement (R)
64
ARGUS Copper Box
Starts loop resistance measurement
ARGUS in the status display.
The loop resistance measurement is not
yet started.
In this example, a loop resistance of
420 Ω is measured.
To perform a loop resistance measurement using another socket combination,
see page 22.
Loop resistance measurement can also return no result (---) e.g. due to excessive
influence. The polarity is reversed up to 20 kΩ.
Opens the Copper Box
menus, see page 20.
Starts loop resistance
measurement.
Display of minimum and
maximum measured values.
The continuing
measurement is paused, see
also page 36.
12 Loop resistance measurement (R)
ARGUS Copper Box
65
Example measurements:
The example shows a measurement of
458 Ω. For a copper cable with a specific
resistance of 160 Ω / km, this indicates a
short-circuited twisted pair with a length
of 1.43 km.
As an explanation, the line can be
deliberately shorted at one end, or is
unintentionally shorted due to a fault in an
unknown location.
A large value such as 20 MΩ or >40 MΩ
indicates that no line loop (intentional
shorting) has been set up.
Loop resistance cannot be measured
then.
12.1 Calculating line length
See page 49 for calculation of line length using the R-values.
13 DC current measurement (I=)
66
ARGUS Copper Box
13 DC current measurement (I=)
DC current measurement enables detection e.g. of feeds, emergency feeds or
line terminations.
Before measuring, make sure that the Copper Box is connected in the
circuit in series.
The measuring range, resolution and accuracy of DC current measurement are described
in the chapter Technical data, see page 14.
DC current measurement settings
ARGUS in the status display.
The current measurement is not yet
started
DC voltage is always measured continu-
ously.
Hotkeys/settings of current measure-
ment s. page 20 and in the sheet under.
Opens the Copper Box
menus, see page 20.
Starts loop current
measurement.
Determines the normal display (current measured value) or min./max display (min./max.
values) is used.
Default: Normal display
Information bell:
Select whether an acoustic signal sounds when a measured value is displayed.
Default: off
13 DC current measurement (I=)
ARGUS Copper Box
67
Starting DC current measurement
ARGUS in the status display.
The DC current measurement is not yet
started.
In this example, a DC current of 57.5 mA
is measured.
To measure DC voltage using a different socket combination, see page 22.
Opens the Copper Box
menu, see page 20.
Starts the DC current
measurement.
Display of minimum and
maximum measured values.
The continuing
measurement is paused, see
also page 36.
14 Longitudinal conversion loss (LCL)
68
ARGUS Copper Box
14 Longitudinal conversion loss (LCL)
LCL measurement (asymmetry attenuation) looks at the balance between the Tip-wire with
respect to Ground (GND) compared to the Ring-wire with respect to Ground. In this
measurement, a tone is fed symmetrically to the wires with respect to Ground at 1 MHz. If a
difference can be measured between the two wires, an asymmetry is present.
A voltage check is performed before the measurement in which the voltage
limits given in the table (see page 11) may not be exceeded.
The length of the measuring leads can significantly affect the measurement,
so you should always perform your measurements using the original
accessories.
The measuring ranges, resolutions and accuracies of the LCL measurement are set out
in the chapter Technical data, see page 14.
LCL measurement settings
ARGUS in the status display.
The LCL measurement is not yet started.
Hotkeys/settings of LCL measurement
see page 20 and in the sheet under.
Opens the Copper Box
menu, see page 20.
Starts the LCL measurement.
Determines the normal display (current measured value) or min./max display (Min./max.
values) is used.
Default: Normal display
Information bell:
Selects whether an acoustic signal sounds when a measured value is displayed.
Default: off
14 Longitudinal conversion loss (LCL)
ARGUS Copper Box
69
Starting LCL measurement
ARGUS in the status display.
The LCL measurement is not yet started.
In this example, an asymmetry
attenuation of 54.1 dB is measured.
It is not possible to change the socket combination for LCL measurement.
Opens the Copper Box
menu, see page 20.
Starts LCL measurement.
Display of minimum and
maximum measured values.
The continuing
measurement is paused, see
also page 36.
14 Longitudinal conversion loss (LCL)
70
ARGUS Copper Box
Example measurements:
If the two wires are symmetric, i.e. identi-
cal in their mechanical properties,
a max. value is measured (in this exam-
ple >65 dB). Such a high value can only
be expected in ideal situations or in the
event of a short circuit between Tip and
Ring or an extremely short open line.
A value greater than 65 dB means that
the result is above the measuring range.
However, a value >40 dB (as in this
example 41.5 dB) is considered suffi-
ciently symmetric.
If a value <40 dB is measured, or even
7.8 dB as in this example, Tip or Ring
may be shorted to Ground.
15 Near-end crosstalk (NEXT)
ARGUS Copper Box
71
15 Near-end crosstalk (NEXT)
ARGUS generates a 1 MHz tone on line pair 1 and measures the crosstalk on the adjacent
line pair 2, which can cause severe performance issues for DSL. A customer's line can
pass through multiple segments. In this case, the measurement must be performed
segment by segment, and different line pairs may need to be used as the measuring lines.
A voltage check is performed before the measurement in which the voltage
limits given in the table (see page 11) may not be exceeded.
The length of the measuring leads can significantly affect the measurement,
so you should always perform your measurements using the original
accessories.
The measuring ranges, resolutions and accuracies of NEXT measurement are set out in
the chapter Technical data, see page 14.
Connection example (towards to the customer):
* Measurement is also
possible from the street
cabinet towards the
customer (segment by
segment).
Connection example (towards to the MDF):
* Measurement is also
possible from the street
cabinet towards the
MDF (segment by
segment).
15 Near-end crosstalk (NEXT)
72
ARGUS Copper Box
NEXT measurement settings
ARGUS in the status display.
The NEXT measurement is not yet
started.
Hotkeys/settings of NEXT measurement
see page 20 and in the sheet under.
Settings
Description
Displaymode:
Determines the normal display (current measured value) or min./max display (min./max.
values) is used.
Default: Normal display
Information bell:
Selects whether an acoustic signal sounds when a measured value is dispalyed.
Default: off
Opens the Copper Box
menu, see page 20.
Starts the DC current
measurement.
15 Near-end crosstalk (NEXT)
ARGUS Copper Box
73
15.1 Starting NEXT measurement
ARGUS in the status display.
The NEXT measurement is not yet
started.
The NEXT measurement is started.
In this example, a crosstalk attenuation
of 23.7 dB is measured.
It is not possible to change the socket combination for NEXT measurement.
Opens the Copper Box
menu, see page 20.
Starts NEXT measurement.
Display of minimum and
maximum measured values.
The continuing
measurement is paused, see
also page 36.
15 Near-end crosstalk (NEXT)
74
ARGUS Copper Box
Example measurements:
If there is no crosstalk between line pair 1
and line pair 2 (adjacent line), a maximum
value is reached (in this example 65 dB).
However, a value >37 dB (as in this
example 37.7 dB) considered sufficient.
The crosstalk to the adjacent line is
tolerable.
If a value <37 dB is measured, or even
5.8 dB as in this example, there may be a
short between the two wire pairs. The
tone is 100 % cross-coupled and severely
interferes with the line.
16 Signature detection/termination detection
ARGUS Copper Box
75
16 Signature detection/termination detection
Signature detection makes it possible to detect e.g. a passive test termination (PPA).
A voltage check is performed before the measurement in which the voltage
limits given in the table (see page 11) may not be exceeded.
Signature
PTT
PTT (inverted)
Circuit diagrams:
Starting signature detection
ARGUS in the status display.
Signature detection is not yet started.
Opens the Copper Box
menu, see page 20.
Starts signature detection.
16 Signature detection/termination detection
76
ARGUS Copper Box
Signature detection can take a few
seconds.
In this example, a passive test termina-
tion (PPA) was detected between Tip and
Ring.
See the table on page 77 for possible
results.
In this example, a PPA and a signature
circuit between Tip and Ring were
detected.
Stops the test.
Starts a new signature
detection.
Starts a new signature
detection.
16 Signature detection/termination detection
ARGUS Copper Box
77
16.1 Possible results
Test
Possible results
Characteristic
features
Remark
PPA
PPA detected
PPA inverted detected
PPA with reversed polarity (inverted)
detected.
Multiple PPAs?
Potentially multiple PPAs connected
in parallel detected.
Multiple inv. PPAs?
Potentially multiple PPAs with
reversed polarity (inverted) con-
nected in parallel detected.
2x PPA/Iso fault?
Potentially double PPA antiparallel
detected or isolation fault.
Signature Signature detected
Signature circuit detected.
Signature?
Signature possibly present:
External voltage, isolation resistance
or additional connected devices can
influence detection such that an
unambiguous result is no longer pos-
sible.
Unknown termination
Cp >500 nF
Unknown termination:
Cancellation, as no further detection
is possible, e.g. NTBA connected,
extremely long line.
Other
No termination detected
Loop?
R <8 kΩ
Cancellation, as no further detection
possible.
Isolation fault?
Rp <1 MΩ
Cancellation, as no further detection
possible.
17 Fast cable check
ARGUS Copper Box
79
17 Fast cable check
The fast cable check checks the correctness of a three-pole connection (Tip/Ring/Ground)
against a loop, in other words whether Tip/Ring/Ground is properly connected for subse-
quent tests.
To achieve this, the cable check measures the resistance. All values below a threshold of
10 kΩ, are interpreted as a loop, and all values greater as an open line. The cable check
can be performed as a standalone test or as part of the auto test. In auto test, an audible
alarm sounds when a fault is detected and the auto test sequence is stopped.
The loop between Tip/Ring/Ground can be realised manually or using an
electronic instrument kit.
Starting cable check
ARGUS in the status display.
The cable check is not yet started.
Hotkeys/settings of cable check see
page 20.
Opens the Copper Box
menu, see page 20.
Starts the cable check.
17 Fast cable check
80
ARGUS Copper Box
17.1
Test
Possible results
Remark
Line
Line OK
Line pair connected correctly
Line open
Line pair not connected correctly, e.g.
open.
Possible results
Opens the Copper Box
menu, see page 20.
Starts the cable check.
Starts a new cable check.
18 Remote kit control (Rem.)
ARGUS Copper Box
81
18 Remote kit control (Rem.)
The Copper Box remote instrument kit control (Rem.) function enables ARGUS to control
commonly available electronic instrument kits, such as the TX915/916.
Before using the remote control function, make sure that the voltage limits
specified in the table on page 11 are not exceeded during the measurement.
Remote control of other electronic instrument kits that may offer additional
functoins is only possible using the commands listed below.
Settings
Description
Remote kit type
TX91x + 900/905 remote
The common and extended 900/905 remote commands
can be acessed in the menu. When a common command is
selceted, both control conmands are transmitted. In the
extended 900/905 remote commands, only these control
commands are transmitted, and the acknowledgement
evaluated.
TX91x (orange)
The menu offers only the TX commands, and only the TX
control commands are transmitted.
900/905 remote (blue)
The menu offers all 900/905 remote commands, only these
commands are transmitted and the achnowledgement
evaluated.
18 Remote kit control (Rem.)
82
ARGUS Copper Box
Starting remote kit control
ARGUS in the status display.
In this example, the command
"Exchange connect" is selected.
Confirm the command.
The command "Exchange connect" is
sent to the remote unit.
Opens the Copper Box
menu, see page 20.
Selects the command for
remote kit control.
18 Remote kit control (Rem.)
ARGUS Copper Box
83
The command has been sent to the
remote unit.
Commands:
Description:
Short circuit a/T-b/R
The instrument kit short-circuits the Tip and Ring wires, e.g. for
loop resistance measurement.
Exchange connect
The instrument kit switches the line through. On a DSL access for
example directly to the DSLAM.
Short circuit a/T-
b/R-/GND
The instrument kit shorts Tip, Ring and Ground, e.g. for resistance
symmetry measurement.
Open circuit
The instrument kit opens a line, e.g. for a capacitance
measurement.
Tone mode
The tone generator in the instrument kit is activated. The tone can
be followed.
Port 1 loop 3 sec.
(90x)
The instrument kit type 900/905 creates a loop on port 1 for
3 sec.
Port 1+3 connect
(90x)
The instrument kit type 900/905 connects port 1 with port 3.
P1+2 & 3+4 connect
(90x)
The instrument kit type 900/905 connects port 1 with port 2 and
port 3 with port 4.
Port 1+4 connect
(90x)
The instrument kit type 900/905 connects port 1 with port 4.
Port 1+4 loop (90x)
The instrument kit type 900/905 creates a loop (short-circuit)
between port 1 and port 4.
Port 4 loop 3 sec.
The instrument kit type 900/905 creates a loop on port 4 for 3 sec.
Port 2+4 connect
(90x)
The instrument kit type 900/905 connects port 2 with port 4.
Port 3+4 connect
(90x)
The instrument kit type 900/905 connects port 3 with port 4
Selects and sends a new
command.
ARGUS Copper Box
19 Appendix
A) Abbreviations
Characters
@
at
Ω
Ohm (unit of electrical resistance)
ΔC
Absolute deviation between CaE and CbE
ΔR
Absolute deviation between Ra and Rb
A
a
a-wire (tip)
A
Ampere
APL
APL (German) copper line termination point
B
b
b-wire (ring)
C
C
1. Celsius
2. Capacitance
Cm
Operating capacitance (mutual capacitance)
Cp
Parallel capacitance
csv
File format (comma-separated values)
CSym
Capacitive symmetry
D
D
Germany
dB
Decibel
DE
Deutsch (German)
DIN
Deutsches Institut für Normung - German Institute for Standardization
DSL
Digital Subscriber Line
DSLAM
Digital Subscriber Line Access Multiplexer
E
E
Ground (earth)
EC
European Community
EMC
Electromagnetic compatibility
EN
European standard
F
F
Farad (unit of electrical capacitance)
G
g
Gram
GND
Ground (earth)
H
Hz
Hertz (unit of frequency)
ARGUS Copper Box
I
I
Electric current
IDC
DC current
ISDN
Integrated Services Digital Network
Iso.
Isolation resistance measurement
L
LCL
Longitudinal conversion loss
M
m
Meter
max
Maximum
min
Minimum
N
NEXT
Near-end crosstalk
NTBA
Network termination for ISDN basic rate access
O
opt.
Optional
P
PPA
Passive test termination
Pol.
Polarity
POTS
Plain Old Telephone Service (PSTN)
Q
QR
Quick response
R
R
1. Resistance (electrical resistance)
2. Ring (b-wire)
Ra
Resistance a-wire
Rb
Resistance b-wire
Rem.
Remote (instrument kit control)
RC
Resistance (R) and capacitance (C)
RE
Ground resistance
Ri
Internal resistance
Ring
b-wire
RoHS
Restriction of Hazardous Substances
Rs
Loop resistance
RSym
Resistance symmetry
S
S/N
Serial number
SHDSL
Single-Pair High-speed Digital Subscriber Line
Sig.
Signatures
Sym
Symmetry
T
ARGUS Copper Box
T
Tip (a-wire)
TAL
Subscriber line
TDR
Time domain reflectometry
TRG
Tip, Ring, Ground
U
U
Voltage
UAC
AC voltage
UDC
DC voltage
USB
Universal Serial Bus
V
V
Volt
V/2
Pulse propagation time
VoP
Velocity of propagation
Vpp
Volts Peak-to-peak
W
WEEE
Waste Electrical and Electronic Equipment
ARGUS Copper Box
B) Software licenses
The ARGUS firmware contains code from open-source packages published under a variety
of licenses (GPL, LGPL, MIT, BSD, etc.).
Additional information can be found on the CD-ROM included with your device (where
ordered, see Software_License.htm) or on the internet at:
http://www.argus.info/web/download/Software_License.
If you are interested in the sources listed under GPL/LGPL, please contact
[email protected]. intec Gesellschaft für Informationstechnik mbH will provide you with a
machine-readable copy of the source texts for a nominal fee to defray the costs of the
physical copying process. This offer is valid for 3 years.
intec Gesellschaft für Informationstechnik mbH
Rahmedestr. 90
D-58507 Lüdenscheid
Tel.: +49 (0) 2351 / 9070-0
Fax: +49 (0) 2351 / 9070-70
www.argus.info/en
[email protected]
ARGUS Copper Box
C) Index
A
Abbreviations ................................................................................................................ 84
Alarm tones ................................................................................................................... 23
Appendix ....................................................................................................................... 84
ARGUS
Dimensions ........................................................................................................... 13
Inputs and outputs ................................................................................................ 13
Weight ................................................................................................................... 13
Asymmetry attenuation ................................................................................................. 68
C
Calculating line length ............................................................................................. 49, 65
Capacitance measurement ........................................................................................... 46
Capacitance measurement examples .......................................................................... 47
Capacitance measurement settings ................................................................. 43, 46, 51
Capacitive symmetry example measurements ............................................................. 45
Capacitive symmetry measurement ............................................................................. 42
Changing socket assignment ........................................................................................ 22
Configuring accesses ................................................................................................... 17
D
DC current measurement ............................................................................................. 66
Declaration of conformity .............................................................................................. 13
default wire types .......................................................................................................... 31
Device specifications .................................................................................................... 13
DIN EN 50419 ............................................................................................................... 12
Disposal ........................................................................................................................ 12
E
Electromagnetic compatibility ....................................................................................... 13
ElektroG ........................................................................................................................ 12
EN60950-1 .................................................................................................................... 13
Example isolation resistance measurements ............................................................... 54
Example resistance symmetry measurements ............................................................. 61
F
Fast cable check ........................................................................................................... 79
H
Humidity ........................................................................................................................ 13
I
Index ............................................................................................................................. 88
Input resistance ............................................................................................................ 38
Instructions for use ......................................................................................................... 9
Instrument kits .............................................................................................................. 81
intec Gesellschaft für Informationstechnik mbH ....................................................... 6, 87
Introduction ..................................................................................................................... 4
Isolation resistance measurement ................................................................................ 51
L
LCL ............................................................................................................................... 68
ARGUS Copper Box
89
LCL measurement examples .........................................................................................70
Loop resistance measurement ......................................................................................62
Loop resistance measurement examples ......................................................................65
M
Measurement voltage ....................................................................................................52
Measuring accuracies ....................................................................................................14
N
NEXT .............................................................................................................................71
NEXT example measurements
..............................................................................................................................74
O
Operating temperature ..................................................................................................13
P
Polarity ...........................................................................................................................52
PPA ...............................................................................................................................76
R
Range limit .....................................................................................................................52
Resistance symmetry ....................................................................................................55
Return of old equipment ................................................................................................12
Rights ..............................................................................................................................2
RoHS conformity ...........................................................................................................13
RoHS Directive ..............................................................................................................12
S
Safety information ............................................................................................................7
Saves the results ...........................................................................................................33
Selecting and activating the Copper Box .......................................................................18
Signature detection ..................................................................................................75, 79
Software licenses ..........................................................................................................87
Special instructions for individual measurements ..........................................................10
Capacitance ..........................................................................................................10
DC current .............................................................................................................10
Isolation resistance ................................................................................................10
Longitudinal conversion loss/NEXT .......................................................................10
Loop resistance .....................................................................................................10
Signature detection/remote kit control ...................................................................10
Voltage ..................................................................................................................10
Specifications ..................................................................................................................7
AC voltage ...............................................................................................................7
Capacitance measurement .....................................................................................7
DC current ...............................................................................................................7
DC voltage ...............................................................................................................7
Isolation resistance measurement ...........................................................................7
Longitudinal conversion loss ...................................................................................7
Loop resistance .......................................................................................................7
Near-end crosstalk (NEXT) .....................................................................................7
Start symmetry measurement .......................................................................................25
Starting capacitance measurement .........................................................................43, 47
Starting DC current measurement .................................................................................67
ARGUS Copper Box
Starting isolation resistance measurement ....................................................................53
Starting symmetry measurement ...................................................................................32
Starts capacitance measurement. .................................................................................59
Starts loop resistance measurement .............................................................................63
Storage temperature ......................................................................................................13
Support ......................................................................................................................6, 87
Susceptibility ..................................................................................................................43
T
Technical features ...........................................................................................................5
U
User safety ....................................................................................................................13
Using the ARGUS Copper Box ......................................................................................18
V
Version ............................................................................................................................1
Voltage measurement examples ...................................................................................39
Voltage thresholds .........................................................................................................11
W
Website ......................................................................................................................6, 87
WEEE ............................................................................................................................12
Wire types ......................................................................................................................30