Benning PV 2 Operating Manual
Text of the manual
Top side of the device
Fig. 2:
Top side of the device
Bild 1:
Gerätefrontseite
Fig. 1:
Appliance front face
Fig. 1:
Front tester panel
Nullabgleich der Sicherheitsmessleitung für
RPE-Messung
Fig. 4:
Null balance of the measuring lead for RPE
measurement
Prüfung des Schutzleiterwiderstandes RPE
Fig. 5:
Testing the protective conductor resistance (RPE)
Bild 6:
Automatische Solarmodul-Messung, ISO-
Messung über Erdleiter
Fig. 6:
Automatic solar panel measurement, ISO
measurement via earthing conductor
Max.
Nicht an
parallel
geschalteten
PV-Strängen
messen!
Do not
measurements
at PV strings
which are
connected in
parallel!
PV-Modul/ -Strang
PV-module/ -string
PV-Modul/ -Strang
PV-module/ -string
Spannungsmessung über 4 mm Prüfbuchsen
Fig. 7:
Voltage measurement via 4 mm test sockets
Bild 8:
AC/ DC-Strommessung mit optionalem
Stromzangenadapter BENNING CC 3
Fig. 8:
AC/ DC current measurement by means
of optional current clamp adapter
BENNING CC 3
Isolationswiderstandsmessung RISO (2-polig)
Fig. 9:
Insulating resistance measurement RISO (2-pin)
Bild 10:
Anzeige der I-U Kennlinie über Android-Gerät
Fig. 10: Displaying the I-V characteristic via an Android device
PV-Modul/ -Strang
PV-modul/ -string
Funkverbindung zum optionalen Einstrahlungs- und Temperaturmessgerät BENNING SUN 2
Fig. 11:
Radio connection to the optional insolation and temperature measuring instrument BENNING SUN 2
Bild 12:
Batterie-/ Sicherungswechsel
Fig. 12:
Battery/ fuse replacement
Max.
PV-Modul/ -Strang
PV-module/ -string
BENNING PV 2
-
Falschpolung
Inhaltsverzeichnis
1. Benutzerhinweise
2. Sicherheitshinweise
3. Lieferumfang
4. Gerätebeschreibung
5. Allgemeine Angaben
6. Umgebungsbedingungen
7. Elektrische Angaben
8. Prüfen mit dem BENNING PV 2
8.1 Vorbereiten der Prüfung
8.1.1
Ein-, Ausschalten des BENNING PV 2
8.1.2
Einstellen der automatischen Abschaltzeit (APO, Auto-Power Off)
8.1.3
Einstellen von Datum und Uhrzeit
8.1.4
Prüfung des Batteriezustandes
8.1.5
Kompensation der Sicherheitsmessleitungen (Nullabgleich)
8.2 Prüfung des Schutzleiterwiderstandes, RPE
8.3 Automatische Messungen am PV-Generator
8.4 Gleich- und Wechselspannungsmessung
8.5 Strommessung mit optionalem AC/ DC-Stromzangenadapter BENNING CC 3
8.6 Isolationswiderstandsmessung (RISO, 2-polig)
8.7 Messwertspeicher
8.7.1
Messwerte speichern
8.7.2
Messwerte aufrufen
8.7.3
Messwertspeicher löschen
8.7.4
Messwertspeicher über USB-Schnittstelle auslesen
8.7.5
Darstellung der I-U Kennlinie über Android-Gerät
8.8 Funkverbindung zu Einstrahlungs- und Temperaturmessgerät BENNING SUN 2
8.8.1
Koppeln mit BENNING SUN 2
8.8.2
Entkoppeln vom BENNING SUN 2
8.8.3
Aktivieren/ Deaktivieren der Funkübertragung des BENNING SUN 2
9. Instandhaltung
9.1 Fehlercodes
9.2 Sicherstellen des Gerätes
9.3 Reinigung
9.4 Batteriewechsel
9.5 Sicherungswechsel
9.6 Kalibrierung
9.7 Ersatzteile
10. Umweltschutz
1. Benutzerhinweise
Diese Bedienungsanleitung ist für ausgebildetes Fachpersonal geschrieben!
Qualifiziertes Personal ist befähigt Risiken zu erkennen und mögliche Gefähr-
dungen zu vermeiden. Es besteht Verletzungsgefahr durch unsachgemäße
Handhabung!
Warnung vor gefährlicher elektrischer Spannung!
Beachten Sie unbedingt alle Sicherheitshinweise!
oder zum Tode
Der Anschluss an den PV-Generator ist ausschließlich gemäß den
Anschlussbildern der Bedienungsanleitung vorzunehmen.
Nicht benötigte Sicherheitsmessleitungen sind von dem BENNING PV 2 zu
trennen.
Der PV-Generator darf die maximale Leerlaufspannung von 1000 V, den maxi-
malen Kurzschlussstrom von 15 A und die maximale DC-Leistung (P = U x I)
von 10 kW nicht überschreiten.
Gemäß DIN EN 62446 sind die Messungen pro PV-Strang durchzuführen!
Eine Messung an parallel geschalteten PV-Strängen oder eine Kontaktierung
am PV-Wechselrichter kann zur Beschädigung des BENNING PV 2 führen!
Die PV-Prüfbuchsen
und
sind ausschließlich für die Kontaktierung mit
PV-Generatoren (PV-Modul, PV-Strang) vorgesehen.
Das Prüfgerät BENNING PV 2 direkt nach beendeter Prüfung vom PV-Generator
trennen.
Messspitzen nicht berühren!
Bei Isolationswiderstandsmessungen können hohe elektrische Spannungen an
den Messspitzen anliegen.
Während der Messung keine Metallteile des Prüfobjektes berühren.
Der PV-Generator muss von der elektrischen Hauptversorgung isoliert sein!
Weder Plus- noch Minuspol des PV-Generators darf geerdet sein!
Über die 4 mm Sicherheitsmessleitungen sind Spannungsmessungen an
Steckdosenstromkreise möglich. Das BENNING PV 2 darf über die 4 mm Prüf-
buchsen
und
nur in Stromkreisen der Überspannungskategorie III mit
max. 300 V AC/DC Leiter gegen Erde benutzt werden. Bei Spannungsmessun-
gen über die 4 mm Prüf buchsen
und
sind vorher die PV-Sicherheitsmess-
leitungen von den PV-Prüfbuchsen
und
zu trennen.
Extreme Vorsicht bei Arbeiten um blanke Leiter oder Hauptleitungsträger. Ein
Kontakt mit Leitern kann einen Elektroschock verursachen.
Beachten Sie, dass Arbeiten an spannungsführenden Teilen und Anlagen
grundsätzlich gefährlich sind. Bereits Spannungen ab 30 V AC und 60 V DC
können für den Menschen lebensgefährlich sein.
Vor jeder Inbetriebnahme überprüfen Sie das Gerät und die Leitungen auf Be-
schädigungen. Ein beschädigtes BENNING PV 2 nicht verwenden! Beschädig-
te Sicherheitsmessleitungen ersetzen!
Verwenden Sie ausschließlich, die im Lieferumfang des BENNING PV 2
enthaltenen Sicherheitsmessleitungen.
Das Prüfgerät BENNING PV 2 ausschließlich gemäß der in dieser Dokumenta-
tion angegebenen, bestimmungsgemäßen Verwendung einsetzen. Bei Nicht-
beachtung kann die Schutzfunktion des BENNING PV 2 beeinträchtigt werden.
Sollte das BENNING PV 2 verwendet werden um das Vorhandensein einer
gefährlichen Spannung nachzuweisen, prüfen Sie das BENNING PV 2 immer
vor und nach der Messung an einer bekannten Spannungsquelle oder
Prüfeinrichtung.
Das BENNING PV 2 nur in trockener Umgebung verwenden.
zu setzen und gegen unbeabsichtigten Betrieb zu sichern.
Um eine Gefährdung auszuschließen
-
berühren Sie die Leitungen nicht an den blanken Messspitzen,
-
stecken Sie die Leitungen in die entsprechend gekennzeichneten Buchsen
am Messinstrument
Wartung:
Das Gerät nicht öffnen, es enthält keine durch den Benutzer wartbaren Kompo-
nenten. Reparatur und Service kann nur durch qualifiziertes Personal erfolgen.
Reinigung:
Das Gehäuse regelmäßig mit einem Tuch und Reinigungsmittel trocken
abwischen. Kein Poliermittel oder Lösungsmittel verwenden.
3. Lieferumfang
+ PV-Prüfbuchse (rot)
– PV-Prüfbuchse (schwarz)
– 4 mm Prüfbuchse (schwarz)
+ 4 mm Prüfbuchse (rot)
-
Displayanzeige:
RPE Spannungs-Polaritätsanzeige
und
Stromzangen-Messung
RPE (Null-Offset),
-
standes.
(Achtung, heiße Oberfläche)
(Polaritätsanzeige)
und
an.
(Achtung, gefährliche Spannung)
(RPE LOCK)
-
Auswahl der Isolationsprüfspannung,
-
messung an.
Error (Fehler),
-
STORE,
RECALL,
Speicherplatzanzeige, Clr
NFC
-
Einstrahlungsänderung,
Batteriesymbol, erscheint bei entladenen Batterien
Kennliniensymbol
5. Allgemeine Angaben
aus.
-
-
6. Umgebungsbedingungen
und schwarze
und schwarze
-
-
7. Elektrische Angaben
-
7.1 Schutzleiterwiderstand RPE
Messbereich
Auflösung
Messgenauigkeit
7.2 PV-Modul/ PV-Strang, Leerlaufspannung, Vo/c
Messbereich
Auflösung
Messgenauigkeit
7.3 PV-Modul/ PV-Strang, Kurzschlussstrom, Is/c
Messbereich
Auflösung
Messgenauigkeit
7.4 Isolationswiderstand, RISO (AUTO-Messung des PV-Generators)
Messbereich
Auflösung
Messgenauigkeit
/ 1000 V
7.5 Isolationswiderstand, RISO (2-polig)
Messbereich
Auflösung
Messgenauigkeit
/ 1000 V
7.6 Spannung über 4 mm Prüfbuchsen
Messbereich
Auflösung
Messgenauigkeit
(DC, AC 50 Hz - 60 Hz)
1 V
7.7 Strom mit AC/ DC Stromzangenadapter BENNING CC 3
Messbereich
Auflösung
Messgenauigkeit
(DC, AC 50 Hz - 60 Hz)
8. Prüfen mit dem BENNING PV 2
8.1 Vorbereiten der Prüfung
-
Vor jeder Inbetriebnahme überprüfen Sie das Gerät, die Leitungen und das
Prüfobjekt auf Beschädigungen.
-
Durch gleichzeitiges Betätigen der
und
APO Auto-Power-
O
und
betätigt werden.
APO Auto-Power O
und
aus.
und betätigen Sie gleichzeitig die
und
. Halten Sie die
in Minuten an.
-
Jede Betätigung der
Minuten.
-
Lassen Sie die
und
aus.
und betätigen Sie gleichzeitig die
und
.
und die
und
Hinweis:
-
-
-
-
erscheint.
Messleitungswiderstände können bis 10 Ohm kompensiert werden.
-Messung
8.2 Prüfung des Schutzleiterwiderstandes, RPE
Der PV-Generator muss von der elektrischen Hauptversorgung isoliert sein!
Weder Plus- noch Minuspol des PV-Generators darf geerdet sein!
Bei Anliegen einer Spannung von > 30 V am Prüfobjekt wird die ermittelte
Spannung angezeigt. Steigt die Messspannung an den Prüfspitzen über 30 V
wird die RPE-Messung blockiert! Wird die RPE-Messung unterhalb von 30 V
gestartet, kann eine niederohmige Spannungsquelle die eingebaute Sicherung
im BENNING PV 2 auslösen!
-
buchse
und
.
speichern.
8.3 Automatische Messungen am PV-Generators
Der PV-Generator darf die maximale Leerlaufspannung von 1000 V, den maxi-
malen Kurzschlussstrom von 15 A und die maximale DC-Leistung (P = U x I)
von 10 kW nicht überschreiten.
Gemäß DIN EN 62446 sind die Messungen pro PV-Strang durchzuführen!
Eine Messung an parallel geschalteten PV-Strängen oder eine Kontaktierung
am PV-Wechselrichter kann zur Überlastung des BENNING PV 2 führen!
Der PV-Generator muss von der elektrischen Hauptversorgung isoliert sein!
Weder Plus- noch Minuspol des PV-Generators darf geerdet sein!
Alle Messleitungen sind sicher mit dem PV-Generator zu kontaktieren.
Verwenden Sie hierzu die beiliegenden Sicherheitsmessleitungen mit
PV-Stecker
sowie
die
Sicherheitsmessleitungen
mit
Prüfspitze
und
Krokodilklemme!
Trennen
Sie
die
kontaktierten
PV-Sicherheitsmessleitungen
während
einer Messung niemals von dem PV-Generator. Eine Trennung kann einen
elektrischen Lichtbogen und zur Überlastung des BENNING PV 2 führen.
Während der Isolationswiderstandsmessung wird der PV-Generator kurzge-
schlossen. Die Messung erfolgt zwischen der roten 4 mm Prüfbuchse und den
kurzgeschlossenen PV-Prüfbuchsen.
Falls die DC-Polarität fehlerhaft ist oder die DC-Spannung im Bereich von < 5 V
bzw. > 1000 V liegt, ist keine automatische PV-Messung möglich.
dem PV-Modul bzw. dem PV-Strang.
-
-
und
wird
bis
- Is/c Kurzschlussstrom
Isolationswiderstand
Mode
: „
- Is/c Kurzschlussstrom
Mode
„
- Is/c Kurzschlussstrom
Isolationswiderstand
Mode
-
5%
I-U Kennlinienmessung wird durchgeführt.
Bewegung
I-U Kennlinienmessung ist beendet.
-
schlussstrom Is/c wird eingeblendet
-
-
. Liegt der Isolationswider-
.
speichern.
VISO
Grenzwert Isolationswiderstand
1000 V
8.4 Gleich- und Wechselspannungsmessung
und
.
-
und
an.
+/–
-
+/–
speichern.
8.5 Strommessung mit optionalem AC/ DC-Stromzangenadapter BENNING CC 3
und
-
angezeigt.
speichern.
8.6 Isolationswiderstandsmessung RISO (2-polig)
Der PV-Generator muss von der elektrischen Hauptversorgung isoliert sein!
Weder Plus- noch Minuspol des PV-Generators darf geerdet sein!
Bei Anliegen einer Spannung von > 30 V am Prüfobjekt wird die ermittelte
Spannung angezeigt und die RISO-Messung wird blockiert! Kontaktieren Sie die
Messspitzen während der RISO-Messung nicht an eine Spannungsquelle, die
eingebaute Sicherung im BENNING PV 2 könnte auslösen!
und
.
angezeigt.
speichern.
ob die Messwerte gespeichert werden sollen.
speichern.
-
cherplatznummer
. Die
Speicherplatznummer
dargestellt.
wird zum vorherigen Speicherplatz gewechselt.
NFC
-
.
und
-
speichers abgebrochen.
-
-
.
Hinweise zum Datentransfer:
NFC
8.8 Funkverbindung zu Einstrahlungs- und Temperaturmessgerät BENNING SUN 2 -
Wireless SUN link
-
-
derlich.
-
und die
-
-
und die
Vor dem Öffnen das BENNING PV 2 unbedingt spannungsfrei machen!
Elektrische Gefahr!
ausschließlich Elektrofachkräften
vorbehalten, die dabei besondere Maßnahmen zur Unfallverhütung treffen müssen.
9.1 Fehlercodes
Fehlercode
Abhilfe
-
-
-
schritten. Die Messung wurde abgebrochen.
-
und betätigen Sie gleichzeitig
9.2 Sicherstellen des Gerätes
und gegen erneute Nutzung zu sichern.
9.3 Reinigung
-
-
-
9.4 Batteriewechsel
Vor dem Öffnen des Gerätes sind alle Messleitungen zu entfernen!
Elektrische Gefahr!
erscheint.
-
Schalten Sie das
-
Leisten Sie Ihren Beitrag zum Umweltschutz! Batterien dürfen nicht in den
Hausmüll. Sie können bei einer Sammelstelle für Altbatterien bzw. Sondermüll
abgegeben werden. Informieren Sie sich bitte bei ihrer Kommune.
9.5 Sicherungswechsel
Vor dem Öffnen des Gerätes sind alle Messleitungen zu entfernen!
Elektrische Gefahr!
und
-
dem Sicherungshalter.
-
Setzen Sie die neue Sicherung ein. Verwenden Sie nur Sicherungen mit gleichem
-
Hinweis:
und
-
9.6 Kalibrierung
9.7 Ersatzteile
10. Umweltschutz
BENNING PV 2
-
Table of contents
1. User notes
2. Safety note
3. Scope of delivery
4. Unit description
5. General information
6. Environment conditions:
7. Electrical specifications
8. Making measurements with the BENNING PV 2
8.1 Preparations for making measurements
8.1.1
Switching the BENNING PV 2 ON/ OFF
8.1.2
Setting the automatic switch-off time (APO, Auto-Power Off)
8.1.3
Setting the date and the time
8.1.4
Testing the battery condition
8.1.5
Compensation of the measuring leads (null balance)
8.2 Testing the protective conductor resistance (RPE)
8.3 Automatic measurements on the PV generator
8.4 AC/ DC voltage measurement
8.5 Current measurement by means of optional AC/ DC current clamp adapter BENNING CC 3
8.6 Insulating resistance measurement RISO (2-pin)
8.7 Measured value memory
8.7.1
Storing measured values
8.7.2
Calling measured values
8.7.3
Deleting the measured value memory
8.7.4
Reading out the measured value memory via the USB interface
8.7.5
Representing the I-V characteristic via an Android device
8.8 Radio connection to insolation / temperature measuring instrument BENNING SUN 2
8.8.1
Coupling with BENNING SUN 2
8.8.2
Decoupling from the BENNING SUN 2
8.8.3
Activating / deactivating the radio transmission of the BENNING SUN 2
9. Maintenance
9.1 Error codes
9.2 Securing the instrument
9.3 Cleaning
9.4 Battery replacement
9.5 Fuse replacement
9.6 Calibration
9.7 Spare parts
10. Environmental notice
1. User notes
This operating manual is intended for qualified technical personnel! Qualified
technical personnel is competent to identify risks and to prevent possible
hazards. Improper handling involves the risk of injury!
Warning of dangerous electric voltage!
Absolutely observe all safety instructions!
Protection class II
2. Safety note
the notes and warnings given in these instructions at all times. Improper handling and non-
injuries or danger to life.
The connection to the PV generator is made exclusively in accordance with the
connection figure of the operating instructions.
Disconnect not required safety test leads from the BENNING PV 2.
The PV module/ PV string must not exceed the maximum opencircuit voltage
of 1000 V DC, the maximum short-circuit current of 15 A and the maximum DC
power (P = U x I) of 10 kW.
According to DIN EN 62446, the measurements have to be made for each PV
string.
Measuring at PV strings connected in parallel or contacting at the PV inverter
might involve damages of the BENNING PV 2!
The PV test sockets
and
are intended exclusively for the connection with
PV generators (PV module, PV string).
Disconnect the BENNING PV 2 from the PV generator directly after the test.
Do not touch the measuring probes!
During insulating resistance measurements, high electric currents might be
applied to the measuring probes.
Do not touch any metal parts of the test object during measurement.
The PV generator must be isolated from the electric power supply!
Neither the positive nor the negative pole of the PV generator must be earthed!
Via the 4 mm safety test leads, voltage measurements on mains supply circuits
are possible. Via the 4 mm test sockets
and
, the BENNING PV 2 must be
used only in electric circuits of overvoltage category III with max. 300 V AC/ DC
for phase-to-earth measurements. For voltage measurements via the 4 mm test
sockets
and
, please disconnect the PV measuring leads from the PV test
sockets
and
before measuring.
WARNING! Be careful when working with bare conductors or main line carrier!
Contact with live conductors will cause an electric shock!
Remember that work on electrical components of all kinds is dangerous. Even
low voltages of 30 V AC and 60 V DC may be dangerous to human life.
Before starting the appliance tester up, always check it for signs of damage.
Do not use a damaged BENNING PV 2! Damaged measuring leads must be
replaced!
Only use safety measuring leads, which are supplied with the BENNING PV 2.
Use the BENNING PV 2 only in compliance with the intended use specified in
this documentation. If the BENNING PV 2 is used in a manner not specified by
this document then the protection provided by the equipment may be impaired.
If the BENNING PV 2 is being used to determine the presence or absence
of hazardous voltages, always prove the operation of voltage measurement
function before and after use by means of a known voltage source or proving
unit.
Use the BENNING PV 2 under dry ambient conditions only.
-
the device is exposed to moisture.
In order to prevent danger
-
do not touch the bare measuring probe tips of the measuring leads,
-
plug the leads into the correspondingly marked jacks at the measuring
instrument
Maintenance:
Do not open the tester, because it contains no components which can be
repaired by the user. Repair and service must be carried out by qualified
personnel only!
Cleaning:
Regularly wipe the housing by means of a dry cloth and cleaning agent. Do not
use any polishing agents or solvents!
3. Scope of delivery
4. Unit description
Digital display
-key
-key
-key
-key
-key
-key
NFC sensor
-key
USB interface
-key
-key
+ PV test socket (red)
- PV test socket (black)
- 4 mm test socket (black)
clip
+ 4 mm test socket (red)
clip
Digital display:
RPE voltage polarity indication
and
Current clamp measurement enabled
RPE null offset
lead resistance
Attention, hot surface!
has disappeared.
Polarity indication
and
Attention
RPE LOCK
Attention
manual in order to avoid danger.
RISO
preset limiting values
Selection of the insulation test voltage
Storage location indication
Clr
NFC
-
Change of insolation
Battery symbol
Curve symbol
5. General information
-
PV 2 to cool down.
6. Environment conditions:
the red
the red
7. Electrical specifications
7.1 Protective conductor resistance RPE
Measuring range
Resolution
Measuring accuracy
7.2 PV module/ PV string, open-circuit voltage, (Vo/c)
Measuring range
Resolution
Measuring accuracy
0.1 V
7.3 PV module/ PV string, short-circuit current, (Is/c)
Measuring range
Resolution
Measuring accuracy
7.4 Insulation resistance RISO (AUTO measurement of the PV generator)
Messbereich
Auflösung
Messgenauigkeit
/ 1000 V
7.5 Insulating resistance, RISO (2-pin)
Measuring range
Resolution
Measuring accuracy
/ 1000 V
7.6 Voltage via 4 mm test sockets
Measuring range
Resolution
Measuring accuracy
(DC, AC 50 Hz - 60 Hz)
1 V
7.7 Current by means of AC/ DC current clamp adapter BENNING CC 3
Measuring range
Resolution
Measuring accuracy
(DC, AC 50 Hz - 60 Hz)
8. Measuring with the BENNING PV 2
8.1 Preparation for measuring
readings and measuring errors.
Before starting the BENNING PV 2, always check the device, the leads and the
test object for damages.
and
APO Auto Power-
O
and
APO Auto-Power-O
and
.
-
Press and hold the
and
. Keep
the
pressed.
second line.
to store the setting.
and
.
-
Press and hold the
and
.
-
Press the
-
Press the
or the
to increase or decrease the value.
and
to store the set-
ting.
Note:
-
-
until the
.
Measuring lead resistances can be compensated by up to 10 Ohm.
measurement
8.2 Testing the protective conductor resistance (RPE)
The PV generator must be isolated from the electric power supply! Neither the
positive nor the negative pole of the PV generator must be earthed!
If a voltage of > 30 V is applied to the test object, the determined voltage will
be displayed. If the measuring voltage at the probe tips exceeds 30 V, the RPE
measurement will be blocked! If the RPE measurement is started below 30 V, a
low-impedance voltage source might trip the built-in fuse of the BENNING PV 2!
and
and connect the test object.
-
Press the
.
-
Press the
to store the measured value to the next storage location available.
and
and connect the test object.
-
Press and hold the
-
Press the
to terminate the continuous measurement.
-
Press the
to store the measured value to the next storage location available.
8.3 Automatic measurements on the PV generator
The PV generator must not exceed the maximum opencircuit voltage of 1000 V
DC, the maximum short-circuit current of 15 A and the maximum DC power (P
= U x I) of 10 kW.
According to DIN EN 62446, the measurements have to be made for each PV
string.
Measuring at PV strings connected in parallel or contacting at the PV inverter
might involve damages of the BENNING PV 2!
The PV generator must be isolated from the electric power supply! Neither the
positive nor the negative pole of the PV generator must be earthed!
Make sure that all measuring lines are safely connected to the PV generator.
For this, use the enclosed safety measuring lines with PV connector as well as
the safety measuring lines with probe tip and alligator clip!
Never disconnect the connected PV safety measuring lines from the PV
generator during measurement! Disconnection might involve an electric arc
and overload of the BENNING PV 2.
During the insulating resistance measurement, the PV generator is shorted.
Measurement is made between the red 4 mm test socket and the shorted PV
test sockets.
In case DC polarity is wrong or DC voltage is within the range of < 5 V or > 1000
V, it is not possible to make an automatic PV measurement.
and
warning
-
Press the
to
Mode
I-V curve measurement is running.
I-V curve measurement is completed.
-
Press the
is pressed.
-
Press the
to store the measured value to the next storage location available.
Limiting value of insulating resistance
1000 V
8.4 AC/ DC voltage measurement
and
.
and
.
will
-
Press the
to store the measured value to the next storage location available.
8.5 Current measurement by means of optional AC/ DC current clamp adapter
BENNING CC 3
and
to select the Mode
.
-
Press the
to store the measured value to the next storage location available.
8.6 Insulating resistance measurement RISO (2-pin)
The PV generator must be isolated from the electric power supply!
Neither the positive nor the negative pole of the PV generator must be earthed!
If a voltage of > 30 V is applied to the test object, the determined voltage will
be displayed and the RISO measurement will be blocked! Do not connect the
measuring probes to a voltage source during RISO measurement, as the built-in
fuse of the BENNING PV 2 might trip!
and
and connect the test object.
-
Press the
-
Press the
-
.
.
-
Press the
to store the measured value to the next storage location available.
and
and connect the test object.
-
Press the
-
Press and hold the
-
.
-
Press the
-
to terminate the continuous measurement.
-
Press the
to store the measured value to the next storage location available.
8.7 Measured value memory
date/ time stamp.
the measured values shall be stored.
-
Press the
to store the measured value to the next storage location available.
-
Press the
to recall the stored measured values with the corresponding storage
location number
.
-
Press the
again to go to the previous storage location.
-
Press the
to go to the next storage location.
. Press the
NFC
-
-
Press the
.
and the
.
hardware can be used.
-
Press the
.
-
Press the
Note:
-
-
and write the
Notes on data transfer:
NFC
is shown
8.8 Radio connection to insolation / temperature measuring instrument BENNING SUN 2 –
wireless SUN link
and the
2
-
Press and hold the
and the
seconds.
above the
-
measured insolation value is outside the measuring range.
-
Note:
Before opening the BENNING PV 2, make quite sure that it is voltage free!
Electrical danger!
may be carried out only by skilled
electricians with special precautions for the prevention of accidents.
9.1 Error codes
Error codes
Remedy
-
reached the maximum admissible temperature. Discon-
and let it cool down.
generator!
oder
etc.
-
again to the next storage location available.
to start the measurement.
9.2 Securing the instrument
-
Incorrect measurement results.
9.3 Cleaning
9.4 Battery replacement
Before opening the BENNING PV 2 ensure that all the test leads have been
disconnected from the BENNING PV 2. Danger of electric shock!
unit
.
Make your contribution to environmental protection! Do not dispose of
discharged batteries in the household garbage. Instead, take them to a
collecting point for discharged batteries and special waste material. Please
inform yourself in your community.
9.5 Fuse replacement
Before opening the BENNING PV 2 ensure that all the test leads have been
disconnected from the BENNING PV 2. Danger of electric shock!
and
driver.
Benning Elektrotechnik & Elektronik GmbH & Co. KG
Münsterstraße 135 - 137
D - 46397 Bocholt
Phone: +49 (0) 2871 - 93 -
- 93 - 429