PUBLICATION NO. 980914-100X
RACAL INSTRUMENTS
United States
(Corporate Headquarters and Service Center)
4 Goodyear Street, Irvine, CA 92618
Tel: (800) 722-2528, (949) 859-8999; Fax: (949) 859-7139
5730 Northwest Parkway Suite 700, San Antonio, TX 78249
Tel: (210) 699-6799; Fax: (210) 699-8857
Europe
(European Headquarters and Service Center)
18 Avenue Dutartre, 78150 LeChesnay, France
Tel: +33 (0)1 39 23 22 22; Fax: +33 (0)1 39 23 22 25
29-31 Cobham Road, Wimborne, Dorset BH21 7PF, United Kingdom
Tel: +44 (0) 1202 872800; Fax: +44 (0) 1202 870810
Via Milazzo 25, 20092 Cinisello B, Milan, Italy
Tel: +39 (0)2 6123 901; Fax: +39 (0)2 6129 3606
Racal Instruments Group Limited, Technologie Park, D-51429 Bergisch Gladbach, Germany
Tel: +49 2204 844205; Fax: +49 2204 844219
[email protected]
[email protected]
[email protected]
http://www.racalinstruments.com
[email protected]
www.racalinstruments.de
PUBLICATION DATE: November 23, 2005
Copyright 2005 by Racal Instruments, Inc. Printed in the United States of America. All rights reserved.
This book or parts thereof may not be reproduced in any form without written permission of the publisher.
THANK YOU FOR PURCHASING THIS RACAL INSTRUMENTS PRODUCT.
For this product, or any other Racal Instruments product that incorporates software drivers, you may access
our web site to verify and/or download the latest driver versions. The web address for driver downloads is:
http://www.racalinstruments.com/downloads
If you have any questions about software driver downloads or our privacy policy, please contact us at
[email protected]
WARRANTY STATEMENT
All Racal Instruments, Inc. products are designed and manufactured to exacting standards and in full
conformance to Racal’s ISO 9001:2000 processes.
This warranty does not apply to defects resulting from any modification(s) of any product or part without Racal
Instruments express written consent, or misuse of any product or part. The warranty also does not apply to
fuses, software, non-rechargeable batteries, damage from battery leakage, or problems arising from normal
wear, such as mechanical relay life, or failure to follow instructions.
This warranty is in lieu of all other warranties, expressed or implied, including any implied warranty of
merchantability or fitness for a particular use. The remedies provided herein are buyer’s sole and exclusive
remedies.
For the specific terms of your standard warranty, or optional extended warranty or service agreement, contact
your Racal customer service advisor. Please have the following information available to facilitate service.
1. Product serial number
2. Product model number
3. Your company and contact information
You may contact your customer service advisor by:
E-Mail:
[email protected]
Telephone:
+1 800 722 3262
(USA)
+44(0) 8706 080134
(UK)
Fax:
+1 949 859 7309
(USA)
+44(0) 1628 662017
(UK)
RETURN of PRODUCT
Authorization is required from Racal Instruments before you send us your product for service or calibration.
Call your nearest Racal Instruments support facility. A list is located on the last page of this manual. If you are
unsure where to call, contact Racal Instruments, Inc. Customer Support Department in Irvine, California, USA
at 1-800-722-3262 or 1-949-859-8999 or via fax at 1-949-859-7139. We can be reached at:
[email protected].
PROPRIETARY NOTICE
This document and the technical data herein disclosed, are proprietary to Racal Instruments, and shall not,
without express written permission of Racal Instruments, be used, in whole or in part to solicit quotations from
a competitive source or used for manufacture by anyone other than Racal Instruments. The information
herein has been developed at private expense, and may only be used for operation and maintenance
reference purposes or for purposes of engineering evaluation and incorporation into technical specifications
and other documents which specify procurement of products from Racal Instruments.
DISCLAIMER
Buyer acknowledges and agrees that it is responsible for the operation of the goods purchased and should
ensure that they are used properly and in accordance with this handbook and any other instructions provided
by Seller. Racal Instruments products are not specifically designed, manufactured or intended to be used as
parts, assemblies or components in planning, construction, maintenance or operation of a nuclear facility, or
in life support or safety critical applications in which the failure of the Racal Instruments product could create
a situation where personal injury or death could occur. Should Buyer purchase Racal Instruments product for
such unintended application, Buyer shall indemnify and hold Racal Instruments, its officers, employees,
subsidiaries, affiliates and distributors harmless against all claims arising out of a claim for personal injury or
death associated with such unintended use.
FOR YOUR SAFETY
Before undertaking any troubleshooting, maintenance or exploratory procedure, read carefully the
WARNINGS and CAUTION notices.
This equipment contains voltage hazardous to
human life and safety, and is capable of
inflicting personal injury.
If this instrument is to be powered from the AC line (mains) through an auto-
transformer, ensure the common connector is connected to the neutral (earth pole)
of the power supply.
Before operating the unit, ensure the conductor (green wire) is connected to the
ground (earth) conductor of the power outlet. Do not use a two-conductor extension
cord or a three-prong/two-prong adapter. This will defeat the protective feature of the
third conductor in the power cord.
Maintenance and calibration procedures sometimes call for operation of the unit with
power applied and protective covers removed. Read the procedures and heed
warnings to avoid “live” circuit points.
Before operating this instrument:
1. Ensure the proper fuse is in place for the power source to operate.
2. Ensure all other devices connected to or in proximity to this instrument are properly grounded or
connected to the protective third-wire earth ground.
If the instrument:
-
fails to operate satisfactorily
-
shows visible damage
-
has been stored under unfavorable conditions
-
has sustained stress
Do not operate until, performance is checked by qualified personnel.
1260-100X User Manual
i
Table of Contents
Chapter 1 ...............................................................................................................................1-1
OVERVIEW......................................................................................................................................1-1
Items Shipped with 1260-100X.....................................................................................................1-1
Part Numbers, Options and Ordering Information .......................................................................1-1
Functional Description..................................................................................................................1-2
Features .......................................................................................................................................1-3
Compatibility with Racal Instruments 1260-Series ...................................................................1-3
Ease of Use ..............................................................................................................................1-4
Maintainability...........................................................................................................................1-4
Dynamically-Configured Analog Bus ........................................................................................1-5
Chapter 2 ...............................................................................................................................2-1
GETTING STARTED .......................................................................................................................2-1
Unpacking and Inspection............................................................................................................2-1
Reshipment Instructions...............................................................................................................2-1
Option-01T ...................................................................................................................................2-2
Setting Configuration Switches ....................................................................................................2-4
Local Bus Daisy-Chain Switches..............................................................................................2-5
Module Address Switch................................................................................................................2-5
Inserting and Removing Plug-Ins .................................................................................................2-6
Installing the 1260-100X into a VXIbus Chassis ..........................................................................2-6
Making Connections to Plug-Ins (Pin-Outs) .................................................................................2-7
Power-On Self-Test......................................................................................................................2-7
Closing and Opening Relays........................................................................................................2-8
1260-100X User Manual
ii
Chapter 3...............................................................................................................................3-1
USING THE 1260-100X...................................................................................................................3-1
Checking the Response of Plug-Ins.............................................................................................3-1
Closing and Opening Relays........................................................................................................3-1
Emergency Reset Function..........................................................................................................3-2
Description................................................................................................................................3-2
Interface....................................................................................................................................3-2
Normal State of External Switch...............................................................................................3-2
Scope of Reset Event...............................................................................................................3-3
Sequence of Events .................................................................................................................3-3
Chapter 4...............................................................................................................................4-1
MAINTENANCE...............................................................................................................................4-1
Performance Verification Procedure ............................................................................................4-1
Purpose ....................................................................................................................................4-1
Equipment Required.................................................................................................................4-1
Software Required....................................................................................................................4-2
Procedure.................................................................................................................................4-2
Chapter 5...............................................................................................................................5-1
PRODUCT SUPPORT.....................................................................................................................5-1
Product Support ...........................................................................................................................5-1
Warranty.......................................................................................................................................5-1
Support Offices.............................................................................................................................5-2
1260-100X User Manual
iii
Appendix A........................................................................................................................... A-1
1260-100X SPECIFICATIONS........................................................................................................ A-1
General........................................................................................................................................ A-1
Mechanical .............................................................................................................................. A-1
Electrical .................................................................................................................................. A-1
Power and Cooling .................................................................................................................. A-2
Environmental.......................................................................................................................... A-2
EMC and Safety....................................................................................................................... A-2
Reliability ................................................................................................................................. A-2
Appendix B........................................................................................................................... B-1
1260-100X THEORY OF OPERATION .......................................................................................... B-1
General........................................................................................................................................ B-1
VXI Interface PCB Subassembly................................................................................................. B-1
Power Filtering......................................................................................................................... B-1
Signal Connections.................................................................................................................. B-2
OPTION-01T/Passport PCB Subassembly ............................................................................. B-2
Backplane/ Mezzanine PCB Subassembly .......................................................................... B-2
Local Bus Signal Routing ........................................................................................................ B-3
Reset Circuit ............................................................................................................................ B-3
Interrupt ................................................................................................................................... B-3
Communications...................................................................................................................... B-4
PLD Programming Connector.................................................................................................. B-5
1260-100X User Manual
iv
Appendix C ..........................................................................................................................C-1
TROUBLESHOOTING....................................................................................................................C-1
Introduction..................................................................................................................................C-1
Troubleshooting Table.................................................................................................................C-1
Communications Troubleshooting...........................................................................................C-2
List of Figures
Figure 1-1, The 1260-100X Shown with Two X-series Adapt-a-Switch Cards Installed..................1-2
Figure 1-2, Ease-of-Use Features ...................................................................................................1-5
Figure 2-1, The Front View –Two Carriers Configuration. ...............................................................2-2
Figure 2-2, 1260-100X Configuration Switches ...............................................................................2-4
1260-100X User Manual
Racal Instruments © 2005
Overview 1-1
Chapter 1
OVERVIEW
ITEMS SHIPPED
WITH 1260-100X
The following items are shipped with each 1260-100X Adapt-a-
Switch Carrier:
ITEM
RACAL PART
NUMBER
QUANTITY
Retaining Plate, Top
457383-001
1
Retaining Plate,
Middle
457383-003
1
Retaining Plate,
Bottom
457383-002
1
Blanking plate
457403
1
Instruction manual
980914-100X
1
The following accessories may be ordered separately:
Blanking Plate
457403
1
EMI Shield
457376
1
Retaining Plate, Top
457383-001
1
Retaining Plate,
Middle
457383-003
1
Retaining Plate,
Bottom
457383-002
1
PART NUMBERS,
OPTIONS AND
ORDERING
INFORMATION
The 1260-100X can be ordered as an assembly or as individual
pieces as needed depending upon the application. The following
Part Numbers are available as standard products:
408007
Includes:
408034 1260-100X Enclosure Assembly
4408035-100X Shipping Kit
1260-100X User Manual
Overview 1-2
Racal Instruments © 2005
Figure 1-1, The 1260-100X Shown with Two X-series Adapt-a-Switch Cards Installed
FUNCTIONAL
DESCRIPTION
The 1260-100X Adapt-a-Switch Carrier (currently without Option
01T) is a two-slot VXI bus module that may contain up to six X-
series Adapt-a-Switch plug-in cards. The X-series Adapt-a-Switch
cards are 3.3” longer and 0.4” wider than the non X-series cards,
thus allowing higher density relay configurations.
You may leave some slots unoccupied for future expansion.
Each plug-in has its own unique configuration of relays or digital
testing capabilities, such as:
1260-100X User Manual
Racal Instruments © 2005
Overview 1-3
•
Switch, for connecting individual pairs of inputs together.
•
Multiplexer (MUX), for connecting one input to any one of
several outputs.
•
Matrix. A matrix has two groups of channels: a row and a
column. The matrix can connect any row channel to any
column channel. The matrix may make multiple independent
connections at the same time.
•
RF
switch,
for
switching
radio-frequency
signals
via
impedance-matched signal paths.
•
Digital I/O, for a wide range of digital test applications.
Features
Compatibility
with Racal
Instruments 1260-
Series
Functionally, it is identical to the 1260-100 but currently without an
Option-01T. It only accepts the X-series Adapt-a-Switch plug-in
cards. The 1260-100X Adapt-a-Switch Carrier is controlled by the
same Option-01T controller installed in another carrier in the same
group as conventional Racal Instruments 1260-Series switch
modules. A single Option-01T Switch Controller Module may
control up to twelve switch modules or plug-ins. This may include
a combination of 1260-Series and Adapt-a-Switch plug-ins.
See Chapter 2: Getting Started for instructions on combining
1260-Series modules and Adapt-a-Switch plug-ins in the same
VXIbus chassis.
1260-100X User Manual
Overview 1-4
Racal Instruments © 2005
Ease of Use
The 1260-100X includes several ease-of-use features (the 1260-
100X model is shown in Figure 1-2):
•
Plug-ins are easy to insert and extract from the front panel of
the 1260-100X, even while it is installed in a chassis.
•
For only the 1260-100X, module addresses, used to uniquely
identify each plug-in, are set by a single DIP switch that is
easily accessible from the back of the 1260-100X. See
Setting Configuration Switches in Chapter 2: Getting
Started for instructions.
•
The 1260-100X ejector handles also function as injectors,
making it easy to install the carrier into a VXIbus chassis.
Maintainability
Maintainability of the 1260-100X is enhanced by the fact that plug-
ins are removable from the front panel. This has the following
benefits:
•
The mean time to repair (MTTR) is minimized, since the carrier
does not need to be disassembled or even removed from the
VXIbus chassis.
•
The switch system may be easily upgraded in the field by
adding or changing plug-ins.
•
Sparing is less costly and more flexible, since individual plug-
ins may be spared rather than the entire 1260-100X module.
1260-100X User Manual
Racal Instruments © 2005
Overview 1-5
Figure 1-2, Ease-of-Use Features
Dynamically-
Configured
Analog Bus
The 1260-100X Adapt-a-Switch Carrier offers an analog bus to
interconnect two or more plug-ins. This allows the creation of
large multiplexers and matrices. The analog bus may be
dynamically configured under software control.
For instructions in using the analog bus, refer to the instruction
manuals for the individual multiplexer and matrix plug-ins.
1260-100X User Manual
Overview 1-6
Racal Instruments © 2005
This page was left intentionally blank.
1260-100X User Manual
Racal Instruments © 2005
Getting Started 2-1
Chapter 2
GETTING STARTED
Unpacking and
Inspection
1. Before unpacking the switching module, check the
exterior of the shipping carton for any signs of damage.
All irregularities should be noted on the shipping bill and
reported.
2. Remove the instrument from its carton, preserving the
factory packaging as much as possible.
3. Inspect the switching module for any defects or damage.
Immediately notify the carrier if any damage is apparent.
4. Have a qualified person check the instrument for safety
before use.
Reshipment
Instructions
1. Use the original packing material when returning the
switching module to Racal Instruments for servicing. The
original shipping carton and the instrument’s plastic foam
will provide the necessary support for safe reshipment.
2. If the original packing material is unavailable, wrap the
switching module in an ESD Shielding bag and use plastic
spray foam to surround and protect the instrument.
3. Reship in either the original or a new shipping carton.
1260-100X User Manual
Getting Started 2-2
Racal Instruments © 2005
1260-100 1260-100X
w/Option-01T
Figure 2-1, The Front View –Two Carriers Configuration.
Option-01T
Presently, the 1260-100X can be used only as a compatible 1260-
type carrier adjacent to a 1260-100 module with an Option 01-T
card installed, or as one of the several 1260-Series modules in
consecutive slots.
The following rules apply when using the Option-01T controller:
1. Each Adapt-a-Switch or full-size VXI switch module has
its own unique module address.
2. Model 1260-100 assigns module addresses 1 through
12, to its module slot, depending on the position of the
Address Range DIP switch.
3. To be controlled by a single Option-01T, a group of
switch modules and carriers must be installed in a
contiguous group of VXIbus chassis slots and the
Option-01T must be installed in the left-most module in
the group.
1260-100X User Manual
Racal Instruments © 2005
Getting Started 2-3
Below are some examples of possible configurations:
Example 1:
A single Option-01T may control all Adapt-a-Switch
plug-ins in one or two 1260-100X Carriers. When it
controls two carriers, the carriers must be in adjacent
slots in the VXIbus chassis, with the Option-01T-
equipped carrier in the left-most slot of the group.
A single Option-01T may control all Adapt-a-Switch
plug-ins in a 1260-101, a 1260-100, or a 1260-100X
Carriers. When it controls two carriers, the carriers must
be in adjacent slots in the VXIbus chassis, with the
Option-01T-equipped carrier in the left-most slot of the
group. A single option_01T will not control more than
one 1260-101.
Example 2:
A single Option-01T may control a group of 1260-Series
switch modules. The modules must be installed in a
contiguous group of slots in the VXIbus chassis. Figure
2-1 shows the configuration which the left most 1260-
100 carrier is with Option 01-T installed and one 1260-
100X carrier.
Example 3:
A single Option-01T may control a mixed group of
1260-Series switch modules and Adapt-a-Switch plug-
ins. The maximum number of items in each group is
twelve. Each 1260-100X carrier counts as six plug-ins
and each 1260-101 carrier counts as two plug-ins,
regardless of the quantity of plug-ins actually installed.
For instance, a single Option-01T may control the
following group:
•
Slot 3:1260-54 RF Switch (conventional one-slot
VXI module), equipped with Option-01T
•
Slots 4 and 5:1260-100 Adapt-a-Switch carrier,
containing three 1260-138 plug-ins and two
1260-150 RF switch plug-ins.
•
Slot 6:1260-22 Power Switch (conventional one-
slot VXI module)
•
Slot 7:1260-22 Power Switch
1260-100X User Manual
Getting Started 2-4
Racal Instruments © 2005
•
Slot
8:
1260-101
Adapt-a-Switch
carrier,
containing one 1260-145.
The above group counts as eleven items: the 1260-54 and
1260-22 modules count as one item each, for a total of
three. The 1260-100X carrier is considered as six items,
regardless of the number of plug-ins installed. The 1260-
101 carrier is considered as two items. Adding these
together, the total number of controlled items is eleven.
This means that there is still room to add one 1260-Series
switch modules to the group, with the entire group
controlled by one Option-01T. The Option-01T must be
installed in the 1260-54 module in slot 3. In the case where
an Option-01T controls plug-ins residing in two adjacent
carriers, the Option-01T must be installed in the carrier on
the left.
For groups that are larger than twelve, additional
Option-01T modules are used.
Setting
Configuration
Switches
Model 1260-100X has its own configuration switches located
on the back panel (See Figure 2-2).
Figure 2-2, 1260-100X Configuration Switches
1260-100X User Manual
Racal Instruments © 2005
Getting Started 2-5
Set these switches as follows:
Local Bus Daisy-
Chain Switches
Currently, the 1260-100X requires another carrier with an Option-
01T installed in the adjacent left slot.
If you are using a 1260-100X Carrier as the last carrier in the VXI
chassis with no switch modules or carriers to its right, set all
twelve local bus daisy-chain switches to the OPEN position. On
the other hand, if you install a 1260-Series switch module or
another 1260-100X Carrier to the immediate right, set all twelve
local bus daisy-chain switches (S1 positions 1-6 and S2 positions
1-6) to the CLOSE position. This allows the Option-01T in the
carrier of the adjacent left slot to control not only the plug-ins in
that carrier, but the switch modules installed to the right, as well.
Module Address
Switch
Although plug-ins and 1260-Series modules do not require logical
addresses (the Option-01T has one instead), they each have a
module address. The Option-01T uses module addresses to
distinguish between plug-ins, and between 1260-Series modules.
Each plug-in or module must have a unique module address from
one through twelve, inclusive. Models 1260-100X assigns
module addresses as follows:
1260-100X Module Address
The module address DIP switch S2 postion 7 (see figure 2-2)
sets the logical addresses for all plug-ins in the 1260-100X
Carrier. The switch has two settings:
•
1-6 (opened) When the switch is set to this position, the
module addresses of the plug-ins are from
1 through 6, with 1 being in the left slot of
the top row. The plug-ins are addressed in
the following pattern:
•
7-12 (closed) When the switch is set to this position, the
module addresses of the plug-ins are from
7 through 12, in the following pattern:
1260-100X User Manual
Getting Started 2-6
Racal Instruments © 2005
Inserting and
Removing Plug-
Ins
CAUTION
If the 1260-100X Carrier is in a VXIbus chassis, make
sure the chassis power is off before inserting or
removing plug-ins.
Use proper anti-static handling procedures when plug-
ins are outside the carrier.
Figure 1-2 shows how to insert or remove a plug-in. Prior to
inserting or removing a plug-in, the retaining plates have to be
removed. For examples, to insert or remove a plug-in in slots 1,
2, or 3, the top and middle retaining plates have to be removed.
To insert a plug-in in slots 4, 5, and 6, the middle and bottom
retaining plates have to be removed. To insert a plug-in, position
the module so that the edges align with the grooves in the carrier
slot. Slide the plug-in all the way into the carrier, then press
firmly to engage its connector with the carrier backplane.
Replace the plug-in retaining plates and tighten the screws
securely.
To remove a plug-in, remove the retaining plates as stated
previously to make them float freely. Since they are captive, they
do not come completely out. Pull firmly and evenly on the front
panel to disengage the plug-in from the backplane, then pull the
plug-in out of the carrier.
INSTALLING THE
1260-100X INTO A
VXIBUS CHASSIS
To install the 1260-100X Carrier into a VXIbus chassis, position the
carrier so that the ridges at the top and bottom align with the
grooves in the chassis slots (see Figure 1-3). As you push the
carrier into the chassis, ensure that the injector/ejector handles are
in the proper positions. The upper handle should be all the way
up, and the lower handle should be all the way down. Push the
carrier into the chassis until it stops. Then use the injector/ejector
handles to engage the carrier with the chassis backplane.
Simultaneously push the upper handle downward and the lower
handle upward until the carrier is fully inserted. Tighten the upper
and lower retaining screws.
1260-100X User Manual
Racal Instruments © 2005
Getting Started 2-7
To remove the 1260-100X Carrier from the VXIbus chassis,
loosen the upper and lower retaining screws. Simultaneously
push the upper injector handle upward and pull the lower handle
downward until the carrier is disengaged from the chassis
backplane. Then slide the carrier out of the chassis.
Making
Connections to
Plug-Ins (Pin-
Outs)
For information about connecting external equipment to the plug-
ins, refer to the individual manuals for the plug-ins. These
include information regarding mating connector part numbers,
pin-outs, and electrical ratings.
Power-On Self-
Test
When the VXIbus chassis is switched on, the Option-01T in the
system automatically runs a self-test and the Fail LED in a carrier
with an Option-01T will illuminate.
When the Option-01T in the system completes self-test, the Fail
LED will extinguish. This occurs within five seconds after chassis
power is turned on.
The self-test includes the following:
1. PROM checksum
2. RAM write/read pattern verification
3. Microprocessor instruction test
4. Non-volatile RAM verification
5. Timer test
This self-test represents 70% fault detection.
1260-100X User Manual
Getting Started 2-8
Racal Instruments © 2005
Closing and
Opening Relays
For general commands to close and open relays, refer to the
section Closing a Relay in Chapter 1: Getting Started of the
Option-01T Instruction Manual. Other commands are specific to
individual plug-ins. For detailed information on module-specific
commands, refer to the instruction manual for the appropriate
plug-in.
1260-100X User Manual
Racal Instruments © 2005
Using the 1260-100X 3-1
Chapter 3
USING THE 1260-100X
Checking the
Response of
Plug-Ins
To verify that the plug-ins are operational, send the MOD:LIST?
command to the Option-01T. Then read back the response of
the Option-01T. The Option-01T responds with a message
specifying which plug-ins and 1260-Series modules are under its
control. If the Option-01T does not respond, or does not properly
list the installed switch products, refer to Appendix C:
Troubleshooting.
Closing and
Opening Relays
To use the message-based mode of communication, refer to
Chapter 2 of the Option-01T Instruction Manual for general
commands. For module-specific commands, refer to the
individual manuals for the plug-ins.
For high-speed operation, use the register-based mode with
Racal
Instruments
VXIplug&play
drivers.
For
detailed
information, refer to the VISA and the VXIplug&play Software
section of Chapter 2: Using the Option-01T in the Option-01T
Instruction Manual.
1260-100X User Manual
Using the 1260-100X 3-2
Racal Instruments © 2005
Emergency Reset
Function
Description
To enhance safety in applications that involve power switching,
certain Adapt-a-Switch plug-ins and conventional 1260-Series
modules offer an emergency reset function, activated by an
external input. This function instantly opens all relays in
response to an external event. This is a hardware operation that
does not require intervention on the part of the Option-01T or
application software.
When using the emergency reset feature, exercise due caution if
any undesirable effects may result from opening all channels.
Interface
Select Adapt-a-Switch plug-ins and 1260-Series modules have a
pair of front-panel-mounted connector pins. You may connect
these to an external device such as a mechanical switch. When
the switch is opened, all relays on the plug-in or module are
immediately opened.
An example of a plug-in that provides the emergency switch
interface is the 1260-120 Power Switch. All other plug-ins and
1260-Series modules that provide the switch interface operate in
the same manner as the 1260-120.
Normal State of
External Switch
If desired, a jumper setting on the plug-ins allows it to
accommodate a normally-closed external switch. Normally-open
is the default. Refer to the individual plug-in manual.
1260-100X User Manual
Racal Instruments © 2005
Using the 1260-100X 3-3
Scope of Reset
Event
The 1260-120 design allows the scope of an emergency reset to
be configured. In the default configuration, the 1260-120 opens
only its own relays in response to an external reset event. If
desired, a jumper setting on the 1260-120 may be changed so
that the effect of an emergency reset is global. With the global
reset configuration, a reset signal propagates through the VXI
Local Bus to open all relays on all switch products that are
connected to the same Option-01T.
All Racal 1260-Series and Adapt-a-Switch plug-ins respond to an
emergency reset condition when the scope is set to global.
Certain plug-ins and 1260-Series modules provide the front-panel
contacts for the external switch, and the circuitry to recognize a
reset event. A switch system may make full use of the
emergency reset feature if it includes at least one plug-in or
1260-Series module that has reset recognition capability.
Sequence of
Events
When a reset event occurs, the following sequence takes place:
1. The plug-in that is connected to the external switch
recognizes the reset event and sets a logic latch.
2. All relays are opened on the switch product that recognized
the reset event. If the scope of the reset function is set to
global, then all relays on all switch products connected to
the same Option-01T are also opened. This happens
independently of software.
3. An interrupt is generated to notify the Option-01T that a
reset event has occurred. This informs the Option-01T that
the relays have opened, so that it remains up-to-date on the
relay status.
4. The relays remain open and do not respond to any
commands from the Option-01T.
5. After the reset condition is removed (the external switch is
returned to its normal position), the relays remain open until
otherwise commanded by the Option-01T.
Other Adapt-a-Switch plug-ins and 1260-Series modules that are
equipped to recognize an emergency reset are configured and
operate the same way as the 1260-120. It is important to note
that a global reset condition opens relays on all plug-ins and
1260-Series modules that are connected to the same
Option-01T, regardless of whether they are equipped with the
interface for the external reset switch.
1260-100X User Manual
Using the 1260-100X 3-4
Racal Instruments © 2005
This page was left intentionally blank.
1260-100X User Manual
Racal Instruments © 2005
Maintenance 4-1
Chapter 4
MAINTENANCE
Performance
Verification
Procedure
Purpose
The following procedure verifies the basic functions of the
1260-100X Adapt-a-Switch Carrier. Test coverage includes:
•
Option-01T power-on self-test
•
Option-01T basic operation in message based mode
•
Option-01T operation in register-based mode
•
1260-100X receive and transmit functions.
Equipment
Required
•
VXIbus chassis
•
Embedded slot 0 controller, or MXI slot 0 controller with
IBM-compatible PC
•
A carrier with an Option-01T installed.
1260-100X User Manual
Maintenance 4-2
Racal Instruments © 2005
Software
Required
•
Microsoft Windows 95, 98, ME, NT, 2000, and XP
•
VXI Resource Manager (RESMAN) by National
Instruments
•
VXI Interactive Control (VIC) by National Instruments
Procedure
1. Ensure that an Option-01T is installed in the
carrier adjacent to the 1260-100X carrier as
outlined in the Option-01T section of Chapter 2:
Getting Started.
2. Set the Option-01T logical address switch to 3.
3. Set the 1260-100X module address switch to the
1-6 position.
4. Install the carrier into an empty pair of slots in a
VXIbus chassis. Remove all other modules from
the chassis except the slot 0 controller and the
carrier with an Option-01T.
5. Insert one or more plug-ins into the 1260-100X
Carrier.
6. Turn the chassis power on. Verify that the Fail
LED on the carrier with an Option 01T illuminates,
then extinguishes within five seconds.
7. Run the National Instruments Resource Manager
program (RESMAN). From the RESMAN output
log file, verify the following:
8. The Option-01T is recognized as DEVICE_3.
9. The self-test for DEVICE_3 passed.
10. 16K-bytes
of
memory
are
allocated
for
DEVICE_3.
11. Start the VXI Interactive Control (VIC) program
(by National Instruments).
1260-100X User Manual
Racal Instruments © 2005
Maintenance 4-3
12. Send
a
*IDN?
command
to
DEVICE_3
(Option-01T).
13. Read the response from the Option-01T. Verify
that the response is of the form:
14. Racal
Instruments,1260A
Option-01T,,
15. where is the serial number, and
is the revision level of the firmware.
16. Click on the Bus Access feature of VIC. Select
the WRITE mode.
17. Calculate the address offset of one of the Adapt-
a-Switch plug-ins. To do this, begin with the
address reported by RESMAN as being the
starting address for DEVICE_3. Add 1. Then
add 40016 times the module address of the plug-
in. Enter this address into the ADDRESS field in
VIC.
18. Select the A24 address mode.
19. In the data field, enter 5516. Click on GO. If the
plug-in is a relay type (not a digital type), you
should hear the relays close.
20. Select the READ mode, set the number of bytes
to at least 1, and click on GO. Verify that the
Option-01T responds with AA16, which is the
one’s compliment of the data written.
21. In the data field, enter AA16. Click on GO. If the
plug-in is a relay type (not a digital type), you
should hear the relays open.
22. Select the READ mode, set the number of bytes
to at least 1, and click on GO. Verify that the
Option-01T responds with 5516, which is the one’s
compliment of the data written.
23. Repeat steps 12 through 17 for each of the
remaining plug-ins.
24. Turn the chassis power switch off.
25. If fewer plug-ins are available than the number of
1260-100X User Manual
Maintenance 4-4
Racal Instruments © 2005
slots, repeat steps 5 through 18 with the plug-ins
in different slots until all carrier slots have been
tested.
26. For the 1260-100X set the module address switch
on the carrier to the 7-12 position. Repeat steps
5 through 20. When calculating the address
offset in step 12, remember that the module
addresses now range from 7 through 12.
1260-100X User Manual
Racal Instruments © 2005
Product Support 5-1
Chapter 5
PRODUCT SUPPORT
PRODUCT
SUPPORT
Racal
Instruments
has a complete Service and Parts
Department. If you need technical assistance or should it be
necessary to return your product for repair or calibration, call
1-800-722-3262. If parts are required to repair the product at
your facility, call 1-949-859-8999 and ask for the Parts
Department.
When sending your instrument in for repair, complete the form in
the back of this manual.
For worldwide support and the office closes to your facility, refer
to the Support Offices section on the following page.
WARRANTY
Use the original packing material when returning the 1260-100X
to Racal Instruments for calibration or servicing. The original
shipping container and associated packaging material will
provide the necessary protection for safe reshipment.
If the original packing material is unavailable, contact Racal
Instruments Customer Service for information.
1260-100X User Manual
Product Support 5-2
Racal Instruments © 2005
SUPPORT
OFFICES
RACAL INSTRUMENTS
United States
(Corporate Headquarters and Service Center)
4 Goodyear Street, Irvine, CA 92618
Tel: (800) 722-2528, (949) 859-8999; Fax: (949) 859-7139
5730 Northwest Parkway Suite 700, San Antonio, TX 78249
Tel: (210) 699-6799; Fax: (210) 699-8857
Europe
(European Headquarters and Service Center)
18 Avenue Dutartre, 78150 LeChesnay, France
Tel: +33 (0)1 39 23 22 22; Fax: +33 (0)1 39 23 22 25
29-31 Cobham Road, Wimborne, Dorset BH21 7PF, United Kingdom
Tel: +44 (0) 1202 872800; Fax: +44 (0) 1202 870810
Via Milazzo 25, 20092 Cinisello B, Milan, Italy
Tel: +39 (0)2 6123 901; Fax: +39 (0)2 6129 3606
Racal Instruments Group Limited, Technologie Park,
D-51429 Bergisch Gladbach, Germany
Tel: +49 2204 844205; Fax: +49 2204 844219
1260-100X User Manual
Racal Instruments © 2005
Product Support 5-3
Repair and Calibration Request Form
To allow us to better understand your repair requests, we suggest you use the following outline
when calling and include a copy with your instrument to be sent to the Racal Repair Facility.
Model
Serial No.
Date
Company Name
Purchase Order #
Billing Address
City
State/Province Zip/Postal Code
Country
Shipping Address
City
State/Province
Zip/Postal Code
Country
Technical Contact
Phone Number ( )
Purchasing Contact
Phone Number ( )
1. Describe, in detail, the problem and symptoms you are having. Please include all set up details,
such as input/output levels, frequencies, waveform details, etc.
2. If problem is occurring when unit is in remote, please list the program strings used and the controller
type.
3. Please give any additional information you feel would be beneficial in facilitating a faster repair
time (i.e., modifications, etc.)
4. Is calibration data required? Yes No (please circle one)
Call before shipping
Ship instruments to nearest support office.
Note: We do not accept
“collect” shipments.
1260-100X User Manual
Product Support 5-4
Racal Instruments © 2005
This page was left intentionally blank.
1260-100X User Manual
Racal Instruments © 2005
1260-100X Specifications A-1
Appendix A
1260-100X SPECIFICATIONS
GENERAL
1260 Series Compatibility Needs to be controlled by an Option
01T installed in another carrier
Annunciators
NONE
Host Interface
VXIbus backplane
Control Type
Message-based.
Register based: VXIbus A24 address
space.
Switching Response Time*
Register-based:
9 μs max.
Message-based:
10 ms typ.
VXI Plug&Play
Compatible drivers for all
1260-Series switching modules
Mechanical
Size
VXIbus C-size, two-slot module
Module Capacity
Six Adapt-a-Switch plug-ins
Weight (empty carrier)
Weight: 4.3 lbs (2.0 kg)
Front-Panel Connectors
Provided by each plug-in switch card
Electrical
Indicators
NONE
RF Shielding
Fixed steel shields between plug-in
slots
Analog Bus
Four two-wire channels
50-ohm impedance differential Pair
2A current capacity
1260-100X User Manual
1260-100X Specifications A-2
Racal Instruments © 2005
Power and
Cooling
Power Consumption
+5V:
1.0A
(without plug-ins)
+12V: 0A
+24V: 0A
Power Available
+5V:
2.5A
(per plug-in)
+12V: 1.0A
+24V: 1.0A
Cooling (worst-case
Airflow:
3.0 liters/sec
plug-in configuration)
Backpressure:
0.7 mm H20
Environmental
Temperature
Operating:
0°C to +55°C
Non-operating:
-40°C to +70°C
Relative Humidity
95% RH non-condensing
Altitude
Operating:
10,000 ft.
Non-operating:
15,000 ft.
Vibration
0.013 in. P-P, 5 Hz to 55 Hz; meets
MIL-T-28800C Type III, Class 5,
Style F
Shock
30g, 11 ms, 1/2 sine wave
Bench Handling
4 inch drop at 45°
EMC and Safety
Emissions
EN61326: 1997+A1:1998, Class A
Susceptibility, Safety
EN61010-1:1993+A2:1995
Reliability
Mean Time Between
646,921hours, MIL-HBK-217,
Failures (MTBF)
ground-benign, 25oC
Mean Time To Repair
Replace plug-in: 5 min.
(MTTR)
Replace other:
<30 min.
*Measured from start of VXIbus cycle until relay coil is fully energized.
1260-100X User Manual
Racal Instruments © 2005
1260-100X Theory of Operation B-1
Appendix B
1260-100X THEORY OF OPERATION
GENERAL
The 1260-100X consists of three PCB subassemblies: VXI
Interface, Option 01T/Passport, and Backplane/Mezzanine PCB
subassemblies.
VXI INTERFACE
PCB SUB-
ASSEMBLY
The VXI Interface PCB assembly serves two functions. It connects
+5V, +12V, and +24V power from the VXIbus backplane to the
1260-100X backplane. It also connects the VXIbus backplane
signals to the
Power Filtering
The power is filtered before reaching Option-01T/Passport
subassembly. The fuses, F1 through F3, are of the surface-mount
type. The fuses used are as follows.
FUSE
MANUFACTURER
MANUFACTURER’S
PART NUMBER
RATING
(AMPS)
RACAL PART
NUMBER
F1
Wickmann USA, Inc.
418 2100000
10
921545
F2
Wickmann USA, Inc.
417 1315000
3.15
921546
F3
Wickmann USA, Inc.
417 1315000
3.15
921546
1260-100X User Manual
1260-100X Theory of Operation B-2
Racal Instruments © 2005
Signal
Connections
Connectors P1, P2, P3, and P4 connect to the VXIbus backplane.
Connectors J8 and J9 pass the VXI bus signals and +5V, +12V
and +24V to the Option-01T/Passport PCB subassembly.
DIP switches S1 pins 1 through 6 and S2 pins 1 through 6 set the
local bus pass through. S2 pin 7 sets the module address of the
plug-ins. If S2 pin 7 is set to OFF position, the module addresses
of the plug-ins are set to operation from 1 through 6. If S2 pin 7 is
set to ON position, the module addresses of the plug-ins are set to
operate from 7 through 12. which connects to the VXI backplane.
The Bridge then connects to these signals using its P2 connector.
OPTION-
01T/Passport
PCB Sub-
assembly
Currently Option-01T controller circuit is not available for 1260-
100X carrier. So the 1260-100X is controlled by another carrier
with Option-01T installed.
Without the Option-01T controller circuit, the Option-01T/Passport
PCB subassembly only provides the pass through local bus
signals, +5V, +12, +24V power and Ground signals to the
Backplane/Mezzanine PCB subassembly.
Backplane/
Mezzanine PCB
Subassembly
Signals
The backplane/Mezzanine PCB subassembly provides bridging
functions between the plug-ins and the Option-01T/Passport PCB
subassembly. All communications signals are connected in
parallel to all six plug-ins, except LBOARDSEL1 through
LBOARDSEL6. These six signals route to their respective plug-in
slots, and identify which plug-in is currently addressed.
The backplane also provides an analog bus. This bus consists of
four pairs of PCB traces that daisy-chain from one plug-in to the
next, connecting to all six slots in parallel. Each pair of traces is
impedance-matched to 50 ohms, and is routed for maximum
isolation and minimum crosstalk.
1260-100X User Manual
Racal Instruments © 2005
1260-100X Theory of Operation B-3
Local Bus Signal
Routing
The Backplane/Mezzanine PCB subassembly provides the bridge
circuitry to interface between the plug-ins and the Option-01T. It
receives the local bus signals from the Option-01T on connector
P2A.
The local bus signals LBUSC00S through LBUSC09S are
differential. U4 and U5 converts LBUSC00S through LBUSC07S
to TTL levels. LBUSC08S and LBUSC09S are a bidirectional
differential pair. U5 and U8 receive and drive these signals,
respectively.
Reset Circuit
U1 is a voltage monitor used to reset the Bridge and plug-ins at
power turn-on. At power turn-on, it drives LLOGICRST low for
approximately 130ms. It then takes LLOGICRST high and leaves
it high, unless +5V power drops below 4.5V.
The signal LBUSC10S is an open-collector line from the
Option-01T. During power turn-on, the Option-01T drives
LBUSC10S low for about 100ms, then allows it to be pulled high.
While LBUSC10S is low, it resets all relay drivers on all plug-ins to
the inactive state, opening all relays. It also clears all outputs on
digital test plug-ins to a known state.
LBUSC10S also serves another function when a plug-in asserts a
global reset during an emergency reset event. When such an
event occurs, the plug-in drives LBUSC10S low. It leaves it low
until the event has passed and the Option-01T firmware clears the
reset latch in the plug-in.
LBUSC10S also connects to all other devices controlled by the
same Option-01T. When any device in this group asserts a global
reset, all devices in the group respond by resetting their drivers.
Interrupt
The signal LBUSC11S is an open-collector line to the Option-01T.
It has two purposes. When a plug-in asserts an emergency reset
condition, it drives the LBUSC11S signal low. Also, the 1260-114
drives the LBUSC11S signal low to announce a synchronous
event. When LBUSC11S goes low, the Option-01T polls the plug-
ins to determine which one generated the interrupt. It then
responds accordingly.
1260-100X User Manual
1260-100X Theory of Operation B-4
Racal Instruments © 2005
Communications
U3 is a programmable logic device (PLD) that handles the
communications between the plug-ins and the Option-01T. It
implements the same state machine used in 1260-Series switch
modules. communications proceed as follows.
When the Option-01T is about to write data to a plug-in, it takes
HLBUS_A/D to a logic 1, indicating that the data to be serially
shifted out will be an address. It also drives the HLBUS_R/W
signal low to indicate that data will be written. It then outputs an
8MHz clock on HLBUS_CLK, and places serial data on
HLBUS_DATAIN, synchronous with HLBUS_CLK. The first 13 bits
of data represent the address to be written to, with the most-
significant bit first. This address is relative to the A24 address
offset assigned to the Option-01T by RESMAN. For example,
address 0 references the first address in the space allocated to the
Option-01T.
The 13 address bits are clocked into a shift register internal to U3.
The parallel outputs of this shift register are connected directly to
all six plug-ins via the 1260-100 backplane.
After sending the address, The Option-01T drives HLBUS_A/D
low, indicating that the next data sent will be data. It then shifts the
serial data that will be written to the address just sent. There are
eight data bits, with the most-significant bit sent first.
After the last data bit has been sent, the signal HLBUS_STROBE
transitions to a logic high for one clock period of HLBUS_CLK.
HLBUS_STROBE is sent to the six plug-ins. The plug-in
containing the addressed device uses HLBUS_STROBE to latch
the data into the device.
To read data from a plug-in, the Option-01T begins by taking
HLBUS_A/D to a logic 1, indicating that the data to be serially
shifted out will be an address. It also drives the HLBUS_R/W
signal high to indicate that data will be read. It then outputs an
8.25 clock on HLBUS_CLK, and shifts the address out on
HLBUS_DATAIN, as it does when writing.
The 13 address bits are clocked into U5 and U6, as when writing.
After sending the address, the Option-01T drives the HLBUS_A/D
low, indicating that it is ready for data. The plug-in containing the
addressed
device
begins
shifting
serial
data
onto
HLBUS_DATAOUT. There are eight data bits, with the most-
significant bit sent first.
The signal HLBUS_STROBE is used only when writing, not when
reading. During reading, it remains at logic 0.
1260-100X User Manual
Racal Instruments © 2005
1260-100X Theory of Operation B-5
PLD
Programming
Connector
J9is a 10-pin header that is connected to the PLD (U3). It uses the
JTAG interface to allow in-system programming of the PLD during
upgrades or troubleshooting.
1260-100X User Manual
1260-100X Theory of Operation B-6
Racal Instruments © 2005
This page was left intentionally blank.
1260-100X User Manual
Racal Instruments © 2005
Troubleshooting C-1
Appendix C
TROUBLESHOOTING
INTRODUCTION
If you have difficulty in using the 1260-100X Adapt-a-Switch
Carrier or its plug-ins, refer to the following troubleshooting table.
If you are still unable to resolve the problem, contact one of the
Racal Customer Support offices listed in Chapter 5: Support
Offices.
TROUBLESHOOTING TABLE
PROBLEM
POSSIBLE CAUSES
WHAT TO DO
No response from plug-in via
VXIbus.
•
Option-01T not installed in
the carrier controlling the
1260-100X, or installed in
incorrect module.
•
Module addresses
conflicting.
•
Plug-in not fully seated.
•
Communication problem with
Option-01T, 1260-100X
Carrier, or plug-in.
•
Option-01T must be installed in
left-most carrier or 1260-Series
module in group. See
Option-01T in Chapter 2:
Getting Started.
•
Please note that currently 1260-
100X does not have an Option-
01T feature. It requires another
carrier with an Option-01T to
control it.
•
Ensure that each plug-in and
1260-Series module has its own
unique module address. See
Module Address Switch in
Chapter 2: Getting Started.
•
Ensure that plug-in is firmly
engaged.
•
See special troubleshooting
procedure below, entitled
“Communications
Troubleshooting”
1260-100X User Manual
Troubleshooting C-2
Racal Instruments © 2005
PROBLEM
POSSIBLE CAUSES
WHAT TO DO
Plug-in does not appear in
VXIplug&play soft panel list.
•
Soft panel driver not
installed.
•
Communication problem with
Option-01T, 1260-100X
Carrier, or plug-in.
•
Install 1260 Soft Front Panel
Driver.
•
See special troubleshooting
procedure below, entitled
“Communications
Troubleshooting”.
Switch modules to the right of
the 1260-100X does not appear
in VXIplug&play soft panel list
(1260-100X plug-ins do
appear).
•
Local Bus daisy-chain
switches not closed.
•
Close all twelve Local Bus
daisy-chain switches on
1260-100X Carrier when
another carrier or 1260-Series
module is installed to its
immediate right.
Relays do not close when
commanded, but plug-in
appears in VXIplug&play soft
panel list.
•
Emergency reset function
activated
•
If external switch is connected,
ensure that it is in the inactive
position. See the Emergency
Reset Function section in
Chapter 3.
Communications
Troubleshooting
Please note that temporarily 1260-100X does not have an Option-
01T feature. It requires another module with an Option-01T to the
left-most of the group.
Try the following procedure if you have trouble operating the plug-
ins. If you still have difficulty, contact Racal Customer Support
Chapter 5: Product Support.
1. Turn off the VXIbus chassis power switch. Wait at least two
seconds, then turn it back on. The Fail LED on the left-most
1260-serier
Carrier
with
the
controlling
Option-01T.
1260-Series module should illuminate, then go out within five
seconds. If it does, then proceed to step 2. If the Fail LED
does not illuminate when power is turned on, then go to step 6.
If the Fail LED illuminates, and stays on after five seconds,
then the Option-01T may be defective; contact Racal Customer
Support.
1260-100X User Manual
Racal Instruments © 2005
Troubleshooting C-3
2. Send a *IDN? command to the Option-01T, then read the
response. It should be in the following format:
Racal Instruments,1260A Option-01T,,
where is the serial number, and
is the revision level of the firmware. If the Option-01T
responds with a message in this form, then proceed to step 3.
Otherwise, go to step 8.
3. Execute the Racal Instruments 1260-Series soft front panel
program. It should display a list of all plug-ins and 1260-Series
modules that are connected to the Option-01T. If it does not
display a list, then go to step 8. Otherwise, proceed with step
4.
4. If the list displayed in step 3 is complete, then go to step 5.
Otherwise, go to step 10.
5. Since the list of plug-ins is complete, the Option-01T is
successfully communicating with all of the plug-ins. If you still
have other communication difficulties, contact Racal Customer
Support.
6. If this fails to correct the problem, then contact Racal Customer
Support.
7. Try to isolate the problem to a single 1260-100X Carrier or
1260-Series module by removing others from the chassis.
Remember the Option-01T is installed in the left-most module
in the group.
For problems with Option-01T, please refer to the user manual of
the controlling carrier with the Option-01T installed.