PDK PDR1000 Operation & User’s Manual

Specifications

Brand: PDK
Category: Measuring Instruments
Model: PDR1000 (Operation & User’s Manual)
File Type: PDF
File Size: —
Pages: 25

Read Online

Read the PDK PDR1000 Operation & User’s Manual online using the built-in PDF viewer.

Text of the manual

Table of Contents

1. General Introduction ……………………………………………………………………………………… 27 2. Abbreviations & Initials ………………………………………………………………………………… 28 3. Introduction ………………………………………………………………………………………………… 28 3.1 Specification …………………………………………………………………………………… 29 3.2 Front / Back panel …………………………………………………………………………… 30 4. Operation ………………………………………………………………………………………………… 32 4.1 Key Operation …………………………………………………………………………………… 32 4.1.1 Power On …………………………………………………………………………… 32 4.1.2 PEAK …………………………………………………………………………………… 32 4.1.3 UNIT …………………………………………………………………………………… 32 4.1.4 ZERO …………………………………………………………………………………… 32 4.1.5 Back Light …………………………………………………………………………… 33 4.2 Menu/Function Select ……………………………………………………………………… 33 4.2.1 Alarm …………………………………………………………………………………… 33 4.2.2 Pressure Switch Test ……………………………………………………………… 35 4.2.3 Calibration …………………………………………………………………………… 38 4.2.4 Baud Rate …………………………………………………………………………… 39 4.2.5 RS232 Communication …………………………………………………………… 40 4.2.6 Auto Power Off …………………………………………………………………… 42 4.2.7 Data Logger ………………………………………………………………………… 43 4.2.8 Storage Data Clear ……………………………………………………………… 45 4.2.9 End …………………………………………………………………………………… 45 5. Maintenance Adjustments and Calibration ……………………………………………………… 46 5.1 Zero Adjustment ……………………………………………………………………………… 46 5.2 Pressure Calibration …………………………………………………………………………… 46 5.3 Analog Output Calibration ………………………………………………………………… 46 6. Troubleshooting ……………………………………………………………………………………………… 47 7. Appendix …………………………………………………………………………………………………… 48 7.1. Warranty Statement ………………………………………………………………………… 48 7.2. Accessory ……………………………………………………………………………………… 48 7.3. Table of pressure unit conversion ……………………………………………………… 49

1. General Introduction

Manual Conventions This manual is intended to provide the user with the basic information necessary to operate a PDR1000 (Digital Pressure Gauge). It also includes a great deal of additional information provided to allow you to optimize PDR1000 use and take full advantage of its many features and functions. “Caution” is used in throughout the manual to identify conditions or actions that could cause harm to the PDR1000 or user of the PDR1000. “Note” is used throughout the manual to identify operating and applications advice and additional explanations. High pressure liquids and gases are potentially hazardous. Energy stored in these liquids and gases can be released unexpectedly and with extreme force. High pressure systems should be assembled and operated only by authorized personnel who have been instructed in proper safety practices. 27

2. Abbreviations & Initials

Units & Words Keys Units & Words Keys kPa kilopascal a absolute MPa megapascal g gauge hPa hectopascal PEAK peak mbar millibar Hi high bar bar Lo low psi pounds per square inch MAX maximum mmH2O millimeter of water MIN minimum mmHg millimeter of mercury Hold Hold inH2O inch of water mA Milliampere inHg inch of mercury Alarm atm atmosphere V Voltage kgf/cm2 kilogram force per square centimeter REM serial interface communication standard 3. Introduction The PDR1000 Series Reference Digital Pressure Gauges (the Product) are high accuracy digital pressure test gauges. PDR1000 has best performance that can measure precision pressure from -100 kPa to 500 MPa without a separate external power supply in anytime any place, including on-site and laboratory. Also PDR1000 equipped solid and stylish aluminum outer casing for durability. PDK produced PDR1000 with PDK’s unique pressure calibration and production technology that has best performance in its class. PDR1000 is digital pressure gauge for precise pressure measurement and test with Built-in various features. Simply use buttons on front panel to operate and select various functions and external power input section, analog output port (1 to 5 VDC), and communication port (RS232) and batteries change socket on rear panel. Combine pressure comparator or hand pump for using pressure calibrator in field and calibration lab for simple pressure measurement and calibration provides an excellent solution. 28

Over Pressure limit

Bust Pressure Pressure Unit Workable Temperature Range Storage Temperature Range Temperature Compensation Range RS232 Communication Power Requirement Media of Use Display Display Speed Analog Output Pressure Port Weight Size Data Logging -100 kPa to 150 MPa 0 to 250 MPa 0 to 500 MPa Gauge : ±0.025% F.S Absolute / Compound : ±0.05% F.S ±0.05% F.S ±0.1% F.S ** Included Nonlinearity, Hysteresis, Repeatability, errors for -10 ~ 50 ℃ Temperature range 150 % of full scale kPa, MPa, kg/cm2, psi, mbar, bar, inHg, inH2O, mmH2O, mmHg, atm -20 ~ 70 ℃ -30 ~ 80 ℃ -10 ~ 50 ℃ Provide Commands in manual, Available to use private communication cable (Option) AA Alkaline battery 3ea, Hours of battery use approximately 1000 hours External power supply(Option, 12~24 VDC), Gas & Liquid (Only Gas available below 100 kPa) 5 Digits, Backlight On/Off, Auto Off 3 times / Second (10 times automatic change if Peak function use) 1 - 5 VDC (Must use external power supply), 0.1% Accuracy (Option : 4-20mA , 0-5VDC , 0-10VDC) 1/4" NPT M, 1/4" PF (G), 1/4" PT (R), 9/16" UNF Cone Threaded (HF4/AF250C, 100 MPa / 150 MPa) 530 g Dia. 110 (W) mm x 38 (D) mm x 150 (H) mm Included Pressure Port 1 time / 1, 3, 5, 30, 60 second, Store Max 3000 data Refer to pressure type and range in catalog. 29

3.2 Front / Back Panel

3.2.1 On / Off Key PDR1000 Power On/Off key 3.2.2 PEAK/MENU Mode key 3.2.3 UNITS / Analog out / Down Key 3.2.4 ZERO / Up Key PEAK MENU UNITS ZERO Photo 1. PDR1000 Front / Rear Panels Photo Figure 1. Buttons Peak shows maximum and minimum pressure value. Menu makes selection and set for Alarm, Switch, CAL, Baud Rate, Auto Power Off, and Data Logger. Saved peak value will be clear when PDR1000 turns off. Change the pressure units and On/Off for Analog output. In configure mode, push Down key to move backward through the menus. (Analog output function is only available when external power supply used) Zeroing pressure offset value. Pressure Switch On/Off, Peak value will be clear when Peak mode. In configure mode, push Up key to move forward through the menus. 30

3.2.5 Back Light / Enter Key

Backlight On/Off Use Enter key to set up the Menu. [ D-Sub 9 Pin Map ] Figure 2. RS-232 Port Table 1. Pin outs 5 6 7 8 9 4 3 2 1 Pin No Description Note 1 GND 2 RX RS232 3 TX 4 Analog Out+ Analog Output 5 GND 6 Ext. Switch In 7 Alarm Out 8 Ext. Power + External Power 9 Ext. Power - 31

4.1 Key Operation

4.1.1 Power On : Push POWER key on front panel to power on. Screen at no pressurize or Zero set. 4.1.2 PEAK / MENU : Peak shows maximum and minimum pressurized value on PDR1000. Push PEAK/MENU key indicates Hi for maximum pressurized value. Push PEAK/MENU key again, indicates Lo for minimum pressurized value. Push PEAK/MENU key again, return to the current pressure value. 4.1.3 UNITS : This function can change engineering units on PDR1000. When push UNIT key engineering units will change as following steps. MPa → mbar → bar → inH2O → mmH2O → inHg → mmHg → kgf/cm² → atm → psi → kPa → MPa To set Analog Out function, push UNIT key for 2 seconds. (It only operates when external power supply connected) 4.1.4 ZERO : This function is zeroing the PDR1000. ZERO PEAK MENU UNITS For Gauge mode push ZERO key for 2 seconds under normal atmosphere condition. For the 32 psi % g 0

4.1.5 Back Light / Enter

: Push to turn on the backlight. Push again to turn off the backlight. 4.2 Menu/Function Select 4.2.1 Alarm Alarm function must need external power supply. (Figure 3 reference) Figure 3. Alarm Setup flowchart Absolute mode, make vacuum state (under 1 × 10-2 torr) then ZERO key for 2 seconds to zeros the display. Zero key can clear the Peak Hi Low values when Peak mode. Also it can select the set up when Pressure Switch mode. Push Enter key for 2 seconds to start or stop for Data Logger function. To set Data Logger function, turn on the Data Logger function on configuration menu. Push PEAK/MENU key for 2 seconds to go to setup and configuration menus. Push power key to go to End Menu or Push PEAK/MENU key to set other function settings after Menu function sets. Analog out, Alarm : These function must need external power supply to operate. 33 ZERO psi % g 0 psi % g 0 Alarm Lo psi % g 0 Alarm Hi psi % g 0 Alarm psi % g 0 Alarm psi % g 0 Alarm psi % g 0 Alarm % 0 psi g PEAK MENU

First screen of Alarm function page. Press UP key to turn Alarm On.

Alarm On screen on above. Press Enter key to save Alarm On status then Alarm Low value setup screen shows with blink. Alarm Low value setup screen on above. Digits on above screen is percentage value of PDR1000 Full Scale. Push ZERO key and UNITS key to change percentage value. After push Enter key to save Alarm Low value, Alarm Hi value setup screen shows with blink. Alarm Hi value setup screen on above. Digits on above screen is percentage value of PDR1000 Full Scale. Push ZERO key and UNITS key to change percentage value. After push Enter key to save Alarm Hi value, Electric Contact Alarm Signal setup screen shows with blinking. 34 psi % g 0 Alarm Lo psi % g 0 Alarm Hi psi % g 0 Alarm psi % g 0 Alarm

Electric Contact Alarm Signal setup screen on above. Push ZERO key to select Normal Close or

Normal Open status set. Push Enter key to save Electric Contact Alarm Signal setup. < Electric Contact Alarm Signal operation mode > 4.2.2 Pressure Switch Test Figure 4. Switch test Setup flowchart < Normal Open (N_O) > 5V Hi Lo 0V 0V 5V + Pressure - Pressure < Normal Close (N_C) > 5V 5V Hi Lo 0V 0V + Pressure - Pressure 35 psi % g 0 Alarm psi % g 0 Alarm psi % g 0 psi % g 0 psi % g 0 psi % g 0 % 0 psi g psi % g 0 PEAK MENU ZERO

Pressure Switch Test function is that receive and indicate Electric contact signal (On/Off) from

Pressure Switch. Push PEAK/MENU key for 2 seconds to go to Menu function then select Switch Menu screen. Push ZERO key to set Switch On or Off. Push Enter key to save Switch status then escape the Switch Menu. < Operation Mode > 1) Operation for External Pressure Switch Off to On Pressure Switch function “ON” screen on above. If PDR1000 received external electric contact signal (Switch turns Off to On) 2) Operation for External Pressure Switch On to Off Switch icon will change from open to close ( ) then pressure value will be hold status with “Hold” icon blinking. Even supply pressure changed, Pressure value will keep hold until push ZERO key. Pressure changes can check via percentage section on the bottom side of screen. This case is that external electric contact open to close – in other words switch status changed from Off to On. The other case is switch turns from On to Off. If it connect with Switch “On” status, it will be “1)” operation because initial status of Switch function is open icon as ( ). If push ZERO key, it will be same screen as on below screen. 36 MPa % g 0 50 MPa Hold % g 0 50

If PDR1000 received external electric contact signal (Switch turns On to Off), Switch icon will

change from close to open ( ) then pressure value will be hold status with “Hold” icon blinking. Even supply pressure changed, Pressure value will keep hold until push ZERO key. Pressure changes can be checked via percentage section on the bottom side of screen. Apply Switch function operation with Switch “On” status then connect External Pressure Switch to PDR1000. If pressure value reach the Pressure Switch set value, display pressure value will be hold then copy down the held pressure value. Push ZERO key to release the held pressure value then depressurize, Electronic contact on Pressure Switch will reverse and locked up the display value. Pressure Switch Hysteresis value is difference between when pressurize then electronic contact changed and when depressurize then electronic contact changed. < Normal Close type > PSV ON OFF Hys + Pressure - Pressure PSV : Pressure safety Valve Hys : Hysteresis Value 37 MPa % g 0 50 MPa Hold % g 0 50

Figure 5. Calibration flowchart

Calibration function can adjust Zero and Full Scale value. Zero Adjustment – Push Enter key for 2 seconds to set Zero value. If Zero set finished, display shows corrected Zero. Zero adjustment should be adjusted within ±10% of full scale. This function is Full Scale pressure value adjustment. Push Enter key for 2 seconds to change Full Scale value. After Full Scale adjustment finished, < Normal Open type > PSV ON OFF Hys + Pressure - Pressure PSV : Pressure safety Valve Hys : Hysteresis Value 38 psi CAL g 0 50 100 CAL psi CAL g 0 (Caution) Do not pressurize quickly during Pressure Switch Test. psi CAL g 0 50 100 CAL psi % g 0 psi CAL g 0 % 0 psi g PEAK MENU

4.2.4 Baud Rate

Figure 6. Baud Rate setup flowchart This function can set Baud Rate for RS232 Communication. display shows current adjusted pressure value. Full Scale adjustment should be adjusted within ±10% of full scale value and displayed full scale value adjust with Pressure controller/calibrator or Piston gauge. Push PEAK/MENU key for 2 seconds to select Menu function. Push again PEAK/MENU key to select Baud Rate function then push ZERO key to choose Baud Rate. 39 psi % g 0 REM % 0 psi g psi % g 0 psi % g 0 REM psi % g 0 REM PEAK MENU ZERO psi % g 0 REM psi % g 0 REM psi % g 0 REM psi % g 0 REM

4.2.5 RS232 Communication

1. Data Request Configuration 5 Byte 1 Byte 3 Byte 1 Byte 4~6 Byte 1 Byte ID BL C1 BL C2 LF 3. Data Indication 5 Byte 1 Byte 3 Byte 1 Byte 4 Byte 1 Byte ID BL C1 BL C2 CR CR [001] GET INFO [001] ID NO : [001] Model Name : Jain- 2 KD Serial_No. : 2020G01-0001 Sensor Range : -2.0 ~ 2.0 K{ETX} < Response Example > 2. Data Response Configuration 1 Byte 5 Byte 1 Byte 1 Byte 1 Byte STX ID D1 CR LF EOT ID will shows as [xxx], D1 has up to 25 Bytes of output data. Baud Rate : 19200 Data Bit Length : 8 bit Stop Bit : 1 bit Parity : None Data flow control : N/A Data Code : ASCII code BL : Blank C1 : Command1 LF : Line Feed CR : Carriage Return STX : Start of text D1 : Data EOT : End of text C2 : Command2

6. Data Logger function

5 Byte 1 Byte 3 Byte 1 Byte 6 Byte 1 Byte ID BL C1 BL C2 CR [001] GET LOGALL CR 4. Continuous sample Pressure value, Temperature value 5 Byte 1 Byte 3 Byte 1 Byte 6 Byte 1 Byte ID BL C1 BL C2 CR [001] GET MDAUTO [001] P = -0.00136, T = 24.95 [001] P = -0.00136, T = 24.95 [001] P = -0.00132, T = 24.95 [001] P = -0.00126, T = 24.95 < Response Example > CR 5. Sample one Pressure value, one Temperature value 5 Byte 1 Byte 3 Byte 1 Byte 6 Byte 1 Byte ID BL C1 BL C2 CR [001] GET MDMANU [001] GET MANU [001] P = -0.00136, T = 0.00, [001] GET MANU [001] P = -0.00122, T = 0.00, [001] GET MANU [001] P = -0.00115, T = 0.00, < Response Example > CR

4.2.6 Auto Power Off

Figure 7. Auto Power Off setup flowchart Push PEAK/MENU key for 2 seconds to select Menu function then push PEAK/MENU key to choose Auto Power Off function. Push ZERO key to set the necessary interval. Default Auto Power Off time set is 10 minutes. Auto Power Off function can set Auto Off. This function can use after Auto Power Off function on. Choose one of 5 min, 10 min, 20 min, and 30 minutes. 42 % 0 psi g psi % g 0 psi % g 0 PEAK MENU psi % g 0 psi % g 0 psi % g 0 psi % g 0 psi % g 0 ZERO

4.2.7 Data Logger

Figure 8. Data logger setup flowchart Push PEAK/MENU key for 2 seconds to select Menu function. Push PEAK/MENU key to select Data Logger function. Push ZERO key to set Data Logger function. Push ENTER key to save Data Logger status. Push ENTER key to go to next Menu when Data Logger off, push ENTER key to set Data Logger function save and time mode when Data Logger On. psi % g 0 psi % g 0 psi % g 0 psi % g 0 psi % g 0 psi % g 0 % 0 psi g psi % g 0 psi % g 0 psi % g 0 psi % g 0 psi % g 0 psi % g 0 psi % g 0 psi % g 0 ZERO

Push ZERO key to save Data Logger Data save time.

Select Data save time and push ENTER key to save then go to Data save page. Figure 9. Data storage address set up Long push ZERO key to select start page when Data Logger function data save. It can choose page 1 to page 30. Choose and select Data save page then push ENTER key to save. To set rest of Menu function, push PEAK/MENU key again to set other various functions. Data Logger function can save 100 sample data per page. Data saved value are average and max. Push ENTER key for 2 seconds to stop Data Logger function. Available to check the data via RS232 Communication. psi % g 0 psi % g 0 psi % g 0 Push PEAK/MENU key a few times to select End Menu or push POWER key to select End Menu. After Menu sets then use Logger function, push ENTER key to 2 seconds then display shows “Log” with blinking. At this moment, unable to use RS232 Communication because PDR1000 is saving pressure value at the same time. psi % g 0 psi % g 0 psi % g 0 psi % g 0 psi % g 0 psi % g 0

4.2.8 Storage Data Clear

Push PEAK/MENU key for 2 seconds to select Menu. Push PEAK/MENU key to select Data Clear function. Push ZERO to Data clear or not. This function is erase saved data on EPROM. 4.2.8 End This function is save all of Menu. This is last function to activate and store for each configured setting. Push ENTER key to save all of function. If user do not want to save functions of menu or change and back to menu, push ZERO key and select No at the End Menu then push ENTER key to back to Menu. 0 psi g 0 psi g % 0 psi g % 0 psi g

5. Maintenance Adjustments and Calibration

5.1 Zero Adjustment Using PDR1000 for a long time, it occurs Zero point change because of yearly variation for pressure sensor and PCB components or other environmental condition. If user wants to set Zero, use Calibration Zero function on CAL Function. 5.2 Pressure Calibration PDR1000 is reference pressure measuring gauge. It needs calibration by authorized organization every year if user wants to maintain PDR1000 guaranteed accuracy. Change the slope value with calibration function to changed pressure correction for PDR1000. 5.3 Analog Output Calibration D-SUB terminal on back side of PDR1000 has various analog output for 0 ~ Full Scale pressure value. If output became incorrect because of long time use, calibration output via RS232 communication program is possible. 46

PDR1000 is reference pressure gauge with high technology.

PDK does not recommend to user open the case to solve the PDR1000 problem. PDK does not guarantee when PDR1000 got damage with user open the case in PDR1000. PDK recommends contacting/send PDR1000 to PDK for suitable and stable troubleshooting. Also recommends using PDR1000 in normal stable condition specified by the manual. SYMPTOM PROBABLE CAUSE SOLUTION Will not power up Low Battery Change Battery Poor Power Board Contact to PDK Poor Main Board Contact to PDK Poor Display Board Contact to PDK Pressure value goes down Poor pressure tube Change tube, leak check Poor Sensor Contact to PDK Check Peak Poor pressure measure Peak Poor Main Board Contact to PDK Poor Sensor Board Contact to PDK Poor Sensor Contact to PDK Poor external equipment Key does not work Poor Membrane Switch Contact to PDK Poor Main Board Contact to PDK Poor Key Cable Contact to PDK Poor external equipment Contact to PDK LCD does not show Poor LCD Module Contact to PDK Poor LCD Cable Contact to PDK Poor external equipment Contact to PDK 47

7.1 Warranty Statement

PDR1000 warranty period of quality is 3 years. PDK guarantee all of case of failure symptom except user’s fault within 3 years. Even when warranty is expired, PDK will help our customer with minimum repair fee. PDK Service Center (e-mail : [email protected]) Address Contact PDK Co., Ltd. 10-6, Expo-ro 339beon-gil, Yuseong-gu, Daejeon, Korea 34122 Tel +82-42-862-6880 Fax +82-42-862-6881 7.2 Accessory User Manual 1 pcs 48

7.3 Table of pressure unit conversion

1 MPa = 1 000 kPa = 10 000 hPa 1 bar = 1 000 mbar 1 inHg = 25.4 mmHg (Torr) 1 inH2O = 2.54 cmH2O = 25.4 mmH2O 1 mmH2O of 4℃ = 1 ÷ 0.998 23 mmH2O of 20 ℃ = 1.001 771 mmH2O of 20 ℃ Density of water is 1.000 00 at 4℃, at 20℃ is 0.998 23. unit atm MPa hPa bar psi kgf/cm2 mmHg mmH2O atm 1 0.101 325 1 013.250 1.013 250 14.695 94 1.0332 27 760.000 0 10 332.27 MPa 9.869 239 1 10 000.00 10.000 00 145.037 7 10.197 16 7 500.622 101 971.6 hPa 0.000 987 0.000 100 1 0.0010 00 0.014 504 0.001 020 0.750 062 10.197 16 Bar 0.986 923 0.100 000 1 000.000 1 14.503 77 1.019 716 750.062 2 10 197.16 psi 0.068 046 0.006 895 68.947 57 0.068 948 1 0.070 307 51.714 97 703.069 6 kgf/cm2 0.967 841 0.098 067 980.665 0 0.980 665 14.223 34 1 735.559 7 10 000.00 mmHg 0.001 316 0.133 322 1.333 223 0.001 333 0.019 337 0.001 360 1 13.595 09 mmH2O 9.678 42E-05 9.806 65E-06 0.098 067 9.806 65E-05 0.001 422 0.000 100 51.714 97 1 49

Other Measuring Instruments Manuals