PCB Piezotronics 357E91 Manual
Text of the manual
The information contained in this document supersedes all similar information that
may be found elsewhere in this manual.
Service – Due to the sophisticated
nature of the sensors and associated
instrumentation
provided
by
PCB
Piezotronics, user servicing or repair is
not recommended and, if attempted,
may void the factory warranty. Routine
maintenance, such as the cleaning of
electrical connectors, housings, and
mounting surfaces with solutions and
techniques that will not harm the
physical material of construction, is
acceptable. Caution should be observed
to ensure that liquids are not permitted
to migrate into devices that are not
hermetically
sealed.
Such
devices
should only be wiped with a dampened
cloth and never submerged or have
liquids poured upon them.
Repair – In the event that equipment
becomes
damaged
or
ceases
to
operate, arrangements should be made
to
return
the
equipment
to
PCB
Piezotronics for repair. User servicing or
repair is not recommended and, if
attempted,
may
void
the
factory
warranty.
Calibration – Routine calibration of
sensors and associated instrumentation
is recommended as this helps build
confidence in measurement accuracy
and
acquired
data.
Equipment
calibration
cycles
are
typically
established by the users own quality
regimen. When in doubt about a
calibration cycle, a good “rule of thumb”
is to recalibrate on an annual basis. It is
also good practice to recalibrate after
exposure to any severe temperature
extreme,
shock,
load,
or
other
environmental influence, or prior to any
critical test.
PCB Piezotronics maintains an ISO-
9001 certified metrology laboratory and
offers calibration services, which are
accredited by A2LA to ISO/IEC 17025,
with full traceability to SI through
N.I.S.T. In addition to the normally
supplied calibration, special testing is
also available, such as: sensitivity at
elevated or cryogenic temperatures,
phase response, extended high or low
frequency response, extended range,
leak
testing,
hydrostatic
pressure
testing, and others. For information on
standard
recalibration
services
or
special testing, contact your local PCB
Piezotronics
distributor,
sales
representative,
or
factory
customer
service representative.
Returning
Equipment
–
Following
these procedures will ensure that your
returned materials are handled in the
most
expedient
manner.
Before
returning
any
equipment
to
PCB
Piezotronics,
contact
your
local
distributor,
sales
representative,
or
factory customer service representative
to obtain a Return Warranty, Service,
Repair, and Return Policies and
Instructions
Materials
Authorization
(RMA) Number. This RMA number
should be clearly marked on the outside
of all package(s) and on the packing
Service, Repair, and Return
Policies and Instructions
list(s) accompanying the shipment. A
detailed account of the nature of the
problem(s) being experienced with the
equipment should also be included
inside the package(s) containing any
returned materials.
A Purchase Order, included with the
returned materials, will expedite the
turn-around of serviced equipment. It is
recommended to include authorization
on the Purchase Order for PCB to
proceed with any repairs, as long as
they do not exceed 50% of the
replacement
cost
of
the
returned
item(s). PCB will provide a price
quotation
or
replacement
recommendation for any item whose
repair costs would exceed 50% of
replacement cost, or any item that is not
economically feasible to repair. For
routine
calibration
services,
the
Purchase
Order
should
include
authorization to proceed and return at
current pricing, which can be obtained
from
a
factory
customer
service
representative.
Contact Information – International
customers should direct all inquiries to
their local distributor or sales office. A
complete list of distributors and offices
can
be
found
at
www.pcb.com.
Customers within the United States may
contact their local sales representative
or
a
factory
customer
service
representative. A complete list of sales
representatives
can
be
found
at
www.pcb.com.
Toll-free
telephone
numbers for a factory customer service
representative,
in
the
division
responsible for this product, can be
found on the title page at the front of this
manual. Our ship to address and
general contact numbers are:
PCB Piezotronics, Inc.
3425 Walden Ave.
Depew, NY14043 USA
Toll-free: (800) 828-8840
24-hour SensorLineSM: (716) 684-0001
Website: www.pcb.com
E-mail: [email protected]
PCB工业监视和测量设备 - 中国RoHS2公布表
PCB Industrial Monitoring and Measuring Equipment - China RoHS 2 Disclosure Table
部件名称
有害物质
铅 (Pb)
汞
(Hg)
镉
(Cd)
六价铬 (Cr(VI))
多溴联苯 (PBB)
多溴二苯醚 (PBDE)
住房
O
O
O
O
O
O
PCB板
X
O
O
O
O
O
电气连接器
O
O
O
O
O
O
压电晶体
X
O
O
O
O
O
环氧
O
O
O
O
O
O
铁氟龙
O
O
O
O
O
O
电子
O
O
O
O
O
O
厚膜基板
O
O
X
O
O
O
电线
O
O
O
O
O
O
电缆
X
O
O
O
O
O
塑料
O
O
O
O
O
O
焊接
X
O
O
O
O
O
铜合金/黄铜
X
O
O
O
O
O
本表格依据 SJ/T 11364 的规定编制。
O: 表示该有害物质在该部件所有均质材料中的含量均在 GB/T 26572 规定的限量要求以下。
X: 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB/T 26572 规定的限量要求。
铅是欧洲RoHS指令2011/65/ EU附件三和附件四目前由于允许的豁免。
CHINA RoHS COMPLIANCE
DOCUMENT NUMBER: 21354
DOCUMENT REVISION: D
ECN: 46162
Component Name
Hazardous Substances
Lead
(Pb)
Mercury
(Hg)
Cadmium
(Cd)
Chromium VI
Compounds
(Cr(VI))
Polybrominated
Biphenyls
(PBB)
Polybrominated
Diphenyl
Ethers (PBDE)
Housing
O
O
O
O
O
O
PCB Board
X
O
O
O
O
O
Electrical
Connectors
O
O
O
O
O
O
Piezoelectric
Crystals
X
O
O
O
O
O
Epoxy
O
O
O
O
O
O
Teflon
O
O
O
O
O
O
Electronics
O
O
O
O
O
O
Thick Film
Substrate
O
O
X
O
O
O
Wires
O
O
O
O
O
O
Cables
X
O
O
O
O
O
Plastic
O
O
O
O
O
O
Solder
X
O
O
O
O
O
Copper Alloy/Brass
X
O
O
O
O
O
This table is prepared in accordance with the provisions of SJ/T 11364.
O: Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit
requirement of GB/T 26572.
X: Indicates that said hazardous substance contained in at least one of the homogeneous materials for this part is above
the limit requirement of GB/T 26572.
Lead is present due to allowed exemption in Annex III or Annex IV of the European RoHS Directive 2011/65/EU.
for use with
PIEZOELECTRIC CHARGE MODE ACCELEROMETERS
SPECIFICATION SHEET, INSTALLATION DRAWING
AND CALIBRATION INFORMATION ENCLOSED
PCB ASSUMES NO RESPONSIBILITY FOR DAMAGE CAUSED TO THIS PRODUCT AS A RESULT OF
PROCEDURES THAT ARE INCONSISTENT WITH THIS OPERATING GUIDE.
1.0 INTRODUCTION
Congratulations on the purchase of a quality PCB charge
mode accelerometer. In order to ensure the highest level
of performance for this product, it is imperative that you
properly familiarize yourself with the correct mounting
and installation techniques before attempting to operate
this device. If, after reading this manual, you have any
additional
questions concerning this sensor or its
application, feel free to call an Application Engineer at
716-684-0001 or the closest PCB representative.
2.0 CHARGE MODE ACCELEROMETERS
Ceramic
shear-structured
accelerometers
offer
high
performance for precision vibration measurements in
high-temperature environments.
The use of ceramic
sensing crystals, operating in the shear mode, reduces
erroneous output due to base strain, thermal transients,
and transverse motion.
Charge mode accelerometers output a strong, high-
impedance charge signal directly from their ceramic
sensing element.
They do not contain built-in signal
conditioning
electronics;
the
signal
is
conditioned
externally by either a laboratory-style charge amplifier
or in-line charge converter prior to being analyzed by a
readout or recording device.
The absence of built-in
electronics permits operation to elevated temperatures of
500 °F (260 °C) for most models.
These accelerometers are ideal for structural testing,
machine monitoring, and vehicular shock and vibration
measurement tasks where high temperatures preclude
the use of accelerometers with built-in microelectronics.
Enclosed is a Specification Sheet, which lists the
complete performance characteristics of the particular
accelerometer.
3.0 OPTIONAL FEATURES
In addition to these standard features, a variety of
options are also available. When listed before the model
number, the following prefix letters indicate that the
sensor
is
manufactured
or supplied
with optional
features: A option- adhesive mount; J option- electrical
ground isolation; M option- metric mounting thread; P
option- positive polarity output (for use with non-
inverting, signal conditioning amplifiers); and W option-
water-resistant cable connection.
4.0 INSTALLATION OVERVIEW
When choosing a mounting method, consider closely
both
the
advantages
and
disadvantages
of
each
technique.
Characteristics like location, ruggedness,
amplitude
range,
accessibility,
temperature,
and
portability are extremely critical.
However, the most
important and often overlooked consideration is the
effect the mounting technique has on the high-frequency
performance of the accelerometer.
Shown hereafter are six possible mounting techniques
and their effects on the performance of a typical
piezoelectric accelerometer.
(Note that not all of the
mounting methods may apply to your particular sensor.)
The mounting configurations and corresponding graph
demonstrate how the high-frequency response of the
accelerometer may be compromised as mass is added to
the system and/or the mounting stiffness is reduced.
Note: The low-frequency response is unaffected by the
mounting technique. Its roll-off behavior is dependent
upon the coupling type or time constant of the external
charge amplifying device.
This characteristic may be
fixed or adjustable, depending on the charge amplifier.
Please
refer
to
the
appropriate
charge
amplifier
specification sheet for information on low-frequency
characteristics.
Figure 1. Assorted Mounting Configurations and Their
Effects on High Frequency
4.1 STUD MOUNT
This mounting technique requires smooth, flat contact
surfaces for proper operation and is recommended for
permanent and/or secure installations. Stud mounting is
also recommended when testing at high frequencies.
Note: Do NOT attempt mounting on curved, rough, or
uneven surfaces, as the potential for misalignment and
limited contact surface may significantly reduce the
sensor’s upper operating frequency range.
Figure 2. Mounting Surface Preparation
STEP 1: First, prepare a smooth, flat mounting surface,
then drill and tap a mounting hole in the center of this
area as shown in Figure 2 and in accordance with the
Installation Drawing for the specific sensor that is
being mounted.
A
precision-machined
mounting
surface
with
a
minimum
finish
of
63
µin
(0.00016
mm)
is
recommended. (If it is not possible to properly prepare
the test structure mounting surface, consider adhesive
mounting as a possible alternative.) Inspect the area,
checking that there are no burrs or other foreign
particles interfering with the contact surface.
STEP 2: Wipe clean the mounting surface and spread
on a light film of grease, oil, or similar coupling fluid
prior to installation.
Figure 3. Mounting Surface Lubrication
Adding
a
coupling
fluid
improves
vibration
transmissibility by filling small voids in the mounting
surface and increasing the mounting stiffness. For semi-
permanent mounting, substitute epoxy or another type of
adhesive.
STEP 3: Screw the mounting stud into the base of the
accelerometer
and
hand-tighten.
Then,
screw
the
sensor/stud assembly into the prepared tapped hole and
tighten
to
the
recommended
mounting
torque
as
indicated on the Installation Drawing.
Note: It is important to use a torque wrench during this
step.
Under-torquing the sensor may not adequately
couple the device; over-torquing may result in stud
failure.
4.2 ADHESIVE MOUNT
Adhesive
mounting
is
often
used
for
temporary
installation or when the test object surface cannot be
adequately prepared for stud mounting. Adhesives like
hot glue and wax work well for temporary mounts; two-
part epoxies and quick-bonding gels provide a more
permanent mount.
Note: Adhesively mounted sensors often exhibit a
reduction in high-frequency range.
Generally, smooth
surfaces and stiff adhesives provide the best frequency
response.
METHOD 1 - Adhesive Mounting Base
This method involves attaching a base to the test
structure, then securing the sensor to the base.
This
allows for easy removal of the accelerometer.
Also,
since the bases are of a “hard-coated” aluminum design,
they provide electrical isolation to eliminate ground
loops and reduce electrical interference that may be
present on the surface of the test object.
STEP 1: Prepare a smooth, flat mounting surface.
A
minimum surface finish of 63 µin (0.00016 mm)
generally works best.
STEP 2: Stud-mount the sensor to the appropriate
adhesive mounting base according to the guidelines set
forth in STEPS 2 and 3 of the Stud Mount Procedure.
STEP 3: Place a small portion of adhesive on the
underside of the mounting base. Firmly press down on
the assembly to displace any extra adhesive remaining
under the base.
Figure 4. Mounting Base: Adhesive Installation
METHOD 2 - Direct Adhesive Mount
For restrictions of space or for convenience, most
sensors (with the exception of integral stud models) can
be adhesive-mounted directly to the test structure.
STEP 1: Prepare a smooth, flat mounting surface.
A
minimum surface finish of 63 µin (0.00016 mm)
generally works best.
STEP 2: Place a small portion of adhesive on the
underside of the sensor. Firmly press down on the top
of the assembly to displace any adhesive. Be aware that
excessive amounts of adhesive can make sensor removal
difficult.
Figure 5. Direct Adhesive Mounting
4.3 HANDHELD OR PROBE TIP MOUNT
This
method
is
NOT
recommended
for
most
applications.
It is generally used only for machinery
monitoring and other portable trending applications.
Both the accuracy and repeatability at low (<5 Hz) and
high frequency (>1 kHz) ranges are questionable.
5.0 CABLING
Care and attention to installation is essential, as the
reliability and accuracy of your system is no better than
that of the output cable. Charge mode sensors require
the use of low noise cables. Cables and connectors must
be kept clean and dry to maintain high insulation
resistance. In the event that the insulation resistance is
compromised, inspect, clean, and bake cables and
connectors to restore insulation resistance.
STEP 1: Ascertain that you have ordered the correct
cable type. Use only PCB Series 003 Low-Noise Cable
(or
equivalent)
with
charge-output
accelerometers.
Some integral-cable units use Series 030 Cable, which is
also low noise. Keep connections clean to maintain
insulation resistance and good low-frequency response.
STEP 2: Connect the cable to the accelerometer.
A
small amount of thread-locking compound placed on the
connector prior to attachment helps secure the cable
during testing.
In harsh environments, the connection
can be sealed with silicone rubber, O-rings, and flexible
heat-shrink tubing.
TYPE 1: Coaxial Connectors
First, plug the male connector of the cable into the
mating female sensor connector.
Then, holding the
sensor stationary, secure the connector in place by
tightening down the attached cable sleeve.
TYPE 2: Pigtail Connections
“Pigtail” connections are convenient for easy field repair
of cable connections. Simply solder the stripped cable
to the exposed pins on the sensor.
(Check the
Installation Drawing to determine signal and ground
In many cases, it is also helpful to protect the
solder joint with heat-shrink tubing or epoxy.
Note: If you do not have the experience or resources to
attach pigtail leads, consult PCB to discuss factory
attachment.
Damage caused by excessive heat is not
covered by warranty.
STEP 3: Route the cable to the signal conditioner,
making
certain
to
stress-relieve
the
sensor/cable
connection and minimize motion by clamping the cable
at regular intervals.
Certain precautions should be used to avoid physical
damage and minimize electrical noise.
For instance,
avoid routing cables near high-voltage wires.
Do not
route cables along floors or walkways where they may
be stepped on or become contaminated. Shielded cables
should have the shield grounded at one end only.
STEP 4: Finally, connect the remaining cable end to the
signal conditioner.
To dissipate charge that may have
accumulated in the cable, short the signal lead to the
ground lead prior to attachment.
6.0 POWERING
A conventional method for conditioning the high-
impedance signal generated by a charge output sensor is
to use a laboratory-style charge amplifier. The charge
amplifier converts the high-impedance charge signal
generated by the sensor to a low-impedance voltage
signal that can then be transmitted to a readout or a
recording device for analysis.
Before connecting the
low-noise cable from the accelerometer to the charge
amplifier, be sure to ground both the charge amplifier
and the cable.
This ensures that an excessive static
charge
that
may
have
accumulated
across
the
accelerometer or cable is harmlessly discharged. Failure
to observe this precaution can result in the destruction of
the input FET of certain amplifiers.
Once each element is connected, allow a few minutes for
the system to thermally stabilize.
Place the switch on
the charge amplifier in the OPERATE position and
proceed with the measurement.
It is often convenient to normalize the accelerometer and
charge amplifier system to a precise sensitivity, such as
10.0 or 100.0 mV/g for ease of data analysis. This is
accomplished
with
most
PCB
laboratory
charge
amplifiers and some miniature in-line units as well.
For fixed sensitivity in-line charge converters, such as
the PCB Series 422, the system sensitivity (mV/g) is
determined as the product of the charge amplifier
sensitivity (mV/pC) and the charge sensitivity of the
accelerometer (pC/g).
Note: When using charge-amplified systems, the noise
floor
of
the
system
is
dependent
on
the
input
capacitance to the charge amplifier.
Since the cable
adds to the capacitance and to minimize the noise
threshold,
keep
the
cable
length
between
the
accelerometer and the charge amplifier to a minimum.
Cable length does not affect the system sensitivity of
charge-amplified systems.
Since charge amplifier operation varies, please contact
the respective signal conditioner manufacturer or check
the product manual for additional information.
7.0 ACCELEROMETER CALIBRATION
Accelerometer calibration provides, with a definable
degree of accuracy, the necessary link between the
physical quantity being measured and the electrical
signal generated by the sensor. In addition, other useful
information
concerning
operational
limits,
physical
parameters, electrical characteristics, or environmental
influences may also be determined. Without this link,
analyzing data becomes a nearly impossible task. PCB
provides a calibration record that documents the exact
characteristics of each sensor. (The type and amount of
data varies depending on the sensor type, contractual
regulations, and other special requirements.)
Under
normal
operating
conditions,
piezoelectric
sensors
are
extremely stable, and
their calibrated
performance characteristics do not change over time.
However,
harsh
environments
or
other
unusual
conditions that cause the sensor to experience dynamic
phenomena outside of its specified operating range may
temporarily or permanently affect the sensor.
This
change manifests itself in a variety of ways, including a
shift of the sensor resonance due to a cracked crystal, or
a temporary loss of low-frequency measuring capability
due to a drop in insulation resistance.
For these reasons, it is recommended that a recalibration
cycle be established for each accelerometer.
This
schedule is unique and is based on a variety of factors,
such
as
extent
of
use,
environmental
conditions,
accuracy requirements, trend information obtained from
previous calibration records, contractual regulations,
frequency of “cross-checking” against other equipment,
manufacturer recommendation, and any risk associated
with incorrect readings. International standards, such as
ISO 10012-1, provide insight and suggested methods for
determining recalibration intervals for most measuring
equipment.
With the above information in mind and
under
“normal” circumstances, PCB
conservatively
suggests a 12- to 24-month recalibration cycle for most
piezoelectric accelerometers.
Note: It is good measurement practice to verify the
performance of each accelerometer with a Handheld
Shaker or other calibration device before and after each
measurement.
The PCB Model 394C06 Handheld
Shaker operates at a fixed frequency and known
amplitude (1.0 g) to provide a quick check of sensor
sensitivity.
7.1 SENSOR RECALIBRATION
Accelerometer
recalibration
services
are
typically
performed by PCB’s internal metrology laboratory.
(Other international and private laboratories are also
available.) The PCB laboratory is certified to ISO 9001,
accredited by A2LA to ISO 17025, complies with ISO
10012-1 (and former MIL-STD-45662A), and uses
equipment directly traceable to N.I.S.T. This assures an
accurate calibration of relevant specifications.
In addition, many companies choose to purchase the
equipment
necessary
to
perform
the
recalibration
procedure themselves. While this may result in both a
savings of time and money, it has also been attributed to
incorrect readings and costly errors.
Therefore, in an
effort to prevent the common mistakes associated with
customer-performed calibration, this document includes
a broad overview of the Back-to-Back Calibration
technique.
This technique provides a quick and easy
method
for
determining
the
sensitivity
of
a
test
accelerometer over a wide frequency range.
7.2 BACK-TO-BACK CALIBRATION THEORY
Back-to-Back Calibration is perhaps the most common
method for determining the sensitivity of piezoelectric
accelerometers.
This method relies on a simple
comparison to a previously calibrated accelerometer,
typically referred to as a reference standard.
Figure 9. Reference Standard Accelerometer
These
high-accuracy
devices,
which
are
directly
traceable to a recognized standards laboratory, are
designed for stability, as well as configured to accept a
test accelerometer. By mounting a test accelerometer to
the
reference
standard
and
then
connecting
this
combination to a suitable vibration source, it is possible
to vibrate both devices and compare the data as shown
in Figure 10. (Test set-ups may be automated and vary,
depending on the type and number of accelerometers
being calibrated.)
Figure 10. Typical Back-to-Back Calibration System
Because the acceleration is the same on both sensors,
the ratio of their outputs (VT/VR) must also be the ratio
of their sensitivities.
With the sensitivity of the
reference standard (SR) known, the exact sensitivity of
the test sensor (ST) is easily calculated by using the
following equation:
ST = SR (VT/VR)
By varying the frequency of the vibration, the sensor
may be calibrated over its entire operating frequency
range.
The
typical
response
of
an
unfiltered
accelerometer is shown in Figure 11.
Figure 11. Typical Test Accelerometer Response
7.3 PCB CALIBRATION PROCEDURE
Numerous precautions are taken at PCB to insure
accurate and repeatable results. This section provides a
brief overview of the primary areas of concern.
Since the Back-to-Back Calibration technique relies on
each sensor experiencing an identical acceleration level,
proper mounting of the test sensor to the reference
standard is imperative. Sensors with mounting holes are
attached directly to the reference standard with a stud
tightened to the recommended mounting torque. A
shouldered mounting stud is typically used to prevent
the stud from “bottoming out” in the hole.
Both mounting surfaces are precision-machined and
lapped to provide a smooth, flat interface according to
the manufacturer’s specification.
A thin layer of
silicone grease is placed between the mating surfaces to
fill
any
imperfections
and
increase
the
mounting
stiffness. The cables are stress-relieved by first routing
them to the shaker head, securing them with tape or
cable ties, then routing them to a nearby stationary
location. This reduces cable motion, which is especially
important when testing charge output sensors and helps
to prevent extraneous noise or stresses from being
imparted into the system. A typical set-up is shown in
Figure 12.
Figure 12. Typical Calibration Set-Up
Adhesively mounted sensors use similar practices.
However, in this case, a small portion of quick-bonding
gel or similar temporary adhesive is used to attach the
test sensor to a reference standard designed with a
smooth, flat mounting surface.
In addition to mounting, the selection of the proper
equipment is critical.
Some of the more important
considerations include: 1) the reference standard must
be
specified
and
previously
calibrated
over
the
frequency and/or amplitude range of interest; 2) the
shaker should be selected to provide minimal transverse
(lateral) motion and minimal distortion; and 3) the
quality of the meters, signal generator, and other devices
should be selected so as to operate within the limits of
permissible error.
7.4 COMMON MISTAKES
Most
calibration
errors
are
caused
by
simply
overlooking some of the fundamental principals of
dynamics. This section attempts to address some of the
more common concerns.
For stud-mount sensors, always mount the accelerometer
directly to the reference standard.
Ensure that the
mounting surfaces are smooth, flat, and free of any
burrs.
Always use a coupling fluid, such as silicone
grease, in the mounting interface to maintain a high
mounting stiffness. Mount the sensor according to the
manufacturer’s recommended mounting torque.
DO
NOT use any intermediate mounting adaptors, as the
mounted resonant frequency may be reduced and
thereby compromise the high-frequency performance. If
necessary, use adaptor studs.
Figure 13. Stud Mounting
For adhesive mount sensors, use a thin, stiff layer of
temporary adhesive such as quick-bonding gel or
superglue. DO NOT use excessive amounts of glue or
epoxy, as the mounting stiffness may be reduced and
compromise high-frequency performance.
It may also
damage the sensor during removal.
Figure 14. Incorrect Adhesive Mounting
Triaxial accelerometers should always be mounted
directly to the reference standard.
Unless absolutely
required, DO NOT use adaptors to re-orient the sensor
along the axis of motion, as the mounting stiffness may
be altered. The vibration at the test sensor sensing
element may differ from the vibration at the reference
standard due to a “cantilever” effect, seen in Figure 15.
Figure 15. Mounting Triaxial Sensors (Incorrect)
Understand Back-to-Back Calibration limitations.
Do
not expect the uncertainty of calibration to be any better
than ±2%. (In fact, the uncertainty may be as high as
±3% or ±4% for frequencies <10 Hz or >2 kHz.) Since
large sensors may affect high-frequency accuracy, verify
that the test sensor does not mass load the reference
standard. Validate your calibration system with another
accelerometer prior to each calibration session. Check
with the manufacturer for exact system specifications.
7.5 CONCLUSIONS
Without
an
adequate
understanding
of
dynamics,
determining what, when, and how to test a sensor is a
difficult task. Therefore, each user must weigh the cost,
time, and risk associated with self-calibration versus the
services of an accredited laboratory.
3425 Walden Avenue, Depew, NY 14043
Toll Free: 888-684-0013 • 24-hour SensorLineSM: 716-684-0001 • FAX: 716-685-3886
E-mail: [email protected] • Website: www.pcb.com
File name: VIB-CHGOPGUIDE-0702
Doc. No. 18293 Rev. NR
CHARGE OUTPUT ACCELEROMETER
Revision: A
ECN #: 46623
Performance
ENGLISH
SI
Sensitivity(± 10 %)
5 pC/g
.51 pC/(m/s²)
Measurement Range
± 1000 g pk
± 9800 m/s² pk
Frequency Range(± 5 %)
2.5 kHz
2.5 kHz
Frequency Range(± 1 dB)
4 kHz
4 kHz
Resonant Frequency
≥ 13.5 kHz
≥ 13.5 kHz
Non-Linearity(per full scale range)
≤ 2.5 %
≤ 2.5 %
Transverse Sensitivity
≤ 5 %
≤ 5 %
Environmental
Overload Limit(Shock)
± 2000 g pk
± 19,600 m/s² pk
Temperature Range(Operating)
-67 to 1200 °F
-55 to 649 °C
Temperature Range(Survivable Limit)
-67 to 1400 °F
-55 to 760 °C
Temperature Range(Cable Termination)
-67 to 900 °F
-55 to 482 °C
Temperature Response
See graph
See graph
[2]
Radiation Exposure Limit(Integrated Neutron Flux)
1 E10 N/cm²
1 E10 N/cm²
Radiation Exposure Limit(Integrated Gamma Flux)
1 E8 rad
1 E8 rad
Electrical
Capacitance
1000 pF
1000 pF
[1][2]
Insulation Resistance(@ 1200°F[649°C])
≥ 10 kohm
≥ 10 kohm
Insulation Resistance(@ 70° F [21°C])
≥ 100 MOhm
≥ 100 MOhm
Electrical Isolation
Signal isolated from case
Signal isolated from case
Physical
Sensing Element
UHT-12™
UHT-12™
Sensing Geometry
Shear
Shear
Housing Material
Inconel
Inconel
Sealing(Sensor Housing)
Hermetic
Hermetic
Sealing(Connector)
Welded Hermetic
Welded Hermetic
Size (Height x Length x Width)
.66 in x 1.16 in x .66 in
16.7 mm x 29.5 mm x 16.7 mm
Weight(without cable)
1.8 oz
50 gm
[2]
Electrical Connector
Integral Hardline Cable
Integral Hardline Cable
Cable Length
10 ft
3.05 m
Cable Type
MI Hardline Cable
MI Hardline Cable
Cable Termination
10-32 Coaxial Jack
10-32 Coaxial Jack
Mounting
Through Hole
Through Hole
All specifications are at room temperature unless otherwise specified.
In the interest of constant product improvement, we reserve the right to change specifications without notice.
ICP® is a registered trademark of PCB Group, Inc.
OPTIONAL VERSIONS
Optional versions have identical specifications and accessories as listed for the standard model
except where noted below. More than one option may be used.
NOTES:
[1] Additional cable length will add approx. 100pF/ft.
[2] Typical.
SUPPLIED ACCESSORIES:
Model 081C108 Mounting Screw, 1/4-28x1, Inconel (1)
Model ACS-1 NIST traceable frequency response (10 Hz to upper 5% point). (1)
Entered: LK
Engineer: GJR
Sales: JC
Approved: BAM
Spec Number:
Date: 3/30/2017
Date: 3/30/2017
Date: 3/30/2017
Date: 3/30/2017
66378
3425 Walden Avenue, Depew, NY 14043
Phone: 716-684-0001
Fax: 716-684-0987
E-Mail: [email protected]
UNLESS OTHERWISE SPECIFIED TOLERANCES ARE:
DIMENSIONS IN MILLIMETERS
[ IN BRACKETS ]
XX ± 0.13
ANGLES 2 DEGREES
3425 WALDEN AVE. DEPEW, NY 14043
(716) 684-0001 E-MAIL: [email protected]
DIMENSIONS IN INCHES
DECIMALS XX ±.01
ANGLES 2 DEGREES
FILLETS AND RADII
.003 - .005
DECIMALS X ± 0.3
FILLETS AND RADII
0.07 - 0.13
INSTALLATION DRAWING
66377
1 OF 1
1.25X
MODEL 357E91
ACCELEROMETER
XXX ±.005
JDM
3/14/17
JDM
3/14/17
BAM
3/14/17
66377
PCB Piezotronics Inc. claims proprietary rights in
the information disclosed hereon. Neither it nor any
reproduction thereof will be disclosed to others
without the written consent of PCB Piezotronics Inc.
REVISIONS
REV
DESCRIPTION
DIN
A
UPDATED MOUNTING HOLE DEPTH
46617
.656 [16.66]
.290 [7.37]
1.16 [29.5]
.265 [6.73]
.72 [18.2]
120 4 3048 102
[
]
1.28 [32.5]
.286 THRU
.450 .250
7.26 THRU
[11.43 6.35
]
MINERAL OXIDE INSULATED
TRIAXIAL HARDLINE CABLE
.090 [2.29]
.25 [6.4]
10-32 COAXIAL CONNECTOR
5.) CABLE MINIMUM BEND RADIUS: .25 [6.35]
4.) RECOMMENDED MOUNTING TORQUE, 2-5 FOOT POUNDS [3-7 NEWTON METERS]
3.) FOR BEST RESULTS, PLACE A THIN LAYER OF SILICONE GREASE (OR EQUIVALENT) ON INTERFACE PRIOR
TO MOUNTING
MOUNTING SURFACE SHOULD BE FLAT TO WITHIN .001[0.03] TIR OVER DIM 'A' WITH A 63[1.61] SURFACE FINISH
OR BETTER FINISH FOR BEST RESULTS
DRILL PERPENDICULAR TO MOUNTING SURFACE TO WITHIN 1
1
2
MOUNTING HOLE PREPARATION:
.218 [5.54] .800 [20.32] MIN
1/4-28 UNF-2B .700 [17.78] MIN
1
2
DIRECTION OF ACCELERATION
TO PROVIDE A NEGATIVE SIGNAL
MARKING LOCATION