HP 8 1 6 3 A / 4 A / 6 A L i g h t w a v e S e r i e s M a i n f r a m e s
HP 8163A Lightwave Multimeter,
HP 8164A Lightwave Measurement
System,
& HP 8166A Lightwave
Multichannel System Programming
Guide
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System Programming Guide, E1299
Notices
This document contains proprietary information that is protected
by copyright. All rights are reserved.
No part of this document may be photocopied, reproduced, or
translated to another language without the prior written consent
of Hewlett-Packard GmbH.
Copyright 1999 by:
Hewlett-Packard GmbH
Herrenberger Str. 130
71034 Böblingen
Germany
Subject Matter
The information in this document is subject to change without
notice.
Hewlett-Packard makes no warranty of any kind with regard to
this printed material, including, but not limited to, the implied
warranties of merchantability and fitness for a particular
purpose.
Hewlett-Packard shall not be liable for errors contained herein or
for incidental or consequential damages in connection with the
furnishing, performance, or use of this material.
Printing History
New editions are complete revisions of the guide reflecting
alterations in the functionality of the instrument. Updates are
occasionally made to the guide between editions. The date on the
title page changes when an updated guide is published. To find
out the current revision of the guide, or to purchase an updated
guide, contact your Hewlett-Packard representative.
Control Serial Number: First Edition applies directly to all
instruments.
Warranty
This Hewlett-Packard instrument product is warranted against
defects in material and workmanship for a period of one year
from date of shipment. During the warranty period, HP will, at
its option, either repair or replace products that prove to be
defective.
For warranty service or repair, this product must be returned to a
service facility designated by HP. Buyer shall prepay shipping
charges to HP and HP shall pay shipping charges to return the
product to Buyer. However, Buyer shall pay all shipping charges,
duties, and taxes for products returned to HP from another
country.
HP warrants that its software and firmware designated by HP for
use with an instrument will execute its programming instructions
when properly installed on that instrument. HP does not warrant
that the operation of the instrument, software, or firmware will
be uninterrupted or error free.
Limitation of Warranty
The foregoing warranty shall not apply to defects resulting from
improper or inadequate maintenance by Buyer, Buyer-supplied
software or interfacing, unauthorized modification or misuse,
operation outside of the environmental specifications for the
product, or improper site preparation or maintenance.
No other warranty is expressed or implied. Hewlett-Packard
specifically disclaims the implied warranties of Merchantability
and Fitness for a Particular Purpose.
Exclusive Remedies
The remedies provided herein are Buyer’s sole and exclusive
remedies. Hewlett-Packard shall not be liable for any direct,
indirect, special, incidental, or consequential damages whether
based on contract, tort, or any other legal theory.
Assistance
Product maintenance agreements and other customer assistance
agreements are available for Hewlett-Packard products. For any
assistance contact your nearest Hewlett-Packard Sales and
Service Office.
Certification
Hewlett-Packard Company certifies that this product met its
published specifications at the time of shipment from the factory.
Hewlett-Packard further certifies that its calibration
measurements are traceable to the United States National
Institute of Standards and Technology, NIST (formerly the
United States National Bureau of Standards, NBS) to the extent
allowed by the Institutes’s calibration facility, and to the
calibration facilities of other International Standards
Organization members.
ISO 9001 Certification
Produced to ISO 9001 international quality system standard as
part of our objective of continually increasing customer
satisfaction through improved process control.
08164-91016 E1299
First Edition:
E1299: December 1999
Second Edition
E0599: May 1999
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System Programming Guide, E1299
3
In this Manual
This manual contains information about SCPI commands which can be used to
program the following instruments:
• HP 8163A Lightwave Multimeter
• HP 8164A Lightwave Measurement System
• HP 8166A Lightwave Multichannel System
The Structure of this Manual
This manual is divided into 5 parts:
• “Introduction to Programming” on page 13 gives a general introduction to
SCPI programming with the HP 8163A Lightwave Multimeter, the HP 8164A
Lightwave Measurement System, and the HP 8166A Lightwave Multichannel
System.
• “Specific Commands” on page 33 lists all instrument specific commands.
• “Instrument Setup and Status” on page 43, “Measurement Operations &
Settings” on page 63, and “Mass Storage, Display, and Print Functions” on
page 121 give fuller explanations of all instrument specific commands.
• “Programming Examples” on page 125 gives some example programs
showing how the SCPI commands can be used with the HP 8163A Lightwave
Multimeter, the HP 8164A Lightwave Measurement System, and the
HP 8166A Lightwave Multichannel System.
• “The HP 816x VXIplug&play Instrument Driver” on page 147, “GPIB
Command Compatibility List” on page 163, and “Error Codes” on page 171
give information about the HP 816x VXIplug&play Instrument Driver,
compatibility issues, and error codes.
Conventions used in this Manual
• All commands and typed text is written in Courier font, for example
INIT[:IMM].
• SCPI commands are written in mixed case: text that you MUST print is written
in capitals; text which is helpful but nor necessary is written in lower case.
So, the command INITiate[:IMMediate] can be entered either as
init[:imm], or as initiate[:immediate]. It does not matter whether
you enter text using capitals or lower-case letters.
• SCPI commands often contain extra arguments in square brackets. These
arguments may be helpful, but they need not be entered.
So, the command INITiate[:IMMediate] can be entered as init or
initiate:imm.
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System Programming Guide, E1299
• A SCPI command which can be either a command or a query is appended with
the text /?.
So, DISPlay:ENABle/? refers to both the command DISPlay:ENABle
and the query DISPlay:ENABle?.
Related Manuals
You can find more information about the instruments covered by this manual in
the following manuals:
• HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide (HP Product
Number 08164-91011).
NOTE
Please note that User Guides no longer contain programming information, and
must now be used in conjunction with this manual.
If you are not familiar with the General Purpose Interface Bus, GPIB, then refer
to the following books:
• ANSI/IEEE-488.1-1978, IEEE Standard Digital Interface for Programmable
Instrumentation, and ANSI/IEEE-488.2-1987, IEEE Standard Codes,
Formats, and Common Commands, published by the Institute of Electrical and
Electronic Engineers.
In addition, the commands not from the IEEE 488.2 standard are defined
according to the Standard Commands for Programmable Instruments (SCPI). For
an introduction to SCPI and SCPI programming techniques, refer to the following
documents:
• Hewlett-Packard Press (Addison-Wesley Publishing Company, Inc.): A
Beginners Guide to SCPI by Barry Eppler.
• The SCPI Consortium: Standard Commands for Programmable Instruments.
To obtain a copy of this manual, contact the following address:
SCPI Consortium Office
Bode Enterprise
2515 Camino del Rio South, Suite 340
San Diego, CA, 92108
USA
Web: http://www.scpiconsortium.org
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System Programming Guide, E1299
5
Contents
In this Manual
3
The Structure of this Manual
3
Conventions used in this Manual
3
Related Manuals
4
Contents
5
Figures
9
Tables
11
Introduction to Programming
13
GPIB Interface
15
Setting the GPIB Address
16
Returning the Instrument to Local Control
17
Message Queues
17
How the Input Queue Works
17
Clearing the Input Queue
18
The Output Queue
18
The Error Queue
18
Programming and Syntax Diagram Conventions
19
Short Form and Long Form
19
Command and Query Syntax
19
Units
20
Data Types
20
Slot and Channel Numbers
21
Laser Selection Numbers
22
Common Commands
22
Common Command Summary
23
Common Status Information
23
The Status Model
25
Status Registers
25
Status System for HP 8163A & HP 8164A
26
Status System for HP 8166A
27
Annotations
28
Status Byte Register
28
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System Programming Guide, E1299
Standard Event Status Register
29
Operation/Questionable Status Summary
29
Operation/Questionable Status Summary Register
29
Operation/Questionable Slot Status
29
Operation Slot Status Register
30
Questionable Slot Status Register
30
Status Command Summary
30
Other Commands
31
Specific Commands
33
Specific Command Summary
35
Instrument Setup and Status
43
IEEE-Common Commands
45
Status Reporting – The STATus Subsystem
50
Interface/Instrument Behaviour Settings – The SYSTem Subsystem59
Measurement Operations & Settings
63
Root Layer Command
65
Measurement Functions – The SENSe Subsystem
68
HP 81635A and HP 81619A- Master and Slave Channels
68
Signal Generation – The SOURce Subsystem
86
Triggering - The TRIGger Subsystem
108
Extended Trigger Configuration
114
Extended Trigger Configuration Example
118
Mass Storage, Display, and Print Functions
121
Display Operations – The DISPlay Subsystem
123
Programming Examples
125
How to Use VISA Calls
127
How to Set up a Fixed Laser Source
129
How to Measure Power using FETCh and READ
131
How to Co-ordinate Two Modules
134
How Power Varies with Wavelength
138
How to Log Results
142
The HP 816x VXIplug&play Instrument Driver
147
Installing the HP 816x Instrument Driver
149
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System Programming Guide, E1299
7
Using Visual Programming Environments
152
Getting Started with HP VEE
152
GPIB Interfacing in HP VEE
153
Getting Started with LabView
154
Getting Started with LabWindows
156
Features of the HP 816x Instrument Driver
156
Directory Structure
157
Opening an Instrument Session
157
Closing an Instrument Session
158
VISA Data Types and Selected Constant Definitions
159
Error Handling
159
Introduction to Programming
161
Example Programs
161
VISA-Specific Information
161
Instrument Addresses
161
Callbacks
161
Development Environments
161
Microsoft Visual C++ 4.0 (or higher) and Borland C++ 4.5 (or higher)161
Microsoft Visual Basic 4.0 (or higher)
161
HP VEE 5.01 (or higher)
162
LabWindows CVI/ (R) 4.0 (or higher)
162
Online Information
162
GPIB Command Compatibility List
163
Compatibility Issues
165
GPIB Bus Compatibility
165
Status Model
165
Preset Defaults
165
Removed Command
166
Obsolete Commands
166
Changed Parameter Syntax and Semantics
167
Changed Query Result Values
168
Timing Behavior
168
Error Handling
169
Command Order
169
Instrument Status Settings
169
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System Programming Guide, E1299
Error Codes
171
GPIB Error Strings
173
Index
179
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System Programming Guide, E1299
9
Figures
Figure 1
Remote Control . . . . . . . . . . . . . . 17
Figure 2
The Event Status Bit . . . . . . . . . . . . . 24
Figure 3
The Registers and Filters for a Node . . . . . . . . . . 25
Figure 4
The Operational/Questionable Status System for HP 8163A & HP 8164A . 27
Figure 5
The Operational/Questionable Status System for HP 8166A . . . . 28
Figure 6
Extended Trigger Configuration . . . . . . . . . . . 116
Figure 7
Setup for Extended Trigger Configuration Example . . . . . . 118
Figure 8
Non-Administrator Installation Pop-Up Box . . . . . . . . 149
Figure 9
Message Screen . . . . . . . . . . . . . . 150
Figure 10
Customizing Your Setup . . . . . . . . . . . . 151
Figure 11
Program Folder Item Options . . . . . . . . . . . 151
Figure 12
Device Configuration . . . . . . . . . . . . . 153
Figure 13
Advanced Device Configuration - Plug&play Driver . . . . . . 153
Figure 14
FP Conversion Options Box. . . . . . . . . . . . 155
Figure 15
Windows 95 and Windows NT VXIPNP Directory Structure . . . . 157
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System Programming Guide, E1299
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System Programming Guide, E1299
11
Tables
Table 1
GPIB Capabilities . . . . . . . . . . . . . . 16
Table 2
Units and allowed Mnemonics . . . . . . . . . . . 20
Table 3
Common Command Summary . . . . . . . . . . . 23
Table 4
Specific Command Summary . . . . . . . . . . . 35
Table 5
Commands that can only be configured using the master channel . . . 69
Table 6
Commands that are independent for both master and slave channels. . . 69
Table 7
Triggering and Power Measurements . . . . . . . . . 108
Table 8
Generating Output Triggers from Power Measurements . . . . . 109
Table 9
Incompatible GPIB Bus Commands . . . . . . . . . . 165
Table 10
Removed Commands . . . . . . . . . . . . . 166
Table 11
Obsolete Commands . . . . . . . . . . . . . 167
Table 12
Commands with Different Parameters or Syntax . . . . . . . 167
Table 13
Queries with Different Result Values . . . . . . . . . 168
Table 14
Timing Behavior Changes . . . . . . . . . . . . 168
Table 15
Error Handling Changes . . . . . . . . . . . . 169
Table 16
Specific Errors. . . . . . . . . . . . . . . 169
Table 17
Overview for Supported Strings . . . . . . . . . . . 173
Table 18
Overview for Unsupported Strings . . . . . . . . . . 177
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System Programming Guide, E1299
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
13
Introduction to
Programming
Introduction to Programming
14
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
This chapter gives general information on how to control your instrument
remotely.
Descriptions for the actual commands for the instruments are given in the
following chapters. The information in these chapters is specific to the HP 8163A
Lightwave Multimeter, HP 8164A Lightwave Measurement System, and
HP 8166A Lightwave Multichannel System and assumes that you are already
familiar with programming the GPIB.
Introduction to Programming
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
15
GPIB Interface
The interface used by your instrument is the GPIB (General Purpose Interface
Bus).
GPIB is the interface used for communication between a controller and an
external device, such as the tunable laser source. The GPIB conforms to IEEE
standard 488-1978, ANSI standard MC 1.1 and IEC recommendation 625-1.
If you are not familiar with the GPIB, then refer to the following books:
• The International Institute of Electrical and Electronics Engineers. IEEE
Standard 488.1-1987, IEEE Standard Digital Interface for Programmable
Instrumentation. New York, NY, 1987
• The International Institute of Electrical and Electronics Engineers. IEEE
Standard 488.2-1987, IEEE Standard Codes, Formats, Protocols and
Common Commands For Use with ANSI/IEEE Std 488.1-1987. New York,
NY, 1987
To obtain a copy of either of these last two documents, write to:
The Institute of Electrical and Electronics Engineers, Inc.
345 East 47th Street
New York, NY 10017
USA.
In addition, the commands not from the IEEE-488.2 standard, are defined
according to the Standard Commands for Programmable Instruments (SCPI).
For an introduction to SCPI, and SCPI programming techniques, please refer to
the following documents:
• Hewlett-Packard Press (Addison-Wesley Publishing Company, Inc.). A
Beginners Guide to SCPI. Barry Eppler. 1991.
• The SCPI Consortium: Standard Commands for Programmable Instruments.
To obtain a copy of this manual, contact the following address:
SCPI Consortium Office
Bode Enterprise
2515 Camino del Rio South, Suite 340
San Diego, CA, 92108
USA
Web: http://www.scpiconsortium.org
The interface of the HP 8163A Lightwave Multimeter and of the HP 8164A
Lightwave Measurement System to the GPIB is defined by the IEEE Standards
488.1 and 488.2.
Table 1 shows the interface functional subset that the instruments implement.
Introduction to Programming
GPIB Interface
16
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
Setting the GPIB Address
There are two ways to set the GPIB address:
• You can set the GPIB address by using the command
“:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess” on page 62.
• You can set the GPIB address from the front panel. See your instrument’s
User’s Guide for more information.
The default GPIB address is 20.
Mnemonic Function
SH1
Complete source handshake capability
AH1
Complete acceptor handshake capability
T6
Basic talker; serial poll; unaddressed to talk if addressed to
listen
L4
Basic listener; unaddressed to listen if addressed to talk; no
listen only
SR1
Complete service request capability
RL1
Complete remote/local capability
PP0
No parallel poll capability
DC1
Device clear capability
DT0
No device trigger capability
C0
No controller capability (Controller capability to be imple-
mented)
Table 1
GPIB Capabilities
Introduction to Programming
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
17
Returning the Instrument to Local
Control
If the instrument is in remote control, a screen resembling
Figure 1 will appear. Press [Local] if you wish to return the instrument to local
control.
Message Queues
The instrument exchanges messages using an input and an output queue. Error
messages are kept in a separate error queue.
How the Input Queue Works
The input queue is a FIFO queue (first-in first-out). Incoming bytes are stored in
the input queue as follows:
1 Receiving a byte:
– Clears the output queue.
– Clears Bit 7 (MSB).
2 No modification is made inside strings or binary blocks. Outside strings and
binary blocks, the following modifications are made:
Figure 1
Remote Control
Introduction to Programming
Message Queues
18
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
– Lower-case characters are converted to upper-case.
– The characters 0016 to 0916 and 0B16 to 1F16 are converted to spaces (2016).
– Two or more blanks are truncated to one.
3 An EOI (End Or Identify) sent with any character is put into the input queue as
the character followed by a line feed (LF, 0A16). If EOI is sent with a LF, only
one LF is put into the input queue.
4 The parser starts if the LF character is received or if the input queue is full.
Clearing the Input Queue
Switching the power off, or sending a Device Interface Clear signal, causes
commands that are in the input queue, but have not been executed to be lost.
The Output Queue
The output queue contains responses to query messages. The instrument transmits
any data from the output queue when a controller addresses the instrument as a
talker.
Each response message ends with a carriage return (CR, 0D16) and a LF (0A16),
with EOI=TRUE. If no query is received, or if the query has an error, the output
queue remains empty.
The Message Available bit (MAV, bit 4) is set in the Status Byte register whenever
there is data in the output queue.
The Error Queue
The error queue is 30 errors long. It is a FIFO queue (first-in first-out). That is,
the first error read is the oldest error to have occurred. A new error is only put into
the queue if it is not already in it.
If more than 29 errors are put into the queue, the message:
-350
is placed as the last message in the queue.
Programming and Syntax Diagram Conventions
Introduction to Programming
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
19
Programming and Syntax
Diagram Conventions
A program message is a message containing commands or queries that you send
to the instruments. The following are a few points about program messages:
• You can use either upper-case or lower-case characters.
• You can send several commands in a single message. Each command must be
separated from the next one by a semicolon (;).
• A command message is ended by a line feed character (LF) or .
• You can use any valid number/unit combination.
In other words, 1500NM,1.5UM and 1.5E-6M are all equivalent.
If you do not specify a unit, then the default unit is assumed. The default unit
for the commands are given with command description in the next chapter.
Short Form and Long Form
The instrument accepts messages in short or long forms.
For example, the message
:STATUS:OPERATION:ENABLE 768
is in long form.
The short form of this message is
:STAT:OPER:ENAB 768
In this manual, the messages are written in a combination of upper and lower
case. Upper case characters are used for the short form of the message.
For example, the above command would be written
:STATus:OPERation:ENABle
The first colon can be left out for the first command or query in your message.
That is, the example given above could also be sent as
STAT:OPER:ENAB 768
Command and Query Syntax
All characters not between angled brackets must be sent exactly as shown.
The characters between angled brackets (<...>) indicate the kind of data that
you should send, or that you get in a response. You do not type the angled
brackets in the actual message.
Introduction to Programming
Programming and Syntax Diagram Conventions
20
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
Descriptions of these items follow the syntax description. The following types of
data are most commonly used:
Other kinds of data are described as required.
The characters between square brackets ([...]) show optional information that
you can include with the message.
The bar (|) shows an either-or choice of data, for example, a|b means either a or
b, but not both simultaneously.
Extra spaces are ignored, so spaces can be inserted to improve readability.
Units
Where units are given with a command, usually only the base units are specified.
The full sets of units are given in the table below.
Data Types
With the commands you give parameters to the instrument and receive response
values from the instrument. Unless explicitly specified these data are given in
ASCII format. The following types of data are used:
• Boolean data may only have the values 0 or 1.
• Integer range is given for each individual command.
• Float variables may be given in decimal or exponential writing (0.123 or
123E-3).
All Float values conform to the 32 bit IEEE Standard, that is, all Float values
are returned as 32-bit real values.
• A string is contained between double quotes ("...") or single quotes (‘...’).
When the instrument returns a string, it is always included in " " and
terminated by .
string
is ascii data. A string is contained between double
quotes ("...") or single quotes (‘...’).
value
is numeric data in integer (12), decimal (34.5) or ex-
ponential format (67.8E-9).
wsp
is a white space.
Unit
Default
Allowed Mnemonics
meters
M
PM, NM, UM, MM, M
decibel
DB
MDB, DB
second
S
NS, US, MS, S
decibel/1mW
DBM
MDBM, DBM
Hertz
HZ
HZ, KHZ, MHZ, GHZ, THZ
Watt
Watt
PW, NW, UW, MW, Watt
meters per second M/S
NM/S, UM/S, MM/S, M/S
Table 2
Units and allowed Mnemonics
Programming and Syntax Diagram Conventions
Introduction to Programming
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
21
• When a register value is given or returned (for example *ESE), the decimal
values for the single bits are added. For example, a value of nine means that bit
0 and bit 3 are set.
• Larger blocks of data are given as Binary Blocks, preceded by
“#”, terminated by ; represents the number
of digits, represents the number of bytes, and is the data
block. For example, for a Binary Block with 1 digit and 6 bytes this is:
#16TRACES.
Slot and Channel Numbers
Each module is identified by a slot number and a channel number. For commands
that require you to specify a channel, the slot number is represented by [n] in a
command and the channel number is represented by [m].
The slot number represents the module’s position in the mainframe. These are:
• from one to two for the HP 8163A,
• from zero to four for the HP 8164A, and
• from one to seventeen for the HP 8166A.
These numbers are displayed on the front panel beside each module slot.
NOTE
The HP 8164A slot for back-loadable tunable laser modules is numbered zero.
Channel numbers apply to modules that have two inputs/outputs, for example, the
HP 81635A Dual Power Sensor.
Modules with two channels, for example, the HP 81635A Dual Power Sensor, use
the channel number to distinguish between these channels.
NOTE
The channel number of single channel modules is always one.
For example, if you want to query slot 1, channel 2 with the command,
“:SENSe[n]:[CHANnel[m]]:POWer:WAVelength?” on page 83, you should
send the command:
• :sens1:chan2:pow:wav?
NOTE
If you do not specify a slot or channel number, the lowest possible number is used
as the default value. This means:
•
Slot 1 for the HP 8163A and HP 8166A mainframes.
•
Slot 0 for the HP 8164A mainframe.
•
Channel 1 for all channels.
Introduction to Programming
Common Commands
22
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
Laser Selection Numbers
The laser selection number, [l], identifies the upper or lower wavelength laser
source for dual wavelength Laser Source modules and Return Loss modules with
two internal laser sources. The lower wavelength source is denoted by 1. The
upper wavelength source is denoted by 2.
NOTE
For Return Loss modules, 0 denotes the use of an external laser source as the
input to your Return Loss module for the following commands:
• “:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:FPDelta[l]” on
page 85,
• “:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:FPDelta[l]?” on
page 85,
• “:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l]” on
page 85, and
• “:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l]?” on
page 86.
Common Commands
The IEEE 488.2 standard has a list of reserved commands, called common
commands. Some of these commands must be implemented by any instrument
using the standard, others are optional.
Your instrument implements all the necessary commands, and some optional
ones. This section describes the implemented commands.
Introduction to Programming
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
23
Common Command Summary
Table 3 gives a summary of the common commands.
NOTE
These commands are described in more detail in “IEEE-Common Commands” on page 45.
Common Status Information
There are three registers for the status information. Two of these are status-
registers and one is an enable-registers. These registers conform to the IEEE
Standard 488.2-1987. You can find further descriptions of these registers under
*ESE, *ESR?, and *STB?.
Command Parameter Function
Page
*CLS
Clear Status Command
page 45
*ESE
Standard Event Status Enable Command
page 45
*ESE?
Standard Event Status Enable Query
page 46
*ESR?
Standard Event Status Register Query
page 46
*IDN?
Identification Query
page 46
*OPC
Operation Complete Command
page 47
*OPC?
Operation Complete Query
page 47
*OPT?
Options Query
page 47
*RST
Reset Command
page 48
*STB?
Read Status Byte Query
page 48
*TST?
Self Test Query
page 49
*WAI
Wait Command
page 49
Table 3
Common Command Summary
Introduction to Programming
Common Commands
24
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
Figure 2 shows how the Standard Event Status Enable Mask (SESEM) and the
Standard Event Status Register (SESR) determine the Event Status Bit (ESB) of
the Status Byte.
The SESR contains the information about events that are not slot specific. For
details of the function of each bit of the SESR, see “Standard Event Status
Register” on page 29.
The SESEM allows you to choose the event that may affect the ESB of the Status
Byte. If you set a bit of the SESEM to zero, the corresponding event cannot affect
the ESB. The default is for all the bits of the SESEM to be set to 0.
The questionable and operation status systems set the Operational Status Bit
(OSB) and the Questionable Status Bit (QSB). These status systems are described
in “The Status Model” on page 25 and “Status Reporting – The STATus
Subsystem” on page 50.
NOTE
Unused bits in any of the registers change to 0 when you read them.
Figure 2
The Event Status Bit
0
1
2
3
4
5
6
7
*STB? returns the Status Byte Register
Status
OSB
ESB
QSB
*ESR? returns the Standard Event Status Register
0
0
1
0
1
2
3
4
5
6
7
Event
1
0
0
0
0
0
Status
Register
0
1
2
3
4
5
6
7
Event
1
1
1
1
1
1
Status
Enable
Mask
*ESE sets the Standard Event Status Enable Mask
&
&
&
&
&
&
&
&
OR
Byte
All bits shown as
are unused
The Status Model
Introduction to Programming
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
25
The Status Model
Status Registers
Each node of the status circuitry has three registers:
• A condition register (CONDition), which contains the current status. This
register is updated continuously. It is not changed by having its contents read.
• The event register (EVENt), which contains details of any positive transitions
in the corresponding condition register, that is, when a bit changes from 0 → 1.
The contents of this register are cleared when it is read. The contents of any
higher-level registers are affected with regard to the appropriate bit.
• The enable register (ENABle), which enables changes in the event register to
affect the next stage of registers.
NOTE
The event register is the only kind of register that can affect the next stage of
registers.
The structures of the Operational and Questionable Status Systems are similar.
Figure 4 describe how the Questionable Status Bit (QSB) and the Operational
Status Bit (OSB) of the Status Byte Register are determined.
Figure 3
The Registers and Filters for a Node
Enable Registers
To the
Condition Register
of the Next Node
Event Registers
Condition Registers
OR
1 1 1 1 1
A positive transition in the condition
causes the corresponding bit of the
to change from 0 → 1.
corresponding event register
register, when a bit changes from 0 → 1,
Introduction to Programming
The Status Model
26
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
The Operational/Questionable Slot Status Event Register (OSSER/QSSER)
contains the status of a particular module slot. A bit changes from 0 → 1 when an
event occurs, for example, when a laser is switched on. For details of the function
of each bit of these registers, see “Operation/Questionable Status Summary
Register” on page 29 and “Operation/Questionable Status Summary Register”
on page 29.
The Operational/Questionable Slot Enable Status Mask (OSESM/QSESM)
allows you to choose the events for each module slot that may affect the
Operational/Questionable Status Event Register (see below). If you set a bit of the
OSESM/QSESM to zero, the occurence of the corresponding event for this
particular module slot cannot affect the Operational/Questionable Status Event
Register. The default is for all the bits of the OSESM/QSESM to be set to 0.
The Operational/Questionable Status Event Summary Register (OSESR/QSESR)
summarizes the status of every module slot of your instrument. If, for any slot,
any bit of the QSSER goes from 0 → 1 AND the corresponding bit of the
QSSEM is 1at the same time, the QSESR bit representing that slot is set to 1.
The Operational/Questionable Status Enable Summary Mask (OSESM/QSESM)
allows you to choose the module slots that may affect the OSB/QSB of the Status
Byte. If any bit of the QSESR goes from 0 → 1 AND the corresponding bit of the
QSESM is 1at the same time, the QSB of the Status Byte is set to 1. If you set a
bit of the OSESM/QSESM to zero, the corresponding module slot cannot affect
the OSB/QSB. The default is for all the bits of the OSESM/QSESM to be set to 0.
The Operational/Questionable Status Enable Summary Mask for the HP 8163A
Lightwave Multimeter and the HP 8164A Lightwave Measurement System
consists of one level. These are described in “The Operational/Questionable
Status System for HP 8163A & HP 8164A” on page 27.
As the HP 8166A Lightwave Multichannel System has 17 module slots, the
Operational/Questionable Status Enable Summary Mask consists of two levels.
This is described in “Status System for HP 8166A” on page 27.
Status System for HP 8163A & HP 8164A
The status system for the HP 8163A Lightwave Multimeter and the HP 8164A
Lightwave Measurement System returns the status of 2 and 5 module slots
respectively. The Operational/Questionable Status Summary Registers consist of
The Status Model
Introduction to Programming
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
27
one level and are described by Figure 4. Any commands that require LEVel1 do
not apply to these mainframes.
Status System for HP 8166A
The status system for the HP 8166A Lightwave Multichannel System returns the
status of 17 module slots. The Operational/Questionable Status Summary
Registers consists of two levels, as described by Figure 5.
Module slots 1 to 14 affect the Level 0 summary register as described in Figure 4.
Bit 0 of the Level 0 summary registers represents the summary of the status of
Figure 4
The Operational/Questionable Status System for HP 8163A & HP 8164A
Operational/Questionable Status
Operational/Questionable Status
Operational/Questionable Status
Condition Summary Register
Event Summary Register
Enable Summary Mask
Register
Status Byte
Status Byte Register
Status Summary
to next
Operational/Questionable
Operational/Questionable
Operational/Questionable
Slot Status Condition
Slot Status Event
Slot Status Enable Mask
&
&&&
OR
level
to next
for a positive
&
&&&
OR
level
Slot 1
transition
for a positive
transition
Register
Register
Register
Introduction to Programming
The Status Model
28
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
module slots 15, 16, and 17. The Level 1 summary registers contain an individual
summary for each of these module slots.
Annotations
Status Byte Register
• Bit 3, the QSB, is built from the questionable event status register and its
enable mask.
• Bit 5, the ESB, is built from the SESR and its SESEM.
• Bit 7, the OSB, is built from the operation event status register and its enable
mask.
• All other bits are unused, and therefore set to 0.
Figure 5
The Operational/Questionable Status System for HP 8166A
Operational/Questionable Status
Operational/Questionable Status
Operational/Questionable Status
Condition Summary Register (Level 0)
Event Summary Register (Level 0)
Enable Summary Mask
Register (Level 0)
Status Byte
Status Byte Register
Status
Summary
to next
&
&&&
OR
level
to next
Operational/Questionable
Operational/Questionable
Operational/Questionable
Slot Status Condition
Slot Status Event
Slot Status Enable Mask
&
&&&
OR
level
to next
for a positive
&
&&&
OR
level
Status
Summary
Slot 15
transition
for a positive
transition
for a positive
transition
Operational/Questionable Status
Operational/Questionable Status
Operational/Questionable Status
Condition Summary Register (Level 1)
Event Summary Register (Level 1)
Enable Summary Mask
Register
Register
Register
Register (Level 1)
for Level 1
for Level 0
The Status Model
Introduction to Programming
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
29
Standard Event Status Register
• Bit 0 is set if an operation complete event has been received since the last call
to *ESR?.
• Bit 1 is always 0 (no service request).
• Bit 2 is set if a query error has been detected.
• Bit 3 is set if a device dependent error has been detected.
• Bit 4 is set if an execution error has been detected.
• Bit 5 is set if a command error has been detected.
• Bit 6 is always 0 (no service request).
• Bit 7 is set for the first call of *ESR? after Power On.
Operation/Questionable Status Summary
• The Operation/Questionable Status Summary consist of a condition and an
event register.
• A "rising" bit in the condition register is copied to the event register.
• A "falling" bit in the condition register has no effect on the event register.
• Reading the condition register is non-destructive.
• Reading the event register is destructive.
• A summary of the event register and its enable mask is set in the status byte.
Operation/Questionable Status Summary Register
• Bits 0 to 4 are built from the OSSER/QSSER and the OSSEM/QSSEM.
• A summary of the event register, the condition register and the enable mask is
set in the status byte.
Operation/Questionable Slot Status
• The Operation/Questionable Slot Status consist of a condition and an event
register.
• A "rising" bit in the condition register is copied to the event register.
• A "falling" bit in the condition register has no effect on the event register.
• Reading the condition register is non-destructive.
• Reading the event register is destructive.
• A summary of the event register, the condition register and the enable mask is
set in the status byte.
Introduction to Programming
The Status Model
30
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
Operation Slot Status Register
• Bit 0 is set if the laser is switched on.
• Bit 1 is set if the Coherence Control is switched on.
• Bit 3 is set if Power Meter zeroing or Tunable Laser module lambda zeroing is
ongoing.
• All other bits are unused, and therefore set to 0.
Questionable Slot Status Register
• Bit 0 is set if excessive power is set by the user for any source module or if
excessive averaging time is set for any Power Meter.
• Bit 1 is set if the last Power Meter zeroing or Tunable Laser module lambda
zeroing failed.
• Bit 2 is set if temperature is out of range.
• Bit 3 is set if laser protection is switched on.
• Bit 4 is set if the module has not settled.
• Bit 5 is set if the module is out of specifications.
• Bit 6 is set if ARA is recommended.
• Bit 7 is set if the duty cycle is out of range.
• All other bits are unused, and therefore set to 0.
Status Command Summary
*STB?
returns status byte, value 0 .. +255
*ESE
sets the standard event status enable mask, parameter 0 .. +255
*ESE?
returns SESE, value 0 .. +255
*ESR?
returns the standard event status register, value 0 .. +255
*OPC
parses all program message units in the message queue.
*OPC? returns 1 if all operations (scan trace printout, measurement) are com-
pleted. Otherwise it returns 0.
*CLS
clears the status byte and SESR, and removes any entries from the error
queue.
*RST
clears the error queue, loads the default setting, and restarts communi-
cation.
NOTE: *RST does NOT touch the STB or SESR. A running measure-
ment is stopped.
*TST?
initiates an instrument selftest and returns the results as a 32 bit LONG.
The Status Model
Introduction to Programming
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
31
Other Commands
*OPT?
returns the installed modules and the slots these modules are installed
in:
For example, *OPT? → 81682A, 81533B, 81532A, ,
Modules 81682A, 81533B, and 81532A are installed in slots 0 to 2 re-
spectively. Slots 3 and 4 are empty.
*WAI
prevents the instrument from executing any further commands until the
current command has finished executing. All pending operations are
completed during the wait period.
*IDN?
identifies the instrument; returns the manufacturer, instrument model
number, serial number, and firmware revision level.
Introduction to Programming
The Status Model
32
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
33
Specific Commands
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
This chapter lists all the instrument specific commands relating to the HP 8163A
Lightwave Multimeter and the HP 8164A Lightwave Measurement System, with
a single-line description.
Each of these summaries contains a page reference for more detailed information
about the particular command later in this manual.
Specific Command Summary
Specific Commands
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
35
Specific Command Summary
The commands are ordered in a command tree. Every command belongs to a node
in this tree.
The root nodes are also called the subsystems. A subsystem contains all
commands belonging to a specific topic. In a subsystem there may be further
subnodes.
All the nodes have to be given with a command. For example in the command
disp:brig
• DISPlay is the subsystem containing all commands for controlling the
display,
• BRIGhtness is the command selecting brightness.
NOTE
If a command and a query are both available, the command ends /?.
So, disp:brig/? means that disp:brig and disp:brig? are both
available.
Table 4 gives an overview of the command tree. You see the nodes, the subnodes,
and the included commands.
Command
Description
Page
:DISPlay
:BRIGhtness/?
Changes or queries the current display brightness.
123
:CONTrast/?
Changes or queries the current display contrast.
123
:ENABle/?
Switches the display on or off or queries whether the
display is on or off.
124
:FETCh[n][:CHANnel[m]][:SCALar]
:POWer[:DC]?
Returns a power value from a sensor.
70
:RETurnloss?
Returns the current return loss value.
70
:RETurnloss?
Returns a return loss value.
70
:INITiate[n]:[CHANnel[m]]
[:IMMediate]
Starts a measurement.
70
:CONTinuous/?
Starts or Queries a single/continuous measurement. 71
:LOCK/?
Switches the lock on/off or returns the current state
of the lock.
65
:OUTPut[n][:CHANnel[m]]
Table 4
Specific Command Summary
Specific Command Summary
36
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
:CONNection/?
Selects or returns Analog Output parameter.
86
:PATH/?
Sets or returns the regulated path.
87
[:STATe]/?
Sets a source’s output terminals to open or closed or
returns the current status of a source’s output termi-
nals.
87
:READ[n][:CHANnel[m]][:SCALar]
:POWer[:DC]?
Reads the current power value from a sensor.
71
:RETurnloss?
Reads the current return loss value.
72
:RETurnloss?
Returns the current return loss value.
72
:SENSe[n][:CHANnel[m]]:CORRection
[:LOSS][:INPut][:MAGNitude]/?
Sets or returns the value of correction data for a sen-
sor.
72
:COLLECT:ZERO
Executes a zero calibration of a sensor module.
73
:COLLECT:ZERO?
Returns the current zero state of a sensor module.
73
:COLLECT:ZERO:ALL
Executes a zero calibration of all sensor modules.
73
:SENSe[n][:CHANnel[m]]:FUNCtion
:PARameter:LOGGing/?
Sets or returns the number of samples and the aver-
aging time, tavg, for logging.
74
:PARameter:MINMax/?
Sets or returns the minmax mode and the window
size.
75
:PARameter:STABility/?
Sets or returns the total time, delay time and the av-
eraging time, tavg, for stability.
76
:RESult?
Returns the data array of the last function.
77
:STATe/?
Enables/disables the function mode or returns
whether the function mode is enabled.
77
:THReshold/?
Sets or returns the threshold value and the start
mode.
78
:SENSe[n][:CHANnel[m]]:POWer
:ATIMe/?
Sets or returns the average time of a sensor.
78
:RANGe[:UPPer]/?
Sets or returns the most positive signal entry expect-
ed for a sensor.
79
:RANGe:AUTO/?
Sets or returns the range of a sensor to produce the
most dynamic range without overloading.
79
:REFerence/?
Sets or returns the reference level of a sensor.
80
UNIT/?
Sets or returns the units used for absolute readings
on a sensor.
82
:WAVelength/?
Sets or returns the wavelength for a sensor.
83
:SENSe[n][:CHANnel[m]]:POWer:REFerence
Command
Description
Page
Table 4
Specific Command Summary (continued)
Specific Command Summary
Specific Commands
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
37
:DISPlay
Sets the reference level for a sensor from the input
power level.
81
:STATe/?
Sets or returns whether sensor results are in relative
or absolute units.
81
:STATe:RATio/?
Sets or returns whether sensor results are displayed
relative to a channel or to an absolute reference.
82
:SENSe[n][:CHANnel[m]]:RETurnloss:CALibration
:FACTory
84
:REFLectance
84
:TERMination
84
:SENSe[n][:CHANnel[m]]:RETurnloss:CORRection
:FPDelta[l]/?
Sets or returns
85
:REFLectance[l]/?
Sets or returns
85
:SENSe[n][:CHANnel[m]]:RETurnloss:CALibration
:FACTory
Selects the factory-set calibration values
84
:COLLect:REFLectance
Sets the Reflection Reference calibration values to
the values currently measured by the chosen return
loss module
84
:COLLect:TERMination
Sets the Reflection Reference calibration values to
the values currently measured by the chosen return
loss module
84
:SENSe[n][:CHANnel[m]]:RETurnloss:CORRection
:FPDelta[n]/?
Sets or returns the front panel delta, that is, the loss
variation value due to the front panel connector
85
:REFLectance[n]/?
Sets or returns Return Loss Reference, the return
loss value of your reference reflector.
85
:SLOT[n]
:EMPTy?
Returns whether the module slot is empty.
65
:IDN?
Returns information about the module.
66
:OPTions?
Returns the module’s options.
66
:TST?
Returns the latest selftest results for a module.
66
:SLOT[n][:HEAD[m]]
:EMPTy?
Returns whether an optical head is connected.
67
:IDN?
Returns information about the optical head.
67
:OPTions?
Returns the optical head’s options.
67
:TST?
Returns the latest selftest results for an optical head. 68
Command
Description
Page
Table 4
Specific Command Summary (continued)
Specific Command Summary
38
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
[:SOURce[n]][:CHANnel[m]]
:MODout/?
Returns the mode of the modulation output.
91
[:SOURce[n]][:CHANnel[m]:]AM
[:INTernal]:FREQuency[l]/?
Sets or returns the frequency of an internal signal
source.
88
:SOURce[l]/?
Sets or returns a source for the modulating system. 89
:STATe[l]/?
Turns Amplitude Modulation of a source on or off or
returns whether Amplitude Modulation is on or off.
90
[:SOURce[n]][:CHANnel[m]:]POWer
[:LEVel][:IMMediate][:AMPLitude]
Sets the laser output power of a source.
94
[:LEVel][:IMMediate][:AMPLitude[l]]? Returns the laser output power of a source.
95
[:LEVel]:RISetime/?
Sets or returns the laser rise time of a source.
95
:ATTenuation[l]/?
Sets or returns the attenuation level for a source.
92
:STATe/?
Sets or returns the state of the source output signal. 96
:UNIT/?
Sets or returns the power units.
96
:WAVelength/?
Sets or returns the wavelength source of a dual-
wavelength source.
99
[:SOURce[n]][:CHANnel[m]:]POWer:ATTenuation
:AUTO/?
Selects Automatic or Manual Attenuation Mode for
a source or returns the selected mode.
93
:DARK/?
Enables/disables ‘dark’ position on a source or re-
turns whether ‘dark’ position is active for a source.
93
[:SOURce[n]][:CHANnel[m]:]READout
:DATA?
Returns number of datapoints returned by the
[:SOURce[n]][:CHANnel[m]]:READout:POINts?
command.
98
:POINts?
Returns the data as a binary stream from either a
lambda logging operation or the maximum power
the laser can produce at each wavelength.
98
[:SOURce[n]][:CHANnel[m]:]WAVelength
[:CW|:FIXED]
Sets the absolute wavelength of a source.
99
[:CW|:FIXED[l]]?
Returns the absolute wavelength of a source.
99
:FREQuency/?
Sets the frequency difference used to calculate a rel-
ative wavelength for a source.
100
:REFerence?
Returns the reference wavelength of a source.
101
[:SOURce[n]][:CHANnel[m]:]WAVelength:CORRection
:ARA
Realigns the laser cavity.
100
Command
Description
Page
Table 4
Specific Command Summary (continued)
Specific Command Summary
Specific Commands
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
39
:ZERO
Exexutes a wavelength zero.
100
[:SOURce[n]][:CHANnel[m]:]WAVelength:REFerence
:DISPlay
Sets the reference wavelength of a source to the val-
ue of the output wavelength.
101
[:SOURce[n]][:CHANnel[m]:]WAVelength:SWEep
:CYCLes/?
Sets or returns the number of cycles.
101
:DWELl/?
Sets or returns the dwell time.
102
:LLOGging/?
Switches lambda logging on or off or queries the
state of lambda logging.
103
:MODE/?
Sets or returns the sweep mode.
103
:PMAX?
Returns the highest permissible power for a wave-
length sweep.
104
:REPeat/?
Sets or returns the repeat mode.
104
:SPEed/?
Sets or returns the speed for continuous sweeping.
105
:STARt/?
Sets or returns the start point of the sweep.
105
:STOP/?
Sets or returns the end point of the sweep.
106
[:STATe]/?
Stops, starts, pauses or continues a wavelength
sweep or returns the the state of a sweep.
106
[:SOURce[n]][:CHANnel[m]:]WAVelength:SWEep:STEP
:NEXT
Performs the next sweep step.
107
:PREVious
Performs the previous sweep step again.
107
[:WIDTh]/?
Sets or returns the width of the sweep step.
107
:SPECial
:REBoot
Reboots the mainframe and all modules.
68
:STATus[n]
:PRESet
Presets all Enable Registers.
54
:STATus:OPERation
[:EVENt]?
Returns the Operational Status Event Summary Reg-
ister.
50
[:EVENt]:LEVel1?
Returns the Operational Status Event Summary Reg-
ister for slots 15 - 17 of the HP 8166A Lightwave
Multichannel System.
52
:CONDition?
Returns the Operational Status Condition Summary
Register.
51
:CONDition:LEVel1?
Returns the Operational Status Condition Summary
Register for slots 15 - 17 of the HP 8166A Light-
wave Multichannel System.
52
Command
Description
Page
Table 4
Specific Command Summary (continued)
Specific Command Summary
40
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
:ENABle/?
Sets or queries the Operational Status Enable Sum-
mary Mask.
51
:ENABle:LEVel1/?
Sets or queries the Operational Status Enable Sum-
mary Mask for slots 15 - 17 of the HP 8166A Light-
wave Multichannel System.
52
:STATusn:OPERation
[:EVENt]?
Returns the Operational Slot Status Event Register
for slot n.
53
:CONDition?
Returns the Operational Slot Status Condition Reg-
ister for slot n.
53
:ENABle/?
Sets or queries the Operation Slot Status Enable
Mask for slot n.
54
:STATus:QUEStionable
[:EVENt]?
Returns the Questionable Status Event Summary
Register.
55
[:EVENt]:LEVel1?
Returns the Questionable Status Event Summary
Register for slots 15 - 17 of the HP 8166A Light-
wave Multichannel System.
57
:CONDition?
Returns the Questionable Status Condition Summa-
ry Register.
56
:CONDition:LEVel1?
Returns the Questionable Status Condition Summa-
ry Register for slots 15 - 17 of the HP 8166A Light-
wave Multichannel System.
57
:ENABle/?
Sets or queries the Questionable Status Enable Sum-
mary Mask.
56
:ENABle:LEVel1/?
Sets or queries the Questionable Status Enable Sum-
mary Mask for slots 15 - 17 of the HP 8166A Light-
wave Multichannel System.
57
:STATusn:QUEStionable
[:EVENt]?
Returns the Questionable Slot Status Event Register
for slot n.
58
:CONDition?
Returns the Questionable Slot Status Condition Reg-
ister for slot n.
58
:ENABle/?
Sets or queries the Questionable Slot Status En-
able Mask for slot n.
59
:SYSTem
:DATE/?
Sets or returns the instrument’s internal date.
59
:ERRor?
Returns the contents of the instrument’s error queue.60
:HELP:HEADers?
Returns a list of GPIB commands.
60
:PRESet
Sets all parameters to their default values.
61
:TIME/?
Sets or returns the instrument’s internal time.
61
Command
Description
Page
Table 4
Specific Command Summary (continued)
Specific Command Summary
Specific Commands
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
41
:VERSion?
Returns the instrument’s SCPI version.
61
:SYSTem:COMMunicate:GPIB
[:SELF]:ADDress/?
Sets or returns the GPIB address.
62
:TRIGger
Generates a hardware trigger.
109,
115
:CONFiguration/?
Sets or returns trigger configuration.
113
:TRIGger:CONFiguration
:EXTended/?
Sets or returns extended trigger configuration.
115
:FPEDal/?
Enables/disables the Input Trigger connector to be
triggered using a Foot Pedal or returns whether the
Input Trigger connector can be triggered using a
Foot Pedal.
115
:TRIGger[n][CHANnel[m]]
:INPut/?
Sets or returns the incoming trigger response .
110
:OUTPut/?
Sets or returns the outgoing trigger response.
112
Command
Description
Page
Table 4
Specific Command Summary (continued)
Specific Command Summary
42
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
43
Instrument Setup and
Status
Instrument Setup and Status
44
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
This chapter gives descriptions of commands that you can use when setting up
your instrument. The commands are split into the following separate subsytems:
• IEEE specific commands that were introduced in “Common Commands” on
page 22.
• STATus subsystem commands that relate to the status model.
• SYSTem subsystem commands that control the serial interface and internal
data.
Instrument Setup and Status
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
45
IEEE-Common Commands
“Common Commands” on page 22 gave a brief introduction to the IEEE-
common commands which can be used with the instruments. This section gives
fuller descriptions of each of these commands.
command:
*CLS
syntax:
*CLS
description:
The CLear Status command *CLS clears the following:
• Error queue
• Standard event status register (SESR)
• Status byte register (STB)
After the *CLS command the instrument is left waiting for the next command. The instru-
ment setting is unaltered by the command, although *OPC/*OPC? actions are cancelled.
parameters:
none
response:
none
example:
*CLS
command:
*ESE
syntax:
*ESE
0 ≤ value ≤ 255
description:
The standard Event Status Enable command (*ESE) sets bits in the Standard Event Status
Enable Mask (SESEM) that enable the corresponding bits in the standard event status reg-
ister (SESR).
The register is cleared:
• at power-on,
• by sending a value of zero.
The register is not changed by the *RST and *CLS commands.
parameters:
The bit value for the register (a 16-bit signed integer value):
Bit
Mnemonic
Decimal Value
7 (MSB)
Power On
128
6
Not Used
0
5
Command Error
32
4
Execution Error
16
3
Device Dependent Error
8
2
Query Error
4
1
Not Used
0
0 (LSB)
Operation Complete
1
response:
none
example:
*ESE 21
Instrument Setup and Status
IEEE-Common Commands
46
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
*ESE?
syntax:
*ESE?
description:
The standard Event Status Enable query *ESE? returns the contents of the Standard
Event Status Enable Mask (see *ESE for information on this register).
parameters:
none
response:
The bit value for the register (a 16-bit signed integer value).
example:
*ESE? → 21
command:
*ESR?
syntax:
*ESR?
description:
The standard Event Status Register query *ESR? returns the contents of the Standard
Event Status Register. The register is cleared after being read.
parameters
none
response
The bit value for the register (a 16-bit signed integer value):
Bit
Mnemonic
Decimal Value
7 (MSB)
Power On
128
6
Not used
0
5
Command Error
32
4
Execution Error
16
3
Device Dependent Error
8
2
Query Error
4
1
Not used
0
0 (LSB)
Operation Complete
1
example:
*ESR? → 21
command:
*IDN?
syntax:
*IDN?
description:
The IDeNtification query *IDN? gets the instrument identification over the interface.
parameters:
none
response:
The identification terminated by :
For example.
HEWLETT-PACKARD
mmmm
ssssssss
rrrrrrrrrr
manufacturer
instrument model number (for example 8164A)
serial number
firmware revision level
NOTE
The HP 8163A, HP8164A, and HP8166A will always return HEWLETT-PACKARD as
the manufacturer. This will not be affected by the transition of these instruments to
Agilent Technologies. This will allow programs that use this string to continue
functioning.
See “:SLOT[n]:HEAD[n]:IDN?” on page 67 for information on module identity strings.
example:
*IDN? → HEWLETT-PACKARD,mmmm,ssssssss,rrrrrrrrrr
Instrument Setup and Status
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
47
command:
*OPC
syntax:
*OPC
description:
The instrument parses and executes all program message units in the input queue and sets
the operation complete bit in the standard event status register (SESR). This command
can be used to avoid filling the input queue before the previous commands have finished
executing.
The following actions cancel the *OPC command (and put the instrument into Operation
Complete, Command Idle State):
•
Power-on
•
the Device Clear Active State is asserted on the interface.
•
*CLS
•
*RST
parameters:
none
response:
none
example:
*OPC
command:
*OPC?
syntax:
*OPC?
description:
The OPeration Complete query *OPC? parses all program message units in the input
queue, sets the operation complete bit in the Standard Event Status register, and places an
ASCII ’1’ in the output queue, when the contents of the input queue have been processed.
The following actions cancel the *OPC? query (and put the instrument into Operation
Complete, Command Idle State):
•
Power-on
•
the Device Clear Active State is asserted on the interface.
•
*CLS
•
*RST
parameters:
none
response:
1 is always returned.
example:
*OPC? → 1
command:
*OPT?
syntax:
*OPT?
description:
The OPTions query *OPT? returns the modules installed in your instrument.
parameters:
none
response:
Returns the part number of all installed modules, separated by commas.
Slots are listed starting with the lowest slot number, that is, slot 0 for the HP 8164A and
Slot 1 for the HP 8163A and HP 8166A.
If any slot is empty or not recognised, two spaces are inserted instead of the module’s part
number. See the example below, where slots 1 and 4 are empty.
example:
*OPT? → 81682A ,
, 81533B, 81532A,
Instrument Setup and Status
IEEE-Common Commands
48
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
*RST
syntax:
*RST
description:
The ReSeT command *RST sets the mainframe and all modules to the reset setting (stan-
dard setting) stored internally.
Pending *OPC? actions are cancelled.
The instrument is placed in the idle state awaiting a command. The *RST command clears
the error queue.
The *RST command is equivalent to the *CLS command AND the syst:preset com-
mand.
The following are not changed:
• GPIB (interface) state
• Instrument interface address
• Output queue
• Service request enable register (SRE)
• Standard Event Status Enable Mask (SESEM)
parameters:
none
response:
none
example:
*RST
command:
*STB?
syntax:
*STB?
description:
The STatus Byte query *STB? returns the contents of the Status Byte register.
parameters:
none
response:
The bit value for the register (a 16-bit signed integer value):
Bit
Mnemonic
Decimal Value
7 (MSB)
Operation Status
128
6
Not used
0
5
Event Status Bit
32
4
Not used
0
3
Questionable Status
8
2
Not used
0
1
Not used
0
0
Not used
0
example:
*STB? → 128
Instrument Setup and Status
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
49
command:
*TST?
syntax:
*TST?
description:
The self-TeST query *TST? makes the instrument perform a self-test and place the results
of the test in the output queue. If the self-test fails, the results are also put in the error
queue. We recommend that you read self-test results from the error queue. No further
commands are allowed while the test is running. After the self-test the instrument is re-
turned to the setting that was active at the time the self-test query was processed. The self-
test does not require operator interaction beyond sending the *TST? query.
parameters:
none
response:
The sum of the results for the individual tests (a 32-bit signed integer value, where 0 ≤
value ≤ 4294967296):
Bits
Mnemonic
Decimal Value
31
Selftest failed on Mainframe
A negative value
18 - 30
Not used
0
17
Selftest failed on Slot 17
131072
16
Selftest failed on Slot 16
65536
15
Selftest failed on Slot 15
32768
14
Selftest failed on Slot 14
16384
13
Selftest failed on Slot 13
8192
12
Selftest failed on Slot 12
4096
11
Selftest failed on Slot 11
2048
10
Selftest failed on Slot 10
1024
9
Selftest failed on Slot 9
512
8
Selftest failed on Slot 8
256
7
Selftest failed on Slot 7
128
6
Selftest failed on Slot 6
64
5
Selftest failed on Slot 5
32
4
Selftest failed on Slot 4
16
3
Selftest failed on Slot 3
8
2
Selftest failed on Slot 2
4
1
Selftest failed on Slot 1
2
0
Selftest failed on Slot 0
1
If 16 is returned, the module in slot 4 has failed.
If 18 is returned, the modules in slots 1 and 4 have failed.
A value of zero indicates no errors.
example:
*TST? → 0
command:
*WAI
syntax:
*WAI
description:
The WAIt command prevents the instrument from executing any further commands until
the current command has finished executing. All pending operations are completed during
the wait period.
parameters:
none
response:
none
example:
*WAI
Instrument Setup and Status
Status Reporting – The STATus Subsystem
50
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
Status Reporting – The STATus
Subsystem
The Status subsystem allows you to return and set details from the Status Model.
For more details, see “The Status Model” on page 25.
command:
:STATus:OPERation[:EVENt][:LEVel0]?
syntax:
:STATus:OPERation[:EVENt][:LEVel0]?
description:
Returns the Operational Status Event Summary Register (OSESR).
parameters:
none
response:
The sum of the results for the slots (a 16-bit signed integer value, where 0 ≤ value ≤
32767):
Bits
Mnemonics
Decimal Value
HP 8163A
HP 8164A
HP 8166A
15
Not used
Not used
Not used
0
14
Not used
Not used
Slot 14 Summary
16384
13
Not used
Not used
Slot 13 Summary
8192
12
Not used
Not used
Slot 12 Summary
4096
11
Not used
Not used
Slot 11 Summary
2048
10
Not used
Not used
Slot 10 Summary
1024
9
Not used
Not used
Slot 9 Summary
512
8
Not used
Not used
Slot 8 Summary
256
7
Not used
Not used
Slot 7 Summary
128
6
Not used
Not used
Slot 6 Summary
64
5
Not used
Not used
Slot 5 Summary
32
4
Not used
Slot 4 Summary
Slot 4 Summary
16
3
Not used
Slot 3 Summary
Slot 3 Summary
8
2
Slot 2 Summary
Slot 2 Summary
Slot 2 Summary
4
1
Slot 1 Summary
Slot 1 Summary
Slot 1 Summary
2
0
Not used
Slot 0 Summary
Level 1 Summary
1
example:
stat:oper? → +0
Status Reporting – The STATus Subsystem
Instrument Setup and Status
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
51
command:
:STATus:OPERation:CONDition[:LEVel0]?
syntax:
:STATus:OPERation:CONDition[:LEVel0]?
description:
Reads the Operational Status Condition Summary Register.
parameters:
none
response:
The sum of the results for the individual slots (a 16-bit signed integer value, where 0 ≤ val-
ue ≤ 32767):
Bits
Mnemonics
Decimal Value
HP 8163A
HP 8164A
HP 8166A
15
Not used
Not used
Not used
0
14
Not used
Not used
Slot 14 Summary
16384
13
Not used
Not used
Slot 13 Summary
8192
12
Not used
Not used
Slot 12 Summary
4096
11
Not used
Not used
Slot 11 Summary
2048
10
Not used
Not used
Slot 10 Summary
1024
9
Not used
Not used
Slot 9 Summary
512
8
Not used
Not used
Slot 8 Summary
256
7
Not used
Not used
Slot 7 Summary
128
6
Not used
Not used
Slot 6 Summary
64
5
Not used
Not used
Slot 5 Summary
32
4
Not used
Slot 4 Summary
Slot 4 Summary
16
3
Not used
Slot 3 Summary
Slot 3 Summary
8
2
Slot 2 Summary
Slot 2 Summary
Slot 2 Summary
4
1
Slot 1 Summary
Slot 1 Summary
Slot 1 Summary
2
0
Not used
Slot 0 Summary
Level 1 Summary
1
example:
stat:oper:cond? → +0
command:
:STATus:OPERation:ENABle[:LEVel0]
syntax:
:STATus:OPERation:ENABle[:LEVel0]
description:
Sets the bits in the Operational Status Enable Summary Mask (OSESM) that enable the
contents of the OSESR to affect the Status Byte (STB).
Setting a bit in this register to 1 enables the corresponding bit in the OSESR to affect bit 7
of the Status Byte.
parameters:
The bit value for the OSESM as a 16-bit signed integer value (0 .. +32767)
The default value is 0.
response:
none
example:
stat:oper:enab 128
command:
:STATus:OPERation:ENABle[:LEVel0]?
syntax:
:STATus:OPERation:ENABle[:LEVel0]?
description:
Returns the OSESM for the OSESR
parameters:
none
response:
The bit value for the operation enable mask as a 16-bit signed integer value (0 .. +32767)
example:
stat:oper:enab? → +128
Instrument Setup and Status
Status Reporting – The STATus Subsystem
52
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
:STATus:OPERation[:EVENt]:LEVel1?
syntax:
:STATus:OPERation[:EVENt]:LEVel1?
description:
Returns the Operational Status Event Summary Register (OSESR) for slots 15 to 17 of the
HP 8166A Lightwave Multichannel System.
parameters:
none
response:
The sum of the results for the slots (a 16-bit signed integer value, where 0 ≤ value ≤
32767):
Bits
Mnemonics
Decimal Value
HP 8166A
15-4
Not used
0
3
Slot 17 Summary
8
2
Slot 16 Summary
4
1
Slot 15 Summary
2
0
Not used
0
example:
stat:oper:level1? → +0
command:
:STATus:OPERation:CONDition:LEVel1?
syntax:
:STATus:OPERation:CONDition:LEVel1?
description:
Returns the Operational Status Condition Summary Register for slots 15 to 17 of the
HP 8166A Lightwave Multichannel System.
parameters:
none
response:
The sum of the results for slots 15 to 17 (a 16-bit signed integer value, where 0 ≤ value ≤
32767):
Bits
Mnemonics
Decimal Value
HP 8166A
15-4
Not used
0
3
Slot 17 Summary
8
2
Slot 16 Summary
4
1
Slot 15 Summary
2
0
Not used
0
example:
stat:oper:cond:level1? → +0
command:
:STATus:OPERation:ENABle:LEVel1
syntax:
:STATus:OPERation:ENABle:LEVel1
description:
Sets the bits in the Operational Status Enable Summary Mask (OSESM) that enable the
contents of the OSESR for slots 15 - 17 of the HP 8166A Lightwave Measurement System
to affect the Status Byte (STB).
Setting a bit in this register to 1 enables the corresponding bit in the OSESR for slots 15 -
17 of the HP 8166A Lightwave Measurement System to affect bit 7 of the Status Byte.
parameters:
The bit value for the OSESM as a 16-bit signed integer value (0 .. +32767)
The default value is 0.
response:
none
example:
stat:oper:enab:level1 128
Status Reporting – The STATus Subsystem
Instrument Setup and Status
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
53
command:
:STATus:OPERation:ENABle:LEVel1?
syntax:
:STATus:OPERation:ENABle:LEVel1?
description:
Returns the OSESM for the OSESR for slots 15 - 17 of the HP 8166A Lightwave Measure-
ment System
parameters:
none
response:
The bit value for the operation enable mask as a 16-bit signed integer value (0 .. +32767)
example:
stat:oper:enab:level1? → +128
command:
:STATusn:OPERation[:EVENt]?
syntax:
:STATusn:OPERation[:EVENt]?
description:
Returns the Operational Slot Status Event Register (OSSER) of slot n.
parameters:
none
response:
The results for the individual slot events (a 16-bit signed integer value, where 0 ≤ value ≤
32767):
Bit
4-15
3
2
1
0
Mnemonic
Not used
Slot n: Zeroing started
Not used
Slot n: Coherence Control has been switched on
Slot n: Laser has been switched on
Decimal Value
0
8
0
2
1
example:
stat1:oper? → +0
command:
:STATusn:OPERation:CONDition?
syntax:
:STATusn:OPERation:CONDition?
description:
Returns the Operational Slot Status Condition Register of slot n.
parameters:
none
response:
The results for the individual slot events (a 16-bit signed integer value, where 0 ≤ value ≤
32767):
Bit
4-15
3
2
1
0
Mnemonic
Not used
Slot n: Zeroing ongoing
Not used
Slot n: Coherence Control is switched on
Slot n: Laser is switched on
Decimal Value
0
8
0
2
1
example:
stat1:oper:cond? → +0
Instrument Setup and Status
Status Reporting – The STATus Subsystem
54
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
:STATusn:OPERation:ENABle
syntax:
:STATusn:OPERation:ENABle
description:
Sets the bits in the Operation Slot Status Enable Mask (OSSEM) for slot n that enable the
contents of the Operation Slot Status Event Register (OSSER) for slot n to affect the OS-
ESR.
Setting a bit in this register to 1 enables the corresponding bit in the OSSER for slot n to
affect bit n of the OSESR.
parameters:
The bit value for the OSSEM as a 16-bit signed integer value (0 .. +32767)
response:
none
example:
stat:oper:enab 128
command:
:STATusn:OPERation:ENABle?
syntax:
:STATusn:OPERation:ENABle?
description:
Returns the OSSEM of slot n
parameters:
none
response:
The bit value for the OSSEM as a 16-bit signed integer value (0 .. +32767)
example:
stat:oper:enab? → +128
command:
:STATus:PRESet
syntax:
:STATus:PRESet
description:
Presets all bits in all the enable masks for both the OPERation and QUEStionable status
systems to 0, that is, OSSEM, QSSEM, OSESM, and QSESM.
parameters:
none
response:
none
example:
stat:pres
Status Reporting – The STATus Subsystem
Instrument Setup and Status
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
55
command:
:STATus:QUEStionable[:EVENt][:LEVel0]?
syntax:
:STATus:QUEStionable[:EVENt][:LEVel0]?
description:
Returns the Questionable Status Event Summary Register (QSESR).
parameters:
none
response:
The sum of the results for the QSESR as a 16-bit signed integer value (0 .. +32767)
Bits
Mnemonics
Decimal
Value
HP 8163A
HP 8164A
HP 8166A
15
Not used
Not used
Not used
0
14
Not used
Not used
Slot 14 Summary
16384
13
Not used
Not used
Slot 13 Summary
8192
12
Not used
Not used
Slot 12 Summary
4096
11
Not used
Not used
Slot 11 Summary
2048
10
Not used
Not used
Slot 10 Summary
1024
9
Not used
Not used
Slot 9 Summary
512
8
Not used
Not used
Slot 8 Summary
256
7
Not used
Not used
Slot 7 Summary
128
6
Not used
Not used
Slot 6 Summary
64
5
Not used
Not used
Slot 5 Summary
32
4
Not used
Slot 4 Summary
Slot 4 Summary
16
3
Not used
Slot 3 Summary
Slot 3 Summary
8
2
Slot 2 Summary
Slot 2 Summary
Slot 2 Summary
4
1
Slot 1 Summary
Slot 1 Summary
Slot 1 Summary
2
0
Not used
Slot 0 Summary
Level 1 Summary
1
example:
stat:ques? → +0
Instrument Setup and Status
Status Reporting – The STATus Subsystem
56
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
:STATus:QUEStionable:CONDition[:LEVel0]?
syntax:
:STATus:QUEStionable:CONDition[:LEVel0]?
description:
Returns the Questionable Status Condition Summary Register.
parameters:
none
response:
The sum of the results for the Questionable Status Condition Summary Register as a 16-
bit signed integer value (0 .. +32767)
Bits
Mnemonics
Decimal
Value
HP 8163A
HP 8164A
HP 8166A
15
Not used
Not used
Not used
0
14
Not used
Not used
Slot 14 Summary
16384
13
Not used
Not used
Slot 13 Summary
8192
12
Not used
Not used
Slot 12 Summary
4096
11
Not used
Not used
Slot 11 Summary
2048
10
Not used
Not used
Slot 10 Summary
1024
9
Not used
Not used
Slot 9 Summary
512
8
Not used
Not used
Slot 8 Summary
256
7
Not used
Not used
Slot 7 Summary
128
6
Not used
Not used
Slot 6 Summary
64
5
Not used
Not used
Slot 5 Summary
32
4
Not used
Slot 4 Summary
Slot 4 Summary
16
3
Not used
Slot 3 Summary
Slot 3 Summary
8
2
Slot 2 Summary
Slot 2 Summary
Slot 2 Summary
4
1
Slot 1 Summary
Slot 1 Summary
Slot 1 Summary
2
0
Not used
Slot 0 Summary
Level 1 Summary
1
example:
stat:ques:cond? → +0
command:
:STATus:QUEStionable:ENABle[:LEVel0]
syntax:
:STATus:QUEStionable:ENABle[:LEVel0]
description:
Sets the bits in the Questionable Status Enable Summary Mask (QSESM) that enable the
contents of the QSESR to affect the Status Byte (STB).
Setting a bit in this register to 1 enables the corresponding bit in the QSESR to affect bit 3
of the Status Byte.
parameters:
The bit value for the questionable enable mask as a 16-bit signed integer value
(0 .. +32767)
The default value is 0.
response:
none
example:
stat:ques:enab 128
command:
:STATus:QUEStionable:ENABle[:LEVel0]?
syntax:
:STATus:QUEStionable:ENABle[:LEVel0]?
description:
Returns the QSESM for the event register
parameters:
none
response:
The bit value for the QSEM as a 16-bit signed integer value (0 .. +32767)
example:
stat:ques:enab? → +128
Status Reporting – The STATus Subsystem
Instrument Setup and Status
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
57
command:
:STATus:QUEStionable[:EVENt]:LEVel1?
syntax:
:STATus:QUEStionable[:EVENt]:LEVel1?
description:
Returns the Questionable Status Event Summary Register (QSESR) for slots 15 to 17 of the
HP 8166A Lightwave Multichannel System.
parameters:
none
response:
The sum of the results for the slots (a 16-bit signed integer value, where 0 ≤ value ≤ 32767):
Bits
Mnemonics
Decimal Value
HP 8166A
15-4
Not used
0
3
Slot 17 Summary
8
2
Slot 16 Summary
4
1
Slot 15 Summary
2
0
Not used
0
example:
stat:ques:level1? → +0
command:
:STATus:QUEStionable:CONDition:LEVel1?
syntax:
:STATus:QUEStionable:CONDition:LEVel1?
description:
Returns the Questionable Status Condition Summary Register for slots 15 to 17 of the
HP 8166A Lightwave Multichannel System.
parameters:
none
response:
The sum of the results for the slots (a 16-bit signed integer value, where 0 ≤ value ≤ 32767):
Bits
Mnemonics
Decimal Value
HP 8166A
15-4
Not used
0
3
Slot 17 Summary
8
2
Slot 16 Summary
4
1
Slot 15 Summary
2
0
Not used
0
example:
stat:ques:cond:level1? → +0
command:
:STATus:QUEStionable:ENABle:LEVel1
syntax:
:STATus:QUEStionable:ENABle:LEVel1
description:
Sets the bits in the Questionable Status Enable Summary Mask (QSESM) that enable the
contents of the QSESR for slots 15 - 17 of the HP 8166A Lightwave Measurement Sys-
tem to affect the Status Byte (STB).
Setting a bit in this register to 1 enables the corresponding bit in the OSESR for slots 15 -
17 of the HP 8166A Lightwave Measurement System to affect bit 7 of the Status Byte.
parameters:
The bit value for the QSESM as a 16-bit signed integer value (0 .. +32767)
The default value is 0.
response:
none
example:
stat:oper:enab:level1 128
Instrument Setup and Status
Status Reporting – The STATus Subsystem
58
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
:STATus:QUEStionable:ENABle:LEVel1?
syntax:
:STATus:QUEStionable:ENABle:LEVel1?
description:
Returns the QSESM for the QSESR for slots 15 - 17 of the HP 8166A Lightwave Mea-
surement System
parameters:
none
response:
The bit value for the QSESM as a 16-bit signed integer value (0 .. +32767)
example:
stat:oper:enab:level1? → +128
command:
:STATusn:QUEStionable[:EVENt]?
syntax:
:STATusn:QUEStionable[:EVENt]?
description:
Returns the questionable status of slot n - the Questionable Slot Status Event Register
(QSSER).
parameters:
none
response:
The results for the individual slot events (a 16-bit signed integer value, where 0 ≤ value ≤
32767):
Bit
8-15
7
6
5
4
3
2
1
0
Mnemonic
Not used
Slot n: Duty cycle has been out of range
Slot n: ARA has been recommended
Slot n: Module has been out of specification
Slot n: Module has settled unsuccessfully
Slot n: Laser protection has been on
Slot n: Temperature has been out of range
Slot n: A Zeroing operation has failed
Slot n: Excessive Value has occurred
Decimal Value
0
128
64
32
16
8
4
2
1
Every nth bit is the summary of slot n.
example:
stat1:oper? → +0
command:
:STATusn:QUEStionable:CONDition?
syntax:
:STATusn:QUEStionable:CONDition?
description:
Returns the Questionable Slot Status Condition Register for slot n.
parameters:
none
response:
The results for the individual slot events (a 16-bit signed integer value, where 0 ≤ value ≤
32767):
Bit
8-15
7
6
5
4
3
2
1
0
Mnemonic
Not used
Slot n: Duty cycle is out of range
Slot n: ARA recommended
Slot n: Module is out of specification
Slot n: Module has not settled
Slot n: Laser protection on
Slot n: Temperature out of range
Slot n: Zeroing failed
Slot n: Excessive Value
Decimal Value
0
128
64
32
16
8
4
2
1
Every nth bit is the summary of slot n.
example:
stat1:ques:cond? → +0
Interface/Instrument Behaviour Settings – The SYSTem Subsystem
Instrument Setup and Status
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
59
Interface/Instrument
Behaviour Settings – The
SYSTem Subsystem
The SYSTem subsystem lets you control the instrument’s serial interface. You can
also control some internal data (like date, time, and so on).
command:
:STATusn:QUEStionable:ENABle
syntax:
:STATusn:QUEStionable:ENABle
description:
Sets the bits in the Questionable Slot Status Enable Mask (QSSEM) for slot n that enable
the contents of the Questionable Slot Status Register (QSSR) for slot n to affect the QS-
ESR.
Setting a bit in this register to 1 enables the corresponding bit in the QSSER for slot n to
affect bit n of the QSESR.
parameters:
The bit value for the QSSEM as a 16-bit signed integer value (0 .. +32767)
response:
none
example:
stat:ques:enab 128
command:
:STATusn:QUEStionable:ENABle?
syntax:
:STATusn:QUEStionable:ENABle?
description:
Returns the QSSEM for slot n
parameters:
none
response:
The bit value for the QSSEM as a 16-bit signed integer value (0 .. +32767)
example:
stat:ques:enab? → +128
command:
:SYSTem:DATE
syntax:
:SYSTem:DATE,,
description:
Sets the instrument’s internal date.
parameters:
• the first value is the year (four digits),
• the second value is the month, and
• the third value is the day.
response:
none
example:
syst:date 1999, 1, 12
Instrument Setup and Status
Interface/Instrument Behaviour Settings – The SYSTem Subsystem
60
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
:SYSTem:DATE?
syntax:
:SYSTem:DATE?
description:
Returns the instrument’s internal date.
parameters:
none
response:
The date in the format year, month, day (16-bit signed integer values)
example:
syst:date? → +1999,+1,+12
command:
:SYSTem:ERRor?
syntax:
:SYSTem:ERRor?
description:
Returns the next error from the error queue (see “The Error Queue” on page 18).
Each error has the error code and a short description of the error, separated by a comma,
for example 0, "No error".
Error codes are numbers in the range -32768 and +32767.
Negative error numbers are defined by the SCPI standard. Positive error numbers are de-
vice dependent.
parameters:
none
response:
The number of the latest error, and its meaning.
example:
syst:err? → -113,"Undefined header"
command:
:SYSTem:HELP:HEADers?
syntax:
:SYSTem:HELP:HEADers?
description:
Returns a list of GPIB commands.
parameters:
none
response:
Returns a list of GPIB commands
example:
syst:help:head? → Returns a list of all GPIB commands
Interface/Instrument Behaviour Settings – The SYSTem Subsystem
Instrument Setup and Status
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
61
command:
:SYSTem:PRESet
syntax:
:SYSTem:PRESet
description:
Sets the mainframe and all installed modules to their standard settings. This command has
the same function as the Preset hardkey.
The following are not affected by this command:
• the GPIB (interface) state,
• the backlight and contrast of the display,
• the interface address,
• the output and error queues,
• the Service Request Enable register (SRE),
• the Status Byte (STB),
• the Standard Event Status Enable Mask (SESEM), and
• the Standard Event Status Register (SESR).
parameters:
none
response:
none
example:
SYST:PRES
command:
:SYSTem:TIME
syntax:
:SYSTem:TIME,,
description:
Sets the instrument’s internal time.
parameters:
• the first value is the hour (0 .. 23),
• the second value is the minute, and
• the third value is the seconds.
response:
none
example:
syst:time 20,15,30
command:
:SYSTem:TIME?
syntax:
:SYSTem:TIME?
description:
Returns the instrument’s internal time.
parameters:
none
response:
The time in the format hour, minute, second. Hours are counted 0...23 (16-bit signed inte-
ger values).
example:
syst:time? → +20,+15,+30
command:
:SYSTem:VERSion?
syntax:
:SYSTem:VERSion?
description:
Returns the SCPI revision to which the instrument complies.
parameters:
none
response:
The revision year and number.
example:
syst:vers? → 1995.0
Instrument Setup and Status
Interface/Instrument Behaviour Settings – The SYSTem Subsystem
62
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
syntax:
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess
description:
Sets the GPIB address.
parameters:
The GPIB Address
response:
none
example:
SYST:COMM:GPIB:ADDR 20
command:
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess?
syntax:
:SYSTem:COMMunicate:GPIB[:SELF]:ADDRess?
description:
Returns the GPIB address.
parameters:
none
response:
The GPIB Address
example:
SYST:COMM:GPIB:ADDR? → +20
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
63
Measurement
Operations & Settings
Measurement Operations & Settings
64
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
This chapter gives descriptions of commands that you can use when you are
setting up or performing measurements. The commands are split up into the
following subsystems:
• Root layer commands that take power measurements, configures triggering,
and return information about the mainframe and it’s slots
• SENSe subsystem commands that control Power Sensors, Optical Head
Interface Modules, and Return Loss Modules.
• SOURce subsystem commands that control Laser Source modules, DFB
source modules, Tunable Laser modules, and Return Loss Modules with
internal laser sources.
• TRIGger subsystem commands that control triggering.
Root Layer Command
Measurement Operations & Settings
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
65
Root Layer Command
The commands in the Slot subsystem allow you to query the following:
• a particular slot, for example, using slot1:empt?,
• or, an Optical Head attached to an Optical Head Interface Module, for
example, an Optical Head Interface Module in slot1 with an Optical Head
attached to channel 2, using slot1:head2:empt?.
command:
:LOCK
syntax:
:LOCK,
description:
Switches the lock off and on.
High power lasers cannot be switched on, if you switch the lock on. High power lasers are
switched off immediately when you switch the lock on.
parameters:
A boolean value:
0 or OFF: switch lock off
1 or ON: switch lock on
is the four-figure lock password.
response:
none
example:
lock 1,1234 - 1234 is the default password
command:
:LOCK?
syntax:
:LOCK?
description:
Returns the current state of the lock.
parameters:
none
response:
A boolean value:
0: lock is switched off
1: lock is switched on
example:
lock? → 1
command:
:SLOT[n]:EMPTy?
syntax:
:SLOT[n]:EMPTy?
description:
Returns whether the module slot is empty.
parameters:
none
response:
A boolean value:
0: there is a module in the slot
1: the module slot is empty
examples:
slot1:empt? → 0
There is a module in slot1
affects:
Independent of module type
Measurement Operations & Settings
Root Layer Command
66
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
:SLOT[n]:IDN?
syntax:
:SLOT[n]:IDN?
description:
Returns information about the module.
parameters:
none
response:
HEWLETT-PACKARD:
mmmm:
ssssssss:
rrrrrrrrrr:
manufacturer
instrument model number (for example 81533B)
serial number
date of firmware revision
example:
slot1:idn? →
HEWLETT-PACKARD, 81533B,3411G06054,07-Aug-98
NOTE
• The HP 81640A/80A/82A/89A Tunable Laser modules will always return
HEWLETT-PACKARD as the manufacturer.
• All other HP 8163A Series modules return Agilent Technologies as the
manufacturer.
• The HP 8153A Series modules will always return HEWLETT-PACKARD as the
manufacturer.
See “*IDN?” on page 46 for information on mainframe identity strings.
affects:
Independent of module type
command:
:SLOT[n]:OPTions?
syntax:
:SLOT[n]:OPTions?
description:
Returns information about a module’s options.
parameters:
none
response:
A string.
example:
slot1:opt? → NO CONNECTOR OPTION, NO INSTRUMENT OPTIONS
affects:
Independent of module type
command:
:SLOT[n]:TST?
syntax:
:SLOT[n]:TST?
description:
Returns the latest selftest results for a module.
NOTE
This command does not perform a selftest. Use selfTeST command, *TST? on page 59, to
perform a selftest.
parameters:
none
response:
Returns an error code and a short description of the error.
example:
slot:tst? → +0,"self test OK"
affects:
Independent of module type
Root Layer Command
Measurement Operations & Settings
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
67
command:
:SLOT[n]:HEAD[n]:EMPTy?
syntax:
:SLOT[n]:HEAD[n]:EMPTy?
description:
Returns whether an optical head is connected.
parameters:
none
response:
A boolean value:
0: there is a module in the slot
1: the module slot is empty
examples:
slot1:head:empt? → 0
An optical head is connected to the optical
head interface module in slot 1
NOTE
• All other HP 8163A Series Optical Heads return Agilent Technologies as the
manufacturer.
• The HP 8153A Series Optical Heads will always return HEWLETT-PACKARD as the
manufacturer.
See “*IDN?” on page 46 for information on mainframe identity strings.
affects:
Optical heads
command:
:SLOT[n]:HEAD[n]:IDN?
syntax:
:SLOT[n]:HEAD[n]:IDN?
description:
Returns information about the optical head.
parameters:
none
response:
HEWLETT-PACKARD:
mmmm:
ssssssss:
rrrrrrrrrr:
manufacturer
instrument model number (for example 81520A)
serial number
date of firmware revision
example:
slot1:head:idn? →
HEWLETT-PACKARD, 81520A,3411G06054,07-Aug-98
affects:
Optical heads
command:
:SLOT[n]:HEAD[m]:OPTions?
syntax:
:SLOT[n]:HEAD[m]:OPTions?
description:
Returns information about an optical head’s options.
parameters:
none
response:
A string.
example:
slot1:head:opt? → NO CONNECTOR OPTION, NO INSTRUMENT
OPTIONS
affects:
Optical heads
Measurement Operations & Settings
Measurement Functions – The SENSe Subsystem
68
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
Measurement Functions – The
SENSe Subsystem
The SENSe subsystem lets you control measurement parameters for a Power
Sensor, an Optical Head Interface module, or a return loss module.
HP 81635A and HP 81619A- Master and
Slave Channels
For the HP 81635A Dual Power Sensor and HP 81619A Dual Optical Head
Interface module, channel 1 is the master channel and channel 2 is the slave
channel. The master and slave channels share the same software and hardware
triggering system. For some commands, setting parameters for the master channel
sets the parameters for the slave channel. In these cases, you may only set
parameters for the slave channel by setting master channel parameters.
The commands listed in Table 5 can only be configured using the master channel.
The commands listed in Table 6 are independent for both master and slave
channels.
command:
:SLOT[n]:HEAD[m]:TST?
syntax:
:SLOT[n]:HEAD[m]:TST?
description:
Returns the latest selftest results for an optical head.
NOTE
This command does not perform a selftest. Use selfTeST command, “*TST?” on
page 49, to perform a selftest.
parameters:
none
response:
Returns an error code and a short description of the error.
example:
slot:head:tst? → +0,"self test OK"
affects:
Optical heads
command:
:SPECial:REBoot
syntax:
:SPECial:REBoot
description:
Reboots the mainframe and all modules.
parameters:
none
response:
none
example:
spec:reb
Measurement Functions – The SENSe Subsystem
Measurement Operations & Settings
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
69
Command
Page
:INITiate[n]:[CHANnel[m]][:IMMediate]
70
:INITiate[n]:[CHANnel[m]]:CONTinuous/?
71
:READ[n][:CHANnel[m]][:SCALar]:POWer[:DC]?
71
:SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO
73
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:LOGGing/?
74
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:MINMax/?
75
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:STABility/?
76
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe/?
77
:SENSe[n]:[CHANnel[m]]:POWer:ATIME/?
78
:SENSe[n]:[CHANnel[m]]:POWer:RANGe:AUTO/?
79
:TRIGger[n][:CHANnel[m]]:INPut/?
110
:TRIGger[n][:CHANnel[m]]:INPut:REARm/?
111
:TRIGger[n][:CHANnel[m]]:OUTPut/?
112
:TRIGger[n][:CHANnel[m]]:OUTPut:REARm/?
112
Table 5
Commands that can only be configured using the master channel
Command
Page
:FETCh[n][:CHANnel[m]][:SCAlar]:POWer[:DC]?
70
:SENSe[n]:[CHANnel[m]]:CORRection[:LOSS][:INPut] [:MAGNitude]/?
72
:SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO?
73
:SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO:ALL
73
:SENSe[n][:CHANnel[m]]:FUNCtion:RESult?
77
:SENSe[n]:[CHANnel[m]]:POWer:RANGe[:UPPer]/?
79
:SENSe[n]:[CHANnel[m]]:POWer:REFerence/?
80
:SENSe[n]:[CHANnel[m]]:POWer:REFerence:DISPlay
81
:SENSe[n]:[CHANnel[m]]:POWer:REFerence:STATe/?
81
:SENSe[n]:[CHANnel[m]]:POWer:REFerence:STATe:RATio/?
82
:SENSe[n]:[CHANnel[m]]:POWer:UNIT/?
82
:SENSe[n]:[CHANnel[m]]:POWer:WAVelength/?
83
Table 6
Commands that are independent for both master and slave channels
Measurement Operations & Settings
Measurement Functions – The SENSe Subsystem
70
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
:FETCh[n][:CHANnel[m]][:SCAlar]:POWer[:DC]?
syntax:
:FETCh[n]:[CHANnel[m]][:SCAlar]:POWer[:DC]?
description:
Reads the current power meter value. It does not provide its own triggering and so must
be used with either continuous software triggering (see
“:INITiate[n]:[CHANnel[m]]:CONTinuous?” on page 71) or a directly preceding imme-
diate software trigger (see “:INITiate[n]:[CHANnel[m]][:IMMediate]” on page 70).
It returns the power meter value the previous software trigger measured. Any subsequent
FETCh command will return the same value, if there is no subsequent software trigger.
parameters:
none
response:
The current power meter value as a float value in dBm,W or dB.
NOTE
If the reference state is absolute, units are dBm or W.
If the reference state is relative, units are dB.
example:
fetc1:pow? → +6.73370400E-04
affects:
All power meters and return loss modules
dual sensors:
Master and slave channels are independent.
command:
:FETCh[n][:CHANnel[m]][:SCAlar]:RETurnloss?
syntax:
:FETCh[n]:[CHANnel[m]][:SCAlar]:RETurnloss?
description:
Reads the current return loss value. It does not provide its own triggering and so must be
used with either continuous software triggering (see
“:INITiate[n]:[CHANnel[m]]:CONTinuous?” on page 71) or a directly preceding imme-
diate software trigger (see “:INITiate[n]:[CHANnel[m]][:IMMediate]” on page 70).
It returns the return loss value the previous software trigger measured. Any subsequent
FETCh command will return the same value, if there is no subsequent software trigger.
parameters:
none
response:
The current power meter value as a float value in dB.
example:
fetc1:ret? → +6.73370400E-00
affects:
All return loss modules
command:
:INITiate[n]:[CHANnel[m]][:IMMediate]
syntax:
:INITiate[n]:[CHANnel[m]][:IMMediate]
description:
Initiates the software trigger system and completes one full trigger cycle, that is, one mea-
surement is made.
parameters:
none
response:
none
example:
init
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel is also affected.
Measurement Functions – The SENSe Subsystem
Measurement Operations & Settings
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
71
command:
:INITiate[n]:[CHANnel[m]]:CONTinuous
syntax:
:INITiate[n]:[CHANnel[m]]:CONTinuous
description:
Sets the software trigger system to continuous measurement mode.
parameters:
A boolean value:
0 or OFF: do not measure continuously
1 or ON: measure continuously
response:
none
example:
init2:cont 1
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel is also affected.
command:
:INITiate[n]:[CHANnel[m]]:CONTinuous?
syntax:
:INITiate[n]:[CHANnel[m]]:CONTinuous?
description:
Queries whether the software trigger system operates continuously or not
parameters:
none
response:
A boolean value:
0 or OFF: measurement is not continuous
1 or ON: measurement is continuous
example:
init2:cont? → 1
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel parameters are identical.
command:
:READ[n][:CHANnel[m]][:SCALar]:POWer[:DC]?
syntax:
:READ[n]:[CHANnel[m]][:SCALar]:POWer[:DC]?
description:
Reads the current power meter value. It provides its own software triggering and does not
need a triggering command.
If the software trigger system operates continuously (see
“:INITiate[n]:[CHANnel[m]]:CONTinuous?” on page 71), this command is identical to
“:FETCh[n][:CHANnel[m]][:SCAlar]:POWer[:DC]?” on page 70.
If the software trigger system does not operate continuously, this command is identical to
generating a software trigger (“:INITiate[n]:[CHANnel[m]][:IMMediate]” on page 70)
and then reading the power meter value.
NOTE
The power meter must be running for this command to be effective.
parameters:
none
response:
The current power meter reading as a float value in dBm, W or dB.
NOTE
If the reference state is absolute, units are dBm or W.
If the reference state is relative, units are dB.
example:
read1:pow? → +1.33555600E-006
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel is also triggered.
To read a simultaneous result from the slave channel, send
“:FETCh[n][:CHANnel[m]][:SCAlar]:POWer[:DC]?” on page 70 directly after this
command.
Measurement Operations & Settings
Measurement Functions – The SENSe Subsystem
72
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
:READ[n][:CHANnel[m]][:SCALar]:RETurnloss?
syntax:
:READ[n]:[CHANnel[m]][:SCALar]:RETurnloss?
description:
Reads the current return loss value. It provides its own software triggering and does not
need a triggering command.
If the software trigger system operates continuously (see
“:INITiate[n]:[CHANnel[m]]:CONTinuous?” on page 71), this command is identical to
“:FETCh[n][:CHANnel[m]][:SCAlar]:RETurnloss?” on page 70.
If the software trigger system does not operate continuously, this command is identical to
generating a software trigger (“:INITiate[n]:[CHANnel[m]][:IMMediate]” on page 70)
and then reading the power meter value.
NOTE
The return loss module must be running for this command to be effective.
parameters:
none
response:
The current power meter reading as a float value in dB.
example:
read1:ret? → +1.33555600E-000
affects:
All return loss modules
command:
:SENSe[n]:[CHANnel[m]]:CORRection[:LOSS][:INPut][:MAGNitude]
syntax:
:SENSe[n]:[CHANnel[m]]:CORRection[:LOSS][:INPUT][:MAGNitude]
[DB|MDB]
description:
Enters a calibration value for a module.
parameters:
The calibration factor as a float value
If no unit type is specified, decibels (dB) is implied.
response:
none
example:
sens1:corr 10DB
affects:
All power meters
dual sensors:
Master and slave channels are independent.
command:
:SENSe[n]:[CHANnel[m]]:CORRection[:LOSS][:INPut][:MAGNitude]?
syntax:
:SENSe[n]:[CHANnel[m]]:CORRection[:LOSS][:INPUT][:MAGNitude]?
description:
Returns the calibration factor for a module.
parameters:
none
response:
The calibration factor as a float value. Units are in dB, although no units are returned in
the response message.
example:
sens1:corr? → +1.00000000E+000
affects:
All power meters
dual sensors:
Master and slave channels are independent.
Measurement Functions – The SENSe Subsystem
Measurement Operations & Settings
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
73
NOTE
Setting parameters for the logging function sets some parameters, including
hidden parameters, for the stability and MinMax functions and vice versa. You
must use the :SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:LOGGing
command:
:SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO
syntax:
:SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO
description:
Zeros the electrical offsets for a power meter or return loss module.
parameters:
none
response:
none
example:
sens1:corr:coll:zero
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel is also zeroed.
command:
:SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO?
syntax:
:SENSe[n]:[CHANnel[m]]:CORREction:COLLect:ZERO?
description:
Returns the status of the most recent zero command.
parameters:
none
response:
0:
any other number:
zero succeeded without errors.
remote zeroing failed (the number is the error code returned
from the operation).
example:
sens1:corr:coll:zero? → 0
affects:
All power meters and return loss modules
dual sensors:
Master and slave channels are independent.
command:
:SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO:ALL
syntax:
SENSe[n]:[CHANnel[m]]:CORRection:COLLect:ZERO:ALL
description:
Zeros the electrical offsets for all installed power meter and return loss modules.
parameters:
none
response:
none
example:
sens:chan:corr:coll:zero:all
affects:
All power meters and return loss modules
dual sensors:
Command is independent of channel.
Measurement Operations & Settings
Measurement Functions – The SENSe Subsystem
74
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command to set parameters before you start a logging function using the
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe command.
NOTE
Setting parameters for the MinMax function sets some parameters, including
hidden parameters, for the stability and logging functions and vice versa. You
must use the :SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:MINMax
command:
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:LOGGing
syntax:
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:LOGGing,
[NS|US|MS|S]
description:
Sets the number of data points and the averaging time for the logging data acquisition
function.
parameters:
Data Points:
Averaging time:
Data Points is the number of samples that are recorded before the
logging mode is completed. Data Points is an integer value.
Averaging time is a time value in seconds.
There is no time delay between averaging time periods. Use
“:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:STABility?”
on page 76 if you want to use delayed measurement.
If you specify no units for the averaging time value in your command, seconds are used as
the default.
NOTE
See “:SENSe[n][:CHANnel[m]]:FUNCtion:STATe” on page 77 for information on start-
ing/stopping a data acquisition function.
NOTE
See “:SENSe[n][:CHANnel[m]]:FUNCtion:RESult?” on page 77 for information on ac-
cessing the results of a data acquisition function.
NOTE
See “Triggering and Power Measurements” on page 108 for information on how trigger-
ing affects data acquisition functions.
response:
none
example:
sens1:func:par:logg 64,1ms
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel is also affected.
command:
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:LOGGing?
syntax:
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:LOGGing?
description:
Returns the number of data points and the averaging time for the logging data acquisition
function.
parameters:
none
response:
Returns the number of data points as an integer value and the averaging time, tavg, as a
float value in seconds.
example:
sens1:func:par:logg? → +64,+1.00000000E-001
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel parameters are identical.
Averaging Time
t
Measurement Running
Measurement Stopped
1
3
5
2
4
6
8
7
9
Measurement Functions – The SENSe Subsystem
Measurement Operations & Settings
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
75
command to set parameters before you start a MinMax function using the
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe command.
NOTE
Setting parameters for the stability function sets some parameters, including
hidden parameters, for the logging and MinMax functions and vice versa. You
must use the :SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:STABility
command:
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:MINMax
syntax:
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:MINMax
CONTinous|WINDow|REFResh,
description:
Sets the MinMax mode and the number of data points for the
MinMax data acquisition function.
parameters:
CONTinous:
WINDow:
REFResh:
continuous MinMax mode
window MinMax mode
refresh MinMax mode
Data Points is the number of samples that are recorded in the memory buffer used by the
WINDow and REFResh modes.
Data Points is an integer value.
See Chapter 3 of the HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measure-
ment System, & HP 8166A Lightwave Multichannel System User’s Guide, for more in-
formation on MinMax mode.
NOTE
See “:SENSe[n][:CHANnel[m]]:FUNCtion:STATe” on page 77 for information on start-
ing/stopping a data acquisition function.
NOTE
See “:SENSe[n][:CHANnel[m]]:FUNCtion:RESult?” on page 77 for information on ac-
cessing the results of a data acquisition function.
NOTE
See “Triggering and Power Measurements” on page 108 for information on how trigger-
ing affects data acquisition functions.
response:
none
example:
sens1:func:par:minm WIND,10
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel is also affected.
command:
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:MINMax?
syntax:
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:MINMax?
description:
Returns the MinMax mode and the number of data points for the MinMax data acquisition
function.
parameters:
none
response:
CONT:
WIND:
REFR:
continuous MinMax mode
window MinMax mode
refresh MinMax mode
The number of data points is returned as an integer value.
example:
sens1:func:par:minm? → WIND,+10
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel parameters are identical.
Measurement Operations & Settings
Measurement Functions – The SENSe Subsystem
76
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command to set parameters before you start a stability function using the
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe command.
command:
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:STABility
syntax:
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:STABility
[NS|US|MS|S],[NS|US|MS|S],[NS|US|MS|S]
description:
Sets the total time, period time, and averaging time for the stability data acquisition function.
parameters:
Total time:
Period time:
Averaging time:
The total time from the start of stability mode until it is completed.
A new measurement is started after the completion of every period time.
A measurement is averaged over the averaging time.
NOTE
The total time should be longer than the period time.
The period time should be longer than the averaging time.
The number of data points is equal to the total time divided by the period time.
Total time, period time, and averaging time are time values in seconds.
If you specify no units in your command, seconds are used as the default.
NOTE
See “:SENSe[n][:CHANnel[m]]:FUNCtion:STATe” on page 77 for information on start-
ing/stopping a data acquisition function.
NOTE
See “:SENSe[n][:CHANnel[m]]:FUNCtion:RESult?” on page 77 for information on ac-
cessing the results of a data acquisition function.
NOTE
See “Triggering and Power Measurements” on page 108 for information on how triggering
affects data acquisition functions.
response:
none
example:
sens1:func:par:stab 1s,0.1s,0.1s
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel is also affected.
command:
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:STABility?
syntax:
:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:STABility?
description:
Returns the total time, period time, and averaging time for the stability data acquisition
function.
parameters:
none
response:
Total time, delay time, and averaging time are float values in seconds.
example:
sens1:func:par:stab? → +1.00000000E+000,
+1.00000000E-001,+1.00000000E-001
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel parameters are identical.
Averaging Time
Period Time
t
Measurement Running
Measurement Stopped
1
2
3
4
5
Measurement Functions – The SENSe Subsystem
Measurement Operations & Settings
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
77
command:
:SENSe[n][:CHANnel[m]]:FUNCtion:RESult?
syntax:
:SENSe[n][:CHANnel[m]]:FUNCtion:RESult?
description:
Returns the data array of the last data acquisition function.
parameters:
none
response:
The last data acquisition function’s data array as a binary block, one measurement value
is a 4-byte-long float in Intel byte order.
example:
sens1:func:res? → returns a data array
affects:
All power meters and return loss modules
dual sensors:
Master and slave channels are independent.
command:
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe
syntax:
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe
LOGGing|STABility|MINMax,STOP|STARt
description:
Enables/Disables the logging, MinMax, or stability data acquisition function mode.
parameters:
LOGGing:
STABility:
MINMax:
STOP:
STARt:
Logging data acquisition function
Stability data acquisition function
MinMax data acquisition function
Stop data acquisition function
Start data acquisition function
NOTE
When you enable a logging data acquisition function for a HP 8163A Series Power Meter
with averaging time of less than 100 ms with input hardware triggering disabled, all GPIB
commands will be ignored for the duration of the function.
See “:SENSe[n][:CHANnel[m]]:FUNCtion:PARameter:LOGGing” on page 74 for more
information on the logging data acquisition function.
NOTE
Stop any function before you try to set up a new function. Some parameters cannot be set
until you stop the function.
response:
none
example:
sens1:func:stat logg,star
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel is also affected.
command:
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe?
syntax:
:SENSe[n][:CHANnel[m]]:FUNCtion:STATe?
description:
Returns the function mode and the status of the data acquisition function.
parameters:
none
response:
NONE
LOGGING_STABILITY
MINMAX
PROGRESS
COMPLETE
No function mode selected
Logging or stability data acquisition function
MinMax data acquisition function
Data acquisition function is in progress
Data acquisition function is complete
example:
sens1:func:stat? → LOGGING_STABILITY,COMPLETE
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel parameters are identical.
Measurement Operations & Settings
Measurement Functions – The SENSe Subsystem
78
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
:SENSe[n][:CHANnel[m]]:FUNCtion:THReshold
syntax:
:SENSe[n][:CHANnel[m]]:FUNCtion:THReshold,
[PW|NW|UW|MW|Watt|DBM]
description:
Sets the start mode and the threshold value.
parameters:
ABOVe:
BELow:
IMMediately:
Threshold Value:
Function starts when power is above the threshold value.
Function starts when power is below the threshold value.
Function starts immediately.
A float value in Watts or dBm.
response:
none
example:
sens1:func:thr IMM,20nw
affects:
All HP 8153A Lightwave Multimeter series power meters and the HP 81534A Return
Loss module
command:
:SENSe[n][:CHANnel[m]]:FUNCtion:THReshold?
syntax:
:SENSe[n][:CHANnel[m]]:FUNCtion:THReshold?
description:
Returns the start mode and the threshold value.
parameters:
none
response:
ABOV:
BEL:
IMM:
Threshold Value:
Function starts when power is above the threshold value.
Function starts when power is below the threshold value.
Function starts immediately.
A float value in Watts or dBm.
example:
sens1:func:thr? → IMM,+2.00000000E-008
affects:
All HP 8153A Lightwave Multimeter series power meters and the HP 81534A Return
Loss module
command:
:SENSe[n]:[CHANnel[m]]:POWer:ATIMe
syntax:
:SENSe[n]:[CHANnel[m]]:POWer:ATIMe[NS|US|MS|S]
description:
Sets the averaging time for the module.
parameters:
The averaging time as a float value in seconds.
If you specify no units in your command, seconds are used as the default.
response:
none
example:
sens1:pow:atim 1s
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel is also affected.
command:
:SENSe[n]:[CHANnel[m]]:POWer:ATIMe?
syntax:
:SENSe[n]:[CHANnel[m]]:POWer:ATIMe?
description:
Returns the averaging time for the module.
parameters:
none
response:
The averaging time as a float value in seconds.
example:
sens1:pow:atim? → +1.00000000E+000
affects:
All power meters and return loss modules
dual sensors:
Can only be sent to master channel, slave channel parameters are identical.
Measurement Functions – The SENSe Subsystem
Measurement Operations & Settings
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
79
command:
:SENSe[n]:[CHANnel[m]]:POWer:RANGe[:UPPer]
syntax:
:SENSe[n]:[CHANnel[m]]:POWer:RANGe[:UPPer][DBM]
description:
Sets the power range for the module.
The range changes at 10 dBm intervals. The corresponding ranges for linear measure-
ments (measurements in Watts) is given below:
Range
+30 dBm
+20 dBm
+10 dBm
0 dBm
−10 dBm
−20 dBm
−30 dBm
−40 dBm
Upper Linear
Power Limit
1999.9 mW
199.99 mW
19.999 mW
1999.9 µW
199.99 µW
19.999 µW
1999.9 nW
199.99 nW
Range
−50 dBm
−60 dBm
−70 dBm
−80 dBm
−90 dBm
−100 dBm
−110 dBm
Upper Linear
Power Limit
19.999 nW
1999.9 pW
199.99 pW
19.999 pW
1.999 pW
0.199 pW
0.019 pW
parameters:
The range as a float value in dBm. The number is rounded to the closest multiple of 10,
because the range changes at 10 dBm intervals. Units are in dBm.
response:
none
example:
sens1:pow:rang -20DBM
affects:
All power meters
dual sensors:
Master and slave channels are independent.
command:
:SENSe[n]:[CHANnel[m]]:POWer:RANGe[:UPPer]?
syntax:
:SENSe[n]:[CHANnel[m]]:POWer:RANGe[:UPPer]?
description:
Returns the range setting for the module
parameters:
none
response:
The range setting as a float value in dBm
(−110 Š≤ value ≤ +30).
example:
sens1:pow:rang? → -2.00000000E+001
affects:
All power meters
dual sensors:
Master and slave channels are independent.
command:
:SENSe[n]:[CHANnel[m]]:POWer:RANGe:AUTO
syntax:
SENSe[n]:[CHANnel[m]]:POWer:RANGe:AUTO
description:
Enables or disables automatic power ranging for the module.
If automatic power ranging is enabled, ranging is automatically determined by the instru-
ment. Otherwise, it must be set by the sensn:pow:rang command.
parameters:
A boolean value:
0 or OFF: automatic ranging disabled
1 or ON: automatic ranging enabled
response:
none
example:
sens1:pow:rang:auto 1
affects:
All power meters
dual sensors:
Can only be sent to master channel, slave channel is also affected.
Measurement Operations & Settings
Measurement Functions – The SENSe Subsystem
80
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
:SENSe[n]:[CHANnel[m]]:POWer:RANGe:AUTO?
syntax:
:SENSe[n]:[CHANnel[m]]:POWer:RANGe:AUTO?
description:
Returns whether automatic power ranging is being used by the module.
parameters:
none
response:
A boolean value:
0: automatic ranging is not being used.
1: automatic ranging is being used.
example:
sens1:pow:rang:auto? → 1
affects:
All power meters
dual sensors:
Can only be sent to master channel, slave channel parameters are identical.
command:
:SENSe[n]:[CHANnel[m]]:POWer:REFerence
syntax:
:SENSe[n]:[CHANnel[m]]:POWer:REFerence
TOMODule|TOREF,PW|NW|UW|MW|Watt|DBM|DB|MDB
description:
Sets the sensor reference value.
parameters:
TOMODule:
TOREF:
Sets the reference value in dB used if you choose measurement rela-
tive to another channel
Sets the reference value in Watts or dBm if you choose measurement
relative to a constant reference value
The reference as a float value.
NOTE
You must append a unit type
• dB if you use TOMODule or
• Watts or dBm if you use TOREF.
NOTE
The two reference values are completely independent. When you change the reference
mode using the command
“:SENSe[n]:[CHANnel[m]]:POWer:REFerence:STATe:RATio?” on page 82, the
instrument uses the last reference value entered for the selected reference mode.
response:
none
example:
sens1:pow:ref tomod,-40DB
affects:
All power meters
dual sensors:
Master and slave channels are independent.
command:
:SENSe[n]:[CHANnel[m]]:POWer:REFerence?
syntax:
:SENSe[n]:[CHANnel[m]]:POWer:REFerence?TOMODule|TOREF
description:
Returns the sensor reference value.
parameters:
TOMODule:
TOREF:
Returns the reference value in dB used if you choose measurement
relative to another channel
Returns the reference value in Watts or dBm if you choose measure-
ment relative to a constant reference value
response:
The reference as a float value.
example:
sens1:pow:ref? toref → +1.00000000E-006
affects:
All power meters
dual sensors:
Master and slave channels are independent.
Measurement Functions – The SENSe Subsystem
Measurement Operations & Settings
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
81
command:
:SENSe[n]:[CHANnel[m]]:POWer:REFerence:DISPlay
syntax:
:SENSe[n]:[CHANnel[m]]:POWer:REFerence:DISPlay
description:
Takes the input power level value as the reference value.
parameters:
none
response:
none
example:
sens1:pow:ref:disp
affects:
All power meters
dual sensors:
Master and slave channels are independent.
command:
:SENSe[n]:[CHANnel[m]]:POWer:REFerence:STATe
syntax:
:SENSe[n]:[CHANnel[m]]POWer:REFerence:STATe
description:
Sets the measurement units to relative or absolute units.
parameters:
A boolean value:
0 or OFF: absolute
1 or ON: relative
response:
none
example:
sens1:pow:ref:stat 1
affects:
All power meters
dual sensors:
Master and slave channels are independent.
command:
:SENSe[n]:[CHANnel[m]]:POWer:REFerence:STATe?
syntax:
:SENSe[n]:[CHANnel[m]]POWer:REFerence:STATe?
description:
Inquires whether the current measurement units are relative (dB) or absolute (Watts or
dBm).
parameters:
none
response:
A boolean value:
0: absolute
1: relative
example:
sens1:pow:ref:stat? → 1
affects:
All power meters
dual sensors:
Master and slave channels are independent.
Measurement Operations & Settings
Measurement Functions – The SENSe Subsystem
82
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
:SENSe[n]:[CHANnel[m]]:POWer:REFerence:STATe:RATio
syntax:
:SENSe[n]:[CHANnel[m]]POWer:REFerence:STATe:RATio
|255|TOREF,
description:
Selects the reference for the module.
parameters:
slot number:
255 or TOREF:
channel number:
an integer value representing the slot number you want to reference
results are displayed relative to an absolute reference
an integer value representing the channel number you want to ref-
erence
NOTE
If you want to reference another power sensor channel, use an integer value
corresponding to the slot for the first parameter and an integer value corresponding to the
channel for the second value.
If you want to use an absolute reference, use TOREF as the first parameter and any
integer value as the second parameter.
response:
none
examples:
sens1:pow:ref:stat:rat 2,1
sens1:pow:ref:stat:rat TOREF,1
References channel 2.1
References an absolute reference
affects:
All power meters
dual sensors:
Master and slave channels are independent.
command:
:SENSe[n]:[CHANnel[m]]:POWer:REFerence:STATe:RATio?
syntax:
:SENSe[n]:[CHANnel[m]]POWer:REFerence:STATe:RATio?
description:
Returns the reference setting for the module.
parameters:
none
response:
results are displayed relative to an absolute reference or to the current power reading from
another channel.
examples:
sens1:pow:ref:stat:rat? → +255,+0
sens1:pow:ref:stat:rat? → +2,+1
results are displayed relative
to an absolute reference
results are displayed relative
to channel 2.1
affects:
All power meters
dual sensors:
Master and slave channels are independent.
command:
:SENSe[n]:[CHANnel[m]]:POWer:UNIT
syntax:
:SENSe[n]:[CHANnel[m]]:POWer:UNITDBM|0|Watt|1
description:
Sets the sensor power unit
parameters:
An integer value:
0: dBm
1: Watt
or DBM or Watt
response:
none
example:
sens1:pow:unit 1
affects:
All power meters
dual sensors:
Master and slave channels are independent.
Measurement Functions – The SENSe Subsystem
Measurement Operations & Settings
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
83
command:
:SENSe[n]:[CHANnel[m]]:POWer:UNIT?
syntax:
:SENSe[n]:[CHANnel[m]]:POWer:UNIT?
description:
Inquires the current sensor power unit
parameters:
none
response:
An integer value:
0: Current power units are dBm.
1: Current power units are Watts.
example:
sens1:pow:unit? → +1
affects:
All power meters
dual sensors:
Master and slave channels are independent.
command:
:SENSe[n]:[CHANnel[m]]:POWer:WAVelength
syntax:
:SENSe[n]:[CHANnel[m]]:POWer:WAVelength|MIN|MAX|DEF
[PM|NM|UM|MM|M]
description:
Sets the sensor wavelength.
parameters:
The wavelength as a float value in meters.
Also allowed are:
MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is
half the sum of, the minimum programmable value and
the maximum programmable value
response:
none
example:
sens1:pow:wav 1550nm
affects:
All power meters and return loss modules
dual sensors:
Master and slave channels are independent.
command:
:SENSe[n]:[CHANnel[m]]:POWer:WAVelength?
syntax:
:SENSe[n]:[CHANnel[m]]:POWer:WAVelength?[MIN|MAX|DEF]
description:
Inquires the current sensor wavelength.
parameters:
none
Also allowed are:
MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is
half the sum of, the minimum programmable value and
the maximum programmable value
response:
The wavelength as a float value in meters.
example
sens1:pow:wav? → +1.55000000E-006
affects:
All power meters and return loss modules
dual sensors:
Master and slave channels are independent.
Measurement Operations & Settings
Measurement Functions – The SENSe Subsystem
84
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:FACTory
syntax:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:FACTory
description:
Overwrites the current calibration values with the factory-set calibration settings. See
“:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:REFLectance” on
page 84 and
“:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:TERMination” on
page 84 for information on calibrating your return loss module.
parameters:
none
response:
none
example
sens1:ret:cal:fact
affects:
All return loss modules
command:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:REFLectance
syntax:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:REFLectance
description:
Start the calibration and save the calibration values for a defined reflectance reference
measurement. See
“:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l]” on page 85 for
information on setting the return loss value of your reference reflector.
parameters:
none
response:
none
example
sens1:ret:cal:coll:refl
affects:
All return loss modules
command:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:TERMination
syntax:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CALibration:COLLect:TERMination
description:
Start the calibration and save the calibration values for a defined termination reference
measurement. See
“:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l]” on page 85 for
information on setting the return loss value of your reference reflector.
parameters:
none
response:
none
example
sens1:ret:cal:coll:term
affects:
All return loss modules
Measurement Functions – The SENSe Subsystem
Measurement Operations & Settings
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
85
command:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:FPDelta[l]
syntax:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:FPDelta[l][dB]
description:
Sets the front panel delta, that is, the loss correction value, for example, due to the front
panel connector. Twice this value is added to the measured Return Loss.
NOTE
Use [l] to set the front panel delta for an external source or the upper or lower wavelength
laser source of a dual return loss module.
An external laser source is denoted by 0. 0 is the default value of [l].
A lower wavelength source is denoted by 1.
An upper wavelength source is denoted by 2.
parameters:
Sets the front panel delta as a float value in dB
response:
none
example
sens1:ret:cal:corr:fpd 0.08DB
affects:
All return loss modules
command:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:FPDelta[l]?
syntax:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:FPDelta[l]?
description:
Returns the front panel delta, that is, the loss correction value, for example, due to the
front panel connector. Twice this value is added to the measured Return Loss.
NOTE
Use [l] to query the front panel delta for an external source or the upper or lower wave-
length laser source of a dual return loss module.
An external laser source is denoted by 0. 0 is the default value.
A lower wavelength source is denoted by 1.
An upper wavelength source is denoted by 2.
parameters:
Returns the front panel delta as a float value in dB
response:
none
example
sens1:ret:cal:corr:fpd? → +8.00000000E-002
affects:
All return loss modules
command:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l]
syntax:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l][dB]
description:
Sets the Return Loss Reference, the return loss value of your reference reflector.
For example, the HP 81000BR reference reflector provides an accurate and stable 0.18 dB
reference.
NOTE
Use [l] to set the return loss value of your reference reflector for an external source or the
upper or lower wavelength laser source of a dual return loss module.
An external laser source is denoted by 0. 0 is the default value of [l].
A lower wavelength source is denoted by 1.
An upper wavelength source is denoted by 2.
parameters:
Sets the Return Loss Reference as a float value in dB
response:
none
example
sens1:ret:cal:corr:refl 0.18DB
affects:
All return loss modules
Measurement Operations & Settings
Signal Generation – The SOURce Subsystem
86
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
Signal Generation – The
SOURce Subsystem
The SOURce subsystem allows you to control a laser source module, DFB source
module, tunable laser module, or a return loss module that has an internal source.
command:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l]?
syntax:
:SENSe[n]:[CHANnel[m]]:RETurnloss:CORRection:REFLectance[l]?
description:
Returns the Return Loss Reference, the return loss value of your reference reflector.
For example, the HP 81000BR reference reflector provides an accurate and stable 0.18 dB
reference.
NOTE
Use [l] to query the return loss value of your reference reflector for an external source or
the upper or lower wavelength laser source of a dual return loss module.
An external laser source is denoted by 0. 0 is the default value of [l].
A lower wavelength source is denoted by 1.
An upper wavelength source is denoted by 2.
parameters:
none
response:
Returns the Return Loss Reference as a float value in dB
example
sens1:ret:cal:corr:refl? → +1.80000000E-001
affects:
All return loss modules
command:
:OUTPut[n][:CHANnel[m]]:CONNection
syntax:
OUTPut[n][:CHANnel[m]]:CONNectionMOD|VPP
description:
Sets the analog output parameter.
parameters:
MOD:
VPP:
The modulation frequency modulates the analog output.
Output Voltage is proportional to optical power.
response:
none
example:
outp1:conn mod
affects:
All tunable laser modules
command:
:OUTPut[n][:CHANnel[m]]:CONNection?
syntax:
OUTPut[n][:CHANnel[m]]:CONNection?
description:
Returns the analog output parameter.
parameters:
none
response:
MOD:
VPP:
The modulation frequency modulates the analog output.
Output Voltage is proportional to optical power.
example:
outp1:conn? → MOD
affects:
All tunable laser modules
Signal Generation – The SOURce Subsystem
Measurement Operations & Settings
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
87
command:
:OUTPut[n][:CHANnel[m]]:PATH
syntax:
:OUTPut[n][:CHANnel[m]]:PATH
description:
Sets the regulated path.
parameters:
HIGHpower:
LOWSse:
BHRegulated:
BLRegulated:
The High Power output is regulated.
The Low SSE output is regulated.
Both outputs are active but only the High Power output is Regulated.
Both outputs are active but only the Low SSE output is Regulated.
response:
none
example:
output1:path high
affects:
All tunable laser modules
command:
:OUTPut[n][:CHANnel[m]]:PATH?
syntax:
:OUTPut[n][:CHANnel[m]]:PATH?
description:
Returns the regulated path.
parameters:
none
response:
HIGH:
LOWS:
BHR:
BLR:
The High Power output is regulated.
The Low SSE output is regulated.
Both outputs are active but only the High Power output is Regulated.
Both outputs are active but only the Low SSE output is Regulated.
example:
output1:path? → HIGH
affects:
All tunable laser modules
command:
:OUTPut[n][:CHANnel[m]][:STATe]
syntax:
:OUTPut[n][:CHANnel[m]][:STATe]OFF|ON|0|1
description:
Switches the laser current off and on.
The laser emits light only when the current is on. Set the state to OFF or 0 to switch the
laser current off. Set the state to ON or 1 to switch the laser current on. The default is for
the laser current to be off.
parameters:
0 or OFF:
1 or ON:
switch laser current off
switch laser current on
response:
none
example:
outp 1
affects:
All laser source, DFB source, tunable laser modules and return loss modules with an in-
ternal source
command:
:OUTPut[n][:CHANnel[m]][:STATe]?
syntax:
:OUTPut[n][:CHANnel[m]][:STATe]?
description:
Returns the current state of the laser current.
parameters:
none
response:
A boolean value:
0 – laser current off
1 – laser current on
example:
outp? → 1
affects:
All laser source, DFB source, tunable laser modules and return loss modules with an inter-
nal source
Measurement Operations & Settings
Signal Generation – The SOURce Subsystem
88
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
[:SOURce[n]][:CHANnel[m]]:AM[:INTernal]:FREQuency[l]
syntax:
[:SOURce[n]][:CHANnel[m]]:AM[:INTernal]:FREQuency[l]
[THZ|GHZ|MHZ|KHZ|HZ]
description:
Sets the frequency of the amplitude modulation of the laser output.
parameters:
The frequency as a float value in Hz.
Also allowed are: MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half the sum of,
the minimum programmable value and the maximum programmable
value
The default units are HZ, although KHZ, MHZ, GHZ, and THZ can also be specified.
The resolution of the frequency is always 1 Hz.
NOTE
Use [l] to set the modulation frequency of the upper or lower wavelength laser source of a
dual-wavelength laser source or a return loss module with an internal dual-wavelength la-
ser source. The default value of [l] is 1, the lower wavelength source. The upper wave-
length source is denoted by 2.
response:
none
example:
sour2:am:freq 270hz
affects:
All laser source, DFB source, and tunable laser modules
command:
[:SOURce[n]][:CHANnel[m]]:AM[:INTernal]:FREQuency[l]?
syntax:
[:SOURce[n]][:CHANnel[m]]:AM[:INTernal]:FREQuency[l]? [MIN|DEF|MAX]
description:
Returns the frequency of the amplitude modulation as a float value in Hertz.
parameters:
MIN: minimum modulation frequency
MAX: maximum modulation frequency
DEF: This is not the preset (*RST) default value but is half the sum of, the minimum
modulation frequency and the maximum modulation frequency
NOTE
Use [l] to query the modulation frequency of the upper or lower wavelength laser source
of a dual-wavelength laser source or a return loss module with an internal dual-wave-
length laser source. The default value of [l] is 1, the lower wavelength source. The upper
wavelength source is denoted by 2.
response:
modulation frequency relevant to the current value or specified parameter (if MIN, MAX,
or DEF are chosen as a parameter).
example:
sour2:am:freq? min → +2.00000000E+002
All laser source, DFB source, and tunable laser modules
Signal Generation – The SOURce Subsystem
Measurement Operations & Settings
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
89
command:
[:SOURce[n]][:CHANnel[m]]:AM:SOURce[l]
syntax:
[:SOURce[n]][:CHANnel[m]]:AM:SOURce[l] INT|INT1|INT2|EXT|0|1|2
description:
Selects the type or source of the modulation of the laser output.
parameters:
0, INT1, or INTernal:
1, COHCtrl, or INT2:
2, AEXTernal, or EXT:
3 or DEXTernal:
4 or LFCohctrl:
5 or WVLLocking:
6 or BACKplane:
internal digital modulation
coherence control
external analog modulation
external digital modulation
low frequency coherence control
wavelength locking
external digital modulation using Input Trigger Connector
NOTE
Use [l] to set the modulation source of the upper or lower wavelength laser source of a
dual-wavelength laser source or a return loss module with an internal dual-wavelength la-
ser source. The default value of [l] is 1, the lower wavelength source. The upper wave-
length source is denoted by 2.
response:
none
example:
sour2:am:sour int
affects:
All laser source, DFB source, and tunable laser modules can use internal digital modula-
tion
All other modulation modes are only available with tunable laser modules.
command:
[:SOURce[n]][:CHANnel[m]]:AM:SOURce[l]?
syntax:
[:SOURce[n]][:CHANnel[m]]:AM:SOURce[l]?
description:
Returns the type or source of the modulation of the laser output.
parameters:
0:
1:
2:
3:
4:
5:
6:
internal digital modulation
coherence control
external analog modulation
external digital modulation
low frequency coherence control
wavelength locking
external digital modulation using Input Trigger Connector
NOTE
Use [l] to query the modulation source of the upper or lower wavelength laser source of a
dual-wavelength laser source or a return loss module with an internal dual-wavelength la-
ser source. The default value of [l] is 1, the lower wavelength source. The upper wave-
length source is denoted by 2.
response:
none
example:
sour2:am:sour? → +0
affects:
All laser source, DFB source, and tunable laser modules can use internal digital modula-
tion
All other modulation modes are only available with tunable laser modules.
Measurement Operations & Settings
Signal Generation – The SOURce Subsystem
90
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
[:SOURce[n]][:CHANnel[m]]:AM:STATe[l]
syntax:
[:SOURce[n]][:CHANnel[m]]:AM:STATe[l] OFF|ON|0|1
description:
Enables and disables amplitude modulation of the laser output.
parameters:
A boolean value:
OFF or 0: modulation disabled (default)
ON or 1: modulation enabled.
NOTE
Use [l] to enable/disable amplitude modulation for the upper or lower wavelength laser
source of a dual-wavelength laser source or a return loss module with an internal dual-
wavelength laser source. The default value of [l] is 1, the lower wavelength source. The
upper wavelength source is denoted by 2.
NOTE
When the internal modulation is selected, the Modulation Output on the front panel
outputs a version of the modulating signal that has the same frequency and phase as the
modulating signal, but has a fixed, TTL-level amplitude. You can use this to synchronize
your external measuring equipment to your instrument.
To allow for your possible synchronization requirements, there are two ways in which the
signal can be output. Either the signal is combined with the laser-ready signal, so that the
output is kept low when there is no optical signal being output (for example, while the
laser is settling after a change of wavelength). Or the modulation signal is output all the
time. This is set by the :SOURCE:MODOUT command (see
“[:SOURce[n]][:CHANnel[m]]:MODout” on page 91).
NOTE
When you enable lambda logging, see
“[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:LLOGging” on page 103, and
modulation simultaneously, a sweep can not be started, see
“[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:[STATe]” on page 106.
response:
none
example:
sour2:am:stat 0
affects:
All laser source, DFB source, and tunable laser modules
command:
[:SOURce[n]][:CHANnel[m]]:AM:STATe[l]?
syntax:
[:SOURce[n]][:CHANnel[m]]:AM:STATe[l]?
description:
Returns the current state of modulation.
NOTE
Use [l] to query the current state of modulation of the upper or lower wavelength laser
source of a dual-wavelength laser source or a return loss module with an internal dual-
wavelength laser source. The default value of [l] is 1, the lower wavelength source. The
upper wavelength source is denoted by 2.
parameters:
none
response:
A boolean value:
0: modulation is disabled
1: modulation is enabled
example:
sour2:am:stat? → 0
affects:
All laser source, DFB source, and tunable laser modules
Signal Generation – The SOURce Subsystem
Measurement Operations & Settings
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
91
command:
[:SOURce[n]][:CHANnel[m]]:MODout
syntax:
[:SOURce[n]][:CHANnel[m]]:MODoutFRQ|FRQRDY|0|1
description:
Sets the modulation output
parameters:
FRQ or 0:
FRQRDY or 1:
modulation signal is output all the time
modulation is combined with the laser-ready signal.
In this case, the output is kept low when no optical signal is output
(for example, while the laser is settling after a change of wavelength).
response:
none
example:
sour2:mod 0
affects:
All tunable laser modules
command:
[:SOURce[n]][:CHANnel[m]]:MODout?
syntax:
[:SOURce[n]][:CHANnel[m]]:MODout?
description:
Returns the mode of the modulation output.
parameters:
none
response:
0:
1:
modulation signal is output all the time
modulation is combined with the laser-ready signal.
In this case, the output is kept low when no optical signal is output (for example,
while the laser is settling after a change of wavelength).
example:
sour2:mod? → 0
affects:
All tunable laser modules
Measurement Operations & Settings
Signal Generation – The SOURce Subsystem
92
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]
syntax:
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l][DB|MDB]
description:
Sets the level of attenuation.
parameters:
Any value in the specified range (see the specifications in the appropriate User’s Guide).
Also allowed
(for tunable la-
ser modules
only) are:
MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half the sum of,
the minimum programmable value and the maximum programmable
value
NOTE
Use [l] to set the attenuation level of the upper or lower wavelength laser source of a dual-
wavelength laser source or of a return loss module with an internal dual-wavelength laser
source. The default value of [l] is 1, the lower wavelength source. The upper wavelength
source is denoted by 2.
NOTE
Tunable laser modules with in-built optical attenuators need to be in Manual Attenuation
Mode (see “[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation:AUTO” on page 93)
for this value to have an affect. The output power is a combination of this value and the
laser output power (see
“[:SOURce[n]][:CHANnel[m]]:POWer[:LEVel][:IMMediate][:AMPLitude]” on
page 94).
NOTE
In this respect, this command does not conform to the SCPI standard. The SCPI standard
requires that entering an explicit value for the attenuation switches the attenuation mode
OFF.
The default units are dB.
response:
none
example:
sour2:pow:att 22.32db
affects:
All tunable laser modules with an in-built optical attenuator, all laser source modules, and
return loss modules with an internal source
command:
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]?
syntax:
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]?[MIN|DEF|MAX]
description:
Returns the attenuation level.
When using a tunable laser module with an in-built optical attenuator, the value returned
applies only to the attenuation mode (see
“[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation:AUTO” on page 93).
NOTE
Use [l] to query the attenuation level of the upper or lower wavelength laser source of a
dual-wavelength laser source or of a return loss module with an internal dual-wavelength
laser source. The default value of [l] is 1, the lower wavelength source. The upper wave-
length source is denoted by 2.
parameters:
Also allowed (for
tunable laser mod-
ules only) are:
MIN: minimum amplitude level
MAX: maximum amplitude level
DEF: This is not the preset (*RST) default value but is half the sum
of, the minimum amplitude level and the maximum amplitude level
response:
attenuation level relevant to the current value or specified parameter (if MIN, MAX, or
DEF are chosen as a parameter).
example:
sour2:pow:att? def → +3.10000000+E001
affects:
All tunable laser modules with an in-built optical attenuator, all laser source modules, and
return loss modules with an internal source
Signal Generation – The SOURce Subsystem
Measurement Operations & Settings
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
93
command:
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation:AUTO
syntax:
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation:AUTOOFF|ON|0|1
description:
Selects Automatic or Manual Attenuation Mode.
In Automatic Attenuation Mode, you specify the output power.
In Manual Attenuation Mode, you must specify both the laser output power, and the atten-
uation level.
parameters:
OFF or 0:
ON or 1:
Attenuation Mode
Power Mode
response:
none
example:
sour2:pow:att:auto 1
affects:
All tunable laser modules with an in-built optical attenuator
command:
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation:AUTO?
syntax:
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation:AUTO?
description:
Returns whether the instrument is in Automatic or Manual Attenuation Mode.
parameters:
none
response:
0:
1:
Manual Attenuation Mode
Automatic Attenuation Mode
example:
sour2:pow:att:auto? → 1
affects:
All tunable laser modules with an in-built optical attenuator
command:
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation:DARK
syntax:
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation:DARKOFF|ON|0|1
description:
Sets or unsets the attenuator to ‘dark’ position.
Dark position blocks all light from the laser. You can use this as an alternative to dis-
abling the laser, the advantage of doing this is that you avoid the laser rise time.
This command is available in Attenuation Mode Only.
parameters:
OFF or 0:
ON or 1:
Unsets dark position
Sets dark position
response:
none
example:
sour2:pow:att:dark 1
affects:
All tunable laser modules with an in-built optical attenuator
command:
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation:DARK?
syntax:
[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation:DARK?
description:
Queries whether the attenuator is set to ‘dark’ position (where all light is blocked by the
laser).
parameters:
none
response:
0:
1:
dark position not set
dark position set
example:
sour2:pow:att:dark? → 1
affects:
All tunable laser modules with an in-built optical attenuator
Measurement Operations & Settings
Signal Generation – The SOURce Subsystem
94
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
[:SOURce[n]][:CHANnel[m]]:POWer[:LEVel][:IMMediate][:AMPLitude]
syntax:
[:SOURce[n]][:CHANnel[m]]:POWer[:LEVel][:IMMediate][:AMPLitude]
[PW|NW|UW|MW|Watt|DBM]
description:
Sets the power of the laser output.
NOTE
If an optical attenuator is installed, the power value returned is dependent on whether you
are using power or attenuation mode (see
“[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation:AUTO” on page 93).
• If you are using power mode, the value returned is the output power.
• If you are using attenuation mode, the value returned is the laser output power, and you
must also use the attenuation value to calculate the output power (see
“[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]” on page 92).
The values for the output power that you set in the Power Mode, and the laser output
power that you set in the Attenuation Mode, are stored and used independently.
NOTE
The instrument may not be able to output a signal with the maximum programmable
power, it will output a signal with the maximum power. Use the
“[:SOURce[n]][:CHANnel[m]]:POWer[:LEVel][:IMMediate][:AMPLitude]” on
page 94 to query the power being output.
The default units are DBM or W, depending on the unit selected using the following com-
mand: “[:SOURce[n]][:CHANnel[m]]:POWer:UNIT” on page 96.
parameters:
Any value in the specified range (see the appropriate User’s Guide).
Also allowed are:
MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value, but is half the
sum of the minimum programmable value and the maximum pro-
grammable level
response:
none
example:
sour2:pow 23uW
affects:
All tunable laser and DFB source modules
Signal Generation – The SOURce Subsystem
Measurement Operations & Settings
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
95
command:
[:SOURce[n]][:CHANnel[m]]:POWer[:LEVel][:IMMediate][:AMPLitude[l]]?
syntax:
[:SOURce[n]][:CHANnel[m]]:POWer[:LEVel][:IMMediate][:AMPLitude[l]]?
[MIN|DEF|MAX]
description:
Returns the amplitude level of the output power.
The value returned is the actual amplitude that is output, which may be dif-ferent from the
value set for the output. If these two figures are not the same, it is indicated in the :STA-
Tus:OPERation register.
NOTE
If an optical attenuator is installed, the power value returned is dependent on whether you
are using power or attenuation mode (see
“[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation:AUTO” on page 93).
• If you are using power mode, the value returned is the output power.
• If you are using attenuation mode, the value returned is the laser output power, and you
must also use the attenuation value to calculate the output power (see
“[:SOURce[n]][:CHANnel[m]]:POWer:ATTenuation[l]” on page 92).
The values for the output power that you set in the Power Mode, and the laser output
power that you set in the Attenuation Mode, are stored and used independently.
parameters:
Also allowed
(for tunable laser
modules only) are:
MIN: minimum amplitude level
MAX: maximum amplitude level
DEF: This is not the preset (*RST) default value but is half the
sum of, the minimum amplitude level and the maximum amplitude
level
NOTE
Use [l] to query the amplitude level of the output power of the upper or lower wavelength
laser source of a dual-wavelength laser source or a return loss module with an internal
dual-wavelength laser source. The default value of [l] is 1, the lower wavelength source.
The upper wavelength source is denoted by 2.
response:
Amplitude level relevant to the current value or specified parameter (if MIN, MAX, or
DEF are chosen as a parameter).
example:
sour2:pow? → +8.00000000E-004
affects:
All laser source, DFB source, and tunable laser modules and return loss modules with an
internal source
command:
[:SOURce[n]][:CHANnel[m]]:POWer[:LEVel]:RISetime
syntax:
[:SOURce[n]][:CHANnel[m]]:POWer[:LEVel]:RISetime[NS|US|MS|S]
description:
Sets the laser rise time of the chosen source.
parameters:
Any value in the specified range (see the appropriate User’s Guide).
Also allowed are:
MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half the
sum of, the minimum programmable level and the maximum
programmable level
response:
none
example:
sour2:pow:ris 10ns
affects:
All tunable laser and DFB source modules
Measurement Operations & Settings
Signal Generation – The SOURce Subsystem
96
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
[:SOURce[n]][:CHANnel[m]]:POWer[:LEVel]:RISetime?
syntax:
[:SOURce[n]][:CHANnel[m]]:POWer[:LEVel]:RISetime?[MIN|DEF|MAX]
description:
Returns the laser rise time of the chosen source.
parameters:
Also allowed are:
MIN: minimum programmable value
MAX: maximum programmable value
DEF: This is not the preset (*RST) default value but is half the
sum of, the minimum programmable level and the maximum
programmable level
response:
The rise time as a float value in seconds.
example:
sour2:pow:ris? → +1.00000000E-009
affects:
All tunable laser and DFB source modules
command:
[:SOURce[n]][:CHANnel[m]]:POWer:STATe
syntax:
[:SOURce[n]][:CHANnel[m]]:POWer[1|2]:STATe
description:
Switches the laser of the chosen source on or off.
parameters:
A boolean value:
0: Laser Off
1: Laser On
response:
none
example:
sour2:pow:stat 1
affects:
All laser source, DFB source, and tunable laser modules and return loss modules with an
internal source
command:
[:SOURce[n]][:CHANnel[m]]:POWer:STATe?
syntax:
[:SOURce[n]][:CHANnel[m]]:POWer:STATe?
description:
Queries the laser state of the chosen source.
parameters:
none
response:
A boolean value:
0: Laser Off
1: Laser On
example:
sour2:pow:stat? → 1
affects:
All laser source, DFB source, and tunable laser modules and return loss modules with an
internal source
command:
[:SOURce[n]][:CHANnel[m]]:POWer:UNIT
syntax:
[:SOURce[n]][:CHANnel[m]]:POWer:UNITDBM|0|Watt|1
description:
Sets the power units
parameters:
0 or DBM:
1 or W:
dBm (default)
Watts
response:
none
example:
sour2:pow:unit w
affects:
All tunable laser and DFB source modules
Signal Generation – The SOURce Subsystem
Measurement Operations & Settings
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
97
command:
[:SOURce[n]][:CHANnel[m]]:POWer:UNIT?
syntax:
[:SOURce[n]][:CHANnel[m]]:POWer:UNIT?
description:
Return the current power units
parameters:
0:
1:
dBm
Watts
response:
none
example:
sour2:pow:unit? → +0
affects:
All tunable laser and DFB source modules
command:
[:SOURce[n]][:CHANnel[m]]:POWer:WAVelength
syntax:
[:SOURce[n]][:CHANnel[m]:POWer:WAVelength[LOWer|UPPer|BOTH]
NOTE
For compatibility reasons, WAVelength may be replaced with WAVE.
description:
Sets the wavelength source for a dual-wavelength laser source.
parameters:
LOWer:
UPPer:
BOTH:
The lower wavelength source
The upper wavelength source
Both wavelength sources
response:
none
example:
sour2:pow:wav upp
affects:
All dual-wavelength laser source modules and return loss modules with two internal
sources
command:
[:SOURce[n]][:CHANnel[m]]:POWer:WAVelength?
syntax:
[:SOURce[n]][:CHANnel[m]:POWer:WAVelength?
NOTE
For compatibility reasons, WAVelength may be replaced with WAVE.
description:
Returns the wavelength source for a dual-wavelength laser source.
parameters:
none
response:
LOW
UPP
BOTH
The lower wavelength source
The upper wavelength source
Both wavelength sources
example:
sour2:pow:wav? → LOW
affects:
All dual-wavelength laser source modules and return loss modules with two internal
sources
Measurement Operations & Settings
Signal Generation – The SOURce Subsystem
98
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
[:SOURce[n]][:CHANnel[m]]:READout:DATA?
syntax:
[:SOURce[n]][:CHANnel[m]:READout:DATA?
description:
Returns the data as a binary stream from either a lambda logging operation or the maxi-
mum power the laser can produce at each wavelength.
parameters:
LLOGging:
PMAX:
Returns a binary stream that contains each wavelength step of the lambda
logging operation, see
“[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:LLOGging” on
page 103. Each binary block is an 8-byte long double in Intel byte order.
Returns a binary stream that contains the maximum power the laser can
produce at each wavelength. Each binary block is a 8-byte long double
(the wavelength value) followed by a 4-byte long float (the power value).
The stream is in Intel byte order.
response:
A binary stream in Intel byte order.
example:
sour2:read:data? llog → the data as a binary stream
affects:
All tunable laser and DFB source modules
command:
[:SOURce[n]][:CHANnel[m]]:READout:POINts?
syntax:
[:SOURce[n]][:CHANnel[m]:READout:POINts?LLOGging|PMAX
description:
Returns the number of datapoints that the [:SOURce[n]][:CHANnel[m]]:READout:DA-
TA? command will return.
parameters:
LLOGging:
PMAX:
Returns the number of wavelength steps for a lambda logging operation,
see “[:SOURce[n]][:CHANnel[m]]:WAVelength:SWEep:LLOGging” on
page 103.
Returns the number of datapoints (each datapoint contains a value for
wavelength and power) the [:SOURce[n]][:CHANnel[m]]:READout:DA-
TA? PMAX command will return, number of datapoints depends on the
calibration data for your module.
response:
The number of datapoints as an integer value.
example:
sour2:read:poin? pmax → 120
affects:
All tunable laser and DFB source modules
Signal Generation – The SOURce Subsystem
Measurement Operations & Settings
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
99
command:
[:SOURce[n]][:CHANnel[m]]:WAVelength[:CW|:FIXED]
syntax:
[:SOURce[n]][:CHANnel[m]]:WAVelength[:CW|:FIXED]
[PM|NM|UM|MM|M]
description:
Sets the absolute wavelength of the output.
parameters:
Any wavelength in the specified range (see the specifications in the appropriate User’s
Guide).
The programmable range is larger than the range specified in the User’s Guide. The pro-
grammable range is set individually for each instrument when it is calibrated during pro-
duction.
Also allowed are:
MIN: minimum wavelength value
MAX: maximum wavelength value
DEF: This is not the preset (*RST) default value but is half the
sum of, the minimum wavelength value and the maximum
wavelength value
The default units are M.
response:
none
example:
sour2:wav 1550NM
affects:
All tunable laser and DFB source modules
command:
[:SOURce[n]][:CHANnel[m]]:WAVelength[:CW[l]|:FIXED[l]]?
syntax:
[:SOURce[n]][:CHANnel[m]:WAVelength[:CW[l]|:FIXED[l]]?[[MIN|DEF|MAX]
description:
Returns the wavelength value in meters.
NOTE
Use [l] to query the upper or lower wavelength laser source of a dual-wavelength laser
source. The default value of [l] is 1, the lower wavelength source. The upper wavelength
source is denoted by 2.
parameters:
none
Also allowed
(for tunable laser modules only)
are:
MIN: minimum wavelength
MAX: maximum wavelength
DEF: This is not the preset (*RST) default value but is
half the sum of, the minimum wavelength value and the
maximum wavelength value
response:
The wavelength as a float value in meters.
example:
sour0:wav? → +1.5672030E-006
sour0:wav? min → +1.5500000E-006
sour2:wav:fixed2? → +1.61544494E-006
Returns the current wave-
length value for a tunable
laser module.
Returns minimum wave-
length for a tunable laser
module.
Returns the wavelength val-
ue of the upper wavelength
source of a dual-wavelength
laser source.
affects:
All laser source, DFB source, and tunable laser modules and return loss modules with an
internal source
Measurement Operations & Settings
Signal Generation – The SOURce Subsystem
100
HP 8163A Lightwave Multimeter, HP 8164A Lightwave Measurement System,
& HP 8166A Lightwave Multichannel System User’s Guide, E1299
command:
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ARA
syntax:
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ARA
description:
Realigns the laser cavity.
parameters:
none
response:
none
example:
sour2:wav:corr:ara
affects:
All tunable laser modules except HP 81689A
command:
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ZERO
syntax:
[:SOURce[n]][:CHANnel[m]]:WAVelength:CORRection:ZERO
description:
Executes a wavelength zero.
parameters:
none
response:
none
example:
sour2:wav:corr:zero
affects:
All tunable laser modules except HP 81689A
command:
[:SOURce[n]][:CHANnel[m]]:WAVelength:FREQuency
syntax:
[:SOURce[n]][:CHANnel[m]]:WAVelength:FREQuency
[THZ|GHZ|MHZ|KHZ|HZ]
description:
Sets the frequency difference used to calculate a relative wavelength. The output wave-
length is made up of the reference wavelength and this frequency difference.
The default units for frequency are Hertz.
The output wavelength (λ) is set from the base wavelength (λ0) and the frequency offset
(df). The formula for calculating the output wavelength is:
where c is the speed of light in a vacuum (2.990 x 108 ms-1)
parameters:
The frequency difference is a float value in Hz.
response:
none
example:
sour2:wav:freq -10THZ
affects:
All tunable laser and DFB modules
command:
[:SOURce[n]][:CHANnel[m]]:WAVelength:FREQuency?
syntax:
[:SOURce[n]][:CHANnel[m]]:WAVelength:FREQuency?
description:
Returns the frequency difference used to calculate a relative wavelength.
parameters:
none
response:
Returns the frequency difference as a float value in Hz.
example:
wav:freq? → -1.00000000E+013
affects:
All tunable laser and DFB modules
λ
c
( )
(λodf
)
+ c
(
)
------------------------------ λo
=