Axis 4B/F Series Instruction Manual
Read Online
Read the Axis 4B/F Series Instruction Manual online using the built-in PDF viewer.
Text of the manual
Table of contents:
1.
General description ................................................................................................................................................................ 3
2.
Completation .......................................................................................................................................................................... 4
3.
Technical data ........................................................................................................................................................................ 4
4.
Balance keys and indicators .................................................................................................................................................. 5
5.
Safety principles ..................................................................................................................................................................... 6
6.
Principles of used balance treatment ..................................................................................................................................... 6
7.
Foundation preparation .......................................................................................................................................................... 7
8.
Balance installation ................................................................................................................................................................ 9
9.
Balance start ........................................................................................................................................................................ 10
10.
Operation principles ........................................................................................................................................................ 11
11.
Balance test .................................................................................................................................................................... 12
12.
Balance adjustment ........................................................................................................................................................ 12
13.
Connection witch computer, printer or label printer ....................................................................................................... 12
14.
Basic function.................................................................................................................................................................. 16
14.1
Normal weighing ............................................................................................................................................................. 16
14.2
Weighing with tare .......................................................................................................................................................... 17
14.3
Increased readability ....................................................................................................................................................... 17
15.
Special functions describtion .......................................................................................................................................... 18
15.1
Menu customization function (ACtIV and dEFAUL) ........................................................................................................ 19
15.2
Autozeroing function (AutoZE) ....................................................................................................................................... 20
15.3
Pieces counting (PCS) ................................................................................................................................................... 21
15.4
Printer cooperation settings (PrInt) ................................................................................................................................ 22
15.5
Serial port parameters setting (Port) .............................................................................................................................. 23
15.6
Constant tare (tArE) ........................................................................................................................................................ 24
15.7
Average calculation function (AVErAG) ......................................................................................................................... 26
15.8
Recipe weighing function (RECIPE) .............................................................................................................................. 27
15.9
Percentage weighing function (PErC) ............................................................................................................................ 28
15.10
Anti-disturbance filter function (FILtEr) ..................................................................................................................... 29
15.11
Function for maximum value indication (UP) ............................................................................................................. 30
15.12
Force measuring function (nEWto) ........................................................................................................................... 31
15.13
Function for weighing large animals (LOC) ............................................................................................................... 32
15.14
Checkweighing function (thr) ..................................................................................................................................... 33
15.15
Total weight function (totAL) ...................................................................................................................................... 36
15.16
Function for setting date and time (dAtE) .................................................................................................................. 38
15.17
Function for selecting label number (LAbEL) ............................................................................................................ 39
15.18 External calibration (CALIb) ........................................................................................................................................ 40
15.19 Setting time of stabilization function (Stb) .................................................................................................................. 41
15.20
Entering reference zero function (Zero) ..................................................................................................................... 42
15.21
Automatic switching off the scale function (AutoOF) ................................................................................................. 43
15.22
Density determining function (dEnSIt) ....................................................................................................................... 44
15.23
Statistical calculations function (StAt) ........................................................................................................................ 47
Maintenance and small repairs ....................................................................................................................................................... 49
Declaration of Conformity ............................................................................................................................................................... 50
Appendix A ..................................................................................................................................................................................... 51
1. General description
Balances from 4B/F series in basic version from lacquered steel, with SE-
01/A/18 meter, are designed for operation in foundation structures, in light
environmental conditions.
Balances from 4B/F series from stainless steel, with SE-01/A/18 meter, are
designed for operation in foundation structures, in hard environmental
conditions.
All balances are tested in respect of metrology. According to order, balances
may be verified or calibrated.
EC verification (conformity assessment) of balances is required for special
applications, mentioned in Ministry of Labour and Social Policy decree from
11.12.2003 (trade, tariffs, pharmacy recipes, medical and pharmaceutical
analysis, packing of goods). For other applications it is recommended to replace
verification with calibration.
Verified balances have the following verification features:
- protection stamps, located on balance meter and converters connection box
(installed under the bearing surface).
- Office of Measures marks and green metrological mark, located on the rating
plate.
EC verification is valid for 25 months from 1st of December of year, when the
EC verification took place, under condition of stamp integrity.
Balances classification (PKWiU code) 29.24.23.
Certificates:
Certificate
Certificate
of type approval
DIN EN ISO 9001:2000
no. PL 04 020 no. 78 100 6386
(AXIS management system)
Basic set includes:
1. Bearing surface and balance meter, connected permanently with cable
2. RS232C connector (only 4B/FN balances)
3. Instruction manual
4. Warranty
3. Technical data
Balance type
4B300F(N)
4B600F(N)
4B1500F(N)
4B2000F(N)
4B3000F(N)
Maximum load (Max)
300kg
600kg
1500kg
2000kg
3000kg
Reading graduation (d)
100g
200g
500g
1kg
1kg
Verification graduation (e)
100g
200g
500g
1kg
1kg
Minimum load (Min)
2kg
4kg
10kg
20kg
20kg
Tare set range
-300kg
-600kg
-1500kg
-2000kg
-3000kg
Precision class
III
Operational temperature
-10÷40°C
Weighing time
<3s
Foundation depth
128mm
Platform dimensions: Balance
weight:
1000x1000mm 140kg
●
●
●
1250x1250mm 170kg
●
●
●
●
1500x1500mm 210kg
●
●
●
●
1500x2000mm 300kg
●
Display cable length
4m
Power supply
~230V, 50Hz, 8VA
Meter protection rating
SE-01/A/18 - no IP, SE-01/N/18 - IP65
Sensors protection rating
IP68
4. Balance keys and indicators
key
I/
- ON/OFF switch (standby),
key
→T←
- tare setting (entering of mass subtracted from weighed
mass),
key
B/G
- gross weight indication,
key
→0←
- zeroing,
key
MENU
- special functions menu,
key
- result print,
key
HR
- weight indication resolution increasing,
indicator
→0←
- zero indicator (for unloaded balance),
indicator
- signals weighing result stabilization,
indicator
NET
- net weight (after →T← key use),
indicator
- gross weight (after B/G key use),
indicator
- pieces counting function indicator (indications in pieces)
Additionally in LCD display version:
indicator
MODE
- special function activation menu,
indicator bar
- balance load indicator (0-100%),
indicator
WYŁ
- balance switched off with key (standby),
indicator
B/G
- gross weight (after B/G key use),
indicator
pcs
- indications in pieces
5. Safety principles
It is necessary to become familiar with safety principles
shown below, which are necessary to avoid electric shock
and damage of balance or connected devices.
• Maintain causion while opening the pan. It is suggested to
carefully hold the pan while opening.
• Repairs and necessary adjustments must be performed by qualified
personnel only
• To avoid fire hazard, use only typical supply cable, and supply voltage
must be consistent with technical data.
• For the balance supply use the socket with protective contact.
• Do not use balance when the cover is removed.
• Do not use balance in explosive atmosphere.
• Do not use balance in locations with high humidity, when the cover does
not have special protections for N type.
• In case of damage suspicion, switch the balance off and do not use it until
it is tested in professional service company.
6. Principles of used balance treatment
According to valid regulations regarding natural
environment protection do not place used electronic
devices in containers with common waste.
• Used balance after operation period may be transferred to units authorized
for collection of used electronic equipment, or to place of purchase.
7. Foundation preparation
CAUTION!
Decision of substrate reinforcing and density is taken by building works
designer.
All dimensions are in millimetres.
During foundation preparation follow the principles below:
1. Foundation and feet under balance legs should be made from concrete.
Foundation bottom thickness should not be smaller than 100 mm.
2. Foundation bottom inclination angle (shown in fig. 1 on page 8) should be at
least 3%.
3. Keep the same level for all feet.
4. Level the foundation frame, keeping perpendicularity of sides and equality of
diagonals.
5. Diameter of PVC pipe for cables to connection box should be at least 50 mm.
6. Keep the area free from reinforcing bars as shown in fig. 2 on page 8.
7. Foundation external and internal dimensions, proper for various platform
dimensions are shown in table below.
Balance type Platform dimensions
[mm]
Foundation internal
dimensions
(WxLxH) [mm]
Foundation enternal
dimensions
(WxL) [mm]
4B300FN
800x800
820x820x110
1220x1220
4B300FN
1000x1000
1020x1020x110
1420x1420
4B300FN
1250x1250
1270x1270x110
1670x1670
4B600FN
800x800
820x820x110
1220x1220
4B600FN
1000x1000
1020x1020x110
1420x1420
4B600FN
1250x1250
1270x1270x110
1670x1670
4B1500FN
1250x1250
1270x1270x110
1670x1670
4B1500FN
1500x1500
1520x1520x128
1920x1920
4B2000FN
1250x1250
1270x1270x128
1670x1670
4B2000FN
1500x1500
1520x1520x128
1920x1920
4B3000FN
1500x1500
1520x1520x128
1920x1920
4B3000FN
1500x2000
1520x2020x128
1920x2420
Keep all dimensions, shown in drawings and table.
Careless and inaccurate foundation preparation may make balance
installation impossible or cause its improper operation!
Fig.1 Foundation – crosssection
Fig. 2 Foundation – view from above
300
70
180
180
65
65
300
210
210
Stopy pod nóżki wagi
200
Rama fundamentu
z kotwami
60
Kolektor do odprowadzenia wody
Rura z PCV
na przewód
Obszar wolny
od prętów
zbrojeniowych
Rama fundamentu
z kotwami
Miejsca na kotwy ramy
Feet under balance legs
Foundation frame
with anchors
Water collector
PCV pipe for cable
Places for frame anchors
Area
reinforcement
bar free
Foundation
frame with
anchors
8. Balance installation
1. Balance structure
2. Screws fixing the limiter
3. Screws fixing base plates
4. Limiter
5. Base plate
6. Foundation
7. Balance leg
1. Prepare the foundation (7) according to guidelines (section 6) and engineering
technique principles.
2. Put the balances base plates (5) freely in the foundation.
3. Place limiters plates (4) on plates (5). Connect both plates with M8 screws (2)
so they cannot move in respect to each other, but to allow for limiters (4)
manoeuvring.
4. Unscrew screws fixing the balance upper plate.
5. Put the balance structure (1) in the foundation, so legs (7) fit vertically in the
limiters (4) holes.
6. Place upper plate on balance structure and lightly screw in the fixing screws.
7. Check the upper plate location in the foundation – level in relation to
foundation edge and distance from the edges (fig. 1).
8. When needed, shift the plates (5) or put spacers beneath them, made from sheet
with dimensions of plate.
9. If the position of upper plate is correct, carefully remove it to not move plates
(5). Check if legs (7) are straight.
10. Carefully remove balance structure (1) from foundation to not move plates (5).
11. Fix plates (5) with supplied screws with steel pins.
12. Adjust possible displacements with plates (4) position, so legs are located
centrally and vertically in plates (4) holes. Tightly screw plates (5) and (4).
13. Put balance structure (1) in place.
14. Place the upper plate and fix it with screws, beginning from holes in the upper
plate centre.
Connecting cable must be laid freely!
15. Unpack the balance. Attach meter to the
wall or desktop, choosing one of three
methods.
16. Connect the supply cable plug to socket
with protective contact with unloaded
balance,
which
will
cause
autotests
performing and after indication stabilization
displaying of sero indication.
9. Balance start
Connecting of supply plug to socket of ~230V/50Hz installation with unloaded
balance will cause the following sequence of actions:
Display test
Tests of basic electronic subassemblies,
ended with messages: C1, C-2, ... C-6.
Displaying of software version.
Ready for operation.
1. Before each measurement the balance should be properly zeroed, which is
signalled by “→0←” indicator. If the zero indication is not shown when the
balance is unloaded, or "----" is displayed, press the "→0←" key.
2. The balance enables tare setting in the whole measuring range. It is performed
by pressing "→T←" key.
3. Weighed mass should be placed in the platform centre.
Do not drop weighed objects on the platform.
Do not overload the balance over 20% of maximum load.
4. The weighing result should be read during the " " indicator lighting, which
indicates the result stabilization.
5. When there is no weighing, but the balance should be ready for operation, it may
be switched off by I/ key. It causes the balance reading system deactivation
end entering the standby mode. Balance turning on is performed by pressing "
I/ " key.
6. Protect the balance from dust, aggressive dusts and liquids. For cleaning
purposes use water and dry it.
Maintain causion while opening the pan. It is suggested to
carefully hold the pan while opening!
11. Balance test
During balance operation, in order to confirm its efficiency, it is recommended to
check the weighing precision by putting and object of exactly known weight before
and after series of measurements.
For testing of verified balances use weight standard, having valid standardization
certificates. In case of allowable measurement error exceeding contact authorized
service company to perform balance adjustment.
12. Balance adjustment
Balance adjustment must be performed by authorized service company
only, as it is connected with necessity of seals breaching, required for
warranty purposes.
13. Connection with computer, printer or label printer
The scale is equipped with RS232C or
RS485 (optional), which can be used to
connect external devices such as a computer
or a printer. Detailed way of communication
data is defined by one of protocol:
- LonG – protocol to printer and computer
- ELtron – protocol to label printer
When cooperating with a printer data is send
automatically after result stabilisation, but
next transmission is possible after removing
previously weighted sample.
When protocol Long is chosen scale sends
result of weighting under the influence of
initializing signal from computer, or after
pressing a
key.
When automatic transmission mode is
chosen (function PrInt), a printer data is send
automatically after result stabilisation, but
next transmission is possible after removing
previously weighted sample. The scale sends
following information: successive number of
weighing, date and time (if clock module is
installed )and weighing result.
During transmission communicate Print is displayed. It is also possible by scale to
display text message sent from computer confirming received information.
RS-232C
RS485 (option)
RxD (receiver)
TxD (transmitter)
GND
Case
1
2
3
4
5
6
7
8
9
A
Input
Output
A
B
B
Case
1
2
3
4
5
6
7
8
9
}
}
Scale with Eltron protocol sending record of data after pressed
key.
The scale sends following information: number of label, date and time (if clock
module is installed ) and weighing result. During transmission communicate Label
is displayed.
In special features scale can be equipped in second RS232 interface for additional
functions, ex. continuous transmitting of indications to second display.
Computer must have a special program for cooperation with data from a scale.
Programs offer is available on www.axis.pl/en internet site.
Detailed protocol description
LonG protocol
Communication parameters: 8 bits, 1 stop bit, no parity, baud rate 4800bps,
� Readout of scale indication ( corresponds to pressing
key
Computer→Scale: S I CR LF (53h 49h 0Dh 0Ah),
Scale→Computer: scale response according to description below (16 bytes):
Byte 1
- sign „-” or space
Byte 2
- space
Byte 3÷4 - digit or space
Byte 5÷9 - digit, decimal point or space
Byte 10
- digit
Byte 11
- space
Byte 12
- k, l, c, p or space
Byte 13
- g, b, t, c or %
Byte 14
- space
Byte 15
- CR
Byte 16
- LF
Attention:
Network number different than zero (F..-rS / nr function) changes scale working
mode: communication with a computer is possible after logging the scale in with
02h scale number command. To log the scale out use 03h command.
For example: Using a program to test RS232 interface ( program is available on
www.axis.pl in computer programs section) for scale number 1 please write: $0201
to log in, then SI, and write: $03 to close communication.
� Asking for scale presence in system (testing scale connection with computer):
Computer→Scale: S J CR LF (53h 4Ah 0Dh 0Ah),
Scale→Computer: M J CR LF (4Dh 4Ah 0Dh 0Ah),
� Displaying a sign on scale display (text message from computer):
Computer→Scale: S N n n X X X X X X CR LF (53h 4Eh 0Dh 0Ah),
nn-displaying time in seconds; XXXXXX- signs to display
Scale→Computer: M N CR LF (4Dh 4Eh 0Dh 0Ah),
� Scale tarring (calling →T← key press) :
Computer→Scale: S T CR LF (53h 54h 0Dh 0Ah),
Scale→Computer: without response,
� Scale zeroing (calling →0← key press):
Computer→ Scale: S Z CR LF (53h 5Ah 0Dh 0Ah),
Scale →Computer: without response,
� Scale turning on / off (calling I/ key press):
Computer→ Scale: S S CR LF (53h 53h 0Dh 0Ah),
Scale →Computer: without response,
� Entering to special function menu (calling MENU key press):
Computer→ Scale: S F CR LF (53h 46h 0Dh 0Ah),
Scale →Computer: without response,
� Setting low threshold value (option):
Computer→ Scale: S L D1...DN CR LF (53h 4Ch D1...DN 0Dh 0Ah)
D1...DN – threshold value, maximum 8 characters („-” – negative value, digits,
dot – decimal separator), number of digits after dot should be the same as on
scale display,
Scale →Computer: without response,
� Example:
⋅ in order to set low threshold 1000g in scale B1.5 (d=0.5g) the following order
should be sent:
S L 1 0 0 0 . 0 CR LF (53h 4Ch 31h 30h 30h 30h 2Eh 30h 0Dh 0Ah),
⋅ in order to set low threshold 100kg in scale B150 (d=50g) the following order
should be sent:
S L 1 0 0 . 0 0 CR LF (53h 4Ch 31h 30h 30h 2Eh 30h 30h 0Dh 0Ah),),
� Setting high threshold value (option):
Computer→ Scale: S H D1...DN CR LF (53h 48h D1...DN 0Dh 0Ah),
D1...DN – threshold value (see )
Scale →Computer: without response.
Connecting cable WK-1 (scale – computer / 9-pin interface):
Protocol ELTRON description
Communication parameters: 8bits, 1stop, no parity, 9600bps,
� After use
key scale send:
Scale→Label printer : EPL-2 language instruction to initialize print of label:
US
- control instruction
FR"0001"
- number of labels
?
- starting instruction
hh:mm
- 5 signs: hours:minutes
yyyy.mm.dd
- 10 signs: year.month.day
mass + unit
- 10 signs: scale indicate + mass unit
P1
- ending instruction
Attention:
1. It is possible to place a constant signs (company name, product name).
2. The default number of label’s is 0001 (label number 1).Different numbers of
label are possible by using a LAbEL function.
3. Label formula must be saved in printer – label forma must be designed in
computer (by Zebra Designer program) and saved in printer.
4. Parameters and protocol of transmission must be compatible with printer label.
14. Basic function
To make clear how to manage with each function, in further part of instruction
descriptions are replaced with pictures.
- put a load on the pan
- remove the load from the pan
- press the key when indication is
displayed
- forced change
- automatic change
14.1 Normal weighing
Press →0← key (→T← key in non-
legalized scales), which zeros the
scale, operates only when the pan is
empty .
Weighing result should be read when
the indicator "
" lights.
0.00
2.30
0
2.30
T
14.2 Weighing with tare
The scale is equipped with tare equal
to its range.
Joint value tare and mass net can not
cross a maximum of scale.
To display gross weight press B/G
key.
Note:
Press B/G key to return to net weight
indication.
14.3 Increased readability
Press HR key to display the weighing
result (for 5s.) with the highest
readability possible. This function is
especially helpful in scales with legal
verification with d=e.
The weighing result with increased
readability
can
be
used
for
informational purposes only and
cannot be printed or sent to a
computer with
key.
2.30
0.00
40.01
42.31
T
NET
NET
2.30
2.29
2.30
Special functions describtion
List of available functions:
�
menu customization function (ACtIV),
�
removal of all function from menu (dEFAUL).
�
autozeroing function (AutoZE),
�
pieces counting function (PCS),
�
function for setting serial port working (PrInt)
�
function for setting serial port (Port)
�
entering tare function (tArE)
�
recipe weighing function (rECIPE),
�
weighing large animals function (LOC)
�
force measuring function (nEWto)
�
maximum value indication function (UP)
�
anti-disturbance filter function (FILtEr)
�
average calculating function (AVErA)
�
percentage weighing function (PErC),
�
extended calibration function (CALIb)
�
setting time of stabilisation function (Stb)
�
selecting label number function (LabEL)
�
automatic switching off scale function (AutoOF)
�
entering reference zero (Zero)
�
determining solids and liquids density function (dEnSIt)
�
calculator for good packaging control (tP) - option*
�
statistical calculations (StAt)- option*
�
paperweight calculation function (PAP) - option*
function with additional equipment require:
� options with the clock:
- setting current date and time function (dAtE)
- total weight function (totAL)
� options with the transoptors connectors:
- checkweighing function (thr)
* Functions offered with special version of scale software (with limited possibility of using other
special functions).
User create own menu by choosing function in ACtIV function (described in chapter 14.1).
15.1 Menu customization function (ACtIV and dEFAUL)
Among available user functions it is
possible to select these, which should be
displayed after pressing MENU key. It
allows avoiding displaying whole list of
available
functions,
which
makes
operation time longer.
Operation sequence shown in the
pictures on the left causes adding
function for setting serial interface
RS232C parameters (Port) to function
menu.
After switching on ACTIV function a
dot is displayed on the right side (to
distinguish from regular menu). Chosen
functions are displayed with a dot on
the left side.
In every moment, it is possible to
restore primary (manufacture) settings
choosing dEFAULt special function.
In order to remove function from menu
in the last operation in place of selecting
Port -1 choose Port -0.
0.00
ACtIV
Port -1
Port -0
PCS .
.
.
Port
T
T
T
15.2 Autozeroing function (AutoZE)
When F..-Aut function is activated,
the scale automatically ensures stable
zero indication if the pan is empty or
if zero indication was acquired by
pressing →T← key.
To leave the function press MENU
key, then with →T← key chose
AutoZE and Aut-0.
Note:
Autozeroing function is activated
automatically for 10 min. after
switching-on.
0.01
AutoZE
Aut-1
Aut-0
0.00
T
T
15.3 Pieces counting (PCS)
This function enables to count
identical pieces, e.g. turnbuckles or
buttons.
A measurement is performed in two
phases:
- first phase - single piece weight
calculation on the basis of defined
pieces amount (5, 10, 20, 50, 100,
200 or 500 pieces),
- second phase – pieces counting.
It is advised that single piece weight
is not less than one reading unit and
sample weight used in first phase is
bigger than 100 reading units.
To leave function press MENU key
and then with →Τ← key chose
PCS and PCS-0.
Note:
1. Err-3 communicate signalises that
a sample was not put on the pan.
The same
communicate
appears
if single piece weight is less than one
reading unit (it is possible to count
pieces but measuring error is bigger).
2. To chose previously used pieces
amount select " _ _ " in first phase
(in case no value was chosen, error
communicate appears
3. During pieces counting →T← key
function does not change.
4.In scales equipped with LCD
display, weighing unit is visible and
"" sign is replaced with "pcs ".
0
0.00
23.40
PCS
PCS -0
PCS-1
PCS - --
5
5
12
12
T
T
T
T
15.4 Printer cooperation settings (PrInt)
Activate the function for automatic
serial
port
working
mode
(cooperation with a printer.
After activation the scale prints
a header.
Weighing
result
with
a successive measurement number is
printed automatically after result
stabilisation (without using
key).
To select computer cooperation mode
(
key activated and weighing
results without successive numbers)
press MENU key, then with →T←
key chose PrIn-0 and PrIn-1.
0.00
PrInt
PrIn-1
PrIn-0
0.00
1.05
PrInt
T
T
15.5 Serial port parameters setting (Port)
The function enables to set the
following transmission parameters
(standard parameters underlined:
- transmission protocol (Prot):
LonG - printer,
ELtron – label printer,
- transmission speed (bAud: 1200,
4800, 9600, ...),
- the number of bits in a byte (bit: 7, 8),
- parity control (PArItY: 0, 1; Odd: 0,
1),
- network number when working in
multistand computer system (when
working as a single scale the value
should be “0”,
- continuous transmission – without
using
key, approx. 10 results per
second (SEnd: 0, 1).
Protocol Eltron automaticly activated
function LAbEL.
To
set
desired
transmission
parameters activate Port function,
choose appropriate parameter and
press →T← key to accept needed
parameter value. The example at the
left presents how to set transmission
speed value to 9600bps.
To leave the function choose out
option.
0.00
Port
Port-1
Port-0
baud
-9600
-4800
bIt
nr
out
Odd
PArItY
Send
T
T
T
T
T
T
T
T
T
T
15.6 Constant tare (tArE)
This function enables to measure gross weight of a sample placed in a container of a known
weigh value (stored in the memory) and to display calculated net weight of the sample. Tare
value is recalled from the memory with →0← key when the pan is empty. Tare value may be
entered using the keypad or by sampling container weight from the pan.
Operation sequence:
The following options are possible:
- tAr-0 – leave the function,
- tAr-1 – activate the function with
the previous tare value,
- tAr-2 – sample tare value from the
pan,
- tAr-3 – enter tare value with keys:
→0←,
, →T← and MENU
- tAr-4 – printout a setting value of
tare
If the function is active, NET
indicator is displayed.
Options tAr-1 enables to activate the
function with previous tare value
after leaving the function with tAr-0
option.
Note:
Tare value is stored in memory also
after unplugging the scale from the
mains.
-
1.7
0.00
tArE
1.
1
tAr -3
1.0
T
T
T
NET
...
...
1.70
Weighing with constant tare:
When tAr function is activated, press
→0← key to zero the indication and
to recall tare value from the memory.
Tare value is displayed with ”-” sign.
B/G
key
enables
for
instant
switching between net and gross
weight.
Note:
When the pan is empty →T← key
does not operate – to tare the scale
use →0← key.
-1.70
13.70
10.00
0
B/G
NET
.
15.7 Average calculation function (AVErAG)
The function allows for calculating average value of performed measurement
series. During series of measurements successive results are registered
automatically when scale indication is stabilised.
Press MENU key and select AUErAG
pressing →T← key.
The
following
will
be
shown
successively on display:
- AUEr-0 – leaving function,
- AUEr-1 – measurement with average
calculation..
Select AUEr-1. It will allow weighing
with simultaneous storing results into
summing
register
for
average
calculation
Measurement registration is performed
automatically in the moment when
scale indication becomes stable. Short
time displaying of „ - - ” denotes that
load can be taken off and new one can
be put on. Results above scale Min are
registered
only.
Number
of
measurements is limited to 9999.
In order to read average value
key
should be used.
- first
pressing
causes
displaying
number of measurements (n).
- second displaying causes displaying
average value (=).
- third pressing allows continuing
average calculation.
If a printer is connected to scale the
following report will be printed :
In order to finish calculation press
MENU key, and then select AUEr i
AUEr-0.
Date: ... Time. ...
MEASUREMENS No = ...
AVERAGE VALUE = ...
0.00
0.00
AUErAG
AUEr- 1
1.00
2.00
2
1.50
__
0.00
0.00
MENU
T
T
n
15.8 Recipe weighing function (RECIPE)
This function allows for separate
weighing of several ingredients in
one container with a possibility of
control total weight of all weighed
components.
The function has the following
options:
- rECI-0 – leave the function with
possibility of reading to read total
weight,
- rECI -1 – start recipe weighing
- rECI -2 – continue previous recipe.
When preparing a recipe successive
ingredients (A, B, C, etc.) are
weighed each time starting from zero
indication. In order to allow this after
weighing of each ingredient tare the
scale.
After weighing of several ingredients
reading total weight is possible
(despite scale taring). In order to do
that press MENU key, select rECIPE
function once more and use rECI -0
option.
Sign „
__ ” signals total weight
indication. Recipe is finished when
→T← key is pressed.
When „
__” sign is displayed recipe
continuing is possible. rECI -2 option
is used for that.
Note:
Sign „≡” on the left side of display
informs
about
rECIPE
function
activity.
0.00
rECIPE
rECI- 0
rECI- 1
0.11
0.12
0.13
0.00
0.00
0.36
A
A+ B+ C
B
C
MENU
T
T
T
T
rECIPE
rECI- 0
MENU
T
T
_
_
_
_
_
_
_
_
_
_
_
_
_
_
_
_
_
15.9 Percentage weighing function (PErC)
This
function
allows
displaying
weighing result in percents.
A measurement is performed in two
phases:
-
first
phase
–
weighing
a
reference sample (100%),
-
second
phase
–
measuring
specific sample as a percentage
of the reference sample.
Weighing result is displayed in
different format, depending on the
weight value of reference sample. For
weight values of reference sample
0÷3,5% of weighing range result is
displayed in format 100, for range
3,5÷35% - in format 100.0, and
above 35% - in format 100.00.
„%” sign is replaced with „
__ ”
indicator.
The function has the following
options:
- PErC-0 – disable the function,
- PErC-1– set current scale indication
as 100% and activate percentage
weighing,
-
PErC-2 – continue
percentage
weighing after exiting to normal
weighing.
Note:
During percentage weighing →T←
key has its normal function.
0.00
1.70
PErC
PErC-0
PErC-1
100.00
95.00
100%
5%
T
T
_
_
_
_
15.10 Anti-disturbance filter function (FILtEr)
This function allows using digital filter
with
selected
intensivity
during
weighing. Filter reduces the influence
of mechanical vibrations (air blasts,
base vibrations) on measurement result.
Press MENU key and select FILtEr
pressing →T← key.
The following options will be shown
successively on display:
- FIL-0 – work without
- FIL-1 - filter I (weak)
- FIL-2 - filter II (medium)
- FIL-3 - filter III (sharp)
- FIL-4 - filter IV (very sharp)
Select on of four filters. This will cause
starting weighing with selected filter.
In order to go back to normal weighing
use MENU key once more and choose
FILtEr and FIL-0.
0.00
0.00
FILtEr
FIL- 0
FIL- 1
FIL- 2
FIL- 4
T
T
15.11 Function for maximum value indication (UP)
This function allows holding on display maximum value shown by the scale in a
while.
Before measurement scale should be tared.
After using MENU key and selecting UP
function the highest mass result will be hold
on display.
Pressing →T← key will cause result
zeroing.
Note:
Autozeroing
function
and
the
stabilisation
indicator
are
deactivated when UP function is
running.
Weighing
result
is
continuously
averaged
from
5
measurements.
0.00
UP
UP- 0
UP- 1
1.00
1g
1g
10.00
10.00
T
T
Force measuring function (nEWto)
Function activation causes displaying
result in force units (N).
Press MENU key.
Using →T← key choose NEWto
function, and then NEW-1.
Note: 1N≈0,1019kg
0.0
nEWto
nEW -1
nEW -0
0.000
T
T
15.13 Function for weighing large animals (LOC)
The function allows weighing animal moving on the scale.
Press MENU key.
When LOC function is displayed press
→T← key.
The following options appear on
display successively:
- LOC-0 – leave the function,
- LOC-1 – automatic weighing after
loading the scale,
- LOC-2 – the measurement initiated
manually by pressing
key.
When LOC-1F..-1 is displayed press
→T← key.
Tare the scale using →T← key if
necessary and place the animal on the
pan.
Wait until the weighing result is
averaged – scale display will be
blinking. Then scale will show stable
averaged result and will send it through
serial port. Final result is displayed on
the display and send via serial port to
computer or printer.
The result remains on display for about
30 second.
Important notes:
1. The loads less than Min are not averaged.
2. In the case when placing the animal takes more than 5s, it is advised to use
LOC-2 option (measurement initiated manually). It will allow performing
measurement in right moment pressing
key.
0.00
LOC
LOC -0
LOC -1
0.25
0.25
PrInt
MENU
T
T
15.14 Checkweighing function (thr)
This function allows comparing weighing result with two programmed reference
values: lower and upper threshold. Comparison result is signalled with indicators
(MIN, OK, MAX) and sound signal generated when threshold values are exceeded.
If comparison result is:
- smaller than lower threshold – the scale signals MIN (yellow colour),
- between threshold values - the scale signals OK (green colour, with the short
sound signal),
- greater than upper threshold - the scale signals MAX (red colour, long sound
signal).
The checkweighing results can be use to control:
- optical indicator (Indication mode),
- batching devices (Batching mode).
-
Standard scale is set for cooperation with optical indicator.
On outputs P1-P3 (Relays socket) short-circuit states appear as result of
comparison scale indication with threshold values.
On the chart below output states are shown during increasing load on the scale for
both working modes:
Indication mode: Batching mode:
P3
P1
P2
P3
zero
zero
treshold
treshold
thr I
thr I
thr I
thr II
thr II
thr II
P1
P2
In Batching mode on P1 (thr I) and P2 (thr II) outputs short-circuit impulses
appears for time of 0,5s. On P3 (zero) output short-circuit state appears when
indication does not exceed threshold value signalling zero load.
Relays connection diagram:
Relays output is the open collector transoptor output with load capacity 25mA /
24V. Transmitter inputs must be protected with diodes, e.g. 1N4148.
It is advised to use MS3K/P electronic board (sold separately), consisting of
RM96P transmitters, with DC24V input voltage and AC250V, 3A output.
Important notes:
1. After switching the scale on, both thresholds are set to maximum values.
2. When setting upper threshold value, pay attention that its value is not below
lower threshold value.
3. Setting lower and upper threshold value is possible after sending appropriate
orders from computer, what is described in scale user manual.
Imax < 25mA
Press MENU key and choose thr
pressing →T← key.
The following options are displayed
successively:
- thr-0 – deactivate the function,
- thr-1 – activate the function,
- thr-P – check last threshold values
(press
key several times),
- thr-t – choose Relays socket mode:
0 – exit to weighing
1 – Batching mode
2 – Indication mode.
Choose thr-1option using →T← key.
The following options for entering
thresholds are displayed:
- SEt-0
- go
to
weighing
with
signalling threshold excess,
- SEt-1 - set lower threshold value,
- SEt-2 - set upper threshold value,
-SEt-3 - set zero signalisation threshold.
Using →T← key select SEt-1 option.
Set lower threshold value using the
following keys:
→0← - digit increase,
- decimal point,
→T← - move to next digit,
MENU - finish.
Then select SEt-2 option and enter
upper threshold value.
Choosing Set-0 option will cause
starting work with signalisation of
exceeding thresholds and zero.
To change Relays socket mode use thr-t
option. Default option is Indication.
To leave the function, press MENU key
and then choose thr and thr-0 options.
0.00
thr
thr -1
thr -0
SEt-1
SEt-0
SEt-2
5
5 -
50
100
62.00
MENU
MENU
T
T
T
T
T
0
0
T
15.15 Total weight function (totAL)
The function allows calculating total weight for series of measurements, which can be grater than
scale capacity. It allows calculating total weight as well as average value.
Press MENU key.
When totAL is displayed press →T← key.
The
following
options
appear
successively:
- tot-P - report printout without clearing
total register,
- tot-0 - clearing total register, report
printout and leaving the function,
- tot-1 - working with receipt printout
after each measurement,
- tot-2
-
working
without
receipt
printout.
Press →T← key when tot-1 is displayed.
Perform measurement series pressing
key for storing results into total register.
In order to print and display results enter
to function choosing total and tot-P
option from menu
The results are display in the following
sequence:
- total weight (≡)
- number of registered measurements (n),
- average value (=),
regarding
that
moving
to
display
successive result is performed after
pressing
key.
In order to go back to total weighing
without zeroing total register press
key third time.
0.00
0.00
totAL
totAL
tot- P
tot- 1
10.00
20.00
2
15.00
0.00
30.00
MENU
T
T
T
T
n
_
_
_
_
_
To leave the function with clearing total register, select total function from menu
and choose tot-0 option. When It will cause the scale prints the communicate
informing about clearing registers.
The form of receipt after each measurement:
Report form:
Note:
When the scale has not an internal clock, Date and Time do not appear
on printout.
Maximum number of measurements 99 999.
Maximum total load 99 999 000d.
The weighing unit of the total value from the register (Total) is the same as the
weighing unit stated on the keypad or is 1000 times greater, what is signalled by
“o” indicator at the left of the display.
If the registered value is too big to be displayed, “E” communicate appears on the
display.
If the number of series is too high and cannot be displayed, “Err1”communicate
appears on the display.
Date: ... Time. ...
measurement no weight
measurement no weight
Date: ... Time. ...
TOTAL WEIGHT =
NUMBER OF SAMPLES =
AVERAGE VALUE =
15.16 Function for setting date and time (dAtE)
The function allows setting current
date and time of scale internal clock
and mode of its use.
The function has the following
options:
- dAt-0 – deactivate date and time
during printout of current weighing
result,
- dAt-1 – activate date and time
during printout of current indication
(
key),
- dAt-2 - change current date and
time.
The example at the left presents how
to set current date and time using
dAt-2 option.
On successive positions digits are
changing automatically or manually
using →0← key several times.
In order to choose appropriate digit
and move to the next position use
→T← key.
After setting proper date and time it
should be activated with dAt-1
option.
Time format: h gg – mm
(gg - hour, m - minute).
Date format: d mm – dd
(m - month, d - day).
Year format: r - rr
(r - two last year digits).
0.00
dAtE
dAt- 2
dAt- 0
h13 - 18
h13 - 18
h13 - 19
h13 - 9
1
0.00
d04 - 05
r - 06
r - - 09
T
T
T
0
T
T
T
T
15.17 Function for selecting label number (LAbEL)
The function appears in scales with ELTRON data transmission protocol. This
protocol allows printing scale indication and optionally date and time on label
printer, as variable texts. Other data, e.g. company address, product name, its bar
code can appear on label as constant fields. Label forms used by user, named as a
numeric value (max. 4 digits) should be previously stored in printer memory
according to printer user manual. Choosing label form is performed by entering
label number using LAbEL function.
Press MENU key.
When LAbEL is displayed press
→T← key.
The
following
options
appear
successively on display:
- LAb-0 – leave without changes,
- LAb-1 – enter label number,
Using →T← key select LAb- 1.
For entering label number the
following keys should be used:
→0← - increase digit,
→T← - move to next digit,
MENU – finish entering.
After putting load on and pressing
key data is sent to a printer.
Format of data sent to label printer:
US (55 53 0D 0A)
FR"0001" (46 52 22 30 30 30 31 22 0D
0A)
? (3F 0D 0A)
00:00 (30 30 3A 30 30 0D 0A)
2000.00.00 (32 30 30 30 2E 30 30 2E 30
30 0D 0A)
-
27
5.00
LAbEL
2
LAb -1
20
T
T
T
...
...
5.00
15.18 External calibration (CALIb)
Note: This function is enabled in non legalized scales only.
Calibration of sensitivity should be make when aaccuracy of scale is not
satisfactory. Could use standard mass equal of maximum range of scale (Max).
Operation sequence:
Press MENU key to call functions
menu and to choose CALIbr with
→T← key.
The
following
options
appear
successively on display:
- CAL-0 – leave without calibration,
- CAL-1 – quick calibration – without
confirms by MENU key,
- CAL–2 – calibration with confirms
- out – leave without changes
Press MENU when communicate
CAL-2 is display.
Confirm readiness to calibration by
press MENU key – pan must be
empty.
When communicate C is display put
a standard mass (equal of maximum)
to a pan and press MENU key.
Wait to end of calibration process.
m
0.00
CAL
0.00
m
0.00
CALIb
CAL-2
CAL-0
...
T
T
15.19 Setting time of stabilization function (Stb)
The function allows changing stabilisation time of scale indication and connected
with it the time of waiting for starting result printout on a printer connected to the
scale.
Press MENU key.
When Stb is displayed press →T← key.
The
following
options
appear
successively on display:
- Stb-0 – deactivate the function,
- Stb-1 – the longest stabilisation time,
- Stb-2 – long stabilisation time,
- Stb-3 – medium stabilisation time,
- Stb-4 – shorter stabilisation time,
- Stb-5– the shortest stabilisation time.
After choosing one of the options
weighing with activated filter is started..
In order to go back to normal work of
the scale choose Stb–0 option.
0.00
0.00
Stb
Stb-5
T
T
15.20 Entering reference zero function (Zero)
Note: This function is enabled in non legalized scales only.
ZER function allows entering new value of reference zero (value referred to empty
pan) without need of contacting with authorised service centre.
Press MENU key while Err-b is
displayed.
Zero
function
will
be
activated
automatically.
After CodE communicate disappears
insert access code to function (in new
scale 1234). The following options
appear successively on display:
ZEr-0 – activate function,
ZEr-1 – enter new zero value,
ZEr-2 – enter new protecting code.
Using →T← key, choose ZEr-1 and
direct result from A/C converter will
appear on scale display.
When the pan is empty press →0← key.
Wait for finishing zeroing process.
In order to change access code use
ZEr-2
option.
Entering
value
is
performed similarly as with ZEr-1
option.
It is also possible to use function in
standard way:
Activate Zero function using ACtIV
function.
Press MENU key.
When ZEro is displayed press →T←
key and so on.
Err-b
CodE
ZEr-1
T
-
1234
1
12
12
40007
- -
0.00
T
T
...
...
15.21 Automatic switching off the scale function (AutoOF)
The function is helpful in scales supplied from accumulator. The function causes
scale to switch off automatically after c.a. 10 minutes of not using it. Switching
function on causes last entered zero and tare values are remembered in scale
memory. After next start-up of the scale these values are restored.
Press MENU key.
When AutoOF is displayed press →T←
key.
The
following
options
appear
successively on display:
Auto-0 – activate function,
Auto-1 – deactivate function.
After function activation scale will
control changes of its indications. If the
scale will not be used and scale
indications will not be changing, after
time of c.a. 10 minutes the scale will
switch off remembering its settings
(zero and tare).
Switching on the scale is performed
after pressing I/ key.
10 min.
0.00
0.00
AutoOF
Auto -1
T
T
15.22 Density determining function (dEnSIt)
The function allows determining
solid density basing on its weight in
the air and in liquid of known
density using the formula below:
m1
g= * g liquid
m1- m2+ m3
where: m1-weight in the air
m2- weight in liquid
m3- hanger weight
g liquid – liquid density
By default:
g liquid = 1g/cm3
(for distilled water).
When
using
liquid
other
than
distilled water, choose dEnSIt from
menu and use dEn-2 option to enter
liquid
density
taking
into
consideration its temperature.
To enter value use the following keys:
→0←
- digit increase,
- decimal point,
→T← - move to next digit,
MENU - finish.
The measurement is performed in
three phases:
-measurement in the air,
-measurement in liquid,
-hanger weighing
To read density determination result,
enter dEnSIt function menu and use
dEn-P option. First pressing
key
causes
displaying
successive
measurement
number.
Second
pressing
key causes displaying
and printing result, and then going to
the next density measurement.
dEnSIt
dEnSIt
dEn -1
dEn -P
--
T
T
T
T
0.00
0.00
0.00
0.00
100.00
51.00
1.00
2.0000
1
n
If a printer is connected to the scale, printout of solid density measurement results
will be performed in the following form:
It is recommended to use pan hanging below balance, weighing in the air and in
liquid is performed then in the following way.
Phase I: measurement in the air.
Phase II: measurement in liquid.
Date: ... Time. ...
MEASUREMENT No. =
WEIGHT in air
= g
WEIGHT in liquid
= g
HANGER WEIGHT
= g
LOAD DENSITY
= g/cm3
LIQUID DENSITY
= g/cm3
Successive results will be printed in table
with indication of their values using "*" character
referred to limit values.
1. After performing last measurement
"END" text will appear and summarising report of
control results will be printed:
2. In order to finish working with the function and reset
results register, press MENU key, and when "F..-tP"
and "F..-0" is displayed press →T← key.
Note: Activating TP function causes that indicators
signal exceeding limit values Qn-2T1 and Qn+2T1.
PLACE OF
CONTROL: .................................
Date: ... Time: ...
CONTROL NO.: ................................
SCALE TYPE
:
...
FACTORY NO.
:
...
BATCH SIZE
:
...
VALUE Qn
:
...
VALUE Qn-T1
:
...
VALUE Qn-2T1
:
...
Qn-2T1 Qn Qn+2T1
... g
*
... g
*
...............
Date: ... Time:
...
MEASUREMENT COUNT =
...
QUALIFYING AVERAGE =
...
STANDARD DEVIAT. S
=
...
*HISTOGRAM*
Qn-2T1
+n2T1
RESULT :
...
CONTROLLER :
_________
15.23 Statistical calculations function (StAt)
Attention: Function is available on demand and it replaces other special functions.
This function evaluates from series of measurements (max 500) statistical parameters of
weighting process. Adding successively measurements to register is automatic and it occur after
the scale is loaded and its indications stabilize.
After each loading printout is made with: number of measurements, result, date and time.
Next measurement is made after taking off earlier load.
For the obtained measurements series the scale evaluates:
- n
- number of samples
- sum x
- sum of all samples
= ∑
nx
sum _ x
- x
-average value (sum x)/n
- min
-minimal value from n samples
- max
-maximal value from n samples
- max-min
-maximal value minus minima value
- S
-standard deviation
2)
(
)1
(
1
∑
−
−
=
n
n
x
x
n
S
- srel
-variance factor
x
srel = S
Statistical calculations results can be printed.
Order of operations:
1. Press MENU key.
2. When F..-StA is displayed press →T← key.
The following options are displayed:
-
F..-P – statistical data printout,
-
F..-0 - out of function, register zeroing, statistic data printout,
-
F..-1 - enter or continue function,
3. Press →T← key when F..-1 is displayed.
4. Put on successively objects on pan, (remove after indication stabilization) in order to add them
to measurement register.
5. In order to obtain printed statistical results from measurements series press MENU key.
When sign F..-StA is displayed, press →T← key.
When F..-P is displayed press again →T← key.
This will cause printout of calculated statistics and
histogram :
LSL - allowable lower value,
USL - allowable upper value,
A, B, C, .. – measurement intervals,
nA ... – amount of measurements in A interval;
measurement is in A interval if it is bigger or
equal to A interval threshold and smaller
than B interval threshold.
nB ... - amount of measurements in B interval;
measurement is in B interval if it is bigger or
equal to B interval threshold and smaller
than C interval threshold.
Thresholds are printed under histogram.
-NG - amount of measurements under allowable
lower value
+NG - amount of measurements above allowable
upper value
To finish work with this function and zeroing result
register press F key, then during "F..-StA" and
"F..-0" is displayed, press →T← key. This will
cause printing message about register zeroing.
Cooperation between statistics function with computer and printer.
The scale can be equipped with two RS232C connections marked as RS232C-I (computer) and
RS232C-II (printer). After each printer data printout, identical set of data is send to computer.
After S A CR LF (53h 49h 0Dh 0Ah) initialization signal is sent by computer, the scale sends to
computer statistical data contained in histogram.
Data: ... Hour. ...
SAMPLES
=
...
TOTAL MASS
=
...
AVER MASS
=
...
MIN MASS
=
...
MAX MASS =
...
MAX – MIN
=
...
S
=
...
SREL
=
...
*HISTOGRAM*
LSL
...
USL
...
DIV
...
-NG
...
LSL
A
nA
B
nB
C
nC
D
nD
E
nE
F
nF
G
nG
H
nH
I
nI
J
nJ
USL
+NG
...
A
~
...
B
~
...
C
~
...
D
~
...
E
~
...
F
~
...
G
~
...
H
~
...
I
~
...
J
~
...
...
Measurement made by : ____________
Maintenance and small repairs
1. Keep the balance clean.
2. Take care, for during the operation some impurities may enter under balance
base and into sensors vicinity. In case of impurities finding remove them.
3. In case of improper operation due to short power failure, switch off the balance
by unplugging the power cable from the socket, and plug it again after a few
seconds.
4. The message “Err-b”, displayed after switching on the unloaded balance,
indicates the balance sensor mechanical damage.
5. All repairs must be performed by the authorized personnel.
6. To perform the repair contact the nearest service point.
Emergency messages:
Message
Reason
Recommendation
C-1 ... 6 (over
1min.)
autotest negative result
contact the service
balance loaded during turning on
remove load from the
balance
balance sensor mechanical failure
contact the service
l
balance mechanical failure
contact the service
H
balance overload
remove load from the
balance
balance mechanical failure
contact the service
indicator does
not work
balance unstable,
base vibrations,
air blasts
place the balance in
location, assuring the
indications stability
balance damage
contact the service
- - - - - -
tare setting not finished
contact the service
- -
tare setting unsuccessful
(too low load or B/G pressed)
zero the balance or pres
B/G again
- -
zeroing with too high load
set the tare
Declaration of Conformity
We:
AXIS Spółka z o.o. 80-125 Gdańsk, ul.Kartuska 375B declare with full responsibility, that the
balances:
4B300F, 4B600F, 4B1500F, 4B2000F,4B3000F i 4B6000F
oraz 4B300FN, 4B600FN, 4B1500FN, 4B2000FN,4B3000FN i 4B6000FN
marked with CE mark are consistent with:
1. Standard PN-EN 61010-1:2004 Safety requirements for electrical equipment for
measurement, control and laboratory use. General requirements harmonized with the directive
73/23/EEC (Low Voltage Directive),
2. Standard PN-EN 55022:2000 Electromagnetic compatibility (EMC) – Information
technology equipment – Radio disturbance characteristics – Limits and methods of
measurement, and PN-IEC 61000-4-3 Electromagnetic compatibility (EMC) – Part 4-3:
Testing and measurement techniques - Radio-frequency electromagnetic field immunity test,
harmonized with 89/336/EEC directive (electromagnetic compatibility).
Moreover, balances with the following information on rating plates:
- sticker with two digits of year of conformity assessment and
with number of notified unit, performing the assessment,
- green metrological marking M,
- protective mark from notified unit
are made according to certificate of approval EC no. PL 04 020 and have EC verification,
confirming conformity with:
3. Standard PN-EN 45501 Specification for metrological aspects of non-automatic weighing
instruments, issued in December 1999, harmonized with the 90/384/EEC council directive,
changed by 93/68/EEC council directive).
Additional information
- Conformity evaluation for the Council Directive 73/23/EWG and 89/336/EWG were carried
out by Laboratorium Badawcze Oddziału Instytutu Elektrotechniki w Gdańsku, accredited to
PCA.
- Approval certificate no. PL 04 020, issued by Central Office Of Measures in Warsaw
(Notified Unit No. 1440).
Gdańsk, 9th of January, 2008
Authorized by the Director AXIS Sp. z o.o.:
Production manager mgr inż. Jan Kończak
_________________
Signature
08
Information’s concerning double-range scale (options)
1. General description
Double-range scale’s have capability of work with greatest accuracy in bottom
measuring range part. Weighing of smallest mass is more precise.
This type of scale’s have two measurement range:
-
Max1 - 50 % of maximum load (mostly),
-
Max2 – 100% of maximum load,
and adequate reads digit: d1 i d2 (d1 < d2).
Double-range options causes change of scale’s operation:
-
after turn on (in small mass range 0- Max1) scale displays result with reading
unit d1,
-
when the load pass the Max1 scale changes reading unit on d2; from this
moment scale displays result with reading unit d2 on all measure range
(0- Max2),
return to unit d1 is succeed after zeroing the scale (→0← key), or when all mass is
removed from pan (indicator "→0←").
2. Double-range scale parameters
Range and graduation values are represented on nominal table on the scale.
52
INSTRUCTION MANUAL
_____________________________________________________________________________________________________________________
NOTES