Keysight Technologies B2900B Series Quick Reference

Text of the manual

Keysight B2900B Series

Precision Source/Measure Unit Quick Reference Preparing the Instrument for Use To position the instrument: 1. Grab the handle by the sides and pull outward. 2. Rotate the handle. 3. Position the instrument. To turn the instrument ON: 1. Connect the power cord from AC input connector to an AC power outlet at your site. 2. Press the line switch. To set the power line cycle: 1. Press the More > System > PLC function keys. 2. Press the 50 Hz or 60 Hz key to specify the power line cycle at the site. To connect the device under test (DUT), optional: 1. B2900 terminals use banana jacks. Prepare test leads that use banana plugs. 2. Connect the test leads as shown in the figures. For using 4-wire connection, set the Sensing Type to 4-WIRE on the Output Connection dialog box. With the default setting, it is set to 2-WIRE. The dialog box is opened by pressing the Config > Source > Connection function keys. Latest Information To get the latest firmware, software, manuals, and support information, go to www.keysight.com. You can then search them by product number. 2-wire connection High Sense High Force Low Force Low Sense DUT 4-wire connection High Force Low Force DUT B2901BL/B2910BL B2901B/B2902B B2911B/B2912B

1. Line switch: Turns the instrument on or off.

2. USB-A connector: Used to connect a USB memory. 3. Auto key: Starts a repeat measurement or aborts the repeat measurement. 4. Trigger key: Starts a single measurement, aborts a repeat measurement, or initiates trigger system. 5. Assist keys: Five keys for setup assistance. Mode, Source, Limit, Measure, More, etc. 6. Numeric/alpha keys: Used to enter the value of setup parameters specified by the field pointer. 7. Rotary knob: In MOVE (blue) status: Turning it moves the field pointer. Pressing it fixes the pointer position. In EDIT (green) status: Turning it changes the field pointer parameter value. Pressing it fixes the value. 8. Left and right keys: In MOVE (blue) status: Moves the field pointer. In EDIT (green) status: Changes the field pointer parameter value. If the field pointer is on a numeric value entry field, pressing the key changes the pointer to a digit pointer. 9. Channel 1 terminals: High Force, Low Force, High Sense, Low Sense, Guard, and chassis ground 10. On/Off switch(es): Used to enable or disable the channel. Turns the channel off if it is in the output status even if it is in the remote status. Two switches on 2-channel models. The switch turns green if the channel is enabled. The switch turns red if the channel is in the high voltage state. 11. View key: Changes the display mode. 12. Cancel/Local key: Cancels the setup operation if the instrument is in the local status. Returns the instrument to the local status if it is in the remote status. 13. Function keys: Six keys for detail setup of several functions. Config, Function, Trigger, Result, File, Program, I/O, System, and More. 14. Channel 2 terminals: Only on 2-channel models. 15. GPIB interface connector: Connects to GPIB interface of an external computer or equipment. 16. USB-B connector: Connects to USB interface. 17. LAN interface connector: Connects to 10/100 Base-T interface. Left LED indicates activity. Right LED indicates link integrity. 18. Digital I/O connector: D-sub 25 pin female connector for general purpose I/O (GPIO). For trigger input/ output, interface to a handler, interface to an interlock circuit, etc. If the interlock terminals are open, the instrument output is limited to ±42 V. 19. AC input connector: AC power cord is connected to this receptacle. 1 3 5 6 7 9 12 11 13 14 10 2 8 4 15 14 16 19 18 17

LXI LAN status indicator

View mode 1 2 3 4 MOVE status EDIT status 1 Press to select MOVE status 2 EDIT status 3 4 Select Press to fix Fix R Editing the Setup Changing the View Mode Status Information

b. Press the Mode assist key.

c. Press the VOLTS (V) assist key. 1: Single 2 Set the Limit value. d. Press the Source assist key. e. Edit the value as follows or in the same way as for 2-b. a. Press the Limit assist key. Limit value is set to +10 mA. b. Edit the value as follows or in the same way as for 1-e. 3. Press the Ch1 switch to enable channel 1. This turns the switch green. Channel 1 starts applying the voltage specified by the Source value. Changing the setup changes channel output immediately. Performing Measurement Disabling the Channel Press the Ch1 switch to disable channel 1. This turns off the switch light. Measured value a. Press the View key to display the Single view. Press to start a single measurement. Press to start a repeat measurement. * You can change the measurement parameter by using the Measure assist key. B2900 applies a constant voltage of +500 mV by the following procedure. 1. Set the source mode and the Source value. Source is set to +500 mV. Applying DC Output This is the measurement example for a 1 kW resistor. (1) Change the value (2) Press the unit assist key to fix value (1) Move the digit pointer (2) Change the value (3) Press the rotary knob to fix the value

Performing Low Current Measurements

For performing low current measurements accurately, use the triaxial cable. It can reduce the affect of leakage current, external noise, and such caused by the extended measurement path. For using the triaxial cables, the Keysight N1297A/N1297B banana to triaxial adapter is required. This adapter can convert the B2900 source/ measure terminals to the triaxial connectors. 1. Press the key to display the Single view. 2. Set the Source mode, Limit value, and measurement mode as shown in "Applying DC Output". 3. Set the sweep parameters. a. Press the More assist key to change the assist keys. b. Press the Show Sweep assist key to display the Sweep setup parameters. c. Press the rotary knob to change the pointer status to EDIT (green). d. Press the LINEAR SINGLE assist key to set the linear single sweep mode. The source shape indicator shows the staircase icon, and the status changes to MOVE (blue). e. Use the rotary knob, arrow keys, or numeric/alpha keys to set the Start (sweep start), Stop (sweep stop), and Step (sweep step) or Points (number of sweep steps) values. 4. Press the key to display the Graph view. 5. Press the Ch1 switch to enable channel 1. This turns the switch green. Channel 1 starts applying the voltage specified by the Source value. 6. Press the key to start a single sweep measurement. The measurement results will be displayed on the graph. 7. Press the Auto Scale assist key to fit the trace in the graph scale. 8. Press the Ch1 switch to disable channel 1. This turns off the switch light. Performing Sweep Measurements This is the measurement example for a 1 kW resistor. B2900 applies a staircase sweep voltage and measures the current at each step voltage by the following procedure. G: Graph N1297A Bad connection Good connection (no space) Note for attaching the adapter on the B2900 terminals: Push the adapter in firmly until it locks in-place (< 1 mm spacing). If there is some space as shown in “Bad connection”, the contact is not enough.

1. Channel number. 1 or 2.

2. Latest measurement data 3. Over voltage/current status indicator (OV or OC) 4. High capacitance mode indicator (HC) 5. Remote sensing (4-wire connection) status indicator 6. Source function. VOLTS or AMPS. 7. Source shape indicator. DC, pulse, sweep, or pulsed sweep. DC does not show the indicator. 8. Source output value, 5½ digit resolution on B2901B/B2902B/B2901BL/B2910BL, and 6½ digit resolution on B2911B/B2912B 9. Limit (Compliance) value 10. Measurement speed. AUTO, SHORT, MEDIUM, NORMAL, LONG, or MANUAL. 11. Aperture time in seconds or PLC (power line cycle, number of power line cycles per measurement). Only for the MANUAL speed. 12. Primary measurement data 13. Secondary measurement data or limit test result Pass or Fail Single View Range Setup Sweep Setup Pulse Setup Trigger Setup Source Volts: Spot Measure Amps Source Amps: Spot Measure Volts Measure Ohms Constant voltage output range Current measurement range Constant current output range Voltage measurement range Resistance measurement range Sweep Parameters Start Stop Points Step Sweep mode. LINEAR SINGLE, LINEAR DOUBLE, LOG SINGLE, or LOG DOUBLE Sweep start value Sweep stop value Number of sweep steps Sweep step value Pulse Peak Delay Width Pulse output ON or OFF Pulse peak value Pulse delay time Pulse width Trigger Count Delay Period Trigger Trigger type. AUTO, SYNC, TIMER, or MANUAL Trigger count (number of triggers) Trigger delay time Trigger period Trigger source. AUTO, BUS, TIMER, INTn (n=1 or 2), LAN, or EXTm (m=1 to 14) Pulse base value is the same as the Source output value. (Available for B2901B/B2902B/B2911B/B2912B) (More > Show Sweep) (More > Show Pulse) (More > Show Trigger) 3

1. Graph display status ON or OFF. Only on 2-channel models. [n] is for channel n.

2. Y-axis data type I (A), V (V), R (W), P (W), or MATH 3. Y-axis scale LINEAR or LOG 4. X-axis data type I (A), V (V), R (W), P (W), MATH, t (s), V1, or V2 5. X-axis scale LINEAR or LOG 6. Graph maximum value 7. Graph minimum value 8. Channel 1 and/or 2 source output value, limit value, or none (controlled by the Ch n Source, Ch n Limit, or Hide Ch n assist key) 9. Channel 1 and/or 2 Y-axis data at the active X-cursor position. ----.---- is displayed for the no-data position. 10. Cursor data (controlled by the Show Cursors or Hide Cursors assist key) First line Positions and distance (e.g. I1, I2, ΔA) of Y-cursors 1 and 2 Second line Positions and distance (e.g. t1, t2, Δt) of X-cursors 1 and 2 1 2 3 4 5 6 7 8 9 10 Graph View 1. Displays status ON or OFF, on B2911B Displays status Ch 1, Ch 2, or OFF, on B2912B Two lines can be displayed on the graph. [1] indicates the graph setup for line 1. [2] indicates the graph setup for line 2. 2. Y-axis data type I, V, R, or P. 3. Y-axis scale per division A/div., V/div., W/div., or W/div. 4. X-axis scale per division s/div. 5. X-axis minimum value (minimum timestamp) 6. Y-axis offset values for line 1 and 2 7. X-axis maximum value (maximum timestamp) 8. See Graph View. 9. See Graph View. 10. See Graph View. Roll View 1 2 3 4 5 6 7 8 9 10 Displays the time domain graph for plotting the channel 1 and/or 2 measurement data. (Only on B2911B/B2912B) Displays the graph for plotting the channel 1 and/or 2 measurement or math result.

© Keysight Technologies 2020

Printed in Malaysia Part Number B2910-90100 What is the B2900 SMU? An SMU combines the capabilities of a current source, a voltage source, a current meter, and a voltage meter along with the capability to switch easily between these various functions into a single instrument. This gives it the ability to evaluate the IV characteristics of devices across all four measurement quadrants without the need for any additional equipment. In addition to its DC operation mode, B2901B/B2902B/ B2911B/B2912B also have the ability to perform pulsed measurements in order to prevent device self-heating from distorting the measurement results. - A - V Voltage source Current meter Current source Voltage meter 2-Wire and 4-Wire Connections To simplify the connections, use 2-wire connection by connecting the Force terminals only and opening the Sense terminals. For low resistance measurements and high current measurements, use 4-wire connection. Connecting the Force and Sense lines together at the terminal of the DUT minimizes the measurement error caused by the residual resistance of the test leads or cables. Grounded and Floating Measurements With the default setting, Low Force terminal is connected to the chassis ground. However, it can be disconnected from the ground for floating measurements. This setup is effective for differential voltage measurements which usually require two channels as shown in the Grounded 2 connection. DUT Grounded 1 DUT Grounded 2 High Force Low Force DUT Floating SMU High Force Low Force SMU High Force Low Force SMU High Force Low Force SMU For the floating measurements, set the Low Terminal State to FLOATING on the Output Connection dialog box. With the default setting, it is set to GROUNDED. The dialog box is opened by pressing the Config > Source > Connection function keys. With the FLOATING setting, the Low Force and Low Sense terminals can be connected to the maximum of ±250 V. WARNING: Potentially hazardous voltages of up to ±250 V may be present at the Low Force and Low Sense terminals. To prevent electrical shock, use accessories that comply with IEC 61010-031. The terminals and the extended conductors must be isolated by using insulation caps, sleeves, etc. DUT 4-Wire DUT High Force Low Force SMU 2-Wire High Force Low Force SMU High Sense Low Sense