Evikon PluraSens E2618-Cl2 Operation & User’s Manual

Text of the manual

Heavy greenish-yellow gas with a pungent, irritating odour. Highly toxic and

corrosive. Chlorine is used in chemical industry and for bleaching and desinfection. Chemical formula Cl2 Molar weight 70.9 Relative gas density (to air) 2.44 Conversion 1 ppm = 2.90 mg/m3 Boiling point -34.04 °C Odour Pungent, irritating smell Odour detection threshold Some individuals will not notice the odour until it is more than three times the exposure limit. Hazards Nonflammable, but a strong oxidizer, may react explosively with many common chemicals. Potent irritant of the eyes, mucous membrane, skin and respiratory system. Death can occur within minutes after exposure of 400 to 1000 ppm. Chronic exposure of 1 ppm can cause a moderate, but permanent, reduction in pulmonary function. Exposure limits (Directive 2006/15/EC) TWA not specified STEL 1.5 mg/m3 /0.5 ppm IDLH (NIOSH) 10 ppm Conversion of ppm to mg/m3 is calculated for 25°C and 1 atm. Installation guidelines (See Installation and connections section for general information.) Chlorine is considerably heavier than air and tends to sink. The sensor should be placed at ca. 15 cm above the floor. Calibration E2618-Cl2 transmitters have been calibrated by Manufacturer with standard gas mixtures before delivery. Provided that the sensor is used under moderate conditions,the recommended interval for field recalibration is 6 months. Please contact your dealer for more information. Maintenance Do not perform any maintenance operation with the power on. Clean the device with soft damp cloth. Do not use any abrasive cleaning agents. Do not immerse the device into water or any cleaning media. Delivery set –Transmitter E2618 (wall mount or duct mount version) – Mounting accessories: 4 screws with plastic dowels for wall mount version rubber flange for duct mount version. fixing clamp for remote probe version E2618-Cl2_UM_EN Rev 16.07.2019 Specifications Sensor type electrochemical cell Sampling method diffusion Typical detection range 0...10 ppm Maximum overload 100 ppm Resolution / digital unit 0.1 ppm Response time T90 <60s Sensor lifetime > 2 years Calibration interval 6 months Signal update every 1 second Load resistance RL < (Us - 2 V) / 22 mA for 4-20 mA RL > 250 kOhm for 0-10 V mode Digital interface RS485, Modbus RTU protocol no galvanic isolation Power supply options 11...30 VDC or 24 VAC Power consumption < 2 VA Analog outputs 2 × 4-20 mA / 0-10 V, user settable Outputs assignment OUT1 2 gas; OUT2 2 gas NOTE For LCD version only OUT1 is available. Cable connections screwless spring loaded terminals Enclosure grey ABS, wall or duct mount, protection class IP65 Dimensions H82 × W80 × D55 mm CE marking according to 2014/30/EU and EN61326-1 requirements Operating conditions -20...+50 °C, 15...90 %RH, atmospheric pressure ±10% non ATEXrated areas, non-aggressive atmosphere NOTE The sensor may exhibit considerable sensitivity to hydrogensulfide, nitrogen dioxidу and sulfur dioxide LCD option Operating temperature 0...+50 °C Display dimensions 72 × 36 mm Number of digits 3.5 7-segment Character height 14 mm Other features Backlight Other options Remote probe Protection IP65, default cable length 3.0 m; max height 80 mm, max diameter 65 mm Warranty This product is warranted to be free from defects in material and workmanship for a period of one year from the date of original sale. During this warranty period Manufacturer will, at its option, either repair or replace product that proves to be defective. This warranty is void if the product has been operated in conditions outside ranges specified by Manufacturer or damaged by customer error or negligence or if there has been an unauthorised modification. Evikon MCI OÜ Teaduspargi 7, Tartu [email protected] Tel. +372 733 6310 50411 Estonia www.evikon.eu PluraSens® Chlorine Transmitter E2618-Cl2 User Manual

E2618-series transmitters belong to the PluraSens® family of multifunctional

measurement instruments. The instruments utilise gas sensors of various types with excellent repeatability, stability and long lifetime The devices are supplied either in duct-mount or wall-mount version. The wall mount version of the device is available with remote probe. The remote probe is connected to the main unit with shielded cable. Default connection cable length is 3 m. E2618 series provides two independent analog outputs OUT1 and OUT2, user- selectable to 4-20 mA or 0-10 V, proportional either to gas concentration or temperature. RS485 Modbus RTU digital communication interface allows easy instrument configuration and integration into various automation systems. A design with LCD indicator is available as an option. The version of your transmitter is marked on the package. Safety requirements Always adhere to the safety provisions applicable in the country of use. Do not perform any maintenance operation with the power on. Do not let water or foreign objects inside the device. Operating conditions The device should be used in explosion-safe (non ATEX -rated) indoor areas, without aggressive gases in the atmosphere. See Specification table for more details. Installation and connections There are no precise rules or standards to follow when installing the gas detectors. The following points must be taken into account: application (air quality control or leakage detection), properties of the space under investigation (room geometry, direction and velocity of air flows etc), detected gas (relative density to air, whether the gas is flammable, or toxiс, or oxygen displacing), safety: strong vibrations, mechanical shock, and the sources of strong electromagnetic interference should be avoided, the device should be accessible for maintenance and repair. For early leakage detection install the sensor as close as possible to the potential leakage sources (flanges, valves, pressure reducers, pumps, etc), taking into consideration other points listed above. For general area monitoring without definite leakage sources, the detectors should be distributed evenly in the room. For personal safety control the detectors are installed in the breathing zone (at the height of the head of people or animals). Recommended sensor position is vertical, pointing downwards. See Installation guidelines section on the next page for more information. 1. Wall mount version: Unscrew four lid screws and detach the lid from the transmitter. Attach the device to a wall with screws passing through mounting holes (see dimensional drawing below). Duct mount version: Cut hole with a diameter of 36...45 mm in the air duct at the chosen mounting place. Place the rubber flange aligning the holes in the flange and the air-duct and fix the flange with four self-tapping screws. Pass the sensor probe through the flange and adjust it to the appropriate depth. Unscrew four lid screws and detach the lid from the transmitter. 2. Use two M16 cable glands to pass the cables of the power supply and of the external devices. Attach the power cable to the device without turning it on. Using the connection diagram below, connect the analog outputs and digital interface terminals to the relevant devices according to your tasks. The screwless quick connect spring terminals on the E2618 series devices are suitable for a wide range of wires with cross-section 0,2...1,5 mm2. We recommend to strip the wire end by 8...9 mm and tin it, or to use the wire end sleeves. To connect the wire, insert the wire end into terminal hole. To disconnect, push the spring loaded terminal lever, pull the wire out, and release the lever. Use twisted pair cable, e.g. LiYY TP 2×2×0,5 mm2 or CAT 5, to connect the device to RS485 network. Use one pair for A and B wires and the second pair for common 0 V and power +U wires to connect the transmitter to Fieldbus network. Respect polarity. Overall length of all connections via RS485 interface should not exceed 1200 m. The type of each analog output can be independently changed between 4- 20 mA and 0-10 V with jumpers J1 (OUT1) and J2 (OUT2). With closed jumper the output is 0-10 V, with open jumper the output is 4-20 mA. By default both outputs OUT1 and OUT2 are assigned to gas concentration. The device has built-in temperature sensor which may be tied to any of the outputs. The output assignments and scales can be changed by Modbus commands. We recommend to set the difference between the upper and bottom limits of the output scale not narrower than 20% of detection range (for CO detectors the scales down to 5% of range are allowed). In any case, do not set the output scale below the tenfold resolution of the device. NOTE If you use a version with LCD, only OUT1 is available. 3. Turn on the power. The sensor heating up may take up to five minutes after switching on. During this period relays, analog outputs and Modbus interface are off. A LED placed on the PCB of the device allows to control the connection process. The LED response to different processes is presented in the table below. Process LED mode Sensor heating period Blinking 0.5 Hz (50% on, 50% off) Sensor absence or malfunction Blinking 0.5 Hz (90% off, 10% on) Modbus response The signal is modulated with short on-off pulses, even single Modbus cycle is traceable* Normal measurement Continuous light 4. Make sure that the transmitter is properly mounted, the external devices connected, power on and control LED is constantly lit. Place the lid back and fix it with the screws. The device is ready to use. It is recommended to keep the device on constantly, except for periods of maintenance and calibration, deplacement etc. Sensor probe handling The wall mount version of the transmitter is available with remote probe (see drawing below for dimensions). The remote probe is connected to the main unit with shielded cable. Default connection cable length is 3 m.The sensor probes of all types are equipped with a hydrophobic microporous PTFE filter to protect the sensor from dust, dirt and water drops. The filter may be replaced if it gets strongly contaminated. To replace the PTFE filter, unscrew the M25 nut and remove the old filter. Place a new filter into the nut and tighten it again. NB! Never stab or press the filter near its centre where the sensor is located since this may damage the sensor. The recommended orientation of sensor probe is vertical with the sensor tip pointing downwards. This prevents possible accumulation of condensed water on the sensor protection filter. The horizontal orientation is also suitable. Avoid upward position of the sensor tip. Configuring Gas transmitters E2618 share all functionalities of the PluraSens® multifunctional transmitter platform. The features and options include: - digital output change rate limiting filter - digital integrating (averaging) filter - temperature measurement channel with internal sensor - free assignment of each analog output to chosen parameter - flexible setting of analog output scales for each output - output shift and slope adjustment for calibration - Modbus controlled forced state option for analog outputs . E2618 can be configured through its RS485 interface by Modbus RTU commands. A standard configuration kit includes Model E1087 USB-EIA485 converter and a software pack. Please contact your Seller or the Manufacturer for more information. Return to default settings To reset the device’s Slave ID, baudrate and sbit number to factory settings, proceed as follows: 1. De-energize the device 2. Connect the J3 jumper 3. Turn on the device 4. De-energize the device 5. Disconnect the J3 jumper 6. Turn on the device RS485 communication interface See Annex 1 36 77 Ø35...36 Ø50 120° Ø 65 80 3 27 Ø34 max Hole 35 OUT1 OUT2 0V A B +U J3 J1 J2 Fieldbus Controller Supply Power Control LED Input 1 4-20 mA / 0-10 V Input 4-20 mA / 0-10 V 2 0V / GND 0V +24V J1: OUT1 type (open: 4-20 mA; closed: 0-10 V) J : OUT type 2 2 (open: 4-20 mA; closed: 0-10 V) J : 3 return to factory settings SENSOR 55 80 130 82 Ø4 mm M25 M16 50 70 200 M16 80 Ø35 2 0 3 M25 35 82 55

E2618_UM_EN Annex 1. Modbus RTU Communication Reference

RS485 communication interface Databits: 8 Parity: none Stop bits: 1 or 2 Protocol: Modbus RTU Supported Modbus functions: 03 - read multiple registers 06 - write single register Communication parameters Parameter Permitted values Default Supported baudrates 1200, 2400, 4800, 9600, 19200, 38400, 57600 9600 Data bits 8 8 Parity none none Stop bits 1, 2 1 Protocol Modbus RTU Modbus functions 03 - read multiple registers 06 - write single register Error codes 01 - illegal function 02 - illegal data address 03 - illegal data value 04 - slave device failure (details of last error 04 can be read from register 0x0008) Modbus holding registers Register addresses are shown 0-based, Addr in hexadecimal, Reg in decimal format. Modbus holding register numbers MHR are shown in decimal 1-based format, and may be addressed either from 00001 or 40001 base. Addr Reg / MHR RW Description Supported values (dec) Default 0x0001 1 R Hardware version - 0x0002 2 R Software version - 0x0003 3 R Product serial number 1...65535 - 0x0004 4 RW Slave ID (net address) * 1...247 ** 1 0x0005 5 RW Baudrate * 1200, 2400, 4800, 9600, 19200, 38400, 57600 9600 0x0006 6 /40007 RW Response delay, ms 1...255 10 0x0007 7 / 40008 RW Stop bits * 1: no parity bit, 1 stop bit (default after factory reset) 2: no parity bit, 2 stop bits 3: odd parity, 1 stop bit 4: even parity, 1 stop bit NOTE: 3 and 4 are available starting from the Sofrware version 0x218 (dec. 536) 1 0x0008 8 /40009 R Last error code 1...255 - 0x0011 17 / 40018 RW Restarts counter write ‘42330’ to restart device (no response will follow) - 0x0097 151 /40152 R Sensor type code 0...65535 0 0x0098 152 /40153 R Output units code 0 - ppm, 1 - ‰, 2 - % 0 * - The new value is applied after restart. ** - Broadcast slave ID 0 can be used to assign a new ID to device with unknown ID. When addressing by ID 0 the device shall be the only Modbus instrument in the network. The device will not respond to Master command when addressed by ID 0. *** - This value is dynamic and not kept in EEPROM after restart

E2618_UM_EN Annex 1. E2618 series Modbus holding registers (part 2)

Register addresses are shown 0-based, Addr in hexadecimal, Reg in decimal format. Modbus holding register numbers MHR are shown in decimal 1-based format, and may be addressed either from 00001 or 40001 base. Addr Reg / MHR RW Description Supported values (dec) Default 0x00A2 162 / 40163 RW Zero adjustment for temperature data, °C × 100 -32000...+32000 (-320,00...+320,00 °C) 0 0x00A5 165 / 40166 RW Zero adjustment for gas data, ADC -32000...+32000 ADC units 0 0x00A6 166 / 40167 RW Slope adjustment for gas data 1...65535 512 0x00A7 167 / 40168 RW Change rate limit for gas data, ppm (‰ for O2) / s 1...32000, 0 - no limit 0 0x00A8 168 / 40169 RW Integrating filter time constant, s 1...32000 (seconds), 0 - no filter 0 0x00C9 201 / 40202 RW Parameter tied to analog output 1 0-none 1- temperature 2- gas concentration 9- forced Modbus control, value set in MHR / 40204 2 0x00CA 202 / 40203 RW Parameter tied to analog output 2 0-none 1- temperature 2- gas concentration 9- forced Modbus control, value set in MHR / 40205 2 0x00CB 203 / 40204 RW Forced value for analog output 1*** 0...1000 (0,0%...100,0% of output scale) 0 0x00CC 204 / 40205 RW Forced value for analog output 2*** 0...1000 (0,0%...100,0% of output scale) 0 0x00FF 255 / 40256 RW Sensor, analog outputs, LED and buzzer status bit[0]=0/1 - sensor present/absent, read-only! bit[1]=0/1 - analog outputs deactivated/activated, bit[2]= 0/1 - in case the sensor is absent, turn signaling off/on analog output1, bit[3]=0/1 - in case the sensor is absent, turn on signaling with low current/high current on analog output1; if bit[2]==0 this bit will be ignored, bit[4]=0/1 - in case of sensor absent, turn signaling off/on analog output2 bit[5]=0/1 - in case of sensor absent, turn on signaling with low current/high current on analog output2; if bit[4]==0 this bit will be ignored, bit[6]=0/1 - current/voltage output detected on output1, read-only! bit[7]=0/1 - current/voltage output detected on output2, read-only! bit[8]=0/1 - LED deactivated/activated (always 0 for E2618), bit[9]=0/1 - buzzer deactivated/activated (always 0 for E2618) - 0x0100 256 / 40257 R Raw temperature data, °C×100 signed integer, -4000...+8500 (-40,00...+85,00 °C) 0x0101 257 / 40258 R Raw gas sensor data ADC data 0...4095 0x0102 258 / 40259 R Measured temperature, °C×100 signed integer, -4000...+12500 (-40,00...+125,00 °C) 0x0103 259 / 40260 R Gas concentration, ppm / ‰ signed integer, -32000...+32000 (ppm / ‰) 0x0105 261 / 40262 RW 0% value for analog output 1 signed integer, -32000...+32000 (ppm / ‰) 0 0x0106 262 / 40263 RW 100% value for analog output 1 signed integer, -32000...+32000 (ppm / ‰) 1000 0x0107 263 / 40264 RW 0% value for analog output 2 signed integer, -32000...+32000 (ppm / ‰) 0 0x0108 264 / 40265 RW 100% value for analog output 2 signed integer, -32000...+32000 (ppm / ‰) 1000 * - The new value is applied after restart. ** - Broadcast slave ID 0 can be used to assign a new ID to device with unknown ID. When addressing by ID 0 the device shall be the only Modbus instrument in the network. The device will not respond to Master command when addressed by ID 0. *** - This value is dynamic and not kept in EEPROM after restart