Dwyer Instruments CSG Operation & User’s Manual
Read Online
Read the Dwyer Instruments CSG Operation & User’s Manual online using the built-in PDF viewer.
Text of the manual
Accuracy at 25°C
0.0 to 20.0mA DC in 1 mA steps
0.0 to 10.0V DC in 1V steps
Current
+/- 0.05mA DC
Voltage:
+/- 0. 0SV DC
Load Impedance
Current:
300 ohms maximum
Voltage:
1000 ohms minimum
External mA Transmitter Source
Voltage:
12V minimum to 30V peak maximum
RAMPING
Manual
Auto
Values:
Step Time:
DISPLAY
Anywhere in range
User adjustable Min. & Max. values
anywhere within range
User adjustable from 1 to 20
seconds, in 1 second increments
Numerical
LCD, 00.0 to 99.9, 0.59"H (15mm)
Annunciators
"RAMP", "OVER LOAD"', "mA", "V'.
''TIME". and symbol for low battery
Blue Backlight
Energized with momentary side
switch
ENVIRONMENT
Operating
POWER
9V Battery
Low Battery
120 V AC
32 to 122°F (0 to 50°C)
NEDA1604. JIS006P. or IEC6F22
(Duracell MN1604 or equivalent)
Up to 30 hours continuous use
Indicator comes on below
approximately 7 volts
Plug-in 9V DC output supply for
bench-top use
Auto Power Off
User adjustable delay time from
MECHANICAL
Dimensions
Weight
1 to 20 minutes after last key press
Can also be disabled
5.5"H x 2S-W x 1.4"0
(140H x 63.5W x 35.60 mm)
6 oz (170 g)
USER'S GUIDE
ModelCSG
Current & Voltage Calibrator
l'lw9e
20□
RAMP
P6'ffl
mA
•
¥i1.1::
-----✓)
• • •
mA
mA
·,n.ilb
.. i.;•
;.,.,,i.
··, ,·pq·
;"4J,1•.,1 .,.
.. \·
!
Copyright @ 2006. All rights reserved including the
right of reproduction in whole or in part in any form
Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176
TestEquipmentDepot.com
Make sure that the battery is installed or the AC adapter is
plugged into the EXT POWER jack. The AC adapter disconnects the
internal battery.
To turn the CSG on, press and hold the upper red ON/OFF but
ton for at least 2 seconds. The LCD annunciators will cycle and then hold
steady when the CSG is turned on.
To tum off, press and hold the ON/OFF button for 5 seconds
until "TIME" annunciator is displayed.
SELECTING OUTPUT MODE
Plug the test lead assembly into the appropriate output jack
as shown below. If the lest leads are plugged into the mA XMTR or mA
OUTPUT jack, or no leads are plugged in. ·mA" will tum on. If the load
resistance is too high or the leads are left open, "OVERLOAD" will tum on.
If the leads are plugged into the VOLTS OUTPUT jack. ··v will turn on. If
the load resistance is loo low or the leads are shorted. "mA" may stay on
or turn on.
Fig 1. Voltage Output
Fig. 2 mA SourƘe Output
LOAD
LOAD
+
+
._ ______
....., _
__,
VOLTS
BLACK
BLACK
mA
OUTPUT
Fig 3. Simulate a 2-Wiro mA T ransmitter Loop With External 12 to 30V
DC Supply. A similar connection can be used to extend output range of
CSG beyond 300 ohms A 24V source will allow CSG to source into an
impedance of >1000 ohms.
EXTERNAL
LOOP SUPPLY
+ 12 to 30V _
mA
XMTR
+
LOAD
BLACK
MANUAL SELECTION OF OUTPUT LEVEL
To increase the output. press the 'fr key. To reduce the output,
press the \ key. New output value will be displayed on the display.
AUTO RAMPING
The output is preprogrammed lo ramp between 4 to 20 mA(or 2
to 10V) and back in 1.0 mA(or 1.0 V) steps, with 5 seconds between each
step. See below to change the ramp end points and step time interval.
To start ramping the output, press the RAMP/TIME key for 2
seconds. "RAMP" will tum on and the output will start ramping within the
minimum and maximum values programmed into the CSG. It will start at
the output value manually set on the CSG before the key is pressed.
To stop the ramping and return to manual mode. press the
RAMP/TIME key again.
CHANGING THE AUTO RAMPING END POINTS AND STEP TIME
Press and hold the RAMP/TIME key. and immediately (within 2
seconds) press the -0- key without releasing the RAMP/TIME key until
"RAMP" and "TIME· tum on. "mA" will tum off. Set the time duration for
eacJ;i step increment from 1.0 to 20.0 seconds, in 1.0 second increments
by using the 'fr and \ keys.
To store the desired time interval. press the RAMP/TIME key
again. "TIME" will turn off and the appropriate "mA" or -V will flash.
"RAMP" will remain on.
Set the first end point by using the 'fr and \ keys. To store
the first end point. press the RAMP/TIME key again. The appropriate "mA"
or ·v· and "RAMP" will flash.
Set the second end point by using the 'fr and \ keys. To
store the second end point and exit the programming mode, press the
RAMP/TIME key again. T he second value stored will be displayed along
with the appropriate "mA" or ·v.
SETTING THE AUTO POWER OFF DELAY TIME
With the unit ON, press and hold the RAMP/TIME key, and im
mediately (within 2 seconds) press the 'fr
key without releasing the
RAMP/TIME key until "TIME" turns on. All other annunciators are off. Set
the time delay from the last key press until the unit automatically turns
OFF by using the 'fr and -0- keys The time delay can be set from
1 O to 20.0 minutes In 1.0 minute ,ncremenls. If the delay time Is set lo
00.0 minutes. the Auto Power Off Delay function is disabled This is useful
when the unit is used on a bench with the AC adapter
To store the desired time delay. press the RAMP/TIME key
again ... TIME" will turn off and the appropriate "mA" or ·v· will turn on