Andatech SOBERLIVE FRX Operation & User’s Manual
Text of the manual
Andatech SOBERLIVE FRX
Facial Recognition Breathalyser with Temperature Screening
User Manual
Version 1.5. Updated December 2020.
Thank you for purchasing the Andatech® Soberlive FRX breathalyser.
An essential for workplaces, the Andatech® Soberlive FRX is an all-in-one breathalyser, temperature
scanner and facial recognition ID to screen employees and visitors before they even enter your
premises. With the ability to be connected to any access control system, the Soberlive FRX can also
automatically restrict access to those who are running a fever or are over your workplace policy's
alcohol limit. The first line of defense to protect workplace safety, the Soberlive FRX saves valuable
time, eliminates direct physical contact and reduces the need for extra manpower when it comes to
temperature and alcohol screening.
We hope you enjoy owning your Andatech® Soberlive FRX breathalyser. If you have any questions
regarding the unit or want to speak with us further, please visit andatech.com.au or call 1300 800 200
(between 9am and 5.30pm Monday to Friday).
CONTENTS
INTRODUCTION
2
CONTENTS
2
DEVICE FEATURES
4
LCD SCREEN
6
SPECIFICATIONS
7
DEVICE INSTALLATION
8
GUIDELINES
8
WALL-MOUNTING THE DEVICE
8
ANDALINK
9
SETUP INSTRUCTIONS / ANDALINK MANUAL
9
ADMIN SETUP
9
ANDALINK SYNC
9
FORCE SYNC
9
ABOUT TEMPERATURE CALIBRATION
10
TEMPERATURE CALIBRATION PROCEDURE
10
ADMIN TEMPERATURE CALIBRATION (ON DEVICE STARTUP)
10
INTERVAL TEMPERATURE CALIBRATION
10
MANUAL TEMPERATURE CALIBRATION
10
CALIBRATION OPERATORS
11
2
Alcohol Test Percentage
15
Recognition (Device screen will show ‘Blow’)
16
Temperature
18
Enrolment
20
Device Info
20
Record
21
Date & Time
22
Language
22
Voice
23
Customize
24
Printer
25
Temperature Calibration
25
Calibration interval
25
Manual calibration
26
Show alcohol test reading
26
TROUBLESHOOTING / WARNING DISPLAY
27
FACIAL RECOGNITION GUIDELINES
29
FAQs
29
BREATHALYSER DISCLAIMERS
30
MEASUREMENTS & CONDITIONS FOR ALCOHOL TESTING
31
LIABILITY FOR PROPER FUNCTION OR DAMAGE
31
WARRANTY
31
BREATHALYSER CALIBRATION & SERVICE
31
HOW TO REMOVE/REPLACE SENSOR FOR CALIBRATION
32
WARRANTY DETAILS
33
3
1 x Soberlive FRX
1 x Power Cable
1 x RS232 Cable
2 x Straw Adapters
5 x Sampling Cups
2 x Keys (unlock to connect to bracket)
2 x Screws
2 x Wall Plugs
4
Indication of BAC
0.00 - 0.40% BAC
Display
LCD screen
Units of Measure
BAC: %BAC, g/100mL,
‰, mg/100mL, g/L
BrAC: mg/L, ug/100mL,
mg/mL, mg%, mcg/L,
g/210L
Infrared
Temperature
Sensor
Accuracy within ±0.3°C,
< 1 second response rate
Warm up time
30 - 60 seconds
Camera
100W pixels
Response time
5 seconds
Connection Port
USB
Sample time
4 seconds of continuous
blow
Network
WIFI, 4G (optional),
Ethernet
Recovery time
30 seconds after a
positive test
Calibration
Unlimited tests, every 6
months
Accuracy
±5% at 0 - 0.10% BrAC
±8% at 0.10 - 0.15%
BrAC
±10% at >0.15% BrAC
Ambient
Conditions
Operation: -10 to 50°C
Storage: -20 to 50°C
Humidity: 20% to 95%RH
Memory (built-in)
10,000 registered faces,
2,000 test results
Power supply
AC 100-240V 50-60Hz,
DC 12V 3A
Andalink memory
(cloud storage)
20,000 registered faces,
5,000 test results
(upgrade options
available)
Weight
1243g
Sensor
Replaceable fuel cell
sensor
Dimensions
20cm x 22cm x 8.8cm
Mouthpieces
Replaceable sampling
cup (passive testing) or
straw adaptor (specific
testing)
Warranty
1 year
Install the device only indoors, in a covered environment.
●
Install the device in a location with sufficient lighting
●
Do not use the facial recognition feature under a strong light.
●
Do not expose the device to the sun for a prolonged period of time.
●
Do not use in locations that are subject to high humidity. Excessive humidity may cause
damage to the device's components.
●
Do not install in locations with large temperature fluctuations throughout the day as this may
affect temperature calibration and body temperature results.
●
Choose an installation location that has a regular, room temperature - away from direct
sunlight, excessive heat or cold air (e.g. directly facing an air conditioner). Excessively hot or
cold temperatures may affect body temperature readings.
●
Use our provided power adapter- connecting it to an improper power source may damage
your equipment and lead to fire and/or damage.
●
Keep the camera clean for optimum photos and facial recognition.
●
Device should be installed facing directly forward with as little tilt or inclination as possible to
ensure the best body temperature results.
WALL-MOUNTING THE DEVICE
To mount on the wall, use the appropriate
screws (2) and wall anchors (2) for the type
of wall material you are mounting the device
to.
Drill two holes on the wall (depth: 50mm,
distance between: 34mm). Put the wall
anchors inside the hole, and push the screw
inside the wall anchors. Reserve around a
10mm gap from the wall, and fix the device
into place, then hang the device.
7
The Andalink is a secure, cloud-based software that bridges the Soberlive FRX with a complete data
management system. With Andalink, company and OH&S managers can register users and faces, as
well as view, download, and manage reports for their employees' alcohol and temperature results.
SETUP INSTRUCTIONS / ANDALINK MANUAL
For instructions on how to set up your Andalink account for your Soberlive FRX, please visit
andatech.com.au/andalink-manual to see the updated Andalink manual and log in with the
credentials provided in the letter with your device during purchase.
ADMIN SETUP
Important: Before operating the Soberlive FRX, please create at least one (1) staff that has the user
role Admin for each location/device.
By default, an Admin staff is required to conduct the Admin Temperature Calibration on the device
during startup.
You can disable this requirement or add a new Admin user on the device under Advanced Setting.
You can also add new Admin users online on Andalink (refer to Andalink manual above).
ANDALINK SYNC
Test results from the device are synced to Andalink immediately (as long as there is an active internet
connection on the device).
All other changes from Andalink to the device are synced every 10 minutes.
FORCE SYNC
To Force Sync the device with Andalink, connect a mouse to the device.
Then click and hold the cursor on the version number (displayed on the bottom right of
the device screen).
The server (cloud) icon will start rotating and display a downward arrow icon to indicate
that it is syncing.
The Force Sync is complete when the icon stops rotating. The number beside the icon will be updated
to show the updated number of registered faces synced from Andalink, and a pop-up message will
appear in the Notification Area that the Force Sync is complete.
8
ABOUT TEMPERATURE CALIBRATION
The Temperature Calibration is required to evaluate and adjust the precision and accuracy of the
Soberlive FRX's temperature measurements.
TEMPERATURE CALIBRATION PROCEDURE
To perform a Temperature Calibration, a calibration operator stands in
front of the Soberlive FRX and faces the camera.
The device will then take their temperature and use that as the
guideline for 'normal body temperature'.
Note: It is recommended that the calibration operator first check their
body temperature (e.g. with a separate accurate, reliable handheld
thermometer) to make sure that it is normal (within 36~36.5°C) before
performing the Temperature Calibration.
ADMIN TEMPERATURE CALIBRATION (ON DEVICE
STARTUP)
Each time the device is switched on, it takes 10 minutes to warm up
and will then require a Temperature Calibration.
This can only be performed by an Admin user or by any user (if enabled under Advanced Settings).
Note: The device will auto-restart daily at 4AM (device time) by default - this setting can be changed in
Device Settings. Any software updates are done automatically during restart as long as your device is
connected to the internet.
INTERVAL TEMPERATURE CALIBRATION
During normal operation, the device will also require a
Temperature Calibration at regular intervals (recommended /
default setting: 60 minutes)
When the device requires Temperature Calibration, the Thermal
Imaging Display will turn completely orange and the Notification
Box will show 'Please contact calibration operator' if anyone
attempts to take a test.
A normal test cannot be taken until a calibration operator
performs the Temperature Calibration.
MANUAL TEMPERATURE CALIBRATION
If a temperature calibration is needed outside of its regular
interval period (e.g. temperatures seem too high or too low, or is
measuring inconsistently), a manual temperature calibration can
be performed.
9
To do this, connect a mouse to the device and enter the Device Settings Menu > Temperature
Calibration and click on 'Start Manual Calibration'. Then, follow the steps under 'Temperature
Calibration Procedure'.
CALIBRATION OPERATORS
By default, only a user with an Admin role (set in Andalink) can
conduct the Interval and Manual Temperature Calibration.
The calibration operator for these can be changed to a non-Admin (any user) under Advanced
Settings (Temperature) by un-ticking the checkbox for 'Temperature calibration by operator only'.
The calibration operator for Admin Temperature Calibration (on device startup) can also be changed
to any user under Advanced Settings (Temperature).
Refer to Advanced Settings (Temperature) for more information.
10
The operating temperature range is between -10°C to 50°C. Using the device within this
temperature range will limit condensation from a user’s breath to ensure an accurate reading
from a user’s breath sample.
2.
It is recommended that the user does not consume anything for 20 minutes before the breath
test. Following this procedure will eliminate possible misreadings from mouth alcohol and
food.
3.
The device can be used with either a sampling cup or a straw adapter and regular drinking
straws (not provided). Set the alcohol test type to either Quick Test Mode or Standard Test
Mode under Advanced Settings (Blow) and install either the sampling cup or straw adapter
onto the device before using.
a.
The sampling cup is recommended for Quick Test Mode. The sampling cup should be
cleaned with a non-alcohol disinfectant and wiped dry between each test to prevent
cross contamination or alcohol residue from affecting the next reading.
b.
The straw adaptor is recommended for Standard Test Mode. Users should be
provided with individually wrapped regular drinking straws (not provided) to take a
test. Each user should use a new straw for each test and dispose of the used straw
hygienically after testing.
4.
After switching on the device, allow 10 minutes for the device to warm up before performing
the Admin Temperature Calibration.
11
Follow the device's instructions shown in the Notification Area when taking a test.
1.
'Face camera. Keep face inside the frame'
Stand in front of the device, face the camera and position face within the Facial Recognition
Frame shown on the LCD display.
2.
Device will analyse the results and display them in the Notification Area.
a.
If the temperature reading is above the Body Temperature Limit, an alert will sound
and the test procedure ends.
b.
If the temperature reading is below the Body Temperature Limit, the test will proceed
to the next step.
Note: Body Temperature Limit is set under Advanced Settings.
3.
Device will then prompt the user to take an alcohol breath test. Follow the instructions in the
Notification Area:
4.
Device will analyse the results and display them in the Notification Area along with a voice
alert (if enabled) as “Pass/Fail” “Name/Visitor” “Body temperature” “Alcohol reading” .
Note: Pass/Fail alcohol limit is 0.00 by default and cannot be changed. Any result over 0.00
will be a 'Fail'.
5.
Results are synced to Andalink immediately after each test (if device is connected to the
internet).
6.
A new user can take another test by repeating from Step 1.
Note: Check ‘Test Delay’ under Advanced Settings menu for the time delay before the next
test can be taken.
12
'Please Blow'
Quick test mode - Blow fast and hard into the
sampling cup from a distance of 3-5 cm without
touching the sampling cup
Standard test mode - Insert a drinking straw into the
straw adaptor and blow slow and steadily into the
straw
'Blowing'
(only displays during standard test
mode)
Continue blowing into the straw
'Blow completed. Please Wait'
Stop blowing and wait for the results
To access the Device Settings Menu, plug in a USB
mouse into the device and long press (click + hold) the
icon on the bottom left of the screen.
Calibration
Access is restricted to authorised calibration centres only.
Network
There are 3 methods of connection: Ethernet, Wifi, and 4G (additional module)
Click on any method of network connection to activate or deactivate it. Once a network preference
has been selected and activated, the corresponding icon will turn blue.
13
Ethernet
Plug in ethernet cable to
connect to the network
automatically.
Wifi
Select SSID and input
password to connect to WIFI
network.
4G
With the device turned off,
insert SIM card into the SIM
slot. Then, turn the device on.
Enter the password 978165 to access the Advanced
Setting menu for Recognition, Temperature and
Enrollment.
Alcohol Test Percentage
Set the percentage of registered users to randomly test for
alcohol.
or example: When this is set to 100%, all registered users
will be tested for their alcohol reading. If this is set to 50%,
then 50% of all registered users will be randomly tested for
an alcohol reading.
14
Recognition (Device screen will show ‘Blow’)
●
Disable Alcohol Test - Tick this to disable alcohol testing
on the device. Untick to enable alcohol testing on the
device.
●
Quick test mode - Tick this for fast testing using a
sampling cup. Only a short breath sample is required to get
a result. The provided sampling cup is recommended for
use with this mode.
●
Standard test mode - Tick this for more accurate alcohol
testing. A longer breath sample is required to get a result.
The provided straw adaptor is recommended for use with
this mode, and users should be provided with clean,
regular drinking straws (not provided) for individual testing.
○
Min.blow time(s)
■
Set the minimum blow time required for
sampling (from 3.0s and above).
■
Applicable only for Standard test mode.
Note: In both Quick and Standard test modes, the alcohol test readings will be shown (e.g.
0.02%BAC) unless this setting is disabled under 'Show alcohol test reading' in the Settings
menu.
●
Tail Sampling
Read the FAQs section for an explanation of Tail Sampling.
○
Tick to enable taking the test sample only from the last few seconds of the blow.
○
Untick to disable tail sampling.
●
Blank Test
○
Tick to enable an air blank test before each new test. The device does this to ensure
the mouthpiece and sensor are clear of alcohol from any previous tests that may
affect the next reading results.
○
Untick to disable blank test.
●
Pass Limit
○
Set the pass limit for the alcohol test result on this device. Any test result above this
set limit will be recorded as a “FAIL” in the device and Andalink.
○
Important: Please ensure that the limit is set the same in Andalink (see KB article:
Ensure set alcohol / temperature limits are the same in Andalink and device)
15
Enable Mask Detection
○
Tick to enforce all users to wear a mask. If a user is
detected as not wearing a mask, device will notify
'PLEASE WEAR A MASK' and sound an alert.
Users cannot take a test unless a mask is detected.
■
Note: If Mask Detection is enabled, the setting 'Visitors Allowed' must also be
enabled as the device cannot conduct any facial recognition on registered
staff wearing masks.
○
Untick to disable mask detection.
●
Visitors Allowed -
○
Tick this to enable non-registered users (visitors) to use the device.
■
Device will display the user ID as 'Visitor'.
■
Andalink will store visitor's picture and test results. User ID will be recorded
as 'Guest Mode: User not registered'.
■
Registered users will be able to use the device as normal (user's name will
be recorded).
○
Untick this to allow only registered users to use the device.
■
If a non-registered user uses the device when visitors are not allowed, it will
display “FAIL” and the visitor cannot take a test on the device.
●
Without Recognition
○
Untick this to enable the device's facial
recognition feature (default setting).
○
Tick this to disable the device's facial
recognition feature under one of the modes
below:
■
Controlled Mode - Allows an external controller (e.g. external card reader) to
control the device. In this case, special software may be used, e.g. to read
the card first before the alcohol test, etc. Please contact us for information.
■
Master Mode - Disable facial recognition for all users. Registered and
non-registered users can use the device but all users will be identified as
'Visitor' on the device or 'Guest' in Andalink test records.
●
Turn off calibration alarm
○
Untick this to get notified when the breathalyser sensor is due for calibration.
○
Tick this to disable the notification when the breathalyser sensor is due for calibration.
16
After making changes, click
to save the settings.
Temperature
●
Body Temperature -
○
Tick to enable temperature scanning for all
users
○
Untick to disable temperature scanning
completely
●
Body Temperature Scale -
○
Tick to select either °C or °F for body
temperature reading.
●
No body temperature limit -
○
Tick to disable all temperature alerts.
■
Temperature pass limit is 0 by default
(all temperature readings will be
'Pass')
○
Untick to set a body temperature limit and
alert sound:
■
Body temperature limit - Set a body temperature limit (e.g. 38.0). Any
detected body temperature above this set limit will sound an alert.
■
Important: Please ensure that the limit is set the same in Andalink (see KB
article: Ensure set alcohol / temperature limits are the same in Andalink and
device)
●
Do not display temperature reading -
○
Tick to hide the temperature reading when temperature is scanned.
○
Untick to show the temperature reading (e.g. 37.0°C) when temperature is scanned.
●
Locked after startup for 10 mins before calibration
○
Tick to lock the device for 10 minutes after device startup. No test or Temperature
Calibration can be performed during this time.
○
Untick to leave the device unlocked and allow testing and Temperature Calibration
after device startup.
●
Manual temperature calibration by operator only - Controls the calibration operator setting
for Manual Temperature Calibration (under 'Temperature Calibration' in Settings menu).
17
Tick to allow only Admin to conduct the Manual Temperature Calibration (default
setting).
○
Untick to allow anyone to perform the Manual Temperature Calibration.
●
Calibration by first measurement result and Startup calibration only - Controls the
settings for Interval Temperature Calibration.
After making changes, click
to save the settings.
18
Only Admin Calibration Operators can conduct
Interval Temperature Calibration
Anyone can conduct Interval Temperature
Calibration
Anyone can conduct Interval Temperature
Calibration but only during device Startup
Use this step to enrol anyone, including admin role users
for temperature calibration, into the device or Andalink.
●
Connect an external keyboard to the device
●
Enter the staff’s details into the system. All fields
are required (Staff ID or Username, Position,
First and Last Names, and Role).
●
Click on the ‘Server Enroll’ button to submit the
enrolment details.
●
Position the registered staff’s face in the device
camera so that it fills the green frame on the
screen. Please follow the Facial Recognition
Guidelines mentioned in this manual.
●
Once successfully enrolled, the device will
display ‘Enrolment successful’.
●
The device will sync the new user’s details to
Andalink during the next sync (about 10 minutes).
Alternatively, force sync the device to Andalink by
long clicking the version number (displayed on
the bottom right of the device screen).
Device Info
Shows the Device Information, including software version
number, serial numbers, last and next calibration dates, total
numbers of tests, total number of remaining tests and location.
19
2,000 test records can be stored along with the photo of the users.
Click on
to scroll or
to search for a record.
Click on a record to view details of the test record.
Click on
to print the test record (compatible printer must be connected) or click
to
return to view test records.
20
Date & Time
Set the device date and time here.
Language
Set device language preferences here.
Select the preferred language from the drop down menu
then click
to save.
21
Control the voice and audio alert volume here.
Voice recorded alerts
Controls the voice recorded alerts such as 'Visitor / Temperature
Reading / Pass'.
Drag the red dot on the bar to adjust the volume level.
To mute or disable all voice alerts on the device, click on the
icon.
The icon will change to
when muted/disabled. Click it again to unmute or enable voice alerts.
Sound alerts
Controls the sound alerts (alarm sounds when temperature is high etc.).
Click on the drop down arrow to change the volume level.
To mute or disable all sound alerts on the device, click on the
icon.
The icon will change to
when muted/disabled. Click it again to unmute or re-enable sound
alerts.
After making changes, click
to save the settings.
Note: The Temperature Calibration alarm sound is controlled under Advanced Settings > Test Delay
22
Set a customized Boot Animation and background picture for the device.
Tick the checkbox for 'Customize' and follow the instructions on the screen to upload the files.
Boot Animation Video - Save the file name as 'StartVideo.mp4' (resolution 800x1280, 10 seconds)
Background Image - Save the file name as 'Background.jpg' (resolution 800x1280)
Load the files to a USB drive's root directory, insert into the device USB slot, then click on Download
to Device.
After making changes, click
to save the settings.
23
Test records can be printed out automatically or manually
after each test.
Tick the checkboxes to select the applicable settings for the
printer.
After making changes, click
to save the settings.
Note: The printer is an optional accessory and not included
with the device. Contact Andatech for purchase options.
Temperature Calibration
Please read 'About Temperature Calibration' for more
information.
Calibration interval
By default, the device will require a Temperature Calibration
every 60 minutes. This is the recommended interval.
Although not advisable, you can select a different
temperature calibration interval from the drop down menu.
After changing the interval, click on the 'Start manual
calibration' button to save the new calibration interval
setting and conduct a manual temperature calibration.
A calibration operator needs to be physically present to
conduct the temperature calibration on the device. Refer to
'Temperature Calibration Procedure' for instructions.
24
If a temperature calibration is needed outside of its regular interval period (e.g. temperatures seem
too high or too low, or is measuring inconsistently), a manual temperature calibration can be
performed.
To do this, just click on the 'Start manual calibration' button.
A calibration operator needs to be physically present to conduct the temperature calibration on the
device. Refer to 'Temperature Calibration Procedure' for instructions.
Show alcohol test reading
Tick to display the Quick and Standard alcohol test results in numerical digits (default setting) e.g.
'0.02%BAC'.
Untick to display alcohol test results as Pass or Fail only.
This setting will apply to test results recorded both on the device and in Andalink.
Note: The Pass limit is 0.00 by default and cannot be changed. Any alcohol reading over 0.00 will be
considered a Fail and sound an alert.
25
TROUBLESHOOTING / WARNING DISPLAY
26
Issue / Warning Display
Possible Cause(s)
Solution(s)
Blow interrupt, please
blow
User stopped blowing or did not blow
hard enough.
Please try blowing into the sampling hole
again using a strong, steady breath until
the screen displays 'Blow completed.
Please Wait'.
Alcohol test failed
A user's face was recognised and
the Soberlive FRX was ready to take
a test, but the user failed to provide a
sample.
Face the Soberlive FRX camera again to
take a test.
Fail!
User is not registered as staff and
‘Visitors Allowed’ is disabled.
Register user as staff in Andalink or
enable ‘Visitors Allowed’ under
‘Advanced Settings (Blow)’.
Please contact
calibration operator
and/or
Thermal Imaging
Display is orange
The device requires Temperature
Calibration.
Perform a temperature calibration
following the instructions under
‘Temperature Calibration Procedure’.
Body Temperature:
XX.X°C Fail!
The detected body temperature was
higher than the set limit.
Check and change the body temperature
limit (e.g. 38.0°C) under ‘Advanced
Settings (Temperature)’ if it is too low.
Body Temperature is
too high (e.g. 38.0°C)
but the user passed
No Body Temperature Limit was set
(default limit is 0°C - any
temperature will Pass) or was set too
low (e.g. 32.0°C).
Go to Advanced Settings (Temperature)’
and set a body temperature limit (e.g.
37.5°C).
Device not recognising
or detecting user or
face correctly
User’s face is not within the Facial
Recognition Frame.
Position user so that face is within the
Facial Recognition Frame.
User is not following Facial
Recognition Guidelines.
Ensure users are following Facial
Recognition Guidelines.
Also, ensure that the user's photo in
Andalink is following Facial Recognition
Guidelines.
User is not registered in Andalink or
device.
Register the user in Andalink or device.
User was recently registered in
Andalink but not yet synced to the
device.
Ensure the device is connected to the
internet. Wait 10 minutes or Force Sync
to sync user to device.
User is not registered and Visitors
Mode is disabled.
Register user or enable Visitors Mode
under ‘Advanced Settings (Blow)’.
Mask Detection is enabled but
‘Visitors Allowed’ is disabled.
Either disable ‘Mask Detection’ or enable
both ‘Mask Detection and ‘Visitors
Allowed’ under ‘Advanced Settings
(Blow)’.
Users are being
detected as Visitor /
Guest although already
registered as staff in
Andalink
User was recently registered in
Andalink but not yet synced to
device.
Ensure device is connected to the
internet. Wait 10 minutes or Force Sync
to sync user to device.
Facial Recognition was disabled.
Untick the ‘Without Recognition’ setting
under ‘Advanced Settings (Blow)’ to
enable Facial Recognition feature.
Mask Detection is enabled.
Disable Mask Detection under
‘Advanced Settings (Blow)’.
Device is not
responding or the
display screen is frozen
Device requires a restart
Restart the device.
Temperature readings
are highly inaccurate or
inconsistent
Device installation did not follow
guidelines.
Ensure device installation follows
guidelines under ‘Device Installation’,
including:
-
Device is not tilting too much
upwards or downwards
-
Room does not have large
temperature fluctuations or is
constantly too hot or too cold
-
Heater or air conditioner is not
close or blowing directly onto
device
Perform a Manual Temperature
Calibration. Ensure that the calibration
operator has a normal body temperature
(e.g. 36~36.5°C).
Room temperature has large
temperature fluctuations.
Temperature calibration performed
incorrectly.
Temperature calibration performed
when the environment was too hot or
too cold.
Calibration operator did not have
normal body temperature.
1. Can a user delete or edit their details or test results?
No, this cannot be done on the device or in Andalink by any user (including by Admins or Managers).
2. How instant are the Soberlive FRX's test results, changes and updates synced to Andalink?
They sync automatically every 10 minutes. You can do a Force Synchronisation to sync it instantly.
3. What will happen if the Wi-Fi or internet is down in my location?
The Soberlive FRX will continue to operate and save test results offline. Once it is connected to the
internet or Wi-Fi network again, it will sync the results to the Andalink.
4. What is tail sampling and when should I enable or disable it?
Tail sampling takes the breath sample from the end of the blow. This is common in evidential testing
devices to make sure that the breath sample is a deep lung sample to provide an accurate alcohol
test result.
If you have enabled Standard Test mode and have users providing the breath sample using a straw,
you can enable Tail Sampling to get a more accurate alcohol test result.
However, if you are using the Soberlive FRX as a screening device and just need a pass or fail result
(Quick Test mode and using a sampling cup), then you can disable Tail Sampling.
BREATHALYSER DISCLAIMERS
●
This product must be used strictly in accordance with the instructions in this manual.
●
This product is intended to provide the user with an indication of alcohol on the breath.
However, neither the manufacturer nor the supplier gives any warranty (express or implied)
that the product can give any indication of blood alcohol level.
●
A person's BAC (Blood Alcohol Concentration) can only be accurately measured by the
administration of a blood test. This product only measures breath alcohol content which, in
turn, only gives an approximate indication of blood alcohol concentration.
●
In these circumstances the user of this product should not rely on the information provided
thereby as an accurate indication of their blood alcohol concentration or of their level of
intoxication or of their fitness to drive and in particular should not take any action in reliance
on such information as the product can provide no more than an indication of alcohol on the
breath.
●
If the user of this product makes any decision or takes any action based on the use of the
product such decision or such action shall be at the sole risk of the user and neither the
manufacturer nor the supplier assumes any responsibility for such decision or action.
●
In particular, neither the manufacturer nor the supplier shall be in any way liable if the user of
the product asserts that the product led them to believe they were fit to operate a motor
vehicle and are subsequently found not to be so fit.
29
MEASUREMENTS & CONDITIONS FOR ALCOHOL TESTING
●
Use a clean sampling cup or straw for each test. Sampling cups should be replaced after
each positive BAC test result or when it is dirty.
●
Do not allow samples containing smoke, tobacco or other chemical products to enter the unit,
as damage to the fuel cell can occur. Cigarette smoke may also increase the blood alcohol
concentration value displayed. Allow one minute after smoking before testing.
●
There must be an interval of at least 15 minutes after alcohol consumption as alcohol residue
in the mouth may give a false reading. Such residue may also be left by aromatic drinks,
mouthwash, medicines and drops, as well as by belching and vomiting. Rinsing your mouth
with water or non-alcoholic drinks does not substitute for an interval.
●
A person's blood alcohol concentration (BAC) can continue to rise for up to 2 hours after
drinking alcohol. If a result is close to the designated BAC, it is recommended to wait and
retest after 20 minutes.
●
It can take 10 hours or more for a person's BAC to return to zero after a high BAC has been
reached. A further test should be carried out later in the day or the following morning.
LIABILITY FOR PROPER FUNCTION OR DAMAGE
The liability for proper functioning of the instrument is irrevocably transferred to the owner or operator
in the event the instrument is serviced or repaired by personnel not employed or authorised by
Andatech. In such an event, the warranty will be void.
30
WARRANTY
The manufacturer warrants for 1 year from the
date of purchase, the product to be free from
defects in workmanship or material (excluding
recalibration) under normal use. The
manufacturer’s obligation under this warranty is
limited to replacing, adjusting or repairing the
unit if returned along with proof of purchase.
This warranty is void if the unit has been
tampered with, damaged or abused in any way.
Visit my.andatech.com.au to register your
warranty online and receive calibration
reminders.
BREATHALYSER
CALIBRATION & SERVICE
Andatech recommends the Andatech® Soberlive
FRX to be calibrated every 6 months. The
Soberlive FRX cannot be used to take a test
once it has passed its calibration due date.
To register your product warranty, receive
calibration reminders and book in a calibration,
please go to my.andatech.com.au or contact:
Andatech
Hotline: 1300 800 200
Website: www.andatech.com.au
HOW TO REMOVE/REPLACE SENSOR FOR CALIBRATION
The Soberlive FRX has a replaceable sensor that can be removed and returned for the calibration
service without sending the whole device back. A new sensor can then be replaced into the device
easily. Follow the instructions below:
31
Register your warranty online now at my.andatech.com.au
Congratulations on your purchase! It is our aim to provide you with quality products that you can trust. Our
products come with guarantees that cannot be excluded under the Australian Consumer Law.
You are entitled to a replacement or a refund for a major failure and for compensation for any other reasonably
foreseeable loss or damage. You are also entitled to have the goods repaired or replaced if the goods fail to be of
acceptable quality and the failure does not amount to a major failure.
The Guarantee
The product is guaranteed to be free from defects in workmanship and parts for a period of 12 months from the
date of purchase for domestic use and 3 months for commercial use. Defects that occur within this warranty
period, under normal use and care, will be repaired, replaced or refunded at our discretion, solely at our option
with no charge for parts and labour. The benefits conferred by this warranty are in addition to all rights and
remedies in respect of the product that the consumer has under the Trade Practices Act and similar state laws.
Proof of Purchase
This warranty is valid for the original purchase and is not transferable. Please keep your purchase docket or
receipt as proof of purchase and as proof of date on which the purchase was made. The purchase docket (or a
copy) or receipt must be presented with the warranty when making a claim under this warranty.
Service during the Warranty Period
To claim for warranty, contact us on 1300 800 200 or [email protected] to get an RA# (Return
Authorisation Number). Ensure that the RA# is clearly stated on the outside of the packaging and that the product
is properly packaged so that no damage occurs to the product during transit. Shipping of the product back to us
for warranty will be at our cost. A product return without the RA# or proof of purchase will not be accepted.
Extent of Warranty
This warranty is limited to defects in workmanship or parts. All defective products or parts will be repaired or
replaced. This warranty does not cover manuals and packaging.
Normal Wear and Tear
This warranty does not cover normal wear and tear to the product or parts.
Exclusions
This warranty does not cover:
●
Calibration costs
●
Any defect caused by an accident, misuse, abuse, improper operation, lack of reasonable care,
unauthorised modification, loss of parts, tampering or attempted repair by a person not authorised by
the distributor.
●
Any product that has been damaged by a lightning strike either directly or indirectly or a main power
surge or liquid ingress.
●
The product if it is located outside of Australia.
●
Any damage caused by improper power input or improper cable connection.
Andatech Pty. Ltd.
PO Box 3038, Nunawading VIC 3131, Australia.
Phone: 1300 800 200 Fax: 1300 883 802 Email: [email protected]